The new Agilent 1290 Infinity Quaternary LC

Size: px
Start display at page:

Download "The new Agilent 1290 Infinity Quaternary LC"

Transcription

1 The new Agilent 1290 Infinity Quaternary LC How binary pump-like performance can enhance your lab efficiency Sebastian Krahe Produktspezialist HPLC Agilent Technologies Page 1

2 The 1290 Infinity Quaternary LC System Leveraging the Performance of the 1290 Infinity Binary LC System 1290 Infinity DAD with 60mm cell (& HDR optional) 1290 Infinity TCC with broadest valve selectivity 1290 Infinity Autosampler (with Flexible Cube for lowest carry-over) 1290 Infinity Quaternary Pump Page 2

3 Innovation: 1290 Infinity Quaternary Pump Quaternary Pump with Binary Pump-like Performance 1290 Infinity Quaternary Pump Boost Performance Highest accuracy and precision for composition and flow, with exceptionally low delay volume Save Time Accelerates transfer of existing HPLC methods to UHPLC Reduce Costs Outstanding UHPLC performance at HPLC-like costs of ownership Page 3

4 1290 Infinity Quaternary Pump Various tools enabling the high performance Inlet-Weaver Switchable Jet Weaver Mixer 4-channel Degasser Multi-Purpose- Valve MCGV (Multi- Channel-Gradient- Valve) Pump Head In-line Filter Page 4

5 Multipurpose valve functions for highest comfort Standard Application for standard Delay Volume Page 5

6 Agilent 1290 Infinity Quaternary Pump Specifications & Benefits Power Range For any kind of analysis Composition Accuracy and Precision < 0.15 % RSD or 0.02 min SD ± 0.4 % (1-99 % Composition B) High RT precision in gradient runs Flow Accuracy and Precision < 0.07 % RSD or 0.01 min SD ± 1.0 % or 10 µl High RT precision in isocratic runs Composition Range 1-99 % Wide analytical range Delay Volume < 350 µl For fast quaternary gradients Page 6

7 Composition Accuracy and Precision Step Gradient Analysis with Tracer Different mixtures (H2O, MeOH, ACN) at different pressures and different flow rates Full application range! Page 7

8 Flow and Composition Accuracy and Precision Isocratic analysis of Glucocorticoids 1: Hydrocortisone 2: Beclomethasonedi-proprionate 3: Hydrocortisone acetate Column: Agilent ZORBAX Eclipse Plus C18, 4.6 x 100 mm, 5 µm (p/n ) Mobile phase: Water/ Acetonitrile = 30/70 Flow rate: 1 ml/min Stop time: 5 min Injection vol.: 3 µl (needle wash for 3 sec) Column Temp.: 30 C DAD: 254/10 nm, (ref 400/ 80 nm), 20 Hz, slit 4 nm Page 8

9 Flow and Composition Accuracy and Precision Isocratic analysis of Glucocorticoids mau RT of compounds very sensitive regarding changes in mobile phase composition! Water/ACN 29:71 Water/ACN 31:69 Water/ACN 30:70 min Page 9

10 Accuracy and Precision Isocratic analysis of Glucocorticoids Specification (< 0.07 %) mau : Hydrocortisone 2: Beclomethasonedi-proprionate : Hydrocortisone acetate min Page 10

11 Accuracy and Precision Dynamically vs. pre-mixed solvents Isocratic composition by dynamic mixing: A/B = 70:30 Isocratic composition by premixed solvent: Which method is better? RSD RT(%) A/B = 70:30 Page 11

12 Accuracy and Precision Dynamically vs. pre-mixed solvents Retention time precision Single operator, single bottle Dynamic mixing For illustration only, data can be found in Agilent Application Note EN Page 12

13 Accuracy and Precision Dynamically vs. pre-mixed solvents Retention time precision Single operator, single bottle Dynamic mixing Single operator, multiple bottles For illustration only, data can be found in Agilent Application Note EN Page 13

14 Accuracy and Precision Dynamically vs. pre-mixed solvents Retention time precision Single operator, single bottle Dynamic mixing Single operator, multiple bottles Multiple operators, multiple bottles For illustration only, data can be found in Agilent Application Note EN Page 14

15 Accuracy and Precision Gradient Analysis of Clopidole Clopidole Clopidole is used as prophylactics and/or therapeutics in the prevention or treatment of coccidiosis in the poultry and rabbit industries. It is added to finished feed in a concentration range of 100 to 125mg/kg. Column: Agilent ZORBAX SB-C18, 4.6 x 250 mm, 3.5 µm (p/n ) Mobile phase: A = Water/ACN (95/5) % TFA B = Water/Acetonitrile(40/60) % TFA Gradient: % B in 15min Flow rate: 1 ml/min Stop time: 15 min Injection vol.: 1 µl Column Temp.: 30 C DAD: 263/10 nm, (ref. 400/ 80 nm), 10 Hz Page 15

16 Accuracy and Precision Dynamically vs. pre-mixed solvents Gradient from 0 to 100 % with premixed solvent: H 2 O +0.1% TFA % B ACN +0.1% TFA Gradient from 5 to 60 % B by dynamic mixing: H 2 O +0.1% TFA ACN +0.1% TFA 5 60 % B Which method is better? RSD RT(%) Page 16

17 Accuracy and Precision Dynamically vs. pre-mixed solvents Dynamic mixing 5 60 % B Premixed, % B mau mau min min Page 17

18 Accuracy and Precision Dynamically vs. pre-mixed solvents Retention time precision RSD RT (%) Specification (< 0.15 %) 5 60 % B Dynamic mixing % B Premixed Page 18

19 Agilent 1290 Infinity Quaternary Pump Specifications & Benefits Power Range For any kind of analysis Composition Accuracy and Precision < 0.15 % RSD or 0.02 min SD ± 0.4 % (1-99 % Composition B) High RT precision in gradient runs Flow Accuracy and Precision < 0.07 % RSD or 0.01 min SD ± 1.0 % or 10 µl High RT precision in isocratic runs Composition Range 1-99 % Wide analytical range Delay Volume < 350 µl For fast quaternary gradients Page 19

20 Ternary/Quaternary Gradients Blend Assist You need different concentrations of modifiers in your analysis, would like to have just one stocksolution and do online dilution to profite from the quaternary mixing capability of your pump? Here is a simple tool BlendAssist! Water Water 1% TFA Desired method conditions - example: 1. 5 to 95% gradient of ACN with 0.1% TFA in Water and 0.08% TFA in ACN % gradient of ACN with 0.5% TFA in Water and 0.4% TFA in ACN Without BlendAssist you need to either pre-mix the required solvents or by using stock-solutions of TFA in Water and ACN to program complex gradients (%A, B, C, D). ACN ACN 1% TFA With BlendAssist: just program your binary organic/aqueous gradient and define the dilution factor! Page 20

21 Ternary/Quaternary Gradients Blend Assist, different Phosphate buffer concentrations H 3 C CH 3 N H 3 C CH 3 N O NH 2 C H 3 N O OH C H 2 N Disopyramide Quinidine Column: Agilent ZORBAX Eclipse Plus C18, 4.6 x 100 mm, 1.8 μm Mobile phase: A = Water B = Water/Acetonirile 20: M NaH 2 PO 4, ph=4.6 Gradient: var. Flow rate: 1.5 ml/min Stop time: 8.5 min Injection vol.: 1 µl Column Temp.: 40 C DAD: 270/10 nm, (ref. 360/ 100 nm), 20 Hz Page 21

22 Ternary/Quaternary Gradients Blend Assist, different Phosphate buffer concentrations mau Quinidine Procainamide Acetyl- Procainamide 0.05 M buffer M buffer Disopyramide 0.01 M buffer min Page 22

23 Ternary/Quaternary Gradients Blend Assist - simple Tool for Online-Dilution of Modifiers and Gradient Set-Up BlendAssist - Solvent/modifier channel definition Page 23

24 1290 Infinity Quaternary Pump Conclusion Full 1290 Infinity LC Power range Highest composition precision and accuracy, solvent independent and across pressure and flow rate range Full application range with binary-like pump performance! Blend Assist for Buffer/Additive Blending ISET-ready! Full flexibility! Page 24

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump Performance characteristics of the Agilent 129 Infinity Quaternary Pump Technical Overview Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract This Technical Overview presents Proof

More information

Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 1290 Infinity LC System with Blend Assist

Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 1290 Infinity LC System with Blend Assist Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 129 Infinity LC System with Technical Overview Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract During

More information

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India Effective use of pharmacopeia guidelines to reduce cost of chromatographic analysis Optimized, cost-effective HPLC analysis of atorvastatin by varying column dimensions within the USP allowed limts

More information

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India Reducing analysis time and solvent consumption for isocratic USP assay methods with current and proposed USP guidelines using the Agilent 129 Infinity LC System An efficient way to reduce cost of analysis

More information

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times The Secrets of Rapid HPLC Method Development Choosing Columns for Rapid Method Development and Short Analysis Times Rapid Analysis Is More Than Run Time It is developing a method to meet a goal and developing

More information

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent Infinity LC System with Diode Array Detector Application Note Author Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany [mau]

More information

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India QbD Based Method Development on an Agilent 129 Infinity UHPLC System Combined with a Seamless Method Transfer to HPLC Using Intelligent System Emulation Technology Application Note Pharmaceutical QA/QC

More information

Transferring HPLC methods from the HP 1090 Series to the Agilent 1100 Series HPLC system

Transferring HPLC methods from the HP 1090 Series to the Agilent 1100 Series HPLC system Transferring HPLC methods from the HP 1090 Series to the Agilent 1100 Series HPLC system Technical Note Introduction The HP 1090 Series HPLC system, introduced in 1983, is regarded to be an outstanding

More information

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 126 Infinity Binary LC system with UV detection Application Note Food Author Srividya Kailasam Agilent Technologies,

More information

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity Agilent 12 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity A case study with degraded omeprazole drug product Application Note Small Molecule

More information

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster Welcome to our E-Sear: Choosing HPLC Columns for Faster Analysis Smaller and Faster High Throughput/Fast LC Requires. Short columns 0 mm or shorter Small particle sizes. µm Rapid Resolution or new.8 µm

More information

Practical Applications of Method Translation Using the Agilent Method Translation Tool Rita Steed Inside Application Engineer Agilent Technologies

Practical Applications of Method Translation Using the Agilent Method Translation Tool Rita Steed Inside Application Engineer Agilent Technologies Practical Applications of Method Translation Using the Agilent Method Translation Tool Rita Steed Inside Application Engineer Agilent Technologies Title Objectives Demonstrate Agilent Method Translation

More information

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases Agilent 129 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases Technical Overview Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract The Agilent

More information

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note Developing a fast, generic method for rapid resolution LC with quadrupole MS detection Combining the Agilent 6140 quadrupole MS and multimode source with the Agilent 1200 Series Rapid Resolution LC system

More information

Agilent LC & LC-MS Cannabis Testing

Agilent LC & LC-MS Cannabis Testing Agilent LC & LC-MS Cannabis Testing Potency Analysis for Cannabinoids Mike Adams 1, A. Roth 1, Joan Stevens 2, Karen Kaikaris 1, & Sue D Antonio 2 1 CWC LABS, 2 Agilent Technologies, UV detection with

More information

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Figure 1. Structures for Vitamin B2 and Vitamin B1. CH 461 & CH 461H F 18 Name Experiment 2C Integrated Laboratory Experiment DETERMINATION OF RIBOFLAVIN: A COMPARISON OF TECHNIQUES PART C. HIGH PERFORMANCE LIQUID CHROMATOGRAPHY The purpose of this experiment

More information

Application Note. Agilent Application Solution Analysis of acaricides in honey. Authors. Abstract. Food

Application Note. Agilent Application Solution Analysis of acaricides in honey. Authors. Abstract. Food Agilent Application Solution Analysis of acaricides in honey Application Note Food Authors mau 12 1 A.G. Huesgen Agilent Technologies, Inc. Waldbronn, Germany 8 6 4 2 2 4 6 8 1 12 14 16 18 Abstract Varroosis

More information

High Resolution Fast LC

High Resolution Fast LC High Resolution Fast LC Easier Than You Think Rita Steed LC Columns Application Engineer May 9, 2013 What is High Resolution Fast LC? Maintain Resolution with Faster Run Time Increased Resolution with

More information

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography Technical Overview 2D-LC Conventional 1D-LC Abstract This Technical Overview presents

More information

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Figure 1. Structures for Vitamin B2 and Vitamin B1. CH 461 & CH 461H F 15 Name Experiment 2C Integrated Laboratory Experiment DETERMINATION OF RIBOFLAVIN: A COMPARISON OF TECHNIQUES PART C. HIGH PERFORMANCE LIQUID CHROMATOGRAPHY The purpose of this experiment

More information

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis High-Resolution Sampling D-LC for Pharmaceutical Impurity Analysis Detection of Impurities Hidden Under the API Peak at Relevant Levels Application Note Small Molecule Pharmaceuticals and Generics Authors

More information

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis Automated Switching Between D-LC and Comprehensive D-LC Analysis The Agilent 90 Infinity D-LC Solution Technical Overview Author Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany Abstract This

More information

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 129 Infinity LC system Application Note Food and Beverages Author Srividya Kailasam

More information

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS Application Note Traditional Chinese Medicine Author Rongjie Fu Agilent Technologies

More information

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation Technical Overview The Agilent InfinityLab D-LC Solution with Active Solvent Modulation Achieving Improved Resolution and Sensitivity for Challenging Combinations of Separation Conditions Author Sonja

More information

Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system

Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system Application Note Food Testing & Agriculture Authors Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany Keiko

More information

Automated method development using Agilent 1100 Series HPLC systems, Agilent ChemStation and ChromSword software. Application

Automated method development using Agilent 1100 Series HPLC systems, Agilent ChemStation and ChromSword software. Application Automated method development using Agilent 1100 Series HPLC systems, Agilent ChemStation and ChromSword software Application Sergey Galushko Ralf Honsberg Abstract The development of an HPLC method is

More information

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Application Note Authors Kevin McCann, Sameer Nene, Doug McIntyre, Edmond Neo, Dennis Nagtalon, and

More information

Successfully Scaling and Transferring HPLC and UPLC Methods

Successfully Scaling and Transferring HPLC and UPLC Methods Successfully Scaling and Transferring HPLC and UPLC Methods Esa Lehtorinne Esa_Lehtorinne@waters.com Tel: +358-9-5659 6288 Fax: +358-9-5659 6282 Waters Finland Kutomotie 16 00380 Helsinki 2013 Waters Corporation

More information

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform -xt SYSTEM Prep and Load Platform -xt Specifications for PAL-xt Systems Valid for PAL-xt System models only Revised March 2013 V5 PAL is a registered trademark of CTC Analytics AG Switzerland -xt SYSTEM

More information

LC and LC/MS Column Selection Flow Chart

LC and LC/MS Column Selection Flow Chart LC and LC/MS Column Selection Flow Chart To use the column selection diagram below, simply follow the path for your analyte and mobile phase. At the far right, follow your final column selection to the

More information

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy Analytical Quantification of Deoxygenated Compounds in Catalytic Reaction Samples as an Evaluation Technique to Detere Reaction Conversion from a Bioderived Oil in Novel Biofuel Testing Application Note

More information

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates Using the Agilent 9 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates Technical Overview Author Florian Rieck Agilent Technologies, Inc. Waldbronn, Germany Abstract This Technical

More information

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Application Note Clinical Research Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Authors Yanan Yang 1, Victor Mandragon 2, and Peter Stone 1 1

More information

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc. Agilent Application Solution Transfer of a USP method for tolazamide from normal phase HPLC to SFC using the Agilent 126 Infinity Hybrid SFC/UHPLC System Improving peak shape and providing wider UV selectivity

More information

Getting More Resolution from Your LC Michael Woodman Columns and Consumables September 10, 2008

Getting More Resolution from Your LC Michael Woodman Columns and Consumables September 10, 2008 Getting More Resolution from Your LC Michael Woodman Columns and Consumables September 10, 2008 Page 1 What are the Parameters Affecting Resolution? Isocratic Separations: physical and chemical separation

More information

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

Prep 150 LC System: Considerations for Analytical to Preparative Scaling Andrew Aubin and Jo-Ann Jablonski Waters Corporation, Milford, MA, USA APPLICATION BENEFITS The Prep 150 LC System is an affordable, highly reliable system for preparative chromatography and is suitable

More information

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application Food Author Michael Woodman Agilent Technologies, Inc. 285 Centerville Road Wilgton, DE 1988-161 USA Abstract Soft drinks,

More information

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone Mia Summers and Kenneth J. Fountain Waters Corporation, 34 Maple St.,

More information

HPLC Method Development: Standard Practices and New Columns

HPLC Method Development: Standard Practices and New Columns HPLC Method Development: Standard Practices and New Columns Edward Kim Application Engineer Agilent Technologies, Inc. March 8, 29 What are Some Standard Method Development Practices?. Follow preferred

More information

Detection of Trace Level Pharmaceuticals in Drinking Water by Online SPE Enrichment with the Agilent 1200 Infinity Series Online-SPE Solution

Detection of Trace Level Pharmaceuticals in Drinking Water by Online SPE Enrichment with the Agilent 1200 Infinity Series Online-SPE Solution Detection of Trace Level Pharmaceuticals in Drinking Water by nline SPE Enrichment with the Agilent Infinity Series nline-spe Solution Application Note Environmental Author Edgar Naegele Agilent Technologies,

More information

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer APPLICATION NOTE 72385 Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer Authors Stephan Meding, Katherine Lovejoy, Martin Ruehl Thermo Fisher Scientific,

More information

Fast methods for the determination of ibuprofen in drug products

Fast methods for the determination of ibuprofen in drug products APPLICATION NOTE 779 Fast s for the determination of ibuprofen in drug products Authors Sylvia Grosse, Mauro De Pra, Frank Steiner, Thermo Fisher Scientific, Germering, Germany Keywords Pharmaceutical,

More information

LC800. Smart HPLC. Until now UHPLC From now Smart HPLC

LC800. Smart HPLC. Until now UHPLC From now Smart HPLC LC800 Smart HPLC Until now UHPLC From now Smart HPLC Smart HPLC Leads to Ultimate Performance Patent Pending S LC800 is a completely new and unique concept, designed for maximum performance in high resolution

More information

Too Polar for Reversed Phase What Do You Do?

Too Polar for Reversed Phase What Do You Do? Too Polar for Reversed Phase What Do You Do? June 20, 2013 Mark Powell Columns and Consumables Technical Support Page 1 C8 or C18 Doesn t Always Do the Job Typical reversed phase conditions involve water/buffer

More information

Peptide and protein analysis by capillary HPLC Optimization of chromatographic and instrument parameters. Application. Angelika Gratzfeld-Huesgen

Peptide and protein analysis by capillary HPLC Optimization of chromatographic and instrument parameters. Application. Angelika Gratzfeld-Huesgen Peptide and protein analysis by capillary HPLC Optimization of chromatographic and instrument parameters Application Angelika Gratzfeld-Huesgen Abstract This application has been verified using an Agilent

More information

HPLC Method Development with Eclipse Plus: Standard Practices and New Columns. Agilent Technologies

HPLC Method Development with Eclipse Plus: Standard Practices and New Columns. Agilent Technologies HPLC Method Development with Eclipse Plus: Standard Practices and New Columns Agilent Technologies What are Some Standard Method Development Practices?. Follow preferred method development scheme and do

More information

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography Application Note Inorganic Ions, Water Testing, Minerals, Metals, Basic Chemicals Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography Authors Anne Mack, Adam Bivens

More information

(HILIC) Bill Champion Agilent Technologies, Inc opt 3/opt3/opt 2 HILIC - Agilent Restricted

(HILIC) Bill Champion Agilent Technologies, Inc opt 3/opt3/opt 2 HILIC - Agilent Restricted Hydrophilic Interaction Chromatography (HILIC) Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com Page 1 HILIC A method of recent attention Bill Champion

More information

How proteins separate on reverse-phase HPLC

How proteins separate on reverse-phase HPLC 1 Reverse Phase How proteins separate on reverse-phase HPLC RP chromatography separates proteins through the interaction of the hydrophobic foot of the protein with a nonpolar surface of the particle RP

More information

Taking Full Advantage of UHPLC with Agilent 6460A Triple Quadrupole MS. Outline

Taking Full Advantage of UHPLC with Agilent 6460A Triple Quadrupole MS. Outline Taking Full Advantage of UHPLC with Agilent 0A Triple Quadrupole MS Dr. Anabel Fandino R&D Application Scientist LC/MS Division Agilent Technologies Santa Clara, CA utline Ultra High Definition LC/MS using

More information

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Application Note Doping Control Authors Prof. Maria Kristina Parr Freie Universität Berlin Institute

More information

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China Rapid Deteration of Eight Related Aromatic Acids in the p-phthalic Acid Mother Liquid Using an Agilent 126 Infinity LC System and an Agilent Poroshell 12 SB-C18 Column Application Note Author Xinlei Yang

More information

The Agilent 1290 Infinity II Multi Method Solution

The Agilent 1290 Infinity II Multi Method Solution The Agilent 129 Infinity II Multi Method Solution Switch between seven applications on a single LC system by using automatic column and solvent selection Application ote Food Testing & Agriculture Author

More information

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection APPLICATION NOTE 72686 Determination of underivatized aflatoxins,, G2, and in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection Authors Sylvia Grosse, Mauro De Pra, Frank

More information

An Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining

An Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining An Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining Dave Weil, Ph.D. and Jim Lau, Ph.D. Typical Method Conditions: 1260 UHPLC

More information

Routine MS Detection for USP Chromatographic Methods

Routine MS Detection for USP Chromatographic Methods Daniel S. Root, Thomas E. Wheat, and Patricia McConville Waters Corporation, Milford, MA, USA APPLICATION BENEFITS This approach enables mass spectral analysis of peaks directly from unmodified USP HPLC

More information

HPLC Praktikum Skript

HPLC Praktikum Skript HPLC Praktikum Skript Assistants: Gianluca Bartolomeo HCI D330, 3 46 68, bartolomeo@org.chem.ethz.ch Sahar Ghiasikhou HCI E330, 2 29 29, ghiasikhou@org.chem.ethz.ch 1. Introduction In chromatographic techniques,

More information

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany Analysis of Pesticides by Supercritical Fluid Chromatography/ Mass Spectrometry Optimizing the Agilent 126 Infi nity Analytical SFC System in Combination with the Agilent 646 Triple Quadrupole LC/MS Application

More information

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract. Separation of Explosives in EPA 833: Column Choices Optimize Speed, Resolution, and Solvent Use Application Environmental Authors John W. Henderson Jr. and William J. Long Agilent Technologies, Inc. 28

More information

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide What is the MassHunter METLIN Metabolite PCD/PCDL? 2 Where to find more information 2 Kit Content 3 Installation 5 Before you start 5 Install MassHunter

More information

Choosing Columns and Conditions for the Best Peak Shape

Choosing Columns and Conditions for the Best Peak Shape Choosing Columns and Conditions for the Best Peak Shape What is Good Peak Shape and Why is it Important? Good peak shape can be defined as a symmetrical or gaussian peak and poor peak shape can include

More information

Application Note. Agilent Application Solution Analysis of PAHs in soil according to EPA 8310 method with UV and fluorescence detection.

Application Note. Agilent Application Solution Analysis of PAHs in soil according to EPA 8310 method with UV and fluorescence detection. Agilent Application Solution Analysis of PAHs in soil according to EPA 3 method with UV and fluorescence detection Application Note Environmental Authors Sonja Volk, Angelika Gratzfeld-Huesgen Agilent

More information

High Speed vs. High Resolution Analyses in HPLC: A Critical Performance Comparison of Column Options Using Poppe and Kinetic Plots

High Speed vs. High Resolution Analyses in HPLC: A Critical Performance Comparison of Column Options Using Poppe and Kinetic Plots High Speed vs. High Resolution Analyses in HPLC: A Critical Performance Comparison of Column Options Using Poppe and Kinetic Plots Jason C. Link, William E. Barber, William Long, Anne Mack Agilent Technologies

More information

Comprehensive 2D-LC Analysis of Chinese Herbal Medicine

Comprehensive 2D-LC Analysis of Chinese Herbal Medicine Comprehensive 2D-LC Analysis of Chinese Herbal Medicine The Agilent 1290 Infinity 2D-LC Solution Application Note Small Molecule Pharmaceuticals & Generics Author Sonja Krieger Agilent Technologies, Inc.

More information

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX Technical Note Agilent s SampliQ WAX provides Applications for strongly acidic, acidic and neutral compounds Excellent

More information

HPLC Separation Robustness and Ruggedness Assessing the Effects of Experimental Variables During Method Development

HPLC Separation Robustness and Ruggedness Assessing the Effects of Experimental Variables During Method Development HPLC Separation Robustness and Ruggedness Assessing the Effects of Experimental Variables During Method Development Rita Steed Application Engineer Agilent Technologies, Inc. November 19, 2015 Method Development

More information

HPLC Method Development: Standard Practices and New Columns. Agilent Technologies February 2007

HPLC Method Development: Standard Practices and New Columns. Agilent Technologies February 2007 HPLC Method Development: Standard Practices and New Columns Agilent Technologies February 7 What are Some Standard Method Development Practices?. Follow preferred method development scheme and do hands-on

More information

Eliminating Unwanted Variability in LC Methods: More Uptime, More Sample Throughput

Eliminating Unwanted Variability in LC Methods: More Uptime, More Sample Throughput Eliminating Unwanted Variability in LC Methods: More Uptime, More Sample Throughput HPLC Separation Robustness and Ruggedness Group/Presentation Title June 20, Agilent 2017Month Restricted ##, 200X Page

More information

Agilent 1200 Series Rapid Resolution LC and Rapid Resolution LC/MS Optimization Guide

Agilent 1200 Series Rapid Resolution LC and Rapid Resolution LC/MS Optimization Guide Agilent 1200 Series Rapid Resolution LC and Rapid Resolution LC/MS Optimization Guide Authors: Angelika Gratzfeld-Huesgen, Michael Frank, Christian Gotenfels Table of Contents Introduction 5 1. Theory

More information

Application. Authors. Abstract. Introduction. Hydrocarbon Processing

Application. Authors. Abstract. Introduction. Hydrocarbon Processing Fast Analysis Method for Rubber Chemical Antidegradants Using Rapid Resolution Liquid Chromatography (RRLC) Systems with Rapid Resolution High Throughput (RRHT) Columns Application Hydrocarbon Processing

More information

On the Advantages and Costs of Two-dimensional HPLC (2D-LC)

On the Advantages and Costs of Two-dimensional HPLC (2D-LC) On the Advantages and Costs of Two-dimensional HPLC (2D-LC) Georges Guiochon, Krisztìan Horvàth, Jacob Fairchild CoSMoS, Boston, 08/3-5/2009 1 What is Chromatography About? Separating Components of Mixtures

More information

Quantification of Furocoumarins

Quantification of Furocoumarins Version 8 (12/04/2013) Quantification of Furocoumarins HPLC/DAD Procedure Scope The European Cosmetics Directive 76/768/EEC, which is about to be adopted in the European Cosmetic Regulation, in Annex II

More information

Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions

Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions Today, ultra-high-performance liquid chromatography (UHPLC) has taken a firm foothold in the analytical laboratory.

More information

HPLC Column Troubleshooting: Is It Really The Column? LC Columns and Consumables. Robyn C. Woolley Online Application Engineer July 24, 2007.

HPLC Column Troubleshooting: Is It Really The Column? LC Columns and Consumables. Robyn C. Woolley Online Application Engineer July 24, 2007. HPLC Column Troubleshooting: Is It Really The Column? LC Columns and Consumables Robyn C. Woolley Online Application Engineer July 24, 2007 Page 1 Troubleshooting in HPLC ma U 20 00 15 00 10 005 0 0 0

More information

Qualification Specifications for Agilent 1100/1200 HPLC Systems

Qualification Specifications for Agilent 1100/1200 HPLC Systems Agilent 1100/1200 HPLC Systems 35 Vantage Point Drive, Rochester NY 14624 Phone: 800-828-1470 Fax: 800-395-0543 Transcat.com Test Set Points/ Range Acceptance Criteria Flow Rate Accuracy and Precision

More information

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction Application Note Consumer Products Authors Devon C. Zimmerman, Henry F. Rossi III, Jacqueline Rizzo, Daniel W. Keating,

More information

Impact of Instrument Dispersion on Performance of HPLC Capillary Columns

Impact of Instrument Dispersion on Performance of HPLC Capillary Columns Impact of Instrument Dispersion on Performance of HPLC Capillary Columns Wendy Roe, Richard A. Henry, and Hillel Brandes Supelco, Div. of Sigma-Aldrich Bellefonte, PA 16823 USA T413131 sigma-aldrich.com/analytical

More information

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-exyl Columns to Separate Estrogens Application Note Pharmaceutical, Environmental Authors John W. enderson Jr. and William J. Long Agilent

More information

for Acclaim Mixed-Mode HILIC-1 Column

for Acclaim Mixed-Mode HILIC-1 Column for Acclaim Mixed-Mode HILIC-1 Column Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 1 of 17 Product Manual for ACCLAIM Mixed-Mode HILIC-1 Column 5µm, 4.6 x 250mm, P/N 066844 5µm, 4.6 x 150mm, P/N

More information

ION CHROMATOGRAPHY SYSTEM S 150

ION CHROMATOGRAPHY SYSTEM S 150 ION CHROMATOGRAPHY SYSTEM S 150 WATER ANALYSIS ENVIRONMENTAL ANALYSIS ANION & CATION ANALYSIS ION CHROMATOGRAPHY IIon Chromatography is an analytical separation technique based on ionic interactions. Dissolved

More information

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS Application Note Drug Discovery Authors Srividya Kailasam Agilent

More information

Application Note. Edgar Naegele. Abstract

Application Note. Edgar Naegele. Abstract Agilent MassHunter Fast computer aided analysis of LC/ESI-TF data from complex natural product extracts Part 1: Analysis of Agilent 6210 TF data with the Molecular Feature Extractor in MassHunter Workstation

More information

Barry E. Boyes, Ph.D. Consumables and Accessories Business Unit May 10, 2000

Barry E. Boyes, Ph.D. Consumables and Accessories Business Unit May 10, 2000 Barry E. Boyes, Ph.D. Consumables and Accessories Business Unit May 0, 000 HPLC Column Troubleshooting What Every HPLC User Should Know :00 a.m. EST Telephone Number: 86-650-06 Chair Person: Tim Spaeder

More information

Dilution(*) Chromatography

Dilution(*) Chromatography WA20264 Poster # 184, HPLC 2002, Montreal, 4-5 June 2002 At-Column Column-Dilution for Preparative Dilution(*) Chromatography Cecilia Mazza, Jie Cavanaugh, Ziling Lu,Tom Sirard,Tom Wheat and Uwe Neue Waters

More information

Comprehensive Polymer Analysis Strategies

Comprehensive Polymer Analysis Strategies Comprehensive Polymer Analysis Strategies Mike Woodman, LC Applications Scientist 10 February, 2016 1 Review of GPC/SEC Polymer Analysis Portfolio The Agilent GPC/SEC portfolio is a complete GPC solution

More information

Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software

Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software Application ote Drug discovery and development: Metabolite Identifi cation Authors Yuqin Dai,

More information

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note Authors Siji Joseph Agilent Technologies India Pvt. Ltd.

More information

William E. Barber, Ph.D. Applications Chemist October

William E. Barber, Ph.D. Applications Chemist October William E. Barber, Ph.D. Applications Chemist ctober 0 00 Secrets of Good Peak Shape in HPLC Time: :00 p.m. CET Telephone Number: + 44 0 76 088 Chairperson: John Vis The Secrets of Good Peak Shape in HPLC

More information

Validated HPLC Methods

Validated HPLC Methods Validated HPLC Methods To Tweak or Not to Tweak Slide 4 Adjustments of operating conditions to meet system suitability requirements may be necessary. USP 23 p1776 Slide 5 001816S1.PPT What is the line

More information

Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX)

Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX) Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX) Application Note Forensic Toxicology Authors Bellah. Pule,

More information

Acclaim Mixed-Mode WCX-1

Acclaim Mixed-Mode WCX-1 Acclaim Mixed-Mode WCX-1 Product Manual for the Acclaim Mixed-Mode WCX-1 Column Page 1 of 28 PRODUCT MANUAL for the Acclaim Mixed-Mode WCX-1 Columns 4.6 x 150 mm, P/N (068353) 4.6 x 250 mm, P/N (068352)

More information

Tips and Tricks of Faster LC Analysis Without Capital Investment It may be easier than you think!

Tips and Tricks of Faster LC Analysis Without Capital Investment It may be easier than you think! Tips and Tricks of Faster LC Analysis Without Capital Investment It may be easier than you think! Agilent Technologies, Inc. Fall 28 Page 1 There Are Many Reasons to Reduce Analysis Time Run More Samples/Day

More information

HPLC analysis of VFAs, Furfural and HMF

HPLC analysis of VFAs, Furfural and HMF HPLC analysis of VFAs, Furfural and HMF Introduction The following chromatographic analysis is developed in order to analyse volatile fatty acids and carbohydrate degradation products; furfural, hydroxymethylfurfural

More information

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION Andrew J. Aubin and Tanya L. Jenkins Waters Corporation, Milford, MA, USA INTRODUCTION Benzocaine (4-Aminobenzoic

More information

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India Analysis of fumonisin, FB1 and FB2 mycotoxins in corn food and feed samples using the Agilent 1120 Compact LC System coupled to the Agilent 6140 Single Quadrupole LC/MS System Application Note Food Safety

More information

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS) Liquid Chromatography Mass Spectrometry (LC/MS) Prelab Questions: Questions to be answered before doing the experiment. The answers are due at the beginning of each experiment without exception (the questions

More information

Overcome Top Challenges in Analytical Method Development and Validation

Overcome Top Challenges in Analytical Method Development and Validation Overcome Top Challenges in Analytical Method Development and Validation Q. Chan Li, Ph.D. Analytical Development US, Boehringer Ingelheim Pharmaceuticals, Inc. IVT Analytical Method Validation, San Diego,

More information

Improving HPLC Column Selection and System Performance

Improving HPLC Column Selection and System Performance Improving HPLC Column Selection and System Performance Richard A. Henry Penn State University T410170 sigma-aldrich.com LC Particle Innovation Leads the Way* Particle Size and Architecture (porosity) More

More information

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners APPLICATION NOTE 21734 Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners Authors Aaron Lamb, Joanne Jones, Thermo Fisher Scientific, Runcorn, UK Keywords Environment, food

More information