High Precision Pb Isotope Techniques from the WHOI NEPTUNE PIMMS

Size: px
Start display at page:

Download "High Precision Pb Isotope Techniques from the WHOI NEPTUNE PIMMS"

Transcription

1 High Precision Pb Isotope Techniques from the WHOI NEPTUNE PIMMS S.R. Hart, R.K. Workman, L. Ball and J. Blusztajn Woods Hole Oceanographic Institution Woods Hole MA 02543, USA WHOI Plasma Facility Open File Report 10

2 1. NEPTUNE NUTS AND BOLTS NBS 981 NEBULIZER 0.12 ml/minute. LEAD SENSITIVITY volts/ppm. TYPICAL RUN INTENSITY volts. RUN LENGTH 9 to 18 minutes. 204Hg CORRECTION ppm of mass 204. ABUNDANCE SENSITIVITY ppm of 205Tl at mass 204; for Pb/Tl = 3, correction on 204Pb ~ 35 ppm. EFFECT OF Pb/Tl RATIO less than 70 ppm from Pb/Tl = 2.5 to Pb/Tl = 12. INTERNAL PRECISION (2s) ppm on 20X/204 ratios. EXTERNAL PRECISION (2s, n = 38) 80 ppm on and 58 ppm on 206/207 Pb. SUMMARY - EXTERNAL PRECISION (2-Sigma) RATIO NBS 981 SAMOA BASALTS 206/ ppm 101 ppm 207/ / / / of 7

3 2. Pb Isotopes by NEPTUNE PIMMS 8000 Thallium Tailing, NEPTUNE Counts per Second Hg 1.9 ppm of 205 Tl 205 Tl = 22 volts MASS, amu Figure 1. The abundance sensitivity of the NEPTUNE at mass 204 is ~ 1.7 ppm/amu (down-mass) and ~0.8 ppm/amu (up-mass); this leads to ~1.9 ppm from combined 203Tl and 205Tl. For Pb/Tl ratios greater than 3, the effect on 204Pb is less than 40 ppm. By using baselines at 0.5 amu, tailing effects are insignificant (< 10 ppm for 20X/204 Pb ratios). 3 of 7

4 205/203 Tl Thallium Mass Bias versus Time 7/30/02-8/05/02 3/22/02-4/01/02 8/29/02-8/30/ Run Number Figure 2. The variation of mass bias with time on the NEPTUNE is quite small, as shown by the time variation of 205Tl/203Tl for Pb runs spiked with NBS 997. The average mass bias for 37 runs over a period of 6 months is +0.79%/amu; the standard deviation of these runs (1s) is 860 ppm Pb/ 206 Pb in NBS981 versus Pb/Tl Ratio D 207Pb/206Pb A C B A - 0.5N HCl B - 1% Seawater C - 33 ppb Pb D - ± 50 ppm Pb/205Tl Figure 3. Because of the excellent abundance sensitivity and mass bias stability of the NEPTUNE, there is virtually no effect of varying Pb/Tl ratios on the external precision. For example, there is no statistical variation of 207Pb/206Pb over a Pb/Tl range from 2.5 to 12 (208Pb/205Tl = ). Based on initial tests, there is also no obvious matrix effect between HNO3 (typical run solution), HCl and weak seawater. 4 of 7

5 SRM 981 NEPTUNE Runs [ Diamonds - 3/22/02; Squares -7/30/02; Triangles - 8/29/02 ] fractionation line ±100ppm in 206/ ±100ppm in 207/ ±50ppm in 207/ Figure 4. External precision of 37 runs of the NBS 981 standard (14 days, 5 sessions, 6 months) is 80 ppm for 206Pb/204Pb and 94 ppm for 207Pb/204Pb respectively (2s). The data are well aligned along a fractionation line, showing that the external precision is principally limited by the Tl massbias correction. However, these runs also encompass the use of a variety of baseline protocols during development of the technique; these contribute to variability along a 204Pb error line (which is flatter but sub-parallel to the fractionation line). Use of a consistent baseline protocol is expected to improve external precision for ratios involving 204Pb. For illustration, an error polygon is drawn for ±100ppm errors in 206/204 and 207/204, and ±50ppm in 207/ SRM COMPARISON WITH LITERATURE VALUES Diamonds - TIMS Squares - PIMMS grand mean NBS Certified value ppm/amu ppm/amu WHOI NEPTUNE fractionation line Figure 5. A comparison of NEPTUNE data on NBS 981 with other literature values, both TIMS and PIMMS, shows that all data is aligned closely along a fractionation line, but there is ~ 480ppm/amu spread along the fractionation line. Some of this relates to the error uncertainties assigned to the absolute values of the NBS Pb and Tl standards (180ppm/amu and 210ppm/amu respectively) that are used to calibrate fractionation effects. 5 of 7

6 3. SAMPLE ANALYSES Savai'i Basalts - Replicate Pb Analyses S15 ± 100ppm in 207/204 S25 S Figure 6. External precision of repeat runs on three Samoan basalts; filled diamonds are single runs on leached samples; diamonds connected by bars are runs on different days of a single chemistry (unleached samples), repeat solutions, 2s = 90 ppm for 207Pb/204Pb. Runs on duplicate chemistries are not as precise, as there is obvious sample heterogeneity (see especially S25). Sample heterogeneity issues may now be the limiting factor in Pb isotope studies, given the large improvement in mass spec precision afforded by PIMMS and double/triple spike TIMS techniques Western Samoan Lavas Savai'i TIMS Savai'i Neptune Upolu TIMS Upolu Neptune Figure 7. High-precision NEPTUNE Pb data for basalts from the Upolu and Savai i shields, compared to literature data for Upolu shield basalts and post-erosional basalts from Savai i. The new Upolu data form a tight array with a normal positive slope, as a lower bound to the existing field; there is no evidence for a large dispersion in 207Pb/204Pb as seen in the existing data. There is no existing data for Savai i shield basalts, but our new data bound the field for post-erosional basalts, with much less scatter. The negative slope is very unusual for OIB lavas, and reflects mixing with a high, probably ancient, D7/4 Pb component. 6 of 7

7 4. ACKNOWLEDGMENTS The NSF/EAR Instrumentation and Facilities Program was instrumental in facilitating the acquisition of a NEPTUNE multi-collector ICP/MS at WHOI. The robust health of analytical geochemistry in the US is due largely to this NSF Program, carefully nurtured over several decades by Dan Weill s vision and leadership. Our enduring thanks to Dan for this contribution. We also appreciate the patient and detailed help given to us by Wafa Abouchami and Steve Galer, during our conversion to the Mainz Pb chemistry. NEPTUNE steerage at WHOI has been a joint collaboration involving also Ken Sims, Bernhard Peucker-Ehrenbrink, Roger Francois and Simon Thorrold. For reference, also see Hart et al., EOS 83, F20, of 7

Uncertainty in Measurement of Isotope Ratios by Multi-Collector Mass Spectrometry

Uncertainty in Measurement of Isotope Ratios by Multi-Collector Mass Spectrometry 1 IAEA-CN-184/168 Uncertainty in Measurement of Isotope Ratios by Multi-Collector Mass Spectrometry R. Williams Lawrence Livermore National Laboratory Livermore, California U.S.A. williams141@llnl.gov

More information

Precise Pb isotope analysis of igneous rocks using fully-automated double spike thermal ionization mass spectrometry (FA -DS- TIMS)

Precise Pb isotope analysis of igneous rocks using fully-automated double spike thermal ionization mass spectrometry (FA -DS- TIMS) JAMSTEC-R IFREE Special Issue, November 2009 Precise Pb isotope analysis of igneous rocks using fully-automated double spike thermal ionization mass spectrometry (FA -DS- TIMS) Takashi Miyazaki 1*, Nobuyuki

More information

Low level Os isotopic measurements using multiple ion counting

Low level Os isotopic measurements using multiple ion counting APPLICATION NOTE 30355 Low level Os isotopic measurements using multiple ion counting Authors Introduction Jean Louis Birck, Delphine Limmois, Institut de Physique du Globe de Paris, Sorbonne Paris Cité,

More information

Evaluating the precision of Pb isotope measurement by mass spectrometry

Evaluating the precision of Pb isotope measurement by mass spectrometry Evaluating the precision of Pb isotope measurement by mass spectrometry Journal: Journal of Analytical Atomic Spectrometry Manuscript ID: JA-PER-0--000.R Article Type: Paper Date Submitted by the Author:

More information

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS Application note Environmental Authors Steve Wilbur, Craig Jones Agilent Technologies,

More information

G 3. Matthew G. Jackson Woods Hole Oceanographic Institution, Woods Hole, Massachusetts , USA

G 3. Matthew G. Jackson Woods Hole Oceanographic Institution, Woods Hole, Massachusetts , USA Geosystems G 3 AN ELECTRONIC JOURNAL OF THE EARTH SCIENCES Published by AGU and the Geochemical Society Article Volume 10,Number3 11 March 2009 Q03006, doi: ISSN: 1525-2027 Click Here for Full Article

More information

G 3. AN ELECTRONIC JOURNAL OF THE EARTH SCIENCES Published by AGU and the Geochemical Society

G 3. AN ELECTRONIC JOURNAL OF THE EARTH SCIENCES Published by AGU and the Geochemical Society Geosystems G 3 AN ELECTRONIC JOURNAL OF THE EARTH SCIENCES Published by AGU and the Geochemical Society Article Volume 9, Number 4 1 April 2008 Q04003, doi:10.1029/2007gc001651 ISSN: 1525-2027 Click Here

More information

ARTICLE IN PRESS. Measuring 87 Sr/ 86 Sr variations in minerals and groundmass from basalts using LA-MC-ICPMS

ARTICLE IN PRESS. Measuring 87 Sr/ 86 Sr variations in minerals and groundmass from basalts using LA-MC-ICPMS DTD 5 Chemical Geology xx (2004) xxx xxx www.elsevier.com/locate/chemgeo Measuring 87 Sr/ 86 Sr variations in minerals and groundmass from basalts using LA-MC-ICPMS Frank C. Ramos a, *, John A. Wolff b,

More information

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS Enhancing the productivity of food sample analysis with the Agilent 77x ICP-MS Application note Foods testing Authors Sebastien Sannac, Jean Pierre Lener and Jerome Darrouzes Agilent Technologies Paris,

More information

Nu Plasma II - Collector Configuration

Nu Plasma II - Collector Configuration Nu Plasma II - Collector Configuration Nu Plasma II - Collector Configuration U-Th-Pb mass array 238U 232Th 207Pb 206Pb 208Pb 202Hg 204Pb, 204Hg Peak Alignment/Coincidence 238 U 232 Th 208 Pb 207 Pb 206

More information

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis The Agilent 77x ICP-MS Advantage for Drinking Water Analysis Application Note Environmental Authors Steve Wilbur Agilent Technologies, Inc. 338 146th Place SE, Suite 3, Bellevue Washington, 987 USA Introduction

More information

Comparison of Th, Sr, Nd and Pb isotopes in oceanic basalts: Implications for mantle heterogeneity and magma genesis

Comparison of Th, Sr, Nd and Pb isotopes in oceanic basalts: Implications for mantle heterogeneity and magma genesis Earth and Planetary Science Letters 245 (2006) 743 761 www.elsevier.com/locate/epsl Comparison of Th, Sr, Nd and Pb isotopes in oceanic basalts: Implications for mantle heterogeneity and magma genesis

More information

Measurement of 234 U/ 238 U and 230 Th/ 232 Th in volcanic rocks using the Neptune MC-ICP-MS

Measurement of 234 U/ 238 U and 230 Th/ 232 Th in volcanic rocks using the Neptune MC-ICP-MS PAPER www.rsc.org/jaas Journal of Analytical Atomic Spectrometry Measurement of 234 U/ 238 U and 230 Th/ 232 Th in volcanic rocks using the Neptune MC-ICP-MS Lary Ball,* a Kennneth W. W. Sims a and Johannes

More information

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS ICP-MS Environmental Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS Application Note Steve Wilbur Analysis of arsenic and selenium in seawater at trace

More information

ELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 1) Module 4

ELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 1) Module 4 ELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 1) Module 4 Tatiana Trejos, M.Sc Florida International University Department of Chemistry and Biochemistry International Forensic Research Institute Outline

More information

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Application Note Semiconductor Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Tokyo, Japan Introduction

More information

Error Analysis. The big questions are: - What value best represents a set of measurements and how reliable is it?

Error Analysis. The big questions are: - What value best represents a set of measurements and how reliable is it? The big questions are: - What value best represents a set of measurements and how reliable is it? An ICP analysis yields LOTS of data typically, 9 individual measurements for each analyte for each sample

More information

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast Application Note Environmental Authors Austin Schultz and Jake Unnerstall, Elemental Scientific, Omaha NE, USA Steve Wilbur,

More information

Use of ICP-MS in analysing radioisotopes. Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark

Use of ICP-MS in analysing radioisotopes. Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark Use of ICP-MS in analysing radioisotopes Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark Inductively Coupled Plasma Mass Spectrometry (ICP-MS) History ICP-AES

More information

2. QUALITY CONTROL IN MASS SPECTROMETRY

2. QUALITY CONTROL IN MASS SPECTROMETRY IAEA-SM-367/5/03 Evaluation of Uncertainties for Pu and U Measurements Achieved in the On-Site Laboratory by Thermal Ionisation Mass Spectrometry during Two Years of Operation E. Zuleger, K. Mayer, L.

More information

When good zircons go bad Redistribution of Radiogenic Pb in Granulite Grade Zircon, Snowbird Tectonic Zone, Canada

When good zircons go bad Redistribution of Radiogenic Pb in Granulite Grade Zircon, Snowbird Tectonic Zone, Canada When good zircons go bad Redistribution of Radiogenic Pb in Granulite Grade Zircon, Snowbird Tectonic Zone, Canada Nicole Rayner, Bill Davis, Tom Pestaj Geological Survey of Canada, SHRIMP Lab 2 J.C. Roddick

More information

Samples compliant with the WATER

Samples compliant with the WATER ENVIRONMENTAL analysis A routine method for the Quantitative Measurement of trace metals in WATER Samples compliant with the WATER Quality Standard EN ISO 17294 on the Agilent 7000 Series ICP-MS. Solutions

More information

GSA Data Repository Denyszyn, et al., 2018, A bigger tent for CAMP: Geology,

GSA Data Repository Denyszyn, et al., 2018, A bigger tent for CAMP: Geology, GSA Data Repository 2018306 Denyszyn, et al., 2018, A bigger tent for CAMP: Geology, https://doi.org/10.1130/g45050.1 SPPLEMENTARY FILE: Methods and Data Geochemistry Methods Bulk-rock compositions of

More information

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) Application Note Environmental Authors Steve Wilbur Agilent Technologies,

More information

ICP/MS Multi-Element Standards

ICP/MS Multi-Element Standards Standards Ultra Pure Matrix Special Packaging Traceability to National Reference Materials AccuStandard s ICP/MS Standards are formulated to meet the needs of this very special instrument. As matrix effect

More information

of numerous north-south trending Miocene and volcanic centres and lava flows. For

of numerous north-south trending Miocene and volcanic centres and lava flows. For DATA REPOSITORY DR009071 Duggen et al. SAMPLES AND ANALYTICAL METHODS Lavas were collected from a ca. 10 km by 100 km volcanic field consisting of numerous north-south trending Miocene and volcanic centres

More information

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Application note Energy and fuels Authors Jenny Nelson, Agilent Technologies, USA Ed McCurdy, Agilent Technologies, UK Introduction

More information

Automating Sample Purification MC-ICPMS. Elemental Scientific ICP ICPMS AA

Automating Sample Purification MC-ICPMS. Elemental Scientific ICP ICPMS AA prepfast-mc Automating Sample Purification MC-ICPMS Elemental Scientific ICP ICPMS AA 1 prepfast-mc Brief and system schematics...2 Uranium Isotopes...3 Strontium Isotopes...6 Boron Isotopes...9 Brief

More information

Evidence for the return of subducted continental crust in Samoan lavas

Evidence for the return of subducted continental crust in Samoan lavas Evidence for the return of subducted continental crust in Samoan lavas Matthew G. Jackson 1, Stanley R. Hart 2, Anthony A.P. Koppers 3,4, Hubert Staudigel 3, Jasper Konter 3, Jerzy Blusztajn 2, Mark Kurz

More information

PSRD: A Complication in Determining the Precise Age of the Solar System

PSRD: A Complication in Determining the Precise Age of the Solar System January 21, 2010 A Complication in Determining the Precise Age of the Solar System --- The presence of short-lived isotope Curium-247 in the early Solar System complicates the job of dating the earliest

More information

U Exploration Using Pb Isotopes : Opportunities In Partial Extraction Geochemistry

U Exploration Using Pb Isotopes : Opportunities In Partial Extraction Geochemistry U Exploration Using Pb Isotopes : Opportunities In Partial Extraction Geochemistry Graham Carr (CSIRO) And Neil Rutherford (Rutherford Mineral Resource Consultants) Introduction Research Undertaken in

More information

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive Chapter 9: End-of-Chapter Solutions 1. The following comparison provides general trends, but both atomic absorption spectroscopy (AAS) and atomic absorption spectroscopy (AES) will have analyte-specific

More information

Institute for Research on Earth Evolution, Japan Agency for Marine-Earth Science and Technology, 2-15 Natsushima-cho, Yokosuka , Japan

Institute for Research on Earth Evolution, Japan Agency for Marine-Earth Science and Technology, 2-15 Natsushima-cho, Yokosuka , Japan Geochemical Journal, Vol. 47, pp. 369 to 384, 2013 NOTE Pb isotope analyses of silicate rocks and minerals with Faraday detectors using enhanced-sensitivity laser ablation-multiple collector-inductively

More information

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode Lead isotope analysis: Removal of Hg isobaric interference from using ICP-QQQ in MS/MS mode Application note Authors Geochemistry and isotope analysis Glenn Woods Agilent Technologies, LDA UK Ltd., Stockport,

More information

Thermo Scientific Neptune XT MC-ICP-MS

Thermo Scientific Neptune XT MC-ICP-MS Thermo Scientific Neptune XT MC-ICP-MS Neptune XT MC-ICP-MS A new edition of the market leading Thermo Scientific Neptune Series MC-ICP-MS, capturing the best of technology for high-precision isotope ratio

More information

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode The ultratrace determination of iodine in aqueous samples using the 7700x ICP-MS with oxygen reaction mode Application note Nuclear Authors Kazumi Nakano, Yasuyuki Shikamori, Naoki Sugiyama and Shinichiro

More information

ICP-OES Application Note Number 35

ICP-OES Application Note Number 35 ICP-OES Application Note Number 35 Rapid measurement of major, minor and trace levels in soils using the Varian 730-ES Vincent Calderon Varian, Inc. Introduction As part of the global strategy for sustainable

More information

TRACE ELEMENT ANALYSIS OF DIAMOND BY LAM ICPMS: STANDARDISATION, RESULTS AND DIRECTIONS

TRACE ELEMENT ANALYSIS OF DIAMOND BY LAM ICPMS: STANDARDISATION, RESULTS AND DIRECTIONS TRACE ELEMENT ANALYSIS OF DIAMOND BY LAM ICPMS: STANDARDISATION, RESULTS AND DIRECTIONS W.L. Griffin 1, 3, Sonal Rege 1, Rondi M. Davies 1, 2, Simon Jackson 1, Suzanne Y. O Reilly 1 1.ARC National Key

More information

A new variable dispersion double-focusing plasma mass spectrometer with performance illustrated for Pb isotopes

A new variable dispersion double-focusing plasma mass spectrometer with performance illustrated for Pb isotopes International Journal of Mass Spectrometry 181 (1998) 51 58 A new variable dispersion double-focusing plasma mass spectrometer with performance illustrated for Pb isotopes N.S. Belshaw a, *, P.A. Freedman

More information

G 3. AN ELECTRONIC JOURNAL OF THE EARTH SCIENCES Published by AGU and the Geochemical Society

G 3. AN ELECTRONIC JOURNAL OF THE EARTH SCIENCES Published by AGU and the Geochemical Society Geosystems G 3 AN ELECTRONIC JOURNAL OF THE EARTH SCIENCES Published by AGU and the Geochemical Society Article Volume 4, Number 10 4 October 2003 8621, doi:10.1029/2002gc000443 ISSN: 1525-2027 Aberrant

More information

PERSPECTIVE STABLE ISOTOPE RATIO MS

PERSPECTIVE STABLE ISOTOPE RATIO MS Attom High Resolution ICP-MS PERSPECTIVE STABLE ISOTOPE RATIO MS Nu Plasma II Multi-Collector ICP-MS Panorama HR Stable Isotope Ratio MS Astrum Glow Discharge MS Nu Plasma 1700 Multi-Collector ICP-MS Horizon

More information

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell Trace elemental analysis of distilled alcoholic beverages using the Agilent 77x ICP-MS with octopole collision/ reaction cell Application note Food testing Author Glenn Woods Agilent Technologies Cheadle

More information

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution APEX-ACM Ca Ratios Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution Abstract High resolution ICP-MS is used

More information

Fergus Keenan Shona McSheehy Julian Wills

Fergus Keenan Shona McSheehy Julian Wills Novel QCell Technology for Inference Removal in ICP-MS - Combining Low Mass Filtration with Kinetic Energy Discrimination Fergus Keenan Shona McSheehy Julian Wills 1 The world leader in serving science

More information

Yongsheng He (1)*, YangWang (1), ChuanweiZhu (1), ShichunHuang (2) and Shuguang Li (1) Vol. 41 N p

Yongsheng He (1)*, YangWang (1), ChuanweiZhu (1), ShichunHuang (2) and Shuguang Li (1) Vol. 41 N p Vol. 41 N 2 06 17 p.283 302 Mass-Independent and Mass-Dependent Ca Isotopic Compositions of Thirteen Geological Reference Materials Measured by Thermal Ionisation Mass Spectrometry Yongsheng He (1)*, YangWang

More information

Chem 321 Lecture 4 - Experimental Errors and Statistics 9/5/13

Chem 321 Lecture 4 - Experimental Errors and Statistics 9/5/13 Chem 321 Lecture 4 - Experimental Errors and Statistics 9/5/13 Student Learning Objectives Experimental Errors and Statistics The tolerances noted for volumetric glassware represent the accuracy associated

More information

CERTIFICATE OF. Single Analyte Mass Spec Solution ± 0.1 µg/ml g/ml (measured at 20 ± 4 ºC)

CERTIFICATE OF. Single Analyte Mass Spec Solution ± 0.1 µg/ml g/ml (measured at 20 ± 4 ºC) CERTIFICATE OF ANALYSIS 300 Technology Drive Christiansburg, VA 24073. USA inorganicventures.com tel: 800.669.6799. 540.585.3030 fax: 540.585.3012 info@inorganicventures.com 1.0 ACCREDITATION / REGISTRATION

More information

Th- 234 U model ages of some uranium standard reference materials

Th- 234 U model ages of some uranium standard reference materials Proc. Radiochim. Acta 1, 31 35 (2011) / DOI 10.1524/rcpr.2011.0005 by Oldenbourg Wissenschaftsverlag, München Th- U model ages of some uranium standard reference materials By R. W. Williams and A. M. Gaffney

More information

Additional Analytical Methods. Detrital zircon samples were collected from nine fine-, medium-, and coarse-grained

Additional Analytical Methods. Detrital zircon samples were collected from nine fine-, medium-, and coarse-grained GSA Data Repository 2016214 Leary, R.J., DeCelles, P.G., Quade, J., Gehrels, G.E., and Waanders, G., 2016, The Liuqu Conglomerate, southern Tibet: Early Miocene basin development related to deformation

More information

New Samoan lavas from Ofu Island reveal a hemispherically heterogeneous high 3 He/ 4 He mantle

New Samoan lavas from Ofu Island reveal a hemispherically heterogeneous high 3 He/ 4 He mantle Available online at www.sciencedirect.com Earth and Planetary Science Letters 264 (2007) 360 374 www.elsevier.com/locate/epsl New Samoan lavas from Ofu Island reveal a hemispherically heterogeneous high

More information

Metals Determinations with High Accuracy and Traceable Uncertainty

Metals Determinations with High Accuracy and Traceable Uncertainty 32 nd FACSS/51 st ICASS Conference Metals Determinations with High Accuracy and Traceable Uncertainty T. M. Rettberg, S. Evans Norris, J. MacIntosh VHG Labs 276 Abby Road Manchester, NH 03103 USA Quantitative

More information

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED Application Note: 40852 Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED - Interference-free analysis of 29 elements in environmental

More information

Trace Elements - Definitions

Trace Elements - Definitions Trace Elements - Definitions Elements that are not stoichiometric constituents in phases in the system of interest For example, IG/MET systems would have different trace elements than aqueous systems Do

More information

Sample Analysis Design PART II

Sample Analysis Design PART II Sample Analysis Design PART II Sample Analysis Design Generating high quality, validated results is the primary goal of elemental abundance determinations It is absolutely critical to plan an ICP-MS analysis

More information

EAS4561/6560 Isotope Geochemistry

EAS4561/6560 Isotope Geochemistry 1. Use Dodson s equation (equation 2.37) to calculate the closure temperatures of biotite for the cases of a slowly cooled intrusion discussed in Section 2.3.1, namely at 10 /Ma and 100 /Ma. Use the data

More information

Actinides in Human Urine by Alpha Pulse Height Analysis (PHA)

Actinides in Human Urine by Alpha Pulse Height Analysis (PHA) Actinides in Human Urine by Alpha Pulse Height Analysis (PHA) Brian K. Culligan Fellow Scientist April 20, 2012 Health Physics Society Meeting Aiken SC SRNS-L4600-2012-00040 1 Outline Basic Principals

More information

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Application note Pharmaceutical Authors Miao Jing, Yingping Ni, Yanping Wang and Zhixu Zhang Agilent Technologies, China Introduction

More information

Sediment Transport and Strata Formation in the Adriatic Sea

Sediment Transport and Strata Formation in the Adriatic Sea Sediment Transport and Strata Formation in the Adriatic Sea Wayne R. Geyer James D. Irish Peter A. Traykovski Woods Hole Oceanographic Institution Woods Hole, MA 02543 Tel. (508) 289-2868, Fax: (508) 457-2194,

More information

PERSPECTIVE STABLE ISOTOPE RATIO MS

PERSPECTIVE STABLE ISOTOPE RATIO MS Nu Plasma II Multi-Collector ICP-MS Aspect HR Stable Isotope Ratio MS PERSPECTIVE STABLE ISOTOPE RATIO MS Astrum Glow Discharge MS Nu Plasma 1700 Multi-Collector ICP-MS Panorama HR Stable Isotope Ratio

More information

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Application note Semiconductor Authors Michiko Yamanaka, Kazuo Yamanaka and Naoki

More information

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS Authors Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS Application Note Food safety Craig Jones and Jenny Nelson Agilent Technologies, USA Introduction In the U.S., marijuana remains

More information

MM800 (a) Ion Exchange and ICP/MS of Uranium in Water. 1.0 Scope and Application

MM800 (a) Ion Exchange and ICP/MS of Uranium in Water. 1.0 Scope and Application Analytical/Inorganic MM800 (a) Ion Exchange and ICP/MS of Uranium in Water 1.0 Scope and Application This procedure can be used to determine U concentration or isotopic-ratio composition in groundwater

More information

Exploring new fields of Applications using Multi Collector Magnetic Sector MS

Exploring new fields of Applications using Multi Collector Magnetic Sector MS The world leader in serving science Exploring new fields of Applications using Multi Collector Magnetic Sector MS Sylveer Bergs Luca Simonotti Thermo Fisher Scientific Modena 28. & 29 Mai 2009 Introduction

More information

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Tran T. Nham Introduction

More information

Finnigan TRITON Finnigan NEPTUNE. High Resolution Multicollector Mass Spectrometers

Finnigan TRITON Finnigan NEPTUNE. High Resolution Multicollector Mass Spectrometers m a s s s p e c t r o m e t r y Finnigan TRITON Finnigan NEPTUNE High Resolution Multicollector Mass Spectrometers Analyze Detect Measure Control Finnigan TRITON / Finnigan NEPTUNE High Resolution Multicollector

More information

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3 Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3 Author Application Environmental Tetsuo Kubota Agilent Technologies, Japan Introduction Regulations concerning activities such

More information

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company Why ICP Mass Spectrometry? Ultra-trace multi-element analytical

More information

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Application Note Pharmaceutical Authors Rebeca Santamaria-Fernandez, SheilaMerson,

More information

Hydra-C Application Note: 1076

Hydra-C Application Note: 1076 Hydra-C Application Note: 1076 The Determination of Mercury in Samples by U.S. EPA SOW 846 Method 7473 using the Hydra-C Mercury Analyzer Introduction The accurate determination of mercury in various samples

More information

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT#

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT# 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

PUBLICATIONS. Geochemistry, Geophysics, Geosystems

PUBLICATIONS. Geochemistry, Geophysics, Geosystems PUBLICATIONS Geochemistry, Geophysics, Geosystems RESEARCH ARTICLE Key Points: Ofu Island in Samoa has the highest 3He/4He ratio measured in the Pacific Major and trace element data for Ofu are presented

More information

Single Analyte Custom Grade Solution ± 4 µg/ml g/ml (measured at 20 ± 1 ºC) Acidimetric NIST SRM 84L Lot Number: 84L

Single Analyte Custom Grade Solution ± 4 µg/ml g/ml (measured at 20 ± 1 ºC) Acidimetric NIST SRM 84L Lot Number: 84L 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C Book 5 - Laboratory Policies and Procedures C03-1 2009-06-15 1 Scope 1.1 This method describes a general procedure for the determination of leachable arsenic, selenium, cadmium, antimony, and barium in

More information

DETERMINATION OF TEMPERATURE DEPENDENT STABLE

DETERMINATION OF TEMPERATURE DEPENDENT STABLE DETERMINATION OF TEMPERATURE DEPENDENT STABLE STRONTIUM ISOTOPE ( 88 SR/ 86 SR) FRACTIONATION VIA BRACKETING STANDARD MC-ICP-MS J. Fietzke & A. Eisenhauer Leibniz-Institut für Meereswissenschaften (IfM-GEOMAR),

More information

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Comparative analysis of RNA- Seq data with DESeq2

Comparative analysis of RNA- Seq data with DESeq2 Comparative analysis of RNA- Seq data with DESeq2 Simon Anders EMBL Heidelberg Two applications of RNA- Seq Discovery Eind new transcripts Eind transcript boundaries Eind splice junctions Comparison Given

More information

FRONTIER RESEARCH ON EARTH EVOLUTION, VOL. 2. Research Program for Data and Sample Analyses, Institute For Research on Earth Evolution (IFREE) 2

FRONTIER RESEARCH ON EARTH EVOLUTION, VOL. 2. Research Program for Data and Sample Analyses, Institute For Research on Earth Evolution (IFREE) 2 Evaluation of silica-gel activator in order to find the optimal silica-gel activator for lead isotope measurement by thermal ionization mass spectrometer (TIMS) Takashi Miyazaki 1, Tomoyuki Shibata 2,

More information

Examples of Method Validation Studies Conducted in Different Economies

Examples of Method Validation Studies Conducted in Different Economies Examples of Method Validation Studies Conducted in Different Economies Template for group discussion National Metrology Laboratory, Malaysia 8th APMP TCQM-DEC MiC Workshop, Kobe, Japan December 2011 Description

More information

Automated Determination of PPQ Levels of Thorium in High Purity Copper

Automated Determination of PPQ Levels of Thorium in High Purity Copper PPQ Levels of Th in High Purity Cu TRUFAST ICPMS Automated Determination of PPQ Levels of Thorium in High Purity Copper Using the ESI TRUFAST System and ICPMS Detection By Nathan Saetveit, PhD and Dan

More information

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information: 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Oddo-Harkins rule of element abundances

Oddo-Harkins rule of element abundances Page 1 of 5 Oddo-Harkins rule of element abundances To instructors This is a simple exercise that is meant to introduce students to the concept of isotope ratios, simple counting statistics, intrinsic

More information

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry APPLICATION NOTE Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry AN43227 Antonella Guzzonato 1, Shona McSheehy Ducos

More information

Single Analyte Custom Grade Solution. CGWH2O1 M2-W H2O µg/ml ea: Tungsten Ammonium Metatungstate

Single Analyte Custom Grade Solution. CGWH2O1 M2-W H2O µg/ml ea: Tungsten Ammonium Metatungstate 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

Rapid Detection of Americium-241 in Food by Inductively-Coupled Plasma Mass Spectrometry

Rapid Detection of Americium-241 in Food by Inductively-Coupled Plasma Mass Spectrometry Rapid Detection of Americium-241 in Food by Inductively-Coupled Plasma Mass Spectrometry Zhichao Lin, Kathryn Emanuele, Stephanie Healey, and Patrick Regan Analytical Branch Winchester Engineering and

More information

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction. Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS Application Environmental Author Steven Wilbur Agilent Technologies, Inc. 3380 146th Place, SE, Suite 300 Bellevue, WA

More information

Appendix 1. Supplementary data presented here include isotopic and concentration data for

Appendix 1. Supplementary data presented here include isotopic and concentration data for 267 Appendix 1 Chronology of Pluton Emplacement and Regional Deformation in the Southern Sierra Nevada Batholith, California-Supplementary Data and Discussions Saleeby, J., Division of Geological and Planetary

More information

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT# Calculated See Sec. 4.

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT# Calculated See Sec. 4. 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

LAACHER SEE REVISITED: HIGH SPATIAL RESOLUTION ZIRCON DATING IMPLIES RAPID FORMATION OF A ZONED MAGMA CHAMBER -

LAACHER SEE REVISITED: HIGH SPATIAL RESOLUTION ZIRCON DATING IMPLIES RAPID FORMATION OF A ZONED MAGMA CHAMBER - LAACHER SEE REVISITED: HIGH SPATIAL RESOLUTION ZIRCON DATING IMPLIES RAPID FORMATION OF A ZONED MAGMA CHAMBER - DATA REPOSITORY ANALYTICAL PROCEDURES Ion microprobe U-Th measurements Th-U dating was performed

More information

What can isotopes tell us about mantle dynamics? Sujoy Mukhopadhyay. Harvard University

What can isotopes tell us about mantle dynamics? Sujoy Mukhopadhyay. Harvard University What can isotopes tell us about mantle dynamics? Sujoy Mukhopadhyay Harvard University The mantle zoo Hofmann, 1997 187 Os/ 188 Os 0.168 0.156 0.144 0.132 EM1 Hawaii Pitcairn DMM peridotites Shield Basalts

More information

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors.

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors. Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology Part 2 of a 3 part series on Environmental Analysis Application Environmental Authors

More information

V. LAB REPORT. PART I. ICP-AES (section IVA)

V. LAB REPORT. PART I. ICP-AES (section IVA) V. LAB REPORT The lab report should include an abstract and responses to the following items. All materials should be submitted by each individual, not one copy for the group. The goal for this part of

More information

Oceanic plateau subduction during closure of Bangong-Nujiang Tethys: Insights from Central Tibetan volcanic rocks

Oceanic plateau subduction during closure of Bangong-Nujiang Tethys: Insights from Central Tibetan volcanic rocks GSA Data Repository Item 2018295 Hao, L.-L., Wang, Q., Zhang, C., Ou, Q., Yang, J.-H., Dan, W., and Jiang, Z.-Q., 2018, Oceanic plateau subduction during closure of the Bangong-Nujiang Tethyan Ocean: Insights

More information

IRMS. perspective. Isotope Ratio Mass Spectrometry.

IRMS. perspective. Isotope Ratio Mass Spectrometry. IRMS perspective Isotope Ratio Mass Spectrometry www.nu-ins.com Perspective The Nu Instruments Perspective geometry forms the basis for two state of the art IRMS instruments. Designed for flexibility,

More information

Th) ) dating of micro-baddeleyite

Th) ) dating of micro-baddeleyite U-Pb (and U-ThU Th) ) dating of micro-baddeleyite 30 μm Axel K. Schmitt UCLA SIMS, NSF National Ion Microprobe Facility Collaborators: T. Mark Harrison (UCLA) Kevin Chamberlain (University of Wyoming)

More information

Supplemental files. Supplemental file 1 analytical method

Supplemental files. Supplemental file 1 analytical method GSA Data Repository Item 2016363 Liu, H., Wang, Y., Cawood, P.A., and Guo, X., 2016, Episodic slab-rollback and back-arc extension in the Yunnan- Burma region: Insights from Cretaceous Nb-enriched and

More information

Geochemical analysis unveils frictional melting process in a

Geochemical analysis unveils frictional melting process in a GSA Data Repository 219116 1 2 3 Geochemical analysis unveils frictional melting process in a subduction zone fault Tsuyoshi Ishikawa and Kohtaro Ujiie 4 Supplemental Material 6 7 8 9 METHOD TABLES (Tables

More information

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS Tomoko Vincent 1, Simon Lofthouse 2, Daniel Kutscher 1 and Shona McSheehy Ducos

More information

Discovery of a large 2.4 Ma Plinian eruption of Basse-Terre, Guadeloupe, from the marine sediment record

Discovery of a large 2.4 Ma Plinian eruption of Basse-Terre, Guadeloupe, from the marine sediment record GSA Data Repository 2016031 Discovery of a large 2.4 Ma Plinian eruption of Basse-Terre, Guadeloupe, from the marine sediment record Martin R. Palmer et al. Methods Geochemistry 15-25 mg of volcanic grains

More information

BESIII Collaboration Meeting in Winter of 2015 BEAM ENERGY MEASUREMENT SYSTEM

BESIII Collaboration Meeting in Winter of 2015 BEAM ENERGY MEASUREMENT SYSTEM BESIII Collaboration Meeting in Winter of 2015 BEAM ENERGY MEASUREMENT SYSTEM Nickolai Muchnoi Budker INP, Novosibirsk December 12, 2015 Nickolai Muchnoi BES-III Winter Collaboration Meeting December 12,

More information