Reliably measure trace-level contaminants and meet regulatory requirements with confidence
|
|
- Aron Doyle
- 5 years ago
- Views:
Transcription
1 Agilent J&W GC Column Portfolio For Environmental Applications Reliably measure trace-level contaminants and meet regulatory requirements with confidence From testing semi-volatile contaminants in drinking water to quantifying waterborne pollutants analyzing increasingly small quantities of active solutes must be done more reliably, more efficiently, and with higher quality results than ever before. You simply cannot afford interferences or reduced sensitivity caused by column bleed or activity. For starters, having to redo a run or verify suspect analytes wastes valuable resources, hinders productivity, and hurts your bottom line. Even worse, unreliable results could have catastrophic implications in terms of environmental safety.
2 Agilent s J&W GC column portfolio was specifically designed and application tested to help you achieve the lowest possible detection limits for difficult analytes Backed by 0 years of GC column innovation and applications expertise, Agilent J&W GC columns deliver low column bleed and low column activity for your sensitive, trace-level applications regardless of detector type. So you can cut through the noise and stop worrying about peak tailing for active compounds. Agilent J&W Ultra Inert GC columns Perform consistent trace-level analysis of pesticides, phenols, and other active compounds By itself, low column bleed increases the signal-to-noise ratio; however, your results will be flawed if any analyte is adsorbed by active sites on the column. Similarly, if a well-deactivated column has a high bleed rate, some of the analyte-generated signal could be smothered by the bleed signal. Again, your results will be flawed. Ultra Inert GC Capillary columns give you both low bleed and low activity with: The industry s highest degree of column inertness for sharper peaks, better signal-to-noise ratios, and longer column life The lowest column bleed in the business for increased detector sensitivity, faster baseline stabilization, and reduced instrument downtime Minimal compound adsorption for more accurate quantification Superior column-to-column consistency for improved productivity and reliable, reproducible results In addition, all Ultra Inert GC columns are individually tested against Agilent s exclusive Ultra Inert test probe mixture, so you know they can meet the inertness demands of today s environmental applications. Agilent J&W High Efficiency Capillary GC columns Decrease your run times by 0% or more, without compromising resolution Now it is possible to achieve high efficiency, high throughput, and high resolution without the high costs. High Efficiency Capillary GC columns can reduce your sample run time by 0% or more (compared to conventional GC), so you can get the reliable results you need using the resources you have. They are ideal for applications that require faster run times, and offer: The flexibility to choose between helium and hydrogen carrier gases. You can stay with a helium carrier if you wish to simplify method development, or switch to a hydrogen carrier if faster analysis is desired The ability to separate samples using less carrier gas, which can lead to longer intervals between cylinder changes, increased uptime, and a lower cost per sample As an added benefit, High Efficiency Capillary GC columns are compatible with all standard-pressure capillary GC and GC/MS instruments without expensive high-pressure modifications.
3 Agilent J&W DB-EUPAH columns Be certain that you are conforming to stringent European (EU) regulatory requirements DB-EUPAH GC columns are application designed, optimized, and tested for analyzing + EU-regulated priority PAHs. Engineered and manufactured to the industry s tightest QC specifications for column bleed, inertness, sensitivity, and efficiency, DB-EUPAH GC columns deliver: Optimal performance for all + EU-regulated priority PAHs Baseline resolution for difficult-to-separate critical isomer pairs Accurate analysis of high-boiling PAHs, such as dibenzopyrene isomers Exceptional thermal stability, low column bleed at elevated temperatures, and consistent column inertness Every DB-EUPAH GC column is also individually tested with a two-part, application-specific QC procedure including a test probe mixture that incorporates EU regulatory methods. Application-specific columns Tailored to individual methods and classes of compounds To supplement our standard GC column families, Agilent offers a wide selection of GC columns that have been designed, manufactured, and tested for particular samples and applications. Many of these columns have also been written into regulatory methods, making it easier for your lab to meet compliance requirements. Here are some examples of our most popular columns for environmental applications Analyte Semi-volatiles CLP Pesticides (dual column configuration) Column DB-ms Ultra Inert HP-ms Ultra Inert DB-ms HP-ms DB-. DB-3ms or DB-ms (primary) DB-XLB (confirmation) Pesticides DB-3ms DB-XLB DB-ms Ultra Inert HP-ms Ultra Inert DB-ms HP-ms PAHs PCBs Volatile organic compounds (VOCs) Volatiles and residual solvents DB-ms DB-EUPAH DB-XLB DB-3ms DB-VRX DB- 3
4 ACTIvE SEMI-volATIlES Superior performance vs. the competition: The examples that follow prove that Agilent J&W GC columns deliver reliable results and low column bleed for a broad range of benign and difficult sample types. US EPA Method 80 Short Mix Semi-volatile analysis using methods similar to US EPA Method 80 is becoming increasingly important in environmental laboratories worldwide. Note the good peak shapes of the active analytes in the semi-volatile set below. These demonstrate the superior column inertness performance of the Agilent J&W DB-ms Ultra Inert column. Column: Carrier: Oven: Injection: Detector: Sample: Suggested Supplies DB-ms Ultra Inert -3UI 30 m x 0. mm, 0. µm Helium constant flow 30 cm/s 0 C ( min) to 00 C ( C min), 0 C to 0 C ( min), C/min. to 30 C (8 min) Split/splitless; 0 C, 3. ml/min. total flow, purge flow 0 ml/min. on at 0. min., gas saver flow 80 ml/min. on at 3.0 min. MSD source at 300 C, quadrapole at 80 C, transfer line at 0 C, full scan m/z µl splitless injection, ng each component on column Liner: Direct connect, dual taper, deactivated, mm ID, G-8000 Syringe: 0 µl tapered, FN 3-s//HP, N-nitrosodimethylamine. Aniline 3., dichoronezene-d., dichlorobenzene. Naphthalene-D8. Acenapthene-D0.,-dinitrophenol 8. -nitrophenol. -methyl-,-dinitrophenol 0. Pentachlorophenol. -aminobiphenyl. Phenanthrene- D0 3. Benzidine. Chrysene-D. 3,3 -dichlorobenzidine. Benzo [b] fluoranthene. Benzo [k] fluoranthene 8. Perylene-D EPA80
5 ACTIvE SEMI-volATIlES Agilent J&W HP-ms Ultra Inert GC columns outperform Restek Rxi-ms columns in the recovery of active compounds In these examples, the Ultra Inert GC column provides excellent peak shape for acids and bases, while the Restek Rxi-ms column shows poor peak shape for some active compounds. (Acidic analytes are highlighted). Sample: 0. ng on column loading of Short Mix Components with ISTD Column: Agilent J&W HP-ms Ultra Inert 0 m x 0.8 mm x 0.8 µm Carrier: Helium 3 cm/s, Ramped flow; 0. ml/min (0. min) to.3 ml/min ( ml/min Oven: 3 C (. min) to 80 C (0 C/min), C/min to 00 C, 8 C/min to C ( min) Injection: 0. µl, splitless. 80 C purge flow 30 ml/min at 0. min MSD: Transfer line 0 C, Source 300 C, Quad 80 C Abundance n-nitrosodimethylamine. Aniline 3.,-Dichlorobenzene-d. Benzoic acid. Naphthaline-d8. Acenaphithene-d0.,-Dinitrophenol 8. -Nitrophenol. -Me-,-dinitrophenol 0. -Aminobiphenyl. Pentachlorophenol. Phenanthrene-d0 3. Benzidine. Chrysene-d. 3,3 -Dichlorobenzidine. Benzo[b]fluoranthene. Benzo[k]fluoranthene 8. Perylene-d Restek Rxi-ms Column Shows Poor Peak Shapes for Semi-volatile Acids Sample: 0. ng on column loading of Short Mix Components with ISTD Column: Rxi-ms 0 m x 0.8 mm x 0.8 µm Carrier: Helium 3 cm/s, Ramped flow; 0. ml/min (0. min) to.3 ml/min ( ml/min Oven: 3 C (. min) to 80 C (0 C/min), C/min to 00 C, 8 C/min to C ( min) Injection: 0. µl, splitless. 80 C purge flow 30 ml/min at 0. min MSD: Transfer line 0 C, Source 300 C, Quad 80 C Abundance n-nitrosodimethylamine. Aniline 3.,-Dichlorobenzene-d. Benzoic acid. Naphthaline-d8. Acenaphithene-d0.,-Dinitrophenol 8. -Nitrophenol. -Me-,-dinitrophenol 0. -Aminobiphenyl. Pentachlorophenol. Phenanthrene-d0 3. Benzidine. Chrysene-d. 3,3 -Dichlorobenzidine. Benzo[b]fluoranthene. Benzo[k]fluoranthene 8. Perylene-d
6 ClP Pesticides When it comes to speed and resolution, Agilent J&W High Efficiency GC columns outperform a leading competitor and we can prove it Rapid ClP (Contract laboratory Program) Pesticide Analysis: A side-by-side column comparison Agilent, DB-ms primary column Agilent part no. - Hz Here, Agilent s DB-ms primary analysis column resolved all peaks of interest in less than minutes with sharp symmetry and minimal baseline drift. Conversely, Restek s primary analysis column resolved only 0 of peaks and displayed evidence of peak tailing. Agilent, DB-XlB confirmatory column Agilent part no. - Hz Agilent s DB-XlB confirmatory analysis column resolved 0 peaks of interest in less than minutes (the remaining peaks were close to being baseline resolved and were sufficient for peak confirmation.) Although Restek s confirmatory column resolved all peaks of interest, there is evidence of peak tailing as well as an unacceptable level of temperature-dependent baseline drift. Compare that to Agilent s results, which show sharp, symmetrical peaks and minimal temperature-dependent baseline drift. Restek primary column Hz Restek confirmatory column Co-elution Peak tailing Baseline rise 3 Experimental Conditions: Carrier Oven Injection Detector Hydrogen ( cm/sec at 0º C, ramped at ml/min to 0 cm/sec at. minutes) 0º C (0.3 min); 0º C/min to 0º C; 30º C/min to 8º C (0.8 min); 38.8º C/min to 300º C (. min) Split/splitless; 0º C, pulsed splitless (3 psi for 0. min, purge flow of 0 ml/min on at minute, gas saver flow 0 ml/min on 3 minutes) µecd 30º C; nitrogen makeup; constant column + makeup flow 0 ml/min Hz Peak tailing Baseline rise 3 GCHE00. Tetrachloro-m-xylene. Alpha BHC 3. Gamma BHC. Beta BHC. Delta BHC. Heptachlor. Aldrin 8. Heptachlor Epoxide. Gamma Chlordane 0. Alpha Chlordane. Endosulfan I., DDE 3. Dieldrin. Endrin., DDD. Endosulfan II., DDT 8. Endrin Aldehyde. Endosulfan Sulfate 0. Methoxychlor. Endrin Ketone. Decachlorobiphenyl
7 PAHs Agilent J&W DB-EUPAH GC columns clearly surpass the competition in detecting dangerous PAHs Performance Comparison for + EU-Regulated Priority PAHs In this chromatogram, all + EU-regulated priority PAHs are well resolved with the DB-EUPAH column. Challenging Benzo[b,k,j] fluoranthene isomers are baseline resolved, allowing accurate quantitation of each isomer. Baseline resolution is also achieved for three critical pairs: benz[a]anthracene and cyclopenta[c,d]pyrene, cyclopenta[c,d]pyrene and chrysene, and indeno[,,3-cd]pyrene and dibenzo[a,h]anthracene. Agilent J&W DB-EUPAH 0 m x 0.8 mm, 0. µm Abundance Abundance Rs. Rs. 3 Rs. Rs. 8 Rs Time EU_PAHs Here, the Restek Rxi- column shows significantly higher bleed than the DB-EUPAH column, even at 30 C. As a result, the signal-to-noise ratios are less than half of those achieved by the DB-EUPAH columns. The Rxi- column s excess bleed at higher temperatures makes trace-level detection difficult and unreliable for the four late-eluting dibenzopyrene isomers. With its higher upper temperature limit, superior thermal stability and greater column inertness, the Agilent J&W DB-EUPAH column improves peak shape and sensitivity. This translates to consistently lower detection limits a must for analyzing EU-priority PAHs. Rxi- 0 m x 0.8 mm, 0.8 µm Time Analyte S/N ratio 3. Dibenzo[a,l]pyrene.8. Dibenzo[a,e]pyrene.. Dibenzo[a,i]pyrene.. Dibenzo[a,h]pyrene 8. Analyte S/N ratio 3. Dibenzo[a,l]pyrene.. Dibenzo[a,e]pyrene 3.. Dibenzo[a,i]pyrene.. Dibenzo[a,h]pyrene 3. Column : Agilent J&W DB-EUPAH 0 m x 0.8 mm, 0. µm (Agilent Part No. -) Column : Restek Rxi- 0 m x 0.8 mm, 0.8 µm Instrument: Agilent 80N/B MSD Sampler: Agilent 83B,.0 µl syringe (Agilent Part No. 8-3) 0. µl splitless injection, injection speed µl/min Carrier: Helium, ramped flow.0 ml/min (0. min), ml/min to. ml/min Inlet: 3 C splitless, purge flow 0 ml/min at 0.8 min Oven: C (0.8 min) to 00 C ( C/min),. C/min to C, 3 C/min to C, C/min to 300 C, 0 C/min to 30 C (. min) Detector: MSD source at 300 C, quadrupole at 80 C, transfer line at 330 C, Scan range 0-0 AMU. Benzo[c]fluorene. Benz[a]anthracene 3. Cyclopenta[c,d]pyrene. Chrysene. -Methylchrysene. Benzo[b]fluoranthene. Benzo[k]fluoranthene 8. Benzo[j]fluoranthene. Benz[a]pyrene 0. Indeno[,,3-cd]pyrene. Dibenzo[a,h]anthracene. Benzo[g,h,i]perylene 3. Dibenzo[a,l]pyrene. Dibenzo[a,e]pyrene. Dibenzo[a,i]pyrene. Dibenzo[a,h]pyrene
8 PAHs Agilent, DB-ms primary column Agilent part no. - EU and US-EPA regulated PAH separation on an Agilent J&W DB-ms 0 x 0.8 µm (Agilent P/N -) column Naphthalene. Acenaphthylene 3. Acenaphthene. Fluorene. Phenanthrene. Anthracene. Fluoranthene 8. Pyrene. Benzo[c]fluorene 0. Cyclopenta[c,d]pyrene. Benz[a]anthracene. Chrysene 3. -Methylchrysene. Benzo[b]fluoranthene. Benzo[k]fluoranthene. Benzo[j]fluoranthene. Benzo[a]pyrene 8. Indeno[,,3-cd]pyrene. Dibenz[a,h]anthracene 0. Benzo[g,h,i]perylene. Dibenzo[a,l]pyrene. Dibenzo[a,e]pyrene 3. Dibenzo[a,i]pyrene. Dibenzo[a,h]pyrene 8
9 Confirmed speed and accuracy: See how an Agilent J&W High Efficiency DB-VRX column resolved VOCs in under 8 minutes volatiles High Speed voc, EPA Method 80 EPA Method 80 with P&T sample introduction is one of the most widely used water analysis methods. As the chromatogram below demonstrates, DB-VRX columns ensure the fewest chromatographic coelutions and the highest degree of mass spectral integrity for VOC analysis. Column: DB-vRX - 0 m x 0.8 mm,.00 µm Carrier: Helium at cm/sec (. ml/min) Oven: C for 3.0 minutes -0 C at 3 C/min 0- C at 0 C/min C for 0. min Sampler: Purge and Trap (Tekmar 300) Purge: min Trap: Vocarb 3000 Preheat: C Desorb: 0 C for min Bake: 0 C for 0 min Line & valve: 00 C Injection: Split, 0 C Split ratio 0: Detector: Agilent 3 MSD, Scan range: 3-0 amu Scan rate: 3. scans/sec Quad temperature: 0 C Source temperature: 00 C Transfer line temp: 00 C Sample: ml Halogentated and aromatic analytes at 0 ppb Internal standards at 0 ppb Polar analytes (i.e., ethers, alcohols and ketones at ppb) Suggested Supplies: Septum: mm Advanced Green septa, 83- Liner: Direct,. mm ID, Seal: Gold plated seal, , 0 3, 3 3, 8-0, , 3-3 3, 3, 3,, , 0 3, , 0 03, ,, 8 8, 8 0, 3 8 8, , 8 3. Dichlorodifluoromethane. Chloromethane 3. Hydroxypropionitrile. Vinyl chloride. Bromomethane. Chloroethane. Ethanol 8. Acetonitrile. Acrolein 0. Trichlorofluoromethane. Isopropyl alcohol. Acetone 3. Ethyl ether.,-dichloroethene. tert-butyl alcohol. Acrylonitrile. Methylene chloride 8. Allyl chloride. Allyl alcohol 0. -Propanol. Propargyl alcohol. trans-,-dichloroethene 3. MTBE.,-Dichloroethane. Propionitrile. -Butanone. Diisopropyl ether 8. cis-,-dichloroethene. Methacrylonitrile 30. Bromochloromethane 3. Chloroform 3.,-Dichloropropane 33. Ethyl acetate 3. Ethyl-tert-butyl ether 3. Methyl acrylate 3. Dibromofluoromethane (IS) 3. Isobutanol 38. Dichloroethane-d (IS) 3. Pentafluorobenzene 0.,-Dichloroethane.,,-Trichloroethane. -Chlorobutane 3. Crotonaldehyde. -Chloroethanol.,-Dichloropropene. -Butanol. Carbon tetrachloride 8. Chloroacetonitrile. Benzene 0. tert-amylmethyl ether. Fluorobenzene (IS). -Pentanone 3. Dibromomethane.,-Dichloropropane. Trichloroethene. Bromodichloromethane. -Nitropropane ,-Dioxane. Epichlorohydrin 0. Methyl methacrylate. cis-,3-dichloropropene. Propiolactone 3. Bromoacetone. Pyridine. trans-,3-dichloropropene.,,-trichloroethane. Toluene-d8 (IS) 8. Toluene.,3-Dichloropropane 0. Paraldehyde. Ethyl methacrylate. Dibromochloromethane 3. 3-Chloropropionitrile.,-Dibromoethane. Tetrachloroethene.,,,-Tetrachloroethane. -Chlorohexane 8. Chlorobenzene. Ethylbenzene 80. Bromoform 8. m-xylene 8. p-xylene 83. trans-dichlorobutene 8.,3-Dichloro--propanol 8. Styrene 8.,,,-Tetrachloroethane 8. o-xylene 88.,,3-Trichlropropane 8. cis-dichlorobutene 0. -Bromofluorobenzene (IS). Isopropylbenzene. Bromobenzene 3. Propylbenzene. -Chlorotoluene. -Chlorotoluene GCEV003.,3,-Trimethylbenzene. Pentachloroethane 8. tert-butylbenzene.,,-trimethylbenzene 00. sec-butylbenzene 0.,3-Dichlorobenzene 0. Benzylchloride 03.,-Dichlorobenzene-d (IS) 0.,-Dichlorobenzene 0. Isopropyltoluene 0.,-Dichlorobenzene 0. Butylbenzene 08.,-Dibromo-3-chloropropane 0. Hexachloroethane 0. Nitrobenzene.,,-Trichlorobenzene. Naphthalene 3. Hexachlorobutadiene.,,3-Trichlorobenzene
10 volatiles Productivity-boosting power: Fast GC separation of 3 VOCs using an Agilent J&W High Efficiency DB- Column Fast voc Analysis Agilent DB- columns are the proven choice for analyzing VOCs in many environmental laboratories. Their stationary phase and optimized film thickness provide the resolving power needed to meet method requirements. Column: DB m x 0.8 mm,.00 µm Carrier: Helium at 3 cm/sec, (constant flow mode) Oven: 3 C for min 3-00 C at /min 00 C for 0. min 0-00 C at /min Sampler: Purge and trap (Tekmar LSC 3000) Purge: Helium for min at 0 ml/min Trap: Preheat: 0 C Desorb: 0 C for min Line & valve: 00 C Detector: MSD, 0 C transfer line Full scan 3-0 amu 3. scans per second Sample: 0 ppb per component in ml water Suggested Supplies: Septum: mm Advanced Green septa, 83- Liner: Direct,. mm ID, Seal: Gold plated seal, , Dichlorofluoromethane. Chloromethane 3. Vinyl chloride. Bromomethane. Chloroethane. Trichlorofluoromethane.,-Dichloroethene 8. Methylene chloride. trans-,-dichloroethene 0.,-Dichloroethane.,-Dichloropropane. Bromochloromethane 3. Chloroform.,,-Trichloroethane. Carbon tetrachloride. Benzene. Fluorobenzene 8. Trichloroethene.,-Dichloropropane 0. Dibromomethane. Bromodichloromethane. cis-,3-dichloropropene 3. Toluene. trans-,3-dichloropropene.,,-trichloroethane. Tetrachloroethene. Dibromochloromethane 8.,-Dibromomethane. Chlorobenzene 30.,,,-Tetrachloroethane 3. Ethylbenzene 3. m-xylene 33. p-xylene 3. o-xylene 3. Bromoform 3. Isopropylbenzene 3. Bromofluorobenzene 38. Bromobenzene 3. n-propylbenzene 0. -Chlorotoluene.,3,-Trimethylbenzene. tert-butylbenzene 3.,,-Trimethylbenzene. sec-butylbenzene.,3-dichlorobenzene. -Isopropyltoluene.,-Dichlorobenzene 8.,-Dichlorobenzene.,-Dibromo-3-chloropropane 0.,,-Trichlorobenzene. Hexachlorobutadiene. Napthalene 3.,,3-Trichlorobenzene 0
11 PCBs Ultimate confidence for key environmental applications PCBs by EPA Method 808 Agilent DB-3ms columns utilize arylene phase technology, giving them improved thermal stability through a stiffening of the polymer backbone. In this example, DB-3ms columns completely resolved CLP pesticides in less than minutes. Column: DB-3ms m x 0.3 mm, 0. µm Carrier: Helium at cm/sec (EPC in constant flow mode) Oven: 0 C for 0. min 0-30 C at C/min 30 C for min Injection: Splitless, 0 C 30 sec purge activation time Detector: µecd, 30 C Nitrogen makeup gas (column + makeup flow = 30 ml/min constant flow) Sample: 0 pg per component DB-3ms Suggested Supplies: Septum: mm Advanced Green septa, 83- Liner: Splitless, single taper, deactivated, mm ID, 8-33 Syringe: 0 µl tapered, FN 3-s//HP, 8-. IUPAC. Tetrachloro-m-xylene (IS/SS) 3. IUPAC. IUPAC 8. IUPAC 3. IUPAC. IUPAC 8. IUPAC. IUPAC 0 0. IUPAC 8. IUPAC 0. IUPAC 3. IUPAC 3. IUPAC. IUPAC 3. IUPAC 8. IUPAC IUPAC 80. IUPAC 0 0. IUPAC 0. Decachlorobiphenyl (IS/SS) IS/SS - Internal Standard/ Surrogate Standard CHloRINATED HyDRoCARBoNS Analysis of EPA. chlorinated solvents, trihalomethanes, and disinfection by-products using an Agilent J&W HP-ms Ultra Inert capillary GC column as the primary column and a DB-30 column for confirmation. The excellent peak shape of the chloral hydrate, along with the resolution between bromodichloromethane and trichloroethylene, emphasize the superior column inertness performance of the HP-ms Ultra Inert column. Column : Agilent J&W HP-ms Ultra Inert 30 m 0. mm.0 µm (Agilent p/n 0S-33UI) Hz min Column : Agilent J&W DB m 0. mm.0 µm (Agilent p/n -333) Hz min
12 Easy ordering guide Use the charts below to select the column you need based on your application. Environmental/EPA Methods Analyte Type EPA Method Reference Common Sample Preparation Detector Types Sample Matrix Recommended Agilent Column volatiles Trihalomethanes 0 Purge and trap, direct injection, headspace Volatile Organic Compounds (VOCs) Purgeable Halogenated Organics Purgeable Aromatic Organics Volatile Organic Compounds (VOCs) Using MSD Volatile Organic Compounds (VOCs) Using 3 MSD 0., 80, CLP-Volatiles Purge and trap, direct injection, headspace 0, 800 Purge and trap, headspace for screening 03., 0, 800.,, 80, 80, CLP-VOCs.,, 80, 80, CLP-VOCs Purge and trap, headspace for screening Purge and trap, direct injection, headspace Purge and trap, direct injection, headspace EDB and DBCP 0., 80 Microextraction with Hexane Acrylonitrile and Acrolein 03, 80, 803 Purge and trap, liquid extraction, sonication ELCD, ECD PID, ELCD PID, ELCD PID MSD MSD MSD FID, NPD Drinking water DB-VRX, 30 m x 0. mm,. µm, -3 DB-, 30 m x 0. mm,. µm, -33 HP-, 30 m x 0.3 mm,. µm, 0Z-3 Drinking water, DB-VRX, m x 0. mm,. µm, - waste water, DB-, m x 0. mm,. µm, -3 DB-0., 0 m x 0.3mm, 3.0 µm,-a Waste water, solid waste water, DB-VRX, m x 0. mm,. µm, - DB-, m x 0. mm,. µm, -3 Drinking water, waste water, DB-VRX, 30 m x 0. mm,. µm, -3 DB-, 30 m x 0. mm,. µm, -33 Drinking water, DB-VRX, 0 m x 0. mm,.0 µm, - waste water, DB-, 0 m x 0. mm,.0 µm, -3 HP-VOC, 0 m x 0.0 mm,.0 µm, 0R-30 Drinking water, waste water, DB-VRX, 0 m x 0.8 mm,.00 µm, - DB-, 0 m x 0.8 mm,.00 µm, -3 Drinking water, DB-VRX, 30 m x 0. mm,. µm, -3 DB-, 30 m x 0. mm,. µm, -33 Waste water, DB-VRX, 30 m x 0. mm,. µm, -3 DB-, 30 m x 0. mm,. µm, -33 Semivolatiles Semivolatile Organic Compounds,, 80 Liquid extraction, sonication, soxhlet extraction, SPE MSD Drinking water, waste water, DB-.,30 m x 0. mm, 0.0 µm, -3 DB-.,0 m x 0.8 mm, 0.3 µm, - HP-ms Ultra Inert, 30 m x 0. mm, 0. µm, 0S-33UI DB-ms Ultra Inert, 30 m x 0. mm,0. µm, -3UI HP-ms, 30 m x 0. mm, 0.0 µm, 0S-33 Phenols 8, 0, 800, 80 Liquid extraction, sonication, soxhlet extraction, derivatization ECD, FID Waste water, DB-ms Ultra Inert, 30 m x 0. mm, 0. µm, -3UI DB-ms, 30 m x 0. mm, 0. µm, -3 DB-XLB, 30 m x 0. mm, 0. µm, -3 DB-ms, 30 m x 0.3 mm,.0 µm, -3
13 Environmental/EPA Methods Analyte Type EPA Method Reference Common Sample Preparation Detector Types Sample Matrix Recommended Agilent Column Semivolatiles (Continued) Phthalate Esters 0, 0, Liquid extraction, ECD, Drinking water, DB-ms Ultra Inert, 30 m x 0. mm, 0. µm, -3UI 800, 80 sonication, soxhlet FID waste water, extraction, SPE DB-ms, 30 m x 0. mm, 0. µm, -3 DB-ms, 30 m x 0.3 mm,.0 µm, -3 DB-08, 30 m x 0.3 mm, 0.0 µm, -83 Benzidines 0 Liquid extraction ECD Waste water DB-ms Ultra Inert, 30 m x 0. mm, 0. µm, -3UI DB-ms, 30 m x 0. mm, 0. µm, -3 DB-ms, 30 m x 0.3 mm,.0 µm, -3 DB-08, 30 m x 0.3 mm, 0.0 µm, -83 Nitrosamines 0, 800 Liquid extraction, NPD Waste water, DB-ms Ultra Inert, 30 m x 0. mm, 0. µm, -3UI sonication, soxhlet extraction, SPE DB-ms, 30 m x 0. mm, 0. µm, -3 DB-ms, 30 m x 0.3 mm,.0 µm, -3 Nitroaromatics 0, 800 Liquid extraction, ECD, Waste water, HP-ms Ultra Inert, 30 m x 0. mm, 0. µm, 0S-33UI and Isophorone sonication, soxhlet FID extraction, SPE HP-ms, 30 m x 0. mm, 0.0 µm, 0S-33 DB-ms, 30 m x 0.3 mm,.0 µm, -3 DB-08, 30 m x 0.3 mm, 0.0 µm, -83 Polynuclear Aromatic 0, 800 Liquid extraction, FID Waste water, DB-ms Ultra Inert, 30 m x 0. mm, 0. µm, -3UI Hydrocarbons (PAHs) sonication, soxhlet extraction, SPE DB-ms, 30 m x 0. mm, 0. µm, -3 DB-ms, 30 m x 0.3 mm, 0. µm, 3-3 DB-ms, 30 m x 0. mm, 0. µm, -03 Chlorinated Hydrocarbons, 80, Liquid extraction, ECD Waste water, DB-ms Ultra Inert, 30 m x 0.3 mm, 0.0 µm, 3-3UI 8 sonication, soxhlet extraction, SPE DB-ms, 30 m x 0.3 mm, 0.0 µm, 3-3 HP-ms, 30 m x 0.3 mm, 0.0 µm, 0S-3 DB-, 30 m x 0.3 mm, 0.0 µm, 3-03E Chlorinated Disinfection Byproducts,.A Liquid extraction, derivatization ECD Drinking water HP-ms Ultra Inert,30 m x 0. mm x.0 µm, 0S-33UI DB-30, 30 m x 0. mm x.0 µm, -333 DB-ms, 30 m x 0. mm,.00 µm, -33 DB-, 30 m x 0. mm,.00 µm, -033 DB-, 30 m x 0. mm,.00 µm, -033 Halogenated Acetic Acids,.,. Liquid extraction, derivatization ECD Drinking water DB-3ms, 30 m x 0.3 mm, 0. µm, DB-XLB, 30 m x 0.3 mm, 0.0 µm, 3-3 3
14 Environmental/EPA Methods Analyte Type EPA Method Reference Common Sample Preparation Detector Types Sample Matrix Recommended Agilent Column Pesticides, Herbicides, and PCBs Organochlorine Pesticides and PCBs 08., 08, 808A, 808, CLP-Pesticides Phenoxy Acid Herbicides,, 80, 8 N- and P-Containing Pesticides and Herbicides PCB Congeners Using MSD 0,,,, 80, 8A Liquid extraction, derivatization Liquid extraction, derivatization Liquid extraction, derivatization Liquid extraction, derivatization ECD Drinking water DB-ms Ultra Inert, 30 m x 0. mm, 0. µm, -3UI DB-ms Ultra Inert, 30 m x 0. mm, 0. µm, -03UI DB-3ms, 30 m x 0.3 mm, 0. µm, DB-XLB, 30 m x 0.3 mm, 0.0 µm, 3-3 DB-XLB, 0 m x 0.8 mm, 0.8 µm, - ECD Drinking water DB-3ms, 30 m x 0.3 mm, 0. µm, DB-XLB, 30 m x 0.3 mm, 0.0 µm, 3-3 NPD, Drinking water DB-ms Ultra Inert, 30 m x 0. mm, 0. µm, -3UI ELCD, FPD HP-ms Ultra Inert, 30 m x 0. mm, 0. µm, 0S-33UI DB-3ms, 30 m x 0. mm, 0. µm, -383 DB-ms, 30 m x 0. mm, 0. µm, -3 MSD DB-3ms, 30 m x 0.3 mm, 0. µm, DB-XLB, 30 m x 0.3 mm, 0.0 µm, 3-3 Ultra Inert ms Capillary GC Columns ID (mm) length (m) Film (µm) Part No. DB-ms Ultra Inert UI 0.3-3UI UI.00-3UI 0. -UI UI 0.0-3UI.00-33UI UI UI.00-3UI UI UI UI UI HP-ms Ultra Inert S-UI S-3UI S-33UI 0.0 0S-33UI.00 0S-33UI S-3UI S-3UI.00 0S-3UI Ultra Inert ms Capillary GC Columns ID (mm) length (m) Film (µm) Part No. DB-ms Ultra Inert UI UI UI UI UI UI HP-ms Ultra Inert S-UI S-3UI S-33UI 0.0 0S-33UI.00 0S-33UI S-UI 0. 0S-UI S-3UI.00 0S-3UI
15 High Efficiency Capillary GC Columns Stationary Phase Part No. ( cage) ID (mm) length (m) Film (µm) Part No. ( cage) DB E E -0A E E HP- 0Z Z-E DB-ms E HP-ms 0S S-E DB E E E HP- 0J J-E DB-ms E HP-ms 0S S-E DB-XLB E DB-3ms E DB DB-ms E HP-0+ 0L DB High Efficiency Capillary GC Columns (continued) Stationary Phase Part No. ( cage) ID (mm) length (m) Film (µm) Part No. ( cage) DB DB E DB DB DB-WAX E E E HP-INNOWax 0N N-E DB DB-VRX E DB DB-EUPAH GC Columns Stationary Phase Part No. ID (mm) length (m) Film (µm) Temp limit ( C) DB-EUPAH to 30/30 DB-EUPAH -L to 30/30 DB-EUPAH to 30/30 + PAH standard 0 µg/ml 0-08 From sample preparation to separation Agilent columns and supplies ensure a lifetime of peak instrument performance As the world s largest GC and GC/MS instrument provider, Agilent is uniquely positioned to offer you the widest selection of chromatography columns, parts and supplies, including: Innovative GC columns Premium non-stick inlet septa and non-stick liner O-rings Certified vials, caps and septa Proprietary deactivated inlet liners Pre-conditioned ferrules in ultra-clean packaging Metal injection-molded inlet gold seals Gold Standard autosampler syringes Gas purifiers, and more All are engineered or selected by our experienced instrument design teams, manufactured to our demanding specifications, and tested under a variety of strict conditions. They are also backed by unmatched technical support on the Web, by phone or in person plus a 0-day warranty from the date of shipment.
16 Visit to learn more about meeting your most challenging environmental application demands For more information Find an Agilent customer center in your country: U.S. and Canada: Europe: Asia Pacific: With the 00 addition of Varian Inc., Agilent now offers an even greater range of instrumentation, as well as the most comprehensive columns and supplies portfolio in the market. Agilent is here to provide the technology and the Measure of Confidence you need to be successful. This information is subject to change without notice. Agilent Technologies, Inc. 00 Printed in USA, July, EN The Mea sure of Confidence
Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1
Application Note US EPA Method 524.2 with the Teledyne Tekmar Lumin P&T Concentrator and Agilent 7890B GC / 5977A MS Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Abstract US EPA Method
More informationUS EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1806 P+T VOC ENV US EPA Method 524.2 with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS Amy Nutter, Applications Chemist; Teledyne
More informationValidation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS
Validation of USEPA Method 524.2 Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS Application Note By: Nathan Valentine Abstract The US EPA developed Method 524.2¹, Measurement
More informationSolid Phase Micro Extraction of Flavor Compounds in Beer
Solid Phase Micro Extraction of Flavor s in Beer ANNE JUREK USEPA 524.2 Method Validation Using the Evolution Purge and Trap Concentrator and the Centurion WS Autosampler Application Note Environmental
More informationAnalysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler
Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the Introduction Although analysis of VOCs by purge and trap is considered a mature technique, advances in
More informationUS EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS
Application Note US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS Author Amy Nutter Applications Chemist, Teledyne Tekmar Abstract US EPA Method 8260 in conjunction
More informationValidation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters
Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters Application Note Abstract Water quality laboratories across the
More informationValidation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler
Validation of USEPA Method 524.2 Using a Stratum PTC and the New AQUATek 100 Autosampler Application Note Abstract Automation is the key to increasing laboratory productivity and minimizing costs. It is
More informationAnalysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler
Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler Application Note By: Anne Jurek Abstract In order to determine the amount
More informationA Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas
A Comparison of Volatile Organic Compound When Using Nitrogen as a Gas Application Note By: Anne Jurek Abstract For many years Helium has been the gas of choice for purging Volatile Organic Compounds (VOCs).
More informationRoger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1
Application Note US EPA Method 524.4 with the Tekmar Lumin P & T Concentrator and Thermo Scientific TRACE 1310 GC and ISQ LT MS System Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1 Abstract
More informationExploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time
Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Introduction Purge and Trap (P&T) concentrators
More informationOptimizing. Abstract: is standardd. procedures. altered to
Optimizing Standard Preparation ANNE JUREK Abstract: Standardd preparation can often be a time consuming, tedious process. The opportunity for human error or inconsistencies between individual preparation
More informationOptimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator
Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator Application Note Introduction Environmental laboratories have utilized the Purge & Trap sample concentration technique for over
More informationAnalysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler
Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler Application Note By: Teri Dattilio Introduction Purge and Trap concentration
More informationDetection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS
Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Application Note Environmental Author Xiaohua Li Agilent Technologies (Shanghai)
More informationValidation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions
Validation of Volatile Organic Compound by USEPA Method 8260C Application Note Abstract In order to determine the concentration of volatile organic compounds (VOCs) in water and soil matrices the USEPA
More informationUsing Hydrogen as An Alternative Carrier Gas for US EPA 8260
Using Hydrogen as An Alternative Carrier Gas for US EPA 8260 Application Note Abstract Due to regional shortages and increasing costs of helium, the preferred carrier gas in gas chromatography, alternative
More informationMaximizing Production While Minimizing Costs
JSB is an authorised partner of Maximizing Production While Minimizing Costs Anne Jurek #123 Abstract: The need to increase productivity and reduce cost is becoming of greater concern to environmental
More informationMethanol Extraction of high level soil samples by USEPA Method 8260C
Methanol Extraction of high level soil samples by USEPA Method 8260C Abstract In order to determine the concentration of Volatile Organic Compounds (VOCs) in soil and waste samples the USEPA developed
More informationValidation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS
Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS Application Note By Tyler Trent Abstract In order to determine the concentration of volatile organic compounds
More informationMaximizing Sample Throughput In Purge And Trap Analysis
Maximizing Sample Throughput In Purge And Trap Analysis LINDSEY PYRON ANNE JUREK INTRODUCTION There are several demands and requirements imposed on chemists performing volatile organic analysis (VOC) in
More informationOREGON Environmental Laboratory Accreditation Program ORELAP Fields of Accreditation ALS Environmental - Simi Valley
MATRIX Reference Code Analyte Code Description ASTM C471M- 14 2014 ASTM D5504-12 2012 3964 Orthorhombic Cyclooctasulfur 4842 1-Propanethiol 6113 2,5-Dimethylthiophene 4544 2-Ethylthiophene 4843 2-Propanethiol
More informationA Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260
JSB is an authorised partner of A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260 Anne Jurek #101 Introduction: The United States Environmental Protection Agency (USEPA)
More informationOptimizing Standard Preparation
JSB is an authorised partner of Optimizing Standard Preparation Anne Jurek #106 Abstract: Standard preparation can often be a time consuming, tedious process. The opportunity for human error or inconsistencies
More information2017 EPA Method Update Rule and EPA Method 624.1
Introduction Method 624 is for the determination of volatile organic compounds in industrial discharges and other liquid environmental samples by gas chromatography combined with mass spectrometry (GC/MS).
More informationHelium conservation in volatile organic compound analysis using U.S. EPA Method 8260C
APPLICATION NOTE 10441 Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C Authors Andrea Caruso, 1 Tommaso Albertini, 1 Jacob A. Rebholz 2 ; 1 Thermo Fisher Scientific,
More informationAnalysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.
Optimal Conditions for USEPA Method 8260 Analysis Anne Jurek Introduction: Over the past decade, the need for environmental laboratoriess using purge and trap systems to report at or below the Method Detection
More informationImproved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source
Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source Application Note Environmental Authors Peter Gautschi and Harry Prest Senior Application Scientist
More informationSolid Phase Micro Extraction of Flavor Compounds in Beer
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Automed Sampling of Methanol Extractions Applicion Note Environmental Author Anne Jurek Applicions Chemist EST Analytical Cincinni, OH
More informationNon-Potable Water A2LA Certificate
COLOR Color Biochemical Oxygen Demand (BOD); Carbonaceous DEMAND BOD; Chemical Oxygen Demand (COD); Total Organic Carbon (TOC) HEXAVALENT CHROMIUM INORGANIC NUTRIENTS LOW-LEVEL MERCURY METALS I & II METALS
More informationU.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS
ENVIRONMENTAL application note U.S. EPA VOLATILE ORGANICS METHOD 524.2 USING PURGE AND TRAP GC/MS Cheri Coody, Public Health Laboratory, Mississippi State Health Department, Jackson, MS USA Michael J.
More informationEPA TO-17 Volatile Organic Compounds
EPA TO-17 Volatile Organic Compounds Method TO-17 is used to analyze samples for volatile organic compounds collected on multi-bed sorbent tubes, which are thermally desorbed and cryo-focused on the capillary
More informationEd George and Anaïs Viven Varian, Inc.
Application Note SI-02408 Evaluation of EPA Method 524.3: a New Draft Method for the Analysis of VOCs in Drinking Water using GC/MS and V:Results GC/MS Software Ed George and Anaïs Viven Varian, Inc. Introduction
More informationOptimizing of Volatile Organic Compound Determination by Static Headspace Sampling
JSB is an authorised partner of Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling Anne Jurek Introduction: #134 Static headspace sampling has always been a viable option
More informationINNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical
INNOVATIVE PRODUCTS, SUPERIOR SUPPORT Presenter: Anne Jurek, Senior Applications Chemist,, pp, EST Analytical Air pollution is a growing problem due to the global economy and industrial development in
More informationDate Issued: April 01, 2013 Expiration Date: June 30, 2013
State of Florida Department of Health~Bureau of Public Health Laboratories I his is to certify that ALS ENVIRONMENTAL - SIMI VALLEY 2655 PARK CENTER DRIVE, SUITE A SIMI VALLEY, CA 93065 has complied with
More informationA Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes
A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes Application Note Environmental Authors Ken Lynam and Doris Smith Agilent Technologies, Inc. 2850 Centerville Road Wilmington,
More information317 Elm Street Milford, NH (603) Fax (603)
317 Elm Street Milford, NH 03055 (603) 673-5440 Fax (603) 673-0366 Sales@chemservelab.com Friday, October 06, 2017 Domenic Carapella Greenplay Organics, LLC 6 Hawthorne Ave Merrick NY 11566 Project Name:
More informationSystem and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.
ANNE JUREK Optimal Conditions for USEPA Method 8260B Analysis Using the EST Analyticall Sampling System and the Shimadzu GCMS-QP2010s Introduction: The USEPA developedd Method 8260B for the determination
More informationPerformance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices
Application Note Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices Abstract By: Ed Price In today s laboratories, increased efficiency and productivity are of
More informationMethod 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8
application Note Gas Chromatography/ Mass Spectrometry Authors Ruben Garnica Dawn May PerkinElmer, Inc. Shelton, CT USA Method by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares
More informationApplication Note. Abstract. Authors. Environmental
GC/MS Analysis of European Union (EU) Priority Polycyclic Aromatic Hydrocarbons (PAHs) using an Agilent J&W DB-EUPAH GC Column with a Column Performance Comparison Application Note Environmental Authors
More informationMulti-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms
Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms Application Note Author Laura Provoost Agilent Technologies, Inc. Introduction In multi-residue analysis different groups
More informationDetermination of Volatile Organic Compounds in Air
JSB is an authorised partner of Determination of Volatile Organic s in Air Anne Jurek #126 Introduction Air pollution is a growing problem due to the global economy and industrial development in many countries.
More informationPAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices
PAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices Food Quality and Environmental Author Ken Lynam Agilent Technologies, Inc. 280 Centerville Road Wilmington, DE 19808 Abstract
More informationSolid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS
Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS LECO Corporation; Saint Joseph, Michigan USA Key Words: GC-TOFMS,
More informationAnalytical Trap Comparison for USEPA Method 8260C
Analytical Trap Comparison for USEPA Method 8260C Application Note Abstract Purge and trap concentration is a technique that is used for the analysis of Volatile Organic Compounds (VOCs). The major component
More informationOptimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s
JSB is an authorised partner of Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s Anne Jurek #120 Introduction: The USEPA developed
More informationAnalysis of Low Level Volatile Organic Compounds in Air Anne Jurek
JSB is an authorised partner of Analysis of Low Level Volatile Organic s in Air Anne Jurek #111 Introduction The New Jersey Department of Environmental Protection (NJDEP) developed a new low level TO-15
More informationApplication News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS
HS-20 & GCMS-QP2010Ultra Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS Introduction Volatile Organic Compounds (VOCs) refer to a group
More informationAUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL
AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL Challenges of Water Monitoring Volatile Organic Compounds (VOCs) can have negative health impacts even at ppb levels VOC concentrations can
More informationSemivolatile Organics Analysis Using an Agilent J&W HP-5ms Ultra Inert Capillary GC Column
Semivolatile Organics Analysis Using an Agilent J&W HP-5ms Ultra Inert Capillary GC Column Application Note Environmental Author Doris Smith and Kenneth Lynam Agilent Technologies, Inc. 285 Centerville
More informationMcCAMPBELL ANALYTICAL INC Willow Pass Road Pittsburg CA
Comparison of VOC : s, Target Lists, MAI Acenaphthene 83-32-9 Acetalaldehyde (Ethanal) 75-07-0 Acetone 67-64-1 Acetonitrile 75-05-8 Acetophenone 96-86-2 Acrolein (Propenal) 107-02-8 Acrylamide 79-06-1
More informationSelection of a Capillary
Selection of a Capillary GC Column - Series 3 Mark Sinnott Application Engineer March 19, 2009 Page 1 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector
More informationRely on Rxi -PAH Columns
Rely on Rxi -PAH Columns to Ensure Successful Polycyclic Aromatic Hydrocarbon (PAH) Analysis Optimized Efficiency, Selectivity, and Robustness Let You: Report accurate results with speed and confidence.
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares
More informationSUPELCO. Improved Performance for Cyanopropylphenyldimethylpolysiloxane (SPB-1701, SPB-1301) Capillary Columns
Improved Performance for Cyanopropylphenyldimethylpolysiloxane (SPB-1701, SPB-1301) Capillary Columns L.M. Sidisky, Yizeng Ni, Greg Baney, Rodney George, Katherine K. Stenerson 01-0026 T401044 CFK Discussion
More informationBIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: CTRA Project: 6454 T Lab Order: A 6454 Aqueous 3/1/2013.
Date: 08-Mar-13 Work Order Sample Summary Lab Sample ID Client Sample ID Matrix Collection Date Date Received 1303004-01A 6454 Aqueous 3/1/2013 3/1/2013 Page 1 of 1 Date: 08-Mar-13 CASE NARRATIVE Samples
More informationBIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: Cascade Thornapple River Assoc. Project: Water Analysis Lab Order:
Date: 31-Jul-18 Work Order Sample Summary Lab Sample ID Client Sample ID Matrix Collection Date Date Received 1807123-01A Dam & Burger Water 7/25/2018 7/25/2018 1807123-02A 48th & RR Water 7/25/2018 7/25/2018
More informationEvaluating CLP and EPA Methods for Pesticides in Water Using Agilent J&W DB-CLP1/DB-CLP2 GC Columns
Evaluating CLP and EPA Methods for Pesticides in Water Using Agilent J&W DB-CLP/DB-CLP GC Columns Application Note Environmental Authors Doris Smith and Ken Lynam Agilent Technologies, Inc. Abstract The
More informationAnalysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System
Application Note Residual Solvent Analysis of USP Method Residual Solvents on the Agilent 889 GC System Author Lukas Wieder, Jie Pan, and Rebecca Veeneman Agilent Technologies, Inc. 8 Centerville Road
More informationMeasuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS
application Note Gas Chromatography Author Heidi Grecsek PerkinElmer, Inc. Shelton, CT 06484 USA Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS Introduction
More informationSelection of a Capillary GC Column
Selection of a Capillary GC Column Mark Sinnott Application Engineer March 13, 2008 Page 1 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector Electrometer
More informationReduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System
Reduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System Application Note 19080203 Keywords Dual Purge-and-Trap Eclipse Sample Concentrator Gas Chromatography GC GC/MS Model 4560 Sample
More informationSo Many Columns! How Do I Choose? Daron Decker Chromatography Technical Specialist
So Many Columns! How Do I Choose? Daron Decker Chromatography Technical Specialist GC Columns Wall Coated Open Tubulars Liquid phase coated capillaries Internal Diameter 0.05 0.53mm Length 5m 100m Porous
More informationNational Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds
Permit Guidance 12 Final National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds Statutory references: ORC 6111.01, 6111.03, 6111.04 Rule references: OAC 3745-1-05,
More informationOptimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note Environmental Author Anne Jurek Applications
More informationTarget Compound Results Summary
Phone/ (86)88-4800 / (86)88-471 http://www.emsl.com to1lab@emsl.com Customer ID: EMSL0A K12312.D 28 cc 0 0/01/2017 K12327.D 21 cc 600 Target Compound Results Summary Propylene 11-07-1 42.08 ND.0 ND 8.6
More informationKSA Environmental Laboratory Inc.
February 26, 07 KSA Environmental Laboratory Inc. Hydrologic Associates - Miami Attn: Brad Waller 8925 S.W. 148th Street Suite 212 Miami, FL 33176 RE: Parkland KSA Workorder: Q0012 Dear Brad Waller, Enclosed
More informationCALA Directory of Laboratories
CALA Directory of Laboratories Membership Number: 2628 Laboratory Name: Caduceon Environmental Laboratories (Richmond Hill) Parent Institution: Caduceon Enterprises Inc. Address: 110 West Beavercreek Rd.
More informationDetection of Environmental Contaminants Caused by the Oil Spill in the Gulf of Mexico by GC and Hplc
Detection of Environmental Contaminants Caused by the Oil Spill in the Gulf of Mexico by GC and Hplc Sky Countryman, Ngoc Nguyen, Jeff Layne, Kory Kelly, and Zeshan Aqeel Phenomenex, Inc., 411 Madrid Ave.,
More informationAgilent J&W GC Columns For Environmental Applications MEASURE TRACE-LEVEL CONTAMINANTS AND MEET REGULATORY REQUIREMENTS WITH CONFIDENCE
Agilent J&W GC Columns For Environmental Applications MEASURE TRACE-LEVEL CONTAMINANTS AND MEET REGULATORY REQUIREMENTS WITH CONFIDENCE Reliably and efficiently analyze increasingly small quantities of
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares
More informationThermal Desorption Technical Support
Thermal Desorption Technical Support Note 86a: US EPA Method TO-17 for Monitoring Air Toxics in Ambient Air Using Sorbent Tubes and Automated, Cryogen-free Thermal Desorption Application Note Environmental,
More informationLow Bleed Stationary Phases for Gas Chromatography
Low Bleed Stationary Phases for Gas Chromatography Slide 4 Advantages of Low Bleed Phase Lower bleed compared to standard phases Better signal to noise (sensitivity) Higher upper temperature - shorter
More informationMethods and Analytes Requiring Laboratory Certification Attachment 3
s and s Requiring Laboratory Certification Attachment 3 Current as of July 1, 2011 Laboratory Certificates need to accurately reflect the methods performed and reported by the laboratory, including the
More informationTable 1 Soil and Groundwater Sampling Summary Table
Table Soil and Groundwater Sampling Summary Table Phase II Investigation Consolidated Edison - Former kent Avenue Generating Station 0 Kent Avenue Brooklyn, NY Sample ID Date Installed Date Sample Collected
More informationRtx -CLPesticides Columns Reduce Analysis Time; Reduce System Maintenance
Columns Reduce Analysis Time; Reduce System Maintenance 0 YEARS Turning Visions into Reality www.restek.com 00-- --00 and Fused Silica Columns Detect and confirm 0 common organochlorine pesticides simultaneously.
More informationFaster Semivolatiles Analysis with a Scaled- Down Method and GC Accelerator Kit
Faster Semivolatiles Analysis with a Scaled- Down Method and GC Accelerator Kit Analysis times for semivolatile compounds can limit sample throughput and decrease overall lab productivity. You can significantly
More informationANALYTICAL REPORT. Results relate only to the items tested and the sample(s) as received by the laboratory. Page 1 of 5
ANALYTICAL REPORT TestAmerica Laboratories, Inc. TestAmerica Savannah 5102 LaRoche Avenue Savannah, GA 31404 Tel: (912)354-7858 Client For: Imperial Paints LLC PO BOX 489 Fairforest, South Carolina 29336
More informationPractical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software
Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software High Efficiency GC Columns Page 1 Variables for Shortening Run Times Stationary Phase Shorten Column Length
More informationVolatile Organic Compounds in Water PBM
Organic Constituents and Compounds Revision Date: July 10, 2017 Volatile Organic Compounds in Water PBM Parameter Analytical Method Introduction Method Summary MDL(s) and EMS Analyte Codes Volatile Organic
More informationBuilding 280A and Building 450 Soil Sampling Results Richmond Field Station Site Berkeley Global Campus at Richmond Bay, Richmond, California
TETRA TECH, INC. February 5, 2016 Lynn Nakashima Berkeley Regional Office 700 Heinz Avenue, Suite 200C Berkeley, California 94710 Subject: Building 280A and Building 450 Soil Sampling Results Richmond
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: 6701 Aberdeen Avenue, Suite #9, Lubbock, TX 79424
More informationApplication Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ
Application Note Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ -3 Application Note Abstract This application note demonstrates the
More informationHeadspace Technology for GC and GC/MS: Features, benefits & applications
Headspace Technology for GC and GC/MS: Features, benefits & applications Karima Baudin Oct 2015 Why use Headspace? Very Simple no to minimum sample prep Robust enhance uptime Non-detectable carry-over
More informationAppendix 2 Data Distributions for Contaminants in Water and Sediment Samples in Response to the Deepwater Horizon Oil Spill, 2010
University of Nebraska - Lincoln DigitalCommons@University of Nebraska - Lincoln Publications of the US Geological Survey US Geological Survey 2013 Appendix 2 Data Distributions for Contaminants in Water
More information2. September 2015 Sub-Concrete Slab Soil Investigation
To: Pam Barnett, RACER Ref. No.: 11109615 From: John-eric Pardys/kf/1 Date: October 30, 2015 Re: Summary of Sub-Concrete Slab Soil Investigation RACER Fredericksburg Facility Spotsylvania County, Virginia
More informationQuantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor
Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor Application Note Pesticides Authors Charlie Spanjers and Paul Dauenhauer University of Minnesota, Twin Cities
More informationStudy of Residual Solvents in Various Matrices by Static Headspace
Application Note Abstract United States Pharmacopeia (USP) chapter is a widely used method for identifying and quantifying Organic Volatile Impurities (OVI) used in the production of pharmaceuticals.
More informationANALYTICAL REPORT. Job Number: Job Description: Transform Complete
ANALYTICAL REPORT Job Number: 640-4848-1 Job Description: Transform Complete For: American Master*Tech Scientific, Inc. 1330 Thurman Street PO BOX 2539 Lodi, CA 95240 Attention: Mr. Adam Kowalczyk Chad
More informationElectronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers
Electronic Supplementary Material (ESI) for Environmental Science: Processes & Impacts. This journal is The Royal Society of Chemistry 2017 Electronic Supplementary Material Experimentally Validated Mathematical
More informationRapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air
Environmental Applications Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air By Jason S. Herrington Abstract The following study evaluated the efficacy of using Nutech s 8900DS preconcentrator
More informationIntroducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products.
Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products. CHRISTOPHER M. ENGLISH, CHRISTOPHER S. COX, FRANK
More informationResidual solvents analysis using an Agilent Intuvo 9000 GC system
Residual solvents analysis using an Intuvo GC system Technology advantage: simplified dual column analysis with Flow Chip modularity Introduction USP defines a method for analysis of residual solvents
More informationThe following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.
5/26/2017 Ms. Anna Kelley Hamilton Co. Environmental Services 250 Wm. Howard Taft Cincinnati OH 45219 Project Name: Sun Coke Project #: Workorder #: 1705303 Dear Ms. Anna Kelley The following report includes
More informationANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident
ANALYTICAL RESULTS Prepared by: Eurofins Lancaster Laboratories Environmental 2425 New Holland Pike Lancaster, PA 760 Prepared for: Mobil Pipeline Company PO Box 446 Houston TX 7720-446 July 30, 203 Project:
More information