The Importance of Field Strength in the Low Field Portion of a Differential Ion Mobility Spectrometry Waveform
|
|
- Marjory Holland
- 5 years ago
- Views:
Transcription
1 The Importance of Field Strength in the Low Field Portion of a Differential Ion Mobility Spectrometry Waveform Jesus I. Martinez-Alvarado, Brandon G. Santiago, Gary L. Glish University of North Carolina, Chapel Hill, Introduction: Differential ion mobility spectrometry (DIMS) is a gas-phase separation technique that can be coupled to mass spectrometry to improve signal-to-noise ratios. DIMS separations use the dependence of ion mobility on electric field strength to separate ions. Through application of an asymmetric waveform alternating between high and low electric field strengths, the difference between high and low field mobilities is sampled. Addition of solvent vapors to the DIMS carrier gas has been shown to alter the differential ion mobility of ions and enhance the selectivity of DIMS separations. The primary explanation in the literature to explain solvent effects is the dynamic cluster-decluster model. This model rationalizes the change in differential ion mobility, and thus the compensation field required to pass the ion through DIMS, by stating that an ion travels in a solvated state during a low field portion of the separation waveform and then undergoes rf heating during the high field portion of the waveform, causing it to travel as a bare ion. Herein, we investigate the dynamic cluster-decluster model by varying the amount of time spent in the low field portion of the DIMS waveform. Experimental Methods: Experiments were performed using a Bruker HCT Quadrupole Ion Trap mass spectrometer for detection. The DIMS device consists of two planar stainless steel electrodes 4 mm wide and 1 mm long, separated by a.3 mm gap. DIMS scans were performed using a dispersion field strength of 37.5 kv/cm, with the waveform parameters as described below. The electrospray desolvation gas was used as both desolvation gas and carrier gas through DIMS by diverting the flow through the housing of the DIMS assembly. The desolvation gas flow was set to 5. L/min and either 2 or 3 C in the instrument control software. Solvent vapor from LC-MS grade methanol was introduced at a rate of.3 ml/min to the desolvation gas via a Swagelok tee connected to the output of a Hitachi LC 2 pump. Electrospray ionization was used to ionize.1% triethylamine in 9/1 MeOH/EtOH at a flow rate of 2 μl/min. Results and Discussion: The dynamic cluster-decluster model is predicated around the idea that the ion travels in a solvated state during the low field portion of the separation waveform, and after undergoing rf heating travels as a bare ion during the high field portion of the waveform. In DIMS literature, low field is typically described as fields <1 kv/cm in strength (although some claim this value to be as low as 2.5 kv/cm). Due of the power requirements of using a rectangular waveform at the required voltages for DIMS, other
2 alternatives are typically used. One such alternative is a bisinusoidal waveform, generated by coupling two sinusoidal waveforms. A drawback of using this type of waveform is that as the amplitude of the high field portion of the waveform is raised, the absolute amplitude of the low field portion of the waveform is also increased. It was observed that at field strengths where the low field portion of the waveform was >1 kv/cm the addition of methanol vapor still affected the differential ion mobility of various analyte ions. To further examine this previous observation the form parameter of the bisinusoidal waveform was varied, with example waveforms shown in Figure 1. The form parameter is defined as the amplitude of the first harmonic divided by the summed amplitude of the two harmonics, with greater form parameter yielding less time spent in the low field. At a desolvation gas setting of 3 C, it can be observed in Figure 2a that methanol changes the required compensation field to pass triethylamine ions through the DIMS device at all form parameters. A form parameter of.7 yields no net time in the low field at a dispersion field of 37.3 kv/cm, yet it is apparent that the differential ion mobility has been changed based on the shift in required compensation field. This shift is also observed at a desolvation gas setting of 2 C (Figure 2b). Additionally, Figure 2 shows that at both temperature settings form parameters generating greater time in the low field yield greater changes in the required compensation field. As expected, this effect is more prominent at 2 C where the lower effective temperature of the ions makes clustering more favorable. Further manipulation of the time spent in the low field was accomplished through varying the phase shift of the two sinusoidal waveforms relative to each other. Figure 3 depicts the changes to the waveform shape upon varying the phase shift and how time in the low field was increased. Although drawing conclusions from how the phase shift affected the required compensation field is beyond the scope of this work, it can be observed that at both temperatures and all phase shifts the use of methanol led to a change in required compensation field. As with form parameter, a greater change in required compensation field was observed at a temperature setting of 2 C, presumably due to the more favorable clustering kinetics present at lower ion effective temperatures. Summary and Conclusions: The form parameter and phase shift of a bisinusoidal waveform used for DIMS separations were varied to manipulate the duration of the waveform that could be described as low field. At form parameters and phase shifts that gave no or minimal net time in the low field, it was observed that methanol still changed the required compensation field. This result would not be expected based on the dynamic clusterdecluster theory. However, in agreement with dynamic cluster-decluster theory is the fact that form parameters yielding increased time in the low field showed greater changes in required compensation field. It was also shown that lower temperatures increased the shift in required compensation field caused by dopant vapors, a change that could be explained by more favorable ion kinetics for clustering at lower
3 temperatures. These results suggest another factor beyond the dynamic clusterdecluster theory plays a role in how solvent vapors affect differential ion mobility. Figure 1. Representative waveforms depicting how changing the form parameter changes the amount of time spent in the low field. Here the form parameter is changed from.57 (left) to.77 (right), and time in the low field is highlighted in yellow
4 Compensation Field (V/cm) Compensation Field (V/cm) 4 3 (a) Triethylamine 3 No MeOH Triethylamine.3 ml/min MeOH 2 (b) Form Paramter Triethylamine with no methanol Form Parameter Triethylamine Figure 2. Plots with depicting.3 ml/min the effects methanol of form parameter on required compensation field with no methanol is used (red) and when methanol vapor is added to the carrier gas (blue) at a dispersion field of 37.3 kv/cm. Experiments were performed at desolvation gas temperature settings of 3 C (a) and 2 C (b)
5 Figure 3. Representative waveforms depicting how changing the phase shift changes the amount of time spent in the low field. Time spent in the low field increases from 1 (left) to 4 (right), and is highlighted in yellow
6 Compensation Field (V/cm) Compensation Field (V/cm) 4 3 (a) Triethylamine with no methanol Waveform Triethylamine 3 with.3 ml/min methanol 2 2 (b) Triethylamine with no methanol Waveform Trielthylamine with.3 ml/min methanol Figure 4. Plots depicting the effects of phase shift on required compensation field with no methanol is used (red) and when methanol vapor is added to the carrier gas (blue) at a dispersion field of 37.3 kv/cm. Experiments were performed at desolvation gas temperature settings of 3 C (a) and 2 C (b)
B American Society for Mass Spectrometry, 2017 J. Am. Soc. Mass Spectrom. (2017) 28:2160Y2169 DOI: /s RESEARCH ARTICLE
B American Society for Mass Spectrometry, 2017 J. Am. Soc. Mass Spectrom. (2017) 28:2160Y2169 DOI: 10.1007/s13361-017-1726-8 RESEARCH ARTICLE Variables Affecting the Internal Energy of Peptide Ions During
More informationUse of the SFC-MS Splitter Kit G
Use of the SFC-MS Splitter Kit G4309-68715 Technical Note Use of the SFC- MS Splitter Kit Using the SFC- MS Splitter Kit to couple an Agilent 1260 Infinity Analytical Supercritical Fluid Chromatography
More informationAssay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer
38 Current Trends in Mass Spectrometry November 6 Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer This article demonstrates the improved
More informationIntroduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph
SCIETIFIC ISTRUMET EWS 2016 Vol. 7 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation Introduction of Hitachi Chromaster 5610 MS Detector for High Performance
More informationTANDEM MASS SPECTROSCOPY
TANDEM MASS SPECTROSCOPY 1 MASS SPECTROMETER TYPES OF MASS SPECTROMETER PRINCIPLE TANDEM MASS SPECTROMETER INSTRUMENTATION QUADRAPOLE MASS ANALYZER TRIPLE QUADRAPOLE MASS ANALYZER TIME OF FLIGHT MASS ANALYSER
More informationFAIMS Technology at a Glance
FAIMS Technology at a Glance Field asymmetric ion mobility spectrometry (FAIMS), also known as differential mobility spectrometry (DMS), is a gas detection technology that separates and identifies chemical
More informationSupporting Information. Use of Mass Spectrometric Vapor Analysis to Improve Canine Explosive Detection Efficiency
Supporting Information Use of Mass Spectrometric Vapor Analysis to Improve Canine Explosive Detection Efficiency Ta-Hsuan Ong, Ted Mendum, Geoff Geurtsen, Jude Kelley, Alla Ostrinskaya, Roderick Kunz Chemical,
More informationChapter 1. Introduction
Chapter 1. Introduction 1-1 1.1. Overview In the past twenty years, charged droplets and strong electric fields have quietly revolutionized chemistry. In combination with an atmospheric-sampling mass spectrometer,
More informationChemistry Instrumental Analysis Lecture 37. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 37 Most analytes separated by HPLC are thermally stable and non-volatile (liquids) (unlike in GC) so not ionized easily by EI or CI techniques. MS must be at
More informationMonday March
Monday March 12 2012 600-5 ASHLEY T WILKS & Billy Boyle Pittcon 2012 Orange County Convention Center Orlando, Fl Developments in Ultra FAIMS Instrumentation for Standalone and Hyphenated Applications Principles
More informationTANDEM DIFFERENTIAL MOBILITY SPECTROMETER:
TANDEM DIFFERENTIAL MOBILITY SPECTROMETER: AN IONIZATION DETECTOR FOR GC WITH HIGH SPEED, SELECTIVITY, SMALL SIZE, AND LOW COST G.A. Eiceman & M. Menlyadiev Department of Chemistry & Biochemistry New Mexico
More informationM M e M M H M M H. Ion Sources
Ion Sources Overview of Various Ion Sources After introducing samples into a mass spectrometer, the next important step is the conversion of neutral molecules or compounds to gas phase ions. The ions could
More informationMass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.
Mass Analyzers After the production of ions in ion sources, the next critical step in mass spectrometry is to separate these gas phase ions according to their mass-to-charge ratio (m/z). Ions are extracted
More information(Refer Slide Time 00:09) (Refer Slide Time 00:13)
(Refer Slide Time 00:09) Mass Spectrometry Based Proteomics Professor Sanjeeva Srivastava Department of Biosciences and Bioengineering Indian Institute of Technology, Bombay Mod 02 Lecture Number 09 (Refer
More informationSeparation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers
Application Note Forensic and Toxicology - Criminalistics Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers Using the Agilent 126 Infinity II SFC and Detection by Coupled
More informationMass Spectrometry. Fundamental LC-MS. Electrospray Ionisation Instrumentation
Mass Spectrometry Fundamental LC-MS Electrospray Ionisation Instrumentation i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be
More informationChemistry Instrumental Analysis Lecture 34. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 34 From molecular to elemental analysis there are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry
More informationClinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for
Clinical Toxicology Page of 0 Materials and Methods Biomass Component Extraction: The uneaten cooked plant specimen was prepared for chemical analysis as follows. The sample was frozen, diced, pulverized
More informationAnalysis of Polar Metabolites using Mass Spectrometry
Analysis of Polar Metabolites using Mass Spectrometry TransMed Course: Basics in Clinical Proteomics and Metabolomics. Oct 10-19, 2012 dd.mm.yyyy Vidya Velagapudi, Ph.D, Adjunct Professor Head of the Metabolomics
More informationLIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC
LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC INTRODUCTION Principle: LC/MS is a technique that combines physical separation capabilities of liquid
More informationTraditional Herbal Medicine Structural Elucidation using SYNAPT HDMS
Traditional Herbal Medicine Structural Elucidation using SYNAPT HDMS with Time-Aligned Parallel (TAP) Fragmentation Kate Yu, Jose Castro-Perez, and John Shockcor Waters Corporation, Milford, MA, U.S. INTRODUCTION
More information1. The range of frequencies that a measurement is sensitive to is called the frequency
CHEM 3 Name Exam 1 Fall 014 Complete these problems on separate paper and staple it to this sheet when you are finished. Please initial each sheet as well. Clearly mark your answers. YOU MUST SHOW YOUR
More informationisobaric Barbiturates in Serum using LDTD- MS/MS combined with differential mobility
Ultra-Fast Separation Ultra-Fast and Quantification Separation and of isobaric Barbiturates in Serum using LDTD- Quantification of isobaric MS/MS combined with differential mobility spectrometry Barbiturates
More informationMass Spectrometry in MCAL
Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS GC seperates small, volatile, non-polar material MS is detection devise (Agilent 320-MS TQ Mass Spectrometer) Full scan monitoring SIM single ion monitoring
More informationExtraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis
Application Note AN89.V.1 Extraction of Methylmalonic Acid from Serum Using ISOLUTE SAX Page 1 Extraction of Methylmalonic Acid from Serum Using ISOLUTE SAX Prior to LC-MS/MS Analysis Sample Preparation
More informationFundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics
Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass
More informationMS/MS .LQGVRI0606([SHULPHQWV
0DVV6SHFWURPHWHUV Tandem Mass Spectrometry (MS/MS) :KDWLV0606" Mass spectrometers are commonly combined with separation devices such as gas chromatographs (GC) and liquid chromatographs (LC). The GC or
More informationMechanistic Insights into Chloride Ion Detection from the Atmospheric-Pressure Afterglow of an Argon Inductively Coupled Plasma
Electronic Supplementary Material (ESI) for JAAS. This journal is The Royal Society of Chemistry 2018 Supplementary Information Mechanistic Insights into Chloride Ion Detection from the Atmospheric-Pressure
More informationTranslational Biomarker Core
Translational Biomarker Core Instrumentation Thermo Scientific TSQ Quantum Triple Quadrupole Mass Spectrometers. There are two TSQ Quantum Ultra AM instruments available in the TBC. The TSQ Quantum Ultra
More informationApplication Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany
Analysis of Pesticides by Supercritical Fluid Chromatography/ Mass Spectrometry Optimizing the Agilent 126 Infi nity Analytical SFC System in Combination with the Agilent 646 Triple Quadrupole LC/MS Application
More informationChemistry 311: Topic 3 - Mass Spectrometry
Mass Spectroscopy: A technique used to measure the mass-to-charge ratio of molecules and atoms. Often characteristic ions produced by an induced unimolecular dissociation of a molecule are measured. These
More informationBiasing Brownian motion from thermal ratchets
Mayra Vega MAE 216- Statistical Thermodynamics June 18, 2012 Introduction Biasing Brownian motion from thermal ratchets Brownian motion is the random movement of small particles however, by understanding
More informationThis document is a preview generated by EVS
TECHNICAL REPORT RAPPORT TECHNIQUE TECHNISCHER BERICHT CEN/TR 15641 August 2007 ICS 67.050 English Version Food analysis - Determination of pesticide residues by LC- MS/MS - Tandem mass spectrometric parameters
More informationMiniaturized Sensor for the Detection of Environmental Pollutants Alexander Graf 1, Ronald Stübner 1, Christian Kunath 1, Sebastian Meyer 1, Harald Schenk 1 1 Fraunhofer Institute for Photonic Microsystems
More informationToxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.
1 Supporting Information 2 3 Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish. 4 5 John Moreman, Okhyun
More informationPressure & temperature sampling options for Hiden gas analysis systems
QIC Series Gas Analysers Pressure & temperature sampling options for Hiden gas analysis systems Gas & Vapour Inlets STANDARD QIC QUARTZ INLET CAPILLARY: Sample pressure range 2 bar to 100 mbar Operates
More informationFereshteh Zandkarimi, Samanthi Wickramasekara, Jeff Morre, Jan F. Stevens and Claudia S. Maier
Chapter 2 Electrospray Ionization Traveling Wave Ion Mobility Spectrometry Mass Spectrometry for the Analysis of Plant Phenolics: An Approach for Separation of Regioisomers Fereshteh Zandkarimi, Samanthi
More informationSelecting an LC/MS Interface Becky Wittrig, Ph.D.
Selecting an LC/MS Interface Becky Wittrig, Ph.D. RESTEK CORPORATION LC/MS Interfaces I. Background of LC/MS I. Historical Perspective II. Reasons for use II. Interfaces I. Transport devices II. Particle
More informationBioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)
Liquid Chromatography Mass Spectrometry (LC/MS) Prelab Questions: Questions to be answered before doing the experiment. The answers are due at the beginning of each experiment without exception (the questions
More informationRapid Thermal Modulation Ion Spectrometry (RTMIS) Part 1
Rapid Thermal Modulation Ion Spectrometry (RTMIS) Part 1 Underlying Separation Principles and Model of Operation *Ashley Wilks, Matthew Hart, John Somerville, Andrew Kohel, David Ruiz-Alonso, Paul Boyle,
More informationBecause of its high sensitivity and selectivity, liquid
Characterization of a Temperature-Controlled FAIMS System David A. Barnett, Michael Belford, Jean-Jacques Dunyach, and Randy W. Purves* Thermo Fisher Scientific, San Jose, California, USA High-field asymmetric
More informationSecondary Ion Mass Spectroscopy (SIMS)
Secondary Ion Mass Spectroscopy (SIMS) Analyzing Inorganic Solids * = under special conditions ** = semiconductors only + = limited number of elements or groups Analyzing Organic Solids * = under special
More information2 THEORY. 2.1 Introduction. As with all analytical techniques a knowledge of the fundamental theory is necessary for a
THEORY.1 Introduction As with all analytical techniques a knowledge of the fundamental theory is necessary for a full interpretation of the results obtained. Building upon the introduction this chapter
More informationRapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS
Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS Application Note Drug Discovery Authors Srividya Kailasam Agilent
More informationDifferential Mobility Spectrometry of Derivatized Steroid Hormones: Examining the Relationships between Structures and Solvation
Differential Mobility Spectrometry of Derivatized Steroid Hormones: Examining the Relationships between Structures and Solvation Chang Liu, J. Larry Campbell, Yves Le Blanc, Subhakar Dey, Subhasish Purkayastha,
More informationMaximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System
Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Application Note Authors Kevin McCann, Sameer Nene, Doug McIntyre, Edmond Neo, Dennis Nagtalon, and
More informationSupplementary Figure 1. Structures of substrates tested with 1. Only one enantiomer is shown.
Supplementary Figure 1. Structures of substrates tested with 1. Only one enantiomer is shown. Supplementary Figure 2. CD spectra obtained using 1 and (R)-3 (blue) and (S)-3 (red) Supplementary Figure 3.
More informationUse of High Speed/High Resolution Size-Based Chromatographic Separation of Polymeric Mixtures with Offline Infrared Detection
Use of High Speed/High Resolution Size-Based Chromatographic Separation of Polymeric Mixtures with Michael O Leary 1, Jennifer Gough 1, Damian Morrison 1, Alain Creissen 2 1 Waters Corporation, Milford,
More informationMass Spectrometry. What is Mass Spectrometry?
Mass Spectrometry What is Mass Spectrometry? Mass Spectrometry (MS): The generation of gaseous ions from a sample, separation of these ions by mass-to-charge ratio, and measurement of relative abundance
More informationIon-mobility mass spectrometry of polyoxometalate Keplerate clusters and their supramolecular assemblies
SUPPLEMENTARY INFORMATION Table of Contents Summary 1. General materials and methods 2. High-resolution mass spectra 3. Ion-mobility mass spectra 4. Fragmentation of clusters in the TWIMS drift tube 5.
More informationQuantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.
Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS Application Food Analysis Authors Sheher Mohsin Agilent Technologies, Inc. Schaumburg, Il USA Michael Zumwalt
More informationThe genealogical relationships between ions in a
Determination of the Dissociation Kinetics of a Transient Intermediate Michael R. Asam and Gary L. Glish University of North Carolina at Chapel Hill, Hill, North Carolina, USA Tandem mass spectrometry
More informationAccurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies
Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF Application Note Donghui Yi and Christine Miller Jon D. Williams, GlaxoSmithKline Introduction Mass spectrometry (MS) has become a core technology
More informationMolecular Mass Spectrometry
Molecular Mass Spectrometry Mass Spectrometry: capable of providing information about (1) Elemental composition of samples of matter: atomic mass (2) Structures of inorganic, organic, and biological molecules
More informationSupporting Information
1 Supporting Information 2 3 Discovery and implications of C 2 and C 3 perfluoroalkyl sulfonates in aqueous film forming foams (AFFF) and groundwater 4 Krista A. Barzen-Hanson a and Jennifer A. Field b*
More informationThe Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results
The Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results A. Kürten 1, J. Curtius 1 and S. Borrmann 1,2 1 Johannes Gutenberg-University Mainz, Germany 2 Max Planck
More informationFast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma
Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma Jon Bardsley 1, James Goldberg 2 1 Thermo Fisher Scientific, Runcorn, UK; 2 Thermo Fisher Scientific, West Palm Beach,
More informationMASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps
ION ANALYZERS MASS ANALYSER sample Vacuum pumps Mass analysers - separate the ions according to their mass-to-charge ratio MASS ANALYSER Separate the ions according to their mass-to-charge ratio in space
More informationHPLC solvents and mobile phase additives
HPLC solvents and mobile phase additives The following is intended to be a guide for the choice of solvent and mobile phase additives. Of course the choice of solvents for LC will be dictated primarily
More informationHigh-Throughput Protein Quantitation Using Multiple Reaction Monitoring
High-Throughput Protein Quantitation Using Multiple Reaction Monitoring Application Note Authors Ning Tang, Christine Miller, Joe Roark, Norton Kitagawa and Keith Waddell Agilent Technologies, Inc. Santa
More informationMS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies
MS n Analysis With Fast Polarity Switching in the Agilent Series LC/MSD Trap SL Application Note Christine Miller Introduction During mass spectrometric analysis, many compounds respond better to a particular
More informationAB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences
AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching s by Removal of Isobaric Interferences Differential Mobility Used as a Tool to Address Selectivity Challenges Adrian M. Taylor
More informationFourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS)
Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS) The first steps along the way to FT- ICR-MS date back to the 1930s, when Ernest Lawrence, at the University of California - Berkeley,
More informationTypes of Analyzers: Quadrupole: mass filter -part1
16 Types of Analyzers: Sector or double focusing: magnetic and electric Time-of-flight (TOF) Quadrupole (mass filter) Linear ion trap Quadrupole Ion Trap (3D trap) FTICR fourier transform ion cyclotron
More informationTEPZZ A_T EP A1 (19) (11) EP A1 (12) EUROPEAN PATENT APPLICATION. (51) Int Cl.:
(19) TEPZZ 6 6697A_T (11) EP 2 626 697 A1 (12) EUROPEAN PATENT APPLICATION (43) Date of publication: 14.08.2013 Bulletin 2013/33 (1) Int Cl.: G01N 30/32 (2006.01) G01N 30/20 (2006.01) (21) Application
More informationQuadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365
Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365 Effortless Performance The LCMS-9030 quadrupole time-of-flight (Q-TOF) mass spectrometer integrates the world s fastest
More informationLecture 8: Mass Spectrometry
intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene experiment CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can we get from MS spectrum?
More informationUtilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification
Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography Tips and Techniques to Optimize ELSD and MS based Purification Utilizing ELSD and MS as Secondary Detectors for Prep
More informationLC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode
LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note Authors Siji Joseph Agilent Technologies India Pvt. Ltd.
More informationThermo Finnigan LTQ. Specifications
IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Finnigan LTQ Specifications
More informationCEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis
Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511 524) (Harris, Chapt.
More informationLive Webinar : How to be more Successful with your ACQUITY QDa Detector?
Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Q&A Transcript ---------------- Q - How do you generate multiple charges reproductively? A - If you use the same settings on the
More informationCEE 772: Instrumental Methods in Environmental Analysis
Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511-524) (Harris, Chapt.
More informationLecture 8: Mass Spectrometry
intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can be obtained from a MS spectrum?
More informationYun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION
Quantitation by High Resolution Mass Spectrometry: Using Target Enhancement and Tof-MRM to Achieve Femtogram-level On-column Sensitivity for Quantitation of Drugs in Human Plasma Yun W. Alelyunas, Mark
More informationMAX300-BIO Bioreactor and Fermentation Gas Analyzer. Providing Solutions for Gas Analysis
+ MAX300-BIO Bioreactor and Fermentation Gas Analyzer Providing Solutions for Gas Analysis + Quadrupole Mass Spectrometry (QMS) For 50 years Extrel has supplied Quadrupole Mass Spectrometers systems and
More informationSupporting Information for the paper: Structure-Reactivity Correlations in Metal Atom Substitutions of Monolayer- Protected Noble Metal Alloy Clusters
Supporting Information for the paper: Structure-Reactivity Correlations in Metal Atom Substitutions of Monolayer- Protected Noble Metal Alloy Clusters Kumaranchira Ramankutty Krishnadas, Debasmita Ghosh,
More informationINTRODUCTION OVERVIEW
COMPARISON OF CCS(N 2 ) MEASUREMENTS OBTAINED FROM TWO DIFFERENT T-WAVE ION S WITH DIRECT MEASUREMENTS USING A DRIFT TUBE ION Kevin Giles, Martin Palmer, Keith Richardson, Nick Tomczyk Waters MS Technologies
More informationIDENTIFICATION OF ORGANOMETALLIC COMPOUNDS USING FIELD DESORPTION IONIZATION ON THE GCT
IDETIFICATIO OF ORGAOMETALLIC COMPOUDS USIG FIELD DESORPTIO IOIZATIO O THE GCT David Douce 1, Michael Jackson 1, Robert Lewis 1, Peter Hancock 1, Martin Green 1 and Stuart Warriner 2 1 Waters Corporation,
More informationNanoporous GaN-Ag Composite Materials Prepared by Metal-Assisted Electroless Etching
Supporting Information for Nanoporous GaN-Ag Composite Materials Prepared by Metal-Assisted Electroless Etching for Direct Laser Desorption-Ionization Mass Spectrometry Bei Nie a,, Barrett K. Duan a and
More informationChemistry Instrumental Analysis Lecture 35. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 35 Principle components: Inlet Ion source Mass analyzer Ion transducer Pumps Signal processor Mass analyzers Quadrupole Time of Flight Double Focusing Ion
More informationDEPARTMENT OF ELECTRONICS AND INSTRUMENTATION ENGINEERING SRM NAGAR, KATTANKULATHUR-603203 EI 2302 ANALYTICAL INSTRUMENTS QUESTION BANK UNIT I COLORIMETRY AND SPECTROPHOTOMETRY Part A 1. State Lambert
More informationField-Free Atmospheric Ionization: Atmospheric Pressure Chemical Ionization & Electrospray
Field-Free Atmospheric Ionization: Atmospheric Pressure Chemical Ionization & Electrospray Edward W. Sheehan 1, Ross C. Willoughby 1, Robert Classon 2, and Dan Dodgen 2 1 Chem-Space Associates, Inc., Pittsburgh,
More informationInstrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai
303351 Instrumental Analysis Mass Spectrometry Lecturer:! Somsak Sirichai Mass Spectrometry What is Mass spectrometry (MS)? An analytic method that employs ionization and mass analysis of compounds in
More informationIdentification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry
Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry Jonathan Williams Waters Corporation, Milford, MA, USA A P P L I C AT
More informationFirst described by W. Paul and H. Steinwedel in
Effect of an Axial Magnetic Field on the Performance of a Quadrupole Mass Spectrometer Sarfaraz U. Syed, Jeyan Sreekumar, Boris Brkic, John R. Gibson, and Stephen Taylor Department of Electrical Engineering
More information--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.
Translated English of Chinese Standard: GB5009.17-2014 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD OF GB THE PEOPLE S REPUBLIC OF CHINA National Food Safety Standard-Determination
More informationSupplementary Figure 1. A photographic image of directionally grown perovskite films on a glass substrate (size: cm).
Supplementary Figure 1. A photographic image of directionally grown perovskite films on a glass substrate (size: 1.5 4.5 cm). 1 Supplementary Figure 2. Optical microscope images of MAPbI 3 films formed
More informationAsymmetrical features of mass spectral peaks produced by quadrupole mass filters
RAPID COMMUNICATIONS IN MASS SPECTROMETRY Rapid Commun. Mass Spectrom. 2003; 17: 1051 1055 Published online in Wiley InterScience (www.interscience.wiley.com). DOI: 10.1002/rcm.1017 Asymmetrical features
More informationQuattro Micro - How does it work?
Quattro Micro - How does it work? 1 Introduction This document is designed to familiarise you with the principles behind how the Quattro Micro works. The level of this document is designed as Level One
More informationMolecular Mass Spectrometry
Molecular Mass Spectrometry Mass Spectrometry: capable of providing information about (1) Elemental composition of samples of matter: atomic mass (2) Structures of inorganic, organic, and biological molecules
More informationMass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed
Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS) Mass Analyzers
More informationDetermination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)
Determination of ormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell P Column (EPA 539) Application Note Environmental Authors Rong-jie Fu and Chen-ao (Andy) Zhai Agilent Technologies
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment LCQ Deca XP Plus Improved Ion Optics for Greater Sensitivity and Precision
More informationPROPERTY STUDY ON EMATS WITH VISUALIZATION OF ULTRASONIC PROPAGATION
More Info at Open Access Database www.ndt.net/?id=18576 PROPERTY STUDY ON EMATS WITH VISUALIZATION OF ULTRASONIC PROPAGATION T. Yamamoto, T. Furukawa, I. Komura Japan Power Engineering and Inspection Corporation,
More informationAccurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS
Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS Application Note Authors E. Michael Thurman and Imma Ferrer Center for Environmental
More informationMS Goals and Applications. MS Goals and Applications
MS Goals and Applications 3 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Applied Biosystems QStar Pulsar i Features of the API QSTAR Pulsar i The
More informationINDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY Edited by AKBAR MONTASER George Washington University Washington, D.C. 20052, USA WILEY-VCH New York Chichester Weinheim Brisbane Singapore Toronto CONTENTS
More informationTrapping in 2-D The Radio Frequency Quadrupole
Trapping in -D The Radio Frequency Quadrupole The Radio Frequency Quadrupole (RFQ) uses time dependent electric fields with alternating polarity to trap ions in two dimensions. These devices are generally
More information