NEW X-RAY DETECTORS FOR XRF ANALYSIS. Jan S. Iwanczyk & Bradley E. Patt Photon Imaging, Inc., Northridge, CA 91324

Size: px
Start display at page:

Download "NEW X-RAY DETECTORS FOR XRF ANALYSIS. Jan S. Iwanczyk & Bradley E. Patt Photon Imaging, Inc., Northridge, CA 91324"

Transcription

1 951 NEW X-RAY DETECTORS FOR XRF ANALYSIS Jan S. Iwanczyk & Bradley E. Patt Photon Imaging, Inc., Northridge, CA Abstract The use of miniaturized XRF instrumentation for in-vivo applications imposes specific requirements on the choice of detector technologies that can be used, and on the design and construction of the instrumentation. One critical issue is the temperature of operation of the detector. Substantial cooling cannot be tolerated because of the difficulties of extracting the heat from a small probe. Another issue is the bias voltage of the detector, which impacts the safety design. Issues related to complexity of the device, ease of use, and options for sterilization between uses are also important considerations impacting the choice of detector technology. In this paper we present recent advances in detector technologies which open up new possibilities for in-vivo applications of XRF. In particular we discuss the characteristics of HgI,, CdTe, Si (pi-n), and Silicon Drift Detectors (SDD s) including their basic material properties, suitable detector structures, and state-of-the-art performance. The usefulness of the various technologies for in-vivo X-ray spectroscopy applications is discussed. Introduction There has been a continuous effort to develop non-cryogenic energy-dispersive compound semiconductor x-ray detectors for XRF applications. With advances in material characteristics and detector design, excellent results have been achieved with HgI,, CdZnTe, and CdTe (p-i-n) x-ray detectors. Resolution of better than 200 ev FWHM for HgI, detectors (at 0 C) [l] and better than 250 ev FWHM for both CdZnTe and CdTe(p-i-n) detectors (at -40 C) [2,3] have been measured at 5.9 kev. The above results were obtained with the use of a miniaturized, low-power thermoelectric coolers and dry nitrogen atmosphere. The active area of these detectors was about 10 mm2. With the availability of high quality, high resistivity silicon and better surface passivation techniques, encouraging results have also been obtained recently with Si (p-i-n) detectors. We have obtained energy resolution better than 220 ev FWHM at 5.9 kev with a device operating at about -40 C [3]. The active thickness of the device was about 300 pm. Thus the device has limited efficiency (less than 63% absorption) above 13 kev. In addition, the small thickness of the Si (p-i-n) structures limits their active area to only 6-7 mm2. Larger active areas would have too high detector capacitance resulting in increased electronic noise. This limitation can be overcome by using a novel structure, namely the silicon drift detector (SDD) structure [4]. This structure has a small charge-collection anode and cathodes that are designed to focus the charge (electrons) to the small anode. The device capacitance (anode capacitance) is minimized and is

2 This document was presented at the Denver X-ray Conference (DXC) on Applications of X-ray Analysis. Sponsored by the International Centre for Diffraction Data (ICDD). This document is provided by ICDD in cooperation with the authors and presenters of the DXC for the express purpose of educating the scientific community. All copyrights for the document are retained by ICDD. Usage is restricted for the purposes of education and scientific research. DXC Website ICDD Website -

3 952 independent of the detector size because the collection anode can be kept very small. SDD structures with an energy resolution of 145 ev (FWHM) at 5.9 kev have been shown [5] at an operating temperature of -20 C. However, these devices have a limited active area of 3 mm2. This paper will describe our efforts to obtain large SDD structures (close to 1 cm ) for X-ray spectroscopy applications. High Z Compound Semiconductors High atomic number (high-z) and wide band-gap compound semiconductors have been found to be an attractive choice for room-temperature x-ray spectrometers. The broad band-gap of these materials allows for a low detector leakage current and with a construction designed to keep detector capacitance below 1 pf, the detector electronic noise can be very low. Because of the high Z of the detector material, x-rays are absorbed very close to the entrance electrode. This feature helps to create a single-charge-collection situation which is very important for compound materials such as HgI,, CdTe and CdZnTe where electrons are much more mobile than holes. When all x-ray interactions take place near the negatively biased electrode, the generated holes travel only a minuscule distances, while the electrons must cross the detector s entire active thickness. In this case, the contribution of the holes to the induced pulse amplitude may be neglected and essentially only the electrons contribute to the spectral response. Over the past several years, a combination of low leakage currents and relatively good electron collection have resulted in energy resolution in the vicinity of 200 ev (FWHM) at 5.9 kev for HgI,, CdTe and CdZnTe compound detectors. The construction of the spectrometers and experimental details are given elsewhere [2,3]. Silicon Drift Detector (SDD) Basic Operation The basic configuration of the silicon drift x-ray detector is shown in Fig. 1. The p+ electrodes on the front and the back side are reverse-biased, so that the bulk of the detector is fully depleted, producing a parabolic-shaped potential valley within the detector from the front to the back side. The bottom of the drift valley corresponds to the minimum potential energy for electrons, and is where the electrons are transported. By applying a radial potential gradient across the back surface of the detector (with decreasing potential from V, to V,), the drift valley can be forced to tilt along the detector toward the n+ collection anode as shown in Fig. 1. The detector structure thus focuses the electron collection to the small anode with the induced signal occurring predominantly between the anode and the adjacent focusing electrode. In the type of structure depicted in Fig. 1, the anode capacitance is minimized (and is independent on the detector size) because the collection anode can be very small (80 pm diameter). Our approach, described below, has several new features incorporated into the design. These include: p+/n+ concentric rings on the back side, and equipotential bias on the front side.

4 cm J I I., I v2 p+ enpance electrode,- / I 1 L Figure 1: \ I p electrode HV termination guard node Cross section of cylindrical drift detector (Basic concept) The investigation of the various contacting configurations, biasing voltages, dimensions of the structure, and resulting electric field distributions has been accomplished by computer simulation using 2D simulation tools such as SILVACO and MEDIC1 physical device simulator packages VI. The theoretical noise limit of the SDD was determined from calculations discussed in [4] to be about 8 em RMS for the optimal case with integrated FET and - 60 em RMS for the case with an external FET. Reduction of Surface leakage current One of the major components of the leakage current in drift detectors is due to surface leakage current. Surface leakage is due to electron-hole pair generation at the silicon/oxide interface. Because of the large number of midgap states at the interface, the surface generation component is typically much larger than the bulk component and diffusion current. Surface leakage is strongly dependent on the starting material, the processing conditions, and the operating temperature. The oxidation process is probably the most critical. TCA (trichloroethane) oxidation, which we have used, has been reported to lower generation rates [6]. We measured rates of around 60 na/cm2. One way to control the surface leakage current is to redirect it to a guard anode, preventing it from reaching the main anode [7]. This technique works well with a diffusion spiral [7], because the non-diffused region forms its own spiral which directs charge to a guard anode. In our previous work we introduced a new approach developed to suppress surface generation current. Instead of diverting surface leakage current to a separate guard anode, we have implemented an alternating p+/n+ ring structure to minimize the depleted surface area, which is responsible for the current. If the surface of the detector between p+ diffusion rings is doped n+, it will not be depleted, and only those areas between the n+ and p+ rings will be depleted. MEDIC1 simulations and experimental data show that with this technique the amount of depleted surface area is significantly reduced and the surface leakage generation rates become comparable with the bulk generation rates (1 &cm*) [8]. The additional advantage of the new structure is the robustness of the design to process defects. With the guard anode technique, any defect

5 954 along the surface current path, resistance variation or large local current generation could render the guard anode ineffective. With the p+/n+ ring design, however, a defect in either p+ or n+ region is likely to be benign. X-Ray Measurements Electronic Noise: Preliminary X-ray measurements were made with a p+/n+ ring SDD structure. The SDD was mounted in a test chamber on a ceramic substrate and slightly cooled with a Peltier cooler to about -10 C to reduce the electronic noise and achieve less than 75 pa leakage current. The test chamber, with a 125 pm thick Beryllium window, was evacuated to prevent water condensation. Electronics with an external FET were used in these measurements. The use of an external FET elevates the noise over what can be achieved optimally with an integrated FET. Figure 2 shows an Fe-55 source spectrum when the whole SDD active area was uniformly illuminated. Energy resolution of 540 ev (FWHM) was obtained at 5.9 kev and electronic noise of 500 ev was measured by the pulser method. Charge Collection Uniformity: An experiment was designed to evaluate the charge collection uniformity over the whole active SDD area. A 0.6 mm diameter collimator was placed directly above the SDD constricting the x-ray beam from an Fe-55 source illuminating the detector to a 2 mm radius. The collimated source was moved above the detector using a precision X-Y stage and the spectral response and the count rate were measured for each point on a square 1 mm step grid. The measured energy resolution was identical (within the statistical error) over the whole active area and equal to that taken with the whole SDD exposed (i.e. without the source collimation). The theoretical and measured (normalized to unity) count rate along the X-axis is shown in Fig. 3. The theoretical count rate characteristic response is the convolution of the detector geometry and the shadow of the collimator on the detector. The observed distribution agrees well with the expected count rate :- Fe-55 (entire SDD) 5000 I- g 4000 I- $ 3000 I- 540eVFWHM --, : :- Pulser IO 12 Energy(keV) Figure 2. Fe-55 source x-ray spectrum 1 g d & : d Distance (mm) Figure 3. Count rate measured along X-axis. Theoretical calculation (solid line) and measured response (filled circles).

6 955 The above results show excellent electron collection from the full detector volume regardless of the drift distance (shortest at the device center and longest at its periphery). Previously reported results have only achieved results for very small drift detector volume (fraction of the full detector volume). Thus, to the best of our knowledge, this is first demonstration of large active volume x-ray SDD at this performance level by anyone. X-Ray Spectra: Figure 4 presents the X-ray fluorescence (XRF) spectra of characteristic K X- ray lines of Titanium (4.5 kev), Iron (6.4 kev), Nickel (7.47 kev), and Copper (8.04 kev). The XRF spectra were taken using a laboratory X-ray excitation source and detection system consisting of X-ray generator tube (with Rhodium anode and Beryllium exit window), target holder and an attachment for the detector test chamber. The first peak in all of the spectra in Fig. 4 is from the XRF target while the second peak is a pulser-induced peak used to measure the electronic noise of the system. The electronic noise level was 500 ev FWHM with the external FET. This agrees well with the theoretical noise (460 ev) calculated using the model described in [4] for total input capacitance of about 3 pf at 1.5 microseconds peaking time. The gaussian spread of the detector signal (electron cloud) in the time domain estimated using measurement of the rise time was found to be less than 35 ns for all drift distances. This number agrees well with the results of theoretical calculations of the drifting electron cloud shape evolution [9] ~ Energy (kev) Energy (kev) Figure 4: X-ray fluorescence (XRF) spectra of characteristic K X-ray lines of Titanium (4.5 kev), Iron (6.4 kev), Nickel (7.47 kev), and Copper (8.04 kev). The first peak in all spectra is from XRF target while the second peak is the pulser-induced peak.

7 956 Discussion Use of a particular detector technology for a given practical application depends on many factors. The most important factors include a) energy resolution; b) spectral quality (peak symmetry, tailing, peak-to-background ratio, etc.); c) detection efficiency; d) maximum detector active area, e) maximum count rate, r) detector and input electronics operating temperature; g) stability and reliability; i) ability to reduce weight, size, and power consumption; and j) cost. The requirements for the design and construction of instrumentation for medical applications impose additional restrictions. These additional requirements include a) ease of use (miniaturization), b) maximum solid angle and efficiency to minimize both radiation dose and measurement time, c) low operating bias voltages for safety reasons, and d) capability for sterilization. Currently all the detector technologies that we have discussed (HgI,, CdTe, CdZnTe, Si (p-i-n), and SDD) offer very similar spectral characteristics for the detection of x-rays. Thus the practical selection of a particular detector technology for a given application is dictated by the other requirements. One of these important criterion is the stopping power (inverse of the photoelectric linear attenuation coefficient). Si diodes of thickness 300 pm have a good detection efficiency up to 10 kev (900/,). For higher energies the detection efficiency drops very rapidly making them impractical for the majority of applications above 10 kev. CdTe and HgI, have significantly higher detection efficiency due to their high average atomic numbers. CdTe and HgI, have similar detection efficiencies up to 88 kev with certain differences related specifically to the x-ray absorption of material constituents and their absorption edges. Above 88 kev (Mercury K-absorption edge) the stopping power of HgI, is significantly higher than CdTe due to a higher Z number. Thus, for example, at 140 kev, which corresponds to the gamma-ray energy of the most widely used isotope in nuclear medicine: 99mTc, an easily attainable thickness of < 3 mm of mercuric iodide ensures efficiency of 90% whereas for CdTe almost 7 mm thick detectors are needed in order to obtain the same efficiency. The other differentiating factor in the above detector technologies is the operating temperature. In the case of HgI,, an operating temperature of 0 C is sufficient to obtain excellent spectral results. Cooling below -5 C gives very little benefit since properly fabricated detectors can attain leakage currents of <lo-l3 A at 0 C. Further cooling only reduces input FET noise and eventually enhances charge collection (through an increase in the mobilities of carriers) that can only be seen with thick detectors. SDDs can be operated between -10 C and -20 C. In the case of CdTe, CdZnTe, and Si(p-i-n) the best results are obtained at -40 C. The primary reason for the lower temperature is to keep the dark leakage current below lo-l3 A. The difference in operating temperature has a large impact on the design of hand held portable instrumentation. Whereas truly miniature systems are possible using -5 C detector technologies, systems using -40 C detector technologies will be significantly more bulky, primarily due to additional battery requirements [3]. Future improvements will rely on use of more efficient battery technologies and switched temperature regulators. Another issue of importance for any practical application is the stability and reliability of the detector technology. Silicon technology, of course, has achieved the highest degree of maturity;

8 957 however, CdTe and HgI, technologies have made great advances with improvements in lattice perfection, a decrease in impurities, and better detector surface passivation and encapsulation. SDDs when completely developed including integrated input FETs offer large active areas >lcm*, unparalleled high count rates >I O6 counts per second due to the theoretical minimum of electronic noise at short peaking times (0.1 ps), and high energy resolution. Currently attainable x-ray spectroscopy performance with the technological approaches discussed above is comparable to that of cryogenically cooled Si[Li] or Ge systems without the need for liquid nitrogen. The parallel development of miniaturized, low-noise, low-power processing electronics allows one to construct very convenient instruments for a variety of applications. The reduced weight, power, and size of the spectrometers makes them extremely attractive as portable x-ray florescence instruments for medical research, environmental monitoring, geological exploration, marine mineral analysis, archeometry, space exploration, and industrial materials quality assurance. ACKNOWLEDGEMENTS This work was supported by the Jet Propulsion Laboratory Contract No and by the National Institutes of Health, Grant No. lr43 CA75815 and Grant No. 2R44 RR The authors wish to thank Dr. Julie D. Segal of Stanford University for her contribution to the simulation and design of SDD structures. The valuable technical assistance of Mr. J. Menjivar and Mrs. M. Wells of Photon Imaging, Inc. is greatly appreciated. REFERENCES J.S. Iwanczyk, Y.J. Wang, J.G. Bradley, J.M. Conley, A.L. Albee, and T.E. Economou, IEEE Trans. Nucl. Sci. NS-36 (1) (1989) A. Niemela and H. Sipila, IEEE Trans. on Nucl. Sci. Vol. 41, No. 4, (1988) J.S. Iwanczyk, B.E. Patt, Y.J. Wang, and A.Kh. Khusainov, Nucl. Instr. & Meth. in Phys. Res. A 380 (1996) J.S. Iwanczyk, B.E. Patt, G. Vilkelis, L. Rehn, J. Metz, B. Hedman & K. Hodgson, Nucl. Instr. & Meth. in Phys. Res. A380 (1996) RONTEC GmbH Catalog, June W. Chen et al., Fabrication of Large Area Si Cylindric Drift Detectors, IEEE Trans. on Nucl. Sci., Vol. 41, No. 4, (1994) P. Rehak et al, IEEE Trans. on Nucl. Sci., Vol. 36, No. 1 (1989) S. Holland, Nucl Instr. andmeth. in Phys. Res., Vol. A305, (1991) J.D. Segal, J.D. Plummer, and C. Kenney, Proceedings 1996 IEEE Nuclear Science Symposium, pp

VORTEX A NEW HIGH PERFORMANCE SILICON DRIFT DETECTOR FOR XRD AND XRF APPLICATIONS

VORTEX A NEW HIGH PERFORMANCE SILICON DRIFT DETECTOR FOR XRD AND XRF APPLICATIONS Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 332 VORTEX A NEW HIGH PERFORMANCE SILICON DRIFT DETECTOR FOR XRD AND XRF APPLICATIONS Shaul Barkan,

More information

IN c/1vfl XRF SPECTROSCOPY: SUMMARY OF WORKSHOP II

IN c/1vfl XRF SPECTROSCOPY: SUMMARY OF WORKSHOP II 958 IN c/1vfl XRF SPECTROSCOPY: SUMMARY OF WORKSHOP II Contributors: Q. Ao, North Carolina State University R. Gardner, North Carolina State University J. Iwanczyk, Photon Imaging, Inc. L. Wielopolski,

More information

Semiconductor X-Ray Detectors. Tobias Eggert Ketek GmbH

Semiconductor X-Ray Detectors. Tobias Eggert Ketek GmbH Semiconductor X-Ray Detectors Tobias Eggert Ketek GmbH Semiconductor X-Ray Detectors Part A Principles of Semiconductor Detectors 1. Basic Principles 2. Typical Applications 3. Planar Technology 4. Read-out

More information

SILICON DRIFT DETECTORS FOR HIGH RESOLUTION, HIGH COUNT RATE X-RAY SPECTROSCOPY AT ROOM TEMPERATURE

SILICON DRIFT DETECTORS FOR HIGH RESOLUTION, HIGH COUNT RATE X-RAY SPECTROSCOPY AT ROOM TEMPERATURE Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 53 SILICON DRIFT DETECTORS FOR HIGH RESOLUTION, HIGH COUNT RATE X-RAY SPECTROSCOPY AT ROOM TEMPERATURE

More information

Chem 481 Lecture Material 3/20/09

Chem 481 Lecture Material 3/20/09 Chem 481 Lecture Material 3/20/09 Radiation Detection and Measurement Semiconductor Detectors The electrons in a sample of silicon are each bound to specific silicon atoms (occupy the valence band). If

More information

Gamma and X-Ray Detection

Gamma and X-Ray Detection Gamma and X-Ray Detection DETECTOR OVERVIEW The kinds of detectors commonly used can be categorized as: a. Gas-filled Detectors b. Scintillation Detectors c. Semiconductor Detectors The choice of a particular

More information

DEVELOPMENT OF A NEW POSITRON LIFETIME SPECTROSCOPY TECHNIQUE FOR DEFECT CHARACTERIZATION IN THICK MATERIALS

DEVELOPMENT OF A NEW POSITRON LIFETIME SPECTROSCOPY TECHNIQUE FOR DEFECT CHARACTERIZATION IN THICK MATERIALS Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 59 DEVELOPMENT OF A NEW POSITRON LIFETIME SPECTROSCOPY TECHNIQUE FOR DEFECT CHARACTERIZATION IN

More information

Semiconductor Detectors

Semiconductor Detectors Semiconductor Detectors Summary of Last Lecture Band structure in Solids: Conduction band Conduction band thermal conductivity: E g > 5 ev Valence band Insulator Charge carrier in conductor: e - Charge

More information

Silicon Drift Detectors: Understanding the Advantages for EDS Microanalysis. Patrick Camus, PhD Applications Scientist March 18, 2010

Silicon Drift Detectors: Understanding the Advantages for EDS Microanalysis. Patrick Camus, PhD Applications Scientist March 18, 2010 Silicon Drift Detectors: Understanding the Advantages for EDS Microanalysis Patrick Camus, PhD Applications Scientist March 18, 2010 EDS Detector Requirements Detect whole energy range of x-rays 50 ev

More information

Praktikum zur. Materialanalytik

Praktikum zur. Materialanalytik Praktikum zur Materialanalytik Energy Dispersive X-ray Spectroscopy B513 Stand: 19.10.2016 Contents 1 Introduction... 2 2. Fundamental Physics and Notation... 3 2.1. Alignments of the microscope... 3 2.2.

More information

BENEFITS OF IMPROVED RESOLUTION FOR EDXRF

BENEFITS OF IMPROVED RESOLUTION FOR EDXRF 135 Abstract BENEFITS OF IMPROVED RESOLUTION FOR EDXRF R. Redus 1, T. Pantazis 1, J. Pantazis 1, A. Huber 1, B. Cross 2 1 Amptek, Inc., 14 DeAngelo Dr, Bedford MA 01730, 781-275-2242, www.amptek.com 2

More information

Development and characterization of 3D semiconductor X-rays detectors for medical imaging

Development and characterization of 3D semiconductor X-rays detectors for medical imaging Development and characterization of 3D semiconductor X-rays detectors for medical imaging Marie-Laure Avenel, Eric Gros d Aillon CEA-LETI, DETectors Laboratory marie-laure.avenel@cea.fr Outlines Problematic

More information

Efficiency and Attenuation in CdTe Detectors

Efficiency and Attenuation in CdTe Detectors Efficiency and Attenuation in CdTe Detectors Amptek Inc. Bob Redus, May 5, 00 Amptek s XR-00T-CdTe is a high performance x-ray and gamma ray detector system. Like Amptek s other XR00 products, a detector

More information

Compton Camera. Compton Camera

Compton Camera. Compton Camera Diagnostic Imaging II Student Project Compton Camera Ting-Tung Chang Introduction The Compton camera operates by exploiting the Compton Effect. It uses the kinematics of Compton scattering to contract

More information

Measurement of material uniformity using 3-D position sensitive CdZnTe gamma-ray spectrometers

Measurement of material uniformity using 3-D position sensitive CdZnTe gamma-ray spectrometers Nuclear Instruments and Methods in Physics Research A 441 (2000) 459}467 Measurement of material uniformity using 3-D position sensitive CdZnTe gamma-ray spectrometers Z. He *, W.Li, G.F. Knoll, D.K. Wehe,

More information

Diffraction: spreading of waves around obstacles (EM waves, matter, or sound) Interference: the interaction of waves

Diffraction: spreading of waves around obstacles (EM waves, matter, or sound) Interference: the interaction of waves Diffraction & Interference Diffraction: spreading of waves around obstacles (EM waves, matter, or sound) Interference: the interaction of waves Diffraction in Nature What is Interference? The resultant

More information

RADIOACTIVE SAMPLE EFFECTS ON EDXRF SPECTRA

RADIOACTIVE SAMPLE EFFECTS ON EDXRF SPECTRA 90 RADIOACTIVE SAMPLE EFFECTS ON EDXRF SPECTRA Christopher G. Worley Los Alamos National Laboratory, MS G740, Los Alamos, NM 87545 ABSTRACT Energy dispersive X-ray fluorescence (EDXRF) is a rapid, straightforward

More information

AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD

AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN 1097-0002 219 AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD Takao Moriyama 1), Atsushi Morikawa 1), Makoto

More information

Time-Resolved μ-xrf and Elemental Mapping of Biological Materials

Time-Resolved μ-xrf and Elemental Mapping of Biological Materials 296 Time-Resolved μ-xrf and Elemental Mapping of Biological Materials K. Tsuji 1,2), K. Tsutsumimoto 1), K. Nakano 1,2), K. Tanaka 1), A. Okhrimovskyy 1), Y. Konishi 1), and X. Ding 3) 1) Department of

More information

LASER-COMPTON SCATTERING AS A POTENTIAL BRIGHT X-RAY SOURCE

LASER-COMPTON SCATTERING AS A POTENTIAL BRIGHT X-RAY SOURCE Copyright(C)JCPDS-International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Vol.46 74 ISSN 1097-0002 LASER-COMPTON SCATTERING AS A POTENTIAL BRIGHT X-RAY SOURCE K. Chouffani 1, D. Wells

More information

Advances in Field-Portable XRF

Advances in Field-Portable XRF Advances in Field-Portable XRF Volker Thomsen and Debbie Schatzlein Field-portable x-ray fluorescence (XRF) allows us to take the laboratory to the sample. The latest generation of such handheld x-ray

More information

Analysis of Background Events in Silicon Drift Detectors

Analysis of Background Events in Silicon Drift Detectors Analysis of Background Events in Silicon Drift Detectors T. Eggert a, P. Goldstrass a J. Kemmer a A. Pahlke b a Ketek GmbH, Gustav-Heinemann-Ring 125, D-81739 München, Germany b MPI Halbleiterlabor, Otto-Hahn-Ring

More information

XRF books: Analytical Chemistry, Kellner/Mermet/Otto/etc. 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures:

XRF books: Analytical Chemistry, Kellner/Mermet/Otto/etc. 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures: 1 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures: XRF spectroscopy 1 exam question. Notes on: www.nuigalway.ie/nanoscale/3rdspectroscopy.html XRF books: Analytical Chemistry,

More information

Advantages / Disadvantages of semiconductor detectors

Advantages / Disadvantages of semiconductor detectors Advantages / Disadvantages of semiconductor detectors Semiconductor detectors have a high density (compared to gas detector) large energy loss in a short distance diffusion effect is smaller than in gas

More information

DETECTORS. I. Charged Particle Detectors

DETECTORS. I. Charged Particle Detectors DETECTORS I. Charged Particle Detectors A. Scintillators B. Gas Detectors 1. Ionization Chambers 2. Proportional Counters 3. Avalanche detectors 4. Geiger-Muller counters 5. Spark detectors C. Solid State

More information

NEW CORRECTION PROCEDURE FOR X-RAY SPECTROSCOPIC FLUORESCENCE DATA: SIMULATIONS AND EXPERIMENT

NEW CORRECTION PROCEDURE FOR X-RAY SPECTROSCOPIC FLUORESCENCE DATA: SIMULATIONS AND EXPERIMENT Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 266 NEW CORRECTION PROCEDURE FOR X-RAY SPECTROSCOPIC FLUORESCENCE DATA: SIMULATIONS AND EXPERIMENT

More information

Development of semiconductor imaging detectors for a Si/CdTe Compton camera

Development of semiconductor imaging detectors for a Si/CdTe Compton camera SLAC-PUB-12926 October 27 Development of semiconductor imaging detectors for a Si/CdTe Compton camera Shin Watanabe a,, Shin ichiro Takeda a,b, Shin-nosuke Ishikawa a,b, Hirokazu Odaka a,b, Masayoshi Ushio

More information

Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector

Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector C. Fiorini, member, IEEE, A. Gola, M. Zanchi, A. Longoni, P. Lechner, H. Soltau, L. Strüder Abstract In this work

More information

Chapter 4 Scintillation Detectors

Chapter 4 Scintillation Detectors Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 4-1 4.1. Basic principle of the scintillator Chapter 4 Scintillation Detectors Scintillator Light sensor Ionizing radiation Light (visible,

More information

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY Copyright -International Centre for Diffraction Data 2010 ISSN 1097-0002 CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY B. Chyba, M. Mantler, H. Ebel, R. Svagera Technische Universit Vienna,

More information

Boron-based semiconductor solids as thermal neutron detectors

Boron-based semiconductor solids as thermal neutron detectors Boron-based semiconductor solids as thermal neutron detectors Douglas S. McGregor 1 and Stan M. Vernon 2 1 S.M.A.R.T. Laboratory, Department of Nuclear Engineering and Radiological Sciences, University

More information

PERFORMANCE IMPROVEMENT OF CZT DETECTORS BY LINE ELECTRODE GEOMETRY

PERFORMANCE IMPROVEMENT OF CZT DETECTORS BY LINE ELECTRODE GEOMETRY Applications of Nuclear Techniques (CRETE3) International Journal of Modern Physics: Conference Series Vol. 27 (24) 4644 (8 pages) The Authors DOI:.42/S294546446 PERFORMANCE IMPROVEMENT OF CZT DETECTORS

More information

Semiconductor-Detectors

Semiconductor-Detectors Semiconductor-Detectors 1 Motivation ~ 195: Discovery that pn-- junctions can be used to detect particles. Semiconductor detectors used for energy measurements ( Germanium) Since ~ 3 years: Semiconductor

More information

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K.

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K. 783 SCOPE AND LIMITATIONS XRF ANALYSIS FOR SEMI-QUANTITATIVE Introduction Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K. Historically x-ray fluorescence spectrometry has

More information

Energetic particles and their detection in situ (particle detectors) Part II. George Gloeckler

Energetic particles and their detection in situ (particle detectors) Part II. George Gloeckler Energetic particles and their detection in situ (particle detectors) Part II George Gloeckler University of Michigan, Ann Arbor, MI University of Maryland, College Park, MD Simple particle detectors Gas-filled

More information

Silicon Drift Detectors for gamma-ray detection: 15 years of research (and collaboration between Politecnico and INAF-Milano)

Silicon Drift Detectors for gamma-ray detection: 15 years of research (and collaboration between Politecnico and INAF-Milano) Silicon Drift Detectors for gamma-ray detection: 15 years of research (and collaboration between Politecnico and INAF-Milano) Outline The Silicon Drift Detector (SDD) Gamma-ray detectors based on scintillators

More information

X-ray spectrometry with Peltier-cooled large area avalanche photodiodes

X-ray spectrometry with Peltier-cooled large area avalanche photodiodes Nuclear Instruments and Methods in Physics Research B 213 (24) 267 271 www.elsevier.com/locate/nimb X-ray spectrometry with Peltier-cooled large area avalanche photodiodes L.M.P. Fernandes, J.A.M. Lopes,

More information

EEE4106Z Radiation Interactions & Detection

EEE4106Z Radiation Interactions & Detection EEE4106Z Radiation Interactions & Detection 2. Radiation Detection Dr. Steve Peterson 5.14 RW James Department of Physics University of Cape Town steve.peterson@uct.ac.za May 06, 2015 EEE4106Z :: Radiation

More information

Lecture 2. Introduction to semiconductors Structures and characteristics in semiconductors. Fabrication of semiconductor sensor

Lecture 2. Introduction to semiconductors Structures and characteristics in semiconductors. Fabrication of semiconductor sensor Lecture 2 Introduction to semiconductors Structures and characteristics in semiconductors Semiconductor p-n junction Metal Oxide Silicon structure Semiconductor contact Fabrication of semiconductor sensor

More information

Lecture 2. Introduction to semiconductors Structures and characteristics in semiconductors

Lecture 2. Introduction to semiconductors Structures and characteristics in semiconductors Lecture 2 Introduction to semiconductors Structures and characteristics in semiconductors Semiconductor p-n junction Metal Oxide Silicon structure Semiconductor contact Literature Glen F. Knoll, Radiation

More information

Final report on DOE project number DE-FG07-99ID High Pressure Xenon Gamma-Ray Spectrometers for Field Use

Final report on DOE project number DE-FG07-99ID High Pressure Xenon Gamma-Ray Spectrometers for Field Use Final report on DOE project number DE-FG07-99ID13772 High Pressure Xenon Gamma-Ray Spectrometers for Field Use Principle Investigator: Glenn K. Knoll Co-investigator: David K. Wehe, Zhong He, University

More information

Lecture 2. Introduction to semiconductors Structures and characteristics in semiconductors

Lecture 2. Introduction to semiconductors Structures and characteristics in semiconductors Lecture 2 Introduction to semiconductors Structures and characteristics in semiconductors Semiconductor p-n junction Metal Oxide Silicon structure Semiconductor contact Literature Glen F. Knoll, Radiation

More information

ABNORMAL X-RAY EMISSION FROM INSULATORS BOMBARDED WITH LOW ENERGY IONS

ABNORMAL X-RAY EMISSION FROM INSULATORS BOMBARDED WITH LOW ENERGY IONS 302 ABNORMAL X-RAY EMISSION FROM INSULATORS BOMBARDED WITH LOW ENERGY IONS M. Song 1, K. Mitsuishi 1, M. Takeguchi 1, K. Furuya 1, R. C. Birtcher 2 1 High Voltage Electron Microscopy Station, National

More information

The annealing of interstitial carbon atoms in high resistivity n-type silicon after proton irradiation

The annealing of interstitial carbon atoms in high resistivity n-type silicon after proton irradiation ROSE/TN/2002-01 The annealing of interstitial carbon atoms in high resistivity n-type silicon after proton irradiation M. Kuhnke a,, E. Fretwurst b, G. Lindstroem b a Department of Electronic and Computer

More information

ELECTRIC FIELD INFLUENCE ON EMISSION OF CHARACTERISTIC X-RAY FROM Al 2 O 3 TARGETS BOMBARDED BY SLOW Xe + IONS

ELECTRIC FIELD INFLUENCE ON EMISSION OF CHARACTERISTIC X-RAY FROM Al 2 O 3 TARGETS BOMBARDED BY SLOW Xe + IONS 390 ELECTRIC FIELD INFLUENCE ON EMISSION OF CHARACTERISTIC X-RAY FROM Al 2 O 3 TARGETS BOMBARDED BY SLOW Xe + IONS J. C. Rao 1, 2 *, M. Song 2, K. Mitsuishi 2, M. Takeguchi 2, K. Furuya 2 1 Department

More information

Hard X- and g-ray measurements with a large volume coplanar grid CdZnTe detector

Hard X- and g-ray measurements with a large volume coplanar grid CdZnTe detector Nuclear Instruments and Methods in Physics Research A 563 (26) 242 248 www.elsevier.com/locate/nima Hard X- and g-ray measurements with a large volume coplanar grid CdZnTe detector Alan Owens a,, T. Buslaps

More information

MONTE-CARLO MODELING OF SILICON X-RAY DETECTORS

MONTE-CARLO MODELING OF SILICON X-RAY DETECTORS 274 MONTE-CARLO MODELING OF SILICON X-RAY DETECTORS Brian Cross (1), Greg Bale (2), Barrie Lowe (2) and Rob Sareen (2) (1) CrossRoads Scientific, 414 Av. Portola, El Granada, CA 94018-1823, USA. (2) Gresham

More information

Review of Semiconductor Drift Detectors

Review of Semiconductor Drift Detectors Pavia October 25, 2004 Review of Semiconductor Drift Detectors Talk given by Pavel Rehak following a presentation on 5 th Hiroshima Symposium of Semiconductor Tracking Detectors Outline of the Review Principles

More information

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors 5. Radiation Microsensors Radiation µ-sensors convert incident radiant signals into standard electrical out put signals. Radiant Signals Classification

More information

On the development of compound semi-conductor thallium bromide detectors for astrophysics

On the development of compound semi-conductor thallium bromide detectors for astrophysics On the development of compound semi-conductor thallium bromide detectors for astrophysics A. Owens *1, M. Bavdaz 1, I. Lisjutin 2, A. Peacock 1, S. Zatoloka 2 1 Astrophysics Division, ESA/ESTEC, Postbus

More information

Detectors for High Resolution Gamma-ray Imaging Based on a Single CsI(Tl) Scintillator Coupled to an Array of Silicon Drift Detectors

Detectors for High Resolution Gamma-ray Imaging Based on a Single CsI(Tl) Scintillator Coupled to an Array of Silicon Drift Detectors 1 Detectors for High Resolution Gamma-ray Imaging Based on a Single CsI(Tl) Scintillator Coupled to an Array of Silicon Drift Detectors C. Fiorini, A. Longoni, F. Perotti, C. Labanti, E. Rossi, P. Lechner,

More information

Single Photon detectors

Single Photon detectors Single Photon detectors Outline Motivation for single photon detection Semiconductor; general knowledge and important background Photon detectors: internal and external photoeffect Properties of semiconductor

More information

USABILITY OF PORTABLE X-RAY SPECTROMETER FOR DISCRIMINATION OF VALENCE STATES

USABILITY OF PORTABLE X-RAY SPECTROMETER FOR DISCRIMINATION OF VALENCE STATES Copyright (c)jcpds-international Centre for Diffraction Data 00, Advances in X-ray Analysis, Volume 45. 409 ISSN 1097-000 USABIITY OF POTABE X-AY SPECTOMETE FO DISCIMINATION OF VAENCE STATES I.A.Brytov,.I.Plotnikov,B.D.Kalinin,

More information

PERFORMANCE OF A ROOM TEMPERATURE GAS PROPORTIONAL SCINTILLATION COUNTER IN X-RAY ANALYSIS OF METALLIC ALLOYS EXCITED WITH ALPHA PARTICLES

PERFORMANCE OF A ROOM TEMPERATURE GAS PROPORTIONAL SCINTILLATION COUNTER IN X-RAY ANALYSIS OF METALLIC ALLOYS EXCITED WITH ALPHA PARTICLES 249 PERFORMANCE OF A ROOM TEMPERATURE GAS PROPORTIONAL SCINTILLATION COUNTER IN X-RAY ANALYSIS OF METALLIC ALLOYS EXCITED WITH ALPHA PARTICLES F. I. G. M. Borges, S. J. C. do Carmo, T. H. V. T. Dias, F.

More information

Outline. Introduction, motivation Readout electronics, Peltier cooling Input J-FETsJ

Outline. Introduction, motivation Readout electronics, Peltier cooling Input J-FETsJ Progress in low energy X-rayX spectroscopy using semi-insulating insulating GaAs detectors F. Dubecký 1, B. Zaťko 1, P. Boháček 1, L. Ryć 2, E. Gombia 2, and V. Nečas 3 1 IEE SAS, Bratislava, Slovakia

More information

3.1 Introduction to Semiconductors. Y. Baghzouz ECE Department UNLV

3.1 Introduction to Semiconductors. Y. Baghzouz ECE Department UNLV 3.1 Introduction to Semiconductors Y. Baghzouz ECE Department UNLV Introduction In this lecture, we will cover the basic aspects of semiconductor materials, and the physical mechanisms which are at the

More information

SEM. Chemical Analysis in the. Elastic and Inelastic scattering. Chemical analysis in the SEM. Chemical analysis in the SEM

SEM. Chemical Analysis in the. Elastic and Inelastic scattering. Chemical analysis in the SEM. Chemical analysis in the SEM THE UNIVERSITY Chemical Analysis in the SEM Ian Jones Centre for Electron Microscopy OF BIRMINGHAM Elastic and Inelastic scattering Electron interacts with one of the orbital electrons Secondary electrons,

More information

A dual scintillator - dual silicon photodiode detector module for intraoperative gamma\beta probe and portable anti-compton spectrometer

A dual scintillator - dual silicon photodiode detector module for intraoperative gamma\beta probe and portable anti-compton spectrometer University of Wollongong Research Online Faculty of Engineering - Papers (Archive) Faculty of Engineering and Information Sciences 2008 A dual scintillator - dual silicon photodiode detector module for

More information

Investigation of the Asymmetric Characteristics and Temperature Effects of CdZnTe Detectors

Investigation of the Asymmetric Characteristics and Temperature Effects of CdZnTe Detectors 2068 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 5, OCTOBER 2005 Investigation of the Asymmetric Characteristics and Temperature Effects of CdZnTe Detectors Benjamin W. Sturm, Student Member, IEEE,

More information

Edgeless sensors for full-field X-ray imaging

Edgeless sensors for full-field X-ray imaging Edgeless sensors for full-field X-ray imaging 12 th iworid in Cambridge July 14 th, 2010 Marten Bosma 12 th iworid, Cambridge - July 14 th, 2010 Human X-ray imaging High spatial resolution Low-contrast

More information

Chapter 8 Hyper-Pure Germanium Detector

Chapter 8 Hyper-Pure Germanium Detector Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-1 Chapter 8 Hyper-Pure Germanium Detector 8.1. Introduction Silicon semiconductor detectors described in the previous chapter have depletion

More information

Horst Ebel, Robert Svagera, Christian Hager, Maria F.Ebel, Christian Eisenmenger-Sittner, Johann Wernisch, and Michael Mantler

Horst Ebel, Robert Svagera, Christian Hager, Maria F.Ebel, Christian Eisenmenger-Sittner, Johann Wernisch, and Michael Mantler DETECTION OF SUBMONOLAYERS BY MEASUREMENT OF THE TOTAL ELECTRON YIELD (TEY) OF X-RAY EXCITED ELECTRON EMISSION Horst Ebel, Robert Svagera, Christian Hager, Maria F.Ebel, Christian Eisenmenger-Sittner,

More information

Chapter 8 Hyper-Pure Germanium Detector

Chapter 8 Hyper-Pure Germanium Detector Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 8-1 8.1. Introduction Chapter 8 Hyper-Pure Germanium Detector Silicon semiconductor detectors described in the previous chapter have depletion

More information

Position Sensitive Germanium Detectors for the Advanced Compton Telescope

Position Sensitive Germanium Detectors for the Advanced Compton Telescope Position Sensitive Germanium Detectors for the Advanced Compton Telescope R.A. Kroeger 1, W.N. Johnson 1, J.D. Kurfess 1, B.F. Phlips, P.N. Luke 3, M. Momayezi 4, W.K. Warburton 4 1 Naval Research Laboratory,

More information

Radioactivity. Lecture 6 Detectors and Instrumentation

Radioactivity. Lecture 6 Detectors and Instrumentation Radioactivity Lecture 6 Detectors and Instrumentation The human organs Neither humans nor animals have an organ for detecting radiation from radioactive decay! We can not hear it, smell it, feel it or

More information

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 19 Chapter 12 There are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry X-ray Techniques include:

More information

Development of High-Z Semiconductor Detectors and Their Applications to X-ray/gamma-ray Astronomy

Development of High-Z Semiconductor Detectors and Their Applications to X-ray/gamma-ray Astronomy Development of High-Z Semiconductor Detectors and Their Applications to X-ray/gamma-ray Astronomy Taka Tanaka (SLAC/KIPAC) 9/19/2007 SLAC Advanced Instrumentation Seminar Outline Introduction CdTe Diode

More information

Chapter 3 Gas Filled Detectors

Chapter 3 Gas Filled Detectors Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 3-1 Chapter 3 Gas Filled Detectors 3.1. Ionization chamber A. Ionization process and charge collection The interactions of charged particles

More information

FACTORS AFFECTING IN-LINE PHASE CONTRAST IMAGING WITH A LABORATORY MICROFOCUS X-RAY SOURCE

FACTORS AFFECTING IN-LINE PHASE CONTRAST IMAGING WITH A LABORATORY MICROFOCUS X-RAY SOURCE Copyright JCPDS-International Centre for Diffraction Data 26 ISSN 197-2 FACTORS AFFECTING IN-LINE PHASE CONTRAST IMAGING WITH A LABORATORY MICROFOCUS X-RAY SOURCE 31 K. L. Kelly and B. K. Tanner Department

More information

Chemical Analysis in TEM: XEDS, EELS and EFTEM. HRTEM PhD course Lecture 5

Chemical Analysis in TEM: XEDS, EELS and EFTEM. HRTEM PhD course Lecture 5 Chemical Analysis in TEM: XEDS, EELS and EFTEM HRTEM PhD course Lecture 5 1 Part IV Subject Chapter Prio x-ray spectrometry 32 1 Spectra and mapping 33 2 Qualitative XEDS 34 1 Quantitative XEDS 35.1-35.4

More information

CHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE

CHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE CHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE 218 Chris M. Sparks 1, Elizabeth P. Hastings 2, George J. Havrilla 2, and Michael Beckstead 2 1. ATDF,

More information

InAs avalanche photodiodes as X-ray detectors

InAs avalanche photodiodes as X-ray detectors Home Search Collections Journals About Contact us My IOPscience InAs avalanche photodiodes as X-ray detectors This content has been downloaded from IOPscience. Please scroll down to see the full text.

More information

A monolithic array of silicon drift detectors coupled to a single scintillator for γ-ray imaging with sub-millimeter position resolution

A monolithic array of silicon drift detectors coupled to a single scintillator for γ-ray imaging with sub-millimeter position resolution 1 A monolithic array of silicon drift detectors coupled to a single scintillator for γ-ray imaging with sub-millimeter position resolution C.Fiorini a*, A.Longoni a, F.Perotti b, C.Labanti c, E.Rossi c,

More information

Introduction to Semiconductor Physics. Prof.P. Ravindran, Department of Physics, Central University of Tamil Nadu, India

Introduction to Semiconductor Physics. Prof.P. Ravindran, Department of Physics, Central University of Tamil Nadu, India Introduction to Semiconductor Physics 1 Prof.P. Ravindran, Department of Physics, Central University of Tamil Nadu, India http://folk.uio.no/ravi/cmp2013 Review of Semiconductor Physics Semiconductor fundamentals

More information

GEM: A new concept for electron amplification in gas detectors

GEM: A new concept for electron amplification in gas detectors GEM: A new concept for electron amplification in gas detectors F. Sauli, Nucl. Instr. & Methods in Physics Research A 386 (1997) 531-534 Contents 1. Introduction 2. Two-step amplification: MWPC combined

More information

n N D n p = n i p N A

n N D n p = n i p N A Summary of electron and hole concentration in semiconductors Intrinsic semiconductor: E G n kt i = pi = N e 2 0 Donor-doped semiconductor: n N D where N D is the concentration of donor impurity Acceptor-doped

More information

Diffractometer. Geometry Optics Detectors

Diffractometer. Geometry Optics Detectors Diffractometer Geometry Optics Detectors Diffractometers Debye Scherrer Camera V.K. Pecharsky and P.Y. Zavalij Fundamentals of Powder Diffraction and Structural Characterization of Materials. Diffractometers

More information

Alpha-particle Stopping Powers in Air and Argon

Alpha-particle Stopping Powers in Air and Argon Alpha-particle Stopping Powers in Air and Argon Mohammad QH 1* and Maghdid HA 2 1 Department of Physics, College of Science, Raparin University, Sulaimanyah, Iraqi Kurdistan, Iraq 2 Department of Physics,

More information

EE495/695 Introduction to Semiconductors I. Y. Baghzouz ECE Department UNLV

EE495/695 Introduction to Semiconductors I. Y. Baghzouz ECE Department UNLV EE495/695 Introduction to Semiconductors I Y. Baghzouz ECE Department UNLV Introduction Solar cells have always been aligned closely with other electronic devices. We will cover the basic aspects of semiconductor

More information

Radionuclide Imaging MII Detection of Nuclear Emission

Radionuclide Imaging MII Detection of Nuclear Emission Radionuclide Imaging MII 3073 Detection of Nuclear Emission Nuclear radiation detectors Detectors that are commonly used in nuclear medicine: 1. Gas-filled detectors 2. Scintillation detectors 3. Semiconductor

More information

THE spectroscopic performance of large volume CdZnTe

THE spectroscopic performance of large volume CdZnTe 3098 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 6, DECEMBER 2004 Analysis of Detector Response Using 3-D Position-Sensitive CZT Gamma-Ray Spectrometers Feng Zhang, Student Member, IEEE, Zhong He,

More information

Chemical Engineering 412

Chemical Engineering 412 Chemical Engineering 412 Introductory Nuclear Engineering Lecture 26 Radiation Detection & Measurement II Spiritual Thought 2 I would not hold the position in the Church I hold today had I not followed

More information

Advances in the Micro-Hole & Strip Plate gaseous detector

Advances in the Micro-Hole & Strip Plate gaseous detector Nuclear Instruments and Methods in Physics Research A 504 (2003) 364 368 Advances in the Micro-Hole & Strip Plate gaseous detector J.M. Maia a,b,c, *, J.F.C.A. Veloso a, J.M.F. dos Santos a, A. Breskin

More information

Development of Gamma-ray Monitor using CdZnTe Semiconductor Detector

Development of Gamma-ray Monitor using CdZnTe Semiconductor Detector Development of Gamma-ray Monitor using CdZnTe Semiconductor Detector A. H. D. Rasolonjatovo 1, T. Shiomi 1, T. Nakamura 1 Y. Tsudaka 2, H. Fujiwara 2, H. Araki 2, K. Matsuo 2, H. Nishizawa 2 1 Cyclotron

More information

Practical 1P4 Energy Levels and Band Gaps

Practical 1P4 Energy Levels and Band Gaps Practical 1P4 Energy Levels and Band Gaps What you should learn from this practical Science This practical illustrates some of the points from the lecture course on Elementary Quantum Mechanics and Bonding

More information

Lecture 18. New gas detectors Solid state trackers

Lecture 18. New gas detectors Solid state trackers Lecture 18 New gas detectors Solid state trackers Time projection Chamber Full 3-D track reconstruction x-y from wires and segmented cathode of MWPC z from drift time de/dx information (extra) Drift over

More information

Semiconductor Detectors are Ionization Chambers. Detection volume with electric field Energy deposited positive and negative charge pairs

Semiconductor Detectors are Ionization Chambers. Detection volume with electric field Energy deposited positive and negative charge pairs 1 V. Semiconductor Detectors V.1. Principles Semiconductor Detectors are Ionization Chambers Detection volume with electric field Energy deposited positive and negative charge pairs Charges move in field

More information

INFLUENCE OF GROWTH INTERRUPTION ON THE FORMATION OF SOLID-STATE INTERFACES

INFLUENCE OF GROWTH INTERRUPTION ON THE FORMATION OF SOLID-STATE INTERFACES 122 INFLUENCE OF GROWTH INTERRUPTION ON THE FORMATION OF SOLID-STATE INTERFACES I. Busch 1, M. Krumrey 2 and J. Stümpel 1 1 Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany

More information

Solid State Detectors

Solid State Detectors Solid State Detectors Most material is taken from lectures by Michael Moll/CERN and Daniela Bortoletto/Purdue and the book Semiconductor Radiation Detectors by Gerhard Lutz. In gaseous detectors, a charged

More information

Practical 1P4 Energy Levels and Band Gaps

Practical 1P4 Energy Levels and Band Gaps Practical 1P4 Energy Levels and Band Gaps What you should learn from this practical Science This practical illustrates some of the points from the lecture course on Elementary Quantum Mechanics and Bonding

More information

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters ) September 17, 2018 Reference literature (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters 13-14 ) Reference.: https://slideplayer.com/slide/8354408/ Spectroscopy Usual Wavelength Type of Quantum

More information

CVD Diamond History Introduction to DDL Properties of Diamond DDL Proprietary Contact Technology Detector Applications BDD Sensors

CVD Diamond History Introduction to DDL Properties of Diamond DDL Proprietary Contact Technology Detector Applications BDD Sensors Diamond Detectors CVD Diamond History Introduction to DDL Properties of Diamond DDL Proprietary Contact Technology Detector Applications BDD Sensors Kevin Oliver CEO Alex Brown Sales & Marketing 20 May,

More information

New perspectives in X-ray detection of concealed illicit materials brought by CdTe/CdZnTe spectrometric detectors

New perspectives in X-ray detection of concealed illicit materials brought by CdTe/CdZnTe spectrometric detectors New perspectives in X-ray detection of concealed illicit materials brought by CdTe/CdZnTe spectrometric detectors Jean-Marc Dinten, Jean-Louis Amans, Loïck Verger, Olivier Peyret CEA-LETI, MINATEC, Recherche

More information

IMPLEMENTATION OF THE MONTE CARLO-LIBRARY LEAST- SQUARES APPROACH TO ENERGY DISPERSIVE X-RAY FLUORESCENCE ANALYSIS

IMPLEMENTATION OF THE MONTE CARLO-LIBRARY LEAST- SQUARES APPROACH TO ENERGY DISPERSIVE X-RAY FLUORESCENCE ANALYSIS 227 IMPLEMENTATION OF THE MONTE CARLO-LIBRARY LEAST- SQUARES APPROACH TO ENERGY DISPERSIVE X-RAY FLUORESCENCE ANALYSIS Fusheng Li, Weijun Guo, and Robin P. Gardner Center for Engineering Applications of

More information

New trends in CdTe detectors for X and γ-ray applications

New trends in CdTe detectors for X and γ-ray applications New trends in CdTe detectors for X and γ-ray applications Olivier Limousin CEA Saclay / DSM / DAPNIA Service d Astrophysique France New developments in photodetection, Beaune 2002 / Solid state detectors

More information

Chapter Seven (Nuclear Detectors)

Chapter Seven (Nuclear Detectors) Al-Mustansiriyah University College of Science Physics Department Fourth Grade Nuclear Physics Dr. Ali A. Ridha Chapter Seven (Nuclear Detectors) Ionizing radiation is rarely detected directly. Instead,

More information

ION-EXCHANGE FILMS FOR ELEMENT CONCENTRATION IN X-RAY FLUORESCENCE ANALYSIS WITH TOTAL REFLECTION OF THE PRIMARY BEAM.

ION-EXCHANGE FILMS FOR ELEMENT CONCENTRATION IN X-RAY FLUORESCENCE ANALYSIS WITH TOTAL REFLECTION OF THE PRIMARY BEAM. 822 ION-EXCHANGE FILMS FOR ELEMENT CONCENTRATION IN X-RAY FLUORESCENCE ANALYSIS WITH TOTAL REFLECTION OF THE PRIMARY BEAM. Abstract A.P.Morovov, L.D.Danilin, V.V.Zhmailo, Yu.V.Ignatiev, A.E.Lakhtikov,

More information

Status Report: Charge Cloud Explosion

Status Report: Charge Cloud Explosion Status Report: Charge Cloud Explosion J. Becker, D. Eckstein, R. Klanner, G. Steinbrück University of Hamburg Detector laboratory 1. Introduction and Motivation. Set-up available for measurement 3. Measurements

More information

Defining quality standards for the analysis of solid samples

Defining quality standards for the analysis of solid samples Defining quality standards for the analysis of solid samples Thermo Scientific Element GD Plus Glow Discharge Mass Spectrometer Redefine your quality standards for the elemental analysis of solid samples

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement June 2008 Tom Lewellen Tkldog@u.washington.edu Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron

More information