Luminescence Spectroscopy Excitation is very rapid (10-15 s). Vibrational relaxation is a non-radiational process. It involves vibrational levels of
|
|
- Milton Jones
- 5 years ago
- Views:
Transcription
1 Luminescence Spectroscopy Excitation is very rapid (10-15 s). Vibrational relaxation is a non-radiational process. It involves vibrational levels of the same electronic state. The excess of vibrational energy is released by the excited molecule in the form of thermal or vibrational motion to the solvent molecules. It takes between and s. Internal conversion is a non-radiational process. It is the crossover between states of the same multiplicity. It occurs when the high vibrational levels of the lower electronic state overlap with the low vibrational levels of the high electronic state. It can occur between excited states (S 2 and S 1 ) or between excited and ground states (S 1 and S 0 ). Between excited states is rapid (10-12 s). Fluorescence is a radiational transition between S 1 and S 0. It occurs from the ground vibrational level of S 1 to various vibrational levels in S 0. It requires to 10-6 s to occur. External conversion is a nonradiative process in which excited states transfer their excess energy to other species, such as solvent or solute molecules. Intersystem crossing is a non-radiative process between electronic states of different multiplicities. It requires a change in electronic spin and, therefore, it has a much lower probability to occur than spin-allowed transitions. Because the time scale is similar to the one for fluorescence ( s), it competes with fluorescence for the deactivation of the S 1 state Phosphorescence is a radiational deactivation process between electronic states of different multiplicity, i.e. T 1 to S 0. It usually takes between 10-4 to 10s to occur because the process is spin forbidden. 139
2 Intensity Intensity of fluorescence (or phosphorescence) emission as a function of fluorophor (or phosphor) concentration The power of fluorescence (or phosphorescence) radiation (F or P) is proportional to the radiant power of the excitation beam that is absorbed by the system: F = K (P 0 P) (1) The transmittance of the sample is given by Beer s law: P/P 0 = 10 -ebc (2) Re-arranging Eq. 1: F = K P 0 (1 P / P 0 ) Substituting Eq. 2 above: F = K (1 10 -ebc ) (3) The exponential term in Eq. 3 can be expanded as a Maclaurin series to: F = K.P 0.[2.303ebc (2.303ebc) 2 + (2.303ebc) 3 - ] (4) 2! 3! For diluted solutions, i.e ebc < 0.05, all of the subsequent terms in the brackets become negligible with respect to the first, so: F = K.P ebc or F = P 0.K.ebc (5) A plot of fluorescence (or phosphorescence) intensity as a function of concentration should be linear up to a certain concentration. Three are the main reasons for lack of linearity at high concentrations: a) 2.303ebc > 0.05 b) self-quenching c) self-absorption P 0 P F Detector LDR Concentration 140
3 Fluorescence quantum yield The slope of the calibration curve is equal to P 0.K.ebc. K is also known as the fluorescence quantum yield (f F ). The fluorescence quantum yield is the ratio between the number of photons emitted as fluorescence and the number of photons absorbed: f F = # of fluorescence photons # of absorbed photons Fluorescence quantum yields may vary between zero and unity: 0 < f F < 1 The higher the quantum yield, the stronger the fluorescence emission. In terms of rate constants, the fluorescence quantum yield is expressed as follows: f F = kf kf + ki + kec + kic + kpd +kd kf kic kec ki Consider a dilute solution of a fluorescent species A whose concentration is [A] (in mol.l -1 ). A very short pulse of light at time 0 will bring a certain number of molecules A to the S 1 excited state by absorption of photons: A + hν A* The excited molecules then return to S 0, either radiatively or non-radiatively, or undergo intersystem crossing. As in classical kinetics, the rate of disappearance of excited molecules is expressed by the following differential equation: -d[a*] / dt = (kf + knr) [A*] knr = ki + kec + kic + kpd + kd are the rate constants of competing processes that ultimately reduce the intensity of fluorescence. k units = s
4 Fluorescence lifetime Integration of equation: -d[a*] / dt = (kf + knr) [A*] yields the time evolution of the concentration of excited molecules [A*]. Let [A*] 0 be the concentration of excited molecules at time 0 resulting from the pulse light excitation. Integration leads to: [A*] = [A*] 0 exp(-t / t) where t is the lifetime of the excited state S 1. The fluorescence lifetime is then given by: t = 1 / kf + knr Typical fluorescence lifetimes are in the ns range. Typical phosphorescence lifetimes are in the ms to s range. The fluorescence (or phosphorescence lifetime) is the time needed for the concentration of excited molecules to decrease to 1/e of its original value. The fluorescence lifetime correlates to the fluorescence quantum yield as follows: f f = kf. t 142
5 Excitation and emission spectra Excitation spectra appear in the same wavelength region as absorption spectra. However, it is important to keep in mind that absorption spectra are not the same as excitation spectra. Fluorescence and phosphorescence spectra appear at longer wavelength regions than excitation spectra. Phosphorescence spectra appear at longer wavelength regions than fluorescence spectra. In some cases, it is possible to observe vibrational transitions in room-temperature fluorescence spectra. There are several parameters (many wavelengths and two lifetimes) for compound identification. 143
6 Fluorescence and Structure General rule: Most fluorescent compounds are aromatic. An increase in the extent of the p-electron system (i.e. the degree of conjugation) leads to a shift of the absorption and fluorescence spectra to longer wavelengths and an increase in the fluorescence quantum yield. Example: Aromatic hydrocarbon Naphthalene Anthracene Naphthacene Fluorescence ultraviolet blue green Pentacene red The lowest-lying transitions of aromatic hydrocarbons are of the p p* type, which are characterized by high molar absorption coefficients and relatively high quantum fluorescence quantum yields. No fluorescence Fluorescence 144
7 Substituted Aromatic Hydrocarbons The effect of substituents on the fluorescence characteristics of aromatic hydrocarbons varies with the type of substituent. Heavy atoms: the presence of heavy atoms (e.g. Br, I, etc.) results in fluorescence quenching (internal heavy atom effect) because of the increase probability of intersystem crossing (ISC). Electron- donating : -OH, -OR, -NH 2, -NHR, -NR 2 This type of substituent generally induces and increase in the molar absorption coefficient and a shift in both absorption and fluorescence spectra. Electron-withdrawing substituents: carbonyl and nitro-compounds Carbonyl groups: There is no general rule. Their effect depends on the position of the substituent group in the aromatic ring. Nitro groups: No detectable fluorescence. 145
8 Rigidity Rigidity usually enhances fluorescence emission ph of solution Additional resonance forms lead to a more stable first excited state 146
9 Quenching: nonradiative energy transfer from an excited species to other molecules. Quenching results in deactivation of S 1 without the emission of radiation causing a decrease in fluorescence intensity. Types of quenching: dynamic, static and others. Dynamic quenching: or collisional quenching requires contact between the excited species and the quenching agent. Dynamic quenching is a diffusion controlled process. As such, its rate depends on the temperature and viscosity of the sample. High temperatures and low viscosity promote dynamic quenching. For dynamic quenching with a single quencher, the Stern-Volmer expression is valid: F 0 /F = 1 + K q [Q] Where F 0 and F are the fluorescence intensities in the absence and the presence of quencher, respectively. [Q] is the concentration of quencher (mols.l -1 ) and K q is the Stern-Volmer quenching constant. The Stern-Volmer constant is defined as: K q = kq / kf + ki + kic where kq is the rate constant for the quenching process (diffusion controlled). Static Quenching: the quencher and the fluorophor in the ground state form a complex. The complex is no fluorescent (dark complex). The Stern-Volmer equation is still valid but Kq in this case is the equilibrium constant for complex formation: A + Q <=> AQ. In static quenching, the fluorescence lifetime of the fluorophor is not affected. In dynamic quenching, the fluorescence lifetime of the fluorophor is affected. So, lifetime can be used to distinguish between dynamic and static quenching. Quenching Fluorescence quenching of quinine sulfate as a function of chloride concentration Oxygen sensors O 2 is paramagnetic, i.e. its natural electronic configuration is the triplet state. When O 2 interacts with the fluorophor, it promotes conversion and deactivation of excited fluorophor molecules. Upon interaction, O 2 goes into the singlet state (diamagnetic). 147
10 Instrumentation Typical spectrofluorometer (or spectrofluuorimeter) configuration Recording excitation and emission spectra Recording synchronous fluorescence spectra 148
11 Spectrofluorimeter capable to correct for source wavelength dependence Spectrofluorimeter with the ability to record total luminescence 149
12 Measuring phosphorescence Using a spectrofluorimeter with a continuous excitation source: be careful with fluorescence and second order emission! Rotating-can phosphoroscope: good-old approach to discriminate against fluorescence. A. True representation B. Approximate representation t e = exposure time; t d = shutter delay time; t t = shutter transit time; t C = time for one cycle of excitation and observation; t E = t e + t t ; t D = t d + t t Spectrofluorimeters with a pulsed source: employ a gated PMT for fluorescence discrimination. Excitation source on => PMT off = fluorescence decays Excitation source off => PMT on phosphorescence measurement 150
13 Introduction to Chromatographic Separations Analysis of complex samples usually involves previous separation prior to compound determination. Two main separation methods instrumentation are available: based on Chromatography Electrophoresis Chromatography is based on the interaction of chemical species with a mobile phase (MP) and a stationary phase (SP). The MP and the SP are immiscible. The sample is transported by the MP. The interaction of species with the MP and the SP separates chemical species in zones or bands. The relative chemical affinity of chemical species with the MP and the SP dictates the time the species remain in the SP. MP + sample SP Detector Two general types of chromatographic techniques exist: Planar: flat SP, MP moves through capillary action or gravity Column: tube of SP, MP moves through gravity or pressure. 151
14 Classification of Chromatographic Methods Chromatographic methods can be classified on the type of MP and SP and the kinds of equilibrium involved in the transfer of solutes between phases: Column 1: Type of MP Column 2: Type of MP and SP Column 3: SP Column 4: Type of equilibrium Concentration profiles of solute bands A and B at two different times in their migration down the column. 152
15 Migration Rates of Solutes Two-component chromatogram illustrating two methods for improving separation: (a) Original chromatogram with overlapping peaks; (b) improvement brought about an increase in band separation; (c) improvement brought about by a decrease in the widths. Distribution constant or partition ratio or partition coefficient: It describes the partition equilibrium of an analyte between the SP and the MP. If K = K c and SP = S and MP = M K c = c S / C M = n S / V S n M / V M Where V S and V M are the volumes of the two phases and n S and n M are the moles of A in SP and MP, respectively. 153
16 Retention Time Retention time is a measured quantity. From the figure: t M = time it takes a non-retained species (MP) to travel through the column = dead or void time. t R = retention time of analyte. The analyte has been retained because it spends a time t S in the SP. The retention time is then given by: t R = t S + t M The average migration rate (cm/s) of the solute through the column is: Where L is the length of the column. The average linear velocity of the MP molecules is: 154
17 Relationship Between Retention Time and Distribution Constant The Rate of Solute Migration: The Retention Factor Retention factor = capacity factor = k A = k A. However: k A K C or k A K A Where V S and V M are the volumes of SP and MP in the column. Knowing that: => An equation can be derived to obtain K C of A (K A ) as a function of experimental parameters. => 155
18 Band Broadening and Column Efficiency The shape of an analyte zone eluting from a chromatographic column follows a Gaussian profile. Some molecules travel faster than the average. The time it takes them to reach the detector is: t R Dt. Some molecules travel slower than the average molecule. The time it takes them to reach the detector is: t R + Dt The selectivity factor can be measured from the chromatogram. Selectivity factors are always greater than unity, so B should be always the compound with higher affinity by the SP. Selectivity factors are useful parameters to calculate the resolving power of a column. So, the Gaussian provides an average retention time (most frequent time) and a time interval for the total elution of an analyte from the column. The magnitude of Dt depends on the width of the peak. 156
19 Considering that: v_ = L / t R => L = v_ x or L ± DL = v_ x [t R ± Dt] The equation above correlates chromatographic peaks with Gaussian profiles to the length of the column. Both Dt and DL correspond to the standard deviation of a Gaussian peak: Dt DL s If s = Dt, s units are in minutes. If s = DL, s units are in cm. The width of a peak is a measure of the efficiency of a column. The narrower the peak, the more efficient is the column. => The efficiencies of chromatographic columns can be compared in terms of number of theoretical plates. 157
20 The Plate Theory The plate theory supposes that the chromatographic column contains a large number of separate layers, called theoretical plates. Separate equilibrations of the sample between the stationary and mobile phase occur in these "plates". The analyte moves down the column by transfer of equilibrated mobile phase from one plate to the next. It is important to remember that theoretical plates do not really exist. They are a figment of the imagination that helps us to understand the processes at work in the column. As previously mentioned, theoretical plates also serve as a figure of merit to measuring column efficiency, either by stating the number of theoretical plates in a column (N) or by stating the plate height (H); i.e. the Height Equivalent to a Theoretical Plate. The number of theoretical plates is given by: N = L / H If the length of the column is L, then the HETP is: H = s 2 / L Note: For columns with the same length (same L): The smaller the H, the narrower the peak. The smaller the H, the larger the number of plates. => column efficiency is favored by small H and large N. It is always possible to using a longer column to improve separation efficiency. 158
21 Experimental Evaluation of H and N Calling s in time units (minutes or seconds) as t: => s / t = cm / s Considering that: v_ = L / t R = cm /s We can write: L / t R = s / t or t = s = s L / t R v_ If the chromatographic peak is Gaussian, approximately 96% of its area is included within ± 2s. This area corresponds to the area between the two tangents on the two sides of the chromatographic peak. The width of the peak at its base (W) is then equal to: W = 4t in time units or W = 4s in length units. Substituting t = W / 4 in the equation above we obtain: s = LW / 4t R Substituting s = W / 4 in the same equation we obtain: t = Wt R / 4L Substituting s = LW / 4t R in the HEPT equation: H = LW 2 / 16t R 2 Substitution of this equation in N gives: N = 16 (t R / W) 2 These two equations allow one to estimate H and N from experimental parameters. If one considers the peak of the width at the half maximum (W 1/2 ): N = 5.54 (t R / W 1/2 ) 2 In comparing columns, N and H should be obtained with the same compound! 159
22 Kinetic Variables Affecting Column Efficiency The plate theory provides two figures of merit (N and H) for comparing column efficiency but it does not explain band broadening. Table 26-2 provides the variables that affect band broadening in a chromatographic column and, therefore, affect column efficiency. The effect of these variables in column efficiency is best explained by the theory of band broadening. This theory is best represented by the van Deemter equation: H = A + B / u+ C S. u + C M. U where H is in cm and u is the velocity of the mobile phase in cm.s -1. The other terms are explained in Table Table 26-3: f(k) and f (k) are functions of k l and g are constants that depend on the quality of the packing. B is the coefficient of longitudinal diffusion. C S and C M are coefficients of mass transfer in stationary and mobile phase, respectively. Before we try to understand the meaning of the van Deemter equation, a better understanding of chromatographic columns is needed. 160
23 Some Characteristics of Gas Chromatography (GC) Columns Two general types => Open Tubular Columns (OTC) or Capillary Columns => Packed Columns OTC: => Wall Coated Open Tubular (WCOT) Columns => Support Coated Open Tubular (SCOT) Columns => The most common inner diameters for capillary tubes are 0.32 and 0.25mm. Packed Columns: => Glass tubes with 2 to 4mm inner diameter. => Packed with a uniform, finely divided packing material of solid support, coated with a thin layer (005 to 1mm) of liquid stationary phase. Solid Support Material in OTC and Packed Columns Diatomaceous Earth: skeletons of species of single-celled plants that once inhabited ancient lakes and seas. 161
24 Some Characteristics of HPLC Columns Only packed columns are used in HPLC. Current packing consists of porous micro-particles with diameters ranging from 3 to 10mm. The particles are composed of silica, alumina, or an ion-exchange resin. Silica particles are the most common. Thin organic films are chemically or physically bonded to the silica particles. The chemical nature of the thin organic film determines the type of chromatography. SP for normal-phase chromatography Typical microparticle SP for partition chromatography 162
25 Another Look at the van Deemter Equation H = A + B/u + C S u + C M u The multi-path term A or Eddy-Diffusion: => This term accounts for the multitude of pathways by which a molecule (or ion) can find its way through a packed column. A = 2ld p where: l is a geometrical factor that depends on the shape of the particle: 1 l 2. d p is the diameter of the particle. Using packing with spherical particles of small diameters should reduce eddy -diffusion. Injection Detector 163
26 The Longitudinal Diffusion Term B/u: Longitudinal diffusion is the migration of solute from the concentrated center of the band to the more diluted regions on either side of the analyte zone. B/u = 2gD M /u where: g is a constant that depends on the nature of the packing and it varies from 0.6 g 0.8. D M is the diffusion coefficient in the mobile phase. D M a T / m, where T is the temperature and m is the viscosity of the mobile phase. As u infinite, B/u zero. So, the contribution of longitudinal diffusion in the total plate height is only significant at low MP flow rates. Its contribution is potentially more significant in GC than HPLC because of the relatively high column temperatures and low MP viscosity (gas). Initial band Diffusion of the band with time The initial part of the curve is predominantly due to the B/u term. Because the term B/u in GC is larger than HPLC, the overall H in GC is about 10x the overall H in HPLC. 164
27 The Stationary-Phase Mass Transfer Term C S u: C S = mass transfer coefficient in the SP. C S a d f 2 / D S where d f is the thickness of the SP film and D S is the diffusion coefficient in the SP. => Thin-film SP and low viscosity SP (large D S ) provide low mass transfer coefficients in the SP and improve column efficiency. The Mobile-Phase Mass Transfer Term C M u: C M = mass transfer coefficient in the MP. C M a d p 2 / D M where d p is the diameter of packing particles and D M is the diffusion coefficient in the MP. The contribution of mass transfer in the SP and MP on the overall late height depends on the flow velocity of the mobile phase. Both phenomena play a predominant role at high MP flow rates. Liquid SP coated on solid support Cartoon with example of SP mass-transfer in a liquid SP: different degrees of penetration of analyte molecules in the liquid layer of SP lead to band-broadening Porous in silica particle Cartoon with example of MP masstransfer: stagnant pools of MP retained in the porous of silica particles lead to bandbroadening. 165
28 Summary of Methods for Reducing Band Broadening Packed Columns: Most important parameter that affects band broadening is the particle diameter. If the SP is liquid, the thickness of the SP is the most important parameter. Capillary Columns: No packing, so there is no Eddy-diffusion term. Most important parameter that affects band broadening is the diameter of the capillary. Gaseous MP (GC): The rate of longitudinal diffusion can be reduced by lowering the temperature and thus the diffusion coefficient. The effect of temperature is mainly noted at low flow rate velocities where the term B/u is significant. Temperature has little effect on HPLC. Effect of particle diameter in GC Effect of particle diameter in HPLC Note: m = mm 166
29 Optimization of Column Performance Optimization experiments are aimed at either reducing zone broadening or altering relative migration rates of components. The time it takes for chromatographic analysis is also an important parameter that should be optimized without compromising chromatographic resolution. Column Resolution: It is a quantitative measure of the ability of the column to separate two analytes. It can be obtained from the chromatogram with the equation: In terms of retention factors k A and k B for the two solutes, the selectivity factor and the number of theoretical plates of the column: From the last equation we can obtain the number of theoretical plates needed to achieve a given resolution: 167
30 For compounds with similar capacity factors, i.e. k A k B : The time it takes to achieve a separation can be predicted with the formula: Where k = k A + k B / 2 168
31 The General Elution Problem The general elution problem occurs in the separation of mixtures containing compounds with widely different distribution constants. The best solution to the general elution problem is to optimize eluting conditions for each compound during the chromatographic run. In HPLC, this is best accomplished by changing the composition of the mobile phase (Gradient Elution Chromatography). In GC, this is best accomplished by changing the temperature of the column during the chromatographic run. 169
32 Qualitative and Quantitative Analysis in GC and HPLC Both are done with the help of standards. Qualitative analysis, i.e. compound identification is done via retention time. The retention time from the pure standard is compared to the retention time of the analyte in the sample. Note: retention times are experimental parameters and as such are prone to standard deviation. Quantitative analysis is done via the calibration curve method (or external standard method) or the internal standard method. Calibration curve or external standard method: the procedure is the same as usual. The calibration curve can be built plotting the peak height or the peak area versus standard concentration. The same volume of sample was injected in each case, but Sample B has a much smaller peak. Since the t R at the apex of both peaks is 2.85 minutes, this indicates that they are both the same compound, (in this example, acrylamide (ID)). The Area under the peak ( Peak Area Count ) indicates the concentration of the compound. This area value is calculated by the Computer Data Station. Notice the area under the Sample A peak is much larger. In this example, Sample A has 10 times the area of Sample B. Therefore, Sample A has 10 times the concentration, (10 picograms) as much acrylamide as Sample B, (1 picogram). Note, there is another peak, (not identified), that comes out at 1.8 min. in both samples. Since the area counts for both samples are about the same, it has the same concentration in both samples. 170
33 Internal Standards An internal standard is a known amount of a compound that is added to the unknown. The signal from analyte is compared to the signal from the standard to find out how much analyte is present. This method compensates for instrumental response that varies slightly from run to run and deteriorates reproducibility considerably. This is the case of mobile phase flow rate variations in chromatographic analysis. Internal standards are also desirable in cases where the possibility of loosing sample during analysis exists. This is the case of sample separation in the chromatographic column. The internal standard should be chosen according to the analyte. Their chemical behavior with regards to the SP and MP should be similar. How to use an internal standard?: a mixture with the same known amount of standard and analyte is prepared to measure the relative response of the detector for the two species. The factor (F) is obtained from the relative response of the detector. Once the relative response of the detector has been found, the analyte concentration is calculated according to the formula: Area of analyte signal Concentration of analyte = F x Area of standard signal Concentration of standard A known amount of standard is added to the unknown X. The relative response is measured to obtain the detector s response factor F. 171
34 Example of Internal Standards In a preliminary experiment, a solution containing M X and 0.066M S gave peak areas of A X = 423 and A S = 347. Note that areas are measured in arbitrary units by the instrument s computer. To analyze the unknown, 10.0mL of 0.146M S were added to 10.0mL of unknown, and the mixture was diluted to 25.0mL in a volumetric flask. This mixture gave a chromatogram with peak areas A X = 553 and A S = 582. Find the concentration of X in the unknown. First use the standard mixture to find the response factor: A X / [x] = F x {A S / [S]} Standard mixture: 423 / = F {347 / } => F = In the mixture of unknown plus standard, the concentration of S is: [S] = (0.146M)(10.0mL / 25.0mL) = M where: 10.0mL / 25.0mL is the dilution factor Using the known response factor and S concentration of the diluted sample in the equation above: 553 / [X] = (582 / ) =>[X] = M 172
35 Liquid Chromatography Size exclusion or gel: polystyrene-divinylbenzene Silica with various porous sizes 173
36 Instrumentation 174
37 Pumping Systems Pumping systems that allow to change the MP composition during the chromatographic run provide better separation of compounds with wide range of k factors. 175
38 Detectors UV-VIS absorption cell for HPLC 176
Introduction to Chromatographic Separations
Introduction to Chromatographic Separations Analysis of complex samples usually involves previous separation prior to compound determination. Two main separation methods instrumentation are available:
More informationChapter 15 Molecular Luminescence Spectrometry
Chapter 15 Molecular Luminescence Spectrometry Two types of Luminescence methods are: 1) Photoluminescence, Light is directed onto a sample, where it is absorbed and imparts excess energy into the material
More informationAnalytical Chemistry
Analytical Chemistry Chromatographic Separations KAM021 2016 Dr. A. Jesorka, 6112, aldo@chalmers.se Introduction to Chromatographic Separations Theory of Separations -Chromatography Terms Summary: Chromatography
More informationIntroduction to Chromatography
Introduction to Chromatography Dr. Sana Mustafa Assistant Professor Department of Chemistry, Federal Urdu University of Arts, Science & Technology, Karachi. What is Chromatography? Derived from the Greek
More informationChromatography. Gas Chromatography
Chromatography Chromatography is essentially the separation of a mixture into its component parts for qualitative and quantitative analysis. The basis of separation is the partitioning of the analyte mixture
More informationHPLC Background Chem 250 F 2008 Page 1 of 24
HPLC Background Chem 250 F 2008 Page 1 of 24 Outline: General and descriptive aspects of chromatographic retention and separation: phenomenological k, efficiency, selectivity. Quantitative description
More informationA Fluorometric Analysis of Quinine in Tonic Water
A Fluorometric Analysis of Quinine in Tonic Water Introduction In this Laboratory Exercise, we will determine the amount of quinine in Tonic Water using a fluorometric analysis. Fluorescence Spectroscopy
More informationMolecular Luminescence Spectroscopy
Molecular Luminescence Spectroscopy In Molecular Luminescence Spectrometry ( MLS ), molecules of the analyte in solution are excited to give a species whose emission spectrum provides information for qualitative
More informationIntroduction to Chromatographic Separations (Chapter 1) Many determinations involve separation followed by analysis chromatography electrophoresis
Introduction to Chromatographic Separations (Chapter 1) Many determinations involve separation followed by analysis chromatography electrophoresis Chromatography: sample transported by mobile phase electrostatic
More informationLABORATORY OF ELEMENTARY BIOPHYSICS
LABORATORY OF ELEMENTARY BIOPHYSICS Experimental exercises for III year of the First cycle studies Field: Applications of physics in biology and medicine Specialization: Molecular Biophysics Fluorescence
More informationChromatographic Analysis
Chromatographic Analysis Distribution of Analytes between Phases An analyte is in equilibrium between the two phases [S 1 ] [S 2 ] (in phase 1) (in phase 2) AS [S2 ] K 2 A S [S1 ] 1 AS, A 1 S Activity
More informationSeparation Methods Based on Distributions in Discrete Stages (02/04/15)
Separation Methods Based on Distributions in Discrete Stages (02/04/15) 1. Chemical Separations: The Big Picture Classification and comparison of methods 2. Fundamentals of Distribution Separations 3.
More informationMolecular Luminescence. Absorption Instrumentation. UV absorption spectrum. lg ε. A = εbc. monochromator. light source. Rotating mirror (beam chopper)
Molecular Luminescence Absorption Instrumentation light source I 0 sample I detector light source Rotating mirror (beam chopper) motor b sample I detector reference I 0 UV absorption spectrum lg ε A =
More informationRemember - Ions are more soluble in water than in organic solvents. - Neutrals are more soluble in organic solvents than in water.
IN-CLASS PROBLEMS SEPARATION SCIENCE CROMATOGRAPHY UNIT Thomas Wenzel, Bates College In-class Problem Set - Extraction 1. Devise a way to separate the materials in the following sample by performing an
More informationSinglet. Fluorescence Spectroscopy * LUMO
Fluorescence Spectroscopy Light can be absorbed and re-emitted by matter luminescence (photo-luminescence). There are two types of luminescence, in this discussion: fluorescence and phosphorescence. A
More informationWhat is Chromatography?
What is Chromatography? Chromatography is a physico-chemical process that belongs to fractionation methods same as distillation, crystallization or fractionated extraction. It is believed that the separation
More informationCh 313 FINAL EXAM OUTLINE Spring 2010
Ch 313 FINAL EXAM OUTLINE Spring 2010 NOTE: Use this outline at your own risk sometimes a topic is omitted that you are still responsible for. It is meant to be a study aid and is not meant to be a replacement
More informationChemistry Instrumental Analysis Lecture 26. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 26 Rate Theory Focuses on the contributions of various kinetic factors to zone or band broadening. Column Dispensivity, H, is assumed to be the sum of the individual
More informationWhat the Einstein Relations Tell Us
What the Einstein Relations Tell Us 1. The rate of spontaneous emission A21 is proportional to υ 3. At higher frequencies A21 >> B(υ) and all emission is spontaneous. A 21 = 8π hν3 c 3 B(ν) 2. Although
More informationInformation given in these slides are, either in part or all, recollection from the followings:
Information given in these slides are, either in part or all, recollection from the followings: http://bionmr.unl.edu/courses/chem421-821/lectures/chapter-2... http://faculty.atu.edu/abhuiyan/course/chem
More informationInstrumental Analysis II Course Code: CH3109. Chromatographic &Thermal Methods of Analysis Part 1: General Introduction. Prof. Tarek A.
Instrumental Analysis II Course Code: CH3109 Chromatographic &Thermal Methods of Analysis Part 1: General Introduction Prof. Tarek A. Fayed What is chemical analysis? Qualitative analysis (1) Chemical
More informationChromatography and other Separation Methods
Chromatography and other Separation Methods Probably the most powerful class of modern analytical methods for analyzing mixture of components---and even for detecting a single component in a complex mixture!
More informationLuminescence transitions. Fluorescence spectroscopy
Luminescence transitions Fluorescence spectroscopy Advantages: High sensitivity (single molecule detection!) Measuring increment in signal against a dark (zero) background Emission is proportional to excitation
More informationChromatographic Separation
What is? is the ability to separate molecules using partitioning characteristics of molecule to remain in a stationary phase versus a mobile phase. Once a molecule is separated from the mixture, it can
More informationCHROMATOGRAPHY. The term "chromatography" is derived from the original use of this method for separating yellow and green plant pigments.
CHROMATOGRAPHY The term "chromatography" is derived from the original use of this method for separating yellow and green plant pigments. THEORY OF CHROMATOGRAPHY: Separation of two sample components in
More informationChem 454 instrumental Analysis Exam 1 February 6 th, 2008
Chem 454 instrumental Analysis Exam 1 February 6 th, 2008 1 Name: 1] A glass electrode was immersed into a solution of ph 4.33 gave a response of 677.1 mv. This electrode was used to measure a sample solution
More information3) In CE separation is based on what two properties of the solutes? (3 pts)
Final Exam Chem 311 Fall 2002 December 16 Name 1) (3 pts) In GC separation is based on the following two properties of the solutes a) polarity and size b) vapor pressure and molecular weight c) vapor pressure
More informationLiquid Chromatography
Liquid Chromatography 1. Introduction and Column Packing Material 2. Retention Mechanisms in Liquid Chromatography 3. Method Development 4. Column Preparation 5. General Instrumental aspects 6. Detectors
More informationCHEM 429 / 529 Chemical Separation Techniques
CHEM 429 / 529 Chemical Separation Techniques Robert E. Synovec, Professor Department of Chemistry University of Washington Lecture 1 Course Introduction Goal Chromatography and Related Techniques Obtain
More informationChapter 26. An Introduction to Chromatographic Separations. Chromatography
Chapter 26 An Introduction to Chromatographic Separations Chromatography 1 Chromatography-Model as Extraction Chromatography-Model as Extraction 2 Chromatography Planar Chromatography-Types paper chromatography
More informationOpen Column Chromatography, GC, TLC, and HPLC
Open Column Chromatography, GC, TLC, and HPLC Murphy, B. (2017). Introduction to Chromatography: Lecture 1. Lecture presented at PHAR 423 Lecture in UIC College of Pharmacy, Chicago. USES OF CHROMATOGRAPHY
More informationChapter 26: An Introduction to Chromatographic Separations
Chapter 26: An Introduction to Chromatographic Separations Column Chromatography Migration Rates Distribution Contstants Retention Times Selectivity Factor Zone Broadening & Column Efficiency Optimizing
More informationAbstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.
Chromatography Primer Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments. At its heart, chromatography is a technique
More informationChem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS
Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS 1. List two advantages of temperature programming in GC. a) Allows separation of solutes with widely varying retention factors in a reasonable
More informationPHOTOCHEMISTRY NOTES - 1 -
- 1 - PHOTOCHEMISTRY NOTES 1 st Law (Grotthus-Draper Law) Only absorbed radiation produces chemical change. Exception inelastic scattering of X- or γ-rays (electronic Raman effect). 2 nd Law (Star-Einstein
More informationReflection = EM strikes a boundary between two media differing in η and bounces back
Reflection = EM strikes a boundary between two media differing in η and bounces back Incident ray θ 1 θ 2 Reflected ray Medium 1 (air) η = 1.00 Medium 2 (glass) η = 1.50 Specular reflection = situation
More informationHigh Performance Liquid Chromatography
High Performance Liquid Chromatography What is HPLC? It is a separation technique that involves: Injection of small volume of liquid sample Into a tube packed with a tiny particles (stationary phase).
More informationSeparations---Chromatography and Electrophoresis
Separations---Chromatography and Electrophoresis Chromatography--one of most diverse and important analytical methods-- Used initially primarily to purify species With advent of sensitive detectors---now
More informationFluorescence (Notes 16)
Fluorescence - 2014 (Notes 16) XV 74 Jablonski diagram Where does the energy go? Can be viewed like multistep kinetic pathway 1) Excite system through A Absorbance S 0 S n Excite from ground excited singlet
More informationCHAPTER 1. Introduction, Chromatography Theory, and Instrument Calibration
1 1 1 1 1 1 CHAPTER 1 Introduction, Chromatography Theory, and Instrument Calibration 1.1 Introduction Analytical chemists have few tools as powerful as chromatography to measure distinct analytes in complex
More informationChemistry Instrumental Analysis Lecture 3. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 3 Quantum Transitions The energy of a photon can also be transferred to an elementary particle by adsorption if the energy of the photon exactly matches the
More informationHigh Pressure/Performance Liquid Chromatography (HPLC)
High Pressure/Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) is a form of column chromatography that pumps a sample mixture or analyte in a solvent (known as the
More informationLEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT. Thomas Wenzel, Bates College. In-class Problem Set Extraction.
LEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT Thomas Wenzel, Bates College In-class Problem Set Extraction Problem #1 1. Devise a scheme to be able to isolate organic acids, bases
More informationChapter 17: Fundamentals of Spectrophotometry
Chapter 17: Fundamentals of Spectrophotometry Spectroscopy: the science that deals with interactions of matter with electromagnetic radiation or other forms energy acoustic waves, beams of particles such
More information1. Photoreduction of Benzophenone in 2-Propanol
1. Photoreduction of Benzophenone in 2-Propanol Topic: photochemistry, photophysics, kinetics, physical-organic chemistry Level: undergraduate physical chemistry Time: 2 x 2 hours (separated by ~24 hours)
More informationPRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC
PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC CHROMATOGRAPHY Laboratory technique for the Separation of mixtures Chroma -"color" and graphein
More informationAnalytical Technologies in Biotechnology Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee
Analytical Technologies in Biotechnology Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Module - 3 Chromatographic methods Lecture - 2 Basic Concepts in Chromatography
More informationCHEM Outline (Part 15) - Luminescence 2013
CHEM 524 -- Outline (Part 15) - Luminescence 2013 XI. Molecular Luminescence Spectra (Chapter 15) Kinetic process, competing pathways fluorescence, phosphorescence, non-radiative decay Jablonski diagram
More informationCHAPTER 13 Molecular Spectroscopy 2: Electronic Transitions
CHAPTER 13 Molecular Spectroscopy 2: Electronic Transitions I. General Features of Electronic spectroscopy. A. Visible and ultraviolet photons excite electronic state transitions. ε photon = 120 to 1200
More information高等食品分析 (Advanced Food Analysis) I. SPECTROSCOPIC METHODS *Instrumental methods: 1. Spectroscopic methods (spectroscopy): a) Electromagnetic radiation
*Instrumental methods: 1. Spectroscopic methods (spectroscopy): a) Electromagnetic radiation (EMR): γ-ray emission X-Ray absorption, emission, fluorescence and diffraction Vacuum ultraviolet (UV) absorption
More informationCourse goals: Course goals: Lecture 1 A brief introduction to chromatography. AM Quality parameters and optimization in Chromatography
Emqal module: M0925 - Quality parameters and optimization in is a separation technique used for quantification of mixtures of analytes Svein.mjos@kj.uib.no Exercises and lectures can be found at www.chrombox.org/emq
More informationChapter 6 Photoluminescence Spectroscopy
Chapter 6 Photoluminescence Spectroscopy Course Code: SSCP 4473 Course Name: Spectroscopy & Materials Analysis Sib Krishna Ghoshal (PhD) Advanced Optical Materials Research Group Physics Department, Faculty
More informationChromatography. Mrs. D. MEENA MPharm PA & QA KTPC
Chromatography Mrs. D. MEENA MPharm PA & QA KTPC INTRODUCTION ANALYTICAL TECHNIQUES Analytical chemistry involves separating, identifying and determining the relative amount of the components in a sample
More informationChemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets
Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets Liquid chromatography was developed by Tswett in early 1900 s and was shown to be a powerful separation
More informationChapter 17: Fundamentals of Spectrophotometry
Chapter 17: Fundamentals of Spectrophotometry Spectroscopy: the science that deals with interactions of matter with electromagnetic radiation or other forms energy acoustic waves, beams of particles such
More informationCork Institute of Technology. Summer 2005 Instrumental Analysis (Time: 3 Hours) Section A
Cork Institute of Technology Higher Certificate in Science in Applied Biology Award (National Certificate in Science in Applied Biology Award) Answer FIVE questions; answer Section A, TWO questions from
More informationExperiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks
Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks bjective: The purpose of this experiment is to quantify the caffeine content of a diet soda sample using Ultra-High Performance
More informationChapter 27: Gas Chromatography. Principles Instrumentation Detectors Columns and Stationary Phases Applications
Chapter 27: Gas Chromatography Principles Instrumentation Detectors Columns and Stationary Phases Applications GC-MS Schematic Interface less critical for capillary columns Several types of Mass Specs
More informationExcited State Processes
Excited State Processes Photophysics Fluorescence (singlet state emission) Phosphorescence (triplet state emission) Internal conversion (transition to singlet gr. state) Intersystem crossing (transition
More informationChromatography- Separation of mixtures CHEM 212. What is solvent extraction and what is it commonly used for?
Chromatography- Separation of mixtures CHEM 212 What is solvent extraction and what is it commonly used for? How does solvent extraction work? Write the partitioning coefficient for the following reaction:
More informationIntroduction to Gas Chromatography
Introduction to Gas Chromatography 31-1 Objectives To know what is chromatography To understand the mechanism of compound separation To know the basic of gas chromatography system 31-2 Chromatography Definition
More informationSkoog/Holler/Crouch Chapter 26 Principles of Instrumental Analysis, 6th ed. CHAPTER 26
Skoog/Holler/Crouch Chapter 26 Principles of Instrumental Analysis, 6th ed. Instructor s Manual CHAPTE 26 26-1. (a) Elution is a process in which species are washed through a chromatographic column by
More information2. a) R N and L N so R L or L R 2.
1. Use the formulae on the Some Key Equations and Definitions for Chromatography sheet. a) 0.74 (remember that w b = 1.70 x w ½ ) b) 5 c) 0.893 (α always refers to two adjacent peaks) d) 1.0x10 3 e) 0.1
More informationHarris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY
Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY Chapter 23. Gas Chromatography What did they eat in the year 1,000? GC of Cholesterol and other lipids extracted from
More informationHigh Performance Liquid Chromatography
Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN
More informationHigh Performance Liquid Chromatography
Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 1 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN
More informationElectronic Spectra of Complexes
Electronic Spectra of Complexes Interpret electronic spectra of coordination compounds Correlate with bonding Orbital filling and electronic transitions Electron-electron repulsion Application of MO theory
More informationGas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5
Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH 5730-6730 Analysis Topic 5 Chromatography! Group of separation techniques based on partitioning (mobile phase/stationary phase). Two immiscible
More information1. Cyclic voltammetry involves the measurement of a diffusion controlled at an electrode in which the is controlled. (4 points)
Chem 454 First Exam Feb. 20, 2002 1. Cyclic voltammetry involves the measurement of a diffusion controlled at an electrode in which the is controlled. (4 points) 2. (5 points) A. Sketch a cyclic voltammogram
More informationVolatile organic compounds (VOCs):
Volatile organic compounds (VOCs): Organic chemicals with a high vapour pressure at room temperature. High vapour pressure results from a low boiling point. The World Health Organization (WHO) defined
More informationLAB #3: FLUROESCENCE SPECTROSCOPY AND ELECTRON TRANSFER (This lab is adapted from the U of MN Phsyical Chemistry lab manual)
Chemistry 372 Gustavus Adolphus College A. Purpose LAB #3: FLUROESCENCE SPECTROSCOPY AND ELECTRON TRANSFER (This lab is adapted from the U of MN Phsyical Chemistry lab manual) In this experiment, you will
More informationInstrumental Chemical Analysis
L2 Page1 Instrumental Chemical Analysis Chromatography (General aspects of chromatography) Dr. Ahmad Najjar Philadelphia University Faculty of Pharmacy Department of Pharmaceutical Sciences 2 nd semester,
More informationCh.28 HPLC. Basic types of Liquid Chromatography Partition (LLC) Adsorption (LSC) Ion Exchange (IC) Size Exclusion (SEC or Gel Chromatography)
Ch.28 HPLC 28.1 Basic types of Liquid Chromatography Partition (LLC) Adsorption (LSC) Ion Exchange (IC) Size Exclusion (SEC or Gel Chromatography) High Performance (Pressure) LC Glass column st.steel (high
More informationInstrumental Analysis
Chem 454 Name: Instrumental Analysis Exam I February 5, 1999 80 possible points 1] 5 points Which of the following samples would be suitable for analysis by a calibration curve technique using a potentiometric
More informationLecture 3: Light absorbance
Lecture 3: Light absorbance Perturbation Response 1 Light in Chemistry Light Response 0-3 Absorbance spectrum of benzene 2 Absorption Visible Light in Chemistry S 2 S 1 Fluorescence http://www.microscopyu.com
More informationERT320 BIOSEPARATION ENGINEERING CHROMATOGRAPHY
ERT320 BIOSEPARATION ENGINEERING CHROMATOGRAPHY CHROMATOGRAPHY Week 9-10 Reading Assignment: Chapter 7. Bioseparations Science & Engineering, Harrison, R; Todd, P; Rudge, S.C and Petrides, D,P CHROMATOGRAPHY
More informationHPLC. High Performance Liquid Chromatography (HPLC) Harris Chapter 25
High Performance Liquid Chromatography (HPLC) Harris Chapter 25 12/1/2005 Chem 253 - Chapter 25 1 HPLC Separation of nonvolatile or thermally unstable compounds. If the analyte/sample can be found to be
More information2] The plate height in chromatography is best described as 2
9 Chromatography. General Topics 1] Explain the three major components of the van Deemter equation. Sketch a clearly labeled diagram describing each effect. What is the salient point of the van Deemter
More informationLiquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure.
High performance liquid chromatography (HPLC) is a much more sensitive and useful technique than paper and thin layer chromatography. The instrument used for HPLC is called a high performance liquid chromatograph.
More informationLecture- 08 Emission and absorption spectra
Atomic and Molecular Absorption Spectrometry for Pollution Monitoring Dr. J R Mudakavi Department of Chemical Engineering Indian Institute of Science, Bangalore Lecture- 08 Emission and absorption spectra
More informationGas Chromatography. Vaporization of sample Gas-solid Physical absorption Gas-liquid Liquid immobilized on inert solid
Gas Chromatography Vaporization of sample Gas-solid Physical absorption Gas-liquid Liquid immobilized on inert solid Principles Instrumentation Applications 18-1 Retention Volumes Volumes rather than times
More informationChapter content. Reference
Chapter 7 HPLC Instrumental Analysis Rezaul Karim Environmental Science and Technology Jessore University of Science and Technology Chapter content Liquid Chromatography (LC); Scope; Principles Instrumentation;
More informationFluorescence 2009 update
XV 74 Fluorescence 2009 update Jablonski diagram Where does the energy go? Can be viewed like multistep kinetic pathway 1) Excite system through A Absorbance S 0 S n Excite from ground excited singlet
More informationChapter 23 Introduction to Analytical Separations
Chapter 23 Introduction to Analytical Separations Homework Due Monday April 24 Problems 23-1, 23-2, 23-7, 23-15, 23-27, 23-29, 23-32 Analytical Separations: Universal approach to analyzing complex mixtures
More informationFluorescence Workshop UMN Physics June 8-10, 2006 Quantum Yield and Polarization (1) Joachim Mueller
Fluorescence Workshop UMN Physics June 8-10, 2006 Quantum Yield and Polarization (1) Joachim Mueller Quantum yield, polarized light, dipole moment, photoselection, dipole radiation, polarization and anisotropy
More informationChromatography & instrumentation in Organic Chemistry
Chromatography & instrumentation in Organic Chemistry What is Chromatography? Chromatography is a technique for separating mixtures into their components in order to analyze, identify, purify, and/or quantify
More informationAn Introduction to Chromatographic Separations
An Introduction to Chromatographic Separations Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451
More informationI. Proteomics by Mass Spectrometry 1. What is an internal standard and what does it accomplish analytically?
Name I. Proteomics by Mass Spectrometry 1. What is an internal standard and what does it accomplish analytically? Internal standards are standards added intentionally to all samples, standards and blanks.
More informationDetermination of Caffeine by HPLC
Determination of Caffeine by HPLC Introduction It was a long history before real high performance liquid chromatography (HPLC) had evolved. The very first indication of a chromatographic separation was
More informationChromatography Outline
Chem 2001 Summer 2004 Outline What is? The Chromatogram Optimization of Column Performance Why Do Bands Spread? Gas High-Performance Liquid Ion-Exchange 2 What is? In chromatography, separation is achieved
More informationChem Homework Set Answers
Chem 310 th 4 Homework Set Answers 1. Cyclohexanone has a strong infrared absorption peak at a wavelength of 5.86 µm. (a) Convert the wavelength to wavenumber.!6!1 8* = 1/8 = (1/5.86 µm)(1 µm/10 m)(1 m/100
More informationModel Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy
Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy Section I Q1. Answer (i) (b) (ii) (d) (iii) (c) (iv) (c) (v) (a) (vi) (b) (vii) (b) (viii) (a) (ix)
More informationsingle-molecule fluorescence resonance energy transfer
single-molecule fluorescence resonance energy transfer (2) determing the Förster radius: quantum yield, donor lifetime, spectral overlap, anisotropy michael börsch 26/05/2004 1 fluorescence (1) absorbance
More informationA very brief history of the study of light
1. Sir Isaac Newton 1672: A very brief history of the study of light Showed that the component colors of the visible portion of white light can be separated through a prism, which acts to bend the light
More informationSpectroscopy Problem Set February 22, 2018
Spectroscopy Problem Set February, 018 4 3 5 1 6 7 8 1. In the diagram above which of the following represent vibrational relaxations? 1. Which of the following represent an absorbance? 3. Which of following
More informationLC Technical Information
LC Technical Information Method Transfer to Accucore.6 μm Columns Containing solid core particles, which are engineered to a diameter of.6μm and a very narrow particle size distribution; Accucore HPLC
More informationBiochemistry. Biochemical Techniques HPLC
Description of Module Subject Name Paper Name 12 Module Name/Title 13 1. Objectives 1.1. To understand the basic concept and principle of 1.2. To understand the components and techniques of 1.3. To know
More informationDirect, Reliable, and Correct Oxygen Measurement
Direct, Reliable, and Correct Oxygen Measurement February, 2015 Oxygen-dependent quenching of phosphorescence has been used extensively for the last 25 years as a reliable method for measuring oxygen.
More informationtwo slits and 5 slits
Electronic Spectroscopy 2015January19 1 1. UV-vis spectrometer 1.1. Grating spectrometer 1.2. Single slit: 1.2.1. I diffracted intensity at relative to un-diffracted beam 1.2.2. I - intensity of light
More informationFluorescence Excitation and Emission Fundamentals
Fluorescence Excitation and Emission Fundamentals Fluorescence is a member of the ubiquitous luminescence family of processes in which susceptible molecules emit light from electronically excited states