Carbonyl-Reactive Tandem Mass Tag (TMT) Reagents for Mass Spectrometry-Based Quantitative Glycomics

Size: px
Start display at page:

Download "Carbonyl-Reactive Tandem Mass Tag (TMT) Reagents for Mass Spectrometry-Based Quantitative Glycomics"

Transcription

1 Carbonyl-Reactive Tandem Mass Tag (TMT) Reagents for Mass Spectrometry-Based Quantitative Glycomics Sergei I. Snovida, 1 Rosa Viner, 2 John C. Rogers 1 1 Thermo Fisher Scientific, Rockford, IL; 2 Thermo Fisher Scientific, San Jose, CA

2 Overview Purpose: To demonstrate applications of the new carbonyl-reactive aminoxytmt reagent in quantitative analysis of carbohydrates by direct infusion and liquid chromatography-coupled mass spectrometry. Methods: released from standard glycoproteins, including several monoclonal antibodies, were individually labeled with the isobaric aminoxytmt reagents, and combined together for quantitative analysis by mass spectrometry. Results: Relative quantitation of the labeled glycans from different samples was done by measuring relative peak intensities of the TMT reporter ions at 2 or 3 levels. Introduction Aberrant glycosylation profiles may reflect abnormal physiological state of a cell, an organ, or an organism as a whole, and may be indicative of a disease state or cancer. Thus, glycans may be used as biomarkers. Additionally, many biotherapeutic drugs are glycoproteins and their activity and efficacy greatly depends on the type and extent of glycosylation. These compounds must be characterized in terms of their glycosylation profiles in both the development and quality control phases of drug development and manufacture. Accurate quantitation of glycans remains elusive due to the lack of a comprehensive selection of available standards, poor ionization efficiency of carbohydrates relative to other classes of biomolecules, and broad structural heterogeneity of glycomic samples. Recently, we introduced a set of isobaric carbonyl-reactive TMT reagents, Thermo Scientific aminoxytmt (Tandem Mass Tag ) Label Reagents (Figure 1), which can react with the reducing end of glycans to form a stable oxime product. The six compounds of the Thermo Scientific aminoxytmtsixplex Reagent Set have the same mass (i.e., isobaric) and chemical structure (carbonyl-reactive aminoxy group, spacer arm, and mass reporter). However, the specific distribution of 13 C and N isotopes on either side of the / fragmentation site in each reagent results in a unique reporter mass ( Da) in the low region of 2 spectra. This set of reporter ions is used to measure the relative abundance of labeled molecules in a combined (multiplexed) sample representing six different treatment conditions, time points, or replicates. For glycobiology applications, the reagents enable quantitative profiling of glycoforms and discovery of glycan biomarkers; they provide improved ionization of glycans for increased sensitivity and increased analytical throughput by sample multiplexing. In this work, we showcase the use of these reagents for quantitative glycomics by combining our multiplexed TMT-based approach with HILIC LC- technique to enable more-sensitive analysis with improved glycome coverage. Methods Sample Preparation Bovine thyroglobulin (Sigma-Aldrich) and several monoclonal antibodies (Thermo Fisher Scientific) were reduced, alkylated, and digested with -grade trypsin (Thermo Fisher Scientific). Digest mixture was then treated with PNGase F (New England Biolabs) to release N-linked glycans. Following deglycosylation, released glycans were separated from peptides using Oasis HLB (Waters) solid phase extraction columns. After drying, glycan samples were labeled with aminoxytmt reagents (Thermo Fisher Scientific) according to manufacturer s protocol. Liquid Chromatography LC- experiments were performed in HILIC mode on a Thermo Scientific Accucore Amide column (2.6µm, Å; µm mm). Mobile phases were.1% formic acid in water (A) and.1% formic acid in acetonitrile (B). Separation was carried out with a 4 nl/min flowrate using different, sample-dependent gradients. Mass Spectrometry For direct infusion experiments, labeled glycan samples were dissolved in % acetonitrile solution containing µm NaOH. Samples were analyzed on Thermo Scientific Orbitrap Velos Pro ion trap and Thermo Scientific Orbitrap Fusion Tribrid mass spectrometers in the positive ion mode. Data Analysis All data was processed manually. FIGURE 2. Positive ion mode aminoxytmt-labeled high-m unlabeled B (Na.34 aminoxytmt Reagent-labeled * Carbonyl-Reactive Tandem Mass Tag (TMT) Reagents for Mass Spectrometry-Based Quantitative Glycomics

3 blf_ng_tmt6_2_1_infusion_ot 193_6res #1-36 RT:.-.13 AV: 36 NL:.8E T: FT + p ESI Full ms2 193.@hcd. [.-2.] Results General Features of aminoxytmt-labeled Glycans Two types of the reagent are available, Thermo Scientific aminoxytmtzero Label Reagents, the light version of the reagent recommended for method development work, and the heavy, isobaric aminoxytmt sixplex set (Figure 1). When a labeled glycan is subjected to / fragmentation by, mass reporter region of the label molecule is cleaved, leading to an intense peak in the low region of / spectrum. For the aminoxytmtsixplex set, each member of the set produces a mass reporter ion with a unique, thus allowing for relative quantitation of glycoforms across up to six different samples (Figure 3). The carbonyl-reactive alkoxyamine group reacts rapidly with aldehydes and ketones to yield a stable oxime product, which is resistant to hydrolysis over a broad range of experimental conditions. Comparison of the fragmentation spectra of the sodiated precursors of unlabeled and aminoxytmt-labeled glycans (Figure 2) shows that the same structurally relevant product ions are produced for both the labeled and the unlabeled glycans, suggesting that structural information is not lost or diminished after labeling with the aminoxytmt reagent. Therefore, in addition to the mass reporter ions relevant for quantitation, information which can be derived from the / spectra of the unlabeled glycans is still present here. FIGURE 1. Chemical structure of aminoxytmt reagents and labeling reaction scheme. AminoxyTMT Reagent Structure AminoxyTMTsixplex Isobaric Reagents FIGURE 3. Workflow for quantitativ Std Lc-fluorescence Data #1 #2 Standard Reference Sample #3 #4 3_mAbs_12 T: IT + p # TMT Reporter Ion Channels + Reaction: + glycan aminoxytmt reagent Stable oxime product FIGURE 2. Positive ion mode fragmentation of native unlabeled and aminoxytmt-labeled high-mannose glycans NL: 1.16E4 RNAseB_NG_2ug_prot_ unlabeled eq_ml_641#9 RT:.1 [M+2Na] 2+ AV: 1 T: IT + p ESI Full [M+Na] + ms2 64.@hcd. [.-2.] Y (Na +,2 ) A NL:.17E3 blf_rnaseb_btg_tmt_pos_78 # RT:.9 AV: 1 T: 78.@hcd [.-2.] 1.36 C (Na + ) aminoxytmt Reagent-labeled * [M+ Na-reporter-CO] , A Trap- 64. Trap Thermo Scientific Poster Note PN AS-EN-614S 3

4 blf_ng_tmt6_2_1_infusion_ot 193_6res #1-36 RT:.-.13 AV: 36 NL:.8E T: FT + p ESI Full ms2 193.@hcd. [.-2.] FIGURE 3. Workflow for quantitative glycomics multiplexed experiment. Lc-fluorescence Data Unknown Samples Standard Reference Sample Std #1 #2 #3 #4 # TMT 126 TMT 127 TMT 128 TMT 129 TMT 13 TMT 131 RT: Std #1 #2 # #4 TMT Reporter Ion Channels _mAbs_ #4 RT:.11 AV: 1 NL: 2.41E T: IT + p ESI Full ms [7.-.] Sixplex Sample Time (min) 8 7 / 6 # Actual ratios :2:::2: E 3.9E of native unlabeled and NL: 1.16E4 RNAseB_NG_2ug_prot_ eq_ml_641#9 RT:.1 [M+2Na] 2+ AV: 1 T: IT + p ESI Full [M+Na] + ms2 64.@hcd. [.-2.] ,2 A NL:.17E3 blf_rnaseb_btg_tmt_pos_78 # RT:.9 AV: 1 T: 78.@hcd [.-2.] C (Na + ) [M+ Na-reporter-CO] +,2 A Trap- 64. Trap Multiplexing Glycomic Samples - Overview In a typical experiment, up to six separate samples may be analyzed together, which allows lower overall analysis time and minimizes the variations in quantitative reproducibility, as the number of independent sample handling steps is reduced in this workflow (Figure 3). After labeling with aminoxytmt reagents, unreacted reagents are quenched with acetone and samples are combined. An additional clean-up step using HILIC solid phase extraction material is used to separate the quenched reagent from the labeled glycans. During analysis, precursor ions corresponding to the glycoforms of interest are identified and are subjected to / analysis by fragmentation. Relative peak intensities of the reporter ions observed in the / spectra are proportional to the relative abundance levels of the selected glycoform in different samples in the set. Several peaks observed in the low region of / spectra of the labeled glycans may interfere with accurate quantitative analysis if a low resolution mass spectrometer is used (Figure 3). These peaks are some of the product ions of HexNAc fragmentation, and are thus present in / spectra of most glycoforms. Although this may be an issue for low resolution instruments, these interferences are easily resolved and do not pose a problem for accurate quantitation at the lowest resolution settings of a Thermo Scientific Orbitrap mass spectrometer. If a Velos Pro ion trap mass spectrometer is used for the analysis, application of a targeted 3 method (Figure 4) solves this interference problem. First, precursor of interest is isolated and fragmented using CID or fragmentation mode. In the second step, abundant Y 1 ions, which still contain intact aminoxytmt label, are isolated for 3. 3 spectra contain clean mass reporter ions and their ratios are now as expected for the sample set. Multiplexing Glycomic Samples - Examples Several experiments showcasing the use of these reagents for quantitative analysis of glycans were performed using antibodies as glycan sources. In the first experiment (Figure ), 1 µg of two anti c-myc monoclonal antibodies, each coming from a different source, were separately deglycosylated using PNGase F glycosidase, glycans were isolated and each sample was labeled with a different heavy version of the isotopic aminoxytmt reagent. After quenching and clean-up, the samples were combined and analyzed by direct infusion using a Velos Pro mass spectrometer. Anti-c-Myc hybridoma Extracted glycan profiles IgGs Re Mouse Culture Relative quantitation of the three by comparing relative peak inten spectra for each glycoform. After produced in the cell culture cont than the antibodies obtained from 4 Carbonyl-Reactive Tandem Mass Tag (TMT) Reagents for Mass Spectrometry-Based Quantitative Glycomics

5 NL: 1.3E4 btg_ng_native_oxytmt_1_2 hilic_acnsamp le_1_1ul# RT: AV: 42 F: IT + p NSI d Full ms @hcd8. [1.-2.] 1to1_mix_native_TMT_pos #18 RT:.11 AV: 1 NL: 1.82E T: IT + p ESI E Full ms [.-2.] ltiplexed experiment Unknown Samples #2 #3 #4 # TMT 128 TMT 129 TMT 13 TMT 131 Sixplex Sample FIGURE 4. Targeted 3 acquisition approach using a Velos Pro ion trap instrument. RT: Time (min) Actual ratios 1:2:::2: E4 3.9E2 97. DDA/targeted 98. NL: 9.19E6 Base Peak btg_ng_native_oxy TMT_1_2 HILIC_ ACNsample_1_1 ul_ Y 1 -ion ( or CID) clean reporter ions 3 () [M+3H] In another example (Figure 6), 1 deglycosylated using PNGase F g pool was divided into two equal pa the heavy isobaric aminoxytmtsix the individual samples were comb three separate samples in a duplic infusion to reveal that two out with respect to glycosylation profile antibody having substantially highe FIGURE 6. Quantitative compari antibodies with replicates. NL: 7.4E4 btg_ng_native_oxytmt_1_2 hilic_acnsamp le_1_1ul# RT: AV: 42 F: IT + p NSI d Full ms @hcd8. [1.-2.] NL: 3.88E2 btg_ng_native_oxytmt_1_2 HILIC_AC Nsample_1_1uL_ # RT: AV: 147 F: IT + p NSI d Full ms3 891.@hcd @hcd1. [1.-2.] NL: 1.93E6 btg_ng_native_oxytmt_1_2 HILIC_AC Nsample_1_1uL_ # RT:.-81. AV: 148 T: IT + p NSI E Full ms [4.-2.] NL: 3.88E2 btg_ng_native_oxytmt_1_2 HILIC_AC Nsample_1_1uL_ # RT: AV: 147 F: IT + p NSI d Full ms3 891.@hcd @hcd1. [1.-2.] TMT 126 TMT 129 TMT 127 TMT 13 TMT 128 TMT _mAbs_1231 T: IT + p E 1 9 FIGURE. Comparing glycosylation profiles of two monoclonal antibodies from different sources. Anti-c-Myc hybridoma IgGs 1/1 Deglycosylate, label, mix, clean 1/1 GF G1F G2F Combined Sample Extracted glycan profiles Reporter ion relative intensities NL: 7.76E2 native_tmt128_131_pos_8 94# RT:.1 AV: 1 T: 894.@hcd. [.-2.] NL: 2.E2 native_tmt128_131_pos_ # RT:.11 AV: 1 T: @hcd. [.-2.] NL: 1.79E1 native_tmt128_131_pos_1 6#4 RT:.9 AV: 1 T: 6.@hcd. [.-2.] Mouse Culture Relative quantitation of the three main glycoforms, GF, G1F, and G2F, was performed by comparing relative peak intensities of the mass reporter peaks in the / spectra for each glycoform. After signal normalization, we found that the antibody produced in the cell culture contained higher amounts of the G1F and G2F glycoforms than the antibodies obtained from the mouse. Thermo Scientific Poster Note PN AS-EN-614S

6 1to1_mix_native_TMT_pos #18 RT:.11 AV: 1 NL: 1.82E T: IT + p ESI E Full ms [.-2.] _mAbs_123123_894 #1 RT:.1 AV: 1 NL:.69E3 T: IT + p ESI Full ms @hcd. [.-.] _mabs_123123_9 # RT:.24 AV: 1 NL:.48E T: IT + p ESI Full ms2 9.2@hcd [.-.] _mabs_123123_ #8 RT: AV: 11 NL: 1.22E3136 T: IT + p ESI Full ms2 6.2@hcd. [.-.] d [M+3H] 3+ In another example (Figure 6), 1 µg of three different monoclonal antibodies were deglycosylated using PNGase F glycosidase, glycans were isolated, and each sample pool was divided into two equal parts. Each part was labeled with a different version of the heavy isobaric aminoxytmtsixplex reagent, and after a quenching and clean-up, the individual samples were combined to yield a single sixplex sample consisting of three separate samples in a duplicate. The sample was then analyzed by direct infusion to reveal that two out of three different antibodies analyzed were identical with respect to glycosylation profiles of the three main glycoforms, with the third antibody having substantially higher abundance of the G1F and G2F glycoforms. NL: 1.93E6 btg_ng_native_oxytmt_1_2 HILIC_AC Nsample_1_1uL_ # RT:.-81. AV: 148 T: IT + p NSI E Full ms [4.-2.] Y 1 -ion 2 ( or CID) FIGURE 6. Quantitative comparison of glycosylation of several monoclonal antibodies with replicates. NL: 7.4E4 btg_ng_native_oxytmt_1_2 hilic_acnsamp le_1_1ul# RT: AV: 42 F: IT + p NSI d Full ms @hcd8. [1.-2.] er 3 () NL: 3.88E2 btg_ng_native_oxytmt_1_2 HILIC_AC Nsample_1_1uL_ # RT: AV: 147 F: IT + p NSI d Full ms3 891.@hcd @hcd1. [1.-2.] GF G1F TMT 126 TMT 129 TMT 127 TMT 13 TMT 128 TMT 131 Combined Sample GF 3_mAbs_ #4 RT:.11 AV: 1 NL: 2.41E6 T: IT + p ESI Full ms [7.-.] G1F G2F es G2F NL: 7.76E2 native_tmt128_131_pos_8 94# RT:.1 AV: 1 T: @hcd [.-2.] NL: 2.E2 native_tmt128_131_pos_ # RT:.11 AV: 1 T: @hcd [.-2.] NL: 1.79E1 native_tmt128_131_pos_ #4 RT:.9 AV: 1 T: @hcd. [.-2.] F, G1F, and G2F, was performed porter peaks in the / n, we found that the antibody ts of the G1F and G2F glycoforms Conclusion Increase sample analysis throughput with aminoxytmt reagents Overall improvement in sensitivity with aminoxytmt label approach Structural elements are preserved in labeled glycans LC-UV/fluorescence and TMT-based quantitation can be complementary Ideal for analysis of biotherapeutics and biomarker discovery Acknowledgements We would like to thank the following colleagues and collaborators for critical advice and helpful discussions: Dr. Ryan Bomgarden, Dr. Chris Etienne, Dr. Kay Opperman, Dr. Venky Shivalingappa, and Dr. Julian Saba (Thermo Fisher Scientific); Dr. Kay-Hooi Khoo, Dr. Ming-Yi Ho, Dr. Shui-Hua Wang, and Dr. Chia-Wei Lin (Academia Sinica, Taipei, Taiwan); Dr. Ian Pike and Dr. Karsten Kuhn (Proteome Sciences); Dr. Bernhard Küster, Dr. Hannes Hahne, and Dr. Patrick Neubert (Technische Universität München, Munich, Germany); Dr. Lingjun Li, Dr. Xuefei Zhong, and Dr. Yan Liu (University of Wisconsin, Madison, WI, USA); Dr. Yehia Mechref, Shiyue Zhou, and Yunli Hu (Texas Tech University, Lubbock, TX, USA); Dr. Shonali Paul and Dr. Sanjib Meitei (Premier Biosoft). PO64133-EN 614S 6 Carbonyl-Reactive Tandem Mass Tag (TMT) Reagents for Mass Spectrometry-Based Quantitative Glycomics

7 214 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. Tandem Mass Tag and TMT are registered trademarks of Proteome Sciences plc. Sigma-Aldrich is a registered trademark of Sigma-Aldrich Co. LLC. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa Australia Austria Belgium Canada China (free call domestic) Denmark Europe-Other Finland France Germany India Italy Japan Latin America Middle East Netherlands New Zealand Norway Russia/CIS Thermo Fisher Scientific, San Jose, CA USA is ISO 91:28 Certified. Singapore Spain Sweden Switzerland UK USA PN EN-614S

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS Jason Lai, Jia Wang, Brad Hart, Pavel Aronov, Kristine Van Natta, Marta Kozak, Jorge Valdivia, Andy Jacobs, Haibo

More information

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput Yi Zhang 1, Zhiqi Hao 1, Markus Kellmann 2 and Andreas FR. Huhmer

More information

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer Application ote: 54 Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap rbitrap Hybrid Mass Spectrometer Kate Comstock, Yingying Huang; Thermo Fisher Scientific,

More information

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap Dirk Nolting and Andreas Wieghaus Thermo Fisher Scientific, Bremen, Germany Overview Purpose: Characterize the rate

More information

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry O. Scheibner, 1 P. van Baar, 2 F. Wode, 2 U. Dünnbier, 2 K. Akervik, 3 J. Humphries, 3 M. Bromirski 1

More information

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3 Second-Generation Electron Transfer Dissociation (ETD) on the Thermo Scientific Orbitrap Fusion Mass Spectrometer with Improved Functionality, Increased Speed, Christopher Mullen, 1 Lee Earley, 1 Jean-Jacques

More information

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use Shih-Tse Jason Lai, Jia Wang, Brad Hart, Kristine Van Natta, Marta Kozak, Jorge Valdivia, Haibo Wang,

More information

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer Reiko Kiyonami, Mary Blackburn, Andreas FR Hühme: Thermo Fisher

More information

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer Olaf Scheibner and Maciej Bromirski Thermo Fisher Scientific, Bremen, Germany Overview Purpose: Improve the performance

More information

Increasing the Multiplexing of Protein Quantitation from 6- to 10-Plex with Reporter Ion Isotopologues

Increasing the Multiplexing of Protein Quantitation from 6- to 10-Plex with Reporter Ion Isotopologues Increasing the Multiplexing of Protein Quantitation from 6- to 1-Plex with Reporter Ion Isotopologues Rosa Viner, 1 Ryan Bomgarden, 2 Michael Blank, 1 John Rogers 2 1 Thermo Fisher Scientific, San Jose,

More information

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications Xiaodong Liu and Christopher Pohl; Thermo Fisher Scientific, Sunnyvale, CA Overview Review mixed-mode column technology

More information

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method Xiang He and Marta Kozak ThermoFisher Scientific, San Jose, CA,

More information

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis Tim Stratton 1, Yingying Huang 1, Katianna Pihakari 1, Ian Acworth 2, and Michael Weber 2 1 Thermo Fisher Scientific,

More information

HILIC Method Development in a Few Simple Steps

HILIC Method Development in a Few Simple Steps HILIC Method Development in a Few Simple Steps Monica Dolci, Luisa Pereira, Dafydd Milton and Tony Edge Thermo Fisher Scientific, Runcorn, Cheshire, UK Overview This poster presents a systematic approach

More information

of mass spectrometry

of mass spectrometry Thermo Scientific 253 Ultra High resolution isotope ratio MS Discover a new world of mass spectrometry Paleoclimatology Atmospheric science Biogeochemistry Petrology Discover the isotopic anatomy of molecules

More information

Improving Intact Antibody Characterization by Orbitrap Mass Spectrometry

Improving Intact Antibody Characterization by Orbitrap Mass Spectrometry Improving Intact Antibody Characterization by Orbitrap Mass Spectrometry Kai Scheffler, 1 Eugen Damoc, 2 Mathias Müller, 2 Martin Zeller, 2 Thomas Moehring 2 Thermo Fisher Scientific, Dreieich 1 and Bremen,

More information

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites Robert Heintz, Mark Wall, Jennifer Ramirez, Stephan Woods Thermo Fisher Scientific, Madison WI Overview Purpose:

More information

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis A. Caruso, M. Santoro, P. Magni, S. Pelagatti, and R. Facchetti Thermo Fisher Scientific, Milan,

More information

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery Metabolite Stability Screening and Hotspot Metabolite Identification by Combining High-Resolution, Accurate-Mass Nonselective and Selective Fragmentation Tim Stratton, Caroline Ding, Yingying Huang, Dan

More information

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS Kate Comstock, Tim Stratton, Hongxia (Jessica) Wang, and Yingying Huang

More information

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS Daniel Kutscher, 1 Jörg Bettmer, 2 Torsten Lindemann, 1 Shona McSheehy-Ducos, 1 Lothar Rottmann 1 1 Thermo Fisher Scientific, Germany 2

More information

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems Yan Liu, Zhongqing Lu, and Chris Pohl; Thermo Fisher Scientific, Sunnyvale, CA USA Overview

More information

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 000 Elemental Analyzer Dr. Liliana Krotz and Dr. Guido Giazzi Thermo Fisher Scientific, Milan, Italy

More information

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry Olaf Scheibner, 1 Patrizia van Baar, 2 Florian Wode, 2 Uwe Dünnbier, 2 Kristi Akervik, 3 Jamie Humphrie,

More information

High-Field Orbitrap Creating new possibilities

High-Field Orbitrap Creating new possibilities Thermo Scientific Orbitrap Elite Hybrid Mass Spectrometer High-Field Orbitrap Creating new possibilities Ultrahigh resolution Faster scanning Higher sensitivity Complementary fragmentation The highest

More information

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency Anthony Edge, 1 Luisa Pereira, 1 Dafydd Milton 1 and Andrew Shalliker 2 1 Thermo Fisher Scientific,

More information

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3 Second-Generation Electron Transfer Dissociation (ETD) on the Thermo Scientific Orbitrap Fusion Mass Spectrometer with Improved Functionality, Increased Speed, and Improved Robustness of Data Christopher

More information

Relative quantification using TMT11plex on a modified Q Exactive HF mass spectrometer

Relative quantification using TMT11plex on a modified Q Exactive HF mass spectrometer POSTER NOTE 6558 Relative quantification using TMT11plex on a modified mass spectrometer Authors Tabiwang N. Arrey, 1 Rosa Viner, 2 Ryan D. Bomgarden, 3 Eugen Damoc, 1 Markus Kellmann, 1 Thomas Moehring,

More information

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery Application Note: 3 Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery Yoko Yamagishi, Thermo Fisher Scientific, C-2F 3-9

More information

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer Improved Screening for 2 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer Mary lackburn, Jia Wang, Jonathan eck, Charles Yang, Dipankar Ghosh, Thermo Fisher Scientific, San Jose, C,

More information

Monitoring Protein PEGylation with Ion Exchange Chromatography

Monitoring Protein PEGylation with Ion Exchange Chromatography Monitoring Protein PEGylation with Ion Exchange Chromatography Peter Yu, Deanna Hurum, Jinyuan (Leo) Wang, 2 Terry Zhang, 2 and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA; 2 Thermo Fisher Scientific,

More information

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma LC-MS/MS Method for the Determination of Diclofenac in Human Plasma J. Jones, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20569 Key Words SPE, SOLA, Accucore RP-MS, diclofenac, Core

More information

MetWorks Metabolite Identification Software

MetWorks Metabolite Identification Software m a s s s p e c t r o m e t r y MetWorks Metabolite Identification Software Enabling Confident Analysis of Metabolism Data Part of Thermo Fisher Scientific MetWorks Software for the Confident Analysis

More information

Direct Analysis using Paper-Spray Mass Spectrometry: Method Development for the Rapid Screening of Drugs of Abuse for Forensic Toxicology

Direct Analysis using Paper-Spray Mass Spectrometry: Method Development for the Rapid Screening of Drugs of Abuse for Forensic Toxicology Direct Analysis using Paper-Spray Mass Spectrometry: Method Development for the Rapid Screening of Drugs of Abuse for Forensic Toxicology Maria C. Prieto Conaway, 1 Nicholas E. Manicke, 2 Marta Kozak 1

More information

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis Alessandro Santato, 1 Doug Hamilton, 1 Jan Wijbrans, 2 Claudia Bouman 1 1 Thermo Fisher Scientific, Bremen, Germany 2 VU University Amsterdam,

More information

A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzcloud Library

A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzcloud Library A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzcloud Library Junhua Wang, 1 David A. Peake, 1 Robert Mistrik, 2 Yingying Huang 1 1 Thermo Fisher Scientific

More information

Targeted protein quantification

Targeted protein quantification Targeted Quantitative Proteomics Targeted protein quantification with high-resolution, accurate-mass MS Highly selective Very sensitive Complex samples HR/AM A more complete quantitative proteomics picture

More information

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Andrea Caruso and Massimo Santoro, Thermo Fisher Scientific, Milan, Italy Application Note 1316 Key

More information

Electron Transfer Dissociation of N-linked Glycopeptides from a Recombinant mab Using SYNAPT G2-S HDMS

Electron Transfer Dissociation of N-linked Glycopeptides from a Recombinant mab Using SYNAPT G2-S HDMS Electron Transfer Dissociation of N-linked Glycopeptides from a Recombinant mab Using SYNAPT G2-S HDMS Jonathan P. Williams, Jeffery M. Brown, Stephane Houel, Ying Qing Yu, and Weibin Chen Waters Corporation,

More information

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter Luisa Pereira, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20755 Key Words Solid core, fused core, superficially

More information

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches Hydrophilic Interaction Liquid Chromatography: Method Development Approaches Monica Dolci, Luisa Pereira, and Tony Edge Thermo Fisher Scientific, Runcorn, Cheshire, UK Abstract This poster summarizes the

More information

Thermo Scientific Pesticide Explorer Collection. Start-to-finish. workflows for pesticide analysis

Thermo Scientific Pesticide Explorer Collection. Start-to-finish. workflows for pesticide analysis Thermo Scientific Pesticide Explorer Collection Start-to-finish workflows for pesticide analysis Comprehensive Pesticide Analysis Solutions Pesticide Explorer Collection Selection table Lab Profile Routine

More information

for XPS surface analysis

for XPS surface analysis Thermo Scientific Avantage XPS Software Powerful instrument operation and data processing for XPS surface analysis Avantage Software Atomic Concentration (%) 100 The premier software for surface analysis

More information

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil Che Jinshui, 1 Deng Guifeng, 1 Liang Lina, 1 and Aaron Kettle, 2 1 Thermo Fisher Scientific (China)

More information

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS Thermo Scientific ConFlo IV Universal Interface Continuous Flow Interface Isotope Ratio MS 3 ConFlo IV Universal Interface for Continuous Flow Isotope Ratio MS The development of Continuous Flow carrier

More information

High-Resolution Accurate-Mass (HRAM) Phthalate Screening using Direct Analysis in Real Time (DART) Ambient Ionization

High-Resolution Accurate-Mass (HRAM) Phthalate Screening using Direct Analysis in Real Time (DART) Ambient Ionization igh-resolution Accurate-Mass (RAM) Phthalate Screening using Direct Analysis in Real Time (DART) Ambient Ionization Catharina Crone, 1 Markus Kellmann, 1 Yue Xuan, 1 Elizabeth Crawford 2 1 Thermo Fisher

More information

Quan/Qual Analyses. Unmatched Confidence for. Thermo Scientific Q Exactive Orbitrap LC-MS/MS System. Identify Quantify Confirm

Quan/Qual Analyses. Unmatched Confidence for. Thermo Scientific Q Exactive Orbitrap LC-MS/MS System. Identify Quantify Confirm 26.1645 165.91 33.267 Thermo Scientific Q Exactive rbitrap LC-MS/MS System Unmatched Confidence for Quan/Qual Analyses Identify Quantify Confirm Quanfirmation \ kwän-f r-'m -shän\ n 1 : the ability to

More information

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer Defining quality standards for the analysis of solid samples Redefine your quality standards for the elemental analysis of solid samples

More information

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration Zhang Xuebin 1, Yu Chongtian 1, Liang Lina 1, Hans-Joachim Huebschmann 2, Thermo Fisher Scientific Shanghai, 1 China,

More information

XPS Surface Characterization of Disposable Laboratory Gloves and the Transfer of Glove Components to Other Surfaces

XPS Surface Characterization of Disposable Laboratory Gloves and the Transfer of Glove Components to Other Surfaces Application Note: 52287 XPS Surface Characterization of Disposable Laboratory Gloves and the Transfer of Glove Components to Other Surfaces Brian R. Strohmeier, Thermo Fisher Scientific, Madison, WI, USA

More information

Fast, Effective XPS Point Analysis of Metal Components

Fast, Effective XPS Point Analysis of Metal Components Application Note: 52297 Fast, Effective XPS Point Analysis of Metal Components Chris Baily and Tim Nunney, Thermo Fisher Scientific, East Grinstead, West Sussex, UK Key Words K-Alpha Auto-Analysis Multi-Spectrum

More information

The Raman Spectroscopy of Graphene and the Determination of Layer Thickness

The Raman Spectroscopy of Graphene and the Determination of Layer Thickness Application Note: 52252 The Raman Spectroscopy of Graphene and the Determination of Layer Thickness Mark Wall, Ph.D., Thermo Fisher Scientific, Madison, WI, USA Key Words DXR Raman Microscope 2D Band D

More information

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity Monica Dolci, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20544 Key Words Hydrophilic, HILIC,

More information

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Thunyarat Phesatcha, Suparerk Tukkeeree, Jeff Rohrer 2 Thermo

More information

Improved 6- Plex TMT Quantification Throughput Using a Linear Ion Trap HCD MS 3 Scan Jane M. Liu, 1,2 * Michael J. Sweredoski, 2 Sonja Hess 2 *

Improved 6- Plex TMT Quantification Throughput Using a Linear Ion Trap HCD MS 3 Scan Jane M. Liu, 1,2 * Michael J. Sweredoski, 2 Sonja Hess 2 * Improved 6- Plex TMT Quantification Throughput Using a Linear Ion Trap HCD MS 3 Scan Jane M. Liu, 1,2 * Michael J. Sweredoski, 2 Sonja Hess 2 * 1 Department of Chemistry, Pomona College, Claremont, California

More information

Quantitation of TMT-Labeled Peptides Using Higher-Energy Collisional Dissociation on the Velos Pro Ion Trap Mass Spectrometer

Quantitation of TMT-Labeled Peptides Using Higher-Energy Collisional Dissociation on the Velos Pro Ion Trap Mass Spectrometer Application Note: 520 Quantitation of TMT-Labeled Peptides Using Higher-Energy Collisional Dissociation on the Velos Pro Ion Trap Mass Spectrometer Roger G. Biringer, Julie A. Horner, Rosa Viner, Andreas

More information

Key Words Q Exactive Focus, Orbitrap, veterinary drugs, small molecule HRAM quantitation, small molecule HRAM screening, UHPLC, vdia

Key Words Q Exactive Focus, Orbitrap, veterinary drugs, small molecule HRAM quantitation, small molecule HRAM screening, UHPLC, vdia Variable Data-Independent Acquisition (vdia) Delivers High Selectivity and Sensitivity in Combined Targeted and Untargeted Analyses for Small Molecules Olaf Scheibner 1, Markus Kellmann 1, Charles Yang

More information

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Mass Spectral Studies of Polypropylene Chromatographic Well Plates Mass Spectral Studies of Polypropylene Chromatographic Well Plates Brian King, Thermo Fisher Scientific, Runcorn, Cheshire, UK Detlev Lennartz, Thermo Fisher Scientific, Langerwehe, Germany White Paper

More information

Introduction to Fourier Transform Infrared Spectroscopy

Introduction to Fourier Transform Infrared Spectroscopy molecular spectroscopy Introduction to Fourier Transform Infrared Spectroscopy Part of Thermo Fisher Scientific Introduction What is FT-IR? FT-IR stands for Fourier Transform InfraRed, the preferred method

More information

Introduction to Fourier Transform Infrared Spectroscopy

Introduction to Fourier Transform Infrared Spectroscopy Introduction to Fourier Transform Infrared Spectroscopy Introduction What is FTIR? FTIR stands for Fourier transform infrared, the preferred method of infrared spectroscopy. In infrared spectroscopy, IR

More information

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer APPLICATION NOTE 72385 Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer Authors Stephan Meding, Katherine Lovejoy, Martin Ruehl Thermo Fisher Scientific,

More information

Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS

Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS Kelly Munro, 1 Anthony Edge, 2 Claudia Martins, 3 David Cowan 4 and Leon Barron 1 1 Analytical & Environmental

More information

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma Jon Bardsley 1, James Goldberg 2 1 Thermo Fisher Scientific, Runcorn, UK; 2 Thermo Fisher Scientific, West Palm Beach,

More information

An Evaluation of Various High-Resolution, Accurate-Mass Scan Modes for In Vitro Drug Discovery Screening

An Evaluation of Various High-Resolution, Accurate-Mass Scan Modes for In Vitro Drug Discovery Screening An Evaluation of Various High-Resolution, Accurate-Mass Scan Modes for In Vitro Drug Discovery Screening Jonathan McNally, 1 Nicholas Duczak Jr., 1 Patrick Bennett, 1 Francois Espourteille, and Maciej

More information

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany Generic Method Approach for Pharmaceutical Drug Discovery and Development using Reversed-Phase Hydrophilic Interaction Liquid Chromatography with Universal Charged Aerosol Detection Frank Steiner, Michael

More information

Detection of Mycotoxins in Corn Meal Extract Using Automated Online Sample Preparation with Liquid Chromatography-Tandem Mass Spectrometry

Detection of Mycotoxins in Corn Meal Extract Using Automated Online Sample Preparation with Liquid Chromatography-Tandem Mass Spectrometry Detection of Mycotoxins in Corn Meal Extract Using Automated Online Sample Preparation with Liquid Chromatography-Tandem Mass Spectrometry Yang Shi, Catherine Lafontaine, Timothy B. Haney, and François

More information

( )( ) Selectivity Choices in Reversed-Phase Fast LC. Introduction. R s = 1 a 1 k 4 a 1 + k

( )( ) Selectivity Choices in Reversed-Phase Fast LC. Introduction. R s = 1 a 1 k 4 a 1 + k Selectivity Choices in Reversed-Phase Fast LC Luisa Pereira, Monica Dolci, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20543 Key Words Accucore, Hypersil GOLD, Syncronis, Column Chemistry,

More information

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes Joe DiBussolo, Marta Kozak Thermo Fisher Scientific, San Jose, CA Application

More information

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA LECTURE-25 Quantitative proteomics: itraq and TMT TRANSCRIPT Welcome to the proteomics course. Today we will talk about quantitative proteomics and discuss about itraq and TMT techniques. The quantitative

More information

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection Ian Acworth, Bruce Bailey, Paul Gamache, and John Waraska Thermo Fisher Scientific, Chelmsford, MA, USA Introduction

More information

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection Terri Christison, Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA Overview

More information

protein quantitation

protein quantitation Thermo Scientific Pierce Reagents for Quantitative Proteomics protein quantitation using mass spectrometry Specialized reagents for discovery and targeted analysis table of contents verview 1 Discovery

More information

A Platform Method for Pharmaceutical Counterion Analysis by HPLC

A Platform Method for Pharmaceutical Counterion Analysis by HPLC A Platform Method for Pharmaceutical Counterion Analysis by HPLC Xiaodong Liu, Mark Tracy, and Christopher Pohl; Thermo Fisher Scientific, Sunnyvale, CA, USA Overview Results and Simple, fast, economical

More information

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories APPLICATION NOTE 65200 Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories Authors Neloni Wijeratne, Claudia Martins, Mary

More information

Enhancement of Linearity and Response in Charged Aerosol Detection

Enhancement of Linearity and Response in Charged Aerosol Detection Enhancement of Linearity and Response in Charged Aerosol Detection Christopher Crafts, Marc Plante, Bruce Bailey, Ian Acworth, Thermo Fisher Scientific, Chelmsford, MA, USA Overview Purpose: This work

More information

Quantitative and Qualitative Confirmation of Pesticides in Beet Extract Using a Hybrid Quadrupole-Orbitrap Mass Spectrometer

Quantitative and Qualitative Confirmation of Pesticides in Beet Extract Using a Hybrid Quadrupole-Orbitrap Mass Spectrometer Quantitative and Qualitative Confirmation of Pesticides in Beet Extract Using a Hybrid Quadrupole-Orbitrap Mass Spectrometer Charles Yang and Dipankar Ghosh, Thermo Fisher Scientific, San Jose, CA Olaf

More information

MS-based proteomics to investigate proteins and their modifications

MS-based proteomics to investigate proteins and their modifications MS-based proteomics to investigate proteins and their modifications Francis Impens VIB Proteomics Core October th 217 Overview Mass spectrometry-based proteomics: general workflow Identification of protein

More information

[ Care and Use Manual ]

[ Care and Use Manual ] Oasis HLB DISKS I. Introduction Oasis HLB is a Hydrophilic-Lipophilic-Balanced, water-wettable, reversed-phase sorbent for all your SPE needs. It is made from a specific ratio of two monomers, the hydrophilic

More information

IN QUALITATIVE ANALYSIS,

IN QUALITATIVE ANALYSIS, IN QUALITATIVE ANALYSIS, YOU VE ALWAYS HAD TO CHOOSE BETWEEN ACCURACY AND THROUGHPUT. NOW THERE S NO REASON TO ever have to. WATERS QUALITATIVE ANALYSIS solutions Waters ACQUITY UPLC System with the Waters

More information

Charles Yang, Dipankar Ghosh, Jonathan Beck Thermo Fisher Scientific, San Jose, CA, USA

Charles Yang, Dipankar Ghosh, Jonathan Beck Thermo Fisher Scientific, San Jose, CA, USA Utilization of High Resolution LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using a Q Exactive Benchtop Orbitrap Platform Charles Yang, Dipankar Ghosh, Jonathan Beck Thermo

More information

Screening Method for 30 Pesticides in Green Tea Extract Using Automated Online Sample Preparation with LC-MS/MS

Screening Method for 30 Pesticides in Green Tea Extract Using Automated Online Sample Preparation with LC-MS/MS Application Note: 514 Screening Method for 3 Pesticides in Green Tea Extract Using Automated Online Sample Preparation with LC-MS/MS Yang Shi, Catherine Lafontaine, Thermo Fisher Scientific, Franklin,

More information

Key Words Nanoparticles, spicp-ms, ICP-MS, Qtegra ISDS Software, AF4-ICP-MS

Key Words Nanoparticles, spicp-ms, ICP-MS, Qtegra ISDS Software, AF4-ICP-MS Nanoparticle Characterization Via Single Particle Inductively Coupled Plasma Mass Spectrometry (spicp-ms) Using a Dedicated Plug-in for Qtegra ISDS Software Technical Note 43279 Daniel Kutscher, Julian

More information

Analysis of a Verapamil Microsomal Incubation using Metabolite ID and Mass Frontier TM

Analysis of a Verapamil Microsomal Incubation using Metabolite ID and Mass Frontier TM Application Note: 320 Analysis of a Verapamil Microsomal Incubation using Metabolite ID and Mass Frontier TM Key Words Metabolism Study Structure Elucidation Metabolite ID Mass Frontier Chromatography

More information

Automated and accurate component detection using reference mass spectra

Automated and accurate component detection using reference mass spectra TECHNICAL NOTE 72703 Automated and accurate component detection using reference mass spectra Authors Barbara van Cann 1 and Amit Gujar 2 1 Thermo Fisher Scientific, Breda, NL 2 Thermo Fisher Scientific,

More information

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION A purpose-built collection of the best-inclass components for forensic toxicology Whether you re challenged to present reliable and secure forensic sample

More information

Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry

Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry Jon Hao, Rong Ye, and Mason Tao Poochon Scientific, Frederick, Maryland 21701 Abstract Background:

More information

Exploring Mixed-Mode Chromatography: Column Chemistry, Properties, and Applications

Exploring Mixed-Mode Chromatography: Column Chemistry, Properties, and Applications Exploring Mixed-Mode Chromatography: Column Chemistry, Properties, and Applications Xiaodong Liu, Mark Tracy, Udayanath Aich, and Christopher Pohl; Thermo Fisher Scientific, Sunnyvale, CA, USA Overview

More information

Thermo Scientific HyperSep Dispersive SPE Products. Efficient sample preparation and clean-up using the QuEChERS Method

Thermo Scientific HyperSep Dispersive SPE Products. Efficient sample preparation and clean-up using the QuEChERS Method Thermo Scientific HyperSep Dispersive SPE Products Efficient sample preparation and clean-up using the QuEChERS Method The QuEChERS method overcomes the problems associated with time consuming, expensive

More information

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS anion-exchange column is designed for the fast analysis of inorganic

More information

Measurement of Metabolic Stability Using SIM and Identification of Metabolites by Data Dependent Full-Scan MS/MS and CNL Scanning

Measurement of Metabolic Stability Using SIM and Identification of Metabolites by Data Dependent Full-Scan MS/MS and CNL Scanning Application Note: 349 Measurement of Metabolic Stability Using SIM and Identification of Metabolites by Data Dependent Full-Scan MS/MS and CNL Scanning Peter B Ehmer, Ethirajulu Kantharaj, Katie De Wagter,

More information

Complete Materials Deformulation Using TGA-IR

Complete Materials Deformulation Using TGA-IR Application Note: 51694 Complete Materials Deformulation Using TGA-IR Michael Bradley, Thermo Fisher Scientific, Madison, WI, USA Key Words Feedstock Supplies Finished Products Multi-component Search OMNIC

More information

EXPRESSION MORE PEPTIDES. MORE PROTEINS. MORE REPRODUCIBILITY. MORE QUANTIFICATION.

EXPRESSION MORE PEPTIDES. MORE PROTEINS. MORE REPRODUCIBILITY. MORE QUANTIFICATION. A New Era in Protein Quantification [ systems BIOLOGY Solutions ] [ systems BIOLOGY Solutions ] EXPRESSION MORE PEPTIDES. MORE PROTEINS. MORE REPRODUCIBILITY. MORE QUANTIFICATION. The Waters Expression

More information

SILAC and TMT. IDeA National Resource for Proteomics Workshop for Graduate Students and Post-docs Renny Lan 5/18/2017

SILAC and TMT. IDeA National Resource for Proteomics Workshop for Graduate Students and Post-docs Renny Lan 5/18/2017 SILAC and TMT IDeA National Resource for Proteomics Workshop for Graduate Students and Post-docs Renny Lan 5/18/2017 UHPLC peak chosen at 26.47 min LC Mass at 571.36 chosen for MS/MS MS/MS MS This is a

More information

Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry

Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry Cristian Cojocariu and Paul Silcock Thermo Fisher Scientific, Runcorn,

More information

Reagents. Affinity Tag (Biotin) Acid Cleavage Site. Figure 1. Cleavable ICAT Reagent Structure.

Reagents. Affinity Tag (Biotin) Acid Cleavage Site. Figure 1. Cleavable ICAT Reagent Structure. DATA SHEET Protein Expression Analysis Reagents Background The ultimate goal of proteomics is to identify and quantify proteins that are relevant to a given biological state; and to unearth networks of

More information

This document is a preview generated by EVS

This document is a preview generated by EVS TECHNICAL REPORT RAPPORT TECHNIQUE TECHNISCHER BERICHT CEN/TR 15641 August 2007 ICS 67.050 English Version Food analysis - Determination of pesticide residues by LC- MS/MS - Tandem mass spectrometric parameters

More information

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma Ken Meadows, Thermo Fisher Scientific, Runcorn, UK Application Note 20789 Key Words

More information

Relative Quantitation of TMT-Labeled Proteomes Focus on Sensitivity and Precision

Relative Quantitation of TMT-Labeled Proteomes Focus on Sensitivity and Precision Relative Quantitation of TMT-Labeled Proteomes Focus on Sensitivity and Precision R. Viner 1, M. Scigelova 2, M. Zeller 2, M. Oppermann 2, T. Moehring 2 and V. Zabrouskov 1 1 Thermo Fisher Scientific,

More information

Fast Protein and Peptide Separations Using Monolithic Nanocolumns and Capillary Columns

Fast Protein and Peptide Separations Using Monolithic Nanocolumns and Capillary Columns Application Note 3 Fast Protein and Peptide Separations Using Monolithic Nanocolumns and Capillary Columns INTRODUCTION Polymeric monolithic stationary phases offer an alternative to the classical microparticulate

More information

Peptide Targeted Quantification By High Resolution Mass Spectrometry A Paradigm Shift? Zhiqi Hao Thermo Fisher Scientific San Jose, CA

Peptide Targeted Quantification By High Resolution Mass Spectrometry A Paradigm Shift? Zhiqi Hao Thermo Fisher Scientific San Jose, CA Peptide Targeted Quantification By High Resolution Mass Spectrometry A Paradigm Shift? Zhiqi Hao Thermo Fisher Scientific San Jose, CA Proteomics is Turning Quantitative Hmmm.. Which ones are my targets?

More information