1 Intelligent RP LC Column Selection & New Technologies Peru ctober 2016 Ricardo Martínez 2016 Waters Corporation 1

2 Agenda LC Fundamentals Columns Quality Particles & Ligands Column Efficiency ph in LC New Tech Columns Summary & Questions 2016 Waters Corporation 2

3 Escitalopram (ral Solution) Mobile Phase: Buffer 6.1g of monobasic potassium phosphate. To each L of this solution add 1.5 ml of triethylamine. Adjust with phosphoric acid to ph of 2.5. Mobile phase ACN/Buffer (32:68). Column: 4.6 x 250 mm, 5 um, column packed with L Waters Corporation 3

4 2016 Waters Corporation 4

5 Chromatogram Characteristics Resolution Sensitivity Precision Time Analysis Method mplicity Cost 2016 Waters Corporation 5

6 Tswett's Experiment (1903) Tall glass open column filled with sand-like particles (alumina, or chalk) Ground-up plant extract Poured into the column and saw colored bands develop as the extract percolated down thru the column Different compounds had separated Different Colored Bands Greek Chroma Graphy -- color -- writing/study of Note: Tswett in Russian means Color 2016 Waters Corporation 6 6

7 Chromatography Applications pharmaceutical proteomics clinical food safety environmental 2016 Waters Corporation 7

8 HPLC System for Purification 2016 Waters Corporation 8

9 Identification and Quantitation Compound Identification Based on Retention Time Acrylamide at 2.85 minutes 2016 Waters Corporation 9

10 2998 Photodiode Array Detector Features: Spectral Analysis Library Matching Spectral analysis algorithms can distinguish closely related compounds providing reliable library matching. Benefit: Peak Identification Peak 3 - Butalbarbital Library Match - Butalbarbital 2016 Waters Corporation 10

11 LC/MS/MS Instrument 2016 Waters Corporation 11

12 AU Imp. F Imp. A Imp. G API Imp Imp. H Imp. C Imp. D Imp. B PDA-MS Detection 0.04 ACQUITY Arc System Minutes Imp. F API Imp. A Imp. G UV Spectra MS Spectra Apex Apex Apex Apex 2016 Waters Corporation 12

13 Identification and Quantitation 10X Area Count Sample A has 10X the concentration of Acrylamide 2016 Waters Corporation 13

14 Normal vs Reversed-Phase Chromatography Normal Phase Reversed-Phase Stationary Phase Un-bonded lica (Polar) Surface Non-Polar Ligand (C18) Bonded to lica Surface Packing Polarity Polar Non-Polar Mobile Phase Polarity Non-Polar Polar Elution rder Most Non-Polar First Most Polar First Effect of Increasing Mobile Phase Polarity Reduces Retention Time Decreases Retention Time 2016 Waters Corporation 14

15 Waters Column Product History µbondapak Styragel Spherisorb DeltaPak PrepPak 1958 SymmetryShield XTerra Symmetry ACQUITY UPLC BEH SunFire Columns XTerraPrep XBridge ACQUITY UPLC BEH Amide ACQUITY UPLC BEH Glycan XBridge Amide XSelect HSS HPLC Columns ACQUITY UPLC HSS C 18 and HSS C 18 SB ACQUITY UPLC HSS Cyano & PFP columns XSelect TM HSS Cyano & PFP columns XP 2.5 µm Columns Nova-Pak ProteinPak TM Pico-Tag TM AccQTag TM Symmetry 300 Atlantis Atlantis HILIC lica Prep BD Intelligent Speed TM BioSuite NanoEase Atlantis T3 ACQUITY UPLC HSS T3 AccQTag TM Ultra BEH130 Columns BEH300 Columns XBridge HILIC ACQUITY UPLC BEH200 SEC XSelect CSH HPLC columns ACQUITY CSH Columns Viridis SFC Columns ProteinPak High Rs IEX ACQUITY UPLC BEH125 SEC ACQUITY UPC 2 Columns CRTECS Columns ACQUITY UPLC BEH450 SEC ACQUITY APC Columns CSH130 Columns 2016 Waters Corporation 15

16 Quality Systems Minimize Risk with Dependable Column Performance Method Validation kits provide three batches of chromatographic media [derived from different base particles] to judge the quality, reliability and consistency of an analytical method Waters uniquely positioned as an industry partner to minimize risk 2016 Waters Corporation 16

17 Benchmarking System Performance Principle of benchmarking: Use of a Quality Control Reference Material (QCRM) to evaluate or check key performance criteria by comparison with data generated when the system is known to be in good working order Importance of benchmarking Routine use on the analytical system and control charting the data allows for an understanding of the capability of your system to provide reliable results and is a useful troubleshooting tool. Typical Criteria 1. Retention time range or reproducibility 2. Peak area range or reproducibility 3. Peak tailing range 4. Peak resolution 5. Response 6. System Pressure 2016 Waters Corporation 17

18 What is a lanol Group? H H H H Comes from the silica gel particle (substrate) used to make reversed-phase packing materials 2016 Waters Corporation 18

19 lica Gel Pore Structure -H = lanol 2016 Waters Corporation 19

20 Making a Bonded Phase Material: Monofunctional Synthesis H + H 3 C Cl C CH 3 C8 lane Ligand C C C C C C CH 3 lica Gel Surface H 3 C C CH 3 C C C C C C + HCl CH Waters Corporation 20

21 C18 Bonded lica Gel Pore H 3 C CH 3 CCCCCCCCCCCCCCCCCC 25 Å Steric Hindrance 2016 Waters Corporation 21

22 C18 Bonded and Fully Endcapped lica Gel Pore H 3 C CH 3 CCCCCCCCCCCCCCCCCC 25 Å Steric Hindrance H 3 C CH 3 CH 3 Endcap (2 nd Bonding Step) What do you still see? lanols! 2016 Waters Corporation 22

23 Surface of a lica Gel Bonded-Phase Packing Material H 3 C C 8 alkyl chains H 3 C H 3 C H 3 C H 2 C CH 2 H 2 C CH 2 H 2 C CH 2 H 2 C CH 2 H 2 C CH 2 H 2 C CH 2 H 2 C CH 2 H 2 C CH 2 CH 2 CH 2 H 2 C H 2 C H 2 C residual silanol CH 2 CH 3 CH 2 H 3 C CH H H 3 C H 3 C 3 C 3 CH CH 3 3 H H H H CH 2 CH 2 H 2 C CH 2 endcap CH 2 H 3 C CH 3 CH 3 CH 3 H 3 C CH H 3 H H Note: ~ 30-50% of the surface silanols remain even with high bonding densities 2016 Waters Corporation 23

24 Why Do You See Poor Peak Shape? Answer: Ionization of lanols Surface silanol charge changes with mobile phase ph H H+ Behaves as a Cation Exchanger (ph 2) (ph 7) Result: Strong interaction (cation exchange) between ionized surface silanols (negative charge) and ionized basic analytes (positive charge) lanols are weak acids 2016 Waters Corporation 24

25 RP-Ion exchange Hydrophobic Interaction with Bonded Phase Mobile Phase ph < 3 - H - - H - - H - H H HN (CH 3 ) 2 Ion exchange Interaction with Charged tes High lanol Activity (CH) 2 HN 3 Mobile - Phase ph > 5 N Substrate Protonated - no charge Substrate De-protonated -- Negative Charge Base RP Cation X Base Increased retention and PR Peak Shape SAME CLUMN 2016 Waters Corporation 25

26 AU Why not just run at ph s less than 3? Selectivity ph Minutes A nortriptyline amitriptyline = A ph A A 2016 Waters Corporation 26

27 Different lica Gel Manufacturing Processes Precipitation of sodium silicate: (irregular particles)-µbondapak Sodium silicate polymerized to silicic acid (silica gel) then ground to fine particles SolGel 1 Process: (spherical particles)- Nova-Pak Colloidal silica spheres fused into chromatographic particle lgel 1 Process: (spherical particles)-symmetry Pure silanes polymerized to form chromatographic particle 2016 Waters Corporation 27

28 Types of Surface lanols Found on lica Gel Vicinal (Bridged) H H Geminal (lanediol) H H H Lone (most active) H 2016 Waters Corporation 28

29 Acidic Compounds:Two Different C18 s (Different Brands Same Ligand, cont.) Suprofen Ketoprofen and Tolmetin Note: Different selectivity due to different silica particle. Naproxen Col Brand A DS Suprofen Ketoprofen Naproxen Tolmetin Col Brand B DS Minutes 2016 Waters Corporation 29

30 Aluminum in the lica Gel Lattice Bronsted Acid 3D top view of silica particle surface with silanols pointing upward H H Al H Metal available for chelation 2016 Waters Corporation 30

31 Correlation Between Base Tailing and Aluminum Content of lica Gel Analyte: Chlorpheniramine Mobile Phase: Acetonitrile/KH 2 P 4 ph 3.0 (20:80) Tailing Factor "High Purity lica Gel" Region Aluminum Content, ppm 2016 Waters Corporation 31

32 Peak Shapes of Chelating Agent (Hinokitiol) Low metals M = metal M n+ High metals Minutes 2016 Waters Corporation 32

33 Test Compounds for the Measure of Residual lanol Activity N + Amitriptyline pka = 9.4 Basic Compound Acenaphthene Neutral Compound 2016 Waters Corporation 33

34 Impact of Choosing Two Different C18 Brands on Peak Shape for a Base Note: Different brands (different silica particle) have different silanol behaviors. Waters SunFire C18 Amitriptyline Waters Spherisorb DS Minutes Note: ph plays a role in peak shape 2016 Waters Corporation 34

35 Pore ze Most silica gel packings are porous >99% of the surface area is contained within the particle (not on the surface)- Where the chromatography happens. Rules of Thumb The smaller the pore size, the greater the surface area. o (100 Å approx. 300 m 2 /gram) o (300 Å approx. 100 m 2 /gram) The greater the surface area, the greater the retention. A typical 15 cm column holds a surface area of ~ square meters 2016 Waters Corporation 35

36 lica Gel Pore Structure Analyte MW Pore ze Recommendation < 3, Å (6-13 nm) 3,000 10, Å ( nm) >10, ,000 Å ( nm) Very Large Non-porous Analyte molecule Analyte needs to fit inside a pore to interact with chromatographic surface -H = lanol Note: Pore ze is a distribution 2016 Waters Corporation 36

37 Selectivity ln [α] amitriptyline/acenaphthene) Reversed-Phase Column Selectivity Chart 3.6 Waters Spherisorb S5 P Waters Spherisorb S5CN Nova-Pak CN HP ACQUITY UPLC BEH Phenyl XBridge Phenyl Inertsil Ph-3 Hydrophobicity (ln [k] acenaphthene) ACQUITY UPLC HSS PFP XSelect HSS PFP ACQUITY UPLC HSS C18 SB XSelect HSS C18 SB Waters Spherisorb DS1 Resolve C18 Hypersil Phenyl ACQUITY UPLC CSH Fluoro-Phenyl ACQUITY UPLC BEH C18 XSelect CSH Fluoro-Phenyl XBridge C18 Waters Spherisorb DS2 ACQUITY UPLC HSS CN µbondapak C18 ACQUITY UPLC HSS T3 XSelect HSS CN YMC-Pack YMC J'sphere XSelect HSS T3 Phenyl DS L80 Nucleosil C18 Hypersil CPS Cyano Inertsil CN-3 Nova-Pak Phenyl CRTECS C18+ Hypersil BDS Phenyl YMC J'sphere DS M80 Chromolith Nova-Pak YMC J'sphere DS H80 XTerra YMCbasic RP-18 C18 YMC-Pack DS AQ Phenyl YMC-Pack CN Nova-Pak Luna YMC-Pack Pro C4 Atlantis dc18 C8 Phenyl Hexyl Zorbax XDB C18 YMC-Pack Pro C8 Atlantis T3 ACQUITY UPLC BEH C8 XTerra MS C8 ACT Ace C18 Symmetry C8 YMC-Pack DS-A XBridge C8 Luna Luna C18 (2) C8 (2) YMC-Pack Inertsil DS-3 Pro C18 ACQUITY UPLC CSH Phenyl-Hexyl SunFire C18 SunFire C8 XTerra MS C18 XSelect CSH Phenyl-Hexyl Symmetry C18 SymmetryShield RP8 Zorbax SB C18 XTerra RP18 ACQUITY UPLC BEH Shield RP18 SymmetryShield RP18 XBridge Shield RP18 ACQUITY UPLC CSH ACQUITY UPLC HSS C18 XTerra RP8 XSelect CSH C18 XSelect HSS C18 YMC-Pack PolymerC CRTECS C Waters Corporation 37

38 Creating a New Particle Inorganic (licon) Advantages Mechanically strong High efficiency Predictable retention Disadvantages Limited ph range Tailing peaks for bases Chemically unstable Polymer (Carbon) Wide ph range No ionic interactions Chemically stable Mechanically soft Low efficiency Unpredictable retention Hybrid (licon-carbon) Particle Technology 2016 Waters Corporation 38

39 2016 Waters Corporation 39

40 1 st Generation Hybrid (Methylloxane/lica) Particles Methyl Groups on Hybrid Surface (Better Peak Shape) and in Hybrid Particle (High ph Life-time) Waters Patented technology US Patent: 6,686,035 B2 Date of Patent: Feb. 3, 2004 MethylPolyethoxysilane (MPES) Tetraethoxysilane (TES) Methyltriethoxysilane (MTES) 2016 Waters Corporation 40

41 2016 Waters Corporation 41

42 2016 Waters Corporation 42

43 Ethylene Bridged Hybrid [BEH] Particle U.S. Patent No. 6,686,035 B2 Bridged Ethanes within a silica matrix Anal. Chem. 2003, 75, Waters Corporation 43

44 2016 Waters Corporation 44

45 The Resolution Equation N α 1 k Rs 4 α k 1 Mechanical Contributions Ultra-low dispersion system perate at optimal linear velocity Particle morphology Small particles Well-packed columns Chemical/Physical Contributions Complementary bonded phases Multiple particle substrates Ability to utilize high ph Increase retentivity 2016 Waters Corporation 45

46 The Resolution Equation N α 1 k Rs 4 α k 1 Mechanical Contributions Ultra-low dispersion system perate at optimal linear velocity Particle morphology Small particles Well-packed columns Chemical/Physical Contributions Complementary bonded phases Multiple particle substrates Ability to utilize high ph Increase retentivity 2016 Waters Corporation 46

47 Column Length and Mechanical Separating Power* Columns contain the same packing material, same particle size and same mobile phase, only one is twice as long Additional column length does provide a better separation. However, several costs are incurred: more time (2X) for the analysis, use more solvent, increased back pressure and the longer column costs more to buy. A better approach, would be to try a different particle chemistry/mobile phase combination or a smaller particle size that can create the separation in less time. * This is also called Efficiency 2016 Waters Corporation 47

48 Particle ze and Mechanical Separating Power* Columns contain the same packing material chemistry, are the same length with the same mobile phase. ne column has particles which are a third the size. Smaller particle sizes provide for a better separation with the same run time. However, back pressure will increase. * This is also called Efficiency 2016 Waters Corporation 48

49 N = Theoretical Plates A Measure of Efficiency 2 Variances Peak Width/Band Spread = 2 extra column + 2 column Analyte Band Flowing Through Detector Cell The narrower the peak (W), the higher the Plate Count, N Elution Volume Tangents 2016 Waters Corporation 49

50 Calculation of Column Efficiency 5 Method N V R plates V R N plates 4.4% height W 1.03 Good Column 4.4% height W 1.44 Bad Column 2016 Waters Corporation 50

51 Waters Particle Technology 60 μm Human Hair (very fine hair) 5 μm Analytical Particles (can fit 12 across hair) Images are on same scale (Bar = 10 μm) 1.7 μm ACQUITY UPLC Particles ptimal Particle ze Distribution For Max Efficiency at a given Pressure 2016 Waters Corporation 51

52 Plate Count (4 sigma) Efficiency vs. Flow Rate Acenaphthene, 2.1 x 50 mm columns 70/30 MeCN/H 2, 30 C, 254 nm µm ACQUITY UPLC BEH C µm XBridge C µm Superficially Porous C Flow Rate (ml/min) 2016 Waters Corporation 52

53 Plate Height (H, 4 sigma) van Deemter Curves Plate Height (Non-reduced, H) 25 Acenaphthene, 2.1 x 50 mm columns 70/30 MeCN/H 2, 30 C, 254 nm 1. 7 µm ACQUITY UPLC BEH C µm XBridge C µm Superficially Porous C Linear Velocity, u (cm/sec) 2016 Waters Corporation 53

54 UPLC Technology & The Fundamental Resolution Equation Rs N 4 Physical ( α 1)( k ) α k 1 Chemical In UPLC systems, increasing N (efficiency) is the primary focus Selectivity and retentivity are the same as in HPLC Resolution, Rs, is proportional to the square root of N Rs N If N 3x, Rs 1.7x 2016 Waters Corporation 54

55 AU AU Constant Column Length Flow Rate Proportional to Particle ze µm, 0.6 ml/min, 7656 psi Minutes 4.8 µm, 0.2 ml/min, 354 psi Minutes 2016 Waters Corporation Reality 1.5X Resolution 2.6X Faster 1.4X Sensitivity 22X Pressure Theory 1.7X Resolution 3X Faster 1.7X Sensitivity 25X Pressure 2.1 x 50 mm columns Very High System Pressure

56 Resolution (and Speed) Constant Column Length Plates, Flow Rate and Particle ze Rs N Smaller Particle ptimal Larger Particle Smaller Particle ze *Increased N, *Higher, optimal u *Increased pressure Flow Rate {ml/min} ptimal flow rate is inversely proportional to dp F opt 1 dp Isocratic analysis time is inversely proportional to F dp 3X, N 3X, Rs 1.7X, T 3X (e.g., 5 μm to 1.7 μm) (Rs α N) 2016 Waters Corporation 56

57 Speed Increases Constant L/dp Efficiency, N, is directly proportional to column length, L, and inversely proportional to particle size, dp: N L dp For same N and, therefore, same Rs dp 3X, L 3X, N = 1X, Rs = 1X, (e.g., 5 μm to 1.7 μm) (e.g., 150 mm to 50 mm) F 3X, T 9X (i.e., F increases 3X, L decreases 3X) Efficiency & Resolution Remain Unchanged 2016 Waters Corporation 57

58 AU AU Length Proportional to Particle ze milar L/dp µm, 30 mm, 0.6 ml/min Minutes 4.8 µm, 100 mm, 0.2 ml/min Minutes 2.1 mm ID columns Reality Same Resolution 8X Faster 2.5X Sensitivity 11X Back Pressure Theory Same Resolution 9X Faster 3X Sensitivity 9X Back Pressure Manageable Backpressure Increase 2016 Waters Corporation 58

59 Ratio of Column Length-to-Particle ze: Maintaining Separation Power Typical HPLC Column 4.6 x 150 mm, 5 µm Column length (L) = 150 mm = 150,000 µm d p = 5 µm L = 150,000 = 30,000 d p 5 Separation Index Application Example Efficiency (N) L/d p Easy Content Uniformity 5,000 15,000 Moderately Challenging Related compound assay 12,000 30,000 Difficult Impurity profiling 20,000 50,000 Extremely Difficult Metabolite identification 35,000 85, Waters Corporation 59

60 30 mm 50 mm 30 mm 75 mm 50 mm 30 mm 100 mm 75 mm 50 mm 30 mm 150 mm 100 mm 75 mm 50 mm L/dp Ratio 150 mm 100 mm 75 mm 250 mm 150 mm 100 mm 250 mm 150 mm Ratio of Column Length to Particle ze Resolution Capability mm, 1.7 µm 75 mm, 2.5 µm 100 mm, 3.5 µm 150 mm, 5 µm L/d p ~ 30,000 Extremely Difficult Difficult Impurity Moderately Challenging Easy particle size (µm) Waters Corporation 60

61 Improved Productivity Well packed columns containing smaller particles have more intrinsic efficiency than columns packed with larger particles. N α L dp The nce Typical 4.6 x 100 mm, 3.5 µm Column L = 100,000 µm = 28,571 d p 3.5 µm Replaced with 2.1 x 50 mm, 1.6 µm solid-core Column L = 50,000 µm = 31,250 d p 1.6 µm 2016 Waters Corporation 61

62 Proper Method Transfer ACQUITY CSH Phenyl-Hexyl 50 mm, 1.7 µm α 3,5 = UPLC XSelect CSH Phenyl-Hexyl 100 mm, 3.5 µm α 3,5 = HPLC XSelect CSH Phenyl-Hexyl 150 mm, 5 µm α 3,5 = HPLC/Prep Waters Corporation 62

63 Defining the LC Categories by their System Dispersion How are these categories differentiated? Chromatographic Resolution Increases verall Run Time Decreases Method Sensitivity Increases 2016 Waters Corporation 63

64 What is at the Root of the Performance Differences across the LC Categories? Dispersion n. Broadening of an analyte band due to both on-column effects (diffusion and mass transfer kinetics which are both dependent on particle size and linear velocity) and system effects (tubing internal diameter (I.D.) and length, connections, detector flow cell volumes, etc.) True separation performance is governed by the system dispersion paired with a flow rate range that yields the highest possible efficiency for a given analytical column 2016 Waters Corporation 64

65 Maximizing the Separation Performance based on System Dispersion Determine your systems Dispersion Matching the right LC system to the right column will yield the best chromatographic results 2016 Waters Corporation 65

66 Maximizing the Separation Performance based on System Dispersion Select the appropriate particle size to match the HPLC, UHPLC, or UPLC system 2016 Waters Corporation 66

67 Maximizing the Separation Performance based on System Dispersion Pair the particle size with the column i.d. that best matches the dispersion of your chromatographic system 2016 Waters Corporation 67

68 Maximizing the Separation Performance based on System Dispersion Select a flow rate that gives the optimal linear velocity to maximize efficiency for your column characteristics (van Deemter) 2016 Waters Corporation 68

69 Maximizing the Separation Performance based on System Dispersion The system must be able to operate at the typical back pressures associated with the selected column 2016 Waters Corporation 69

70 Maximizing the Separation Performance based on System Dispersion The relative cost/per analysis will decrease as you move from HPLC to UHPLC to UPLC due to shorter run times at lower flow rates 2016 Waters Corporation 70

71 Maximizing the Separation Performance based on System Dispersion HPLC UHPLC UPLC 2016 Waters Corporation 71

72 4.6 x 75 mm, 2.7 µm AU AU AU 3.0 x 75 mm, 2.7 µm AU AU AU 2.1 x 50 mm, 1.6 µm AU AU AU Dispersion Impact on Performance: Isocratic Separations on HPLC, UHPLC and UPLC 1.00 HPLC 1.00 UHPLC 1.00 UPLC Minutes Minutes Minutes Minutes Minutes Minutes * * Minutes Minutes 2016 Waters Corporation Minutes * Strong solvent effects

73 4.6 x 75 mm, 2.7 µm AU AU AU 3.0 x 75 mm, 2.7 µm AU AU AU 2.1 x 50 mm, 1.6 µm AU AU AU Dispersion Impact on Performance: Isocratic Separations on HPLC, UHPLC and UPLC 1.00 HPLC 1.00 UHPLC 1.00 UPLC Minutes Minutes Minutes Minutes Minutes Minutes * * Minutes Minutes 2016 Waters Corporation Minutes * Strong solvent effects

74 What Causes Poor Peak Shape with Basic Compounds? lanol Interactions with Basic Compounds 2016 Waters Corporation 74

75 Impact of poor peak shape in methods development Integration errors Reduced resolution Reduced sensitivity Minutes 2016 Waters Corporation 75

76 Influence of Poor Peak Shape on Integration Tailing Factor = 1.00 Recovered Peak Areas 99.9 % 99.8 % 99.6 % Tailing Factor = 1.58 Recovered Peak Areas 97.8 % 95.3 % 92.3 % 2016 Waters Corporation 76

77 Influence of Poor Peak Shape on Resolution Rs = (t r Peak 2 t r Peak 1)/ 0.5 (w 4.4% Peak 1 + w 4.4% Peak2) Rs = ( )/ 0.5 ( ) Rs = 0.72 Note: Baseline resolution = Minutes 2016 Waters Corporation 77

78 Tamoxifen: Influence of Poor Peak Shape on Sensitivity AU C18 Brand A gnal/noise Ratio A 11.0 B 6.5 AU Minutes 5 10 Minutes C18 Brand B Note: gnal/noise Requirements LQ>10 LD> Waters Corporation 78

79 Factors Affecting the Separation Stationary Phase Factors Type of bonded ligand Ligand Density Concentration of the ligand on the surface of the particle Ligand High Density Low Density Mobile Phase Factors rganic strength adjusted to create appropriate elution times 2016 Waters Corporation 79

80 Surface of a lica Gel Bonded-Phase Packing Material H 3 C H H 2 C H 3 C H H 3 C H 2 C H 2 C H 2 C H 3 C H 2 C H 2 C CH 2 Polar analytes are not able to energetically fit between ligands can t interact with surface or ligands CH 2 CH 2 CH 2 CH 2 CH 2 CH 2 CH 2 CH 2 CH 2 CH 2 H 2 C H 2 C H 2 C CH 2 CH 2 CH 2 Residual silanol Endcap H 2 C H 2 C H 2 C CH 2 CH 2 CH 2 CH 3 CH H 3 3 C H 3 C CH 3 CH H 3 C CH 3 H 3 C 3 CH 3 CH H 3 3 C H H H H H H C 8 alkyl chains CH 3 H 3 C H CH 3 H 3 C H 2 C CH 2 Polar analytes easily interact with surface and ligands H 3 C H 2 C CH 2 Residual silanols H 3 C H 2 C H 2 C H 2 C H 3 C H 2 C CH 2 CH 2 CH 2 CH 2 CH 3 CH 2 H 3 C H 3 C CH 3 CH 3 H CH 3 CH H 3 C 3 H H H H H High Coverage High Ligand Density Low Coverage Low Ligand Density Endcap CH 3 CH Waters Corporation 80

81 Surface of a lica Gel Bonded-Phase Packing Material H 2 C Non-polar portion of analyte interacts with bonded phase H 2 C (hydrophobic) H 3 C H H 2 C H 2 C CH 2 CH 2 CH 2 CH 2 Alkyl chains H CH 3 C 3 H N H NH CH 3 Residual silanols H 3 C CH 3 H Polar portion of analyte interacts with silanols (hydrogen bonding and/or ion exchange) CH 3 CH 3 H Low Coverage Low Ligand Density ~ Waters Corporation 81 N H NH 2 N H H 3 C H H 2 C H 2 C H 2 C H 2 C CH 2 CH 2 CH 2 CH 2 CH 3 H H 3 C Endcap H CH 3 CH 3

82 What is De-Wetting Hydrophobic Collapse? A: Loss of reversed-phase retention Low organic or pure aqueous mobile phase is expelled from pores (de-wetted) C18 lica-gel Wetted Pore De-Wetted Pore Note: Retentivity is a function of the surface area and ligand density. However, if the surface is non-wetted, then the effective chromatographic surface area is reduced > 95%. Thus reducing the retentivity of the analyte => poor capture = De-wetting (Hydrophobic collapse) Remember! almost all of the surface area is in the pores! 2016 Waters Corporation 82

83 AU AU AU AU HSS T3 Versus HSS C18: 100% Aqueous Conditions XSelect HSS T3 5 Initial Pks Minutes Peak i.d.: 1) thiourea 2) 5-Fluorocytosine 3) adenine 4) thymidine-5 -monophosphate 5) thymine Conditions: 10mM Ammonium Formate ph 3; 0.2mL/min; 30ºC; 2.1 x 50 mm Waters Corporation 83 5 After dewetting XSelect HSS C18 Initial After dewetting

84 Excellent Peak Shape and Retention with mple MS Compatible Mobile Phases The popular void marker Uracil is well retained and is actually peak 3 Conditions Column: Atlantis TM dc x 150 mm, 5 µm Mobile Phase A: H 2 Mobile Phase B: ACN Mobile Phase C: 100 mm CH 3 CNH 4, ph 5.0 Flow Rate: 1.0 ml/min Gradient: Time Profile (min) %A %B %C Injection Volume: 10 µl Temperature: 30 o C 1 Detection: 254 nm Instrument:: Alliance V 0 = 1.83 TM 2695, min 2996 PDA Compounds: 1. Cytosine 2. 5-Fluorocytosine 3. Uracil 4. 5-Fluorouracil 5. Guanine 6. Thymine 7. Adenine 5 6 Note excellent peak shape for strong polar base Adenine at ph Minutes At ph 5.0, the strong polar base Adenine still exhibits excellent peak shape (reason: fully endcapped) Grumbach, Diehl 2016 Waters Corporation 84

85 Perfect Balance Between Polar and Non-Polar Compound Retention More retention for polar compounds 7 Equal or less retention for non-polar compounds V 0 = 0.98 min Atlantis dc Minutes Conventional C 18 7 Back Marker V 0 = 0.98 min Minutes 2016 Waters Corporation 85

86 Increased Polar Compound Retention with ACQUITY UPLC HSS T3 Columns Compounds: Conc (mg/ml) 1. Norepinephrine Epinephrine Dopamine ,4- Dihydroxyphenylacetic acid Serotonin (5-HT) Hydroxy-3-indoleacetic acid Hydroxy-3-methoxyphenylacetic acid (HVA) Conditions: Columns: As Indicated Mobile Phase A: 10mM CH 3 CNH 4, ph 5.0 Mobile Phase B: ACN Flow Rate: ml/min Isocratic Mobile Phase Composition: 2% B Injection Volume: 0.7 µl Temperature: 30 o C Detection: 280 nm Instrument: ACQUITY UPLC TM System with ACQUITY UPLC 2996 PDA 7 ACQUITY UPLC BEH C Minutes NEW ACQUITY UPLC HSS T3 6 7 Increased Retention Minutes 2016 Waters Corporation 86

87 Polar Retention: Why Does HSS T3 Work? Dominant retention mechanism is reversed-phase (van der Waals forces hydrophobic attraction) Retention maximized using 100% aqueous mobile phases Retention maximized by using reduced C 18 coverage o Polar analytes can fit between C 18 ligands and interact with pores of material o ptimized particle morphology (i.e. pore diameter/volume) Secondary interactions due to residual silanols that are more accessible due to reduced C 18 coverage Cation-exchange interactions Hydrogen bonding interactions 2016 Waters Corporation 87

88 Different C18 Bonding Levels: Carbon Load Measured as weight of bonded phase per weight of silica (w/w %) Typical range for C18 ligand : 6 to 22% Carbon Load Retention 2016 Waters Corporation 88

89 Different Bonding Levels Packing Column Volume (ml) lica Density (g/ml) Mass of Packing (g) Carbon Load (%) Carbon in Column (g) Nova-Pak C18 µbondapak C Waters Corporation 89

90 (ln [a] amitriptyline/acenaphthene) Column Selectivity Comparison Chart (Methanol) Decreasing lanol Activity % with 1.42 µmoles/m 2 µbondapak C18 Waters Spherisorb DS2 11.5% with 2.84 µmoles/m Nova-Pak C18 7.3% with 2.80 µmoles/m Increasing Hydrophobicity Symmetry C18 20% with 3.20 µmoles/m (ln [k] acenaphthene) 2016 Waters Corporation 90

91 (ln [a] amitriptyline/acenaphthene) Column Selectivity Comparison Chart Decreasing lanol Activity Sph1DS-Low 9% Waters Spherisorb DS2 µbondapak C18 14% Sph2DS-Med Sph3DS-High 0.8 Nova-Pak C18 22% 0.3 Increasing Hydrophobicity Symmetry C (ln [k] acenaphthene) 2016 Waters Corporation 91

92 (ln [a] amitriptyline/acenaphthene) Selectivity Chart NTE: Do you see two columns that look like they are out of place? Hint: upper left quadrant C18 s Resolve C18 Waters Spherisorb DS µbondapak C18 Waters Spherisorb DS2 YMC J'sphere DS L80 Nucleosil C18 YMC J'sphere DS M80 YMCbasic Hypersil DS YMC J'sphere DS H80 YMC-Pack Pro C4 Inertsil C8 Nova-Pak C18 Nova-Pak C8 Kromasil C8 YMC-Pack DS AQ YMC-Pack Pro C8 YMC-Pack DS-A Symmetry C8 Hypersil BDS C8 YMC-Pack Pro C18 Kromasil C18 Inertsil DS-3 Hypersil BDS C18 Inertsil DS-2 Symmetry C (ln [k] acenaphthene) 2016 Waters Corporation 92

93 Base-Deactivated Packings What are they? A class of reversed-phase packing materials that exhibit improved peak shape for basic compounds---especially at ph s 4-7 How are they made? Various techniques in silica gel synthesis and/ or in the bonding process that yields a particle surface with less silanol interaction with basic analytes 2016 Waters Corporation 93

94 Base-Deactivated Packings a. Metal Impurities (silica gel) a. Lower concentration of Al 3+ and Fe 3+ b. Ligand Density c. Mixed C18 and Amino Ligands d. Embedded polar ligands (advanced ligands) 2016 Waters Corporation 94

95 Acid, Base, Neutral Chromatographic Evaluation Test at ph 3.0 H N + H 3 C H CH 3 N + Cl + H H N H HC - C H H CH 3 NH 2 Chlorpheniramine pk a = 3.6, 9.2 Propranolol pk a = 9.5 Maleic Acid pk a = 1.9, 6.3 Toluamide Base: Positive Charge Base: Positive Charge Acid: Negative Charge Neutral: No Charge Mobile Phase: 50 mm H3P4-KH2P4 ph 3.0/Acetonitrile (80/20) 2016 Waters Corporation 95

96 Waters Spherisorb DS2 Standard C18 Mobile Phase: 50 mm H3P4-KH2P4 ph 3.0/Acetonitrile (80/20) Maleic Acid Toluamide Chlorpheniramine Propranolol Minutes 2016 Waters Corporation 96

97 Repulsion of Cations with Technique C - + NH 3 (CH 2 ) 11 CH 3 Reduced silanol effect on basic analyte resulting in better peak shape H 3 C H N CH 3 + Repulsion of cationic base resulting in shorter retention times H N + Cl Waters Waters Corporation 97

98 Propranolol Waters Spherisorb DS2 versus Waters Spherisorb DSB (same silica gel) Acenaphthene Base-Deactivated DSB Technique C Acenaphthene Standard C18 Ligand DS2 Amitriptyline Minutes Waters Corporation 98

99 Maleic Acid Toluamide Technique C: Pros and Cons (Better for Bases Poor for Acids) Mobile Phase: 50 mm H3P4-KH2P4 ph 3.0/Acetonitrile (80/20)? Waters Spherisorb DSB Base Deactivated Waters Spherisorb DS2 Chlorpheniramine Propranolol Minutes 2016 Waters Corporation 99

100 Maleic Acid Toluamide Technique C: Pros and Cons Mobile Phase: 50 mm H3P4-KH2P4 ph 3.0/Acetonitrile (80/20) Maleic Acid Waters Spherisorb DSB Base Deactivated Waters Spherisorb DS2 Propranolol Minutes 2016 Waters Corporation 100

101 Making an Embedded Polar Bonded Phase Material: Two-Step Synthesis H H H H H H H H H + Cl Cl N H NH 2 NH 2 C NH 2 Cl + Amide (CH 2 ) n Cl CH 3 C Amination Step with trifunctional silane (CH 2 ) n CH 3 Acylation Step Note: Residual amine on particle surface due to steric hindrance (incomplete acylation) Waters Corporation 101

102 Characteristics of Two-Step Synthesis Process Inexpensive process: aminopropyl silane and acyl chlorides readily available Incomplete secondary reaction due to steric hinderance resulting in the presence of an aminopropyl group Packing materials show anion exchange properties that are undesirable (unusual retention of acidic compounds) Batch to batch reproducibility issues 2016 Waters Corporation 102

103 Toluamide Chlorpheniramine Propranolol Two-Step Synthesis Embedded Polar Bonded Phase Column Mobile Phase: 50 mm H3P4-KH2P4 ph 3.0/Acetonitrile (80/20) Note: Great Peak Shape for the Bases Maleic Acid? Minutes Note: Technique D using Two-Step Synthesis Process with residual amines Waters Corporation 103

104 Chlorpheniramine Propranolol Two-Step Synthesis Embedded Polar Bonded Phase Column Mobile Phase: 50 mm H3P4-KH2P4 ph 3.0/Acetonitrile (80/20) Maleic Acid Minutes 2016 Waters Corporation 104

105 Toluamide Embedded Polar Ligand: Two Step Synthesis Note: milar results to Technique C: Mixed C18 and Amino Ligands Mobile Phase: 50 mm H3P4-KH2P4 ph 3.0/Acetonitrile (80/20) Chlorpheniramine Propranolol Supelcosil LC-ABZ+Plus Technique D: Two-Step Maleic Acid Waters Spherisorb DSB Technique C Minutes 2016 Waters Corporation 105

106 Making an Embedded Polar Bonded Phase Material: ne-step Synthesis H + Monofunctional silane H 3 C Cl CH 3 C N H (CH 2 ) 7 CH 3 lica Gel Surface H 3 C CH 3 C N H (CH 2 ) 7 + HCl Carbamate group built into starting silane CH 3 Note: No residual amines on silica gel surface 2016 Waters Corporation 106

107 Possible Mechanisms of Embedded Polar Ligand Me Me C H N C 8 H 17 Analyte interacting with polar group H X R Me H Me HN C 8 H 17 Analyte competing with polar group for silanol H X R Me Me H H C H N C 8 H 17 H X R Polar group increases H 2 concentration in surface layer 2016 Waters Corporation 107

108 Embedded Polar Ligand: Possible Mechanism Polar group increases water concentration in surface layer - H 3 C CH 3 H H C N H (CH 2 ) 7 CH Waters Corporation 108

109 Water Layer Embedded Polar Ligand: Possible Mechanism Reason 1 Water layer increases dielectric constant, reducing ionic interactions Particle Reduced retention of bases Reduced peak tailing Both reasons given describe why the Shield RP18 boning provides excellent peak shape observed for bases Reason 2 The carbamate group may hydrogen bond with silanol groups, shielding analytes from interacting with them 2016 Waters Corporation 109

110 Maleic Acid Toluamide Chlorpheniramine Propranolol SymmetryShield RP18 Mobile Phase: 50 mm H3P4-KH2P4 ph 3.0/Acetonitrile (80/20) Minutes Technique D: ne-step process (great peak shape for acids and bases) 2016 Waters Corporation 110

111 Embedded Polar Ligand versus Linear Alkyl Ligand on lica Gel Amitriptyline SymmetryShield RP18 TF USP = 1.1 Symmetry C18 TF USP = Minutes Note: Reduced retention, 25 > 15 min. and improved peak shape for the base Waters Corporation 111

112 Acidic Compounds:Embedded Polar versus Alkyl Linear Ligands S S K T K N T N F F D In D In Ib Symmetry C18 Suprofen = S Ketoprofen = K Naproxen = N Tolmetin = T Fenoprofen = F Diclofenac = D Indomethacin = In Ibuprofen = Ib SymmetryShield RP18 Ib Minutes Note: No change in mobile phase Waters Corporation 112

113 Chromatographic Impact of 1990 s Embedded Polar Ligands Reduced retention and decreased tailing of basic compounds Unique selectivity Improved water wettability of the chromatographic bed (hydrogen bonding of water to the particle surface) and reduced risk of hydrophobic collapse 2016 Waters Corporation 113

114 Percent Un-Ionized Retention Factor (k) Reversed-Phase Retention Map: The Impact of ph on Ionizable Compounds Maximum acidic compound retention range Maximum basic compound retention range Acid Neutral Base ph ph Range for lica Particles ph Range for lica Particles ph Range for Hybrid Particles 2016 Waters Corporation 114

115 Capacity Factor (k) Relating Retention Maps to Chromatography Reversed-Phase Retention Map Neutral B A N ph Acid Base ph 2016 Waters Corporation 115

116 Capacity Factor (k) Relating Retention Maps to Chromatography Reversed-Phase Retention Map Acid Base Neutral ph B A N ph A,B N ph Waters Corporation 116

117 Capacity Factor (k) Relating Retention Maps to Chromatography Reversed-Phase Retention Map Acid Base Neutral ph A B A,B A B N N N ph 2 ph 5 ph Minutes Note: ph is a powerful selectivity tool 2016 Waters Corporation 117

118 Capacity Factor (k) Impact of ph on the Retention of a Zwitterionic Compound Aqueous pk a = 4.3 Aqueous pk a = Negative Charge Positive Charge B + A Dual Charge B A H N H + B + A - ph Fexofenadine (Antihistamine) CH 3 H CH 3 C Waters Corporation 118

119 Tailing Factor Peak Shape ver Wide ph Range (Strong Base -- Amitriptyline pka 9.4) 4 Low ligand density and high metal content silica gel: Conventional C High ligand density and high purity silica gel: Modern C Buffer ph Ideal Behavior Pure Polymer No lanols 2016 Waters Corporation 119

120 AU AU AU The Importance of Mobile Phase ph: Rapid Method Development ph Minutes ph Minutes ph Minutes Using a wide mobile phase ph range is an effective approach to change compound selectivity Increase selectivity for: Acids (Green/Brown) Bases (Red/Yellow) Neutrals (Peaks 2 & 5) are largely unaffected by mobile phase ph 2016 Waters Corporation 120

121 General Retention and Mobile Phase ph Rules To achieve stable chromatographic retention times Stay outside the ± 2 ph units from the analytes pk a value Always measure ph before adding the organic solvent 2016 Waters Corporation 121

122 Retention Time Variability Influence of ph Non-Column Influences: ph Neutrals: No Influence Acids: Reduced Retention with Increasing ph Bases: Increased Retention with Increasing ph Up to 10% Change in Retention per 0.1 ph Unit (largest shift within +/- 1 ph unit of pka) 2016 Waters Corporation 122

123 log (k') Impact n Retention Factor (k) By Changing Mobile Phase ph -- HPLC In REVERSED-PHASE Mode 100 Acid (Un-Ionized) 10 Note: Column Particle,Temperature and % rganic Held Constant ph Limit of lica gel Base (Un-Ionized) Neutral Base (Ionized) ph Acid (Ionized) Range of poor reproducibility due to steep slopes - small ph change -- large retention change 2016 Waters Corporation 123

124 ph Test for Robustness Poor Results Conditions Column: XTerra RP x 100 mm, 5 µm Mobile Phase A: Mobile Phase B: Flow Rate: 20mM Ammonium Acetate (ph 5.0 to 5.8) or 20mM Ammonium Bicarbonate (ph 6.8 to 7.0) ACN 0.5 ml/min Isocratic Mobile Phase Composition: 40% A; 60% B Injection volume: 10µL Temperature: Detection: Instrument: 30 o C 230 nm Alliance TM 2690, 996 PDA ph ph ph Method developed at ph = 5.0 Zone of reproducibility risk for ph control What is peak 1? 2? 3? ph Grumbach, Diehl 2016 Waters Corporation 124

125 ph Test for Robustness Poor Results Conditions Column: XTerra RP x 100 mm, 5 µm Mobile Phase A: Mobile Phase B: Flow Rate: 20mM Ammonium Acetate (ph 5.0 to 5.8) or 20mM Ammonium Bicarbonate (ph 6.8 to 7.0) ACN 0.5 ml/min Isocratic Mobile Phase Composition: 40% A; 60% B Injection volume: 10µL Temperature: Detection: Instrument: 30 o C 230 nm Compounds 1. p-toluamide -neutral 2. Lidocaine - base 3. Ibuprofen- acid Alliance TM 2690, 996 PDA Method developed at ph = 5.0 Zone of reproducibility risk for ph control What is peak 1? 2? 3? ph ph ph ph Grumbach, Diehl 2016 Waters Corporation 125

126 log (k') Impact on Retention Factor (k) on Robustness By Changing Mobile Phase ph Range - HPLC In Pure REVERSED-PHASE Mode 100 ph Limit of lica gel 10 Acid (Un-Ionized) Base (Un-Ionized) Neutral 1 Base (Ionized) Acid (Ionized) ph Greatly improved reproducibility due to flattening of curves Robustness Note: Column Particle, Temperature and % rganic Held Constant 2016 Waters Corporation 126

127 ph Test for Robustness Good Results 2016 Waters Corporation 127

128 log (k) Impact of Retention Factor (k) on Prep LC In Pure REVERSED-PHASE Mode ph Limit of lica gel Acid (Un-Ionized) Base (Un-Ionized) Neutral 1 Base + (Ionized) Acid - (Ionized) Maximum Loading for Prep is when analyte is in the UN-INIZED Form ACID at Low ph BASE at High ph ph Note: Column Particle, Temperature and % rganic Held Constant 2016 Waters Corporation 128

129 Superior Mass Load Capacity without Sacrificing Purity or Speed! ph 10 Loadability Study a Mixture of Three Basic Compounds XBridge 82 x In Un-ionized form! 2016 Waters Corporation 129

130 (ln [a] amitriptyline/acenaphthene) Basic Compounds W aters Spherisorb DS1 Resolve C µbondapak C18 W aters Spherisorb DS2 1.3 YMC J' sphere DS L80 Amine additives required with buffers N ucleosil C18 YMC J' sphere DS M80 Hypersil DS YMC J' sphere DS H N ova-pak C18 YMC-Pack DS AQ 0.3 Hypersil BDS C18 YMC-Pack Pro C18 Buffers required Kromasil C18 Inertsil DS-3 Inertsil DS-2 Symmetry C (ln [k] acenaphthene) 2016 Waters Corporation 130

131 Impact of rganic Solvent Composition on Retention Factor (k), ph, and pk a k k HA,0 e B HA c k 1 10 A,0 pk e B a c HA c ph 10 c pk a c ph c rganic Solvent Composition c Retention Factor k Note: k, pk a and ph are a function of the organic solvent composition Slope of relationship between Log k and % rganic B 2016 Waters Corporation 131

132 Impact of rganic Concentration on the pk a of the Analyte In general: Basic Compounds: pk a will decrease with the addition of an organic solvent Acidic Compounds: pk a will increase with the addition of an organic solvent The specific change in pk a will be compound dependent 2016 Waters Corporation 132

133 Amitriptyline + NH CH 3 N CH 3 CH 3 CH 3 Ionized Un-ionized pka (H 2 0) 9.4 apparentpk a (20% ACN) 8.5 apparentpk a (30% ACN) 8.3 apparentpk a (40% ACN) 8.0 Note: you cannot determine the degree or the direction of the pk a shift with out experimentally running each organic concentration 2016 Waters Corporation 133

134 Naproxen CH 3 CH CH 3 C - H 3 C H 3 C Un-ionized Ionized pka (H 2 0) 4.2 apparent pk a (20% ACN) 4.6 apparent pk a (30% ACN) 4.8 Note: you cannot determine the degree or the direction of the pk a shift with out experimentally running each organic concentration 2016 Waters Corporation 134

135 What is the practical application of this information? For new methods development: aqueous buffers at ph 2 (for acidic analytes) and ph 10 (for basic analytes) when mixed with organic solvents may provide stable chromatographic regions where most analytes will be unionized Caution: Must consider Un-ionized analyte hydrophobicity rganic solvent strength (Methanol versus THF) 2016 Waters Corporation 135

136 ph Limitations of lica Based Packing Materials Hydrolysis of Bonded Ligand Dissolution of silica particle ph 2016 Waters Corporation 136

137 Hydrolysis of a Bonded Phase Material: Monofunctional Ligand H + H 3 C Cl C CH 3 C C C C C C CH 3 H 3 C C CH 3 C C C C C C CH 3 Low ph Mobile Phase (hydrolysis of ligand) + HCl H + H 3 C H C CH 3 C C C C C C CH 3 lane Bleed 2016 Waters Corporation 137

138 Making a Bonded Phase Material: Multifunctional Synthesis H H H + Cl Cl C8 Trichlorosilane Ligand CH 3 Cl H H CH 3 Multi-Point Attachment - usually not 3 + HCl 2016 Waters Corporation 138

139 Making a Bonded-Phase Material: Trifunctional Synthesis Notice: need to break 3 bonds before H H H lane bleeds CH 3 CH 3 CH 3 1 and 2-point attachments to silica gel surface More hydrolytically stable Risks: Notice creation of additional silanols from the trifunctional synthesis Attachment of silanes may not be attached to silica gel surface Potential for poor batch-to-batch reproducibility 2016 Waters Corporation 139

140 Peak Shape: Monofunctional lane versus Trifunctional lane Risk: trifunctional silane can increase silanols lica gel particle Monofunctional C18 amitriptyline TF = lica gel particle Trifunctional C18 amitriptyline TF = Minutes 2016 Waters Corporation 140

141 Characteristics of Multifunctional Synthesis Multiple bonds to silica gel surface More resistant to hydrolysis at ph s <2 More difficult for reproducible synthesis Higher silanol activity Capable of higher ligand density 2016 Waters Corporation 141

142 Changes in Ligand Density Due to Bonding Technique Changes in ligand density due to ligand length and multifunctional silanes Ligand Monofunctional Synthesis (Ligand Density) Trifunctional Synthesis (Ligand Density) C µm/m µm/m 2 C8 3.5 µm/m µm/m 2 C4 3.8 µm/m µm/m 2 Note: Values from different column manufacturers data sheets 2016 Waters Corporation 142

143 Trifunctional Synthesis: Higher Ligand Densities, Greater Hydrolytic Stability H CH 3 H H H Bonded but not actually CH 3 reacted to substrate CH 3 H H H CH 3 CH 3 CH 3 Risks: Notice creation of additional silanols from the trifunctional synthesis Attachment of silanes may not be attached to silica gel surface Potential for poor batch-to-batch reproducibility 2016 Waters Corporation 143

144 Solubility of lica in Water (ppm) Mobile Phase ph and Column Life-time Even when bonded with ligands it will still dissolve lica Solubility Curve lica ph 2-8 Polymer ph 2-12 At ph >8 silica dissolves ph Elevated Temperature causes more rapid failure ph > 8 causes VIDING for traditional silica/ bonded particles 2016 Waters Corporation 144

145 Effects of High ph Mobile Phases Surface Modified lica Particles Hybrid Particles Complete dissolution of lica Catastrophic column failure Short lifetimes Slow rate of surface dissolution Incorporated methyl groups uncovered slows rate of dissolution Longer column lifetimes 2016 Waters Corporation 145

146 Mechanical -- Column Packing Material Flush Well packed column 2016 Waters Corporation 146

147 Good Plate Count Results (Instrument K, Column K, Connections K) Well packed column 2016 Waters Corporation 147

148 Triethylamine.temperature.pH 2016 Waters Corporation 148

149 Column Bed Collapse Mechanical Poor Plate Count Instrument K, Connections K Voided column - Channeling Packing Material Settled or dissolved 2016 Waters Corporation 149

150 Column Collapse (voiding) (due to shock / high ph {dissolution of particle}) All Peaks Distorted Voids - high back pressure, distorted and/or double peaks 2016 Waters Corporation 150

151 Extended ph Range of XBridge/BEH Columns mplifies Methods Development (ne Column Instead of Three) Low ph silica gel column High ph polymeric column Intermediate ph silica gel column Broad ph range Xbridge/BEH columns ph 2016 Waters Corporation 151

152 Different Ligands: Different Selectivity Changes in hydrophobicity Longer alkyl chain will provide greater retention Changes in silanol activity Affect peak asymmetry and influences secondary interactions Changes in hydrolytic stability Longer column lifetimes with greater number of ligand attachment points to the particle surface Changes in ligand density Influences sample loadability 2016 Waters Corporation 152

153 Initial Phases Chosen for Selectivity 2016 Waters Corporation 153

154 Two Fully-Scalable LC Column Platforms Family designed and optimized for ph stability Family designed and optimized for selectivity 1.7 [UPLC], 2.5 XP, 3.5, 5 and 10 µm 1.7 [UPLC], 2.5 XP, 3.5, 5 and 10 µm CSH 1.8 [UPLC], 2.5, 3.5 and 5 µm HSS 2016 Waters Corporation 154

155 2016 Waters Corporation 155

156 XBridge/BEH Family Family designed and optimized for ph stability Excellent general-purpose HPLC and UPLC columns for small molecule separations Broadest range of compound classes Widest ph range (1 12) Widest temperature range Ultra-low MS bleed Direct scalability between UPLC Technology, analytical HPLC and preparative HPLC 1.7 [UPLC], 2.5 XP, 3.5, 5 and 10 µm 2016 Waters Corporation 156

157 Ethylene Bridged Hybrid [BEH] Chemistries Rugged, reproducible, fully-scalable column chemistries for reversed-phase and HILIC separations BEH C18: First column choice, widest ph range, LC/MS BEH C8: For hydrophobic compounds, widest ph range, LC/MS BEH Shield RP18: Embedded carbamate group, alternate selectivity BEH Phenyl: Most stable phenyl column, wide ph range BEH HILIC: Unbonded, rugged BEH particle for HILIC separation of very polar bases BEH Amide: General-purpose HILIC column for very polar compounds such as sugars, saccharides, carbohydrates, etc Waters Corporation 157

158 Industry Leading ph Stability Column lifetimes in acidic mobile phases Shorter bar equals longer lifetime under low ph conditions Test conditions: 1% TFA in water (ph 1.0) at 80 o C. Comparative separations may not be representative of all applications Waters Corporation 158

159 Industry Leading ph stability Column lifetimes in alkaline mobile phases High ph mobile phases (>ph 10) rapidly dissolve silica-based stationary phases nly modern hybrid-based sorbents, [BEH] and [CSH], extend the usable mobile phase ph range from 1-12 Comparative separations may not be representative of all applications Waters Corporation 159

160 2016 Waters Corporation 160

161 2016 Waters Corporation 161

162 XSelect HSS Family Highest pressure tolerance; first silica-based particle (1.8 μm) designed for UPLC applications Alternate selectivities (vs. BEH & CSH particle columns) Increased retention 1.8 [UPLC], 2.5 XP, 3.5 and 5 µm HSS 2016 Waters Corporation 162

163 HSS Chemistries For general-purpose C 18 columns for low to neutral ph reversedphase separations HSS C18: Superior peak shape & acid stability HSS C18 SB: Alternate Selectivity for Basic (SB) compounds HSS T3: Enhanced reversed-phase retention of polar compounds HSS CN: Stable, normal phase-compatible, reproducible CN chemistry HSS PFP: Exceptional selectivity and retention for positional isomers, halogenated compounds and polar basic compounds 2016 Waters Corporation 163

164 2016 Waters Corporation 164

165 Charged Surface Hybrid [CSH] Particles Step 1 Step 2 Step 3 CHARGED SURFACE HYBRID PRCESS 2016 Waters Corporation 165

166 Column Selectivity Choices Caffeic Acid Derivatives ,4 5 5 BEH C 18 BEH C 8 BEH Shield RP18 BEH Phenyl 5 CSH Phenyl- Hexyl CSH Fluoro-Phenyl 5 CSH C 18 Conditions : Columns: 2.1 x 50 mm Mobile Phase A: 0.1% CF 3 CH in H 2 Mobile Phase B: 0.08% CF 3 CH in ACN Flow Rate: 0.5 ml/min Gradient: Time Profile Curve (min) %A %B Injection Volume: 1.0 µl Sample Diluent: 50:50 H 2 : MeH with 0.05% CF 3 CH Sample Conc.: 100 µg/ml Temperature: 40 o C Detection: 330 nm Sampling rate: 40 pts/sec Time Constant: 0.1 Instrument: Waters ACQUITY UPLC, with ACQUITY UPLC TUV HSS CN HSS PFP HSS T3 HSS C 18 HSS C 18 SB Compounds 1. Caftaric acid 2. Chlorogenic acid 3. Cynarin 4. Echinacoside 5. Cichoric acid Minutes 2016 Waters Corporation Comparative separations may not be representative of all applications 166

167 Selectivity Comparison ACQUITY UPLC BEH C pyrenesulfonic acid 2. Flavone 3. Imipramine 4. Fenoprofen 5. Amitriptyline 6. Diclofenac 7. ctanophenone ACQUITY UPLC CSH C ACQUITY UPLC CSH Phenyl-Hexyl ACQUITY UPLC CSH Fluoro-Phenyl All columns 2.1 x 50 mm, 1.7 µm Minutes 2016 Waters Corporation 167

168 NT All Peaks are Created Equally! Most common peak shapes Gaussian peak shape due to single interaction with the surface Linear isotherm Exponential tail due to secondary interactions Acidic silanols Shark fin peak shape due to mass overload Convex isotherm 2016 Waters Corporation 168

169 Conventional and CSH Technology High Purity C 18 columns Prednisone Caffeine C 18 Columns 1.7 µm BEH 1.7 µm CSH milar peak shape for neutrals Prednisone Caffeine Amitriptyline Metoprolol Expected P c With CSH Technology & Small Particles! Amitriptyline Metoprolol Peak Capacity Dramatically Improved Peak Shape for Bases! 2016 Waters Corporation 169

170 AU AU AU Benefits of CSH Technology: Loading Comparison ACQUITY UPLC CSH C x 50 mm, 1.7 µm Quetiapine (base) Propiophenone (neutral) Fully porous silica C x 50 mm, 1.8 µm Quetiapine (base) Propiophenone (neutral) Competitive Core-Shell C x 50 mm, 1.7 µm Quetiapine (base) Propiophenone (neutral) Minutes 2016 Waters Corporation Comparative separations may not be representative of all applications 170

171 AU Imipramine AU Imipramine CSH Technology: Influence of Sample Loading on Trace Impurity Detection ACQUITY CSH C µm 0.1 % formic acid 2.0 % 1.0 % 0.5 % 0.1 % Amitriptyline Minutes Superficially Porous C µm 0.1 % formic acid bservations CSH Technology enables superior peak shape and efficiency in low ionic strength mobile phases Improved sensitivity for trace level impurity analysis % 1.0 % 0.5 % 0.1 % Minutes Imipramine concentration held constant at 0.5 mg/ml; 0.1% formic acid mobile phase 2016 Waters Corporation 171

172 AU Imipramine AU Imipramine AU Imipramine CSH Technology: Influence of Sample Loading on Trace Impurity Detection A 0.1% impurity ACQUITY CSH C % formic acid B 0.1% impurity Superficially porous C % formic acid C 0.1% impurity Superficially porous C % TFA Minutes 2016 Waters Corporation 172

173 Intensity Imipramine Intensity Imipramine CSH Technology: Influence of Sample Loading on Trace Impurity Detection 8x10 8 Total Ion Chromatogram 0.1% impurity (m/z 278.3) ACQUITY UPLC CSH C % formic acid 4x Minutes 8x10 8 Total Ion Chromatogram UV Trace Superficially porous C % TFA 4x % impurity Minutes 2016 Waters Corporation 173

174 Implementing Mobile Phase ph Switching: Monitoring Column Performance Performance in 0.1% formic acid/acetonitrile gradient MAINTAINS Retention Time and Peak Shape After High ph Exposure. AU XSelect CSH C µm Before high ph exposure After high ph exposure AU Metoprolol min 20% RT change 64% peak height loss Amitriptyline 25% RT change 81% peak height loss neutral phthalates Hybrid Coated C 18 3 µm Before high ph exposure After high ph exposure Metoprolol Amitriptyline neutral phthalates min 2016 Waters Corporation 174

175 Why Haven t I Seen This Before? Remember: when using low ph buffers as part of your method development protocols (e.g., ammonium formate, sodium phosphate, etc.) these ph switching effects do not occur ph switching effects only occur when using acidic, low ionic strength additives ALL high ph-capable columns EXCEPT ACQUITY CSH and XSelect Columns exhibit this behavior 2016 Waters Corporation 175

176 XSelect CSH Family Family designed and optimized for selectivity Designed for LC/MS Improved peak shape for basic analytes under low-ionic strength acidic mobile phases Reduces overload effect Rapid equilibration; prevents retention time drift with changes in ph Multiple particle substrates to solve multiple chromatographic problems 1.7 [UPLC], 2.5 XP, 3.5, 5 and 10 µm CSH 2016 Waters Corporation 176

177 What Column(s) Should I Choose? CSH Columns Use additives AND buffers Impurity Profile work Isolation/purification Prefer to work at low ph with occasional high ph work Switch back/forth between low & high ph (additives) LC/MS laboratory Seeking additional ACQUITY UPLC column selectivities Designed for Selectivity 2016 Waters Corporation 177

178 CSH Chemistries For general-purpose C 18 columns for low to neutral ph reversedphase separations CSH C18: Rapid ph switching for method development CSH Phenyl-Hexyl: High performance phenyl column, best in additive-based mobile phases, different selectivity vs BEH Phenyl CSH Fluoro-Phenyl: Truly different selectivity, optimized for low ph separations, enhanced retention of acidic compounds 2016 Waters Corporation 178

179 Widest Selection of Unique Particle ffering BEH Technology HSS Technology CSH Technology Unparalleled ph stability Mobile phase and temperature versatility Seamless scalability UPLC to HPLC Maximum retention Particle and ligand selectivity Seamless scalability UPLC to HPLC Exceptional loading capacity Superior peak shape for basic analytes Seamless scalability UPLC to HPLC Waters is the NLY company that has fully porous and solid-core sub-2-µm particle columns 2016 Waters Corporation 179

180 extended Performance Columns 2.5 mm 2016 Waters Corporation 180

181 Addressing the Challenges: 2.5 µm extended Performance Columns Addressing the challenges Utilize XP 2.5 µm columns to improve productivity or achieve higher resolution Compatible with bandspread volumes of UHPLC systems Implement XP 2.5 µm columns to decrease backpressure or increase flow rate 2016 Waters Corporation 181

182 Flexible ptions to Improve LC Productivity 2016 Waters Corporation 182

183 Allowable Adjustments in Chromatography <621> Variable Allowable Changes Particle ze -50% Column Length ±70% Flow Rate ±50% Column ID Any allowed Injection Volume Any reduction Column Temperature ±10% Mobile Phase ph ±0.2 unit System Suitability. USP 37. Pag Waters Corporation 183

184 USP Assay: Levonorgestrel and Ethinyl Estradiol Isocratic HPLC Method System Suitability Requirements Column: Injection Volume: Flow Rate: Mobile Phase: Detection Wavelength: XBridge C x 150 mm 5 μm (L7) 50 μl 1.00 ml/min 35:15:45 ACN:MeH:Water UV 230 nm USP Resolution: NLT 2.5 Peak Area Precision: NMT 2.0% RSD Questions Can we transfer this simple method? What combination of column and system? What compromises (if any) must be made? What conditions provide the greatest benefits? 2016 Waters Corporation 184

185 AU ethinyl estradiol levonorgestrel riginal Isocratic USP Assay: 4.6 x 150 mm, 5 μm (Alliance HPLC) Flow Rate: 1.00 ml/min Pressure: 1400 psi Run time: 8.00 min System Suitability Requirements USP Resolution: NLT 2.5 Peak Area Precision: NMT 2.0% RSD USP Res: Minutes L/dp = N = 1X Rs = 1X Run time = 1X Pressure = 1X N Remember: Rs 2016 Waters Corporation 185 L dp N F opt 1 Run Time L F 1 dp

186 AU ethinyl estradiol levonorgestrel AU ethinyl estradiol levonorgestrel Isocratic Method Transferred to 4.6 mm XP Columns (Alliance HPLC) x 75 mm XP 2.5 μm Flow Rate: 1.00 ml/min Pressure: 2600 psi Run time: 4.00 min USP Res: 6.5 Could not run at SCALED flow rate due to system pressure limitations Results: Rs = 0.9X Run time = 0.5X Pressure = 2X Minutes 4.6 x 50 mm XP 2.5 μm Flow Rate: 2.00 ml/min Pressure: 3600 psi Run time: 1.30 min USP Res: > 50% flow rate change not <621> compliant LWER L/dp column (20000) to reduce pressure: Results: Rs = 0.7X Run time = 0.16X Pressure = 2.9X Waters Corporation Minutes 186

187 ethinyl estradiol AU levonorgestrel AU ethinyl estradiol levonorgestrel Isocratic Method Transferred to ACQUITY UPLC H-Class System x 75 mm XP, 2.5 μm Flow Rate: 0.42 ml/min Pressure: 4400 psi Run Time: 2.00 min USP Res: 7.9 Results: Rs = 1.1X Run time = 0.25X Pressure = 3.1X Minutes 2.1 x 50 mm UPLC, 1.7 μm Flow Rate: 0.61 ml/min Pressure: 7700 psi Run Time: 0.90 min. Maximum benefits obtained on ACQUITY UPLC System USP Res: 7.0 Results: Rs = 1X Run time = 0.1X Pressure = 5.5X Minutes 2016 Waters Corporation 187

188 Column Parameters Column length More length increases separation power Increases runtime Column ID Smaller ID increases sensitivity Reduces injection volume Particle size Smaller particle size equals more efficiency Increases backpressure L/dp ratio A measure of separation power Equal values of L/dp are equivalent efficiency 2016 Waters Corporation 188

189 Summary Waters is committed to manufacturing high-quality consumables that ensure reproducible results year after year XBridge BEH Columns designed for ph stability and method flexibility XSelect HSS and CSH Columns provide maximum retention, selectivity and excellent peak shape for basic analytes 2.5 µm XP columns bridge the gap between HPLC and UPLC Compatible with HPLC, UHPLC and UPLC system platforms Maximize HPLC productivity 2016 Waters Corporation 189

190 Column ptions 2016 Waters Corporation 190

191 System Comparisons Breeze 2 HPLC Alliance HPLC ACQUITY Arc HPLC/UHPLC ACQUITY H-Class UPLC ACQUITY I-Class UPLC ACQUITY M-Class UPLC 1.6 µm for ultimate efficiency at UPLC backpressure 1.7/1.8 µm for high efficiency and selectivity 2.7 µm for maximum HPLC performance XP 2.5 µm for HPLC selectivity 5 µm for low backpressure and legacy methods 1.8 µm for high efficiency, retention and selectivity 2016 Waters Corporation 191

192 Resources that Enable Success: Technology Primers Waters Corporation 192

193 Questions 2016 Waters Corporation 193

194 Escitalopram (ral Solution) Escitalopram pk a =9.78 Mobile Phase: Buffer 6.1g of monobasic potassium phosphate. To each L of this solution add 1.5 ml of triethylamine. Adjust with phosphoric acid to ph of 2.5. Mobile phase ACN/Buffer (32:68). Column: 4.6 x 250 mm, 5 um, column packed with L Waters Corporation 194

195 RP-Ion exchange Hydrophobic Interaction with Bonded Phase Mobile Phase ph < 3 - H - - H - - H - H H HN (CH 3 ) 2 Ion exchange Interaction with Charged tes High lanol Activity (CH) 2 HN 3 Mobile - Phase ph > 5 N Substrate Protonated - no charge Substrate De-protonated -- Negative Charge Base RP Cation X Base Increased retention and PR Peak Shape SAME CLUMN 2016 Waters Corporation 195

196 Tailing Factor Peak Shape ver Wide ph Range (Strong Base -- Amitriptyline pka 9.4) 4 Low ligand density and high metal content silica gel: Conventional C High ligand density and high purity silica gel: Modern C Buffer ph Ideal Behavior Pure Polymer No lanols 2016 Waters Corporation 196

197 log (k') Impact on Retention Factor (k) on Robustness By Changing Mobile Phase ph Range - HPLC In Pure REVERSED-PHASE Mode 100 ph Limit of lica gel 10 Acid (Un-Ionized) Base (Un-Ionized) Neutral 1 Base (Ionized) Acid (Ionized) ph Greatly improved reproducibility due to flattening of curves Robustness Note: Column Particle, Temperature and % rganic Held Constant 2016 Waters Corporation 197

198 Column ptions 2016 Waters Corporation 198

199 Dibucaine Reported pk a = 8.9 Eluent: 1.2 g sodium lauryl sulfate (sodium dodecyl sulfate), 0.2g sodium acetate, 2.0 ml triethylamine in 300 ml water, adjust ph to 5.6 and add 700 ml methanol Column: 3.9 x 300 mm L1. Tailing Factor NMT Waters Corporation 199

200 RP-Ion exchange Hydrophobic Interaction with Bonded Phase Mobile Phase ph < 3 - H - - H - - H - H H HN (CH 3 ) 2 Ion exchange Interaction with Charged tes High lanol Activity (CH) 2 HN 3 Mobile - Phase ph > 5 N Substrate Protonated - no charge Substrate De-protonated -- Negative Charge Base RP Cation X Base Increased retention and PR Peak Shape SAME CLUMN 2016 Waters Corporation 200

201 log (k') Impact n Retention Factor (k) By Changing Mobile Phase ph -- HPLC In REVERSED-PHASE Mode 100 Acid (Un-Ionized) 10 Note: Column Particle,Temperature and % rganic Held Constant ph Limit of lica gel Base (Un-Ionized) Neutral Base (Ionized) ph Acid (Ionized) Range of poor reproducibility due to steep slopes - small ph change -- large retention change 2016 Waters Corporation 201

202 Column ptions 2016 Waters Corporation 202

203 Quinidine pka 8.56 (at 25 C) Solution A: Add 35 ml of methanesulfonic acid to 20 ml of glacial acetic acid, and dilute with water to 500 ml Solution B: Dissolve 10 ml of Diethylamine in water to obtain 100 ml of solution Mobile Phase Acetonitrile, Solution A, Solution B, and Water (10:1:1:40) Column 4.6 x 250 mm, 5 um, L Waters Corporation 203

204 Column ptions 2016 Waters Corporation 204

205 Tioconazole (Imidazole Antifungal Basic Compound) and Related Compounds A, B, and C Predicted pka = 6.19 Structures for Related Compounds A-C Unknown Mobile Phase: (Note- Prepare the Mobile phase fresh daily.) Mix 400 ml of acetonitrile, 400 ml of methanol, and 280 ml of water. Add 2 ml of ammonium hydroxide and mix. Column: A 4 mm x 10-cm pre-column that contains packing L4 (silica gel), installed between the pump and the injector (replaced daily) and a 5-mm x 25-cm analytical column that contains packing L Waters Corporation 205

206 log (k') Impact on Retention Factor (k) on Robustness By Changing Mobile Phase ph Range - HPLC In Pure REVERSED-PHASE Mode 100 ph Limit of lica gel 10 Acid (Un-Ionized) Base (Un-Ionized) Neutral 1 Base (Ionized) Acid (Ionized) ph Greatly improved reproducibility due to flattening of curves Robustness Note: Column Particle, Temperature and % rganic Held Constant 2016 Waters Corporation 206

207 Column ptions 2016 Waters Corporation 207

208 Succinylcholine Nicotinic acetilcholine receptor agonist Mobile Phase: Prepare a solution in water containing 3.85 g per L of sodium 1-pentanesulfonate anhydrous, 2.9 g per L of sodium chloride, and 1% (v/v) of 1 N sulfuric acid. Prepare a filtered and degassed mixture of Buffer solution and acetonitrile (95:5) Column: A 4.6-mm x 25-cm, 5 um, analytical column that contains packing L Waters Corporation 208

209 Surface of a lica Gel Bonded-Phase Packing Material H 3 C H H 2 C H 3 C H H 3 C H 2 C H 2 C H 2 C H 3 C H 2 C H 2 C CH 2 Polar analytes are not able to energetically fit between ligands can t interact with surface or ligands CH 2 CH 2 CH 2 CH 2 CH 2 CH 2 CH 2 CH 2 CH 2 CH 2 H 2 C H 2 C H 2 C CH 2 CH 2 CH 2 Residual silanol Endcap H 2 C H 2 C H 2 C CH 2 CH 2 CH 2 CH 3 CH H 3 3 C H 3 C CH 3 CH H 3 C CH 3 H 3 C 3 CH 3 CH H 3 3 C H H H H H H C 8 alkyl chains CH 3 H 3 C H CH 3 H 3 C H 2 C CH 2 Polar analytes easily interact with surface and ligands H 3 C H 2 C CH 2 Residual silanols H 3 C H 2 C H 2 C H 2 C H 3 C H 2 C CH 2 CH 2 CH 2 CH 2 CH 3 CH 2 H 3 C H 3 C CH 3 CH 3 H CH 3 CH H 3 C 3 H H H H H High Coverage High Ligand Density Low Coverage Low Ligand Density Endcap CH 3 CH Waters Corporation 209

210 Surface of a lica Gel Bonded-Phase Packing Material H 2 C Non-polar portion of analyte interacts with bonded phase H 2 C (hydrophobic) H 3 C H H 2 C H 2 C CH 2 CH 2 CH 2 CH 2 Alkyl chains H CH 3 C 3 H N H NH CH 3 Residual silanols H 3 C CH 3 H Polar portion of analyte interacts with silanols (hydrogen bonding and/or ion exchange) CH 3 CH 3 H Low Coverage Low Ligand Density ~ Waters Corporation 210 N H NH 2 N H H 3 C H H 2 C H 2 C H 2 C H 2 C CH 2 CH 2 CH 2 CH 2 CH 3 H H 3 C Endcap H CH 3 CH 3

211 Column ptions 2016 Waters Corporation 211

212

Transferring Yesterdays Methods to Tomorrows Technology: The Evolution from HPLC to UPLC TM

Transferring Yesterdays Methods to Tomorrows Technology: The Evolution from HPLC to UPLC TM Transferring Yesterdays Methods to Tomorrows Technology: The Evolution from HPLC to UPLC TM Eric S. Grumbach Thomas E. Wheat Chuck Phoebe Joe Arsenault Jeffrey R. Mazzeo Diane M. Diehl UPLC TECHNOLOGY

More information

Successfully Scaling and Transferring HPLC and UPLC Methods

Successfully Scaling and Transferring HPLC and UPLC Methods Successfully Scaling and Transferring HPLC and UPLC Methods Esa Lehtorinne Esa_Lehtorinne@waters.com Tel: +358-9-5659 6288 Fax: +358-9-5659 6282 Waters Finland Kutomotie 16 00380 Helsinki 2013 Waters Corporation

More information

UPLC Method Development and Validation

UPLC Method Development and Validation UPLC Method Development and Validation Rev. 2 2008 Waters Corporation Challenges of Method Development Methods are developed throughout the drug development process Samples vary in complexity Redundancy

More information

LC and LC/MS Column Selection Flow Chart

LC and LC/MS Column Selection Flow Chart LC and LC/MS Column Selection Flow Chart To use the column selection diagram below, simply follow the path for your analyte and mobile phase. At the far right, follow your final column selection to the

More information

Product Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns

Product Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns Product Bulletin HPLC Columns ACE LC/MS and Rapid Analysis HPLC Columns 0 mm, 30 mm, 35 mm and 50 mm column lengths.0,., 3.0, 4.0 and 4.6 mm column diameters Configured for High Sample Throughput Specially

More information

Validation of ACQUITY UPLC Methods

Validation of ACQUITY UPLC Methods Tanya Jenkins Andrew Aubin Dr. Michael Swartz Waters Corporation 34 Maple Street Milford, MA, 01757,USA Validation of ACQUITY UPLC Methods Agenda Test Method Validation Why ACQUITY UPLC? Redevelop or Convert

More information

Comparison Guide to C18 Reversed Phase HPLC Columns

Comparison Guide to C18 Reversed Phase HPLC Columns Comparison Guide to C18 Reversed Phase HPLC Columns Comparison Data on Commonly Used C18 Phases Stationary Phase Specifications Phases Compared According to Relative Hydrophobicity Phases Compared According

More information

ProntoSIL C18-EPS Reversed-Phase HPLC Columns

ProntoSIL C18-EPS Reversed-Phase HPLC Columns ProntoSIL C8-EPS Reversed-Phase HPLC Columns Provides excellent separation of polar compounds Better peak shape for acids and bases Stabilized bonded phase for rugged, robust HPLC methods More retentive

More information

State-of-the-art C18 HPLC Columns

State-of-the-art C18 HPLC Columns An HPLC GL Sciences Newest and Most Advanced ODS Phase-New For 00 State-of-the-art C HPLC s Improved Peak Shapes and Heights Enhancing Sensitivity High Resolution Fast Equilibration Compatible with 00%

More information

New 5-micron HALO-5 columns based on Fused- Core particle technology boost the performance of HPLC. Compared to other HPLC columns, HALO-5 columns

New 5-micron HALO-5 columns based on Fused- Core particle technology boost the performance of HPLC. Compared to other HPLC columns, HALO-5 columns New 5-micron HALO-5 columns based on Fused- Core particle technology boost the performance of HPLC. Compared to other HPLC columns, HALO-5 columns have: the highest plate number versus any other 5-micron

More information

Too Polar for Reversed Phase What Do You Do?

Too Polar for Reversed Phase What Do You Do? Too Polar for Reversed Phase What Do You Do? June 20, 2013 Mark Powell Columns and Consumables Technical Support Page 1 C8 or C18 Doesn t Always Do the Job Typical reversed phase conditions involve water/buffer

More information

penta-hilic UHPLC COLUMNS

penta-hilic UHPLC COLUMNS penta-hilic UHPLC COLUMNS penta-hilic Highly retentive, proprietary penta-hydroxy-ligand Excellent peak shape for polar compounds with a variety of functional groups: acids, bases, zwitterions strong and

More information

Reversed Phase Solvents

Reversed Phase Solvents Part 1. General Chromatographic Theory Part 2. verview of HPLC Media Part 3. The Role of the Mobile Phase in Selectivity Part 4. Column Care and Use Reversed Phase Solvents 2 Solvents for RP Chromatography

More information

Hypersil BDS Columns TG 01-05

Hypersil BDS Columns TG 01-05 TG 0-0 Hypersil BDS Columns Introduction Hypersil BDS columns have gained a reputation over the years as one of the most robust, reproducible and reliable HPLC column brands available. This Technical Guide

More information

COMPARISON GUIDE. to C18 Reversed Phase HPLC Columns. Comparison Data on 60 Commonly Used C18 Phases. Stationary Phase Specifications

COMPARISON GUIDE. to C18 Reversed Phase HPLC Columns. Comparison Data on 60 Commonly Used C18 Phases. Stationary Phase Specifications COMPARISON GUIDE to C18 Reversed Phase HPLC Columns Comparison Data on 60 Commonly Used C18 Phases Stationary Phase Specifications Phases Compared According to Relative Hydrophobicity Phases Compared According

More information

Packings for HPLC. Packings for HPLC

Packings for HPLC. Packings for HPLC Summary of packings for HPLC In analytical HPLC, packings with particle sizes of 3 to 10 µm are preferred. For preparative separation tasks, also particles with diameters larger than 10 µm are applied.

More information

LC Technical Information

LC Technical Information LC Technical Information Method Transfer to Accucore.6 μm Columns Containing solid core particles, which are engineered to a diameter of.6μm and a very narrow particle size distribution; Accucore HPLC

More information

Choosing Columns and Conditions for the Best Peak Shape

Choosing Columns and Conditions for the Best Peak Shape Choosing Columns and Conditions for the Best Peak Shape What is Good Peak Shape and Why is it Important? Good peak shape can be defined as a symmetrical or gaussian peak and poor peak shape can include

More information

penta-hilic UHPLC COLUMNS

penta-hilic UHPLC COLUMNS penta-hilic UHPLC COLUMNS Highly retentive, proprietary penta-hydroxy-ligand penta-hilic Excellent peak shape for polar compounds with a variety of functional groups: acids, bases, zwitterions strong and

More information

Inertsil Technical Library

Inertsil Technical Library HPLC COLUMN INERTSIL TECHNICAL LIBRARY H P L C C o l u m n Inertsil Technical Library Inertsil ODS-4, C8-4 Comparison of Performance List of Compared Columns Experimental Explanation, Analytical Conditions

More information

Part 2. Overview of HPLC Media

Part 2. Overview of HPLC Media Part 1. General Chromatographic Theory Part 2. Overview of HPLC Media Part 3. The Role of the Mobile Phase in Selectivity Part 4. Column Care and Use 1 HPLC Particle Technology Core-Shell Particle Fully

More information

Chemistry Instrumental Analysis Lecture 28. Chem 4631

Chemistry Instrumental Analysis Lecture 28. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 28 High Performance Liquid Chromatography () Instrumentation Normal Phase Chromatography Normal Phase - a polar stationary phase with a less polar mobile phase.

More information

Method Development Kits

Method Development Kits Method Development Kits sales representative for ordering information or contact...77 Method Development Kits First column choice for Pharma Development USP L C Method Development Kits Maximize retention

More information

William E. Barber, Ph.D. Applications Chemist October

William E. Barber, Ph.D. Applications Chemist October William E. Barber, Ph.D. Applications Chemist ctober 0 00 Secrets of Good Peak Shape in HPLC Time: :00 p.m. CET Telephone Number: + 44 0 76 088 Chairperson: John Vis The Secrets of Good Peak Shape in HPLC

More information

Kromasil. for your analytical HPLC. Kromasil. The way to peak performance in liquid chromatography

Kromasil. for your analytical HPLC. Kromasil. The way to peak performance in liquid chromatography Kromasil for your analytical HPLC Kromasil The way to peak performance in liquid chromatography Kromasil is known, worldwide, for its high performance and excellent total economy in preparative and industrial

More information

ACE SuperC18. ACE UHPLC and HPLC Columns. Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability

ACE SuperC18. ACE UHPLC and HPLC Columns. Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability ACE SuperC8 TM Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability Exploit selectivity at low, intermediate and high ph Use with LC/MS compatible buffers between ph.. High efficiency µm, µm,

More information

InertSustainSwift C8

InertSustainSwift C8 Physical Properties Silica Particle Size Surface Area Pore Size Pore Volume Bonded Phase End-capping Carbon Loading ph Range USP Code :ES (Evolved Surface) Silica Gel :.9 μm, μm, μm :00 m /g :00 Å (0 nm)

More information

P R O D U C T B U L L E T I N. Fused-Core particle technology for hyper-fast and super-rugged HPLC columns

P R O D U C T B U L L E T I N. Fused-Core particle technology for hyper-fast and super-rugged HPLC columns P R O D U C T B U L L E T I N Fused-Core particle technology for hyper-fast and super-rugged HPLC columns HALO column packings are not made the typical way. Instead, the particles packed into HALO columns

More information

Basic Principles for Purification Using Supercritical Fluid Chromatography

Basic Principles for Purification Using Supercritical Fluid Chromatography Basic Principles for Purification Using Supercritical Fluid Chromatography Jo-Ann M. Jablonski, Christopher J. Hudalla, Kenneth J. Fountain, Steven M. Collier, and Damian Morrison Waters Corporation, Milford,

More information

ACE Excel SuperC18. ACE UHPLC and HPLC Columns. Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability

ACE Excel SuperC18. ACE UHPLC and HPLC Columns. Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability ACE Excel SuperC8 TM Ultra-Inert UHPLC and HPLC s with Extended ph Stability Exploit selectivity at low, intermediate and high ph Use with LC/MS-compatible buffers between ph.. High efficiency µm, µm,

More information

A C18 Silica Column With Exceptional Temperature and ph Stability

A C18 Silica Column With Exceptional Temperature and ph Stability A C18 lica Column With Exceptional Temperature and ph Stability Brian A. Jones 1, Stephanie J. Marin 1, Jody Clark 1, Nathan L. Porter 1, J. Andreas Lippert 2, and Todd M. Johnson 2 1. Selerity Technologies,

More information

An HPLC column offering moderate retentivity with superb peak shape Inertsil ODS Sprint

An HPLC column offering moderate retentivity with superb peak shape Inertsil ODS Sprint An HPLC column offering moderate retentivity with superb peak shape Inertsil ODS Sprint An important new column in the Inertsil series Bonded-Phase Structure Perfect balance of retention for both polar

More information

ACE SuperC18. ACE UHPLC and HPLC Columns. Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability

ACE SuperC18. ACE UHPLC and HPLC Columns. Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability ACE SuperC8 TM Ultra-Inert UHPLC and HPLC s with Extended ph Stability Exploit selectivity at low, intermediate and high ph Use with LC/MS compatible buffers between ph.. High efficiency µm, µm, µm and

More information

STUDY OF LOADABILITY AND SELECTIVITY OF PHARMACEUTICAL COMPOUNDS ON RPLC COLUMNS

STUDY OF LOADABILITY AND SELECTIVITY OF PHARMACEUTICAL COMPOUNDS ON RPLC COLUMNS STUDY OF LOADABILITY AND SELECTIVITY OF PHARMACEUTICAL COMPOUNDS ON RPLC COLUMNS Fang Xia, Jie Y. Cavanaugh, Uwe Neue, Jeffrey R. Mazzeo, Diane M. Diehl Waters Corporation, Milford, MA USA OVERVIEW- INTRODUCTION-

More information

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application Pharmaceuticals Authors John W. Henderson Jr., Nona Martone, and Cliff Woodward

More information

Sunniest C18 Sunniest RP-AQUA Sunniest C8 Patent pending

Sunniest C18 Sunniest RP-AQUA Sunniest C8 Patent pending Patent pending Website : www.chromtech.net.au E-mail : info@chromtech.net.au telno : 0 97 0... in AUSTRALIA A Novel Bonding Technique (patent pending) An unique trifunctional silyl-reagent was developed

More information

HICHROM. Chromatography Columns and Supplies. LC COLUMNS Zorbax. Catalogue 9. Hichrom Limited

HICHROM. Chromatography Columns and Supplies. LC COLUMNS Zorbax. Catalogue 9. Hichrom Limited HICHROM Chromatography Columns and Supplies LC COLUMNS Zorbax Catalogue 9 1 The Markham Centre, Station Road Theale, Reading, Berks, RG7 4PE, UK Tel: +44 (0)118 930 3660 Fax: +44 (0)118 932 3484 Email:

More information

There s Great Chemistry Between Us.

There s Great Chemistry Between Us. There s Great Chemistry Between Us. Introduction HPLC Columns Theory, Technology and Practice Uwe D. Neue, Waters Corporation, Milford, MA published by John Wiley & Sons High performance liquid chromatography

More information

columns Acclaim Mixed-Mode WCX-1 for Separating Basic Molecules

columns Acclaim Mixed-Mode WCX-1 for Separating Basic Molecules columns Acclaim Mixed-Mode WCX- for Separating Basic Molecules The Acclaim Mixed-Mode WCX- is a novel, high-efficiency, silica-based column specially designed for separating various basic analytes. This

More information

ANALYSIS OF BASIC COMPOUNDS

ANALYSIS OF BASIC COMPOUNDS COMPARISON OF Fully AND Superficially POROUS PARTICLE COLUMNS FOR THE ANALYSIS OF BASIC COMPOUNDS Kenneth J. Fountain, Jane Xu, Zhe Yin, Pamela C. Iraneta, Diane M. Diehl Waters Corporation GOAL Compare

More information

[ HPLC COLUMNS ] Continuing the Legacy of HPLC Column Performance

[ HPLC COLUMNS ] Continuing the Legacy of HPLC Column Performance [ HPLC CLUMNS ] Continuing the Legacy of HPLC Column Performance HPLC Columns Creating Exceptional Chromatography Waters reputation is based on chromatography, but we do not create chromatography you do.

More information

Inertsil ODS-EP Technical Information

Inertsil ODS-EP Technical Information Technical Information Advantages of The selectivity is completely different from those of conventional columns such as ODS column due to its specific polar group in the stationary phase. Durability can

More information

Overview Applications of NUCLEODUR HPLC Phases

Overview Applications of NUCLEODUR HPLC Phases verview Applications of HPLC Phases Phase Specification Characteristics* Stability Structure A B C octadecyl phase, high density coating multi-endcapping 8% C USP L ph stability -, suitable for LC/MS (Si-

More information

2010 Waters Corporation

2010 Waters Corporation UPLC Column Positioning Jurmala September 3-5 Anne Dyrdal CBU 2010 Waters Corporation Waters Column Product History Styragel 1958 µbondapak ACQUITY UPLC BEH SunFire Columns Spherisorb SymmetryShield XTerra

More information

Reversed Phase Chromatography: Mobile phase Considerations

Reversed Phase Chromatography: Mobile phase Considerations Reversed Phase: Mobile Phase Considerations Dr. www.forumsci.co.il/hplc Example of a Reversed Phase Method with UV Detection 1 Chromatographic Process B+A Mobile phase Elution through the Column-movie

More information

Changing the way you think about HPLC

Changing the way you think about HPLC 3521 Silverside Road Suite 1-K Quillen Building Wilmington, DE 19810 USA info@advanced-materials-tech.com Changing the way you think about HPLC Changing the way you think about HPLC New 5-micron HALO-5

More information

Thermo Scientific Syncronis HPLC Columns. Remarkable separations guaranteed time after time

Thermo Scientific Syncronis HPLC Columns. Remarkable separations guaranteed time after time Thermo Scientific Syncronis HPLC Columns Remarkable separations guaranteed time after time Syncronis HPLC and UHPLC Columns Remarkable separations guaranteed time after time When developing a new method,

More information

InertSustainSwift C8

InertSustainSwift C8 HPLC, LC/MS Columns InertSustainSwift TM C8 New! InertSustainSwift C8 is an octyl group (C8) bonded column offering the same extreme inertness to any type of compounds just like InertSustainSwift C8, which

More information

Comparison of different aqueous mobile phase HPLC techniques

Comparison of different aqueous mobile phase HPLC techniques June 009 ewsletter Pharmaceutical Analysis: What is Your Problem? SIELC Technologies, Inc., Prospect Heights, IL 0070 Pharmaceutical analysis involves liquid chromatography of various compounds, from active

More information

Uptisphere CS Evolution Core Shell columns for fast & highly efficient identification & quantification of small molecules

Uptisphere CS Evolution Core Shell columns for fast & highly efficient identification & quantification of small molecules Uptisphere CS Evolution Core Shell columns for fast & highly efficient identification & quantification of small molecules www.interchim.com www.blog.interchim.com www.forum.interchim.com www.facebook.com/interchim

More information

New Product Update. HPLC Columns

New Product Update. HPLC Columns New Product Update HPLC Columns ZORBAX MicroBore HPLC Columns ZORBAX Extend-C18 Columns ZORBAX Bonus-RP Columns Inertsil ODS-2 Cartridge Columns ZORBAX Carbohydrate Analysis Columns ZORBAX Eclipse dsdna

More information

Optimisil HPLC & UPLC Column catalogue

Optimisil HPLC & UPLC Column catalogue Optimisil HPLC & UPLC Column catalogue Analytical Hub Plot No: 37, P & T Colony, Vikrampuri, Secunderabad 500009 Tel: 040-64621313, Mobile: +91 7799931313 Website: www.analyticalhub.com email: contact@analyticalhub.com

More information

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX Technical Note Agilent s SampliQ WAX provides Applications for strongly acidic, acidic and neutral compounds Excellent

More information

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

Thermo Scientific Accucore XL HPLC Columns. Technical Manual Thermo Scientific Accucore XL HPLC Columns Technical Manual Thermo Scientific Accucore XL HPLC Columns Based on Core Enhanced Technology using µm solid core particles, Accucore XL HPLC columns allow users

More information

Alltech Alltima HP Introduction

Alltech Alltima HP Introduction Alltech Introduction High-Stability, High-Purity, High-Performance, Low-Bleed Columns for Demanding Applications Better Peak Symmetry high-purity silica eliminates peak tailing problems Long Column Life

More information

( )( ) Selectivity Choices in Reversed-Phase Fast LC. Introduction. R s = 1 a 1 k 4 a 1 + k

( )( ) Selectivity Choices in Reversed-Phase Fast LC. Introduction. R s = 1 a 1 k 4 a 1 + k Selectivity Choices in Reversed-Phase Fast LC Luisa Pereira, Monica Dolci, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20543 Key Words Accucore, Hypersil GOLD, Syncronis, Column Chemistry,

More information

High Performance Liquid Chromatography

High Performance Liquid Chromatography Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN

More information

High Performance Liquid Chromatography

High Performance Liquid Chromatography Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 1 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN

More information

Acclaim Mixed-Mode WCX-1

Acclaim Mixed-Mode WCX-1 Acclaim Mixed-Mode WCX-1 Product Manual for the Acclaim Mixed-Mode WCX-1 Column Page 1 of 28 PRODUCT MANUAL for the Acclaim Mixed-Mode WCX-1 Columns 4.6 x 150 mm, P/N (068353) 4.6 x 250 mm, P/N (068352)

More information

ACE Ultra Inert Base Deactivated HPLC Columns

ACE Ultra Inert Base Deactivated HPLC Columns ACE Ultra Inert Base Deactivated HPLC Columns Contents Page ACE HPLC Columns...- Independent Comparison of HPLC Columns #... Independent Comparison of HPLC Columns #... ACE 00Å HPLC Columns Specifications...-

More information

viridis Columns Bringing a New Dimension to SFC Combining state-of-the-art media manufacturing with industry leading column technology, Viridis

viridis Columns Bringing a New Dimension to SFC Combining state-of-the-art media manufacturing with industry leading column technology, Viridis [ viridis columns ] viridis Columns Bringing a New Dimension to SFC Combining state-of-the-art media manufacturing with industry leading column technology, Viridis Supercritical Fluid Chromatography (SFC)

More information

Introduction to Chromatography

Introduction to Chromatography Introduction to Chromatography Dr. Sana Mustafa Assistant Professor Department of Chemistry, Federal Urdu University of Arts, Science & Technology, Karachi. What is Chromatography? Derived from the Greek

More information

Ion Exchange Columns ---- Inertsil AX 078 Inertsil CX 080. HPLC Column Catalog

Ion Exchange Columns ---- Inertsil AX 078 Inertsil CX 080. HPLC Column Catalog Ion Exchange ---- Inertsil AX 078 Inertsil CX 080 HPLC Column Catalog 077 Inertsil AX Physical Properties Silica : Series High Purity Silica Gel : μm Surface Area : 0 m /g Pore Size : 00 Å (0 nm) Pore

More information

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency Luna 2.5 µm C18(2)-HST Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency Table of Contents Part 1 Theory 1.1 Abstract...3 1.2 Introduction...3 Part 2 Set Up 2.1

More information

Routine MS Detection for USP Chromatographic Methods

Routine MS Detection for USP Chromatographic Methods Daniel S. Root, Thomas E. Wheat, and Patricia McConville Waters Corporation, Milford, MA, USA APPLICATION BENEFITS This approach enables mass spectral analysis of peaks directly from unmodified USP HPLC

More information

HICHROM. Chromatography Columns and Supplies. LC COLUMNS HALO and HALO-5. Catalogue 9. Hichrom Limited

HICHROM. Chromatography Columns and Supplies. LC COLUMNS HALO and HALO-5. Catalogue 9. Hichrom Limited HICHROM Chromatography Columns and Supplies LC COLUMNS HALO and HALO- Catalogue 9 The Markham Centre, Station Road Theale, Reading, Berks, RG PE, UK Tel: + (0)8 90 0 Fax: + (0)8 9 8 Email: sales@hichrom.co.uk

More information

for Acclaim Trinity TM P1

for Acclaim Trinity TM P1 for Acclaim Trinity TM P Product Manual for Acclaim TRINITY P Columns Page of 4 Product Manual for Acclaim Trinity TM P Columns 07556 Acclaim Trinity P, 3µm, 4.6 x 00mm 07556 Acclaim Trinity P, 3µm, 4.6

More information

Shodex TM ODP2 HP series columns

Shodex TM ODP2 HP series columns HPLC Columns Shodex TM ODP2 HP series columns Better retention of highly polar substances Technical notebook No. 6 Contents 1. Introduction 1-1. Specifications 1-2. Eluent Compatibility of ODP2 HP Series

More information

HICHROM. Chromatography Columns and Supplies. LC COLUMNS SIELC Primesep. Catalogue 9. Hichrom Limited

HICHROM. Chromatography Columns and Supplies. LC COLUMNS SIELC Primesep. Catalogue 9. Hichrom Limited HICHROM Chromatography Columns and Supplies LC COLUMNS SIELC Primesep Catalogue 9 The Markham Centre, Station Road Theale, Reading, Berks, RG7 4PE, UK Tel: +44 (0)8 90 660 Fax: +44 (0)8 9 484 Email: sales@hichrom.co.uk

More information

Technical Guide Thermo Scientific Syncronis HPLC Columns

Technical Guide Thermo Scientific Syncronis HPLC Columns Technical Guide Thermo Scientific Syncronis HPLC Columns Consistent, predictable separations, Column after column, time after time Thermo Scientific Syncronis Columns When developing a new method, one

More information

Developing Analytical Chromatographic Methods for Pharmaceutical Stability Investigations

Developing Analytical Chromatographic Methods for Pharmaceutical Stability Investigations Developing Analytical Chromatographic Methods for Pharmaceutical Stability Investigations Warren B Potts III Senior Director Americas Marketing Warren_potts@waters.com 2017 Waters Corporation 1 Outline

More information

Liquid Chromatography

Liquid Chromatography Liquid Chromatography 1. Introduction and Column Packing Material 2. Retention Mechanisms in Liquid Chromatography 3. Method Development 4. Column Preparation 5. General Instrumental aspects 6. Detectors

More information

Small Molecule Selectivity Sampler

Small Molecule Selectivity Sampler Small Molecule Selectivity Sampler HALO 90 Å C18, AQ-C18, BIPHENYL Fused-Core Particle Technology Why Choose Fused-Core Particle Technology? Fused-Core Out-Performs Totally Porous Particles. HALO particles

More information

Analysis - HPLC A.136. Primesep 5 µm columns B.136

Analysis - HPLC A.136. Primesep 5 µm columns B.136 Primesep 5 µm columns Primesep columns feature double functionality of the bonding i.e : alkyl chain with anionic or cationic group, chelating group. This feature creates unique selectivities when using

More information

How proteins separate on reverse-phase HPLC

How proteins separate on reverse-phase HPLC 1 Reverse Phase How proteins separate on reverse-phase HPLC RP chromatography separates proteins through the interaction of the hydrophobic foot of the protein with a nonpolar surface of the particle RP

More information

Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks

Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks bjective: The purpose of this experiment is to quantify the caffeine content of a diet soda sample using Ultra-High Performance

More information

for Acclaim Mixed-Mode HILIC-1 Column

for Acclaim Mixed-Mode HILIC-1 Column for Acclaim Mixed-Mode HILIC-1 Column Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 1 of 17 Product Manual for ACCLAIM Mixed-Mode HILIC-1 Column 5µm, 4.6 x 250mm, P/N 066844 5µm, 4.6 x 150mm, P/N

More information

1.7 m Fortis UHPLC Columns

1.7 m Fortis UHPLC Columns Strength in Technology 1.7 m Fortis Columns Ultra High Pressure Chromatography Improve your Performance 8 Chemistry Choices Increase Efficiency Increase Speed Improve Resolution Greater Sensitivity Lower

More information

for Acclaim Mixed-Mode WCX-1

for Acclaim Mixed-Mode WCX-1 for Acclaim Mixed-Mode WCX- Product Manual for the Acclaim Mixed-Mode WCX- Column Page of 8 Product Manual for Acclaim Mixed-Mode WCX- Columns Acclaim Mixed-Mode WCX-, 3µm, Analytical column, 3.0 x 50mm

More information

Orosil HPLC Columns. OroSil HPLC columns are designed for the separation of polar, semi-polar, and nonpolar compounds at low to medium ph.

Orosil HPLC Columns. OroSil HPLC columns are designed for the separation of polar, semi-polar, and nonpolar compounds at low to medium ph. Orosil OroSil HPLC columns are designed for the separation of polar, semi-polar, and nonpolar compounds at low to medium ph. Excellent organic base selectivity with very low asymmetry values Compatible

More information

Dilution(*) Chromatography

Dilution(*) Chromatography WA20264 Poster # 184, HPLC 2002, Montreal, 4-5 June 2002 At-Column Column-Dilution for Preparative Dilution(*) Chromatography Cecilia Mazza, Jie Cavanaugh, Ziling Lu,Tom Sirard,Tom Wheat and Uwe Neue Waters

More information

InertSustain AQ-C18. HPLC, LC/MS Columns. Maximizing retention for highly polar compounds in reversed phase methods with highly aqueous mobile phases

InertSustain AQ-C18. HPLC, LC/MS Columns. Maximizing retention for highly polar compounds in reversed phase methods with highly aqueous mobile phases HPLC, LC/MS Columns InertSustain AQ-C8 Maximizing retention for highly polar compounds in reversed phase methods with highly aqueous mobile phases Physical Properties Silica : ES (Evolved Surface) Silica

More information

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times The Secrets of Rapid HPLC Method Development Choosing Columns for Rapid Method Development and Short Analysis Times Rapid Analysis Is More Than Run Time It is developing a method to meet a goal and developing

More information

HPLC Column Selection: Solve the Separation Mystery. The world leader in serving science

HPLC Column Selection: Solve the Separation Mystery. The world leader in serving science HPLC Column Selection: Solve the Separation Mystery The world leader in serving science verview Introduction Historical verview in Stationary Phases Basics in Column Selection Resolution, Efficiency and

More information

QUALITY HPLC COLUMNS PARTICLE PORE PAGE SIZE SIZE. C18 Bidentate C18 Extremely long column life and rugged 4um 100Å 3

QUALITY HPLC COLUMNS PARTICLE PORE PAGE SIZE SIZE. C18 Bidentate C18 Extremely long column life and rugged 4um 100Å 3 Cogent HPLC Columns Manufactured by Table of Contents QUALITY HPLC COLUMNS PARTICLE PORE PAGE SIZE SIZE C18 Bidentate C18 Extremely long column life and rugged 4um 100Å 3 C8 Bidentate C8 Extremely long

More information

HPLC Column Classification

HPLC Column Classification HPLC Column Classification Introduction 1978 USP XIX Fourth supplement : L1 designation for C18 column 1980 USP XX : 7 columns were classified -Classified HPLC column to 56 descriptions -More than 220

More information

Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents

Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents No. 089 SEPARATION REPORT Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS Table of Contents 1. Introduction 1 2. Column Specification 1 3. Features of Packing Materials

More information

CHROMATOGRAPHY. The term "chromatography" is derived from the original use of this method for separating yellow and green plant pigments.

CHROMATOGRAPHY. The term chromatography is derived from the original use of this method for separating yellow and green plant pigments. CHROMATOGRAPHY The term "chromatography" is derived from the original use of this method for separating yellow and green plant pigments. THEORY OF CHROMATOGRAPHY: Separation of two sample components in

More information

RP C18 column with feature of a silanol group

RP C18 column with feature of a silanol group RP C8 column with feature of a silanol group lanol Activity Controlled C8 Column Sunrise ctacosyl C8) Sunrise ctadecyl C8) Sunrise ctadecyl-sac C8-SAC has an interaction of silanol groups ChromaNik Sunrise

More information

C18 Column. Care & Use Sheet

C18 Column. Care & Use Sheet C18 Column Care & Use Sheet HALO Description HALO C18 is a high-speed, high-performance liquid chromatography column based on a new Fused-CoreTM particle design. The Fused-Core particle provides a thin

More information

UPLC for Synthetic Peptides: A User`s Perspective

UPLC for Synthetic Peptides: A User`s Perspective Tides 2007 Conference Las Vegas, May 23, 2007 UPLC for Synthetic Peptides: A User`s Perspective Holger Hermann, Jens Donath* Lonza AG, Switzerland Welcome to Lonza! One of the leading CMO Production facilities

More information

LC Column Troubleshooting. Do you have an equivalent column for my LC column?

LC Column Troubleshooting. Do you have an equivalent column for my LC column? LC Column Troubleshooting Do you have an equivalent column for my LC column? LC columns from different suppliers may have very different retention properties even if the bonding is the same. Although there

More information

High Performance Liquid Chromatography

High Performance Liquid Chromatography High Performance Liquid Chromatography What is HPLC? It is a separation technique that involves: Injection of small volume of liquid sample Into a tube packed with a tiny particles (stationary phase).

More information

Kromasil Eternity Designed for long life. For use in HPLC and UHPLC

Kromasil Eternity Designed for long life. For use in HPLC and UHPLC Kromasil Eternity Designed for long life For use in HPLC and UHPLC The long-lasting phase for demanding applications Imagine working under virtually no ph restrictions during method development. That s

More information

Bringing uhplc Performance to the Separation of Peptides

Bringing uhplc Performance to the Separation of Peptides P e p t i d e E S - C 1 8 U H P L C C o l u m n s Bringing uhplc Performance to the Separation of Peptides Compatible with UHPLC and conventional HPLC equipment. Sumpfstr. 3, CH-6300 Zug; Fax: 041 748

More information

Silanol. Groups Liquid Chromatography

Silanol. Groups Liquid Chromatography Abstract umber: 0-8P Study of Secondary Interaction Based on Residual lanol Groups for Reversed-Phase Liquid Chromatography II orikazu agae Ph.D, Kouji Yamamoto, Chiaki Kadota Chromaik Technologies Inc.

More information

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC USP Method Transfer and Routine Use Analysis of Tablets from HPLC to UPLC Aparna Chavali, Tanya Jenkins and Patricia McConville Waters Corporation, 34 Maple Street, Milford, MA USA APPLICATION BENEFITS

More information

Biochemistry. Biochemical Techniques HPLC

Biochemistry. Biochemical Techniques HPLC Description of Module Subject Name Paper Name 12 Module Name/Title 13 1. Objectives 1.1. To understand the basic concept and principle of 1.2. To understand the components and techniques of 1.3. To know

More information

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster Welcome to our E-Sear: Choosing HPLC Columns for Faster Analysis Smaller and Faster High Throughput/Fast LC Requires. Short columns 0 mm or shorter Small particle sizes. µm Rapid Resolution or new.8 µm

More information

State - of - the -art C18 HPLC Columns

State - of - the -art C18 HPLC Columns An HPLC GL Sciences Newest and Most Advanced ODS Phase -New For State - of - the -art C HPLC s Improved Peak Shapes and Heights Enhancing Sensitivity High Resolution Fast Equilibration Compatible with

More information