Basic chromatographic parameters and optimization in LC

Size: px
Start display at page:

Download "Basic chromatographic parameters and optimization in LC"

Transcription

1 AM0925 Assignment Basic chromatographic parameters and optimization in LC Introduction This is a computer exercise where you will apply a simulator of reversed phase LC to study the influence of chromatographic conditions on retention, efficiency, selectivity and resolution. The first thing you have to do is to download the HPLC simulator and become familiar with the interface. The HPLC simulator can be downloaded from Press "Launch HPLC Simulator" Note! If the HPLC simulator is blocked by Java security settings. See solution to the problem in the appendix. Select thereafter open with Java Web Start Launcher The window should look approximately as shown in Figure 1 below. Figure 1. The HPLC Simulator In the panel to the left you can select different conditions. The conditions you need to modify during the exercise are: Mobile phase composition, Solvent B : Either methanol or acetonitrile, which in addition to water are the two most common solvents in reversed phase LC. Mobile phase composition, Solvent B fraction : Percent of mobile phase that is solvent B. Chromatographic properties, Temperature : The temperature of the column. 1

2 Chromatographic properties, Flow rate : The flow through the column in ml/min. Column properties, Stationary phase : Waters Acquity BEH-C18 or Agilent Zorbax SB-C18. Note that changing the column also mean changing the analytes. There are nine analytes in the example with the Waters column and five analytes in the example with the Agilent column. Column properties, Particle size: Average diameter of the particles in the column (μm) All other settings should always be kept at the default values when you do the exercise. You can reset to the default settings by choosing [File]+[Reset To Default Settings]. When you select a new column the settings will also be changed to the default values for the column. Before you start the exercise it may be a good idea to play around with different settings to get used to the user interface and see how the different parameters influence the chromatography. Part A, van Deemter curves The purpose of this part is to investigate how the flow rate and the column temperature influence the parameters in the van Deemter curve. Start by doing the following: Reset all settings to default values as explained above. Select one of the two columns. Select one of the two options (acetonitrile or methanol) for solvent B. Set the temperature to 25 C. Select any compound in the peak parameters list. Adjust solvent B fraction so that the retention factor, k (column 3 in the peak parameters list) is somewhere between 4 and 6. Thereafter you verify that the particle size is set to 3 μm and you record the following for your notes: the applied column, selected solvent B, solvent B fraction, the selected compound, and the retention factor. You should now vary the flow rate and record the following parameters: retention time, peak width and the backpressure. The backpressure is the pressure that must be applied by the pump to achieve the required flow rate and is found in the panel to the left under Chromatographic Properties. The two other values are reported in the list of peak parameters as [tr (min)] and [σ total (s)]. You should record values for the following flow rates, 0.5, 1.0, 2.0, 4.0 and 8.0 ml/min. Organize the recorded values in a spreadsheet (Excel, Calc or similar) as shown below. Flow rate [ml/min] Backpressure [Bar] Retention time, t R [min] Peak width, σ [s] Thereafter you increase the particle size to 5 μm and repeat the process. Increase thereafter the temperature to 75 C and repeat the process with both 3 and 5 μm particles. In the end you should have four tables (2 particle sizes 2 temperatures). When you change the temperature the retention factor will also change. Record k for 75 C in addition to the one that you have for 25 C. 2

3 From the recorded retention times and peak widths you thereafter calculate the plate number, N, and the plate height, H. The plate height should be calculated in mm. Note that the retention times are given in minutes and the peak widths are given in seconds. They must therefore be converted to the same scale. Also note that peak widths are reported as the standard deviation, σ, which is 1/4 of the peak width at baseline, w b. The plate number can be calculated directly from σ or from w b by the following equation: N=( t R σ ) 2 =16( t R 2 w b) The plate height, H, is the column length, L, divided by the plate number: H=L/ N To verify that you calculate correctly, you can use the following example. If the retention time is 3.22 min and the peak width reported as 1σ is 2.06 s, the plate number is 8791, and since the column length is 100 mm, the plate height is mm. When you have calculated the plate height for all the recorded data the next step is to solve the van Deemter equation and to evaluate how the flow rate and the temperature influence the three constants, A, B and C, which explain the multiple path effect, longitudinal diffusion and resistance to mass transfer, respectively. The van Deemter equation is given below. (1) (2) H=A+ B u +C u (3) The parameter u is the mobile phase velocity. This is the value reported as chromatographic flow velocity under Column properties in the HPLC simulator. In LC the column flow is proportional to the mobile phase velocity, and can therefore replace u. The van Deemter equation is not a standard equation that can be solved by the built-in equations in Excel or Calc, but it can be solved by a tool called "Solver", which is found in both programs (in Excel it may have to be activated, see link if necessary). In LibreOffice Calc you find the Solver under the [Tools] menu. In Excel it is usually found in the rightmost group under the [Data] menu. Depending on which spreadsheet program you use, you should download the "CURVES_LC.ods" (LibreOffice) or the "CURVES_LC.xls" (Excel) file that you find in the module for the exercise at Mitt UiB. The following example shows how you create the van Deemter curves and find the constants A, B and C in LibreOffice Calc. The spreadsheet for calculating van Deemter curves and the settings for Solver is shown in Figure 2. 3

4 Solver found here G13 (SSE) is the target cell Minimize target Your calculated plate heights Cells C6:E6 (A,B,C) can be changed van Deemter curve Figure 2. The CURVES_LC.odsˮ file and the settings for Solver in LibreOffice. The seven columns in the table are the following: A) The column flow. B) The observed plate heights, H obs. Your calculated values should be entered here. Use [Paste Special] and [Paste as numbers] when you copy the data into the sheet. C) The A term. D) The constant for the B term (blue) and the effect of the constant at the applied flow (B/flow). E) The constant for the C term (blue) and the effect of the constant at the applied flow (C flow). F) The calculated plate height, H calc. When the model is fitted, H calc should be similar to H obs. The values in this column are the sum of the values in columns C-E. G) Squared errors, SE. This is the squared value of the difference between observed and predicted plate height (H calc H obs ) 2. The sum of the squared errors (SSE) is given at the bottom of the table and this value explains how well the model fits the observed data. The value should be as low as possible. For each of the four sets of experiments you should calculate the van Deemter curve and record the values of the three constants, A, B, and C. You should also copy the figures of the van Deemter curves for your report. When you paste your values into the first column the model will no longer fit the data and it must therefore be updated. You can try to adjust the three constants manually to get a better fit, but in the end you should apply the Solver using the following procedure: Select the cell with the value for SSE Go to the [Tools] menu and select [Solver] ([Data] menu in Excel). In Solver, verify that the target cell is G13 or change it. Select Optimize result to minimum Put the cursor in the field next to "by changing cells" and select the three cells with blue text in the spreadsheet. 4

5 Thereafter you press the [Solve] button. Press [OK] and [Keep result]. In LibreOffice it can be an idea to press [Continue] a few times before pressing [OK] until you see that you have a stable result. Your observed values should now be approximately at the red curve and SSE should be zero, which show that the model now explains the observations. Record the values for the three constants and copy the plot before you continue with the next data set. Reporting Introduction Give a brief explanation of the purpose of the exercise and the theory of band broadening in packed columns (approximately 1/2 page). The van Deemter equation and the equations for plate number, plate height and optimal velocity/flow rate should be given and explained. Explain that flow were used instead of velocity and that these two parameters are proportional in LC. Explain that the HPLC simulator was applied and how the van Deemter equation was solved. Experimental Under experimental, report the applied conditions: Results The selected column Column dimensions (length, diameter, particle size) The selected compound Solvent A Solvent B Solvent B fraction The retention factor at 25 and 75 C Explain that the parameters you varied were the temperature and the column flow. Include the van Deemter plots and a table of the A, B and C constants for the four experiments. Make a figure that illustrates how the constants vary with particle size and the temperature. The mobile phase velocity (or flow rate in this case) that minimizes the plate height, and therefore maximizes the plate number, is often referred to as the optimal mobile phase velocity, u opt. Since this is a minimum in the curve, it is found where the derivative of the van Deemter equation is zero, and it can be calculated by Equation 4. u opt = B C (4) Use this equation to calculate the optimal flow rate for all four experiments. Thereafter insert u opt in the van Deemter equation and calculate the plate heights, H, at these flow rates. Verify that your calculations are in accordance with the plots. The calculated values should be included in the report. Discuss how the particle size and the temperature affect the three constants of the van Deemter equations, and how this also affects u opt and the plate height at u opt. As long as the retention factor, k, is constant, the time it takes to elute a compound is inversely proportional to the flow. Make a plot of retention time (y-axis) against flow rate (x-axis) for the two experiments with 3 μm particles (plots for 5 μm particles will be identical). Discuss what this relationship means for the time it takes to perform the separation. A power function (ax b ) can be fitted through the data points. 5

6 Finally you should make a plot of the recorded backpressures (y-axis) against flow rates (x-axis) for all four experiments. Conventional HPLC pumps can typically deliver pressures up to bar, while pumps for UHPLC (Ultra High Performance LC) typically can deliver pressures up to bar. However, it is usually a good advice to have good margins, and to keep the pressure below 50% of the maximum allowed pressure when a method is developed. Discuss which limitations the different particle sizes and temperatures set on the flow rates that can be applied. Part B, Method optimization challenge In part B of the exercise you should try to develop a good method for the separation of the five compounds in the problem with the Agilent Zorbax column. Start by resetting to default settings ([File] menu) and select this column. Your goal is that the retention time of the last compound should be as low as possible, the resolution, R s, between all peaks should be at least 1.5, and the backpressure should not exceed 300 bar. Solvent B should be set to acetonitrile. The resolution is calculated by Equation 5. R s =2 t R (B) t R (A) w b (B) +w b (A) =0.5 t R (B) t R (A) σ (B) +σ (A) (5) w b is peak widths at baseline (4σ) and t R is retention times. A and B denote respectively the first and the last of the two peaks. You are allowed to vary the following parameters: The solvent B fraction. The temperature, but only in the range 30 to 60 C. The flow rate. The particle size can be set to 3, 4 or 5 μm. All other parameters should be kept at the default values for the Agilent Zorbax column. Note that the choice of solvent B, the solvent B fraction and the temperature all have large effects on the selectivity and can change the elution order of the compounds. Also pay attention to what part A of the exercise tells about the effects of the particle size and temperature on the constants in the van Deemter equation, the optimal flow rate and the backpressure. It may also help to consider what the Purnell equation tells about the three factors that are necessary to achieve separation, efficiency, selectivity and retention. The equation is given below. R s = N B 4 (α 1 α )( k B (6) k B +1) α is the separation factor that explains the selectivity, and this is the ratio of the two retention factors, k B /k A, where A and B refer to the first and last of the two peaks, respectively. If N (efficiency), α (selectivity), or k (retention) is zero, there will be no separation. This can for instance be illustrated by setting the solvent B fraction to 100%; the retention factors drops to nearly zero and all peaks are overlapping. The equation also tells that if the selectivity is high, it may still be possible to get sufficient separation even if the efficiency and the retention is fairly low. In a system where the parameters have large influence on the selectivity (as in this case) it is often a good idea to first find a combination of solvent composition and temperature that gives a good selectivity. When you have found the solution you are satisfied with, record the values of all the parameters you have varied. Record t R of the last eluting compound and the backpressure. For the two peaks with the poorest resolution you record the retention time (t R ), the retention factor (k) and the peak 6

7 width (σ). If there are three peaks that seem equally spaced you record these values for all three. Finally, you press the [Copy] button under the chromatogram, and paste the chromatogram into a document. Before you do all this, remember to verify that you have the required resolution by using Equation 5. To help you determine visually whether you have enough resolution, examples of peak resolutions of 1.0, 1.5 and 2 are given in Figure 3. R s = 1.0 R s = 1.5 R s = 2.0 Figure 3: Examples of different resolutions Reporting Introduction Give a brief explanation of the problem and explain the strategy that you followed to solve it. List the applied equations for calculation of resolution (Equations 5 and 6) and explain them. Experimental Report the following parameters for the solution that you found: Results The choice of solvent B (methanol or acetonitrile). The solvent B fraction. The temperature The flow rate The particle size Include the copy of the chromatogram for the optimized conditions and report the retention time for the last compound. For the two most closely eluting peaks you should report the following parameters: Retention time Peak width (σ or w b ) Retention factor Calculated plate number, N For these two peaks you should prove that the criterion of R s 1.5 is fulfilled by applying Equation 5. Report the value with two decimals. You should also report the backpressure to prove that the other criterion is fulfilled. In addition you should calculate the resolution for the two peaks by Equation 6. For the plate number, N, you apply the average for the two peaks. The Purnell equation is an estimate that assumes for instance equal peak widths of the two peaks, while equation 5 defines resolution. You may get slightly different values from the two equations and it is therefore acceptable that R s calculated by equation 6 is slightly below 1.5, as long as Equation 5 gives an acceptable value. 7

8 Appendix Circumventing that application is blocked by Java When starting the HPLC simulator you may get an error message saying that the application is blocked by Java Security. See Figure 4. Figure 4: Warning message from Java stating that Java Bindings for OpenGL is blocked To get around this problem, do the following: On Windows: Go to the Windows start menu, select [All programs], find the 'Java' folder, open it and run 'Configure Java'. Alternatively, go to the start button, type 'java' in the 'search programs and files' field and start the 'Configure Java' application. On Mac: Open System preferences. Select Java. In the 'Java control panel', choose 'Security'. You should get a window similar to the one below. Ensure that Security level is set to 'High' Press [Edit Site List] nex to the list of exceptions. Press [Add] in the new window. Add and to the list and press [OK] followed by [Continue]. Ensure that the web adresses are added to the exception list, as shown in Figure 5. Press [Apply] followed by [OK]. Try to launch the HPLC simulator again. 8

9 Figure 5: and added to the exception site list.. 9

Exercise in gas chromatography

Exercise in gas chromatography AM0925 Exercise in gas chromatography Introduction The purpose of this exercise is to investigate the effect of different carrier gases (helium and nitrogen) on the Golay equation and the optimal carrier

More information

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency Luna 2.5 µm C18(2)-HST Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency Table of Contents Part 1 Theory 1.1 Abstract...3 1.2 Introduction...3 Part 2 Set Up 2.1

More information

[S016. CHROMATOGRAPHY]

[S016. CHROMATOGRAPHY] Phyto-Analysis Sheet Number : 16 Prof. Dr. Talal Aburjai Page 1 of 9 How to read the chromatogram? Comes from any automated chromatography. The chromatograms show the 0 t (t m ) which indicates the solvent

More information

2] The plate height in chromatography is best described as 2

2] The plate height in chromatography is best described as 2 9 Chromatography. General Topics 1] Explain the three major components of the van Deemter equation. Sketch a clearly labeled diagram describing each effect. What is the salient point of the van Deemter

More information

Analytical Chemistry

Analytical Chemistry Analytical Chemistry Chromatographic Separations KAM021 2016 Dr. A. Jesorka, 6112, aldo@chalmers.se Introduction to Chromatographic Separations Theory of Separations -Chromatography Terms Summary: Chromatography

More information

HPLC Background Chem 250 F 2008 Page 1 of 24

HPLC Background Chem 250 F 2008 Page 1 of 24 HPLC Background Chem 250 F 2008 Page 1 of 24 Outline: General and descriptive aspects of chromatographic retention and separation: phenomenological k, efficiency, selectivity. Quantitative description

More information

High Speed vs. High Resolution Analyses in HPLC: A Critical Performance Comparison of Column Options Using Poppe and Kinetic Plots

High Speed vs. High Resolution Analyses in HPLC: A Critical Performance Comparison of Column Options Using Poppe and Kinetic Plots High Speed vs. High Resolution Analyses in HPLC: A Critical Performance Comparison of Column Options Using Poppe and Kinetic Plots Jason C. Link, William E. Barber, William Long, Anne Mack Agilent Technologies

More information

Chromatographic Separation

Chromatographic Separation What is? is the ability to separate molecules using partitioning characteristics of molecule to remain in a stationary phase versus a mobile phase. Once a molecule is separated from the mixture, it can

More information

Biochemistry. Biochemical Techniques HPLC

Biochemistry. Biochemical Techniques HPLC Description of Module Subject Name Paper Name 12 Module Name/Title 13 1. Objectives 1.1. To understand the basic concept and principle of 1.2. To understand the components and techniques of 1.3. To know

More information

Easy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator

Easy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator Technical Note 75 Easy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator Introduction The goal of every chromatographic optimization is a method that sufficiently resolves all

More information

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS) Liquid Chromatography Mass Spectrometry (LC/MS) Prelab Questions: Questions to be answered before doing the experiment. The answers are due at the beginning of each experiment without exception (the questions

More information

Fall 2012 Due In Class Friday, Oct. 19. Complete the following on separate paper. Show your work and clearly identify your answers.

Fall 2012 Due In Class Friday, Oct. 19. Complete the following on separate paper. Show your work and clearly identify your answers. CHEM 322 Name Fall 2012 Due In Class Friday, Oct. 19 Complete the following on separate paper. Show your work and clearly identify your answers. General Separations 1. Describe the relative contributions

More information

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology Gas Chromatography/Mass Spectrometry Author Harry Prest 1601 California Avenue

More information

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Application Note Authors Kevin McCann, Sameer Nene, Doug McIntyre, Edmond Neo, Dennis Nagtalon, and

More information

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Figure 1. Structures for Vitamin B2 and Vitamin B1. CH 461 & CH 461H F 18 Name Experiment 2C Integrated Laboratory Experiment DETERMINATION OF RIBOFLAVIN: A COMPARISON OF TECHNIQUES PART C. HIGH PERFORMANCE LIQUID CHROMATOGRAPHY The purpose of this experiment

More information

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India Effective use of pharmacopeia guidelines to reduce cost of chromatographic analysis Optimized, cost-effective HPLC analysis of atorvastatin by varying column dimensions within the USP allowed limts

More information

Open Column Chromatography, GC, TLC, and HPLC

Open Column Chromatography, GC, TLC, and HPLC Open Column Chromatography, GC, TLC, and HPLC Murphy, B. (2017). Introduction to Chromatography: Lecture 1. Lecture presented at PHAR 423 Lecture in UIC College of Pharmacy, Chicago. USES OF CHROMATOGRAPHY

More information

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents Marco Karsten, Bas, Dolman, Giovanni Maio, Frank Steiner, Holger Franz, Frank Arnold and Remco Swart LC Packings,

More information

Signal Processing First Lab 11: PeZ - The z, n, and ˆω Domains

Signal Processing First Lab 11: PeZ - The z, n, and ˆω Domains Signal Processing First Lab : PeZ - The z, n, and ˆω Domains The lab report/verification will be done by filling in the last page of this handout which addresses a list of observations to be made when

More information

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump Performance characteristics of the Agilent 129 Infinity Quaternary Pump Technical Overview Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract This Technical Overview presents Proof

More information

Introduction to Capillary GC

Introduction to Capillary GC Introduction to Capillary GC LC Columns and Consumables Simon Jones Chromatography Applications Engineer February 20, 2008 Page 1 Introduction to Capillary GC t r K c?? Kβ k = - tr t m? t m R s Page 2

More information

How to Troubleshoot a Failed System Suitability Test

How to Troubleshoot a Failed System Suitability Test Volume 34 Number 12, 881 932 Volume 34 Number 12 December 216 www.chromatographyonline.com LCGC NORTH AMERICA How to Troubleshoot a Failed System Suitability Test December 216 Preventing Carryover in GC

More information

Introduction to Chromatographic Separations

Introduction to Chromatographic Separations Introduction to Chromatographic Separations Analysis of complex samples usually involves previous separation prior to compound determination. Two main separation methods instrumentation are available:

More information

High Performance Liquid Chromatography

High Performance Liquid Chromatography High Performance Liquid Chromatography What is HPLC? It is a separation technique that involves: Injection of small volume of liquid sample Into a tube packed with a tiny particles (stationary phase).

More information

Basic Principles for Purification Using Supercritical Fluid Chromatography

Basic Principles for Purification Using Supercritical Fluid Chromatography Basic Principles for Purification Using Supercritical Fluid Chromatography Jo-Ann M. Jablonski, Christopher J. Hudalla, Kenneth J. Fountain, Steven M. Collier, and Damian Morrison Waters Corporation, Milford,

More information

Using Microsoft Excel

Using Microsoft Excel Using Microsoft Excel Objective: Students will gain familiarity with using Excel to record data, display data properly, use built-in formulae to do calculations, and plot and fit data with linear functions.

More information

Introduction to Chromatography

Introduction to Chromatography Introduction to Chromatography Dr. Sana Mustafa Assistant Professor Department of Chemistry, Federal Urdu University of Arts, Science & Technology, Karachi. What is Chromatography? Derived from the Greek

More information

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Figure 1. Structures for Vitamin B2 and Vitamin B1. CH 461 & CH 461H F 15 Name Experiment 2C Integrated Laboratory Experiment DETERMINATION OF RIBOFLAVIN: A COMPARISON OF TECHNIQUES PART C. HIGH PERFORMANCE LIQUID CHROMATOGRAPHY The purpose of this experiment

More information

LC Technical Information

LC Technical Information LC Technical Information Method Transfer to Accucore.6 μm Columns Containing solid core particles, which are engineered to a diameter of.6μm and a very narrow particle size distribution; Accucore HPLC

More information

Chromatographic Analysis

Chromatographic Analysis Chromatographic Analysis Distribution of Analytes between Phases An analyte is in equilibrium between the two phases [S 1 ] [S 2 ] (in phase 1) (in phase 2) AS [S2 ] K 2 A S [S1 ] 1 AS, A 1 S Activity

More information

Chapter 26. An Introduction to Chromatographic Separations. Chromatography

Chapter 26. An Introduction to Chromatographic Separations. Chromatography Chapter 26 An Introduction to Chromatographic Separations Chromatography 1 Chromatography-Model as Extraction Chromatography-Model as Extraction 2 Chromatography Planar Chromatography-Types paper chromatography

More information

Automated method development using Agilent 1100 Series HPLC systems, Agilent ChemStation and ChromSword software. Application

Automated method development using Agilent 1100 Series HPLC systems, Agilent ChemStation and ChromSword software. Application Automated method development using Agilent 1100 Series HPLC systems, Agilent ChemStation and ChromSword software Application Sergey Galushko Ralf Honsberg Abstract The development of an HPLC method is

More information

Chromatography and other Separation Methods

Chromatography and other Separation Methods Chromatography and other Separation Methods Probably the most powerful class of modern analytical methods for analyzing mixture of components---and even for detecting a single component in a complex mixture!

More information

v Prerequisite Tutorials GSSHA WMS Basics Watershed Delineation using DEMs and 2D Grid Generation Time minutes

v Prerequisite Tutorials GSSHA WMS Basics Watershed Delineation using DEMs and 2D Grid Generation Time minutes v. 10.1 WMS 10.1 Tutorial GSSHA WMS Basics Creating Feature Objects and Mapping Attributes to the 2D Grid Populate hydrologic parameters in a GSSHA model using land use and soil data Objectives This tutorial

More information

LC and LC/MS Column Selection Flow Chart

LC and LC/MS Column Selection Flow Chart LC and LC/MS Column Selection Flow Chart To use the column selection diagram below, simply follow the path for your analyte and mobile phase. At the far right, follow your final column selection to the

More information

CHEM340 Tutorial 4: Chromatography

CHEM340 Tutorial 4: Chromatography CHEM340 Tutorial 4: Chromatography 1. The data in the table below was obtained from a chromatogram obtained with a 10 cm liquid chromatography column. Under the conditions used, the compound uracil is

More information

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011 ?? Kβ? Page 1 Typical GC System Gas supply Injector Detector Data handling GAS Column Oven Page 2 CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention

More information

Information Dependent Acquisition (IDA) 1

Information Dependent Acquisition (IDA) 1 Information Dependent Acquisition (IDA) Information Dependent Acquisition (IDA) enables on the fly acquisition of MS/MS spectra during a chromatographic run. Analyst Software IDA is optimized to generate

More information

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46>

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46> High Performance Liquid Chromatography HPLC-9 Ultra-high Speed Analysis of Ibuprofen within USP Allowed Limits by Nexera Method Scouting In recent years, high-throughput analytical techniques have

More information

LIQUID CHROMATOGRAPHY

LIQUID CHROMATOGRAPHY LIQUID CHROMATOGRAPHY RECENT TECHNIQUES HPLC High Performance Liquid Chromatography RRLC Rapid Resolution Liquid Chromatography UPLC Ultra Performance Liquid Chromatography UHPLC Ultra High Pressure Liquid

More information

NINE CHOICE SERIAL REACTION TIME TASK

NINE CHOICE SERIAL REACTION TIME TASK instrumentation and software for research NINE CHOICE SERIAL REACTION TIME TASK MED-STATE NOTATION PROCEDURE SOF-700RA-8 USER S MANUAL DOC-025 Rev. 1.3 Copyright 2013 All Rights Reserved MED Associates

More information

Electric Fields and Equipotentials

Electric Fields and Equipotentials OBJECTIVE Electric Fields and Equipotentials To study and describe the two-dimensional electric field. To map the location of the equipotential surfaces around charged electrodes. To study the relationship

More information

Successfully Scaling and Transferring HPLC and UPLC Methods

Successfully Scaling and Transferring HPLC and UPLC Methods Successfully Scaling and Transferring HPLC and UPLC Methods Esa Lehtorinne Esa_Lehtorinne@waters.com Tel: +358-9-5659 6288 Fax: +358-9-5659 6282 Waters Finland Kutomotie 16 00380 Helsinki 2013 Waters Corporation

More information

Virtual Beach Making Nowcast Predictions

Virtual Beach Making Nowcast Predictions Virtual Beach 3.0.6 Making Nowcast Predictions In this module you will learn how to: A. Create a real-time connection to Web data services through EnDDaT B. Download real-time data to make a Nowcast prediction

More information

OECD QSAR Toolbox v.4.1. Tutorial illustrating new options for grouping with metabolism

OECD QSAR Toolbox v.4.1. Tutorial illustrating new options for grouping with metabolism OECD QSAR Toolbox v.4.1 Tutorial illustrating new options for grouping with metabolism Outlook Background Objectives Specific Aims The exercise Workflow 2 Background Grouping with metabolism is a procedure

More information

Phenyl-Hexyl. UHPLC Columns. Alternate, complementary selectivity to C18 and C8 bonded phases

Phenyl-Hexyl. UHPLC Columns. Alternate, complementary selectivity to C18 and C8 bonded phases Phenyl-Hexyl UHPLC Columns Alternate, complementary selectivity to C8 and C8 bonded phases Particularly recommended for compounds containing aromatic groups Excellent bonded phase stability for durable,

More information

HPLC Praktikum Skript

HPLC Praktikum Skript HPLC Praktikum Skript Assistants: Gianluca Bartolomeo HCI D330, 3 46 68, bartolomeo@org.chem.ethz.ch Sahar Ghiasikhou HCI E330, 2 29 29, ghiasikhou@org.chem.ethz.ch 1. Introduction In chromatographic techniques,

More information

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments. Chromatography Primer Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments. At its heart, chromatography is a technique

More information

OECD QSAR Toolbox v.4.0. Tutorial on how to predict Skin sensitization potential taking into account alert performance

OECD QSAR Toolbox v.4.0. Tutorial on how to predict Skin sensitization potential taking into account alert performance OECD QSAR Toolbox v.4.0 Tutorial on how to predict Skin sensitization potential taking into account alert performance Outlook Background Objectives Specific Aims Read across and analogue approach The exercise

More information

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis Application Note AN89.V.1 Extraction of Methylmalonic Acid from Serum Using ISOLUTE SAX Page 1 Extraction of Methylmalonic Acid from Serum Using ISOLUTE SAX Prior to LC-MS/MS Analysis Sample Preparation

More information

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract. Separation of Explosives in EPA 833: Column Choices Optimize Speed, Resolution, and Solvent Use Application Environmental Authors John W. Henderson Jr. and William J. Long Agilent Technologies, Inc. 28

More information

Watershed Modeling Orange County Hydrology Using GIS Data

Watershed Modeling Orange County Hydrology Using GIS Data v. 10.0 WMS 10.0 Tutorial Watershed Modeling Orange County Hydrology Using GIS Data Learn how to delineate sub-basins and compute soil losses for Orange County (California) hydrologic modeling Objectives

More information

Understanding Your Spectra Module. Agilent OpenLAB CDS ChemStation Edition

Understanding Your Spectra Module. Agilent OpenLAB CDS ChemStation Edition Understanding Your Spectra Module Agilent OpenLAB CDS ChemStation Edition Notices Agilent Technologies, Inc. 1994-2012, 2013 No part of this manual may be reproduced in any form or by any means (including

More information

What is Chromatography?

What is Chromatography? What is Chromatography? Chromatography is a physico-chemical process that belongs to fractionation methods same as distillation, crystallization or fractionated extraction. It is believed that the separation

More information

OECD QSAR Toolbox v.4.1. Tutorial on how to predict Skin sensitization potential taking into account alert performance

OECD QSAR Toolbox v.4.1. Tutorial on how to predict Skin sensitization potential taking into account alert performance OECD QSAR Toolbox v.4.1 Tutorial on how to predict Skin sensitization potential taking into account alert performance Outlook Background Objectives Specific Aims Read across and analogue approach The exercise

More information

Introduction to Chromatographic Separations (Chapter 1) Many determinations involve separation followed by analysis chromatography electrophoresis

Introduction to Chromatographic Separations (Chapter 1) Many determinations involve separation followed by analysis chromatography electrophoresis Introduction to Chromatographic Separations (Chapter 1) Many determinations involve separation followed by analysis chromatography electrophoresis Chromatography: sample transported by mobile phase electrostatic

More information

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent Infinity LC System with Diode Array Detector Application Note Author Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany [mau]

More information

Analytical Technologies in Biotechnology Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Analytical Technologies in Biotechnology Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Analytical Technologies in Biotechnology Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Module - 3 Chromatographic methods Lecture - 2 Basic Concepts in Chromatography

More information

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates Using the Agilent 9 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates Technical Overview Author Florian Rieck Agilent Technologies, Inc. Waldbronn, Germany Abstract This Technical

More information

Using Tables and Graphing Calculators in Math 11

Using Tables and Graphing Calculators in Math 11 Using Tables and Graphing Calculators in Math 11 Graphing calculators are not required for Math 11, but they are likely to be helpful, primarily because they allow you to avoid the use of tables in some

More information

1 Introduction & Objective. 2 Warm-up. Lab P-16: PeZ - The z, n, and O! Domains

1 Introduction & Objective. 2 Warm-up. Lab P-16: PeZ - The z, n, and O! Domains DSP First, 2e Signal Processing First Lab P-6: PeZ - The z, n, and O! Domains The lab report/verification will be done by filling in the last page of this handout which addresses a list of observations

More information

New 5-micron HALO-5 columns based on Fused- Core particle technology boost the performance of HPLC. Compared to other HPLC columns, HALO-5 columns

New 5-micron HALO-5 columns based on Fused- Core particle technology boost the performance of HPLC. Compared to other HPLC columns, HALO-5 columns New 5-micron HALO-5 columns based on Fused- Core particle technology boost the performance of HPLC. Compared to other HPLC columns, HALO-5 columns have: the highest plate number versus any other 5-micron

More information

Tutorial. Getting started. Sample to Insight. March 31, 2016

Tutorial. Getting started. Sample to Insight. March 31, 2016 Getting started March 31, 2016 Sample to Insight CLC bio, a QIAGEN Company Silkeborgvej 2 Prismet 8000 Aarhus C Denmark Telephone: +45 70 22 32 44 www.clcbio.com support-clcbio@qiagen.com Getting started

More information

BCMB/CHEM 8190 Lab Exercise Using Maple for NMR Data Processing and Pulse Sequence Design March 2012

BCMB/CHEM 8190 Lab Exercise Using Maple for NMR Data Processing and Pulse Sequence Design March 2012 BCMB/CHEM 8190 Lab Exercise Using Maple for NMR Data Processing and Pulse Sequence Design March 2012 Introduction Maple is a powerful collection of routines to aid in the solution of mathematical problems

More information

Liquid Chromatography

Liquid Chromatography Liquid Chromatography 1. Introduction and Column Packing Material 2. Retention Mechanisms in Liquid Chromatography 3. Method Development 4. Column Preparation 5. General Instrumental aspects 6. Detectors

More information

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter Luisa Pereira, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20755 Key Words Solid core, fused core, superficially

More information

Chapter 26: An Introduction to Chromatographic Separations

Chapter 26: An Introduction to Chromatographic Separations Chapter 26: An Introduction to Chromatographic Separations Column Chromatography Migration Rates Distribution Contstants Retention Times Selectivity Factor Zone Broadening & Column Efficiency Optimizing

More information

TALLINN UNIVERSITY OF TECHNOLOGY, INSTITUTE OF PHYSICS 6. THE TEMPERATURE DEPENDANCE OF RESISTANCE

TALLINN UNIVERSITY OF TECHNOLOGY, INSTITUTE OF PHYSICS 6. THE TEMPERATURE DEPENDANCE OF RESISTANCE 6. THE TEMPERATURE DEPENDANCE OF RESISTANCE 1. Objective Determining temperature coefficient of metal and activation energy of self-conductance of semiconductor sample. 2. Equipment needed Metal and semiconductor

More information

Changing the way you think about HPLC

Changing the way you think about HPLC 3521 Silverside Road Suite 1-K Quillen Building Wilmington, DE 19810 USA info@advanced-materials-tech.com Changing the way you think about HPLC Changing the way you think about HPLC New 5-micron HALO-5

More information

LECTURE 2. Advanced Separation Science Techniques Present and Future Separation Tools

LECTURE 2. Advanced Separation Science Techniques Present and Future Separation Tools LECTURE 2 Advanced Separation Science Techniques Present and Future Separation Tools Jack Henion, Ph.D. Emeritus Professor, Analytical Toxicology Cornell University Ithaca, NY 14850 Lecture 2, Page 1 Contents

More information

CHEM 429 / 529 Chemical Separation Techniques

CHEM 429 / 529 Chemical Separation Techniques CHEM 429 / 529 Chemical Separation Techniques Robert E. Synovec, Professor Department of Chemistry University of Washington Lecture 1 Course Introduction Goal Chromatography and Related Techniques Obtain

More information

Fast methods for the determination of ibuprofen in drug products

Fast methods for the determination of ibuprofen in drug products APPLICATION NOTE 779 Fast s for the determination of ibuprofen in drug products Authors Sylvia Grosse, Mauro De Pra, Frank Steiner, Thermo Fisher Scientific, Germering, Germany Keywords Pharmaceutical,

More information

Patrick: An Introduction to Medicinal Chemistry 5e MOLECULAR MODELLING EXERCISES CHAPTER 17

Patrick: An Introduction to Medicinal Chemistry 5e MOLECULAR MODELLING EXERCISES CHAPTER 17 MOLECULAR MODELLING EXERCISES CHAPTER 17 Exercise 17.6 Conformational analysis of n-butane Introduction Figure 1 Butane Me Me In this exercise, we will consider the possible stable conformations of butane

More information

Chromatography Outline

Chromatography Outline Chem 2001 Summer 2004 Outline What is? The Chromatogram Optimization of Column Performance Why Do Bands Spread? Gas High-Performance Liquid Ion-Exchange 2 What is? In chromatography, separation is achieved

More information

Separation Methods Based on Distributions in Discrete Stages (02/04/15)

Separation Methods Based on Distributions in Discrete Stages (02/04/15) Separation Methods Based on Distributions in Discrete Stages (02/04/15) 1. Chemical Separations: The Big Picture Classification and comparison of methods 2. Fundamentals of Distribution Separations 3.

More information

Skyline Small Molecule Targets

Skyline Small Molecule Targets Skyline Small Molecule Targets The Skyline Targeted Proteomics Environment provides informative visual displays of the raw mass spectrometer data you import into your Skyline documents. Originally developed

More information

Getting started with BatchReactor Example : Simulation of the Chlorotoluene chlorination

Getting started with BatchReactor Example : Simulation of the Chlorotoluene chlorination Getting started with BatchReactor Example : Simulation of the Chlorotoluene chlorination 2011 ProSim S.A. All rights reserved. Introduction This document presents the different steps to follow in order

More information

High Temperature Ultra Fast Liquid Chromatography. Peter W. Carr, Jon Thompson, Dwight Stoll, and Adam Schallinger

High Temperature Ultra Fast Liquid Chromatography. Peter W. Carr, Jon Thompson, Dwight Stoll, and Adam Schallinger High Temperature Ultra Fast Liquid Chromatography Peter W. Carr, Jon Thompson, Dwight Stoll, and Adam Schallinger Conclusions 1. To do fast LC, use a WEAK eluent and a HOT column.. Use a highly retentive

More information

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput MAY 2005 SEPARATION SCIENCE REDEFINED 31 Assay Transfer from HPLC to UPLC for Higher Analysis Throughput A typical HPLC assay was transferred and optimized for a Waters ACQUITY UPLC system to achieve both

More information

Chemistry Instrumental Analysis Lecture 26. Chem 4631

Chemistry Instrumental Analysis Lecture 26. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 26 Rate Theory Focuses on the contributions of various kinetic factors to zone or band broadening. Column Dispensivity, H, is assumed to be the sum of the individual

More information

DSP First. Laboratory Exercise #10. The z, n, and ˆω Domains

DSP First. Laboratory Exercise #10. The z, n, and ˆω Domains DSP First Laboratory Exercise #10 The z, n, and ˆω Domains 1 Objective The objective for this lab is to build an intuitive understanding of the relationship between the location of poles and zeros in the

More information

Introduction to Computer Tools and Uncertainties

Introduction to Computer Tools and Uncertainties Experiment 1 Introduction to Computer Tools and Uncertainties 1.1 Objectives To become familiar with the computer programs and utilities that will be used throughout the semester. To become familiar with

More information

How proteins separate on reverse-phase HPLC

How proteins separate on reverse-phase HPLC 1 Reverse Phase How proteins separate on reverse-phase HPLC RP chromatography separates proteins through the interaction of the hydrophobic foot of the protein with a nonpolar surface of the particle RP

More information

Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks

Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks bjective: The purpose of this experiment is to quantify the caffeine content of a diet soda sample using Ultra-High Performance

More information

M > ACN > > THF

M > ACN > > THF Method Development in HPLC Dr. Amitha Hewavitharana School of Pharmacy University of Queensland Method Development in HPLC References: D A Skoog Principles of instrumental analysis, 3 rd Edition Chapters

More information

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure.

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure. High performance liquid chromatography (HPLC) is a much more sensitive and useful technique than paper and thin layer chromatography. The instrument used for HPLC is called a high performance liquid chromatograph.

More information

2. a) R N and L N so R L or L R 2.

2. a) R N and L N so R L or L R 2. 1. Use the formulae on the Some Key Equations and Definitions for Chromatography sheet. a) 0.74 (remember that w b = 1.70 x w ½ ) b) 5 c) 0.893 (α always refers to two adjacent peaks) d) 1.0x10 3 e) 0.1

More information

HPLC. GRATE Chromatography Lab Course. Dr. Johannes Ranke. September 2003

HPLC. GRATE Chromatography Lab Course. Dr. Johannes Ranke. September 2003 HPLC GRATE Chromatography Lab Course Dr. Johannes Ranke Organisation The groups Start at 9:00 am End at 18:00 pm at the latest Friday, 19th we will finish at 2:00 pm Thursday, 11th: Lecture at 08:15 am

More information

Mnova Software for Analyzing Reaction Monitoring NMR Spectra

Mnova Software for Analyzing Reaction Monitoring NMR Spectra Mnova Software for Analyzing Reaction Monitoring NMR Spectra Version 10 Chen Peng, PhD, VP of Business Development, US & China Mestrelab Research SL San Diego, CA, USA chen.peng@mestrelab.com 858.736.4563

More information

Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions

Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions Today, ultra-high-performance liquid chromatography (UHPLC) has taken a firm foothold in the analytical laboratory.

More information

An area chart emphasizes the trend of each value over time. An area chart also shows the relationship of parts to a whole.

An area chart emphasizes the trend of each value over time. An area chart also shows the relationship of parts to a whole. Excel 2003 Creating a Chart Introduction Page 1 By the end of this lesson, learners should be able to: Identify the parts of a chart Identify different types of charts Create an Embedded Chart Create a

More information

Transferring Yesterdays Methods to Tomorrows Technology: The Evolution from HPLC to UPLC TM

Transferring Yesterdays Methods to Tomorrows Technology: The Evolution from HPLC to UPLC TM Transferring Yesterdays Methods to Tomorrows Technology: The Evolution from HPLC to UPLC TM Eric S. Grumbach Thomas E. Wheat Chuck Phoebe Joe Arsenault Jeffrey R. Mazzeo Diane M. Diehl UPLC TECHNOLOGY

More information

OECD QSAR Toolbox v.3.3. Step-by-step example of how to build a userdefined

OECD QSAR Toolbox v.3.3. Step-by-step example of how to build a userdefined OECD QSAR Toolbox v.3.3 Step-by-step example of how to build a userdefined QSAR Background Objectives The exercise Workflow of the exercise Outlook 2 Background This is a step-by-step presentation designed

More information

MassHunter TOF/QTOF Users Meeting

MassHunter TOF/QTOF Users Meeting MassHunter TOF/QTOF Users Meeting 1 Qualitative Analysis Workflows Workflows in Qualitative Analysis allow the user to only see and work with the areas and dialog boxes they need for their specific tasks

More information

Spring 2010 updated March 24 Determination of Aromatics in Gasoline by Gas Chromatography- Mass Spectrometry: Comparison of Grades and Brands

Spring 2010 updated March 24 Determination of Aromatics in Gasoline by Gas Chromatography- Mass Spectrometry: Comparison of Grades and Brands Chem F413W Spring 2010 updated March 24 Determination of Aromatics in Gasoline by Gas Chromatography- Mass Spectrometry: Comparison of Grades and Brands Introduction Gasoline is a complex mixture of volatile

More information

Agilent ESI and APCI sources: for polar to non-polar compounds

Agilent ESI and APCI sources: for polar to non-polar compounds 1 Agilent 6400 Series Triple Quadrupole Users Workshop 1 Agilent ESI and APCI sources: for polar to non-polar compounds Nebulizer Pressure Corona current Nebulizer Pressure Vaporizer Vcap Vcap Drying Gas

More information

CHAPTER 1. Introduction, Chromatography Theory, and Instrument Calibration

CHAPTER 1. Introduction, Chromatography Theory, and Instrument Calibration 1 1 1 1 1 1 CHAPTER 1 Introduction, Chromatography Theory, and Instrument Calibration 1.1 Introduction Analytical chemists have few tools as powerful as chromatography to measure distinct analytes in complex

More information

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India Reducing analysis time and solvent consumption for isocratic USP assay methods with current and proposed USP guidelines using the Agilent 129 Infinity LC System An efficient way to reduce cost of analysis

More information

Theory and Instrumentation of GC. Chromatographic Parameters

Theory and Instrumentation of GC. Chromatographic Parameters Theory and Instrumentation of GC Chromatographic Parameters i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this

More information

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India QbD Based Method Development on an Agilent 129 Infinity UHPLC System Combined with a Seamless Method Transfer to HPLC Using Intelligent System Emulation Technology Application Note Pharmaceutical QA/QC

More information