Agilent s New 7000B Quadrupole GC/MS/MS:

Size: px
Start display at page:

Download "Agilent s New 7000B Quadrupole GC/MS/MS:"

Transcription

1 Agilent s New 7000B Quadrupole GC/MS/MS: Evolving technology for higher performance tools Terry L. Sheehan, Ph.D. GC/MS/MS Product Manager Agilent Technologies Santa Clara, CA Page 1

2 Triple Quadrupole MS/MS (also called tandem quadrupole) is changing the way scientists think about GC/MS Page 2

3 Transmission Quadrupole MS & MS/MS Sensitivity and Selectivity Scale Least Sensitive Most Sensitive Fast 12,500 u/s (5975C) Fast 6,250 u/s (7000) Targets & Non-targets < < Typical < 2,000 u/s Targets & Non-targets Dwell ms Dwell 1-50 ms Targets only *Multiple Reaction Monitoring Also called SRM for Selected Reaction Monitoring (similar to SIM Selected Ion Monitoring) Page 3

4 Transmission Quadrupole MS & MS/MS Sensitivity and Selectivity Scale Least Sensitive Fast 12,500 u/s (5975C) Fast 6,250 u/s (7000) Targets & Non-targets Least Selective < < Typical < 2,000 u/s Targets & Non-targets Most Sensitive Dwell ms Dwell 1-50 ms Targets only Most Selective Unit mass resolution < < Unit mass + AMDIS deconvolution Targets only per DBL Page 4

5 Summary of Relative Performance Factors GC/MSD Scan GC/MSD SIM GC/MSD Scan-DRS GC/TQ MS/MS S from analyte S from matrix N from neutrals = = = Ultra low N from matrix N from "bleed" Low Low Lower 0 MDL (clean) Very Low Lowest Lower Lowest MDL (very dirty) Low Low Lower Lowest Quant Error Low Low Lower Lowest MDL = S analyte N neutral + N matrix + N GC bleed Quant Error = S analyte S matrix + S matrix But scan is essential for non-targets (unknowns)! Page 5

6 Fundamental Benefits of MS/MS for GC Higher Selectivity to Identify Target Analytes in Complex Matrices Consistently High Sensitivity for Target Analytes in Complex Matrices Page 6

7 MS/MS Eliminates Scan and SIM Interferences Single Quad MS selectivity proportional to spectral resolution no selectivity against ions with same m/z interference analyte unit mass resolution Page 7

8 MS/MS Eliminates Scan and SIM Interferences Single Quad MS selectivity proportional to spectral resolution no selectivity against ions with same m/z interference analyte Product 2 unit mass resolution Product 3 Product 1 analyte Product Ion Spectrum Page 8

9 MS/MS Eliminates Scan and SIM Interferences Single Quad MS selectivity proportional to spectral resolution no selectivity against ions with same m/z interference analyte Triple Quad MS High probability that one or more of the product ions will be a unique dissociation product of the precursor only Product 2 unit mass resolution Product 3 Product 1 interference analyte Product Ion Spectrum Page 9

10 MS/MS Ensures Lowest Detection Limits EI: spectrum of analyte can also include ions from matrix, column bleed, gases, etc. Q1 SIM isolate precursor before CID chemical noise from these other ions is eliminated Page 10

11 MS/MS Ensures Lowest Detection Limits EI: spectrum of analyte can also include ions from matrix, column bleed, gases, etc. Product 2 Q1 SIM isolate precursor before CID Product 3 Product 1 chemical noise from these other ions is eliminated CID + Q2 SIM Lower m/z Product Ions measured against zero chemical noise for consistently lower LOQ and LODs Page 11

12 SIM Limited by Unit Mass Resolution in diesel fuel GC/MS Single Quad SIM Interfering matrix peaks = chemical noise C 20 Eicosane 100 Hexachlorobenzene Cl Cl Cl Cl Cl Cl (mainlib) Eicosane (mainlib) Benzene, hexachloro Page 12

13 SIM Limited by Unit Mass Resolution in diesel fuel GC/MS Single Quad SIM Interfering matrix peaks = chemical noise Cl Cl Cl Cl Cl Cl (mainlib) Eicosane (mainlib) Benzene, hexachloro Page 13

14 SIM Limited by Unit Mass Resolution in diesel fuel GC/MS Single Quad SIM Interfering matrix peaks = chemical noise At unit mass resolution SIM, HCB is buried by the more intense peaks from C 20 to C 22 hydrocarbons Page 14

15 MS/MS Succeeds Where MS Fails in diesel fuel GC/MS Single Quad SIM Interfering matrix peaks = chemical noise GC/MS Triple Quad SRM A chromatographer s dream: single peak on flat baseline Page 15

16 Pesticide Analysis: Comparing SIM to MS/MS SIM EIC (246) Apple Cabbage Ginseng X Orange X Spinach Page 16

17 Pesticide Analysis: Comparing SIM to MS/MS SIM EIC (246) MRM (2 transitions) Apple Cabbage Ginseng > 176.1, > x10 4 Ratio= A c q u isitio n T im e (m in ) > 176.1, > x10 3 Ratio= A c q u is itio n T im e (m in ) > 176.1, > x10 4 Ratio= Counts Counts Counts 0 S/N = 448 S/N = 241 S/N = 446 X Orange Acquisition Time (min) > 176.1, > x10 3 Ratio= Counts S/N = X Spinach A c q uisitio n T ime (min ) > 176.1, > x10 4 Ratio= Counts A c q u is itio n T im e (m in ) S/N = 260 Page 17

18 7000B MS/MS: High Performance Components The best technologies from the industry leading 5975C Series GC/MSD Heated monolithic gold plated quartz quadrupole Reliable, inert, high performance ion source High Signal/Low Noise Triple-Axis Detector PCI/NCI ion source and the 6400 Series LC/MS/MS Linear acceleration enhanced hexapole collision cell MassHunter software New inert, high temp, extraction EI source Split Flow Turbo Gold plated quartz, hyperbolic quadrupoles New Ramped Iris Triple-Axis Detector High pressure hexapole collision cell with Patent Pending Helium Quench Gas Hinged access to source and detector Page 18

19 7000B MS/MS: Hot (200ºC) Quartz Analyzers High temperature analyzers stay clean even with highly concentrated, complex matrices Gold plated quartz, hyperbolic quadrupoles Page 19

20 Proprietary Heated Quartz Quadrupole High temperature (200ºC) operation results in greater reliability with dirty samples MS/MS methods typically have more dirty samples and benefit even more from the higher temp quad analyzer Virtually eliminates the need to clean quads as a maintenance issue More stable tunes and methods over a longer period of time in real world sample environments Page 20

21 Hyperbolic, Gold-coated, Quartz - Proprietary! Why quartz? Low coefficient of linear expansion (expansivity; α) dl = α L 0 dt [t = ºC; dl = change in length (m); L 0 = length at 0 ºC] α varies greatly for different materials (m/m K x 10-6 ) Stainless steel 304 = 17.3; Stainless steel 316 = 16.0 Inconel (Ni alloy) = 12.6; Monel (Ni/Cu alloy) = 13.5 Glass pyrex = 4.0; Quartz, fused = 0.59 For change from 25 to 150 ºC, which material expands the LEAST? But what is the analytical value? Low thermal expansion means dimensions/alignments do not change After maintenance, the quartz quad can be baked at 200 ºC Under operation, the quartz quad can be run at 150, 180 or even 200 ºC Heat is the friend of GC/MS methods true for the source and analyzer! Page 21

22 More Information on the Agilent Website Detailed Video Page 22

23 7000B MS/MS: Collision Cell with Helium Quench Highly efficient dissociation of precursor ions Maximum transitions per second without product ion carryover (cross-talk) High pressure hexapole collision cell with Patent Pending Helium Quench Gas Reduction of metastable helium from EI source Page 23

24 Transmision Hexapole Cell: Better Transmission Characteristics Mass Range Transmission Quadrupole Hexapole Octapole m/z GC/MS m/z=1050 Quadrupoles are the best mass filters (analyzers) Hexapoles and octapole are the best transmission devices Page 24

25 500 Transitions/sec: Why Is This Important? Narrow GC peaks (width <1-2 s) Sufficient data points (15) to perform acceptable/accurate quantification Many compounds to monitor in single run (multiresidue) when coeluting peaks are quite frequent Regulated environment where QC checks are required to ensure data validity Confirmatory transitions according to the 96/23/CE directive Page 25

26 Measured Collision Cell Clearing Time 1 msec No product ion signal = no cross-talk at 500 transitions per second Page 26

27 Collision Cell Process: Typical Description Collision Cell 1 ml/min N 2 Collision Gas Quad Analyzer Quad Analyzer Source Precursor Ions In collision induced dissociation Product Ions Out Detector This is a good description for LC/MS/MS, but it is not complete for GC/MS/MS Page 27

28 Collision Cell Process: Full Description for GC/MS Source He* + Quad Analyzer Collision Cell Precursor Ions In 1 ml/min N 2 Collision Gas collision induced dissociation Quad Analyzer Product Ions Out At the detector, metastable helium generates neutral noise. + He* Detector A high population of highly energetic helium metastables are produced in an EI source; since metastable helium is not charged, it can pass through mass analyzer field into the collision cell and through to the HED-EM In GC/MS, neutral noise is buried in much higher chemical noise. In GC/MS/MS, chemical noise is greatly reduced so neutral noise is a critical source of noise. Page 28

29 Collision Cell Process with Helium Quench Gas Collision Cell 1 ml/min N 2 Collision Gas Quad Analyzer Quad Analyzer Source He* + Precursor Ions In collision induced dissociation Product Ions Out + He* Detector He Buffer Gas He* + He 2 He + heat Transmission of metastable helium to the detector is greatly reduced; the Triple-Axis detector further reduces neutral noise for ultra-low neutral noise. Page 29

30 Benefit from Helium Quench tested with hexachlorobenzene +MRM( > ) 1pgHCB_std_cond_UnkGain_2090EMV_He0pct_05.d x103 Noise(RMS) =20.08; SNR(6.470min) =341.0 Same Signal Noise Decreased Helium Quench Gas OFF M RM( > )1pgHCB_std_cond_UnkG ain_2090em V_He0pct_05.d S)=20.08;SNR(6.470m in)=341.0 x102 Noise(RM S/N 341: Countsvs. AcquisitionTime(min) MRM( > ) 1pgHCB_std_cond_UnkGain_2090EMV_He20pct_04.d x103 Noise(RMS) =6.03; SNR(6.465min) = Countsvs.AcquisitionTim e(m in) Same Signal Helium Quench Gas ON S/N 1294: Page Countsvs. AcquisitionTime(min) In MS apps, neutral noise is not visible it is buried by more intense chemical noise. Since MS/MS eliminates chemical noise, neutral noise is visible and must be minimized.

31 7000B MS/MS: New Higher Sensitivity Source Increase signal for precursor ions using an extractor source design New inert, high temp, extraction EI source Page 31

32 Lab-Proven Inert Source Solid, inert materials, not a coating that can be damaged by cleaning Stay-Clean (350ºC) design extends maintenance intervals with dirty samples Source tune parameters in the Autotune file Dual filaments reduces maintenance intervals 1000 s of satisfied MSD users Partition between source and analyzer compartments support dual stage pumping Page 32

33 7000B Extraction Source External View - Assembled Like the 7000A source, the 7000B extraction source is more open than the MSD source looks similar to the CI source At first look, it does not look any different than the 7000A source Page 33

34 7000B Extraction Source Expanded View of Components Component #7 is a new insulator And there is new connection to the lens #11 Design uses combination of push (repeller) and pull (extractor lens) to increase ion intensity Page 34

35 7000B Increased Signal Extractor source cutaway Extractor Source w/ Ramped Iris Detector Push-Pull Principle: Repeller and Extractor Standard Source & Detector mode Page 35

36 Page 36 Why Is the New Extractor Source More Valuable for Triple Quad MS/MS than Single Quad MSD?

37 Extractor Source Increases Signal for All Ions EI: spectrum of analyte can also include ions from matrix, column bleed, gases, etc. Single Quad MS Chemical Noise Limited Analyte Signal Increased Chemical Noise Increased S/N does not improve Page 37

38 Extractor Source Increases Signal for All Ions EI: spectrum of analyte can also include ions from matrix, column bleed, gases, etc. MS/MS still eliminates chemical noise Product 2 Q1 SIM isolate precursor before CID Product 3 Product 1 chemical noise from these other ions is eliminated CID + Q2 SIM Page 38

39 Why Is the New Extractor Source Valuable for MS/MS? Single Quad MS Chemical Noise Limited Analyte Signal Increased Chemical Noise Increased S/N does not improve Triple Quad MS/MS Product Ion Signal Limited Product Ion Signal Increased Chemical Noise Does NOT Increase S/N (and LOD) are improved Page 39

40 Unique Entrance Lens Eliminates Pre-filter Improved lens The last lens, which extends into the heated monolithic quadrupole, optimizes coupling of the source to the analyzer. Voltage on this lens is dynamically ramped synchronously with analyzer voltage to focus maximum ion signal for every m/z into the center of the quadrupole field. Page 40

41 7000B MS/MS: New Ramped-Iris Detector Ramped-iris improves transfer of ion from the exit of the second analyzer into the Triple-Axis detector New Ramped Iris Triple-Axis Detector Page 41

42 High Performance Triple-Axis Detector with New Dynamically Ramped Iris Maximum reduction of neutral noise The combination of helium quench gas and the Triple-Axis design delivers ultra low noise Long life and high linearity Triple channel electron multiplier Gain Normalization corrects tune file for detector aging Consistent long term method sensitivity Design proven in the 5975C MSD hyperbolic quartz transmission quadrupole analyzer Y ion beam steering rod shield for secondary particles X Offset Z high energy dynode triple channel electron multiplier Page 42

43 Ramped-Iris Triple-Axis Detector Improved high mass peak shape Improved mid to high mass sensitivity Mass Iris (V) Signal Increase X X X X X X Page 43

44 7000B GC/MS/MS Specifications Installation Checkout Specifications EI SRM > 500:1 for 100 fg OFN PCI SRM (CH4) > 50:1 for 100 fg BZP Typical Precision EI MRM Pk Area RSD < 8% for 100 fg OFN Typical Sensitivity EI Scan > 300:1 for 1 pg OFN NCI SIM (CH4) > 200:1 for 10 fg OFN OFN = octafluoronaphthalene BZP = benzophenone Page 44

45 Column Backflush: Many Wonderful Advantages for GC/MS/MS of Complex Matrices Page 45

46 What Real Value Does High Performance EPC Have If Matrix Carryover Destroys Rt Precision? Norm Runs without Backflushing Retention times shift ~4-5 sec during 10 runs m in Norm Runs with Backflushing m in Page 46

47 Productivity: CFT Backflush More analyses per hour... Shorter analysis cycle time (time from injection to injection) Reduced requirement for sample preparation with consistently better results... Quantitative: accuracy, precision, detection limits, linear range Qualitative: response ratios, confirmatory retention times, library searches (eliminate false negatives and false positives) while using less operator time for each analysis. Less time performing maintenance on column and ion source Page 47

48 Multiple Backflush Configurations Multiple Backflush Devices Backflush from the start of the column Backflush guard column only Backflush from the end of the column Backflush entire column Backflush from the middle of the column Backflush a percentage of the column Example: at tee between two 15 m columns New Pressure Controlled Tee Kit Purged Ultimate Union Page 48

49 Backflush Mid-Column (Backflush Mode) Split/Splitless Injection Port Pressure / Flow Controller 7890A GC Vent Decrease inlet pressure during backflush Pressure Controlled Tee 5975C MSD or 7000B MS/MS 15-m HP-5ms (0.25mm id x0.25um) 15-m HP-5ms (0.25mm id x0.25um) Page 49

50 Backflush with Purged Ultimate Union: Lipid Peroxidation Products in Blood 4-hydroxy-2,3-nonenal (HNE), an indicator of oxidative stress, and its metabolite, 1,4- dihydroxynonene (DHN) PCI with NH 3 App Note EN Late eluting peaks were quickly contaminating the source Page 50

51 Backflush with Purged Ultimate Union: Loss of Response without Backflush with backflush without backflush App Note EN Page 51

52 Backflush with Purged Ultimate Union: Loss of Response without Backflush with backflush These were the first backflush results; the system is getting clearer without backflush App Note EN Page 52

53 Break for Questions For questions, at break please dial *1 on your phone, or type onto the Q&A box at any time during the presentation. Page 53 Page 53 Pesticide Analysis with New Agilent Jet Stream Technology December 12, 2008

54 MassHunter Software Page 54

55 MassHunter Software Common Workflow for all Mass Spec. applications ChemStation Single Quad LC/MS ChemStation Single Quad GC/MS (MSD) LC/MS Acquisition Module Translator Translator Triple Quad LC/MS, LC/TOF, LC/QTOF MassHunter Quant and Qual modules GC/MS Acquisition Module Triple Quad GC/MS MassHunter format Page 55

56 MassHunter Data Analysis Applications One software for all your Agilent MS applications Quantitative Analysis Target Compound Analysis High throughput quantitation Qualitative Analysis Discovery & Method Development Find features Identify compounds Page 56

57 Quantitative Analysis View Results Information Page 57

58 Quantitative Analysis View Results Information Compound Information Page 58

59 Quantitative Analysis View Results Information Compound Information Calibration Information Page 59

60 Results Information: All Results for batch in one view Easy visualization for all results Customize columns for desired data Easy Navigation Compound and Calibration views change when new compound is selected Reduce data by filtering on sample type, compound, time segment, and outlier Page 60

61 Results Information: Detailed compound view Detailed compound information is easily seen Page 61

62 Results Information: Highlight outliers Compounds with results outside specified ranges are highlighted 30 metrics are available Custom Calculation is possible More to be added i.e. Peak metrics Page 62

63 Results Information: Highlight outliers Compounds with results outside specified ranges are highlighted 30 metrics are available Custom Calculation is possible More to be added i.e. Peak metrics Page 63

64 Results Information : Simple Metrics Plot Columns Ability to see trends in a batch Ability to diagnose quantitation problems easily Missing ISTD amount Retention time drift Page 64

65 Compound Information: All relevant information in one view Compound Integration Qualifier Information Spectrum Information ISTD Integration ISTD Qualifier Information ISTD Spectrum Information Page 65

66 Calibration Curve Views Can include ISTD, QC, and CC data points on curve Point information is available as pop-up QC point Page 66

67 Calibration Curve Assistant Display all possible Calibration Curve fit options Review R2, Standard error, or Max % Residual Can sort, filter, or browse options Page 67

68 Calibration Curve Assistant Filter possible calibration curves based on requirements Compare selected calibration fit (black) with proposed (blue). Linear All points Ignore origin Page 68

69 Reporting and Data Transfer Use of XML results format Easy data exchange LIMS data exchange Copy/Paste views from MassHunter Microsoft Excel reporting engine Powerful Features Excel Macros and formulas can be created for custom result manipulation Flexible Reports and Layouts Custom Layouts and graphics properties. Preconfigured report template examples Page 69

70 Productivity of GC/MS/MS Sample complexity forces the change from MS to MS/MS Could not meet required LOD/LOQ in real samples Could not avoid interferences from matrix ions (isobaric ions) Page 70

71 Productivity of GC/MS/MS Sample complexity forces the change from MS to MS/MS Could not meet required LOD/LOQ in real samples Could not avoid interferences from matrix ions (isobaric ions) Practical aspects of productivity MS/MS selectivity for less sample preparation MS/MS sensitivity for less injected on-column In turn, less maintenance MS/MS selectivity for faster analysis cycles MS/MS selectivity lessens the requirement for column selectivity Less bakeout or backflush may be needed Caution: selectivity is a type of blindness and can be misleading MS/MS selectivity for faster and easier peak integration What can be easier than a flat baseline and a single peak Page 71

72 Break for Questions For questions, at break please dial *1 on your phone, or type onto the Q&A box at any time during the presentation. Page 72 Page 72 Pesticide Analysis with New Agilent Jet Stream Technology December 12, 2008

73 Thank You for Attending Today s eseminar Terry L. Sheehan, Ph.D. GC/MS/MS Product Manager Agilent Technologies, Santa Clara, CA Page 73

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies Page 1 Agilent GC/MS Portfolio Smallest footprint, single GC column 5975VL with 6850GC Industry leading

More information

Highly Sensitive and Rugged GC/MS/MS Tool

Highly Sensitive and Rugged GC/MS/MS Tool Highly Sensitive and Rugged GC/MS/MS Tool For Pesticide Multiresidue Analysis in Food Samples Agilent 7 Series Triple Quadrupole GC/MS. The world s first MS/MS designed specifically for GC Analysis Introduction

More information

Workflow Using Hyphenated Mass Spectrometry Techniques

Workflow Using Hyphenated Mass Spectrometry Techniques Pesticide Analysis Workflow Using Hyphenated Mass Spectrometry Techniques Chinkai (Kai) Meng, Ph.D. Senior Applications Chemist Wilmington, Delaware USA chin_meng@agilent.com com February 12, 2009 Page

More information

Agilent 5977A Series GC/MSD System

Agilent 5977A Series GC/MSD System Agilent 5977A Series GC/MSD System Data Sheet 5977A GC/MSD Overview The Agilent 5977A Series GC/MSD builds on a 45-year tradition of leadership and innovation, bringing together the technologies of the

More information

Pesticide Analysis Workflow Using Hyphenated Mass Spectrometry Techniques

Pesticide Analysis Workflow Using Hyphenated Mass Spectrometry Techniques Pesticides e-seminar Series - Pesticide Analysis Workflow Using Hyphenated Mass Spectrometry Techniques Date: 30 April 2009 Time : 10am Singapore Time For audio, please dial one of the phone numbers listed:

More information

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses Technical Overview Authors Abstract Peter Stone, Thomas Glauner, Frank Kuhlmann, Tim Schlabach and Ken Miller

More information

Agilent MassHunter Quantitative Data Analysis

Agilent MassHunter Quantitative Data Analysis Agilent MassHunter Quantitative Data Analysis Presenters: Howard Sanford Stephen Harnos MassHunter Quantitation: Batch and Method Setup Outliers, Data Review, Reporting 1 MassHunter Quantitative Analysis

More information

Agilent 6400 Series Triple Quadrupole LC/MS/MS Users Session

Agilent 6400 Series Triple Quadrupole LC/MS/MS Users Session Agilent 6400 Series Triple Quadrupole LC/MS/MS Users Session QQQ Method Development and Optimization MassHunter Quant: Method setup Peak detection optimization Quant troubleshooting David Presser Application

More information

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application Food Safety, Environmental Authors Chin-Kai Meng and Bruce Quimby Agilent Technologies, Inc. 2850 Centerville

More information

Taking Full Advantage of UHPLC with Agilent 6460A Triple Quadrupole MS. Outline

Taking Full Advantage of UHPLC with Agilent 6460A Triple Quadrupole MS. Outline Taking Full Advantage of UHPLC with Agilent 0A Triple Quadrupole MS Dr. Anabel Fandino R&D Application Scientist LC/MS Division Agilent Technologies Santa Clara, CA utline Ultra High Definition LC/MS using

More information

USP<467> residual solvents

USP<467> residual solvents Application Note Pharmaceuticals USP residual solvents Applying the Agilent 977A MSD with the Agilent 797A headspace sampler and Agilent 7890B GC Authors Roger L. Firor and Mike Szelewski Agilent Technologies,

More information

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems Technical Overview Introduction All Ions MS/MS is a technique that is available for Agilent high resolution

More information

EGASUS. 4D GCxGC TOFMS. Delivering the Right Results

EGASUS. 4D GCxGC TOFMS. Delivering the Right Results P EGAU 4D GCxGC TFM Delivering the Right Results LEC s P EGAU 4D GCxGC-TFM The Ultimate Analytical Instrument Pioneering GCxGC technology, combined with the Pegasus TFM and ChromaTF software provides today

More information

Useful GC/MS Sensitivity for Improved Environmental Analyses. GC/MS Agilent Restricted

Useful GC/MS Sensitivity for Improved Environmental Analyses. GC/MS Agilent Restricted Useful Sensitivity for Improved Environmental Analyses Page 1 Definitions Sensitivity Analytical definition: slope of the plot of amount of analyte (x-axis) versus signal response (y-axis) Commonly used

More information

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Requirements of an Automated Sample Delivery System in Today s Realm of Ever Increasing Sensitivity Demands Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek Abstract

More information

Agilent ESI and APCI sources: for polar to non-polar compounds

Agilent ESI and APCI sources: for polar to non-polar compounds 1 Agilent 6400 Series Triple Quadrupole Users Workshop 1 Agilent ESI and APCI sources: for polar to non-polar compounds Nebulizer Pressure Corona current Nebulizer Pressure Vaporizer Vcap Vcap Drying Gas

More information

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7B Triple Quadrupole GC/MS Application Note Authors Dale R. Walker and Fred Feyerherm Agilent Technologies, Inc.

More information

Agilent MassHunter Quantitative Data Analysis

Agilent MassHunter Quantitative Data Analysis Agilent MassHunter Quantitative Data Analysis Presenters: Howard Sanford Stephen Harnos MassHunter Quantitation: Batch Table, Compound Information Setup, Calibration Curve and Globals Settings 1 MassHunter

More information

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application Environmental Analysis Author Mike Szelewski Agilent Technologies, Inc. 2850 Centerville

More information

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs Application Note Author Harry Prest, Ph.D. Senior Chemist in GC-MS Agilent Technologies, Inc. 5301 Stevens Creek Blvd Santa Clara,

More information

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Application Note Clinical Research Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Authors Yanan Yang 1, Victor Mandragon 2, and Peter Stone 1 1

More information

Enhanced Sensitivity for Biomarker Characterization in Petroleum Using Triple Quadrupole GC/MS and Backflushing

Enhanced Sensitivity for Biomarker Characterization in Petroleum Using Triple Quadrupole GC/MS and Backflushing Enhanced Sensitivity for Biomarker Characterization in Petroleum Using Triple Quadrupole GC/MS and Backflushing Application Note Environmental Authors Melissa Churley Agilent Technologies, Inc. 5310 Stevens

More information

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology Agilent s New 8800 ICP-QQQ Transforming ICP-MS Technology 1 8800 ICP-QQQ Key Product Hardware High matrix introduction (HMI) technology Dual conical Extraction and Omega lens focuses ions across the mass

More information

Electron Ionization Source Parameters in GC-MS

Electron Ionization Source Parameters in GC-MS Electron Ionization Source Parameters in GC-MS Inge De Dobbeleer Regional Marketing Manager GC and GC-MS, EMEA Thermo Fisher Scientific, Breda/The Netherlands The world leader in serving science Content

More information

Making Sense of Differences in LCMS Data: Integrated Tools

Making Sense of Differences in LCMS Data: Integrated Tools Making Sense of Differences in LCMS Data: Integrated Tools David A. Weil Agilent Technologies MassHunter Overview Page 1 March 2008 How Clean is our Water?... Page 2 Chemical Residue Analysis.... From

More information

Self Cleaning Ion Source and other Innovations in Mass Spectrometry. Kirk Lokits (GCMS Applications Chemist) Blacksburg, VA February 2 nd, 2017

Self Cleaning Ion Source and other Innovations in Mass Spectrometry. Kirk Lokits (GCMS Applications Chemist) Blacksburg, VA February 2 nd, 2017 Self Cleaning Ion Source and other Innovations in Mass Spectrometry Kirk Lokits (GCMS Applications Chemist) Blacksburg, VA February 2 nd, 2017 1 Jet Clean! Outcome JetClean self-cleaning ion source technology

More information

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC Applying the Agilent 5977A MSD to the Analysis of USP Residual Solvents with the 7697A Headspace Sampler and 7890B GC Application Note Pharmaceuticals Authors Roger L. Firor and Mike Szelewski Agilent

More information

WADA Technical Document TD2003IDCR

WADA Technical Document TD2003IDCR IDENTIFICATION CRITERIA FOR QUALITATIVE ASSAYS INCORPORATING CHROMATOGRAPHY AND MASS SPECTROMETRY The appropriate analytical characteristics must be documented for a particular assay. The Laboratory must

More information

Agilent 6400 Series Triple Quadrupole Users Workshop. MH QQQ Users workshop 2/21/2014 1

Agilent 6400 Series Triple Quadrupole Users Workshop. MH QQQ Users workshop 2/21/2014 1 Agilent 6400 Series Triple Quadrupole Users Workshop 1 Agilent 6400 Series Triple Quadrupole Users Workshop QQQ Method Development and Optimization MassHunter Quantitation: Batch and Method setup QC s,

More information

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note Authors Siji Joseph Agilent Technologies India Pvt. Ltd.

More information

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Application Note Authors Kevin McCann, Sameer Nene, Doug McIntyre, Edmond Neo, Dennis Nagtalon, and

More information

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION Quantitation by High Resolution Mass Spectrometry: Using Target Enhancement and Tof-MRM to Achieve Femtogram-level On-column Sensitivity for Quantitation of Drugs in Human Plasma Yun W. Alelyunas, Mark

More information

Translational Biomarker Core

Translational Biomarker Core Translational Biomarker Core Instrumentation Thermo Scientific TSQ Quantum Triple Quadrupole Mass Spectrometers. There are two TSQ Quantum Ultra AM instruments available in the TBC. The TSQ Quantum Ultra

More information

Agilent All Ions MS/MS

Agilent All Ions MS/MS Agilent All Ions MS/MS Workflow Overview A Determine fragment ions for LC/MS Quant method B Develop final Quant method Develop LC/MS Qualitative Analysis method Process data with Find by Formula Build

More information

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology Gas Chromatography/Mass Spectrometry Author Harry Prest 1601 California Avenue

More information

Author. Abstract. Introduction

Author. Abstract. Introduction Improved Performance for the Analysis of Aromatics in Gasoline by ASTM Method D5769 Using the Agilent 5973 inert Gas Chromatography/Mass Spectrometry System Application Author James D. McCurry Agilent

More information

GC-APCI IMS of Diesel

GC-APCI IMS of Diesel GC-APCI IM of Diesel Application Note Energy and Chemicals Authors heher Bano Mohsin, David Wong, and F. Robert Ley Agilent Technologies, Inc. Abstract This Application Note describes the use of ion mobility

More information

Guide to Peptide Quantitation. Agilent clinical research

Guide to Peptide Quantitation. Agilent clinical research Guide to Peptide Quantitation Agilent clinical research Peptide Quantitation for the Clinical Research Laboratory Peptide quantitation is rapidly growing in clinical research as scientists are translating

More information

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors Trace-Level Analysis of Melamine in Milk Products on Agilent 789A/5975C GC/MSD Using a ew Agilent J&W DB-5ms Ultra Inert Column and SampliQ SCX Cartridges Application ote Gas Chromatography/Mass Spectrometry/Food

More information

The Emergence of Benchtop Tandem Quadrupole Mass Spectrometers as a Routine Tool in Quantitative GC/MS/MS Applications

The Emergence of Benchtop Tandem Quadrupole Mass Spectrometers as a Routine Tool in Quantitative GC/MS/MS Applications The Emergence of Benchtop Tandem Quadrupole Mass Spectrometers as a Routine Tool in Quantitative GC/MS/MS Applications Doug Stevens Waters Corp. Introduction The range of GC/MS instrument types has expanded

More information

Improved Detection of Food Contamination using GC/QQQ and GC/QTOF

Improved Detection of Food Contamination using GC/QQQ and GC/QTOF Improved Detection of Food Contamination using GC/QQQ and GC/QTOF Philip L. Wylie, Ph.D. Sr. Applications Chemist for Food Safety Agilent Technologies December 1, 011 1 Some of the Food and Feed Contaminants

More information

Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software

Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software NEMC, Washington D.C. August 6 th 2012 Nicola Watson Environmental Specialist Markes International

More information

László Tölgyesi LC/MS scientist EFSC Demo Group. International Symposium on Food Safety and Quality: Applications of Nuclear and Related Techniques

László Tölgyesi LC/MS scientist EFSC Demo Group. International Symposium on Food Safety and Quality: Applications of Nuclear and Related Techniques Minimizing matrix effects in multiresidue pesticide analysis using a new sensitivity-enhanced triple quadrupole instrument in combination with extensive sample dilution László Tölgyesi LC/MS scientist

More information

Multi-residue analysis of pesticides by GC-HRMS

Multi-residue analysis of pesticides by GC-HRMS An Executive Summary Multi-residue analysis of pesticides by GC-HRMS Dr. Hans Mol is senior scientist at RIKILT- Wageningen UR Introduction Regulatory authorities throughout the world set and enforce strict

More information

MassHunter TOF/QTOF Users Meeting

MassHunter TOF/QTOF Users Meeting MassHunter TOF/QTOF Users Meeting 1 Qualitative Analysis Workflows Workflows in Qualitative Analysis allow the user to only see and work with the areas and dialog boxes they need for their specific tasks

More information

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany Analysis of Pesticides by Supercritical Fluid Chromatography/ Mass Spectrometry Optimizing the Agilent 126 Infi nity Analytical SFC System in Combination with the Agilent 646 Triple Quadrupole LC/MS Application

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting Application Note Authors Angela Smith Henry and Bruce Quimby Agilent

More information

Agilent G3212 GC-APCI Source

Agilent G3212 GC-APCI Source Agilent G3212 GC-APCI Source Quick Start Guide Where to find information 2 Getting Started 3 Step 1. Start the Data Acquisition program for the GC and the Q-TOF 3 Step 2. Prepare the GC and Q-TOF for data

More information

Agilent 6495 Triple Quadrupole LC/MS System EXPERIENCE A NEW LEVEL OF CONFIDENCE

Agilent 6495 Triple Quadrupole LC/MS System EXPERIENCE A NEW LEVEL OF CONFIDENCE Agilent 49 Triple Quadrupole LC/MS System EXPERIENCE A NEW LEVEL OF CONFIDENCE AGILENT 49 TRIPLE QUADRUPOLE LC/MS SYSTEM WORK SMARTER, DO MORE ifunnel new Q1 Optics new Ion Detector IMPROVED SENSITIVITY

More information

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Application Note Food Authors Jianqiu Mi, Zhengxiang Zhang, Zhixu Zhang,

More information

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Automated Handling Techniques for the Analysis of Elevated Volatile Organic Compound (VOC) Concentrations in Soils Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek

More information

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note Abstract Area and retention time are the two primary measurements in gas chromatography. The precision with which

More information

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry PO-CON111E Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry Pittcon 201 1120-2 Brahm Prakash, William Lipps, Di Wang, Shilpi Chopra, Nicole

More information

UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation

UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation Technical Overview Introduction Inert flow path technology Modern GC and GC/MS instrumentation is an important analytical tool

More information

The 5975 inert MSD Benefits of Enhancements in Chemical Ionization Operation Technical Note

The 5975 inert MSD Benefits of Enhancements in Chemical Ionization Operation Technical Note The 5975 inert MSD Benefits of Enhancements in Chemical Ionization Operation Technical Note Authors Chris Sandy Agilent Technologies Wokingham, Berkshire, UK Jean-Francois Garnier Agilent Technologies

More information

brochure brief 5973Network Mass Selective Detector General Description and Configurations

brochure brief 5973Network Mass Selective Detector General Description and Configurations brochure brief 5973Network Mass Selective Detector This publication provides general technical information about the 5973Network MSDs to accompany the specifications and brochure. The MSD series was originally

More information

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK Multi-residue GC-MS analysis Richard Fussell CSL York, UK e-mail: r.fussell@csl.gov.uk Outline of Presentation Background Gas Chromatography Mass Spectrometry - Options Capabilities and Applications Summary

More information

The Tool Box Getting More Productivity from Your GC and GC/MSD

The Tool Box Getting More Productivity from Your GC and GC/MSD The Tool Box Getting More Productivity from Your GC and GC/MSD M Ramesh Application Engg GC/GCMS Agilent Technologies India Pvt. Ltd. Productivity Tool Box for GC & GC/MSOverview 5975C MSD Agilent Restricted

More information

Gas Chromatograph Mass Spectrometer GCMS-TQ8040 C146-E251

Gas Chromatograph Mass Spectrometer GCMS-TQ8040 C146-E251 Gas Chromatograph Mass Spectrometer GCMS-TQ8040 C146-E251 with Smart Technologies Smart Productivity, Smart Operation, Smart Performance Finally, a triple quadrupole GCMS Smart enough for everyday use

More information

Die Nadel im Heuhaufen

Die Nadel im Heuhaufen Die Nadel im Heuhaufen Workflow zur Identifizierung unerwarteter Komponenten in LC Q-Tof Daten Umwelt & Lebensmittel Seminar Tour Andreas Reimann Produktspezialist LC-MS Agilent Technologies Instrumentation

More information

Agilent MassHunter Profinder: Solving the Challenge of Isotopologue Extraction for Qualitative Flux Analysis

Agilent MassHunter Profinder: Solving the Challenge of Isotopologue Extraction for Qualitative Flux Analysis Agilent MassHunter Profinder: Solving the Challenge of Isotopologue Extraction for Qualitative Flux Analysis Technical Overview Introduction Metabolomics studies measure the relative abundance of metabolites

More information

Agilent Technologies LCMS Portfolio

Agilent Technologies LCMS Portfolio Agilent Technologies LCMS Portfolio Hardware and Software Solutions for the Most Challenging Applications Jean-Francois Roy MS Product Specialist Agilent Technologies 1 Raising the Standard in UHPLC Highest

More information

Trim Time Not Columns Multiresidue Pesticides Analysis with an Intuvo GC/TQ

Trim Time Not Columns Multiresidue Pesticides Analysis with an Intuvo GC/TQ Trim Time Not Columns Multiresidue Pesticides Analysis with an Intuvo GC/TQ Courtney Milner, Ph.D. Agilent Technologies Santa Clara, CA ASMS 2018 VIP Session 1 Agenda Pesticide analysis trends Targeted

More information

Gas Chromatograph Mass Spectrometer GCMS-TQ8050 C146-E326

Gas Chromatograph Mass Spectrometer GCMS-TQ8050 C146-E326 Gas Chromatograph Mass Spectrometer GCMS-TQ8050 C146-E326 with Smart Abilities Enhanced Sensitivity Leads to New GC-MS/MS Possibilities The GCMS-TQ8050 features a new highly efficient detector and three

More information

Thermo Finnigan LTQ. Specifications

Thermo Finnigan LTQ. Specifications IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Finnigan LTQ Specifications

More information

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada) Quantitation and Identification of Pharmaceuticals and Personal Care Products (PPCP) in Environmental Samples using Advanced TripleTOF MS/MS Technology André Schreiber AB SCIEX Concord, Ontario (Canada)

More information

Mass Spectrometry in MCAL

Mass Spectrometry in MCAL Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS GC seperates small, volatile, non-polar material MS is detection devise (Agilent 320-MS TQ Mass Spectrometer) Full scan monitoring SIM single ion monitoring

More information

Introducing the Agilent G6410A Triple Quadrupole Mass Spectrometer

Introducing the Agilent G6410A Triple Quadrupole Mass Spectrometer Introducing the Agilent G6410A Triple Quadrupole Mass Spectrometer Michael Zumwalt Quad Mass Filter (Q1) Quad Mass Filter (Q3) Ion source Collision Cell Overview of QQQ Hardware QQQ Mechanical Layout Common

More information

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol. Determination of Volatile Petroleum Hydrocarbons ANNE JUREK Abstract: Due to current events, the importance of determining volatile petroleum hydrocarbons in both soils and waters has become an issue.

More information

Comprehensive Pesticide Analysis in Juice Using a Combination of GC/MS and LC/MS Methods

Comprehensive Pesticide Analysis in Juice Using a Combination of GC/MS and LC/MS Methods Comprehensive Pesticide Analysis in Juice Using a Combination of GC/MS and LC/MS Methods Application Note Food Safety Authors Fadwa Al-Taher, Rima Juskelis, Yang Chen, Jack Cappozzo Institute for Food

More information

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401 Gas Chromatograph Mass Spectrometer No. GCMS-1401 Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS Introduction Development of methods for analysis of drugs of abuse has become a high

More information

Accurate Mass for GC/MSD & LC/MSD

Accurate Mass for GC/MSD & LC/MSD Accurate Mass for GC/MSD & LC/MSD A Presentation at Agilent Hospitality VIP Event Yongdong Wang, PhD Cerno Bioscience Norwalk, CT, USA June 6, 2018 Agenda Cerno s Novel MS Calibration: Why & How Application

More information

Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365

Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365 Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365 Effortless Performance The LCMS-9030 quadrupole time-of-flight (Q-TOF) mass spectrometer integrates the world s fastest

More information

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Application Note Doping Control Authors Prof. Maria Kristina Parr Freie Universität Berlin Institute

More information

Advantages of Reversed Sandwich Injection for Pesticide Residue Analysis

Advantages of Reversed Sandwich Injection for Pesticide Residue Analysis Advantages of Reversed Sandwich Injection for Pesticide Residue Analysis Application Note Author Jessica Westland Agilent Technologies, Inc. Abstract This application note demonstrates the use of the Agilent

More information

MassHunter Software Overview

MassHunter Software Overview MassHunter Software Overview 1 Qualitative Analysis Workflows Workflows in Qualitative Analysis allow the user to only see and work with the areas and dialog boxes they need for their specific tasks A

More information

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF Analysis of Biomarkers in Crude Oil Using the Agilent 7 GC/Q-TOF Application Note Petrochemical and Environmental Authors Frank David Research Institute for Chromatography, Kennedypark 6, B-85 Kortrijk,

More information

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays : Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays Mark Wrona, Jayne Kirk, and Yun Alelyunas Waters Corporation, Milford, MA, USA APPLICATION BENEFITS Ability

More information

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific LTQ Orbitrap Velos

More information

Finnigan LCQ Advantage MAX

Finnigan LCQ Advantage MAX www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Finnigan LCQ Advantage MAX The Finnigan LCQ Advantage MAX ion trap mass spectrometer

More information

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70 Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70 Application Note Abstract Geosmin and 2-Methylisoborneol are organic compounds that have a distinct scent. These compounds

More information

Analysis of Polar Metabolites using Mass Spectrometry

Analysis of Polar Metabolites using Mass Spectrometry Analysis of Polar Metabolites using Mass Spectrometry TransMed Course: Basics in Clinical Proteomics and Metabolomics. Oct 10-19, 2012 dd.mm.yyyy Vidya Velagapudi, Ph.D, Adjunct Professor Head of the Metabolomics

More information

Theory and Practical Consequences.

Theory and Practical Consequences. Theory and Practical Consequences. Number of hits depending on reachable mass accuracy 250 hits @ ±10 ppm 137 hits @ ±5 ppm 78 hits @ ±3 ppm 29 hits @ ±1 ppm 3 hits @ ±0.1 ppm 1 hit @ ± 0.05 ppm Intens.

More information

Introduction to Capillary GC

Introduction to Capillary GC ?? Kβ? Page 1 Typical GC System Gas supply Injector Detector Data handling GAS Column Oven Page 2 CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention

More information

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers Application Note Forensic and Toxicology - Criminalistics Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers Using the Agilent 126 Infinity II SFC and Detection by Coupled

More information

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Sensitive Detection of 2-MIB and Geosmin in Drinking Water Sensitive Detection of -MIB and Geosmin in Drinking Water Application Note Environmental Author Yean-Woong You Agilent Technologies, Inc. Seoul, Korea Abstract An automated SPME extraction method for easy

More information

Optimizing Conditions for Volatiles Analysis

Optimizing Conditions for Volatiles Analysis July 2016 SPEXCertiPrep.com/ECS Optimizing Conditions for Volatiles Analysis by Mark A. Ferry GC/MS Technical Consultant MFerry@SPEX.com A publication by Environmental Consulting & Supplies a Division

More information

Rapid, Robust, and Sensitive Detection of 11-nor-D 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair

Rapid, Robust, and Sensitive Detection of 11-nor-D 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair Rapid, Robust, and Sensitive Detection of -nor-d 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair Application Note Forensic Toxicology/Doping Control Authors David Engelhart Omega Labs, Inc. Mogadore,

More information

Agilent TOF Screening & Impurity Profiling Julie Cichelli, PhD LC/MS Small Molecule Workshop Dec 6, 2012

Agilent TOF Screening & Impurity Profiling Julie Cichelli, PhD LC/MS Small Molecule Workshop Dec 6, 2012 1 Agilent TOF Screening & Impurity Profiling Julie Cichelli, PhD LC/MS Small Molecule Workshop Dec 6, 2012 Review: Technology for Accurate Mass Analysis: TOF LC/MS Mass measurements accurate to several

More information

Quattro Micro - How does it work?

Quattro Micro - How does it work? Quattro Micro - How does it work? 1 Introduction This document is designed to familiarise you with the principles behind how the Quattro Micro works. The level of this document is designed as Level One

More information

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring High-Throughput Protein Quantitation Using Multiple Reaction Monitoring Application Note Authors Ning Tang, Christine Miller, Joe Roark, Norton Kitagawa and Keith Waddell Agilent Technologies, Inc. Santa

More information

An Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining

An Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining An Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining Dave Weil, Ph.D. and Jim Lau, Ph.D. Typical Method Conditions: 1260 UHPLC

More information

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors. Sensitive Femtogram Determination of Aflatoxins B 1, B 2, G 1 and G 2 in Food Matrices using Triple Quadrupole LC/MS Application Note Food Safety Authors Yang Chen and Jack Cappozzo National Center for

More information

GCXGC-qMS with Total Flow Modulation for Volatile secondary metabolites analysis from organically-grown apples by HS-SPME

GCXGC-qMS with Total Flow Modulation for Volatile secondary metabolites analysis from organically-grown apples by HS-SPME GCXGC-qMS with Total Flow Modulation for Volatile secondary metabolites analysis from organically-grown apples by HS-SPME Users Meeting RIVA DEL GARDA 20 May 2014 Eligio Sebastiani - SRA Instruments SpA

More information

Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS

Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS Application Note # CA 284780 Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS Abstract The analysis of semi-volatile organic compounds

More information

Finnigan TRACE DSQ GC/MS

Finnigan TRACE DSQ GC/MS m a s s s p e c t r o m e t r y Finnigan TRACE DSQ GC/MS Discover Unmatched Sensitivity, Performance and Value Full-scan sensitivity for the demanding environmental analysis Proven performance for difficult

More information

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source Application Note Environmental Authors Peter Gautschi and Harry Prest Senior Application Scientist

More information

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 646 Triple Quadrupole LC/MS system with Triggered MRM Application Note Authors Thomas Glauner Agilent Technologies,

More information

Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode

Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode PO-CON1610E Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode Pittcon 2016 1110-2 Brahm Prakash, William Lipps, Di Wang, Shilpi Chopra, Nicole

More information