Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS

Size: px
Start display at page:

Download "Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS"

Transcription

1 Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS Application note Food safety Authors Bing Yue China National Center For Food Safety Risk Assessment, Beijing, China Juane Song, Ke Xie, Zhao-jun Yun and Jian-qiu Mi Agilent Technologies, Beijing, China Introduction Arsenic (As) is a well-known toxic element that can be present in the environment and foodstuffs. Consequently it is a highly regulated element that is widely monitored. However, As speciation analysis rather than total As analysis is important because its biological toxicity is highly dependent on its chemical form. For example, if a seaweed sample is known to contain a high level of As, but mainly in the form of AsB, then there is no potential risk from its consumption as AsB is non-toxic. Among the five major species of arsenic,, MMA, As(III) DMA and AsB, only the two inorganic forms (As(III) and ) are toxic. They are also carcinogenic to humans, so need to be measured, especially in foodstuffs. Arsenic in rice is of special concern in countries where it is consumed daily. Rice plants absorb As from soil and water. Since rice requires a lot of water to grow compared to other plants, the potential for arsenic accumulation is increased. In this study, we present a rapid method for the determination of the toxic As species in rice using an Agilent 1260 HPLC system equipped with an Agilent ZORBAX SB-Aq column coupled to an Agilent 7900 ICP-MS.

2 Experimental Instrumentation An Agilent 1260 HPLC system comprising a quaternary pump, autosampler and vacuum degasser was coupled to an Agilent 7900 CP-MS. The column exit was simply connected to the nebulizer of the ICP-MS using PFA tubing. The Agilent ZORBAX SB-Aq (part number , 4.6 mm id x 250 mm, 5 μm) reversed phase column was maintained at room temperature throughout the analysis. The mobile phase was 20 mm citric acid and 5 mm sodium hexane sulfonate, adjusted to ph = 4.3 with sodium hydroxide. Two injection volumes, 5 μl or 100 μl, were evaluated. HPLC and ICP MS parameters are given in Table 1. The Agilent 7900 ICP-MS with the Octopole Reaction System (ORS 4 ) collision/reaction cell (CRC) provides sensitive and specific analysis of As in the presence of multiple interferences. In some regulatory guidance, food samples must be extracted using HCl to simulate gastric conditions. In this case, the formation of polyatomic ions based on ArCl and CaCl may cause spectral interferences on the sole isotope of As at m/z 75. However, operating the 7900 ICP-MS ORS 4 in helium mode effectively removes these matrix-based polyatomic interferences on As. Table 1. Operating conditions of the Agilent 1260 HPLC and 7900 ICP-MS ICP-MS parameters RF power Sampling depth Carrier gas Makeup gas Cell gas KED Isotopes monitored Dwell time HPLC parameters Mobile phase Column Flow rate Temperature Injection volume Run time 1550 W 8 mm 0.6 L/min 0.5 L/min He at 4 ml/min 3 V 75 As, 35 Cl 1.0 s/ 75 As; 0.1 s/ 35 Cl Isocratic, 20 mm C 6 H 8 O 7 and 5 mm C 6 H 13 NaO 3 S, adjusted to ph 4.3 with NaOH Agilent ZORBAX SB-Aq 1.2 ml/min Ambient 5 μl or 100 μl 4 min Chemicals and reagents Stock standards of, MMA, As(III), DMA and AsB were purchased from NIMC (National Institute of Metrology China, Beijing, China). High purity citric acid ( 99.5%), sodium hexane-sulfonate ( 98%) and sodium hydroxide ( 99.99%) were purchased from Sigma Aldrich Chemicals (UK). Ultra-pure grade HNO 3 (Sigma Aldrich, UK) was used for the sample preparation. Sample preparation The polished rice samples were obtained from different locations in the following provinces of China: Jiangsu, Fujian, Liaoning, Chongqing and Jiangxi. The polished rice samples were crushed and sieved through a 30 mesh strainer to make rice flour g of the rice flour sample was weighed into a polypropylene centrifuge tube (50 ml). 15 ml of 0.15 M HNO 3 was added and the sample was vortexed for 30 seconds. The tightly capped tube was left to stand overnight at room temperature; it was then placed in an oven at 90 C for 2.5 hours where it was agitated for 30 seconds, every 30 minutes. After the extraction had finished, the sample was cooled in a refrigerator at 4 C. The cooled rice flour suspension was centrifuged at 8000 rpm for 15 minutes and then the supernatant was passed through a 0.22 μm polyether sulphide (PES) filter. The filtrate was stored at 4 C and analyzed within 24 hours to minimize any species inter-conversion. Results and discussion Separation and quantification Calibration standard solutions containing the five As species at different concentrations (1 ppb, 5 ppb, 10 ppb, 50 ppb, 100 ppb and 200 ppb) were prepared and analyzed. The chromatogram for the 10 ppb calibration standard (5 μl injection) shows the five As species are well separated in 4 minutes (Figure 1). The results were used to create calibration curves for the As species, as shown in Figure 2. 2

3 Table 2. Detection limits (ppb) for, MMA, As(III), DMA and AsB with 5μL and 100 μl injections Injection volume MMA As(III) DMA AsB (μl) Figure 1. Chromatogram of a 10 ppb As (conc. as As element) species mixture standard (5 μl injection) Figure 3 shows the chromatograph of a 50 ppt As species standard mixture (100 μl injection) demonstrating the ultra-trace analysis capability of the method. Figure 3. Chromatogram of 50 ppt As species mixture (100 μl injection) Certified Reference Material (CRM) analysis The method was applied to the analysis of two rice flour CRMs; NIST 1568a and NIST 1568b. Though only total As content is certified for NIST 1568a, (290 ± 30 ppb), the CRM has been well-characterized in the literature. The two CRMs were extracted following the aforementioned sample preparation method and analyzed for the five As species. Figure 2. Calibration graphs for, MMA, As(III),DMA and AsB showing excellent linearity. Detection limits Detection limits (DLs) for each arsenic species were calculated as the concentration equivalent to three times the peak-to-peak baseline noise (S/N) of the standard chromatogram. The results are summarized in Table 2. As shown, the DL of the five As species were between ppb and ppb with a 5 μl injection and ppb and ppb with a 100 μl injection. Tables 3 and 4 summarize the quantified results of the inorganic As species, MMA, DMA and the total As content in the two CRMs, the certified values or reported values in the literature. For this study, total As was calculated from the sum of inorganic As, MMA, DMA and AsB. As shown, the quantified inorganic As (ias) in NIST 1568b and total As in both CRMs are in good agreement with the certified values and within the standard deviation. The quantified results for each species are also in good agreement with reported values in the literature [1-4]. 3

4 Table 3. Quantification of As species in NIST 1568b Rice Flour a This study Certified Value MMA As(III) DMA ias b Total As 62.0 ± ± ± ± ± ± ± ± ±14 a value is mean ± SD of ten injections, b inorganic arsenic: sum of As(III) and Table 4. Comparison of quantified As species in NIST 1568a Rice Flour determined in this study compared to values from the literature a. The certified total As content of NIST 1568a is 290 ± 30 ppb. MMA As(III) DMA ias b Total As Ref. [1] 44±2 12±0.8 52±1 173± ±2 Ref. [2] 50.3± ± ± ± ±9.9 Ref. [3] 53.7± ± ± ± Ref. [4] 13.6± ± ± ±3.8 This 64±5 14±2 57±4 140± ±4 study a value is mean ± SD of ten injections, b inorganic arsenic: sum of As(III) and Spiked sample recovery Spiked sample recovery was checked for the five As species. A mixture of, MMA, As(III), DMA and AsB was spiked into three rice sample extracts at 1 ppb, 5 ppb and 10 ppb. Three replicates were prepared (spike 1, 2 and 3). As shown in Table 5, good recoveries ranging from 85% to 110% were achieved for all species, at all concentrations, in each of the spiked solutions. Precision Precision of the method was checked by repeatedly analyzing a rice sample, with ten separate injections. Table 6 shows the concentration results for all the As species, indicating the excellent reproducibility for the ten separate analyses. The %RSDs for all As species were less than 4%, and less than 2% for inorganic As. Table 6. Results of reproducibility of rice # 4 (n = 10) with 5 μl injection Sample MMA As(III) DMA AsB ias N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D Average % RSD 1.3% 0.9% 3.9% 0.5% Table 5. Results of spiked sample recovery with 5 μl injection Sample name Spike Conc. MMA As(III) DMA AsB Conc.(ppb) Rec.(%) Conc.(ppb) Rec.(%) Conc.(ppb) Rec.(%) Conc.(ppb) Rec.(%) Conc.(ppb) Rec.(%) N.D N.D. Spike % % % % % Spike 2 1 ppb % % % % % Spike % % % % % N.D N.D. Spike % % % % % Spike 2 5 ppb % % % % % Spike % % % % % N.D N.D. Spike % % % % % Spike 2 10 ppb % % % % % Spike % % % % % 4

5 Commercial rice sample analysis The As concentration was determined in the 20 different commercial rice samples and the results are summarized in Table 7. As shown,, As(III) and DMA were found in all 20 rice samples. The inorganic As and DMA content varied significantly among the rice samples, likely reflecting the As concentration in the growing environment of the original plant. Although the concentration of toxic inorganic As varied among the samples, all of the rice samples contained lower limit than the upper level of 0.2 mg/kg recommended by Codex Alimentarius Commission. Table 7. Quantitative results (μg/kg) for all five As species in 20 rice samples Sample MMA As(III) DMA AsB ias name N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D N.D Conclusions The separation of five arsenic species in polished rice was carried out using an Agilent 1260 Infinity LC coupled to an Agilent 7900 ICP-MS. The method is rapid - all five species including the toxicologically relevant inorganic forms As(III) and were determined in 4 minutes, with excellent sensitivity, accuracy and precision. Any potential interferences on As, such as ArCl +, were removed by the ORS 4 cell operating in helium mode. Inorganic arsenic and DMA species were detected in all of the twenty rice samples at various concentrations. However, none of the samples contained more than Codex Alimentarius Commission s recommended limit of 0.2 mg/kg. The validated methodology outlined here is suitable for the rapid determination of, MMA, As(III), DMA and AsB at trace levels in rice and other food samples. References 1. Narukawa T., Inagaki K., Kuroiwa T., Chiba K, The extraction and speciation of arsenic in rice flour by HPLC-ICP-MS [J], Talanta 77, 2008, Huang J H, Ilgen G, Blank value, adsorption, preconcentration, and preservation for arsenic speciation of the environmental water samples [J]. Analytica Chimica Acta, 2004, 512: Caruso J.A, B Hymer C, Heitkemper D.T., An evaluation of extraction techniques for arsenic species from freeze-dried apple sample [J]. Analyst, 2001, 126: Toni LIorente-Mirandes et al., A fully validated method for the determination of arsenic species in rice and infant cereal products, Pure Appl. Chem., 2012, vol. 84, No. 2,

6 Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc Published July 15, 2015 Publication number: EN

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Application Note Food Authors Jianqiu Mi, Zhengxiang Zhang, Zhixu Zhang,

More information

DISCLAIMER: This method:

DISCLAIMER: This method: Inorganic arsenic determination in fresh mussels using water bath extraction and anion exchange chromatography-inductively coupled plasma mass spectrometry DISCLAIMER: This method: - has to be considered

More information

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Application Note Food Testing and Agriculture Author Chen-Hao (Andy) Zhai and Rong-jie Fu Agilent Technologies (Shanghai)

More information

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS ICP-MS Environmental Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS Application Note Steve Wilbur Analysis of arsenic and selenium in seawater at trace

More information

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors Determination of major, minor and trace elements in rice flour using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Application note Food testing Authors John Cauduro Agilent Technologies,

More information

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS Enhancing the productivity of food sample analysis with the Agilent 77x ICP-MS Application note Foods testing Authors Sebastien Sannac, Jean Pierre Lener and Jerome Darrouzes Agilent Technologies Paris,

More information

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Application note Semiconductor Authors Michiko Yamanaka, Kazuo Yamanaka and Naoki

More information

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS Application note Semiconductor Authors Junichi Takahashi Agilent Technologies, Japan Introduction

More information

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 126 Infinity Binary LC system with UV detection Application Note Food Author Srividya Kailasam Agilent Technologies,

More information

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Application note Pharmaceutical Authors Miao Jing, Yingping Ni, Yanping Wang and Zhixu Zhang Agilent Technologies, China Introduction

More information

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry APPLICATION NOTE Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry AN43227 Antonella Guzzonato 1, Shona McSheehy Ducos

More information

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS Application ote Food Testing & Agriculture Author Andy Zhai Agilent Technologies, Inc. Shanghai Co. Ltd.

More information

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) Application Note Environmental Authors Steve Wilbur Agilent Technologies,

More information

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode The ultratrace determination of iodine in aqueous samples using the 7700x ICP-MS with oxygen reaction mode Application note Nuclear Authors Kazumi Nakano, Yasuyuki Shikamori, Naoki Sugiyama and Shinichiro

More information

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors Trace-Level Analysis of Melamine in Milk Products on Agilent 789A/5975C GC/MSD Using a ew Agilent J&W DB-5ms Ultra Inert Column and SampliQ SCX Cartridges Application ote Gas Chromatography/Mass Spectrometry/Food

More information

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS Application note Environmental Authors Steve Wilbur, Craig Jones Agilent Technologies,

More information

Simultaneous Arsenic and Chromium Speciation by HPLC/ICP-MS in Environmental Waters

Simultaneous Arsenic and Chromium Speciation by HPLC/ICP-MS in Environmental Waters Simultaneous Arsenic and Chromium Speciation by HPLC/ICP-MS in Environmental Waters Introduction Many elements can exist in a variety of oxidation states which have differing impacts on health and the

More information

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Application note Energy and fuels Authors Jenny Nelson, Agilent Technologies, USA Ed McCurdy, Agilent Technologies, UK Introduction

More information

Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan Kim, and Daniel R. Wiederin

Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan Kim, and Daniel R. Wiederin Elemental Scientific prepfast IC: Inline Autodilution Method to Eliminate Species Interconversion for LC-ICPMS Based Applications Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan

More information

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539) Determination of ormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell P Column (EPA 539) Application Note Environmental Authors Rong-jie Fu and Chen-ao (Andy) Zhai Agilent Technologies

More information

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell Trace elemental analysis of distilled alcoholic beverages using the Agilent 77x ICP-MS with octopole collision/ reaction cell Application note Food testing Author Glenn Woods Agilent Technologies Cheadle

More information

Analysis of domestic sludge using the Agilent 4200 MP-AES

Analysis of domestic sludge using the Agilent 4200 MP-AES Analysis of domestic sludge using the Agilent 4200 MP-AES Application note Environmental Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction Managing the treatment and disposal

More information

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 66 Triple Quadrupole LC/MS System Application Note Environmental Authors Don Noot, Matthew Noestheden, and Ralph Hindle Vogon

More information

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Application note Geochemistry, metals, mining Authors John Cauduro Agilent Technologies, Mulgrave, Australia Introduction

More information

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity Agilent 12 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity A case study with degraded omeprazole drug product Application Note Small Molecule

More information

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis The Agilent 77x ICP-MS Advantage for Drinking Water Analysis Application Note Environmental Authors Steve Wilbur Agilent Technologies, Inc. 338 146th Place SE, Suite 3, Bellevue Washington, 987 USA Introduction

More information

A Comparison of Three Methods for Arsenic Speciation in Biological Tissues. May Nguyen Brooks Rand Labs Seattle, WA

A Comparison of Three Methods for Arsenic Speciation in Biological Tissues. May Nguyen Brooks Rand Labs Seattle, WA A Comparison of Three Methods for Arsenic Speciation in Biological Tissues May Nguyen Brooks Rand Labs Seattle, WA Select Arsenic Species Inorganic Species arsenite [As(III)] arsenate [As(V)] Organic Species

More information

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction. Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS Application Environmental Author Steven Wilbur Agilent Technologies, Inc. 3380 146th Place, SE, Suite 300 Bellevue, WA

More information

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Application Note Semiconductor Authors Junichi Takahashi Agilent Technologies Tokyo, Japan Abstract Ammonium hydroxide

More information

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Elemental analysis of river sediment using the Agilent 4200 MP-AES Elemental analysis of river sediment using the Agilent 4200 MP-AES Application note Environmental: Soils, sludges & sediments Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction

More information

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS Application Note Food Testing Author John Cauduro Agilent Technologies, Australia Introduction Trace

More information

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note Authors Siji Joseph Agilent Technologies India Pvt. Ltd.

More information

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast Application Note Environmental Authors Austin Schultz and Jake Unnerstall, Elemental Scientific, Omaha NE, USA Steve Wilbur,

More information

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Adulteration in Dietary Supplements and Botanicals Application Note Food Testing & Agriculture Authors James Neal-Kababick

More information

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes Application Note Food Safety Author Sunil Kulkarni Agilent Technologies

More information

LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges

LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges Application ote Food Testing & Agriculture Author Andy Zhai Agilent Technologies Shanghai Co. Ltd. Abstract This application

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

Advances in Interference Removal for Accurate Arsenic Analysis in Food and Beverages

Advances in Interference Removal for Accurate Arsenic Analysis in Food and Beverages Advances in Interference Removal for Accurate Arsenic Analysis in Food and Beverages National Environmental Monitoring Conference 2013 Austin TX Steve Wilbur and Amir Liba Agilent Technologies Let s Define

More information

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3 Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3 Author Application Environmental Tetsuo Kubota Agilent Technologies, Japan Introduction Regulations concerning activities such

More information

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS APPLICATION NOTE 43326 Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS Authors Introduction Tomoko Vincent1, Simon Lofthouse2,

More information

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ Application Note Consumer products Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ Effective removal of tungsten-based interferences on five Hg isotopes using MS/MS Authors

More information

Characterization of Arsenic Species in Apple Juice using a NexSAR HPLC-ICP-MS Speciation Analysis Ready Solution

Characterization of Arsenic Species in Apple Juice using a NexSAR HPLC-ICP-MS Speciation Analysis Ready Solution APPLICATION NOTE HPLC-ICP-MS Author: Eve M. Kroukamp PerkinElmer, Inc. Woodbridge, ON, Canada Characterization of Arsenic Species in Apple Juice using a NexSAR HPLC-ICP-MS Speciation Analysis Ready Solution

More information

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors. Sensitive Femtogram Determination of Aflatoxins B 1, B 2, G 1 and G 2 in Food Matrices using Triple Quadrupole LC/MS Application Note Food Safety Authors Yang Chen and Jack Cappozzo National Center for

More information

Samples compliant with the WATER

Samples compliant with the WATER ENVIRONMENTAL analysis A routine method for the Quantitative Measurement of trace metals in WATER Samples compliant with the WATER Quality Standard EN ISO 17294 on the Agilent 7000 Series ICP-MS. Solutions

More information

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode Lead isotope analysis: Removal of Hg isobaric interference from using ICP-QQQ in MS/MS mode Application note Authors Geochemistry and isotope analysis Glenn Woods Agilent Technologies, LDA UK Ltd., Stockport,

More information

Savillex Technical Note Performance of the DST-1000 Acid Purification System

Savillex Technical Note Performance of the DST-1000 Acid Purification System Savillex Technical Note Performance of the DST-1000 Acid Purification System Summary The Savillex DST-1000 acid purification system is a sub-boiling still manufactured from PFA that produces high purity

More information

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System Application Note Foods Author Glenn Woods ICP-MS Specialist Agilent Technologies UK Ltd. Lakeside Business

More information

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS Authors Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS Application Note Food safety Craig Jones and Jenny Nelson Agilent Technologies, USA Introduction In the U.S., marijuana remains

More information

Using TOF for Screening and Quantitation of Sudan Red Colorants in Food Application

Using TOF for Screening and Quantitation of Sudan Red Colorants in Food Application Using TOF for Screening and Quantitation of Sudan Red Colorants in Food Application Food Authors Yanyan Fang Agilent Technologies, Inc. 412 Ying Lun Road Pu Dong, Shanghai 200131 China Michael Zumwalt

More information

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Application Note Semiconductor Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Tokyo, Japan Introduction

More information

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS Application note Metals Analysis & Production Authors Naoki Sugiyama and Mineko Omori Agilent Technologies,

More information

The analysis of organic acid content of additives, premix, feed, and water.

The analysis of organic acid content of additives, premix, feed, and water. The analysis of organic acid content of additives, premix, feed, and water. Contents Foreword Introduction Warnings 1. Scope 2 1.1 LOD and LOQ 3 2. Normative References 3 3. Definitions 3 3.1 Feed (or

More information

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS Accurate analysis of neptunium in a uranium matrix, using ICP-QQQ with MS/MS Application note Nuclear, environmental Authors Garry Duckworth, Springfields Fuels Ltd, K, Glenn Woods, Agilent Technologies,

More information

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS Application note Semiconductor Author Naoki Sugiyama Agilent Technologies, Tokyo,

More information

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System Application Note Authors Wenwen Wang, Zhe Cao Agilent Technologies, Inc. No.3, Wang Jing Bei Lu, Chao Yang District,

More information

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction Application Note Consumer Products Authors Devon C. Zimmerman, Henry F. Rossi III, Jacqueline Rizzo, Daniel W. Keating,

More information

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application Environmental Author Neal Julien Midwest Research Institute

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009. Translated English of Chinese Standard: GB5009.17-2014 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD OF GB THE PEOPLE S REPUBLIC OF CHINA National Food Safety Standard-Determination

More information

Analytical Approaches to Arsenic

Analytical Approaches to Arsenic Analytical Approaches to Arsenic (and Selenium) Analysis in Foods Brian Jackson Trace Element Analysis Core Dartmouth College Arsenic in food Arsenic levels in rice and rice products FDA data WHO data

More information

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode Application note Semiconductor Author Katsuo Mizobuchi Masakazu

More information

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD Application Note Environmental Author Suli Zhao, Andy Zhai Agilent Technologies Co., Ltd. 412 Yinglun Road

More information

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS Tomoko Vincent 1, Simon Lofthouse 2, Daniel Kutscher 1 and Shona McSheehy Ducos

More information

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract. Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS Application Food Analysis Authors Sheher Mohsin Agilent Technologies, Inc. Schaumburg, Il USA Michael Zumwalt

More information

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India QbD Based Method Development on an Agilent 129 Infinity UHPLC System Combined with a Seamless Method Transfer to HPLC Using Intelligent System Emulation Technology Application Note Pharmaceutical QA/QC

More information

Application Note. Abstract. Authors. Environmental

Application Note. Abstract. Authors. Environmental Determination of Acidic Herbicides using an Agilent 6460 Triple Quadrupole LC/MS Equipped with Agilent Jet Stream Technology and Direct Aqueous Injection, for Potable and Environmental Samples Application

More information

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS APPLICATION NOTE 43374 Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS Authors Tomoko Vincent, Product Specialist, Thermo Fisher Scientific Keywords

More information

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Sensitive Detection of 2-MIB and Geosmin in Drinking Water Sensitive Detection of -MIB and Geosmin in Drinking Water Application Note Environmental Author Yean-Woong You Agilent Technologies, Inc. Seoul, Korea Abstract An automated SPME extraction method for easy

More information

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 646 Triple Quadrupole LC/MS system with Triggered MRM Application Note Authors Thomas Glauner Agilent Technologies,

More information

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

Pesticides Analysis Using the Agilent 5977A Series GC/MSD Pesticides Analysis Using the Agilent 5977A Series GC/MSD Application Note Food Testing and Agriculture Authors Jia-jia Wu and Wen-wen Wang Agilent Technologies Co. Ltd (China) Beijing, People s Republic

More information

A Comparison of Arsenic Speciation

A Comparison of Arsenic Speciation A Comparison of Arsenic Speciation Methods for Biological Tissues Tamas Ugrai (tamas@brooksrand.com) Authors: T. Ugrai, T. Ellis, M. Nguyen, P. Petrov, J. Creswell, A. Carter, and M. Briscoe Brooks Rand

More information

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Application Note Pharmaceutical Authors Rebeca Santamaria-Fernandez, SheilaMerson,

More information

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India Analysis of fumonisin, FB1 and FB2 mycotoxins in corn food and feed samples using the Agilent 1120 Compact LC System coupled to the Agilent 6140 Single Quadrupole LC/MS System Application Note Food Safety

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS Application Note Traditional Chinese Medicine Author Rongjie Fu Agilent Technologies

More information

Limit-test of NDMA and NDEA in Sartans. by GC-MS (Liquid-direct-injection)

Limit-test of NDMA and NDEA in Sartans. by GC-MS (Liquid-direct-injection) Limit-test of NDMA and NDEA in Sartans 1 Purpose and scope by GC-MS (Liquid-direct-injection) The method is used to test the permissible limits of N-nitrosodimethylamine and N- nitrosodiethylamine in various

More information

DEVELOPMENT AND VALIDATION OF A HPLC METHOD FOR IN-VIVO STUDY OF DICLOFENAC POTASSIUM

DEVELOPMENT AND VALIDATION OF A HPLC METHOD FOR IN-VIVO STUDY OF DICLOFENAC POTASSIUM IJPSR (2013), Vol. 4, Issue 2 (Research Article) Received on 28 September, 2012; received in revised form, 24 November, 2012; accepted, 23 January, 2013 DEVELOPMENT AND VALIDATION OF A HPLC METHOD FOR

More information

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors.

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors. Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology Part 2 of a 3 part series on Environmental Analysis Application Environmental Authors

More information

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 77A mini TD and 70A GC Application Note Environmental Authors Tingting Bu, Xiaohua Li Agilent Technologies (Shanghai) Co.,

More information

Determination of iopromide in environmental waters by ion chromatography-icp-ms

Determination of iopromide in environmental waters by ion chromatography-icp-ms Determination of iopromide in environmental waters by ion chromatography-icp-ms Application note Environmental Authors Armando Durazo, Tarun Anumol, and Shane A. Snyder Department of Chemical & Environmental

More information

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry Application Note Food safety Authors Ismael Flores and Carlos Sepulveda Agrolab México Km 7 Carretera Pachuca-Actopan

More information

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Thermo Scientific icap RQ ICP-MS: Typical limits of detection TECHNICAL NOTE 43427 Thermo Scientific icap RQ ICP-MS: Typical limits of detection Author Tomoko Vincent Keywords BEC, interference removal, KED, LOD Introduction Inductively Coupled Plasma Mass Spectrometry

More information

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs Application note Authors Food safety Kazuhiro Sakai 1, Junichi

More information

Supporting Information

Supporting Information 1 Supporting Information 2 3 Discovery and implications of C 2 and C 3 perfluoroalkyl sulfonates in aqueous film forming foams (AFFF) and groundwater 4 Krista A. Barzen-Hanson a and Jennifer A. Field b*

More information

Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS

Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS Application note Environmental Author Sarah Hill and Rebeca Santamaria-Fernandez

More information

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS Kate Comstock, Tim Stratton, Hongxia (Jessica) Wang, and Yingying Huang

More information

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ Application Note Semiconductor Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Japan Introduction Hydrochloric

More information

Determination of Alkaloids in Goldenseal Using Agilent Bond Elut Plexa Solid Phase Extraction Sorbent for Cleanup and HPLC-DAD Analysis

Determination of Alkaloids in Goldenseal Using Agilent Bond Elut Plexa Solid Phase Extraction Sorbent for Cleanup and HPLC-DAD Analysis Determination of Alkaloids in Goldenseal Using Agilent Bond Elut Plexa Solid Phase Extraction Sorbent for Cleanup and HPLC-DAD Analysis Application Note Dietary Supplement Author Kediemetse Mothibedi,

More information

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Application Note Clinical Research Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Authors Yanan Yang 1, Victor Mandragon 2, and Peter Stone 1 1

More information

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION Andrew J. Aubin and Tanya L. Jenkins Waters Corporation, Milford, MA, USA INTRODUCTION Benzocaine (4-Aminobenzoic

More information

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note Environmental Author Chin-Kai Meng Agilent Technologies 85 Centerville Road Wilmington, DE 988-6 USA Abstract Chlorinated

More information

Automated Sample Preparation in Quality Control of Eye-Drop Formulation

Automated Sample Preparation in Quality Control of Eye-Drop Formulation Automated Sample Preparation in Quality Control of Eye-Drop Formulation Application Note Authors Bart Tienpont, Maria Rambla Alegre, Frank David, and Pat Sandra Research Institute for Chromatography Kennedypark

More information

Determination of Butyltin Compounds in Environmental Samples by Isotope Dilution GC-MS. Application. Authors. Abstract. Introduction.

Determination of Butyltin Compounds in Environmental Samples by Isotope Dilution GC-MS. Application. Authors. Abstract. Introduction. Determination of Butyltin Compounds in Environmental Samples by Isotope Dilution GC-MS Application Environmental Authors Giuseppe Centineo, Pablo Rodríguez-González, and J. Ignacio García Alonso Department

More information

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry Application Note Environmental Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry Authors Ye Kong and Zhe Cao Agilent Technologies,

More information

Dartmouth Trace Element Analysis Core: Lab methods

Dartmouth Trace Element Analysis Core: Lab methods Dartmouth Trace Element Analysis Core: Lab methods General All prices are based on analysis of up to 14 elements of interest to the investigator. More analytes can be included in the run after discussion

More information

materials analysis Solutions for Your Analytical Business Markets and Applications Programs

materials analysis Solutions for Your Analytical Business Markets and Applications Programs materials analysis DETERMINATION OF DIISOCYANATES IN PLASTICS USED IN TEXTILE PRODUCTS USING THE AGILENT 6410B LC/MS/MS Solutions for Your Analytical Business Markets and Applications Programs Author Dr.

More information

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS Technical Overview Abstract Inductively coupled plasma mass spectrometry

More information

Anaylsis of Pesticide Residues in Rice Using Agilent Bond Elut QuEChERS AOAC Kit by LC-MS/MS Detection

Anaylsis of Pesticide Residues in Rice Using Agilent Bond Elut QuEChERS AOAC Kit by LC-MS/MS Detection Anaylsis of Pesticide Residues in Rice Using Agilent Bond Elut QuEChERS AOAC Kit by LC-MS/MS Detection Application Note Food Safety Author Andy Zhai Agilent technologies Co., Ltd. 412 Yinglun Road Shanghai,

More information

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120 LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120 Application Note Food Testing & Agriculture Author Irina Dioumaeva Agilent Technologies, Inc. Abstract

More information

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Thermo Scientific icap TQ ICP-MS: Typical limits of detection TECHNICAL NOTE 43287 Thermo Scientific icap TQ ICP-MS: Typical limits of detection Authors Tomoko Vincent Keywords BEC, LOD, SQ-KED, TQ mass shift, TQ on mass, typical performance Introduction Inductively

More information

Quantification of Furocoumarins

Quantification of Furocoumarins Version 8 (12/04/2013) Quantification of Furocoumarins HPLC/DAD Procedure Scope The European Cosmetics Directive 76/768/EEC, which is about to be adopted in the European Cosmetic Regulation, in Annex II

More information