Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers

Size: px
Start display at page:

Download "Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers"

Transcription

1 Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N O T E Summary The far UV region (generally covering the range from 190 nm to 150 nm) has great potential experimental interest. All compounds have electronic transitions in this region, including the first allowed benzenoid transition, the amide absorption bands, the first transition of the olefinic double bond, and the n-σ* absorption bands of water, alcohols, ammonia, and sulfur compounds. 1 In particular, spectra of steroids, nucleic acids, proteins, and polypeptides can be found in this region. The purpose of this application note is to investigate the far UV performance of the new PerkinElmer LAMBDA 850 and 950 UV/Vis and UV/Vis/NIR spectrophotometers. Both of these instruments are high research grade systems, incorporating double monochromators, and offer ultra high optical resolution to < 0.05 nm and a wide dynamic range to 8 A. Both units include complete far UV capability and allow scanning to 175 nm. The use of nitrogen purging, selection of appropriate solvents and cuvettes, selection of instrumental parameters, and validating experimental results will be discussed. Examples of far UV application data will also be presented. Introduction Usually, UV/Vis spectrophotometers with conventional optics using a monochromator and photomultipliers as detectors have a working range from 190 nm to 900 nm. The wavelength range below 190 nm down to 150 nm is part of the far ultraviolet range and is usually only accessible with dedicated instruments prepared for transmission measurements in a vacuum. Most of the compounds analyzed by UV/Vis spectroscopy have an absorption maximum above 200 nm. Nevertheless, there are a number of applications where there is either only a low absorbance above 200 nm, as in the analysis of sugars, or the absorbance increase below 200 nm is used for the determination of impurities, e.g. in highly pure isooctane or alkylpyridines.

2 Instrument requirements for far UV measurements The far UV range is made available in the PerkinElmer LAMBDA 850 and 950 spectrophotometers through incorporation of hardware not typically present in a conventional spectrophotometer. In particular, the LAMBDA 850 and 950 include as standard: Nitrogen purge ready, sealed optical system Special water-free Suprasil W for all windows High efficiency purge fittings with multiple lines routed to the monochromators and the source, sample, and detector compartments An exclusive far UV photomultiplier (R 6872) with fused silica optical window The latest generation high energy deuterium sources incorporating fused silica optical windows Fused silica photomultiplier and deuterium source Photomultipliers usually incorporate a quartz window for surface protection, which contributes to energy decrease below 200 nm. A photomultiplier with a fused silica window has a noise level reduced by a factor of 3 to 5 below 200 nm, when compared to a normal photomultiplier with a quartz window. Similarly, deuterium lamps with fused silica windows give two times higher radiation density at 190 nm and about five times higher radiation density at 180 nm when compared to a standard quartz window deuterium lamp. Nitrogen purging The constituents of air start to absorb below 200 nm. The UV cutoff of ozone starts at 295 nm, of oxygen at 200 nm and of nitrogen at 170 nm. 2 Thus, for working below 200 nm, it is necessary to remove the oxygencontaining compounds from the light path to increase energy and decrease noise level. The effect of nitrogen purging on the energy level between 200 nm and 180 nm is shown in Figure 1. The bottom energy curve is without purge, the top curve is with purge. Because of the relatively high diffusion coefficients of gases, it is necessary to have a permanent flush of nitrogen. The flow velocity is dependent on the size and the sealing of the optical system. An initial nitrogen flow rate of 20 l/min is recommended for 20 to 30 minutes, and then the flow rate can be dropped to about 6 to 7 l/min. Only nitrogen gas of the highest purity should be used. Instruments specifically designed to accommodate far UV measurements, such as the PerkinElmer LAMBDA 850 and 950 UV/Vis and UV/Vis/NIR spectrophotometers, have sealed optical compartments. Radiation from the source compartment enters the monochromators through a water-free fused silica window and the sample compartment area is sealed with water-free fused silica windows. Thus, the entire optical path from the source lamp to the detector can be purged largely free of water vapor and oxygen. To facilitate effective purging, the PerkinElmer LAMBDA series spectrophotometers incorporate a high efficiency purge system, with multiple purge lines routed to the monochromators, the source, the sample, and the detector compartments. Stray radiation Usually, stray radiation in an optical system increases toward shorter wavelengths because of higher scattering. The LAMBDA 850 and 950 spectrophotometers are high-performance, double monochromator systems, with a specified stray radiation level of less than % transmission Figure 1. Energy spectrum with and without nitrogen purging. Data collected at 10 nm/min with a 0.1 nm slit. 2

3 at 220 nm measured according to ASTM recommendation E 387. Potassium chloride solution can be used to determine stray radiation levels below 200 nm. Stray light values below 200 nm can typically increase to about 0.01% transmission. Experiments designed for the far UV should take into consideration the lower dynamic range due to the higher stray radiation present. It is generally recommended to keep solutions below 2 A when measuring in the far UV. Choice of solvent and cuvette Many solvents absorb greatly in the range of 200 and 175 nm. Most solvents are completely opaque in the far UV range in a standard 10 mm pathlength cuvette. For example, distilled water measured in a 10 mm pathlength cuvette will totally cut off at 185 nm. Therefore, narrow pathlength cuvettes are required or thin film demountable liquid cells with fused silica or CaF 2 windows. A maximum pathlength of 1 mm is recommended with the solvents that are transparent in the far UV. Common far solvents with the wavelength of their absolute cut-offs include water (170 nm), deuterium oxide (166 nm), methyl alcohol (182 nm), trifluoroethanol (165 nm), hexafluoro-2-propanol (164 nm), acetonitrile (173 nm) and pentane (167 nm). 1 A general recommendation before conducting an experiment in the far UV is to fill a 1 mm silica cuvette with the solvent of choice, and scan this against a purged air baseline. The solvent absorption observed will be a consideration in the dynamic range allowed for the analyte. If the solvent exceeds 2 A, it should probably not be used, or a narrower pathlength cell needs to be used. Choice of bandpass The quality of spectra obtained in the far UV range can be optimized by using broader bandpasses. In particular, energy is extremely low from 180 to 175 nm. The absorbance noise data presented in Table 1 shows the effect of slit width. The bandpass chosen for any particular experiment will be a compromise between the resolution required and maximizing the signal-to-noise. The general recommendation is to select the widest possible bandpass to achieve the resolution required for the compound being studied. Typical far UV corrected baselines are presented in Figure 2 using a 4 nm bandpass with nitrogen purging. Effectively purged baselines will normally be within +/ A for a 2 nm bandpass and +/ A for a 4 nm bandpass. The far UV experiment checklist Purge the instrument with nitrogen at approximately 20 l/min for at least 20 minutes prior to use. After 20 minutes, the nitrogen flow can be reduced to about 6-7 l/min. Use 1 mm pathlength silica cuvettes to start. Create a method with a 4 nm bandpass, a 1 nm data interval, and a 1.44 second integration time (40 nm/min scan speed). Select a suitable far UV solvent for the analyte. If possible, use pure distilled water. Against an air background, determine the absorbance of the solvent in a 1 mm cell. If it is over 2 A near the peak for the analyte, select a different solvent or use a demountable liquid cell with a shorter pathlength. The sample compartment will flood with oxygen when the compartment lid is opened. The Table 1. Absorbance RMS noise level at a 2 nm and a 4 nm slit as a function of wavelength. Wavelength (nm) Slit Figure 2. Expanded scale overlay of 5 corrected baselines versus air with nitrogen purge. Conditions were 40 nm/min scan speed, 1.44 second integration, and 4 nm slit (LAMBDA 950). 3

4 cover should be opened only for the shortest time possible. The sample compartment should be purged for about 5 to 10 minutes after inserting a sample. Optionally, to reduce the time between samples, an additional purge line can be routed to the sample compartment to accelerate the oxygen purge. If possible, keep the absorbance of the sample solution at 2 A or less as this will minimize any stray light effects. Validating experimental results A real problem for a researcher studying the far UV is the validation of results. Because of the difficulty in obtaining true spectra in the far UV, with demanding instrumental and technique requirements, there is a suspicion that many of the deep UV publications contain spectra not totally free of instrumental artifacts, and often the spectra are difficult to reproduce. Most importantly, insufficient or inefficient nitrogen purging creates high oxygen backgrounds, greatly reducing the dynamic range in the far UV, and causing spectra to be affected by stray light. Spectra affected by stray light are invalid, often yielding irreproducible spectra with shifted peak maximums. Therefore, validating experimental results in the far UV is critical. The following procedures are recommended for validating experimental results: Run replicates on the same sample to see if the peaks overlay consistently. If time permits, repeat the experiment with a new corrected baseline, running the same sample. Run the sample at different times during the purge process. The spectra should overlay with acceptable precision. Acquire fresh corrected baselines before each sample measurement for the best precision. The oxygen backgrounds can fluctuate slightly with time. Prepare and run different concentrations of the sample, and keep the absorbance at the peak at < 2 A. The spectra and peak maximums should be the same for the different concentration. To reduce the possibility that a stray light induced peak is being observed, validate the experimental peaks by scanning a number of neutral density screens bracketing the absorbance range of the results (Figure 3). The screens should give a flat spectral response to 175 nm against a fresh corrected purged baseline. If peaks are observed in the corrected baseline, or in the neutral density screen spectra, then most likely the purging was not adequate, and oxygen is still present. Applications Among the possible applications in the far UV range with conventional UV/Vis spectrophotometers are the study of gases with large absorption coefficients in this range, 3 and the high sensitivity of organic compounds in solution such as sugars, substituted benzenes and pyridines. A spectrum of a 10 ppm solution of benzene in water is shown in Figure 4. Method parameters were a 40 nm/min scan speed, a 1.44 second integration time, and a 4 nm slit. Figure 5 shows two overlaid spectra of deoxyribonucleic (calf thymus, Sigma D 1501) acquired at different concentrations. The nearly identical spectra help validate the experiment by assuring stray light artifacts are not a factor. The transmission characteristics and purity of deep UV transmitting materials can also be studied. Figure 6 shows two overlaid spectra of different calcium fluoride lens materials. Figure 3. Overlay spectra of four neutral density screens scanned into the deep UV with nitrogen purging. A flat spectral response can help validate the experimental results by assuring stray light artifacts are not present. Figure 4. Spectrum of 10 ppm benzene in water scanned in a 1 mm cell. Conditions were 40 nm/min scan speed, 1.44 second integration, and 4 nm slit. 4

5 Conclusion High quality far ultraviolet data can be obtained with a correctly configured UV/Vis spectrophotometer, using appropriate experimental materials and procedures. The PerkinElmer LAMBDA 850 and 950 UV/Vis and UV/Vis/NIR double monochromator spectrophotometers are ideal systems for conducting far UV experiments. These research grade units include the hardware required for these measurements, as standard. This includes sealed optical systems, high efficiency multiple purge lines, water-free Suprasil windows, and a special extended range photomultiplier. Also, the ultra-low stray light characteristics of the LAMBDA 850 and 950 allows publication quality spectra, free of artifacts, to be collected to 175 nm. References 1 Malcolm F. Fox, Applied Spectroscopy, Volume 27, Number 3, 1973, pg W. Boehme, G. Naundorf, Applied UV Spectroscopy 13E, Bodenseewerk Perkin-Elmer + Co. GmbH, Uberlingen, West Germany B.A. Thompson et al., Journal of Geophysical Research, Vol. 68, No. 24, December Figure 5. Overlaid spectra of DNA in distilled water (calf thymus sodium salt) scanned in the deep UV in a 1 mm pathlength cuvette at approximate concentrations of 15 (lower) and 35 (upper) µg/ml. Figure 6. Overlaid spectra of two calcium fluoride crystals scanned into the deep UV with nitrogen purge. The differences observed are caused by impurities. PerkinElmer Life and Analytical Sciences 710 Bridgeport Avenue Shelton, CT USA Phone: (800) or (+1) For a complete listing of our global offices, visit /lasoffices 2004 PerkinElmer, Inc. All rights reserved. The PerkinElmer logo and design are registered trademarks of PerkinElmer, Inc. LAMBDA is a trademark and PerkinElmer is a registered trademark of PerkinElmer, Inc. or its subsidiaries, in the United States and other countries. Suprasil is a registered trademark of Heraeus Holding GmbH. All other trademarks not owned by PerkinElmer, Inc. or its subsidiaries that are depicted herein are the property of their respective owners. PerkinElmer reserves the right to change this document at any time without notice and disclaims liability for editorial, pictorial or typographical errors _05 Printed in USA

Traceable UV-Vis Reference Materials

Traceable UV-Vis Reference Materials UV-Vis Spectroscopy Traceable UV-Vis Reference Materials For Regulatory Compliance Introduction Many laboratories, particularly in the pharmaceutical industry, are under increasing pressure to show proof

More information

Design Considerations for a Variable Angle Absolute Reflectance Accessory For the LAMBDA 950/850/650 UV/Vis/NIR and UV/Vis Spectrophotometers

Design Considerations for a Variable Angle Absolute Reflectance Accessory For the LAMBDA 950/850/650 UV/Vis/NIR and UV/Vis Spectrophotometers Design Considerations for a Variable Angle Absolute Reflectance Accessory For the LAMBDA 950/850/650 UV/Vis/NIR and UV/Vis Spectrophotometers UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N

More information

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS FIELD APPLICATION REPORT ICP MASS SPECTROMETRY Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS Authors: Kenneth R. Neubauer, Ph.D. Perkin Elmer Life and Analytical Sciences 710

More information

Instrumental Limitations for High Absorbance Measurements in Noise

Instrumental Limitations for High Absorbance Measurements in Noise TECHNICAL NOTE Instrumental Limitations for High Absorbance Measurements in Noise UV/Vis/NIR Spectroscopy Author: Jeffrey L. Taylor, Ph.D. PerkinElmer, Inc. Shelton, CT Spectrophotometer Energy Using a

More information

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation Introduction Fire investigation involves many different types of analyses from crime scene

More information

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry 13A Measurement Of Transmittance and Absorbance Absorption measurements based upon ultraviolet and visible radiation

More information

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography Introduction ASTM International publishes a number of different methods for the determination of volatile compounds in water-based coatings.

More information

Evaluating Labsphere s new UV-2000

Evaluating Labsphere s new UV-2000 Ciba Grenzach GmbH Optical Spectroscopy Sabrina Siegle Dr. Wolfgang Schlenker Tel. +49 76 24 / 12-28 13 E-mail wolfgang.schlenker@ciba.com May 26, 9 Evaluating Labsphere s new UV- Abstract Labsphere s

More information

FLUORESCENCE APPLICATIONS

FLUORESCENCE APPLICATIONS FLUORESCENCE APPLICATIONS CORRECTION OF EMISSION SPECTRA USING THE PERKINELMER MODEL LS-50 LUMINESCENCE SPECTROMETER Spectra collected on luminescence spectrometers consist of true spectral information

More information

Nucleic Acid Quantitation in Microplates

Nucleic Acid Quantitation in Microplates PHARMACEUTICAL application note Nucleic Acid Quantitation in Microplates HansWilly Mueller, Ph.D. J. Fenton Williams, Ph.D. Introduction Many molecular biology applications require the determination of

More information

C101-E107B UV Shimadzu UV-VIS-NIR Spectrophotometer

C101-E107B UV Shimadzu UV-VIS-NIR Spectrophotometer C101-E107B UV-3600 Shimadzu UV-VIS-NIR Spectrophotometer High sensitivity, high resolution, and an ultra-low stray-light level achieved with the latest technology lead the way to new solutions. UV-3600

More information

Simultaneous Arsenic and Chromium Speciation by HPLC/ICP-MS in Environmental Waters

Simultaneous Arsenic and Chromium Speciation by HPLC/ICP-MS in Environmental Waters Simultaneous Arsenic and Chromium Speciation by HPLC/ICP-MS in Environmental Waters Introduction Many elements can exist in a variety of oxidation states which have differing impacts on health and the

More information

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Analytical Technologies in Biotechnology Prof. Dr. Ashwani K Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Module - 6 Spectroscopic Techniques Lecture - 2 UV-Visible Spectroscopy

More information

Compact Knowledge: Absorbance Spectrophotometry. Flexible. Reliable. Personal.

Compact Knowledge: Absorbance Spectrophotometry. Flexible. Reliable. Personal. L A B O R A T O R Y C O M P E T E N C E Compact Knowledge: Absorbance Spectrophotometry Flexible. Reliable. Personal. The interaction of light with molecules is an essential and well accepted technique

More information

3 - Atomic Absorption Spectroscopy

3 - Atomic Absorption Spectroscopy 3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,

More information

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at 2001 Spectrometers Instrument Machinery Movies from this presentation can be access at http://www.shsu.edu/~chm_tgc/sounds/sound.html Chp20: 1 Optical Instruments Instrument Components Components of various

More information

Spectroscopy. Page 1 of 8 L.Pillay (2012)

Spectroscopy. Page 1 of 8 L.Pillay (2012) Spectroscopy Electromagnetic radiation is widely used in analytical chemistry. The identification and quantification of samples using electromagnetic radiation (light) is called spectroscopy. Light has

More information

Chem 310 rd. 3 Homework Set Answers

Chem 310 rd. 3 Homework Set Answers -1- Chem 310 rd 3 Homework Set Answers 1. A double line labeled S 0 represents the _ground electronic_ state and the _ground vibrational_ state of a molecule in an excitation state diagram. Light absorption

More information

UV-Vis optical fiber assisted spectroscopy in thin films and solutions

UV-Vis optical fiber assisted spectroscopy in thin films and solutions UV-Vis optical fiber assisted spectroscopy in thin films and solutions Description UV-Visible absorption and transmission spectra provide fundamental information for all experiments related to the attenuation

More information

25 Instruments for Optical Spectrometry

25 Instruments for Optical Spectrometry 25 Instruments for Optical Spectrometry 25A INSTRUMENT COMPONENTS (1) source of radiant energy (2) wavelength selector (3) sample container (4) detector (5) signal processor and readout (a) (b) (c) Fig.

More information

Eppendorf UVette. Instructions for use

Eppendorf UVette. Instructions for use Instructions for use Instructions Eppendorf see Fig. Tab. p. p. UVette for use Copyright 2014 Eppendorf AG, Hamburg, Germany. All rights reserved, including graphics and images. No part of this publication

More information

Integrating Spheres in Molecular Spectrophotometry

Integrating Spheres in Molecular Spectrophotometry Integrating Spheres in Molecular Spectrophotometry Theory and Practice 2012 Perkin Elmer 2012 Perkin Elmer General Sphere Theory 3 Integrating Spheres Types of Sphere Measurements Total Reflectance (Specular

More information

Whole Tablet Measurements Using the Spectrum One NTS Tablet Autosampler System

Whole Tablet Measurements Using the Spectrum One NTS Tablet Autosampler System Whole Tablet Measurements Using the Spectrum One NTS Tablet Autosampler System A P P L I C A T I O N N O T E Introduction Recent advances in NIR technology have changed the ways in which both the pharmaceutical

More information

9/28/10. Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Valence Electronic Structure. n σ* transitions

9/28/10. Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Valence Electronic Structure. n σ* transitions Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Electromagnetic Spectrum - Molecular transitions Widely used in chemistry. Perhaps the most widely used in Biological Chemistry.

More information

NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation.

NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation. NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation. SCANNING UV VISIBLE SPECTROPHOTOMETER Features SCANNING UV VISIBLE SPECTROPHOTOMETER

More information

Fluorescence Spectrophotometry

Fluorescence Spectrophotometry Chemistry 422L Manual Page 27 I. Introduction Fluorescence Spectrophotometry Ru(bpy) 3 2+, where bpy = 2, 2' bipyridine, has been one of the most widely studied metal complexes in recent years. Interest

More information

INTRODUCTION The fundamental law of spectrophotometry is known as the Beer-Lambert Law or Beer s Law. It may be stated as: log(po/p) = A

INTRODUCTION The fundamental law of spectrophotometry is known as the Beer-Lambert Law or Beer s Law. It may be stated as: log(po/p) = A S2. INTRODUCTION TO ULTRA-VIOLET / VISIBLE SPECTROSCOPY AIM 1. To become familiar with the operation of a conventional scanning ultra-violet spectrophotometer 2. To determine suitable cells and solvents

More information

CONFOCHECK. Innovation with Integrity. Infrared Protein Analysis FT-IR

CONFOCHECK. Innovation with Integrity. Infrared Protein Analysis FT-IR CONFOCHECK Infrared Protein Analysis Innovation with Integrity FT-IR CONFOCHECK: FT-IR System for Protein Analytics FT-IR Protein Analysis Infrared spectroscopy measures molecular vibrations due to the

More information

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology APPLICATION NOTE Gas Chromatography Author: David Scott PerkinElmer, Inc. Shelton, CT Residual Solvents in Pharmaceuticals by USP Chapter Methodology Introduction The synthesis of active pharmaceutical

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS for the Perkin Elmer Model 650-10S Fluorescence Spectrophotometer Re: Operating Instructions PE 650 2 The following flow diagram should help to guide you through the steps necessary

More information

Degree of labeling (DOL)

Degree of labeling (DOL) Degree of labeling (DOL) A written explanation for the determination of the degree of labeling when using NuLink reagents. The DOL is the average number of reagent molecules that have been covalently attached

More information

Investigating Transition Metal Complexes

Investigating Transition Metal Complexes Exercise 4 Investigating Transition Metal Complexes 4 Introduction Colour is a well known property of the transition metals. The colour produced as parts of the visible spectrum are due to electron transitions

More information

Instrumental Technique: Cuvette. Md Rabiul Islam

Instrumental Technique: Cuvette. Md Rabiul Islam Instrumental Technique: Cuvette Md Rabiul Islam 16-7-2016 What is cuvette? A cuvette is a small tube of circular or square cross section, sealed at one end, made of plastic, glass, or fused quartz (for

More information

Optical density measurements automatically corrected to a 1-cm pathlength with PathCheck Technology

Optical density measurements automatically corrected to a 1-cm pathlength with PathCheck Technology APPLICATION NOTE Optical density measurements automatically corrected to a 1-cm pathlength with PathCheck Technology Introduction UV/VIS spectrophotometers and microplate readers differ fundamentally in

More information

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters ) September 17, 2018 Reference literature (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters 13-14 ) Reference.: https://slideplayer.com/slide/8354408/ Spectroscopy Usual Wavelength Type of Quantum

More information

INTERNATIONAL OLIVE COUNCIL

INTERNATIONAL OLIVE COUNCIL INTERNATIONAL OLIVE COUNCIL COI/T.20/Doc. No 19/Rev. 4 2017 ENGLISH Original: ENGLISH Príncipe de Vergara, 154 28002 Madrid España Telef.: +34 915 903 638 Fax: +34 915 631 263 - e-mail: iooc@internationaloliveoil.org

More information

Chirascan 6-Cell Peltier Cell Holder: Rapid Optimisation of Buffer Conditions for Stabilising Protein Therapeutics

Chirascan 6-Cell Peltier Cell Holder: Rapid Optimisation of Buffer Conditions for Stabilising Protein Therapeutics CHIRASCAN SERIES APPLICATION NOTE Chirascan 6-Cell Peltier Cell Holder: Rapid Optimisation of Buffer Conditions for Stabilising Protein Therapeutics Abstract: The selection of buffer conditions that maximise

More information

MOLEBIO LAB #4: Using a Spectrophotometer

MOLEBIO LAB #4: Using a Spectrophotometer Introduction: Spectrophotometry MOLEBIO LAB #4: Using a Spectrophotometer Many kinds of molecules interact with or absorb specific types of radiant energy in a predictable fashion. For example, when while

More information

S2. INTRODUCTION TO ULTRA-VIOLET / VISIBLE SPECTROSCOPY

S2. INTRODUCTION TO ULTRA-VIOLET / VISIBLE SPECTROSCOPY S2. INTRODUCTION TO ULTRA-VIOLET / VISIBLE SPECTROSCOPY PURPOSE 1. To become familiar with the operation of a conventional scanning ultra-violet spectrophotometer 2. To determine suitable cells and solvents

More information

Microplate Spectrophotometer

Microplate Spectrophotometer Microplate Spectrophotometer Features Benefits Applications Bali, October 1999 Optical System Xenon Flashing Lamp Order Sorting Filters Monochromator Data Light collection Light Transfer Why are these

More information

Microdetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats

Microdetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats Application Note 24 Microdetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats INTRODUCTION Most spectrophotometric

More information

Background Information:

Background Information: Measurement of Light in Radiation Damaged Quartz Bars for Q weak Katie Kinsley Advisors: Dave Mack, Julie Roche Thomas Jefferson National Accelerator Facility Newport News, VA Summer 2007 (final edit Sept

More information

Outline of Recombinant DNA technology. Application of UV spectroscopy in recombinant DNA technology

Outline of Recombinant DNA technology. Application of UV spectroscopy in recombinant DNA technology NIKHIL.K.POTDUKHE Outline of UV spectrophotometer Outline of Recombinant DNA technology Application of UV spectroscopy in recombinant DNA technology References Lambert law: When a beam of light is allowed

More information

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency. Light We can use different terms to describe light: Color Wavelength Frequency Light is composed of electromagnetic waves that travel through some medium. The properties of the medium determine how light

More information

RamanStation 400: a Versatile Platform for SERS Analysis

RamanStation 400: a Versatile Platform for SERS Analysis FIELD APPLICATION REPORT Raman Spectroscopy Author: Dean H. Brown PerkinElmer, Inc. Shelton, CT USA RamanStation 400 RamanStation 400: a Versatile Platform for SERS Analysis Introduction Surface Enhanced

More information

IMPCA METHANOL REFERENCE SPECIFICATIONS

IMPCA METHANOL REFERENCE SPECIFICATIONS IMPCA INTERNATIONAL METHANOL PRODUCERS & CONSUMERS ASSOCIATION IMPCA METHANOL REFERENCE SPECIFICATIONS a.i.s.b.l. International Methanol Producers & Consumers Association i.v.z.w. Avenue de Tervueren 270

More information

Trace Analyses in Metal Matrices Using the ELAN DRC II

Trace Analyses in Metal Matrices Using the ELAN DRC II www.perkinelmer.com Trace Analyses in Metal Matrices Using the ELAN DRC II Introduction Analyses of matrices containing high levels of metals present a challenge for ICP-MS. First, the concentrations of

More information

Spectroscopy Primer. for ultraviolet and visible absorbance spectroscopy. by Stephanie Myers Summer 2015

Spectroscopy Primer. for ultraviolet and visible absorbance spectroscopy. by Stephanie Myers Summer 2015 Spectroscopy Primer for ultraviolet and visible absorbance spectroscopy by Stephanie Myers Summer 2015 Abstract: An overview of uv vis absorbance spectroscopy including Beer s Law, calibration curves,

More information

SPECTROPHOTOMETERS. Visible and UV-Visible

SPECTROPHOTOMETERS. Visible and UV-Visible Visible and UV-Visible The Manifacturer ONDA Spectrophotometers are manifactured by a company with over than ten years of experience in development of UV/Vis instruments, Single beam and Double beam, with

More information

FLAME PHOTOMETRY AIM INTRODUCTION

FLAME PHOTOMETRY AIM INTRODUCTION FLAME PHOTOMETRY AIM INTRODUCTION Atomic spectroscopy is based on the absorption, emission or fluorescence process of light by atoms or elementary ions. Information for atomic scale is obtained in two

More information

CH0302 Process Instrumentation. Lecture 9 Composition Analysis

CH0302 Process Instrumentation. Lecture 9 Composition Analysis CH0302 Process Instrumentation Lecture 9 Composition Analysis Department of Chemical Engineering School of Bioengineering SRM University Kattankulathur 603203 4/3/15 Chemical Engineering 1 Industrial significance

More information

09/05/40 MOLECULAR ABSORPTION METHODS

09/05/40 MOLECULAR ABSORPTION METHODS MOLECULAR ABSORPTION METHODS Absorption spectroscopy refers to spectroscopic techniques that measure the absorption of radiation, as a function of wavelength ( absorption spectrum ), due to its interaction

More information

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy A. Evaluation of Analytical Parameters in Atomic Absorption Spectroscopy Objective The single feature that contributes most to making atomic

More information

Experimental Design and Data Collection

Experimental Design and Data Collection A. Checks to run to make sure the instrument is in good working condition B. Decide which experiment type is most appropriate, equilibrium or velocity C. Buffer considerations D. Speed selection and length

More information

Introduction to Electromagnetic Radiation and Radiative Transfer

Introduction to Electromagnetic Radiation and Radiative Transfer Introduction to Electromagnetic Radiation and Radiative Transfer Temperature Dice Results Visible light, infrared (IR), ultraviolet (UV), X-rays, γ-rays, microwaves, and radio are all forms of electromagnetic

More information

Understanding Your Spectra Module. Agilent OpenLAB CDS ChemStation Edition

Understanding Your Spectra Module. Agilent OpenLAB CDS ChemStation Edition Understanding Your Spectra Module Agilent OpenLAB CDS ChemStation Edition Notices Agilent Technologies, Inc. 1994-2012, 2013 No part of this manual may be reproduced in any form or by any means (including

More information

MASTERING THE VCE 2014 UNIT 3 CHEMISTRY STUDENT SOLUTIONS

MASTERING THE VCE 2014 UNIT 3 CHEMISTRY STUDENT SOLUTIONS MASTERING THE VCE 2014 UNIT 3 CHEMISTRY STUDENT SOLUTIONS FOR ERRORS AND UPDATES, PLEASE VISIT WWW.TSFX.COM.AU/VCE-UPDATES QUESTION 45 QUESTION 46 Answer is A QUESTION 47 The number of protons in the element.

More information

Beer's Law and Data Analysis *

Beer's Law and Data Analysis * OpenStax-CNX module: m15131 1 Beer's Law and Data Analysis * Mary McHale This work is produced by OpenStax-CNX and licensed under the Creative Commons Attribution License 2.0 1 Beer's Law and Data Analysis

More information

METHANOL REFERENCE SPECIFICATIONS

METHANOL REFERENCE SPECIFICATIONS IMPCA INTERNATIONAL METHANOL PRODUCERS & CONSUMERS ASSOCIATION Avenue de Tervuren 149 1150 Brussels, Belgium Tel : (32/2) 7418683 Fax : (32/2) 7418684 e-mail : impca@epca.be IMPCA METHANOL REFERENCE SPECIFICATIONS

More information

Spectrophotometry. Dr. Shareef SHAIK ASST. PROFESSOR Pharmacology

Spectrophotometry. Dr. Shareef SHAIK ASST. PROFESSOR Pharmacology Spectrophotometry Dr. Shareef SHAIK ASST. PROFESSOR Pharmacology Content Introduction Beer-Lambert law Instrument Applications Introduction 3 Body fluids such as blood, csf and urine contain organic and

More information

SolidSpec-3700 SolidSpec-3700DUV

SolidSpec-3700 SolidSpec-3700DUV C101-E101C SolidSpec-3700 SolidSpec-3700DUV Shimadzu UV-VIS-NIR Spectrophotometer UV-VIS-NIR Spectrophotometer SolidSpec-3700 Standard model SolidSpec-3700DUV Enhanced model for deep UV Measurement to

More information

The DialPath Solution an Easier Way to Analyze Liquids By FTIR. Yanqia Wang, Ph.D. Application Engineer - FTIR Agilent Technologies May 7, 2015

The DialPath Solution an Easier Way to Analyze Liquids By FTIR. Yanqia Wang, Ph.D. Application Engineer - FTIR Agilent Technologies May 7, 2015 The DialPath Solution an Easier Way to Analyze Liquids By FTIR Yanqia Wang, Ph.D. Application Engineer - FTIR Agilent Technologies May 7, 2015 Outline Background of FTIR Transmission Technology Traditional

More information

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications Bedanta Kr. Deka, D. Thakuria, H. Bora and S. Banerjee # Department of Physicis, B. Borooah College, Ulubari,

More information

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES Chemistry 524--Final Exam--Keiderling Dec. 12, 2002 --4-8 pm -- 238 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils are permitted plus one 8.5 x 11 sheet

More information

MOHAWK COLLEGE OF APPLIED ARTS AND TECHNOLOGY CHEMICAL AND ENVIRONMENTAL TECHNOLOGY DEPARTMENT. Lab Report ROOM NO: FE E309

MOHAWK COLLEGE OF APPLIED ARTS AND TECHNOLOGY CHEMICAL AND ENVIRONMENTAL TECHNOLOGY DEPARTMENT. Lab Report ROOM NO: FE E309 MOHAWK COLLEGE OF APPLIED ARTS AND TECHNOLOGY CHEMICAL AND ENVIRONMENTAL TECHNOLOGY DEPARTMENT Lab Report ROOM NO: FE E309 EXPERIMENT NO : 4 TITLE : UV Spectrophotometric Analysis of DNA Submitted by Class

More information

UV-Visible Absorption Analysis of a Two Component System

UV-Visible Absorption Analysis of a Two Component System UV-Visible Absorption Analysis of a Two Component System 1. Purpose This procedure will determine the amounts of vanillin and p-hydroxybenzaldehyde in an unknown mixture using a UV-Vis spectrophotometer.

More information

GENERAL PHARMACOPOEIA MONOGRAPH

GENERAL PHARMACOPOEIA MONOGRAPH MINISTRY OF HEALTH OF THE RUSSIAN FEDERATION GENERAL PHARMACOPOEIA MONOGRAPH Spectrophotometry in the ultraviolet GPM.1.2.1.1.0003.15 and visible spectral regions Replaces the SPRF X GPM, SPRF XI GPM,

More information

Real-time ppb CO 2 Impurity Detection by an Advanced FTIR- UVF System

Real-time ppb CO 2 Impurity Detection by an Advanced FTIR- UVF System Real-time ppb CO 2 Impurity Detection by an Advanced FTIR- UVF System Presented at the BevTech Conference, Albuquerque, NM 2018 by Charles M. Phillips Ph.D., Max Analytical Technologies Mark Taylor, Vice

More information

CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis Updated: 31 August 2014 Print version (Harris, Chapt. 1) (pp.1-20) CEE 772: Instrumental Methods in Environmental Analysis Lecture #3 Statistics: Detection Limits Spectroscopy: Beer s Law & Electronic

More information

Introduction to Fourier Transform Infrared Spectroscopy

Introduction to Fourier Transform Infrared Spectroscopy Introduction to Fourier Transform Infrared Spectroscopy Introduction What is FTIR? FTIR stands for Fourier transform infrared, the preferred method of infrared spectroscopy. In infrared spectroscopy, IR

More information

1. Preliminary qualitative analysis of unknown substances (liquid or solid).

1. Preliminary qualitative analysis of unknown substances (liquid or solid). Name of Procedure: Ultraviolet Spectroscopy Suggested Uses: 1. Preliminary qualitative analysis of unknown substances (liquid or solid). 2. Quantitative analysis of known compounds. Apparatus Used to Perform

More information

Jasco V-670 absorption spectrometer

Jasco V-670 absorption spectrometer Laser Spectroscopy Labs Jasco V-670 absorption spectrometer Operation instructions 1. Turn ON the power switch on the right side of the spectrophotometer. It takes about 5 minutes for the light source

More information

Chapter 15 Molecular Luminescence Spectrometry

Chapter 15 Molecular Luminescence Spectrometry Chapter 15 Molecular Luminescence Spectrometry Two types of Luminescence methods are: 1) Photoluminescence, Light is directed onto a sample, where it is absorbed and imparts excess energy into the material

More information

THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS

THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS INTRODUCTION In recent years, researchers involved in many unrelated scientific disciplines have acquired an interest in accurately

More information

Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS)

Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS) Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS) The UV-VIS spectrometer is a portable ultraviolet light and visible light spectrophotometer, combining a spectrometer and a light source/cuvette

More information

PHOTOELECTRON SPECTROSCOPY IN AIR (PESA)

PHOTOELECTRON SPECTROSCOPY IN AIR (PESA) PHOTOELECTRON SPECTROSCOPY IN AIR (PESA) LEADERS IN GAS DETECTION Since 1977 Model AC-3 Features: Atmospheric pressure operation (unique in the world) Estimate work function, ionization potential, density

More information

Infrared Spectroscopy

Infrared Spectroscopy Reminder: These notes are meant to supplement, not replace, the laboratory manual. Infrared Spectroscopy History and Application: Infrared (IR) radiation is simply one segment of the electromagnetic spectrum

More information

Paper: 12, Organic Spectroscopy Module: 5, Applications of UV spectroscopy

Paper: 12, Organic Spectroscopy Module: 5, Applications of UV spectroscopy Subject Chemistry Paper No and Title Module No and Title Module Tag Paper 12: Organic Spectroscopy Applications of UV-visible Spectroscopy CHE_P12_M5 TABLE OF CONTENTS 1. Learning Outcomes 2. Introduction

More information

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber CH217 Fundamentals of Analytical Chemistry Module Leader: Dr. Alison Willows Electromagnetic spectrum Properties of electromagnetic radiation Many properties of electromagnetic radiation can be described

More information

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD) The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD) Basically all compounds which are less volatile than the mobile phase can be detected. Detection is based on a Universal property of

More information

Basics of UV-Visible Spectroscopy *

Basics of UV-Visible Spectroscopy * OpenStax-CNX module: m34525 1 Basics of UV-Visible Spectroscopy * Brittany L. Oliva-Chatelain Andrew R. Barron This work is produced by OpenStax-CNX and licensed under the Creative Commons Attribution

More information

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS APPLICATION NOTE ICP - Mass Spectrometry Author Kenneth Ong PerkinElmer, Inc. Singapore Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS Introduction The control of impurity

More information

C101-E145 TALK LETTER. Vol. 17

C101-E145 TALK LETTER. Vol. 17 C101-E145 TALK LETTER Vol. 17 UV-VIS Spectroscopy and Fluorescence Spectroscopy (Part 1 of 2) ------- 02 Applications: Spectrofluorophotometers Used in a Variety of Fields ------- 06 Q&A: Is there a way

More information

Strategies to compare far-uv Circular Dichroism spectra of similar proteins using Chirascan -plus ACD

Strategies to compare far-uv Circular Dichroism spectra of similar proteins using Chirascan -plus ACD Chirascan Series Application note Strategies to compare far-uv Circular Dichroism spectra of similar proteins using Chirascan -plus ACD Abstract Circular Dichroism (CD) spectroscopy is an extremely sensitive

More information

TIE-35: Transmittance of optical glass

TIE-35: Transmittance of optical glass PAGE 1/12 0 Introduction Optical glasses are optimized to provide excellent transmittance throughout the total visible range from 400 to 800 nm Usually the transmittance range spreads also into the near

More information

FLS980 Series Reference Guide

FLS980 Series Reference Guide FLS980 Series Reference Guide Integrating Sphere for Measurements of Fluorescence Quantum Yields and Spectral Reflectance Revision 1 Copyrights Copyright 2016 Edinburgh Instruments Ltd. All rights reserved.

More information

SPECTROPHOTOMETRY AND SPECTROMETRY - CONCEPT AND APPLICATIONS

SPECTROPHOTOMETRY AND SPECTROMETRY - CONCEPT AND APPLICATIONS SPECTROPHOTOMETRY AND SPECTROMETRY - CONCEPT AND APPLICATIONS Renjini A 1, Dani Dileep 2 1 Assistant Professor, Department of ECE, Rajadhani Institute of Engineering and Technology, Kerala, India 2 PG

More information

VUV ANALYTICS VGA-100 GC DETECTOR Gas Chromatograph Agilent 6890 equipped with a 7683 model autosampler Restek 30m x 0.25mm x 0.

VUV ANALYTICS VGA-100 GC DETECTOR Gas Chromatograph Agilent 6890 equipped with a 7683 model autosampler Restek 30m x 0.25mm x 0. System Description The VUV PIONA + Analyzer is a VGA-100 Gas Chromatography (GC) detector configured for PIONA compound analysis by the VUV PIONA + method. The VUV PIONA + method uses relatively simple

More information

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD) Application Note 19010803 A Newly Approved ASTM Standard For Analysis of in Using a Pulsed Flame Photometric Detector (PFPD) Keywords ASTM Standard D4735-96 FPD Flame Photometric Detector Model 5380 PFPD

More information

levels. The signal is either absorbance vibrational and rotational energy levels or percent transmittance of the analyte

levels. The signal is either absorbance vibrational and rotational energy levels or percent transmittance of the analyte 1 In this chapter, absorption by molecules, rather than atoms, is considered. Absorption in the ultraviolet and visible regions occurs due to electronic transitions from the ground state to excited state.

More information

PAPER No. 12: ORGANIC SPECTROSCOPY MODULE No. 4: Basic principles and Instrumentation for IR spectroscopy

PAPER No. 12: ORGANIC SPECTROSCOPY MODULE No. 4: Basic principles and Instrumentation for IR spectroscopy Subject Chemistry Paper No and Title Module No and Title Module Tag Paper 12: Organic Spectroscopy Module 4: Basic principles and Instrumentation for IR spectroscopy CHE_P12_M4_e-Text TABLE OF CONTENTS

More information

Studies to develop certified reference material to calibrate spectrophotometer in ultraviolet region

Studies to develop certified reference material to calibrate spectrophotometer in ultraviolet region Studies to develop certified reference material to calibrate spectrophotometer in ultraviolet region Felipe Carvalho da Conceição, Paulo Paschoal Borges and Juliana Freitas Santos Gomes 2 National Institute

More information

Advanced Analytical Chemistry

Advanced Analytical Chemistry 84.514 Advanced Analytical Chemistry Part III Molecular Spectroscopy (continued) Website http://faculty.uml.edu/david_ryan/84.514 http://www.cem.msu.edu/~reusch/virtualtext/ Spectrpy/UV-Vis/spectrum.htm

More information

Ultraviolet-Visible and Infrared Spectrophotometry

Ultraviolet-Visible and Infrared Spectrophotometry Ultraviolet-Visible and Infrared Spectrophotometry Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451

More information

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace_Rev3 Introducing

More information

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer APPLICATION NOTE 72385 Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer Authors Stephan Meding, Katherine Lovejoy, Martin Ruehl Thermo Fisher Scientific,

More information

RS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers

RS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers RS DYNAMICS ECOPROBE 5 Portable IR/PID Gas Analyzer PID + IR PID and IR Analyzers General ECOPROBE 5 has two autonomous analyzers in one case. The combination of analyzers provides a set of data designed

More information

Advanced Pharmaceutical Analysis

Advanced Pharmaceutical Analysis Lecture 2 Advanced Pharmaceutical Analysis IR spectroscopy Dr. Baraa Ramzi Infrared Spectroscopy It is a powerful tool for identifying pure organic and inorganic compounds. Every molecular compound has

More information

BioDrop DUO dsdna Application Note

BioDrop DUO dsdna Application Note BioDrop DUO dsdna Application Note Using a BioDrop DUO spectrophotometer to measure the concentration of low volume samples of dsdna Micro-volume measurement of DNA is a routine application in many life

More information

UV / VIS Spectrophotometer EMCLAB Instruments GmbH

UV / VIS Spectrophotometer EMCLAB Instruments GmbH A Professional Manufacturer UV / VIS Spectrophotometer EMCLAB Instruments GmbH EMCLAB Instruments GmbH Bismarckstrasse 120 47057 Duisburg Germany Phone: +49 203 3064042 Fax: +49 203 3064044 E-mail: info@emc-lab.de

More information