Total Dissolved Residue (Total Dissolved Solids, TDS) in Aqueous Matrices

Size: px
Start display at page:

Download "Total Dissolved Residue (Total Dissolved Solids, TDS) in Aqueous Matrices"

Transcription

1 Total Dissolved Residue (Total Dissolved Solids, TDS) in Aqueous Matrices Environmental Express 2345A Charleston Regional Parkway Charleston, SC

2 Table of Contents 1. Scope and Application Summary of Method Definitions Interferences Safety Equipment and Supplies Reagents and Standards Sample Collection Preservation and Storage Quality Control... 6 Table 1: 40 CFR part Quality Control Requirements Procedure Data Analysis and Calculations Method Performance Pollution Prevention Waste Management References Tables, Diagrams, Flowcharts, and Validation Data Table A1: Environmental Express TDS Stable-Weigh Tests Blank Table A2: Environmental Express TDS Stable-Weigh Tests 200 mg Table A3: Environmental Express TDS Stable-Weigh Tests 100 mg Table A4: Environmental Express TDS Stable-Weigh Tests 20 mg Table A5: Environmental Express TDS Stable-Weigh Tests 20 mg-evaporation Dish.. 21 Table A6: StableWeigh and Porcelain Evaporation Dish Precision Table A7: StableWeigh Method Blank Page

3 Table of Figures Figure 1: Initial Demonstration of Capability Calculation... 7 Figure 2: LRB Calculation... 7 Figure 3: LFB Calculation... 7 Figure 4: RPD Calculation... 8 Figure 5: Total Solids Calculation Figure 6: StableWeigh 200 mg Initial Demonstration of Capability Figure 7: StableWeigh 100 mg Initial Demonstration of Capability Figure 8: StableWeigh vs. Porcelain Evaporation Dish Averages Page

4 1. Scope and Application 1.1. This method is based upon Standard Methods 2540 for the determination of Total Dissolved Residue in aqueous sample matrices This method is for use in the Environmental Protection Agency s (EPA s) data gathering and monitoring programs under the Clean Water Act, the Resource Conservation and Recovery Act, the Comprehensive Environmental Response, Compensation and Liability Act, and the Safe Drinking Water Act This method is used to determine Total Dissolved Residue (Total Dissolved Solids, TDS) This method is applicable to drinking, surface, and saline waters, domestic and industrial wastes 1.5. The concentration range for this method is determined by the minimum and maximum residue mass allowed ( mg) and the volume of sample needed to achieve the desired residue mass. Samples for TDS have no upper volume limit specified but are limited by the size of the evaporation vessel and the practicality of repeated additions of sample volume Each laboratory that uses the method must demonstrate the ability to generate acceptable results using the procedure. 2. Summary of Method 2.1. For TDS the entire aliquot of sample is evaporated in a pre-weighed vessel and dried to constant at 180 ± 2 C. The mass of the residue is determined by difference in mass of the vessel. 3. Definitions 3.1. Laboratory Fortified Blank (LFB) - An aliquot of reagent water or other blank matrix to which known quantities of the method analytes and all the preservation compounds are added. The LFB is processed and analyzed exactly like a sample, and its purpose is to determine whether the methodology is in control, and whether the laboratory is capable of making accurate and precise measurements Laboratory Matrix/Duplicate (LM/LMD) also called Matrix / Duplicate (M/D): An aliquot of an environmental sample and a duplicate which are analyzed and its purpose is to determine whether the sample matrix contributes bias to the analytical results Laboratory Reagent Blank (LRB) - A volume of reagent water or other blank matrix that is processed exactly as a sample including exposure to all glassware and equipment, that 2

5 are used in the analysis batches. The LRB is used to determine if the method analytes or other interferences are present in the laboratory environment, the reagents, or the apparatus Minimum Reporting Level (MRL) - The minimum concentration that can be reported by a laboratory as a quantitated value for a method analyte in a sample following analysis. This concentration must not be any lower than the concentration of the lowest calibration standard for that instrument Total Dissolved Residue or Total Dissolved Solids is defined as all material that is not volatile at 180 C Water Sample: For the purpose of this method, a sample taken from one of the following sources: drinking water, surface water, storm runoff, industrial or domestic wastewater. 4. Interferences 4.1. Samples with multiple phases or those that settle into multiple layers present a problem with homogeneity. Ensure that such samples are well mixed during use. Vigorous shaking through inversion will provide a sufficient mixing for most samples. Samples that settle rapidly or resist mixing may require constant stirring with a magnetic stirrer. Use a wide bore pipette in such instances. Draw sample from the midpoint between the wall and the vortex. Avoid using propeller style mixers as these may shear particles resulting in underreported values Some samples may form a crust during drying. This crust can prevent some evaporation of water during the drying process. If a water trapping crust continues to be problematic reduce the amount of sample used to minimize the crust being formed or take other steps to allow water to evaporate. Document additional steps being used as well as the precautions that are taken to avoid loss of residue from the dried sample Organic compounds such as oil and grease may present problems in obtaining a stable final. Such compounds are often visible to the naked eye and well present as a constantly decreasing mass after each drying event. Develop a Quality Plan to positively identify these types of samples. Obtain a value for HEM for such samples, if possible and consult with the final data user how to use that value in conjunction with the data obtained from the analysis. 5. Safety 5.1. When working with and around ovens or other evaporation equipment the materials in use will become hot. Use appropriate heat resistant gloves and other necessary PPE to protect the analysts from exposure to heat. Different temperatures may require different PPE to be effective. 3

6 5.2. Use all standard PPE (gloves, safety glasses, lab coats, etc.) as required by your Laboratory Safety and Chemical Hygiene Plan. Pay particular attention to samples of unusual appearance or odor and request additional information from the sample source as necessary to identify unknown or unexpected hazards This method does not address all safety issues associated with its use. The laboratory is responsible for maintaining a safe work environment and a current awareness file of OSHA regulations regarding the safe handling of the chemicals specified in this method. A reference file of safety data sheets (SDS s) must be available to all personnel involved in these analyses. 6. Equipment and Supplies Note: Brand names, suppliers and part numbers are cited for illustrative purposes only. No endorsement is implied. Equivalent performance may be achieved using equipment and materials other than those specified here, but demonstration of equivalent performance that meets the requirements of this method is the responsibility of the laboratory Evaporation Vessel StableWeigh TDS Vessel from Environmental Express, or equivalent. This vessel is not suitable for use in volatile solids determination. In such instances use ceramic, high silica glass, or platinum evaporating dishes of appropriate dimensions. When using the StableWeigh product, other accessories are recommended or required for ease of use StableWeigh modular rack recommended. Used to transport and hold StableWeigh Vessels to, from, and in ovens and desiccators StableWeigh weighing bracket recommended. Used to support vessels on the balance Static electricity dissipation device. Best results have been achieved using the Mettler Toledo balance deionizer or similar device. Mounted anti-static ionizing cartridges or brushes can be effective if care is taken to ensure the entire vessel is appropriately treated Drying oven for use at C and C Dessicator with dessicant must include some indicator to determine when replacement of the desiccant is necessary. This can be indicating desiccant or other appropriate means Analytical balance capable of accurate and precise weighing to 0.1 mg Class A measuring devices of appropriate volume to measure necessary amounts of sample. 4

7 6.9. Glass Fiber Filter Discs without organic binder, and 1.5 m nominal pore size. Unprepared filters are available or use Washed & Dried filters from Environmental Express, or equivalent, to skip the filter preparation steps in the procedure Filtration apparatus, StableWeigh Filling Station Use a filter funnel of the appropriate size to hold the volume of sample and diameter of filter being used. The StableWeigh Filling Station (single place or six place) will allow the analyst to filter samples directly into the vessel. 7. Reagents and Standards 7.1. The various forms of residue are determined based solely on physical characteristics. No specific chemicals or reagents are required for this analysis No specific standard is required. The TDS Standard from Environmental Express or equivalent may be used. Each purchased standard may have slightly different target values. Consult the certificate of analysis for the assigned true value of each standard. Alternatively, follow the directions in Sample Collection Preservation and Storage 8.1. Always refer to the latest guidelines in 40 CFR part 136 for sampling guidelines, preservation guidelines, and holding time. Those instructions supersede any guidance given in this document Collect sample in a plastic or glass bottle. A recommended volume is a minimum of 250 ml Store samples at 6 C when not in use Store samples for no more than 7 days from collection. 5

8 9. Quality Control 9.1. Each laboratory that uses this method is required to develop and implement a formal Quality Manual. The minimum requirements of this program are outlined in 40 CFR part Table 1: 40 CFR part Quality Control Requirements 40 CFR QC Requirement Required (Yes/No) and Notes Demonstration of Capability (DOC), Method Detection Limit (MDL), Laboratory reagent blank (LRB), Laboratory fortified blank (LFB), also referred to as a spiked blank, or laboratory control sample (LCS), Matrix spike (MS), matrix spike duplicate (MSD), or laboratory fortified blank duplicate (LFBD) for suspected difficult matrices, Internal standard/s, surrogate standard/s (for organic analysis) or tracer (for radiochemistry), Calibration (initial and continuing), Control charts (or other trend analyses of quality control results), Corrective action (root cause analyses), Specific frequency of QC checks, QC acceptance criteria, and Definitions of a batch (preparation and analytical) Initial and Ongoing Demonstration of Capability Required MDL Spike and MDL Blank not required for methods defined test. Required daily or with each batch of 20 or fewer samples. LFB is required daily or with each batch of 20 or fewer samples. Duplicates daily or with each batch of 20 or fewer samples. Not required. Calibrate balance as per manufacturer and regulatory requirements. Required. Required. Required. Required. Required. Note: Daily: Each day samples are analyzed in the laboratory Initial Demonstration of Capability (IDC) 6

9 9.3. Before new analysts run any samples, verify their capability with the method. Run a LFB at least four times. Calculate the standard deviation of the four samples. All 4 LFB values must fall within the LFB recovery limits for the analyst to pass the IDC LFB s initial recovery limits = Mean ± ( 5.84 Standard Deviation) Where: 5.84 = the two-sided Student s t factor for three degrees of freedom Laboratory Reagent Blank (LRB): Figure 1: Initial Demonstration of Capability Calculation 9.5. High quality laboratory water that is analyzed as a sample. Analyzed at a frequency of no less than once per sample batch. Develop control chart limits to determine acceptable recovery. Perform Root Cause Analysis as per the laboratory Quality Manual in recovery is unacceptable ExperimentalValue = LRB ExperimentalValue = LRB Concentration determined experimentally 9.6. Laboratory Fortified Blank (LFB): LFB can : Figure 2: LRB Calculation Be purchased from a 3 rd party supplier such as Environmental Express. Follow supplier s directions or: Prepare in a 1000 ml volumetric flask put g (± g) of Celite and g (± g) of Sodium Chloride. Dilute to volume with DI water and mix well. Transfer 10 ml of the LFB to a StableWeigh vessel and dry. Recovery will be g/l A maximum recovery range of 80% to 120% is expected. Develop control chart limits to determine acceptable recovery. Perform Root Cause Analysis as per the laboratory Quality Manual in recovery is unacceptable. Experimental Value *100= Percent Recovery LFB Expected Value Experimental Value = LFB Concentration determined experimentally Expected Value = Known LFB concentration Figure 3: LFB Calculation 9.7. Laboratory Matrix/ Laboratory Matrix Duplicate (LM/LMDD): 7

10 9.8. A duplicate set of TS samples are analyzed. LM/LMD are analyzed at a frequency of no less than once per sample batch. A Relative Percent Difference (RPD) will be calculated for each LM/LMD set. Develop control chart limits for RPD to determine acceptable recovery. Perform Root Cause Analysis as per the laboratory Quality Manual in recovery is unacceptable. Absolute Value ( LM-LMD) *100 = RPD LM+LMD 2 LM = Concentration determined for LM LMD = Concentration determined for LM duplicate 9.9. Ongoing Demonstration of Capability Figure 4: RPD Calculation Samples consisting of the LRB, LFB, and LM/LMD will be run each batch of samples. These results will be charted on control charts to determine ODC for the laboratory. Develop control chart limits to determine acceptable recovery. Perform Root Cause Analysis as per the laboratory Quality Manual in recovery is unacceptable Balance Check. An appropriate check should be evaluated at the beginning and end of each weighing sequence. The checks will demonstrate lack of drift in the balance. Reasonable choices would be within the same order of magnitude as the expected mass of the residue plus StableWeigh vessel used for the samples. Develop control chart limits to determine acceptable check. Perform Root Cause Analysis as per the laboratory Quality Manual in check is unacceptable Sample Batch A group of samples which behave similarly with respect to the sampling or the testing procedures being employed and which are processed as a unit. For QC purposes, if the number of samples in a group is greater than 20, then each group of 20 samples or less will all be handled as a separate batch. A batch cannot span between laboratory work days (24 hrs.). New batches must be started each laboratory work day Sample Batch: Typical sample analysis sequence Balance Check LRB LFB 8

11 LM LMD Samples Balance Check Other Equipment Checks Other equipment used in this analysis, such as ovens, balances, and thermometers, must be maintained and checked as required by the laboratory Quality Manual Control charts and trend analysis will be performed on all areas identified in the Quality Control section in this method and in the laboratories Quality Manual. 10. Procedure Note: This method is entirely empirical. Acceptable results can be obtained only by strict adherence to all details Select an appropriate number of StableWeigh TDS vessels for the required number of samples and QC. Assign each vessel to a sample and note the vessel tare in the appropriate analysis data log. StableWeigh vessels must not be used if the sample aliquot is to be analyzed for volatile solids If ProWeigh Washed and Dried filters are used, skip the following preparation steps and proceed to Insert a filter into the filtration apparatus and apply vacuum Wash the filter with three successive 20 ml portions of reagent water. Continue to apply vacuum to remove all traces of wash water from the filter Discard wash water Choose a sample size expected to yield between 2.5 and 700 mg of residue. Mix the sample well through appropriate means (repeated inversion, stirring with magnetic stirrer or other) and measure out the appropriate sample volume. If the vessel is not of sufficient size to hold the volume required, successive portions of sample may be added after evaporation Place the filter and StableWeigh TDS Vessel in the StableWeigh filling station. Apply vacuum and add the sample to the filter holder. No more than 10 minutes should be required for sample filtration. Larger filter sizes or smaller sample volumes may be necessary to fulfill this time requirement. After filtration is complete wash the filter with three successive 10 ml portions of reagent water to ensure complete transfer of dissolved constituents into the filtrate. Continue suction until all visible water has been removed from the filter. 9

12 10.8. Remove the StableWeigh TDS vessel from the filling station and transfer it to the oven rack Evaporate the sample to dryness using a block digester, oven, steam bath, or other appropriate method. Set the evaporation temperature low enough that the samples do not boil and splatter Dry the evaporated sample and vessel in a drying oven at 180 ± 2 C for at least one hour Remove the vessels from the oven and place in a dessicator until they have reached balance temperature. Cooling time for StableWeigh vessels can be reduced by transferring the vessels to a cool rack prior to placing in the dessicator Weigh each vessel on the balance and record the to 0.1 mg Repeat steps at least one more time and as many times as necessary to obtain consecutive s that differ by less than 0.5 mg. QC samples must go through the drying/cooling/weighing cycle as long as other samples require them. Other samples may be removed from the cycle once they have shown stability. When using crucibles, keeping the dessicator time constant with each cycle will help to reduce the number of cycles needed to achieve constant. 11. Data Analysis and Calculations Calculations for Total Dissolved Solids ( ) 12. Method Performance A-B x1000 = Total DissolvedSol ids (mg/l) C A = Weight of dried sample + StableWeigh B = Weight of StableWeigh C = volume of sample(ml) Figure 5: Total Solids Calculation See Standard Methods 2540 for performance data on the original analysis method Comparison data validating the use of the StableWeigh vessel in comparison to the crucibles required for TDS are found in Tables/Appendices. 10

13 13. Pollution Prevention The main source of pollution from this method is from unknown samples. Follow your laboratory Waste Management Program for the identification and disposal of potentially hazardous samples. 14. Waste Management It is the laboratory's responsibility to comply with all federal, state, and local regulations governing waste management, particularly the hazardous waste identification rules and land disposal restrictions, and to protect the air, water, and land by minimizing and controlling all releases from fume hoods and bench operations. Compliance with all sewage discharge permits and regulations is also required Use the least amount of sample possible to generate valid results, particularly for TS. This will minimize the amount of waste generated by the analysis For further information on waste management, consult "The Waste Management Manual for Laboratory Personnel," and "Less is Better: Laboratory Chemical Management for Waste Reduction," both available from the American Chemical Society's Department of Government Relations and Science Policy, th Street N.W., Washington DC, References Eaton, A. D., Clesceri, L. S., Baird, R. B., Rice, E. W., American Public Health Association., American Water Works Association., & Water Environment Federation. (2012). Standard Methods for the Examination of Water and Wastewater, 22 nd edition. Washington, DC: American Public Health Association 11

14 16. Tables, Diagrams, Flowcharts, and Validation Data 12

15 Analyst Edward F. Askew Date 4/14/2016 Sample # Bag # Volume (ml) Table A1: Environmental Express TDS Stable-Weigh Tests Blank Initial 1 Blank 2 3 Two Consecutive Weights Difference used Percent Change % % % % % % % % % % % % % % % % % % % 13

16 Analyst Edward F. Askew Date 4/14/2016 Sample # Bag # Volume (ml) Table A1: Environmental Express TDS Stable-Weigh Tests Blank Initial 1 Blank 2 3 Two Consecutive Weights Difference used Percent Change % % % % % % Average % Standard Deviation % 14

17 Analyst Edward F. Askew Date 4/18/2016 Sample # Bag # Volume (ml) Table A2: Environmental Express TDS Stable-Weigh Tests 200 mg Initial mg per 50 ml 2 3 Two Consecutive Weights Difference used Solids Recovered Weight Percent Recovery % % % % % % % % % % % % % % % % % % % 15

18 Analyst Edward F. Askew Date 4/18/2016 Sample # Bag # Volume (ml) Table A2: Environmental Express TDS Stable-Weigh Tests 200 mg Initial mg per 50 ml 2 3 Two Consecutive Weights Difference used Solids Recovered Weight Percent Recovery % % % % % % Average % Standard Deviation % 16

19 Analyst Edward F. Askew Date 4/20/2016 Sample # Bag # Volume (ml) Table A3: Environmental Express TDS Stable-Weigh Tests 100 mg Initial mg per 50 ml 2 3 Two Consecutive Weights Difference used Solids Recovered Weight Percent Recovery % % % % % % % % % % % % % 14 NA % % % % % % 17

20 Analyst Edward F. Askew Date 4/20/2016 Sample # Bag # Volume (ml) Table A3: Environmental Express TDS Stable-Weigh Tests 100 mg Initial mg per 50 ml 2 3 Two Consecutive Weights Difference used Solids Recovered Weight Percent Recovery % % % % % % Average % Standard Deviation % 18

21 Analyst Edward F. Askew Date 4/24/2016 Sample # Bag # Volume (ml) Table A4: Environmental Express TDS Stable-Weigh Tests 20 mg Initial mg per 50 ml 2 3 Two Consecutive Weights Difference used Solids Recovered Weight Percent Recovery % % % % % % % % % % % % % % % % % % % 19

22 Analyst Edward F. Askew Date 4/24/2016 Sample # Bag # Volume (ml) Table A4: Environmental Express TDS Stable-Weigh Tests 20 mg Initial mg per 50 ml 2 3 Two Consecutive Weights Difference used Solids Recovered Weight Percent Recovery % % % % % % Average % Standard Deviation % 20

23 Table A5: Environmental Express TDS Stable-Weigh Tests 20 mg-evaporation Dish Analyst Edward F. Askew Date 4/22/ mg per 50 ml (Evaporation Dish) Sample # Evaporation Dish # Volume (ml) Initial Two Consecutive Weights Difference used Solids Recovered Weight Percent Recovery % % % % % % % % % % % % % % % % % % % 21

24 Table A5: Environmental Express TDS Stable-Weigh Tests 20 mg-evaporation Dish Analyst Edward F. Askew Date 4/22/ mg per 50 ml (Evaporation Dish) Sample # Evaporation Dish # Volume (ml) Initial Two Consecutive Weights Difference used Solids Recovered Weight Percent Recovery % % % % % 25 A % Average % Standard Deviation % 22

25 Table A6: StableWeigh and Porcelain Evaporation Dish Precision TDS Mass Vessel Average Standard Deviation % RSD Duplicate Relative Percent Difference 200 mg StableWeigh % 1.00% 100 mg StableWeigh % 2.50% 20 mg StableWeigh % 9.05% 20 mg Porcelain Evaporation Dish % 15.31% Table A7: StableWeigh Method Blank Sample # Blank Weight Difference

26 Figure 6: StableWeigh 200 mg Initial Demonstration of Capability 24

27 Figure 7: StableWeigh 100 mg Initial Demonstration of Capability 25

28 Figure 8: StableWeigh vs. Porcelain Evaporation Dish Averages 26

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Method 1061 Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Company Method 1061 Revision 1. December 015 Organic Carbon in Finished

More information

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water Hach Method 1041 Spectrophotometric Measurement of Free Chlorine (Cl ) in Finished Drinking Water Hach Company Method 1041 Revision 1. November 015 Spectrophotometric Measurement of Free Cl in Finished

More information

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Nitrate Method 10206 Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Method 10206 Revision 2.2 January 15, 2013 Spectrophotometric

More information

ALLOWAY METHOD OUTLINE

ALLOWAY METHOD OUTLINE ALLOWAY METHOD OUTLINE Standard Laboratory Method SM4500-Cl -G Parameter Residual Chlorine & Free Chlorine Method DPD Colorimetric Test Kit Date Issued Originator: Section Supervisor: QA Manager Date:

More information

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method: ALLOWAY METHOD OUTLINE Standard Laboratory Method: SM Parameter: Methylene Blue Method: Colorimetric Reporting Level: Reference: 0.05 mg/l Standard Methods for the Examination Of Water and Wastewater;

More information

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1 Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall Corvallis,

More information

Timberline Ammonia-001. Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis

Timberline Ammonia-001. Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis Timberline Ammonia-001 Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis Prepared By: Dr. Edward F. Askew Date: June 24, 2011 Table of Contents 1. SCOPE

More information

Hach Company TNTplus TM Phosphorus Spectrophotometric Measurement of Phosphorus in Water and Wastewater

Hach Company TNTplus TM Phosphorus Spectrophotometric Measurement of Phosphorus in Water and Wastewater Hach Company TNTplus TM Phosphorus Spectrophotometric Measurement of Phosphorus in Water and Wastewater Hach Company TNTplus TM Phosphorus Method 10209/10210/843/844/845 November 2015 Revison 2 Spectrophotometric

More information

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Organics Revision Date: July 19, 2017 Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Parameter Perfluorinated Alkyl Acids (Perfluorobutane Sulphonate (PFBS), Perflourooctane Sulphonate (PFOS),

More information

Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2

Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2 Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall Corvallis,

More information

Is the laboratory s pledge or declaration of the quality of the results produced. to produce data compliant with the Safe Drinking Water Act (SDWA)

Is the laboratory s pledge or declaration of the quality of the results produced. to produce data compliant with the Safe Drinking Water Act (SDWA) QA/QC Is the laboratory s pledge or declaration of the quality of the results produced. to produce data compliant with the Safe Drinking Water Act (SDWA) Is a description of the policies, procedures, techniques

More information

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2 Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State

More information

Approximate Volatile Acids by Titration

Approximate Volatile Acids by Titration SOP AMBL-101-A Page 1 of 5 Standard Operating Procedure AMBL-101-A Prepared: April 12, 2006 Revised: July 16, 2014 Prepared by: Terry E. Baxter Reviewed by: Approximate Volatile Acids by Titration METHOD

More information

LOW-RANGE TOTAL CHLORINE

LOW-RANGE TOTAL CHLORINE LOW-RANGE TOTAL CHLORINE Thermo Orion Method AC2072 Revision 4 3/14/2002 Accepted by U.S. EPA on March 20, 2002 for wastewater compliance monitoring Accepted by U.S. EPA on September 31, 2002 for drinking

More information

METHOD DETERMINATION OF SULFATE BY AUTOMATED COLORIMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

METHOD DETERMINATION OF SULFATE BY AUTOMATED COLORIMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division METHOD 375.2 DETERMINATION OF SULFATE BY AUTOMATED COLORIMETRY Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division Revision 2.0 August 1993 ENVIRONMENTAL MONITORING SYSTEMS

More information

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP.

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP. NEW JERSEY DEPARTMENT OF ENVIRONMENTAL PROTECTION OFFICE OF QUALITY ASSURANCE ENVIRONMENTAL LABORATORY CERTIFICATION PROGRAM ON-SITE LABORATORY EVALUATION RADIOCHEMISTRY PROCEDURES Gross Alpha-Gross Beta

More information

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University. Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared

More information

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows Nitrate DOC316.53.01067 Cadmium Reduction Method Method 8192 0.01 to 0.50 mg/l NO 3 N (LR) Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA)

Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA) Methodology Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA) (Cartridge Part #A002603) 1.0 Scope and Application 1.1 This method is used to determine the concentration of nitrate

More information

METHOD DETERMINATION OF PHOSPHORUS BY SEMI-AUTOMATED COLORIMETRY

METHOD DETERMINATION OF PHOSPHORUS BY SEMI-AUTOMATED COLORIMETRY METHOD 365.1 DETERMINATION OF PHOSPHORUS BY SEMI-AUTOMATED COLORIMETRY Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division Revision 2.0 August 1993 ENVIRONMENTAL MONITORING

More information

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,: sop-c-102 Determination of Chemical Oxygen Demand Revision 6 Approval: Laborato ry Manager/LQAO/RS O Date Z/zs Date Z-z* I A Renewat date: 7lt:,: Effective date: 3, tnitiats: wu\ Texas lnstitute for Applied

More information

Test Method: CPSC-CH-E

Test Method: CPSC-CH-E UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 10901 DARNESTOWN RD GAITHERSBURG, MD 20878 Test Method: CPSC-CH-E1001-08 Standard Operating Procedure

More information

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows Silver DOC316.53.01134 Colorimetric Method Method 8120 0.02 to 0.70 mg/l Ag Powder Pillows Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows

More information

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE 1.0 SCOPE AND APPLICATION 1.1 This method can be used for measuring total solubilized nitrate in drinking

More information

METHOD DETERMINATION OF AMMONIA NITROGEN BY SEMI-AUTOMATED COLORIMETRY

METHOD DETERMINATION OF AMMONIA NITROGEN BY SEMI-AUTOMATED COLORIMETRY METHOD 350.1 DETERMINATION OF AMMONIA NITROGEN BY SEMI-AUTOMATED COLORIMETRY Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division Revision 2.0 August 1993 ENVIRONMENTAL MONITORING

More information

Percent Solids Determination

Percent Solids Determination Title: Percent Solids Page 1 of 10 Percent Solids Determination References: Method 2540G, Standard Methods For the Examination of Water and Wastewater, APHA 18 th edition, 1992. 1. Scope and Application

More information

Standard Operating Procedure for the Analysis of Nitrate/Nitrite and Nitrite in Fresh Waters CCAL 31B.1

Standard Operating Procedure for the Analysis of Nitrate/Nitrite and Nitrite in Fresh Waters CCAL 31B.1 Standard Operating Procedure for the Analysis of Nitrate/Nitrite and Nitrite in Fresh Waters CCAL 31B.1 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson

More information

Chloride by Flow Injection Analysis (FIA)

Chloride by Flow Injection Analysis (FIA) Methodology Chloride by Flow Injection Analysis (FIA) (Cartridge Part #A001568) 1.0 Scope and Application 1.1 This method is used for the determination of chloride in drinking water, surface water, saline

More information

Hexavalent Chromium by Flow Injection Analysis (FIA)

Hexavalent Chromium by Flow Injection Analysis (FIA) Methodology Hexavalent Chromium by Flow Injection Analysis (FIA) (Cartridge Part #A002831) 1.0 Scope and Application 1.1 This method is used for the determination of hexavalent chromium in water, including

More information

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8030 is used to determine the concentration of the following volatile organic compounds: Compound Name

More information

ALLOWAY. Method Outline. Standard Laboratory Method: Total Fluoride via Distillation. Distillation then Specific Ion. Date Issued: 01/04/82

ALLOWAY. Method Outline. Standard Laboratory Method: Total Fluoride via Distillation. Distillation then Specific Ion. Date Issued: 01/04/82 ALLOWAY Method Outline Standard Laboratory Method: SM4500-F - B and SM4500-F-C Parameter: Total Fluoride via Distillation Method: Distillation then Specific Ion Date Issued: 01/04/82 Date Revised: 04/09/02,

More information

QAM-I-116 Preparation of Labware

QAM-I-116 Preparation of Labware 1. Applicability and Purpose i. This procedure applies to all labware (glassware and plasticware) used for analysis in the laboratory, and sampling bottles and equipment used in field sample operations.

More information

Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission

Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission 0 chapter Sodium and Potassium Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission Spectroscopy 67 S. S. Nielsen (ed.), Food Analysis Laboratory Manual Springer

More information

Standard Operating Procedure VUSP M

Standard Operating Procedure VUSP M Villanova University Date Nov 2009 Page 1 of 17 Villanova University Villanova Urban Stormwater Partnership Watersheds Laboratory Standard Operating Procedure VUSP M Nitrate by Discrete Analysis Systea

More information

Alkalinity, USEPA by Flow Injection Analysis (FIA)

Alkalinity, USEPA by Flow Injection Analysis (FIA) Methodology Alkalinity, USEPA by Flow Injection Analysis (FIA) (Cartridge Part #A002745) 1.0 Scope and Application 1.1 This method is used for the determination of alkalinity in drinking water, surface

More information

METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE)

METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE) METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE) 1.0 SCOPE AND APPLICATION 1.1 This method is applicable to ground water, drinking, surface, and saline waters, and domestic and industrial wastes.

More information

Standard Operating Procedure for: Total Phosphorus (3010R02 Total P.doc) Missouri State University. and

Standard Operating Procedure for: Total Phosphorus (3010R02 Total P.doc) Missouri State University. and Standard Operating Procedure for: Total Phosphorus (3010R02 Total P.doc) Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared by: Date: OEWRI Quality Assurance

More information

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

Tex-620-J, Determining Chloride and Sulfate Contents in Soil Contents in Soil Contents: Section 1 Overview...2 Section 2 Sample Preparation...3 Section 3 Ion Chromatography Method...5 Section 4 Wet Chemical Method...9 Section 5 Archived Versions...15 Texas Department

More information

Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and

Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared by: OEWRI Laboratory Manager

More information

Method April 1995

Method April 1995 N-Hexane Extractable Material (HEM) and Silica Gel Treated N-Hexane Extractable Material (SGT-HEM) by Extraction and Gravimetry (Oil and Grease and Total Petroleum Hydrocarbons) April 1995 U.S. Environmental

More information

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials Nitrogen, Total DOC316.53.01085 Persulfate Digestion Method Method 10072 2 to 150 mg/l N (HR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific information

More information

QA/QC in the Wastewater Laboratory. Steve Roberts Ohio EPA Division of Environmental Services 05/11/2016

QA/QC in the Wastewater Laboratory. Steve Roberts Ohio EPA Division of Environmental Services 05/11/2016 QA/QC in the Wastewater Laboratory Steve Roberts Ohio EPA Division of Environmental Services 05/11/2016 What is Quality Assurance? Per Standard Methods for the Examination of Water and Wastewater, 21 st

More information

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3 , MR, 8171 DOC316.53.01069 Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3 N) Powder Pillows or AccuVac Ampuls Scope and Application: For water, wastewater and seawater Test preparation

More information

METHOD DETERMINATION OF TOTAL CYANIDE BY SEMI-AUTOMATED COLORIMETRY

METHOD DETERMINATION OF TOTAL CYANIDE BY SEMI-AUTOMATED COLORIMETRY METHOD 335.4 DETERMINATION OF TOTAL CYANIDE BY SEMI-AUTOMATED COLORIMETRY Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division Revision 1.0 August 1993 ENVIRONMENTAL MONITORING

More information

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge Standard Operating Procedure for GLNPO Specific Conductance: Marvin Palmer United States Environmental Protection Agency Great Lakes National Program Office Region 5 77 West Jackson 9th Floor Chicago,

More information

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Nitrate, MR DOC316.53.01069 Cadmium Reduction Method Method 8171 0.1 to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Scope and application: For water, wastewater

More information

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY.

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY. 1 REVISOR 4740.2100 4740.2100 QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY. Subpart 1. Scope. This part applies to laboratories performing testing under the inorganic chemistry, metals,

More information

ASTM Designation: D Standard Test Method for Determination of Iodine Number of Activated Carbon

ASTM Designation: D Standard Test Method for Determination of Iodine Number of Activated Carbon ASTM Designation: D4607-94 Standard Test Method for Determination of Iodine Number of Activated Carbon 1. Scope 1.1 This test method covers the determination of the relative activation level of unused

More information

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction II. Quality Control A. Introduction ENCO adheres to strict quality control practices in order to assure our clients that the data provided are accurate and reliable. We are required by the EPA to analyze

More information

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials Nitrogen, Total DOC316.53.01086 Persulfate Digestion Method Method 10071 0.5 to 25.0 mg/l N (LR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific

More information

DR/4000 PROCEDURE NITRATE. 2. The display will show:

DR/4000 PROCEDURE NITRATE. 2. The display will show: Method 8192 Powder Pillows DR/4000 PROCEDURE Cadmium Reduction Method LR (0 to 0.50 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit for program number

More information

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows Potassium DOC316.53.01127 Tetraphenylborate Method Method 8049 0.1 to 7.0 mg/l K Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific information

More information

SECTION D.2 AMMONIA NITROGEN

SECTION D.2 AMMONIA NITROGEN SECTION D.2 AMMONIA NITROGEN CEDR Method Code: NH4F L01 a) Scope and Application i) This method describes the determination of low-level ammonia nitrogen concentrations in filtered samples taken from fresh

More information

SOP-C-120 Determination of ph

SOP-C-120 Determination of ph i. Identification of the method a. SM 4500-H + (approved 2011) and EPA 9045C (approved 1997) ii. Applicable matrix or matrices a. Waters or aqueous solutions and suspensions iii. Limits of detection and

More information

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Glossary of Common Laboratory Terms

Glossary of Common Laboratory Terms Accuracy A measure of how close a measured value is to the true value. Assessed by means of percent recovery of spikes and standards. Aerobic Atmospheric or dissolved oxygen is available. Aliquot A measured

More information

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure Troy, NY 12180 Date: 12/12/2017 Page 1 of 7 USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure 1. Scope and Application 1.1 Analytes Dissolved inorganic

More information

Safety in the Chemistry Laboratory

Safety in the Chemistry Laboratory Safety in the Chemistry Laboratory CHAPTER1 Safety must be everyone s primary concern in the chemistry lab. Understanding and following all safety rules in the organic chemistry lab is critical to your

More information

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT METHOD 9040B ph ELECTROMETRIC MEASUREMENT 1.0 SCOPE AND APPLICATION 1.1 Method 9040 is used to measure the ph of aqueous wastes and those multiphase wastes where the aqueous phase constitutes at least

More information

CCME Reference Method for the Canada-Wide Standard for Petroleum Hydrocarbons (PHC) in Soil - Tier 1 Method

CCME Reference Method for the Canada-Wide Standard for Petroleum Hydrocarbons (PHC) in Soil - Tier 1 Method Rating Guide Appendix for the CCME Reference Method for the Canada-Wide Standard for Petroleum (PHC) in Soil - Tier 1 Method (Note: Checklist incorporates requirements from Dec 2000 version + Addendum

More information

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY 1.0 SCOPE AND APPLICATION 1.1 This digestion procedure is used for the preparation

More information

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater. Nitrogen, Ammonia DOC316.53.01078 USEPA 1 Nessler Method 2 Method 8038 0.02 to 2.50 mg/l NH 3 N Reagent Solution Scope and application: For water, wastewater and seawater. Distillation is required for

More information

Substances and Mixtures:Separating a Mixture into Its Components

Substances and Mixtures:Separating a Mixture into Its Components MiraCosta College Introductory Chemistry Laboratory Substances and Mixtures:Separating a Mixture into Its Components EXPERIMENTAL TASK To separate a mixture of calcium carbonate, iron and sodium chloride

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical Oxygen Demand, Chemical DOC316.53.01103 USEPA Reactor Digestion Method Method 10211 1 to 60 mg/l COD (ULR) TNTplus 820 Scope and application: For wastewater, process water, surface water, and cooling water.

More information

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters) Chromium, Total DOC316.53.01034 Alkaline Hypobromite Oxidation Method 1 Method 8024 0.01 to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters) Scope and application: For water and wastewater.

More information

United States Office of Water Environmental Protection Agency (4303) February 2010

United States Office of Water Environmental Protection Agency (4303) February 2010 United States Office of Water Environmental Protection Agency (4303) February 2010 Method 1664, Revision B: n-hexane Extractable Material (HEM; Oil and Grease) and Silica Gel Treated n-hexane Extractable

More information

Chemical Reactions: The Copper Cycle

Chemical Reactions: The Copper Cycle 1 Chemical Reactions: The Copper Cycle ORGANIZATION Mode: pairs assigned by instructor Grading: lab notes, lab performance and post-lab report Safety: Goggles, closed-toe shoes, lab coat, long pants/skirts

More information

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD)

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD) 683 METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD) NOTE: This method does not include all of the specifications (e.g., equipment and

More information

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION METHOD 80 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION 1.0 SCOPE AND APPLICATION 1.1 Method 80 may be used to determine the concentration of acetonitrile (CAS No. 75-05-8) in aqueous

More information

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution Chloride DOC316.53.01017 Mercuric Thiocyanate Method Method 8113 0.1 to 25.0 mg/l Cl Reagent Solution Scope and application: For water and wastewater. Test preparation Instrument-specific information Table

More information

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

DR/4000 PROCEDURE NITRATE. Using Powder Pillows DR/4000 PROCEDURE Method 8171 Powder Pillows or AccuVac Ampuls Cadmium Reduction Method MR (0 to 5.0 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit

More information

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C Book 5 - Laboratory Policies and Procedures C03-1 2009-06-15 1 Scope 1.1 This method describes a general procedure for the determination of leachable arsenic, selenium, cadmium, antimony, and barium in

More information

Standard Operating Procedure VUSP N

Standard Operating Procedure VUSP N Villanova University Date Oct 2011 Page 1 of 14 Villanova University Villanova Urban Stormwater Partnership Watersheds Laboratory Standard Operating Procedure VUSP N Nitrite, USEPA by Discrete Analysis

More information

Method Total Kjeldahl Nitrogen in Water and Biosolids by Automated Colorimetry with Preliminary Semi-automatic Digestion. DRAFT January 2001

Method Total Kjeldahl Nitrogen in Water and Biosolids by Automated Colorimetry with Preliminary Semi-automatic Digestion. DRAFT January 2001 EPA-821-R-01-007 January 2001 Method 1688 Total Kjeldahl Nitrogen in Water and Biosolids by Automated Colorimetry with Preliminary Semi-automatic Digestion DRAFT January 2001 U.S. Environmental Protection

More information

Total Kjeldahl Nitrogen by Gas Diffusion and Flow Injection Analysis (FIA)

Total Kjeldahl Nitrogen by Gas Diffusion and Flow Injection Analysis (FIA) Methodology Total Kjeldahl Nitrogen by Gas Diffusion and Flow Injection Analysis (FIA) (Cartridge Part #A002380) 1.0 Scope and Application 1.1 This method is used for the determination of Total Kjeldahl

More information

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation Phosphorus, Total DOC316.53.01121 USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method 8190 0.06 to 3.50 mg/l PO 3 4 (0.02 to 1.10 mg/l P) Test N Tube Vials Scope and application: For water,

More information

GRIGNARD REACTION Synthesis of Benzoic Acid

GRIGNARD REACTION Synthesis of Benzoic Acid 1 GRIGNARD REACTION Synthesis of Benzoic Acid In the 1920 s, the first survey of the acceleration of chemical transformations by ultrasound was published. Since then, many more applications of ultrasound

More information

LEAD (Colorimetric) 2. Muffle Furnace: Equipped with pyrometer and capable of operating at controlled temperatures up to 500 C

LEAD (Colorimetric) 2. Muffle Furnace: Equipped with pyrometer and capable of operating at controlled temperatures up to 500 C LEADX.01-1 LEAD (Colorimetric) PRINCIPLE SCOPE Organic matter in the sample is destroyed by ignition in the presence of sulfuric acid. The residue is dissolved in dilute acid, and the lead is complexed

More information

METHOD 13B - DETERMINATION OF TOTAL FLUORIDE EMISSIONS FROM STATIONARY SOURCES (SPECIFIC ION ELECTRODE METHOD)

METHOD 13B - DETERMINATION OF TOTAL FLUORIDE EMISSIONS FROM STATIONARY SOURCES (SPECIFIC ION ELECTRODE METHOD) 874 METHOD 13B - DETERMINATION OF TOTAL FLUORIDE EMISSIONS FROM STATIONARY SOURCES (SPECIFIC ION ELECTRODE METHOD) NOTE: This method does not include all of the specifications (e.g., equipment and supplies)

More information

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS PAGE: 1 of 18 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Cyanide: The Molecule and Its Analysis A Simple Compound with Complex Problems Part 2: Cyanide Sample Collection, Pretreatment and Preservation

Cyanide: The Molecule and Its Analysis A Simple Compound with Complex Problems Part 2: Cyanide Sample Collection, Pretreatment and Preservation Cyanide: The Molecule and Its Analysis A Simple Compound with Complex Problems Part 2: Cyanide Sample Collection, Pretreatment and Preservation Edward F. Askew PhD Askew Scientific Consulting The first

More information

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) 1.0 SCOPE AND APPLICATION 1.1 Method 7060 is an atomic absorption procedure approved for determining the concentration of arsenic in wastes,

More information

Shaker Table Extraction

Shaker Table Extraction Title: Shaker Table Extraction Page 1 of 11 Shaker Table Extraction References: This standard operating procedure (SOP) is a performance-based method. This SOP describes the procedure as developed by Alpha

More information

DETERMINATION OF NITRATE IN BRACKISH WATERS BY FLOW INJECTION ANALYSIS NITRATE REDUCTASE

DETERMINATION OF NITRATE IN BRACKISH WATERS BY FLOW INJECTION ANALYSIS NITRATE REDUCTASE QuikChem Method 31-107-04-5-A DETERMINATION OF NITRATE IN BRACKISH WATERS BY FLOW INJECTION ANALYSIS NITRATE REDUCTASE (LOW FLOW METHOD) Written by Scott Tucker Applications Group Revision Date: 11 February

More information

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right. Barium DOC316.53.01315 Turbidimetric Method 1 Method 10251 2 to 100, 20 to 1000, 200 to 10,000 mg/l Ba (spectrophotometers) Powder Pillows 2 to 80, 20 to 800, 200 to 8000 mg/l Ba (colorimeters) Scope and

More information

Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES Version 2.

Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES Version 2. Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES 701.0 Version 2.3 July 2018 1. SCOPE AND APPLICATION This method is used for the determination

More information

Method to 0.50 mg/l NH 3 N Powder Pillows

Method to 0.50 mg/l NH 3 N Powder Pillows , 8155 Salicylate Method 1 Scope and Application: For water, wastewater and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) DOC316.53.01077 Method 8155 0.01 to 0.50 mg/l NH 3 N Powder Pillows

More information

Carbon (Organic, and Inorganic)

Carbon (Organic, and Inorganic) DATE: January 2011 Carbon (Organic, and Inorganic) 1. Application This method covers the determination of total organic carbon (TOC), and inorganic carbon (IC) concentrations in soil, plant tissues and

More information

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8032 is used to determine trace amounts of acrylamide monomer (CAS No. 79-06-1) in aqueous matrices. This method may be

More information

Working Copy. tlrq /rb I 'B-16. Determination of Gross Alpha/Beta Activity. Revision 0. Laboratory Manag erllqao/rso. Effective Date. lnitials.

Working Copy. tlrq /rb I 'B-16. Determination of Gross Alpha/Beta Activity. Revision 0. Laboratory Manag erllqao/rso. Effective Date. lnitials. SOP-RC.1O1 Revision 0 Laboratory Manag erllqao/rso Effective Date Date Date Renewal Date: lnitials. tlrq /rb I 'B-16 Texas lnstitute for Applied Environmental Research i. Identification of the method a.

More information

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

Scope and application: For wastewater, seawater, drinking water, surface water and process water. Cyanide DOC316.53.01495 USEPA 1 Pyridine Barbituric Acid Method 2 Method 10265 0.01 to 0.60 mg/l CN TNTplus 862 Scope and application: For wastewater, seawater, drinking water, surface water and process

More information

Method 548. July 1990 EPA EMSL-Ci. J.W. Hodgeson

Method 548. July 1990 EPA EMSL-Ci. J.W. Hodgeson Determination of Endothall in Drinking Water by Aqueous Derivatization, Liquid-Solid Extraction, and Gas Chromatography with Electron-Capture Detection July 1990 EPA EMSL-Ci J.W. Hodgeson Determination

More information

Method to 0.50 mg/l NH 3 N Powder Pillows

Method to 0.50 mg/l NH 3 N Powder Pillows , 8155 Salicylate Method 1 Scope and Application: For water, wastewater and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) DOC316.53.01077 Method 8155 0.01 to 0.50 mg/l NH 3 N Powder Pillows

More information

enable measurement. This method separates these isotopes effectively.

enable measurement. This method separates these isotopes effectively. Analytical Procedure URANIUM IN WATER 1. SCOPE 1.1. This is a method for the separation and measurement of uranium in water. After completing this method, source preparation for measurement of uranium

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical Oxygen Demand, Chemical DOC316.53.01104 USEPA Reactor Digestion Method Method 10212 250 to 15,000 mg/l COD (UHR) TNTplus 823 Scope and application: For wastewater and process waters; digestion is required.

More information

STANDARD OPERATING PROCEDURE No. 52 STATIC ACID GENERATION (NAG) TEST

STANDARD OPERATING PROCEDURE No. 52 STATIC ACID GENERATION (NAG) TEST Questa Rock Pile Stability Study SOP 52v6 Page 1 STANDARD OPERATING PROCEDURE No. 52 STATIC ACID GENERATION (NAG) TEST REVISION LOG Revision Number Description Date 52v0 Original SOP by STM 6/9/2004 52v1

More information

Basic Equipments and Instruments used in Chemistry laboratory: Balance: It is an instrument for measuring mass.

Basic Equipments and Instruments used in Chemistry laboratory: Balance: It is an instrument for measuring mass. Basic Equipments and Instruments used in Chemistry laboratory: Balance: It is an instrument for measuring mass. Pipettes: They are used to transfer of known volumes of liquids from one container to another.

More information

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials Nitrate, HR DOC316.53.01068 Chromotropic Acid Method Method 10020 0.2 to 30.0 mg/l NO 3 N (HR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific information

More information

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction 1. Application 1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction Parameters and Codes: EDB and DBCP, whole water recoverable, O-3120-90 Parameter (µg/l)

More information