Comparison of Neutron Sensitive Scintillators for use with a Solid- State Optical Detector

Size: px
Start display at page:

Download "Comparison of Neutron Sensitive Scintillators for use with a Solid- State Optical Detector"

Transcription

1 Comparison of Neutron Sensitive Scintillators for use with a Solid- State Optical Detector Sharmistha Mukhopadhyay 1,Christopher Stapels 1, Eric B. Johnson 1, Eric C. Chapman 1, Paul S. Linsay 1, Thomas H. Prettyman 2, Michael R. Squillante 1, and James F. Christian 1 1 Radiation Monitoring Devices, Inc.; 44 Hunt Street; Watertown, MA Planetary Science Institute; 17 E. Fort Lowell, Suite 16, Tucson, AZ ABSTRACT The detection of illicit nuclear sources and SNM requires an ubiquitous network of sensors. While 3 He proportional counters are excellent neutron detectors, there is an insufficient global supply of 3 He to create the required number of detectors. Alternatives to 3 He must be efficient, insensitive to gamma radiation, easily manufactured, rugged, and inexpensive to enable the procurement of a large numbers of sensors. The use of a high sensitivity solid-state optical detector coupled to scintillation materials, loaded with a neutron absorber such as 6 Li or 1 B, can fulfill these design constraints. In this work, we compare the properties of neutron-sensitive scintillation materials utilizing Monte Carlo simulations and experiments. Cs 2 LiLaBr 6 :Ce is compared to commercially available boron-loaded plastic scintillators and 3 He tubes. The scintillators are compared for neutron detection efficiency, limitations on size, gamma-rejection ratio, neutron detection limits, manufacturing cost, and availability for mass-production. Keywords: Neutron Detector, CMOS SSPM, Scintillation Materials, Detection Efficiency. 1. INTRODUCTION We are building a compact neutron detector with new scintillation material coupled to a solid-state photo multiplier tube (SSPM), which will provide an alternative to 3 He tubes and PMT-based detectors. 3 He proportional counters are capable of detecting neutrons but there is an insufficient global supply to create the large number of detectors required for the application. Existing detectors like boron loaded plastic scintillators efficiently detect neutrons; however, the electronequivalent energy of the 1 B(n,α) reaction is sufficiently low (93 kev) that gamma rays can obscure the neutron signature. The reaction peak for the new scintillators is at high energy, enabling higher precision for neutron measurements than any previous loaded scintillator. The neutron detector is based on a scintillators coupled to the solid-state photo multipliers (SSPM). In this work, we investigated the use Cs 2 LiLaBr 6 (CLLB), an emerging scintillation material [1], which is capable of detecting both neutron and gamma radiation. Similar measurements were made with commercially available boron loaded plastic scintillator (BLP). We will discuss some of the measurements and simulations that we performed for selecting the material and size for the prototype detector. We characterized the neutron source used for our measurements to calibrate the scintillator segments. From the experiments and simulations, we can select CLLB crystals for building our first prototype neutron detector to achieve a performance comparable to that of a helium tube. The CMOS SSPM can be mass-produced with commercially available CMOS process. Reliability, reproducibility and performance are enhanced for SSPM devices fabricated using CMOS technology. The signal processing can be done on the chip because CMOS enables the monolithic integration of the SSPM with other read out elements. This advantage further reduces cost and simplifies the detector system. The scintillation light produced as a result of the capture of a thermal neutron in CLLB corresponds to a light flash from the full-energy absorption of a 3.2-MeV gamma ray. This allows for a simple discrimination method using the pulse Hard X-Ray, Gamma-Ray, and Neutron Detector Physics XI, edited by Ralph B. James, Larry A. Franks, Arnold Burger, Proc. of SPIE Vol. 7449, 74491P 29 SPIE CCC code: X/9/$18 doi: / Proc. of SPIE Vol P-1

2 height to separate neutrons from the background gamma rays, which are prominent at energies below 3 MeV. Based on Monte Carlo calculations and experiments, CLLB can compete with 3 He in terms of counting efficiency and provides adequate gamma ray suppression for the target applications. Advantages of coupling CLLB to an SSPM for detecting neutrons include the capability of integrating signal-processing electronics with the SSPM as a mass-producible detector platform. 2. MATERIALS AND METHOD 2.1 Neutron Source The neutron source used for the measurements is an AmBe source with a neutron flux ~1 6 neutrons/ s. The source was sealed in a lead pig, which is shielded with polyethylene to thermalize the source neutrons. The source assembly was inserted in a portable plastic box shown in Figure 1. Back Side 4 PlasticBox Front Polyethylene sheets Side 3 Source Front Tungsten Figure 1. Block diagram of the source covered with polyethylene sheets on all three sides. There is a layer of tungsten sheet at the front of the source to absorb gamma radiations. We measured the source gamma rays with a NaI detector. Peaks identified in the spectra are the 511-keV annihilation peak, the 2.2 MeV from the thermal neutron capture with hydrogen, and the 4.4 MeV from the source. These gamma ray signatures are consistent with expected emission given in [2]. The spectra measured did not show any gamma ray peak around 3.2 MeV. 2.2 Simulations of the detector response The neutron and gamma ray output of the source was modeled using the Monte Carlo radiation transport code MCNPX [3]. MCNPX was used to calculate the neutron flux and dose as well as the gamma ray background needed for source design and characterization. The code was also used to model the response of detectors used for source characterization ( 3 He and NaI) and to calculate the neutron detection efficiency and detection limits for different CLLB-based designs in comparison to other solid state neutron detectors and 3 He. The intrinsic efficiency of a detector as a function of neutron energy and incident direction is defined as the ratio of the number of pulses recorded to the number of neutrons incident on a detector. The counting rate is given by the integral over all incident energies and directions of the product of the intrinsic efficiency, the projected area of the detector, and the flux of incident neutrons. 2.3 Experimental Characterization of Neutron Detectors BLP and CLLB: The neutron spectra were measured with BLP and CLLB coupled to the PMT using the moderated AmBe source. The block diagram of the experimental set up is shown in Figure 2. We first measured the neutron spectrum with a 3x3x6 mm 3 BLP crystal coupled to the PMT and then with a 3x3x3 mm 3 CLLB crystal coupled to a PMT. These measurements were done to compare the performance of the scintillation material in terms of efficiency and gamma ray rejection ratio. Side-by-side measurements were made with helium tube to compare the performance of the scintillator coupled to PMT with the standard helium tube (Reuter Stokes # P ). The helium tube has an active length of 2.2 cm and radius of.64 cm. Proc. of SPIE Vol P-2

3 M o d e r a to r S O U R C E D e te c to r Figure 2. Experimental set up used to measure scintillation materials coupled to a PMT. 3x3x3 mm 3 and 3x3x6 mm 3 CLLB: We measured the neutron source with 3x3x3 mm3 and 3x3x6 mm3 CLLB coupled to a PMT, and a helium tube as shown in Figure 3. The experimental set up is similar to Figure 2. The CLLB crystal is hygroscopic and therefore we have packaged the crystal in a glass cuvette as shown in Figure 3. The measurements were to see the effect of sensitivity on CLLB crystal size. Packaging Material Teflon Tape CLLB Helium Tube in glass cuvette CLLB Glass Cuvette Figure 3. On the left the photograph shows a 3x3x3mm3 CLLB and a helium tube kept side by side for neutron flux measurement. The right hand side is the diagram of the packaged CLLB crystal. The CLLB crystal is inserted in a glass cuvette of height 3 mm, internal width of 3mm and external width of 5.5 mm. All five sides of the packaged crystal is wrapped with Teflon tape. 2.4 Experimental Characterization of the SSPM SSPMs are fabricated from an array of Geiger photodiodes readout in parallel so that the output signal is proportional to the incident light intensity [4]. A Geiger photodiode is similar to an avalanche photodiode, but these diodes are operated above the reverse breakdown. A large gain (>16) is associated with each photon-initiated avalanche in the photodiode, providing good signal-noise ratio. These SSPMs are fabricated using a commercially available CMOS (complimentary metal oxide semiconductor) process, which provide cost-effective detector-on-a-chip solutions, high sensitivity measurements, and the very fast time response needed for neutron measurements [5]. A photograph of the CMOS SSPM is shown in Figure 4. Proc. of SPIE Vol P-3

4 Figure 4. Photograph of a CMOS SSPM detector. The size is approximately 6.6 mm x 6.6 mm. The fill factor is greater than 47%, and the number of 5μm x 5 μm pixels is In the SSPM neutron characterization measurements, the scintillator is coupled to the SSPM, as shown in Figure 5. The open face of the CLLB sample is 3x3 mm2, packaged in a glass cuvette. The active area of the SSPM is 3x3 mm2, but it has wire bond connections at one side that extend above the surface of the chip. A 3x3 mm2 wave-guide was created to couple the crystal without damaging the wire bonds. Figure 5. Photographs showing the CLLB on a PMT and SSPM prototype. 3. RESULTS 3.1 Scintillation Material Simulations Intrinsic efficiency: Intrinsic efficiency for neutrons has been simulated for selected scintillation materials are shown in Figure 6. The capture cross section for neutrons is inversely proportional to the neutron velocity [2]. This trend is reflected in Figure 6. The intrinsic efficiency for thermal neutrons (up to.4 ev) is greater than neutrons with higher energy. For Li, there is resonance at 1 MeV and hence the intrinsic efficiency increases at this energy. The counting rate that we expect to measure with the source a foot away from the neutron detector is also tabulated in Figure 6. The ratio of the reaction rate (events/source neutron/ cm3) in the helium tube and 3x3x3 mm 3 CLLB for the detectors placed a foot away from the source has been simulated to be 29. We measured the neutron spectra with the helium tube and the CLLB crystal a foot away from the source and the ratio of the measured count rate is 33. Thus, we can say that the measured data is in agreement with the simulated data with in 15%. This agreement with result and simulation confirms the detector characteristic simulated in Figure 6. Proc. of SPIE Vol P-4

5 Intrinsic Efficiency E-3 1E-4 1E-5 3 He (5mm) 3 CLLB En. BLP (5mm) 3 CLLB (3mm) 3 CLLB Energy(MeV) Detector Helium tube CLLB_En CLLB CLLB BLP Area (cm 2 ) Count Rate (Hz) Count Rate (Hz) Norm to Area Figure 6. Left hand side shows the intrinsic efficiency calculated for helium tube (black curve), 5x5x5 mm 3 CLLB crystal (red curve), 5x5x5 mm 3 95% enriched CLLB crystal (green curve), 3x3x3 mm 3 CLLB crystal (blue curve), 1x1x5 mm 3 BLP crystal (dark yellow curve) and lithium-loaded glass (magenta). CLLB En. stands for enriched CLLB in the figure. The table in the right hand side shows the expected count rate for helium tube of area 1.27 cm 2, CLLB crystals of area.25 and.9 cm 2, and 1 cm 2 of a BLP crystal. It can be seen that the intrinsic efficiency of the 5x5x5 mm 3 CLLB with 95% Li enriched is comparable to the helium tube. From the simulations, we can also see that the intrinsic efficiency for BLP is comparable to the helium tube. The neutron peak corresponds to 93-keV for BLP, and at this energy, the gamma-ray background is very high. The source quantification limit for BLP scintillator is higher than enriched CLLB as discussed next. Neutron Quantification Limits for Loaded Scintillators: CLLB crystals that we are examining for building our prototype detector are capable of detecting both neutron and gamma radiations. The neutron peak corresponds to 3.2 MeV in the pulse height spectra and at this region the gamma ray background is relatively small. For our AmBe source we have simulated the gamma background from the decay of 241 Am and the production of high energy gamma rays by the Be(α,n) reaction using MCNPX. The minimum neutron production rate needed to produce a measurable signal in the presence of a background is defined as the source quantification limit, which depends on the detector composition, distance from the source, shielding materials, and gamma ray background. Here, the quantification limit is arbitrarily taken to be the source neutron production rate that results in a relative precision in the net counting rate for the neutron peak of less than 3%: σ N < 3%, N Where N is the net counting rate and σ is the associated uncertainty (1 standard deviation). The calculated source N quantification limits for selected loaded is given in Table 1. The source quantification is given by Proc. of SPIE Vol P-5

6 S Q ε + bδp(1 + f ) = 2 2 ( σ / N ) ε t N where ε is the counts/source neutron, b is the background counts/kev, f = Δ p /Δ b, σ N / N is the quantification limit, and t is the counting time. For our calculation we have used f=2.simulated neutron spectra is shown in Figure 7. (Counts/keV) N Δ l Δ p Δ u Pulse height, E (kev) N=Total neutron counts over the background Δ l = Background peak region of interest on the left side of the neutron peak Δ p = Neutron peak region of interest. Δ u = Background peak region of interest on the right side of the neutron peak Δ b = Background peak region of interest on the right side of the neutron peak Figure 7. Example of typical pulse height spectra used to simulate the source quantification limit, S Q. Table 1. Shows quantification limits determined by Monte Carlo for the measurements of the RMD neutron source with 1x1x5 mm 3 scintillators. The counting time was 6s. Scintillation Material Peak Energy (MeV) Resolution (%FWHM at peak) Efficiency (counts/source neutron) (ε) Background (counts/mev/source neutron) (b) Peak region of interest (Δ p ) CLLB(Li 3.2 Enriched) 3% 6.22E E CLLB(Li 3.2 Natural) 3% 7.41E E BLP.93 25% 5.81E E Source Quantification limit (S Q ) (neutrons/sec ) From the simulation results we see that the CLLB with enriched Li provides the best performance of the materials evaluated, as the S Q is about 3 neutrons/sec. Due to very high equivalent energy for the reaction peak, CLLB is relatively insensitive to variations in gamma ray background. Proc. of SPIE Vol P-6

7 3.2 Scintillation Material Measurements BLP and CLLB: Figure 8 shows the pulse-height spectra collected by the BLP and the CLLB crystal for 72 sec. The neutron count rates measured for the BLP crystal is 1.77 Hz and that for the CLLB crystal is.3 Hz. The neutron peak is fitted to a Gaussian shape and the background counts are calculated with in an interval of 3-channel width. Taking the ratio of the neutron counts to the total counts we get the gamma rejection ratio of about 78% for BLP and 82% for CLLB. Counts 7 Neutron Peak in the 3x3x6 mm3 BLP crystal Counts 3 Neutron Peak in the 3x3x3 mm3 CLLB crystal Energy(keV) Energy(keV) 4 5 Figure 8. Neutron spectra measure with 3x3x6 mm 3 BLP crystal and a 3x3x3 mm 3 CLLB crystal. The neutron peaks are fitted in red. The spectra were taken for 72 sec in the presence of the moderated AmBe source. The spectrum measured with the helium tube is shown in Figure 9. The helium tube is sensitive to neutrons and from the spectra we measure the neutron flux on the helium tube to be 87 neutrons/s cm 2. The neutron flux on the BLP crystal measured from Figure 8 is 119 counts/s cm 2. Since BLP can detect both neutrons and gamma rays and the natural background as well as the source gamma ray background is high at energies below 5 kev, the 93-keV peak measured in the BLP spectra does not originate only from the neutron interaction with the boron. The higher count rate in the BLP crystal is due to the presence of gamma events with in the neutron peak. There is no distinction in the BLP crystal for the natural background and other source producing 93-keV gamma-equivalent peak. From the simulation shown in Figure 6, we expect similar neutron flux measurement by the BLP and the helium tube but our measurements show that the count rates measured in the BLP crystal is higher than helium tube count rate. This also shows the gamma events have been counted with the neutron events. To improve the gamma-ray rejection ratio, therefore, one has to use a coincident measurement with bismuth germinate (BGO), or some other dense scintillation material that will efficiently detect the gamma rays associated with the capture of a thermal neutron. Due to the geometric constraints, the intrinsic efficiency will generally decrease for a coincidence measurement and also this detector system will consist of two scintillation materials, which will make the cost of the detection system, go higher. The advantage of using the CLLB crystal is that the high equivalent gamma energy allows for easy pulse height discrimination (PHD). Using only PHD discrimination we have achieved the gamma discrimination ratio to be 82% for the CLLB and 78% for the BLP. Proc. of SPIE Vol P-7

8 Counts 5 Neutron Spectra in the helium tube Channel Figure 9. Neutron spectra measured with the helium tube. The dashed green curve is the gamma signature. 3x3x3 mm 3 and 3x3x6 mm 3 CLLB: Similar measurements were made with different volume CLLB crystals and with a helium tube. From the measured data and normalizing the total neutron counts to volume, we get the intrinsic efficiency of the 3x3x3 mm 3 CLLB cube to be 3% relative to the helium tube. The neutron count rate measured by the CLLB is.3 Hz. From this measurement, and assuming the incident neutron flux to be 11 n/cm 2 s, we estimate the sensitivity, which is the ratio of the neutrons detected by the detector to the total neutron flux from the source, to be.3. Simulations indicate that the intrinsic efficiency increases with the crystal size; therefore, the use of a larger crystal can improve the intrinsic efficiency. The quantities associated with binning the data for sensitivity characterization and pulse-amplitude discrimination are tabulated in Table 2. The results show that the sensitivity of the 3x3x6 mm 3 crystal is 4% more than the 3x3x3 mm 3 crystal. Table 2. Sensitivity calculated from measurements made with 3x3x3 mm 3 and 3x3x6 mm 3. Total counts represent the counts under the neutron peak. Bgnd. Counts or the background counts are gamma ray events under the neutron peaks. Crystals Total Counts Bgnd. Counts Sensitivity (S) 3x3x3 mm x3x6 mm SSPM Characterization Figure 1 shows the spectrum measured with a LYSO crystal coupled to the CMOS SSPM, irradiated with a 22 Na source. The spectrum is compared to that measured with the same LYSO crystal coupled to a PMT, and we see that the peak resolution for SSPM is comparable. Proc. of SPIE Vol P-8

9 Counts kev FW HM Device PMT 13% 9% SSPM 13% 8% 5.5 x 5.5 x 5.5 mm 3 LYSO Source: 22 Na Room Temperature 1274 kev Energy (kev) Figure 1. Plot of the pulse height spectra from a LYSO crystal coupled to the SSPM, exposed to 511-keV and 1275-keV gamma rays from 22 Na. The optical properties of the individual pixels have been measured. The detection efficiency (DE) is maximized around nm wavelengths. The emission spectra for the CLLB crystal show two peaks at 392 nm and 419 nm [1]. The DE for 419 nm at 2V excess bias is approximately 15%. We have tested our CLLB samples on SSPM segments. The measured spectra with PMT and SSPM are shown in Figure 11. We can see that the performance of SSPM is comparable to the PMT tube; however, the different optical coupling used with the SSPM affects the measurement. 4 SSPM PMT 4 Int Gamma Equivalent Energy (kev) Figure 11. Comparison of CLLB spectra on PMT and SSPM. The SSPM is a 3 mm x 3 mm device operated at 2 V excess bias and.5 ms shaping time. The PMT is operated at 6 V bias and 5 shaping time. 4. CONCLUSION We have made measurements with CLLB and BLP scintillators. From the measured spectra and simulation we can see the disadvantage of using the BLP is that the neutron peak corresponds to 93-keV where the gamma background is relatively high and since BLP can detect both neutron and gamma ray events it is not possible to distinguish between the using a single crystal. Thus, to improve the gamma rejection ratio we need two scintillation materials to set up coincidence measurements that will eliminate the gamma events. From the simulated efficiency plots we see that enriched 5x5x5 mm 3 CLLB crystal is good choice of scintillation material. Our aim is to make a compact, portable, hand held solid-state neutron detector whose performances is comparable to helium tubes. We will use only neutron signature to detect threatening events. Thus, we propose to make a detector with a scintillators coupled to an SSPM, capable of detecting neutron events. The AmBe source that has been used for measurements was characterized, which provided us information for calibrating the detector system. Measurements with 3x3x3 mm 3 CLLB show that the efficiency measured agrees well with the calculated results. For our prototype detector we would like to use a bigger crystal. We made measurement with 3x3x3 mm 3 and 3x3x6 mm 3 crystals. Our results show that we do gain in sensitivity by using bigger crystal. We conclude that CLLB is better than BLP and comparable to helium for detecting neutrons. Proc. of SPIE Vol P-9

10 ACKNOWLEDGMENTS This work has been supported in part by DNDO, Grant Number HSHQDC-8-C-169. REFERENCES [1] J.Glodo, E. V. D. v. Loef, and K. S. Shah, "Elpasolite Crystals for Gamma and Neutron Detection." IEEE, Dresden, Germany, October, 28. [2] G. F. Knoll, "Radiation Detection and Measurements," pp. 546, 2. [3] D. B. E. Pelowitz, "MCNPX User s Manual Version 2.5.," Los Alamos Natl. Lab. Doc. LA-CP-5 369, Los Alamos Natl. Lab., Los Alamos, N. M., 25. [4] P. Buzhan, B. Dolgoshein, L. Filatov, A. Ilyin, V. Kantzerov, V. Kaplin, A. Karakash, F. Kayumov, S. Klemin, E. Popova, and S. Smirnov, "Silicon photomultiplier and its possible applications," Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment Proceedings of the 3rd International Conference on New Developments in Photodetection, vol. 54, pp , 23. [5] C. J. Stapels, W. G. Lawrence, and J. F. Christian, "CMOS Solid-State Photomultiplier for Detecting Scintillation Light in Harsh Environments," SNIC Symposium, pp. 1-8, 26. Proc. of SPIE Vol P-1

V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber

V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber CYRIC Annual Report 2001 V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber Kawata N. Baba M. Kato M.*, Miura T.**, and Yamadera A.***, Cyclotron and Radioisotope Center, Tohoku

More information

EEE4106Z Radiation Interactions & Detection

EEE4106Z Radiation Interactions & Detection EEE4106Z Radiation Interactions & Detection 2. Radiation Detection Dr. Steve Peterson 5.14 RW James Department of Physics University of Cape Town steve.peterson@uct.ac.za May 06, 2015 EEE4106Z :: Radiation

More information

Unit 2. Instrumentation. Experts Teaching from Practical Experience

Unit 2. Instrumentation. Experts Teaching from Practical Experience Unit 2 Instrumentation Experts Teaching from Practical Experience Gas-Filled Detectors Gas-filled detectors measure the charge released when radiation interacts with the gas Three types: Ion Chambers,

More information

Introduction to Radiation Monitoring

Introduction to Radiation Monitoring Introduction to Radiation Monitoring Iain Darby Honorary Research Fellow, University of Glasgow iain.darby@glasgow.ac.uk https://at.linkedin.com/in/idarby https://www.facebook.com/iain.darby.662 Outline

More information

Introduction to neutron rem meters

Introduction to neutron rem meters LA-UR-15-28285 Introduction to neutron rem meters Tom McLean, LANL CSU neutron class Fort Collins, CO Oct. 27-29 2015 Introduction: talk outline Discussion (brief) of neutron remmeters Gas proportional

More information

PoS(TIPP2014)033. Upgrade of MEG Liquid Xenon Calorimeter. Ryu SAWADA. ICEPP, the University of Tokyo

PoS(TIPP2014)033. Upgrade of MEG Liquid Xenon Calorimeter. Ryu SAWADA. ICEPP, the University of Tokyo ICEPP, the University of Tokyo E-mail: sawada@icepp.s.u-tokyo.ac.jp The MEG experiment yielded the most stringent upper limit on the branching ratio of the flavorviolating muon decay µ + e + γ. A major

More information

Chapter 4 Scintillation Detectors

Chapter 4 Scintillation Detectors Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 4-1 4.1. Basic principle of the scintillator Chapter 4 Scintillation Detectors Scintillator Light sensor Ionizing radiation Light (visible,

More information

Alpha particle scintillation detector based on micro pixel avalanche photodiode and LYSO crystal

Alpha particle scintillation detector based on micro pixel avalanche photodiode and LYSO crystal Alpha particle scintillation detector based on micro pixel avalanche photodiode and LYSO crystal G.S. Ahmadov, F.I. Ahmadov Institute of Radiation Problems of ANAS, Baku, Azerbaijan C. Granja, S. Pospíšil

More information

Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector

Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector C. Fiorini, member, IEEE, A. Gola, M. Zanchi, A. Longoni, P. Lechner, H. Soltau, L. Strüder Abstract In this work

More information

The Best Gamma-Ray Detector

The Best Gamma-Ray Detector /sec The Best Gamma-Ray Detector What is the best detector for gamma-ray spectroscopy? Amptek, Inc. provides signal processing electronics which can be used with a variety of detectors (some available

More information

MCNP analysis and optimization of a double crystal phoswich detector

MCNP analysis and optimization of a double crystal phoswich detector EURONS I3 506065 JRA9 RHIB Report made during stay IEM-CSIC Madrid January 2008 MINISTERIO DE ASUNTOS EXTERIORES Y DE COOPERACIÓN AECI VICESECRETARÍA GENERAL MCNP analysis and optimization of a double

More information

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? 2. Briefly discuss dead time in a detector. What factors are important

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement Third Edition Glenn F. Knoll Professor of Nuclear Engineering and Radiological Sciences University of Michigan Ann Arbor, Michigan John Wiley & Sons, Inc. New York/Chichester/Weinheim/Brisbane/Toronto/Singapore

More information

Timing and Energy Response of Six Prototype Scintillators

Timing and Energy Response of Six Prototype Scintillators Timing and Energy Response of Six Prototype Scintillators CCM Kyba 1, J Glodo 2, EVD van Loef 2, JS Karp 1, KS Shah 2 1 University of Pennsylvania 2 Radiation Monitoring Devices SCINT 2007 June 7, 2007

More information

He-3 Neutron Detectors

He-3 Neutron Detectors Application He-3 Neutron Detectors General Considerations, Applications: He-3 filled proportional counters are standard neutron detectors and are most suitable for the detection of thermal neutrons. Larger

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement Fourth Edition Glenn F. Knoll Professor Emeritus of Nuclear Engineering and Radiological Sciences University of Michigan Ann Arbor, Michigan WILEY John Wiley & Sons,

More information

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies CHARACTERIZATION OF PHOSWICH WELL DETECTORS FOR RADIOXENON MONITORING Wolfgang Hennig 1, Hui Tan 1, William K. Warburton 1, Anthony Fallu-Labruyere 1, Konstantin Sabourov 1, Justin I. McIntyre 2, Matthew

More information

RADIATION DETECTION AND MEASUREMENT

RADIATION DETECTION AND MEASUREMENT RADIATION DETECTION AND MEASUREMENT SECOND EDITION GLENN F. KNOLL Professor of Nuclear Engineering The University of Michigan Ann Arbor, Michigan WILEY JOHN WILEY & SONS New York Chichester Brisbane Toronto

More information

Design and production of Scintillation Detectors

Design and production of Scintillation Detectors Design and production of Scintillation Detectors Components for nuclear radiation detectors Fields of application: * Medicine * Industry * Science * Security Scintillation Detectors : Instruments using

More information

PHYS 3446 Lecture #12

PHYS 3446 Lecture #12 PHYS 3446 Lecture #12 Wednesday, Oct. 18, 2006 Dr. 1. Particle Detection Ionization Detectors MWPC Scintillation Counters Time of Flight 1 Announcements Next LPCC Workshop Preparation work Each group to

More information

SCINTILLATION DETECTORS AND PM TUBES

SCINTILLATION DETECTORS AND PM TUBES SCINTILLATION DETECTORS AND PM TUBES General Characteristics Introduction Luminescence Light emission without heat generation Scintillation Luminescence by radiation Scintillation detector Radiation detector

More information

Gamma and X-Ray Detection

Gamma and X-Ray Detection Gamma and X-Ray Detection DETECTOR OVERVIEW The kinds of detectors commonly used can be categorized as: a. Gas-filled Detectors b. Scintillation Detectors c. Semiconductor Detectors The choice of a particular

More information

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation S. Moriuchi, M.Tsutsumi2 and K. Saito2 Nuclear safety technology Center,

More information

New perspectives in X-ray detection of concealed illicit materials brought by CdTe/CdZnTe spectrometric detectors

New perspectives in X-ray detection of concealed illicit materials brought by CdTe/CdZnTe spectrometric detectors New perspectives in X-ray detection of concealed illicit materials brought by CdTe/CdZnTe spectrometric detectors Jean-Marc Dinten, Jean-Louis Amans, Loïck Verger, Olivier Peyret CEA-LETI, MINATEC, Recherche

More information

Neutron Induced Nuclear Counter Effect in Hamamatsu Silicon APDs and PIN Diodes

Neutron Induced Nuclear Counter Effect in Hamamatsu Silicon APDs and PIN Diodes Neutron Induced Nuclear Counter Effect in Hamamatsu Silicon APDs and PIN Diodes Rihua Mao, Liyuan Zhang, Ren-yuan Zhu California Institute of Technology Introduction Because of its immunity to magnetic

More information

Prospects for achieving < 100 ps FWHM coincidence resolving time in time-of-flight PET

Prospects for achieving < 100 ps FWHM coincidence resolving time in time-of-flight PET Prospects for achieving < 100 ps FWHM coincidence resolving time in time-of-flight PET, 28-Feb-2012, ICTR-PHE, Geneva, Switzerland 1 Time-of-flight PET Colon cancer, left upper quadrant peritoneal node

More information

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN 316 Effective atomic number of composite materials by Compton scattering - nondestructive evaluation method Kiran K U a, Ravindraswami K b, Eshwarappa K M a and Somashekarappa H M c* a Government Science

More information

CALIBRATION OF SCINTILLATION DETECTORS USING A DT GENERATOR Jarrod D. Edwards, Sara A. Pozzi, and John T. Mihalczo

CALIBRATION OF SCINTILLATION DETECTORS USING A DT GENERATOR Jarrod D. Edwards, Sara A. Pozzi, and John T. Mihalczo CALIBRATION OF SCINTILLATION DETECTORS USING A DT GENERATOR Jarrod D. Edwards, Sara A. Pozzi, and John T. Mihalczo Oak Ridge National Laboratory Oak Ridge, TN 37831-6010 PO Box 2008 Ms6010 ABSTRACT The

More information

Development of a Hard X-Ray Polarimeter for Solar Flares and Gamma-Ray Bursts

Development of a Hard X-Ray Polarimeter for Solar Flares and Gamma-Ray Bursts Development of a Hard X-Ray Polarimeter for Solar Flares and Gamma-Ray Bursts M.L. McConnell, D.J. Forrest, J. Macri, M. McClish, M. Osgood, J.M. Ryan, W.T. Vestrand and C. Zanes Space Science Center University

More information

arxiv: v2 [physics.ins-det] 8 Feb 2013

arxiv: v2 [physics.ins-det] 8 Feb 2013 Preprint typeset in JINST style - HYPER VERSION arxiv:1302.0278v2 [physics.ins-det] 8 Feb 2013 Investigation of gamma ray detection performance of thin LFS scintillator with MAPD readout E.Guliyev a, F.Ahmadov

More information

Neutron Detection. n interactions with matter n detection High/low energy n detectors

Neutron Detection. n interactions with matter n detection High/low energy n detectors Neutron Detection Example of n detection: Well logging Reservoir/Formation Evaluation Brief introduction to neutron generation Continuous sources Large accelerators Pulsed neutron generators n interactions

More information

Fast Neutron and Gamma-Ray Detectors for the CSIRO Air Cargo Scanner

Fast Neutron and Gamma-Ray Detectors for the CSIRO Air Cargo Scanner Fast Neutron and Gamma-Ray Detectors for the CSIRO Air Cargo Scanner J.E. Eberhardt, A.J. McEwan, D. Milinkovic, V. Sharp, * and J.R. Tickner CSIRO Minerals, Private Mail Bag 5, Menai NSW 2234 Australia

More information

Scintillation Detectors

Scintillation Detectors Scintillation Detectors J.L. Tain Jose.Luis.Tain@ific.uv.es http://ific.uv.es/gamma/ Instituto de Física Corpuscular C.S.I.C - Univ. Valencia Scintillation detector: SCINTILLATION MATERIAL LIGHT-GUIDE

More information

DETECTORS. I. Charged Particle Detectors

DETECTORS. I. Charged Particle Detectors DETECTORS I. Charged Particle Detectors A. Scintillators B. Gas Detectors 1. Ionization Chambers 2. Proportional Counters 3. Avalanche detectors 4. Geiger-Muller counters 5. Spark detectors C. Solid State

More information

The Neutron Diagnostic Experiment for Alcator C-Mod

The Neutron Diagnostic Experiment for Alcator C-Mod PFC/JA-9-16 The Neutron Diagnostic Experiment for Alcator C-Mod C. L. Fiore, R. S. Granetz Plasma Fusion Center Massachusetts Institute of Technology -Cambridge, MA 2139 May, 199 To be published in Review

More information

Calibration of the GNU and HSREM neutron survey instruments

Calibration of the GNU and HSREM neutron survey instruments Calibration of the GNU and HSREM neutron survey instruments Neutron Users Club Meeting National Physical Laboratory 20 th October 2015 J. S. Eakins 1, L. G. Hager 1, J. W. Leake 2, R. S. Mason 2 and R.

More information

Radionuclide Imaging MII Detection of Nuclear Emission

Radionuclide Imaging MII Detection of Nuclear Emission Radionuclide Imaging MII 3073 Detection of Nuclear Emission Nuclear radiation detectors Detectors that are commonly used in nuclear medicine: 1. Gas-filled detectors 2. Scintillation detectors 3. Semiconductor

More information

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory Recent Activities on Neutron Standardization at the Electrotechnical Laboratory K. Kudo, N. Takeda, S. Koshikawa and A. Uritani Quantum Radiation Division, National Metrology Institute of Japan (NMIJ)

More information

Photon Instrumentation. First Mexican Particle Accelerator School Guanajuato Oct 6, 2011

Photon Instrumentation. First Mexican Particle Accelerator School Guanajuato Oct 6, 2011 Photon Instrumentation First Mexican Particle Accelerator School Guanajuato Oct 6, 2011 Outline The Electromagnetic Spectrum Photon Detection Interaction of Photons with Matter Photoelectric Effect Compton

More information

Energetic particles and their detection in situ (particle detectors) Part II. George Gloeckler

Energetic particles and their detection in situ (particle detectors) Part II. George Gloeckler Energetic particles and their detection in situ (particle detectors) Part II George Gloeckler University of Michigan, Ann Arbor, MI University of Maryland, College Park, MD Simple particle detectors Gas-filled

More information

DEVELOPMENT OF ADVANCED NEUTRON INDUCED PROMPT GAMMA- RAY ANALYSIS SYSTEM FOR SURVEY OF ANTI-PERSONNEL MINES

DEVELOPMENT OF ADVANCED NEUTRON INDUCED PROMPT GAMMA- RAY ANALYSIS SYSTEM FOR SURVEY OF ANTI-PERSONNEL MINES DEVELOPMENT OF ADVANCED NEUTRON INDUCED PROMPT GAMMA- RAY ANALYSIS SYSTEM FOR SURVEY OF ANTI-PERSONNEL MINES T. Iguchi 1, J. Kawarabayashi 1, K. Watanabe 1, K. Nishimura 2, T. Handa 2 and H. Sawamura 2

More information

Compton suppression spectrometry

Compton suppression spectrometry Compton suppression spectrometry In gamma ray spectrometry performed with High-purity Germanium detectors (HpGe), the detection of low intensity gamma ray lines is complicated by the presence of Compton

More information

STATUS OF ATLAS TILE CALORIMETER AND STUDY OF MUON INTERACTIONS. 1 Brief Description of the ATLAS Tile Calorimeter

STATUS OF ATLAS TILE CALORIMETER AND STUDY OF MUON INTERACTIONS. 1 Brief Description of the ATLAS Tile Calorimeter STATUS OF ATLAS TILE CALORIMETER AND STUDY OF MUON INTERACTIONS L. E. PRICE Bldg 362, Argonne National Laboratory, Argonne, IL 60439, USA E-mail: lprice@anl.gov (For the ATLAS Tile Calorimeter Collaboration)

More information

DESIGN OF NEUTRON DOSE RATE METER FOR RADIATION PROTECTION IN THE EQUIVALENT DOSE

DESIGN OF NEUTRON DOSE RATE METER FOR RADIATION PROTECTION IN THE EQUIVALENT DOSE DESIGN OF NEUTRON DOSE RATE METER FOR RADIATION PROTECTION IN THE EQUIVALENT DOSE Hiroo Sato 1 and Yoichi Sakuma 2 1 International University of Health and Welfare, Kitakanemaru 2600-1, Ohtawara 324-8501

More information

Characterizing, Simulating, and Eliminating Vibration Induced Counts in Measurement While Drilling Gamma Ray Detectors. authors

Characterizing, Simulating, and Eliminating Vibration Induced Counts in Measurement While Drilling Gamma Ray Detectors. authors Abstract: Characterizing, Simulating, and Eliminating Vibration Induced Counts in Measurement While Drilling Gamma Ray Detectors authors K. Kimmich and C. Grodsinsky 1 A Measurement While Drilling (MWD)

More information

Researchers at Texas A&M University have designed an integrated circuit

Researchers at Texas A&M University have designed an integrated circuit ORIGINAL ARTICLE A neutron detector using integrated circuits Shah MD 1, Marianno CM 1, Khatri SP 2, Fisher BD 1 Shah MD, Marianno CM, Khatri SP, et al. A neutron detector using integrated circuits. J

More information

Detector technology. Aim of this talk. Principle of a radiation detector. Interactions of gamma photons (gas) Gas-filled detectors: examples

Detector technology. Aim of this talk. Principle of a radiation detector. Interactions of gamma photons (gas) Gas-filled detectors: examples Aim of this tal Detector technology WMIC Educational Program Nuclear Imaging World Molecular Imaging Congress, Dublin, Ireland, Sep 5-8, 202 You can now the name of a bird in all the languages of the world,

More information

Recent Advances of Planar Silicon APD Technology

Recent Advances of Planar Silicon APD Technology Recent Advances of Planar Silicon APD Technology M. McClish 1, R. Farrell 1, R. Myers 1, F. Olschner 2, G. Entine 1, K.S. Shah 1 1 Radiation Monitoring Devices Inc., Watertown, MA 2 Cremat Inc., Newton,

More information

Determination of the boron content in polyethylene samples using the reactor Orphée

Determination of the boron content in polyethylene samples using the reactor Orphée Determination of the boron content in polyethylene samples using the reactor Orphée F. Gunsing, A. Menelle CEA Saclay, F-91191 Gif-sur-Yvette, France O. Aberle European Organization for Nuclear Research

More information

Charge collection in PET detectors

Charge collection in PET detectors University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2007 Charge collection in PET detectors Tony Young University of

More information

High-Resolution Gamma-Ray and Neutron Detectors For Nuclear Spectroscopy

High-Resolution Gamma-Ray and Neutron Detectors For Nuclear Spectroscopy High-Resolution Gamma-Ray and Neutron Detectors For Nuclear Spectroscopy Thomas Niedermayr, I. D. Hau, S. Terracol, T. Miyazaki, S. E. Labov and S. Friedrich Former colleagues: M. F. Cunningham, J. N.

More information

RESPONSE FUNCTION STUDY FOR ENERGY TO LIGHT CONVERSION IN ORGANIC LIQUID SCINTILLATORS

RESPONSE FUNCTION STUDY FOR ENERGY TO LIGHT CONVERSION IN ORGANIC LIQUID SCINTILLATORS RESPONSE FUNCTION STUDY FOR ENERGY TO LIGHT CONVERSION IN ORGANIC LIQUID SCINTILLATORS S. Prasad *, A. Enqvist, S. D. Clarke, S. A. Pozzi, E. W. Larsen 1 Department of Nuclear Engineering and Radiological

More information

Neutron Spectrometry Using a 7 Li Enriched CLYC Scintillation Detector

Neutron Spectrometry Using a 7 Li Enriched CLYC Scintillation Detector Neutron Spectrometry Using a 7 Li Enriched CLYC Scintillation Detector Alexander Luke Miller, Rachid Machrafi, Nafisah Khan Faculty of Energy Systems and Nuclear Engineering, University of Ontario Institute

More information

Modular Survey Spectrometer and Compton Imager

Modular Survey Spectrometer and Compton Imager 1 Modular Survey Spectrometer and Compton Imager Audrey MacLeod Ionizing Radiation Standards Measurement Science and Standards National Research Council Canada On behalf of the SCoTSS collaboration (Silicon

More information

A Monte Carlo Simulation for Estimating of the Flux in a Novel Neutron Activation System using 252 Cf Source

A Monte Carlo Simulation for Estimating of the Flux in a Novel Neutron Activation System using 252 Cf Source IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-4861.Volume 7, Issue 3 Ver. II (May. - Jun. 2015), PP 80-85 www.iosrjournals.org A Monte Carlo Simulation for Estimating of the Flux in a Novel Neutron

More information

Results from Silicon Photo-Multiplier neutron irradiation test

Results from Silicon Photo-Multiplier neutron irradiation test 1 Results from Silicon Photo-Multiplier neutron irradiation test R. Faccini Sapienza Università di Roma and INFN, Sezione di Roma, Italy D. Pinci INFN, Sezione di Roma, Italy E-mail: davide.pinci@roma1.infn.it

More information

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator 88 The 1 st NPRU Academic Conference SCI-O11 Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator P. Limkitjaroenporn and W.Chewpraditkul Radiation Physics Laboratory,

More information

Key words: avalanche photodiode, soft X-ray detector, scintillation γ-ray detector, imaging device PACS: 07.85;95.55.A;85.60.D

Key words: avalanche photodiode, soft X-ray detector, scintillation γ-ray detector, imaging device PACS: 07.85;95.55.A;85.60.D We have studied the performance of large area avalanche photodiodes (APDs) recently developed by Hamamatsu Photonics K.K, in high-resolution X-rays and γ- rays detections. We show that reach-through APD

More information

arxiv: v1 [physics.ins-det] 29 Jun 2011

arxiv: v1 [physics.ins-det] 29 Jun 2011 Investigation of Large LGB Detectors for Antineutrino Detection P. Nelson a,, N. S. Bowden b, a Department of Physics, Naval Postgraduate School, Monterey, CA 99, USA b Lawrence Livermore National Laboratory,

More information

Design of a Lanthanum Bromide Detector for TOF PET

Design of a Lanthanum Bromide Detector for TOF PET Design of a Lanthanum Bromide Detector for TOF PET A. Kuhn, S. Surti, Member, IEEE, J. S. Karp, Senior Member, IEEE, P. S. Raby, K. S. Shah, A. E. Perkins, Member, IEEE, G. Muehllehner, Fellow Member,

More information

PoS(ICHEP2016)474. SoLid: Search for Oscillations with a Lithium-6 Detector at the SCK CEN BR2 reactor

PoS(ICHEP2016)474. SoLid: Search for Oscillations with a Lithium-6 Detector at the SCK CEN BR2 reactor SoLid: Search for Oscillations with a Lithium-6 Detector at the SCK CEN BR2 reactor University of Bristol E-mail: dan.saunders@bristol.ac.uk The disappearance of reactor antineutrinos into a new neutral

More information

Radioactivity and Ionizing Radiation

Radioactivity and Ionizing Radiation Radioactivity and Ionizing Radiation QuarkNet summer workshop June 24-28, 2013 1 Recent History Most natural phenomena can be explained by a small number of simple rules. You can determine what these rules

More information

Hands on LUNA: Detector Simulations with Geant4

Hands on LUNA: Detector Simulations with Geant4 : Detector Simulations with Geant4 Gran Sasso Science Institute E-mail: axel.boeltzig@gssi.infn.it Andreas Best Laboratori Nazionali del Gran Sasso E-mail: andreas.best@lngs.infn.it For the evaluation

More information

Radiation Dose, Biology & Risk

Radiation Dose, Biology & Risk ENGG 167 MEDICAL IMAGING Lecture 2: Sept. 27 Radiation Dosimetry & Risk References: The Essential Physics of Medical Imaging, Bushberg et al, 2 nd ed. Radiation Detection and Measurement, Knoll, 2 nd Ed.

More information

Radioactivity. Lecture 6 Detectors and Instrumentation

Radioactivity. Lecture 6 Detectors and Instrumentation Radioactivity Lecture 6 Detectors and Instrumentation The human organs Neither humans nor animals have an organ for detecting radiation from radioactive decay! We can not hear it, smell it, feel it or

More information

Chemical Engineering 412

Chemical Engineering 412 Chemical Engineering 412 Introductory Nuclear Engineering Lecture 26 Radiation Detection & Measurement II Spiritual Thought 2 I would not hold the position in the Church I hold today had I not followed

More information

"Neutron Flux Distribution"

Neutron Flux Distribution TECHNICAL UNIVERSITY DRESDEN Institute of Power Engineering Training Reactor Reactor Training Course Experiment "Neutron Flux Distribution" Instruction for Experiment Neutron Flux Distribution Content:

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement June 2008 Tom Lewellen Tkldog@u.washington.edu Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron

More information

XENON Dark Matter Search. Juliette Alimena Columbia University REU August 2 nd 2007

XENON Dark Matter Search. Juliette Alimena Columbia University REU August 2 nd 2007 XENON Dark Matter Search Juliette Alimena Columbia University REU August 2 nd 2007 Evidence of Dark Matter Missing mass in Coma galaxy cluster (Fritz Zwicky) Flat rotation curves of spiral galaxies (Vera

More information

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY ABSORBED DOSE: The amount of energy absorbed, as a result of radiation passing through a material, per unit mass of material. Measured in rads (1 rad

More information

A POSITION SENSITIVE ALPHA PARTICLE DETECTOR BASED ON A LYSO CRYSTAL AND A MICRO-PIXEL AVALANCHE PHOTODIODE

A POSITION SENSITIVE ALPHA PARTICLE DETECTOR BASED ON A LYSO CRYSTAL AND A MICRO-PIXEL AVALANCHE PHOTODIODE A POSITION SENSITIVE ALPHA PARTICLE DETECTOR BASED ON A LYSO CRYSTAL AND A MICRO-PIXEL AVALANCHE PHOTODIODE Ahmadov G.S. a,b,c, Ahmadov F.I. b,c, Kopatch Yu.N. a, Telezhnikov S.A. a, Nuriyev S.M. a,b,

More information

A new scintillator detector for nuclear physics experiments: the CLYC scintillator

A new scintillator detector for nuclear physics experiments: the CLYC scintillator A new scintillator detector for nuclear physics experiments: the CLYC scintillator Franco Camera 1 and Agnese Giaz 2 1 Università di Milano and INFN sezione di Milano 2 INFN sezione di Milano (current

More information

Timing resolution measurement of a 3 Lanthanum Bromide detector

Timing resolution measurement of a 3 Lanthanum Bromide detector Timing resolution measurement of a 3 Lanthanum Bromide detector L. Galli, M. De Gerone, S. Dussoni, D. Nicolò, A.Papa, F. Tenchini, G. Signorelli INFN Pisa, INFN Genova, PSI Villigen 1 th Pisa Meeting

More information

INORGANIC crystal scintillators are widely used in high

INORGANIC crystal scintillators are widely used in high 29 IEEE Nuclear Science Symposium Conference Record N32-5 Gamma Ray Induced Radiation Damage in and LSO/LYSO Crystals Rihua Mao, Member, IEEE, Liyuan Zhang, Member, IEEE, and Ren-Yuan Zhu, Senior Member,

More information

HIGH-PRESSURE HELIUM-3 SCINTILLATION POSITION-SENSITIVE DETECTOR OF THERMAL NEUTRONS

HIGH-PRESSURE HELIUM-3 SCINTILLATION POSITION-SENSITIVE DETECTOR OF THERMAL NEUTRONS HIGH-PRESSURE HELIUM-3 SCINTILLATION POSITION-SENSITIVE DETECTOR OF THERMAL NEUTRONS A. Bolozdynya 1, A. Bolotnikov, J. Richards Constellation Technology Corporation, 7887 Bryan Dairy Rd., Largo, FL 33777

More information

A Comparison between Channel Selections in Heavy Ion Reactions

A Comparison between Channel Selections in Heavy Ion Reactions Brazilian Journal of Physics, vol. 39, no. 1, March, 2009 55 A Comparison between Channel Selections in Heavy Ion Reactions S. Mohammadi Physics Department, Payame Noor University, Mashad 91735, IRAN (Received

More information

Characterization and Monte Carlo simulations for a CLYC detector

Characterization and Monte Carlo simulations for a CLYC detector Characterization and Monte Carlo simulations for a CLYC detector A. Borella 1, E. Boogers 1, R.Rossa 1, P. Schillebeeckx 1 aborella@sckcen.be 1 SCK CEN, Belgian Nuclear Research Centre JRC-Geel, Joint

More information

Application of large-area avalanche photodiodes to energy-dispersive x-ray fluorescence analysis

Application of large-area avalanche photodiodes to energy-dispersive x-ray fluorescence analysis X-RAY SPECTROMETRY X-Ray Spectrom. 2; 3: 64 69 Application of large-area avalanche photodiodes to energy-dispersive x-ray fluorescence analysis L. M. P. Fernandes, J. A. M. Lopes, J. M. F. dos Santos and

More information

PET. Technical aspects

PET. Technical aspects PET Technical aspects 15 N 15 O Detector 1 β+ Detector 2 e- Evolution of PET Detectors CTI/Siemens 15 N 15 O Detector block 1 β+ Detector block 2 x e- x y y location line of response Constant fraction

More information

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors 5. Radiation Microsensors Radiation µ-sensors convert incident radiant signals into standard electrical out put signals. Radiant Signals Classification

More information

Development of a Dedicated Hard X-Ray Polarimeter Mark L. McConnell, James R. Ledoux, John R. Macri, and James M. Ryan

Development of a Dedicated Hard X-Ray Polarimeter Mark L. McConnell, James R. Ledoux, John R. Macri, and James M. Ryan Development of a Dedicated Hard X-Ray Polarimeter Mark L. McConnell, James R. Ledoux, John R. Macri, and James M. Ryan Space Science Center University of New Hampshire Durham, NH AAS-HEAD Mt. Tremblant,

More information

Factors Affecting Detector Performance Goals and Alternative Photo-detectors

Factors Affecting Detector Performance Goals and Alternative Photo-detectors XENON Experiment - SAGENAP Factors Affecting Detector Performance Goals and Alternative Photo-detectors Department of Physics Brown University Source at http://gaitskell.brown.edu Gaitskell Review WIMP

More information

Novel Plastic Microchannel-Based Direct Fast Neutron Detection

Novel Plastic Microchannel-Based Direct Fast Neutron Detection Novel Plastic Microchannel-Based Direct Fast Neutron Detection D. Beaulieu, P. de Rouffignac, D. Gorelikov, H. Klotzsch, J. Legere*, J. Ryan*, K. Saadatmand, K. Stenton, N. Sullivan, A. Tremsin Arradiance

More information

A Prototype of LaBr3:Ce in situ Gamma-Ray Spectrometer for Marine Environmental Monitoring

A Prototype of LaBr3:Ce in situ Gamma-Ray Spectrometer for Marine Environmental Monitoring A Prototype of LaBr3:Ce in situ Gamma-Ray Spectrometer for Marine Environmental Monitoring 1 Key Laboratory of Particle & Radiation Imaging (Tsinghua University), Ministry of Education Department of Engineering

More information

Nuclear Cross-Section Measurements at the Manuel Lujan Jr. Neutron Scattering Center

Nuclear Cross-Section Measurements at the Manuel Lujan Jr. Neutron Scattering Center 1 Nuclear Cross-Section Measurements at the Manuel Lujan Jr. Neutron Scattering Center M. Mocko 1, G. Muhrer 1, F. Tovesson 1, J. Ullmann 1 1 LANSCE, Los Alamos National Laboratory, Los Alamos NM 87545,

More information

Figure 1. Decay Scheme for 60Co

Figure 1. Decay Scheme for 60Co Department of Physics The University of Hong Kong PHYS3851 Atomic and Nuclear Physics PHYS3851- Laboratory Manual A. AIMS 1. To learn the coincidence technique to study the gamma decay of 60 Co by using

More information

28th Seismic Research Review: Ground-Based Nuclear Explosion Monitoring Technologies DESIGN OF A PHOSWICH WELL DETECTOR FOR RADIOXENON MONITORING

28th Seismic Research Review: Ground-Based Nuclear Explosion Monitoring Technologies DESIGN OF A PHOSWICH WELL DETECTOR FOR RADIOXENON MONITORING DESIGN OF A PHOSWICH WELL DETECTOR FOR RADIOXENON MONITORING W. Hennig 1, H. Tan 1, A. Fallu-Labruyere 1, W. K. Warburton 1, J. I. McIntyre 2, A. Gleyzer 3 XIA, LLC 1, Pacific Northwest National Laboratory

More information

A new neutron monitor for pulsed fields at high-energy accelerators

A new neutron monitor for pulsed fields at high-energy accelerators A new neutron monitor for pulsed fields at high-energy accelerators Marlies Luszik-Bhadra *, Eike Hohmann Physikalisch-Technische Bundesanstalt, Bundesallee 100, D-38116, Braunschweig, Germany. Abstract.

More information

Detectors for High Resolution Gamma-ray Imaging Based on a Single CsI(Tl) Scintillator Coupled to an Array of Silicon Drift Detectors

Detectors for High Resolution Gamma-ray Imaging Based on a Single CsI(Tl) Scintillator Coupled to an Array of Silicon Drift Detectors 1 Detectors for High Resolution Gamma-ray Imaging Based on a Single CsI(Tl) Scintillator Coupled to an Array of Silicon Drift Detectors C. Fiorini, A. Longoni, F. Perotti, C. Labanti, E. Rossi, P. Lechner,

More information

Scintillation detectors

Scintillation detectors 25 de dx Scintillation detectors excitation L25.pdf P627 YK 3/14/2012 detectable photons also by UV, or molecular collisions, chem. reactions, bubbles. etc. Detector building requirements (sometimes controversial):

More information

A dual scintillator - dual silicon photodiode detector module for intraoperative gamma\beta probe and portable anti-compton spectrometer

A dual scintillator - dual silicon photodiode detector module for intraoperative gamma\beta probe and portable anti-compton spectrometer University of Wollongong Research Online Faculty of Engineering - Papers (Archive) Faculty of Engineering and Information Sciences 2008 A dual scintillator - dual silicon photodiode detector module for

More information

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION OBJECTIVE The primary objective of this experiment is to use an NaI(Tl) detector, photomultiplier tube and multichannel analyzer software system

More information

Scintillation Detector

Scintillation Detector Scintillation Detector Introduction The detection of ionizing radiation by the scintillation light produced in certain materials is one of the oldest techniques on record. In Geiger and Marsden s famous

More information

The 46g BGO bolometer

The 46g BGO bolometer Nature, 3 The g BGO bolometer 1 Photograph of the heat [g BGO] and light [Ge; =5 mm] bolometers: see Fig. 1c for description Current events: Amplification gains: 8, (heat channel) &, (light channel). The

More information

Development of a Silicon PIN Diode X-Ray Detector

Development of a Silicon PIN Diode X-Ray Detector 1 E-mail: jkabramov@gmail.com \ Chonghan Liu E-mail: kent@smu.edu Tiankuan Liu E-mail: liu@smu.edu Jingbo Ye E-mail: yejb@smu.edu Xiandong Zhao E-mail: xiandongz@smu.edu X-ray detectors currently on the

More information

SiPM & Plastic Scintillator

SiPM & Plastic Scintillator SiPM & Plastic Scintillator Silicon photomultiplier coupled to plastic scintillator Lodovico Lappetito SiPM_PlasticScint_ENG - 28/04/2016 Pag. 1 Table of contents Introduction... 3 Plastic Scintillators...

More information

Studies of Hadron Calorimeter

Studies of Hadron Calorimeter Studies of Hadron Calorimeter Zhigang Wang Institute of High Energy Physics 2012.10.17 in IHEP Outline 1,The Dark Matter Calorimeter 2,The Hadron Calorimeter(HCAL) 3, Summary 1,Dark Matter Calorimeter

More information

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors Radiation Detectors 1 How do we detect ionizing radiation? Indirectly, by its effects as it traverses matter? What are these effects? Ionization and excitation of the atoms and molecules Heat 2 Types of

More information

THE COMPTON EFFECT Last Revised: January 5, 2007

THE COMPTON EFFECT Last Revised: January 5, 2007 B2-1 THE COMPTON EFFECT Last Revised: January 5, 2007 QUESTION TO BE INVESTIGATED: How does the energy of a scattered photon change after an interaction with an electron? INTRODUCTION: When a photon is

More information

Chapter 11: Neutrons detectors

Chapter 11: Neutrons detectors Chapter 11: Neutrons detectors 1 Contents Principles of neutrons detection Slow neutron detection methods Fast neutron detection methods 2 Introduction Neutrons are uncharged particles cannot be directly

More information