Determination of Silicate in Seawater by Inductively Coupled Plasma Atomic Emission Spectrometry

Size: px
Start display at page:

Download "Determination of Silicate in Seawater by Inductively Coupled Plasma Atomic Emission Spectrometry"

Transcription

1 Journal of Oceanography Vol. 48, pp. 283 to Determination of Silicate in Seawater by Inductively Coupled Plasma Atomic Emission Spectrometry KAZUO ABE and YASUNORI WATANABE Seikai National Fisheries Research Institute, 49 Kokubu-cho, Nagasaki 850, Japan (Received 4 January 1992; in revised form 23 March 1992; accepted 31 March 1992) Inductively Coupled Plasma Atomic Emission Spectrometry (ICP) was adopted for direct analysis of silicate in seawater, eliminating necessity for pre-treatment. Via this method, the determination of silicate is rapid and easy compared with conventional methods of colorimetry. In seawater analysis, a decrease of sensitivity of about 24% was observed due to interference by coexisting elements, mainly Na. Examination of the analytical conditions revealed a detection limit of 0.3 µm Si, and precisions of approximately 3.2, 2.0 and 1.3% for Si levels of 4, 15 and 94 µm, respectively. Using this method, silicate determination in the East China Sea was attempted. 1. Introduction Silicate is one of the important parameters which control marine productivity. Many reports have been published on its behavior (e.g. Banahan and Goering, 1986; Demaster et al., 1983; Nelson and Gordon, 1982). Silicate levels in seawater range from 0 to maximum concentrations of about 240 µm in the Bering Sea (Tsunogai et al., 1979). Determinations are ordinarily conducted by colorimetric methods. Atkins (1923, 1926) applied the yellow silicomolybdic acid method to the analysis of seawater and Robinson and Thompson (1948) described the method with minor modifications. Briefly, this method involves the reaction of molybdate with silicate in seawater, resulting in the formation of silicomolybdate complexes having a yellow color. Mullin and Riley (1955) employed the addition of a reducing solution to the silicomolybdate complex to induce a blue color, hereby enhancing sensitivity. Generally, the formation of the silicomolybdate complex requires several hours. AAS (Atomic Absorption Spectrometry) can be adopted for direct analysis of silicate, but seawater samples are not suitable for analysis by the equipment, due to interference by alkali and alkaline earth salts. Separation of a specific element from these salts should be conducted prior to analysis. In the case of Si determination, the colorimetric methods noted above require the addition of chemical reagents to sample solutions for the formation of complexes. It is sometimes very difficult to obtain silicate-free water or to remove reagent contamination. Consequently, an accurate blank value can not be determined. Separation of a specific element from alkali and alkali earth salts usually requires some kind of treatment, including the addition of chemical reagents. Thus, the risk of reagent contamination should be considerable. The Inductively Coupled Plasma Atomic Emission Spectrometry (ICP) method has been used since 1964 (Greenfield et al., 1964). This is a very sensitive method for multi-element determination, and can be applied to seawater analysis. Akagi et al. (1985) have studied the determination of some trace metals in coastal seawater using this method via coprecipitation with gallium hydroxide. This method has an advantage in that seawater and can be introduced to the

2 284 K. Abe and Y. Watanabe inside of the plasma torch without separating alkali and alkaline earth salts. In this paper, a rapid and simple method for the determination of silicate in seawater by ICP was examined. Subsequently, this method was adapted through water samples from the East China Sea. 2. Experiment 2.1 Apparatus and reagents Analysis of silicate were conducted using the Jarrell-Ash ICAP (Inductively Coupled Argon Plasma Atomic Emission Spectrophotometer)-575 MARK II. Specifications of this equipment are shown in Table 1. After setting analytical conditions, and making background corrections for wave length spectra in accordance with the standard solution profile, sample or test solutions were introduced via the nebulizer inside the plasma torch. Intensity was measured at 17 mm above work coil and determination of silicate in seawater was made by standard additional methods. All reagents used in this study were of special grade, and a Si standard solution was prepared from a 1000 ppm standard solution (Na 2 SiO 3 in 0.2 M Na 2 CO 3 ) for AAS analysis (Wako Pure Chemical Industries, Ltd.) by diluting with distilled water, prior to each analysis. 2.2 Analytical conditions Analytical conditions were examined using a test solution (about 10 µm Si). The possible effects of gas flow rate (coolant, auxiliary, carrier) and spectral interferences from major elements present in seawater (Na, Mg, K and Ca) were examined in order to determine suitable conditions. Furthermore, parameters such as voltage applied to work coil, integration time for calculation, and output power were checked. Values of silicate determined under these analytical conditions were also confirmed by colorimetric methods. 2.3 Field observation Surface water samples were collected via a clean, plastic bucket at the stations (St. 1 St. 6) shown in Fig. 1 on October 24 25, In addition, vertical sampling was carried out with Go- Flo sampling bottles at St. 7 (see Fig. 1) on 25 October After collection, all samples were transferred to carefully acid-washed plastic centrifuge tubes and frozen immediately to circumvent adsorption to tube walls and/or chemical change until laboratory analysis. Table 1. Operational parameters. Spectrometer Jarrell-Ash ICAP-575 MARK II frequency MHz output power (max) 2.5 kw Nebulizer Cross-flow type Monochrometer 0.75 m Zerny-Terner type grating 1800 grooves/mm reciprocal linear dispersion 0.79 nm/mm entrance slit width 25 µm exit slit width 25 µm Wave length nm Torch Fassel type, Quartz

3 Determination of Silicate in Seawater 285 Fig. 1. Sampling stations in the East China Sea. 3. Results and Discussion The profile for the Si standard solution analyzed by ICP in this laboratory is shown in Fig. 2. The detected signal (intensity) shown in this figure is composed not only of emission spectra line originating from the specific element, but also from other components that are not proportional to the concentration of an objective element. If the intensity (concentration) of a certain element is sufficiently high in comparison to the intensity of these other components, this phenomenon can be ignored. A background correction will effectively compensate for this interference. In Fig. 2 the symbol, *, corresponds to the intensity from interferential components. Thus, subtracting signals (intensity) from these components gives a true intensity which depends only on the specific element. L and H show the low and high points of the wave length, respectively, for the background correction sphere. Figure 3 shows the effect of flow rate of coolant gas for Si. Open and closed circles correspond to distilled water and seawater, respectively. Coolant gas is necessary for cooling the quartz torch and preventing air contamination of the argon plasma. When the flow rate is low, emission of NO, NH, or N 2 may be generated due to air contamination. This can cause spectrum interference, so background correction can not be made easily, and a melting of the quartz torch may possibly occur. As shown in Fig. 3, only a small change of intensity is observed with variations of flow rate of the coolant gas. For the two reasons noted above, flow rates above 12 l/min should be applied. The auxiliary gas is necessary for the argon plasma to float above the torch and prevents from sticking salts or carbon inside the torch. Therefore, when analysis of a solution containing highly concentrated salts (e.g. seawater) is carried out, the flow rate should be kept high. Although decreasing intensity is observed with increasing flow rate during Si analysis (Fig. 4), flow rates of more than 0.5 l/min are necessitated.

4 286 K. Abe and Y. Watanabe Fig. 2. Profile of Si standard solution. The symbol * indicates background intensity. L and H correspond to the low and high points of background correction. Fig. 3. Flow rate of coolant gas and the Si intensity. Open and closed circles correspond to distilled and seawater, respectively. Carrier gas is used to introduce an aerosol (sample) from the nebulizer into the plasma. This flow rate is a main indicator of sensitivity and precision. Variation in intensity as a function of flow rate is shown in Fig. 5. Open and closed circles indicate distilled and seawater, respectively. The highest sensitivity for Si determination was 0.40 l/min. Figure 6 shows the effects of coexistence elements on the determination of silicate. In this study the effect of Na, Mg, Ca and K were examined because of their abundance in seawater. The Y and X axes indicate relative intensity compared with that of distilled water without any kind

5 Determination of Silicate in Seawater 287 Fig. 4. Effects of auxiliary gas in Si analysis. Open circles indicate distilled water and closed circles, seawater. Fig. 5. Influence of carrier gas on Si analysis. Open and closed circles indicate distilled and seawater, respectively. of metals, and concentrations of added coexistence elements, respectively. An arrow in Fig. 6 indicates the concentration level of each element in seawater. Interference in determination of Si was observed for every coexistence element examined in this study. At seawater levels, the effect of Na interference is great (about 80%). In Fig. 7 the effects of these elements regarding sensitivity in Si determination are illustrated. In this experiment, filtered seawater (surface

6 288 K. Abe and Y. Watanabe Fig. 6. Interference of coexistence elements on the determination of silicate. Fig. 7. Effects of coexistence elements for sensitivity of Si determination. seawater of the Kuroshio current), distilled water, and mixed solution, i.e. seawater: distilled water = 1:1 were used. Standard additional methods were adapted, so it is not necessary that the three lines in Fig. 7 start from the origin. Sensitivities (slope) in Fig. 7 are 355, 270, and 300 intensity/1 µm Si for distilled water, seawater, and mixed solution, respectively. These results indicate a decrease in sensitivity (slope) of about 24% in seawater. This should be considered as due to interference from coexistence elements in seawater, which is consistent with the results in Fig. 6. Therefore, for determination of Si in seawater, calibration curves should be determined

7 Determination of Silicate in Seawater 289 Fig. 8. Relationship between applied voltage and lowest limit of detection for Si determination. Table 2. Effects of output power (kw). Output power (kw) R. Intensity C. V. (%) Table 3. Effects of integration time. Integration time (s) C. V. (%) using seawater (standard additional methods should be carried out). Figure 8 shows the relationship between applied voltage and the detection limit of the methodology, defined as twice the standard deviation of the blank signal. Results for voltage ranging from 450 to 850 V are depicted. Only slight variations were observed, indicating that any value of applied voltage in this range is acceptable. Though larger intensity can be obtained with increase in applied voltage, in this study, 700 V was adopted for the reasons noted above. Effects of output power (kw) and

8 290 K. Abe and Y. Watanabe Table 4. Precision of the methodology. Si µm µm µm µ = 3.57 ± µ = 14.9 ± µ = 94.1 ± 1.19 (3.21%) (2.03%) (1.26%) Table 5. Comparison of Si analysis with colorimetric methods µm µm ± (8.23%) This study µm µm ± (6.21%) Colorimetry integration time are shown in Tables 2 and 3. These experiments were carried out using distilled water. Though the coefficient of variation (CV) increased with additional output power, obtained relative intensities (each intensity was normalized by the value at the output power, 1.8 kw) increased. The CV values for various integration times showed smaller fluctuations with longer integration time. In this study, 1.6 kw and 4 s were adopted for output power and integration time, respectively. In Table 4, the precision, values of about 3.2, 2.0, and 1.3% are shown as errors in a single determination at Si levels of 4, 15, and 94 µm, respectively. The analytical detection limit of our methodology, as defined in Fig. 8, was 0.30 µm Si. Table 5 shows a results comparison of Si analysis between the methods used in this study and a previous colorimetric method (Sugawara, 1969). In this experiment, filtered surface seawater from the Kuroshio current was also analyzed, and the results show good correspondence between the two methods; therefore, the applicability of our method for silicate analysis detailed in this study was confirmed. Isshiki et al. (1991) reported form of dissolved silicon in seawater, resulting in the concentrations obtained by the two methods, ICP and colorimetry were the same within experimental error. This also supports the applicability of determination of Si in seawater via ICP method. Figures 9 and 10 show concentrations of silicate as a function of longitude and water depth, respectively. Each numeral corresponds to a station in Fig. 1 (Fig. 9). All samples were filtered

9 Determination of Silicate in Seawater 291 Fig. 9. Silicate concentration as a function of longitude. Fig. 10. Vertical distribution of silicate in the East China Sea. through 0.45 µm membrane filter prior to analysis. In Fig. 9, a decrease in concentration from the western station to the eastern station was observed. The highest value of 19.0 µm at E and the lowest of 1.59 µm were obtained at the eastern station (surface). The vertical profile in Fig. 10 indicates a depletion of silicate in the surface water (1.56 µm) and increase with depth

10 292 K. Abe and Y. Watanabe to a maximum of about 75 µm near the sea floor. Input from the land or the river, horizontal and vertical convection, or biological activity can be considered to be sources that regulate these distributions. The sensitivity and dynamic range of methods of silicate analysis detailed in this report are well-suited for application to water samples from this area. In the East China Sea, 0.5 and 1.3 µm were determined via ICP in surface water from the Okinawa Trough region (Isshiki et al., 1991). Our result shows higher value, 19.0 µm at E, 32.5 N (in surface water). Further quantitative study should be carried out. Acknowledgements The authors are indebted to the officers and crew of R/V Xiang Yang Hong 09 (People s Republic of China) for their help in the sampling. We also wish to thank the staff of Seikai National Fisheries Research Institute, Nagasaki, Japan for their useful comments. References Akagi, T., K. Fuwa and H. Haraguchi (1985): Simultaneous multi-element determination of trace metals in sea water by inductively-coupled plasma atomic emission spectrometry after coprecipitation. Anal. Chim. Acta, 177, Atkins, W. R. G. (1923): The silica content of some natural waters and of culture media. J. Marine Biol. Assoc. U.K., 13, Atkins, W. R. G. (1926): Seasonal changes in the silica content of natural waters in relation to the phytoplankton. J. Marine Biol. Assoc. U.K., 14, Banahan, S. and J. J. Goering (1986): The production of biogenic silica and its accumulation on the southeastern Bering Sea shelf. Continent. Shelf Res., 5, Demaster, D. J., G. B. Knapp and C. A. Nittrouer (1983): Biological uptake and accumulation of silica on the Amazon continental shelf. Geochim. Cosmochim. Acta, 47, Greenfield, S., I. L. Jones and C. T. Berry (1964): High-pressure plasmas as spectroscopic emission sources. Analyst, 89, Isshiki, K., Y. Sohrin and E. Nakayama (1991): Form of dissolved silicon in seawater. Mar. Chem., 32, 1 8. Mullin, J. B. and J. P. Riley (1955): The colorimetric determination of silicate with reference to sea and natural waters. Anal. Chim. Acta, 12, Nelson, D. M. and L. I. Gordon (1982): Production and pelagic dissolution of biogenic silica in the Southern Ocean. Geochim. Cosmochim. Acta, 46, Robinson, R. J. and T. G. Thompson (1948): The determination of silicate in sea water. J. Mar. Res., 7, Sugawara, K. (1969): On the preparation of CSK standards for marine nutrient analysis. SCOR-UNESCO, ICES IOC, Tokyo. Tsunogai, S., M. Kusakabe, H. Iizumi, I. Koike and A. Hattori (1979): Hydrographic features of the deep water of the Bering Sea The Sea of Silica. Deep-Sea Res., 26,

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Application Note Semiconductor Authors Junichi Takahashi Agilent Technologies Tokyo, Japan Abstract Ammonium hydroxide

More information

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive Chapter 9: End-of-Chapter Solutions 1. The following comparison provides general trends, but both atomic absorption spectroscopy (AAS) and atomic absorption spectroscopy (AES) will have analyte-specific

More information

Atomic Emission Spectroscopy

Atomic Emission Spectroscopy Atomic Emission Spectroscopy Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451 Saudi Arabia Building:

More information

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS APPLICATION NOTE ICP - Mass Spectrometry Author Kenneth Ong PerkinElmer, Inc. Singapore Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS Introduction The control of impurity

More information

PRINCIPLE OF ICP- AES

PRINCIPLE OF ICP- AES INTRODUCTION Non- flame atomic emission techniques, which use electrothermal means to atomize and excite the analyte, include inductively coupled plasma and arc spark. It has been 30 years since Inductively

More information

UNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy.

UNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy. UNIVERSITI SAINS MALAYSIA Second Semester Examination Academic Session 2004/2005 March 2005 KAA 502 Atomic Spectroscopy Time: 3 hours Please make sure this paper consists of FIVE typed pages before answering

More information

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS Determination of Total Bromine and Iodine Page 2 of 5 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes an inductively coupled plasma optical emission spectrometry method

More information

a. An emission line as close as possible to the analyte resonance line

a. An emission line as close as possible to the analyte resonance line Practice Problem Set 5 Atomic Emission Spectroscopy 10-1 What is an internal standard and why is it used? An internal standard is a substance added to samples, blank, and standards. The ratio of the signal

More information

Analysis Repeatability of Trace and Major Elements in a Water Sample

Analysis Repeatability of Trace and Major Elements in a Water Sample Analysis Repeatability of Trace and Major Elements in a Water Sample Agnès COSNIER HORIBA Scientific Longjumeau, France Keywords: environment Elements: Al, As, B, Ba, Ca, Cd, Co, Cr, Cu, Fe, Hg, K, Li,

More information

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques Chapter 9 Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques Emission Spectroscopy Flame and Plasma Emission Spectroscopy are based upon those particles that are

More information

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy A. Evaluation of Analytical Parameters in Atomic Absorption Spectroscopy Objective The single feature that contributes most to making atomic

More information

DETERMINATIONS OF THE POLLUTION LEVEL OF THE ENVIRONMENT WITH HEAVY METALS

DETERMINATIONS OF THE POLLUTION LEVEL OF THE ENVIRONMENT WITH HEAVY METALS 7 th INTERNATIONAL MULTIDISCIPLINARY CONFERENCE Baia Mare, Romania, May 17-18, 2007 ISSN-1224-3264 DETERMINATIONS OF THE POLLUTION LEVEL OF THE ENVIRONMENT WITH HEAVY METALS Mariana Dobra, Vasile Viman,

More information

Nutrient Analysis, Nitrate, Nitrite, Silicate, Phosphate and Ammonium

Nutrient Analysis, Nitrate, Nitrite, Silicate, Phosphate and Ammonium Nutrient Analysis, Nitrate, Nitrite, Silicate, Phosphate and Ammonium SUMMARY: The phytoplankton macro nutrients nitrate, nitrite, silicate, phosphate and ammonium are analyzed in seawater using a colorimetric

More information

NATURE AN POLUUTION DEGREE OF WATER ASSETS STUDIED BY INDUCTIVELY COUPLED PLASMA-OPTICAL EMISSION SPECTROSCOPY

NATURE AN POLUUTION DEGREE OF WATER ASSETS STUDIED BY INDUCTIVELY COUPLED PLASMA-OPTICAL EMISSION SPECTROSCOPY NATURE AN POLUUTION DEGREE OF WATER ASSETS STUDIED BY INDUCTIVELY COUPLED PLASMA-OPTICAL EMISSION SPECTROSCOPY I.Chindea, C.Mânzat*, S.Cintă-Pânzaru** Axente-Sever Theoretical HighSchool, Mediaş, Romania

More information

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES Application Note - AN1508 Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES Introduction Page 1 The combination of hydride generation with

More information

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid Application Semiconductor Authors Abe G. Gutiérrez Elemental Scientific 2440 Cuming St Omaha, NE 68131 USA abe@icpms.com

More information

Standard Methods for the Examination of Water and Wastewater

Standard Methods for the Examination of Water and Wastewater 4500-B BORON*#(1) 4500-B A. Introduction 1. Occurrence and Significance Boron (B) is the first element in Group IIIA of the periodic table; it has an atomic number of 5, an atomic weight of 10.81, and

More information

Test Method: CPSC-CH-E

Test Method: CPSC-CH-E UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 10901 DARNESTOWN RD GAITHERSBURG, MD 20878 Test Method: CPSC-CH-E1001-08 Standard Operating Procedure

More information

INVESTIGATION OF ICP-OES ANALYSIS FOR DETERMINATION OF TRACE LEAD IN LEAD-FREE ALLOY

INVESTIGATION OF ICP-OES ANALYSIS FOR DETERMINATION OF TRACE LEAD IN LEAD-FREE ALLOY C1_C0011 1 INVESTIGATION OF ICP-OES ANALYSIS FOR DETERMINATION OF TRACE LEAD IN LEAD-FREE ALLOY Janya Buanuam,* Thareerut Woratanmanon, Temporn Sookawee Regional Failure Analysis and Reliability Laboratory,

More information

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution APEX-ACM Ca Ratios Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution Abstract High resolution ICP-MS is used

More information

Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy

Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy Hiroshi Onodera Application & Research Center, JEOL Ltd. Introduction um, PBB and PBDE) are subject to usage restrictions in Europe.

More information

Spectroscopy Division, and * Central Workshop, Bhabha Atomic Research Centre, Trombay, Mumbai , India

Spectroscopy Division, and * Central Workshop, Bhabha Atomic Research Centre, Trombay, Mumbai , India SaÅdhanaÅ, Vol. 25, Part 1, February 2000, pp. 57±69. # Printed in India Design, fabrication and performance evaluation of a 22-channel direct reading atomic emission spectrometer using inductively coupled

More information

Homework 5: Background Ocean Water Properties & Stratification

Homework 5: Background Ocean Water Properties & Stratification 14 August 2008 MAR 110 HW5: Ocean Properties 1 Homework 5: Background Ocean Water Properties & Stratification The ocean is a heterogeneous mixture of water types - each with its own temperature, salinity,

More information

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors Determination of major, minor and trace elements in rice flour using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Application note Food testing Authors John Cauduro Agilent Technologies,

More information

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP Prodigy ICP Application Note: 1035 U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP Introduction This Application Note describes the capability of the Teledyne Leeman Labs Prodigy High

More information

Atomic Absorption Spectrophotometry. Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon

Atomic Absorption Spectrophotometry. Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon Atomic Absorption Spectrophotometry Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon Defination In analytical chemistry, Atomic absorption spectroscopy is a

More information

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS Application note Semiconductor Author Naoki Sugiyama Agilent Technologies, Tokyo,

More information

Multielement Determination of Rare Earth Elements in Geochemical Samples by Liquid Chromatography/ Inductively Coupled Plasma Mass Spectrometry

Multielement Determination of Rare Earth Elements in Geochemical Samples by Liquid Chromatography/ Inductively Coupled Plasma Mass Spectrometry ANALYTICAL SCIENCES JANUARY 1999, VOL. 15 1999 The Japan Society for Analytical Chemistry 17 Multielement Determination of Rare Earth s in Geochemical Samples by Liquid Chromatography/ Inductively Coupled

More information

SCOPE 35 Scales and Global Change (1988)

SCOPE 35 Scales and Global Change (1988) 1. Types and origins of marine sediments 2. Distribution of sediments: controls and patterns 3. Sedimentary diagenesis: (a) Sedimentary and organic matter burial (b) Aerobic and anaerobic decomposition

More information

Optimizing Analytical Performance in ICP-OES Applications

Optimizing Analytical Performance in ICP-OES Applications application note Optimizing Analytical Performance in ICP-OES Applications Introduction To obtain the best possible performance from an analytical instrument, it is necessary to optimize the operating

More information

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer Application note Agriculture Authors Annie Guerin INRA, Laboratoire d Analyses

More information

Sodium Chloride - Analytical Standard

Sodium Chloride - Analytical Standard Sodium Chloride - Analytical Standard Determination of Total Mercury Former numbering: ECSS/CN 312-1982 & ESPA/CN-E-106-1994 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes

More information

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC) METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC) 1.0 SCOPE AND APPLICATION 1.1 Method 7196 is used to determine the concentration of dissolved hexavalent chromium [Cr(VI)] in EP/TCLP characteristic extracts

More information

Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission

Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission 0 chapter Sodium and Potassium Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission Spectroscopy 67 S. S. Nielsen (ed.), Food Analysis Laboratory Manual Springer

More information

OES - Optical Emission Spectrometer 2000

OES - Optical Emission Spectrometer 2000 OES - Optical Emission Spectrometer 2000 OES-2000 is used to detect the presence of trace metals in an analyte. The analyte sample is introduced into the OES-2000 as an aerosol that is carried into the

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009. Translated English of Chinese Standard: GB5009.17-2014 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD OF GB THE PEOPLE S REPUBLIC OF CHINA National Food Safety Standard-Determination

More information

SILICA, Low Range (0 to 1.60 mg/l)

SILICA, Low Range (0 to 1.60 mg/l) SILICA, Low Range (0 to 1.60 mg/l) Heteropoly Blue Method * Method 8186 For water and seawater 1. Enter the stored program number for low range silica (SiO 2 ). Press: PRGM The display will show: PRGM?

More information

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry APPLICATION NOTE Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry AN43227 Antonella Guzzonato 1, Shona McSheehy Ducos

More information

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS Application note Semiconductor Authors Junichi Takahashi Agilent Technologies, Japan Introduction

More information

Specifically colorimetric recognition of calcium, strontium, barium. ions using 2-mercaptosuccinic acid-functionalized gold nanoparticles

Specifically colorimetric recognition of calcium, strontium, barium. ions using 2-mercaptosuccinic acid-functionalized gold nanoparticles Electronic Supporting Information (ESI) for Specifically colorimetric recognition of calcium, strontium, barium ions using 2-mercaptosuccinic acid-functionalized gold nanoparticles and its use in reliable

More information

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy Atomic line widths: Narrow line widths reduce the possibility of spectral overlap and thus interferences. The band width at half height is used to indicate width. This is also sometimes called the effective

More information

EXPERIMENT 09 OBSERVATION OF SPECTRA

EXPERIMENT 09 OBSERVATION OF SPECTRA EXPERIMENT 09 OBSERVATION OF SPECTRA INTRODUCTION: In physics, as in very other area of study, one of the most valuable questions a student can learn to ask is, How do they know that? Thus, when you read

More information

BATTERY INDUSTRY STANDARD ANALYTICAL METHOD

BATTERY INDUSTRY STANDARD ANALYTICAL METHOD BATTERY INDUSTRY STANDARD ANALYTICAL METHOD For the Determination of Mercury, Cadmium and Lead in Alkaline Manganese Cells Using AAS, ICP-AES and "Cold Vapour" European Portable Battery Association (EPBA)

More information

Sample Preparation of Electronic Device Components for Hexavalent Chromium Analysis by IEC Method :2017

Sample Preparation of Electronic Device Components for Hexavalent Chromium Analysis by IEC Method :2017 Page 1 of 3 Abstract The increasing use of consumer and electronic device components worldwide has drawn increased attention to their impact on the environment. The correct disposal of these materials

More information

INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY

INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY Edited by AKBAR MONTASER George Washington University Washington, D.C. 20052, USA WILEY-VCH New York Chichester Weinheim Brisbane Singapore Toronto CONTENTS

More information

EXPERIMENT 7. Determination of Sodium by Flame Atomic-Emission Spectroscopy

EXPERIMENT 7. Determination of Sodium by Flame Atomic-Emission Spectroscopy EXPERIMENT 7 Determination of Sodium by Flame Atomic-Emission Spectroscopy USE ONLY DEIONIZED WATER (NOT DISTILLED WATER!) THROUGHOUT THE ENTIRE EXPERIMENT Distilled water actually has too much sodium

More information

samples before and after the Fukushima Daiichi Nuclear Power Plant accident

samples before and after the Fukushima Daiichi Nuclear Power Plant accident Supporting Information 135 Cs activity and 135 Cs/ 137 Cs atom ratio in environmental samples before and after the Daiichi Nuclear Power Plant accident Guosheng Yang 1,2, Hirofumi Tazoe 1, Masatoshi Yamada

More information

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Application note Energy and fuels Authors Jenny Nelson, Agilent Technologies, USA Ed McCurdy, Agilent Technologies, UK Introduction

More information

Study of the Production Processes of Marine Biogenic Methane and Carbonyl Sulfide Using Stable Isotope Analysis

Study of the Production Processes of Marine Biogenic Methane and Carbonyl Sulfide Using Stable Isotope Analysis Western Pacific Air-Sea Interaction Study, Eds. M. Uematsu, Y. Yokouchi, Y. W. Watanabe, S. Takeda, and Y. Yamanaka, pp. 117 121. by TERRAPUB 2014. doi:10.5047/w-pass.a02.002 Study of the Production Processes

More information

high temp ( K) Chapter 20: Atomic Spectroscopy

high temp ( K) Chapter 20: Atomic Spectroscopy high temp (2000-6000K) Chapter 20: Atomic Spectroscopy 20-1. An Overview Most compounds Atoms in gas phase high temp (2000-6000K) (AES) (AAS) (AFS) sample Mass-to-charge (ICP-MS) Atomic Absorption experiment

More information

GB/T Translated English of Chinese Standard: GB/T

GB/T Translated English of Chinese Standard: GB/T Translated English of Chinese Standard: GB/T18204.26-2000 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD OF THE GB PEOPLE S REPUBLIC OF CHINA GB/T 18204.26-2000 Methods for determination

More information

SOIL ORGANIC CONTENT USING UV-VIS METHOD

SOIL ORGANIC CONTENT USING UV-VIS METHOD Test Procedure for SOIL ORGANIC CONTENT USING UV-VIS METHOD TxDOT Designation: Tex-148-E Effective Date: March 2016 1. SCOPE 1.1 This method determines the soil organic content based on the amount of humic

More information

LEAD(II) AND COBALT(III) HEPTHYLDITHIOCARBAMATES AS NEW COFLOTATION REAGENTS FOR PRECONCENTRATION OF CADMIUM BEFORE ITS ETAAS DETERMINATION

LEAD(II) AND COBALT(III) HEPTHYLDITHIOCARBAMATES AS NEW COFLOTATION REAGENTS FOR PRECONCENTRATION OF CADMIUM BEFORE ITS ETAAS DETERMINATION LEAD(II) AND COBALT(III) HEPTHYLDITHIOCARBAMATES AS NEW COFLOTATION REAGENTS FOR PRECONCENTRATION OF CADMIUM BEFORE ITS ETAAS DETERMINATION TRAJČE STAFILOV, GORICA PAVLOVSKA AND KATARINA ČUNDEVA Institute

More information

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Application Note Semiconductor Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Tokyo, Japan Introduction

More information

METHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE)

METHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE) METHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE) 1.0 SCOPE AND APPLICATION 1.1 This automated method is applicable to ground water, drinking and surface waters, and domestic and industrial wastes

More information

Determination the elemental composition of soil samples

Determination the elemental composition of soil samples 4. Experiment Determination the elemental composition of soil samples Objectives On this practice you will determine the elemental composition of soil samples by Inductively Coupled Plasma Optical Emission

More information

Ocean and Climate I.

Ocean and Climate I. Ocean and Climate I http://www.gerhardriessbeck.de/ Physical Characteristics of the Ocean Surface area: 3.61 10 14 m 2 Mean depth: 3.7 km Ocean volume: 3.2 10 17 m 3 Mean density: 1.035 10 3 kg/m 3 Ocean

More information

SECTION D.2 AMMONIA NITROGEN

SECTION D.2 AMMONIA NITROGEN SECTION D.2 AMMONIA NITROGEN CEDR Method Code: NH4F L01 a) Scope and Application i) This method describes the determination of low-level ammonia nitrogen concentrations in filtered samples taken from fresh

More information

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY 1.0 SCOPE AND APPLICATION 1.1 This digestion procedure is used for the preparation

More information

Handbook of Inductively Coupled Plasma Spectrometry

Handbook of Inductively Coupled Plasma Spectrometry Handbook of Inductively Coupled Plasma Spectrometry Second Edition MICHAEL THOMPSON, BSc, PhD, ARCS, CChem., FRSC Department of Chemistry Birkbeck College University of London J. NICHOLAS WALSH, BSc, PhD

More information

O Sodium Bicarbonate, NaHCO 3 Ammonium Molybdate Tetrahydrate, (NH 4. O 4 Sodium Chloride, NaCl Ascorbic Acid, C 6 H 8 O 6

O Sodium Bicarbonate, NaHCO 3 Ammonium Molybdate Tetrahydrate, (NH 4. O 4 Sodium Chloride, NaCl Ascorbic Acid, C 6 H 8 O 6 Scope Summary Interferences Performance Specifications This method is used for the determination of silica in estuarine and coastal waters (seawater) according to USEPA Method 366.0. This method can also

More information

Evaluation, Control and Diagnosis of an ICP Through Simple Experiments

Evaluation, Control and Diagnosis of an ICP Through Simple Experiments ICP OPTICAL ATOMIC EMISSION SPECTROSCOPY Evaluation, Control and Diagnosis of an ICP Through Simple Experiments Odile Hirsch HORUIBA Scientific, Longjumeau, France Keywords: general 1 Introduction To undertake

More information

and their Maneuverable Application in Water Treatment

and their Maneuverable Application in Water Treatment Hierarchical Films of Layered Double Hydroxides by Using a Sol-Gel Process and their Maneuverable Application in Water Treatment Yufei Zhao, Shan He, Min Wei,* David G. Evans, Xue Duan State Key Laboratory

More information

ANALYTICAL SCIENCES MAY 2016, VOL The Japan Society for Analytical Chemistry

ANALYTICAL SCIENCES MAY 2016, VOL The Japan Society for Analytical Chemistry ANALYTICAL SCIENCES MAY 2016, VOL. 32 535 2016 The Japan Society for Analytical Chemistry Estimation Using an Enhancement Factor on Non Local Thermodynamic Equilibrium Behavior of High-lying Energy Levels

More information

Question. What caused the recent explosive eruptions of hot ash and gas at Kilauea s Halema uma u crater:

Question. What caused the recent explosive eruptions of hot ash and gas at Kilauea s Halema uma u crater: OCN 201 Deep Sea Sediments Question What caused the recent explosive eruptions of hot ash and gas at Kilauea s Halema uma u crater: A. The interaction of lava with seawater B. Drainage of the lava lake

More information

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode The ultratrace determination of iodine in aqueous samples using the 7700x ICP-MS with oxygen reaction mode Application note Nuclear Authors Kazumi Nakano, Yasuyuki Shikamori, Naoki Sugiyama and Shinichiro

More information

Making Sediments: Biogenic Production, Carbonate Saturation and Sediment Distributions

Making Sediments: Biogenic Production, Carbonate Saturation and Sediment Distributions Making Sediments: Biogenic Production, Carbonate Saturation and Sediment Distributions OCN 623 Chemical Oceanography Reading: Libes, Chapters 15 and 16 Outline I. Deep sea sedimentation Detrital sediments

More information

9/13/10. Each spectral line is characteristic of an individual energy transition

9/13/10. Each spectral line is characteristic of an individual energy transition Sensitive and selective determination of (primarily) metals at low concentrations Each spectral line is characteristic of an individual energy transition 1 Atomic Line Widths Why do atomic spectra have

More information

International Journal of Scientific & Engineering Research, Volume 8, Issue 1, January ISSN

International Journal of Scientific & Engineering Research, Volume 8, Issue 1, January ISSN International Journal of Scientific & Engineering Research, Volume 8, Issue 1, January-2017 131 Flame Photometric Estimation of Sodium and Potassium Ion Present In Water Sample of Darna and Godavari River.

More information

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION Written by Pamela Doolittle, University of Wisconsin- Madison, pssemrad@wisc.edu 2014 In this experiment you will determine the phosphorus (P) concentration

More information

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ Application Note Consumer products Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ Effective removal of tungsten-based interferences on five Hg isotopes using MS/MS Authors

More information

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Application Note Food Authors Jianqiu Mi, Zhengxiang Zhang, Zhixu Zhang,

More information

Novel fluorescent cationic benzothiazole dye response to G-quadruplex aptamer as a novel K + sensor

Novel fluorescent cationic benzothiazole dye response to G-quadruplex aptamer as a novel K + sensor Electronic Supplementary Material (ESI) for Analyst. This journal is The Royal Society of Chemistry 2017 Novel fluorescent cationic benzothiazole dye response to G-quadruplex aptamer as a novel K + sensor

More information

Determination of the inorganic ion composition of standing surface water

Determination of the inorganic ion composition of standing surface water 2. Experiment Determination of the inorganic ion composition of standing surface water Objectives All the biologically important inorganic chemical parameters of standing surface water are called halobity.

More information

Use of ICP-MS in analysing radioisotopes. Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark

Use of ICP-MS in analysing radioisotopes. Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark Use of ICP-MS in analysing radioisotopes Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark Inductively Coupled Plasma Mass Spectrometry (ICP-MS) History ICP-AES

More information

Lab 4 Major Anions In Atmospheric Aerosol Particles

Lab 4 Major Anions In Atmospheric Aerosol Particles Georgia Institute of Technology School of Earth and Atmospheric Sciences EAS 4641 Spring 2008 Lab 4 Major Anions In Atmospheric Aerosol Particles Purpose of Lab 4: This experiment will involve determining

More information

ENVG FALL ICP-MS (Inductively Coupled Plasma Mass Spectrometry) Analytical Techniques

ENVG FALL ICP-MS (Inductively Coupled Plasma Mass Spectrometry) Analytical Techniques ENVG 60500 FALL 2013 ICP-MS (Inductively Coupled Plasma Mass Spectrometry) Analytical Techniques HISTORY In the 1940s, arc and high-voltage spark spectrometry became widely utilized for metal analysis

More information

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma THE PLASMA Inductively Coupled Plasma Mass Spectrometry (ICP-MS) What is a Plasma? - The magnetic field created by a RF (radio frequency) coil produces a current within a stream of Argon (Ar) gas, which

More information

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Tran T. Nham Introduction

More information

Bifurcation Current along the Southwest Coast of the Kii Peninsula

Bifurcation Current along the Southwest Coast of the Kii Peninsula Journal of Oceanography, Vol. 54, pp. 45 to 52. 1998 Bifurcation Current along the Southwest Coast of the Kii Peninsula JUNICHI TAKEUCHI 1, NAOTO HONDA 2, YOSHITAKA MORIKAWA 2, TAKASHI KOIKE 2 and YUTAKA

More information

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Application Note Pharmaceutical Authors Rebeca Santamaria-Fernandez, SheilaMerson,

More information

PROTOCOL FOR P FRACTIONATION

PROTOCOL FOR P FRACTIONATION 1 of 5 PROTOCOL FOR P FRACTIONATION KUZYAKOV LAB 1 20.09.2016 Deejay Maranguit Yakov Kuzyakov Rev. Date Prep. Modify. Check. Approv. 2 of 5 Equipment and Materials 1. 50ml centrifuge tubes with screw caps

More information

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Application note Semiconductor Authors Michiko Yamanaka, Kazuo Yamanaka and Naoki

More information

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Elemental analysis of river sediment using the Agilent 4200 MP-AES Elemental analysis of river sediment using the Agilent 4200 MP-AES Application note Environmental: Soils, sludges & sediments Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction

More information

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation Phosphorus, Total DOC316.53.01121 USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method 8190 0.06 to 3.50 mg/l PO 3 4 (0.02 to 1.10 mg/l P) Test N Tube Vials Scope and application: For water,

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB/T Translated English of Chinese Standard: GB/T

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB/T Translated English of Chinese Standard: GB/T Translated English of Chinese Standard: GB/T6682-2008 www.chinesestandard.net Sales@ChineseStandard.net ICS 71.040.30 G 60 NATIONAL STANDARD OF THE PEOPLE'S REPUBLIC OF CHINA GB Replacing GB/T 6682-1992

More information

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application Semiconductor Author Junichi Takahashi Koichi Yono Agilent Technologies, Inc. 9-1, Takakura-Cho, Hachioji-Shi,

More information

Ocean Sciences 101 The Marine Environment OCEA 101 THE MARINE ENVIRONMENT MID-TERM EXAM

Ocean Sciences 101 The Marine Environment OCEA 101 THE MARINE ENVIRONMENT MID-TERM EXAM OCEA 101 THE MARINE ENVIRONMENT MID-TERM EXAM Part I. Multiple Choice Questions. Choose the one best answer from the list, and write the letter legibly in the blank to the left of the question. 2 points

More information

Physiography Ocean Provinces p. 1 Dimensions p. 1 Physiographic Provinces p. 2 Continental Margin Province p. 2 Deep-Ocean Basin Province p.

Physiography Ocean Provinces p. 1 Dimensions p. 1 Physiographic Provinces p. 2 Continental Margin Province p. 2 Deep-Ocean Basin Province p. Physiography Ocean Provinces p. 1 Dimensions p. 1 Physiographic Provinces p. 2 Continental Margin Province p. 2 Deep-Ocean Basin Province p. 2 Mid-Ocean Ridge Province p. 3 Benthic and Pelagic Provinces

More information

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows Nitrate DOC316.53.01067 Cadmium Reduction Method Method 8192 0.01 to 0.50 mg/l NO 3 N (LR) Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Characteristics of Variations of Water Properties and Density Structure around the Kuroshio in the East China Sea

Characteristics of Variations of Water Properties and Density Structure around the Kuroshio in the East China Sea Journal of Oceanography, Vol. 54, pp. 605 to 617. 1998 Characteristics of Variations of Water Properties and Density Structure around the Kuroshio in the East China Sea EITAROU OKA and MASAKI KAWABE Ocean

More information

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample. METHOD 9200 NITRATE 1.0 SCOPE AND APPLICATION 1.1 This method is applicable to the analysis of ground water, drinking, surface, and saline waters, and domestic and industrial wastes. Modification can be

More information

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode Application note Semiconductor Author Katsuo Mizobuchi Masakazu

More information

ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW

ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW Inductively Coupled Plasma Mass Spectrometry (ICP-MS) What is a Plasma? - The magnetic field created by a RF (radio frequency) coil produces

More information

Formation of self-inhibiting copper(ii) nanoparticles in an autocatalytic Fenton-like reaction

Formation of self-inhibiting copper(ii) nanoparticles in an autocatalytic Fenton-like reaction Formation of self-inhibiting copper(ii) nanoparticles in an autocatalytic Fenton-like reaction Kazim R Naqvi, Jennifer Marsh, Victor Chechik,* Department of Chemistry, University of York, YO1 5DD, UK The

More information

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer is powerful simultaneous full

More information

ab83360 Ammonia Assay Kit

ab83360 Ammonia Assay Kit Version 8 Last updated 18 December 2017 ab83360 Ammonia Assay Kit For the measurement of total ammonia and ammonium levels in various samples This product is for research use only and is not intended for

More information

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials Nitrate, HR DOC316.53.01068 Chromotropic Acid Method Method 10020 0.2 to 30.0 mg/l NO 3 N (HR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific information

More information

The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES

The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Authors Christine M. Rivera Doug Shrader

More information