Lasentec Product Group

Size: px
Start display at page:

Download "Lasentec Product Group"

Transcription

1 Document Number: Window Reference Procedure - M500/600 P Probes, version 6.0 and higher FBRM CI Software The Window Reference Procedure is the standard calibration procedure used to reference the Lasentec FBRM instrument focal point to the window surface. This procedure is for an M-series process P probe that is being used with version 6.0 and higher of the Lasentec Control Interface (CI) software. See the table below for the appropriate window reference procedure to use with each probe and software type: Probe Type Software Window Reference Procedure M-series lab L probe Version 6.0 and higher M-series process P probe Version 6.0 and higher Instruction Sheet Instruction Sheet (this document) M-series process P probe Version 5.4 Instruction Sheet a M-series lab L probe Version 5.4 Instruction Sheet b This procedure is required to calculate the Lasentec FBRM focal point position, which is necessary to ensure measurement-to-measurement repeatability. The measured statistic #/Sec, No Wt 0-5 (Statistic #5) will reach a maximum value (also called the peak value) when the focal point is located at the window reference position. The Window Reference Procedure should be performed: If the PVC Reference Procedure indicates the Lasentec FBRM instrument is out of calibration For non-standard probe orientations (see Orienting the Probe below) Orienting the Probe At the factory, the probe is calibrated in a vertical position with the window pointing down, as it is in the fixed beaker stand. The vertical (window down) window reference position must be known before the Window Reference Procedure can be run. Important! If the orientation of the process installation is greater than ±30 from a vertical (window down) position, this Window Reference Procedure should be performed with the Lasentec FBRM probe in the same orientation as in the process installation (i.e., horizontal, vertical, or at a given angle) Rev F Page 1 of 11

2 Reading the Micrometer The micrometer, located towards the back of the Lasentec probe body on an M-series process probe, is used to adjust the focal point position. As shown in Figure 1, the scale on the micrometer spindle provides the relative position of the focal point in millimeters to the nearest 0.5 mm. The scale on the rotating dial provides the relative position of the focal point in millimeters to the nearest 0.01 mm (10 µm). The combination of the micrometer spindle and rotating dial readings indicate the focal point position. MICROMETER SPINDLE ROTATING DIAL EACH HASH MARK REPRESENTS 0.01mm (10µm) LINEAR SCALE EACH HASH MARK REPRESENTS 0.5mm MICROMETER POSITION = mm ( mm mm ) CLOCKWISE = LOWER NUMBER Figure 1 Reading the Micrometer Turning the rotating dial clockwise causes the micrometer reading to decrease. As the rotating dial is turned clockwise, it moves to the left (or down if the probe is in a vertical position), hiding more and more of the incremental hash marks on the micrometer spindle until a 0.00 micrometer position (all hash marks covered) is achieved. Establish the correct perspective of the micrometer before making adjustments. All directions (clockwise or counterclockwise) are written as if the Lasentec probe were in a vertical position (e.g., in a fixed beaker stand), with the user looking down at the micrometer in the direction of the probe tip. The table below shows the distance moved in both millimeters (mm) and microns (µm) as related to the hash marks on the rotating dial and its number of rotations. This information will be important as you make the coarse, mid-range, and fine adjustments in the Locating the Window Reference Position step below Rev F Page 2 of 11

3 Distance (mm) Distance (µm) Increments (hash marks on the Rotating Dial) Movement of the Rotating Dial full turns full turn /5 of a turn /50 of a turn ½ 1/100 of a turn Coating the Window Surface Using the Lasentec Marker Before locating the window reference position, coat the Lasentec FBRM probe window with erasable ink. To coat the window surface: 1. Using the Lasentec Dry Erase Marker provided with the Lasentec hardware, coat the surface of the clean, dry Lasentec FBRM probe window with a uniform, thick layer of ink. While most black, dry-erasable markers will do, it is best to use the marker provided by Mettler-Toledo AutoChem to ensure measurement-to-measurement conformity. 2. Allow two minutes for the ink to dry. Configuring version 6.0 and higher of the FBRM CI Software to Locate the Window Reference Position Before locating the window reference position, verify that the Acquisition Module of version 6.0 and higher of the FBRM CI software is running and configured per the directions below. With the CI software, the micrometer position can be adjusted while the Lasentec instrument output is monitored. To configure the software to locate the window reference position: 1. From the Setup Menu, select Meas. Config. 2. For Measurement Duration, select 5 sec (or 10 sec if using a low-speed optical line driver). 3. For Time Averaging, select No Time Averaging. 4. For Channel Grouping, select µm 200 Linear Ch. 5. Click the Open File for Save button. 6. Open file C:\Fbrm6?\Backup\SN-X-mmddyy-WRP 1 and click Save. 1 SN = the instrument s serial number X = electronics type (F or C) mmddyy = current date Rev F Page 3 of 11

4 7. For When Saving to File, save every v Measurements, set it to save every 1 measurement. 8. Press the NOT Measuring: Press to Start button. 9. From the Setup Menu, select Stat. Config. 10. Press the Load Stats. Config button. 11. From the C:\Fbrm6?\Statistics Setup Files folder, select the default0-100-lin.sta statistics setup (the measured statistic #/sec, No Wt 0-5 will reach a maximum value when the focal point is located at the window reference position) and click Open. 12. From the Display Menu, select Stats. 13. In the Show column, make sure there is an X next to the #/sec, No Wt 0-5 statistic (Statistic #5). This will create a trend plot in the trend display once you begin measuring data. 14. Deselect all other statistics by clicking on any existing Xs in the Show column. Deselecting a statistic will remove the trend from the trend display. 15. Under X-axis Scale (located below the trend display), select Record. 16. Under Buffer Length, type in To autoscale the X- and Y-axes, push both toggle switches (under the Trend Display graph) to the right. 18. From the Setup Menu, select Annotation. 19. Make sure the box next to Save with All Records contains an X. 20. Click the Y-axis applies to: box (located below the trend display) and select the #5 statistic. 21. Press the Measuring: Press to Stop button. WRP = window reference procedure (A file name of 550-F WRP means the file is the calibration procedure for the unit with serial number 550, using F electronics, taken on January 1, 2001.) Rev F Page 4 of 11

5 Locating the Window Reference Position Locating the window reference position takes less than 25 minutes. It is important that this entire process be carried out continuously without interruption. If interrupted, begin again from step 1 of the coarse adjustment below. Important! Establish the correct perspective of the micrometer before making adjustments. All directions (clockwise or counterclockwise) are written as if the Lasentec FBRM probe were in a vertical position (e.g., in a fixed beaker stand), with the user looking down at the micrometer in the direction of the probe tip. To eliminate the effects of hysteresis, you must always reach the micrometer position on an M-series process probe turning counterclockwise. If you pass the desired micrometer position, move the rotating dial at least 500 µm (one full turn) clockwise and then find the correct point moving counterclockwise. Coarse Adjustment The micrometer position is moved in 0.5 mm increments (one full turn of the rotating dial) to pass the peak value for Statistic #5. Several coarse adjustments may be required before the peak value is passed. To perform coarse adjustment of the micrometer: 1. Grasp the rotating dial on the micrometer and turn it clockwise until the 0 on the rotating dial is lined up with the vertical line on the micrometer spindle and the left (or bottom) edge of the rotating dial is lined up with the 0 line on the micrometer spindle. This is referred to as zeroing the micrometer. Depending on where the micrometer was set, you may need to make several full turns of the rotating dial before you reach the zero position. 2. In the Acquisition Module of the FBRM CI software, make sure Annotation is selected from the Setup Menu, then enter the current micrometer position in the Annotation field (i.e., micrometer = 0.00). 3. From the Display Menu, make sure Stats is selected and an X is displayed in the Show box next to Statistic #5. (You will need to be able to see the Current Value box for Statistic #5 when you begin adjusting the micrometer.) 4. Press the NOT Saving, Press to AutoSave button to begin saving to the data file. 5. Press the NOT Measuring, Press to Start button to begin measuring. 6. Wait until the # Meas Since Start box displays 5 (i.e., 5 measurements have been taken), then click the Saving Data: Press to Stop button to stop saving and the Measuring: Press to Stop button to terminate the measurement. 7. Look at the Current Value displayed for Statistic #5. If the micrometer position is still at 0.00 (i.e., you have not yet made any coarse adjustments), the value displayed may still be After a Rev F Page 5 of 11

6 few coarse adjustments this value will begin to increase and data will display in the trend display graph. 8. Slowly turn the rotating dial one full turn to the left (50 hash marks). 9. Type this micrometer position (+0.5 mm greater than the last recorded measurement) in the Annotation field (e.g., micrometer = +0.5). 10. Repeat steps 4 through 9 (coarse adjustment) while watching the trend display graph. It may take several coarse adjustments before data will display in this graph and counts will begin to register in the Current Value box. Continue to watch the trend display graph, performing another coarse adjustment after every five measurements. Once counts begin to register you ll notice that they fluctuate up and down by small increments with each measurement and increase dramatically with each adjustment of the micrometer position. If the trend line disappears off the top of the graph at any point, click the Scale All Trends box to reset the Y-axis range. Once the peak value is reached, counts displayed in the Current Value box will begin to decrease and a definite peak will be seen in the trend display graph. Once you see the peak, perform three additional coarse adjustments (1.5 mm) to verify there is no higher peak, then proceed to step To determine the final micrometer position of the coarse adjustment, minimize the Acquisition Module and open the Data Review Module. 12. From the Setup Menu, select Meas Config. 13. Click the Open File button. 14. Select the data file created while the window reference position was being located (C:\Fbrm6?\Backup\SN-X-mmddyy-WRP) and click Open. 15. For Time Averaging, select No Time Averaging. 16. Click the Apply Time Averaging box. 17. For Channel Grouping, select µm 200 Linear Ch. 18. From the Setup Menu, select Stat. Config. 19. Press the Load Stats. Config button. 20. From the C:\Fbrm6?\Statistics Setup Files folder, select the default0-100-lin.sta statistics setup (the measured statistic #/sec, No Wt 0-5 will reach a maximum value when the focal point is located at the window reference position) and click Open. 21. From the Display Menu, select Stats/Trend. 22. In the Show column, make sure there is an X next to the #/sec, No Wt statistic (Statistic #5). This will create a trend plot in the trend display. 23. Deselect all other statistics by clicking on any existing Xs in the Show column. Deselecting a statistic will remove the trend from the trend display. 24. From the Y Axis applies to: box, select the #5 statistic. 25. Under X-axis Scale (located below the trend display), select Record Rev F Page 6 of 11

7 26. Click the Scale All Trends box to reset the Y-axis. 27. Make sure the Lock Cursor box is not checked (i.e., an X is not displayed), select the + from the trend toolbar (located below the Trend Display), and move the Current Record crosshair on the trend display graph to the peak value measurement of the trended statistic as determined by the fine adjustment (right peak). 28. From the Setup Menu, select Annotation. 29. Look at the Annotation field for the micrometer reading at this maximum value. This is the peak micrometer position of the coarse adjustment (e.g., 2.5 mm). If you still have no counts once the micrometer position has reached 2.0 mm, or if you have not seen a dramatic decrease in value once the micrometer position has reached 5.5 mm, re-coat the probe window with the Lasentec Dry Erase Marker (make certain the window is completely covered with a thick layer of ink) and repeat the coarse adjustment from step 1. Mid-Range Adjustment Adjust or move the micrometer position in 0.02 mm increments (2 hash marks on the rotating dial) in the mid-range adjustment to refine the peak value for Statistic #5. You may make several mid-range adjustments before the peak value is refined. Remember that to eliminate the effects of hysteresis, you must always reach the micrometer position going counterclockwise. If you pass the desired micrometer position, move the micrometer clockwise at least 500 microns (one full turn) and then rotate counterclockwise to find the correct point. To perform mid-range adjustment of the micrometer: 1. With the software set to the Acquisition Module, moving the micrometer clockwise, turn the rotating dial back to the peak micrometer position established in the coarse adjustment (e.g., 2.5 mm), then rotate clockwise 0.5 mm (500 microns or one full turn), and then counterclockwise 0.4 mm (400 microns). 2. Turn the rotating dial two hash marks counterclockwise to increase the micrometer position by 0.02 mm. 3. Type over the previous entry to record this new micrometer position (+0.02 greater than the last recorded measurement) in the Annotation field (e.g., micrometer = 2.52 mm). 4. Press the NOT Saving, Press to AutoSave button to begin saving to the data file. 5. Press the NOT Measuring, Press to Start button to begin measuring. 6. Wait until the # Meas Since Start box displays 5 (i.e., 5 measurements have been taken), then click the Saving Data: Press to Stop button to stop saving and the Measuring: Press to Stop button to terminate the measurement Rev F Page 7 of 11

8 7. Repeat steps 2 through 6 at least three times. Look at the trend display graph. The counts should be increasing (i.e., the trend line going up). If they are not, move the rotating dial to the clockwise to a position 0.2 mm less than the micrometer position found in the coarse adjustment (e.g., 2.5 mm) and repeat the mid-range adjustment from step 2. If the trend line has disappeared off the top of the graph, click the Scale All Trends button to reset the Y-axis. Every time the trend display graph shows counts higher than the peak value from the coarse adjustment, this new value should be considered the new peak and steps 2 through 6 should be repeated three additional times. 8. Once the micrometer position has been successfully moved 0.03 mm (30 microns) beyond a distinct maximum value of Statistic #5, record this micrometer position in the Annotation field (e.g., 2.72 mm). This is considered the peak micrometer position of the mid-range adjustment. Fine Adjustment Increase the micrometer position in 0.005mm increments, one-half of a has mark on the rotating dial, to confirm the peak value for Statistic #5. Several fine adjustments may need to be made before confirming the peak value. Remember that to eliminate the effects of hysteresis, you must always reach the micrometer position going counterclockwise. If you pass the desired micrometer position, move the micrometer clockwise at least 500 microns (one full turn) and then find the correct point moving counterclockwise. To perform the fine adjustment of the micrometer: 1. Decrease the micrometer position by 0.5 mm from the peak mid-range adjustment micrometer position by turning the rotating dial one full turn (500 microns) clockwise, then increase the micrometer by 0.45 mm (450 microns) turning counterclockwise. 2. Type over the previous entry to record the new micrometer position (-0.08 less than the midrange peak value) in the Annotation field. 3. Press the NOT Saving, Press to AutoSave button to begin saving to the data file. 4. Press the NOT Measuring, Press to Start button to begin measuring. 5. Wait until the # Meas Since Start box displays 5 (i.e., 5 measurements have been taken), then click the Saving Data: Press to Stop button to stop saving and the Measuring: Press to Stop button to terminate the measurement. 6. Look at the trend display graph. If the trend line has disappeared off the top of the graph, click the Scale All Trends button to reset the Y-axis. 7. Increase the micrometer position by slowly turn the rotating dial counterclockwise one half of one hash mark. 8. Type over the previous entry to record the new micrometer position ( greater than the last recorded measurement) in the Annotation field (e.g., micrometer = mm). 9. Repeat steps 3 through 8 until the peak value established in the mid-range adjustment is passed by mm. You will see gradual stair-step increases before the line in the trend display graph gradually decreases. After the peak, the counts will likely fluctuate up and down again, but will not reach as high as they did at the peak Rev F Page 8 of 11

9 If counts do fluctuate so multiple peaks are observed, re-coat the window with the Dry Erase Marker and repeat the fine adjustment from step Once the peak value from the mid-range adjustment has been passed by mm, open the Data Review Module to determine the exact micrometer position of the maximum peak value using the trend display graph. 11. From the Setup Menu, select Meas Config. 12. Click the Open File button. 13. Select the data file created while the window reference position was being located (C:\Fbrm6?\Backup\ SN-X-mmddyy-WRP) and click Open. 14. For Time Averaging, select No Time Averaging. 15. Click the Apply Time Averaging box. 16. For Channel Grouping, select µm 200 Linear Ch. 17. From the Setup Menu, select Stat. Config. 18. Press the Load Stats. Config button. 19. From the C:\Fbrm6?\Statistics Setup Files folder, select the default0-100-lin.sta statistics setup (the measured statistic #/sec, No Wt 0-5 will reach a maximum value when the focal point is located at the window reference position) and click Open. 20. From the Display Menu, select Stats/Trend. 21. In the Show column, make sure there is an X next to the #/sec, No Wt 0-5 statistic (Statistic #5). This will create a trend plot in the trend display. 22. Deselect all other statistics by clicking on any existing Xs in the Show column. Deselecting a statistic will remove the trend from the trend display. 23. From the Y Axis applies to: box, select the #5 statistic. 24. Under X-axis Scale (located below the trend display), select Record. 25. Click the Scale All Trends box to reset the Y-axis. 26. The trend display should look similar to Figure 2 below, with two distinct peaks displayed. The left peak is from the mid-range adjustment and the right peak is from the fine adjustment. 27. Make sure the Lock Cursor box is not checked (i.e., an X is not displayed), select the + from the trend toolbar (located below the Trend Display), and move the Current Record crosshair on the trend display graph to the peak value measurement of the trended statistic as determined by the fine adjustment (right peak). 28. From the Setup Menu, select Annotation Rev F Page 9 of 11

10 29. Look at the Annotation field for the micrometer reading at this maximum value. The micrometer position corresponding to this value is the window reference position (e.g., 2.65 mm). In some cases, the maximum value of the trended statistic may be observed at more than one consecutive micrometer position. If a plateau is observed, the last micrometer position encountered at the maximum value of the statistic (i.e., the right-hand side of the plateau in the trend display) should be used as the window reference position (WRP). Figure 2 Example of Trend Display in the Data Review Module Window Reference Procedure Repeatability While all measurements of the window reference position should be within a range of ±5 microns (i.e., ±0.005 mm or one-half of an increment on the rotating dial), due to inconsistencies in ink application, you should not expect actual counts achieved to be the same each time. Contact a Mettler-Toledo Technology and Application Consultant if there is a greater than ±5-micron discrepancy in the measurement of the window reference position Rev F Page 10 of 11

11 Setting the Micrometer at the Standard Focal Point Position ( P Probes) The focal point position defines the location of the focal point relative to the external surface of the Lasentec probe window. The standard focal point position for an M-series process probe is 20 microns inside the sapphire window surface. Unless there is a specific, application-based need for an alternate position, use the standard focal point position for all applications. To set the micrometer at the required focal point position for an M-series process P probe: 1. Turn the rotating dial one full turn clockwise from the window reference position (e.g., if the WRP was at 2.65 mm, turn the micrometer clockwise to 2.15 mm). 2. Next, turn the rotating dial counterclockwise one full turn plus 20 microns (2 increments on the rotating dial). (e.g., for a WRP of 2.65 mm, turn the micrometer counterclockwise from 2.15 mm to 2.67 mm) to reach the standard focal point position. To eliminate the effects of hysteresis, you must always reach the final micrometer position going counterclockwise. If you pass the required focal point position, repeat steps 1 and 2 above. Cleaning the Probe Window For the purpose of this procedure, the probe window is considered clean if counts are less than 150 per second and not repeated in a single channel when the probe tip is immersed in de-ionized (DI) water. If acetone or other solvents are required to clean the probe window, the solvent should be thoroughly rinsed away with DI water. For other applications, see the section called Cleaning the Probe Window in the hardware manual Rev F Page 11 of 11

Lasentec Product Group

Lasentec Product Group Prepared by: Gregor Hsiao Approved on: ECN# 3506 Effective Date: 1/14/2008 Title: Window Reference Procedure for Version 6.0.0 and Higher Purpose: Lasentec D600 Probes and FBRM/Raman Probes using version

More information

Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base

Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base INTRODUCTION Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base Chemists frequently make use of the equivalent weight (eq. wt.) as the basis for volumetric calculations.

More information

iworx Sample Lab Experiment GB-2: Membrane Permeability

iworx Sample Lab Experiment GB-2: Membrane Permeability Experiment GB-2: Membrane Permeability Exercise 1: Movement of Small Positive Ions Across a Membrane Aim: To determine if small, positively charged, hydrogen ions can move across a membrane from a region

More information

Physics Labs with Computers, Vol. 1 P14: Simple Harmonic Motion - Mass on a Spring A

Physics Labs with Computers, Vol. 1 P14: Simple Harmonic Motion - Mass on a Spring A Activity P14: Simple Harmonic Motion - Mass on a Spring (Force Sensor, Motion Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Harmonic motion P14 SHM.DS P19 SHM Mass on a Spring

More information

pka AND MOLAR MASS OF A WEAK ACID

pka AND MOLAR MASS OF A WEAK ACID Experiment 10 pka AND MOLAR MASS OF A WEAK ACID Adapted by the Chemistry Faculty of Eastern Michigan University from EQUL 305,written by Richard C. Bell, Lebanon Valley College, published by Chemical Education

More information

Microscale Acid-Base Titration

Microscale Acid-Base Titration Microscale Acid-Base Titration Experiment 36 A titration is a process used to determine the volume of a solution needed to react with a given amount of another substance. In this experiment, you will titrate

More information

Rate Law Determination of the Crystal Violet Reaction. Evaluation copy

Rate Law Determination of the Crystal Violet Reaction. Evaluation copy Rate Law Determination of the Crystal Violet Reaction Computer 30 In this experiment, you will observe the reaction between crystal violet and sodium hydroxide. One objective is to study the relationship

More information

Acid-Base Titration. Evaluation copy

Acid-Base Titration. Evaluation copy Acid-Base Titration Computer 7 A titration is a process used to determine the volume of a solution that is needed to react with a given amount of another substance. In this experiment, your goal is to

More information

Magnetic Fields. Experiment 1. Magnetic Field of a Straight Current-Carrying Conductor

Magnetic Fields. Experiment 1. Magnetic Field of a Straight Current-Carrying Conductor General Physics Lab Department of PHYSICS YONSEI University Lab Manual (Lite) Magnetic Fields Ver.20181029 NOTICE This LITE version of manual includes only experimental procedures for easier reading on

More information

Evaluation copy. Ground Temperature Measurements. computer OBJECTIVES MATERIALS PRE-LAB QUESTIONS

Evaluation copy. Ground Temperature Measurements. computer OBJECTIVES MATERIALS PRE-LAB QUESTIONS Ground Temperature Measurements Computer P3 Throughout the daylight hours the sun heats up the air and the soil. Project 1 in this manual investigates the changes in air temperature during this process.

More information

ph electrode Instruction Manual For use with the RAH-210 & RPH-250

ph electrode Instruction Manual For use with the RAH-210 & RPH-250 ph electrode Instruction Manual For use with the RAH-210 & RPH-250 The information in this manual was current at the time of printing. The most current versions of all Hydro Instruments manuals can be

More information

Protocol Particle size distribution by centrifugal sedimentation (CPS)

Protocol Particle size distribution by centrifugal sedimentation (CPS) Protocol Particle size distribution by centrifugal sedimentation (CPS) 1. Method The CPS Disc Centrifuge separates particles by size using centrifugal sedimentation in a liquid medium. The sedimentation

More information

Acid-Base ph Titration Introduction

Acid-Base ph Titration Introduction Electronic Supplementary Material (ESI) for Chemistry Education Research and Practice. This journal is The Royal Society of Chemistry 2016 Appendix B: Example of Traditional Investigation Acid-Base ph

More information

The Determination of an Equilibrium Constant

The Determination of an Equilibrium Constant LabQuest 10 The equilibrium state of a chemical reaction can be characterized by quantitatively defining its equilibrium constant, Keq. In this experiment, you will determine the value of Keq for the reaction

More information

Determining the Concentration of a Solution: Beer s Law. Evaluation copy. Figure 1

Determining the Concentration of a Solution: Beer s Law. Evaluation copy. Figure 1 Determining the Concentration of a Solution: Beer s Law Computer 17 The primary objective of this experiment is to determine the concentration of an unknown copper (II) sulfate solution. You will use a

More information

Using Conductivity to Find an Equivalence Point

Using Conductivity to Find an Equivalence Point Experiment 25 PRE LAB DISCUSSION In this experiment, you will monitor conductivity during the reaction between sulfuric acid, and barium hydroxide in order to determine the equivalence point. From this

More information

Fog Monitor 100 (FM 100) Extinction Module. Operator Manual

Fog Monitor 100 (FM 100) Extinction Module. Operator Manual Particle Analysis and Display System (PADS): Fog Monitor 100 (FM 100) Extinction Module Operator Manual DOC-0217 Rev A-1 PADS 2.7.3, FM 100 Extinction Module 2.7.0 5710 Flatiron Parkway, Unit B Boulder,

More information

1iI1E. The Determination of 0 an Equilibrium Constant [LU. Computer

1iI1E. The Determination of 0 an Equilibrium Constant [LU. Computer Computer The Determination of 0 an Equilibrium Constant Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium constant,

More information

Activity P20: Conservation of Mechanical Energy (Force Sensor, Photogate)

Activity P20: Conservation of Mechanical Energy (Force Sensor, Photogate) Name Class Date Activity P20: Conservation of Mechanical Energy (Force Sensor, Photogate) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Energy P20 Mechanical Energy.DS P23 Cons. Mechanical

More information

TITRATION CURVES INTRODUCTION. Read and/or review Sections 4.10 and 16.7 in your textbook.

TITRATION CURVES INTRODUCTION. Read and/or review Sections 4.10 and 16.7 in your textbook. 1 TITRATION CURVES Copyright: Department of Chemistry, University of Idaho, Moscow, ID 83844-2343. 2013. INTRODUCTION Read and/or review Sections 4.10 and 16.7 in your textbook. In an acid - base titration,

More information

Exploring the Poles (Without Leaving Your Classroom!)

Exploring the Poles (Without Leaving Your Classroom!) Exploring the Poles (Without Leaving Your Classroom!) Computer 37 Magnets have north and south poles. Do you think that the poles of differently shaped magnets are in different places? In this activity,

More information

Shown below is a sample titration curve for a diprotic acid. Note the two equivalence points.

Shown below is a sample titration curve for a diprotic acid. Note the two equivalence points. EXPERIMENT 9 Titration Curve for a Polyprotic Acid INTRODUCTION Other than by strength and concentration, another way of classifying acids involves the number of H + ions an acid can donate. A monoprotic

More information

Kinetics of Crystal Violet Bleaching

Kinetics of Crystal Violet Bleaching Kinetics of Crystal Violet Bleaching Authors: V. C. Dew and J. M. McCormick* From Update March 12, 2013 with revisions Nov. 29, 2016 Introduction Chemists are always interested in whether a chemical reaction

More information

Experiment P05: Position, Velocity, & Acceleration (Motion Sensor)

Experiment P05: Position, Velocity, & Acceleration (Motion Sensor) PASCO scientific Physics Lab Manual: P05-1 Experiment P05: Position, Velocity, & Acceleration (Motion Sensor) Concept Time SW Interface Macintosh file Windows file linear motion 30 m 500 or 700 P05 Position,

More information

LAB 2 - ONE DIMENSIONAL MOTION

LAB 2 - ONE DIMENSIONAL MOTION Name Date Partners L02-1 LAB 2 - ONE DIMENSIONAL MOTION OBJECTIVES Slow and steady wins the race. Aesop s fable: The Hare and the Tortoise To learn how to use a motion detector and gain more familiarity

More information

Titration of Acids and Bases

Titration of Acids and Bases Exercise 3 Page 1 Illinois Central College CHEMISTRY 132 Titration of Acids and Bases Name: Equipment 1-25 ml burette 1-pH electrode 1-50 and 1-150 ml beaker 1-stir plate and stir bar 1-Vernier Interface

More information

LABORATORY INVESTIGATION

LABORATORY INVESTIGATION LABORATORY INVESTIGATION Diffusion Through a Dialysis Membrane Transport of substances into and out of the cell is necessary in order to sustain life. Substances transported into the cell are used for

More information

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS) Liquid Chromatography Mass Spectrometry (LC/MS) Prelab Questions: Questions to be answered before doing the experiment. The answers are due at the beginning of each experiment without exception (the questions

More information

Measurements of a Table

Measurements of a Table Measurements of a Table OBJECTIVES to practice the concepts of significant figures, the mean value, the standard deviation of the mean and the normal distribution by making multiple measurements of length

More information

University of Minnesota Nano Center Standard Operating Procedure

University of Minnesota Nano Center Standard Operating Procedure University of Minnesota Nano Center Standard Operating Procedure Equipment Name: Zeta Potential Analyzer Model: Stabino Location: PAN 185 Badger Name: Not on Badger Revision Number: 0-Inital release Revisionist:

More information

Determining the Concentration of a Solution: Beer s Law

Determining the Concentration of a Solution: Beer s Law Determining the Concentration of a Solution: Beer s Law LabQuest 11 The primary objective of this experiment is to determine the concentration of an unknown nickel (II) sulfate solution. You will be using

More information

for a Weak Acid 1-ring stand stir plate and stir bar ph Probe ml burette 1-50 ml and ml beaker Drop counter ph 7.00 buffer 0.

for a Weak Acid 1-ring stand stir plate and stir bar ph Probe ml burette 1-50 ml and ml beaker Drop counter ph 7.00 buffer 0. Exercise 4 Page 1 Illinois Central College CHEMISTRY 132 Name: Determination of for a Weak Acid Equipment Objectives 1-ring stand stir plate and stir bar ph Probe 1-25.00 ml burette 1-50 ml and 1-250 ml

More information

Lab 1a Wind Tunnel Testing Principles & Drag Coefficients of Golf balls

Lab 1a Wind Tunnel Testing Principles & Drag Coefficients of Golf balls Lab 1a Wind Tunnel Testing Principles & Drag Coefficients of Golf balls OBJECTIVES - To perform air flow measurement using the wind tunnel. - To compare measured and theoretical velocities for various

More information

Standardizing a Solution of Sodium Hydroxide. Evaluation copy

Standardizing a Solution of Sodium Hydroxide. Evaluation copy Standardizing a Solution of Sodium Hydroxide Computer 6 It is often necessary to test a solution of unknown concentration with a solution of a known, precise concentration. The process of determining the

More information

2 (aq) [FeSCN [Fe 3JSCN] Figure 1

2 (aq) [FeSCN [Fe 3JSCN] Figure 1 The Determination of an Equilibrium Constant Computer Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium constant,

More information

Lab 12 - Conservation of Momentum And Energy in Collisions

Lab 12 - Conservation of Momentum And Energy in Collisions Lab 12 - Conservation of Momentum And Energy in Collisions Name Partner s Name I. Introduction/Theory Momentum is conserved during collisions. The momentum of an object is the product of its mass and its

More information

Hooke s Law. Equipment. Introduction and Theory

Hooke s Law. Equipment. Introduction and Theory Hooke s Law Objective to test Hooke s Law by measuring the spring constants of different springs and spring systems to test whether all elastic objects obey Hooke s Law Equipment two nearly identical springs,

More information

AlphaLab Gaussmeter Model VGM -- Instructions

AlphaLab Gaussmeter Model VGM -- Instructions AlphaLab Gaussmeter Model VGM -- Instructions Power: As shipped, a standard 9 volt battery is installed. Battery life is about 10 hours (25 ma drain; 6.5 volt dropout). Low battery appears when about one

More information

Experiment 2: THE DENSITY OF A SOLID UNKNOWN AND CALIBRATION WITH DATASTUDIO SOFTWARE

Experiment 2: THE DENSITY OF A SOLID UNKNOWN AND CALIBRATION WITH DATASTUDIO SOFTWARE Experiment 2: THE DENSITY OF A SOLID UNKNOWN AND CALIBRATION WITH DATASTUDIO SOFTWARE Concepts: Density Equipment Calibration Approximate time required: 90 minutes for density 90 minutes for two thermometers

More information

Experiment C-10 Titration of a Strong Acid and a Strong Base

Experiment C-10 Titration of a Strong Acid and a Strong Base 1 Experiment C-10 Titration of a Strong Acid and a Strong Base Objectives To study the titration process. To follow changes in the ph during the titration process while adding a strong base to a strong

More information

Exercise 2-4. Titration of a Buffer Solution EXERCISE OBJECTIVES

Exercise 2-4. Titration of a Buffer Solution EXERCISE OBJECTIVES Exercise 2-4 Titration of a Buffer Solution EXERCISE OBJECTIVES To define the terms buffer solution and buffer capacity; To titrate a buffer solution with a weak acid solution; To plot a graph using the

More information

EXPERIMENT 15. USING CONDUCTIVITY TO LOOK AT SOLUTIONS: DO WE HAVE CHARGED IONS OR NEUTRAL MOLECULES? rev 7/09

EXPERIMENT 15. USING CONDUCTIVITY TO LOOK AT SOLUTIONS: DO WE HAVE CHARGED IONS OR NEUTRAL MOLECULES? rev 7/09 EXPERIMENT 15 USING CONDUCTIVITY TO LOOK AT SOLUTIONS: DO WE AVE CARGED IONS OR NEUTRAL MOLECULES? rev 7/09 GOAL After you complete this experiment, you should have a better understanding of aqueous solutions

More information

Sample Alignment (2D detector) Part

Sample Alignment (2D detector) Part Sample Alignment (2D detector) Part Contents Contents 1 How to set Part conditions...1 1.1 Setting conditions... 1 1.2 Customizing scan conditions and slit conditions... 6 2 Sample alignment sequence...13

More information

Zetasizer Nano-ZS User Instructions

Zetasizer Nano-ZS User Instructions Zetasizer Nano-ZS User Instructions 1. Activate the instrument computer by logging in to CORAL. If needed, log in to the local instrument computer Username: zetasizer. Password: zetasizer. 2. Instrument

More information

DustMon A Standard Operating Procedure From the Microtrac Particle Analysis Laboratory (PAL) SL-AN-44 Revision C

DustMon A Standard Operating Procedure From the Microtrac Particle Analysis Laboratory (PAL) SL-AN-44 Revision C DustMon A Standard Operating Procedure From the Microtrac Particle Analysis Laboratory (PAL) SL-AN-44 Revision C Introduction This operating procedure is provided to assist the user of the Dustmon in obtaining

More information

Conservation of Momentum Using PASCO TM Carts and Track to Study Collisions in One Dimension

Conservation of Momentum Using PASCO TM Carts and Track to Study Collisions in One Dimension 14 Conservation of Conservation of Using PASCO TM Carts and Track to Study s in One Dimension When two objects collide momentum is transferred between them. p is defined as the product of mass and velocity

More information

Acid-Base Titration. Sample

Acid-Base Titration. Sample Acid-Base Titration Computer 7 A titration is a process used to determine the volume of a solution that is needed to react with a given amount of another substance. In this experiment, your goal is to

More information

Jasco V-670 absorption spectrometer

Jasco V-670 absorption spectrometer Laser Spectroscopy Labs Jasco V-670 absorption spectrometer Operation instructions 1. Turn ON the power switch on the right side of the spectrophotometer. It takes about 5 minutes for the light source

More information

The Determination of an Equilibrium Constant

The Determination of an Equilibrium Constant The Determination of an Equilibrium Constant Computer 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium

More information

Force vs time. IMPULSE AND MOMENTUM Pre Lab Exercise: Turn in with your lab report

Force vs time. IMPULSE AND MOMENTUM Pre Lab Exercise: Turn in with your lab report IMPULSE AND MOMENTUM Pre Lab Exercise: Turn in with your lab report Newton s second law may be written r r F dt = p where F is the force and p is the change in momentum. The area under the force vs. time

More information

Name Date Period. 1. If drops of ACID are added to a ph buffer, then the ph of the buffer will [increase / decrease / stay the same].

Name Date Period. 1. If drops of ACID are added to a ph buffer, then the ph of the buffer will [increase / decrease / stay the same]. Name Date Period ACIDS AND BASES Organisms are often very sensitive to the effect of s and s in their environment. They need to maintain a stable internal ph in order to survive even in the event of environmental

More information

Senior astrophysics Lab 2: Evolution of a 1 M star

Senior astrophysics Lab 2: Evolution of a 1 M star Senior astrophysics Lab 2: Evolution of a 1 M star Name: Checkpoints due: Friday 13 April 2018 1 Introduction This is the rst of two computer labs using existing software to investigate the internal structure

More information

Lab 1 Uniform Motion - Graphing and Analyzing Motion

Lab 1 Uniform Motion - Graphing and Analyzing Motion Lab 1 Uniform Motion - Graphing and Analyzing Motion Objectives: < To observe the distance-time relation for motion at constant velocity. < To make a straight line fit to the distance-time data. < To interpret

More information

Activity P27: Speed of Sound in Air (Sound Sensor)

Activity P27: Speed of Sound in Air (Sound Sensor) Activity P27: Speed of Sound in Air (Sound Sensor) Concept Speed of sound DataStudio P27 Speed of Sound 1.DS Equipment Needed Qty Other Qty Sound Sensor (CI-6506B) 1 Tape, duct 1 roll Base and Support

More information

Spectrometer User s Guide

Spectrometer User s Guide Spectrometer User s Guide (Order Codes: V-SPEC, SPRT-VIS, SP-VIS, SP-UV-VIS, ESRT-VIS) The spectrometer is a portable light spectrophotometer, combining a spectrometer and a light source/cuvette holder.

More information

Experiment P30: Centripetal Force on a Pendulum (Force Sensor, Photogate)

Experiment P30: Centripetal Force on a Pendulum (Force Sensor, Photogate) PASCO scientific Physics Lab Manual: P30-1 Experiment P30: (Force Sensor, Photogate) Concept Time SW Interface Macintosh File Windows File centripetal force 30 m 500 or 700 P30 Centripetal Force P30_CENT.SWS

More information

The Decomposition of Hydrogen Peroxide. Evaluation copy

The Decomposition of Hydrogen Peroxide. Evaluation copy The Decomposition of Hydrogen Peroxide Computer 12 The decomposition of hydrogen peroxide in aqueous solution proceeds very slowly. A bottle of 3% hydrogen peroxide sitting on a grocery store shelf is

More information

NE01 - Centripetal Force. Laboratory Manual Experiment NE01 - Centripetal Force Department of Physics The University of Hong Kong

NE01 - Centripetal Force. Laboratory Manual Experiment NE01 - Centripetal Force Department of Physics The University of Hong Kong Background Introduction Laboratory Manual Experiment Department of Physics The University of Hong Kong Circular Motion is one of the simplest forms of 2-dimensional motion in which the locus of the object

More information

How to perform 2D NMR Experiments on the Varian/Agilent VNMRS 500 when using the Chempack interface

How to perform 2D NMR Experiments on the Varian/Agilent VNMRS 500 when using the Chempack interface How to perform 2D NMR Experiments on the Varian/Agilent VNMRS 500 when using the Chempack interface 1 June 3, 2014 To start: 1. Insert your sample 2. If running any 2D spectrum or even just a 1D 13 C spectrum

More information

Evaluation copy. Acids and Bases. computer OBJECTIVES MATERIALS

Evaluation copy. Acids and Bases. computer OBJECTIVES MATERIALS Acids and Bases Computer 2 Organisms are often very sensitive to the effect of s and s in their environment. They need to maintain a stable internal ph in order to survive even in the event of environmental

More information

Additivity of Heats of Reaction: Hess s Law

Additivity of Heats of Reaction: Hess s Law Additivity of Heats of Reaction: Hess s Law Experiment In this experiment, you will use a Styrofoam-cup calorimeter to measure the heat released by three reactions. One of the reactions is the same as

More information

Lab 5 Enthalpy of Solution Formation

Lab 5 Enthalpy of Solution Formation Chemistry 3202 Lab 5 Enthalpy of Solution Formation Page 1 of 9 Lab 5 Enthalpy of Solution Formation Introduction This lab activity will introduce you to the measurement of energy change associated with

More information

Acid-Base Titration Curves Using a ph Meter

Acid-Base Titration Curves Using a ph Meter Acid-Base Titration Curves Using a ph Meter Introduction: In this experiment you will use a ph sensor to collect volume and ph data as you titrate two acids with sodium hydroxide. You will obtain titration

More information

Acid-Base Titration Curves Using a ph Meter

Acid-Base Titration Curves Using a ph Meter Acid-Base Titration Curves Using a ph Meter Introduction: In this experiment you will use a ph sensor to collect volume and ph data as you titrate two acids with sodium hydroxide. You will obtain titration

More information

Become a Microprobe Power User Part 2: Qualitative & Quantitative Analysis

Become a Microprobe Power User Part 2: Qualitative & Quantitative Analysis Become a Microprobe Power User Part 2: Qualitative & Quantitative Analysis Mike Spilde Spring IOM Seminar February 5, 2008 Qualitative Analysis Why use qualitative scans? Elemental ID (especially trace

More information

Conductometric Titration & Gravimetric Determination of a Precipitate

Conductometric Titration & Gravimetric Determination of a Precipitate Conductometric Titration & Gravimetric Determination of a Precipitate Experiment 9 In this experiment, you will monitor conductivity during the reaction between sulfuric acid, H2SO4, and barium hydroxide,

More information

Cart on a Ramp. Evaluation Copy. Figure 1. Vernier Dynamics Track. Motion Detector Bracket

Cart on a Ramp. Evaluation Copy. Figure 1. Vernier Dynamics Track. Motion Detector Bracket Cart on a Ramp Computer 3 This experiment uses an incline and a low-friction cart. If you give the cart a gentle push up the incline, the cart will roll upward, slow and stop, and then roll back down,

More information

Polar alignment in 5 steps based on the Sánchez Valente method

Polar alignment in 5 steps based on the Sánchez Valente method 1 Polar alignment in 5 steps based on the Sánchez Valente method Compared to the drift alignment method, this one, allows you to easily achieve a perfect polar alignment in just one step. By "perfect polar

More information

Connect the Vernier spectrometer to your lap top computer and power the spectrometer if necessary. Start LoggerPro on your computer.

Connect the Vernier spectrometer to your lap top computer and power the spectrometer if necessary. Start LoggerPro on your computer. Connect the Vernier spectrometer to your lap top computer and power the spectrometer if necessary. Start LoggerPro on your computer. The screen shown in Fig. 1 may be displayed. If status line displays

More information

Experiment P43: RC Circuit (Power Amplifier, Voltage Sensor)

Experiment P43: RC Circuit (Power Amplifier, Voltage Sensor) PASCO scientific Vol. 2 Physics Lab Manual: P43-1 Experiment P43: (Power Amplifier, Voltage Sensor) Concept Time SW Interface Macintosh file Windows file circuits 30 m 700 P43 P43_RCCI.SWS EQUIPMENT NEEDED

More information

EXPERIMENT 9 SALTWATER CONDUCTANCE: The Effect of Concentration

EXPERIMENT 9 SALTWATER CONDUCTANCE: The Effect of Concentration EXPERIMENT 9 SALTWATER CONDUCTANCE: The Effect of Concentration Introduction According to the Theory of Ionization proposed by S. Arrhenius, about 1880, ionic compounds dissolve in water forming cations

More information

Old Dominion University Physics 112N/227N/232N Lab Manual, 13 th Edition

Old Dominion University Physics 112N/227N/232N Lab Manual, 13 th Edition RC Circuits Experiment PH06_Todd OBJECTIVE To investigate how the voltage across a capacitor varies as it charges. To find the capacitive time constant. EQUIPMENT NEEDED Computer: Personal Computer with

More information

Chemical Kinetics I: The Dry Lab. Up until this point in our study of physical chemistry we have been interested in

Chemical Kinetics I: The Dry Lab. Up until this point in our study of physical chemistry we have been interested in Chemical Kinetics I: The Dry Lab Up until this point in our study of physical chemistry we have been interested in equilibrium properties; now we will begin to investigate non-equilibrium properties and

More information

Acids and Bases. Figure 1

Acids and Bases. Figure 1 DataQuest 9 Organisms are often very sensitive to the effect of s and s in their environment. They need to maintain a stable internal ph in order to survive even in the event of environmental changes.

More information

Physics Spring 2006 Experiment 4. Centripetal Force. For a mass M in uniform circular motion with tangential speed v at radius R, the required

Physics Spring 2006 Experiment 4. Centripetal Force. For a mass M in uniform circular motion with tangential speed v at radius R, the required Centripetal Force I. Introduction. In this experiment you will study the centripetal force required for a mass in uniform circular motion. You will determine the centripetal forces required for different

More information

Chemical Equilibrium: Finding a Constant, Kc

Chemical Equilibrium: Finding a Constant, Kc Chemical Equilibrium: Finding a Constant, Kc Computer 20 The purpose of this lab is to experimentally determine the equilibrium constant, K c, for the following chemical reaction: Fe 3+ (aq) + SCN (aq)

More information

Investigation #2 TEMPERATURE VS. HEAT. Part I

Investigation #2 TEMPERATURE VS. HEAT. Part I Name: Investigation #2 Partner(s): TEMPERATURE VS. HEAT These investigations are designed to help you distinguish between two commonly confused concepts in introductory physics. These two concepts, temperature

More information

Chemistry MultiMeasure Sensor PS-2170

Chemistry MultiMeasure Sensor PS-2170 ä 5 Instruction Manual 012-09523A Chemistry MultiMeasure Sensor PS-2170 ph/ise/orp PRESSURE CHEMISTRY TEMPERATURE SENSOR VOLTAGE 1 2 3 4 PS-2170 5 6 7 cc 8 10 15 20 25 30 35 40 45 50 55 60 Included Equipment

More information

ACID-BASE TITRATION (MICROSCALE)

ACID-BASE TITRATION (MICROSCALE) ACID-BASE TITRATION (MICROSCALE) LAB PH 4.PALM From Science with Handhelds, Vernier Software & Technology, 2002. INTRODUCTION Acids and bases represent a major class of chemical substances. We encounter

More information

Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS)

Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS) Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS) The UV-VIS spectrometer is a portable ultraviolet light and visible light spectrophotometer, combining a spectrometer and a light source/cuvette

More information

The Determination of an Equilibrium Constant

The Determination of an Equilibrium Constant The Determination of an Equilibrium Constant Chemistry 102 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium

More information

Chemical Equilibrium: Finding a Constant, Kc

Chemical Equilibrium: Finding a Constant, Kc Chemical Equilibrium: Finding a Constant, Kc Experiment 20 The purpose of this lab is to experimentally determine the equilibrium constant, K c, for the following chemical reaction: Fe 3+ (aq) + SCN (aq)

More information

( ) ( ) = q o. T 12 = τ ln 2. RC Circuits. 1 e t τ. q t

( ) ( ) = q o. T 12 = τ ln 2. RC Circuits. 1 e t τ. q t Objectives: To explore the charging and discharging cycles of RC circuits with differing amounts of resistance and/or capacitance.. Reading: Resnick, Halliday & Walker, 8th Ed. Section. 27-9 Apparatus:

More information

Activity P60: Inverse Square Law Nuclear (Nuclear Sensor, Rotary Motion Sensor)

Activity P60: Inverse Square Law Nuclear (Nuclear Sensor, Rotary Motion Sensor) Name Class Date Activity P60: Inverse Square Law Nuclear (Nuclear Sensor, Rotary Motion Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Radioactivity P60 Nuclear Inv Sqr Law.DS P60

More information

HESS S LAW: ADDITIVITY OF HEATS OF REACTION

HESS S LAW: ADDITIVITY OF HEATS OF REACTION HESS S LAW: ADDITIVITY OF HEATS OF REACTION From Chemistry with Calculators, Vernier Software & Technology In this experiment, you will use a Styrofoam-cup calorimeter to measure the heat released by three

More information

Elastic Properties of Solids Exercises I II for one weight Exercises III and IV give the second weight

Elastic Properties of Solids Exercises I II for one weight Exercises III and IV give the second weight Elastic properties of solids Page 1 of 8 Elastic Properties of Solids Exercises I II for one weight Exercises III and IV give the second weight This is a rare experiment where you will get points for breaking

More information

Static and Kinetic Friction

Static and Kinetic Friction Experiment 12 If you try to slide a heavy box resting on the floor, you may find it difficult to get the box moving. Static friction is the force that is acting against the box. If you apply a light horizontal

More information

IncuCyte ZOOM NeuroTrack Fluorescent Processing

IncuCyte ZOOM NeuroTrack Fluorescent Processing IncuCyte ZOOM NeuroTrack Fluorescent Processing The NeuroTrack TM Software Module (Cat No 9600-0011) is used to measure the processes of neurons in monoculture or with fluorescent labeling in co-culture.

More information

Kinematics Lab. 1 Introduction. 2 Equipment. 3 Procedures

Kinematics Lab. 1 Introduction. 2 Equipment. 3 Procedures Kinematics Lab 1 Introduction An object moving in one dimension and undergoing constant or uniform acceleration has a position given by: x(t) =x 0 +v o t +1/2at 2 where x o is its initial position (its

More information

This sensor is now also available with a sturdier flat electrode which enables testing even smaller samples than the conventional electrode.

This sensor is now also available with a sturdier flat electrode which enables testing even smaller samples than the conventional electrode. ph sensor with electrode ENPH-A016 ph sensor with flat electrode ENPH-F052 The ph sensor replaces the traditional ph meter adding powerful capabilities such as collecting the ph data and displaying the

More information

17 Isolation and Epoxidation of a Natural Product: R-(+)-Limonene

17 Isolation and Epoxidation of a Natural Product: R-(+)-Limonene Experiment 17 Isolation and Epoxidation of a Natural Product: R-(+)-Limonene Natural products are compounds produced by living organisms. Recently a great deal of exploration has been done involving the

More information

The Effect of Alcohol on Biological Membranes

The Effect of Alcohol on Biological Membranes TEACHER INFORMATION The Effect of Alcohol on Biological Membranes Experiment 1. The student pages with complete instructions for using a Colorimeter or Spectrometer with LabQuest App and Logger Pro (computers)

More information

The Phase Change Lab: Freezing and Melting of Water

The Phase Change Lab: Freezing and Melting of Water The Phase Change Lab: Freezing and Melting of Water Experiment 3 Freezing temperature is the temperature at which a substance turns from a liquid to a solid. Melting temperature is the temperature at which

More information

Transpiration. Evaluation copy

Transpiration. Evaluation copy Transpiration Computer 9 Water is transported in plants, from the roots to the leaves, following a decreasing water potential gradient. Transpiration, or loss of water from the leaves, helps to create

More information

Harmonic Motion. Mass on a Spring. Physics 231: General Physics I Lab 6 Mar. 11, Goals:

Harmonic Motion. Mass on a Spring. Physics 231: General Physics I Lab 6 Mar. 11, Goals: Physics 231: General Physics I Lab 6 Mar. 11, 2004 Names: Harmonic Motion Goals: 1. To learn about the basic characteristics of periodic motion period, frequency, and amplitude 2. To study what affects

More information

Acid-Base Titration. Computer OBJECTIVES

Acid-Base Titration. Computer OBJECTIVES Acid-Base Titration Computer 7 A titration is a process used to determine the volume of a solution that is needed to react with a given amount of another substance. In this experiment, your goal is to

More information

HMS-5000 Manual. Product Name: HMS-5000 Hall Effect Measurement System with variable temperature from 80K to 350K. - Manual version: ver 5.

HMS-5000 Manual. Product Name: HMS-5000 Hall Effect Measurement System with variable temperature from 80K to 350K. - Manual version: ver 5. HMS-5000 Manual Product Name: HMS-5000 Hall Effect Measurement System with variable temperature from 80K to 350K - Manual version: ver 5.01- www.ecopia21.co.kr - Table of contents - 1. Hardware Installation

More information

General Physics I Lab. M1 The Atwood Machine

General Physics I Lab. M1 The Atwood Machine Purpose General Physics I Lab In this experiment, you will learn the basic operation of computer interfacing and use it in an experimental study of Newton s second law. Equipment and components Science

More information

Heat of Combustion: Magnesium

Heat of Combustion: Magnesium Heat of Combustion: Magnesium Experiment 21 In Experiment 17, you learned about the additivity of reaction heats as you confirmed Hess s Law. In this experiment, you will use this principle as you determine

More information

Ka of Unknown Acid In this experiment you will determine the Ka of an unknown acid by titration with the sodium hydroxide.

Ka of Unknown Acid In this experiment you will determine the Ka of an unknown acid by titration with the sodium hydroxide. Ka of Unknown Acid In this experiment you will determine the Ka of an unknown acid by titration with the sodium hydroxide. Because you will be titrating an unknown acid again, you will be using many of

More information