Determinations of Heat Stable Amine Salts and Amines in Amine Scrubbing Solutions Using Suppressed Conductivity and Charge Detection

Size: px
Start display at page:

Download "Determinations of Heat Stable Amine Salts and Amines in Amine Scrubbing Solutions Using Suppressed Conductivity and Charge Detection"

Transcription

1 Determinations of Heat Stable mine Salts and mines in mine Scrubbing Solutions Using Suppressed Conductivity and Charge Detection Terri Christison, Monika Verma, Pranathi Perati, Linda Lopez Thermo Fisher Scientific, Sunnyvale, C, US

2 Overview Results Heat stable amine (HS) salts, such as organic acids, inorganic anions, and sulfur based salts, in amine scrubber solutions were separated by anion-exchange on a µm particle high-efficiency capillary column. Charge Detection This method utilizes two different but complementary detectors, suppressed conductivity and charge detection. Charge detection provides higher relative responses for multi-charged anions, such as found in heat stable amine salt solutions, than chloride and sulfate. Using these two detectors with different responses were used to confirm peak purity and the presence of co-eluting peaks, thereby improving reporting accuracy. The Dionex QD Ch dissociated compo the charge respon proportionally high FIGURE. Mecha. Introduction n Sour crude natural gas (methane) is typically treated with amine rich scrubber solutions to neutralize carbon dioxide and remove hydrogen sulfide gas impurities before it can be sold as a pure natural gas product., When the neutralizing capacity is deemed inefficient, the amine solutions are regenerated and stripped of elemental sulfur. Dissolved salts (heat stable amine salts) remain, building up over time resulting in higher maintenance costs, and higher incidents of corrosion., nalysis of both the amine concentrations and the heat stable amine salts are needed to ensure a pure product and an efficient scrubbing process. The Thermo Scientific Dionex QD Charge Detector promotes complete dissociation of many weakly dissociated compounds as a result, the charge responses of singlycharged, doubly-charged, and triply-charged ions are proportionally higher than conductivity. CD and QD detector provided different responses were used to confirm peak identity and the presence of unknown co-eluting peaks. Methods Figures 5 demon used for improved Figure shows an detection (R) whic and R) clearly def Samples / Sample Preparation Sour gas amine samples were diluted -fold with deionized water. Some amine samples were treated with Thermo Scientific Dionex OnGuard II RP and Dionex OnGuard II Na cartridges to remove interfering cations and organic compounds. Figure 5 illustrates C concentrations concentration. Instrumentation FIGURE. Using Thermo system. Scientific Dionex ICS-5+ HPIC Ion Chromatography dual capillary Example : QD co DP Dual Pump, EG Eluent Generator, DC Detector Chromatography, and Thermo Scientific Dionex IC Cube modules Thermo Scientific Dionex Detectors: CD Conductivity Detector, QD Charge Detector with Thermo Scientific Dionex QDC Cell utosampler: Thermo Scientific Dionex S-P utosampler Data nalysis: Thermo Scientific Dionex Chromeleon Chromatography Data System Conditions: Thermo Scientific Dionex IonPac GHC-µm, Dionex IonPac SHC-µm,. 5 mm Eluent Source: Thermo Scientific Dionex EGC KOH cartridge, capillary mm KOH (-7 8 min), mm (8 8 min), 85 mm (8 8 min), 85 mm (8 5 min) Flow Rate:.5 ml/min Column Temp.: C Inj. Volume:. µl : Suppressed conductivity, Thermo Scientific Dionex CES nion Capillary Eleyctolytic Suppressor, recycle mode; m : Charge detection, Dionex QDC cell, 6 V Stable mine Salts diagram and minesfor in mine Scrubbing Solutions Using Conductivity and Charge Detection Determinations of Heat FIGURE. Flow the IC system using CDSuppressed and QD detectors. FIGURE 5. Using Example : CD co

3 Thermo Scientific Dionex Detectors: CD Conductivity Detector, QD Charge Detector with Thermo Scientific Dionex QDC Cell utosampler: Thermo Scientific Dionex S-P utosampler Data nalysis: Thermo Scientific Dionex Chromeleon Chromatography Data System Conditions: Thermo Scientific Dionex IonPac GHC-µm, Dionex IonPac SHC-µm,. 5 mm Eluent Source: Thermo Scientific Dionex EGC KOH cartridge, capillary mm KOH (-7 8 min), mm (8 8 min), 85 mm (8 8 min), 85 mm (8 5 min) Flow Rate:.5 ml/min Column Temp.: C Inj. Volume:. µl : Suppressed conductivity, Thermo Scientific Dionex CES nion Capillary Eleyctolytic Suppressor, recycle mode; m : Charge detection, Dionex QDC cell, 6 V FIGURE. Flow diagram for the IC system using CD and QD detectors. FIGURE 5. Using C Example : CD co H High-Pressure Non-Metallic Pump Eluent Generator (OH ) Data Management Charge Detector Separation Column CR-TC Waste Electrolytic Eluent Suppressor Conductivity Detector Figure shows the Dionex IC Cube module and Dionex QDC Charge Detector cell. The Dionex IC Cube is designed to minimize the flow path needed for a capillary flow rate system by containing the columns and consumables in close proximity.. FIGURE. Dionex IC Cube with capillary modules Thermo Scientific Dionex CRD 8 Carbonate Removal Device cartridge Dionex CES Suppressor Dionex Conductivity Detector (CD) Figure 6 shows the acids, and sulfur-b were normalized to which are triply charged nitrite. Th response when no FIGURE 6. Separ Guard and Separation Column tray Dionex IC Cube -Port Injection Valve Thermo Scientific Dionex Degas cartridge Dionex QDC Charge Detector Cell -.5 Thermo Scientific Poster Note PN78_HPLC E_5/S

4 Results ons, and sulfur xchange on a µm pressed relative mine salt solutions, nfirm peak purity ccuracy. Charge Detection The Dionex QD Charge Detector promotes complete dissociation of many weakly dissociated compounds by drawing a current at a fixed potential (Figure ). s a result, the charge responses of singly-charged, doubly-charged, and triply-charged ions are proportionally higher than conductivity. FIGURE. Mechanism of QD Charged Detector FIGURE 7 8: n. crubber solutions es before it can be s deemed tal sulfur. resulting in higher th the amine pure product and ete dissociation of es of singlygher than sed to confirm ome amine RP and Dionex mpounds. 6V Cathode node H FIGURE 7: CD v Y- + + Y- 5 + H 5 6 nion-exchange Membrane Cation-Exchange Membrane Figures 5 demonstrate how the different peak responses by the CD and QD can be used for improved data analysis. Figure shows an example where the peaks have with the same response by CD detection (R) which could be either C or C concentrations; but with QD detection (R and R) clearly defines the concentrations. Figure 5 illustrates another example were QD detection (R) correspond to both C and C concentrations but is clarified by CD detection where R and R confirm C concentration. FIGURE. Using CD and QD detection for peak analysis. ual capillary Figures 7 9 show neutralizing sam was pretreated w compounds and Dionex sample p without introduci QD chromatgram comparison. In F boxes) not notice M Figure 9 show th contamination, s FIGURE 9. 6 Example : QD confirmation. phy, and Thermo CD QD Charge Detector QD or R R R -.5 tem C C C C C-µm, Conclusio capillary 85 mm FIGURE 5. Using CD and QD detection for peak analysis. Example : CD confirmation. Dionex essor, recycle QD R CD or Determinations of Heat Stable mine Salts and mines in mine Scrubbing Solutions Using Suppressed Conductivity and Charge Detection R ors. R Heat stable am organic acids Here we demo stable amine s This method h capillary-size (HPIC) capilla Using differen

5 Conclusion pillary 85 mm FIGURE 5. Using CD and QD detection for peak analysis. Detector cell. pillary flow y. QD This method has capillary-size Dio (HPIC) capillary or R R CD R s. ty Here we demon stable amine sal Example : CD confirmation. Dionex sor, recycle rge Detector Heat stable amin organic acids an C C Using different d detector provide result in improve References C C Figure 6 shows the separation of a standard containing 5 inorganic anions, organic acids, and sulfur-based anions typically found as heat stable amine salts. QD responses were normalized to the nitrite peak. Peak which is doubly charged and Peaks and which are triply charged have higher QD response than strongly ionized singlely charged nitrite. Thiocyanate is a singly charged ion which had a comparable relative response when normalized to nitrite. FIGURE 6. Separation of mixed standard using CD and QD detection Cube 8 Scientific egas µ C tector Cell Minutes 5 Dionex IonPac S-HC-µm with guard Capillary (. 5 mm,. 5 mm) Eluent Source: Dionex, EGC-KOH (Capillary) cartridge Potassium hydroxide mm (-7 8 min), mm (8 8 min), 85 mm (8 8 min), 85 mm (8 5 min) Flow Rate:.5 ml/min Inj. Volume:. µl Column Temp.: C : Suppressed conductivity, Dionex CES suppressor, recycle mode : Dionex QD Charge Detection, 6 V Sample Prep.:. Natural Gas.or. Rooney, P.C.; Corrosivity, Pa Thermo Fisher in Methyldietha Sunnyvale, C lication%&% and%rfi Methyldiethano Peaks:. Fluoride. mg/l 9. Carbonate. cetate.. Sulfite. Propionate.. Sulfate*. Formate.. Oxalate 5. Chloride*. Phosphate 6. Nitrite.. Thiosulfate 7. romide. 5. Thiocyanate 8. Nitrate mg/l ll trademarks are the This information is no the intellectual proper * CD response for chloride and sulfate was truncated at. QD chromatogram was normalized at the nitrite peak. Thermo Scientific Poster Note PN78_HPLC E_5/S 5

6 y weakly ). s a result, ged ions are Figures 7 9 show the determination of heat stable amine salts (HS) in amine neutralizing samples. Prior to injection, the diluted scrubber sample in Figures 7 and 8 was pretreated with Dionex OnGuard RP and Dionex OnGuard Na to remove organic compounds and cation contaminants that were affecting the chromatography. These Dionex sample preparation cartridges are designed to remove select contaminants without introducing IC-detectable ionic contaminants. In Figure 7, the chloride peak in the QD chromatgram was normalized to the same peak in the CD chromatogram for easier comparison. In Figure 8, an expanded QD chromatogram reveals additional peaks (in boxes) not noticeable by CD, thereby providing additional information and peak purity. FIGURE 7 8: nion determinations in -fold diluted amine scrubber solution. FIGURE 7: CD versus QD FIGURE 8: Expanded scale of QD detection. dditional peaks are revealed µ d QD can be Minutes *oxes denote detection of additional peaks se by CD D detection (R o both C and rm C µ 8 6 Conditions:: see Figure 6. : Suppressed conductivity, Dionex CES suppressor, recycle mode : QD Charge Detection, 6 V Sample Prep.: -fold dilution with deionized water, Dionex OnGuard RP and OnGuard Na treatment Peaks (mg/l):. cetate 9. Propionate.8. Formate. Unknown -5. Chloride.9 6. Unknown -7. Carbonate -8. Sulfate.9 9. Oxalate. Unknown. Phosphate. Thiosulfate. Thiocyanate * QD was normalized to the chloride peak Figure 9 show the CD and QD chromatograms of an amine solution with less HS contamination, sampled at an overhead actuator. FIGURE 9. -fold Dilution of an amine solution from an overhead actuator µ Minutes 5 Dionex IonPac S-HC-µm with guard Capillary (. 5 mm,. 5 mm) Eluent Source: Dionex EGC-KOH (Capillary) cartridge Potassium hydroxide, mm (-7 8 min), mm (8 8 min), 85 mm (8 8 min), 85 mm (8 5 min) Flow Rate:.5 ml/min Inj. Volume:. µl Column Temp.: C : Suppressed conductivity, Dionex CES suppressor, recycle mode : QD Charge Detection, 6 V Sample Prep.: Diluted -fold with deionized water Peaks:. cetate. Carbonate. Sulfite. Sulfate 5. Thiosulfate mg/l * QD was normalized to the sulfate peak Conclusion Heat stable amine salts are challenging samples for analysis with many sulfur species, organic acids and high concentrations of inorganic anions. Here we demonstrated an IC method needed by the gas industry to analyze heat stable amine salts in amines samples. This method has high efficiency separations made possible on a µm particle capillary-size Dionex IonPac S-HC-µm column and facilitated by a high-pressure (HPIC) capillary IC system. 6 Determinations of Heat Stable mine Salts and mines in mine Scrubbing Solutions Using Suppressed Conductivity and Charge Detection Using different detectors, such as the CD conductivity detector and the QD charge

7 . Sulfate 5. Thiosulfate * QD was normalized to the sulfate peak Conclusion Heat stable amine salts are challenging samples for analysis with many sulfur species, organic acids and high concentrations of inorganic anions. Here we demonstrated an IC method needed by the gas industry to analyze heat stable amine salts in amines samples. This method has high efficiency separations made possible on a µm particle capillary-size Dionex IonPac S-HC-µm column and facilitated by a high-pressure (HPIC) capillary IC system. Using different detectors, such as the CD conductivity detector and the QD charge detector provide advantages in improved peak identification and peak detection which result in improved reporting accuracy and information. References ons, organic s. QD responses d Peaks and zed singlely able relative n. Natural Gas.org Home Page. (accessed pril 7, ).. Rooney, P.C.; acon, T.R.; DuPart, M.S. Effect of Heat Stable Salts on MDE Corrosivity, Parts and. Hydrocarbon Processing, 996, Mar, 95 ; 997, pr, Thermo Fisher Scientific Technical Note : Separation of Heat Stable mine Salts in Methyldiethanolamine (MDE) Solutions. Thermo Scientific P/N TN79_E, Sunnyvale, C,. [Online] lication%&%technical%notes/chromatography/ion%chromatography/ic% and%rfic%systems/tn--separation-heat-stable-mine-saltsmethyldiethanolamine-solutions-tn79-e.pdf (accessed pril 7, ). S-HC-µm with guard 5 mm,. 5 mm) KOH (Capillary) cartridge droxide n), mm (8 8 min), 8 min), min) d conductivity, suppressor, recycle Charge Detection, 6 V arbonate lfite lfate* alate osphate iosulfate iocyanate mg/l ll trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. d sulfate was truncated was normalized at the nitrite PO78_E 5/S Thermo Fisher Scientific Inc. ll rights reserved. ISO is a trademark of the International Standards Organization. ll other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. frica + 5 ustralia ustria elgium razil Canada China (free call domestic) Denmark Europe-Other + 5 Finland France Germany India Italy Japan Korea Latin merica Middle East + 5 Netherlands New Zealand Norway Thermo Fisher Scientific, Sunnyvale, C US is ISO 9:8 Certified. Russia/CIS + 5 Singapore Sweden Switzerland Taiwan UK/Ireland US PN78_E 5/S

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection Terri Christison, Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA Overview

More information

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC TECHNICAL NOTE 122 Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC Authors Terri Christison and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA,

More information

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems Yan Liu, Zhongqing Lu, and Chris Pohl; Thermo Fisher Scientific, Sunnyvale, CA USA Overview

More information

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Thunyarat Phesatcha, Suparerk Tukkeeree, Jeff Rohrer 2 Thermo

More information

Utility of the Charge Detector in Ion Chromatography Applications

Utility of the Charge Detector in Ion Chromatography Applications Utility of the Charge Detector in Ion Chromatography Applications Mrinal K. Sengupta, Sheetal Bhardwaj, Kannan Srinivasan, Chris Pohl, and Purnendu K. Dasgupta Thermo Fisher Scientific, Sunnyvale, California,

More information

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection Barbara Shao, Terri Christison, Fei Pang, Cathy Tanner, and Frank Hoefler, Thermo Fisher Scientific, Sunnyvale,

More information

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography Yan Liu, Zhongqing Lu, and Chris Pohl, Thermo Fisher Scientific, Sunnyvale, CA USA Overview Purpose: In this work, new high-pressure

More information

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System arbara Shao, Frank Hoefler, Maria Rey, and Linda Lopez, Thermo Fisher Scientific,

More information

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection Determination of Tartaric cid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection Suparerk Tukkeeree, Chanita Chantarasukon, and Jeff Rohrer Thermo Fisher Scientific, angkok,

More information

Dionex IonPac AS28-Fast-4µm column

Dionex IonPac AS28-Fast-4µm column CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS-Fast-4µm Columns Product Specifications The Thermo Scientific Dionex IonPac AS-Fast-4µm column is a high-capacity, hydroxide-selective anionexchange column

More information

Thermo Scientific. Anion-Exchange Column

Thermo Scientific. Anion-Exchange Column CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS9-µm Anion-Exchange Column Product Specifications The Themo Scientific Dionex IonPac AS9-µm high-capacity, hydroxide-selective, anion-exchange column is

More information

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components APPLICATION UPDATE 157 Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components Authors Sumate Pengpumkiat, Weerapong Worawirunwong,

More information

Determination of trace anions in concentrated hydrofluoric acid

Determination of trace anions in concentrated hydrofluoric acid APPLICATION NOTE 78 Determination of trace anions in concentrated hydrofluoric acid Authors Archava Siriraks Thermo Fisher Scientific, Sunnyvale, CA Keywords HF, ICS-5000 +, IonPac AS10, IonPac AC10, ion

More information

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS anion-exchange column is designed for the fast analysis of inorganic

More information

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water Thermo Scientific Dionex Ion Chromatography Solutions Global water safety bromate analysis in drinking water Safe drinking water public health assurance As a vital limited resource required for survival,

More information

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device Determination of Bromate in Bottled Mineral Water Using the CRD 00 Carbonate Removal Device Thunyarat Phesatcha, Weerapong Worawirunwong, and Jeff Rohrer Thermo Fisher Scientific, Bangkok, Thailand; Thermo

More information

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column Chromatography Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS high-capacity, carbonate based anion-exchange column is designed for

More information

Thermo Scientific Dionex IonPac AS25 Anion-Exchange Column. Minutes

Thermo Scientific Dionex IonPac AS25 Anion-Exchange Column. Minutes columns Thermo Scientific Dionex IonPac AS Anion-Exchange Column Isocratic Separation of Sulfur Species and Inorganic Anions Using the Dionex IonPac AS Column. Fluoride. Bromate. Chloride. Nitrite. Bromide.

More information

Inline sample preparation for the determination of anions in sodium hydroxide

Inline sample preparation for the determination of anions in sodium hydroxide APPLICATION UPDATE 72331 Inline sample preparation for the determination of anions in sodium hydroxide Authors Hua Yang and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords Sample preparation,

More information

columns IonPac AS17-C Anion Exchange Column

columns IonPac AS17-C Anion Exchange Column columns IonPac AS-C Anion Exchange Column The IonPac AS-C is a hydroxide-selective anion exchange column designed for fast gradient separation of inorganic anions. The key application for the AS-C column

More information

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory Customer Application Note: 114 Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory Sébastien N. Ronkart, Ph.D.; Société wallonne des eaux, rue de la

More information

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS Daniel Kutscher, 1 Jörg Bettmer, 2 Torsten Lindemann, 1 Shona McSheehy-Ducos, 1 Lothar Rottmann 1 1 Thermo Fisher Scientific, Germany 2

More information

Expanded capabilities in ammonium and amine detection

Expanded capabilities in ammonium and amine detection PRODUCT SPECIFICATIONS Thermo Scientifi c Dionex SC-CERS 500 Salt Converter-Cation Electrolytically Regenerated Suppressor Expanded capabilities in ammonium and amine detection Benefits Broadened ammonium

More information

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000 Evaluation of Various Anion-Exchange Chemistries for the Analysis of Haloacetic Acids in Drinking Water Using -D Matrix Elimination Ion Chromatography and Suppressed Conductivity Detection White Paper

More information

columns IonPac AS18 Anion-Exchange Columns

columns IonPac AS18 Anion-Exchange Columns columns IonPac AS Anion-Exchange Columns The IonPac AS is a hydroxideselective anion-exchange column designed for the determination of inorganic anions and low-molecularweight organic acids including fluoride,

More information

Thermo Scientific Dionex IonPac AS11-HC-4µm Anion-Exchange Column

Thermo Scientific Dionex IonPac AS11-HC-4µm Anion-Exchange Column Chromatography Thermo Scientific Anion-Exchange Column Product Specifications Attain the optimal resolution of organic acids and inorganic anions using the new Thermo Scientific Dionex IonPac AS11-HC-4µm

More information

Thermo Scientific Dionex IonPac AS24A Anion-Exchange Column

Thermo Scientific Dionex IonPac AS24A Anion-Exchange Column CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS4A Anion-Exchange Column Product Specifications Thermo Scientific Dionex IonPac AS4A anion-exchange column is designed for the separation of haloacetic

More information

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA New Stationary Phases for Solid-Phase Extraction Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA Overview Purpose: To develop five new polymer-based stationary

More information

Determination of Silicate in High Purity Water Using Ion Chromatography and AutoPrep

Determination of Silicate in High Purity Water Using Ion Chromatography and AutoPrep pplication Note 70 Determination of Silicate in High Purity Water Using Ion Chromatography and utoprep Introduction The water used in the manufacture of semiconductors and other modern electronic components

More information

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis A. Caruso, M. Santoro, P. Magni, S. Pelagatti, and R. Facchetti Thermo Fisher Scientific, Milan,

More information

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency Anthony Edge, 1 Luisa Pereira, 1 Dafydd Milton 1 and Andrew Shalliker 2 1 Thermo Fisher Scientific,

More information

HILIC Method Development in a Few Simple Steps

HILIC Method Development in a Few Simple Steps HILIC Method Development in a Few Simple Steps Monica Dolci, Luisa Pereira, Dafydd Milton and Tony Edge Thermo Fisher Scientific, Runcorn, Cheshire, UK Overview This poster presents a systematic approach

More information

Determination of silicate in high-purity water using ion chromatography and online sample preparation

Determination of silicate in high-purity water using ion chromatography and online sample preparation PPLICTION NOTE 70 Determination of silicate in high-purity water using ion chromatography and online sample preparation uthors Weerapong Worawirunwong and Jeffrey Rohrer Thermo Fisher Scientific, Bangkok,

More information

Thermo Scientific Dionex IonPac AS26 Anion-Exchange Column

Thermo Scientific Dionex IonPac AS26 Anion-Exchange Column Chromatography Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS column is a high-capacity, hydroxide-selective, anion-exchange column

More information

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap Dirk Nolting and Andreas Wieghaus Thermo Fisher Scientific, Bremen, Germany Overview Purpose: Characterize the rate

More information

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System Application Update 55 Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free System Introduction Hydrogen peroxide is an essential chemical in the fabrication

More information

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry O. Scheibner, 1 P. van Baar, 2 F. Wode, 2 U. Dünnbier, 2 K. Akervik, 3 J. Humphries, 3 M. Bromirski 1

More information

Determination of Organic Acids in Beer Samples Using a High-Pressure Ion Chromatography System

Determination of Organic Acids in Beer Samples Using a High-Pressure Ion Chromatography System Determination of Organic Acids in Beer Samples Using a High-Pressure Ion Chromatography System Terri Christison, Charanjit Saini, and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA Technical

More information

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples Application Note 79 Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples INTRODUCTION Cow s milk is of particular dietary value to infants,

More information

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany Generic Method Approach for Pharmaceutical Drug Discovery and Development using Reversed-Phase Hydrophilic Interaction Liquid Chromatography with Universal Charged Aerosol Detection Frank Steiner, Michael

More information

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil Che Jinshui, 1 Deng Guifeng, 1 Liang Lina, 1 and Aaron Kettle, 2 1 Thermo Fisher Scientific (China)

More information

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Determination of Trace Cations in Power Plant Waters Containing Morpholine Application Note 8 Determination of Trace Cations in Power Plant Waters Containing Morpholine INTRODUCTION Morpholine and ammonium are used as additives in power plant waters. Morpholine acts as a corrosion

More information

Determination of urea in ultrapure water by IC-MS/MS

Determination of urea in ultrapure water by IC-MS/MS APPLICATION NOTE 72482 Determination of urea in ultrapure water by IC-MS/MS Authors Soon Fatt Lee, 1 Fiona Teh Hui Boon, 1 Chris Cheah Hun Teong, 1 and Jeff Rohrer 2 ¹Thermo Fisher Scientific, Singapore

More information

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method Xiang He and Marta Kozak ThermoFisher Scientific, San Jose, CA,

More information

Monitoring Protein PEGylation with Ion Exchange Chromatography

Monitoring Protein PEGylation with Ion Exchange Chromatography Monitoring Protein PEGylation with Ion Exchange Chromatography Peter Yu, Deanna Hurum, Jinyuan (Leo) Wang, 2 Terry Zhang, 2 and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA; 2 Thermo Fisher Scientific,

More information

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications Xiaodong Liu and Christopher Pohl; Thermo Fisher Scientific, Sunnyvale, CA Overview Review mixed-mode column technology

More information

columns IonPac SCS 1 Silica Cation Separator

columns IonPac SCS 1 Silica Cation Separator columns IonPac SCS Silica Cation Separator The SCS Silica Cation Separator is designed for use with nonsuppressed conductivity detection, or single-column ion chromatography (SCIC). This column is particularly

More information

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS Jason Lai, Jia Wang, Brad Hart, Pavel Aronov, Kristine Van Natta, Marta Kozak, Jorge Valdivia, Andy Jacobs, Haibo

More information

Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography

Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography Harpreet Dhillon and John Statler Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 11 Introduction

More information

columns IonPac AS22 Anion-Exchange Column

columns IonPac AS22 Anion-Exchange Column columns IonPac AS Anion-Exchange Column The IonPac AS is a carbonate based anion-exchange column designed for the determination of inorganic anions and low-molecular-weight organic acids including fluoride,

More information

Direct Determination of Small Organic Acids in Sea Water by IC-MS

Direct Determination of Small Organic Acids in Sea Water by IC-MS Direct Determination of Small Organic Acids in Sea Water by IC-MS Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA Application Note Key Words MSQ Plus Single Quadrupole Mass

More information

Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography

Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography Carl Fisher and Linda Lopez Thermo Fisher Scientific, Sunnyvale,

More information

Determination of Adsorbable Organic Halogen in Wastewater

Determination of Adsorbable Organic Halogen in Wastewater Determination of Adsorbable Organic Halogen in Wastewater Jingli Hu, Jeff Rohrer, and Kirk Chassaniol Thermo Fisher Scientific, Sunnyvale, CA The world leader in serving science Outline AOX (Adsorbable

More information

Determination of Inorganic Anions in Drinking Water by Ion Chromatography

Determination of Inorganic Anions in Drinking Water by Ion Chromatography Determination of Inorganic s in Drinking Water by Ion Chromatography Peter Jackson Thermo Fisher Scientific Inc. Application Note Introduction The determination of common inorganic anions in drinking water

More information

Overcoming Challenging Matrices in Ion Chromatography

Overcoming Challenging Matrices in Ion Chromatography Overcoming Challenging Matrices in Ion Chromatography Presented by: Kirk Chassaniol Gulf Coast Conference 2014 October 14, 2014 Thermo Fisher Scientific NA IC Tech Support 1 The world leader in serving

More information

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Andrea Caruso and Massimo Santoro, Thermo Fisher Scientific, Milan, Italy Application Note 1316 Key

More information

A Platform Method for Pharmaceutical Counterion Analysis by HPLC

A Platform Method for Pharmaceutical Counterion Analysis by HPLC A Platform Method for Pharmaceutical Counterion Analysis by HPLC Xiaodong Liu, Mark Tracy, and Christopher Pohl; Thermo Fisher Scientific, Sunnyvale, CA, USA Overview Results and Simple, fast, economical

More information

Determination of Methylamine in Drug Products

Determination of Methylamine in Drug Products Determination of Methylamine in Drug Products Thunyarat Phesatcha, Suparerk Tukkeeree, and Jeff Rohrer 2 Thermo Fisher Scientific, Bangkok, Thailand; 2 Thermo Fisher Scientific, Sunnyvale, CA, USA Application

More information

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer Olaf Scheibner and Maciej Bromirski Thermo Fisher Scientific, Bremen, Germany Overview Purpose: Improve the performance

More information

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites Robert Heintz, Mark Wall, Jennifer Ramirez, Stephan Woods Thermo Fisher Scientific, Madison WI Overview Purpose:

More information

Application Note 187. Brian DeBorba and Jeff Rohrer Thermo Scientific, Sunnyvale, CA, USA

Application Note 187. Brian DeBorba and Jeff Rohrer Thermo Scientific, Sunnyvale, CA, USA Determination of Sub-μg/L Bromate in Municipal and Natural Mineral Waters Using Preconcentration with Two-Dimensional Ion Chromatography and Suppressed Conductivity Detection Brian DeBorba and Jeff Rohrer

More information

anion-exchange column

anion-exchange column columnsionpac ASA Anion-Exchange Column The IonPac ASA is a carbonateselective anion-exchange column designed for the fast separation of inorganic anions, including fluoride, chlorite, bromate, chloride,

More information

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM IC Application Work AW CH6-1048-012011 Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM Branch: Chemical industry; Water, wastewater, environmental protection,

More information

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer Improved Screening for 2 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer Mary lackburn, Jia Wang, Jonathan eck, Charles Yang, Dipankar Ghosh, Thermo Fisher Scientific, San Jose, C,

More information

The determination of trace anions in concentrated phosphoric acid

The determination of trace anions in concentrated phosphoric acid TECHNICAL NOTE The determination of trace anions in concentrated phosphoric acid Authors Edward Kaiser and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords Ion chromatography, IC, IonPac

More information

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches Hydrophilic Interaction Liquid Chromatography: Method Development Approaches Monica Dolci, Luisa Pereira, and Tony Edge Thermo Fisher Scientific, Runcorn, Cheshire, UK Abstract This poster summarizes the

More information

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 000 Elemental Analyzer Dr. Liliana Krotz and Dr. Guido Giazzi Thermo Fisher Scientific, Milan, Italy

More information

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use Shih-Tse Jason Lai, Jia Wang, Brad Hart, Kristine Van Natta, Marta Kozak, Jorge Valdivia, Haibo Wang,

More information

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS Kate Comstock, Tim Stratton, Hongxia (Jessica) Wang, and Yingying Huang

More information

Application of Eluent Generation for Trace Anion Analysis of Borated Waters

Application of Eluent Generation for Trace Anion Analysis of Borated Waters Application of Eluent Generation for Trace Anion Analysis of Borated Waters Edward Kaiser and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 166 Introduction Boric acid is used

More information

Thermo Scientific Dionex eluent suppressors for ion chromatography

Thermo Scientific Dionex eluent suppressors for ion chromatography PRODUCT SPECIFICATIONS Thermo Scientific Dionex eluent suppressors for ion chromatography Suppression was introduced in 97, thereby bringing ion chromatography (IC) to the forefront of modern analytical

More information

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer Application ote: 54 Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap rbitrap Hybrid Mass Spectrometer Kate Comstock, Yingying Huang; Thermo Fisher Scientific,

More information

Thermo Scientific Dionex IonPac AS11 and AS11-HC Anion-Exchange Column

Thermo Scientific Dionex IonPac AS11 and AS11-HC Anion-Exchange Column chromatography Separate a wide variety of inorganic anions and organic acids using the Dionex IonPac AS and AS-HC Anion-Exchange Columns with potassium hydroxide gradients. Product Specifications Thermo

More information

Superior Chromatographic Performance Recommended anion-exchange column for separation of haloacetic acids prior to MS or MS/MS detection

Superior Chromatographic Performance Recommended anion-exchange column for separation of haloacetic acids prior to MS or MS/MS detection columns IonPac AS Anion-Exchange Column The IonPac AS is a highcapacity, hydroxide-selective anion exchange column designed for separation of haloacetic acids (HAAs) and bromate in drinking water prior

More information

of mass spectrometry

of mass spectrometry Thermo Scientific 253 Ultra High resolution isotope ratio MS Discover a new world of mass spectrometry Paleoclimatology Atmospheric science Biogeochemistry Petrology Discover the isotopic anatomy of molecules

More information

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography Application Note 75 Determination of Sulfate and Chloride in Ethanol by Ion Chromatography Ethanol used as a blending agent in gasoline can be contaminated with chloride and sulfate that form plugging

More information

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis Tim Stratton 1, Yingying Huang 1, Katianna Pihakari 1, Ian Acworth 2, and Michael Weber 2 1 Thermo Fisher Scientific,

More information

Concentrator and Trap Columns. Column: IonPac AG15, AS15, 2 mm. Eluent Source: EG40 Concentrator. Temperature: 30 C Column: IonPac AC15 (2 50 mm)

Concentrator and Trap Columns. Column: IonPac AG15, AS15, 2 mm. Eluent Source: EG40 Concentrator. Temperature: 30 C Column: IonPac AC15 (2 50 mm) columns and Trap Columns IonPac AG, AS, mm Eluent Source: EG Temperature: C IonPac AC ( mm) Flow Rate:. m/min Eluent: Potassium hydroxide, Inj. Volume: m, preconcentrated mm from to min, mm mm from min

More information

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer Defining quality standards for the analysis of solid samples Redefine your quality standards for the elemental analysis of solid samples

More information

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth Thermo Fisher Scientific, Chelmsford,

More information

Determination of Aniline and Nitroanilines in Environmental and Drinking Waters by On-Line SPE

Determination of Aniline and Nitroanilines in Environmental and Drinking Waters by On-Line SPE Determination of Aniline and Nitroanilines in Environmental and Drinking Waters by On-Line SPE Li Lang, Xu Qun, and Jeffrey Rohrer Thermo Fisher Scientific, Shanghai, People s Republic of China; Thermo

More information

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath APPLICATIN NTE 9 Determination of an anionic fluorochemical surfactant in a semiconductor etch bath Authors Mark Laikhtman and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA Keywords Ion chromatography,

More information

Determination of trace concentrations of oxyhalides and bromide in municipal and bottled waters

Determination of trace concentrations of oxyhalides and bromide in municipal and bottled waters PPLICTION NOTE 7229 Determination of trace concentrations of oxyhalides and bromide in municipal and bottled waters uthors Jingli Hu and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, C, US Keywords

More information

Determination of trace anions in basic solutions by single pass AutoNeutralization and ion chromatography

Determination of trace anions in basic solutions by single pass AutoNeutralization and ion chromatography APPLIATION NOTE 78 Determination of trace anions in basic solutions by single pass AutoNeutralization and ion chromatography Authors Soon Fatt Lee, Aaron Rose, Terri hristison, and Jeff Rohrer Thermo Fisher

More information

Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography

Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography Kirk Chassaniol Manager of IC Technical Sales Support Thermo Fisher Scientific 1 The world leader in serving

More information

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry APPLICATION NOTE Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry AN43227 Antonella Guzzonato 1, Shona McSheehy Ducos

More information

Determination of Monochloroacetic Acid in Carbocisteine

Determination of Monochloroacetic Acid in Carbocisteine Determination of Monochloroacetic Acid in Carbocisteine Zhong Xinlin, Ye Mingli, Liang Lina, Xu Qun, and Jeffrey Rohrer Thermo Fisher Scientific, Shanghai, People s Republic of China; Thermo Fisher Scientific,

More information

Determination of trace anions in organic solvents using matrix elimination and preconcentration

Determination of trace anions in organic solvents using matrix elimination and preconcentration APPLICATION UPDATE Determination of trace anions in organic solvents using matrix elimination and preconcentration Authors Terri Christison and Jeff Rohrer Thermo Fisher Scientific Sunnyvale, CA Keywords

More information

Determination of chlorine, bromine, and sulfur in polyethylene materials using combustion IC with a carbonate/bicarbonate eluent

Determination of chlorine, bromine, and sulfur in polyethylene materials using combustion IC with a carbonate/bicarbonate eluent APPLICATION UPDATE 72588 Determination of chlorine, bromine, and sulfur in polyethylene materials using combustion IC with a carbonate/bicarbonate eluent Authors Manali Aggrawal and Jeffrey Rohrer Thermo

More information

The Determination of Trace Anions in Concentrated Phosphoric Acid

The Determination of Trace Anions in Concentrated Phosphoric Acid Technical Note The Determination of Trace Anions in Concentrated Phosphoric Acid INTRODUCTION The determination of trace anions in phosphoric acid is hampered by a large excess of phosphate ion. Chloride

More information

Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography

Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography Application Note 9 Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography Introduction The computer, semiconductor, and food industries need analytical

More information

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput Yi Zhang 1, Zhiqi Hao 1, Markus Kellmann 2 and Andreas FR. Huhmer

More information

Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography

Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography Marc Plante, Bruce Bailey, and Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA Overview

More information

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer Reiko Kiyonami, Mary Blackburn, Andreas FR Hühme: Thermo Fisher

More information

Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance

Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance David Thomas, Bruce Bailey, Marc Plante and Ian Acworth Thermo Fisher Scientific,

More information

Determination of trace anions in concentrated glycolic acid

Determination of trace anions in concentrated glycolic acid TECHNICAL NOTE Determination of trace anions in concentrated glycolic acid Authors Edward Kaiser and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA Keywords Ion chromatography, IC, Dionex ICS- +,

More information

( )( ) Selectivity Choices in Reversed-Phase Fast LC. Introduction. R s = 1 a 1 k 4 a 1 + k

( )( ) Selectivity Choices in Reversed-Phase Fast LC. Introduction. R s = 1 a 1 k 4 a 1 + k Selectivity Choices in Reversed-Phase Fast LC Luisa Pereira, Monica Dolci, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20543 Key Words Accucore, Hypersil GOLD, Syncronis, Column Chemistry,

More information

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry Olaf Scheibner, 1 Patrizia van Baar, 2 Florian Wode, 2 Uwe Dünnbier, 2 Kristi Akervik, 3 Jamie Humphrie,

More information

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity Monica Dolci, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20544 Key Words Hydrophilic, HILIC,

More information

The power to increase productivity. Thermo Scientific Dionex Reagent-Free Ion Chromatography (RFIC) System Capabilities

The power to increase productivity. Thermo Scientific Dionex Reagent-Free Ion Chromatography (RFIC) System Capabilities The power to increase productivity Thermo Scientific Dionex Reagent-Free Ion Chromatography (RFIC) System Capabilities RFIC System Technology Since the introduction of eluent generation in 99, Thermo Fisher

More information