Advanced technics in the. (TIC/TOC/TNb)

Size: px
Start display at page:

Download "Advanced technics in the. (TIC/TOC/TNb)"

Transcription

1 Advanced technics in the determination of sum parameters (TI/T/Tb) Andreas Martens, Elementar Analysensysteme Gmb, anau, Germany БЕОГРАД,

2 What is a sum parameter?

3 Definition: sum parameter A composite parameter achieved by the determination of a general characteristic (x) of various chemical compounds (x i ) as sum without specification of a single compound. i i =1 x i = x 1 + x x i

4 sum parameters...are commonly used in water and waste water analyses for the determination of its contamination with organic compounds. Typical sum parameters are: BD x - biological oxygen demand D - chemical oxygen demand T - total organic carbon

5 sum parameters BD x and B are indirect methods to determine the amount of organic carbon by the determination of the biological and chemical oxygen demand during the decomposition of its compounds. Disadvantage: slow insufficient or wrong oxidation depending from biological activity heavy metal containing reagents no automation

6 sum parameters BD 5 - biological oxygen demand after 5 days DI DI Disadvantage: particulates are not sufficiently oxidized Reference: Fonds der chemischen Industrie - Umweltbereich Wasser

7 sum parameters D - chemical oxygen demand DI Disadvantage: higher or fluctuating results, if dichromate is consumed due to the oxidation of sulfur containing compounds Reference: Fonds der chemischen Industrie - Umweltbereich Wasser

8 sum parameters E European tandard Water analysis - Guidlines for the determination of the total organic carbon (T) and dissolved organic carbon (D) The E 1484 approves the determination of T and D as appropriate method to determine the contamination of water with organic compounds.

9 sum parameters E European tandard Determination ti of T and D by: Wet chemical - UV-persulfat method ombustion method

10 Determination of T and D Wet chemical - UV-persulfat method Advantage: large sample volume (up to 20mL) high sensitivity Disadvantage: insufficient oxidation of particulates not applicable for electrolytes not for highly contaminated samples not for solid samples

11 Determination of T and D ombustion method Advantage: quantitative oxidation of all carbon containing compounds low matrix influence applicable for waste water and solids Disadvantage: lower injection volume (up to 3mL) problematically with salts lower sensitivity

12 onclusion The combustion method has the largest potential to be state of the art, when the disadvantages can be minimized.

13 T E European tandard - total carbon TI - total inorganic carbon ( 2 3 ) T - total organic carbon (+ -, -, - ) TI + T = T (additive determination) T = T - TI (difference method) Direct method: T = (T - TI) after external acidification

14 E European tandard T - total organic carbon P - non-purgable organic carbon P (V) - purgable organic carbon T = P + P or TI + P + P = T P = T -TI -P P = 0 P = T

15 Additive determination TI + T = T or TI + P + P = T liqui T II liqui i T trace

16 P direct (after external acidification) o P 850 o salts & ashes arrier Gas

17 TI/P - determination 2 TI >800 o P P >800 o salts & ashes arrier Gas

18 TI/P/ P - determination o P RT o or higher salts & ashes arrier Gas

19 Dynamic 2-zone-furnace Determination ee of TI and T in solids without acidification by the temperature ramp method.

20 Difference method T = T - TI or P = T - TI - P vario T cube

21 vario T cube arrier Gas Flow TI/T analysis

22 vario T cube Acid Dosing TI/T analysis

23 Flushing with sample vario T cube TI/T analysis

24 Dosing (TI-Measurement) vario T cube TI/T analysis

25 TI-Measurement vario T cube TI/T analysis

26 vario T cube T-Measurement TI/T analysis

27 vario T cube TI/T-Waste TI/T analysis

28 Flushing with sample vario T cube P analysis

29 vario T cube combines for the very first time 110 years of experience in elemental analysis and more than 30 years of experience in T analysis

30 vario T cube TE PWER F EXPERIEE ets new standards in T analysis measuring range for T from 2 ppb to ppm optional up to 50mg abs. in solids Makes auto-dilution unnessesary.

31 vario T cube TE PWER F EXPERIEE ets new standards in T analysis with new detectors free of spectral peak area (arbit tr. units) R 2 = 0, ,05 0,1 0,15 0,2 0,25 0,3 0,35 0,4 0,45 T [mg/l] interferences, e.g. of 2 and with extraordinary linearity over the entire detection range.

32 vario T cube TE PWER F EXPERIEE ets new standards in T analysis Also available for T b determination: LD 0.2ppm 200ppm ED 0.2ppm 500ppm ew DIR detectors with outstanding performance in T b determination. Working range: 0.05ppm 50,000ppm

33 vario T cube TE PWER F EXPERIEE ets new standards in T analysis with its integrated automatic sample feeder for liquids

34 vario T cube TE PWER F EXPERIEE ets new standards in T analysis available with 120 positions for up to 300 mg 80 positions for up to 1 g also applicable for liquids and suspensions by capsule sealing technique with its revolutionary automated analysis of solids

35 vario T cube TE PWER F EXPERIEE ets new standards in T analysis with its unique reusable ash/salt crucible for complete matrix separation, e.g. no restrictions on salt content in liquids and solids

36 vario T cube liquid id solid samples samples ombustion tubes

37 vario T cube ombustion tube for solids ombustion tube for liquids ombustion tubes

38 vario T cube TE PWER F EXPERIEE ets new standards in T analysis with no restrictions on the particle size in liquids o need any more for inaccurate suspension method

39 vario T cube TE PWER F EXPERIEE Applications from liquids to... reagent water drinking water waste water... sea water saline solutions electrolytes...

40 vario T cube TE PWER F EXPERIEE... to solids particles suspensions slurries... soils residues solids...

41 vario T cube TE PWER F EXPERIEE Approved methods, once again improved with ID-3 * counter gas membrane drying chemical drying condensation * 3-step intensive drying system

42 vario T cube TE PWER F EXPERIEE omprehensive math data processing tatistics Manual and automatic peak integration Base line correction Peak substraction

43 vario T cube TE PWER F EXPERIEE Automatic ti alibration by fixed volume and different conc. Dilution series Automatic outlier elimination

44 vario T cube TE PWER F EXPERIEE Automatic alibration by fixed concentration and different volumes auto-dilution-effect without dilution errors Multi-point-calibration from a single standard solution

45 vario T cube TE PWER F EXPERIEE Analysis of well water for animals (liquid samples, particles) Method ample T [mg/l] mean [%] well water 76,520 RD (absolute) 0,271 RD (relative) % 0,354

46 vario T cube TE PWER F EXPERIEE Analysis of hard digestible silicon carbid compounds (ib) Method ample T [mg/l] mean [%] ib 30,82 RD (absolute) 0,0683 RD (relative) % 0,0261 vario T cube shows excellent oxidation behaviour even in case of hard digestable compounds like ib ceramics with decomposition temperaturesres of approx o.

47 vario T cube TE PWER F EXPERIEE Analysis of sea water (liquid samples, salt) Method ample T [mg/l] T measurement of n = 50 sea water samples Method ample T [mg/l] T measurement of n = 50 sea water samples. External sample acidification was performed under stirring in the vials of the automatic sample feeder, followed by automatic stripping of 2 with synthetic air. mean [%] orth ea 24,282 mean [%] orth ea 12,209 RD (absolute) 0,160 RD (absolute) 0,074 Excellent for the analysis of brine solutions

48 vario T cube TE PWER F EXPERIEE Analysis of river sediments (solid sample) Method Analysis of 25 dried river sediments (Elbe/Germany) measured in solid mode Method Analysis of 25 dried river sediments (unte/germany) measured in solid mode n T [%] n T [%] mean [%] 25 5,396 mean [%] 25 4,202 RD (absolute) 0,088 RD (absolute) 0,070 Reference value (variomax) 5,35 % Reference value (variomax) 4,22 %

49 vario T cube TE PWER F EXPERIEE TE FAT I Analytical method: atalytic high temperature digestion tandards: I 8245, EPA 415.1, European standard acc. to E 1484, EV 12260, European Pharmacopoeia 6.0, vol.1 (20244) 2008, U..Pharmacopoeia-F, UP ; DI Measured parameters: T; T, TI/P; D; P optional T b Digestion temperature: free selectable up to 1200 o, standard operating temperature in dependency on the chosen catalyst between o

50 vario T cube TE PWER F EXPERIEE TE FAT II Measuring range (ppm): : 0 60,000 ppm (cube) alternatively high% : % (up to 50mg abs. ) Detection limit: : 0 25,000 ppm (trace) : ppm(ld); 500 ppm(e) : 0 50,000 ppm (DIR) : ± 6 µg/l D (cube) : ± 3 µg/l D (trace) : ± 0.02 mg/l D (DIR) Precision: i <1% at > 5 mg/l /L

51 vario T cube TE PWER F EXPERIEE TE FAT III Injection volume (liquid-mode): ample weight (solid-mode): ml (cube) ml (trace) organic substance mg or up to 1 g soil (depends on the sample matrix) Particel size: Duration of analysis: liquid mode: inner tube diameter 0.8 mm solid mode: free selectable in dependency on the chosen capsule size 2-3 min. for T (solids) 3-4 min. per parameter (liquids)

52 vario T cube TE PWER F EXPERIEE TE FAT IV Gas: synthetic air (hydrocarbon 0.1 mg L -1 and 2 1 mg L -1 ) or oxygen (grade 4.5) Gas flow: 200 ml min min -1 at 1 bar pre-pressure pressure Dimensions: Weight: 55 x 42 x 55 cm (L x W x ) basic unit 55 x 42 x 72 cm (L x W x ) basic unit with autosampler for liquids approx. 60 kg Power supply: 100/110/200/230 V, 50/60 z, 1.6 kw

53 vario T cube TE PWER F EXPERIEE Instrument control and data processing peration and control under Windows XP or Windows Vista. All instrument functions are digitally controllable, the comprehensive operation software includes e.g. automatic leak test, extensive fault diagnostics, monitoring of maintenance cycles, sleep- / awake function, statistical evaluation and almost unlimited memory capacity for storage of analytical data and graphics. Integration to data network and LIM. The possibility of remote control and diagnosis via the internet. In full compliance with 21 FR Part 11 (option).

54 ELEMETAR Analysensysteme Gmb Thank you for your attention. ХВАЛА!

55

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple Reliable TOC measurements Stop throwing away your first rep. QbD1200 has virtually eliminated sample-to-sample carryover

More information

The art of elemental analysis

The art of elemental analysis vario MACRO CUBE The art of elemental analysis 100 years ago, in the center of Germany, near Frankfurt, the advanced technology company Heraeus developed and manufactured the world s first elemental analyzer

More information

ECS8020 ORGANIC ELEMENTAL ANALYZER CHNS-O Analyzer

ECS8020 ORGANIC ELEMENTAL ANALYZER CHNS-O Analyzer ECS8020 ORGANIC ELEMENTAL ANALYZER CHNS-O Analyzer ECS 8020 CHNS-O Analyzer Organic Elemental Analysis ECS 8020 is a C-H-N-S-O Elemental Analyzer Model based on the Dumas combustion method. ECS 8020 is

More information

SKALAR. your partner in chemistry automation. Formacs series Total Organic Carbon & Total Nitrogen Analyzers

SKALAR. your partner in chemistry automation. Formacs series Total Organic Carbon & Total Nitrogen Analyzers SKALAR your partner in chemistry automation Formacs series Total Organic Carbon & Total Nitrogen Analyzers Skalar s Formacs HT TOC / TN Analyzer The Formacs HT TOC/TN analyzers provide fast, reliable analysis

More information

Total Organic Carbon Analysis of Solid Samples for Environmental and Quality Control Applications

Total Organic Carbon Analysis of Solid Samples for Environmental and Quality Control Applications Application Note 35140710 Total Organic Carbon Analysis of Solid Samples for Environmental and Quality Control Applications Keywords 1030S Solids Module Aurora 1030 TOC Analyzer High Temperature Catalytic

More information

Total Organic Carbon, Total Nitrogen

Total Organic Carbon, Total Nitrogen Thermalllox TOC--TN The modern alternative to COD and TKN Thermal oxidation with unrivalled sensitivity Aqueous matrices with a solids option TOC, TC, TIC, NPOC, POC, TN b Total Organic Carbon: Total Carbon:

More information

Basic Digestion Principles

Basic Digestion Principles Basic Digestion Principles 1 From Samples to Solutions Direct Analytical Method Solid Sample Problems: Mech. Sample Preparation (Grinding, Sieving, Weighing, Pressing, Polishing,...) Solid Sample Autosampler

More information

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) 1.0 SCOPE AND APPLICATION 1.1 Method 7060 is an atomic absorption procedure approved for determining the concentration of arsenic in wastes,

More information

2. Pharmaceutical industry

2. Pharmaceutical industry 2. Pharmaceutical industry Pharmaceutics is an ancient science that has supported people with remedies to help alleviate pain and heal illnesses. After medication, certain substances are expected to unfold

More information

Innovative. Intuitive. Concentration (ppb) R = Versatile. Hydra II C Mercury Analyzer

Innovative. Intuitive. Concentration (ppb) R = Versatile. Hydra II C Mercury Analyzer Innovative 0.01 0.1 1 10 Intuitive Concentration (ppb) R = 0.99999 Versatile Hydra II C Mercury Analyzer At Teledyne Leeman Labs; atomic spectroscopy is our business our only business. We are industry

More information

3. Chemical industry. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement

3. Chemical industry. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement 3. Chemical industry The most commonly used compound in the chemical industry is water not only as a solvent in processing, but also as an energy carrier in the cooling or heating cycle. As vast amounts

More information

AOAC Official Method 2016.xx. Determination of Total Sulfur in Fertilizers by High Temperature Combustion

AOAC Official Method 2016.xx. Determination of Total Sulfur in Fertilizers by High Temperature Combustion AOAC Official Method 2016.xx Determination of Total Sulfur in Fertilizers by High Temperature Combustion Proposed First Action 2015 (Applicable for measuring total sulfur concentration in solid and liquid

More information

Hydra-C Application Note: 1076

Hydra-C Application Note: 1076 Hydra-C Application Note: 1076 The Determination of Mercury in Samples by U.S. EPA SOW 846 Method 7473 using the Hydra-C Mercury Analyzer Introduction The accurate determination of mercury in various samples

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical Oxygen Demand, Chemical DOC316.53.01103 USEPA Reactor Digestion Method Method 10211 1 to 60 mg/l COD (ULR) TNTplus 820 Scope and application: For wastewater, process water, surface water, and cooling water.

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical Oxygen Demand, Chemical DOC316.53.01104 USEPA Reactor Digestion Method Method 10212 250 to 15,000 mg/l COD (UHR) TNTplus 823 Scope and application: For wastewater and process waters; digestion is required.

More information

The. Family of Mercury Analyzers. Hydra II AA Atomic Absorption Detection Liquid Samples

The. Family of Mercury Analyzers. Hydra II AA Atomic Absorption Detection Liquid Samples The Hydra II Family of Mercury Analyzers At Teledyne Leeman Labs, atomic spectroscopy is our business our only business. We are industry leading innovators with a proven track record providing systems

More information

3. Chemical industry. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement

3. Chemical industry. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement 3. Chemical industry The most commonly used compound in the chemical industry is water not only as a solvent in processing, but also as an energy carrier in the cooling or heating cycle. As vast amounts

More information

DETERMINATION OF CHEMICAL OXYGEN DEMAND COD

DETERMINATION OF CHEMICAL OXYGEN DEMAND COD COMENIUS BILATERAL PARTENARIATE 2012/2014 WATER POLLUTION IN URBAN AREAS: ANALYSIS AND TREATMENT IstitutoTecnico E. Fermi Via San Pelajo, 37 31100 TREVISO ITALY Falu Gymnasium Haraldsbo Kopparvägen, 1

More information

Chemical Oxidation Oxidizing agents

Chemical Oxidation Oxidizing agents Chemical Oxidation CENG 4710 Environmental Control Chemical oxidation is used to detoxify waste by adding an oxidizing agent to chemically transform waste compounds. It is capable of destroying a wide

More information

TRL-TOC-M MODEL MODULAR TC/IC/TOC/TN ANALYZER FOR LIQUID AND SOLID SAMPLES. TRL-TOC-M Model TC/IC/TOC/TN Analyzers are designed to handle:

TRL-TOC-M MODEL MODULAR TC/IC/TOC/TN ANALYZER FOR LIQUID AND SOLID SAMPLES. TRL-TOC-M Model TC/IC/TOC/TN Analyzers are designed to handle: TRL-TOC-M MODEL MODULAR TC/IC/TOC/TN ANALYZER FOR LIQUID AND SOLID SAMPLES TRL-TOC-M Model TC/IC/TOC/TN Analyzers are designed to handle: > Heterogeneous samples like sludge, wood, soil, fertilizers, agricultural

More information

Quality Control Procedures for Graphite Furnace AAS using Avanta Software

Quality Control Procedures for Graphite Furnace AAS using Avanta Software Application Note Quality Control Procedures for Graphite Furnace AAS using Avanta Software Introduction With the arrival of quality standards in the analytical laboratory, an ever increasing emphasis is

More information

ICP-OES Application Note Number 35

ICP-OES Application Note Number 35 ICP-OES Application Note Number 35 Rapid measurement of major, minor and trace levels in soils using the Varian 730-ES Vincent Calderon Varian, Inc. Introduction As part of the global strategy for sustainable

More information

Mercury Stack Monitor SM-4

Mercury Stack Monitor SM-4 Member of the envea Group Mercury Stack Monitor SM-4 Automatic continuous emission monitor (CEM) for mercury Continuous operation Sample dilution directly at stack - works with virtually any sample matrix

More information

Aquagent. The Scharlau comprehensive pyridine-free solutions range for a reliable Karl Fischer Titration

Aquagent. The Scharlau comprehensive pyridine-free solutions range for a reliable Karl Fischer Titration The Scharlau comprehensive pyridine-free solutions range for a reliable Karl Fischer Titration Karl Fischer titration is a well known and globally accepted method for water determination since the beginning

More information

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples A P P L I C A T I O N N ot e Atomic Absorption Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples Introduction The Flow Injection Mercury System (FIMS) is a dedicated system

More information

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Elemental analysis of river sediment using the Agilent 4200 MP-AES Elemental analysis of river sediment using the Agilent 4200 MP-AES Application note Environmental: Soils, sludges & sediments Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction

More information

Fusion UV/Persulfate TOC Analyzer

Fusion UV/Persulfate TOC Analyzer www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Fusion UV/Persulfate TOC Analyzer The Fusion Total Organic Carbon (TOC) Analyzer

More information

ISEmax CAM40/CAS40. Technical Information

ISEmax CAM40/CAS40. Technical Information Technical Information Online measurement of nutrient parameters Ion-selective electrode system for the continuous measurement of ammonium and nitrate Application The ion-selective electrode system works

More information

Simplifying the Process: Automated USP 643 / EP

Simplifying the Process: Automated USP 643 / EP Simplifying the Process: Automated USP 643 / EP 2.2.44 Purified Water and Water For Injection Testing Using A Next Generation TOC Analyzer Application Note By: Brian Wallace and Stephen Lawson Background

More information

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS Application Note Food Testing Author John Cauduro Agilent Technologies, Australia Introduction Trace

More information

SKALAR. your partner in chemistry automation. Primacs SN Protein-Nitrogen Analyzer

SKALAR. your partner in chemistry automation. Primacs SN Protein-Nitrogen Analyzer SKALAR your partner in chemistry automation Primacs SN Protein-Nitrogen Analyzer Applications Skalar has many years of experience developing applications for a variety of industries. Successful integration

More information

Fusion UV/Persulfate TOC Analyzer

Fusion UV/Persulfate TOC Analyzer Fusion UV/Persulfate TOC Analyzer The Fusion Total Organic Carbon (TOC) Analyzer utilizes powerful Ultra Violet (UV) Persulfate oxidation allowing superior carbon liberation from even the most challenging

More information

SKALAR. your partner in chemistry automation. Primacs SNC Total Carbon and Total Nitrogen Analyzer

SKALAR. your partner in chemistry automation. Primacs SNC Total Carbon and Total Nitrogen Analyzer SKALAR your partner in chemistry automation Primacs SNC Total Carbon and Total Nitrogen Analyzer Applications Measuring Carbon and Nitrogen levels is important in soil, plant, animal feed, food samples,

More information

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD) Application Note 19010803 A Newly Approved ASTM Standard For Analysis of in Using a Pulsed Flame Photometric Detector (PFPD) Keywords ASTM Standard D4735-96 FPD Flame Photometric Detector Model 5380 PFPD

More information

Member of the envea Group. Automatic Mercury Analyzer for the Laboratory

Member of the envea Group. Automatic Mercury Analyzer for the Laboratory Member of the envea Group AULA-254 Gold Automatic Mercury Analyzer for the Laboratory Fully automated flow analysis system High performance continuous flow analysis (CFA) technique Cold Vapor Atomic Absorption

More information

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction 1. Application 1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction Parameters and Codes: EDB and DBCP, whole water recoverable, O-3120-90 Parameter (µg/l)

More information

PRINCIPLE OF ICP- AES

PRINCIPLE OF ICP- AES INTRODUCTION Non- flame atomic emission techniques, which use electrothermal means to atomize and excite the analyte, include inductively coupled plasma and arc spark. It has been 30 years since Inductively

More information

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM IC Application Work AW CH6-1048-012011 Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM Branch: Chemical industry; Water, wastewater, environmental protection,

More information

Analysis of domestic sludge using the Agilent 4200 MP-AES

Analysis of domestic sludge using the Agilent 4200 MP-AES Analysis of domestic sludge using the Agilent 4200 MP-AES Application note Environmental Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction Managing the treatment and disposal

More information

costech instruments ECS 4010 Elemental Combustion System CHNS-O

costech instruments ECS 4010 Elemental Combustion System CHNS-O instruments ECS 4010 Elemental Combustion System CHNS-O costech Costech Analytical Tecnologies Inc. 26074 Avenue Hall, Suite 14 Valencia, CA 91355 - USA Phone: (661) 297-2395 Fax: (661) 297-5492 e-mail:

More information

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure Troy, NY 12180 Date: 12/12/2017 Page 1 of 7 USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure 1. Scope and Application 1.1 Analytes Dissolved inorganic

More information

Adsorption of Humic acid on Powdered Activated Carbon (PAC)

Adsorption of Humic acid on Powdered Activated Carbon (PAC) Adsorption of Humic acid on Powdered Activated Carbon (PAC) Department of Civil and Environmental Engineering, MSU, East Lansing, MI, 48824, USA Abstract Removal capacity and rate of Humic Acid (HA) onto

More information

Chapter 3 Electrochemical methods of Analysis-Potentiometry

Chapter 3 Electrochemical methods of Analysis-Potentiometry Chapter 3 Electrochemical methods of Analysis-Potentiometry Electroanalytical chemistry Contents Introduction Galvanic and electrolytic cells Salt bridge Electrode potential and cell potential Indicator

More information

110 Lowell Rd Hudson, NH USA Tel:

110 Lowell Rd Hudson, NH USA Tel: 110 Lowell Rd Hudson, NH 03051 USA Tel: 603-886-8400 www.teledyneleemanlabs.com Product Descriptions and Offerings Teledyne Leeman Labs is a leading innovator of analytical instrumentation for elemental

More information

LCTech GmbH Chromatographie

LCTech GmbH Chromatographie EVA III - an analytical preparation system for all kinds of applications a new system by LCTech that makes work so much easier Overview of Contents: y Broad Spectrum of Application y Automatisation of

More information

881 Compact IC pro / 882 Compact IC plus for ion chromatography. Intelligent. Reliable. Precise.

881 Compact IC pro / 882 Compact IC plus for ion chromatography. Intelligent. Reliable. Precise. Ion analysis 881 Compact IC pro / 882 Compact IC plus for ion chromatography. Intelligent. Reliable. Precise. 881 Compact IC pro and 882 Compact IC plus. Intelligent. Reliable. Precise. Metrohm s top ion

More information

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

Tex-620-J, Determining Chloride and Sulfate Contents in Soil Contents in Soil Contents: Section 1 Overview...2 Section 2 Sample Preparation...3 Section 3 Ion Chromatography Method...5 Section 4 Wet Chemical Method...9 Section 5 Archived Versions...15 Texas Department

More information

Mag-Bind Soil DNA Kit. M preps M preps M preps

Mag-Bind Soil DNA Kit. M preps M preps M preps Mag-Bind Soil DNA Kit M5635-00 5 preps M5635-01 50 preps M5635-02 200 preps January 2013 Mag-Bind Soil DNA Kit Table of Contents Introduction and Overview...2 Kit Contents/Storage and Stability...3 Preparing

More information

2. Pharmaceutical industry

2. Pharmaceutical industry 2. Pharmaceutical industry 2. Pharmaceutical industry 2.1. TOC determination in ultrapure water comparison of the various oxidation techniques 2.2. TOC determination in cleaning validation final Rinse

More information

Product Descriptions and Offerings

Product Descriptions and Offerings Product Descriptions and Offerings Teledyne Leeman Labs is a leading innovator of analytical instrumentation for elemental analysis. Laboratories in industries ranging from environmental science, oil and

More information

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC) METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC) 1.0 SCOPE AND APPLICATION 1.1 Method 7196 is used to determine the concentration of dissolved hexavalent chromium [Cr(VI)] in EP/TCLP characteristic extracts

More information

TOXI-SCREENING KIT. Microbiotest for ultra-rapid on-site. toxicity screening of water STANDARD OPERATIONAL PROCEDURE

TOXI-SCREENING KIT. Microbiotest for ultra-rapid on-site. toxicity screening of water STANDARD OPERATIONAL PROCEDURE TOXI-SCREENING KIT Microbiotest for ultra-rapid on-site toxicity screening of water STANDARD OPERATIONAL PROCEDURE TABLE OF CONTENTS Introduction to the Toxi-Screening microbiotest 2 - Origin 2 - Biological

More information

Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry

Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry NOTICE: This method has been developed by the EURL-CEFAO using the facilities available in its laboratories.

More information

Sampling, Storage and Pre-Treatment Techniques

Sampling, Storage and Pre-Treatment Techniques 1. Sampling Protocol Sample needs to be representative of the body of water (or other matrix) from where it originates. Sampling Considerations A. Location B. Frequency (hourly, daily) C. Spatial and temporal

More information

novaa 800 D Atomic Absorption Spectrometer

novaa 800 D Atomic Absorption Spectrometer Technical Data Atomic Absorption Spectrometer Cpt : +27 (0) 21 905 0476 Jhb : +27 (0) 11 794 Dbn : +27 (0) 31 266 2454 1/7 General The is a compact atomic absorption spectrometer with deuterium background

More information

Thermo Scientific Electrochemistry Products

Thermo Scientific Electrochemistry Products Comparing Performance of DO Methods The world leader in serving science in Water and Seawaters Thermo Scientific Electrochemistry Products Dissolved Oxygen Measurement Oxygen is essential for most of the

More information

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Chemistry Instrumental Analysis Lecture 27. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 27 Gas Chromatography Introduction GC covers all chromatographic methods in which the mobile phase is gas. It may involve either a solid stationary phase (GSC)

More information

lems. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement task.

lems. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement task. TOC the Chemical Industry TOC Measurement Application in Handbook Chemical Industry 3. Chemical industry The most commonly used compound in the chemical industry is water not only as a solvent in processing,

More information

Carbon (Organic, and Inorganic)

Carbon (Organic, and Inorganic) DATE: January 2011 Carbon (Organic, and Inorganic) 1. Application This method covers the determination of total organic carbon (TOC), and inorganic carbon (IC) concentrations in soil, plant tissues and

More information

Removal of suspended and dissolved organic solids

Removal of suspended and dissolved organic solids Removal of suspended and dissolved organic solids Types of dissolved solids The dissolved solids are of both organic and inorganic types. A number of methods have been investigated for the removal of inorganic

More information

NOTICE: This method: The laboratories have to study the instrumental conditions appropriate for their own instrumentation.

NOTICE: This method: The laboratories have to study the instrumental conditions appropriate for their own instrumentation. Determination of Cadmium and Lead in powdered infant formula by Electrothermal Atomization Atomic Absorption Spectrometry with Zeeman Effect Correction NOTICE: This method: - has to be considered only

More information

Product Information. Silicon Carbide Well Plates

Product Information. Silicon Carbide Well Plates Silicon Carbide Well Plates Silicon carbide (SiC) is the perfect material to ensure temperature homogeneity in rectangular multiwell reactors in a microwave field. The high microwave absorbance and excellent

More information

Low Level Phosphorus: A Method Comparison Study

Low Level Phosphorus: A Method Comparison Study Low Level Phosphorus: A Method Comparison Study Katie Adams* Stephanie Le, Theresa McBride, Tom Pearson EPA Region 10 Laboratory August 2012 8/29/2012 1 U.S. Environmental Protection Agency Goal: 1 µg/l

More information

Procedure for the Determination of Permanganate Oxidizable Carbon

Procedure for the Determination of Permanganate Oxidizable Carbon Procedure for the Determination of Permanganate Oxidizable Carbon Overview: Steve Culman, Mark Freeman, Sieglinde Snapp Kellogg Biological Station, Michigan State University, Hickory Corners, MI, 49060

More information

On-line XRF analysis of multiple elements in liquid process streams C-QUAND

On-line XRF analysis of multiple elements in liquid process streams C-QUAND An ideal tool for multiple element analysis using on-line XRF technology, from low ppm to percentage levels C-QUAND On-line XRF analysis of multiple elements in liquid process streams Introduction The

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

Examining the Role of TOC Analyzers in the Pharmaceutical Laboratory

Examining the Role of TOC Analyzers in the Pharmaceutical Laboratory Shim adzu Scientific Instrum ents, 7102 Riverwood Drive, Columbia, MD 21046 Tel: 800.477.1227/410.381.1227; Fax: 410.381.1222; www.ssi.shimadzu.com Examining the Role of TOC Analyzers in the Pharmaceutical

More information

HPLC Winter Webinars Part 2: Sample Preparation for HPLC

HPLC Winter Webinars Part 2: Sample Preparation for HPLC HPLC Winter Webinars Part 2: Sample Preparation for HPLC Jon Bardsley, Application Chemist Thermo Fisher Scientific, Runcorn/UK The world leader in serving science What am I Going to Talk About? What do

More information

BUTYL ALCOHOL in urine by GC-FID Code GC05510

BUTYL ALCOHOL in urine by GC-FID Code GC05510 BUTYL ALCOHOL in urine by GC-FID Code GC00 BIOCHEMISTRY Butanol (or n-butyl alcohol) is an alcohol which at room temperature is presented as a colorless liquid smelling of alcohol. It is an inflammatory

More information

Elemental Scientific. Spike Recovery. Elemental Scientific

Elemental Scientific. Spike Recovery. Elemental Scientific Elemental Scientific prepfast ICP-MS: USP Oral Drugs by Dissolution Automated Inline Calibration and Sample Dilution Authors: Kyle Uhlmeyer and Paul Field (field@icpms.com) Brief The prepfast

More information

ADVANCED SEPARATION TECHNOLOGY APPLICATION FOR NOM REMOVAL FROM A FRESHWATER SUPPLY

ADVANCED SEPARATION TECHNOLOGY APPLICATION FOR NOM REMOVAL FROM A FRESHWATER SUPPLY Costa Mesa, July 27, 2011 -, July 29, 2011 ADVANCED SEPARATION TECHNOLOGY APPLICATION FOR NOM REMOVAL FROM A FRESHWATER SUPPLY Andrea G. Capodaglio,, Arianna Callegari and Philippe Sauvignet 650th Anniversary

More information

Analysis of BTEX in Natural Water with SPME

Analysis of BTEX in Natural Water with SPME Analysis of BTEX in Natural Water with SPME Application Note Environmental Author Giordano Vassalli Sezione Protezione Aria Introduction Benzene, toluene, ethylbenzene and xylene (BTEX) isomers are monocyclic

More information

13. CARBON AND CARBONATE ANALYSES, LEG 9

13. CARBON AND CARBONATE ANALYSES, LEG 9 . CARBON AND CARBONATE ANALYSES, LEG R. E. Boyce and G. W. Bode, Scripps Institute of Oceanography The purpose of these analyses is to determine the amount of acid-soluble and insoluble carbon in the sediment

More information

Drillox Technology. Technology Covered by Patent Application

Drillox Technology. Technology Covered by Patent Application Drillox Technology Technology Covered by Patent Application Presentation Outline Calcium Peroxide Properties The Drillox Technology Principle Comparison with other oxidants Testing protocol Lab studies

More information

DPD Test N Tube Method *

DPD Test N Tube Method * CHLORINE, FREE (0 to 5.00 mg/l) DPD Test N Tube Method * Method 10102 For water, wastewater, and seawater 1. Enter the stored program number for Test N Tube free chlorine (Cl 2 ). Press: PRGM The display

More information

An Innovative Field-Based Analytical Method For Low-Level Detection of Chlorinated Solvents in Groundwater and Soil Samples

An Innovative Field-Based Analytical Method For Low-Level Detection of Chlorinated Solvents in Groundwater and Soil Samples An Innovative Field-Based Analytical Method For Low-Level Detection of Chlorinated Solvents in Groundwater and Soil Samples Abstract Color-Tec is a field-based analytical method developed by Ecology and

More information

Manganese III Digestion Method * (with chloride removal)

Manganese III Digestion Method * (with chloride removal) Method 10067 OXYGEN DEMAND, CHEMICAL (20 to 1,000 mg/l) For water and wastewater Manganese III Digestion Method * (with chloride removal) 1. Enter the stored program number for Manganese III COD. Press:

More information

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT METHOD 9040B ph ELECTROMETRIC MEASUREMENT 1.0 SCOPE AND APPLICATION 1.1 Method 9040 is used to measure the ph of aqueous wastes and those multiphase wastes where the aqueous phase constitutes at least

More information

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Evrim Kilicgedik Product Specialist, Atomic Spectroscopy Agilent Technologies 04.11.2011 2011 The Atomic Spectroscopy

More information

CHEMICAL OXIDATION. The use of oxidizing agents without the need of microorganisms for the reactions to proceed

CHEMICAL OXIDATION. The use of oxidizing agents without the need of microorganisms for the reactions to proceed CHEMICAL OXIDATION The use of oxidizing agents without the need of microorganisms for the reactions to proceed oxidizing agents : O 3, H 2 O 2, Cl 2 or HOCl or O 2 etc catalysts : ph, transition metals,

More information

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Method 1061 Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Company Method 1061 Revision 1. December 015 Organic Carbon in Finished

More information

Kjeltec 2300 Analyzer Unit

Kjeltec 2300 Analyzer Unit Kjeltec 2300 Analyzer Unit Automatic distillation and approved colorimetric titration for analysis of Kjeldahl nitrogen/protein, ammonium- nitrogen, TKN, Direct distillation (DD), volatile acids/ bases

More information

C80. Calvet Calorimeter From ambient to 300 C. A trademark of KEP Technologies group

C80. Calvet Calorimeter From ambient to 300 C. A trademark of KEP Technologies group C80 Calvet Calorimeter From ambient to 300 C A trademark of KEP Technologies group C80 SENSOR The C80 Calorimeter is one of the most powerful, yet flexible calorimeters available. The high precision Calvet

More information

ISO 8245 INTERNATIONAL STANDARD. Water quality Guidelines for the determination of total organic carbon (TOC) and dissolved organic carbon (DOC)

ISO 8245 INTERNATIONAL STANDARD. Water quality Guidelines for the determination of total organic carbon (TOC) and dissolved organic carbon (DOC) INTERNATIONAL STANDARD ISO 8245 Second edition 1999-03-01 Water quality Guidelines for the determination of total organic carbon (TOC) and dissolved organic carbon (DOC) Qualité de l'eau Lignes directrices

More information

S2 PICOFOX. Innovation with Integrity. Spectrometry Solutions TXRF

S2 PICOFOX. Innovation with Integrity. Spectrometry Solutions TXRF S2 PICOFOX Spectrometry Solutions Innovation with Integrity TXRF S2 PICOFOX True Trace Analysis with XRF for the First Time! You need to know the concentration of trace elements in environmental samples?

More information

Product Description and Specification

Product Description and Specification Product Description and Specification Catalog No 7 Synthra [ 11 C]Choline Description Synthra [ 11 C]Choline is a completely automated synthesis system for routine production of [ 11 C]choline. Using in-target

More information

NH 3 -selective electrodes

NH 3 -selective electrodes NH 3 -selective electrodes Manual 8.109.8031EN Metrohm AG CH-9100 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com NH 3 -selective electrodes Manual 8.109.8031EN

More information

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed Date 2017.

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed Date 2017. Final Report Title Characterisation of Job No 2016/11/12-34 AS No. 1234A Client Contact Sample Author report Signed Date 2017 Easy Reach Report 2017 v2.docx 1 of 33 Contents 1. Study Summary Page 3 2.

More information

Thermo Scientific Orion Fluoride Ion Selective Electrodes The standard in fluoride ion analysis EPA compliant

Thermo Scientific Orion Fluoride Ion Selective Electrodes The standard in fluoride ion analysis EPA compliant Thermo Scientific Orion Fluoride Ion Selective Electrodes The standard in fluoride ion analysis EPA compliant Approved ASTM Method for Fluoride in Drinking Water and Wastewater Analyze free fluoride ions

More information

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure.

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure. High performance liquid chromatography (HPLC) is a much more sensitive and useful technique than paper and thin layer chromatography. The instrument used for HPLC is called a high performance liquid chromatograph.

More information

Atomic Absorption Spectroscopy

Atomic Absorption Spectroscopy CH 2252 Instrumental Methods of Analysis Unit IV Atomic Absorption Spectroscopy Dr. M. Subramanian Associate Professor Department of Chemical Engineering Sri Sivasubramaniya Nadar College of Engineering

More information

ACETONE IN URINE BY UV CODE Z42010

ACETONE IN URINE BY UV CODE Z42010 ACETONE IN URINE BY UV CODE Z42010 BIOCHEMISTRY Acetone is a colorless, mobile, flammable liquid. It is readly soluble in water, ethanol, ether etc., and itself serves as an important solvent. Acetone

More information

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry APPLICATION NOTE Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry AN43227 Antonella Guzzonato 1, Shona McSheehy Ducos

More information

AULA 254 AULA Fluorescence

AULA 254 AULA Fluorescence Automatic Mercury Analyzer for the Laboratory MERCURY INSTRUMENTS ANALYTICAL TECH NOLO GIES AULA 254 AULA Fluorescence Fully automated mercury analysis system High performance flow analysis technique Two

More information

This immunoassay kit allows for the in vitro quantitative determination of Aflatoxin M1 concentrations in milk, milk power.

This immunoassay kit allows for the in vitro quantitative determination of Aflatoxin M1 concentrations in milk, milk power. Aflatoxin M1 (AFM1) ELISA Kit This immunoassay kit allows for the in vitro quantitative determination of Aflatoxin M1 concentrations in milk, milk power. This package insert must be read in its entirety

More information

Elemental Scientific. Stability. Elemental Scientific

Elemental Scientific. Stability. Elemental Scientific Elemental Scientific prepfast ICP-MS: USP Oral Drugs by Dilution Automated Inline Calibration and Sample Dilution Authors: Kyle Uhlmeyer and Paul Field (field@icpms.com) Brief The prepfast inline

More information

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC TECHNICAL NOTE 122 Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC Authors Terri Christison and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA,

More information

Instrumental Chemical Analysis. Dr. Abdul Muttaleb Jaber Professor Faculty of Pharmacy Philadelphia University Fall 2012/2013

Instrumental Chemical Analysis. Dr. Abdul Muttaleb Jaber Professor Faculty of Pharmacy Philadelphia University Fall 2012/2013 0510212 Instrumental Chemical Analysis Dr. Abdul Muttaleb Jaber Professor Faculty of Pharmacy Philadelphia University Fall 2012/2013 Chapter 1 Electroanalytical Methods Electroanalytical Chemistry Electroanalytical

More information

Soil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006

Soil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006 Soil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006 Introduction: Capillary electrophoresis (CE) is a relatively new, but rapidly growing separation

More information

NukEx Nucleic Acid Release Reagent

NukEx Nucleic Acid Release Reagent Instruction for Use NukEx Nucleic Acid Release Reagent For general laboratory use. For in vitro use only. Reagent for the enzymatic release of nucleic acid from tissue samples, ticks, insects and swabs.

More information