Original Article Performance Verification Test of High Performance Liquid Chromatography: Practical Example

Size: px
Start display at page:

Download "Original Article Performance Verification Test of High Performance Liquid Chromatography: Practical Example"

Transcription

1 Available online at International Journal of Chromatographic Science Universal Research Publications. All rights reserved Original Article Performance Verification Test of High Performance Liquid Chromatography: Practical Example Lantider Kassaye, Getachew Genete Food, Medicine and Health Care Administration and Control Authority, Food and Drug quality Control Laboratory, Addis Ababa, Ethiopia Received 18 January 2013; accepted 11 March 2013 Abstract The performance of an HPLC system can be evaluated by examining the key functions of the various modules that comprise the system. The pump, auto injector, column oven and the detector are the most important parts of an HPLC system which needs to be verified for the proper functioning. Flow rate accuracy and gradient accuracy for the pump; Precision, Linearity and carry over for the auto sampler; wavelength accuracy and response linearity for the detector and temperature accuracy for the column oven are the most important parameters which need to be considered during HPLC performance verification. In this practical example, all of the modules except column oven (for more than 50 o C) meet the acceptance/refusal criteria suggested by Herman Lam Universal Research Publications. All rights reserved Key words: Verification of HPLC; HPLC; Performance qualification of HPLC Introduction Good analytical results are essential in order to take reliable decisions. Analytical measurements affect the daily lives of every citizen. Sound, accurate and reliable analytical measurements are fundamental to the functioning of modern society. A wrong result can have an enormous social and economic impact [1]. The correctness of measurements and measuring instruments is one of the key prerequisites to ensure the quality of products and services, and the accuracy of the instruments must be consistent with their intended use [2]. Calibration and verification are the most important actions to ensure the correct indication of measuring instruments [2]. Regular calibration of measuring instruments should be carried out in agreement with the implemented quality systems. The industrial metrology ensures the appropriate functioning of measurement instruments used in industry as well in production and testing processes, in order to guarantee the quality of life for citizens and for academic research [3]. Verification is the confirmation, based on evidences (facts, test results) that some specified requirement has been fulfilled. The result from a verification assay will show if the measuring equipment is in agreement with its required specifications, which are generally expressed as tolerances [4]. The verification of measuring instruments includes testing and requires the availability of clear specifications and acceptance/refusal criteria [5]. Verification provides means of checking that the deviations between the values displayed by a measuring instrument and the corresponding known values of measured quantity are under control. High-Performance Liquid Chromatography (HPLC) is one of the foremost analytical techniques widely used in analytical laboratories for the analysis of pharmaceuticals and chemicals [6-8], foods [9-11], cosmetics samples and so on [12, 13]. In order to provide a high level of assurance that the data generated from the HPLC analysis are reliable, the performance of the HPLC system should be monitored at regular intervals. The performance of an HPLC system can be evaluated by examining the key functions of the various modules that comprise the system. The common HPLC performance attributes and the acceptance/refusal criteria are presented in Table 1. The acceptance/refusal criteria values for these attributes are based on the values suggested by Herman Lam in the chapter Performance Verification of the HPLC of his book [14]. 2. Experimental 2.1 Instruments and apparatuses A low pressure quaternary HPLC instrument comprises LC-10ATvp pump, SIL-10ADvp auto sampler, SPD- 10AVvp UV-Vis detector and CTO-10AVP column oven (Shimadzu, Japan) has been verified for its proper functioning. Analytical balance (METTLER Toledo, Switzerland), Calibrated stop watch (Treaceble stopwatch, JUMBO DIGIT, VWR, China) and thermometer (Service Testo, Germany) have been used. 2.2 Materials and chemicals ODS analytical column (150 X 4.6 mm, 5μm, Waters, USA), Class A 25mL volumetric flasks (Pyrex, Germany), 18

2 methanol HPLC grade (BDH, USA), Acetonitrile HPLC grade (J.T. Baker, USA) and caffeine (Fluka, Sigma- Aldrich, China, purity 0.99%) were employed for this work. Table 1: Performance attributes and their acceptance/ refusal criteria for HPLC verification Module Performance Acceptance/refusal Attributes criteria Pump Flow rate accuracy ± 1% of the set flow rate Gradient accuracy ± 1% of the step gradient composition Auto injector Precision RSD 1% Linearity r Carryover < 1% Detector Wave length accuracy ± 2nm Linearity of r Column oven response Temperature accuracy ± 2 o C of the set temperature 2.3 Methods Chromatographic condition (For Auto sampler and Detector verification) A mobile phase containing 85% water and 15% acetonitrile was pumped through the analytical column (ODS, 150 X 4.6mm, 5μm) at a flow rate of 2 ml/minute. The test solutions, caffeine in methanol, were injected and detected at 272nm. The column temperature was ambient Verification Test for different modules of HPLC Pump Module: Flow Rate Accuracy The flow-rate of the pump was set at 2 ml/min. and the time required to fill a 25 ml volumetric flask was measured using the calibrated stopwatch. Gradient Accuracy and Linearity The accuracy and linearity of the gradient solvent delivery has been verified indirectly by monitoring the absorbance change as the binary composition of the two solvents changes from two different channels. The LC gradient has four channels: A, B, C and D. The test was performed for two channels at a time. Channel A is filled with a pure solvent, methanol, while channel B is filled with a solvent containing a UV-active tracer, caffeine. The gradient profile is programmed to pump from solvent A only for two minutes and then to decrease the composition of Solvent A by 10% and to increase solvent B by the same percent and allowing to pump for two minutes at each change. The absorbance change at each composition change is measured and expressed as height H in the plot of absorbance versus solvent composition. The linearity of the gradient delivery was verified by plotting the absorbance at various mobile phase compositions versus the theoretical composition. The entire process was repeated for channels C and D Injector Module: Precision The precision of the injector was demonstrated by making six replicate injections from a sample (caffeine 0.05 mg/ml in methanol). The relative standard deviation (RSD) of the response of the injections was calculated to evaluate the precision. Linearity The linearity of the injector was demonstrated by making injections of 5, 10, 20, 30 and 50 μl of the caffeine solution (0.05 mg/ml in methanol). The response of the injections at each injection volume was plotted against the injection volume. The correlation coefficient of the plot was used in the evaluation of the injection linearity. Carryover Carryover was evaluated by injecting a blank (methanol) after a sample that contains a high concentration of analyte ((0.2 mg/ml of caffeine in methanol). The response of the analyte found in the blank sample expressed as a percentage of the response of the concentrated sample was used to determine the level of carryover UV-Visible Detector Module: Wavelength Accuracy The wavelength verification of the UV range was performed by filling a flow cell with a Caffeine solution (0.015mg/mL in methanol) and the UV spectrum was collected in the range of 200 to 400nm. The obtained scan should have λ max at about 272nm and λ min at about 244nm. Linearity of Response The linearity of the detector response was checked by injecting a series of standard solutions of caffeine (0.05, 0.1, 0.2, 0.4 and 1 mg/ml) to the chromatographic system. From the plot of response versus the concentration of the solutions, the correlation coefficient between sample concentration and response was calculated to determine the linearity Column Oven module: The accuracy of the oven was checked at 30, 40, 50, 60 and 70 o C using the calibrated thermometer. The correction factors obtained from the calibration certificate of the thermometer were considered to get the experimental temperature. Results and Discussion 1. Pump module 1.1. Flow rate Accuracy One of the key performance requirements for the pump module is the ability to maintain accurate and consistent flow of the mobile phase, which will be necessary to provide stable and repeatable interactions between the analytes and the stationary phase [15]. Poor flow-rate accuracy will affect the retention time of the separation. As shown in table 2, the results for flow rate accuracy for all of the four suction channels is in the range of to % and this good result indicates that the pump performance is superior Gradient Accuracy and Linearity For gradient accuracy, the ability of the pump to deliver the mobile phase at different solvent strengths over the time by varying the composition of the mobile phase accurately is fundamental to get the adequate chromatographic separation and reproducibility. As shown in fig 1, the absorbance increased step by step as the composition of the caffeine increased. Similarly, the 19

3 absorbance decreased to zero as the composition of the caffeine decreased to zero. The height increased due to the change in caffeine composition was measured by the LC solution software from the baseline and these heights have been used for the determination of gradient accuracy and gradient linearity. The gradient linearity is determined by graphing the percentage composition of caffeine versus the respective obtained height. As shown in figure 2, the relationship between percentage of the composition of Table 2: Flow rate accuracy test result the four pump channels separately. a directly measured by the stopwatch b converted to minutes Rep Vol. (ml) Theoretical time (min) caffeine and the obtained height is linear with a correlation coefficient of Table 3 and Table 4, shows the result for gradient accuracy for channel A and B and channel C and D respectively. The accuracy is calculated by determining the response factor for each percentage change and comparing it with the average response factor. The obtained accuracies at each level are in agreement with the acceptance/refusal criteria for all channels. Line A Line B Line C Line D a b Accu a b Accur a b Accur a b (min:sec racy acy acy (min) (min:sec) (min) (min:sec) (min) (min:sec) (min) ) : : : : : ; : : : : : : : : : : : : : : Table 3: Gradient accuracy results for channel A and B. Channel A Channel B (min) Height 1 Height 2 Height 3 Average a Response Factor b Acc NA Mean a Response factor = Average height/percentage composition of caffeine b Accuracy = Response factor* 100/Average response factor Table 4: Gradient accuracy results for Channel C and D. a Response factor = Average height/percentage composition of caffeine b Accuracy = Response factor* 100/Average response factor Channel C Channel B (min) Height 1 Height 2 Height 2 Ave. Height a Response Factor b Acc NA Mean Auto injector module 2.1. Precision The ability of the injector to draw the same amount of sample in replicate injections is crucial to the precision and accuracy for peak-area or peak-height comparison for 20 Accur acy external standard quantitation. If the variability of the sample and standard being injected into the column is not controlled tightly, the basic principle of external standard quantitation is seriously compromised. No meaningful comparison between the responses of the sample and the

4 standard can be made. The absolute accuracy of the injection volume is not critical as long as the same amount of standard and sample is injected. Table 5 shows the result for the auto sampler precision test and it proves that the auto sampler is precise as the relative standard deviation for the six injections is less than 1%. Fig 1: Chromatogram for gradient accuracy test of channel A and B Fig 2: Graph which shows the relationship between percentage composition of caffeine and height for channel A and B Table 5: Result for the auto sampler precision Replicate Peak area Average STDVE RSD Linearity Most of the automated LC injectors are capable of varying the injection volume without changing the injection loop. Variable volume of sample will be drawn into a sample injection loop by a syringe or other metering device. The uniformity of the sample loop and the ability of the metering device to draw different amounts of sample in proper proportion will affect the linearity of the injection volume. Most of the employed volumes of injection in HPLC are in the range of 5 to 100 µl. However, in this verification test, the auto sampler linearity has been checked at maximum of 50µL which is the installed loop size for the instrument. As table 6 and fig 3 shows, there is good relationship between volume of injection and response (peak area). The correlation coefficient is which is more than the acceptance/refusal criteria, Table 6: Results for Auto injector linearity with its acceptance/refusal criteria Inj. vol (µl) Peak Area 1 Peak area 2 Peak area 3 Average RSD Correlation coefficient, r Accepted/Not Accepted Accepted Fig 3: Graph for the relationship between volume of injection and peak area 2.3. Carry over Small amounts of analyte may get carried over from the sample injected before and lead to the contamination of the next sample to be injected. The carryover will affect the accuracy of the quantification of the next sample. The carryover assay was carried out by injecting a blank (methanol) after caffeine standard solution (0.2 mg-ml -1 in methanol). The level of carryover was determined by calculating the ratio between the responses (peaks-area) of the caffeine found in the methanol sample and the standard solution. As shown in Table 7, the average percent carry over (n=3) for the auto sampler is 0.26% which is much lower than the acceptance/refusal criteria, 1%. The chromatogram for the sample and blank are shown in fig. 4 and 5. And hence we can say that the cleansing procedure for the auto sampler is good. Moreover, results for precision and linearity prove that the auto sampler is in good condition to use it for quantitative analysis. Table 7: Carry over test results for the auto injector Replicate Peak Area Carry Accepted/Not over Accepted Sample Blank Accepted Accepted Accepted Mean Accepted RSD

5 3. Detector 3.1 Wavelength accuracy Wavelength accuracy is defined as the deviation of the wavelength reading at an absorption or emission band from the known wavelength of the band. The accuracy and sensitivity of the measurement will be compromised if there is a wavelength accuracy problem. Fig 4 shows the UV spectrum for caffeine solution in the range of 190 to 380nm. The obtained λ max and λ min of this spectrum prove that the detector has no problem regarding to wavelength accuracy. Table 8: Summarized results for detector linearity test Conc. Peak Peak Peak Mean RSD (mg/ml) area1 area2 area3 area Correlation coefficient, r = Accept/Not Accept Accepted Fig 5: Graph for Detector linearity Fig 4: UV spectrum for caffeine 3.2 Detector response Linearity The detector linearity is very important when the purpose of work is to carry out quantitative analysis. The linearity of the detector is important to the accuracy for the peak area and peak height comparison between standards and samples and accordingly to the determination of analyte (s) in these samples. Results for detector linearity test have been summarized and graphed in Table 8 and figure 5 respectively. 4. Column Oven Capacity factor, k of an analyte decreases with as temperature increases, and hence the retention of the analysis decreases with temperature [16, 17]. The ability to maintain an accurate column temperature is highly essential to achieve the desired retention time and resolution requirements in the separation process. The temperature accuracy of the column oven is evaluated by placing a calibrated thermometer in the column compartment to measure the actual compartment temperature. As table 9 shows, the column oven is not accurate for temperature more than 50 o C and hence it is highly recommended not to use this HPLC for chromatographic conditions which require column temperature more than 50 o C. Table 9: Results for performance verification of column oven Set Temperature ( o C) Reading ( o C) Correction factor ( o C) Experimental Temperature ( o C) Conclusion The performance of an HPLC system can be evaluated by examining the key functions of the various modules that comprise the system. The common HPLC performance attributes that must be qualified are pump, auto injector, column oven and detector. The obtained verification results for all of the attributes are complying with the acceptance/refusal criteria values as per Herman Lam suggestion. However, the column oven is out of the acceptance/refusal criteria for temperature more than 50 o C. And hence, it is quite logical to use this HPLC instrument for the day to day analytical purpose 1. Prichard, E.; Barwick, V.; Quality Assurance in Analytical Chemistry ; Willey Sommer, K.; Chappel, S.; Kochsiek, M. Calibration and Verification: Two Procedures Having Comparable Objectives and Results ; OIML BULLETIN; vol. XLII, number 1, European Association of National Metrology Institutes (EURAMET); Metrology in Short ; third edition; 22

6 Compañó Beltrán, Ramón; Ríos Castro, Ángel. Garantía de la Calidad en los Laboratorios Analíticos ; Síntesis. 5. Huber, L.; Validation and Qualification in Analytical Laboratories ; Informa Healthcare; B. Palanikumar, A. Thenmozhi, D. Sridharan (2010). A RP-HPLC method for the simultaneous estimation of ceftriaxone sodium and sulbactam sodium injection dosage form. International Journal of Pharmacy and Pharmaceutical Sceiences, Vol 2, A. Narendra, D. Deepika, M. Mathrusri Annapurna (2012). Validated LC method for the estimation of dorzolamide HCl (Carbonic Anhydrase inhibitor) in ophthalmic solutions. E-Journal of chemistry, 9(3), B. P. Nagoria, Amit Maru, Pankaj Muysunic, Subhash Gupta (2011). Method development and its validation for the simultaneous estimation of timolol maleate and dorzolamide HCL in API and ophthalmic solution dosage form by RPHPLC. Journal of Chemical and pharmaceutical Research, 3(4), P. Salo-vaanane, P. Ollilainen, P. Mattila, K. Lehikoinen, E. Salmela-Molsa, V. Piironen (2000). Simultaneous HPLC analysis of fate soluble vitamins in selected animal products after small-scale extraction. Food chemistry, Vol. 71, Macrae (2007). Applications of high pressure liquid chromatography to food analysis. International Journal of food science and Technology, Vol. 15, A. steppuhn, F.L. Wackers (2004). HPLC sugar analysis reveals the nutritional sate and the feeding history of parasitoids. Functional Ecology, 18, Pai-Wen Wu Chieu-Chen Chang, Shin-Shou Chou (2003). Determination of formaldehyde in Cosmetics by HPLC method and Acetylacetone. Journal of food and Drug Analysis, Vol. 11, Wei-Sheng Huang, Cheng-Chinlin, Ming-Chauan Huang, Kuo-Ching Wen (2002). Determination of α- Hydroxyacids in Cosmetics, Journal of Food and Drug Analysis, Vol. 10, Chan, C. et al; Analytical Method Validation and Instrument Performance Verification ; Wiley- Interscience; Guidelines for Calibration in Analytical Chemistry, Part 1: Fundamentals and Single Component Calibration, IUPAC Recommendation, Pure and Applied Chemistry; vol. 70; No. 4; pp , R. G. Wolcott and J. W. Dolan, Column temperature effects in gradient elution, LCGC, 16(12), 1080, J. W. Dolan, The important of temperature, LC-GC, 20(6), 524, Source of support: Nil; Conflict of interest: None declared 23

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (11) 04 QUALIFICATION OF EQUIPMENT ANNEX 1: QUALIFICATION OF HPLC EQUIPMENT Full document title and reference Document type

More information

Qualification Specifications for Agilent 1100/1200 HPLC Systems

Qualification Specifications for Agilent 1100/1200 HPLC Systems Agilent 1100/1200 HPLC Systems 35 Vantage Point Drive, Rochester NY 14624 Phone: 800-828-1470 Fax: 800-395-0543 Transcat.com Test Set Points/ Range Acceptance Criteria Flow Rate Accuracy and Precision

More information

HPLC Praktikum Skript

HPLC Praktikum Skript HPLC Praktikum Skript Assistants: Gianluca Bartolomeo HCI D330, 3 46 68, bartolomeo@org.chem.ethz.ch Sahar Ghiasikhou HCI E330, 2 29 29, ghiasikhou@org.chem.ethz.ch 1. Introduction In chromatographic techniques,

More information

Volume 6, Issue 2, January February 2011; Article-015

Volume 6, Issue 2, January February 2011; Article-015 Research Article DEVELOPMENT AND VALIDATION OF A RP-HPLC METHOD FOR THE DETERMINATION OF DAPOXETINE HYDROCHLORIDE IN PHARMACEUTICAL FORMULATION USING AN EXPERIMENTAL DESIGN Pratik Mehta*, Ujjwal Sahoo,

More information

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 9 DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 9 Determination of drug release during

More information

Reverse Phase High Performance Liquid Chromatography method for determination of Lercanidipine hydrochloride in bulk and tablet dosage form

Reverse Phase High Performance Liquid Chromatography method for determination of Lercanidipine hydrochloride in bulk and tablet dosage form Research Article ISSN: 0974-6943 M.V.Kumudhavalli et al. / Journal of Pharmacy Research 2014,8(11), Available online through http://jprsolutions.info Reverse Phase High Performance Liquid Chromatography

More information

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC.

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC. 7. Stability indicating analytical method development and validation of and in capsule dosage form by HPLC. 7.1 INSTRUMENTS AND MATERIALS USED 7.1.1 INSTRUMENTS 1. Shimadzu LC-2010 CHT with liquid chromatograph

More information

Impact factor: 3.958/ICV: 4.10 ISSN:

Impact factor: 3.958/ICV: 4.10 ISSN: Impact factor: 3.958/ICV: 4.10 ISSN: 0976-7908 99 Pharma Science Monitor 9(4), Oct-Dec 2018 PHARMA SCIENCE MONITOR AN INTERNATIONAL JOURNAL OF PHARMACEUTICAL SCIENCES Journal home page: http://www.pharmasm.com

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION Andrew J. Aubin and Tanya L. Jenkins Waters Corporation, Milford, MA, USA INTRODUCTION Benzocaine (4-Aminobenzoic

More information

Appendix II- Bioanalytical Method Development and Validation

Appendix II- Bioanalytical Method Development and Validation A2. Bioanalytical method development 1. Optimization of chromatographic conditions Method development and optimization of chromatographic parameters is of utmost important for validating a method in biological

More information

A novel approach to the analysis of multivitamin by online supercritical fluid extraction/supercritical fluid chromatography

A novel approach to the analysis of multivitamin by online supercritical fluid extraction/supercritical fluid chromatography PO-CON1618E A novel approach to the analysis of multivitamin by online supercritical fluid extraction/supercritical fluid chromatography Pittcon 216 83-12 Qiang Li, Hongyuan Hao, Taohong Huang, Shin-ichi

More information

Quantitative Analysis of Caffeine in Energy Drinks by High Performance Liquid Chromatography

Quantitative Analysis of Caffeine in Energy Drinks by High Performance Liquid Chromatography Quantitative Analysis of Caffeine in Energy Drinks by High Performance Liquid Chromatography CHEM 329 Professor Vogt TA: Fahad Hasan Allison Poget Date Performed: April 5, 2016 Date Submitted: April 12,

More information

A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS

A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS CH.MOUNIKA*, M.KINNERA Research Article SIR.C.R.REDDY COLLEGE OF PHARMACEUTICAL SCIENCES, ELURU.

More information

Ketorolac tromethamine (KT)[1] is

Ketorolac tromethamine (KT)[1] is RESEARCH ARTICLE Derivative and Simultaneous Equation Methods for the Determination of Fluorometholone and Ketorolac in Ophthalmic Preparations Chandaka Prasanna Kumar*, Bulusu Ravi Teja, Bhupathiraju

More information

Fast methods for the determination of ibuprofen in drug products

Fast methods for the determination of ibuprofen in drug products APPLICATION NOTE 779 Fast s for the determination of ibuprofen in drug products Authors Sylvia Grosse, Mauro De Pra, Frank Steiner, Thermo Fisher Scientific, Germering, Germany Keywords Pharmaceutical,

More information

STABILITY INDICATING RP HPLC METHOD FOR ANALYSIS OF DORZOLAMIDE HCl IN THE BULK DRUG AND IT S PHARMACEUTICAL DOSAGE FORM

STABILITY INDICATING RP HPLC METHOD FOR ANALYSIS OF DORZOLAMIDE HCl IN THE BULK DRUG AND IT S PHARMACEUTICAL DOSAGE FORM International Journal of Pharmacy and Pharmaceutical Sciences ISSN- 0975-1491 Vol 3, Issue 3, 2011 Research Article STABILITY INDICATING RP HPLC METHOD FOR ANALYSIS OF DORZOLAMIDE HCl IN THE BULK DRUG

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis Reference Material - Primary Standard Product Name: Oxycodone 1.0 mg/ml in Methanol Catalogue Number: Lot Number: 37583 LGCAMP0672.10-01 CAS Number: 76-42-6 Molecular Formula: C

More information

Chapter 4: Verification of compendial methods

Chapter 4: Verification of compendial methods Chapter 4: Verification of compendial methods Introduction In order to ensure accurate and reliable test results, the quality control laboratory (QCL) needs to use analytical methods (and accompanying

More information

Development And Validation Of Rp-Hplc Method For Determination Of Velpatasvir In Bulk

Development And Validation Of Rp-Hplc Method For Determination Of Velpatasvir In Bulk International Journal of Engineering Science Invention (IJESI) ISSN (Online): 2319 6734, ISSN (Print): 2319 6726 www.ijesi.org PP. 36-41 Development And Validation Of Rp-Hplc Method For Determination Of

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Dehydroepiandrosterone 3-sulfate, Primary Standard D-065 Page 1 of 6 Catalog Number: D-065 Lot: H 3 C Expiration: Description: February 2019 Dehydroepiandrosterone

More information

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer APPLICATION NOTE 72385 Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer Authors Stephan Meding, Katherine Lovejoy, Martin Ruehl Thermo Fisher Scientific,

More information

Journal of Chemical and Pharmaceutical Research, 2017, 9(1): Research Article

Journal of Chemical and Pharmaceutical Research, 2017, 9(1): Research Article Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research, 2017, 9(1):118-122 Research Article ISSN : 0975-7384 CODEN(USA) : JCPRC5 Development and Validation of High Performance Liquid

More information

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage:

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage: Research Article CODEX: IJRPJK ISSN: 2319 9563 International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage: www.ijrpns.com DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR THE

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis Reference Material - Primary Standard Product Name: Metoprolol Tartrate 1.0 mg/ml in Methanol (as free base) Catalogue Number: LGCAMP0027.00-01 Lot Number: 8525 CAS Number: 56392-17-7

More information

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Determination of Trace Cations in Power Plant Waters Containing Morpholine Application Note 8 Determination of Trace Cations in Power Plant Waters Containing Morpholine INTRODUCTION Morpholine and ammonium are used as additives in power plant waters. Morpholine acts as a corrosion

More information

Research Article METHOD DEVELOPMENT AND VALIDATION FOR SIMULTANEOUS ESTIMATION OF ELBASVIR AND GRAZOPREVIR BY RP-HPLC

Research Article METHOD DEVELOPMENT AND VALIDATION FOR SIMULTANEOUS ESTIMATION OF ELBASVIR AND GRAZOPREVIR BY RP-HPLC ISSN 2395-3411 Available online at www.ijpacr.com 248 Research Article METHOD DEVELOPMENT AND VALIDATION FOR SIMULTANEOUS ESTIMATION OF ELBASVIR AND GRAZOPREVIR BY RP-HPLC Mantha Vebkatesh* and A. Yasodha

More information

Asian Journal of Research in Chemistry and Pharmaceutical Sciences Journal home page:

Asian Journal of Research in Chemistry and Pharmaceutical Sciences Journal home page: Research Article ISSN: 2349 7106 Asian Journal of Research in Chemistry and Pharmaceutical Sciences Journal home page: www.ajrcps.com ESTIMATION OF RAMELTEON IN TABLET DOSAGE FORM BY HPLC M. Jyothsna*

More information

STABILITY INDICATING METHOD OF RELATED IMPURITIES IN VENLAFAXINE HYDROCHLORIDE SUSTAINED RELEASE TABLETS

STABILITY INDICATING METHOD OF RELATED IMPURITIES IN VENLAFAXINE HYDROCHLORIDE SUSTAINED RELEASE TABLETS Issn No: 976-39 RESEARCH ARTICLE STABILITY INDICATING METHOD OF RELATED IMPURITIES IN VENLAFAXINE HYDROCHLORIDE SUSTAINED RELEASE TABLETS CHETLAPALLI SATYA SRINIVAS 1, P.RENUKA DEVI 2 and GAMPA VIJAYAKUMAR*

More information

SIMULTANEOUS RP HPLC DETERMINATION OF CAMYLOFIN DIHYDROCHLORIDE AND PARACETAMOL IN PHARMACEUTICAL PREPARATIONS.

SIMULTANEOUS RP HPLC DETERMINATION OF CAMYLOFIN DIHYDROCHLORIDE AND PARACETAMOL IN PHARMACEUTICAL PREPARATIONS. Ind. J. Anal. Chem Vol. 7 11. 2008 SIMULTANEOUS RP HPLC DETERMINATION OF CAMYLOFIN DIHYDROCHLORIDE AND PARACETAMOL IN PHARMACEUTICAL PREPARATIONS. Authors for correspondence : R. R. Singh1*, M. V. Rathnam,

More information

Determination of Caffeine by HPLC

Determination of Caffeine by HPLC Determination of Caffeine by HPLC Introduction It was a long history before real high performance liquid chromatography (HPLC) had evolved. The very first indication of a chromatographic separation was

More information

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46>

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46> High Performance Liquid Chromatography HPLC-9 Ultra-high Speed Analysis of Ibuprofen within USP Allowed Limits by Nexera Method Scouting In recent years, high-throughput analytical techniques have

More information

KEYWORDS: Acetaminophen, Doxylamine succinate, Dextromethorphan hydrobromide.

KEYWORDS: Acetaminophen, Doxylamine succinate, Dextromethorphan hydrobromide. International Journal of Pharmaceutical Science Invention ISSN (Online): 2319 6718, ISSN (Print): 2319 670X Volume 3 Issue 7 July 2014 PP.08-12 Analytical method development and validation of acetaminophen,

More information

International Journal of Research and Reviews in Pharmacy and Applied science.

International Journal of Research and Reviews in Pharmacy and Applied science. International Journal of Research and Reviews in Pharmacy and Applied science www.ijrrpas.com Corresponding Author NARENDRA DEVANABOYINA* Ch.DURGAKUMARI, L.NAGARAJU, R.SRIDHARANI PRIYA, S.BHAVITHA Department

More information

Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets

Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets Liquid chromatography was developed by Tswett in early 1900 s and was shown to be a powerful separation

More information

Int. J. Pharm. Sci. Rev. Res., 30(2), January February 2015; Article No. 09, Pages: 63-68

Int. J. Pharm. Sci. Rev. Res., 30(2), January February 2015; Article No. 09, Pages: 63-68 Research Article Stability indicating RP-HPLC Method for Determination of FexoFenadine Hydrochloride and its Related Substances in Active Pharmaceutical Substance Abhay Gupta* 1, Dr. Birendra Srivastava,

More information

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity Agilent 12 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity A case study with degraded omeprazole drug product Application Note Small Molecule

More information

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System Margaret Maziarz, Sean M. McCarthy, Michael D. Jones, and Warren B. Potts Waters Corporation, Milford, MA, USA A P

More information

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India QbD Based Method Development on an Agilent 129 Infinity UHPLC System Combined with a Seamless Method Transfer to HPLC Using Intelligent System Emulation Technology Application Note Pharmaceutical QA/QC

More information

Development and Validation of Stability Indicating RP-HPLC Method for the Determination of Anagrelide HCl in Pharmaceutical Formulation

Development and Validation of Stability Indicating RP-HPLC Method for the Determination of Anagrelide HCl in Pharmaceutical Formulation ISSN 0976 3333 Available Online at www.ijpba.info International Journal of Pharmaceutical & Biological Archives 2013; 4(2): 342-346 ORIGINAL RESEARCH ARTICLE Development and Validation of Stability Indicating

More information

Novus International Journal of Analytical Innovations 2012, Vol. 1, No. 3

Novus International Journal of Analytical Innovations 2012, Vol. 1, No. 3 Novus International Journal of Analytical Innovations 2012, Vol. 1, No. 3 ISSN 2278-6953 www.novusscientia.org Accepted on October 22, 2012 RP-HPLC method for simultaneous estimation of Avitriptan and

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis Bromazepam 7-Bromo-1,3-dihydro-5-(2-pyridinyl)-2H-1,4-benzodiazepin-2-one Catalog Number: Br Solution Lot: Expiration Date: Solvent: August 2015 Methanol Volume per Ampule: Not

More information

Validation of Stability-Indicating RP-HPLC Method for the Assay of Ibrutinib in Pharmaceutical Dosage form

Validation of Stability-Indicating RP-HPLC Method for the Assay of Ibrutinib in Pharmaceutical Dosage form Validation of Stability-Indicating RP-HPLC Method for the Assay of Ibrutinib in Pharmaceutical Dosage form 8.1 Introduction Ibrutinib (IBR) is an anticancer drug targeting B-cell malignancies (blood cancer

More information

Method Development and Validation Of Prasugrel Tablets By RP- HPLC

Method Development and Validation Of Prasugrel Tablets By RP- HPLC Method Development and Validation Of Prasugrel Tablets By RP- HPLC K.Sonia*, Ndwabe Hamunyare, K.Manikandan Department of Pharmaceutical Analysis, SRM College of Pharmacy, SRM University, Kattankulathur,

More information

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India Reducing analysis time and solvent consumption for isocratic USP assay methods with current and proposed USP guidelines using the Agilent 129 Infinity LC System An efficient way to reduce cost of analysis

More information

Journal of Chemical and Pharmaceutical Research

Journal of Chemical and Pharmaceutical Research Available on line www.jocpr.com Journal of Chemical and Pharmaceutical Research ISSN No: 0975-7384 CODEN(USA): JCPRC5 J. Chem. Pharm. Res., 2010, 2(5):399-417 Method develpopment and validation of Hydrochloride

More information

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW 75 CHAPTER 3 DEVELOPMENT AND APPLICATION OF STABILITY-INDICATING HPLC METHOD FOR THE DETERMINATION OF NEVIRAPINE AND ITS IMPURITIES IN COMBINATION DRUG PRODUCT 3.1 INTRODUCTION OF DOSAGE FORM AND LITERATURE

More information

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Page 1 of 13 CIPAC/4105/R GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Prepared for CIPAC by Dr M J Tandy*, P M Clarke and B White (UK) The rapid

More information

DEVELOPMENT AND VALIDATION OF A HPLC METHOD FOR IN-VIVO STUDY OF DICLOFENAC POTASSIUM

DEVELOPMENT AND VALIDATION OF A HPLC METHOD FOR IN-VIVO STUDY OF DICLOFENAC POTASSIUM IJPSR (2013), Vol. 4, Issue 2 (Research Article) Received on 28 September, 2012; received in revised form, 24 November, 2012; accepted, 23 January, 2013 DEVELOPMENT AND VALIDATION OF A HPLC METHOD FOR

More information

ANALYTICAL METHOD PROCEDURES

ANALYTICAL METHOD PROCEDURES HPLC ASSAY AND RELATED SUBSTANCE Column Eurospher 100, C18, 25 x 0.40 cm 5µ Mobile Phase Buffer ph 2.0*: Acetonitrile (88:12 v/v) * Buffer ph 2 Potassium dihydrogen phosphate (KH 2 PO 4 ) - 0.68g Hepatane

More information

Pharmacophore 2011, Vol. 2 (4), ISSN Pharmacophore. (An International Research Journal)

Pharmacophore 2011, Vol. 2 (4), ISSN Pharmacophore. (An International Research Journal) Pharmacophore 2011, Vol. 2 (4), 232-238 ISSN 2229 5402 Pharmacophore (An International Research Journal) Available online at http://www.pharmacophorejournal.com/ Original Research Paper SIMULTANEOUS ANALYSIS

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis Reference Material - Primary Standard Product Name: LSD (Lysergic Acid Diethylamide) 1.0 mg/ml in Acetonitrile Catalogue Number: Lot Number: 16905 LGCAMP1346.00-11 O CAS Number:

More information

J Pharm Sci Bioscientific Res (4): ISSN NO

J Pharm Sci Bioscientific Res (4): ISSN NO Development and Validation of Stability Indicating Analytical Method for Simultaneous Estimation of Perindopril and Potassium in Their Combined Marketed Dosage Form ABSTRACT: Gurjeet Kaur*, Nikhil Patel

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Amiodarone-D 4, Primary Standard A-83 Page 1 of 8 Catalog Number: A-83 Lot: Expiration: Description: Packaging: Storage: Shipping: Intended Use: Instructions

More information

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard Certified Reference Material - Certificate of Analysis Methylone, Primary Standard Page 1 of 9 Catalog Number: O Lot: Expiration Date: Description: Packaging: Storage: Shipping: Intended Use: Instructions

More information

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India Effective use of pharmacopeia guidelines to reduce cost of chromatographic analysis Optimized, cost-effective HPLC analysis of atorvastatin by varying column dimensions within the USP allowed limts

More information

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 126 Infinity Binary LC system with UV detection Application Note Food Author Srividya Kailasam Agilent Technologies,

More information

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump Performance characteristics of the Agilent 129 Infinity Quaternary Pump Technical Overview Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract This Technical Overview presents Proof

More information

TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD

TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD APPENDIX V TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD Validating Chromatographic Methods. By David M. Bliesner Copyright 2006 John Wiley & Sons, Inc. 159 160 APPENDIX V Title: Effective: Document No:

More information

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD TO DETERMINE CINITAPRIDE HYDROGEN TARTARATE IN BULK AND PHARMACEUTICAL FORMULATION

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD TO DETERMINE CINITAPRIDE HYDROGEN TARTARATE IN BULK AND PHARMACEUTICAL FORMULATION Research Article ISSN:2230-7346 Journal of Global Trends in Pharmaceutical Sciences Vol.3, Issue 2, pp -619-627, April June 2012 DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD TO DETERMINE CINITAPRIDE HYDROGEN

More information

Journal of Drug Delivery and Therapeutics

Journal of Drug Delivery and Therapeutics Available online on 15.01.2018 at http://jddtonline.info Journal of Drug Delivery and Therapeutics Open Access to Pharmaceutical and Medical Research 2011-17, publisher and licensee JDDT, This is an Open

More information

Pelagia Research Library

Pelagia Research Library Available online at www.pelagiaresearchlibrary.com Der Pharmacia Sinica, 2011, 2 (2): 68-73 ISSN: 0976-8688 CODEN (USA): PSHIBD Method development and validation for determination of methane sulphonic

More information

International Journal of Current Trends in Pharmaceutical Research. International Journal of Current Trends in Pharmaceutical Research

International Journal of Current Trends in Pharmaceutical Research. International Journal of Current Trends in Pharmaceutical Research International Journal of Current Trends in Pharmaceutical Research Journal Home Page: www.pharmaresearchlibrary.com/ijctpr Research Article Open Access Development and Validation Levofloxacin Andambroxol

More information

Preparative Separation of Active Components from Natural Products using Low-pressure Gradient Preparative HPLC

Preparative Separation of Active Components from Natural Products using Low-pressure Gradient Preparative HPLC PO-CON1405E Preparative Separation of Active Components from Natural Products using Low-pressure Gradient Preparative HPLC Pittcon 014 50-7P Kenichiro Tanaka 1, William Hedgepeth 1, Lincoln Grimes, Tsutomu

More information

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application Pharmaceuticals Author Cliff Woodward and Ronald Majors Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808

More information

Intercontinental journal of pharmaceutical Investigations and Research

Intercontinental journal of pharmaceutical Investigations and Research Kranthi K K et al, ICJPIR 2017, 4(1), XXX-XXX Available online at ISSN: 2349-5448 Intercontinental journal of pharmaceutical Investigations and Research ICJPIR Volume 4 Issue 1 Jan Mar- 2017 Research Article

More information

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC USP Method Transfer and Routine Use Analysis of Tablets from HPLC to UPLC Aparna Chavali, Tanya Jenkins and Patricia McConville Waters Corporation, 34 Maple Street, Milford, MA USA APPLICATION BENEFITS

More information

Analytical Method Development and Validation of Lafutidine in Tablet dosage form by RP-HPLC

Analytical Method Development and Validation of Lafutidine in Tablet dosage form by RP-HPLC International Journal of ChemTech Research CODEN( USA): IJCRGG ISSN : 0974-4290 Vol. 3, No.3, pp 1403-1407, July-Sept 2011 Analytical Method Development and Validation of Lafutidine in Tablet dosage form

More information

RESOLUTION OENO 33/2004 DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION

RESOLUTION OENO 33/2004 DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION The GENERAL ASSEMBLY, Considering Article paragraph iv of the agreement establishing the International organisation of vine and wine Upon

More information

Modernizing the USP Monograph for Acetaminophen

Modernizing the USP Monograph for Acetaminophen Modernizing the USP Monograph for cetaophen pplication Note Small Molecule Pharmaceuticals and Generics uthor Rongjie Fu gilent Technologies (Shanghai) Co. Ltd bstract new HPLC method was developed for

More information

IDENTIFICATION OF STEROIDS IN COSMETIC PRODUCTS BY TLC AND HPLC 1 02/12/2005 ACM 007 A. THIN LAYER CHROMATOGRAPHY (TLC)

IDENTIFICATION OF STEROIDS IN COSMETIC PRODUCTS BY TLC AND HPLC 1 02/12/2005 ACM 007 A. THIN LAYER CHROMATOGRAPHY (TLC) Document A. THIN LAYER CHROMATOGRAPHY (TLC) 1. SCOPE AND FIELD OF APPLICATION The method describes the identification of hydrocortisone acetate, dexamethasone, betamethasone, betamethasone 17-valerate

More information

Department of Quality Assurance, Luqman College of Pharmacy, GULBARGA (K.S.) INDIA ABSTRACT

Department of Quality Assurance, Luqman College of Pharmacy, GULBARGA (K.S.) INDIA ABSTRACT Int. J. Chem. Sci.: 12(3), 2014, 871-879 ISSN 0972-768X www.sadgurupublications.com DEVELPMENT AND VALIDATIN F A RAPID RP HPLC METHD FR THE DETERMINATIN F CINITAPRIDE HYDRGEN TARTARATE IN PURE AND ITS

More information

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc. Agilent Application Solution Transfer of a USP method for tolazamide from normal phase HPLC to SFC using the Agilent 126 Infinity Hybrid SFC/UHPLC System Improving peak shape and providing wider UV selectivity

More information

TECHNICAL BRODIFACOUM

TECHNICAL BRODIFACOUM BRODIFACOUM Full specification: Approved 10 December 1999 1. Specification 1.1 Description The material shall consist of brodifacoum together with related manufacturing impurities and shall be in the form

More information

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput MAY 2005 SEPARATION SCIENCE REDEFINED 31 Assay Transfer from HPLC to UPLC for Higher Analysis Throughput A typical HPLC assay was transferred and optimized for a Waters ACQUITY UPLC system to achieve both

More information

Extraction Process Validation of Isatis Radix

Extraction Process Validation of Isatis Radix Pharmacology & Pharmacy,,, - Published Online April in SciRes. http://www.scirp.org/journal/pp http://dx.doi.org/./pp.. Extraction Process Validation of Isatis Radix Junshan Li, Bo Zhang Shineway Pharmaceutical

More information

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy Analytical Quantification of Deoxygenated Compounds in Catalytic Reaction Samples as an Evaluation Technique to Detere Reaction Conversion from a Bioderived Oil in Novel Biofuel Testing Application Note

More information

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

Prep 150 LC System: Considerations for Analytical to Preparative Scaling Andrew Aubin and Jo-Ann Jablonski Waters Corporation, Milford, MA, USA APPLICATION BENEFITS The Prep 150 LC System is an affordable, highly reliable system for preparative chromatography and is suitable

More information

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Replaces: Dated: Author: Date: AM-No.: New New Nils Arne Jentoft 18.06.2014 0 CHANGES This procedure is new. 1 SCOPE This document describes

More information

Development and validation of stability indicating reverse phase high performance liquid chromatography method for Timolol Maleate

Development and validation of stability indicating reverse phase high performance liquid chromatography method for Timolol Maleate International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.6, No.5, pp 1429-1435, Sept-Oct 2014 Development and validation of stability indicating reverse phase high performance

More information

Stability indicating RP-HPLC method for determination of azilsartan medoxomil in bulk and its dosage form

Stability indicating RP-HPLC method for determination of azilsartan medoxomil in bulk and its dosage form IJPAR Vol.3 Issue 4 Oct-Dec-2014 Journal Home page: ISSN: 2320-2831 Research article Open Access Stability indicating RP-HPLC method for determination of azilsartan medoxomil in bulk and its dosage form

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

CHAPTER V ANALYTICAL METHODS ESTIMATION OF DICLOFENAC. Diclofenac (gift sample from M/s Micro Labs Ltd., Pondicherry)

CHAPTER V ANALYTICAL METHODS ESTIMATION OF DICLOFENAC. Diclofenac (gift sample from M/s Micro Labs Ltd., Pondicherry) CHAPTER V ANALYTICAL METHODS ESTIMATION OF DICLOFENAC A UV spectrophotometric method based on the measurement of absorbance at 276nm in phosphate buffer of p H 7.4 was used in the present study of the

More information

H 3 CO H 3 CO S CH 3

H 3 CO H 3 CO S CH 3 FENITROTHION 35 H 3 CO P H 3 CO S O CH 3 NO 2 ISO common name Chemical name Fenitrothion O,O-Dimethyl O-4-nitro-m-tolyl phosphorothioate (IUPAC) O,O-Dimethyl O-(3-methyl-4-nitrophenyl)- phosphorothioate

More information

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler No. SSI-GC- Gas Chromatography No. GC- Analysis of Residual Solvents in Pharmaceuticals (USP) with Shimadzu GC- Plus and HS- Headspace Sampler Introduction Organic solvents are routinely used in manufacturing

More information

Routine MS Detection for USP Chromatographic Methods

Routine MS Detection for USP Chromatographic Methods Daniel S. Root, Thomas E. Wheat, and Patricia McConville Waters Corporation, Milford, MA, USA APPLICATION BENEFITS This approach enables mass spectral analysis of peaks directly from unmodified USP HPLC

More information

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity Janet Hammond Waters Corporation, Wilmslow, UK APPLICATION BENEFITS High

More information

SIMPLE AND SENSITIVE VALIDATED REVERSE PHASE HPLC-UV METHOD FOR THE DETERMINATION OF LYMECYCLINE IN PHARMACEUTICAL DOSAGE FORMS

SIMPLE AND SENSITIVE VALIDATED REVERSE PHASE HPLC-UV METHOD FOR THE DETERMINATION OF LYMECYCLINE IN PHARMACEUTICAL DOSAGE FORMS IJPSR (2012), Vol. 3, Issue 07 (Research Article) Received on 26 March, 2012; received in revised form 25 June, 2012; accepted 28 June, 2012 SIMPLE AND SENSITIVE VALIDATED REVERSE PHASE HPLC-UV METHOD

More information

Pelagia Research Library

Pelagia Research Library Available online at www.pelagiaresearchlibrary.com Der Pharmacia Sinica, 2014, 5(1):34-39 ISSN: 0976-8688 CODEN (USA): PSHIBD Advance simultaneous determination of paracetamol, thiocolchicoside and aceclofenac

More information

Stability-indicating HPLC determination of tolterodine tartrate in pharmaceutical dosage form

Stability-indicating HPLC determination of tolterodine tartrate in pharmaceutical dosage form Indian Journal of Chemical Technology Vol. 13, May 2006, pp. 242-246 Stability-indicating HPLC determination of tolterodine tartrate in pharmaceutical dosage form Vinay Saxena a *, Zahid Zaheer b & Mazhar

More information

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform -xt SYSTEM Prep and Load Platform -xt Specifications for PAL-xt Systems Valid for PAL-xt System models only Revised March 2013 V5 PAL is a registered trademark of CTC Analytics AG Switzerland -xt SYSTEM

More information

Simultaneous estimation of amitryptyline and chlordiazepoxide by RP-HPLC method

Simultaneous estimation of amitryptyline and chlordiazepoxide by RP-HPLC method IJPAR Vol.3 Issue 4 Oct-Dec-2014 Journal Home page: ISSN: 2320-2831 Research article Open Access Simultaneous estimation of amitryptyline and chlordiazepoxide by RP-HPLC method Dr.R.Srinivasan*, K.Lurdhu

More information

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 129 Infinity LC system Application Note Food and Beverages Author Srividya Kailasam

More information

INTERNATIONAL JOURNAL OF RESEARCH IN PHARMACY AND LIFE SCIENCES

INTERNATIONAL JOURNAL OF RESEARCH IN PHARMACY AND LIFE SCIENCES A. Chenthilnathan et al IJRPLS, 2014, 2(2): 185-190 Research Article Available online at www.pharmaresearchlibrary.com/ijrpls ISSN: 2321-5038 INTERNATIONAL JOURNAL OF RESEARCH IN PHARMACY AND LIFE SCIENCES

More information

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine pplication Note Small Molecule Pharmaceuticals, Generics ccurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine igh-resolution sampling D-LC uthors Rongjie Fu and Xinlei

More information

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace_Rev3 Introducing

More information

Journal of Atoms and Molecules

Journal of Atoms and Molecules Research Article Journal of Atoms and Molecules An International Online Journal ISSN 2277 1247 A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF RUXOLITINIB IN FORMULATIONS P.V.V. Satyanarayana*, Alavala

More information

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis Automated Switching Between D-LC and Comprehensive D-LC Analysis The Agilent 90 Infinity D-LC Solution Technical Overview Author Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany Abstract This

More information

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Application Note Doping Control Authors Prof. Maria Kristina Parr Freie Universität Berlin Institute

More information

Sulfotepp impurities in Chlorpyrifos EC formulations

Sulfotepp impurities in Chlorpyrifos EC formulations Page 1 of 16 Method DAS-AM-01-058 Sulfotepp impurities in Chlorpyrifos EC formulations A. ABSTRACT Method DAS-AM-01-058 has been validated for the analysis of the relevant impurity sulfotepp in chlorpyrifos

More information