Portal Monitor Characterization for Internally and Externally Deposited Radionuclides

Size: px
Start display at page:

Download "Portal Monitor Characterization for Internally and Externally Deposited Radionuclides"

Transcription

1 Operational Topic Detection efficiencies for internally and externally deposited radionuclides have been evaluated using a pass-through and static mode portal monitor. Portal Monitor Characterization for Internally and Externally Deposited Radionuclides Matthew Carey, Adam Kryskow, Fred Straccia, and Mark Tries* Abstract: In this evaluation, both the internal and external radionuclide detection efficiencies for a portal monitor were evaluated as a function of photon energy using an anthropomorphic phantom. Pass-through and static measurements were completed using 241 Am, 57 Co, 133 Ba, 137 Cs, 60 Co, 109 Cd, and 54 Mn in various locations both external and internal to the phantom. Other parameters, such as single detector uniformity, total detector uniformity, background linearity, and activity linearity have been analyzed. It was found that the minimum detectable activity for internally deposited 137 Cs in the abdomen was approximately ten times higher for pass-through versus static measurements. Additionally, it was found that the minimum detectable activity for 137 Cs in the abdominal region for both internal and external pass-through scenarios are nearly equivalent. In general, if the expected radionuclide source term is primarily non-transuranic, the pass-through mode offers sufficient sensitivity to identify potential overexposures while providing much greater personnel throughput. However, minimum detectable committed effective doses for transuranics such as 241 Am, show potential for personnel over exposure if the radionuclide mixture contains a significant fraction of transuranics. It is therefore recommended that nuclear facilities evaluate their radionuclide source term in *University of Massachusetts, Lowell, Radiological Sciences, University Avenue, MA Theauthorsdeclarenoconflictsofinterest. order to bound potential personnel doses. Health Phys. 107(Supplement 3):S188 S197; 2014 Key words: operational topics; contamination, internal; counting efficiency; detection limits INTRODUCTION Photon radiation portal monitors are commonly used as an element of radiation protection programs at nuclear facilities worldwide. These radiation monitors are primarily used for personnel monitoring of external contamination, but also have the ability to detect internally deposited radionuclides. Understanding the sensitivity of the radiation portal monitor to both externally and internally deposited radioactivity of single radionuclides or of radionuclide mixtures is an important element in assessing the potential total effective dose equivalent (TEDE) to facility personnel. The Mirion, (5000 Highlands Parkway Suite 150 Smyrna, GA USA) Fast-Track-Fibre (FTF) portal monitor was characterized Matthew Glen Carey is a health physicist at Radiation Safety & Control Services, Inc. Previously, he was a graduate student at the University of Massachusetts Lowell completing his Master s thesis on portal monitor characterization. This research involved Monte Carlo N-Particle (MCNP) radiation transport simulations, point kernel efficiency calculations, physical source measurements, and the comparison thereof. He specializes in using MCNP for solving complex health physics radiation transport problems such as shielding design, activation analysis, and detector response. His address is Matthew_Carey@student.uml.edu. for response to internal and externally deposited radionuclides using an anthropomorphic phantom, as suggested by the portal monitor requirements of the American Nuclear Insurers (ANI) Engineering Inspection Criteria (ANI-EIC) Section (ANI 2008). Detection efficiencies were measured with sources placed at internal and external positions on a phantom, using both static (stationary) and pass-through portal monitor detection modes. The FTF has proprietary algorithms that were not investigated, such as false alarm probability filters, time resolved algorithms, detection of fluctuating background conditions, and handling of special scenarios by pattern recognition. The efficiencies of multiple radionuclides with widely varying photon energies were evaluated in order to determine the FTF s energy response curve. This response curve, in terms of efficiency as a function of energy, allowed for the calculation of the minimum detectable activity (MDA), minimum detectable intake (MDI), and minimum detectable committed effective dose equivalent (MD- CEDE) per alarm for any photon emitting radionuclide mixture as described in Section and Section in ANI-EIC. The FTF was also analyzed for activity linearity, background linearity, single detector uniformity, and total detector uniformity as S188 November 2014

2 The Radiation Safety Journal Vol. 107, suppl 3 November 2014 described in Sections and in ANI-EIC. MATERIALS AND METHODS Portal monitor The FTF portal monitor contains 14 polyvinyl toluene (PVT) detectors with fiber optic coupling to two multichannel photomultiplier units (Mirion Technologies 2011). These PVT detectors run along the sides, bottom, and top of the device. The12sidedetectorsareeach85cm in height, 15 cm in width, 5 cm deep, and include 2.54 cm of lead shielding towards the exterior of the portal monitor, to reduce the background and potential cross-talk between multiple portal monitors if positioned side by side. The bottom and top detectors are 40 cm in width and length and 5 cm deep. The response from these detectors is grouped into eight channels leading to two multichannel photomultiplier units. For all static measurements recorded, except for those contained in the Single Detector Uniformity section, the responses were summed into a single system or total response. Radioactive sources Efficiency versus energy functions were acquired through the use of seven nuclides; five of which, 241 Am, 57 Co, 133 Ba, 137 Cs, 60 Co, are photon test standards recommended in Section 4.6 of N42.35 (ANSI 2006). Additionally, 109 Cd and 54 Mn were used to fill in gaps in the energy spectrum. The sources, with their percentage uncertainty and sigma values, are summarized in Table 1. Anthropomorphic phantom The Lawrence Livermore National Labs (LLNL) phantom was used for the both the internal static and pass-through measurements, as well as the external static measurements. An image of the LLNL phantom positioned inside the FTF Table 1. NIST sources used for evaluation. Nuclide Serial Number Corrected Activity (kbq) Half-life (y) 241 Am ± 3.0% (2.3σ) ± 0.68 (1σ) 109 Cd ± 3.0% (2.3σ) ± (1σ) 57 Co ± 3.0% (2.3σ) ± (1σ) 133 Ba ± 3.0% (2.3σ) ± 0.04 (1σ) 137 Cs ± 3.1% (2.3σ) ± 0.16 (1σ) 137 Cs ± 3.1% (2.3σ) ± 0.16 (1σ) 137 Cs ± 3.1% (2.3σ) ± 0.16 (1σ) 54 Mn ± 3.0% (2.3σ) ± (1σ) 60 Co ± 3.1% (2.3σ) ± (1σ) is shown in Fig. 1. Each source position shown in Fig. 1 is described in Table 2, per ANI-EIC Section The portal monitor was evaluated for internal sensitivity per ANI-EIC Section as well as external sensitivity. Prior to taking measurements, a two point calibration was completed using the 39 kbq 137 Cs source. This process was necessary for the external passthrough analysis described in the Internal and External Efficiencies section below. Static and passthrough measurements were completed for scenarios with an empty portal monitor (i.e., in-air measurements), and with sources positioned inside and outside a LLNL phantom. Due to the LLNL torso phantom not having extremities, external sources were placed on a human during pass-through measurements. Static data were recorded using the FTF s Wait-in mode, which allows for timed static measurements. Pass-through data were recorded using FTF s Fast-Track mode. Detector measurements were used to create photon energy response curves and calculate minimum detectable quantities. Detector linearity Activity linearity response. Section and Section in ANI-EIC requires counting rate to increase linearly with increasing source activity. This linearity was evaluated using three 137 Cs sources with activities of 39, 9.5, and 3.8 kbq as described in Table 1. Background linearity. Background levels were artificially increased using multiple sources positioned around the outside of the portal monitor. Background exposure rates were measured in the center of the FTF using a calibrated Ludlum, (501 Oak St, Sweetwater, TX 79556) 3 97 detector. The total gross counting rate of the portal monitor was evaluated for >linearity as a function of background exposure rate and then used to determine a conversion factor between counts per second (cps) and μr h 1. Effects of background counting rate on lower level discriminator. The lower limit of detection was calculated as a function of background counting rate using a 100 s background count time, 10 ssource count time, and detection probabilities of zα = and zβ =3.09. Single detector uniformity. In order to measure the uniformity of a single detector, a side detector was removed and shielded, such that a source placed on the detector face would not add net counts to any other detector. With this approach, net counts on the detector channel were measured to determine detector uniformity over the detector face. Because the top and bottom detectors are on a dedicated photomultiplier channel, uniformity measurements only required background subtraction. Total detector uniformity. The total response of the system was evaluated at multiple locations, at a 10 cm distance from the inner portal monitor side walls. These measurements were taken from 20 to +20 cm in 10 cm intervals, Operational Radiation Safety S189

3 M. Carey et al. Portal monitor characterization FIG. 1. LLNL phantom within the FTF portal monitor. where 0 cm was taken to be the middle of the portal monitor (i.e., where a person would stand during a static measurement). Vertical measurements were taken at 20 cm intervals from the floor of the portal monitor to a height of 180 cm. Internal and external efficiencies Source locations for internal and external static and passthrough measurements are described in Table 2. The depths shown are given with respect to the surface tissue at each individual location. The width gives the lateral displacement of the source from the phantoms centerline. Right and left are taken from the phantom s frame of reference. A human was used in place of the LLNL phantom for elbow, knee, and foot measurements, as the phantom did not include these extremities. For pass-through measurement, when the FTF alarms, the device displays personnel transit speed (m s 1 ) and maximum and minimum expected activity but no counting data. Therefore, a roundabout method was created to determine the pass-through efficiency of the portal monitor as a function of personnel speed. The initial calibration with 137 Cs was used as the average total detector efficiency. When a pass-through alarm was triggered using the same 137 Cs source, the average of the maximum and minimum expected activities was recorded along with the estimated personnel speed. The average activity was then multiplied by the average detector efficiency for 137 Cs to calculate the expected net counting rate. This methodology was repeated 30 times for each measurement location and fit to a linear regression for counting rate vs. personnel speed. Minimum detectable activities were calculated under the assumption that the average alarm was equal to the alarm threshold. This procedure may be applied to other nuclides with similar photon energies to 137 Cs for external photon efficiency by using a ratio of the efficiency of 137 Cs pass-through to the 137 Cs static measurement. Efficiency calculations The efficiency of a nuclide with a half-life much greater than the counting interval was calculated experimentally in units of counts per nuclear transformation by dividing the net counting S190 November 2014

4 The Radiation Safety Journal Vol. 107, suppl 3 November 2014 Table 2. Source Locations within and on the LLNL Phantom. Region Height (cm) Width (cm) Depth (cm) Position Index Static Internal Left Lung Right Lung Thyroid GI Tract Static External Chest Head Gluteus Knee a Pass-through Internal Abdomen Pass-through External Chest Knee a Abdomen Foot a Elbow a a Used human in place of LLNL phantom. rate by the decay-corrected source activity A as: ε nuc transðcounts=decayþ ¼ R s A : (1) Most photon emitting nuclides emit multiple photon emissions per transformation, the yieldweighted average nuclide energy (also termed effective energy) was calculated by: E nuc ¼ ie i Y i i Y i : (2) The total yield of a nuclide was calculated as the sum of the photons emitted per transformation: Y nuc ¼ Y i : (3) i The effective nuclide efficiency in counts per photon was calculated by: ε nuc photonðcounts=photonþ¼ iε i Y i : (4) i Y i Minimum detectable quantities The efficiency of the portal monitor for different nuclides allowed for the calculation of the MDA, MDI, MD-CEDE per alarm, and MD-DAC-hr per alarm for all nuclides tested. The critical level (L C ) used to evaluate measurements for activity above background was established using a type I (false positive) error probability of five percent. This equates to a normal deviate z α of 1.645, which is used in the equation below (Currie 1968): sffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi L C ¼ z α R b þ R b ; T g (5) T b where T g is the sample counting interval (in this case the personnel counting interval) and T b is the background counting interval. The lower limit of detection (L D ), the smallest net signal that can be detected reliably was established using a type II (false negative) error probability of 0.10%. This equates to a corresponding normal deviate z β of 3.09 which was used in the equation below: L D ¼ Z β 2 þ L C þ 1 2T g 2 vffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi Z 4 Z 2 Z 2 u β β β T 2 þ 4LC þ 4 L 2 T g C g Z 2 α t : (6) The personnel counting interval, T g was equal to 10 s for this evaluation and the background counting interval was 100 s. The MDA for a given nuclide was calculated experimentally as the lower limit of detection divided by the counting efficiency in units of counts per nuclear transformation: MDA nuc ¼ L D ε nuc trans : (7) The MDA for a nuclide mixture was given by: L D MDA mix ¼ ; (8) i f j ε nuc trans j where f j is the nuclide activity fraction for the j th nuclide in the mixture. To determine the MDI for a given nuclide, it was assumed that the detection of the intake will be discovered in less than a half day following an acute inhalation or ingestion intake. The relationship between MDA and MDI was expressed using an intake retention function (IRF)asshownineqn(9): MDI nuc ¼ MDA nuc IRF : (9) For purposes of this evaluation, the delay post intake was assumed to be 0.2 d (4.8 h) and an aerodynamic median activity diameter (AMAD) of1micrometerwasused. NUREG/CR 4884, which is based on ICRP 30 methodology, was used to determine the IRF values for nuclides in the lungs, nasal passages, andgitract(u.s.nrc1987;icrp 1978). The IRF for ingestion was taken as For a mixture of nuclides the MDI was calculated by: MDI mix ¼ MDA mix IRF : (10) The MD-CEDE was calculated from the MDI mix and the most limiting annual limit on intake (ALI) from 10 CFR 20, Appendix B (U. S. NRC 2012). Solubility classes for nuclides 133 Ba and 137 Cs were taken as Type D; 241 Am, 109 Cd, and 54 Mn as Type W; and 57 Co and 60 Co as Type Y. The resultant MD-CEDE represents the minimum internal dose from an acute intake of a photon emitting nuclide mixture (in units of msv) that would correspond to a portal monitor alarm: Operational Radiation Safety S191

5 M. Carey et al. Portal monitor characterization Background linearity. A linear fit (with the y-intercept set to zero) was applied to the total detector counting rates versus background exposure rate data shown in Fig. 3. These data indicate some slight deviations from linearity however, these deviations were considered acceptable (correlation coefficient of ) and are most likely attributable to the nonuniformity of the artificial background field that was created. A conversion factor between total detector counting rate and a uniform background exposure rate was determined as: Background μr hr ¼ R b 1652 ; (13) FIG. 2. Portal monitor response linearity. D MD-CEDE mix ¼ MDI mix ALI fi ALI j j ; (11) where f j is the nuclide activity fraction of the mix, and D ALI is 50 msv for SALI and 500 msv for NALI. The minimum detectable derived air concentration (MD-DAC) in DAC-hrs was calculated as: 2000DAC-hr MD-DAC-hr mix ¼ MDI mix f j SALI j j : (12) The MD-CEDE and MDDAChr for single nuclides are easily inferred from eqn (11) and eqn (12) by setting j equal to one. RESULTS AND DISCUSSION The initial performance parameters of the FTF portal monitor are described in the Detector Linearity section. Detector efficiencies for static and pass-through modes are presented in the following sections for both internal and externally deposited radionuclides. Minimum detectable quantity results such as MDA, MDI, MD-CEDE and MD-DAC-hr are discussed below. rate versus the source activity, as shown in Fig. 2. An unweighted linear function (with the y-intercept set to zero) was fit to the data. The resultant correlation coefficient of indicated the detector was behaving as expected. where 1,652 is the slope of the linear regression line found in Fig. 3. This slope is only accurate for the sources and geometry used to generate the high background field in this experiment. Single detector uniformity. Detector uniformity for the side, top, and bottom type detector is shown in Fig. 4 and Fig. 5. As expected, the detector response was greatest in the center, were the solid angle subtended was the greatest. Total detector response. The measured total detection uniformity in Fig. 6 shows larger normalized values in the center of the horizontal axis, with multiple larger regions along the vertical axis. The increase in response towards the center of the horizontal axis was expected due to the increased total solid angle subtended. The reduced efficiency around 100 cm in height was also expected due to the existence of an air gap between the Detector linearity Activity linearity response. The linearity of the detector was evaluated by plotting the net counting FIG. 3. Detector counting rate as a function of background. S192 November 2014

6 The Radiation Safety Journal Vol. 107, suppl 3 November 2014 Internal static efficiencies Static measurement efficiencies for various internal locations are shown in Fig. 7. Regression parameters for the thyroid, right lung, GI tract and left lung static internal efficiencies have been fit to third order polynomials given in eqns (14) through (17); with correlation coefficients of , , , and , respectively: ε Thyroid ¼ 0:09349 E 3 0:3151 E 2 þ 0:3461E 0:01044 (14) ε Right Lung¼ 0:08009E 3 0:2613E 2 þ 0:2880E 0: (15) ε GI Tract ¼ 0:005043E3 0:08066E 2 þ 0:1657E 0: (16) ε Left Lung¼ 0:08286E 3 0:2820E 2 þ 0:3160E 0:01052: (17) The efficiency curves for the internal efficiency measurements show higher overall efficiencies for the thyroid region, with the GI tract having the lowest efficiency. This is expected as the thyroid has less attenuating material between the source and the detectors, while the GI tract has the most attenuating material, thus the exiting photon has a lower probability of being detected. It should be noted that the left lung efficiencies are slightly higher than the right lung due to the left lung source depth being 5 cm, while the right lung was positioned deeper, at 8 cm. FIG. 4. Normalized single side detector uniformity. upperand lowersidedetectors.it is assumed that Fig. 6 would appear vertically symmetric had measurements been taken at the top of the detector (roughly 200 cm). External static efficiencies Theefficiencyresultsareshown as a function of effective photon energy in Fig. 8. The polynomial regression results for the external static efficiencies are contained in eqns (18) through (21); with correlation coefficients of , , , and , respectively: Operational Radiation Safety S193

7 M. Carey et al. Portal monitor characterization lower efficiency. Two trials were completed for the source on the right knee. A higher efficiency is found when the source leg steps into the portal monitor first, as expected due to the temporary lack of motion as weight is shifted onto the leading leg. FIG. 5. Normalized bottom detector uniformity. Minimum detectable quantities Effects of background counting rate on lower level discriminator. L D as a function of background counting rate is shown in Table 4 and shown in Fig. 10. As one can see,theshapeofthecurveinfig.10 approximates a square root function, as expected from eqns (5) and (6). ε Chest ¼ 0:03021 E 3 0:1146E 2 þ 0:1684E 0:00312(18) ε Gluteus ¼ 0:00255 E 3 0:0822E 2 þ 0:1746E 0:00443 (19) ε Head ¼ 0:02791E 3 0:1473E 2 þ 0:2275E 0:00223 (20) ε Knee ¼ 0:04469 E 3 0:2357E 2 þ 0:3192E 0:01014: (21) Internal pass-through efficiencies Measurement of the internal pass-through efficiency was completed for a source positioned within the abdomen of the LLNL phantom, with the results shown in Fig. 9. Due to the variance in the efficiency results and the uncertainty in entry velocity, a linear regression of the data would result in a very low linear regression correlation coefficient of Therefore, there is no appreciable correlation found between internal pass-through efficiency and entrance velocity. However, the average of the data in Fig. 9 gives an average entry velocity of 0.9 m per second, and an average detector efficiency of 0.8%. External pass-through efficiencies External pass-through efficiency data for each body part from linear regressions of each data set are shown in Table 3. As expected, faster transit times led to a FIG. 6. Normalized measured detector uniformity from detector center. S194 November 2014

8 The Radiation Safety Journal Vol. 107, suppl 3 November 2014 The expected MD-DAC-hr for each nuclide is calculated using eqn (12). This calculation was performed for an inhalation scenario the results of which are shown in Table 8. FIG. 7. Internal static efficiencies. External static minimum detectable activity. The MDA for external source positions was calculated for various locations for a background counting rate of 22,000 cps. These results are contained in Table 9 for the various nuclides used. Internal pass-through minimum detectable activity. An estimate of the MDA for internal pass-through of 137 Cs in the abdomen can be calculated using theaverageofthedatashownin Fig. 9. Using a 1 m per second entry velocity and the minimum efficiency of 0.6% yields an MDA for 137 Cs in the abdomen is found to be Bq. FIG. 8. External static efficiencies. Internal static minimum detectable activity. The average measured internal efficiency for each nuclide was used to calculate the MDA of each nuclide for an average background of 22,000 cps. These results are contained in Table 5 using a false positive probability of 5% and a false negative probability of 0.10% as stated previously. The expected MD-CEDE for the likely solubility class for each nuclide was calculated using eqn (11). For these calculations the left and right lung MDA values were averaged and the thyroid measurements assumed to be the nasal passage (due to proximity). MD-CEDE values are contained for either an inhalation or ingestion scenario in Table 6 and Table 7. The lowest MD-CEDE for each nuclide are bolded in Table 6. FIG. 9. Internal pass-through efficiency of abdomen. External pass-through minimum detectable activity. Measurement of the efficiency as a function of entrance velocity was used to calculate the MDA for 137 Cs for pass-through monitoring. These results are shown for various body locations at different velocities in Table 10. Operational Radiation Safety S195

9 M. Carey et al. Portal monitor characterization Table 3. External pass-through efficiency vs. entrance velocity. Efficiency at given velocity Source location 0.5 (m s 1 ) 1 (m s 1 ) 1.5 (m s 1 ) 2 (m s 1 ) Chest 1.3% 1.2% 1.0% 0.9% Source on right knee, left in first 1.4% 1.4% 1.3% 1.3% Source on right knee, right in first 1.8% 1.8% 1.8% 1.8% Abdomen 1.5% 1.3% 1.2% 1.0% Left foot 2.0% 1.8% 1.5% 1.2% Elbow 2.2% 2.0% 1.7% 1.5% Table 7. Minimum detectable committed effective dose (Sv) from ingestion. Nuclide GI tract 241 Am Cd Co Ba Cs Mn Co Table 4. Background total counting rate vs. lower limit of detection. Background counting rate (cps) Expected background (μr hr 1 ) FIG. 10. Lower limit of detection vs. total background counting rate. Lower limit of detection (cps) 22, , ,075 30, ,161 35, ,241 39, ,316 44, ,386 Table 5. Minimum detectable activity (Bq). Nuclide Right lung Left lung GI Tract Thyroid 241 Am Cd Co Ba Cs Mn Co Table 6. Minimum detectable committed effective dose (Sv) from inhalation. Nuclide Lungs GI Tract Nasal 241 Am Cd Co Ba Cs Mn Co CONCLUSION The FTF portal monitor was found to be operating as expected in terms of activity and background linearity. Uniformity measurements of both the single and total detector uniformity resulted in responses that were mostly due to distance and the solid angle subtended by the detectors, as expected. The FTF was found to have an average static inhalation MDI of approximately Bq for most nuclides (excluding 241 Am) for a measurement of 10 s. This inhalation MDI would result in MD-CEDE values on the order of tens to hundreds of μsv, which is well below regulatory limits. However, for the case of 241 Am, the MD-CEDE for ingestion and especially inhalation is particularly high and therefore shows that nuclides 241 Am can t be sufficiently measured by this technique, which poses a risk for overexposure. Internal static mode measurements found MDA values on the order of 10 4 Bq. External static mode measurements found MDA values for 137 Cs around Bq. The internal pass-through MDA for 137 Cs was found to be approximately Bq. External pass-through MDA results for 137 Cs are on the order of 10 4 Bq and 10 5 Bq for velocities of 0.5 and 2 m s 1,respectively. While the static mode is appreciably more sensitive compared to the pass-through mode, pass-through monitoring may be an acceptable practice to employ for both internal and external source terms. This is, of course, highly dependent on the nuclide mixture and deposition location. S196 November 2014

10 The Radiation Safety Journal Vol. 107, suppl 3 November 2014 Table 8. Minimum DAC-hr from inhalation. Nuclide Right lung Left lung GI tract Thyroid 241 Am Cd Co Ba Cs Mn Co Table 9. External static MDA (Bq) for various source positions. Nuclide Chest Head Gluteus Knee 241 Am Cd Co Ba Cs Mn Co Table 10. Pass-through external surface MDA vs. velocity and source position. Pass-through external surface 0.5 (m s 1 ) 1 (m s 1 ) 1.5 (m s 1 ) 2 (m s 1 ) Chest Source of right knee, left in 1st Source on right knee, right in 1st Abdominal Left foot Elbow In conclusion, understanding the response of portal monitors to internally deposited radioactivity can be an important tool in allowing radiation protection personnel to bound potential internal 137 Cs MDA (Bq) at given velocity doses (Carey et al. 2012). This information could be applied in litigation support; or in evaluation of potential missed doses if unmonitored internal exposures occurred in a facility. Acknowledgments This study was funded in part by the University of Massachusetts Lowell, Radiation Safety & Control Services, Inc., and Mirion Technologies HP Division. REFERENCES American Nuclear Insurers. Engineering inspection criteria for nuclear liability insurance. Section 8.5 Revision 5. West Harford, CT: American Nuclear Insurers; CareyM,DaroisE,StracciaF.Thesensitivity of gamma portal monitors to personnel intakes of radioactivity.ouray:sandrikwadepublishing; Nuclear Decommissioning Report; Mirion Technologies. CheckPoint: Gate and FastTrack Fibre Pedestrian Monitor: Technical handbook. Hamburg, Germany: Mirion Technologies; American National Standards Institute. American National Standard for evaluation and performance of radiation detection portal monitors for use in homeland security; New York: IEEE; N ; Currie LA. Limits for quantative detection and quantitative deternatino. Analytical Chem 40: ; U.S. Nuclear Regulatory Commission. Interpretation of bioassay measurements. Washington, DC: U.S. NRC; NUREG/CR 4884; International Commission on Radiological Protection limits for intakes of radionuclides by workers. Oxford: Pergamon Press; ICRP Publication 30; U.S. Nuclear Regulatory Commission. Annual limits on intake (ALIs) and derived air concentrations (DACs) of radionuclides for occupational exposures. Washington, DC: U.S. Government Printing Office; Operational Radiation Safety S197

Sensitivity of the IRD whole-body counter for in vivo measurements in the case of accidental intakes

Sensitivity of the IRD whole-body counter for in vivo measurements in the case of accidental intakes Sensitivity of the IRD whole-body counter for in vivo measurements in the case of accidental intakes B.M. Dantas, E.A. Lucena and A.L.A. Dantas Laboratório de Monitoração In Vivo Divisão de Dosimetria

More information

Department of Energy Office of Worker Protection Programs and Hazards Management Radiological Control Technical Position RCTP 99-02

Department of Energy Office of Worker Protection Programs and Hazards Management Radiological Control Technical Position RCTP 99-02 Issue: Title 10 Code of Federal Regulations, Part 835 (10 CFR 835), Occupational Radiation Protection, specifies occupational radiation protection requirements for Department of Energy (DOE) activities

More information

CURRENT CAPABILITIES OF THE IRD-CNEN WHOLE BODY COUNTER FOR IN VIVO MONITORING OF INTERNALLY DEPOSITED RADIONUCLIDES IN HUMAN BODY

CURRENT CAPABILITIES OF THE IRD-CNEN WHOLE BODY COUNTER FOR IN VIVO MONITORING OF INTERNALLY DEPOSITED RADIONUCLIDES IN HUMAN BODY International Joint Conference RADIO 2014 Gramado, RS, Brazil, Augustl 26-29, 2014 SOCIEDADE BRASILEIRA DE PROTEÇÃO RADIOLÓGICA - SBPR CURRENT CAPABILITIES OF THE IRD-CNEN WHOLE BODY COUNTER FOR IN VIVO

More information

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the WBC

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the WBC Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the WBC A. Specification Whole body counting method is used to detect the gamma rays emitted by radio nuclides,

More information

CALIBRATION OF IN VITRO BIOASSAY METHODOLOGY FOR DETERMINATION OF 131 I IN URINE

CALIBRATION OF IN VITRO BIOASSAY METHODOLOGY FOR DETERMINATION OF 131 I IN URINE X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN

More information

Performance Characterization of A New Cam System M.J. Koskelo 1, J.C. Rodgers 2, D.C. Nelson 2, A.R. McFarland 3 and C.A. Ortiz 3

Performance Characterization of A New Cam System M.J. Koskelo 1, J.C. Rodgers 2, D.C. Nelson 2, A.R. McFarland 3 and C.A. Ortiz 3 Performance Characterization of A New Cam System M.J. Koskelo 1, J.C. Rodgers 2, D.C. Nelson 2, A.R. McFarland 3 and C.A. Ortiz 3 1 CANBERRA Industries, Meriden, CT 06450 2 Los Alamos National Laboratory,

More information

The FastScan whole body counter: efficiency as a function of BOMAB phantom size and energy modelled by MCNP

The FastScan whole body counter: efficiency as a function of BOMAB phantom size and energy modelled by MCNP The FastScan whole body counter: efficiency as a function of BOMAB phantom size and energy modelled by MCNP Gary H. Kramer and Jeannie Fung Human Monitoring Laboratory, Radiation Protection Bureau, 775

More information

Calibration of a Whole Body Counter and In Vivo measurements for Internal Dosimetry Evaluation in Chile, Two years experience.

Calibration of a Whole Body Counter and In Vivo measurements for Internal Dosimetry Evaluation in Chile, Two years experience. Calibration of a Whole Body Counter and In Vivo measurements for Internal Dosimetry Evaluation in Chile, Two years experience. Osvaldo Piñones O., Sylvia Sanhueza M. Radio medicine Section, Chilean Commission

More information

INTERNAL RADIATION DOSIMETRY

INTERNAL RADIATION DOSIMETRY INTERNAL RADIATION DOSIMETRY Introduction to Internal Dosimetry Importance of External and Internal Exposures by Radiation Type Charged particle radiation (α, p, β) Generally important only for internal

More information

UALR Radiation Safety Office

UALR Radiation Safety Office UALR Radiation Safety Office ETAS-329 501-569 8210 Graduate Institute of Technology University of Arkansas at Little Rock Regulatory Authority Nuclear Regulatory Commission (NRC) EPA, DoE, DoT, OSHA Agreement

More information

Ringhals AB. Routines for whole body counting at Ringhals NPP

Ringhals AB. Routines for whole body counting at Ringhals NPP Routines for whole body counting at Ringhals NPP Structure Equipment - Quick Scan - Whole body counter Marie Carlson M.Sc. Radiology and Dosimetry Radiation Physicist and deputy RP Controller Routines

More information

Radiological Traceability Program (RTP) Radiological and Environmental Science Laboratory (RESL) Svetlana Nour Kenneth Inn Jerry LaRosa Jackie Mann

Radiological Traceability Program (RTP) Radiological and Environmental Science Laboratory (RESL) Svetlana Nour Kenneth Inn Jerry LaRosa Jackie Mann Radiological Traceability Program (RTP) between NIST and the DOE Radiological and Environmental Science Laboratory (RESL) Svetlana Nour Kenneth Inn Jerry LaRosa Jackie Mann 21st Annual Council on Ionizing

More information

UNCORRECTED PROOF. Table of Contents

UNCORRECTED PROOF. Table of Contents 00-Stabin-Prelims SNY001-Stabin (Typeset by spi publisher services, Delhi) vii of xvi June 1, 2007 17:15 Preface xiii Acknowledgments xv Chapter 1. Introduction to Health Physics 1 1.1 Definition of Health

More information

Michael G. Stabin. Radiation Protection and Dosimetry. An Introduction to Health Physics. 4) Springer

Michael G. Stabin. Radiation Protection and Dosimetry. An Introduction to Health Physics. 4) Springer Michael G. Stabin Radiation Protection and Dosimetry An Introduction to Health Physics 4) Springer Table of Contents Preface Acknowledgments Chapter 1. Introduction to Health Physics 1 1.1 Definition of

More information

Regulatory Guide Exposure Pathways, Equations, and Input Requirements

Regulatory Guide Exposure Pathways, Equations, and Input Requirements Regulatory Guide 1.109 Exposure Pathways, Equations, and Input Requirements Ken Sejkora Entergy Nuclear Northeast Pilgrim Station Presented at the 23rd Annual RETS-REMP Workshop Training Session Westminster,

More information

A New Anthropometric Calibration Phantom for In Vivo Measurement of Bone Seeking Radionuclides

A New Anthropometric Calibration Phantom for In Vivo Measurement of Bone Seeking Radionuclides A New Anthropometric Calibration Phantom for In Vivo Measurement of Bone Seeking Radionuclides H. B. Spitz and J. C.Lodwick University of Cincinnati, Department of Mechanical, Industrial & Nuclear Engineering,

More information

RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY

RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY PURPOSE This procedure provides instructions for developing, maintaining, and documenting, radiation safety procedures conducted at the Cyclotron Radiochemistry

More information

General Regression Neural Networks for Estimating Radiation Workers Internal Dose

General Regression Neural Networks for Estimating Radiation Workers Internal Dose General Regression Neural Networks for Estimating Radiation s Internal Dose Eman Sarwat and Nadia Helal Radiation Safety Dep. ENRRA., Egyptian Nuclear and Radiological Regulatory Authority, 3, Ahmed El

More information

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved Assessment Of The Effectiveness Of Collimation Of Cs 137 Panoramic Beam On Tld Calibration Using A Constructed Lead Block Collimator And An ICRU Slab Phantom At SSDL In Ghana. C.C. Arwui 1, P. Deatanyah

More information

Applied Nuclear Science Educational, Training & Simulation Systems

Applied Nuclear Science Educational, Training & Simulation Systems WWW.NATS-USA.COM Applied Nuclear Science Educational, Training & Simulation Systems North American Technical Services Bridging Technology with the Latest in Radiation Detection Systems The Center For Innovative

More information

Determining the Need For External Radiation Monitoring at FUSRAP Projects Using Soil Characterization Data. Todd Davidson

Determining the Need For External Radiation Monitoring at FUSRAP Projects Using Soil Characterization Data. Todd Davidson Determining the Need For External Radiation Monitoring at FUSRAP Projects Using Soil Characterization Data Todd Davidson Introduction According to Regulatory Guide 8.34 Monitoring Criteria and Methods

More information

U.S. EPA Superfund Counts Per Minute (CPM) Electronic Calculator Stuart A. Walker. U.S. Environmental Protection Agency, Washington, DC.

U.S. EPA Superfund Counts Per Minute (CPM) Electronic Calculator Stuart A. Walker. U.S. Environmental Protection Agency, Washington, DC. U.S. EPA Superfund Counts Per Minute (CPM) Electronic Calculator 14336 Stuart A. Walker U.S. Environmental Protection Agency, Washington, DC. 20460 ABSTRACT The U.S. Environmental Protection Agency (EPA)

More information

Rapid Extraction of Plutonium from Urine by Pyrosulfate Fusion and

Rapid Extraction of Plutonium from Urine by Pyrosulfate Fusion and Rapid Extraction of Plutonium from Urine by Pyrosulfate Fusion and PERALS Spectroscopy R.L. Metzger, P.H. Pouquette, and G.W. Klingler Radiation Safety Engineering, Inc. Chandler, AZ Abstract To effectively

More information

Lower Bound of Optimization for the Public Considering Dose Distribution of Radiation due to Natural Background Radiation

Lower Bound of Optimization for the Public Considering Dose Distribution of Radiation due to Natural Background Radiation Lower Bound of Optimization for the Public Considering Dose Distribution of Radiation due to Natural Background Radiation Takatoshi Hattori a* a Central Research Institute of Electric Power Industry, Radiation

More information

NORM and TENORM: Occurrence, Characterizing, Handling and Disposal

NORM and TENORM: Occurrence, Characterizing, Handling and Disposal NORM and TENORM: Occurrence, Characterizing, Handling and Disposal Ionizing Radiation and Hazard Potential John R. Frazier, Ph.D. Certified Health Physicist May 12, 2014 Radiation Radiation is a word that

More information

Calibration & Use of a Capintec CAPTUS 3000 Portable Thyroid Uptake System for Iodine-125 Bioassay Measurements Todd W.

Calibration & Use of a Capintec CAPTUS 3000 Portable Thyroid Uptake System for Iodine-125 Bioassay Measurements Todd W. Calibration & Use of a Capintec CAPTUS 3000 Portable Thyroid Uptake System for Iodine-125 Bioassay Measurements Todd W. Baker, MSPH, CHP Photo image area measures 2 H x 6.93 W and can be masked by a collage

More information

Am-241 as a Metabolic Tracer for Inhaled Pu Nitrate in External Chest Counting

Am-241 as a Metabolic Tracer for Inhaled Pu Nitrate in External Chest Counting Am-24 as a Metabolic Tracer for Inhaled Pu Nitrate in External Chest Counting Nobuhito ISHIGURE, Takashi NAKANO, Hiroko ENOMOTO, Akira KOIZUMI, Harozo IIDA, Kumiko FUKUTSU, Yumi ABE 2, Yuji YAMADA, Katsuhiro

More information

Atoms, Radiation, and Radiation Protection

Atoms, Radiation, and Radiation Protection James E. Turner Atoms, Radiation, and Radiation Protection Third, Completely Revised and Enlarged Edition BICENTENNIAL J 0 1 8 0 Q 71 z m z CAVILEY 2007 1 ;Z z ü ; m r B10ENTENNIAL WILEY-VCH Verlag GmbH

More information

Simulation of Personal Protective Equipment Exposure to Radioactive Particulates. A Master s Level Submission.

Simulation of Personal Protective Equipment Exposure to Radioactive Particulates. A Master s Level Submission. Simulation of Personal Protective Equipment Exposure to Radioactive Particulates M. Roeterink 1*, E.F.G. Dickson 1, P. Bodurtha 1, and E.C. Corcoran 1 1 Royal Military College of Canada, Ontario, Canada

More information

Neutron Detector Gamma Insensitivity Criteria

Neutron Detector Gamma Insensitivity Criteria PNNL-18903 Neutron Detector Gamma Insensitivity Criteria RT Kouzes JR Ely AT Lintereur DL Stephens October 28, 2009 DISCLAIMER This report was prepared as an account of work sponsored by an agency of the

More information

WM2011 Conference, February 27 - March 3, 2011, Phoenix, AZ

WM2011 Conference, February 27 - March 3, 2011, Phoenix, AZ Methodology for Determination of Exposure Point Concentration Using both Systematic and Biased Samples for Radiological Risk and Dose Assessments 11488 Randy Hansen*, Michael Steven Passig*, Mahmudur Rahman**

More information

Neutron Dose near Spent Nuclear Fuel and HAW after the 2007 ICRP Recommendations

Neutron Dose near Spent Nuclear Fuel and HAW after the 2007 ICRP Recommendations Neutron Dose near Spent Nuclear Fuel and HAW after the 2007 ICRP Recommendations Gunter Pretzsch Gesellschaft fuer Anlagen- und Reaktorsicherheit (GRS) mbh Radiation and Environmental Protection Division

More information

Utilization of advanced clutter suppression algorithms for improved spectroscopic portal capability against radionuclide threats

Utilization of advanced clutter suppression algorithms for improved spectroscopic portal capability against radionuclide threats Physical Sciences Inc. VG13-158 Utilization of advanced clutter suppression algorithms for improved spectroscopic portal capability against radionuclide threats Bogdan R. Cosofret, Kirill Shokhirev and

More information

Higher -o-o-o- Past Paper questions o-o-o- 3.6 Radiation

Higher -o-o-o- Past Paper questions o-o-o- 3.6 Radiation Higher -o-o-o- Past Paper questions 2000-2010 -o-o-o- 3.6 Radiation 2000 Q29 Radium (Ra) decays to radon (Rn) by the emission of an alpha particle. Some energy is also released by this decay. The decay

More information

Development of Dose Conversion Coefficients for Radionuclides Produced in Spallation Neutron Sources

Development of Dose Conversion Coefficients for Radionuclides Produced in Spallation Neutron Sources Transmutation Sciences Physics (TRP) Transmutation Research Program Projects 8-11-2011 Development of Dose Conversion Coefficients for Radionuclides Produced in Spallation Neutron Sources Phillip W. Patton

More information

Minimum Detectable Activity Estimates for a Germanium-Detector Based Spectroscopic Portal Monitor

Minimum Detectable Activity Estimates for a Germanium-Detector Based Spectroscopic Portal Monitor Log 121 Minimum Detectable Activity Estimates for a Germanium-Detector Based Spectroscopic Portal Monitor Ronald M. Keyser, Frank Sergent, Timothy R. Twomey, Daniel L. Upp ORTEC 801 South Illinois Avenue

More information

External Dosimetry at NPPs; NVLAP, Noble Gas, and TEDE

External Dosimetry at NPPs; NVLAP, Noble Gas, and TEDE External Dosimetry at NPPs; NVLAP, Noble Gas, and TEDE Joseph L Danek - CHP 2018 Fall Meeting FLORIDA CHAPTER HEALTH PHYSICS SOCIETY 1 Part 1 National Voluntary Laboratory Accreditation Program (NVLAP)

More information

Radioactive Waste Management

Radioactive Waste Management International Journal of Research in Engineering and Science (IJRES) ISSN (Online): 2320-9364, ISSN (Print): 2320-9356 Volume 4 Issue 6 ǁ June. 2016 ǁ PP.67-71 Asma Osman Ibrahim Osman 1, Hamid Mohamed

More information

Evaluation and Measurements of Radioactive Air Emission and Off-Site Doses at SLAC

Evaluation and Measurements of Radioactive Air Emission and Off-Site Doses at SLAC SLAC-PUB-15365 Evaluation and Measurements of Radioactive Air Emission and Off-Site Doses at SLAC I.Chan, J.Liu, H.Tran SLAC National Accelerator Laboratory, M.S. 48, 2575 Sand Hill Road, Menlo Park, CA,

More information

COUNTING ERRORS AND STATISTICS RCT STUDY GUIDE Identify the five general types of radiation measurement errors.

COUNTING ERRORS AND STATISTICS RCT STUDY GUIDE Identify the five general types of radiation measurement errors. LEARNING OBJECTIVES: 2.03.01 Identify the five general types of radiation measurement errors. 2.03.02 Describe the effect of each source of error on radiation measurements. 2.03.03 State the two purposes

More information

Plutonium 239 Equivalency Calculations

Plutonium 239 Equivalency Calculations LLNL-TR-490356 Plutonium 239 Equivalency Calculations J. Wen July 7, 2011 Disclaimer This document was prepared as an account of work sponsored by an agency of the United States government. Neither the

More information

Uncertainty and detection limit in decontamination measurements of gamma emitting radio nuclides in unshielded environments Tom Koivuhuhta

Uncertainty and detection limit in decontamination measurements of gamma emitting radio nuclides in unshielded environments Tom Koivuhuhta Uncertainty and detection limit in decontamination measurements of gamma emitting radio nuclides in unshielded environments Tom Koivuhuhta Master of Science Thesis Supervisor: Mats Isaksson Contents Abstract...

More information

Statistical Uncertainty What, When, Why and How RETS/REMP Workshop Jim Key Key Solutions, Inc.

Statistical Uncertainty What, When, Why and How RETS/REMP Workshop Jim Key Key Solutions, Inc. Statistical Uncertainty What, When, Why and How 2005 RETS/REMP Workshop Jim Key Key Solutions, Inc. www.keysolutionsinc.com What Do I Use and When CL MDA MDC LTL LLD Detection Limits A Posteriori After

More information

JRPR. An Intercomparison of Counting Efficiency and the Performance of Two Whole-Body Counters According to the Type of Phantom.

JRPR. An Intercomparison of Counting Efficiency and the Performance of Two Whole-Body Counters According to the Type of Phantom. Journal of Radiation Protection and Research 2016;41(3):274-281 pissn 2508-1888 eissn 2466-2461 An Intercomparison of Counting Efficiency and the Performance of Two Whole-Body Counters According to the

More information

Radiation Safety In-Service House-Keeping and Security Departments. Petrone Associates LLC Specialists in Applied Medical Physics

Radiation Safety In-Service House-Keeping and Security Departments. Petrone Associates LLC Specialists in Applied Medical Physics Radiation Safety In-Service House-Keeping and Security Departments 1 MEDICAL RADIATION SOURCES Radiation Producing Machines Radioactive Materials 2 Nuclear Medicine Radiation Therapy Research Laboratories

More information

RADON EQUILIBRIUM MEASUREMENT IN THE AIR *

RADON EQUILIBRIUM MEASUREMENT IN THE AIR * RADON EQUILIBRIUM MEASUREMENT IN THE AIR * SOFIJA FORKAPIĆ, DUŠAN MRĐA, MIROSLAV VESKOVIĆ, NATAŠA TODOROVIĆ, KRISTINA BIKIT, JOVANA NIKOLOV, JAN HANSMAN University of Novi Sad, Faculty of Sciences, Department

More information

Radioactivity measurements and risk assessments in soil samples at south and middle of Qatar

Radioactivity measurements and risk assessments in soil samples at south and middle of Qatar Radioactivity measurements and risk assessments in soil samples at south and middle of Qatar A. T. Al-Kinani*, M. A. Amr**, K. A. Al-Saad**, A. I. Helal***, and M. M. Al Dosari* *Radiation and Chemical

More information

Fundamentals of Radionuclide Metrology

Fundamentals of Radionuclide Metrology Fundamentals of Radionuclide Metrology Brian E. Zimmerman, PhD Physical Measurement Laboratory National Institute of Standards and Technology Gaithersburg, MD USA SIM Metrology Workshop Buenos Aires, Argentina

More information

Intercomparison of JAERI Torso Phantom lung sets

Intercomparison of JAERI Torso Phantom lung sets Intercomparison of JAERI Torso Phantom lung sets Gary H. Kramer and Barry M. Hauck Human Monitoring Laboratory, Radiation Protection Bureau, 775 Brookfield Road, Ottawa, Ontario K1A 1C1 Canada (Gary_H_Kramer@hc-sc.gc.ca,

More information

Department of Energy OffIce of Worker Protection Programs and Hazards Management Radiological Control Technical Position RCTP

Department of Energy OffIce of Worker Protection Programs and Hazards Management Radiological Control Technical Position RCTP OffIce of Worker Protection Programs and Hazards Management Issue: The occupational radiation protection regulations established by the (DOE) and Nuclear Regulatory Commission (NRC) are similar in many

More information

Non-Destructive Assay Applications Using Temperature-Stabilized Large Volume CeBr Detectors 14277

Non-Destructive Assay Applications Using Temperature-Stabilized Large Volume CeBr Detectors 14277 Non-Destructive Assay Applications Using Temperature-Stabilized Large Volume CeBr Detectors 14277 D. Nakazawa and F. Bronson Canberra Industries Inc., 800 Research Parkway, Meriden, CT, 06450, USA. ABSTRACT

More information

THE USE OF A THYROID UPTAKE SYSTEM FOR ASSAYING INTERNAL CONTAMINATION FOLLOWING A RADIOACTIVE DISPERSAL EVENT

THE USE OF A THYROID UPTAKE SYSTEM FOR ASSAYING INTERNAL CONTAMINATION FOLLOWING A RADIOACTIVE DISPERSAL EVENT THE USE OF A THYROID UPTAKE SYSTEM FOR ASSAYING INTERNAL CONTAMINATION FOLLOWING A RADIOACTIVE DISPERSAL EVENT A Thesis Presented to The Academic Faculty By Sarah Brashear Scarboro In Partial Fulfillment

More information

AP1000 European 15. Accident Analyses Design Control Document EVALUATION MODELS AND PARAMETERS FOR ANALYSIS OF RADIOLOGICAL CONSEQUENCES OF ACCIDENTS

AP1000 European 15. Accident Analyses Design Control Document EVALUATION MODELS AND PARAMETERS FOR ANALYSIS OF RADIOLOGICAL CONSEQUENCES OF ACCIDENTS APPENDIX 15A EVALUATION MODELS AND PARAMETERS FOR ANALYSIS OF RADIOLOGICAL CONSEQUENCES OF ACCIDENTS This appendix contains the parameters and models that form the basis of the radiological consequences

More information

10 CFR 50 Appendix I Time for a Change?

10 CFR 50 Appendix I Time for a Change? 10 CFR 50 Appendix I Time for a Change? RETS-REMP Workshop Savannah, GA 25-Jun-2014 By Richard Conatser, NRC Background NRC regulations use: ICRP-2 Recommendations, 1959 (Appendix I) ICRP-26 Recommendations,

More information

Radiation Glossary. Radioactive material dispersed in the air in the form of dusts, fumes, particulates, mists, vapors, or gases.

Radiation Glossary. Radioactive material dispersed in the air in the form of dusts, fumes, particulates, mists, vapors, or gases. Activity The rate of disintegration (transformation) or decay of radioactive material. The units of activity are Curie (Ci) and the Becquerel (Bq). Agreement State Any state with which the U.S. Nuclear

More information

DOSE CONVERSION FACTORS. David C. Kocher and Keith F. Eckerman Health and Safety Research Division Oak Ridge National Laboratory

DOSE CONVERSION FACTORS. David C. Kocher and Keith F. Eckerman Health and Safety Research Division Oak Ridge National Laboratory CONP-9203121 2 DOSE CONVERSION FACTORS DE92 011036 David C. Kocher and Keith F. Eckerman Health and Safety Research Division Oak Ridge National Laboratory Pathway Analysis and Risk Assessment for Environmental

More information

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL A. Specification Gamma-spectrometry method is used to identify and determine the activity concentration

More information

10 CFR Amended Text

10 CFR Amended Text Subpart A - General Provisions 835.1 Scope. (a) (b) 10 CFR 835-2007 Amended Text General. The rules in this part establish radiation protection standards, limits, and program requirements for protecting

More information

Canadian Journal of Physics. Determination of Radioactivity Levels of Salt Minerals on the Market

Canadian Journal of Physics. Determination of Radioactivity Levels of Salt Minerals on the Market Determination of Radioactivity Levels of Salt Minerals on the Market Journal: Canadian Journal of Physics Manuscript ID cjp-2017-0775.r1 Manuscript Type: Article Date Submitted by the Author: 29-Nov-2017

More information

Importance of uncertainties in dose assessment to prove compliance with radiation protection standards

Importance of uncertainties in dose assessment to prove compliance with radiation protection standards Importance of uncertainties in dose assessment to prove compliance with radiation protection standards Manfred TSCHURLOVITS, Atominstitute of Austrian Universities, Vienna University of Technology, Stadionallee

More information

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements Specific Accreditation Criteria Calibration ISO/IEC 17025 Annex Ionising radiation measurements January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected

More information

DISMANTLEMENT OF AN UNIT OF METABOLIC TREATMENTS WITH IODINE 131

DISMANTLEMENT OF AN UNIT OF METABOLIC TREATMENTS WITH IODINE 131 DISMANTLEMENT OF AN UNIT OF METABOLIC TREATMENTS WITH IODINE 131 Autores: J.M. Jiménez 1, A.R. Cortés 2, P.Alcorta 2 E-mail: jmjimenez@hsan.osakidetza.net 1 Unidad de Radiofísica. Hospital Santiago Apóstol,

More information

: Y/LB-16,056 OAK RIDGE Y-12 PLANT

: Y/LB-16,056 OAK RIDGE Y-12 PLANT Y-12 Report No.: Y/LB-16,056 OAK RIDGE Y-12 PLANT VERIFICATION OF URANIUM MASS AND ENRICHMENTS OF HIGHLY ENRICHED URANIUM (HEU) USING THE NUCLEAR MATERIALS IDENTIFICATION SYSTEM (NMIS) L. G. Chiang J.

More information

An Experimental Study of the Relative Response of Plastic Scintillators to Photons and Beta Particles

An Experimental Study of the Relative Response of Plastic Scintillators to Photons and Beta Particles An Experimental Study of the Relative Response of Plastic Scintillators to Photons and Beta Particles 1) A Kumar 1 and A.J. Waker 1 Faculty of Energy System and Nuclear Science, UOIT, 2000 Simcoe St. N,

More information

Renewed whole-body counting chamber in STUK

Renewed whole-body counting chamber in STUK Renewed whole-body counting chamber in STUK Seminar DTU Nutech, Roskilde, Denmark Tiina Torvela, Tero Karhunen, Maarit Muikku Environmental Radiation Surveillance and Emergency Preparedness Whole-body

More information

Plutonium Decorporation Following Complex Exposure: Inception

Plutonium Decorporation Following Complex Exposure: Inception Spring 2016 Grad Seminar Presentation WSU Spokane SAC 147 March 4, 2016 Plutonium Decorporation Following Complex Exposure: Inception Sara Dumit, PhD student sara.dumit@wsu.edu United States Transuranium

More information

RADIONUCLIDE INFORMATION

RADIONUCLIDE INFORMATION RADIONUCLIDE INFORMATION Below are some physical properties and practical handling information for radionuclides most commonly used in the research environment. Contact the RSO for additional information

More information

ACTIVATION ANALYSIS OF DECOMISSIONING OPERATIONS FOR RESEARCH REACTORS

ACTIVATION ANALYSIS OF DECOMISSIONING OPERATIONS FOR RESEARCH REACTORS ACTIVATION ANALYSIS OF DECOMISSIONING OPERATIONS FOR RESEARCH REACTORS Hernán G. Meier, Martín Brizuela, Alexis R. A. Maître and Felipe Albornoz INVAP S.E. Comandante Luis Piedra Buena 4950, 8400 San Carlos

More information

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA The Underwater Spectrometric System Based on CZT Detector for Survey of the Bottom of MR Reactor Pool 13461 Victor Potapov, Alexey Safronov, Oleg Ivanov, Sergey Smirnov, Vyacheslav Stepanov National Research

More information

Organ and effective dose rate coefficients for submersion exposure in occupational settings

Organ and effective dose rate coefficients for submersion exposure in occupational settings Radiat Environ Biophys (2017) 56:453 462 DOI 10.1007/s00411-017-0705-6 ORIGINAL ARTICLE Organ and effective dose rate coefficients for submersion exposure in occupational settings K. G. Veinot 1,2 S. A.

More information

RADIATION PROTECTION

RADIATION PROTECTION DEPARTMENT OF NUCLEAR TECHNOLOGY, FACULTY OF ENGINEERING CHULALONGKORN UNIVERSITY PHAYATHAI ROAD, BANGKOK 10330, THAILAND TEL: (662) 218-6772, (662) 218-6784. FAX: (662) 218-6770 E-mail: fnegbr@eng.chula.ac.th

More information

Journal of Radiation Protection and Research

Journal of Radiation Protection and Research 1) SANG BUM HONG et al.: IN SITU MEASUREMENT OF DEPTH DISTRIBUTIONS Journal of Radiation Protection and Research pissn 2508-1888 eissn 2466-2461 http://dx.doi.org/10.14407/jrpr.2016.41.2.173 Paper Received

More information

Physics Experimental Physics Temple University, Spring C. J. Martoff, Instructor

Physics Experimental Physics Temple University, Spring C. J. Martoff, Instructor Physics 4796 - Experimental Physics Temple University, Spring 2010-11 C. J. Martoff, Instructor Physics 4796 Lab Writeup Counting Statistics (or, Is it Radioactive?) 0.1 Purpose of This Lab Exercise: Demonstrate

More information

Activities of the International Commission on Radiological Protection

Activities of the International Commission on Radiological Protection Activities of the International Commission on Radiological Protection Wesley Bolch Advanced Laboratory for Radiation Dosimetry Studies (ALRADS) J. Crayton Pruitt Family Department of Biomedical Engineering

More information

Kinetic Transport Models and Minimum Detection Limits of Atmospheric Particulate Resuspension

Kinetic Transport Models and Minimum Detection Limits of Atmospheric Particulate Resuspension Kinetic Transport Models and Minimum Detection Limits of Atmospheric Particulate Resuspension Shaun Marshall 1, Charles Potter 2, David Medich 1 1 Worcester Polytechnic Institute, Worcester, MA 01609 2

More information

TECHNICAL WORKING GROUP ITWG GUIDELINE ON IN-FIELD APPLICATIONS OF HIGH- RESOLUTION GAMMA SPECTROMETRY FOR ANALYSIS OF SPECIAL NUCLEAR MATERIAL

TECHNICAL WORKING GROUP ITWG GUIDELINE ON IN-FIELD APPLICATIONS OF HIGH- RESOLUTION GAMMA SPECTROMETRY FOR ANALYSIS OF SPECIAL NUCLEAR MATERIAL NUCLE A R FORENSIC S INTERN ATION A L TECHNICAL WORKING GROUP ITWG GUIDELINE ON IN-FIELD APPLICATIONS OF HIGH- RESOLUTION GAMMA SPECTROMETRY FOR ANALYSIS OF SPECIAL NUCLEAR MATERIAL This document was designed

More information

Design and Use of an Interim Noble Gas Effluent Monitor at Columbia Generating Station

Design and Use of an Interim Noble Gas Effluent Monitor at Columbia Generating Station Design and Use of an Interim Noble Gas Effluent Monitor at Columbia Generating Station RETS-REMP & EPRI Groundwater Protection Workshop June 25-27th, 2013 Westminster, CO Eric L. Darois, CHP (RSCS) Jim

More information

Radiological impact of rutile covered welding electrodes

Radiological impact of rutile covered welding electrodes LABORATORIO DE MEDIDAS DE BAJA ACTIVIDAD Universidad del País Vasco / Euskal Herriko Unibertsitatea Radiological impact of rutile covered welding electrodes Margarita Herranz 1, Saroa Rozas 1, Carmen Pérez

More information

WM2016 Conference, March 6 10, 2016, Phoenix, Arizona, USA. Recovering New Type of Sources by Off-Site Source Recovery Project for WIPP Disposal-16604

WM2016 Conference, March 6 10, 2016, Phoenix, Arizona, USA. Recovering New Type of Sources by Off-Site Source Recovery Project for WIPP Disposal-16604 Recovering New Type of Sources by Off-Site Source Recovery Project for WIPP Disposal-16604 Ioana Witkowski, Anthony Nettleton, Alex Feldman Los Alamos National Laboratory INTRODUCTION TO OSRP ACTIVITIES

More information

A new neutron monitor for pulsed fields at high-energy accelerators

A new neutron monitor for pulsed fields at high-energy accelerators A new neutron monitor for pulsed fields at high-energy accelerators Marlies Luszik-Bhadra *, Eike Hohmann Physikalisch-Technische Bundesanstalt, Bundesallee 100, D-38116, Braunschweig, Germany. Abstract.

More information

LAGUNA VERDE NPP DAW DRUMS CLASSIFICATION

LAGUNA VERDE NPP DAW DRUMS CLASSIFICATION ABSTRACT LAGUNA VERDE NPP DAW DRUMS CLASSIFICATION M. Medrano, D. Linares, C. Rodriguez and N. Zarate Gerencia Subsede Sureste National Institute of Nuclear Research Mexico R. Ramirez Radiation Protection

More information

Committed Effective Dose from Thoron Daughters Inhalation

Committed Effective Dose from Thoron Daughters Inhalation Committed Effective Dose from Thoron Daughters Inhalation M.P. Campos and B.R.S. Pecequilo Instituto de Pesquisas Energéticas e Nucleares - Departamento de Radioproteção Ambiental Travessa R, 400 Cidade

More information

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY ABSORBED DOSE: The amount of energy absorbed, as a result of radiation passing through a material, per unit mass of material. Measured in rads (1 rad

More information

Radiation Dose Limits. Radiation Safety Training Module 3

Radiation Dose Limits. Radiation Safety Training Module 3 Module 3 This module will address Western Kentucky University radiation dose limits. Basis for WKU limits Kentucky Administrative Regulations, specifically 902 KAR 100:019 Overview Module 3, Page 2 Annual

More information

2016 No. 410 (W. 128) WATER, ENGLAND AND WALES. The Water Supply (Water Quality) (Amendment) Regulations 2016

2016 No. 410 (W. 128) WATER, ENGLAND AND WALES. The Water Supply (Water Quality) (Amendment) Regulations 2016 W E L S H S T A T U T O R Y I N S T R U M E N T S 2016 No. 410 (W. 128) WATER, ENGLAND AND WALES The Water Supply (Water Quality) (Amendment) Regulations 2016 Made - - - - 21 March 2016 Laid before Parliament

More information

Minimum Detectable and Maximum Missable Activities

Minimum Detectable and Maximum Missable Activities and Maximum Missable The HPA Approach David Williams 18 May 2005 Centre for Radiation, Chemical and Environmental Hazards Radiation Protection Division formerly the National Radiological Protection Board

More information

Investigation Of The Effects Of Variation Of Neutron Source-Detector Distance On The Emitted Neutron Dose Equivalent

Investigation Of The Effects Of Variation Of Neutron Source-Detector Distance On The Emitted Neutron Dose Equivalent ISSN: 9- Vol. Issue, June - Investigation Of The Effects Of Variation Of Neutron Source-Detector Distance On The Emitted Equivalent Igwesi, D. I. Physics and Industrial Physics Department, Faculty of Physical

More information

HALF LIFE. NJSP HMRU June 10, Student Handout CBRNE AWARENESS Module 4 1. Objectives. Student will

HALF LIFE. NJSP HMRU June 10, Student Handout CBRNE AWARENESS Module 4 1. Objectives. Student will June 10, 2004 Radiological/Nuclear Overview 1 Student will demonstrate a knowledge of self protection techniques identify types of radiation and their associated hazards demonstrate a knowledge of terminology

More information

M. Rogozina, M. Zhukovsky, A. Ekidin, M. Vasyanovich. Institute of Industrial Ecology, Ural Branch Russian Academy of Sciences

M. Rogozina, M. Zhukovsky, A. Ekidin, M. Vasyanovich. Institute of Industrial Ecology, Ural Branch Russian Academy of Sciences THORON PROGENY SIZE DISTRIBUTION IN MONAZITE STORAGE FACILITY M. Rogozina, M. Zhukovsky, A. Ekidin, M. Vasyanovich Institute of Industrial Ecology, Ural Branch Russian Academy of Sciences Thoron ( 220

More information

(9C/(9t)t = a(x,t) (92C/3x2)t + b(x,t) (9C/9 x)t + c(x,t)ct + d(x,t)

(9C/(9t)t = a(x,t) (92C/3x2)t + b(x,t) (9C/9 x)t + c(x,t)ct + d(x,t) ABSTRACT Safe management Including disposal of radioactive wastes from the various parts of the nuclear fuel cycle is an important aspect of nuclear technology development. The problem of managing radioactive

More information

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory Recent Activities on Neutron Standardization at the Electrotechnical Laboratory K. Kudo, N. Takeda, S. Koshikawa and A. Uritani Quantum Radiation Division, National Metrology Institute of Japan (NMIJ)

More information

Quantifying Electrostatic Resuspension of Radionuclides from Surface Contamination

Quantifying Electrostatic Resuspension of Radionuclides from Surface Contamination Quantifying Electrostatic Resuspension of Radionuclides from Surface Contamination Shaun Marshall 1, Charles Potter 2, David Medich 1 1 Worcester Polytechnic Institute, Worcester, MA 01609 2 Sandia National

More information

Acoustics and Ionising Radiation Formulation and Strategy. 13 November 2008 Alan DuSautoy

Acoustics and Ionising Radiation Formulation and Strategy. 13 November 2008 Alan DuSautoy Acoustics and Ionising Radiation Formulation and Strategy 13 November 2008 Alan DuSautoy Contents What is the future of Programme Formulation? What is Rolling Formulation? Programme Overview Roadmaps Future

More information

Commissioning of the Beta Secondary Standard (BSS2)

Commissioning of the Beta Secondary Standard (BSS2) Commissioning of the Beta Secondary Standard (BSS2) Speaker / Author: R.W. Thoka* Co-author: S. Jozela* * National Metrology Institute of South Africa (NMISA), Private Bag X 34, Lynnwood Ridge, Pretoria,

More information

Characterization and Monte Carlo simulations for a CLYC detector

Characterization and Monte Carlo simulations for a CLYC detector Characterization and Monte Carlo simulations for a CLYC detector A. Borella 1, E. Boogers 1, R.Rossa 1, P. Schillebeeckx 1 aborella@sckcen.be 1 SCK CEN, Belgian Nuclear Research Centre JRC-Geel, Joint

More information

Dosimetry. Sanja Dolanski Babić May, 2018.

Dosimetry. Sanja Dolanski Babić May, 2018. Dosimetry Sanja Dolanski Babić May, 2018. What s the difference between radiation and radioactivity? Radiation - the process of emitting energy as waves or particles, and the radiated energy Radioactivity

More information

11/23/2014 RADIATION AND DOSE MEASUREMENTS. Units of Radioactivity

11/23/2014 RADIATION AND DOSE MEASUREMENTS. Units of Radioactivity CHAPTER 4 RADIATION UNITS RADIATION AND DOSE MEASUREMENTS 1 Units of Radioactivity 2 1 Radiation Units There are specific units for the amount of radiation you receive in a given time and for the total

More information

Industrial Hygiene: Assessment and Control of the Occupational Environment

Industrial Hygiene: Assessment and Control of the Occupational Environment Industrial Hygiene: Assessment and Control of the Occupational Environment Main Topics Air Pollution Control Analytical Methods Ergonomics Gas and Vapour Sampling General Practice Heat and Cold Stress

More information

Activation of Air and Concrete in Medical Isotope Production Cyclotron Facilities

Activation of Air and Concrete in Medical Isotope Production Cyclotron Facilities Activation of Air and Concrete in Medical Isotope Production Cyclotron Facilities CRPA 2016, Toronto Adam Dodd Senior Project Officer Accelerators and Class II Prescribed Equipment Division (613) 993-7930

More information

DOE S RADIOLOGICAL TRACEABILITY PROGRAM FOR RADIOANALYTICAL MEASUREMENTS: WORK IN PROGRESS

DOE S RADIOLOGICAL TRACEABILITY PROGRAM FOR RADIOANALYTICAL MEASUREMENTS: WORK IN PROGRESS DOE S RADIOLOGICAL TRACEABILITY PROGRAM FOR RADIOANALYTICAL MEASUREMENTS: WORK IN PROGRESS D. E. McCurdy, Duke Engineering & Services, Bolton, MA and J. S. Morton, Department of Energy, Office of Environmental

More information