NEUTRON imaging is a technique for the non-destructive

Size: px
Start display at page:

Download "NEUTRON imaging is a technique for the non-destructive"

Transcription

1 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE Investigating Phase Contrast Neutron Imaging for Mixed Phase-Amplitude Objects Kaushal K. Mishra and Ayman I. Hawari, Member, IEEE Abstract Phase contrast imaging is an imaging modality that has been extensively applied in X-ray imaging and was demonstrated using neutrons over the past few years. In this case, contrast in the image, especially at edges, is enhanced due to phase shifts that take place as the neutron wave passes through regions in the sample that differ in the coherent scattering length density. Usually, a pure phase object approximation is used to formulate the problem, whereas realistic samples represent mixed phase-amplitude objects. In this work, a formulation for mixed phase-amplitude objects with moderate neutron attenuation coefficients and its effect on the neutron image is presented. A computational simulation technique has been devised to study this effect on different types of samples. Using simulations, it is observed that the pure phase object approximation results in over enhancement of edges for a phase-amplitude object, with significant variation (in the case of neutron imaging) depending upon the edge forming material characteristics. The total contrast for the mixed phase-amplitude object is less than the sum of the individual attenuation and phase contrast components. The difference depends on the scalar product of the gradient of the coherent scattering length density and the attenuation coefficient. The presented formulation can aid in predicting and optimizing the performance characteristics of neutron phase contrast imaging experiments. Index Terms Neutron imaging, phase-amplitude objects, phase contrast. I. INTRODUCTION NEUTRON imaging is a technique for the non-destructive testing (NDT) of materials. It is characterized by its sensitivity to low atomic number (Z) elements such as hydrogenous materials. Therefore, it represents a modality that is complementary to more traditional techniques such as X-ray imaging. In recent years, to enhance contrast in the images taken with neutrons, the use of the wave-particle dual nature of neutrons in imaging was demonstrated in what is known as Phase Contrast Neutron Imaging [1] [3]. This is an imaging modality that has been extensively applied in X-ray imaging and electron microscopy [4] [6]. In this case, contrast formation in the image, especially at edges, is enhanced due to phase shifts that take place as the neutron wave passes through the regions in the sample that differ in the coherent scattering length density. The advantage which this technique offers over the usual diffraction imaging techniques is that there is no requirement for the beam to be monochromatic and only spatial coherence is required. Obtaining a monochromatic neutron beam usually requires use of Manuscript received September 11, 2008; revised December 05, Current version published June 17, This work was supported by the U.S. Department of Energy under Grant DE-FG07-03ID The authors are with the Department of Nuclear Engineering, North Carolina State University, Raleigh, NC USA ( kkmishra@ncsu.edu; ayman. hawari@ncsu.edu). Digital Object Identifier /TNS monochromators, like bent sapphire or silicon crystals, or beam choppers and leads to significant reduction in neutron beam intensity. Thus, phase contrast neutron imaging offers a simple and elegant alternative for obtaining the higher contrast needed in the neutron images of samples. Typically, this technique is presented using formulations that imply its suitability for achieving contrast in images of almost pure phase objects (i.e., objects with negligible neutron attenuation). The technique is formulated using the transport of intensity equation and the diffraction propagation function [7] [9]. However, edge enhancement among moderately attenuating objects with similar attenuation coefficients is also desirable and possible. In this work, the pure phase contrast formulation is extended to describe moderately neutron attenuating objects. These objects are referred to as mixed phase-amplitude objects. The extension of the formulation to these objects requires consideration of a phase-amplitude interaction term which mathematically is of the same order of magnitude as the pure phase contrast contributing term. This extension provides a more accurate description of the image contrast formation process as well as extends its applicability to various different objects which are not pure phase. A similar relationship in the context of X-ray phase contrast imaging has been obtained, but the phase-amplitude interaction term was neglected assuming that the variation of the attenuation coefficient is insignificant in most of the X-ray samples [9]. A technique for computationally simulating the images using the obtained formulation has also been developed. Simulations of phase contrast neutron images of various relevant materials forming different edges in the artificially designed phantoms was performed using this technique in order to study their numerical significance in characterizing the edge contrast in neutron images. This investigation can lead to more accurate understanding of edge effects in a phase contrast neutron image. II. PHASE IMAGE FORMULATION A. Mathematical Formulation For a plane neutron wave incident on a thin pure phase object the intensity distribution at a distance from the object plane is given by where is the intensity at the contact plane before the object, is the wavelength of the neutron beam and is the real part of the refractive index that depends on the coherent scattering length densities of the materials present in the object and the wavelength of the beam [10]. A schematic of the setup is shown (1) /$ IEEE

2 1630 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE 2009 B. Simulation Technique For computational simulations, (5) needs to be discritized over a grid and all the parameters involved in it need to be determined. The discretized Laplacian can be written as (6) Fig. 1. A schematic of phase contrast neutron imaging. in Fig. 1. The above equation is well known in the literature of phase contrast imaging. For a mixed phase-amplitude object the transmission function can be written as, where. Here, is the imaginary part of the refractive index that accounts for the attenuation of the neutron beam as it passes through the object. The intensity distribution at a distance for this object can be obtained by convolving the transmission function with the Fresnel propagator given by where. The first order approximation of the obtained intensity distribution can be written as Comparing (3) with (1), the appearance of an additional term, which depends on the scalar product of the gradient of and, can be observed for mixed phase-amplitude objects. This additional term can be referred to as the phase-amplitude interaction effect. It vanishes as becomes zero for a pure phase object reducing (3) to (1). Furthermore, it can also be observed from (3) that the phase-amplitude interaction effect counters the pure phase induced intensity variation. In obtaining (3) it has been assumed that the functions and are at least twice differentiable with respect to and. Although in most of the practical cases we have an edge function which is not differentiable, the above formulation is applicable because the edge function is smoothed by the convolution with the point spread function thus making it differentiable. The final neutron intensity image obtained on the screen is given by For a polychromatic neutron beam and a material with varied composition an approximate approach to obtain the intensity image would be to use the expectation value taken over the neutron energy spectrum as (2) (3) (4) where and the gradient can be descritized using a central differencing scheme. The descritized Laplacian with becomes the regular second order central differencing scheme which is isotropic. Other values of make anisotropic. is a reasonable choice to make in the absence of any information on the anisotropy of in the sample. For a polychromatic neutron beam, calculation of the expected value requires information about the neutron energy spectrum at the image plane. The energy spectrum can be obtained, e.g., through radiation transport calculations. Details on performing such calculations have been already published [11]. These calculations must be performed with the pinhole geometry required for phase contrast imaging. The PSF can be measured either using a gadolinium edge or it can be obtained through a radiation transport calculation using a computer code such as MCNP [12]. The calculation should be performed at the same object to image-plane distance at which the images need to be simulated. C. Beam Requirements The requirement imposed on the neutron beam to obtain the phase induced intensity variation in the image as given by (3) is a high degree of transverse spatial coherence. On the other hand, high degrees of temporal and chromatic coherence are not required. The degree of spatial coherence is a measure of the correlation which exists between the phases of the radiation at two points and is measured by the visibility of fringe patterns in the image. is defined as where, and are the maximum and minimum intensities in the obtained fringe patterns. In the case of perfect coherence it is unity, and for complete incoherence it is zero. For a circular source of uniform intensity and diameter the degree of coherence is given by, where, is the wavelength of the neutron beam, is the first order Bessel function of the first kind, is the distance between the two points whose spatial coherence is of interest (located on the object), is the effective diameter of the source and is the object to image-plane distance. This defines the coherence patch. The radius of the first zero of the above Bessel function is taken as a measure of the width of this patch. This radius is given by (7) where is the expected value of [.] over all wavelengths. (5) (8) which is also referred to as the transverse coherence length (i.e. ). The extent of the observable phase contrast effect is given by the width of this coherence patch which is dependent on beam energy. From (8) it can be observed that

3 MISHRA AND HAWARI: INVESTIGATING PHASE CONTRAST NEUTRON IMAGING FOR MIXED PHASE-AMPLITUDE OBJECTS 1631 the spatial coherence will increase if the expected wavelength of the neutron beam spectrum is increased. Also, the transverse coherence length is directly proportional to the well-known nondimensional quantity of the beam. Upon the substitution of (from (8)) for, the expression is obtained. The degree of coherence for a circular aperture defined by the visibility thus becomes (9) From (9) it can be observed that the degree of spatial coherence does not depend on the wavelength of the beam or the source diameter. Further, it can also be observed from (9) that is dependent only on the ratio and thus for a given, (obtained using (8) or limited by the beam line length) to observe the coherence effect the object-to-image-plane distance must be large enough (within the Fresnel limit). For a thermal neutron beam the wavelength of the neutrons is of the order of. Thus, to achieve a of the order of micrometers (i.e. the order of detection resolution for neutron radiographs) is required which is much higher than the for conventional neutron imaging. To achieve such a high, the neutron source effective diameter d must be reduced to a pinhole. But, the available neutron flux at the image plane which determines the exposure time is inversely proportional to the square of the as given by (10) Thus, a delicate balance between these two counteracting requirements needs to be achieved. To achieve the optimum between maximizing the coherence length and the neutron flux radiation transport simulations can be used as a design tool. The formulation and the simulation technique presented in Sections II-A and II-B respectively will aid in the design by predicting the observable phase contrast effect for any particular object exposed for a fixed exposure time defined by the neutron flux resulting from a given design. Conversely, if the exposure time and hence the flux requirement is dictated by some other factors e.g. frame speed necessary to observe some particular dynamical effect, then the formulation along with the simulation technique can be used to study and anticipate the expected contrast enhancement. This can be used as a decision making tool for choosing this technique in any particular setting and for some specific application. III. PHASE IMAGE SIMULATION Using designed phantoms, simulations of neutron radiographs were performed in order to observe the contribution of each term in (3) separately in the combined neutron image. The characteristics of the neutron imaging beamline located at the North Carolina State University (NCSU) PULSTAR reactor were used for this purpose [11]. Monte Carlo radiation transport calculations were performed using MCNP5 [12] to obtain a of 1.47 (at 6 meter aperture-to-detector distance) and a reasonable neutron flux using a 0.5 mm diameter gadolinium aperture and a 3 inch long sapphire filter. Sapphire acts as a fast neutron filter [13] and thus, helps in increasing the average wavelength of the beam, thereby increasing the transverse Fig. 2. The neutron energy spectrum (d=d(ln(e)) at the image plane with a 3 inch single crystal sapphire filter. coherence length. The beam divergence was reduced from 2 degrees to 0.5 degrees. This provides a beam size close to 10 cm at the image plane. The wavelength used in the simulations is the average wavelength of the PULSTAR imaging beam and is 1.3. The associated energy spectrum is shown in Fig. 2. For such conditions, the exposure time at the PULSTAR reactor is expected to be between three to four hours depending upon the type of sample. The phantoms were designed to depict various edges that are possible in objects characterized by the difference in the transmission functions of the edge forming materials. Phantom 1 was selected to include various engineering, structural, ornamental, nuclear and aerospace materials with moderate but almost similar attenuation contrast. For example, steel is a common structural material, titanium is an aerospace material, nickel-silver composites are used for electrical contacts and electrodes, zirconium and zircaloy is used in nuclear fuel cladding. It is these types of materials on which non destructive testing and non-invasive measurements need to be performed through imaging to locate possible cracks, voids, delaminations in composites, corrosion sites, misalignment etc. Precisely locating the edges constitutes an important step in performing any of the above tasks using the obtained images. Phantom 2 was selected to include organic and biological materials to investigate phase contrast enhancement in such samples. These materials are either present in plants or animal bodies or are very similar to them. For example, ICRP soft tissue and ICRP bone are equivalent to biological tissue and bone, sapphire specks are used in phantoms to simulate micro-calcifications, nylon fibers are used to simulate fibrous structures, acrylic is used to simulate breast tissues [14], urea is used as fertilizer for plants etc. The geometry of the phantoms and the material composition are shown in Fig. 3 and Table I, respectively. To simulate the neutron images of the phantoms, the values were calculated at different object to image plane distances assuming a 0.5 mm diameter point source and a beam divergence of 0.5. The attenuation radiograph simulation of the selected phantoms were performed using

4 1632 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE 2009 For each image simulation particles were run in four cycles with each cycle run on 24 processors for two hours. To further improve the counting statistics pinhole detectors were used to bring the relative error in every pixel inside the field of view of the pinhole down to 0.6%. The magnification of the attenuation images were obtained accurately using the Sobel edge detection algorithm [15]. The calculated magnification and the obtained PSFs were used to obtain the pure phase and the phase-amplitude interaction images of the corresponding phantom at the same object to image plane distance. From (3) it is clear that the final image is the algebraic sum of these three images. The coherent scattering length densities for different materials required for simulating phase images were calculated using the data obtained from the National Institute of Standards and Technology (NIST) database [16]. For the purpose of simulation was used in (6) to approximate the Laplacian performing isotropic filtering. A pixel size of 50 was selected in the image simulation. A. Simulation Results Fig. 3. The geometry of (a) phantom 1 (b) phantom 2. Table I gives the material composition of the phantoms. TABLE I THE MATERIAL COMPOSITION OF THE SIMULATED NEUTRON RADIOGRAPHY PHANTOMS. THE GEOMETRY OF THE PHANTOMS IS SHOWN IN Fig. 3 MCNP5 at the same object to image plane distances. The simulations were executed on a Linux cluster available in the High Performance Computing (HPC) Center at NCSU. The cluster has 612 dual Xeon nodes. Each node has two Xeon processors (mix of single-, dual- and quad-core) that have 2 Gigabytes (GBs) of memory per core and GBs of disk-space. The simulated radiographs of Phantom 1 are shown in Fig. 4. All the radiographs are on the same gray scale. Figs. 4(a), 4(b) and 4(c) depict the attenuation only, pure phase effect and phase-amplitude interaction effect radiographs respectively corresponding to the first, second and the third terms in (3). Fig. 4(d) depicts the total phase effect which is the algebraic sum of the pure phase and the phase-amplitude interaction effects. Figs. 4(e) and 4(f) depict the radiographs of the object considering it as a pure phase object and as a mixed phase-amplitude object respectively. Clearly, the attenuation only radiograph does not provide enough contrast for all the edges of the object to be distinguishable in the image. As is clear from the phase images, the phase effect does not provide different intensities in different materials which may be important for obtaining material makeup information of the object but, it enhances the edges between the materials providing the geometry information of the object. Comparing Fig. 4(b) and Fig. 4(c), it can be observed that the phase-amplitude interaction term provides less contrast than the pure phase term but still is not negligible. This is because the pure phase term involves a higher order derivative which increases the frequency and amplitude of intensity oscillation and hence higher contrast on the gray scale is visible. Also, the magnitude of the phase-amplitude interaction effect depends on the difference in the coherent scattering length densities of the edge forming materials. For example, the nickel-silver edge has more interaction effect than other edges even with a little difference in their attenuation coefficients. The difference in the contrast due to these interaction effects can also be observed when Fig. 4(e) and Fig. 4(f) are compared. This indicates that for some material edges the interaction effect (i.e., third term in (3)) can become really important. Comparing Fig. 4(a) and Fig. 4(f), it can be observed that features like the crack and material intrusions, which were not visible in the attenuation image due to similar attenuation coefficients, become visible in the phase enhanced image because of the difference in their scattering length densities.

5 MISHRA AND HAWARI: INVESTIGATING PHASE CONTRAST NEUTRON IMAGING FOR MIXED PHASE-AMPLITUDE OBJECTS 1633 Fig. 4. Phantom 1 simulation results: (a) normal attenuation radiograph (1st term in Eq. (3)) (b) the pure phase image (2nd term in Eq. (3)) (c) the phase amplitude interaction image (3rd term in Eq. (3)) (d) the phase image with phase-amplitude interaction which is the difference of (b) and (c) as given by Eq. (3) (e) phase contrast radiograph with pure phase approximation (1st and 2nd term in Eq. (3)) (f) Phase contrast radiograph with phase-amplitude interaction (all terms in Eq. (3)). The object to detector distance R=80cm. The grayscale for all six images is the same. Figs. 5(a) and 5(b) depict the normalized intensity profiles taken across the horizontal slice a-a and the vertical slice b-b that are shown in Figs. 4(e) and 4(f). These figures illustrate the sharp intensity changes at the edges because of the phase effect. The quantitative effect of the phase-amplitude interaction term can also be observed in these figures. A significant interaction effect appears at the nickel-silver edge in the profiles. In addition, an edge overlap effect arises in the profiles due to the spread

6 1634 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE 2009 Fig. 5. An intensity profile taken across (a) slice a-a of Fig. 4(e) and Fig. 4(f) and (b) slice b-b of Fig. 4(e) and Fig. 4(f). The profiles compare the pure phase and the mixed phase-amplitude approximations. The circled regions show the edge overlap effect. (c) The shape of the pure phase term with odd symmetry about a material edge in a sample, the phase-amplitude interaction term with even symmetry about the edge in the sample and their difference which is asymmetric about the edge. of the contrast from adjacent silver edges as silver intrusion in the nickel is considerably thin. Moreover, it can be observed that at the edges the mixed phase-amplitude interaction image profile has a higher maximum and a lower minimum, which is due to the asymmetry introduced by the interaction term in (3). This observation can be interpreted with the aid of Fig. 5(c). This figure shows the shape of a generic intensity variation introduced by the pure phase and the phase-amplitude interaction terms of (3) for a material edge in a given sample. As it can be seen, the pure phase consideration leads to an intensity redistribution that has odd symmetry about the mean intensity at the edge. The minimum appears in the material with higher neutron Fig. 6. Phantom 2 simulation results: (a) normal attenuation radiograph (b) phase contrast radiograph with pure phase approximation (c) phase contrast radiograph with phase-amplitude interaction. The object to detector distance R=50cm. scattering length density. The phase-amplitude interaction term leads to an intensity profile with even symmetry about the edge. Thus, the difference of these two terms (as given in (3)) offsets the intensity symmetry by adding intensity on one side and subtracting on the other side of the edge, which results in the profile shown in Fig. 5(c). The simulated radiographs of Phantom 2 are shown in Fig. 6 and Fig. 7. Figs. 6(a), 6(b) and 6(c) depict the attenuation only radiograph, pure phase approximation radiograph and mixed phase-amplitude approximation radiograph respectively. The gray scale of the radiographs is also shown along with the figures. Fig. 6(a) clearly indicates that there is small attenuation contrast present in the sample except at the sapphire specks. On the other hand, Figs. 6(b) and 6(c) indicate the presence of significant phase contrast in these organic and biological materials

7 MISHRA AND HAWARI: INVESTIGATING PHASE CONTRAST NEUTRON IMAGING FOR MIXED PHASE-AMPLITUDE OBJECTS 1635 nevertheless the combined phase-amplitude analysis will always perform better. Phantom 2 constitutes an example where there are some material edges present which have significant phase-amplitude interaction and other edges where it is not significant. In such samples the interaction term will decrease the contrast at those edges that already have high attenuation contrast and thus virtually expand the gray scale for the lesser contrast edges to become more prominent in the digital image. This effect can be observed in Fig. 7(c) where the sapphire edges are suppressed by the interaction term to make other edges more prominent. Thus, the phase-amplitude interaction term can increase the contrast at those edges where it is needed by providing rescaling of the gray scale in the digital images. IV. CONCLUSION Fig. 7. Phantom 2 simulation results: (a) phase image using pure phase approximation (b) phase-amplitude interaction image (c) the phase image with phase-amplitude interaction which is the difference of (a) and (b) as given by Eq. (3). The object to detector distance R=50cm. The images were normalized between zero and one. especially between ICRP soft tissue and urea, which was invisible in attenuation contrast. The presence of such a high phase contrast among these materials makes neutron phase contrast imaging suited for imaging biological and organic samples. Observing the phase images obtained for Phantom 2 in Fig. 7 it is clear that little phase-amplitude interaction is present in these biological materials except at the sapphire-tissue edge. At this edge the interaction is large because of the comparatively large difference in the attenuation coefficient and not so much due to the large difference in the neutron scattering length densities as was the case for the nickel-silver edge in Phantom 1. The effect of this interaction is also clearly visible in the final phase image in Fig. 7(c). Therefore, for these biological materials it can be concluded that for thin samples the pure phase approximation works fine, In the present work a phase contrast formulation for mixed phase-amplitude objects was presented. A phase-amplitude interaction term, which depends on the scalar product of the gradients of the real and imaginary part of the transmission functions of the edge forming materials, needs to be included to extend the pure phase object approximation to mixed phase-amplitude objects. Also, it was observed that the phase-amplitude interaction effect counters the edge contrast effect provided by the pure phase contribution and leads to an asymmetry in the intensity distribution about the edge. A technique for computational simulation of the phase contrast images has also been developed through which contrast mechanisms in various samples could be understood and expected phase contrast could be predicted, which can help in examining this technique for a particular application. Parameters were identified based on which the feasibility and performance of phase contrast imaging experiments can be studied using neutron transport calculations. The transverse coherence length along with the neutron flux at the image plane, given the source characteristics and other constraints, can be optimized through computational simulations. Through computational simulations, effects of the pure phase as well as phase-amplitude interaction terms on contrast enhancement in the images (for different mixed phase-amplitude materials) were investigated. It was observed that the interaction effect is less than the pure phase effect in terms of contrast enhancement; nevertheless it can be significant depending on the type of material edges present in the sample. The interaction effect can also lead to a higher overall contrast in samples having some material edges with significant phase-amplitude interaction and others where it is not significant. This is achieved by decreasing the contrast at those edges that already have high attenuation contrast thereby virtually expanding the gray scale in the digital image making other edges more prominent. ACKNOWLEDGMENT The authors would like to thank Mr. Andrew Cook for providing information on the NCSU PULSTAR reactor. The authors would also like to thank Ms. Emily Miller for her contribution to the design of phantoms.

8 1636 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE 2009 REFERENCES [1] B. E. Allman, P. J. McMahon, K. A. Nugent, D. Paganin, D. L. Jacobson, M. Arif, and S. A. Werner, Phase radiography with neutrons, Nature (Brief Communications), vol. 408, pp , Nov [2] P. J. McMahon, B. E. Allman, K. A. Nugent, D. L. Jacobson, M. Arif, and S. A. Werner, Contrast mechanisms for neutron radiography, Appl. Phys. Lett., vol. 78, no. 7, pp , Feb [3] N. Kardjilov, E. Lehmann, E. Steichele, and P. Vontobel, Phase-contrast radiography with a polychromatic neutron beam, Nucl. Instrum. Methods Phys. Res. A, vol. A527, pp , [4] F. Arfelli et al., Low-dose phase contrast X-ray medical imaging, Phys. Med. Biol., vol. 43, pp , June [5] K. Kanaya and H. Kawakatsu, Experiments on the electron phase microscope, J. Appl. Phys., vol. 29, pp , [6] C. J. Kotre and I. P. Birch, Phase contrast enhancement of X-ray mammography: A design study, Phys. Med. Biol., vol. 44, pp , [7] T. E. Gureyev, A. Roberts, and K. A. Nugent, Partially coherent fields, the transport of intensity equation, and phase uniqueness, J. Opt. Soc. Amer. A, vol. 12, no. 9, pp , Sept [8] D. Paganin and K. A. Nugent, Noninterferometric phase imaging with partially coherent light, Phys. Rev. Lett., vol. 80, pp , Mar [9] A. V. Bronnikov, Theory of quantitative phase-contrast computed tomography, J. Opt. Soc. Amer. A, vol. 19, no. 3, Mar [10] J. M. Cowley, Diffraction Physics. New York: North-Holland, [11] K. K. Mishra, A. I. Hawari, and V. H. Gillette, Design and perfromance of a thermal neutron imaging facility at the North Carolina State University PULSTAR reactor, IEEE Trans. Nucl. Sci., vol. 53, no. 6, pp , Dec [12] X-5 Monte Carlo Team, MCNP A General Monte Carlo N-Particle Transport Code, Version 5, Los Alamos National Laboratory, Los Alamos, NM, Tech. Rep. LA-UR , [13] A. I. Hawari, I. I. Al-Qasir, and K. K. Mishra, Accurate simulation of thermal neutron filter effects in the design of research reactor beam applications, in PHYSOR-2006: Advances in Nuclear Analysis and Simulation, Vancouver, BC, Canada, [14] Mammography Accreditation Phantom Gammex 156 GAMMEX RMI, Middleton, WI, [15] S. J. Lim, Two-Dimensional Signal and Image Processing. Englewood Cliffs, NJ: Prentice-Hall, 1990, pp [16] V. F. Sears, Neutron scattering lengths and cross sections, Neutron News, vol. 3, no. 3, pp , 1992 [Online]. Available: ncnr.nist.gov/resources/n-lengths

THE use of neutrons in imaging has been on-going since

THE use of neutrons in imaging has been on-going since 3904 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 6, DECEMBER 2006 Design and Performance of a Thermal Neutron Imaging Facility at the North Carolina State University PULSTAR Reactor Kaushal K. Mishra,

More information

THE SIMULATION OF THE CODED NEUTRON SOURCE PHASE CONTRAST IMAGING

THE SIMULATION OF THE CODED NEUTRON SOURCE PHASE CONTRAST IMAGING The 12 th International Conference of the Slovenian Society for Non-Destructive Testing»Application of Contemporary Non-Destructive Testing in Engineering«September 4-6, 2013, Portorož, Slovenia More info

More information

Initial Results on the Feasibility of Hybrid X-Ray Microscopy

Initial Results on the Feasibility of Hybrid X-Ray Microscopy CHINESE JOURNAL OF PHYSICS VOL. 43, NO. 5 OCTOBER 2005 Initial Results on the Feasibility of Hybrid X-Ray Microscopy P. K. Tseng, 1 W. F. Pong, 1 C. L. Chang, 1 C. P. Hsu, 1 F. Y. Lin, 2 C. S. Hwang, 2

More information

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER M. Bovi (1), R.F. Laitano (1), M. Pimpinella (1), M. P. Toni (1), K. Casarin(2), E. Quai(2), G. Tromba(2), A. Vascotto(2),

More information

FACTORS AFFECTING IN-LINE PHASE CONTRAST IMAGING WITH A LABORATORY MICROFOCUS X-RAY SOURCE

FACTORS AFFECTING IN-LINE PHASE CONTRAST IMAGING WITH A LABORATORY MICROFOCUS X-RAY SOURCE Copyright JCPDS-International Centre for Diffraction Data 26 ISSN 197-2 FACTORS AFFECTING IN-LINE PHASE CONTRAST IMAGING WITH A LABORATORY MICROFOCUS X-RAY SOURCE 31 K. L. Kelly and B. K. Tanner Department

More information

at Oak Ridge National Laboratory.

at Oak Ridge National Laboratory. 361 Designs for Neutron Radiography at Oak Ridge National Laboratory and Computed Tomography Dudley A. Raine lipv4, Camden R. Hubbard, Paul M. Whaley, and Michael C. Wright3 Oak Ridge National Laboratory

More information

Interference, Diffraction and Fourier Theory. ATI 2014 Lecture 02! Keller and Kenworthy

Interference, Diffraction and Fourier Theory. ATI 2014 Lecture 02! Keller and Kenworthy Interference, Diffraction and Fourier Theory ATI 2014 Lecture 02! Keller and Kenworthy The three major branches of optics Geometrical Optics Light travels as straight rays Physical Optics Light can be

More information

Considerations for Measurements in Support of Thermal Scattering Data Evaluations. Ayman I. Hawari

Considerations for Measurements in Support of Thermal Scattering Data Evaluations. Ayman I. Hawari OECD/NEA Meeting: WPEC SG42 Thermal Scattering Kernel S(a,b): Measurement, Evaluation and Application May 13 14, 2017 Paris, France Considerations for Measurements in Support of Thermal Scattering Data

More information

A Radiation Monitoring System With Capability of Gamma Imaging and Estimation of Exposure Dose Rate

A Radiation Monitoring System With Capability of Gamma Imaging and Estimation of Exposure Dose Rate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 49, NO. 3, JUNE 2002 1547 A Radiation Monitoring System With Capability of Gamma Imaging and Estimation of Exposure Dose Rate Wanno Lee, Gyuseong Cho, and Ho

More information

arxiv: v3 [nucl-ex] 18 May 2018

arxiv: v3 [nucl-ex] 18 May 2018 Observation of Pendellösung Fringes by Using Pulsed Neutrons Shigeyasu Itoh, Masaya Nakaji, Yuya Uchida, Masaaki Kitaguchi, and Hirohiko M. Shimizu Department of Physics, Nagoya University Furo-cho, Chikusa-ku,

More information

DIFFRACTION PHYSICS THIRD REVISED EDITION JOHN M. COWLEY. Regents' Professor enzeritus Arizona State University

DIFFRACTION PHYSICS THIRD REVISED EDITION JOHN M. COWLEY. Regents' Professor enzeritus Arizona State University DIFFRACTION PHYSICS THIRD REVISED EDITION JOHN M. COWLEY Regents' Professor enzeritus Arizona State University 1995 ELSEVIER Amsterdam Lausanne New York Oxford Shannon Tokyo CONTENTS Preface to the first

More information

EDDY-CURRENT nondestructive testing is commonly

EDDY-CURRENT nondestructive testing is commonly IEEE TRANSACTIONS ON MAGNETICS, VOL. 34, NO. 2, MARCH 1998 515 Evaluation of Probe Impedance Due to Thin-Skin Eddy-Current Interaction with Surface Cracks J. R. Bowler and N. Harfield Abstract Crack detection

More information

Modeling of Li-Ion-Batteries to Optimize the Results Gained by Neutron Imaging

Modeling of Li-Ion-Batteries to Optimize the Results Gained by Neutron Imaging Modeling of Li-Ion-Batteries to Optimize the Results Gained by Neutron Imaging M.J. Mühlbauer Dr. A. Senyshyn, Dr. O. Dolotko, Prof. H. Ehrenberg SFB 595 Electrical Fatigue in Functional Materials Contact:

More information

A Brief Introduction to Medical Imaging. Outline

A Brief Introduction to Medical Imaging. Outline A Brief Introduction to Medical Imaging Outline General Goals Linear Imaging Systems An Example, The Pin Hole Camera Radiations and Their Interactions with Matter Coherent vs. Incoherent Imaging Length

More information

5. LIGHT MICROSCOPY Abbe s theory of imaging

5. LIGHT MICROSCOPY Abbe s theory of imaging 5. LIGHT MICROSCOPY. We use Fourier optics to describe coherent image formation, imaging obtained by illuminating the specimen with spatially coherent light. We define resolution, contrast, and phase-sensitive

More information

Supplementary Figure 1: Example non-overlapping, binary probe functions P1 (~q) and P2 (~q), that add to form a top hat function A(~q).

Supplementary Figure 1: Example non-overlapping, binary probe functions P1 (~q) and P2 (~q), that add to form a top hat function A(~q). Supplementary Figures P(q) A(q) + Function Value P(q) qmax = Supplementary Figure : Example non-overlapping, binary probe functions P (~q) and P (~q), that add to form a top hat function A(~q). qprobe

More information

Determining the effective electromagnetic properties of negative-refractive-index metamaterials from internal fields

Determining the effective electromagnetic properties of negative-refractive-index metamaterials from internal fields Determining the effective electromagnetic properties of negative-refractive-index metamaterials from internal fields Bogdan-Ioan Popa* and Steven A. Cummer Department of Electrical and Computer Engineering,

More information

ELEG 479 Lecture #6. Mark Mirotznik, Ph.D. Associate Professor The University of Delaware

ELEG 479 Lecture #6. Mark Mirotznik, Ph.D. Associate Professor The University of Delaware ELEG 479 Lecture #6 Mark Mirotznik, Ph.D. Associate Professor The University of Delaware Summary of Last Lecture X-ray Physics What are X-rays and when are they useful for medical imaging? How are X-rays

More information

Study of the phase contrast for the characterization of the surface of microshell

Study of the phase contrast for the characterization of the surface of microshell 19 th World Conference on Non-Destructive Testing 2016 Study of the phase contrast for the characterization of the surface of microshell Alexandre CHOUX 1,*, Vincent DUTTO 1, Eric BUSVELLE 2, Jean-Paul

More information

Optimisation using measured Green s function for improving spatial coherence in acoustic measurements

Optimisation using measured Green s function for improving spatial coherence in acoustic measurements Ultrasonics 42 (2004) 205 212 www.elsevier.com/locate/ultras Optimisation using measured Green s function for improving spatial coherence in acoustic measurements Matthew Clark *, Steve D. Sharples, Mike

More information

Michelson Interferometer

Michelson Interferometer Michelson Interferometer Objective Determination of the wave length of the light of the helium-neon laser by means of Michelson interferometer subsectionprinciple and Task Light is made to produce interference

More information

Which of the following can be used to calculate the resistive force acting on the brick? D (Total for Question = 1 mark)

Which of the following can be used to calculate the resistive force acting on the brick? D (Total for Question = 1 mark) 1 A brick of mass 5.0 kg falls through water with an acceleration of 0.90 m s 2. Which of the following can be used to calculate the resistive force acting on the brick? A 5.0 (0.90 9.81) B 5.0 (0.90 +

More information

Shielding Design for the Imaging and Medical Beamline at the Australian Synchrotron

Shielding Design for the Imaging and Medical Beamline at the Australian Synchrotron Shielding Design for the Imaging and Medical Beamline at the Australian Synchrotron P. Berkvens and D. Häusermann European Synchrotron Radiation Facility BP 0, Grenoble Cedex 0, France Australian Synchrotron

More information

Quartz-Crystal Spectrometer for the Analysis of Plutonium K X-Rays

Quartz-Crystal Spectrometer for the Analysis of Plutonium K X-Rays Quartz-Crystal Spectrometer for the Analysis of Plutonium K X-Rays Alison V. Goodsell, William S. Charlton alisong@tamu.edu, charlton@ne.tamu.edu Nuclear Security Science & Policy Institute Texas A&M University,

More information

A Study on Radiation Damage in PWO-II Crystals

A Study on Radiation Damage in PWO-II Crystals 2336 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 60, NO. 3, JUNE 2013 A Study on Radiation Damage in PWO-II Crystals Fan Yang, Member, IEEE, Rihua Mao, Member, IEEE, Liyuan Zhang, Member, IEEE, and Ren-yuan

More information

Time of flight SANS data Reduction (nothing really new here, but perhaps presented in unfamiliar ways)

Time of flight SANS data Reduction (nothing really new here, but perhaps presented in unfamiliar ways) Time of flight AN data Reduction (nothing really new here, but perhaps presented in unfamiliar ways) Richard Heenan II Facility, Rutherford Appleton Laboratory richard.heenan@stfc.ac.uk AN2d Collimation

More information

High-Resolution Neutron Diffraction Monochromators for Neutron Diffractometry

High-Resolution Neutron Diffraction Monochromators for Neutron Diffractometry High-Resolution Neutron Diffraction Monochromators for Neutron Diffractometry Pavol Mikula, Nuclear Physics Institute ASCR 25 68 Řež near Prague, Czech Republic NMI3-Meeting, Barcelona, 21 Motivation Backscattering

More information

Applications of MCBEND

Applications of MCBEND Serco Assurance Applications of MCBEND Presentation to NPL Workshop on Monte Carlo codes by Pat Cowan The ANSWERS Software Service Serco Assurance Overview The MCBEND Code Traditional Applications Industrial

More information

Imaging principle and experiment results of an 11 MeV low-energy proton radiography system

Imaging principle and experiment results of an 11 MeV low-energy proton radiography system NUCLEAR SCIENCE AND TECHNIQUES 25, 060203 (2014) Imaging principle and experiment results of an 11 MeV low-energy proton radiography system LI Yi-Ding ( 李一丁 ), 1, ZHANG Xiao-Ding ( 张小丁 ), 1 WEI Tao ( 魏涛

More information

Performance characteristics of the tomography setup at the PSI NEUTRA thermal neutron radiography facility

Performance characteristics of the tomography setup at the PSI NEUTRA thermal neutron radiography facility Performance characteristics of the tomography setup at the PSI NEUTRA thermal neutron radiography facility Vontobel, Peter, Paul Scherrer Institute, Villigen, Switzerland Lehmann, Eberhard, Paul Scherrer

More information

ULTRAFAST LASER PULSE TRAIN RADIATION TRANSFER IN A SCATTERING-ABSORBING 3D MEDIUM WITH AN INHOMOGENEITY

ULTRAFAST LASER PULSE TRAIN RADIATION TRANSFER IN A SCATTERING-ABSORBING 3D MEDIUM WITH AN INHOMOGENEITY Heat Transfer Research 46(9), 861 879 (2015) ULTRAFAST LASER PULSE TRAIN RADIATION TRANSFER IN A SCATTERING-ABSORBING 3D MEDIUM WITH AN INHOMOGENEITY Masato Akamatsu 1,* & Zhixiong Guo 2 1 Graduate School

More information

Modeling of the fringe shift in multiple beam interference for glass fibers

Modeling of the fringe shift in multiple beam interference for glass fibers PRAMANA c Indian Academy of Sciences Vol. 70, No. 4 journal of April 2008 physics pp. 643 648 Modeling of the fringe shift in multiple beam interference for glass fibers A M HAMED 1 Physics Department,

More information

Quantitative Assessment of Scattering Contributions in MeV-Industrial X-ray Computed Tomography

Quantitative Assessment of Scattering Contributions in MeV-Industrial X-ray Computed Tomography 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic More Info at Open Access Database www.ndt.net/?id=16530 Quantitative Assessment of Scattering

More information

Excitations. 15 th Oxford School of Neutron Scattering. Elizabeth Blackburn University of Birmingham. Blackburn et al., Pramana 71, 673 (2008)

Excitations. 15 th Oxford School of Neutron Scattering. Elizabeth Blackburn University of Birmingham. Blackburn et al., Pramana 71, 673 (2008) Excitations Elizabeth Blackburn University of Birmingham Cowley and Woods., Can. J. Phys. 49, 177 (1971) Blackburn et al., Pramana 71, 673 (2008) 15 th Oxford School of Neutron Scattering Excitations Elizabeth

More information

Lecture 19 Optical MEMS (1)

Lecture 19 Optical MEMS (1) EEL6935 Advanced MEMS (Spring 5) Instructor: Dr. Huikai Xie Lecture 19 Optical MEMS (1) Agenda: Optics Review EEL6935 Advanced MEMS 5 H. Xie 3/8/5 1 Optics Review Nature of Light Reflection and Refraction

More information

X-Ray Nanoimaging: Instruments And Methods II (Proceedings Of SPIE)

X-Ray Nanoimaging: Instruments And Methods II (Proceedings Of SPIE) X-Ray Nanoimaging: Instruments And Methods II (Proceedings Of SPIE) If you are searched for a book X-Ray Nanoimaging: Instruments and Methods II (Proceedings of SPIE) in pdf form, then you've come to the

More information

PRINCIPLES OF PHYSICAL OPTICS

PRINCIPLES OF PHYSICAL OPTICS PRINCIPLES OF PHYSICAL OPTICS C. A. Bennett University of North Carolina At Asheville WILEY- INTERSCIENCE A JOHN WILEY & SONS, INC., PUBLICATION CONTENTS Preface 1 The Physics of Waves 1 1.1 Introduction

More information

Transverse Coherence Properties of the LCLS X-ray Beam

Transverse Coherence Properties of the LCLS X-ray Beam LCLS-TN-06-13 Transverse Coherence Properties of the LCLS X-ray Beam S. Reiche, UCLA, Los Angeles, CA 90095, USA October 31, 2006 Abstract Self-amplifying spontaneous radiation free-electron lasers, such

More information

Step index planar waveguide

Step index planar waveguide N. Dubreuil S. Lebrun Exam without document Pocket calculator permitted Duration of the exam: 2 hours The exam takes the form of a multiple choice test. Annexes are given at the end of the text. **********************************************************************************

More information

Shielded Scintillator for Neutron Characterization

Shielded Scintillator for Neutron Characterization Shielded Scintillator for Neutron Characterization A Thesis Submitted in Partial Fulfillment of the Requirements for Graduation with Research Distinction in Engineering Physics By Patrick X. Belancourt

More information

Neutron collimator design of neutron radiography based on the BNCT facility arxiv: v1 [physics.ins-det] 3 May 2013

Neutron collimator design of neutron radiography based on the BNCT facility arxiv: v1 [physics.ins-det] 3 May 2013 CPC(HEP & NP), 2009, 33(X): 1 5 Chinese Physics C Vol. 33, No. X, Xxx, 2009 Neutron collimator design of neutron radiography based on the BNCT facility arxiv:1305.0672v1 [physics.ins-det] 3 May 2013 YANG

More information

Ambiguity of optical coherence tomography measurements due to rough surface scattering

Ambiguity of optical coherence tomography measurements due to rough surface scattering Ambiguity of optical coherence tomography measurements due to rough surface scattering Y. Ashtamker, 1 V Freilikher, 1,* and J C Dainty 2 1 Department of Physics, Bar-Ilan University, Ramat Gan 52900,

More information

Novel method for ultrashort laser pulse-width measurement based on the self-diffraction effect

Novel method for ultrashort laser pulse-width measurement based on the self-diffraction effect Novel method for ultrashort laser pulse-width measurement based on the self-diffraction effect Peng Xi, Changhe Zhou, Enwen Dai, and Liren Liu Shanghai Institute of Optics and Fine Mechanics, Chinese Academy

More information

A new method of detecting interferogram in differential phase-contrast imaging system based on special structured x-ray scintillator screen

A new method of detecting interferogram in differential phase-contrast imaging system based on special structured x-ray scintillator screen A new method of detecting interferogram in differential phase-contrast imaging system based on special structured x-ray scintillator screen Liu Xin 刘鑫, Guo Jin-Chuan 郭金川, and Niu Han-Ben 牛憨笨 College of

More information

Fachbereich II Mathematik - Physik - Chemie

Fachbereich II Mathematik - Physik - Chemie Fachbereich II Mathematik - Physik - Chemie 04/2010 Wolfgang Treimer, Sven-Oliver Seidel, Omid Ebrahimi A compact experimental set up for neutron tomography Ein kompakter Experimentaufbau für Neutronentomographie

More information

Phase contrast x ray imaging in two and three dimensions P. Cloetens

Phase contrast x ray imaging in two and three dimensions P. Cloetens Phase contrast x ray imaging in two and three dimensions P. Cloetens Propagation based phase contrast imaging the forward problem; Fresnel diffraction coherence conditions The inverse problem phase retrieval

More information

SLAC National Accelerator Laboratory FACET & TEST BEAM FACILITIES PROPOSAL

SLAC National Accelerator Laboratory FACET & TEST BEAM FACILITIES PROPOSAL SLAC National Accelerator Laboratory FACET & TEST BEAM FACILITIES PROPOSAL Date: 6/17/13 A. EXPERIMENT TITLE: Development of Electron Radiography for Material Science B. PROPOSERS & REQUESTED FACILITY:

More information

Neutron and Gamma Ray Imaging for Nuclear Materials Identification

Neutron and Gamma Ray Imaging for Nuclear Materials Identification Neutron and Gamma Ray Imaging for Nuclear Materials Identification James A. Mullens John Mihalczo Philip Bingham Oak Ridge National Laboratory Oak Ridge, Tennessee 37831-6010 865-574-5564 Abstract This

More information

Structure Report for J. Reibenspies

Structure Report for J. Reibenspies X-ray Diffraction Laboratory Center for Chemical Characterization and Analysis Department of Chemistry Texas A & M University Structure Report for J. Reibenspies Project Name: Sucrose Date: January 29,

More information

Available online at ScienceDirect. Physics Procedia 69 (2015 )

Available online at  ScienceDirect. Physics Procedia 69 (2015 ) Available online at www.sciencedirect.com ScienceDirect Physics Procedia 69 (2015 ) 392 398 10 World Conference on Neutron Radiography 5-10 October 2014 Au Foil Activation Measurement and Simulation of

More information

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research Prompt gamma measurements for the verification of dose deposition in proton therapy Two Proton Beam Facilities for Therapy and Research Ion Beam Facilities in Korea 1. Proton therapy facility at National

More information

Optical Systems Program of Studies Version 1.0 April 2012

Optical Systems Program of Studies Version 1.0 April 2012 Optical Systems Program of Studies Version 1.0 April 2012 Standard1 Essential Understand Optical experimental methodology, data analysis, interpretation, and presentation strategies Essential Understandings:

More information

Simulation program for reinforced concrete tomography with gamma-rays

Simulation program for reinforced concrete tomography with gamma-rays Simulation program for reinforced concrete tomography with gamma-rays P. Thieberger, M.A.J. Mariscotti and M. Ruffolo Tomografía de Hormigón Armado, S.A., Buenos Aires, Argentina Av. de Mayo 695, 3º "7"

More information

A STUDY OF MULTIPLE REFRACTIVE SCATTERING OF MONOENERGETIC X-RAYS FROM ENSEMBLES OF MONODISPERSE SPHERES. Masterarbeit

A STUDY OF MULTIPLE REFRACTIVE SCATTERING OF MONOENERGETIC X-RAYS FROM ENSEMBLES OF MONODISPERSE SPHERES. Masterarbeit A STUDY OF MULTIPLE REFRACTIVE SCATTERING OF MONOENERGETIC X-RAYS FROM ENSEMBLES OF MONODISPERSE SPHERES Masterarbeit zur Erlangung des akademischen Grades Master of Science in Physics (M.Sc.) dem Fachbereich

More information

SAXS and SANS facilities and experimental practice. Clement Blanchet

SAXS and SANS facilities and experimental practice. Clement Blanchet SAXS and SANS facilities and experimental practice Clement Blanchet SAS experiment Detector X-ray or neutron Beam Sample 2 s Buffer X-rays Roengten, 1895 Electromagnetic wave The electromagnetic spectrum

More information

Electron density and effective atomic number images generated by dual energy imaging with a 320-detector CT system: A feasibility study

Electron density and effective atomic number images generated by dual energy imaging with a 320-detector CT system: A feasibility study Electron density and effective atomic number images generated by dual energy imaging with a 320-detector CT system: A feasibility study Poster No.: C-0403 Congress: ECR 2014 Type: Scientific Exhibit Authors:

More information

arxiv:nucl-ex/ v1 14 Sep 2005

arxiv:nucl-ex/ v1 14 Sep 2005 Measuring the Neutron s Mean Square Charge Radius Using Neutron Interferometry F. E. Wietfeldt, M. Huber Tulane University, New Orleans, LA T. C. Black University of North Carolina, Wilmington, NC arxiv:nucl-ex/0509018

More information

Development of a High-Speed VUV Camera System for 2-Dimensional Imaging of Edge Turbulent Structure in the LHD

Development of a High-Speed VUV Camera System for 2-Dimensional Imaging of Edge Turbulent Structure in the LHD Development of a High-Speed VUV Camera System for 2-Dimensional Imaging of Edge Turbulent Structure in the LHD Masaki TAKEUCHI, Satoshi OHDACHI and LHD experimental group National Institute for Fusion

More information

Analysis of diffraction efficiency of a holographic coupler with respect to angular divergence

Analysis of diffraction efficiency of a holographic coupler with respect to angular divergence Indian J. Phys. 83 (4) 531-538 (009) Analysis of diffraction efficiency of a holographic coupler with respect to angular divergence Mihir Hota and S K Tripathy* National Institute of Science and Technology,

More information

Beyond the Geometric toward the Wave Optical Approach in the Design of Curved Crystal and Multilayer Optics for EDXAS

Beyond the Geometric toward the Wave Optical Approach in the Design of Curved Crystal and Multilayer Optics for EDXAS Beyond the Geometric toward the Wave Optical Approach in the Design of Curved Crystal and Multilayer Optics for EDXAS Vito Mocella CNR IMM Napoli Units, Italy In collaboration with C. Ferrero, C. Morawe,

More information

PROBLEM OF X-RAY SYNCHROTRON BEAM COLLIMATION BY ZONE PLATE

PROBLEM OF X-RAY SYNCHROTRON BEAM COLLIMATION BY ZONE PLATE PROBLEM OF X-RAY SYNCHROTRON BEAM COLLIMATION BY ZONE PLATE A. Kuyumchyan* a, V. Kohn b, A. Snigirev c, I. Snigireva c, M. Grigorev a, S. Kouznetsov a Institute of Microelectronics Technology, RAS, 443,

More information

Supporting Information

Supporting Information Supporting Information Devlin et al. 10.1073/pnas.1611740113 Optical Characterization We deposit blanket TiO films via ALD onto silicon substrates to prepare samples for spectroscopic ellipsometry (SE)

More information

Spatial coherence measurement

Spatial coherence measurement Spatial coherence measurement David Paterson Argonne National Laboratory A Laboratory Operated by The University of Chicago Motivation Third generation synchrotrons produce very bright, partially coherent

More information

Observation of ferroelectric domains in bismuth ferrite using coherent diffraction techniques

Observation of ferroelectric domains in bismuth ferrite using coherent diffraction techniques Observation of ferroelectric domains in bismuth ferrite using coherent diffraction techniques James Vale October 25, 2011 Abstract Multiferroic materials have significant potential for both the scientific

More information

Simple Experimental Design for Calculation of Neutron Removal Cross Sections K. Groves 1 1) McMaster University, 1280 Main St. W, Hamilton, Canada.

Simple Experimental Design for Calculation of Neutron Removal Cross Sections K. Groves 1 1) McMaster University, 1280 Main St. W, Hamilton, Canada. Simple Experimental Design for Calculation of Neutron Removal Cross Sections K. Groves 1 1) McMaster University, 1280 Main St. W, Hamilton, Canada. (Dated: 5 August 2017) This article proposes an experimental

More information

11/10/2014. Chapter 1: Introduction to Medical Imaging. Projection (Transmission) vs. Emission Imaging. Emission Imaging

11/10/2014. Chapter 1: Introduction to Medical Imaging. Projection (Transmission) vs. Emission Imaging. Emission Imaging Chapter 1: Introduction to Medical Imaging Overview of Modalities Properties of an Image: Limitations on Information Content Contrast (both object & image): Brightness difference Sharpness (blur): Smallest

More information

Research on Defect Depth Measurement Algorithm In Digital Radiography Testing

Research on Defect Depth Measurement Algorithm In Digital Radiography Testing 19 th World Conference on Non-Destructive Testing 2016 Research on Defect Depth Measurement Algorithm In Digital Radiography Testing Wenming GUO, Yuliang CHEN School of Software Engineering, Beijing University

More information

Light Propagation in Free Space

Light Propagation in Free Space Intro Light Propagation in Free Space Helmholtz Equation 1-D Propagation Plane waves Plane wave propagation Light Propagation in Free Space 3-D Propagation Spherical Waves Huygen s Principle Each point

More information

Lecture notes 5: Diffraction

Lecture notes 5: Diffraction Lecture notes 5: Diffraction Let us now consider how light reacts to being confined to a given aperture. The resolution of an aperture is restricted due to the wave nature of light: as light passes through

More information

Lasers and Electro-optics

Lasers and Electro-optics Lasers and Electro-optics Second Edition CHRISTOPHER C. DAVIS University of Maryland III ^0 CAMBRIDGE UNIVERSITY PRESS Preface to the Second Edition page xv 1 Electromagnetic waves, light, and lasers 1

More information

Physical Optics. Lecture 7: Coherence Herbert Gross.

Physical Optics. Lecture 7: Coherence Herbert Gross. Physical Optics Lecture 7: Coherence 07-05-7 Herbert Gross www.iap.uni-jena.de Physical Optics: Content No Date Subject Ref Detailed Content 05.04. Wave optics G Complex fields, wave equation, k-vectors,

More information

THREE-DIMENSIONAL INTEGRAL NEUTRON TRANSPORT CELL CALCULATIONS FOR THE DETERMINATION OF MEAN CELL CROSS SECTIONS

THREE-DIMENSIONAL INTEGRAL NEUTRON TRANSPORT CELL CALCULATIONS FOR THE DETERMINATION OF MEAN CELL CROSS SECTIONS THREE-DIMENSIONAL INTEGRAL NEUTRON TRANSPORT CELL CALCULATIONS FOR THE DETERMINATION OF MEAN CELL CROSS SECTIONS Carsten Beckert 1. Introduction To calculate the neutron transport in a reactor, it is often

More information

Focal shift in vector beams

Focal shift in vector beams Focal shift in vector beams Pamela L. Greene The Institute of Optics, University of Rochester, Rochester, New York 1467-186 pgreene@optics.rochester.edu Dennis G. Hall The Institute of Optics and The Rochester

More information

Georgia Institute of Technology. Radiation Detection & Protection (Day 3)

Georgia Institute of Technology. Radiation Detection & Protection (Day 3) Georgia Institute of Technology The George W. Woodruff School of Mechanical Engineering Nuclear & Radiological Engineering/Medical Physics Program Ph.D. Qualifier Exam Spring Semester 2009 Your ID Code

More information

Scattering Lecture. February 24, 2014

Scattering Lecture. February 24, 2014 Scattering Lecture February 24, 2014 Structure Determination by Scattering Waves of radiation scattered by different objects interfere to give rise to an observable pattern! The wavelength needs to close

More information

2. Determine the excess charge on the outer surface of the outer sphere (a distance c from the center of the system).

2. Determine the excess charge on the outer surface of the outer sphere (a distance c from the center of the system). Use the following to answer question 1. Two point charges, A and B, lie along a line separated by a distance L. The point x is the midpoint of their separation. 1. Which combination of charges will yield

More information

Dept. of Physics, MIT Manipal 1

Dept. of Physics, MIT Manipal 1 Chapter 1: Optics 1. In the phenomenon of interference, there is A Annihilation of light energy B Addition of energy C Redistribution energy D Creation of energy 2. Interference fringes are obtained using

More information

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT Proceedings of the Eleventh EGS4 Users' Meeting in Japan, KEK Proceedings 2003-15, p.107-113 ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT K. Noto and

More information

School on Pulsed Neutrons: Characterization of Materials October Neurton Sources & Scattering Techniques (1-2)

School on Pulsed Neutrons: Characterization of Materials October Neurton Sources & Scattering Techniques (1-2) 1866-6 School on Pulsed Neutrons: Characterization of Materials 15-26 October 2007 Neurton Sources & Scattering Techniques (1-2) Guenter Bauer Forschungzentrum Julich GmbH Julich Germany The Abdus Salam

More information

Rad T 290 Worksheet 2

Rad T 290 Worksheet 2 Class: Date: Rad T 290 Worksheet 2 1. Projectile electrons travel from a. anode to cathode. c. target to patient. b. cathode to anode. d. inner shell to outer shell. 2. At the target, the projectile electrons

More information

X-ray phase imaging with a grating interferometer

X-ray phase imaging with a grating interferometer X-ray phase imaging with a grating interferometer Timm Weitkamp, Ana Diaz and Christian David Laboratory for Micro- and Nanotechnology, Paul Scherrer Institut, 5232 Villigen PSI, Switzerland timm.weitkamp@psi.ch

More information

Monte Carlo Characterization of a Pulsed Laser-Wakefield Driven Monochromatic X-Ray Source

Monte Carlo Characterization of a Pulsed Laser-Wakefield Driven Monochromatic X-Ray Source 2009 IEEE Nuclear Science Symposium Conference Record N30-3 Monte Carlo Characterization of a Pulsed Laser-Wakefield Driven Monochromatic X-Ray Source S. D. Clarke, S. A. Pozzi, IEEE Member, N. Cunningham,

More information

custom reticle solutions

custom reticle solutions custom reticle solutions 01 special micro structures Pyser Optics has over 60 years experience in producing high quality micro structure products. These products are supplied worldwide to industries including

More information

Towards Proton Computed Tomography

Towards Proton Computed Tomography SCIPP Towards Proton Computed Tomography L. R. Johnson, B. Keeney, G. Ross, H. F.-W. Sadrozinski, A. Seiden, D.C. Williams, L. Zhang Santa Cruz Institute for Particle Physics, UC Santa Cruz, CA 95064 V.

More information

X-RAY MICRO-TOMOGRAPHY OF PORE-SCALE FLOW AND TRANSPORT. University of California Davis. Dorthe Wildenschild & Annette Mortensen

X-RAY MICRO-TOMOGRAPHY OF PORE-SCALE FLOW AND TRANSPORT. University of California Davis. Dorthe Wildenschild & Annette Mortensen X-RAY MICRO-TOMOGRAPHY OF PORE-SCALE FLOW AND TRANSPORT Jan W. Hopmans University of California Davis Volker Clausnitzer Dorthe Wildenschild & Annette Mortensen ISSUES: Measurements and modeling of water

More information

Principles of Nuclear Magnetic Resonance Microscopy

Principles of Nuclear Magnetic Resonance Microscopy Principles of Nuclear Magnetic Resonance Microscopy Paul T. Callaghan Department of Physics and Biophysics Massey University New Zealand CLARENDON PRESS OXFORD CONTENTS 1 PRINCIPLES OF IMAGING 1 1.1 Introduction

More information

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA The Underwater Spectrometric System Based on CZT Detector for Survey of the Bottom of MR Reactor Pool 13461 Victor Potapov, Alexey Safronov, Oleg Ivanov, Sergey Smirnov, Vyacheslav Stepanov National Research

More information

10. OPTICAL COHERENCE TOMOGRAPHY

10. OPTICAL COHERENCE TOMOGRAPHY 1. OPTICAL COHERENCE TOMOGRAPHY Optical coherence tomography (OCT) is a label-free (intrinsic contrast) technique that enables 3D imaging of tissues. The principle of its operation relies on low-coherence

More information

An experimental study of dual-energy CT imaging using synchrotron radiation

An experimental study of dual-energy CT imaging using synchrotron radiation Nuclear Science and Techniques 24 (2013) 020102 An experimental study of dual-energy CT imaging using synchrotron radiation HAO Jia 1,2 ZHANG Li 1,2 XING Yuxiang 1,2 KANG Kejun 1,2 1 Department of Engineering

More information

AMETHOD of reconstructing the in-plane component of

AMETHOD of reconstructing the in-plane component of 2324 IEEE TRANSACTIONS ON MAGNETICS, VOL. 34, NO. 4, JULY 1998 Magnetization Reconstruction from Differential Phase Contrast Lorentz Microscopy and Magnetic Force Microscopy M. Wdowin, J. J. Miles, B.

More information

Laser Optics-II. ME 677: Laser Material Processing Instructor: Ramesh Singh 1

Laser Optics-II. ME 677: Laser Material Processing Instructor: Ramesh Singh 1 Laser Optics-II 1 Outline Absorption Modes Irradiance Reflectivity/Absorption Absorption coefficient will vary with the same effects as the reflectivity For opaque materials: reflectivity = 1 - absorptivity

More information

ASTM E545 TUTORIAL ON PROPER IMAGE QUALITY INDICATOR USAGE

ASTM E545 TUTORIAL ON PROPER IMAGE QUALITY INDICATOR USAGE ASTM E545 IQI USAGE TUTORIAL Page 1 of 9 ASTM E545 TUTORIAL ON PROPER IMAGE QUALITY INDICATOR USAGE Neutron radiography is a non-destructive testing method similar to the more familiar x- ray, but which

More information

Collimated LaBr 3 detector response function in radioactivity analysis of nuclear waste drums

Collimated LaBr 3 detector response function in radioactivity analysis of nuclear waste drums Nuclear Science and Techniques 4 (13) 63 Collimated LaBr 3 detector response function in radioactivity analysis of nuclear waste drums QIAN Nan 1 WANG Dezhong 1,* WANG Chuan ZHU Yuelong MAUERHOFER Eric

More information

Neutron Dose near Spent Nuclear Fuel and HAW after the 2007 ICRP Recommendations

Neutron Dose near Spent Nuclear Fuel and HAW after the 2007 ICRP Recommendations Neutron Dose near Spent Nuclear Fuel and HAW after the 2007 ICRP Recommendations Gunter Pretzsch Gesellschaft fuer Anlagen- und Reaktorsicherheit (GRS) mbh Radiation and Environmental Protection Division

More information

Progress in Nuclear Science and Technology, Volume 6,

Progress in Nuclear Science and Technology, Volume 6, DOI: 1.15669/pnst.6 Progress in Nuclear Science and Technology Volume 6 (19) pp. 1-16 ARTICLE A study on calculation method of duct streaming from medical linac rooms Takuma Noto * Kazuaki Kosako and Takashi

More information

A comprehensive study on energy absorption and exposure buildup factors for some Vitamins and Tissue Equivalent Materials

A comprehensive study on energy absorption and exposure buildup factors for some Vitamins and Tissue Equivalent Materials International Journal of Scientific and Research Publications, Volume 2, Issue 9, September 2012 1 A comprehensive study on energy absorption and exposure buildup factors for some Vitamins and Tissue Equivalent

More information

Neutron Imaging at Spallation Neutron Sources

Neutron Imaging at Spallation Neutron Sources Neutron Imaging at Spallation Neutron Sources E.H. LEHMANN, A. KAESTNER Paul Scherrer Institut, Deptm. Spallation Neutron Source, Switzerland OUTLINE 1. Introduction: Motivation for Neutron Imaging 2.

More information

Combined neutron and X-ray imaging on different length scales

Combined neutron and X-ray imaging on different length scales More Info at Open Access Database www.ndt.net/?id=18740 Combined neutron and X-ray imaging on different length scales Anders Kaestner, David Mannes, Jan Hovind, Pierre Boillat, and Eberhard Lehmann Neutron

More information

Brewster Angle and Total Internal Reflection

Brewster Angle and Total Internal Reflection Lecture 4: Polarization Outline 1 Polarized Light in the Universe 2 Brewster Angle and Total Internal Reflection 3 Descriptions of Polarized Light 4 Polarizers 5 Retarders Christoph U. Keller, Utrecht

More information

Decoherence in QM physics 491 fall 2009 friday feature M. Gold updated Oct. 1, 2015

Decoherence in QM physics 491 fall 2009 friday feature M. Gold updated Oct. 1, 2015 diffraction experiment Decoherence in QM physics 491 fall 2009 friday feature M. Gold updated Oct. 1, 2015 In reality, it contains the only mystery, the basic peculiarities of all of quantum mechanics.

More information