Spectrophotometric quantification of nucleic acids using the nanovette microliter cell accessory with a DU 730 spectrophotometer

Size: px
Start display at page:

Download "Spectrophotometric quantification of nucleic acids using the nanovette microliter cell accessory with a DU 730 spectrophotometer"

Transcription

1 Application Information Spectrophotometric quantification of nucleic acids using the nanovette microliter cell accessory with a DU 730 spectrophotometer Amy Yoder, Mary Blair, Kenny Abell Beckman Coulter, Inc., Indianapolis, IN, USA Introduction Nucleic acid quantification is one of the most important applications in molecular biology labs. The concentration and purity of double stranded DNA (dsdna), single stranded DNA (ssdna) and RNA are routinely determined utilizing UV/Vis spectroscopy. In UV/Vis spectroscopy, standard cuvettes with 1 cm pathlengths requiring 50 μl to 2000 μl volumes are typically used. In order to remain in the linear range of the spectrophotometer, dilution of the sample is often required which subsequently eliminates its use in downstream applications. The nanovette microliter cell addresses the shortcomings of traditional cuvettes by allowing measurement samples as small as 0.7 μl with concentrations as high as 5,000 ng/μl. This application note presents data on the outstanding linearity, dynamic range, reproducibility, and correlation to a standard cuvette of the nanovette microliter cell when used with Beckman Coulter s DU 730 spectrophotometer. Operation By reducing the measurement pathlength, the nanovette allows sample size to decrease from the typical milliliter to less than a microliter. Pathlength adjustment takes place via interchangeable lids that easily fit onto the top of the microliter cell. A 1 mm pathlength lid provides for a virtual dilution of 1:10; a 0.2 mm pathlength lid provides for a virtual dilution of 1:50. Because a shorter pathlength reduces the measured absorbance of the solution, it is critical that the spectrophotometer offers a wide dynamic range in order to make measurements over a wide range of concentrations. Used with Beckman Coulter s DU 730 spectrophotometer, the nanovette can measure dsdna concentrations ranging from 10 ng/μl to 5,000 ng/μl. Table 1 provides the measurement range of the nanovette when using the DU 730 spectrophotometer for different nucleic acid samples.* Table 1. nanovette Measurement Range with the DU 730 Sample 1 mm Lid 0.2 mm Lid Specific (Virtual Dilution Factor 10) (Virtual Dilution Factor 50) Total Detection Range Factor [ng/μl] [ng/μl] [ng/μl] dsdna , , ,000 ssdna , ,700 ssrna , ,000 Oligo , ,000 * The ranges for ssdna, ssrna and Oligo are based on the results from dsdna. A-11074A

2 The size of a standard cuvette (12.5 mm x 12.5 mm x 53 mm), the nanovette fits easily into the cuvette holder of the DU 730. No alignment is required, although a small step must be attached to the bottom of the microliter cell to obtain the proper aperture center height (both the step and a small screwdriver are included with the nanovette). Once in position, the nanovette cell remains in place during filling, measuring and cleaning. This guarantees a continuously identical position of the aperture in the light beam and no variation in comparison to the reference measurement. Sample volume depends on the particular lid being used: the 0.2 mm pathlength lid accepts from 0.7 μl to 4.0 μl; the 1.0 mm lid accepts from 3.0 μl to 5.0 μl. Detailed operation instructions can be obtained from the nanovette user s guide, available at Experimental For all of the following tests, a stock solution was prepared by dissolving salmon testes DNA in nuclease free sterile water to a concentration of approximately 5,000 ng/μl. This solution was then serially diluted to provide multiple samples ranging from approximately 5,000 ng/μl to 10 ng/μl. For large templates, such as genomic DNA, it is necessary to heat the DNA before measurement to relax the structure in order to obtain a completely homogenous solution. The DNA used in this application was heated at 60 C for at least 15 minutes and then mixed thoroughly prior to measurements. Unless otherwise stated, the sample volume used in each measurement was 1.5 μl with the 0.2 mm pathlength lid and 4 μl with the 1.0 mm To determine the nucleic acid concentration in solution, the absorbance at wavelength 260 nm (A260) is used. The following function, derived from Lambert-Beer s law, is applied: Concentration = Sample Virtual (A260 A320) Specific Dilution Factor Factor In the case of dsdna, the sample specific factor is 50. The virtual dilution factor will be 50 or 10, depending on whether the 0.2 mm pathlength lid or the 1.0 mm pathlength lid, respectively, is used. In this formula, the background absorbance of 320 nm must be subtracted for accurate results. Rather than taking separate measurements, the 320 nm background correction feature was enabled on the DU 730 spectro photometer. This feature automatically measures and subtracts the absorbance at 320 nm from the obtained results. NOTE: Solution homogeneity is critical when measuring low-volume samples on a fiber-optic microcell such as the nanovette. Therefore, when using large templates such as genomic DNA, it is necessary to preheat the sample to 60 C for 15 minutes followed by thoroughly mixing to ensure the accuracy of the data obtained with the nanovette. Linearity To measure linearity, a stock solution was serially diluted to provide samples ranging from approximately 10 ng/μl to 5,000 ng/μl (see experimental). These samples were then measured using both the 1.0 mm pathlength lid and the 0.2 mm pathlength lid and the results plotted as figure 1 and figure 2, respectively. Note that each data point represents the average of 10 measurements. Linearity is excellent, evident by an R 2 of over in both cases. Reproducibility Reproducibility of the nanovette was measured in two different ways. Method 1: Reproducibility using the same sample size 10 measurements each of the varied concentrations were made, with results included in table 2. The sample size for the 0.2 mm lid was 1.5 μl; the sample size for the 1.0 mm lid was 4 μl. 2

3 Figure 1 Figure 2 Table 2. Reproducibility results using same sample size Concentration (ng/μl) Coefficient of Variation 7.77% 1.31% 1.30% 0.83% 0.45% 1.11% Concentration (ng/μl) Coefficient of Variation 2.16% 1.45% 0.49% 1.01% 0.58% 1.67% Method 2: Reproducibility with varied sample size The nanovette accepts sample sizes from 3 μl to 5 μl when using the 1.0 mm pathlength lid, and from 0.7 μl to 4 μl when using the 0.2 mm pathlength lid. In this experiment, 4 measurements of a single concentration were made at various volumes with each The reproducibility data shows excellent reproducibility across a wide concentration, with an expected decrease in reproducibility at the lower end of the measurement range (coinciding with nearing the noise floor of the spectrophotometer). Reproducibility with different volumes is also excellent, with a 0.82% difference from the mean using extreme sample volumes with the 1.0 mm pathlength lid and a 3.41% difference from the mean using the extreme sample volumes with the 0.2 mm Table 3. Reproducibility results with varied sample size Avg. Conc. % diff. Volume (ng/μl) CV from mean 5 μl % 0.31% 4 μl % 0.11% 3 μl % 0.51% Avg. Conc. % diff. Volume (ng/μl) CV from mean 3 μl % 1.78% 2 μl 1, % 0.26% 1 μl 1, % 0.11% 0.7 μl 1, % 1.63% 3

4 Carryover Effects One of the primary benefits of the nanovette is improved throughput; the ability to quickly and easily clean the device is paramount. When cleaning the nanovette, the cell remains inside the spectro - photometer and the measurement window cleaned with a laboratory wipe. For the pathlength lid, a lintfree swab can be used to fully evacuate the area. However, due to the shape of the lid, compressed air is recommended over the lint-free swab. In this application note, the cell and the lid were wiped with a lint free swab followed by drying with compressed air between each sample. Carryover effect was assessed by measuring 3 sets of 3 samples of DNA followed by 3 samples of buffer solution. Results are shown in figure 3 for the 1.0 mm pathlength lid and figure 4 for the 0.2 mm Correlation to a Standard Cuvette To test the correlation of the nanovette with a standard quartz cuvette (pathlength = 10 mm), a dilution series of dsdna was measured over a broad range. 10 measurements with the nanovette were taken at each concentration against 3 measurements with a standard cuvette, with the results plotted as figure 5 for the 1.0 mm pathlength lid and figure 6 for the 0.2 mm Note that in order to compare the data, the dsdna measured in the standard quartz cuvette had to be diluted accordingly to fall within the absorbance range of the instrument ,000 ng/μl ,000 ng/μl Absorbance 260 mm Absorbance 260 mm sample 1 sample 2 sample 3 water 1 water 2 water 3 sample 4 sample 5 sample 6 water 4 water 5 water 6 sample 7 sample 8 sample 9 water 7 water 8 water 9 sample 1 sample 2 sample 3 water 1 water 2 water 3 sample 4 sample 5 sample 6 water 4 water 5 water 6 sample 7 sample 8 sample 9 water 7 water 8 water 9 Figure 3 Figure 4 The data shows the importance of proper cleaning of the nanovette. Empirical results have shown that proper cleaning of the nanovette will result in virtually no carryover Concentration (ng/μl) Standard Cuvette 1.0 mm Lid Concentration (ng/μl) Standard Cuvette 0.2 mm Lid Figure 5 Figure 6 4

5 As expected, the longer 1.0 mm pathlength lid does provide greater accuracy, but both lids provide excellent results over their respective ranges. Conclusion This application note demonstrates the performance of the nanovette microliter cell accessory when used with the DU 730 spectrophotometer. Excellent linearity and reproducibility across a wide concentration range, reproducibility within a wide variety of sample sizes, minimal carryover effect, and correlation to a standard cuvette provides confidence regarding any potential DNA or protein application using the nanovette with the DU 730 spectro - photometer. The nanovette is a useful accessory to the biology laboratory, providing fast, easy, and accurate measurements. Australia, Gladesville (61) Canada, Mississauga (1) China, Beijing (86) Czech Republic, Prague (420) Eastern Europe, Middle East, North Africa, South West Asia: Switzerland, Nyon (41) France, Villepinte (33) Germany, Krefeld (49) Hong Kong (852) India, Mumbai (91) Italy, Cassina de Pecchi, Milan (39) Japan, Tokyo (81) Korea, Seoul (82) Latin America (1) (305) Mexico, Mexico City (001) Netherlands, Mijdrecht (31) Puerto Rico (787) Singapore (65) South Africa/Sub-Saharan Africa, Johannesburg (27) /5 Spain, Madrid (34) Sweden, Bromma (46) Switzerland, Nyon (41) Taiwan, Taipei (886) Turkey, Istanbul (90) UK, High Wycombe (44) USA, Fullerton, CA (1) B Beckman Coulter, Inc. BMR-PRINTED IN U.S.A.

Identification of System Parameters Critical for High-Performance cief

Identification of System Parameters Critical for High-Performance cief Application Information Identification of System Parameters Critical for High-Performance cief Ingrid D. Cruzado-Park, Scott Mack and Chitra K. Ratnayake Beckman Coulter, Inc., Fullerton, CA, USA Introduction

More information

Performance of Semi-micro and Single Drop Analysis by Using. Flash Xenon Lamp Spectrophotometer

Performance of Semi-micro and Single Drop Analysis by Using. Flash Xenon Lamp Spectrophotometer Application Note SpectroArt 2 8/9/27 Performance of Semi-micro and Single Drop Analysis by Using Flash Xenon Lamp Spectrophotometer Aim: To investigate the accuracy, reproducibility and linearity of Wealtec

More information

BioDrop DUO dsdna Application Note

BioDrop DUO dsdna Application Note BioDrop DUO dsdna Application Note Using a BioDrop DUO spectrophotometer to measure the concentration of low volume samples of dsdna Micro-volume measurement of DNA is a routine application in many life

More information

Quick Guide QUICK GUIDE. Activity 1: Determine the Reaction Rate in the Presence or Absence of an Enzyme

Quick Guide QUICK GUIDE. Activity 1: Determine the Reaction Rate in the Presence or Absence of an Enzyme Activity 1: Determine the Reaction Rate in the Presence or Absence of an 1. Find your 15 ml conical tubes labeled Stop Solution, 1.5 mm Substrate, and Buffer. Write your initials on each tube. 2. Label

More information

The Plus for your lab!

The Plus for your lab! BioPhotometer plus NEW! analysis system The Plus for your lab! The easy, rapid and reliable BioPhotometer plus analysis system See it clearly! UVette Precision in plastics The patented* 1 UVette is the

More information

Introduction to Fourier Transform Infrared Spectroscopy

Introduction to Fourier Transform Infrared Spectroscopy Introduction to Fourier Transform Infrared Spectroscopy Introduction What is FTIR? FTIR stands for Fourier transform infrared, the preferred method of infrared spectroscopy. In infrared spectroscopy, IR

More information

[ OSTRO PASS-THROUGH SAMPLE PREPARATION PRODUCT ] The Simpler Way to Cleaner Samples

[ OSTRO PASS-THROUGH SAMPLE PREPARATION PRODUCT ] The Simpler Way to Cleaner Samples [ OSTRO PASS-THROUGH SAMPLE PREPARATION PRODUCT ] The Simpler Way to Cleaner Samples SIMPLE EASY EFFICIENT DEPENDABLE 2 waters.com/ostro 3 SIMPLE The Ostro Pass-Through Sample Preparation Product provides

More information

Optical density measurements automatically corrected to a 1-cm pathlength with PathCheck Technology

Optical density measurements automatically corrected to a 1-cm pathlength with PathCheck Technology APPLICATION NOTE Optical density measurements automatically corrected to a 1-cm pathlength with PathCheck Technology Introduction UV/VIS spectrophotometers and microplate readers differ fundamentally in

More information

Spectrophotometer for Life Science. BioSpec-nano C101-E112D

Spectrophotometer for Life Science. BioSpec-nano C101-E112D Spectrophotometer for Life Science BioSpec-nano C11-E112D BioSpec-nano Spectrophotometer for Life Science Drop-and-Start Analysis Automatic Optical Pathlength Setting & Automatic Wiping Low Carryover Achieved

More information

Nucleic Acid Quantitation in Microplates

Nucleic Acid Quantitation in Microplates PHARMACEUTICAL application note Nucleic Acid Quantitation in Microplates HansWilly Mueller, Ph.D. J. Fenton Williams, Ph.D. Introduction Many molecular biology applications require the determination of

More information

[ Care and Use Manual ]

[ Care and Use Manual ] Oasis HLB DISKS I. Introduction Oasis HLB is a Hydrophilic-Lipophilic-Balanced, water-wettable, reversed-phase sorbent for all your SPE needs. It is made from a specific ratio of two monomers, the hydrophilic

More information

Introduction to Fourier Transform Infrared Spectroscopy

Introduction to Fourier Transform Infrared Spectroscopy molecular spectroscopy Introduction to Fourier Transform Infrared Spectroscopy Part of Thermo Fisher Scientific Introduction What is FT-IR? FT-IR stands for Fourier Transform InfraRed, the preferred method

More information

Nucleic Acid Thermo Scientific NanoDrop Spectrophotometers

Nucleic Acid Thermo Scientific NanoDrop Spectrophotometers Thermo Scientific NanoDrop Spectrophotometers Part of Thermo Fisher Scientific The information in this publication is provided for reference only. All information contained in this publication is believed

More information

Global Data Catalog initiative Christophe Charpentier ArcGIS Content Product Manager

Global Data Catalog initiative Christophe Charpentier ArcGIS Content Product Manager Global Data Catalog initiative Christophe Charpentier ArcGIS Content Product Manager ccharpentier@esri.com We are in a global economy Global Economy Business and Operations are now global Strategic need

More information

Low-volume, High Throughput Workflow for Analysis of Nucleic Acid Samples for Biobanking

Low-volume, High Throughput Workflow for Analysis of Nucleic Acid Samples for Biobanking A p p l i c a t i o n N o t e Low-volume, High Throughput Workflow for Analysis of Nucleic Acid Samples for Biobanking Peter J. Brescia, Jr., Chris Wilson, and Peter Banks, BioTek Instruments, Inc., Winooski,

More information

Applications Note 228 September 2010

Applications Note 228 September 2010 Applications Note 228 September 2010 Reproducible photometric determination of DNA concentrations using the Eppendorf UVette in the Eppendorf BioPhotometer plus Martin Armbrecht, Eppendorf AG, Hamburg,

More information

High-throughput Isolation of Genomic DNA From Buccal Swab on the Eppendorf epmotion 5075 VAC

High-throughput Isolation of Genomic DNA From Buccal Swab on the Eppendorf epmotion 5075 VAC APPLICATION NOTE No. 378 I August 2016 High-throughput Isolation of Genomic DNA From Buccal Swab on the Eppendorf epmotion 5075 VAC Hatim Boughanem, Manuel Macias-Gonzalez, BIOEASY SL, Spain Abstract The

More information

2017 Source of Foreign Income Earned By Fund

2017 Source of Foreign Income Earned By Fund 2017 Source of Foreign Income Earned By Fund Putnam Emerging Markets Equity Fund EIN: 26-2670607 FYE: 08/31/2017 Statement Pursuant to 1.853-4: The fund is hereby electing to apply code section 853 for

More information

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis A. Caruso, M. Santoro, P. Magni, S. Pelagatti, and R. Facchetti Thermo Fisher Scientific, Milan,

More information

Microdetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats

Microdetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats Application Note 24 Microdetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats INTRODUCTION Most spectrophotometric

More information

IN QUALITATIVE ANALYSIS,

IN QUALITATIVE ANALYSIS, IN QUALITATIVE ANALYSIS, YOU VE ALWAYS HAD TO CHOOSE BETWEEN ACCURACY AND THROUGHPUT. NOW THERE S NO REASON TO ever have to. WATERS QUALITATIVE ANALYSIS solutions Waters ACQUITY UPLC System with the Waters

More information

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer Reiko Kiyonami, Mary Blackburn, Andreas FR Hühme: Thermo Fisher

More information

viridis Columns Bringing a New Dimension to SFC Combining state-of-the-art media manufacturing with industry leading column technology, Viridis

viridis Columns Bringing a New Dimension to SFC Combining state-of-the-art media manufacturing with industry leading column technology, Viridis [ viridis columns ] viridis Columns Bringing a New Dimension to SFC Combining state-of-the-art media manufacturing with industry leading column technology, Viridis Supercritical Fluid Chromatography (SFC)

More information

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION A purpose-built collection of the best-inclass components for forensic toxicology Whether you re challenged to present reliable and secure forensic sample

More information

NanoDrop One Viewer software NanoDrop One Website. NanoDrop One Website NanoDrop One Viewer software NanoDrop One Website Software System Update Update Update Software, Update Note OK Language Measure

More information

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer Defining quality standards for the analysis of solid samples Redefine your quality standards for the elemental analysis of solid samples

More information

READY TO SCRAP: HOW MANY VESSELS AT DEMOLITION VALUE?

READY TO SCRAP: HOW MANY VESSELS AT DEMOLITION VALUE? READY TO SCRAP: HOW MANY VESSELS AT DEMOLITION VALUE? August 206 VesselsValue Global number of vessels at demolition value At scrap value 7,27 6 Above scrap value,8 84 Number of vessels at demolition value

More information

MiSeq System. Denature and Dilute Libraries Guide

MiSeq System. Denature and Dilute Libraries Guide MiSeq System Denature and Dilute Libraries Guide Overview 3 Protocol A: Standard Normalization Method 4 Protocol B: Bead-Based Normalization Method 6 Denature and Dilute PhiX Control 7 Supplemental Information

More information

Keysight Technologies Oxygen-Free High-Resolution Electrochemical SPM. Application Note

Keysight Technologies Oxygen-Free High-Resolution Electrochemical SPM. Application Note Keysight Technologies Oxygen-Free High-Resolution Electrochemical SPM Application Note Introduction For two decades, scanning probe microscopy (SPM) has provided scientists a unique tool to study in situ

More information

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites Robert Heintz, Mark Wall, Jennifer Ramirez, Stephan Woods Thermo Fisher Scientific, Madison WI Overview Purpose:

More information

Performance and Control of the Agilent Nano Indenter DCM

Performance and Control of the Agilent Nano Indenter DCM Performance and Control of the Agilent Nano Indenter DCM Application Note Introduction With new materials and material applications come new test challenges. As these new challenges emerge in materials

More information

Appendix B: Detailed tables showing overall figures by country and measure

Appendix B: Detailed tables showing overall figures by country and measure 44 country and measure % who report that they are very happy Source: World Values Survey, 2010-2014 except United States, Pew Research Center 2012 Gender and Generations survey and Argentina 32% 32% 36%

More information

Aerospace part number guide

Aerospace part number guide TRELLEBORG SEALING SOLUTIONS Aerospace part number guide YOUR PARTNER FOR SEALING TECHNOLOGY Your Partner for Sealing Technology Trelleborg Sealing Solutions is a major international sealing force, uniquely

More information

Keysight Technologies Measurement Uncertainty of VNA Based TDR/TDT Measurement. Application Note

Keysight Technologies Measurement Uncertainty of VNA Based TDR/TDT Measurement. Application Note Keysight Technologies Measurement Uncertainty of VNA Based TDR/TDT Measurement Application Note Table of Contents Introduction... 3 S-parameter measurement uncertainty of VNA... 4 Simplification of systematic

More information

ab83360 Ammonia Assay Kit

ab83360 Ammonia Assay Kit Version 8 Last updated 18 December 2017 ab83360 Ammonia Assay Kit For the measurement of total ammonia and ammonium levels in various samples This product is for research use only and is not intended for

More information

Operation Manual. SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

Operation Manual. SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Operation Manual SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC. 4.18 Foreword

More information

Nano Spectrophotometer LNS-A10

Nano Spectrophotometer LNS-A10 Nano Spectrophotometer LNS-A10 Nano spectrophotometer with full spectrum light covers wide wavelength range of 200 to 800nm. It requires only 0.5-2.0µl micro-volume sample for measurement which takes place

More information

Instrumental Technique: Cuvette. Md Rabiul Islam

Instrumental Technique: Cuvette. Md Rabiul Islam Instrumental Technique: Cuvette Md Rabiul Islam 16-7-2016 What is cuvette? A cuvette is a small tube of circular or square cross section, sealed at one end, made of plastic, glass, or fused quartz (for

More information

ˆ GDP t = GDP t SCAN t (1) t stat : (3.71) (5.53) (3.27) AdjustedR 2 : 0.652

ˆ GDP t = GDP t SCAN t (1) t stat : (3.71) (5.53) (3.27) AdjustedR 2 : 0.652 We study the relationship between SCAN index and GDP growth for some major countries. The SCAN data is taken as of Oct. The SCAN is of monthly frequency so we first compute its three-month average as quarterly

More information

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 000 Elemental Analyzer Dr. Liliana Krotz and Dr. Guido Giazzi Thermo Fisher Scientific, Milan, Italy

More information

High-throughput Quantification of DNA for NGS Library Prep with the Zephyr G3 Workstation and the VICTOR Nivo Plate Reader

High-throughput Quantification of DNA for NGS Library Prep with the Zephyr G3 Workstation and the VICTOR Nivo Plate Reader TECHNICAL APPLICATION NOTE High-throughput Quantification of DNA for NGS Library Prep with the Zephyr G3 Workstation and the VICTOR Nivo Plate Reader NGS Automation Image or Color Block Area Next generation

More information

protein interaction analysis bulletin 6300

protein interaction analysis bulletin 6300 protein interaction analysis bulletin 6300 Guide to SPR Data Analysis on the ProteOn XPR36 System Ruben Luo, Bio-Rad Laboratories, Inc., 2000 Alfred Nobel Drive, Hercules, CA 94547 Kinetic Analysis To

More information

ab83360 Ammonia Assay Kit

ab83360 Ammonia Assay Kit Version 9 Last updated 7 February 2019 ab83360 Ammonia Assay Kit For the measurement of total ammonia and ammonium levels in various samples View kit datasheet: www.abcam.com/ab83360 (use www.abcam.cn/ab83360

More information

C101-E112. BioSpec-nano. Shimadzu Spectrophotometer for Life Science

C101-E112. BioSpec-nano. Shimadzu Spectrophotometer for Life Science C101-E112 BioSpec-nano Shimadzu Spectrophotometer for Life Science Power of small. BioSpec-nano BioSpec-nano Shimadzu Spectrophotometer for Life Science Quick and Simple Nucleic Acid Quantitation Drop-and-Click

More information

Bio 120 Lab 5: Quantitative Analysis

Bio 120 Lab 5: Quantitative Analysis Bio 120 Lab 5: Quantitative Analysis Remember from the measurement lab, concentration is the amount of a solute, also called an analyte, in a given amount of solution. We need a way to measure the concentration

More information

Estapor Microspheres n 11. A technical newsletter published by Merck Chimie SAS - France

Estapor Microspheres n 11. A technical newsletter published by Merck Chimie SAS - France Estapor Microspheres n 11 A technical newsletter published by Merck Chimie SAS - France The Importance of Quality Methods in the Precise Sizing of Microsphere s Diameter CILAS Richard VIDAL, Cécile GENIES*

More information

TL6 Ultra Micro-volume Spectrophotometer

TL6 Ultra Micro-volume Spectrophotometer TL6 Ultra Micro-volume Spectrophotometer TL6 is the latest Ultra Micro-volume Spectrophotometer of PERSEE ANALYTICS. It is a small portable instrument for fast analysis of nucleic acid and protein concentrations

More information

ia PU BLi s g C o M Pa K T Wa i n CD-1576

ia PU BLi s g C o M Pa K T Wa i n CD-1576 M h M y CD-1576 o M Pa g C n ar ia PU BLi s in K T Wa i n ed National Geography Standards National Geography Standards Teachers leading discussions while completing units and activities is a prerequisite

More information

Do Policy-Related Shocks Affect Real Exchange Rates? An Empirical Analysis Using Sign Restrictions and a Penalty-Function Approach

Do Policy-Related Shocks Affect Real Exchange Rates? An Empirical Analysis Using Sign Restrictions and a Penalty-Function Approach ISSN 1440-771X Australia Department of Econometrics and Business Statistics http://www.buseco.monash.edu.au/depts/ebs/pubs/wpapers/ Do Policy-Related Shocks Affect Real Exchange Rates? An Empirical Analysis

More information

Fast, Effective XPS Point Analysis of Metal Components

Fast, Effective XPS Point Analysis of Metal Components Application Note: 52297 Fast, Effective XPS Point Analysis of Metal Components Chris Baily and Tim Nunney, Thermo Fisher Scientific, East Grinstead, West Sussex, UK Key Words K-Alpha Auto-Analysis Multi-Spectrum

More information

Keysight Technologies Instrumented Indentation Testing with the Keysight Nano Indenter G200. Application Note

Keysight Technologies Instrumented Indentation Testing with the Keysight Nano Indenter G200. Application Note Keysight Technologies Instrumented Indentation Testing with the Keysight Nano Indenter G200 Application Note Introduction The scale of materials and machined components continues to decrease with advances

More information

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS Thermo Scientific ConFlo IV Universal Interface Continuous Flow Interface Isotope Ratio MS 3 ConFlo IV Universal Interface for Continuous Flow Isotope Ratio MS The development of Continuous Flow carrier

More information

for XPS surface analysis

for XPS surface analysis Thermo Scientific Avantage XPS Software Powerful instrument operation and data processing for XPS surface analysis Avantage Software Atomic Concentration (%) 100 The premier software for surface analysis

More information

Measurement of L-Malic Acid in Wines Using the SpectraMax Plus 384 Microplate Reader

Measurement of L-Malic Acid in Wines Using the SpectraMax Plus 384 Microplate Reader now part of MDS Analytical Technologies Measurement of L-Malic Acid in Wines Using the SpectraMax Plus 384 Microplate Reader SpectraMax Application Note #18 By Theresa Heredia and Gillian Sutherland from

More information

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems Yan Liu, Zhongqing Lu, and Chris Pohl; Thermo Fisher Scientific, Sunnyvale, CA USA Overview

More information

Microplate Spectrophotometer

Microplate Spectrophotometer Microplate Spectrophotometer Features Benefits Applications Bali, October 1999 Optical System Xenon Flashing Lamp Order Sorting Filters Monochromator Data Light collection Light Transfer Why are these

More information

Microbial DNA qpcr Multi-Assay Kit Clostridium perfringens Pathogenicity

Microbial DNA qpcr Multi-Assay Kit Clostridium perfringens Pathogenicity Microbial DNA qpcr Multi-Assay Kit Clostridium perfringens Pathogenicity Cat. no. 330043 BBID-1507ZR-3 For real-time PCR-based, application-specific microbial identification or profiling The Clostridium

More information

EXPERIMENT #3 A Beer's Law Study

EXPERIMENT #3 A Beer's Law Study OBJECTVES: EXPERMENT #3 A Beer's Law Study To operate a Spectronic 20 To convert from percent transmission to absorbance units To plot absorbance versus wavelength and find max To plot absorbance versus

More information

Agilent EEsof EDA.

Agilent EEsof EDA. Agilent EEsof EDA This document is owned by Agilent Technologies, but is no longer kept current and may contain obsolete or inaccurate references. We regret any inconvenience this may cause. For the latest

More information

For the rapid, sensitive and accurate measurement of Caspase 9 activity in cell and tissue lysates.

For the rapid, sensitive and accurate measurement of Caspase 9 activity in cell and tissue lysates. ab65608 Caspase 9 Assay Kit (Colorimetric) Instructions for Use For the rapid, sensitive and accurate measurement of Caspase 9 activity in cell and tissue lysates. This product is for research use only

More information

Introduction to Spectroscopy: Analysis of Copper Ore

Introduction to Spectroscopy: Analysis of Copper Ore Absorbance Introduction to Spectroscopy: Analysis of Copper Ore Introduction The goal of this lab is to determine the unknown concentration of two different copper solution samples, taken from fictitious

More information

MOHAWK COLLEGE OF APPLIED ARTS AND TECHNOLOGY CHEMICAL AND ENVIRONMENTAL TECHNOLOGY DEPARTMENT. Lab Report ROOM NO: FE E309

MOHAWK COLLEGE OF APPLIED ARTS AND TECHNOLOGY CHEMICAL AND ENVIRONMENTAL TECHNOLOGY DEPARTMENT. Lab Report ROOM NO: FE E309 MOHAWK COLLEGE OF APPLIED ARTS AND TECHNOLOGY CHEMICAL AND ENVIRONMENTAL TECHNOLOGY DEPARTMENT Lab Report ROOM NO: FE E309 EXPERIMENT NO : 4 TITLE : UV Spectrophotometric Analysis of DNA Submitted by Class

More information

INSTRUCTION MANUAL. VWR myspec spectrophotometer

INSTRUCTION MANUAL. VWR myspec spectrophotometer VWR myspec spectrophotometer INSTRUCTION MANUAL European Catalogue Numbers: VWR myspec 732-2533 VWR myspec Touch 732-2534 VWR myspec Twin 732-2535 VWR myspec Twin Touch 732-2536 Version: 2.0 Issued: 29

More information

Product Data. Brüel & Kjær B. Sound Intensity Calibrator Type 3541

Product Data. Brüel & Kjær B. Sound Intensity Calibrator Type 3541 Product Data Sound Intensity Calibrator Type 3541 USES: Sound intensity and particle velocity calibrations Sound pressure calibration Measurement of pressure-residual intensity index FEATURES: Intensity

More information

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Andrea Caruso and Massimo Santoro, Thermo Fisher Scientific, Milan, Italy Application Note 1316 Key

More information

Keysight Technologies Young s Modulus of Dielectric Low-k Materials. Application Note

Keysight Technologies Young s Modulus of Dielectric Low-k Materials. Application Note Keysight Technologies Young s Modulus of Dielectric Low-k Materials Application Note Introduction In digital circuits, insulating dielectrics separate the conducting parts (wire interconnects and transistors)

More information

Spectrometer User s Guide

Spectrometer User s Guide Spectrometer User s Guide (Order Codes: V-SPEC, SPRT-VIS, SP-VIS, SP-UV-VIS, ESRT-VIS) The spectrometer is a portable light spectrophotometer, combining a spectrometer and a light source/cuvette holder.

More information

The Raman Spectroscopy of Graphene and the Determination of Layer Thickness

The Raman Spectroscopy of Graphene and the Determination of Layer Thickness Application Note: 52252 The Raman Spectroscopy of Graphene and the Determination of Layer Thickness Mark Wall, Ph.D., Thermo Fisher Scientific, Madison, WI, USA Key Words DXR Raman Microscope 2D Band D

More information

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry O. Scheibner, 1 P. van Baar, 2 F. Wode, 2 U. Dünnbier, 2 K. Akervik, 3 J. Humphries, 3 M. Bromirski 1

More information

Experiment 1 Chemical Kinetics 1

Experiment 1 Chemical Kinetics 1 Experiment 1 Chemical Kinetics 1 Purpose: Determine the rate law for the reaction of the dye crystal violet with hydroxide. Reading: Brown, et. al., Chemistry The Central Science, sections 14.1 14.4. Introduction

More information

Introduction to Spectroscopy: Analysis of Copper Ore

Introduction to Spectroscopy: Analysis of Copper Ore Introduction to Spectroscopy: Analysis of Copper Ore Introduction The goal of this lab is to determine the unknown concentration of two different copper solution samples, taken from fictitious mining sites

More information

Design Considerations for a Variable Angle Absolute Reflectance Accessory For the LAMBDA 950/850/650 UV/Vis/NIR and UV/Vis Spectrophotometers

Design Considerations for a Variable Angle Absolute Reflectance Accessory For the LAMBDA 950/850/650 UV/Vis/NIR and UV/Vis Spectrophotometers Design Considerations for a Variable Angle Absolute Reflectance Accessory For the LAMBDA 950/850/650 UV/Vis/NIR and UV/Vis Spectrophotometers UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N

More information

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1 Experiment 13H 08/03/2017 AHRM THE REACTION OF RED FOOD COLOR WITH BLEACH 1 PROBLEM: Determine the rate law for the chemical reaction between FD&C Red Dye #3 and sodium hypochlorite. LEARNING OBJECTIVES:

More information

ab Alkaline Phosphatase Assay Kit (Luminometric)

ab Alkaline Phosphatase Assay Kit (Luminometric) Version 1 Last updated 24 April 2018 ab233466 Alkaline Phosphatase Assay Kit (Luminometric) For the measurement of Alkaline Phosphatase activity in solutions and cells. This product is for research use

More information

ab MDA Assay Kit (competitive ELISA)

ab MDA Assay Kit (competitive ELISA) Version 1 Last updated 30 August 2018 ab238537 MDA Assay Kit (competitive ELISA) For the quantitative measurement of MDA in protein samples such as purified protein, plasma, serum and cell lysate. This

More information

The Australia Telescope. The Australia Telescope National Facility. Why is it a National Facility? Who uses the AT? Ray Norris CSIRO ATNF

The Australia Telescope. The Australia Telescope National Facility. Why is it a National Facility? Who uses the AT? Ray Norris CSIRO ATNF The Australia Telescope National Facility The Australia Telescope Ray Norris CSIRO ATNF Why is it a National Facility? Funded by the federal government (through CSIRO) Provides radio-astronomical facilities

More information

MOLEBIO LAB #4: Using a Spectrophotometer

MOLEBIO LAB #4: Using a Spectrophotometer Introduction: Spectrophotometry MOLEBIO LAB #4: Using a Spectrophotometer Many kinds of molecules interact with or absorb specific types of radiant energy in a predictable fashion. For example, when while

More information

CONFOCHECK. Innovation with Integrity. Infrared Protein Analysis FT-IR

CONFOCHECK. Innovation with Integrity. Infrared Protein Analysis FT-IR CONFOCHECK Infrared Protein Analysis Innovation with Integrity FT-IR CONFOCHECK: FT-IR System for Protein Analytics FT-IR Protein Analysis Infrared spectroscopy measures molecular vibrations due to the

More information

International Standardization for Measurement and Characterization of Nanomaterials

International Standardization for Measurement and Characterization of Nanomaterials 2012/SOM2/SCSC/CON/008 International Standardization for Measurement and Characterization of Nanomaterials Submitted by: Japan 9 th Conference on Standards and Conformance in the Field of Innovations and

More information

Application Note TABLE OF CONTENTS

Application Note TABLE OF CONTENTS Application Note 25 Making optical density measurements automatically corrected to a 1 cm pathlength in the SPECTRAmax PLUS microplate spectrophotometer (The three P s of Pathcheck : Principle, Procedures

More information

AN2970 Application note

AN2970 Application note Application note Principles of capacitive touch and proximity sensing technology The objective of this document is to present the principles of capacitive sensing and charge transfer used in STMicroelectronics

More information

Introduction to Spectroscopy: Analysis of Copper Ore

Introduction to Spectroscopy: Analysis of Copper Ore Introduction to Spectroscopy: Analysis of Copper Ore Using a Buret and Volumetric Flask: 2.06 ml of solution delivered 2.47 ml of solution delivered 50.00 ml Volumetric Flask Reading a buret: Burets are

More information

The DialPath Solution an Easier Way to Analyze Liquids By FTIR. Yanqia Wang, Ph.D. Application Engineer - FTIR Agilent Technologies May 7, 2015

The DialPath Solution an Easier Way to Analyze Liquids By FTIR. Yanqia Wang, Ph.D. Application Engineer - FTIR Agilent Technologies May 7, 2015 The DialPath Solution an Easier Way to Analyze Liquids By FTIR Yanqia Wang, Ph.D. Application Engineer - FTIR Agilent Technologies May 7, 2015 Outline Background of FTIR Transmission Technology Traditional

More information

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III) Chem 1B Saddleback College Dr. White 1 Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III) Objectives To use spectroscopy to relate the absorbance of a colored solution to its concentration. To

More information

Degree of labeling (DOL)

Degree of labeling (DOL) Degree of labeling (DOL) A written explanation for the determination of the degree of labeling when using NuLink reagents. The DOL is the average number of reagent molecules that have been covalently attached

More information

ApplicationNOTE ULTRA TRACE ANALYSIS OF DIOXINS AND FURANS IN HUMAN ADIPOSE TISSUE USING SFE-LC EXTRACTION/CLEANUP AND THE WATERS AUTOSPEC ULTIMA NT

ApplicationNOTE ULTRA TRACE ANALYSIS OF DIOXINS AND FURANS IN HUMAN ADIPOSE TISSUE USING SFE-LC EXTRACTION/CLEANUP AND THE WATERS AUTOSPEC ULTIMA NT Introduction Legislative limits for dioxins and furans are ever decreasing 1, providing a greater challenge for the dioxin analyst. The extraction and clean-up processes used by the analyst can be imperative

More information

SPECTROPHOTOMETERS. Visible and UV-Visible

SPECTROPHOTOMETERS. Visible and UV-Visible Visible and UV-Visible The Manifacturer ONDA Spectrophotometers are manifactured by a company with over than ten years of experience in development of UV/Vis instruments, Single beam and Double beam, with

More information

IEEE Transactions on Image Processing EiC Report

IEEE Transactions on Image Processing EiC Report IEEE Transactions on Image Processing EiC Report Clem Karl 9/23/2013 Appreciation of AE Effort AEs are on the front lines of TIP AE effort makes the journal what it is Thank you! 2 Outline TIP Overview

More information

ab Maleimide Quantification Kit - Colorimetric

ab Maleimide Quantification Kit - Colorimetric ab112140 Maleimide Quantification Kit - Colorimetric Instructions for Use For detecting maleimide in physiology solutions spectrophotometrically. This product is for research use only and is not intended

More information

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Thunyarat Phesatcha, Suparerk Tukkeeree, Jeff Rohrer 2 Thermo

More information

Introduction to Spectroscopy: Analysis of Copper Ore

Introduction to Spectroscopy: Analysis of Copper Ore Introduction to Spectroscopy: Analysis of Copper Ore Thousands of years ago, copper was abundant enough in quantity that it could be found on the Earth s surface. Prospecting for copper then was relatively

More information

MORE LIGHTS, COLOR, ABSORPTION!

MORE LIGHTS, COLOR, ABSORPTION! Name Partner(s) Section Date MORE LIGHTS, COLOR, ABSORPTION! PRE-LAB QUERIES 1. The terms absorption and transmittance are often used when describing the interaction of light with matter. Explain what

More information

Rate law Determination of the Crystal Violet Reaction Using the Isolation Method

Rate law Determination of the Crystal Violet Reaction Using the Isolation Method Rate law Determination of the Crystal Violet Reaction Using the Isolation Method Introduction A common challenge in chemical kinetics is to determine the rate law for a reaction with multiple reactants.

More information

ab Uricase Assay Kit (Fluorometric) 1

ab Uricase Assay Kit (Fluorometric) 1 Version 1 Last updated 26 April 2018 ab234042 Uricase Assay Kit (Fluorometric) For the measurement of Uricase activity in various biological samples/preparations. This product is for research use only

More information

of mass spectrometry

of mass spectrometry Thermo Scientific 253 Ultra High resolution isotope ratio MS Discover a new world of mass spectrometry Paleoclimatology Atmospheric science Biogeochemistry Petrology Discover the isotopic anatomy of molecules

More information

Spectrophotometric Determination of an Equilibrium Constant

Spectrophotometric Determination of an Equilibrium Constant Spectrophotometric Determination of an Equilibrium Constant v021214 Objective To determine the equilibrium constant (K c ) for the reaction of iron (III) ion with thiocyanate (SCN - ) to form the thiocyanatoiron(iii)

More information

Experiment 1 (Part A): Plotting the Absorption Spectrum of Iron (II) Complex with 1,10- Phenanthroline

Experiment 1 (Part A): Plotting the Absorption Spectrum of Iron (II) Complex with 1,10- Phenanthroline Experiment (Part A): Plotting the Absorption Spectrum of Iron (II) Complex with,0- Phenanthroline Background The first step of an analytical spectrophotometric procedure for quantitative determination

More information

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS Daniel Kutscher, 1 Jörg Bettmer, 2 Torsten Lindemann, 1 Shona McSheehy-Ducos, 1 Lothar Rottmann 1 1 Thermo Fisher Scientific, Germany 2

More information

OCTOBER Almond Industry Position Report Crop Year /01-10/31 Kernel Wt /01-10/31 Kernel Wt.

OCTOBER Almond Industry Position Report Crop Year /01-10/31 Kernel Wt /01-10/31 Kernel Wt. Almond Industry Position Report OCTOBER 2018 2018-2019 Crop Year Salable Forecast Million Lbs. 2018-2019 08/01-10/31 Kernel Wt. 2017-2018 08/01-10/31 Kernel Wt. % Change 1. Carryin August 1, 2018 359.0

More information

Keysight Technologies Measuring Substrate-Independent Young s Modulus of Low-k Films by Instrumented Indentation. Application Note

Keysight Technologies Measuring Substrate-Independent Young s Modulus of Low-k Films by Instrumented Indentation. Application Note Keysight Technologies Measuring Substrate-Independent Young s Modulus of Low-k Films by Instrumented Indentation Application Note Introduction In digital circuits, insulating dielectrics separate the conducting

More information