ACQUITY UPLC System. Application Notebook December 2005

Size: px
Start display at page:

Download "ACQUITY UPLC System. Application Notebook December 2005"

Transcription

1

2

3

4 4 ACQUITY UPLC SYSTEM DECEMBER 25 ACQUITY UPLC System Application Notebook December Recent Developments in Columns for UPLC Eric S. Grumbach, Thomas E. Wheat, and Jeffrey R. Mazzeo Transfer of the USP Human Insulin Related Compounds HPLC Method to the ACQUITY UPLC System Tanya L. Jenkins and Patricia McConville Validation of a UPLC Method for a Benzocaine, Butamben, and Tetracaine Hydrochloride Topical Solution Andrew J. Aubin and Tanya L. Jenkins Meeting Analytical Challenges with UPLC MS John Van Antwerp, Richard DePinto, and Dale Jansen Enabling Significant Improvements for Peptide Mapping with UPLC Jeffrey R. Mazzeo, Thomas E. Wheat, Beth L. Gillece-Castro, and Ziling Lu Analysis of Prostaglandins with the Quattro Premier XE Kate Yu and Donald Kwet

5

6 6 ACQUITY UPLC SYSTEM DECEMBER 25 Recent Developments in Columns for UPLC Eric S. Grumbach, Thomas E. Wheat, and Jeffrey R. Mazzeo Waters Corporation, Milford, Massachusetts Additional UPLC chemistries provide flexibility for methods development. Ultra Performance LC (UPLC ) enables chromatographers to meet the challenges of developing separations that completely characterize the constituents of samples by providing significant improvements in resolution, speed, and sensitivity. These improvements are achieved through the use of columns with 1.7 m particle packings and a matching family of low-dispersion, high-speed instrumentation developed simultaneously to provide full compatibility between chemistry and instrumentation. Recently, additional bonded phases have been developed to focus on improving resolution by manipulating selectivity. A 1.7 m particle-packed column provides significant improvements in resolution due to higher efficiencies. Separation sample components, however, still requires a bonded phase that provides both retention and selectivity. Four bonded phases are available for UPLC separations: Waters ACQUITY UPLC Bridged Ethyl-siloxane/silica Hybrid (BEH) C 18 and C 8 (straight chain alkyl columns), ACQUITY UPLC BEH Shield RP 18 (embedded polar group column), and ACQUITY UPLC BEH Phenyl (phenyl group tethered to the silyl functionality with a C 6 alkyl). Each provides a different combination of hydrophobicity, silanol activity, hydrolytic stability, and chemical interaction with the analytes. Experimental Conditions Columns: ACQUITY UPLC BEH C 18, C 8, Shield RP 18, Phenyl Column Dimensions: mm, 1.7 m Mobile Phase A: Water Mobile Phase B: Methanol Isocratic Conditions: 72 A; 28 B Flow Rate:.5 ml/min Injection Volume: 5. L Weak Needle Wash: 28 methanol Sample: Nitroaromatics Sample Diluent: 3 methanol Sample Concentration: 1 g/ml Temperature: 5 C Detection: 254 nm Sampling Rate: 2 pts/s Time Constant:.1 Instrument: Waters ACQUITY UPLC System with PDA Detector Results The analysis of nitroaromatics from EPA Method 833 is often performed on multiple stationary phases to resolve two sets of critical pairs. The high efficiency of UPLC columns provides the baseline separation of 14 nitroaromatics on the two linear alkane bonded phases (Figure 1). Compared to the C 18 column, the shorter chain length C 8 bonded phase is less hydrophobic, and, therefore, less retentive. Although selectivity differences seldom result from chain length differences, a change in peak elution has occurred. Embedded polar group columns often proved a selectivity that is unique compared to linear alkanes. The ACQUITY UPLC BEH Shield RP 18 column includes an embedded carbamate group that shows preferential retention of hydrogen-bond donors. Additional attributes include low silanol activity and aqueous compatibility. Columns with phenyl ligands provide yet another alternate selectivity. Due to the bonding orbital interactions, phenyl columns provide unique and specific selectivity with aromatic compounds and other analytes with similar electrons. Conclusions Stability over a broad ph operating range (ph 1 12) combined with the several available bonded phases, provides flexibility for methods development. This flexibility enables methods development to be more efficient, allowing products to be brought to market faster. The power of these ultra-efficient columns is combined with a low-dispersion Ultra Performance LC system to successfully transfer existing HPLC methods or to develop new, fast chromatographic methods that offer substantial improvements in resolution, sensitivity, and sample throughput.

7 DECEMBER 25 ACQUITY UPLC SYSTEM 7 Figure 1: Multiple ACQUITY UPLC bonded phases with different selectivity provide flexibility for methods development. Analytes: (1) HMX; (2) RDX; (3) 1,3,5-TNB; (4) 1,3-DNB; (5) NB; (6) Tetryl; (7) TNT; (8) 2-Am-4,6-DNT; (9) 4-Am-2,6-DNT; (1) 2,4-DNT; (11) 2,6-DNT; (12) 2-NT; (13) 4-NT; (14) 3-NT.

8 8 ACQUITY UPLC SYSTEM DECEMBER 25 Transfer of the USP Human Insulin Related Compounds HPLC Method to the ACQUITY UPLC System Tanya L. Jenkins and Patricia McConville Waters Corporation, Milford, Massachusetts The process of method transfer can be tedious and time-consuming, especially for a validated method. As a result, many LC methods developed using dated technology are still in use today. However, with the incentive of significant potential improvements in chromatographic separation speed and resolution, taking the necessary steps to transfer an existing protocol over to newer, better performing technology becomes less daunting. The transfer of the USP (United States Pharmacopeia) method for the related compounds of human insulin is particularly challenging, as it often requires re-optimization for each analysis. In this application note, we outline the successful method transfer steps from the USP-recommended mm, 5 m L1 column run on a traditional HPLC system to a Waters ACQUITY UPLC BEH mm, 1.7 m column on a Waters ACQUITY Ultra Performance LC System. To optimize the new method, experimental van Deemter curves were generated to determine the optimal flow rate. Due to the high molecular weight of the human insulin molecule ( 58), diffusion into the pores of the packing material is slower. Therefore, the optimal flow rate for the separation is relatively slow and its impact on run time was considered for the final method. Instrument considerations, such as detection parameters for sensitivity and injector carryover performance, were optimized to maximize the benefits of the ACQUITY UPLC system for this method. The resulting UPLC method reduced the overall run time by 6, with improved resolution. The data meets or exceeds acceptance criteria. Human insulin is a relatively small protein containing 51 amino acids in two polypeptide chains. The drug product is a biopharmaceutical synthesized through recombinant DNA technology. The insulin made by these cells is identical to the insulin made by the human pancreas. The USP method for the related compounds of human insulin is a 68-min separation involving the isocratic elution of the main component (within a specified retention window of min) and the major related compound (A21 desamido insulin), followed by the gradient elution of the high MW impurities (1). The run is followed by a re-equilibration step which is typically 23 min (5 column and 3 system volume). The elution time of the main insulin peak is extremely sensitive to the amount of organic modifier in the mobile phase. Adjustments to the gradient table are necessary for each analysis to ensure that the main insulin peak is eluted within the specified retention window to achieve sufficient resolution with A21 desamido insulin peak. A change of less than.5 of the organic modifier can cause the peak to elute outside of this retention window, making the on-line mixing and gradient performance of the instrumentation critical for this application. The method also utilizes high salt concentrations at low ph. Small changes in either of these parameters impact peak shape and selectivity. Elution times are also sensitive to changes in temperature. A difference of 1 2 C can push the retention of the main insulin peak outside of the specified window. Injector carryover is also an important consideration. Sample concentrations of up to 4 mg/ml are used to measure impurities and managing carryover can be challenging for HPLC injectors. A method is desired which will reduce the run time. Time is crucial since samples and standards are only useable for up to 12 h. The current USP method with a total analysis time of 91 min is long, considering the optimization time required before samples can be analyzed. However, any reduction of analysis time cannot sacrifice the performance of the method. Precision and carryover must still meet requirements. Method Scaling Equations The HPLC to UPLC method transfer process can be streamlined by using a series of equations to geometrically scale the original method to the new column dimensions (2). These equations account for changes in gradient times, flow rate, and injection volume. They do not compensate for changes in system volume, column selectivity or column load (mass or volume). To start, the gradient steps are scaled from the HPLC column to the UPLC column:

9 DECEMBER 25 ACQUITY UPLC SYSTEM 9 Figure 1: The ACQUITY UPLC System Console Calculator. Geometrically Scaled Gradient Time = UPLC Column Length HPLC Column Length Next, the flow rate is scaled taking into account the difference in the internal diameters of the columns: Geometrically Scaled Flow Rate = (UPLC Column Diameter) (HPLC Column Diameter) 2 HPLC Gradient Time 2 HPLC Flow Rate The flow rate should then be further scaled to take into account the new optimal linear velocity of the separation using the smaller 1.7 m particles. For 2.1 mm i.d. columns, appropriate starting points are typically 65 L/min for small molecules and 1 L/min for high MW compounds. To keep the column volumes proportional, the gradient steps should be readjusted for the new flow rate: UPLC Gradient Time = Scaled Flow Rate x Gradient Time UPLC Flow Rate The injection volume is scaled taking into account the volume of the columns: UPLC Injection Volume = UPLC Column Volume HPLC Column Volume HPLC Injection Volume Alternatively, the ACQUITY UPLC System Console Calculator (Figure 1) will scale the method automatically using the same scaling principles outlined here.

10 1 ACQUITY UPLC SYSTEM DECEMBER 25 Absorbance (AU) Insulin A Time (min) Figure 2: Chromatogram of the 68-min USP HPLC method for the human insulin related compounds assay. Original USP HPLC Method System: Alliance 2695 XC Separations Module; 2996 Photodiode Array Detector; Empower Software Column: Symmetry 3 C 18, 4.6 mm 25 mm, 5 m Detection: 214 nm, 1.2 nm digital bandwidth, 1 pt/s, 1. s time constant Eluents: Buffer:.2 M Sodium Sulphate, ph 2.3; A: 82/18 Buffer/Acetonitrile; B: 5/5 Buffer/Acetonitrile Gradient: 36 min isocratic at 24.5 B min linear gradient to 64 B min isocratic at 64 B min linear gradient to 24.5 B RT Window: min Flow Rate: 1 ml/min Temp: 35 C Sample: USP Human Insulin Reference Standard; 3.75 mg/ml in.1 M Hydrochloric Acid Volume: 2 L Wash: Extended Wash Cycle with 6:3:1 of.1 Phosphoric Acid: Acetonitrile: Isopropyl Alcohol Final Transferred UPLC Method System: ACQUITY UPLC System with Photodiode Array Detector; Empower Software Column: ACQUITY UPLC BEH C 18, 2.1 mm 1 mm, 1.7 m Detection: 214 nm, 12. nm digital bandwidth, 5 pts/s,.3 s time constant Eluents: Buffer:.2 M Sodium Sulphate, ph 2.3; A: 82/18 Buffer/Acetonitrile B: 5/5 Buffer/Acetonitrile Gradient: 14.4 min isocratic at 26 B min linear gradient to 64 B min isocratic at 64 B min linear gradient to 26 B RT Window: 6 1 min Flow Rate: 28 L/min Temp: 35 C Sample: USP Human Insulin Reference Standard; 1.25 mg/ml in.1 M Hydrochloric Acid Volume: 1.8 L characterized full loop Strong Wash: 2 L of 6:3:1 of.1 Phosphoric Acid: Acetonitrile: Isopropyl Alcohol Weak Wash: 12 L of.1 M Hydrochloric Acid

11 DECEMBER 25 ACQUITY UPLC SYSTEM Insulin Absorbance (AU) A Time (min) Figure 3: Chromatogram of the 27-min UPLC method for the human insulin related compounds assay. 25, Number of theoretical plates per meter 2, 15, 1, 5, Large molecule MW = 588 Small molecule MW = Flow rate (µl/min) Figure 4: Dependence of the number of theoretical plates on the flow rate.

12 12 ACQUITY UPLC SYSTEM DECEMBER Absorbance (AU) High concentration standard Absorbance (AU) Blank Time (min) Time (min) Figure 5: Carryover of human insulin on the ACQUITY UPLC system. Results and Discussion The USP method was scaled geometrically and then optimized, according to the USP specifications, to have the main insulin peak elute within the required retention window. This process involved increasing the organic modifier composition by 1.5 for the isocratic portion of the separation. The scaled injection volume and sample concentration were adjusted such that the height of the main insulin peak was within the linear range of the detector. The resulting UPLC method had a run time of 27.2 min (a 6 time reduction compared to the HPLC method), with a total analysis time of 37.5 min (compared to the 91-min HPLC analysis a time reduction of 54 min). To determine the optimal flow rate for the UPLC column, a plot of the number of theoretical plates versus flow rate was generated (Figure 4). For small molecules, the optimal flow rate for a 2.1 mm i.d. column packed with 1.7 m particles is typically around 65 L/min. However, for larger molecules the optimal flow rate is lower, and for the human insulin method it was under 1 L/min. Since the goal of this method transfer is to improve run time, the final method was chosen to operate above the optimal flow rate. Even when operating above the optimum flow rate for the separation, improvements in resolution were observed. The relative peak widths for the UPLC separation (Figure 3) were narrower than those of the orginal HPLC method (Figure 2). Improved resolution was attained with an impurity in the tail of the main insulin peak, as well as with some of the other lower level impurities. The HPLC method had resolution between the main insulin peak and the A21 desamido insulin peak of 5., while the UPLC method had a resolution of 6.8. Signal-to-noise for the A21 desamido insulin peak was 1:1. The narrow peaks generated by the UPLC separation required the detection parameters to be re-optimized. The data rate was increased and the filtering constant decreased to be more compatible with the chromatography and to maximize sensitivity. The performance of the human insulin method on the ACQUITY UPLC system was excellent (Table I), with precision values far below typical requirements. Retention time reproducibility was under.5 at.23. The area reproducibility of the main insulin peak was below 1 at.21 and the A21 desamido insulin peak was well below 1 at only 1.6. The dual wash capabilities of the ACQUITY UPLC system efficiently removes residual sample from the system, resulting in negligible carryover. The wash cycle does not contribute to injection cycle time as it occurs post-injection, while the sample is running. To measure carryover, a set of high concentration standards (1 mg/ml) was used. The injection sequence was 2 blanks, 3 standards at.5 (carryover specification), 6 high concentration standards, and 3 blanks. Table II shows the first injection after the blank had a carryover of.4 and by the third blank injection, no insulin could be detected. Figure 5 shows the high concentration standard compared to the first blank injection.

13 DECEMBER 25 ACQUITY UPLC SYSTEM 13 Table I: Reproducibility values for the human insulin UPLC method. Peak Retention Time RSD Peak Area RSD Human Insulin A21 Desamido Insulin Table II: Carryover of human insulin in subsequent blank injections. Injection Average Peak Height mau Carryover Standard Blank Blank Blank 3 none detected none detected Conclusions We have shown that challenging USP HPLC methods can be successfully transferred to the Waters ACQUITY UPLC system, with numerous chromatographic performance benefits. Transfer of the human insulin method was accomplished by simple scaling with an adjustment to the sample concentration and injection volume. The gradient table was adjusted for the new method according to the USP guidelines. The new UPLC method offered a 6 reduction in run time over the USP HPLC method (a 54- min reduction in total analysis time) with improved resolution. UPLC reproducibility for both peak area and retention time was very good, especially for such a chromatographically sensitive sample. In addition, sample carryover was easily managed and minimized by the ACQUITY UPLC system. This exercise clearly demonstrates the advantages that can be realized by transferring from HPLC to UPLC methodology. References (1) The United States Pharmacopeia USP28, the National Formulary NF23 (United States Pharmacopeial Convention, Inc., 24), pp (2) E. Grumbach, T. Wheat, and J. Mazzeo, An Efficient HPLC to UPLC Method Transfer Protocol, poster presented at HPLC 25, June 25, Stockholm, Sweden (Waters Literature Code WA41923).

14 14 ACQUITY UPLC SYSTEM DECEMBER 25 Validation of a UPLC Method for a Benzocaine, Butamben, and Tetracaine Hydrochloride Topical Solution Andrew J. Aubin and Tanya L. Jenkins Waters Corporation, Milford, Massachusetts Benzocaine (4-Aminobenzoic acid ethyl ester), Butamben (Butyl 4-aminobenzoate), and Tetracaine (4 (Butylamino)benzoic acid 2-(dimethylamino)ethyl ester) are topical anesthetics. The formulated mixture of these three active ingredients is indicated for the production of anesthesia of all accessible mucous membranes except the eyes. In this application note, we demonstrate the successful development and validation of an Ultra Performance LC (UPLC ) method for the analysis of these compounds. Experimental Materials Benzocaine, butamben, tetracaine HCl, benzalkonium chloride, cetyl dimethyl ethyl ammonium bromide, ammonium bicarbonate, and ammonium hydroxide were purchased from Sigma-Aldrich Co. (St Louis, Missouri). Water was purified with a Milli-Q system (Millipore, Billerica, Massachusetts). Acetonitrile was purchased from Mallinckrodt Baker, Inc. (Phillipsburg, New Jersey) and was of HPLC grade. The formulated topical solution was purchased from a local pharmaceutical supplier and was labeled to contain 14. benzocaine and 2. each of butamben and tetracaine HCl as well as the inactive ingredients benzalkonium chloride (.5) and cetyl dimethyl ethyl ammonium bromide (.5). All bottles of the topical solutions were of the same production lot. UPLC Apparatus and Conditions The Waters ACQUITY UPLC System consisted of a Binary Solvent Manager, Sample Manager, and an ACQUITY UPLC Tunable UV (TUV) Detector. System control, data collection, and data processing were accomplished using Waters Empower 2 Chromatography Data Software. Separations were performed at 4 C using an ACQUITY UPLC Bridged Ethyl-siloxane/silica Hybrid (BEH) C m, 2.1 mm 5 mm column. Elution Absorbance (AU) Benzocaine.29 min Butamben.65 min Tetracaine 1.5 min Time (min) Figure 1: UPLC separation of Benzocaine, Butamben, and Tetracaine.

15 DECEMBER 25 ACQUITY UPLC SYSTEM 15 Table I: Detector timetable. Time range (min) Wavelength (nm) Compound RT (min) Benzocaine Butamben Tetracaine 1.5 Table II: Working standard solutions. Standard Benzocaine Stock (ml) Butamben Tetracaine HCl Stock (ml) Approximate concentrations of working solutions (mg/ml) Benzocaine Butamben Tetracaine HCl Level Level Level Level Level x min Intensity Intensity Intensity min 2.x min 3.x m/z Figure 2: Mass spectra of benzocaine (top), butamben (middle), and tetracaine (bottom).

16 16 ACQUITY UPLC SYSTEM DECEMBER 25 was accomplished using a 1. mm ammonium bicarbonate mobile phase (adjusted to ph 1. with ammonium hydroxide) filtered through at.45 m membrane, and acetonitrile (6:4) at 1. ml/min., one microliter injections were used for both standards and samples. The sample manager was equipped with a 5 L loop and used the Partial Loop Needle Overfill mode with 1 L air gaps. To minimize carryover, 12 L of the weak wash (6:4 water/acetonitrile) and 4 L of the strong wash (1:9 water/acetonitrile) were used. Detection wavelengths were optimized for each compound and collected on a timed event basis (Table I). A sampling rate of 2 points/s and a filter constant of.1 s were used throughout. The total run time was 1.5 min (Figure 1). Preparation of Standard Solutions Stock Solutions A standard solution (2. ml) of benzocaine at 1. mg/ml and a standard solution (2. ml) containing.15 mg/ml each of butamben and tetracaine hydrochloride was prepared by dissolving appropriate amounts of each in the diluent (a mixture of 1. mm Table III: Specificity results. Name RT (min) Resolution Purity Angle Purity Threshold Unknown Benzocaine Unknown Butamben Tetracaine Table IV: Linearity results. Name R^2 Y Intercept Difference Equation Benzocaine Y 2.49e 6X 5.47e 3 Butamben Y 4.71e 6 X 5.39e 2 Tetracaine Y 4.24e 6X 2.9e 2 91, 819, 728, 637, 546, Area 455, 364, 273, 182, 91, Amount Figure 3: Calibration curve for benzocaine.

17 DECEMBER 25 ACQUITY UPLC SYSTEM ,. 192,. 168,. Butamben 144,. 12,. Area 96,. Tetracaine 72,. 48,. 24, Amount Figure 4: Calibration curves for butamben and tetracaine. Table V: Repeatability at the 5 level (n 6). RT (min) Benzocaine Butamben Tetracaine Mean Std. Dev RSD Amount Mean Std. Dev RSD Table VI: Repeatability at the 1 level (n 6). RT (min) Benzocaine Butamben Tetracaine Mean Std. Dev RSD Amount Mean Std. Dev RSD

18 18 ACQUITY UPLC SYSTEM DECEMBER Residual deviation () Standard level Figure 5: Residuals plot for standard curves of benzocaine ( ), butamben ( ), and tetracaine ( ). Table VII: Repeatability at the 15 level (n 6). RT (min) Benzocaine Butamben Tetracaine Mean Std. Dev RSD Amount Mean Std. Dev RSD ammonium bicarbonate at ph 1. and acetonitrile (6:4)). Working Solutions Working standard solutions (5 levels, 1. ml of each) were prepared by diluting appropriate amounts of stock solutions in the diluent to give concentrations ranging from.1.3 mg/ml for benzocaine and mg/ml for butamben and tetracaine hydrochloride (Table II). Topical Solution Sample Preparation About 3 mg of Topical Solution was accurately weighed and transferred to a 2 ml volumetric flask. Approximately 18 ml of diluent was added and mixed on a wrist action shaker for 5 min. The sample was diluted to final volume and mixed thoroughly. An aliquot of the sample was transferred to a 2-mL vial and capped. Chromatographic Analysis Procedure Equal volumes (1. L) of the 5 standard preparations and the topical solution sample preparation were separately injected into the chromatograph, chromatograms recorded, peak areas were measured for the 3 major peaks. A calibration curve was then created based on the 5 standard preparations. The amounts were calculated, in mg/ml, for benzocaine, butamben and tetracaine hydrochloride in the portion of topical solution sample preparation by comparing the area of each peak to the linear regression line of the calibration curve, based on the five standard preparations. The amount was calculated, by weight of benzocaine, butamben, and tetracaine hydrochloride in the topical solution using the formula: Concentration (mg/ml) (2 ml Weight Topical Solution in Volumetric (mg)) 1

19 DECEMBER 25 ACQUITY UPLC SYSTEM 19 The current monograph from the USP states that Benzocaine (14), Butamben (2), and Tetracaine Hydrochloride (2) Topical Solution should contain not less than 9. and not more than 11. of the labeled amounts of benzocaine, butamben, and tetracaine hydrochloride. Suitability Criteria Based on the chromatograms of standard preparations, the retention times (in minutes) are about.3 for benzocaine,.65 for butamben, and 1.5 for tetracaine; the resolution, R, between the benzocaine peak and the butamben peak and between the butamben peak and the tetracaine peak is not less than 8; and the relative standard deviation for retention time of replicate injections is not more than 1. for each of the 3 analyte peaks. The relative standard deviation for amount of replicate injections is not more than 2. for each of the 3 analyte peaks. Assay Validation and Results Specificity To demonstrate specificity, injections of individual reference standard, a diluent blank, and a solution that contains approximate concentration of labeled nonactive ingredients were evaluated. The reference standards and diluent blanks were prepared using the same protocols as the assay preparation to ensure no contamination is added during the sample preparation steps. Samples were evaluated using photodiode array detection and MS detection and peak purity were assessed using Empower 2 software. Acceptance Criteria for Specificity The diluent blanks must show no interference with the peaks of the active ingredients. Resolution between the main sample peaks should be greater than or equal to 2.. The resolution factors between all sample peaks and any other peak should be greater than or equal to 2.. Any co-elution of significant nonactive ingredient, excipient or unknown peaks must be clearly noted in the final reporting of the results. To determine peak purity, the threshold must exceed the peak purity angle for the main sample peaks. Table VIII: Intermediate precision results. Active Benzocaine Butamben Tetracaine Analyst 1 Analyst 2 Diff Analyst 1 Analyst 2 Diff Analyst 1 Analyst 2 Diff Mean Std. Dev RSD Table IX: Reproducibility results. Active Benzocaine Butamben Tetracaine Lab 1 Lab 2 Diff Lab 1 Lab 2 Diff Lab 1 Lab 2 Diff Mean Std. Dev RSD Table X: Tested robustness criteria. Parameter Specified Conditions Modified Conditions Wavelength (nm) 22, 29, , 295, 312, and 215, 285, 32 Flow Rate (ml/min) 1..9,.95, 1.5, 1.1 Column Temperature ( C) 4 38, 39, 41, 42 Injection Volume ( L) 1..8,.9, 1.1, 1.2 Mobile Phase Composition 6/4 5/5, 55/45, 65/35, 7/3 Buffer Concentration mmol 1 8, 9, 11, 12 Buffer ph 1. 9., 9.5, 1.5, 11. Sample Prep Shake Time (min) 5 2, 5, 1

20 2 ACQUITY UPLC SYSTEM DECEMBER 25 Table XI: Method robustness results. Method Condition Benzocaine Butamben Tetracaine Wavelengths 5 nm *Specified Wavelengths Wavelengths 5 nm Flow Rate.9 ml/min Flow Rate.95 ml/min *Flow Rate 1. ml/min Flow Rate 1.5 ml/min Flow Rate 1.1 ml/min Column Temperature 38 C Column Temperature 39 C *Column Temperature 4 C Column Temperature 41 C Column Temperature 42 C Injection Volume.8 L Injection Volume.9 L *Injection Volume 1. L Injection Volume 1.1 L Injection Volume 1.2 L Mobile Phase Composition 5/ Mobile Phase Composition 55/ *Mobile Phase Composition 6/ Mobile Phase Composition 65/ Mobile Phase Composition 7/ Buffer Concentration 8 mmol Buffer Concentration 9 mmol *Buffer Concentration 1 mmol Buffer Concentration 11 mmol Buffer Concentration 12 mmol Buffer ph Buffer ph *Buffer ph Buffer ph Buffer ph Sample Prep Shake Time 2 min *Sample Prep Shake Time 5 min Sample Prep Shake Time 1 min * Conditions as specified in the method. 1 Variation of 2. compared to the result obtained at the specified method conditions.

21 DECEMBER 25 ACQUITY UPLC SYSTEM 21 Table XII: Recovery of spiked assay samples at 3 levels (n 6). Benzocaine Butamben Tetracaine Spiked at 8 of label Spiked at 1 of label Spiked at 12 of label Specificity Results Injections of a diluent blank showed no extraneous peaks. Injections of individual reference standards gave the following retention times: benzocaine.29 min, butamben.65 min, and tetracaine 1.5 min. No other peaks were noted in the chromatograms from the individual standards. A solution that contained the inactive ingredients benzalkonium chloride (.5) and cetyl dimethyl ethyl ammonium bromide (.5) in water was analyzed. No peaks were noted in the regions of the active ingredients. Photodiode array analysis and single quadrupole mass spectrometry detection were applied to a sample. Resolution requirements outlined in the acceptance criteria for specificity were easily met (Table III). Photodiode array peak purity results of the sample showed purity angles less than purity thresholds for the 3 peaks of interest, indicating each peak had a high degree of peak homogeneity. Evaluation of the mass spectral data confirmed spectral identity of the 3 peaks of interest (Figure 2). Based on these results, the method was determined to be specific for the analysis of benzocaine, butamben, and tetracaine. Linearity Serial dilutions from a stock standard solution were made to obtain 5 linearity solutions incorporating levels from 5 to 15 of the nominal analytical concentration of the sample components. Each solution was injected in duplicate. The linear regression of these injections provided the correlation coefficient, y-intercept, and the residual sum of squares. Acceptance Criteria for Linearity The correlation coefficient should not be less than.999 for all the sample components and the y-intercept should be within 2. when compared to the 1 level. The plot of the residuals should indicate linearity. Linearity Results The linearity of all 3 compounds met the criteria described above (Table IV). Linearity plots are shown in Figures 3 and 4. Residuals at all 5 standard levels were all less than 2 and indicated linearity (Figure 5). Precision Repeatability The repeatability of the main sample peaks was measured by making 6 replicate injections of a single standard solution at each of 3 levels (5, 1, and 15). Acceptance Criteria for Precision Repeatability The RSD for amount should not be more than 2. for each of the main sample components at each level. The RSD for retention time should not be more than 1. for each of the main sample components at each level. Precision Repeatability Results The repeatability for all three compounds met the acceptance criteria (Tables V VII). Precision Intermediate Precision To establish the effects of random events on the precision of the analytical procedure, 2 analysts prepared and analyzed 6 sample assay preparations from 1 batch and 2 preparations each from 2 additional batches. Each analyst prepared their own standards and solutions, used a column from a different lot, and used different systems to evaluate the sample solutions. Acceptance Criteria for Precision-Intermediate Precision The RSD obtained for the 1 preparations for active ingredient by both analysts should not be more than 2.. The mean analyst 1 result for each batch tested should not differ from the mean analyst 2 result for sample peaks by more than 2.. Precision Intermediate Precision Results Intermediate precision results for both analysts met the criteria described above, demonstrating acceptable intermediate precision for the assay. Results are summarized in Table VIII. Precision Reproducibility Reproducibility is assessed by means of an inter-laboratory trial. Analysts from 2 different labs (different from the analysts involved in the intermediate precision) prepared and analyzed 6 sample assay preparations from one batch and 2 preparations each from 2 additional batches. Each analyst prepared their own standards, solutions, used a column from a different lot, and different systems to evaluate the sample solutions. Acceptance Criteria for Precision-Reproducibility The RSD obtained for the 6 preparations by both analysts should not be more than 2.. The mean lab 1 result for each batch tested should not differ from the mean lab 2 result for the sample peaks by more than 2.. Precision Reproducibility Results Precision reproducibility results from both laboratories met the criteria described above demonstrating acceptable reproducibility for the assay. Results are summarized in Table IX.

22 22 ACQUITY UPLC SYSTEM DECEMBER 25 Robustness Robustness is the ability of a method to remain unaffected by small changes in conditions. If the changes are within the limits that produce acceptable chromatography, the method is considered robust. Six replicate injections of a standard and a sample solution were performed under both the specified and modified conditions shown in Table X. Acceptance Criteria For Robustness The average values obtained at the modified conditions should not differ by more than 2. for the sample components from those values obtained using the specified conditions. Robustness Results The results for the various method modifications are summarized in Table XI. For changes applied to the method, calculated amounts of the active ingredients did not vary by more that 2 (relative to the specified method conditions), except in the case of column temperature, where the results from the 38 C column temperature fell outside of this range for all 3 active ingredients (Table XI). These results demonstrate that the method was robust, although care must be taken to ensure the specified column temperature. Accuracy/Recovery The 3 analytes were spiked into a blank sample matrix at 3 levels: 8 of label, 1 of label, and 12 of label. These spiked samples were prepared according to topical solution sample preparation directions and analyzed according to the chromatographic analysis procedure. Amounts were determined using the same quantitation procedure as was used in the final method procedure. The percent recovery was then calculated. Acceptance Criteria For Accuracy/Recovery The measured value of the sample peaks in the spiked placebos should be within 6 2. of the spiked value. Accuracy/Recovery Results The percent recovery values are summarized in Table XII. Recovery values for all 3 compounds at the 3 spike levels were within 2 of the expected value. Conclusions A linear, accurate, selective, robust, and reproducible UPLC method for the determination of Benzocaine, Butamben, and Tetracaine Hydrochloride Topical Solution was successfully validated. By taking advantage of ACQUITY UPLC system performance and BEH C m columns, the entire method validation process was shortened from weeks to days, yielding a rugged method with superior resolution and speed. Acknowledgments The technical assistance of Katherine Hynes and the participation of Mark Benvenuti and Michael D. Jones are gratefully acknowledged.

23 DECEMBER 25 ACQUITY UPLC SYSTEM 23 Meeting Analytical Challenges with UPLC MS John Van Antwerp, Richard DePinto, and Dale Jansen Waters Corporation, Milford, Massachusetts The utilization of single quadrupole mass spectrometers for drug discovery and development has increased significantly over the past five years, particularly for Open Access walk-up and Mass-Directed Autopurification environments. The ongoing demand for higher throughput and increased speed, sensitivity, and resolution continues to drive chemists to seek out the latest in high performance technologies. Recently, with the introduction of the Waters ACQUITY Ultra Performance LC System, many chemists are considering the addition of this breakthrough chromatographic technology to further improve their overall data quality and productivity. To demonstrate, we have utilized the ACQUITY UPLC System with a Waters ZQ 2 Mass Detector for the analysis of a representative selection of pharmaceutical compounds. The results address a common concern when peaks are very narrow (as with UPLC), and the resulting MS spectral quality from such fast analyses. In this applications note, we show the outstanding performance of the ZQ 2 with narrow peaks in both an ESI mode (Electrospray Ionization) and in an ESCi multi-mode ionization (both ESI and APCI) in the analysis of a multi-component drug test mixture (Figure 1). Objectives When comparing UPLC MS to traditional HPLC MS analysis, the following high performance benefits can be expected: Higher sample throughput Enhanced qualitative detection and sensitivity Increased selectivity and resolution More quality information obtained in less time Added capabilities from a single benchtop system Improved overall efficiency and productivity Method 1 (See Figures 1 2) LC Conditions LC System: ACQUITY UPLC System Column: ACQUITY UPLC BEH C 18, 2.1 mm 5 mm, 1.7 m Flow Rate:.75 ml/min Intensity 1.1x1 8 1.x1 8 9.x1 7 8.x1 7 7.x1 7 6.x1 7 5.x1 7 4.x1 7 3.x1 7 1.x1 7 1.x1 7 Ephedrine Caffeine Doxylamine Lidocaine -.62 Acetanilide Propraolol Diphenhydramine -.86 Bromazapam Trimipramine Oxybutynin Terfenadine Diazapam Time (min) Figure 1: UPLC MS enables accelerated, high-resolution separations with second typical peak widths.

24 24 ACQUITY UPLC SYSTEM DECEMBER Terfendine Trimipramine Propranolol Lidocaine Caffeine Figure 2: Individual spectra for five compounds in the test mix, where [M H] ions are easily discernable. Compound name: terfenadine Correlation coefficient r = , r 2 = Calibration curve: * x Response type: External Std, Area Curve type: Linear, Origin: Exclude, Weighing" 1/x, Axis trans: None 26, 24, 22, 2, 18, 16, Response 14, 12, 1, Quantify Compound Summary Report 8, 6, 4, Compound 2: Terfenadine Name Type Std. Conc RT Area µg /ml Dev UPLC_ZQ_12 Standard UPLC_ZQ_13 Standard UPLC_ZQ_14 Standard UPLC_ZQ_15 Standard UPLC_ZQ_16 Standard UPLC_ZQ_17 Standard UPLC_ZQ_18 Standard UPLC_ZQ_19 Standard UPLC_ZQ_2 Standard , Concentration (µg/ml) Figure 3: Excellent linearity is demonstrated for Terfenadine.

25 DECEMBER 25 ACQUITY UPLC SYSTEM 25 std UPLC_ZQ_ pg/µl each 1 pg on column 2-s peak widths Trimipramine : SIR of 2 Channels ES UPLC_ZQ_12 1: SIR of 2 Channels ES Terfenadine UPLC_ZQ_12 1: SIR of 2 Channels ES TIC Time Figure 4: UPLC MS limits of detection demonstrated with 1 pg on column for the analysis of Trimipramine and Terfenadine. Injection: 2 L Mobile Phase A:.1 formic acid in H 2 O Mobile Phase B:.1 formic acid in acetonitrile Gradient: 5 to 1 B in.5 min MS Conditions MS System: ZQ 2 Mass Detector Scan Range: 15 7 Da Scan Time:.11 s Dwell Time: 5 ms Interscan Delay:.2 s. Cycle Time:.13 s Ionization Mode: ESI Method 2 (See Figures 3 5) LC Conditions LC System: ACQUITY UPLC System Column: ACQUITY UPLC BEH C 18, 2.1 mm 5 mm, 1.7 m Flow Rate:.85 ml/min., 4 L split flow to MS Injection: 2 L Mobile Phase A:.1 formic acid in H 2 O Mobile Phase B:.1 formic acid in acetonitrile Gradient: 1 B to 9 B in 3 min MS Conditions MS System: ZQ 2 Mass Detector Scan Range: 15 7 Da Scan Time:.11 s Dwell Time: 5 ms Interchannel Delay:.2 s Interscan Delay:.2 s Cycle Time:.14 s Ionization Mode: ESI SIR Mode: m/z 295 (for Trimipramine) and m/z 472 (for Terfenadine) Results MS Spectral Quality Narrow Peaks Separations were accomplished in less then 3 s and MS spectral quality allows easy identification of all compounds (Figure 2). This is an indication of what can be achieved with UPLC and a single quadrupole MS in a high throughput laboratory when faster chromatography is the objective. Excellent Linearity Any analytical method requires a linear response relative to analyte concentration. In Figure 3, excellent linearity of response for Terfenadine is demonstrated for the UPLC MS method. Lower Limits of Detection The detectability of low amounts (pg) is also required. The combination of the enhanced resolution and peak concentration afforded by ACQUITY UPLC separations and the sensitivity and selectivity of the ZQ 2 mass detection enables the reliable detection of minor constituents (Figure 4).

26 26 ACQUITY UPLC SYSTEM DECEMBER 25 Quantify Compound Summary Report Compound 1: trimpiramine # Name Sample Text Type Std. Conc RT Area ug/ml Dev UPLC_ZQ_21 std 7 Standard UPLC_ZQ_22 std 7 Standard UPLC_ZQ_23 std 7 Standard UPLC_ZQ_24 std 7 Standard UPLC_ZQ_25 std 7 Standard UPLC_ZQ_26 std 7 Standard Ave: Std dev: RSD: Compound 2: terfenadine # Name Sample Text Type Std. Conc RT Area ug/ml Dev UPLC_ZQ_21 std 7 Standard UPLC_ZQ_22 std 7 Standard UPLC_ZQ_23 std 7 Standard UPLC_ZQ_24 std 7 Standard UPLC_ZQ_25 std 7 Standard UPLC_ZQ_26 std 7 Standard Ave: Std dev: RSD: Figure 5: UPLC MS reproducibility for Trimipramine and Terfenadine. Precision of Quantitation The quality of UPLC separations coupled with the high scan rate capabilities of the ZQ yield exceptional analysis precision (Figure 5). Conclusion The robust UPLC MS platform yields exceptional throughput, speed, and sensitivity for routine analyses. As the expectation to produce more quality data in less time becomes standard, the unique combined capabilities of the ACQUITY UPLC System with the ZQ 2 Mass Detector provide more resolution faster, thus allowing overall throughput to be increased without compromising data quality. This superior data is achieved without the need to resort to very high analytical flow rates, which compromises sensitivity. As a result of this optimized performance, this system solution delivers the uptime, throughput and productivity necessary for any high throughput analytical laboratory.

27 DECEMBER 25 ACQUITY UPLC SYSTEM 27 Enabling Significant Improvements for Peptide Mapping with UPLC Jeffrey R. Mazzeo, Thomas E. Wheat, Beth L. Gillece-Castro, and Ziling Lu Waters Corporation, Milford, Massachusetts Peptide mapping continues to be the preferred technique for the comprehensive characterization of biopharmaceutical products. Its applications include: the identification of proteins based on the elution pattern of peptide fragments, the determination of post-translational modifications, the confirmation of genetic stability, and the analysis of protein sequences when interfaced to a mass spectrometer. In a peptide map, it is necessary to resolve each peptide fragment into a single peak. Therefore, peptide mapping represents a significant chromatographic challenge, due to the inherent complexity of protein digests. In addition to the large number of peptide fragments that are generated from the enzymatic digestion of a protein, the number of alternative peptide structures (e.g., post-translational modifications, oxidations, etc.) can be significant. The capabilities of Ultra Performance LC (UPLC ) technologies make higher resolution peptide mapping possible. This application note demonstrates the advantages of UPLC for peptide mapping. Experimental LC Conditions LC System: Waters ACQUITY UPLC System Columns: UPLC: ACQUITY UPLC Bridged Ethyl-siloxane/silica Hybrid (BEH) C 18 ; and mm, 1.7 m HPLC: BioSuite PA-A mm, 3 m; and BioSuite PA-B mm, 3.5 m Flow Rate:.1 or.3 ml/min Mobile Phase: A:.2 TFA or.1 FA in H 2 O B:.18 TFA or.1 FA in ACN Gradient: Linear gradients from 5 5 B, times as indicated Temperature: 4 C Injection Volume: 2 or 5 L Detection: ESI MS Samples: MassPREP Peptide Standard Mixture; MassPREP Enolase Digest; Tryptic digest of -1 acid glycoprotein MS Conditions MS System: Waters Micromass Q-Tof micro Mass Spectrometer Software: MassLynx 4. Software, SP µl/min 2.1 mm column 25 µl/min 2.1 mm column.2 Increasing Resolution H (cm).15.1 Optimum resolution 3.5 µm packing 1.7 µm packing u (cm/s) Increasing Flow Rate Figure 1: van Deemter plot of a 15 Da peptide.

28 28 ACQUITY UPLC SYSTEM DECEMBER µl/min Peak 1 3 µl/min 33 µl 1 µl/min 2 µl 2 45 µl 33 µl 3 39 µl 25 µl TIC µl/min µl 27 µl 27 µl 3 µl 36 µl 27 µl 19 µl 18 µl 18 µl 19 µl 2 µl 18 µl TIC µl 24 µl Time (min) Figure 2: Effect of flow rate on the MassPREP peptide standard mixture. Both chromatograms are plotted to the same intensity scale. Results and Discussion The chromatographic benefits of UPLC are largely derived from reduced band broadening that is, in turn, a consequence of reduced diffusion distances in small particles. This process is quantitatively described in the van Deemter equation that relates height equivalent of a theoretical plate to linear velocity. This relationship is shown graphically in Figure 1 for a peptide of 15 MW on 3.5 m and 1.7 m packings. The minimum in the curve corresponds to the maximum efficiency, and greatest resolving power, for each particle size. At linear velocities, or flow rates, above and below the optimum, resolving power declines. As expected, the smaller particles have higher resolving power at a higher linear velocity. In quantitative terms, the 3.5 m particles have a minimum plate height of 8.11 m at a linear velocity of.17 mm/s. In contrast, a minimum plate height of 3.94 m is observed at.33 mm/s with the 1.7 m particles. In practical terms, these principles suggest that the small particles used in UPLC could double the resolving power in a peptide mapping experiment and could simultaneously reduce the separation time because the optimum is achieved at a higher linear velocity. For the 3.5 m particle, the optimum linear velocity corresponds to a flow rate of about 24 L/min. on a 2.1 mm i.d. column. In practice, such a flow rate would never be used for a peptide map because the separation times would be far too long. It is common practice to operate at a higher flow rate, typically about 25 L/min. on 2.1 mm columns. This linear velocity of about 1.7 mm/s corresponds to a plate height of about 21 m. This loss of resolution with a 1-fold increase in separation speed has come to be an accepted compromise. For 1.7 m particles, resolution is much better preserved at the higher flow rate. These chromatographic principles suggest several ways to approach improving peptide maps using UPLC. First, the smaller particle packing will improve both resolution and sensitivity by reducing diffusion-related band broadening. Second, the reduced plate height is consistent with obtaining the same or better resolution with shorter columns and higher flow rates. Third, the compromise between separation time and resolution will be more favorable with the smaller particles. The influence of volumetric flow rate on peptide separation performance with 2.1 mm i.d. columns was investigated. A standard peptide mixture was separated on a UPLC column run at 1 L/min and at 3 L/min, as shown in Figure 2. Flow rate, or linear velocity, was the only variable because the gradient change/column volume was the same, ensuring that

29 DECEMBER 25 ACQUITY UPLC SYSTEM ± ± ± ±.24 TIC 3.54± Time (min) Figure 3: UPLC retention time reproducibility for peptide mapping Traditional C µm 2.1 x 1 mm TIC ACQUITY UPLC BEH C µm 2.1 x 1 mm TIC Time (min) Figure 4: Suitability of UPLC for peptide mapping.

30 3 ACQUITY UPLC SYSTEM DECEMBER 25 1 BEH C µm 2.1 x 5 mm TIC Traditional C µm 2.1 x 1 mm, 3 Å TIC Time (min) 4. Figure 5: Reduced column length for increased speed. the chromatographic selectivity is constant. In experimental terms, a 75-min gradient was used at 1 L/min and a 25-min gradient at 3 L/min. To compare the results, peak volumes, calculated by multiplying the flow rate by peak width at the base, are shown in TIC the inset. Running at 1 L/min. provides, on average, about a 1/3 reduction in peak volume. 1 As expected from the chromatographic principles described above, peptide peak volumes, and, therefore, resolution and sensitivity, are optimum at 1 L/min. The flow rate is low compared to what is traditionally used for TIC peptide mapping with 2.1 mm i.d. HPLC columns. The commonly accepted flow rate represents a compromise between resolution and run time 1 but it also reflects instrumental limits in reproducibly pumping liquid at flow rates less than 15 L/min. with accurate and precise gradients. The ACQUITY UPLC instrument performs extremely well at a flow rate of 1 L/min. in gradient mode. This performance is demonstrated by the overlay of 6 gradient runs of a peptide standard, shown in Figure 3. The average and standard deviations of retention times are listed for each peak Deamidated T Time (min) Figure 6: Separation of a deamidated peptide from its native form. T-16 T-16 The chromatographic principles also suggest ways to reduce the run time of a peptide map. Peptide maps run by HPLC often require cycle times as long as 3 5 h to separate all the peptides within the digest, especially for large proteins like antibodies. While faster peptide maps would be desirable, it is critically important that resolution not

31 DECEMBER 25 ACQUITY UPLC SYSTEM FA Time TFA Figure 7: Effect of mobile phase modifier in UPLC BEH material. be compromised. The test results must provide the same level of information. The van Deemter equation predicts that plate height will be 2 4 fold less with 1.7 m particles than with 3.5 m particles. The same resolving power can, therefore, be obtained with a shorter column. These experiments focus on the physical aspects of the separation as they relate to band broadening. Successful peptide mapping depends, of course, on the interaction among the peptides, the mobile phase, and the column surface chemistry. In Figure 4, the separation of a tryptic digest of enolase is shown with a 3.5 m C 18 HPLC column with 3Å pores, typical of the most common peptide separation columns, and a 1.7 m UPLC column. Conditions are the same for both columns. In the UPLC separation, more peaks are observed. The overall resolution and sensitivity are higher. In the UPLC map, there are several small peaks that are difficult to discern with the HPLC run. This result demonstrates that UPLC offers higher resolution and sensitivity when compared to HPLC under the same gradient conditions. As is always observed when comparing two different column chemistries, the separations are not identical in every detail. The overall appearance of the chromatograms is, however, similar. This suggests that the selectivity of the UPLC column is suitable for peptide mapping. To show how UPLC can resolve the same number of peaks in a peptide map as HPLC but in less time, the separation of an enolase digest was done on a 5 mm length UPLC column with a 2-min gradient and on a 1 mm length HPLC column with a 4-min gradient, both with flow rates of 1 L/min. These chromatograms are shown in Figure 5. The UPLC separation shows the same number of peaks and a similar overall elution pattern as the HPLC separation, but in half the time. UPLC offers the potential to reduce cycle times for peptide maps. The higher resolution and sensitivity with UPLC is particularly important when the peptide map is used to detect modified peptides. Higher resolution ensures that modified peptides are resolved from the unmodified form, as well as from other peptides in the digest. Higher sensitivity means that modified peptides can be detected at lower levels. For example, in Figure 6, UPLC is used to separate a deamidated peptide from its unmodified form. UPLC should be the technique of choice for detecting all the peptides in a sample. Peptide mapping is frequently interfaced with electrospray ionization mass spectrometry (ESI MS) to provide additional information about the eluting peptides, including molecular weight and sequence. MS can also identify modified peptides and glycosylation sites. Therefore, it is important that a peptide mapping technique work well under conditions that are favorable for ESI MS. TFA is commonly used as an acidic additive for peptide maps with UV detection, but it can lead to suppression of ionization and reduced sensitivity in ESI MS. Formic acid is more desirable for LC MS, as it causes less ion suppression than TFA. However, many reversed phase columns used for peptide mapping show lower retention and broader peaks with formic acid than with TFA. In Figure 7, the separation of several peptides with formic acid is compared with TFA on a UPLC column with MS detection. With formic acid, the peak heights are

32 32 ACQUITY UPLC SYSTEM DECEMBER 25 1 SIC Time Figure 8: A glycopeptide separation. about 3-fold higher. There is only a slight reduction in retention, corresponding to a difference of a few percent organic at the point of elution, and a slight increase in peak width. This result indicates that the UPLC columns perform extremely well under conditions that are best for ESI MS. Glycosylation is an important post-translational modification that plays a critical role in determining the efficacy and safety of a therapeutic protein. Glycosylation can be analyzed on the intact protein by mass spectrometry, as released glycans or as glycopeptides in LC MS peptide maps. When glycosylation is characterized with LC MS of the glycopeptides, the site of attachment can be directly determined and structural information can be obtained through MS MS experiments. This approach is limited, however, by the poor chromatographic peak shape of glycopeptides and incomplete resolution of glycoforms with HPLC peptide mapping. The poor peak shape has been attributed to the large size of the glycopeptides and their heterogeneous structure. Figure 8 shows the UPLC MS separation of a tryptic digest of -1 acid glycoprotein. The MS detection was performed with a Q-Tof mass spectrometer, which is well suited for glycopeptides due to its extended mass range. Data is plotted as a selected ion chromatogram for m/z 657, a signature ion for glycopeptides resulting from carbohydrate fragments. The glycopeptides are detected as sharp, symmetrical peaks with UPLC. These characteristics are important for minimizing spectral overlap of different glycoforms of the same peptide. UPLC with ESI TOF mass spectrometry will be a powerful tool for studying the glycosylation state of proteins. that of common reversed phase HPLC peptide mapping columns and can be easily transferred to alternative modifiers that give better sensitivity in ESI MS. UPLC with ESI TOF is especially suitable for the separation of glycopeptides. The ACQUITY UPLC System is clearly proving to be the next-generation tool for peptide mapping. Conclusions UPLC facilitates notable improvements in peptide mapping when compared to HPLC. Better resolution is obtained in combination with generally increased sensitivity. Run time can be reduced without compromising resolution by reducing column length and by increasing flow rate. Selectivity is comparable to

33 DECEMBER 25 ACQUITY UPLC SYSTEM 33 Analysis of Prostaglandins with the Quattro Premier XE Kate Yu and Donald Kwet Waters Corporation, Milford, Massachusetts The accurate and precise measurement of candidate pharmaceuticals in biological fluids, environmental pollutants in soils and water or the measurement of toxins in various food matrices requires highly sensitive and specific assays. Over the past 1 years, LC MS and LC MS MS using multiple reaction monitoring (MRM) has become the technique of choice for the trace analysis of analytes in complex mixtures. This has been made possible by the advent of atmospheric pressure ionization sources, such as ESI and APCI. The highly sensitive and specific nature of the tandem quadrupole mass spectrometer facilitates simple method development and highly selective assays. The Waters Micromass Quattro Premier XE mass spectrometer with its T-Wave (1) technology has been specifically designed with high sensitivity analysis in mind. This new mass spectrometer incorporates enhanced electronics and a new high sensitivity detector for improved performance. The superior negative ion sensitivity of the Quattro Premier XE is illustrated using the analysis of three prostaglandins (D 2, E 2, and F 2a ), which are acidic steroidal compounds. Increasing Negative Ion Sensitivity The negative ion sensitivity of the Quattro Premier XE was compared to that of the Quattro Premier. The prostaglandin compounds were analyzed by LC MS MS using the Waters ACQUITY UPLC System. A 5 L aliquot of the sample was loaded onto an ACQUITY UPLC BEH mm C 18, 1.7 m column. The column was eluted with a gradient of 1 95 B over 2 min at.8 ml/min, where A 1 acetonitrile in.1 formic acid and B 8/2 acetonitrile/methanol. The column eluent was monitored by negative ion ESI, using the MRM transitions for prostaglandins D 2 and E 2, and the MRM transition for prostaglandin F 2a. The resulting XIC chromatograms for prostaglandin F 2a using the Quattro Premier XE and Quattro Premier are displayed in Figure Here we can see that with the Quattro Premier XE peak area of the analyte is increased by a factor of approximately 1 compared to the Quattro Premier. A 9-fold increase in peak area was obtained for prostaglandins E 2 and 1-fold for D 2 (data not shown). Improving the Limit of Detection To further illustrate the superior performance of the Quattro Premier XE, the Limit of Detection (LOD) was compared to that of the Quattro Premier for both Prostaglandin D 2 and Prostaglandin E 2. The resulting data is displayed in Figure 2. It is shown that with the same on-column loading, the peak-topeak signal-to-noise value obtained from the Quattro Premier increased from 9.4 to 47.1 for Prostaglandin D 2 (left) and from PGF2a, 353.5>193.1 Peak area = 6 PtP S/N = MRM of 1 Channel ES e Peak area = 59 PtP S/N = MRM of 1 Channel ES e Time Peak area increased 9.8 times Peak-to-Peak Signal-to-Noise (PtP S/N) increased 5.4 times Figure 1: Comparative response for prostaglandin F 2a on the Quattro Premier (top) and Quattro Premier XE (bottom).

34 34 ACQUITY UPLC SYSTEM DECEMBER 25 PGD2, > MRM of 1 Channel ES- PGE2, > MRM of 1 Channel ES- 2.63e3 3.2e3 PtP S/N = 9.36 QP PtP S/N = MRM of 1 Channel ES MRM of 1 Channel ES- 2.63e3 3.2e3 PtP S/N = PtP S/N = QP-XE Time Peak area increased 9.7 times Time Peak area increased 8.8 times Figure 2: Comparing the LOD of the Quattro Premier (top) and Quattro Premier XE (bottom) to 35.7 for Prostaglandin E 2 (right). Therefore, from this data we can ascertain that the new Quattro Premier XE produces a 3- to 5-fold increase in signal-to-noise and a 1-fold increase in peak area sensitivity. Conclusion The new Quattro Premier XE mass spectrometer provides the ultimate in sensitivity in both positive and negative ion modes. This performance is further enhanced by the T-Wave-enabled fast scanning and fast MRM switching capabilities. The Quattro Premier XE, when combined with the superior chromatographic performance of the ACQUITY UPLC makes this the ideal LC MS MS platform for high sensitivity, high throughput analysis. Notes (1) The traveling wave device described here is similar to that described by Kirchner in US Patent 5,26,56 (1993).

35

36 721417EN

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION Andrew J. Aubin and Tanya L. Jenkins Waters Corporation, Milford, MA, USA INTRODUCTION Benzocaine (4-Aminobenzoic

More information

DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR

DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR Tanya Jenkins Waters Corporation, Milford, MA, USA INTRODUCTION ACQUITY UPLC PDA LINEARITY There are two characteristics that

More information

Validation of ACQUITY UPLC Methods

Validation of ACQUITY UPLC Methods Tanya Jenkins Andrew Aubin Dr. Michael Swartz Waters Corporation 34 Maple Street Milford, MA, 01757,USA Validation of ACQUITY UPLC Methods Agenda Test Method Validation Why ACQUITY UPLC? Redevelop or Convert

More information

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC USP Method Transfer and Routine Use Analysis of Tablets from HPLC to UPLC Aparna Chavali, Tanya Jenkins and Patricia McConville Waters Corporation, 34 Maple Street, Milford, MA USA APPLICATION BENEFITS

More information

Routine MS Detection for USP Chromatographic Methods

Routine MS Detection for USP Chromatographic Methods Daniel S. Root, Thomas E. Wheat, and Patricia McConville Waters Corporation, Milford, MA, USA APPLICATION BENEFITS This approach enables mass spectral analysis of peaks directly from unmodified USP HPLC

More information

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput MAY 2005 SEPARATION SCIENCE REDEFINED 31 Assay Transfer from HPLC to UPLC for Higher Analysis Throughput A typical HPLC assay was transferred and optimized for a Waters ACQUITY UPLC system to achieve both

More information

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 9 DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 9 Determination of drug release during

More information

UPLC Method Development and Validation

UPLC Method Development and Validation UPLC Method Development and Validation Rev. 2 2008 Waters Corporation Challenges of Method Development Methods are developed throughout the drug development process Samples vary in complexity Redundancy

More information

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System Margaret Maziarz, Sean M. McCarthy, Michael D. Jones, and Warren B. Potts Waters Corporation, Milford, MA, USA A P

More information

Transferring Yesterdays Methods to Tomorrows Technology: The Evolution from HPLC to UPLC TM

Transferring Yesterdays Methods to Tomorrows Technology: The Evolution from HPLC to UPLC TM Transferring Yesterdays Methods to Tomorrows Technology: The Evolution from HPLC to UPLC TM Eric S. Grumbach Thomas E. Wheat Chuck Phoebe Joe Arsenault Jeffrey R. Mazzeo Diane M. Diehl UPLC TECHNOLOGY

More information

Successfully Scaling and Transferring HPLC and UPLC Methods

Successfully Scaling and Transferring HPLC and UPLC Methods Successfully Scaling and Transferring HPLC and UPLC Methods Esa Lehtorinne Esa_Lehtorinne@waters.com Tel: +358-9-5659 6288 Fax: +358-9-5659 6282 Waters Finland Kutomotie 16 00380 Helsinki 2013 Waters Corporation

More information

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC.

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC. 7. Stability indicating analytical method development and validation of and in capsule dosage form by HPLC. 7.1 INSTRUMENTS AND MATERIALS USED 7.1.1 INSTRUMENTS 1. Shimadzu LC-2010 CHT with liquid chromatograph

More information

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity Janet Hammond Waters Corporation, Wilmslow, UK APPLICATION BENEFITS High

More information

Peptide Isolation Using the Prep 150 LC System

Peptide Isolation Using the Prep 150 LC System Jo-Ann M. Jablonski and Andrew J. Aubin Waters Corporation, Milford, MA, USA APPLICATION BENEFITS The Prep 150 LC System, an affordable, highly reliable system for preparative chromatography, is suitable

More information

Volume 6, Issue 2, January February 2011; Article-015

Volume 6, Issue 2, January February 2011; Article-015 Research Article DEVELOPMENT AND VALIDATION OF A RP-HPLC METHOD FOR THE DETERMINATION OF DAPOXETINE HYDROCHLORIDE IN PHARMACEUTICAL FORMULATION USING AN EXPERIMENTAL DESIGN Pratik Mehta*, Ujjwal Sahoo,

More information

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

Prep 150 LC System: Considerations for Analytical to Preparative Scaling Andrew Aubin and Jo-Ann Jablonski Waters Corporation, Milford, MA, USA APPLICATION BENEFITS The Prep 150 LC System is an affordable, highly reliable system for preparative chromatography and is suitable

More information

LC Technical Information

LC Technical Information LC Technical Information Method Transfer to Accucore.6 μm Columns Containing solid core particles, which are engineered to a diameter of.6μm and a very narrow particle size distribution; Accucore HPLC

More information

Dilution(*) Chromatography

Dilution(*) Chromatography WA20264 Poster # 184, HPLC 2002, Montreal, 4-5 June 2002 At-Column Column-Dilution for Preparative Dilution(*) Chromatography Cecilia Mazza, Jie Cavanaugh, Ziling Lu,Tom Sirard,Tom Wheat and Uwe Neue Waters

More information

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone Mia Summers and Kenneth J. Fountain Waters Corporation, 34 Maple St.,

More information

Fast methods for the determination of ibuprofen in drug products

Fast methods for the determination of ibuprofen in drug products APPLICATION NOTE 779 Fast s for the determination of ibuprofen in drug products Authors Sylvia Grosse, Mauro De Pra, Frank Steiner, Thermo Fisher Scientific, Germering, Germany Keywords Pharmaceutical,

More information

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION Quantitation by High Resolution Mass Spectrometry: Using Target Enhancement and Tof-MRM to Achieve Femtogram-level On-column Sensitivity for Quantitation of Drugs in Human Plasma Yun W. Alelyunas, Mark

More information

Implementation of Methods Translation between Liquid Chromatography Instrumentation

Implementation of Methods Translation between Liquid Chromatography Instrumentation Implementation of Methods Translation between Liquid Chromatography Instrumentation Michael D. Jones, Peter Alden, Kenneth J. Fountain, and Andrew Aubin Waters Corporation, Milford, MA, U.S. APPLICATION

More information

Chapter 4: Verification of compendial methods

Chapter 4: Verification of compendial methods Chapter 4: Verification of compendial methods Introduction In order to ensure accurate and reliable test results, the quality control laboratory (QCL) needs to use analytical methods (and accompanying

More information

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW 75 CHAPTER 3 DEVELOPMENT AND APPLICATION OF STABILITY-INDICATING HPLC METHOD FOR THE DETERMINATION OF NEVIRAPINE AND ITS IMPURITIES IN COMBINATION DRUG PRODUCT 3.1 INTRODUCTION OF DOSAGE FORM AND LITERATURE

More information

The Theory of HPLC. Quantitative and Qualitative HPLC

The Theory of HPLC. Quantitative and Qualitative HPLC The Theory of HPLC Quantitative and Qualitative HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

Int. J. Pharm. Sci. Rev. Res., 30(2), January February 2015; Article No. 09, Pages: 63-68

Int. J. Pharm. Sci. Rev. Res., 30(2), January February 2015; Article No. 09, Pages: 63-68 Research Article Stability indicating RP-HPLC Method for Determination of FexoFenadine Hydrochloride and its Related Substances in Active Pharmaceutical Substance Abhay Gupta* 1, Dr. Birendra Srivastava,

More information

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID Asian Journal of Chemistry Vol. 21, No. 3 (2009), 1739-1746 Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID PRAVEEN KUMAR BALIYAN*, R.P.

More information

KEYWORDS: Acetaminophen, Doxylamine succinate, Dextromethorphan hydrobromide.

KEYWORDS: Acetaminophen, Doxylamine succinate, Dextromethorphan hydrobromide. International Journal of Pharmaceutical Science Invention ISSN (Online): 2319 6718, ISSN (Print): 2319 670X Volume 3 Issue 7 July 2014 PP.08-12 Analytical method development and validation of acetaminophen,

More information

TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD

TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD APPENDIX V TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD Validating Chromatographic Methods. By David M. Bliesner Copyright 2006 John Wiley & Sons, Inc. 159 160 APPENDIX V Title: Effective: Document No:

More information

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India Reducing analysis time and solvent consumption for isocratic USP assay methods with current and proposed USP guidelines using the Agilent 129 Infinity LC System An efficient way to reduce cost of analysis

More information

LC and LC/MS Column Selection Flow Chart

LC and LC/MS Column Selection Flow Chart LC and LC/MS Column Selection Flow Chart To use the column selection diagram below, simply follow the path for your analyte and mobile phase. At the far right, follow your final column selection to the

More information

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW Paula Hong, John Van Antwerp, and Patricia McConville Waters Corporation, Milford, MA, USA Historically UV detection has been

More information

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India Effective use of pharmacopeia guidelines to reduce cost of chromatographic analysis Optimized, cost-effective HPLC analysis of atorvastatin by varying column dimensions within the USP allowed limts

More information

Development And Validation Of Rp-Hplc Method For Determination Of Velpatasvir In Bulk

Development And Validation Of Rp-Hplc Method For Determination Of Velpatasvir In Bulk International Journal of Engineering Science Invention (IJESI) ISSN (Online): 2319 6734, ISSN (Print): 2319 6726 www.ijesi.org PP. 36-41 Development And Validation Of Rp-Hplc Method For Determination Of

More information

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies Jonathan Danaceau, Erin Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA, USA

More information

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance Michael D. Jones, Marian Twohig, Karen Haas, and Robert S. Plumb Waters Corporation,

More information

HPLC Praktikum Skript

HPLC Praktikum Skript HPLC Praktikum Skript Assistants: Gianluca Bartolomeo HCI D330, 3 46 68, bartolomeo@org.chem.ethz.ch Sahar Ghiasikhou HCI E330, 2 29 29, ghiasikhou@org.chem.ethz.ch 1. Introduction In chromatographic techniques,

More information

penta-hilic UHPLC COLUMNS

penta-hilic UHPLC COLUMNS penta-hilic UHPLC COLUMNS penta-hilic Highly retentive, proprietary penta-hydroxy-ligand Excellent peak shape for polar compounds with a variety of functional groups: acids, bases, zwitterions strong and

More information

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract. Separation of Explosives in EPA 833: Column Choices Optimize Speed, Resolution, and Solvent Use Application Environmental Authors John W. Henderson Jr. and William J. Long Agilent Technologies, Inc. 28

More information

Unexpected Peaks in Chromatograms - Are They Related Compounds, System Peaks or Contaminations? From the Diary of an HPLC Detective

Unexpected Peaks in Chromatograms - Are They Related Compounds, System Peaks or Contaminations? From the Diary of an HPLC Detective Unexpected Peaks in Chromatograms - Are They Related Compounds, System Peaks or Contaminations? From the Diary of an HPLC Detective SHULAMIT LEVIN HPLC in Pharmaceutics Σ Stability Indicating Methods Extra

More information

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency Luna 2.5 µm C18(2)-HST Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency Table of Contents Part 1 Theory 1.1 Abstract...3 1.2 Introduction...3 Part 2 Set Up 2.1

More information

Research Article METHOD DEVELOPMENT AND VALIDATION FOR SIMULTANEOUS ESTIMATION OF ELBASVIR AND GRAZOPREVIR BY RP-HPLC

Research Article METHOD DEVELOPMENT AND VALIDATION FOR SIMULTANEOUS ESTIMATION OF ELBASVIR AND GRAZOPREVIR BY RP-HPLC ISSN 2395-3411 Available online at www.ijpacr.com 248 Research Article METHOD DEVELOPMENT AND VALIDATION FOR SIMULTANEOUS ESTIMATION OF ELBASVIR AND GRAZOPREVIR BY RP-HPLC Mantha Vebkatesh* and A. Yasodha

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

Appendix II- Bioanalytical Method Development and Validation

Appendix II- Bioanalytical Method Development and Validation A2. Bioanalytical method development 1. Optimization of chromatographic conditions Method development and optimization of chromatographic parameters is of utmost important for validating a method in biological

More information

A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS

A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS CH.MOUNIKA*, M.KINNERA Research Article SIR.C.R.REDDY COLLEGE OF PHARMACEUTICAL SCIENCES, ELURU.

More information

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46>

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46> High Performance Liquid Chromatography HPLC-9 Ultra-high Speed Analysis of Ibuprofen within USP Allowed Limits by Nexera Method Scouting In recent years, high-throughput analytical techniques have

More information

Mashhour Ghanem 1 and Saleh Abu-Lafi 2 * ABSTRACT ARTICLE INFO

Mashhour Ghanem 1 and Saleh Abu-Lafi 2 * ABSTRACT ARTICLE INFO . Journal of Applied Pharmaceutical Science Vol. 3 (10), pp. 051-058, October, 2013 Available online at http://www.japsonline.com DOI: 10.7324/JAPS.2013.31009 ISSN 2231-3354 Validation of a Stability-Indicating

More information

Validated HPLC Methods

Validated HPLC Methods Validated HPLC Methods To Tweak or Not to Tweak Slide 4 Adjustments of operating conditions to meet system suitability requirements may be necessary. USP 23 p1776 Slide 5 001816S1.PPT What is the line

More information

RP-HPLC Method Development and Validation of Dapagliflozin in Bulk and Tablet formulation

RP-HPLC Method Development and Validation of Dapagliflozin in Bulk and Tablet formulation 221 IJPAR Volume 2 Issue 4 Oct - Dec -2013 ISSN: 2320-2831 Available Online at: [Research article] RP-HPLC Method Development and Validation of Dapagliflozin in Bulk and Tablet formulation Jeyabaskaran.M

More information

Method Development and Validation Of Prasugrel Tablets By RP- HPLC

Method Development and Validation Of Prasugrel Tablets By RP- HPLC Method Development and Validation Of Prasugrel Tablets By RP- HPLC K.Sonia*, Ndwabe Hamunyare, K.Manikandan Department of Pharmaceutical Analysis, SRM College of Pharmacy, SRM University, Kattankulathur,

More information

Reverse Phase High Performance Liquid Chromatography method for determination of Lercanidipine hydrochloride in bulk and tablet dosage form

Reverse Phase High Performance Liquid Chromatography method for determination of Lercanidipine hydrochloride in bulk and tablet dosage form Research Article ISSN: 0974-6943 M.V.Kumudhavalli et al. / Journal of Pharmacy Research 2014,8(11), Available online through http://jprsolutions.info Reverse Phase High Performance Liquid Chromatography

More information

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research Heather A Brown, 1 Claudia Rossi, 2 and Lisa J Calton 1 1 Waters Corporation, Wilmslow, UK 2

More information

Preparative Chromatography with Improved Loadability

Preparative Chromatography with Improved Loadability 05 LCGC RTH AMERICA VLUME 0 UMBER VEMBER 00 www.chromatographyonline.com Preparative Chromatography with Improved Loadability In combinatorial chemistry and drug development, as well as in the full-scale

More information

A Protocol for High-Throughput Drug Mixture Quantitation: Fast LC MS or Flow Injection Analysis MS?

A Protocol for High-Throughput Drug Mixture Quantitation: Fast LC MS or Flow Injection Analysis MS? 60 LCGC VOLUME 19 NUMBER 1 JANUARY 2001 www.chromatographyonline.com A Protocol for High-Throughput Drug Mixture Quantitation: Fast LC MS or Flow Injection Analysis MS? The authors demonstrate the pros

More information

Analysis of Organic Light Emitting Diode Materials by UltraPerformance Convergence Chromatography Coupled with Mass Spectrometry (UPC 2 /MS)

Analysis of Organic Light Emitting Diode Materials by UltraPerformance Convergence Chromatography Coupled with Mass Spectrometry (UPC 2 /MS) Analysis of Organic Light Emitting Diode Materials by UltraPerformance Convergence Chromatography Coupled with Mass Spectrometry (UPC 2 /MS) Michael D. Jones, Andrew Aubin, Peter J. Lee, and Tim Jenkins

More information

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate Michael D. Jones, Xiang Jin Song, Robert S. Plumb, Peter J. Lee, and

More information

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies APPLICATION BENEFITS Regulatory guidelines and recommendations place a greater emphasis on the detection

More information

Validation of Stability-Indicating RP-HPLC Method for the Assay of Ibrutinib in Pharmaceutical Dosage form

Validation of Stability-Indicating RP-HPLC Method for the Assay of Ibrutinib in Pharmaceutical Dosage form Validation of Stability-Indicating RP-HPLC Method for the Assay of Ibrutinib in Pharmaceutical Dosage form 8.1 Introduction Ibrutinib (IBR) is an anticancer drug targeting B-cell malignancies (blood cancer

More information

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS UCT Part Numbers: CSFASETG203 - CLEAN SCREEN FASt ETG, 200mg / 3mL tube SLETG100ID21-3UM - Selectra ETG HPLC column, 100 x 2.1

More information

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF Application Note Donghui Yi and Christine Miller Jon D. Williams, GlaxoSmithKline Introduction Mass spectrometry (MS) has become a core technology

More information

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD TO DETERMINE CINITAPRIDE HYDROGEN TARTARATE IN BULK AND PHARMACEUTICAL FORMULATION

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD TO DETERMINE CINITAPRIDE HYDROGEN TARTARATE IN BULK AND PHARMACEUTICAL FORMULATION Research Article ISSN:2230-7346 Journal of Global Trends in Pharmaceutical Sciences Vol.3, Issue 2, pp -619-627, April June 2012 DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD TO DETERMINE CINITAPRIDE HYDROGEN

More information

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS Application Note: 56 Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS Yang Shi, Catherine Lafontaine, Matthew Berube, John Fink, François Espourteille Thermo

More information

Biochemistry. Biochemical Techniques HPLC

Biochemistry. Biochemical Techniques HPLC Description of Module Subject Name Paper Name 12 Module Name/Title 13 1. Objectives 1.1. To understand the basic concept and principle of 1.2. To understand the components and techniques of 1.3. To know

More information

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA Use of High Speed/High Resolution Size-Based Chromatographic Separation of Surfactants and Oligomeric Materials with Single Quadrupole Mass Spectrometry Michael O Leary, Jennifer Gough, Tanya Tollifson

More information

Basic Principles for Purification Using Supercritical Fluid Chromatography

Basic Principles for Purification Using Supercritical Fluid Chromatography Basic Principles for Purification Using Supercritical Fluid Chromatography Jo-Ann M. Jablonski, Christopher J. Hudalla, Kenneth J. Fountain, Steven M. Collier, and Damian Morrison Waters Corporation, Milford,

More information

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring High-Throughput Protein Quantitation Using Multiple Reaction Monitoring Application Note Authors Ning Tang, Christine Miller, Joe Roark, Norton Kitagawa and Keith Waddell Agilent Technologies, Inc. Santa

More information

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc. Agilent Application Solution Transfer of a USP method for tolazamide from normal phase HPLC to SFC using the Agilent 126 Infinity Hybrid SFC/UHPLC System Improving peak shape and providing wider UV selectivity

More information

4.1 MATERIALS 4.2 EQUIPMENTS

4.1 MATERIALS 4.2 EQUIPMENTS 4.1 MATERIALS 1. Aceclofenac Moraceae Pharma Ltd., Uttarakhand, India 2. Diclofenac Arti Drugs Ltd., Mumbai, India 3. Paracetamol Moraceae Pharma Ltd., Uttarakhand, India 4. Para-aminophenol Hema Pharmaceuticals

More information

Impact factor: 3.958/ICV: 4.10 ISSN:

Impact factor: 3.958/ICV: 4.10 ISSN: Impact factor: 3.958/ICV: 4.10 ISSN: 0976-7908 99 Pharma Science Monitor 9(4), Oct-Dec 2018 PHARMA SCIENCE MONITOR AN INTERNATIONAL JOURNAL OF PHARMACEUTICAL SCIENCES Journal home page: http://www.pharmasm.com

More information

Product Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns

Product Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns Product Bulletin HPLC Columns ACE LC/MS and Rapid Analysis HPLC Columns 0 mm, 30 mm, 35 mm and 50 mm column lengths.0,., 3.0, 4.0 and 4.6 mm column diameters Configured for High Sample Throughput Specially

More information

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note Authors Siji Joseph Agilent Technologies India Pvt. Ltd.

More information

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Page 1 of 13 CIPAC/4105/R GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Prepared for CIPAC by Dr M J Tandy*, P M Clarke and B White (UK) The rapid

More information

2.1 2,3 Dichloro Benzoyl Cyanide (2,3 DCBC) and survey of. manufactured commonly for the bulk drug industry, few references were

2.1 2,3 Dichloro Benzoyl Cyanide (2,3 DCBC) and survey of. manufactured commonly for the bulk drug industry, few references were . Introduction.,3 Dichloro Benzoyl Cyanide (,3 DCBC) and survey of analytical methods,3-dcbc substance, is the although advanced intermediate,3-dcbc is a of lamotrigine well-known bulk drug chemical manufactured

More information

IN QUALITATIVE ANALYSIS,

IN QUALITATIVE ANALYSIS, IN QUALITATIVE ANALYSIS, YOU VE ALWAYS HAD TO CHOOSE BETWEEN ACCURACY AND THROUGHPUT. NOW THERE S NO REASON TO ever have to. WATERS QUALITATIVE ANALYSIS solutions Waters ACQUITY UPLC System with the Waters

More information

Development and Validation of a HPLC Method for Chlorphenamine Maleate Related Substances in Multicomponents Syrups and Tablets

Development and Validation of a HPLC Method for Chlorphenamine Maleate Related Substances in Multicomponents Syrups and Tablets Development and Validation of a HPLC Method for Chlorphenamine Maleate Related Substances in s Syrups and Tablets Larisa Alagić-Džambić*, Midhat Vehabović, Edina Čekić, Mirsad Džambić Development Department,

More information

Q-TOF PREMIER DYNAMIC RANGE ENHANCEMENT AND ACCURATE MASS MEASUREMENT PERFORMANCE

Q-TOF PREMIER DYNAMIC RANGE ENHANCEMENT AND ACCURATE MASS MEASUREMENT PERFORMANCE Q-TOF PREMIER DYAMIC RAGE EHACEMET AD ACCURATE MASS MEASUREMET PERFORMACE Michael McCullagh, Jose Castro-Perez, and Lisa Calton Waters Corporation, Manchester, UK ITRODUCTIO The use of time-of-flight (TOF)

More information

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer APPLICATION NOTE 72385 Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer Authors Stephan Meding, Katherine Lovejoy, Martin Ruehl Thermo Fisher Scientific,

More information

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis High-Resolution Sampling D-LC for Pharmaceutical Impurity Analysis Detection of Impurities Hidden Under the API Peak at Relevant Levels Application Note Small Molecule Pharmaceuticals and Generics Authors

More information

STABILITY INDICATING METHOD OF RELATED IMPURITIES IN VENLAFAXINE HYDROCHLORIDE SUSTAINED RELEASE TABLETS

STABILITY INDICATING METHOD OF RELATED IMPURITIES IN VENLAFAXINE HYDROCHLORIDE SUSTAINED RELEASE TABLETS Issn No: 976-39 RESEARCH ARTICLE STABILITY INDICATING METHOD OF RELATED IMPURITIES IN VENLAFAXINE HYDROCHLORIDE SUSTAINED RELEASE TABLETS CHETLAPALLI SATYA SRINIVAS 1, P.RENUKA DEVI 2 and GAMPA VIJAYAKUMAR*

More information

A Simple, Novel Validated Stability Indicating RP-HPLC method for estimation of Duloxetine HCl in Capsule Pharmaceutical Formulation

A Simple, Novel Validated Stability Indicating RP-HPLC method for estimation of Duloxetine HCl in Capsule Pharmaceutical Formulation Pharmaceutical Research A Simple, Novel Validated Stability Indicating RP-HPLC method for estimation of Duloxetine HCl in Capsule Pharmaceutical Formulation Manisha Puranik* a, Sailesh Wadher b and Kritika

More information

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Application Note Food Authors Jianqiu Mi, Zhengxiang Zhang, Zhixu Zhang,

More information

SIMULTANEOUS RP HPLC DETERMINATION OF CAMYLOFIN DIHYDROCHLORIDE AND PARACETAMOL IN PHARMACEUTICAL PREPARATIONS.

SIMULTANEOUS RP HPLC DETERMINATION OF CAMYLOFIN DIHYDROCHLORIDE AND PARACETAMOL IN PHARMACEUTICAL PREPARATIONS. Ind. J. Anal. Chem Vol. 7 11. 2008 SIMULTANEOUS RP HPLC DETERMINATION OF CAMYLOFIN DIHYDROCHLORIDE AND PARACETAMOL IN PHARMACEUTICAL PREPARATIONS. Authors for correspondence : R. R. Singh1*, M. V. Rathnam,

More information

Revision Bulletin 27 Jan Feb 2017 Non-Botanical Dietary Supplements Compliance

Revision Bulletin 27 Jan Feb 2017 Non-Botanical Dietary Supplements Compliance Niacin Extended-Release Tablets Type of Posting Posting Date Official Date Expert Committee Reason for Revision Revision Bulletin 27 Jan 2017 01 Feb 2017 Non-Botanical Dietary Supplements Compliance In

More information

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories APPLICATION NOTE 65200 Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories Authors Neloni Wijeratne, Claudia Martins, Mary

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (11) 04 QUALIFICATION OF EQUIPMENT ANNEX 1: QUALIFICATION OF HPLC EQUIPMENT Full document title and reference Document type

More information

Shodex TM ODP2 HP series columns

Shodex TM ODP2 HP series columns HPLC Columns Shodex TM ODP2 HP series columns Better retention of highly polar substances Technical notebook No. 6 Contents 1. Introduction 1-1. Specifications 1-2. Eluent Compatibility of ODP2 HP Series

More information

2007 Waters Corporation. Application of UPLC to Peptide Mapping

2007 Waters Corporation. Application of UPLC to Peptide Mapping 2007 Waters Corporation Application of UPLC to Peptide Mapping Comparison of HPLC and UPLC 7.5e-2 7.0e-2 6.5e-2 6.0e-2 5.5e-2 HPLC 5.0e-2 4.5e-2 4.0e-2 3.5e-2 3.0e-2 2.5e-2 2.0e-2 1.5e-2 1.0e-2 30.00 35.00

More information

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China Rapid Deteration of Eight Related Aromatic Acids in the p-phthalic Acid Mother Liquid Using an Agilent 126 Infinity LC System and an Agilent Poroshell 12 SB-C18 Column Application Note Author Xinlei Yang

More information

UPC 2 Strategy for Scaling from Analytical to Preparative SFC Separations

UPC 2 Strategy for Scaling from Analytical to Preparative SFC Separations UPC Strategy for Scaling from Analytical to Preparative SFC Separations Christopher J. Hudalla, Abhijit Tarafder, Jo-Ann Jablonski, and Kenneth J. Fountain Waters Corporation, Milford, MA, USA APPLICATION

More information

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Application Note Doping Control Authors Prof. Maria Kristina Parr Freie Universität Berlin Institute

More information

contents of the currently official monograph. Please refer to the current edition of the USP NF for official text.

contents of the currently official monograph. Please refer to the current edition of the USP NF for official text. Metformin Hydrochloride Extended-Release Tablets Type of Posting Posting Date Targeted Official Date Notice of Intent to Revise 28 Sept 2018 To Be Determined, Revision Bulletin Expert Committee Chemical

More information

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS UCT Part Numbers: CSFASETG203 - CLEAN SCREEN FASt ETG, 200mg / 3mL tube SLETG100ID21-3UM - Selectra ETG HPLC column, 100 x 2.1

More information

Revision Bulletin 29 Dec Jan 2018 Non-Botanical Dietary Supplements Compliance

Revision Bulletin 29 Dec Jan 2018 Non-Botanical Dietary Supplements Compliance Niacin Extended-Release Tablets Type of Posting Posting Date Official Date Expert Committee Reason for Revision Revision Bulletin 29 Dec 2017 01 Jan 2018 Non-Botanical Dietary Supplements Compliance In

More information

CHAPTER - 3 ANALYTICAL PROFILE. 3.1 Estimation of Drug in Pharmaceutical Formulation Estimation of Drugs

CHAPTER - 3 ANALYTICAL PROFILE. 3.1 Estimation of Drug in Pharmaceutical Formulation Estimation of Drugs CHAPTER - 3 ANALYTICAL PROFILE 3.1 Estimation of Drug in Pharmaceutical Formulation 3.1.1 Estimation of Drugs ANALYTICAL PROFILE 84 3.1 ESTIMATION OF DRUG IN PHARMACEUTICAL FORMULATION. Agrawal A et al

More information

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS Marian Twohig, Michael D. Jones, Dominic Moore, Peter Lee, and Robert Plumb Waters Corporation, Milford, MA, USA APPLICATION

More information

UPLC Intact MASS Analysis Application Kit

UPLC Intact MASS Analysis Application Kit (Part Number: 176001519) CONTENTS I. INTACT MASS ANALYSIS APPLICATION KIT COMPONENTS II. INSTALLING THE PRE-COLUMN TUBING INTO MASSPREP MICRO DESALTING COLUMN III. CONNECTING THE COLUMN TO LCT PREMIER

More information

ProPac WCX-10 Columns

ProPac WCX-10 Columns ProPac WCX-10 Columns Guidance for column use Tips to maximize column lifetime ProPac WCX-10 Column Tips and Tricks This guide provides essential information and invaluable guidelines for mobile phases,

More information

Journal of Drug Delivery and Therapeutics

Journal of Drug Delivery and Therapeutics Available online on 15.01.2018 at http://jddtonline.info Journal of Drug Delivery and Therapeutics Open Access to Pharmaceutical and Medical Research 2011-17, publisher and licensee JDDT, This is an Open

More information

Development of Validated Analytical Method of Mefenamic Acid in an Emulgel (Topical Formulation)

Development of Validated Analytical Method of Mefenamic Acid in an Emulgel (Topical Formulation) Development of Validated Analytical Method of Mefenamic Acid in an Emulgel (Topical Formulation) TEENA OSWAL*, DR.SURYAKANT BHOSALE, DR. SONALI NAIK MET Institute Of Pharmacy Met Complex, Bandra Reclamation,

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Amiodarone-D 4, Primary Standard A-83 Page 1 of 8 Catalog Number: A-83 Lot: Expiration: Description: Packaging: Storage: Shipping: Intended Use: Instructions

More information