1 830 ual N) manual Operating manual

2 Copyright 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of the copyright owner. Trademarks Eppendorf and the Eppendorf logo and are registered trademarks of Eppendorf AG, Hamburg, Germany. BIOMASTER and Mastertip are registered trademarks of Eppendorf AG, Hamburg, Germany. Registered trademarks are not marked in all cases with or in this manual /102013

3 Table of contents 3 Table of contents 1 Operating instructions Using this manual Danger symbols and danger levels Danger symbol Danger levels Symbols used Product description Operating principle Main illustration Safety Intended use Warnings for intended use Information on product liability Operation Setting the volume Tip take-up Aspirating liquid Dispensing liquid Ejecting the Mastertip Maintenance Testing the volume Sterilization Cleaning and decontamination Decontamination before shipment Technical data Test conditions Ordering Information

4 4 Operating instructions 1 Operating instructions 1.1 Using this manual Read this operating manual completely before using the device for the first time. Also observe the instructions for use of the accessories. This operating manual is part of the product. Thus, it must always be easily accessible. Enclose this operating manual when transferring the device to third parties. You will find the current version of the operating manual for all available languages on our webpage under Danger symbols and danger levels The safety instructions of this operating manual indicate the following danger symbols and danger levels: Danger symbol Biohazard Toxic substances Hazard point Material damage Danger levels DANGER WARNING CAUTION Notice Will lead to severe injuries or death. May lead to severe injuries or death. May lead to light to moderate injuries. May lead to material damage. 1.3 Symbols used Depiction Meaning 1. Actions in the specified order 2. Actions without a specified order List Text Display text or software text Additional information

5 Product description 5 2 Product description 2.1 Operating principle The Biomaster is a pipette designed for very sensitive testing in the volume range of 1 μl 20 μl. It can be used with variable settings and without entrainment of residues. Its tip, the Mastertip, can be mounted with one hand and disposes of an integrated piston. By inhibiting the formation of aerosols, the Mastertip excludes sample contamination. The Biomaster is UV-resistant and autoclavable. Mastertips are autoclavable. Please refer to the section on sterilization before autoclaving the Biomaster or the Mastertip (121 C, 20 min). 2.2 Main illustration Control button 2 Volume locking button 3 Mastertip 3

6 6 Safety 3 Safety 3.1 Intended use The Biomaster is a lab device which, in combination with a Mastertip, is intended for dispensing aqueous solutions in the volume range of 1 μl 20 μl. In vivo applications (applications in or on the human body) are not permitted. The Biomaster may only be operated by trained specialists. All users must have read the operating manual carefully and familiarized themselves with the device's mode of operation. 3.2 Warnings for intended use WARNING! Damages to health due to infectious liquids and pathogenic germs. When handling infectious liquids and pathogenic germs, observe the national regulations, the biological security level of your laboratory, the material safety data sheets, and the manufacturer's application notes. Wear personal protective equipment. For full instructions regarding the handling of germs or biological material of risk group II or higher, please refer to the "Laboratory Biosafety Manual" (Source: World Health Organization, current edition of the Laboratory Biosafety Manual). WARNING! Damage to health due to toxic, radioactive or aggressive chemicals. Wear personal protective equipment. Observe the national regulations for handling these substances. Observe the material safety data sheets and manufacturer's application notes. CAUTION! Danger to individuals due to careless use Never point the opening of the device towards yourself or other people. Only initiate liquid dispensing if it is safe to do so. With any dispensing task please ensure that you do not endanger yourself and other persons.

7 Safety Information on product liability In the following cases, the designated protection of the device may be compromised. Liability for any resulting property damage or personal injury is then transferred to the operator: The device is not used in accordance with the operating manual. The device is used outside of its intended use. The device is used with accessories or consumables which are not recommended by Eppendorf. The device is maintained or repaired by people not authorized by Eppendorf. The user makes unauthorized changes to the device.

8 8 Operation 4 Operation 4.1 Setting the volume 1. Press and hold the volume catch. 2. Turn the control button to set the volume. 3. Release the volume catch. The set volume is now secured against unintentional readjustment. 4.2 Tip take-up Prerequisites The Mastertip is placed in a tip box. 1. Attach the Biomaster to one Mastertip. You will feel the Biomaster locking in place in the Mastertip. Keep the Mastertip in the tip box after attaching it to the Biomaster. 2. Firmly press the control button down to the second stop. The piston is now fixed to the Biomaster. 3. Release the control button. The piston will slide back up inside the tip until it reaches its final position. The system is now ready for operation. 4.3 Aspirating liquid 1. Press down the control button to the first stop. 2. Immerse the Mastertip vertically approx. 3 mm into the liquid. 3. Let the control button slide back slowly. 4. Hold the Mastertip against the inner wall of the vessel while pulling it out of the liquid. Wipe off any drops on the outside with a lint-free cellulose tissue. Make sure not to aspirate any liquid from the Mastertip when wiping off the drops. 4.4 Dispensing liquid 1. Hold the Mastertip against the inner wall of the vessel at an angle. 2. Press down the control button to the first stop. 3. Hold down the control button and hold the Mastertip against the vessel wall while pulling it up. 4.5 Ejecting the Mastertip Press down the control button completely to eject the Mastertip. Small amounts of residual liquid at the tip of the Mastertip are not part of the dispensing volume and do not have any effect on precision and accuracy.

9 Maintenance 9 5 Maintenance 5.1 Testing the volume You can test the volume of the Biomaster by weighing the pipetted, demineralized water with a sufficiently sensitive analytical balance: The following components must have the same temperature: demineralized water weighing vessel pipette pipette tips Weighings should be conducted at 20 C 27 C. During the measurement the temperature may not fluctuate by more than ±0.5 C. Divide the weight of the water by its density ( mg/μl at 20 C) to calculate the volume. This procedure is based on ISO Regularly test the precision and accuracy of the Biomaster to avoid dispensing errors. You will find an SOP (Standard Operation Procedure) for testing pipettes on our website at Sterilization NOTICE! Damage to device from incorrect handling. Do not use any additional disinfectants, decontamination agents or sodium hypochlorite during autoclaving or UV exposure. When autoclaving, please ensure that the temperature does not exceed 121 C. Check the suitability of an agent and the manufacturer information about resistance to chemicals before using a disinfectant or decontamination agents. Also observe the material of the pipette. You can fully autoclave the Biomaster (at 121 C, for 20 min). Before doing so, unscrew and loosen the upper part of the device from its lower part by 2 turns. Make sure that the Biomaster has completely dried and cooled down before reassembling the device after autoclaving. The special plastic material of the housing allows a UV light treatment (up to 254 nm). The Mastertips can also be autoclaved at 121 C, 20 min. The temperature and time for autoclaving must be strictly observed. Otherwise, the proper functioning of the device cannot be guaranteed.

10 10 Maintenance 5.3 Cleaning and decontamination NOTICE! Damage to device from unsuitable cleaning fluids or sharp or pointed objects. Unsuitable cleaning fluids can damage the device. Never use corrosive cleaning fluids, strong solvents or abrasive polishes. Check the compatibility with the materials used. Do not clean the device with acetone or organic solvents with a similar effect. Do not clean the device with sharp objects. Proceed as follows: To remove any contamination on the outside, dampen a soft cloth with a mild detergent and wipe the housing. To disinfect the device, wipe the housing with isopropanol (70%). No further maintenance work is required. 5.4 Decontamination before shipment If you wish to return the device to Eppendorf AG or an Eppendorf AG service partner to be checked or repaired, please note the following: CAUTION! Use of a contaminated device may result in personal injuries and damage to the device. Clean and decontaminate the device in accordance with the cleaning instructions before shipping or storage. Hazardous substances are: solutions presenting a hazard to health potentially infectious agents organic solvents and reagents radioactive substances proteins presenting a hazard to health DNA

11 Technical data Please note the information in the document "Decontamination certificate for product returns". You can find it as a PDF file on our webpage 2. Enter the serial number of the Pipette in the decontamination certificate. 3. With the shipment please include the completed decontamination certificate for product returns with the Pipette. 4. Send the pipette to Eppendorf AG or an Eppendorf AG service partner. The address details of Eppendorf AG can be found on the last page of this operating manual. The list of service partners can be found at 6 Technical data 6.1 Test conditions Volume Systematic error Random error 2 μl ±6.0% 4.0% 3 μl ±5.0% 3.0% 5 μl ±4.0% 2.0% 10 μl ±3.0% 1.5% 20 μl ±2.0% 0.8% Test conditions and test evaluation in accordance with ISO 8655, part 6. Test with analytical balance with evaporation protection, approved by the board of weights and measures. Measurement errors do not apply to autoclaved Mastertips. When using less than 2 μl for the test, the sample liquid and the experience of the user have a critical influence on accuracy and precision. Therefore no limits are set for this range. Number of determinations: 10 Use of water in accordance with ISO 3696 Test with a Mastertip Test at 20 C 27 C ±0.5 C Dispensing on the inner wall of the vessel Technical specifications are subject to change.

12 12 Ordering Information 7 Ordering Information Order no. (International) Order no. (North America) Description Biomaster Kit consisting of: 1 Biomaster pipette and 1 box of 96 Mastertips Mastertips tips plus piston, ready-for-use boxes of 96 = 480 pieces Pipette carousel with 6 pipette holder for Research and Reference Pipettes Pipette holder For carousel (replacement) Pipette mount For wall mounting (with adhesive backing) for Research and Reference pipettes

13

14 Evaluate your manual Give us your feedback. Your local distributor: Eppendorf AG Hamburg Germany

1 Product description. Material. 3 Warnings for intended use. Instructions for use ept.i.p.s. Version B. English (EN)

1 Product description. Material. 3 Warnings for intended use. Instructions for use ept.i.p.s. Version B. English (EN) Instructions ept.i.p.s. see Fig. Tab. p. p. for Version use B 1 Product description The pipette tips ept.i.p.s. are single-use pipette tips for aspirating and dispensing liquids with air cushion pipettes.

More information

Eppendorf SmartBlock. Instructions for use

Eppendorf SmartBlock. Instructions for use Instructions for use Instructions Eppendorf see Fig. Tab. p. p. SmartBlock for use Copyright 2016 Eppendorf AG, Germany. All rights reserved, including graphics and images. No part of this publication

More information

ept.i.p.s. pipette tips are available with various degrees of purity, with and without filters, and as special tips.

ept.i.p.s. pipette tips are available with various degrees of purity, with and without filters, and as special tips. Instructions ept.i.p.s. see Fig. Tab. p. p. for General use Lab Product 1 Product description The pipette tips are single-use pipette tips for aspirating and dispensing liquids with air cushion pipettes.

More information

Eppendorf Vis Cuvettes. Instructions for use

Eppendorf Vis Cuvettes. Instructions for use Eppendorf Vis Cuvettes Instructions for use Copyright 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of the copyright owner. Eppendorf, the Eppendorf

More information

Eppendorf UVette. Instructions for use

Eppendorf UVette. Instructions for use Instructions for use Instructions Eppendorf see Fig. Tab. p. p. UVette for use Copyright 2014 Eppendorf AG, Hamburg, Germany. All rights reserved, including graphics and images. No part of this publication

More information

ept.i.p.s. Instructions for use Version B

ept.i.p.s. Instructions for use Version B r N) s use for use Instructions for use Version B Copyright 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of the copyright owner. Eppendorf and

More information

N) justment. Register your instrument! Eppendorf Research plus. Factory Adjustment

N) justment. Register your instrument!  Eppendorf Research plus. Factory Adjustment search ment plus N) justment Register your instrument! www.eppendorf.com/myeppendorf Factory Adjustment Copyright 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior

More information

Electrical safety testing Checklist

Electrical safety testing Checklist ety al Check testing Check - Checklist EN) ctrical Electrical safety testing Checklist SOP - Electrical Check Copyright 2013 Eppendorf AG. All rights reserved, including graphics and images. No part of

More information

Montgomery County Community College Document Number: MET DeKalb Pike Revision Number: 0

Montgomery County Community College Document Number: MET DeKalb Pike Revision Number: 0 Page 1 of 5 Approvals: Preparer: Jason McMillan Reviewer: Dr. Maggie Bryans Date: 08JAN14 Date: 10JAN14 1. Purpose: Operation of Eppendorf Research Plus through to the P-1000. 2. Scope: Applies to the

More information

stance et Helper Eppendorf Pipet Helper Chemical Resistance

stance et Helper Eppendorf Pipet Helper Chemical Resistance stance et Helper Resistance N) Chemical Resistance Copyright 2017 Eppendorf AG, Germany. All rights reserved, including graphics and images. No part of this publication may be reproduced without the prior

More information

PhysioCare Concept improved. The feeling of a new philosophy

PhysioCare Concept improved. The feeling of a new philosophy PhysioCare Concept improved The feeling of a new philosophy Intuitive operation Eppendorf Reference (fix), Reference (adjustable-volume) What's new? TÜV approved user-friendliness Optimized pipetting force

More information

PhysioCare Concept improved. The feeling of a new philosophy

PhysioCare Concept improved. The feeling of a new philosophy PhysioCare Concept improved The feeling of a new philosophy Intuitive operation Eppendorf Reference (fix), Reference (adjustable-volume) What's new? TÜV approved user-friendliness Optimized pipetting force

More information

ept.i.p.s. Version A

ept.i.p.s. Version A ept.i.p.s. Fig. Tab. p. Instructions see p. for for use use ept.i.p.s. Instructions for use Version A 0012 580.460 Copyright 2012 Eppendorf AG, Hamburg. No part of this publication may be reproduced without

More information

The New Legend. Eppendorf Reference 2 Pipette

The New Legend. Eppendorf Reference 2 Pipette The New Legend Eppendorf Reference 2 Pipette 2 Eppendorf Reference 2 Eppendorf Reference 2 3»The new Reference class for modern liquid handling tools.«the name»reference«stands for extraordinary precision

More information

Eppendorf BioSpectrometer 6135/6136/6137

Eppendorf BioSpectrometer 6135/6136/6137 nal Spectrometer Qualification 6135/6136/6137 - English N) rational Qualification - English Register your instrument! www.eppendorf.com/myeppendorf Eppendorf BioSpectrometer 6135/6136/6137 SOP - Operational

More information

Eppendorf Xplorer Eppendorf Xplorer plus. Ordering information. Register your instrument!

Eppendorf Xplorer Eppendorf Xplorer plus. Ordering information. Register your instrument! mation orer /Eppendorf Xplorer plus N) nformation Register your instrument! www.eppendorf.com/myeppendorf Eppendorf Xplorer Eppendorf Xplorer plus Ordering information Copyright 2015 Eppendorf AG, Germany.

More information

The New Legend. Eppendorf Reference 2 Pipette

The New Legend. Eppendorf Reference 2 Pipette The New Legend Eppendorf Reference 2 Pipette 2 Eppendorf Reference 2 Eppendorf Reference 2 3»The new Reference class for modern liquid handling tools.«the name»reference«stands for extraordinary precision

More information

Montgomery County Community College Document Number: MET DeKalb Pike Revision Number: 1

Montgomery County Community College Document Number: MET DeKalb Pike Revision Number: 1 Page 1 of 6 Approvals: Preparer: Jason McMillan Reviewer: Dr. Maggie Bryans Date: 08JAN14 Date: 10JAN14 1. Purpose: To verify the calibration of a single channel pipette. 2. Scope: Covers the cleaning,

More information

Eppendorf Reference : The pipette.

Eppendorf Reference : The pipette. Eppendorf Reference : The pipette. Reference class since 1994 Eppendorf Reference Reference class since 1994 The world of research and science is developing in shorter and shorter cycles. That s all for

More information

Well within Reach. The Eppendorf Pipette Holder System

Well within Reach. The Eppendorf Pipette Holder System Well within Reach The Eppendorf Pipette Holder System 3 Eppendorf Pipette Holder System Pipette Stand 2 and Charger Stand 2 Protect your precious liquid handling device from possible damage, contamination

More information

No 022 December Application-Specific Requirements for Consumables in the Laboratory

No 022 December Application-Specific Requirements for Consumables in the Laboratory Userguide No 022 December 2011 Application-Specific Requirements for Consumables in the Laboratory Kornelia Ewald, Eppendorf AG, Hamburg, Germany Abstract The demands placed on consumables made of plastic

More information

Spa & Wellness for Pipettes

Spa & Wellness for Pipettes epservices for premium performance Spa & Wellness for Pipettes Pipette Performance Plans Treat your pipettes well with premium services from the experts! Total approach philosophy For over 60 years Eppendorf

More information

Eppendorf ept.i.p.s. LoRetention Comparison of low retention pipette tips via simple and fast absorbance measurements in the BioPhotometer plus

Eppendorf ept.i.p.s. LoRetention Comparison of low retention pipette tips via simple and fast absorbance measurements in the BioPhotometer plus APPLICATION NOTE No. 218 I November 2009 Eppendorf ept.i.p.s. LoRetention Comparison of low retention pipette tips via simple and fast absorbance measurements in the BioPhotometer plus Natascha Weiß 1,

More information

Simply Better Pipetting

Simply Better Pipetting Simply Better Pipetting Eppendorf Xplorer and Eppendorf Xplorer plus the electronic pipettes »Work Intuitively.«People who give 100 % every day deserve the best tools and the best equipment. You work

More information

Eppendorf Temperature Verification System. Operating manual /0909

Eppendorf Temperature Verification System. Operating manual /0909 Eppendorf Temperature Verification System 6328 900.066-00/0909 Copyright 2009 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of the copyright owner. eppendorf

More information

Technical Report Exact dispensing of whole blood using the Eppendorf positive displacement system Multipette Xstream /Repeater Xstream

Technical Report Exact dispensing of whole blood using the Eppendorf positive displacement system Multipette Xstream /Repeater Xstream Applications Note 242 September 2011 Technical Report Exact dispensing of whole blood using the Eppendorf positive displacement system /Repeater Natascha Weiß 1, Wei Sze Heng 2, Foong Teng Lu 2 1 Eppendorf

More information

symphony Reference Electrodes

symphony Reference Electrodes DOC022.53.80322 symphony Reference Electrodes 07/2012, Edition 1 User Manual Table of contents Safety information on page 3 Preparation for use on page 4 Specifications on page 3 Measurement on page 5

More information

PhysioCare Concept. Make your Lab a better place

PhysioCare Concept. Make your Lab a better place PhysioCare Concept Make your Lab a better place 2 Eppendorf PhysioCare Concept The Eppendorf PhysioCare Concept The mission of Eppendorf has always been to improve the living conditions of their customers.

More information

96 Well? Done! Eppendorf epmotion 96 semi-automated electronic pipette for fast and precise 96 channel microplate processing

96 Well? Done! Eppendorf epmotion 96 semi-automated electronic pipette for fast and precise 96 channel microplate processing 96 Well? Done! Eppendorf epmotion 96 semi-automated electronic pipette for fast and precise 96 channel microplate processing Try it and you will see how easy and fast your 96-well and 384-well plates get

More information

Larger Display Easier to Use. Generation E. Repeater E3 and E3x: Dispense without compromising accuracy or precision

Larger Display Easier to Use. Generation E. Repeater E3 and E3x: Dispense without compromising accuracy or precision Larger Display Easier to Use Generation E Repeater E3 and E3x: Dispense without compromising accuracy or precision 2 Eppendorf Repeater E3 / E3x > > Repeater E3 and Repeater E3x are the experts for long

More information

Precise and Accurate whole Blood Dispensing with Multipette E3x and Combitips advanced

Precise and Accurate whole Blood Dispensing with Multipette E3x and Combitips advanced APPLICATION NOTE No. 386 Precise and Accurate whole Blood Dispensing with Multipette E3x and Combitips advanced Vincent Dufey¹, Muriel Art¹, Hanaë Henke² ¹Eppendorf Application Technologies S.A., Namur,

More information

Raise the limits. Eppendorf Research plus

Raise the limits. Eppendorf Research plus NEW! NEW! Raise the limits Eppendorf Research plus A completely new experience Eppendorf Research plus Eppendorf sets a new standard in pipetting. The ultra light Eppendorf Research plus pipette meets

More information

December 2010 No 036. The electronic Eppendorf Xplorer pipette Versatile adjustment

December 2010 No 036. The electronic Eppendorf Xplorer pipette Versatile adjustment Userguide December 2010 No 036 The electronic Eppendorf Xplorer pipette Versatile adjustment Kornelia Ewald, Eppendorf AG, Hamburg, Germany Abstract Piston-stroke pipettes with an air cushion are subject

More information

Micro Volume QuEChERS kit

Micro Volume QuEChERS kit 225-37872 Sep. 2018 Small Capacity Pretreatment Kit Micro Volume QuEChERS kit Instruction Manual Read this manual thoroughly before you use the product. Keep this manual for future reference. This page

More information

Latest generation testo air velocity & IAQ probes with cable handle. Instruction manual

Latest generation testo air velocity & IAQ probes with cable handle. Instruction manual Latest generation testo air velocity & IAQ probes with cable handle Instruction manual Contents Contents 1 About this document... 5 2 Safety and disposal... 5 3 System description... 6 4 Description of

More information

3. Chemical Hygiene Plan: Laboratory Standard Operating Procedures. A. Laboratory Specific Information and Signatures

3. Chemical Hygiene Plan: Laboratory Standard Operating Procedures. A. Laboratory Specific Information and Signatures 3. Chemical Hygiene Plan: Laboratory Standard Operating Procedures A. Laboratory Specific Information and Signatures The Chemical Hygiene Plan: Laboratory Standard Operating Procedures (section 3 only),

More information

Cleaning and Decontamination of the ept.i.p.s. Box

Cleaning and Decontamination of the ept.i.p.s. Box USERGUIDE No. 38 Cleaning and Decontamination of the ept.i.p.s. Box Kornelia Ewald 1, Detlef Hempfling 2 1 Eppendorf AG, 2 Eppendorf Instrumente, Hamburg, Germany Introduction Various procedures are available

More information

The Pure Truth. Eppendorf Forensic DNA Grade according to ISO 18385

The Pure Truth. Eppendorf Forensic DNA Grade according to ISO 18385 The Pure Truth Eppendorf Forensic DNA Grade according to ISO 18385 2 Eppendorf Forensic DNA Grade»Eppendorf Forensic DNA Grade, for results you can trust.«the prevention of contamination is one of the

More information

Spa & Wellness for Pipettes

Spa & Wellness for Pipettes epservices for premium performance Spa & Wellness for Pipettes Pipette Performance Plans Treat your Pipettes Well with Premium Services from the Experts! Total approach philosophy For over 60 years Eppendorf

More information

Now also in Eppendorf LoBind. Conical Competence. Eppendorf Conical Tubes 15 ml and 50 ml

Now also in Eppendorf LoBind. Conical Competence. Eppendorf Conical Tubes 15 ml and 50 ml Now also in Eppendorf LoBind Conical Competence Eppendorf Conical Tubes 15 ml and 50 ml 2 Eppendorf Conical Tubes The Eppendorf Tubes Best Performance for Your Samples The Eppendorf Conical Tubes 15 ml

More information

NEW: 5.0 ml screw cap tube. The Missing Link. Eppendorf Tubes 5.0 ml Discover a new sample handling system

NEW: 5.0 ml screw cap tube. The Missing Link. Eppendorf Tubes 5.0 ml Discover a new sample handling system NEW: 5.0 ml screw cap tube The Missing Link 5.0 ml Discover a new sample handling system 5mL_Tube_BRO_GB_2015.indd 1 27.11.15 14:02 2 Combined with the comprehensive line of accessories, 5.0 ml are the

More information

Technology that counts

Technology that counts Application Note No. 102 Version 1.3 Determination of Sperm Density in Ejaculates from Bulls Application This Application Note describes how sperm cell density in ejaculates from bulls can be determined

More information

High-throughput Isolation of Genomic DNA From Buccal Swab on the Eppendorf epmotion 5075 VAC

High-throughput Isolation of Genomic DNA From Buccal Swab on the Eppendorf epmotion 5075 VAC APPLICATION NOTE No. 378 I August 2016 High-throughput Isolation of Genomic DNA From Buccal Swab on the Eppendorf epmotion 5075 VAC Hatim Boughanem, Manuel Macias-Gonzalez, BIOEASY SL, Spain Abstract The

More information

NEW! ep Dualfilter T.I.P.S. Peak of perfection

NEW! ep Dualfilter T.I.P.S. Peak of perfection NEW! ep Dualfilter T.I.P.S. Peak of perfection A generation of tips to set standards Eppendorf pipette tips an investment in quality Eppendorf Totally Integrated Pipetting System Eppendorf Totally Integrated

More information

RADIATION SAFETY GUIDELINES FOR NON-USERS

RADIATION SAFETY GUIDELINES FOR NON-USERS RADIATION SAFETY GUIDELINES FOR NON-USERS This is a Read and Sign Awareness Training document. You should read and sign this document if you: 1. DO NOT work directly with radioactive materials, but 2.

More information

HHPS WHMIS. Rules MSDS Hazard Codes Systems. Biology based. Chemistry based. Safety Symbols. Safety in the Lab. Lab Equipment

HHPS WHMIS. Rules MSDS Hazard Codes Systems. Biology based. Chemistry based. Safety Symbols. Safety in the Lab. Lab Equipment Safety Symbols HHPS WHMIS Safety in the Lab Rules MSDS Hazard Codes Systems Lab Equipment Chemistry based Biology based Safety Symbols We will be discussing two types of Information Systems Hazardous Household

More information

Automated purification of high quality genomic DNA

Automated purification of high quality genomic DNA APPLICATION NOTE No. AA267 I October 2012 Automated purification of high quality genomic DNA from various tissues using the Eppendorf MagSep Tissue gdna Kit on the Eppendorf epmotion M5073 Ulrich Wilkening,

More information

PUBLIC EMPLOYEE HAZARDOUS CHEMICAL PROTECTION AND RIGHT TO KNOW ACT O.C.G.A

PUBLIC EMPLOYEE HAZARDOUS CHEMICAL PROTECTION AND RIGHT TO KNOW ACT O.C.G.A PUBLIC EMPLOYEE HAZARDOUS CHEMICAL PROTECTION AND RIGHT TO KNOW ACT O.C.G.A. 45-22-2 Georgia s Right to Know Law Federal regulations require that all employees be trained on the Hazard Communications Standard

More information

Ionizing Radiation Awareness (Non-User)

Ionizing Radiation Awareness (Non-User) Ionizing Radiation Awareness (Non-User) What is radiation? Radiation comes from particles or rays emitted by unstable elements (radioisotopes) or from x-rays x produced directly or indirectly by high-voltage

More information

stance 4 Repeater M4 Register your instrument! Multipette M4 Repeater M4 Chemical Resistance

stance 4 Repeater M4 Register your instrument!   Multipette M4 Repeater M4 Chemical Resistance stance 4 Repeater M4 Resistance N) Register your instrument! www.eppendorf.com/myeppendorf Chemical Resistance Copyright 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without

More information

The information contained in this SOP is general in nature. Any YouTube videos included are as a compliment to the information presented.

The information contained in this SOP is general in nature. Any YouTube videos included are as a compliment to the information presented. Labelling of Hazardous Chemicals (GHS) in Australian Workplaces Risks & Hazards Hazard Control PPE General Safety Operating Safety Maintenance Safe Work Zones Operating Procedures Select Language The information

More information

Type B xxx/60. Snowplough // USER MANUAL Version 1.0 EN. Art.Nr.:

Type B xxx/60. Snowplough // USER MANUAL Version 1.0 EN. Art.Nr.: USER MANUAL Type B xxx/60 Snowplough 21.12.2012 Version 1.0 EN Art.Nr.: 10040218 Kugelmann Maschinenbau e.k. Gewerbepark 1-5 87675 Rettenbach a.a. GERMANY www.kugelmann.com office@kugelmann.com +49 (0)

More information

Individual Laboratory Safety & Chemical Hygiene Plan

Individual Laboratory Safety & Chemical Hygiene Plan APPENDIX A: CHEMICAL HYGIENE PLAN Part B Individual Laboratory Safety & Chemical Hygiene Plan Principal Investigator Date filed Department Building/Floor/Room E-mail address @Pace.edu Phone Facsimile Laboratory

More information

Department of Environmental Health and Safety Laboratory Decommissioning and Clearance

Department of Environmental Health and Safety Laboratory Decommissioning and Clearance Department of Environmental Health and Safety Laboratory Decommissioning and Clearance 1.0 Purpose The Laboratory Decommissioning and Clearance Policy is designed to assist Principal Investigators who

More information

Devyser Index Plate A. Instructions for Use

Devyser Index Plate A. Instructions for Use Devyser Index Plate A Art. No.: 8-A200 Instructions for Use Page 1 of 17 TABLE OF CONTENTS TABLE OF CONTENTS 2 1. INTRODUCTION TO DEVYSER INDEX PLATE A 3 1.1 Intended use 3 1.2 Kit configuration Devyser

More information

mag nanogram kit Description Kit uses

mag nanogram kit Description Kit uses mag nanogram kit Catalogue number 40701 and 40710 (For research use only. Not for use in diagnostic procedures.) Description mag kits use magnetic separation for the preparation of nucleic acids. Superparamagnetic

More information

2.8-1 SCIENCE EXPERIMENTS ON FILE Revised Edition. Dew Formation

2.8-1 SCIENCE EXPERIMENTS ON FILE Revised Edition. Dew Formation 2.8-1 SCIENCE EXPERIMENTS ON FILE Revised Edition Dew Formation Topic Dew Time 30 minutes! Safety Please click on the safety icon to view safety precautions. Be careful using the thermometer. Be careful

More information

THE UNIVERSITY OF NEWCASTLE SCHOOL of BIOMEDICAL SCIENCES

THE UNIVERSITY OF NEWCASTLE SCHOOL of BIOMEDICAL SCIENCES THE UNIVERSITY OF NEWCASTLE SCHOOL of BIOMEDICAL SCIENCES STANDARD OPERATING PROCEDURE PROCEDURE NO: BS-GDP 018 MOD: 3 rd Issue Page: 1 of 5 Procedure Type: General Discipline Procedure 1. Risk Assessment:

More information

Determination of Chloride using Potentiometry

Determination of Chloride using Potentiometry Determination of Chloride using Potentiometry 1. Purpose This procedure will determine the concentration of chloride ion with a chloride specific ion electrode using potentiometry. 2. Background Potentiometry

More information

Geneva College Hazard Communication Program Presentation

Geneva College Hazard Communication Program Presentation Geneva College Hazard Communication Program Presentation Design 2005, 2012 Zywave, Inc. All rights reserved. Hazard Communication: Agenda In today s session, we will discuss the following: - Our Hazard

More information

Technology that counts

Technology that counts Application Note No. 103 Version 1.2 Determination of Sperm Density in Bull Semen Doses for AI Application This Application Note describes how sperm cell density in bull semen doses for Artificial Insemination

More information

Simply better pipetting!

Simply better pipetting! NEW! Great results, congratulations!! Simply better pipetting! Eppendorf Xplorer the new electronic pipette Simply better pipetting Eppendorf Xplorer People who give 100 % every day deserve the best tool

More information

NukEx Nucleic Acid Release Reagent

NukEx Nucleic Acid Release Reagent Instruction for Use NukEx Nucleic Acid Release Reagent For general laboratory use. For in vitro use only. Reagent for the enzymatic release of nucleic acid from tissue samples, ticks, insects and swabs.

More information

NGS Made Easy. Optimize your NGS library preparation with the epmotion Automated liquid handling system

NGS Made Easy. Optimize your NGS library preparation with the epmotion Automated liquid handling system NGS Made Easy Optimize your NGS library preparation with the epmotion Automated liquid handling system NGS Library Preparation Made Easy and Reliable Next-generation sequencing sample preparation is a

More information

Easy Illumina Nextera DNA FLEX Library Preparation using the epmotion 5075t automated liquid handler

Easy Illumina Nextera DNA FLEX Library Preparation using the epmotion 5075t automated liquid handler WHITE PAPER No. 13 Easy Illumina Nextera DNA FLEX Library Preparation using the epmotion 5075t automated liquid handler Executive Summary Library preparation steps, including DNA extraction, quantification,

More information

Lamp and Control Panel (Lamp Panel)

Lamp and Control Panel (Lamp Panel) www.reinhausen.com Lamp and Control Panel (Lamp Panel) Operating Instructions BA 47/0 EN NOTE! Changes may have been made to a product after going to press with this documentation. We expressly reserve

More information

How do I assess the current condition of the chemical inventory?

How do I assess the current condition of the chemical inventory? Appendix 5.2. Chemical Safety SCIENCE SAFETY IN OUR SCHOOLS Managing Your Chemical Inventory, Part 1 What is chemical management? A chemical management plan identifies, manages, and prevents hazards through

More information

Responsibilities: Effective Date: November Revision Date: February 8, VP, Facilities and Construction Management. Issuing Authority:

Responsibilities: Effective Date: November Revision Date: February 8, VP, Facilities and Construction Management. Issuing Authority: Title: Chemical Hygiene Written Program Effective Date: November 2005 Revision Date: February 8, 2017 Issuing Authority: Responsible Officer: VP, Facilities and Construction Management Director Environmental

More information

WHAT S WRONG WITH THIS PICTURE?

WHAT S WRONG WITH THIS PICTURE? 1606 WHAT S WRONG WITH THIS PICTURE? Hazard Communication & GHS Leader s Guide Aurora Pictures Program Synopsis WHAT S WRONG WITH THIS PICTURE? Hazard Communication & GHS This video holds the viewer's

More information

Ballistic pendulum Operating Instructions Fig. 1: Ballistic pendulum SAFETY PRECAUTIONS

Ballistic pendulum Operating Instructions Fig. 1: Ballistic pendulum SAFETY PRECAUTIONS R Ballistic pendulum 11229.00 PHYWE Systeme GmbH & Co. KG Robert-Bosch-Breite 10 D-37079 Göttingen 6 3.8 3.7 3.6 3.5 Phone +49 (0) 551 604-0 Fax +49 (0) 551 604-107 E-mail info@phywe.de Internet www.phywe.de

More information

Technology that counts

Technology that counts Application Note No. 100 Version 1.3 Determination of Sperm Density in Ejaculates from Boars Application This Application Note describes how sperm cell density in ejaculates from boars can be determined

More information

Safe Method of Use 2 Workshop and Arts Studios

Safe Method of Use 2 Workshop and Arts Studios Safe Method of Use 2 Workshop and Arts Studios Purpose: This Safe Method of Use applies to principal investigators (PIs), sector managers, designated laboratory person (DLPs), technical staff and students

More information

Installation guide 862 MIT / MIR

Installation guide 862 MIT / MIR Installation guide 862 MIT / MIR in combination with: 864 MTT or 863 MRT or (dual) spot element November 2005 Part no. 4416.232_Rev3 Enraf BV PO Box 812 2600 AV Delft Netherlands Tel. : +31 15 2701 100

More information

Health, Safety, Security and Environment

Health, Safety, Security and Environment Document owner and change code Document Owner Aaron Perronne Title HSSE Manager Mark X Change Code Description X N/A First Issue A Typographical/Grammatical correction; formatting change; text clarification-no

More information

C h a p t e r 5 : W o r k p l a c e H a z a r d o u s M a t e r i a l s I n f o r m a t i o n S y s t e m ( W H M I S )

C h a p t e r 5 : W o r k p l a c e H a z a r d o u s M a t e r i a l s I n f o r m a t i o n S y s t e m ( W H M I S ) C h a p t e r 5 : W o r k p l a c e H a z a r d o u s M a t e r i a l s I n f o r m a t i o n S y s t e m ( W H M I S ) Overview WHMIS is a provincial legislative response to provincial employees and employers

More information

Automated pipetting made easy new ep Motion 5073 family members for perfect PCR set-up and nucleic acids purifi cation or general liquid handling

Automated pipetting made easy new ep Motion 5073 family members for perfect PCR set-up and nucleic acids purifi cation or general liquid handling Ready, Set, Prep Automated pipetting made easy new epmotion 5073 family members for perfect PCR set-up and nucleic acids purification or general liquid handling 2 Eppendorf epmotion Eppendorf epmotion

More information

symphony Multisensor Probes

symphony Multisensor Probes DOC022.53.80319 symphony Multisensor Probes 07/2012, Edition 1 User Manual Table of contents Safety information on page 3 Calibration on page 6 Specifications on page 3 Sample requirements on page 8 Product

More information

Objective: Science Classroom Laboratory Safety

Objective: Science Classroom Laboratory Safety Science Classroom Laboratory Safety Objective: Learn safety rules for working with chemicals and participating in a safe manner when carrying out lab procedures. Complete the safety and equipment assignment

More information

INSTALLING TRAK SPORT TIRE CHAINS ON 20 MODEL 3 TIRES

INSTALLING TRAK SPORT TIRE CHAINS ON 20 MODEL 3 TIRES INSTALLING TRAK SPORT TIRE CHAINS ON 20 MODEL 3 TIRES 2018 TESLA, INC. All rights reserved. All information in this document and all MODEL 3 software is subject to copyright and other intellectual property

More information

Laboratory Chemical Safety. Pathology Training

Laboratory Chemical Safety. Pathology Training Laboratory Chemical Safety Pathology Training Objectives Explain the OSHA standard for chemical safety as it relates to laboratory workers Explain Chemical labeling requirements Explain the NFPA diamond

More information

7/8/2013. What is GHS?

7/8/2013. What is GHS? What is GHS? 0 Globally Harmonized System of Classification and Labeling of Chemicals 0 A standardized approach to classifying and labeling chemicals based on hazards 0 A United Nations program meant to

More information

Hazard Communication & Chemical Safety. Based on OSHA Standard

Hazard Communication & Chemical Safety. Based on OSHA Standard Hazard Communication & Chemical Safety Based on OSHA Standard 1910.1200 We use many chemicals We want you to know how to use them safely You will learn about The Hazards of Chemicals Our Written Program

More information

Marquette University Effective: 5/21/02 Hazard Communication Program Updated: 03/11/09 Revision No.: 1 Page: 1

Marquette University Effective: 5/21/02 Hazard Communication Program Updated: 03/11/09 Revision No.: 1 Page: 1 Marquette University Effective: 5/21/02 Hazard Communication Program Updated: 03/11/09 Revision No.: 1 Page: 1 Introduction: The Occupational Safety & Health Administration (OSHA) Hazard Communication

More information

AM 500A 500 WATT AM 1A 1 KILOWATT AM BROADCAST TRANSMITTERS. October, 1999 IM No

AM 500A 500 WATT AM 1A 1 KILOWATT AM BROADCAST TRANSMITTERS. October, 1999 IM No AM 500A 500 WATT AM 1A 1 KILOWATT AM BROADCAST TRANSMITTERS October, 1999 IM No. 597 1112 EQUIPMENT LOST OR DAMAGED IN TRANSIT. When delivering the equipment to you, the truck driver or carrier s agent

More information

Multipette / Repeater M4 allows for fast, precise and sterile liquid transfer in cell culture

Multipette / Repeater M4 allows for fast, precise and sterile liquid transfer in cell culture APPLICATION NOTE No. 350 I May 2015 Multipette / Repeater M4 allows for fast, precise and sterile liquid transfer in cell culture Nathalie Chandelier¹, Vincent Dufey¹, Muriel Art¹, Diana Hübler² ¹Eppendorf

More information

COMMODORE BRASS TELESCOPE ZHUMELL COMMODORE BRASS TELESCOPE

COMMODORE BRASS TELESCOPE ZHUMELL COMMODORE BRASS TELESCOPE 3 0 15 15 E OWNER S MANUAL COMMODORE BRASS TELESCOPE ZHUMELL COMMODORE BRASS TELESCOPE 5 W 60 W 30 W 45 W Zhumell customers know that there are plenty of ways to experience the world. They also understand

More information

ODZ-033. Instruction manual

ODZ-033. Instruction manual RUO ELISA-VIDITEST anti-vimentin IgM ODZ-033 Instruction manual PRODUCER: VIDIA spol. s r.o., Nad Safinou II 365, Vestec, 252 42 Jesenice, Czech Republic Tel.: + 420 261 090 565, www.vidia.cz 1. TITLE

More information

Radioactive Waste Policies

Radioactive Waste Policies Radioactive Waste Policies APRIL 8, 2012 By Maury Riner Approved: April 8, 2010 Radiation Safety Committee TABLE OF CONTENTS 1. Purpose 5 2. Scope 5 3. Radioactive Waste Classification 5 3.1 Waste Definitions

More information

Operating Manual. Turb 355 IR/T. Portable turbidimeter. ba75454e03 09/

Operating Manual. Turb 355 IR/T. Portable turbidimeter. ba75454e03 09/ Operating Manual Turb 355 IR/T Portable turbidimeter ba75454e03 09/2010 41 Accuracy when going to press The use of advanced technology and the high quality standard of our instruments are the result of

More information

Automated Illumina TruSeq DNA PCR-Free library construction with the epmotion 5075t/TMX

Automated Illumina TruSeq DNA PCR-Free library construction with the epmotion 5075t/TMX SHORT PROTOCOL No. 15 I August 2016 Automated Illumina TruSeq DNA PCR-Free library construction with the epmotion 5075t/TMX Introduction This protocol describes the configuration and preprogrammed methods

More information

Hazard Communications

Hazard Communications 1 Hazard Communications 1 2 Hazard Communication Program Table of Contents 1. Purpose of the Hazard Communication Program 2. Access to Written Program 3. Responsibilities 4. Hazard Recognition/Determination

More information

Introduction. Sample kit content

Introduction. Sample kit content Introduction Polymer blend sample preparation TN01062 Preparation of polymer samples available from the Lateral force, Phase imaging, and Force modulation mode kits The SBS-PMMA, SBS-PS, and SBR-PMMA samples

More information

HAZARD COMMUNICATION and GHS. Environmental Health and Safety

HAZARD COMMUNICATION and GHS. Environmental Health and Safety HAZARD COMMUNICATION and GHS Environmental Health and Safety RIGHT-TO-KNOW LAW OSHA created the Hazard Communication Standard in 1983. Applies to all places of employment where employees are exposed to

More information

Microscale Acid-Base Titration

Microscale Acid-Base Titration Microscale Acid-Base Titration Experiment 36 A titration is a process used to determine the volume of a solution needed to react with a given amount of another substance. In this experiment, you will titrate

More information

Guidelines for the Labelling of Chemical Products

Guidelines for the Labelling of Chemical Products UFV 2008/759 Guidelines for the Labelling of Chemical Products Uppsala University Approved by the University Director 29 May 2008, revised 21 December 2015 Contents Labelling of chemical products 3 Responsibility

More information

Standard Operating Procedure

Standard Operating Procedure Standard Operating Procedure Procedure Radioimmunoassay with I Department Location SOP Prepared By: Section 1: Purpose Radioimmunoassays are used for detecting the concentration of a specific antigen or

More information

3B SCIENTIFIC PHYSICS

3B SCIENTIFIC PHYSICS B SCIENTIFIC PHYSICS Instruction sheet /5 MH 0066 Electromagnetic Experiment Set Knurled screws to fasten the cross bar Threaded holes (5x) to mount the cross bar Cross bar 4 Conductor swing 5 Stand 6

More information

Canine brain natriuretic peptide,bnp ELISA Kit

Canine brain natriuretic peptide,bnp ELISA Kit Canine brain natriuretic peptide,bnp ELISA Kit Catalog No: E0541c 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH

More information

Monkey Kidney injury molecule 1,Kim-1 ELISA Kit

Monkey Kidney injury molecule 1,Kim-1 ELISA Kit Monkey Kidney injury molecule 1,Kim-1 ELISA Kit Catalog No: E0785Mo 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH

More information