Merja Gergov, Timo Nenonen, Ilkka Ojanpera and Raimo A. Ketola*

Size: px
Start display at page:

Download "Merja Gergov, Timo Nenonen, Ilkka Ojanpera and Raimo A. Ketola*"

Transcription

1 Journal of Analytical Toxicology 2015;39: doi: /jat/bkv020 Advance Access publication March 5, 2015 Article Compensation of Matrix Effects in a Standard Addition Method for Metformin in Postmortem Blood Using Liquid Chromatography Electrospray Tandem Mass Spectrometry Merja Gergov, Timo Nenonen, Ilkka Ojanpera and Raimo A. Ketola* Department of Forensic Medicine, Faculty of Medicine, University of Helsinki, PO Box 40, FI University of Helsinki, Finland *Author to whom correspondence should be addressed. raimo.ketola@helsinki.fi This work describes a procedure to evaluate matrix effects in a combined dilution and standard addition method (SAM) using liquid chromatography electrospray tandem mass spectrometry. The method was validated and applied to an analysis of metformin in postmortem blood samples. The analytical method included protein precipitation with methanol, followed by liquid chromatographic separation of metformin on Gemini NX-C 18 reversed-phase column using a gradient consisting of methanol and ammonium acetate at ph 3.2. The mass spectrometric analysis was performed with a quadrupole-linear ion trap mass spectrometer equipped with a turbo ion spray interface in a positive ion mode using selected reaction monitoring. Quantitation was performed based on an SAM. Validation for metformin revealed a practical limit of quantification of 0.1 mg/l, a linear range from 0.1 to 3.0 mg/l, average precision 10%, accuracy (bias) 9% and reproducibility 10%. Combined matrix effects were evaluated by k-values (slopes) of calibration plots, postextraction addition approach and a comparison of within- and between-sample precision (relative standard deviation). It was demonstrated that the method contained matrix effects which were fully compensated for using dilution and the SAM. Introduction Proper calibration is essential in quantitative analysis in order to obtain reliable results from complex biological samples. Common calibration methods include external standard, internal standard (ISTD) and matrix-matched calibration methods. The ISTD calibration is commonly used in bioanalysis; however, even ISTDs cannot perfectly correct the variability of sample matrix; therefore, the use of matrix-matched standards is preferable. In a matrix-matched method, the standards are prepared in a similar matrix as the sample itself, such as urine, plasma or blood. If the sample is putrefied, like in some postmortem cases, even the use of matrix-matched standards is not perfect. In putrefied samples, the degree and amount of degradation of cells can vary a lot, as well as the amount and quality of lipids. The fourth choice of calibration method, especially suitable for hydrophilic/polar drugs in putrefied samples, is a standard addition method (SAM), where the actual samples are used to create a calibration plot individually (1). The main advantage of this method is that it can correct the matrix effect because exactly the same sample matrix is present both in calibration standards and the sample itself. A drawback of the method is, for example, that the amount of sample needed is much higher than in other methods because the sample is needed also for calibration standards. Another drawback is that it is more tedious to prepare the calibration plot for each sample separately. However, good results These authors contributed equally. have been obtained with the SAM for biological samples. For example, phosphatidylcholines with varying carbon chain lengths in a rat liver tissue were analyzed by nanoflow LC MS-MS (2). With the SAM, the effect of both sample matrix and the carbon chain length could be effectively compensated. A challenging sample matrix for organic analysis is placenta. The research group of Zafra-Go mez used the SAM for the determination of parabens and benzophenones from placental tissues using LC MS-MS and atmospheric pressure chemical ionization (APCI) technique (3, 4). The methods showed good variability as the inter- and intra-day variabilities were under 13.8 and 5% for parabens and benzophenones, respectively. Kuepper et al. employed an SAM with ion-pairing solid-phase extraction and subsequent HPLC MS-MS analysis for detection of succinylmonocholine from various biological postmortem samples, such as the brain, liver, kidney and vitreous humor (5). The method worked well not only for obtaining reliable quantitative results from the samples but also for confirming possible negative results for some samples. Mitamura et al. also applied an LC MS-MS with the SAM for quantitative determination of pregnenolone-3-sulfate in rat brains (6). This study shows another benefit of the SAM, namely it is possible to detect lower concentrations of analytes than actual instrumental limits of quantitation (LOQ), if the method is precise enough. For example, in their study, the instrumental LOQ was 4 ng/ml; but by the standard addition, it was possible to accurately quantitate concentrations at 0.5 ng/ml (the difference between the actual and the added concentrations). In a comparison of different calibration approaches using multivariate curve resolution for LC MS analysis of biocides from environmental samples, it was noticed that the SAM with ISTDs gave the most precise results for the analytes (7). Peters et al. have presented a comprehensive evaluation of validation procedures and criteria, which could also be used as a guideline in validation of postmortem analytical methods (8). Besides calibration, another critical issue in LC MS analysis is the matrix effect as already stated earlier. Different aspects of matrix effects in LC MS analysis for forensic and clinical toxicology were recently comprehensively reviewed by Peters and Remane (9). The review highlights the importance of studying matrix effects in forensic toxicology during method development. It was concluded that postmortem samples usually possess more matrix effects than clinical samples, and it is practically impossible to remove matrix effects completely in bioanalysis using LC MS. Several procedures are available to measure the degree of matrix effects, such as post-column standard infusion and postextraction addition (PEA) methods (10, 11), of which the latter is now commonly used by researchers as it gives a measure of efficiency of the whole analytical process. However, matrix effects have been very rarely studied in SAMs. Basilicata et al. applied two calibration procedures, the ISTD and # The Author Published by Oxford University Press. All rights reserved. For Permissions, please journals.permissions@oup.com

2 the SAMs, for the determination of urinary benzene of workers in occupational exposure studies (12). The conclusion of the study was that the SAM effectively compensated the matrix effects of individual urine samples, and a three-point standard addition protocol was enough for obtaining reliable quantitative results. The relationship between matrix concentration and suppression of electrospray ionization was investigated by Stahnke et al. using pesticides present in QuEChERS extracts as an example (13). They noticed that dilution of extracts by a factor of reduces ion suppression to,20% if the initial suppression is 80%. Five different methods (matrix-matched calibration, standard addition, post-column standard infusion, extrapolative dilution and post-column flow splitting) were compared in accounting for or overcoming matrix effects in LC ESI MS determination of pesticides from food samples (14). Extrapolative dilution is a new form of dilution method where the sample is first diluted prior to sample treatment and analysis to reduce matrix effect, and the final result is obtained by extrapolating to infinite dilution. In this study, extrapolative dilution and SAMs were found to give results that were not statistically different from the correct values. The main aim of the study was to investigate the efficiency of the method in compensating matrix effects encountered in postmortem blood samples and hence to improve the accuracy of the method. A validation procedure of the method is described here, for the determination of metformin in postmortem blood samples using protein precipitation and dilution combined with standard addition and LC ESI MS-MS. Experimental Reagents Metformin was purchased from Weifa AS (Kragero, Norway). Methanol Fluka LC MS Chromasolv w (99.9%) was purchased from Sigma Chemical (St. Louis, MO, USA) and the other solvents from Merck (Darmstadt, Germany). Ammonium acetate, formic acid and sodium phosphate were supplied by Merck. All solvents and reagents were of HPLC or analytical grade. Laboratory water was purified with a Milli-Q Integral 5 system (Millipore, Bedford, MA, USA). Bovine whole blood (blank blood) and authentic whole blood samples from routine medico-legal autopsies served for method development. The time interval between death and autopsy was normally from 1 to 6 days, thus the degree of putrefaction varied from sample to sample. Sample preparation for validation Whole blood sample (0.5 ml) was initially diluted with laboratory water using a dilution factor of 10. The diluted sample was pipetted into five test tubes (0.5 ml each), four of which were fortified with metformin at concentrations of 0.1, 0.3, 1.0 and 3.0 mg/l for calibration. Samples (0.5 ml) were precipitated with 1 ml of methanol by vortexing for 1 min and centrifuged at 4,330g for 5 min. Finally, 100 ml of the aqueous layer was transferred to an autosampler vial and diluted with 900 ml of laboratory water by vortexing. For validation, blank blood samples were first fortified with metformin to produce metformin concentration of 1, 10 or 100 mg/l prior to dilution. Samples with high concentrations of metformin (10 and 100 mg/l) were diluted before standard additions using higher dilution factors than 10, namely 50 or 500. Thus, in the final extract for the LC MS-MS analysis, the original samples were diluted by an absolute factor of ,000. Instrumental analysis LC separations were carried out with a Shimadzu Prominence HPLC system consisting of two LC-20AD pumps, SIL-20AHT autosampler, DGU-20A5 vacuum degasser and CTO-20A column oven (Shimadzu USA Manufacturing, Inc., Canby, OR, USA). A Phenomenex Gemini NX-C 18 column ( mm; 3 mm) and Phenomenex C 18 guard column (4 2 mm) were used in chromatographic separations (Phenomenex, Torrance, CA, USA). The mass spectrometric analysis was performed using an AB Sciex 3200 Q TRAP w LC MS-MS System instrument equipped with a Turbo V TM source and a TurboIonSpray w probe (AB Sciex, Concord, ON, Canada) in triple quadrupole mode. The software was Analyst The analytical column was stabilized at 408C. The mobilephase gradient consisted of 10 mm ammonium acetate buffer (ph 3.2) and methanol (both containing 0.1% of formic acid) as follows: the methanol proportion was held at 10% during the equilibrium time of 7 min. After injection, the methanol proportion was increased to 30% over 4 min, then to 95% in 3 min, kept at 95% for 1 min to clean the column and finally decreased back to 10% in 2 min. The total flow rate through the column was 0.25 ml/min. The injection volume was 10 ml. The total flow from the LC was directed to the ion source without splitting. Needle voltage was 5.0 kv, declustering potential 30 V and ion spray heater temperature 4508C. The curtain gas (nitrogen) was set at 12 and collision gas (nitrogen) pressure in the collision cell at 5 in the Analyst control software. The nebulizer gas (nitrogen) was set at 60 psi (414 kpa) and turbo heater gas (nitrogen) at 70 psi (482 kpa). A selected reaction monitoring (SRM) mode was used for metformin with transitions m/ z 130! 71 and 130! 85, and the corresponding values for collision energy were 30 and 20 ev. A dwell time of 80 ms was used for the transitions. Quantitation method An SAM was applied for quantification. Metformin was fortified to the samples at known concentrations, and signal intensities of the non-fortified and fortified samples were measured. A calibration plot of the peak areas vs. standard additions of the spikes yielded a straight line with linear regression (Figure 1f). The concentration of metformin was calculated from the point (x-axis) at which the extrapolated linear regression line crosses the concentration axis at zero signal. Results Method validation Validation included assessment of selectivity, linearity, lower and higher limits of quantification, precision, accuracy, reproducibility and combined matrix effects, and it was based on the protocol proposed by Peters et al. (8). Summary of the validation results is presented in Table I. 360 Gergov et al.

3 Figure 1. A negative autopsy blood sample diluted with laboratory water (a dilution factor of 10) with standard additions of (A) 0, (B) 0.1, (C) 0.3, (D) 1.0 and (E) 3.0 mg/l (SRM chromatograms of a transition of m/z 130! 71 for metformin). (F) Calibration plots prepared from the negative sample and a positive sample containing 1.9 mg/l metformin (a dilution factor of 10), and the structure of metformin. Table I Validation Results for Metformin in Postmortem Blood Using a Dilution and SAM with Protein Precipitation and LC ESI MS-MS Parameter Value Accuracy (bias%) Precision (%) Linear range (mg/l) Lower limit of quantification (mg/l) a 0.1 Reproducibility (bias%) a A practical lower LOQ. Selectivity/specificity Selectivity was assured by using a triple quadrupole system in the MS-MS mode. Selectivity was evaluated by analyzing 48 autopsy blood samples known to be negative for metformin but after routine drug screening found to contain other drugs. These other drugs did not give any signal interfering with the SRM transitions of metformin. Possible interfering signals were monitored. Selectivity was proven by the absence of any interfering signals at the monitored SRM traces of the studied 48 authentic autopsy samples (Figure 1A). Linearity Blank blood samples were first fortified at 0 (no fortification), 1, 10 and 100 mg/l concentrations of metformin. Five-point calibration plots were prepared in duplicates at each of the concentrations, using a standard addition set of 0, 0.1, 0.3, 1.0 and 3.0 mg/l of metformin. To evaluate linearity also in authentic sample matrix, calibration plots were prepared in 48 laboratory s routine autopsy samples, known to be negative for metformin. For determination of the linear range, the absolute peak area was plotted against the drug concentration. The following criteria were applied for linear range: linear regression with a correlation coefficient (r 2 )of.0.995, bias from the calibration plot of,20% for all individual calibration points and precision as relative standard deviation (RSD),20%. The linear range both in blank blood and in the authentic blood samples was from 0.1 to 3 mg/ L. The average correlation coefficient (r 2 ) was in blank blood and in the autopsy samples. Higher metformin concentrations, 10 and 100 mg/l, resulted in calibration plots with quadratic regression, and therefore, samples with higher concentration than 3 mg/l must be diluted with laboratory water using a higher dilution factor than 10 to reach the linear range from 0.1 to 3 mg/l. Compensation of Matrix Effects by Standard Addition Method 361

4 Accuracy, precision, recovery and lower and upper limits of quantification (LLOQ and ULOQ) Precision, accuracy, recovery and LLOQ were evaluated using blank blood and six authentic negative autopsy blood samples fortified at metformin concentrations of 1, 10 and 100 mg/l. Five-point calibration plots were prepared using standard addition set of 0, 0.1, 0.3, 1.0 and 3.0 mg/l. Two replicate calibration plots were prepared for all three fortified concentrations. Precision was calculated as within-sample intraday RSD of the replicates, and it was averagely 8, 11 and 8% at metformin concentrations of 1, 10 and 100 mg/l, respectively. Accuracy was expressed as average bias between fortified and measured concentrations, and it was 11, 6 and 8% at metformin concentrations of 1, 10 and 100 mg/l, respectively. In this case, the bias served also as an indicator of the recovery; therefore, the recovery was 89, 94 and 92% at 1, 10 and 100 mg/l, respectively. LLOQ was established as the lowest added concentration fulfilling the following criteria: bias from the calibration plot,20%, precision (RSD),20%, signal-to-noise ratio at least 10 and a symmetrical peak shape. The lowest standard addition was 0.1 mg/l, which served as a practical LLOQ. At LLOQ, bias from the calibration plot varied between 2 and 10% and precision 8 and 15% in the autopsy samples. Signal-to-noise ratio at 0.1 mg/l was larger than 45 in all authentic samples. ULOQ (3 mg/l) was defined as the upper limit of the linear range when the initial dilution factor of 10 was used. Examples of calibrations plots, prepared in autopsy blood samples, are presented in Figure 1. Intra-laboratory reproducibility Metformin was considered stable in blood samples and on a freeze thaw cycle. Therefore, reproducibility was evaluated from the re-analysis of 10 autopsy blood samples after a few days (six samples) to several months (four samples) interval of time from the first analysis. Metformin concentrations in the original analyses varied from 1.3 to 52 mg/l. Reproducibility was calculated as bias of the new analytical result from the original one, and on average, it was 9.6%, ranging from 0.5 to 17.3%. Combined variation from matrix In matrix effect tests, three strategies were applied: (i) variation in slopes (k-values) of calibration plots (10), (ii) PEA approach (9) and (iii) a comparison of within- and between-sample RSD (15). Results of the matrix effect tests are summarized in Table II. Table II Summary of Matrix Effect Results Using (a) Variation in the k-values and r 2 of the Calibration Plots, (b) Post-Extraction Addition and (c) Comparison of Within- and Between-Sample RSDs Calibration plots a Post-extraction addition b Comparison of RSD (%) c k-value RSD (%) 42 ME% RE% PE% Wi/Bw (area) Wi/Bw (conc.) r 2 RSD (%) mg/l /7 8/10 10 mg/l /8 11/6 100 mg/l nd nd nd 4/8 8/11 nd, not determined. a RSD of the k-values and correlation coefficient r 2 of the calibration plots in 48 autopsy samples. b Matrix effect (ME), recovery (RE) and process efficiency (PE) in six autopsy samples. c Within (Wi) and between (Bw) sample RSD of metformin peak areas and concentrations in six autopsy samples. Variation in k-values of the calibration plots The matrix effects were estimated from the k-values of calibration plots in 48 authentic negative autopsy blood samples using a dilution factor of 10 prior to standard addition. Straight calibration plots were obtained by using linear regression. Variability of the slopes served as an indicator of matrix effect. Precision of the slopes should not exceed 3 4% for a method to be considered free of matrix effect. In the case of metformin, comparison of the plots revealed intensive matrix effects, while variability of the k-value was 42% (Figure 2). These samples were analyzed within a time period of 6 months; thus, variation in the performance of the LC-ESI MS-MS instrument (shifts in the signal intensities) was included in the variation in the k-values. However, a comparison of correlation coefficients (r 2 ) revealed very good linearity with an average value of and an RSD of 0.43% (Figure 2). PEA approach The second strategy was to estimate matrix effect (ME), recovery (RE) and overall process efficiency (PE) by preparing three sets of calibration plots: pure metformin fortified in laboratory water (Set 1), authentic autopsy blood samples fortified after sample preparation (Set 2) and the same samples fortified before the sample preparation (Set 3). ME (%), RE (%) and PE (%) were calculated by comparing the corresponding absolute peak areas of metformin as a ratio of Set 2 to Set 1, Set 3 to Set 2 and Set 3 to Set 1, respectively. Six samples from the k-value evaluation were fortified with 1 and 10 mg/l. As metformin concentrations in authentic postmortem blood samples can be much.3 mg/l, all samplesweredilutedpriortostandardadditionandthereby the effect of dilution was studied. Two dilution factors for different analyte concentrations were tested: 10 for 1.0 mg/l and 50 for 10 mg/l. Calculations showed that the ME on absolute peak areas depended strongly on the dilution factor. At 1 mg/l concentration (dilution factor of 10), suppression was detected and ME was 70%, whereas at 10 mg/l concentration (dilution factor of 50), there was no matrix effect (ME 100%). The corresponding RE and PE percentages were 93 and 65% at 1 mg/l of metformin and 79 and 78% at 10 mg/l, respectively. Comparison of within- and between-sample precision For the third approach, six authentic autopsy blood samples were fortified in two replicates with metformin at concentrations of 1, 10 and 100 mg/l. The samples were known to be negative for metformin but after routine drug screening found to contain other drugs. The within-sample precision was calculated as the average RSD of the replicates, whereas the betweensample RSD was calculated from the averaged values of the six samples. Calculations were performed both from the measured concentrations and absolute peak areas. In comparison of withinand between-sample RSD, a significantly higher between-sample RSD was tentatively regarded as an indication of combined effects of matrix and variation in extraction efficiency. Calculation from concentrations revealed a within-sample RSD of 8 11% and a between-sample RSD of 6 11%, whereas the corresponding values from absolute peak areas were and %, respectively. Statistical evaluation was done using one-way analysis of variance (ANOVA) with 0.05 as the critical P-value. Calculations revealed that differences in the within- and between-sample RSD 362 Gergov et al.

5 Figure 2. Calibration plots prepared for 48 authentic negative autopsy blood samples using a dilution factor of 10, standard addition, protein precipitation and LC ESI MS-MS. A transition of m/z 130! 71 was monitored. values were significant in the absolute peak areas, but not in the measured concentrations. Moreover, the differences in RSDs were systematically more significant at higher metformin concentrations and standard additions at which the peak areas were greater. This was considered as an indication of matrix effects in absolute peak areas, which was demonstrated also by k-value and PEA approaches. However, matrix had no significant effect on the measured concentrations, as can be concluded also from the obtained accuracies, which in this test were 11, 6 and 8%, expressed as bias from the fortified metformin concentrations of 1, 10 and 100 mg/l, respectively. Discussion Metformin is a very polar compound (Figure 1), and therefore, a fast, easy and rugged protein precipitation method was applied. Unfortunately, protein precipitation is reported to suffer from serious matrix effects (16). In the present study, protein precipitation was combined with dilution and an SAM for quantification. The accuracy and the matrix effects of this combined analytical method were evaluated in three different ways: comparison of the k-values (slopes) of the calibration plots, the PEA approach, and within- and between-sample precision ( peaks areas and concentrations). All approaches revealed matrix effects to a similar extent, 30 40%. The PEA approach showed a PE of 65 78%, but the accuracy could not be deduced from this experiment. Similarly, the comparison of k-values of the calibration plots did not provide straightforward information on accuracy of the method. The comparison of within- and between-sample precision showed clear matrix effects on the analyte peak areas, but not on final analyte concentrations. This proved that the effect of matrix on quantitative results could be compensated using the SAM, thereby giving good accuracy. This is because each sample has an individual calibration plot, and exactly the same matrix is present both in the sample and in all of the calibration points. This result is in accordance with results obtained by Pere -Trepat et al. (7) andbasilicataet al. (12). In another method for the analysis of metformin from postmortem blood, the matrix effects could be eliminated using matrix-matched calibrants with isotopically labeled metformin as an ISTD, protein precipitation and cation-exchange solid-phase extraction followed by hydrophilic interaction LC ESI MS-MS (17). The remaining ion suppression was,5%. However, there was no information on the decomposition status of the postmortem samples and the variability of matrix effects between individual samples. Recovery of the method presented was evaluated in two ways: fortifying in a traditional way and using the PEA method. Both approaches gave similar recoveries, ranging from 76 to 93%, thus proving that both approaches could be used for reliable determination of recovery. There was no direct correlation observed between the degree of putrefaction and validation results, thus indicating that the dilution and SAM method was suitable also for decomposed blood samples. With the present method, the upper limit of linearity range of metformin concentration was 3 mg/l. Samples with a higher metformin concentration resulted in quadratic calibration plots in which accuracy of the lowest addition (0.1 mg/l) did not fulfill the acceptance criteria (bias from calibration plot 20%). Therefore, blood samples were diluted using a dilution factor of Dilution factor tests, as well as earlier experience from the laboratory s other SAMs, show that determination of analyte concentration is more accurate when higher dilution factors are used. This is apparently because the matrix compounds are also diluted, thus reducing the matrix effects. When the dilution factor was large enough compared with the original concentration of the analyte, the average correlation coefficient r 2 was better than with excellent precision. In routine analysis, where the analyte concentration in a sample is unknown, it is impossible to predict the correct level of dilution factor beforehand. Therefore, critical evaluation of the correlation coefficient r 2 and the accuracy of the lowest additions in the calibration plot is essential. The PEA approach is efficient in determination of ME, RE and PE; however, it leads to a large number of measurements, and Compensation of Matrix Effects by Standard Addition Method 363

6 therefore, it is time-consuming and laborious. In postmortem studies, the matrix is much more variable than in clinical samples; therefore, we consider at least 20 authentic samples with two replicates to be a minimum number of measurements to represent a postmortem matrix. When authentic postmortem samples are used for validation, the sample amount available may also become a restricting factor for the number of experiments. With the SAM method, even a restricted assessment of the matrix effect (one concentration level, a five-point calibration plot, 2 replicates of 20 authentic samples) would include 410 measurements by PEA, but only 200 measurements by comparison of within- and between-sample precision. Thereby, the comparison of the within- and between-sample precision apparently is an easier approach, revealing the most important parameters of validation, namely the combined matrix effect, accuracy and precision. In the case of metformin, observation of significant matrix effects on absolute peak areas, but not on the measured concentrations, clearly demonstrated the benefits of SAM. In general, if dilution is needed, like in this case, it will increase the number of experiments, especially if different dilution factors will be evaluated. Based on the results obtained in this study and by Stahnke et al. (13); however, we suggest that evaluation of only two different dilution factors is enough as long as the difference of these factors is at least 10-fold. In our method, several levels of additions were used, due to a large range of metformin concentrations in authentic samples and still a limited linear measurement range with the instrument. In the analysis of compounds with a broad measurement range, it could be possible to use only one addition level to obtain reliable quantitative results. Conclusions Protein precipitation method is a very quick and easy method, but it results in serious matrix effects for postmortem blood samples. A comparison of different approaches for matrix effect assessment showed that all three approaches studied revealed matrix effects in the analytical method, and PEA approach could also estimate the extent of matrix effects. The comparison of within- and between-sample precision not only revealed the presence of matrix effects in absolute peak areas but also proved that these effects were efficiently compensated by dilution and SAM producing good accuracy and reliable quantitative results. The work load in the comparison of within- and between-sample precision approach was also less than in the PEA approach. References 1. Massart, D.L., Vandeginste, B.G.M., Buydens, L.M.C., De Jong, S., Lewi, P.J., Smeyers-Verbeke, J. (1997) Standard addition method. In Vandeginste, B.G.M., Rutan, S.C. (eds), Handbook of Chemometrics and Qualimetrics, Part A, Elsevier, Amsterdam, The Netherlands, p Ahn, E.J., Kim, H., Chung, B.C., Moon, M.H. (2007) Quantitative analysis of phosphatidylcholine in rat liver tissue by nanoflow liquid chromatography/tandem mass spectrometry. Journal of Separation Science, 30, Jime nez-dı az, I., Vela-Soria, F., Zafra-Gómez, A., Navalo n, A., Ballesteros, O., Navea, N. et al. (2011) A new liquid chromatography tandem mass spectrometry method for determination of parabens in human placental tissue samples. Talanta, 84, Vela-Soria, F., Jiménez-Díaz, I., Rodríguez-Gómez, R., Zafra-Go mez, A., Ballesteros, O., Navalo n, A. et al. (2011) Determination of benzophenones in human placental tissue samples by liquid chromatography tandem mass spectrometry. Talanta, 85, Kuepper,U.,Musshoff,F.,Madea,B.(2011)Applicabilityofsuccinylmonocholine as a marker for succinylcholine administration comparative analysis of samples from a fatal succinylcholine-intoxication versus postmortem control specimens. Forensic Science International, 207, Mitamura, K., Yatera, M., Shimada, K. (1999) Quantitative determination of pregnenolone-3-sulfate in rat brains using liquid chromatography/electrospray ionization-mass spectrometry. Analytical Sciences, 15, Peré-Trepat, E., Lacorte, S., Tauler, R. (2007) Alternative calibration approaches for LC MS quantitative determination of coeluted compounds in complex environmental mixtures using multivariate curve resolution. Analytica Chimica Acta, 595, Peters, F.T., Drummer, O.H., Musshoff, F. (2007) Validation of new methods. Forensic Science International, 165, Peters, F.T., Remane, D. (2012) Aspects of matrix effects in applications of liquid chromatography mass spectrometry to forensic and clinical toxicology a review. Analytical and Bioanalytical Chemistry, 403, Matuszewski, B.K., Constanzer, M.L., Chavez-Eng, C.M. (2003) Strategies for the assessment of matrix effect in quantitative bioanalytical methods based on HPLC-MS/MS. Analytical Chemistry, 75, Matuszewski, B.K. (2006) Standard line slopes as a measure of a relative matrix effect in quantitative HPLC MS bioanalysis. Journal of Chromatography B, 830, Basilicata, P., Miraglia, N., Pieri, M., Acampora, A., Soleo, L., Sannolo, N. (2005) Application of the standard addition approach for the quantification of urinary benzene. Journal of Chromatography B, 818, Stahnke, H., Kittlaus, S., Kempe, G., Alder, L. (2012) Reduction of matrix effects in liquid chromatography electrospray ionization mass spectrometry by dilution of the sample extracts: how much dilution is needed? Analytical Chemistry, 84, Kruve, A., Leito, I. (2013) Comparison of different methods aiming to account for/overcome matrix effects in LC/ESI/MS on the example of pesticide analyses. Analytical Methods, 5, Gergov,M.,Nokua,P.,Vuori,E.,Ojanpera,I.(2009)Simultaneous screening and quantification of 25 opioid drugs in post-mortem blood and urine by liquid chromatography tandem mass spectrometry. Forensic Science International, 186, Chambers, E., Wagrowski-Diehl, D.M., Lu, Z., Mazzeo, J.R. (2007) Systematic and comprehensive strategy for reducing matrix effects in LC/MS/MS analyses. Journal of Chromatography B, 852, Sørensen, L.K. (2012) Determination of metformin and other biguanides in forensic whole blood samples by hydrophilic interaction liquid chromatography electrospray tandem mass spectrometry. Biomedical Chromatography, 26, Gergov et al.

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB. Quantitative analysis of small molecules in biological samples 100 Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB % 0 300 500 700 900 1100 1300 1500 1700 m/z Class Overview Introduction

More information

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Application Note Clinical Research Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Authors Yanan Yang 1, Victor Mandragon 2, and Peter Stone 1 1

More information

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS Application Note EPA Method 535 EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS API 3200 LC/MS/MS System Overview Described here is the

More information

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS Christopher Borton AB SCIEX Golden, Colorado verview Described here is the analysis of

More information

Appendix II- Bioanalytical Method Development and Validation

Appendix II- Bioanalytical Method Development and Validation A2. Bioanalytical method development 1. Optimization of chromatographic conditions Method development and optimization of chromatographic parameters is of utmost important for validating a method in biological

More information

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer 38 Current Trends in Mass Spectrometry November 6 Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer This article demonstrates the improved

More information

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Application Note Food Authors Jianqiu Mi, Zhengxiang Zhang, Zhixu Zhang,

More information

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 129 Infi nity LC, and 646 Triple Quadrupole LC/MS Application Note Clinical Research Authors Linda Côté and Christophe

More information

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note Authors Siji Joseph Agilent Technologies India Pvt. Ltd.

More information

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma LC-MS/MS Method for the Determination of Diclofenac in Human Plasma J. Jones, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20569 Key Words SPE, SOLA, Accucore RP-MS, diclofenac, Core

More information

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System Application Note Authors Mohamed Youssef and Vaughn P. Miller Agilent

More information

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Application Note Food Testing and Agriculture Author Chen-Hao (Andy) Zhai and Rong-jie Fu Agilent Technologies (Shanghai)

More information

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 643 LC/MS/MS Application Note Forensic Toxicology Authors Jason Hudson, Ph.D., James Hutchings, Ph.D., and Rebecca Wagner,

More information

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis Application Note AN89.V.1 Extraction of Methylmalonic Acid from Serum Using ISOLUTE SAX Page 1 Extraction of Methylmalonic Acid from Serum Using ISOLUTE SAX Prior to LC-MS/MS Analysis Sample Preparation

More information

Determination of Pesticide Residues in Oats by Automated. QuEChERS and LC/QQQ. Application Note. Abstract. Introduction

Determination of Pesticide Residues in Oats by Automated. QuEChERS and LC/QQQ. Application Note. Abstract. Introduction Determination of Pesticide Residues in Oats by Automated QuEChERS and LC/QQQ Application Note Abstract The QuEChERS (Quick-Easy-Cheap-Effective-Rugged-Safe) sample extraction method was developed for the

More information

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories APPLICATION NOTE 65200 Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories Authors Neloni Wijeratne, Claudia Martins, Mary

More information

Quantitative analysis of mitragynine in human urine by high performance liquid chromatography-tandem mass spectrometry

Quantitative analysis of mitragynine in human urine by high performance liquid chromatography-tandem mass spectrometry Quantitative analysis of mitragynine in human urine by high performance liquid chromatography-tandem mass spectrometry Shijun Lua, Buu N. Trana, Jamie L. Nelsenb, Kenneth M. Aldousa. Journal of Chromatography

More information

LC-MS/MS Analysis of Phytocannabinoids and their

LC-MS/MS Analysis of Phytocannabinoids and their POSTER NOTE 64922 LC-MS/MS Analysis of Phytocannabinoids and their LC-MS/MS Analysis of Phytocannabinoid Metabolites in Urine, Oral Fluid and Blood Rory M Doyle, Sherry Gregory*, Thermo Fisher Scientific,

More information

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma Jon Bardsley 1, James Goldberg 2 1 Thermo Fisher Scientific, Runcorn, UK; 2 Thermo Fisher Scientific, West Palm Beach,

More information

Improved Automated Sample Preparation for the Analysis of 25-OH-Vitamin D3 by LC/MS/MS

Improved Automated Sample Preparation for the Analysis of 25-OH-Vitamin D3 by LC/MS/MS Improved Sample Preparation for the Analysis of 25-OH-Vitamin D3 by LC/MS/MS Abstract In this work, we demonstrate an improved automated sample preparation for liquid chromatography-tandem mass spectrometry

More information

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS 1.0 Purpose - This procedure specifies the required elements for the extraction and quantitation of THC,

More information

Technical Note. Introduction

Technical Note. Introduction Technical Note Analysis and Characterization of Psilocybin and Psilocin Using Liquid Chromatography - Electrospray Ionization Mass Spectrometry (LC-ESI-MS) with Collision-Induced-Dissociation (CID) and

More information

Supporting Information. Sweetened Swimming Pools and Hot Tubs

Supporting Information. Sweetened Swimming Pools and Hot Tubs Supporting Information Sweetened Swimming Pools and Hot Tubs Lindsay K Jmaiff Blackstock, Wei Wang, Sai Vemula, Benjamin T Jaeger and Xing-Fang Li * Division of Analytical and Environmental Toxicology,

More information

Definitive EtG/EtS LC-MS/MS Analysis:

Definitive EtG/EtS LC-MS/MS Analysis: Definitive / LC- Analysis: A Rugged 4-Min Method for High-Throughput Labs By Justin Steimling and Frances Carroll Abstract Methods for monitoring alcohol consumption biomarkers and are generally limited

More information

Translational Biomarker Core

Translational Biomarker Core Translational Biomarker Core Instrumentation Thermo Scientific TSQ Quantum Triple Quadrupole Mass Spectrometers. There are two TSQ Quantum Ultra AM instruments available in the TBC. The TSQ Quantum Ultra

More information

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching s by Removal of Isobaric Interferences Differential Mobility Used as a Tool to Address Selectivity Challenges Adrian M. Taylor

More information

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Chemistry Instrumental Analysis Lecture 37. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 37 Most analytes separated by HPLC are thermally stable and non-volatile (liquids) (unlike in GC) so not ionized easily by EI or CI techniques. MS must be at

More information

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS UCT Part Numbers: CSFASETG203 - CLEAN SCREEN FASt ETG, 200mg / 3mL tube SLETG100ID21-3UM - Selectra ETG HPLC column, 100 x 2.1

More information

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18 Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18 Column Xin Zhang, Jonathan P. Danaceau, and Erin E. Chambers Waters Corporation, Milford, MA, USA APPLICATION BENEFITS Semi-validated

More information

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Organics Revision Date: July 19, 2017 Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Parameter Perfluorinated Alkyl Acids (Perfluorobutane Sulphonate (PFBS), Perflourooctane Sulphonate (PFOS),

More information

Uncontrolled Copy. Table of Contents. 1. Principle of Assay Specimens Reagents and Materials... 3

Uncontrolled Copy. Table of Contents. 1. Principle of Assay Specimens Reagents and Materials... 3 Table of Contents 1. Principle of Assay... 3 2. Specimens... 3 3. Reagents and Materials... 3 4. Standards, Controls, and Solutions... 4 5. Equipment and Special Supplies... 5 6. Instrumentation and Parameters...

More information

A High Sensitivity Dual Solid Phase Extraction LC-MS/MS Assay for the Determination of the Therapeutic Peptide Desmopressin in Human Plasma

A High Sensitivity Dual Solid Phase Extraction LC-MS/MS Assay for the Determination of the Therapeutic Peptide Desmopressin in Human Plasma White Paper A High Sensitivity Dual Solid Phase Extraction LC-MS/MS Assay for the Determination of the Therapeutic Peptide Desmopressin in Human Plasma Lars Neudert, MSc, Senior Scientist Method Development

More information

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies Jonathan Danaceau, Erin Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA, USA

More information

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539) Determination of ormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell P Column (EPA 539) Application Note Environmental Authors Rong-jie Fu and Chen-ao (Andy) Zhai Agilent Technologies

More information

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS Application Note: 56 Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS Yang Shi, Catherine Lafontaine, Matthew Berube, John Fink, François Espourteille Thermo

More information

Supporting Information

Supporting Information 1 Supporting Information 2 3 Discovery and implications of C 2 and C 3 perfluoroalkyl sulfonates in aqueous film forming foams (AFFF) and groundwater 4 Krista A. Barzen-Hanson a and Jennifer A. Field b*

More information

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish. 1 Supporting Information 2 3 Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish. 4 5 John Moreman, Okhyun

More information

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research Heather A Brown, 1 Claudia Rossi, 2 and Lisa J Calton 1 1 Waters Corporation, Wilmslow, UK 2

More information

Simultaneous analysis for forensic drugs in human blood and urine using ultra-high speed LC-MS/MS

Simultaneous analysis for forensic drugs in human blood and urine using ultra-high speed LC-MS/MS Simultaneous analysis for forensic drugs in human blood and urine using ultra-high speed LC-MS/MS TIAFT 204 Toshikazu Minohata, Keiko Kudo 2, Kiyotaka Usui 3, Noriaki Shima 4, Munehiro Katagi 4, Hitoshi

More information

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes Joe DiBussolo, Marta Kozak Thermo Fisher Scientific, San Jose, CA Application

More information

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS UCT Part Numbers: CSFASETG203 - CLEAN SCREEN FASt ETG, 200mg / 3mL tube SLETG100ID21-3UM - Selectra ETG HPLC column, 100 x 2.1

More information

Tecan AC Extraction Plate

Tecan AC Extraction Plate Tecan AC Extraction Plate Simple sample preparation for the determination of vitamin D in serum by LC-MSMS Introduction A growing number of laboratories have moved from the use of immunoassay methods to

More information

High-sensitivity, high-throughput quantitation of catecholamine metabolites in urine by LC/MS/MS for clinical research

High-sensitivity, high-throughput quantitation of catecholamine metabolites in urine by LC/MS/MS for clinical research PO-CON1655E High-sensitivity, high-throughput quantitation of catecholamine metabolites in urine by LC/MS/MS for clinical research ASMS 2016 MP-079 Atsuhiko TOYAMA 1, Kumie SATOU 2, Yuki NAKAMURA 2, Ichiro

More information

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application.

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application. Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source Application Toxicology Authors Jürgen Wendt Agilent Technologies Sales and Support

More information

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 66 Triple Quadrupole LC/MS System Application Note Environmental Authors Don Noot, Matthew Noestheden, and Ralph Hindle Vogon

More information

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS Application Note Drug Discovery Authors Srividya Kailasam Agilent

More information

Validation of analytical methods. Adrian Covaci Toxicological Center, University of Antwerp

Validation of analytical methods. Adrian Covaci Toxicological Center, University of Antwerp Validation of analytical methods Adrian Covaci Toxicological Center, University of Antwerp Why validate? The objective of validating an analytical method is to demonstrate its suitability for its intended

More information

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Application Note Doping Control Authors Prof. Maria Kristina Parr Freie Universität Berlin Institute

More information

Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS/MS Method

Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS/MS Method Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS/MS Method By Sharon Lupo and Frances Carroll Abstract The analysis of synthetic cannabinoids and their metabolites

More information

Determination of trace anions in concentrated hydrofluoric acid

Determination of trace anions in concentrated hydrofluoric acid APPLICATION NOTE 78 Determination of trace anions in concentrated hydrofluoric acid Authors Archava Siriraks Thermo Fisher Scientific, Sunnyvale, CA Keywords HF, ICS-5000 +, IonPac AS10, IonPac AC10, ion

More information

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Application Note Authors Kevin McCann, Sameer Nene, Doug McIntyre, Edmond Neo, Dennis Nagtalon, and

More information

Tyler Trent, Applications Sales Specialist, Teledyne Tekmar P a g e 1

Tyler Trent, Applications Sales Specialist, Teledyne Tekmar P a g e 1 Application Note AutoMate-Q40 Automated QuEChERS Extraction for Pesticide Residues in Botanicals Tyler Trent, Applications Sales Specialist, Teledyne Tekmar P a g e 1 Abstract QuEChERS is a Quick-Easy-Cheap-

More information

Ultrafast Analysis of Methadone and Metabolite EDDP in Urine by the Agilent RapidFire High-Throughput Mass Spectrometry System

Ultrafast Analysis of Methadone and Metabolite EDDP in Urine by the Agilent RapidFire High-Throughput Mass Spectrometry System Ultrafast Analysis of and Metabolite in Urine by the Agilent RapidFire High-Throughput Mass Spectrometry System Application Note Forensic Toxicology Authors Mohamed Youssef and Vaughn P. Miller Agilent

More information

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies APPLICATION BENEFITS Regulatory guidelines and recommendations place a greater emphasis on the detection

More information

Detection of Phencyclidine in Human Oral Fluid Using Solid Phase Extraction and Liquid Chromatography with Tandem Mass Spectrometric Detection

Detection of Phencyclidine in Human Oral Fluid Using Solid Phase Extraction and Liquid Chromatography with Tandem Mass Spectrometric Detection Detection of Phencyclidine in Human Oral Fluid Using Solid Phase Extraction and Liquid Chromatography with Tandem Mass Spectrometric Detection Application Note Forensic Toxicology Authors Christine Moore,

More information

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 126 Infinity Binary LC system with UV detection Application Note Food Author Srividya Kailasam Agilent Technologies,

More information

SAMHSA-Compliant LC/MS/MS Analysis of Phencyclidine in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120

SAMHSA-Compliant LC/MS/MS Analysis of Phencyclidine in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120 SAMHSA-Compliant LC/MS/MS Analysis of Phencyclidine in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120 Application Note Forensic Toxicology Authors Irina ioumaeva, John M. Hughes Agilent

More information

Low-level Determination of 4-Hydrazino Benzoic Acid in Drug Substance by High Performance Liquid Chromatography/Mass Spectrometry

Low-level Determination of 4-Hydrazino Benzoic Acid in Drug Substance by High Performance Liquid Chromatography/Mass Spectrometry ISSN: 0973-4945; CODEN ECJHAO E- Chemistry http://www.e-journals.net 2010, 7(2), 403-408 Low-level Determination of 4-Hydrazino Benzoic Acid in Drug Substance by High Performance Liquid Chromatography/Mass

More information

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS Kate Comstock, Tim Stratton, Hongxia (Jessica) Wang, and Yingying Huang

More information

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer APPLICATION NOTE 72385 Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer Authors Stephan Meding, Katherine Lovejoy, Martin Ruehl Thermo Fisher Scientific,

More information

Extraction of Aflatoxins and Ochratoxin from Dried Chili Using ISOLUTE. Myco Prior to LC-MS/MS Analysis

Extraction of Aflatoxins and Ochratoxin from Dried Chili Using ISOLUTE. Myco Prior to LC-MS/MS Analysis Application Note AN785 Extraction of Aflatoxins and chratoxin From Dried Chili Using ISLUTE Myco prior to LC-MS/MS Analysis Page Extraction of Aflatoxins and chratoxin from Dried Chili Using ISLUTE Myco

More information

Recovery/bias evaluation

Recovery/bias evaluation Recovery/bias evaluation Ivo Leito University of Tartu ivo.leito@ut.ee 1 Overview What is bias? What are the main issues in bias evaluation? Time frame Random and systematic effects Approaches for bias

More information

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method Xiang He and Marta Kozak ThermoFisher Scientific, San Jose, CA,

More information

isobaric Barbiturates in Serum using LDTD- MS/MS combined with differential mobility

isobaric Barbiturates in Serum using LDTD- MS/MS combined with differential mobility Ultra-Fast Separation Ultra-Fast and Quantification Separation and of isobaric Barbiturates in Serum using LDTD- Quantification of isobaric MS/MS combined with differential mobility spectrometry Barbiturates

More information

Uncontrolled Copy. Table of Contents. 1. Introduction and Principle of Assay Specimens Reagents and Materials...

Uncontrolled Copy. Table of Contents. 1. Introduction and Principle of Assay Specimens Reagents and Materials... Table of Contents 1. Introduction and Principle of Assay... 3 2. Specimens... 3 3. Reagents and Materials... 3 4. Standards, Controls, and Solutions... 4 5. Equipment and Special Supplies... 5 6. Instrumentation

More information

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION Quantitation by High Resolution Mass Spectrometry: Using Target Enhancement and Tof-MRM to Achieve Femtogram-level On-column Sensitivity for Quantitation of Drugs in Human Plasma Yun W. Alelyunas, Mark

More information

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS Application Note Forensic Toxicology Authors Martin Josefsson, and Markus Roman National Board of Forensic Medicine Linköping,

More information

This document is a preview generated by EVS

This document is a preview generated by EVS TECHNICAL REPORT RAPPORT TECHNIQUE TECHNISCHER BERICHT CEN/TR 15641 August 2007 ICS 67.050 English Version Food analysis - Determination of pesticide residues by LC- MS/MS - Tandem mass spectrometric parameters

More information

Abstract. Experimental Sample Preparation

Abstract. Experimental Sample Preparation Determination of Benzoylecgonine in Human Urine Using Automated, In-Line, ITSP Solid Phase Extraction and Liquid Chromatography Mass Spectrometric Detection Thurman L. Allsup 1, Kenneth C. Lewis 1 and

More information

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use Shih-Tse Jason Lai, Jia Wang, Brad Hart, Kristine Van Natta, Marta Kozak, Jorge Valdivia, Haibo Wang,

More information

The Theory of HPLC. Quantitative and Qualitative HPLC

The Theory of HPLC. Quantitative and Qualitative HPLC The Theory of HPLC Quantitative and Qualitative HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

Validation Report 18

Validation Report 18 EURL for Cereals and Feeding stuff National Food Institute Technical University of Denmark Validation Report 18 Determination of pesticide residues in maize for livestock feed by GC-MS/MS and LC-MS/MS

More information

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring High-Throughput Protein Quantitation Using Multiple Reaction Monitoring Application Note Authors Ning Tang, Christine Miller, Joe Roark, Norton Kitagawa and Keith Waddell Agilent Technologies, Inc. Santa

More information

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for Clinical Toxicology Page of 0 Materials and Methods Biomass Component Extraction: The uneaten cooked plant specimen was prepared for chemical analysis as follows. The sample was frozen, diced, pulverized

More information

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors Trace-Level Analysis of Melamine in Milk Products on Agilent 789A/5975C GC/MSD Using a ew Agilent J&W DB-5ms Ultra Inert Column and SampliQ SCX Cartridges Application ote Gas Chromatography/Mass Spectrometry/Food

More information

Sreekanth Kakarla et al. / Journal of Pharmacy Research 2015,9(12), Available online through

Sreekanth Kakarla et al. / Journal of Pharmacy Research 2015,9(12), Available online through Research Article ISSN: 0974-6943 Available online through www.jprsolutions.info Sensitive liquid chromatography tandem mass spectrometry method for the quantification of Ropinirole in human plasma using

More information

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3)

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3) Schedule Assignment Due Date Draft Section of Lab Report Monday 6pm (Jan 27) Quiz for Lab 2 Peer Review of Draft Complete Lab Report 1 Tuesday 9:30am Wednesday 6pm Friday 6pm Summary of Paper 2 Monday

More information

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada) Quantitation and Identification of Pharmaceuticals and Personal Care Products (PPCP) in Environmental Samples using Advanced TripleTOF MS/MS Technology André Schreiber AB SCIEX Concord, Ontario (Canada)

More information

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION Andrew J. Aubin and Tanya L. Jenkins Waters Corporation, Milford, MA, USA INTRODUCTION Benzocaine (4-Aminobenzoic

More information

Figure 1. The supported liquid extraction process using the ISOLUTE SLE plate (single well shown)

Figure 1. The supported liquid extraction process using the ISOLUTE SLE plate (single well shown) Supported Liquid Extraction: Automate those Tiresome Bioanalytical Protocols L. Williams, H. Lodder, S. Merriman, A. Howells, S. Jordan, J. Labadie, M. Cleeve, C. Desbrow, R. Calverley and M. Burke 1 Argonaut

More information

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India Analysis of fumonisin, FB1 and FB2 mycotoxins in corn food and feed samples using the Agilent 1120 Compact LC System coupled to the Agilent 6140 Single Quadrupole LC/MS System Application Note Food Safety

More information

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma Ken Meadows, Thermo Fisher Scientific, Runcorn, UK Application Note 20789 Key Words

More information

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System Application Update 55 Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free System Introduction Hydrogen peroxide is an essential chemical in the fabrication

More information

Anaylsis of Pesticide Residues in Rice Using Agilent Bond Elut QuEChERS AOAC Kit by LC-MS/MS Detection

Anaylsis of Pesticide Residues in Rice Using Agilent Bond Elut QuEChERS AOAC Kit by LC-MS/MS Detection Anaylsis of Pesticide Residues in Rice Using Agilent Bond Elut QuEChERS AOAC Kit by LC-MS/MS Detection Application Note Food Safety Author Andy Zhai Agilent technologies Co., Ltd. 412 Yinglun Road Shanghai,

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

A Comprehensive, Fast, and Sensitive Method for the Quantitation of Synthetic Cannabinoids using the Elute UHPLC Coupled to an EVOQ LC-TQ MS

A Comprehensive, Fast, and Sensitive Method for the Quantitation of Synthetic Cannabinoids using the Elute UHPLC Coupled to an EVOQ LC-TQ MS A Comprehensive, Fast, and Sensitive Method for the Quantitation of Synthetic Cannabinoids using the Elute UHPLC Coupled to an EVOQ LC-TQ MS Rapid and Sensitive Quantitation of Synthetic Cannabinoids in

More information

Preparative Separation of Active Components from Natural Products using Low-pressure Gradient Preparative HPLC

Preparative Separation of Active Components from Natural Products using Low-pressure Gradient Preparative HPLC PO-CON1405E Preparative Separation of Active Components from Natural Products using Low-pressure Gradient Preparative HPLC Pittcon 014 50-7P Kenichiro Tanaka 1, William Hedgepeth 1, Lincoln Grimes, Tsutomu

More information

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity Janet Hammond Waters Corporation, Wilmslow, UK APPLICATION BENEFITS High

More information

Development of analytical methods for Catecholamines, Metanephrines and related metabolites in urine and plasma matrices

Development of analytical methods for Catecholamines, Metanephrines and related metabolites in urine and plasma matrices Development of analytical methods for Catecholamines, Metanephrines and related metabolites in urine and plasma matrices From sample preparation to LC/MS/MS Linda Côté Christophe Deckers Agilent Technologies

More information

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Adulteration in Dietary Supplements and Botanicals Application Note Food Testing & Agriculture Authors James Neal-Kababick

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Applied Biosystems QStar Pulsar i Features of the API QSTAR Pulsar i The

More information

Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX)

Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX) Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX) Application Note Forensic Toxicology Authors Bellah. Pule,

More information

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis A. Caruso, M. Santoro, P. Magni, S. Pelagatti, and R. Facchetti Thermo Fisher Scientific, Milan,

More information

Uncontrolled Copy. Table of Contents. 1. Principle of Assay Specimens Reagents and Materials... 4

Uncontrolled Copy. Table of Contents. 1. Principle of Assay Specimens Reagents and Materials... 4 Table of Contents 1. Principle of Assay... 3 2. Specimens... 3 3. Reagents and Materials... 4 4. Standards, Controls, and Solutions... 4 5. Equipment and Special Supplies... 4 6. Instrumentation and Parameters...

More information

Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis

Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis Application Note AN772 Introduction This application note describes the extraction of cocaine and a full range

More information

Validation and Standardization of (Bio)Analytical Methods

Validation and Standardization of (Bio)Analytical Methods Validation and Standardization of (Bio)Analytical Methods Prof.dr.eng. Gabriel-Lucian RADU 21 March, Bucharest Why method validation is important? The purpose of analytical measurement is to get consistent,

More information

Presentation Basic Introduction to Instrumentation Matrix Effects Challenges

Presentation Basic Introduction to Instrumentation Matrix Effects Challenges Keys to Implementing Mass Spectrometry in the Clinical Laboratory Paul J. Taylor Chromatography Mass Spectrometry Satellite Meeting, AACB 20th September, 2013 Dept of Clinical Pharmacology, Princess Alexandra

More information

Available online Journal of Chemical and Pharmaceutical Research, 2012, 4(6): Research Article

Available online  Journal of Chemical and Pharmaceutical Research, 2012, 4(6): Research Article Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research, 2012, 4(6):3275-3279 Research Article ISSN : 0975-7384 CODEN(USA) : JCPRC5 Development and validation of a matrix solid-phase

More information

Impact factor: 3.958/ICV: 4.10 ISSN:

Impact factor: 3.958/ICV: 4.10 ISSN: Impact factor: 3.958/ICV: 4.10 ISSN: 0976-7908 99 Pharma Science Monitor 9(4), Oct-Dec 2018 PHARMA SCIENCE MONITOR AN INTERNATIONAL JOURNAL OF PHARMACEUTICAL SCIENCES Journal home page: http://www.pharmasm.com

More information

4.1 MATERIALS 4.2 EQUIPMENTS

4.1 MATERIALS 4.2 EQUIPMENTS 4.1 MATERIALS 1. Aceclofenac Moraceae Pharma Ltd., Uttarakhand, India 2. Diclofenac Arti Drugs Ltd., Mumbai, India 3. Paracetamol Moraceae Pharma Ltd., Uttarakhand, India 4. Para-aminophenol Hema Pharmaceuticals

More information