Standard Operating Procedure PCM SOP 1-1. Microscope Calibration, Preparation and Analysis of Air Samples by Phase Contrast Microscopy (PCM)

Size: px
Start display at page:

Download "Standard Operating Procedure PCM SOP 1-1. Microscope Calibration, Preparation and Analysis of Air Samples by Phase Contrast Microscopy (PCM)"

Transcription

1 Standard Operating Procedure PCM SOP 1-1 Microscope Calibration, Preparation and Analysis of Air Samples by Phase Contrast Microscopy (PCM) Scope and Purpose: The scope and purpose of this standard operating procedure is to document the procedures involved in the calibration, preparation and analysis of air samples by Phase Contrast Microscopy (PCM) according to the current NIOSH 7400 Method (Analysis of Asbestos and Other Fibers by PCM), Procedures For Laboratory and Field Activities: Alignment Procedures 1. Turn the power switch on and adjust the voltage control dial until the proper light intensity is reached (should be set between the levels of 7 and 10). 2. Turn the revolving nose piece and rotate the objectives until the 40X objective is in place. 3. Place the microscope slide that contains an amount of background loading (a reference slide is recommended) on top of the mechanical stage. While viewing the slide through the eyepiece, focus on the slide using the fine and coarse adjustment knobs. 4. Once focused, close the field iris diaphragm. Turn the stage condenser slowly up or down until a polygonal ring (either blue to magenta in color) appears as shown below: If the image appears to be uncentered from the field of view, center with the centering screws. 5. Open the diaphragm so the polygonal ring disappears from the field of view. Page 1 of 6 Revised 2015

2 6. Remove the right ocular lens and insert telescopic ocular into its place. View into the telescopic ocular and focus. Two rings will be seen; one dark and one light as seen below: Adjust rings with the phase centering screws until the rings are centered one above the other as seen below: Remove the telescopic ocular and replace the right ocular. Record calibration information in the microscope logbook. 7. Test the resolution of the microscope and analysts using the HSE/NPL Phase Contrast Test Slide (see SOP 1-2 for details and timetable for recalibration of the test slide) on each day client/field samples are analyzed or if microscope was moved or serviced. 8. The Walton-Beckette Graticule must be measured once for the graticule using a stage micrometer (see SOP 1-3 for details and timetable for recalibration of the micrometer slide) on each day the client/field samples are analyzed. The area of the graticule must be documented in the microscope logbook. 9. Clean the microscope with canned/aerosol air. The microscope is now ready for use. 10. Any problems or concerns with the QuickFix must be immediately reported to the Laboratory Director or the QC Coordinator and the sample preparation is stopped until the problem is resolved. The issues with the QuickFix shall be recorded on the IEA Lab Reagent Blank Result Log in the Comments section. 11. Rotometers are calibrated according to the steps spelled out in the NIOSH 582e manual, located in the PCM Laboratory, before each summer session begins. The rotometers are kept in the PCM laboratory and checked out during use. Rotometer calibration is noted in Equipment Section of the laboratory equipment spreadsheet and on each rotometer. Sample Reception: 1. When samples are received for analysis by laboratory personnel, the chain-of-custody form shall be signed and dated. Page 2 of 6 Revised 2015

3 2. The sample cassette(s) and chain-of-custody form(s) are inspected to determine if they pass rejection criteria (see SOP 1-4 for details). If the samples fail, the appropriate person(s) is notified that the samples cannot be prepared and analyzed, the reason for the rejection is noted on the chain of custody form, and the samples will not be analyzed until corrective actions are taken. If samples pass rejection criteria, the analyst is able to proceed to the prepare samples. 3. The Chain of Custody and samples are entered into the IEA Laboratory Information Management System (LIMS), where the uniques samples are stored and a log of samples can be produced. At this point the general information is entered into the LIMS, such as client, building where samples were taken, and what each samples unique sample ID number is. Sample Preparation: 1. For Field Analysis, the environmental conditions must be taken into account before the samples can be prepared and analysis is done, some of these include: good ventilation for the use of the QuickFix acetone vaporizer, clean preparation area, a safe environment, and an area that can be secured. It is encumbent upon the field analyst to ensure and maintain a clean area to prep and analyze samples, if at any time this cannot be maintained, a new area must be found for the sample preparation and analysis. 2. Samples are laid out in numerical order on the prep table. QuickFix acetone vaporizer is turned on and allowed to warm up (vaporizer is ready for use when light turns on). 3. Analyslides are set out in front of the sample cassettes and labeled with the project number, sample number, analyst s initials and analysis date. 4. Grasp the air cassette firmly at its base on the prep table. Depending on the type of cassettes, either the white label is cut with a disposable scalpel at the top of the cassette or the white sticker label is peeled back. With a twisting motion, remove the top of the air cassette. Remove the filter from the cassette with a pair of clean forceps and place in the analyslides according to the corresponding project number and sample number. 5. Any sample filters found torn, wet, overloaded with debris or if the filters show signs of tampering, these filters are placed in their analyslide, and the comments noted on the chainof-custody form. These samples will not be prepared past this point and the project manager will be informed about why the samples could not be analyzed. 6. After samples have been placed in their respective analyslide, frosted microscope slides are labeled with project number and sample number. Slides must be free of dust before use. 7. Remove the cover from the first sample analyslide and cut a pie-shaped portion of the filter ( 1 / 6 to 1 / 4 of the total area of the sample filter). The filters are cut in a rocking motion using a clean surgical scalpel to avoid tearing. Place the filter section on the center of a labeled frosted microscope slide (sample side up) with the tip of the wedge facing the frosted section of the slide. 8. The sample slide shall be placed onto the slide holder of the QuickFix acetone vaporizer (sample filter should be under the injector port of the vaporizer). Inject approximately 250µl of acetone into the injector port of the vaporizer where it will vaporize onto the slide (repeat Page 3 of 6 Revised 2015

4 process if necessary for slide to become clear). Allow slide to dry (usually 3-5 seconds), and remove slide from the slide holder of the vaporizer. Place one drop of Triacetin on the dried filter and cover immediately with a clean glass coverslip. Coverslip should be placed onto the slide in a hinge-like manner to minimize air bubbles. 9. Using a permanent marker, outline the area of the filter on the back side of the microscope slide. Place slides on the microscope slide tray and allow to sit for at least ten minutes before being analyzed. 10. Repeat steps 6 through 8 until all samples have been prepped. 11. After all samples have been prepped, place all analyslides with the remaining filters in a labeled, plastic bag with a zip-lock top and archive. These samples are to be kept out of direct light and extreme temperatures to minimize sample breakdown. Sample Analysis: 1. Type the information into the Laboratory Information Management System (LIMS). The following information must be included and typed into the LIMS (If the electronic LIMS is not available, hand write the information into the PCM Field Worksheet Form and return form to laboratory for entry): - Project number, project name, brief project description and sampling date on top of form - Sample number for the project - Sample description (I=inside, O=outside, H=HEPA, C=clearance, P=personal, AR=adjacent response) - Comments/Location section - Sample time, flow rate and correction factor of the rotameter - Volume of sample(s) - Maximum sample fields to be read - Fiber density - Reporting Limit 2. Each analyst analyzes one (1) reference slide before any client samples are analyzed. The reference result is recorded in the reference entry form and on the bottom of the PCM Worksheet Form, if used. 3. Prepared sample slides are analyzed according to the A Counting Rules as described in the NIOSH 7400 Method (see the current NIOSH 7400 method for details). Sample raw counts are placed on PCM Worksheet Form, if used, and into the LIMS system under Raw Counts section. 4. Calculations for fibers per cubic centimeter (f/cc), 95% confidence interval and reporting limit for each sample are computed by the LIMS. 5. When a field blank s fiber density is less then the reporting limit, less than 7 f/mm 2 (<7 f/mm 2 ) is written instead of calculated number. Page 4 of 6 Revised 2015

5 6. Analysts records the date of analysis in the LIMS and signs their name on the Chain of Custody. Copies of results are given to the proper personnel, and originals are placed in the appropriate areas of the laboratory for quality control analysis. 7. The slides are placed in a slide storage box and the slide box and the associated paperwork are placed in the quality control area of the laboratory. 8. PCM samples are read daily according to a first come first serve basis. If several sets of samples are received at the same time and have the same turn-around time for results, the sets are to be read in the following order of importance: 1. Clearance samples 2. Duration samples 3. Background samples 4. Personal samples Quality Assurance Procedures: 1. Each analyst must perform replicate analysis on at least ten percent (10%) of their total number of slides analyzed on any particular day. A different analyst may perform duplicate analysis on at least five percent (5%) of the total number of samples analyzed by any one analyst. Replicate and duplicate results are entered into their appropriate section of the Laboratory Information Management System (LIMS) 2. Field Blank Samples must be prepped in the same manner as field samples and the counts must be entered/written into the PCM Worksheet. Field blank samples must come from the same box of cassette s that the field sample came from. 3. Laboratory blanks are prepared once per day when client samples are analyzed. A clean, labeled microscope slide is placed into the acetone vaporizer with no air filter. Acetone is vaporized on the slide and dried. It is then removed from the vaporizer and a drop of Triacetin is placed onto the slide along with a clean glass coverslip. Approximately 100 fields are analyzed as a check for contamination of sample reagents. Contamination of the reagents or consumables must be defined as excess debris if the mean is greater than or equal to 5 fibers per 100 graticule fields. If contamination is discovered, reagents and consumables should be replaced under laboratory management s supervision. Results are recorded in the appropriate logbook. 4. Air samples are collected and analyzed annually in the PCM Laboratory or in the sample prep area or more frequently if conditions of the laboratory change. Once analyzed, any sample results with a result of 0.02 f/cc or less is considered acceptable, and no corrective action needs to be taken. If the result is higher than the 0.02 f/cc value, the source of the contamination must be determined and the situation rectified. Sample result information is recorded in the appropriate logbook. Page 5 of 6 Revised 2015

6 Author: Timothy Froehlig Revision Number: 11 Date Created: 8/29/95 Approved by: Date: References: NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, 7400 Method, Asbestos and Other Fibers by PCM, 15 August NIOSH 582 Equivalent Course Manual Institute for Environmental Assessment. Olympus CHS/CHT Biological Microscope Instructional Manual. Page 6 of 6 Revised 2015

Laboratory ID. Laboratory Name. Address. Phone. Lab Manager/Supervisor. QA Officer. Analyst(s) Auditor. Date(s) of Audit

Laboratory ID. Laboratory Name. Address. Phone. Lab Manager/Supervisor. QA Officer. Analyst(s) Auditor. Date(s) of Audit Method(s): Laboratory ID Laboratory Name Address Phone Lab Manager/Supervisor QA Officer Analyst(s) Auditor Date(s) of Audit NEW JERSEY DEPARTMENT OF ENVIRONMENTAL PROTECTION OFFICE OF QUALITY ASSURANCE

More information

Introduction to the Microscope

Introduction to the Microscope Title: Microscope Mania "Micro" (Greek!) refers to tiny, "scope" refers to view or look. Microscopes are tools used to enlarge images of small objects so they can be studied. The compound light microscope

More information

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP.

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP. NEW JERSEY DEPARTMENT OF ENVIRONMENTAL PROTECTION OFFICE OF QUALITY ASSURANCE ENVIRONMENTAL LABORATORY CERTIFICATION PROGRAM ON-SITE LABORATORY EVALUATION RADIOCHEMISTRY PROCEDURES Gross Alpha-Gross Beta

More information

Use of light microscope and stereomicroscope: measuring microscopic

Use of light microscope and stereomicroscope: measuring microscopic Experiment 1 Use of light microscope and stereomicroscope: measuring microscopic objects 1.1 Introduction The microscope is a major tool used by biologists, which was invented about 350 years ago. It is

More information

Laboratory Techniques 100: Back To Basics. Carol Injasoulian Lab Manager City of Bay City April 29,2015

Laboratory Techniques 100: Back To Basics. Carol Injasoulian Lab Manager City of Bay City April 29,2015 Laboratory Techniques 100: Back To Basics Carol Injasoulian Lab Manager City of Bay City April 29,2015 QA/QC for Environmental Measurement What is quality assurance (QA)? Broad plan to maintain quality

More information

Facilities Management

Facilities Management Policy Number: 700.20 Title: Chemical Fume Hood Policy Implementation Date: 2002 Last Audited: August, 2017 Last Revised: October 23rd, 2017 Facilities Management Introduction The laboratory chemical fume

More information

Percent Solids Determination

Percent Solids Determination Title: Percent Solids Page 1 of 10 Percent Solids Determination References: Method 2540G, Standard Methods For the Examination of Water and Wastewater, APHA 18 th edition, 1992. 1. Scope and Application

More information

ALLOWAY METHOD OUTLINE

ALLOWAY METHOD OUTLINE ALLOWAY METHOD OUTLINE Standard Laboratory Method SM4500-Cl -G Parameter Residual Chlorine & Free Chlorine Method DPD Colorimetric Test Kit Date Issued Originator: Section Supervisor: QA Manager Date:

More information

Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and

Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared by: OEWRI Laboratory Manager

More information

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method: ALLOWAY METHOD OUTLINE Standard Laboratory Method: SM Parameter: Methylene Blue Method: Colorimetric Reporting Level: Reference: 0.05 mg/l Standard Methods for the Examination Of Water and Wastewater;

More information

Montgomery County Community College Document Number: MET DeKalb Pike Revision Number: 0

Montgomery County Community College Document Number: MET DeKalb Pike Revision Number: 0 Page 1 of 5 Approvals: Preparer: Jason McMillan Reviewer: Dr. Maggie Bryans Date: 08JAN14 Date: 10JAN14 1. Purpose: Operation of Eppendorf Research Plus through to the P-1000. 2. Scope: Applies to the

More information

Working Copy. P /ru /,6. 8-tt tt. EffectiveDate: I ^, Operation and Calibration of the Tri-Carb Liquid Sclntillation Counter QAM-RI.106.

Working Copy. P /ru /,6. 8-tt tt. EffectiveDate: I ^, Operation and Calibration of the Tri-Carb Liquid Sclntillation Counter QAM-RI.106. QAM-RI.106 Operation and Calibration of the Tri-Carb Liquid Sclntillation Counter Revision: 0 Laboratory Man ager/lqao/rso Date Date EffectiveDate: I ^, P /ru /,6 8-tt tt Renewal Date: lnitials: Texas

More information

Technical Procedure for Glass Refractive Index Measurement System 3 (GRIM 3)

Technical Procedure for Glass Refractive Index Measurement System 3 (GRIM 3) Technical Procedure for Glass Refractive Index Measurement System 3 (GRIM 3) 1.0 Purpose - This technical procedure shall be followed for the operation of the GRIM 3. 2.0 Scope - This procedure applies

More information

Scope: Materials: Labeled 1.5 ml or 2 ml Eppendorf tubes Appropriate pipette tips (see Table 1) Labeled LC vials

Scope: Materials: Labeled 1.5 ml or 2 ml Eppendorf tubes Appropriate pipette tips (see Table 1) Labeled LC vials Scope: This SOP applies to sample protein precipitation for global metabolomics analysis by reverse phase or HILIC- HPLC-MS. Samples include but are not limited to tissue, cells, plasma, serum, and stool.

More information

Cells and Microscopes Biology Concepts of Biology 2.1

Cells and Microscopes Biology Concepts of Biology 2.1 Cells and Microscopes Biology 100 - Concepts of Biology 2.1 Name Instructor Lab Section Objectives: To gain an understanding of how to: Correctly use the compound light microscope Differentiate between

More information

GRAVIMETRIC ANALYSIS OF A CHLORIDE SALT. REFERENCES: Nelson, J., Chemistry: The Central Science, 3 rd edition, Prentice-Hall, 1985

GRAVIMETRIC ANALYSIS OF A CHLORIDE SALT. REFERENCES: Nelson, J., Chemistry: The Central Science, 3 rd edition, Prentice-Hall, 1985 1 GRAVIMETRIC ANALYSIS OF A CHLORIDE SALT REFERENCES: Nelson, J., Chemistry: The Central Science, 3 rd edition, Prentice-Hall, 1985 Typical techniques used in gravimetric analyses by quantitatively determining

More information

IV. DIRECT MEASUREMENT LABS. Introduction

IV. DIRECT MEASUREMENT LABS. Introduction IV. DIRECT MEASUREMENT LABS VI-1 I. Purpose Introduction An industrial hygienist frequently needs to evaluate airborne gas and vapor concentrations "on the spot". In other cases, he needs to track down

More information

CCME Reference Method for the Canada-Wide Standard for Petroleum Hydrocarbons (PHC) in Soil - Tier 1 Method

CCME Reference Method for the Canada-Wide Standard for Petroleum Hydrocarbons (PHC) in Soil - Tier 1 Method Rating Guide Appendix for the CCME Reference Method for the Canada-Wide Standard for Petroleum (PHC) in Soil - Tier 1 Method (Note: Checklist incorporates requirements from Dec 2000 version + Addendum

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES STANDARD OPERATING PROCEDURES for Spectrophotometric Measurement Chl a, Chl b, and Bulk Carotenoids using Methanol Solvent Prepared by EcoAnalysts, Inc. 1420 South Blaine Street, Suite 14 Moscow, ID 83843

More information

Operating Instructions Validation Tool Kit, KT2 For use with USP Apparatus #1 & #2

Operating Instructions Validation Tool Kit, KT2 For use with USP Apparatus #1 & #2 Operating Instructions Validation Tool Kit, KT2 For use with USP Apparatus #1 & #2 P/N VALTOL-KT2 Revision 2.0 October 7, 2014 Prepared by: Quality Lab Accessories, LLC 100 Emlen Way, Suite 108 Telford,

More information

saxon Instruction Manual saxon Grandeur Brass Telescope High quality optics

saxon Instruction Manual saxon Grandeur Brass Telescope High quality optics saxon High quality optics Instruction Manual saxon Grandeur Brass Telescope WARNING! Do not use the telescope to look at the sun without an appropriate solar filter. Looking at or near the sun can result

More information

WACEL AGGREGATE LABORATORY TESTING TECHNICIAN

WACEL AGGREGATE LABORATORY TESTING TECHNICIAN STUDY GUIDE WACEL AGGREGATE LABORATORY TESTING TECHNICIAN August 2016 Study Guide Aggregate Laboratory Testing General: An aggregate laboratory technician shall have sufficient training, education, and

More information

RECON. How to Setup the Telescope to Observe. This guide will show you how to setup your telescope for observing. Written By: Brittany McCrigler

RECON. How to Setup the Telescope to Observe. This guide will show you how to setup your telescope for observing. Written By: Brittany McCrigler RECON How to Setup the Telescope to Observe This guide will show you how to setup your telescope for observing. Written By: Brittany McCrigler 2017 recon.dozuki.com Page 1 of 30 INTRODUCTION This guide

More information

MODULE 2B INDUSTRIAL HYGIENE LABORATORY ACCREDITATION PROGRAM (IHLAP) ADDITIONAL REQUIREMENTS

MODULE 2B INDUSTRIAL HYGIENE LABORATORY ACCREDITATION PROGRAM (IHLAP) ADDITIONAL REQUIREMENTS 2B.1 SCOPE MODULE 2B INDUSTRIAL HYGIENE LABORATORY ACCREDITATION PROGRAM (IHLAP) ADDITIONAL REQUIREMENTS The AIHA - Laboratory Accreditation Programs (AIHA-LAP), LLC s Industrial Hygiene Laboratory Accreditation

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS OPERATING and CHANGING Instructions For Models: 8550, 8555, 8560, 8565 READ THESE INSTRUCTIONS THOROUGHLY BEFORE OPERATING THE LOCK OR ATTEMPTING TO CHANGE THE COMBINATION Note: To assure proper operation,

More information

Technical Procedure for General Laboratory Equipment

Technical Procedure for General Laboratory Equipment Technical Procedure for General Laboratory Equipment 1.0 Purpose - This procedure specifies the required elements for the use of general laboratory equipment. 2.0 Scope - This procedure applies to all

More information

[Unless otherwise stated all tolerances are ±5%] SAMPLES

[Unless otherwise stated all tolerances are ±5%] SAMPLES PHOSPHATASE TEST - FLUOROPHOS ALP TEST SYSTEM IMS #28 [Unless otherwise stated all tolerances are ±5%] SAMPLES 1. Laboratory Requirements (see Cultural Procedures [CP] items 33 & 34) [See current version

More information

MicroCell User manual

MicroCell User manual MicroCell User manual In Vitro Diagnostics Do not re-use, discard after procedure Caution: Consult accompanying documents Catalog number Vitrolife Sweden AB, Box 9080 (Gustaf Werners Gata 2) SE-400 92

More information

AIR-O-CELL CSI. Bioaerosol Sampling Cassette For the collection of viable and non-viable airborne particles. Laboratory & User Manual

AIR-O-CELL CSI. Bioaerosol Sampling Cassette For the collection of viable and non-viable airborne particles. Laboratory & User Manual AIR-O-CELL CSI Bioaerosol Sampling Cassette For the collection of viable and non-viable airborne particles Laboratory & User Manual Air-O-Cell CSI Sampling Cassette The Air-O-Cell Collector for SEM Identification

More information

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY.

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY. 1 REVISOR 4740.2100 4740.2100 QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY. Subpart 1. Scope. This part applies to laboratories performing testing under the inorganic chemistry, metals,

More information

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure Troy, NY 12180 Date: 12/12/2017 Page 1 of 7 USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure 1. Scope and Application 1.1 Analytes Dissolved inorganic

More information

ENVR 416 Aerosol Technology - Laboratory Session Fall 2007

ENVR 416 Aerosol Technology - Laboratory Session Fall 2007 ENVR 416 Aerosol Technology - Laboratory Session Fall 2007 CALIBRATION AND USE OF THE OPTICAL MICROSCOPE The objective of this lab is for you to become familiar with use of the optical microscope to measure

More information

Standard Operating Procedure for Performing a QA Review of an Aldehyde Data Batch

Standard Operating Procedure for Performing a QA Review of an Aldehyde Data Batch North Carolina Division of Air Quality Quality Assurance Project Plan 2.52 Aldehydes by TO-11a Section 4.1 Raleigh Central Office Responsibilities Aldehyde Data Review Section 2.52.4.1 Page 1 of 7 Standard

More information

Rate law Determination of the Crystal Violet Reaction Using the Isolation Method

Rate law Determination of the Crystal Violet Reaction Using the Isolation Method Rate law Determination of the Crystal Violet Reaction Using the Isolation Method Introduction A common challenge in chemical kinetics is to determine the rate law for a reaction with multiple reactants.

More information

STANDARD OPERATING PROCEDURE NO. 70 DENSITY AND VOLUMETRIC WATER CONTENT BY THE SAND REPLACEMENT METHOD

STANDARD OPERATING PROCEDURE NO. 70 DENSITY AND VOLUMETRIC WATER CONTENT BY THE SAND REPLACEMENT METHOD Questa Rock Pile Stability Study 70v4 Page 1 STANDARD OPERATING PROCEDURE NO. 70 DENSITY AND VOLUMETRIC WATER CONTENT BY THE SAND REPLACEMENT METHOD REVISION LOG Revision Number Description Date 70.0 Original

More information

N) manual. Biomaster Operating manual

N) manual. Biomaster Operating manual 830 ual N) manual Operating manual Copyright 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of the copyright owner. Trademarks Eppendorf and the

More information

Appendix 2: Disassembling and Assembling the Telescopes and the Celestron Equatorial Mounts

Appendix 2: Disassembling and Assembling the Telescopes and the Celestron Equatorial Mounts Appendix 2: Disassembling and Assembling the Telescopes and the Celestron Equatorial Mounts Disassembling (classroom) 1. The mount must be locked with the telescope situated horizontally. Note the telescope

More information

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction II. Quality Control A. Introduction ENCO adheres to strict quality control practices in order to assure our clients that the data provided are accurate and reliable. We are required by the EPA to analyze

More information

COMMODORE BRASS TELESCOPE ZHUMELL COMMODORE BRASS TELESCOPE

COMMODORE BRASS TELESCOPE ZHUMELL COMMODORE BRASS TELESCOPE 3 0 15 15 E OWNER S MANUAL COMMODORE BRASS TELESCOPE ZHUMELL COMMODORE BRASS TELESCOPE 5 W 60 W 30 W 45 W Zhumell customers know that there are plenty of ways to experience the world. They also understand

More information

LOOKING AT PLANT STEMS

LOOKING AT PLANT STEMS Activity 4.17 Student Sheet LOOKING AT PLANT STEMS Purpose To look at the structure of xylem vessels, phloem sieve tubes and sclerenchyma fibres. To locate the position of these tissues within the stem.

More information

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7 Troy, NY 12180 Date: 03/14/2012 Page 1 of 7 USGS New York Water Science Center Soil and Low Ionic Strength Water Quality Laboratory Electrometric Measurements and Titration Standard Operating Procedure

More information

AIR MONITORING REPORT

AIR MONITORING REPORT AIR MONITORING REPORT 1500 Hallway Frost Lake Magnet 1505 East Hoyt Avenue St. Paul, MN 55103 December 21, 2005 Nova Client No.: STPAU006/Project No: H05-1309 AIR MONITORING REPORT 1500 Hallway Frost Lake

More information

EXPERIMENT 7. Determination of Sodium by Flame Atomic-Emission Spectroscopy

EXPERIMENT 7. Determination of Sodium by Flame Atomic-Emission Spectroscopy EXPERIMENT 7 Determination of Sodium by Flame Atomic-Emission Spectroscopy USE ONLY DEIONIZED WATER (NOT DISTILLED WATER!) THROUGHOUT THE ENTIRE EXPERIMENT Distilled water actually has too much sodium

More information

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay... Table of Contents 1. Principle of Assay... 3 2. Specimens... 3 3. Reagents and Materials (HPLC grade)... 3 4. Standards, Controls, and Solutions... 4 5. Equipment and Special Supplies... 5 6. Instrumentation...

More information

10 - Celestron Telescope II: Operation

10 - Celestron Telescope II: Operation 10 - Celestron Telescope II: Operation Purpose: Gain more experience setting up a 6 Celestron telescope, familiarize yourself with the software interface, and acquire an image with the CCD camera. Due:

More information

Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures

Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures Villanova University Date: Oct 2011 Page 1 of 9 Villanova University Villanova Urban Stormwater Partnership Watersheds Laboratory Standard Operating Procedure VUSP F Zinc Metal Determination Perkin Elmer

More information

22X Builder s Transit Level Model No Instruction Manual

22X Builder s Transit Level Model No Instruction Manual 2595H 7/29/09 10:15 AM Page 1 22X Builder s Transit Level Model No. 40-6910 Instruction Manual Congratulations on your choice of this 22X Builder s Transit Level. We suggest you read this instruction manual

More information

Experiment 8 Michelson Interferometer

Experiment 8 Michelson Interferometer Experiment 8 Michelson Interferometer Introduction This week s experiment utilizes the Michelson interferometer. You are to measure the wavelength of the green mercury light, the wavelength of the sodium

More information

2. OPERATIONAL CONDITIONS

2. OPERATIONAL CONDITIONS 1. INTRODUCTION This device was designed for modern physics labs of colleges and graduate schools. It demonstrates the influence of a magnetic field on light, known as Zeeman Effect, and reveals the behavior

More information

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV Iron, Total DOC316.53.01338 FerroZine Method 1 Method 10264 1 to 100 µg/l Fe (10-cm cell) Reagent Solution Scope and application: For ultrapure water. 1 Adapted from Stookey, L.L., Anal. Chem., 42(7),

More information

THE HAVERFORD UNIVERSITY S STRAWBRIDGE OBSERVATORY S OF THE 16 MEADE. A quick guide to using the 16 Meade Schmidt-Cassegrain Telescope

THE HAVERFORD UNIVERSITY S STRAWBRIDGE OBSERVATORY S OF THE 16 MEADE. A quick guide to using the 16 Meade Schmidt-Cassegrain Telescope THE S OF THE 16 MEADE HAVERFORD UNIVERSITY S STRAWBRIDGE OBSERVATORY A quick guide to using the 16 Meade Schmidt-Cassegrain Telescope First draft: Scott Engle, Fall 2010; Update: Steve Boughn, Fall 2012

More information

TRANSPIRATION With LabQuest INTRODUCTION

TRANSPIRATION With LabQuest INTRODUCTION TRANSPIRATION With LabQuest LAB 10 From Biology with Vernier Westminster College INTRODUCTION Water is transported in plants from the roots to the leaves, following a decreasing water potential gradient.

More information

UNIT #4: Cell Structure & Func4on

UNIT #4: Cell Structure & Func4on UNIT #4: Cell Structure & Func4on PART ONE: Microscopes Four Different Types of Microscopes Dissec:ng Microscope Uses natural or electrical light source to illuminate an object Lens for magnifica4on Organisms

More information

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Electron Microscopy Sciences INSTRUCTION MANUAL. Stereo Microscopes Non-illuminated Stands

Electron Microscopy Sciences INSTRUCTION MANUAL. Stereo Microscopes Non-illuminated Stands Electron Microscopy Sciences 1 INSTRUCTION MANUAL Stereo Microscopes Non-illuminated Stands Model GL7-207: Binocular, 6.5:1 Zoom Ratio 10X to 65X, smooth ball bearing construction and Model GL7-307: Trinocular,

More information

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water Hach Method 1041 Spectrophotometric Measurement of Free Chlorine (Cl ) in Finished Drinking Water Hach Company Method 1041 Revision 1. November 015 Spectrophotometric Measurement of Free Cl in Finished

More information

Eppendorf UVette. Instructions for use

Eppendorf UVette. Instructions for use Instructions for use Instructions Eppendorf see Fig. Tab. p. p. UVette for use Copyright 2014 Eppendorf AG, Hamburg, Germany. All rights reserved, including graphics and images. No part of this publication

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 7 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Owner's Manual NSL100B BUILDERS LEVEL NSL500B TRANSIT LEVEL

Owner's Manual NSL100B BUILDERS LEVEL NSL500B TRANSIT LEVEL Owner's Manual NSL100B BUILDERS LEVEL NSL500B TRANSIT LEVEL 1 1. CONTENTS 2. Nomenclature page 3 3. Care and Maintenance page 4 4. Using your Instrument 4.1 Setting up your Instrument page 5 4.2 Stadia

More information

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7 Troy, NY 12180 Date: 03/16/2012 Page 1 of 7 USGS District Laboratory, Troy, NY Inductively Coupled Plasma Optical Emission Spectrometry Standard Operating Procedure 1. Scope and Application 1.1 Analytes

More information

Lasentec Product Group

Lasentec Product Group Document Number: 004-0043 Window Reference Procedure - M500/600 P Probes, version 6.0 and higher FBRM CI Software The Window Reference Procedure is the standard calibration procedure used to reference

More information

Biology Cell Unit Homework Packet #1

Biology Cell Unit Homework Packet #1 Name DUE: Biology Cell Unit Homework Packet #1 Hour HW #1 Microscope Questions Exploration Drawing Lab Post-Lab Questions / 5 possible points Homework #1: Microscope Exploration Lab When working with a

More information

22X Builder s Level Model No Instruction Manual

22X Builder s Level Model No Instruction Manual 2594H 7/29/09 10:12 AM Page 1 22X Builder s Level Model No. 40-6900 Instruction Manual Congratulations on your choice of this 22X Builder s Level. We suggest you read this instruction manual thoroughly

More information

National Optical & Scientific Instrument Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210)

National Optical & Scientific Instrument Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) National Optical & Scientific Instrument Inc. 6508 Tri-County Parkway Schertz, Texas 78154 Phone (210) 590-9010 Fax (210) 590-1104 INSTRUCTIONS FOR SHOP MICROSCOPES MODEL NUMBERS 186 187 188 189 National

More information

Observing Root and Stem Structures

Observing Root and Stem Structures Name Class Date Chapter 23 Structure and Function of Seed Plants Observing Root and Stem Structures Introduction The first structures to appear on a germinating seed are the roots. The initial root to

More information

Lasentec Product Group

Lasentec Product Group Prepared by: Gregor Hsiao Approved on: ECN# 3506 Effective Date: 1/14/2008 Title: Window Reference Procedure for Version 6.0.0 and Higher Purpose: Lasentec D600 Probes and FBRM/Raman Probes using version

More information

VCE VET Laboratory Skills GA 2: Written examination

VCE VET Laboratory Skills GA 2: Written examination VCE VET Laboratory Skills GA 2: Written examination GENERAL COMMENTS Students who analysed the information given before providing concise, succinct answers that related to laboratory experience best handled

More information

Inserting grids into capsule Removing capsules from grid box Immunolabeling using template guide Preparing for TEM insertion

Inserting grids into capsule Removing capsules from grid box Immunolabeling using template guide Preparing for TEM insertion Protocol Immunolabeling protocols are easily adapted to mprep/g processing. This document illustrates a typical protocol using a primary antibody and a secondary antibody conjugated to colloidal gold,

More information

QAM-I-116 Preparation of Labware

QAM-I-116 Preparation of Labware 1. Applicability and Purpose i. This procedure applies to all labware (glassware and plasticware) used for analysis in the laboratory, and sampling bottles and equipment used in field sample operations.

More information

Apparatus Needed To Perform Procedure Including Preparation of Reagent:

Apparatus Needed To Perform Procedure Including Preparation of Reagent: Name of Procedure: Polarized Light Microscopy Suggested Uses: The polarized light microscope serves as a tool for the testing of excipients, the use of a microcrystalline test for the testing of drugs

More information

Telescope. 1. Read these instructions carefully and familiarise yourself with the procedure before assembling the unit.

Telescope. 1. Read these instructions carefully and familiarise yourself with the procedure before assembling the unit. Telescope a Simple assembly instructions- Please keep for future reference IMPORTANT 1. Read these instructions carefully and familiarise yourself with the procedure before assembling the unit. 2. Check

More information

Coma Correction. Diagram. Introduction

Coma Correction. Diagram. Introduction Starlight Instruments, LLC 2380 E. Cardinal Drive, Columbia City, Indiana 46725 USA t: 260.244.0020 f.260.244.3077 e: sales@starlightinstruments.com web: starlightinstruments.com Introduction Thank you

More information

NYE COUNTY NUCLEAR WASTE REPOSITORY PROJECT OFFICE TECHNICAL PROCEDURE

NYE COUNTY NUCLEAR WASTE REPOSITORY PROJECT OFFICE TECHNICAL PROCEDURE NYE COUN1Y NYE COUNTY NUCLEAR WASTE REPOSITORY PROJECT OFFICE TECHNICAL PROCEDURE TITLE: Revision: 0 THIN SECTION PREPARATION PROCEDURE : 06-16-00 PROCEDURE No.: SUPERSEDES: Pa e: 1 of 6 TP-8.2 DRAFT,

More information

Glossary of Common Laboratory Terms

Glossary of Common Laboratory Terms Accuracy A measure of how close a measured value is to the true value. Assessed by means of percent recovery of spikes and standards. Aerobic Atmospheric or dissolved oxygen is available. Aliquot A measured

More information

Guideline/SOP: Handling of Laboratory Gross Errors/Data History

Guideline/SOP: Handling of Laboratory Gross Errors/Data History Guideline/SOP: Handling of Laboratory Gross Errors/Data History Introduction Laboratory gross errors are events in the laboratory that may produce erroneous results and can be usually attributed to either

More information

QAM-I-111 Operation and Calibration of the Conductivity Meter

QAM-I-111 Operation and Calibration of the Conductivity Meter 1.0 Applicability and Purpose i. This procedure applies to the operation and calibration of the YSI Model 3200 conductivity meter. This procedure is performed prior to any analysis using this meter. By

More information

Prelab vocabulary (complete before turning in beginning of Lab 2) Name: Date: Gen Bio 1 Lab #1: Microscopes and Measurements

Prelab vocabulary (complete before turning in beginning of Lab 2) Name: Date: Gen Bio 1 Lab #1: Microscopes and Measurements Name: Date: Gen Bio 1 Lab #1: Microscopes and Measurements Prelab Reading Assignment: Pages 94-97 in Campbell 10 th edition. Appendix C and D at the back of your text. Please look at the following websites

More information

x Builders Level Service Manual

x Builders Level Service Manual 40-690 22x Builders Level Service Manual Item Description Pages.0 Overall Instrument Assembly 2. Main Assembly 2.2 Telescope Assembly 3.3 Base Assembly 4.4 Frame Assembly 5 2.0 Calibration 6-8 2. Vial

More information

Radioactive Waste Management Guide

Radioactive Waste Management Guide State University of New York at Stony Brook Radiation Protection Services Radioactive Waste Management Guide EHSD0247 (02/08) Page 1 of 12 www.stonybrook.edu/ehs Radiation Protection Services This document

More information

To gain a working knowledge of unit cells, packing efficiency, coordination numbers and their relationship to each other.

To gain a working knowledge of unit cells, packing efficiency, coordination numbers and their relationship to each other. Solid State Modeling PURPOSE A B To explore some simple solid state structures. To explore unit cell stoichiometry. GOALS To visualize the three-dimensional arrangement of atoms and ions in common solid

More information

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web:  sales: Morehouse 1 Morehouse Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA 17403 PH: 717-843-0081 web: www.mhforce.com sales: edlane@mhforce.com 2 This presentation will cover the calibration

More information

The Basic Unit of Life Lab (Adapted from lab of same name) State Standard

The Basic Unit of Life Lab (Adapted from lab of same name) State Standard NAME: DATE: PERIOD: The Basic Unit of Life Lab (Adapted from lab of same name) State Standard 12.11.04 In this investigation, you will review the history of the microscope, practice the techniques for

More information

Owner's Manual. for AUTOMATIC LEVEL. NBL Series. FOR CUSTOMER SERVICE, PARTS & REPAIR, CALL Toll Free:

Owner's Manual. for AUTOMATIC LEVEL. NBL Series. FOR CUSTOMER SERVICE, PARTS & REPAIR, CALL Toll Free: Owner's Manual for AUTOMATIC LEVEL NBL Series FOR CUSTOMER SERVICE, PARTS & REPAIR, CALL Toll Free: 1-888-247-1960 1 1. CONTENTS 2. Nomenclature page 3 3. Care and Maintenance page 4 4. Using your Instrument

More information

Technology that counts

Technology that counts Application Note No. 102 Version 1.3 Determination of Sperm Density in Ejaculates from Bulls Application This Application Note describes how sperm cell density in ejaculates from bulls can be determined

More information

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction 1. Application 1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction Parameters and Codes: EDB and DBCP, whole water recoverable, O-3120-90 Parameter (µg/l)

More information

Technology that counts

Technology that counts Application Note No. 100 Version 1.3 Determination of Sperm Density in Ejaculates from Boars Application This Application Note describes how sperm cell density in ejaculates from boars can be determined

More information

Test Method Development and Validation as Pertaining to Microtrac Particle Size Measuring Instruments

Test Method Development and Validation as Pertaining to Microtrac Particle Size Measuring Instruments Test Method Development and Validation as Pertaining to Microtrac Particle Size Measuring Instruments Philip E. Plantz, PhD Application Note SL AN 17 Rev B Provided By: Microtrac, Inc. Particle Size Measuring

More information

SKELESCOPE REFLECTOR TELESCOPE WITH TABLE TOP TRIPOD SK-100 INSTRUCTION MANUAL

SKELESCOPE REFLECTOR TELESCOPE WITH TABLE TOP TRIPOD SK-100 INSTRUCTION MANUAL TM SKELESCOPE REFLECTOR TELESCOPE WITH TABLE TOP TRIPOD SK-100 INSTRUCTION MANUAL TABLE OF CONTENTS Introduction Box Contents / Visual Key Diagram Telescope Assembly Setting up Your Telescope Attaching

More information

Instructor s Advance Preparation

Instructor s Advance Preparation INSTRUCTOR'S MANUAL Instructor s Advance Preparation This protocol is designed for 80 workstations of 4 students. Each group will prepare a set of standards, a blank, and 2 milk samples (can be a blind

More information

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution Chloride DOC316.53.01017 Mercuric Thiocyanate Method Method 8113 0.1 to 25.0 mg/l Cl Reagent Solution Scope and application: For water and wastewater. Test preparation Instrument-specific information Table

More information

Responsibilities: Effective Date: November Revision Date: February 8, VP, Facilities and Construction Management. Issuing Authority:

Responsibilities: Effective Date: November Revision Date: February 8, VP, Facilities and Construction Management. Issuing Authority: Title: Chemical Hygiene Written Program Effective Date: November 2005 Revision Date: February 8, 2017 Issuing Authority: Responsible Officer: VP, Facilities and Construction Management Director Environmental

More information

THE DIFFRACTION GRATING SPECTROMETER

THE DIFFRACTION GRATING SPECTROMETER Purpose Theory THE DIFFRACTION GRATING SPECTROMETER a. To study diffraction of light using a diffraction grating spectrometer b. To measure the wavelengths of certain lines in the spectrum of the mercury

More information

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University. Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared

More information

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C Book 5 - Laboratory Policies and Procedures C03-1 2009-06-15 1 Scope 1.1 This method describes a general procedure for the determination of leachable arsenic, selenium, cadmium, antimony, and barium in

More information

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows Potassium DOC316.53.01127 Tetraphenylborate Method Method 8049 0.1 to 7.0 mg/l K Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific information

More information

Using Conductivity to Find an Equivalence Point

Using Conductivity to Find an Equivalence Point Experiment 25 PRE LAB DISCUSSION In this experiment, you will monitor conductivity during the reaction between sulfuric acid, and barium hydroxide in order to determine the equivalence point. From this

More information

Test Method: CPSC-CH-E

Test Method: CPSC-CH-E UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 10901 DARNESTOWN RD GAITHERSBURG, MD 20878 Test Method: CPSC-CH-E1001-08 Standard Operating Procedure

More information

Is the laboratory s pledge or declaration of the quality of the results produced. to produce data compliant with the Safe Drinking Water Act (SDWA)

Is the laboratory s pledge or declaration of the quality of the results produced. to produce data compliant with the Safe Drinking Water Act (SDWA) QA/QC Is the laboratory s pledge or declaration of the quality of the results produced. to produce data compliant with the Safe Drinking Water Act (SDWA) Is a description of the policies, procedures, techniques

More information

NcSTAR SHOOTER II SERIES SCOPE

NcSTAR SHOOTER II SERIES SCOPE NcSTAR SHOOTER II SERIES SCOPE Congratulations on the purchase of your New NcSTAR Shooter II Series Scope! The Shooter II Series of scopes gives you many great options, so you can choose the scope that

More information

DISTRIBUTION LIST To be filled out by Document Author or person requesting document deletion

DISTRIBUTION LIST To be filled out by Document Author or person requesting document deletion DISTRIBUTION LIST To be filled out by Document Author or person requesting document deletion: Once approved, the new or revised document or the notice of deleted document(s) should be distributed to all

More information