Operating Instructions. Titration Excellence Terminal

Size: px
Start display at page:

Download "Operating Instructions. Titration Excellence Terminal"

Transcription

1 Operating Instructions Titration Excellence Terminal

2

3 Table of Contents Contents Page 1. Introduction 1 2. Functional Description Layout of the Terminal Operating the Touchscreen The User Interface The Home Screen Entering Data in the User Interface Sorting Lists The Task Menu (T70/T90 only) Shortcuts and Direct Shortcuts The Analysis Start Screen The Online Screen 9 3. Setup (Setup) Overview of the Menu Structure for the Setup Menu Chemicals Titrant Auxiliary Reagents Calibration Standard Titer Standards Substances Hardware Sensors Pumps Peripherals Titration Stands Auxiliary Instruments User Settings Language Display Sounds Shortcuts Keyboards Global Settings System User Management Performance of Analyses and Resources Values Blanks Auxiliary Values Service & Maintenance MT-Service Import/Export Resetting Factory Settings Software History Firmware Terminal Board Data Drives Burettes Upgrade (only for T70) Update Monitoring the Expiration Date and Life Span of Resources Monitoring the Expiration Date of a Resource Monitoring the Life Span of a Resource 59 METTLER TOLEDO T50 / T70 / T90 I

4 Table of Contents 3.9 Sensor Calibration Linear Calibration Segmented Calibration Methods Starting Methods Starting Methods from the Method Editor Establishing Methods Modifying Methods Deleting Methods Method Syntax Rules for Establishing Methods Possible Number of Method Functions Types of Methods Types and Possible Number of Loops Entering and Deleting Loops Method Functions Within a Loop Method Functions Outside of a Loop Overview of Method Functions Method Function Settings 73 Title 73 Sample 74 Sample (titer) 75 Sample (calibration) 75 Titration stand 76 Rinse 78 Conditioning 79 Pump 80 Park 80 Stir 81 Dispense (normal) 81 Calculation 82 Drain 84 Record 85 End of sample 86 Titer 86 Calibration 87 Auxiliary value 88 Blank value 89 Auxiliary instrument 90 Instruction 93 Measure (normal) 95 Measure (MVT) 98 Titration (EQP) 100 Titration (EP) 105 Titration (2-phase) 109 Titration (EQP detection) 114 Stating 116 Dispense (controlled) Value Ranges from Sensor Measuring Units and Control Band Manual operations (Manual) Stirrer SENSOR Temperature sensor Potentiometric Sensor Polarized Sensor Conductivity Sensor Burette 129 II METTLER TOLEDO T50 / T70 / T90

5 Table of Contents Rinse burette Dispense Manual titration Pump Auxiliary Instrument Sample changer Series templates Sample series Series sequence [T90] Free sample series Sample parameters Processing sequence of an analysis Start of an analysis Processing sequence of an analysis Results All results Add result Statistics Outlier test Recalculate Samples Reevaluate Undo changes Delete all results Evaluate and calculate Analysis data Indexing of method functions Naming conventions for using analysis data in calculations Formulas Mathematical Functions and Operators Using analysis data in formulas Explanatory examples Titration method function Stating method function Titrator evaluation procedure Standard Evaluation Procedure Minimum/Maximum Segmented Asymmetrical Sample formulas Constants within a Content Calculation Appendix Predefined Calibration Standards METTLER Methods Application Modes Method Templates 191 Index 193 METTLER TOLEDO T50 / T70 / T90 III

6 Table of Contents IV METTLER TOLEDO T50 / T70 / T90

7 Introduction 1. Introduction Titration excellence simple, efficient, secure The instruments in METTLER TOLEDO's Titration Excellence Line are modern, modular titrators for use in a wide range of applications. They can be used in quality control as well as in research and development. With their wide array of applications, they can satisfy the most stringent requirements. The titrators in the Titration Excellence Line combine simple, easy-to-use functions with the most flexible, outstanding analytical efficiency. Rapid titrant change is made easy thanks to a new procedure that detects titrant automatically (Burette Plug & Play PnP). The titrator automatically recognizes the titrant needed without the need for any action by the user. Even installing sample changers and additional dosing units makes manual adjustments superfluous. Titration Excellence instruments can be controlled by means of a terminal and/or LabX PC software. The terminal is a large color touch screen that allows for intuitive user guidance that can be adjusted to each user. All functions can be accessed directly from the main screen via customizable shortcuts which, when used on a daily basis, lighten the workload considerably. Operating the titrator using a terminal is described in detail in this operating instructions manual. The various titrator operating instructions available when using LabX PC software is illustrated in separate operating instructions entitled "LabX for Titration Excellence". The following separate installation information describes all of the steps necessary to install and begin operating your Titration Excellence instrument. The "Quick Guide" will then guide you through the first titration process using a practical example. If you have any additional questions, METTLER TOLEDO is always available to assist you. METTLER TOLEDO T50 / T70 / T90 1

8 Introduction 2 METTLER TOLEDO T50 / T70 / T90

9 Functional Description 2. Functional Description 2.1 Layout of the Terminal The control panel of the terminal consists of an integrated touchscreen (d) and four additional buttons (a, b and 2 x c), which are located outside of the touch-sensitive interface of the display: The Reset button (a) ends all tasks that are currently running. The Info button (b) activates the interactive online help for the contents of the current screen. Two Home buttons (c) always take you back to the Home screen (to the left and right of the display). You can press these four buttons any time, regardless of which screen you are currently in. The Reset button is a kind of "EMERGENCY OFF" switch. If the titrator malfunctions or there is an operating error, you can stop all current tasks by pressing the reset button. Afterward, for each task, you can decide whether to end it completely or continue. a b d c METTLER TOLEDO T50 / T70 / T90 3

10 Functional Description 2.2 Operating the Touchscreen With the exception of the four buttons, the titrator is operated exclusively using the integrated touchscreen and the On/Off button on the titrator. Simply touch the touch-sensitive surface of the display with your fingertip or fingernail to select a button or entry element. Never touch the surface of the touchscreen with pointed or sharp objects! Doing so can damage the touchscreen! 2.3 The User Interface The graphical user interface consists of five basic elements (a-e). The title bar (a) at the top of the display indicates the name of the current screen. The task button, located in the top right corner, uses animation to indicate whether a task is running. You can access the task menu anytime by pressing the task button, which displays an overview of all tasks currently running (with the T50, or when only one task is operating, you can get to the current task directly by pressing the task button). From the task menu, you can jump to any task currently running. The navigation bar, located below the title bar, displays the path to the current screen. The scroll bar (d), located on the right side of the screen, will be visible as soon as the contents of the screen are beyond the viewable range. If this occurs, use either the arrows or the area in between them to move the viewable area of the screen up or down. (A maximum of) five blue softkeys (e) are located at the bottom of the display. The function of these buttons varies and depends on the context of the current menu. 4 METTLER TOLEDO T50 / T70 / T90

11 Functional Description The Home Screen The Home screen is the first screen that appears after starting the titrator. You can come back to this screen at anytime by pressing the Home button. The Home screen contains five buttons (c), which take you to the individual submenus: Methods: This button takes you to the Method Editor where you can create and manage methods. (See "4. Methods") Series Templates: In this menu, series of individual samples can be created and managed, e.g. when using a sample changer. (See "6. Series Templates") Results: Manage the results of your analysis here. (See "8. Results") Setup: The hardware and resources the titrator uses are configured in the Setup. Both global and user settings are also assigned here. (See "3. Setup") Manual: This button takes you to manual operations. Stirrers, sensors, pumps, etc. can be operated here independently from analyses. (See "5. Manual Operations (Manual)") There is yet another area (b) that can be configured for every user (with the necessary authorization). 12 shortcuts per user can be stored here. By using these shortcuts (a), defined methods, series and manual operations can be started directly from the Home screen. (See "2.3.5 Shortcuts and Direct Shortcuts") METTLER TOLEDO T50 / T70 / T90 5

12 Functional Description Entering Data in the User Interface. There are different types of input fields in the user interface where you can either enter data or select data from a list. Input fields can also be deactivated and their contents are then displayed as information only and cannot be changed in the corresponding menu. The various types of input fields are identified by an icon to the right of the screen: Text input fields Number input fields Drop-down lists List fields Menu fields Formula fields Deactivated (Info field) Any text comprised of letters, numbers and symbols can be entered into these fields. Numbers, formulas and auxiliary values can be entered into these fields. In some fields an auxiliary value can be selected from the Auxiliary Values list by pressing the "H" button. By selecting these fields, a drop-down list will open from which you can select an entry. By selecting these fields a menu list will open in a new window. By selecting these fields, a new screen will open with a number of additional menu options. A formula must be entered in these fields. You can enter these manually or access a menu list by pressing the "Suggestion" softkey. The values in deactivated input fields are displayed as information only and cannot be edited in the corresponding menu. Deactivated checkbox (No) Activated checkbox (Yes) In addition to the input fields there are checkboxes that can be checked in order to select certain functionalities. Checkboxes can affect the scope of the corresponding menu, i.e. input fields can be hidden or visible depending on whether the checkbox is checked. Limits for the input fields (number of characters for text input fields, minimum and maximum values, and number of decimal points for number input fields) are entered into the respective entry display screen: 6 METTLER TOLEDO T50 / T70 / T90

13 Functional Description Sorting Lists All of the lists displayed in the user interface can be sorted by column alphabetically or numerically in ascending or descending order. To do this, simply touch the header row of the list (b) in the column by which you would like to sort the list. A little arrow (a) in the header row indicates the column by which the list is sorted and whether it is sorted in ascending or descending order The Task Menu (T70/T90 only) The task menu is accessed via the task button (in the top right corner of the display screen) and provides an overview of all tasks that are currently running. Selecting one of the individual tasks from the list will take you back to the respective online screen where you can end the task. METTLER TOLEDO T50 / T70 / T90 7

14 Functional Description Shortcuts and Direct Shortcuts Shortcuts allow you to start methods, series and manual operations from the homepage. You can place up to twelve different shortcuts on the homepage by pressing the "AddToHome" softkey button. "AddToHome" softkeys are located on the methods, series and manual operations start screens. Shortcuts are user-specific, i.e. each individual user can create a maximum of twelve shortcuts for the tasks they personally conduct the most with the titrator. The titrator supports two types of shortcuts. Direct shortcuts which, when selected, start immediately without warning (only if the other settings allow this), and normal shortcuts which take you to the corresponding start screen from which you can start the task. Shortcuts for methods, series or manual operations that bring you the corresponding start screen. Shortcuts for methods, series or manual operations with integrated reference symbols in the icon that start the corresponding task when selected provided the other settings allow for it. Shortcuts are managed in Setup, under the subcategory "User Settings". This is where shortcuts can be deleted or modified, and their position on the Home screen changed. (See "3.3.4 Shortcuts") As soon as the maximum number of (12) shortcuts have been created on the Home screen, the AddToHome button in the start screen for methods, series and manual operations will be deactivated. Caution: Tasks started using the shortcuts can begin immediately without warning. Therefore, always make sure that all tubes are connected to suitable vessels prior to using a shortcut. For an explanation of the settings available when creating a shortcut, see "7.1 Starting an Analysis" The Analysis Start Screen An analysis, whether a single or multiple determination, can be started on the titrator in a number of ways: By using the "Start" softkey in the Method Editor By using the "Start" softkey on the Home screen By using a shortcut (or direct shortcut) on the Home screen By using the "Start" softkey in the "Series" menu By using the "Calibrate" or "Titer" softkeys in the "Setup" menu (to perform a calibration or titer determination) The analysis start screen is always the first screen that appears after using a softkey ("Start", "Calibrate" or "Titer") or a shortcut. 8 METTLER TOLEDO T50 / T70 / T90

15 Functional Description When a direct shortcut is activated, the analysis start screen will not appear and the respective method will start immediately, provided that the other settings allow this. The parameters for the previously used method or series will appear in the analysis start screen so that the same method can immediately be started again. Of course, all of the settings can also be adjusted prior to pressing the "Start" softkey. The type and number of settings displayed in the analysis start screen depends on the type of analysis to be started and the resources used. For an explanation of the settings available when starting an analysis from the analysis start screen, see "7.1 Starting an Analysis" The Online Screen The online screen is displayed when an analysis or manual operation is being performed. The method ID of the current method or the type of manual operation is displayed in the title bar. In the navigation bar below, the sample index, displayed as "Sample 2/5" (second of a total of five samples) and loop index, displayed as "Loop 1/3" (first of three loops) are shown. (The loop index will only be displayed if the method actually contains more than one loop (available only with T70 and T90)). The navigation path will be displayed in the navigation bar while a manual operation is being performed. METTLER TOLEDO T50 / T70 / T90 9

16 Functional Description The remainder of the online screen is divided into a graphical area (left) and a data area (right). The measured curve is displayed in the graphical area during a titration or measurement in which the unit to be displayed on each axis can be chosen from a predefined selection. In addition to the online screen, you can also display a table of measured values during the titration or measurement. Depending on the type of method function being performed, the online screen displays various data such as measured values, remaining run times, volume dispensed, stirrer speed, duration of the analysis or temperature. 10 METTLER TOLEDO T50 / T70 / T90

17 Setup 3. Setup (Setup) In this chapter you will learn how to adjust the titrator settings according to your requirements. In order to perform a titration, the resources that are used in the method (titrants, sensors, peripherals, etc.) must be installed and configured. The setup, which helps you install these resources, is divided into the following areas: Setup area Chemicals Hardware User settings Global settings Values Service & maintenance Description All types of chemicals such as titrants, auxiliary reagents and standards for titer determination and calibration standards can be defined here. In this menu you can define the hardware components of your titration system, such as sensors, titration stands, pumps and auxiliary instruments. User-defined settings, such as language, display settings or shortcuts can be assigned here. System settings can be established here that apply to every user. Blank and auxiliary values can be specified under "Values". Here the user can assign actions to be taken regarding the servicing and maintenance of the titrator. The "Expired resources" button On the overview screen of the setup menu you will find the "Expired resources" button. By pressing this button you go to an overview of all expired resources with the type, name and date of expiry of the respective resource. Resources that were selected during setup in "Monitoring expiration dates" and/or "Monitoring life span" may be expired. See "3.8 Monitoring the Expiration Dates and Life Span of Resources". METTLER TOLEDO T50 / T70 / T90 11

18 Setup 3.1 Overview of the Menu Structure for the Setup Menu Chemicals Titrants Auxiliary reagents Calibration standards Titer standards Substances Hardware Sensors Pump Peripherals Titration stands Auxiliary instruments User settings Language Screen Sounds Shortcuts Keyboards Global settings System User management Performance of analyses and resources Values Blanks Auxiliary values Service & maintenance MT-Service Import/Export Reset to factory settings Software history Firmware Terminal Board data Drives Burettes Upgrade Update 12 METTLER TOLEDO T50 / T70 / T90

19 Setup 3.2 Chemicals All chemicals such as titrants, titer and calibration standards, auxiliary reagents and other substances are configured and managed under the menu, "Chemicals". Lists of the chemicals already defined in the system can be viewed and printed at any time, and new chemicals can be defined and existing ones can be deleted. Auxiliary reagents must be assigned to a pump to which they can be added, and titrants (regardless of type) must each be assigned to a drive Titrant The following settings can be edited and viewed for each titrant: Parameter Explanation Value range Type The type of titrant. Either "General titration" for a classic titrant or "Auxiliary reagent" if you want to use a burette to dispense an auxiliary reagent. Name Select any name for the titrant. Arbitrary Concentration Concentration Titer Determination method Date/Time Performed by Monitoring expiration dates Monitoring life span The concentration of the titrant, in [mol/l]. (For "type" = "general titration" only.) The non-dimensional concentration of an auxiliary reagent. (For "type" = "auxiliary reagent" only.) The titer for the titrant. (For "type" = "general titration" only.) Contains eather the entry "Manual" if the value for the titer was entered manually or the ID of the method used to determine the titer. (For "type" = "general titration" only.) (Info field) Either the date and time the titer was entered manually or the date and time the titer was determined by means of a method. (For "type" = "general titration" only.) (Info field) Name of the user who entered or determined the titer. (For "type" = "general titration" only.) (Info field) Specifies whether to monitor the expiration date during which the titer is used. (See "3.8.1 Monitoring the Expiration Date of a Resource") (For "type" = "general titration" only.) Specifies whether to monitor the life span of the titrant. (See "3.8.2 Monitoring the Life Span of a Resource") (For "type" = "general titration".) General titration Auxiliary reagent Manual Method ID Date/Time User name LabX operator Lot/Batch The lot or batch of the titrant. Enter any designation. Arbitrary Fill rate The filling rate of the burette in percent. 100% stands for maximum filling rate. Burette volume Drive The volume of the burette. Choices are 1, 5, 10 and 20 ml. Defines the drive on which you will use the burette containing the titrant. You can select the "PnP" entry for available but unused PnP burettes PnP METTLER TOLEDO T50 / T70 / T90 13

20 Setup Serial number You can enter the serial number of the burette containing the titrant here. For PnP burettes the system will enter this number automatically, but you will be able to change it. Arbitrary A maximum of 100 titrants can be defined in the titrator Auxiliary Reagents Auxiliary reagents are liquid chemicals that can be used to aid the titration process. Auxiliary reagents must be added using a pump and can be used via the method functions "Pumps" and "Rinsing". The following settings can be viewed and edited: Parameter Explanation Value range Name Select any name for the auxiliary reagent. Arbitrary Pump Pump output You can use this setting to select a pump from the pump list in the setup. The pump output assigned to the selected pump. (As defined in the pump setup.) (Info field) A maximum of 50 auxiliary reagents can be defined in the titrator. Pump list List of available pump outputs. 14 METTLER TOLEDO T50 / T70 / T90

21 Setup Calibration Standard In addition to the predefined calibration standard lists for ph and conductivity sensors shown in the appendix, user-defined calibration standard lists can also be managed. In this menu, you can view and print the predefined lists stored in the titrator, and create additional userdefined calibration standard lists for ph buffers and ISE and conductivity standards. First you have to create a new calibration standard list. The following settings are available for this: Parameter Explanation Value range Type Name Unit Base list Ref. temperature Defines the type of calibration standard list to be created. The type will depend on the type of the sensor to be calibrated. For a calibration list of the "Auto ph" type, the titrator will identify the standards automatically. The type selected will define the value range for the "unit of measure" parameter. Select any name for the new calibration standard list. Enter the unit of measure for the calibration standard, depending on the type of the calibration standard list. To create an auto ph calibration standards list, you must select suitable ph buffers from an existing base list. The reference temperature in [ C] that all the standards or ph buffers of a calibration standard list will reference. ph Auto ph ISE Conductivity Arbitrary ph (for "ph" and "auto-ph" types) pm px ppm (for "ISE" type) ms/cm µs/cm (for "conductivity" type) Calibration standard-list After you have created a calibration standard list, you can add various buffers and standards to this list, depending on the type selected: For a calibration standard list for "ph", see " Adding a ph Calibration Standard (ph Buffer)" For a calibration standard list for "auto-ph", see " Adding a ph Calibration Standard (ph Buffer) to an Auto-pH Calibration Standard List" For a calibration standard list for "ISE", see " Adding an ISE Calibration Standard (ISE Standard)" For a calibration standard list for "conductivity", see " Adding a Conductivity Calibration Standard (Conductivity Standard)" To delete a user-defined calibration standard list from the titrator, you first have to view the settings of the list using the "Info" softkey. From this menu you can delete the calibration standard list from the titrator memory by selecting "delete". A maximum of 20 user-defined calibration standard lists and 10 auto-ph buffer lists can be defined in the titrator. METTLER TOLEDO T50 / T70 / T90 15

22 Setup Adding a ph Calibration Standard (ph Buffer) After you have created a "ph" calibration standard list, you can add various ph buffers to thid list by using "New". To do so, enter the respective ph value of the buffer, based on the reference temperature from the calibration standard list: Parameter Explanation Value range Ref. temperature ph value The reference temperature in [ C] that was defined when creating the calibration standard list. (Info field) The ph value of the ph buffer at the reference temperature To display the temperature influence of a ph buffer, you can enter a maximum of 20 variant pairs for each individual buffer composed of the temperature and corresponding ph value: Select a buffer and press "New". Parameter Explanation Value range Temperature ph value The temperature in [ C], for which the ph value of the ph buffer has been entered. The ph value of the ph buffer at the temperature defined with the setting "Temperature" This is how you can create a complete calibration standard list according to the lists already stored in the titrator Adding a ph Calibration Standard (ph Buffer) to an Auto ph Calibration Standard List For a calibration list of the "Auto ph" type, the titrator will identify the various ph buffer automatically. In order to ensure positive identification, the ph values of the individual solutions have to differ from each other by at least two points. You can add the calibration standard lists of various ph buffers by selecting them from the specified base list. By doing so, the titrator offers only suitable ph buffers in order to ensure that the selected ph buffers always differ from each other by at least two ph points. The temperature influence of the individual ph buffers is also be taken from the base list and can only be viewed but not changed Adding an ISE Calibration Standard (ISE Standard) After you have created an "ISE" calibration standard list, you can add this list to various ISE standards by using "New". To do this, enter the corresponding value for the standard in the desired unit of measure, based on the reference temperature from the calibration: Parameter Explanation Value range Ref. temperature Standard value The reference temperature in [ C] that was defined when creating the calibration standard list. (Info field) The value of the ISE standard at the reference temperature in the desired unit of measure METTLER TOLEDO T50 / T70 / T90

23 Setup To display the temperature influence of an ISE standard, you can enter a maximum of 20 variant pairs for each individual standard composed of the temperature and corresponding standard value: Select a buffer and press "New". Parameter Explanation Value range Temperature Standard value The temperature in [ C] for which the standard value is entered. The ph value of the ISE standard at the temperature defined with the setting "Temperature". Thus you can create a complete calibration standard list for ISE sensors Adding a Conductivity Calibration Standard (Conductivity Standard) (for pm and px) (for ppm) After you have created a "Conductivity" calibration standard list, you can add various conductivity standards to this list by using "New". To do this, enter the conductivity for each standard based on the reference temperature from the calibration standard list: Parameter Explanation Value range Ref. temperature Conductivity The reference temperature in [ C] that was defined when creating the calibration standard list. (Info field) The conductivity value of the standard in the unit selected with "Unit" (for ms/cm) (for µs/cm) To display the temperature influence of a conductivity standard, you can enter a maximum of 20 variant pairs for each individual standard composed of the temperature and corresponding conductivity value: Select a buffer and press "New". Parameter Explanation Value range Temperature Conductivity The temperature for which the conductivity of the standard has been specified. The conductivity value of the standard in the unit selected with "Unit" (for ms/cm) (for µs/cm) This is how you can create a complete calibration standard list for conductivity sensors. METTLER TOLEDO T50 / T70 / T90 17

24 Setup Titer Standards The titer standard required for determining the titer for the titrant you used can also be stored and managed in the titrator. The following parameters can be edited and viewed for each titer standard: Parameter Explanation Value range Type Defines whether it is a liquid or solid standard. Solid Liquid Name Select any name for the standard. Arbitrary Purity Concentration The purity of a solid standard, in percent. (For "type" = "solid" only.) The concentration of a liquid standard, in [mol/l]. (For "type" = "liquid" only.) m The molar mass of a solid standard, in [g/mol] Density Equivalent number The density of a liquid standard, in [g/ml]. (For "type" = "liquid" only.) The equivalent number z of the standard Lot/Batch The number of the standard. Arbitrary Date/Time Performed by Monitoring expiration dates The date and time the standard was created or changed (references the "Lot/Batch" setting). (Info field) Name of the user who entered or determined the titer. (Info field) Specifies whether to monitor the expiration date during which the same lot/batch is used. (See "3.8.1 Monitoring the Expiration Date of a Resource") (For "type" = "general titration" only.) A maximum of 50 titer standards can be defined in the titrator Substances Date/Time User name LabX operator In the menu "Substances", any chemical substances that are required for conducting the analysis can be managed using name, empirical formula, molecular weight and equivalent number. Parameter Explanation Value range Name Select any name for the substance. Arbitrary Empirical formula The total molecular formula of the substance. Arbitrary Molecular weight Equivalent number You can enter the molecular weight of the substance here. The equivalent number of the substance. 1 9 A maximum of 100 substances can be defined in the titrator METTLER TOLEDO T50 / T70 / T90

25 Setup 3.3 Hardware In this menu you can configure all of the hardware components connected to the titrator, such as: Sensors Pumps Peripherals (instruments such as printers or balances) Titration stands Auxiliary instruments (such as valves or filling level sensors) Sensors Sensors to be used with the titrator can be configured and managed in this menu. You can view and print the list of all sensors defined in the titrator, add and delete sensors, and change settings already stored in the titrator. The settings for an individual sensor can also be output to a printer. In addition, the corresponding method for sensor calibration can be accessed from here. A maximum of 50 sensors can be defined in the titrator. Each sensor is associated with a specific type. Each sensor type can deliver measured values in one or more units of measure. The following table provides information regarding which units of measure can be selected for a corresponding sensor type: Sensor type Default unit of measure Eligible units of measure mv Sensor mv mv ph sensor ph ph mv ISE sensor pm pm / px ppm mv Phototrode %T %T A mv Polarized sensor mv mv µa Temperature sensor C C K F Conductivity sensor µs/cm µs/cm ms/cm µs ms Changing the unit of measure for a sensor can cause the calibration parameters and expiration date settings to deactivate and be omitted, the titrator to change the calibration parameters (for temperature sensors), or it can cause the titrator to show another set of calibration parameters (for ISE sensors). METTLER TOLEDO T50 / T70 / T90 19

26 Setup The following settings are initially available when creating a new sensor in the titrator: Parameter Explanation Value range Sensor type The type of the sensor. mv ph Phototrode Polarized Temperature Conductivity ISE Additional settings are available when configuring a sensor depending on the sensor type selected: mv Sensors Parameter Explanation Value range Name Select any name for the sensor. Arbitrary Unit The unit of measure for the sensor to be defined. mv Sensor input The input to which you want to connect the sensor. AB1/SENSOR1 AB1/SEN- SOR2 (more depending on equipment) Serial number You can enter a serial number for the sensor here. Arbitrary Monitoring life span ph Sensors Specifies whether to monitor the life span of the sensor. (See "3.8.2 Monitoring the Life Span of a Resource") (For "type" = "general titration".) Parameter Explanation Value range Name Select any name for the sensor. Arbitrary Unit The unit of measure for the sensor to be defined. mv ph Sensor input The input to which you want to connect the sensor. AB1/SENSOR1 AB1/SENSOR2 (more depending on equipment) Serial number You can enter a serial number for the sensor here. Arbitrary Calibration Defines the calibration type. (Appears for "unit" = "ph" only.) Calibration type "Linear" Zero point Slope (TCalib) Slope (25.0 C) Slope The zero point of the ph sensor in the unit [ph]. (Zero point, slope and corresponding calibration temperature are required to convert the mv signal of the sensor to the "ph" unit.) (Appears for "unit" = "ph" only.) The slope of the sensor at the calibration temperature in [mv/ph]. (Appears for "unit" = "ph" only.) The slope of the ph sensor in [mv/ph], based on a temperature of 25 C. This field will update automatically every time the parameters "Slope [mv/ ph]" or "Calibration Temperature [ C]" are changed. (Appears for "unit" = "ph" only.) (Info field) The slope of the sensor in [%]. The titrator automatically calculates the number in this field using the "Slope [%] = Slope [mv/ph] (25 C)/ * 100" (Appears for "unit" = "ph" only.) (Info field) Linear Segmented METTLER TOLEDO T50 / T70 / T90

27 Setup Calibration type "segmented" Calibration type "segmented" Number of segments Zero point 1 Slope 1 (TCalib) Slope 1 (25 C) Slope 1 Seg. boundaries 1,2 [ph] Zero point 2 Slope 2 (TCalib) Slope 2 (25 C) Slope 2 [%] etc. Internal temperature sensor Calibration temperature Calibration method Date/Time This parameter is adjusted automatically by the calibration method used. All subsequent parameters will be adjusted to match the number of segments. The system records two numbers per segment (zero point and slope). x-1 segment boundaries are entered for x segments. The maximum possible calibration is for 9 points (resulting in 8 segments). (Appears for "unit" = "ph" only) (Info field) The zero point of the sensor for the 1st segment in [ph]. (Appears for "unit" = "ph" only.) (Info field) The sensor gradient in [mv/ph for the 1st segment at calibration temperature. (Appears for "unit" = "ph" only.) (Info field) The ph sensor gradient in [mv/ph] for the 1st Segment, umgerechnet auf eine Temperatur von 25 C. (Appears for "unit" = "ph" only.) (Info field) Slope 1 of the sensor in [%]. The titrator automatically calculates the number in this field using the "Slope 1 [%]=Slope [mv/ph] (25 C)/ * 100" (Appears for "unit" = "ph" only.) (Info field) Specifies the segment boundary between Segments 1 and 2 in [ph]. (Appears for "unit" = "ph" only.) (Info field) The zero point of the sensor for the 2nd segment in [ph]. (Appears for "unit" = "ph" only.) (Info field) The sensor gradient in [mv/ph] for the 2nd segment at calibration temperature. (Appears for "unit" = "ph" only.) The zero point of the sensor for the 2nd segment in [mv/ph], recalculated for a temperature of 25 C (Appears for "unit" = "ph" only.) (Info field) The slope of the sensor in [%]. The titrator automatically calculates the number in this field using the "Slope 2 [%] = Slope [mv/ph] (25 C)/ * 100" (Appears for "unit" = "ph" only.) (Info field) If the sensor has an internal temperature sensor, you can select "yes" here. In this case the system will automatically make an entry in the sensor setup for the internal temperature sensor. The temperature during calibration in [ C]. (Appears for "unit" = "ph" only.) (Info field for segmented calibration) The titrator fills this field in automatically. The entry is either filled in as "Manual" if manual changes were made to the calibration parameters, or it is filled in with the name of the method if the sensor was calibrated with the aid of the "Calibrate" method function. (Appears for "unit" = "ph" only.) (Info field) Date and time of the last calibration. (Appears for "unit" = "ph" only.) (Info field) Manual Name of the calibration method Date/Time METTLER TOLEDO T50 / T70 / T90 21

28 Setup Performed by The name of the user or LabX user logged in during the last calibration. (Appears for "unit" = "ph" only.) (Info field) Logged-in user LabX user Monitoring expiration dates Monitoring life span Specifies whether to monitor the expiration date of the sensor. (See "3.8.1 Monitoring the Expiration Date of a Resource") (For "type" = "general titration" only.) (Appears for "unit" = "ph" only.) Specifies whether to monitor the life span of the sensor. (See "3.8.2 Monitoring the Life Span of a Resource") (For "type" = "general titration".) 22 METTLER TOLEDO T50 / T70 / T90

29 Setup Phototrode Parameter Explanation Value range Name Select any name for the sensor. Arbitrary Unit The unit of measure for the sensor to be defined. %T A mv Sensor input The input to which you want to connect the sensor. AB1/SENSOR1 AB1/SEN- SOR2 (more depending on equipment) Serial number You can enter a serial number for the sensor here. Arbitrary Wavelength The Phototrode DP5 features a selection of 5 fixed wavelengths in [nm] Zero point The zero point of the phototrode in [%T] Slope The slope of the phototrode in [mv/%t] Calibration temperature Calibration method Date/Time Performed by The temperature during calibration in [ C]. (Does not appear for "unit" = "mv".) The phototrode can only be manually calibrated. (Does not appear for "unit" = "mv".) (Info field) Date and time of the last calibration. (Does not appear for "unit" = "mv".) (Info field) The name of the user or LabX user logged in during the last calibration. (Does not appear for "unit" = "mv".) (Info field) Manual Date/Time Logged-in user LabX user Monitoring expiration dates Monitoring life span Specifies whether to monitor the expiration date of the sensor. (See "3.8.1 Monitoring the Expiration Date of a Resource") (For "type" = "general titration" only.) (Does not appear for "unit" = "mv".) Specifies whether to monitor the life span of the sensor. (See "3.8.2 Monitoring the Life Span of a Resource") (For "type" = "general titration".) Polarized Sensors Parameter Explanation Value range Name Select any name for the sensor. Arbitrary Unit The unit of measure for the sensor to be defined. mv µa Sensor input The input to which you want to connect the sensor. AB1/SENSOR2 (more depending on equipment) Serial number You can enter a serial number for the sensor here. Arbitrary Monitoring life span Specifies whether to monitor the life span of the sensor. (See "3.8.2 Monitoring the Life Span of a Resource") (For "type" = "general titration".) METTLER TOLEDO T50 / T70 / T90 23

30 Setup Temperature Sensors Parameter Explanation Value range Name Select any name for the sensor. Arbitrary Unit The unit of measure for the sensor to be defined. C F K Sensor input The input to which you want to connect the sensor. AB1/PT1000 (more depending on equipment) Serial number You can enter a serial number for the sensor here. Arbitrary Zero point The zero point of the ph sensor in the selected unit [ C] [ F] [K] Calibration method The titrator fills this field in automatically. The entry is either filled in as "Manual" if the calibration was conducted manually, or it is filled in with the name of the calibration method if the sensor was calibrated with the aid of the "Calibrate" method function. (Info field) Date/Time Date and time of the last calibration. (Info field) Date/Time Performed by Monitoring expiration dates Monitoring life span The name of the user or LabX user logged in during the last calibration. (Info field) Specifies whether to monitor the expiration date of the sensor. (See "3.8.1 Monitoring the Expiration Date of a Resource") (For "type" = "general titration" only.) Specifies whether to monitor the life span of the sensor. (See "3.8.2 Monitoring the Life Span of a Resource") (For "type" = "general titration".) Manual Name of the calibration method Logged-in user LabX user 24 METTLER TOLEDO T50 / T70 / T90

31 Setup Conductivity Sensors Parameter Explanation Value range Name Select any name for the sensor. Arbitrary Unit The unit of measure for the sensor to be defined. µs/cm ms/cm µs ms Sensor input The input to which you want to connect the sensor. CB1/Conductivity (more depending on equipment) Serial number You can enter a serial number for the sensor here. Arbitrary Internal temperature sensor T compensation Temperature coefficient Reference temperature Cell constant Calibration temperature Calibration method Date/Time If the sensor has an internal temperature sensor, you can select "yes" here. In this case the system will automatically make an entry in the sensor setup for the internal temperature sensor. For the temperature compensation, the titrator converts the conductivity to a defined reference temperature. Linear: The conductivity is linearly converted to a reference temperature. The linearity is described by the temperature coefficient [%/ C]. Non-linear: The conductivity is converted to a reference temperature non-linearly, in accordance with the EN norm No: The conductivity is determined without temperature compensation. (Appears for "unit" = "µs/cm" or "ms/cm" only.) The temperature coefficient in [%/ C] defines the percentage of the change in conductivity for linear temperature compensation for a temperature increase of 1 C. (Appears for "T compensation" = "linear" only.) The reference temperature in [ C] for the temperature compensation. (Appears for "T compensation" = "linear" or "non-linear" only.) The cell constant in [1/cm] can be entered here. The cell constant is required to convert the measured sensor conductivity [ms µs] into the conductivity [ms/cm µs/cm]. The temperature compensation only affects the conductivity, not the conductance. (Appears for "unit" = "µs/cm" or "ms/cm" only.) You can enter the calibration temperature here. This is required during calibration to determine the temperature-dependent conductivity from conductivity standards. (Appears for "unit" = "µs/cm" or "ms/cm" only.) The titrator fills this field in automatically. The entry is either filled in as "Manual" if the calibration was conducted manually, or it is filled in with the name of the calibration method if the sensor was calibrated with the aid of the "Calibrate" method function. (Appears for "unit" = "µs/cm" or "ms/cm" only.) (Info field) Date and time of the last calibration. (Appears for "unit" = "µs/cm" or "ms/cm" only.) (Info field) Linear Non-linear No Manual Name of the calibration method Date/Time METTLER TOLEDO T50 / T70 / T90 25

32 Setup Performed by Monitoring expiration dates Monitoring life span The name of the user or LabX user logged in during the last calibration. (Appears for "unit" = "µs/cm" or "ms/cm" only.) (Info field) Specifies whether to monitor the expiration date of the sensor. (See "3.8.1 Monitoring the Expiration Date of a Resource") (For "type" = "general titration" only.) (Appears for "unit" = "µs/cm" or "ms/cm" only.) Specifies whether to monitor the life span of the sensor. (See "3.8.2 Monitoring the Life Span of a Resource") (For "type" = "general titration".) Logged-in user LabX user Temperature compensation is only conducted for the following units of measure: µs/cm and ms/cm (conductivity). Temperature compensation is not conducted for units of measure µs and ms (conductance). 26 METTLER TOLEDO T50 / T70 / T90

33 Setup Ion-selective Sensors (ISE Sensors) Parameter Explanation Value range Name Select any name for the sensor. Arbitrary Ion charge The ion charge of the ion to be determined with the ISE sensor Unit The unit of measure for the sensor to be defined. pm px ppm mv Sensor input The input to which you want to connect the sensor. AB1/SENSOR1 AB1/SENSOR2 (more depending on equipment) Serial number You can enter a serial number for the sensor here. Arbitrary Calibration Calibration type "segmented" Number of segments Zero point 1 Slope 1 (TCalib) Slope 1 (25 C) Defines the calibration type. (Does not appear for "unit" = "mv".) This parameter is adjusted automatically by the calibration method used. All subsequent parameters will be adjusted to match the number of segments. The system records two numbers per segment (zero point and slope). x-1 segment boundaries are entered for x segments. The maximum possible calibration is for 9 points (resulting in 8 segments). (Does not appear for "unit" = "mv".) (Info field) The zero point of the sensor in the selected unit for the first segment. (Does not appear for "unit" = "mv".) (Info field) The sensor gradient in [mv/unit] for the 1st segment at calibration temperature. (Does not appear for "unit" = "mv".) (Info field) The ISE-sensor gradient in [mv/unit] for the 1st Segment, umgerechnet auf eine Temperatur von 25 C. (Does not appear for "unit" = "mv".) (Info field) Seg. limit 1, 2 Specifies the boundary between Segments 1 and 2 in the selected unit. (Does not appear for "unit" = "mv".) (Info field) Zero point 2 Slope 2 (TCalib) Slope 2 (25 C) etc. The zero point of the sensor in the selected unit for the second segment. (Does not appear for "unit" = "mv".) (Info field) The sensor gradient in [mv/unit] for the 2nd segment at calibration temperature. (Does not appear for "unit" = "mv".) (Info field) The gradient of the ISE sensor for the 2nd segment in [mv/unit], recalculated for a temperature of 25 C (Does not appear for "unit" = "mv".) (Info field) Segmented Linear METTLER TOLEDO T50 / T70 / T90 27

34 Setup Calibration type "Linear" Zero point Slope (TCalib) Slope (25.0 C) Calibration temperature Calibration method Date/Time Performed by Monitoring expiration dates Monitoring life span The zero point of the ISE sensor in the selected unit. (Zero point, slope and corresponding calibration temperature are required to convert the mv signal of the sensor to the "ph" unit.) (Does not appear for "unit" = "mv".) The slope of the sensor at the calibration temperature in [mv/unit]. (Does not appear for "unit" = "mv".) The slope of the ISE sensor in [mv/unit], based on a temperature of 25 C. This field will update automatically every time the parameters "Slope [mv/ ph]" or "Calibration Temperature [ C]" are changed. (Does not appear for "unit" = "mv".) (Info field) The temperature in [ C] during calibration. (Does not appear for "unit" = "mv".) (Info field for linear calibration) The titrator fills this field in automatically. The entry is either filled in as "Manual" if the calibration was conducted manually, or it is filled in with the name of the calibration method if the sensor was calibrated with the aid of the "Calibrate" method function. (Does not appear for "unit" = "mv".) (Info field) Date and time of the last calibration. (Does not appear for "unit" = "mv".) (Info field) The name of the user or LabX user logged in during the last calibration. (Does not appear for "unit" = "mv".) (Info field) Specifies whether to monitor the expiration date of the sensor. (See "3.8.1 Monitoring the Expiration Date of a Resource") (For "type" = "general titration" only.) (Does not appear for "unit" = "mv".) Specifies whether to monitor the life span of the sensor. (See "3.8.2 Monitoring the Life Span of a Resource") (For "type" = "general titration".) Manual Name of the calibration method Date/Time Logged-in user LabX user There are two separate calibration settings for ISE sensors: one for the units of measure pm and px, and one for the unit of measure ppm. When calibrating an ISE sensor in ppm units, the sensor's slope and zero point are specified in px or pm units. 28 METTLER TOLEDO T50 / T70 / T90

Compact Titration Compact Titrator G20. One Click Titration Simple & Dependable

Compact Titration Compact Titrator G20. One Click Titration Simple & Dependable Compact Titration Compact Titrator G20 One Click Titration Simple & Dependable Simplicity The Intelligent Helper for Simple Titrations One Click Titration Intuitive method adaptation Accurate results simply

More information

Compact Titration. Compact Titrator G20. One Click Titration Simple & Dependable

Compact Titration. Compact Titrator G20. One Click Titration Simple & Dependable Compact Titration Compact Titrator G20 One Click Titration Simple & Dependable Simplicity The Intelligent Helper for Simple Titrations One Click Titration Intuitive method adaptation Accurate results simply

More information

Automatic titrators, TitroMatic

Automatic titrators, TitroMatic 112 TitroMatic These are the fourth generation of CRISON titrators. CRISON has over 25 years experience in this field and knows that, for certain applications, the performance and cost of large titrator

More information

User Manual. Compact Titrator G10S/G20S

User Manual. Compact Titrator G10S/G20S User Manual Compact Titrator G10S/G20S Table of Contents 1 Introduction 3 2 Safety information 4 2.1 Definition of signal words and warning symbols... 4 2.2 Product-specific safety notes... 4 3 Design

More information

Operating Instructions. METTLER TOLEDO DL77 / DL70ES / DL67 Titrators RESET EXIT RUN

Operating Instructions. METTLER TOLEDO DL77 / DL70ES / DL67 Titrators RESET EXIT RUN Operating Instructions METTLER TOLEDO DL77 / DL70ES / DL67 Titrators SEL RESET EXIT SEL RUN 1. INSTALLATION 2. EDITOR 3. ANALYSIS 4. AUXILIARY FUNCTIONS 5. DOCUMENTATION 6. USER LEVEL 7. REMOTE CONTROL

More information

Titration Excellence. Excellence Titrators T50 T70 T90. One Click Titration Simple, Efficient and Secure

Titration Excellence. Excellence Titrators T50 T70 T90. One Click Titration Simple, Efficient and Secure Titration Excellence Excellence Titrators T50 T70 T90 One Click Titration Simple, Efficient and Secure Titration Excellence Tailored Exactly to Your Needs Perfect Modularity Successful companies take the

More information

Easy and reliable KF titrations From 1 ppm to 100%

Easy and reliable KF titrations From 1 ppm to 100% AGC Karl Book Fischer 20% Titrators Cyan Volumetric KF Titrators DL31, DL38 Coulometric KF Titrators DL32, DL39 Oven sample changer Stromboli Easy and reliable KF titrations From 1 ppm to 100% Karl Fischer

More information

User Manual. Titration Excellence T5/T7/T9

User Manual. Titration Excellence T5/T7/T9 User Manual Titration Excellence T5/T7/T9 Table of Contents 1 Introduction 3 2 Safety information 4 2.1 Definition of signal words and warning symbols... 4 2.2 Product-specific safety notes... 4 3 Design

More information

From 1 ppm to 100% Karl Fischer water content determination

From 1 ppm to 100% Karl Fischer water content determination METTLER TOLEDO Karl Fischer Titrators DL39 & DL32 Coulometers DL38 & DL31 Volumeters Oven sample changer Stromboli From 1 ppm to 100% Karl Fischer water content determination Karl Fischer water content

More information

Automated Chloride Determination in Concrete

Automated Chloride Determination in Concrete ETTLER TOLEDO Automated Chloride Determination in Concrete Application 729 High chloride content in concrete may cause degradation and therefore the content needs to be checked on a regular basis. This

More information

рн PRO Controller The device is used to control and regulate ph level in hydroponic systems and water preparing units automatically USER MANUAL

рн PRO Controller The device is used to control and regulate ph level in hydroponic systems and water preparing units automatically USER MANUAL рн PRO Controller The device is used to control and regulate ph level in hydroponic systems and water preparing units automatically Complete Set ph PRO Controller - 1 pcs. ph electrode - 1 pcs. Calibration

More information

PAA-ANALYZERS PRINCIPLE OF OPERATION

PAA-ANALYZERS PRINCIPLE OF OPERATION PAA-ANALYZERS PRINCIPLE OF OPERATION The PAA process analyzer comprises a control unit and an analyzing section which includes the reaction chamber and measuring cell, reagent pumps and associated valves

More information

Operating Instructions. EasyPlus Titration Easy ph / Easy Ox / Easy Cl / Easy Pro

Operating Instructions. EasyPlus Titration Easy ph / Easy Ox / Easy Cl / Easy Pro Operating Instructions EasyPlus Titration Easy ph / Easy Ox / Easy Cl / Easy Pro Table of Contents 1 Introduction 5 2 Safety Notes 6 2.1 Definition of signal warnings and symbols 6 2.2 Product Specific

More information

TitriSoft 2.5. Content

TitriSoft 2.5. Content Content TitriSoft 2.5... 1 Content... 2 General Remarks... 3 Requirements of TitriSoft 2.5... 4 Installation... 5 General Strategy... 7 Hardware Center... 10 Method Center... 13 Titration Center... 28

More information

PREPARATION FOR CHEMISTRY LAB: FLUORIDE IN WATER

PREPARATION FOR CHEMISTRY LAB: FLUORIDE IN WATER 1 Name: Lab Instructor: PREPARATION FOR CHEMISTRY LAB: FLUORIDE IN WATER On these problems, the solvent is water and the solution is sufficiently dilute so that the density of the solution is the same

More information

MULTIFUNCTION COMPUTER METER CX-742 USERS MANUAL

MULTIFUNCTION COMPUTER METER CX-742 USERS MANUAL MULTIFUNCTION COMPUTER METER CX-742 USERS MANUAL CONTENTS 1. BASIC INFORMATION 3 2. USING THE INFORMATION 4 3. THE PROFILE OF THE METER 5 4. TECHNICAL DATA 7 5. SCREENS 8 6. USING THE KEYBOARD 10 7. THE

More information

Lightcloud Application

Lightcloud Application Controlling Your Lightcloud System Lightcloud Application Lightcloud Application Navigating the Application Devices Device Settings Organize Control Energy Scenes Schedules Demand Response Power Up State

More information

Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base

Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base INTRODUCTION Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base Chemists frequently make use of the equivalent weight (eq. wt.) as the basis for volumetric calculations.

More information

Titration Workstations

Titration Workstations Titration Workstations User s Guide D21T043 D21T043 Printed by Radiometer Analytical SAS 2008-09J Safety information Please read this entire manual before unpacking, setting up, or operating this equipment.

More information

Karl-Fischer Titration the method for determining water

Karl-Fischer Titration the method for determining water Karl-Fischer Titration the method for determining water Quite some experienced analyst will be unpleasantly reminded by the pyridine smell, when hearing the name Karl Fischer. However, modern reagents

More information

ION570. Reference Manual. Ion Analyser D21M073

ION570. Reference Manual. Ion Analyser D21M073 ION570 Ion Analyser Reference Manual D21M073 D21M073 Printed by Radiometer Analytical SAS 2008-07E Contents Contents...3 Introduction...13 Read me!...14 Practical examples...17 Programming electrodes...19

More information

Planning Softproviding Meat User Documentation

Planning Softproviding Meat User Documentation Great ideas are always simple Softproviding simply makes them happen. Planning Softproviding Meat User Documentation Version: 1.00 Date: 24 August 2017 Release: v5.50 Softproviding AG Riehenring 175 CH-4058

More information

ON SITE SYSTEMS Chemical Safety Assistant

ON SITE SYSTEMS Chemical Safety Assistant ON SITE SYSTEMS Chemical Safety Assistant CS ASSISTANT WEB USERS MANUAL On Site Systems 23 N. Gore Ave. Suite 200 St. Louis, MO 63119 Phone 314-963-9934 Fax 314-963-9281 Table of Contents INTRODUCTION

More information

Semi-automatic titrators PH-Burette. A great ph-meter with integrated burette. Incredible price! Infinite possibilities

Semi-automatic titrators PH-Burette. A great ph-meter with integrated burette. Incredible price! Infinite possibilities Semi-automatic titrators PH-Burette. A great ph-meter with integrated burette. Incredible price! Infinite possibilities instruments 5 YEARS G U A R A N T E E Quality at a reasonable price The optimum combination

More information

Using Conductivity to Find an Equivalence Point

Using Conductivity to Find an Equivalence Point Experiment 25 PRE LAB DISCUSSION In this experiment, you will monitor conductivity during the reaction between sulfuric acid, and barium hydroxide in order to determine the equivalence point. From this

More information

Reagents - Install/Replace. Select method/sequence. KF Reagent calibration Running KF titration methods. Running SAC sequences

Reagents - Install/Replace. Select method/sequence. KF Reagent calibration Running KF titration methods. Running SAC sequences Short-For t-form Reminders Reagents - Install/Replace Select method/sequence KF Reagent calibration Running KF titration methods Running SAC sequences Maintenance recommendations Storage recommendations

More information

Creating Empirical Calibrations

Creating Empirical Calibrations 030.0023.01.0 Spreadsheet Manual Save Date: December 1, 2010 Table of Contents 1. Overview... 3 2. Enable S1 Calibration Macro... 4 3. Getting Ready... 4 4. Measuring the New Sample... 5 5. Adding New

More information

Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM / / 2014

Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM / / 2014 Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM8004 80003223 / 00 05 / 2014 Contents 1 Preliminary note...3 1.1 Symbols used...3 2 Safety instructions...4 3 Functions and features...5

More information

Titrex ACT titrators have been designed to simply and precisely satisfy and perform the widest range of potentiometric titrations: acid-base, redox,

Titrex ACT titrators have been designed to simply and precisely satisfy and perform the widest range of potentiometric titrations: acid-base, redox, Titrex ACT titrators have been designed to simply and precisely satisfy and perform the widest range of potentiometric titrations: acid-base, redox, argentometric, complexometric titrations Compact and

More information

NINE CHOICE SERIAL REACTION TIME TASK

NINE CHOICE SERIAL REACTION TIME TASK instrumentation and software for research NINE CHOICE SERIAL REACTION TIME TASK MED-STATE NOTATION PROCEDURE SOF-700RA-8 USER S MANUAL DOC-025 Rev. 1.3 Copyright 2013 All Rights Reserved MED Associates

More information

Titration of a strong acid with a strong base with Cobra4

Titration of a strong acid with a strong base with Cobra4 Titration of a strong acid with a strong base with Cobra4 TEC Related topics Strong and weak acids and bases, ph value, titration curves, equivalence point, potentiometry. Principle Hydrochloric acid is

More information

Microscale Acid-Base Titration

Microscale Acid-Base Titration Microscale Acid-Base Titration Experiment 36 A titration is a process used to determine the volume of a solution needed to react with a given amount of another substance. In this experiment, you will titrate

More information

MEASUREMENT: PART II

MEASUREMENT: PART II 1 MEASUREMENT: PART II Copyright: Department of Chemistry, University of Idaho, Moscow, ID 83844-2343, 2013. INTRODUCTION Read and/or review Section 1.7 and Figure 7.5 in your textbook. The first part

More information

ISIS/Draw "Quick Start"

ISIS/Draw Quick Start ISIS/Draw "Quick Start" Click to print, or click Drawing Molecules * Basic Strategy 5.1 * Drawing Structures with Template tools and template pages 5.2 * Drawing bonds and chains 5.3 * Drawing atoms 5.4

More information

M E R C E R W I N WA L K T H R O U G H

M E R C E R W I N WA L K T H R O U G H H E A L T H W E A L T H C A R E E R WA L K T H R O U G H C L I E N T S O L U T I O N S T E A M T A B L E O F C O N T E N T 1. Login to the Tool 2 2. Published reports... 7 3. Select Results Criteria...

More information

PC Control / Touch Control

PC Control / Touch Control PC Control / Touch Control Version 6.0 / 5.840.0150 New Features Manual 8.840.8007EN Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com

More information

Notifications and Accretions: Facility Role

Notifications and Accretions: Facility Role Notifications and Accretions: Facility Role Goal: Learn to view and resolve notification and accretion discrepancies in CROWNWeb. Estimated Time: 25 to 30 minutes PDF: Download a screen reader compatible

More information

Operating Instructions. EasyPlus Titration Easy KFV

Operating Instructions. EasyPlus Titration Easy KFV Operating Instructions EasyPlus Titration Easy KFV Table of Contents 1 Introduction 5 2 Safety Notes 6 2.1 Definition of Signal Warnings and Symbols 6 2.2 Product Specific Safety Notes 6 3 Design and

More information

Simple Operation Manual

Simple Operation Manual 苦 SET UP Simple Operation Manual (1) Check the level of buffer, purified water and wash solution Left: Buffer Center: Wash solution Right: Purified water Change buffer or refill them (W.S. and P.W.) if

More information

TECDIS and TELchart ECS Weather Overlay Guide

TECDIS and TELchart ECS Weather Overlay Guide 1 of 24 TECDIS and TELchart ECS provides a very advanced weather overlay feature, using top quality commercial maritime weather forecast data available as a subscription service from Jeppesen Marine. The

More information

TECHNICAL MANUAL 820 LX / 910 LX / 1300 LX

TECHNICAL MANUAL 820 LX / 910 LX / 1300 LX TECHNICAL MANUAL 820 LX / 910 LX / 1300 LX LANCER reserves the right to modify, constantly its documentation for its improvement. The values of adjustments indicated in the displays of this manual are

More information

SuperCELL Data Programmer and ACTiSys IR Programmer User s Guide

SuperCELL Data Programmer and ACTiSys IR Programmer User s Guide SuperCELL Data Programmer and ACTiSys IR Programmer User s Guide This page is intentionally left blank. SuperCELL Data Programmer and ACTiSys IR Programmer User s Guide The ACTiSys IR Programmer and SuperCELL

More information

ph electrode Instruction Manual For use with the RAH-210 & RPH-250

ph electrode Instruction Manual For use with the RAH-210 & RPH-250 ph electrode Instruction Manual For use with the RAH-210 & RPH-250 The information in this manual was current at the time of printing. The most current versions of all Hydro Instruments manuals can be

More information

ION570. Reference Manual. Ion Analyser D21M073

ION570. Reference Manual. Ion Analyser D21M073 ION570 Ion Analyser Reference Manual D21M073 D21M073 Printed by Radiometer Analytical SAS 2005-02C Contents Contents...3 Introduction...13 Read me!...14 Practical examples...17 Programming electrodes...19

More information

This page intentionally left blank.

This page intentionally left blank. Vela User s Guide The Vela User s Guide is not intended as a replacement for the Operator s Manual. You must become completely familiar with the Vela Operator s Manual before using the Vela ventilator.

More information

BASIC TECHNOLOGY Pre K starts and shuts down computer, monitor, and printer E E D D P P P P P P P P P P

BASIC TECHNOLOGY Pre K starts and shuts down computer, monitor, and printer E E D D P P P P P P P P P P BASIC TECHNOLOGY Pre K 1 2 3 4 5 6 7 8 9 10 11 12 starts and shuts down computer, monitor, and printer P P P P P P practices responsible use and care of technology devices P P P P P P opens and quits an

More information

Thermo Scientific Orion Star T900 Series Laboratory Titrators. Potentiometric titrators for performing automated ph, redox, or ion titrations

Thermo Scientific Orion Star T900 Series Laboratory Titrators. Potentiometric titrators for performing automated ph, redox, or ion titrations Thermo Scientific Orion Star T900 Series Laboratory Titrators Potentiometric titrators for performing automated ph, redox, or ion titrations Thermo Scientific Orion Star T900 Series Titrators Automate

More information

TitraLab 580 and 585. KF Titration Workstations. Reference Manual D21T083

TitraLab 580 and 585. KF Titration Workstations. Reference Manual D21T083 580 and 585 KF Titration Workstations Reference Manual D21T083 D21T083 Printed by Radiometer Analytical SAS 2008-02B Contents Contents...3 Introduction...11 General Information...12 Starting up instructions...13

More information

INE-PHS-3E ph Meter Operating Manual

INE-PHS-3E ph Meter Operating Manual INE-PHS-3E ph Meter Operating Manual PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC.12.16 ADVICE: Read this instruction

More information

White Paper. Perform Conductivity Measurements In Compliance with USP <645>

White Paper. Perform Conductivity Measurements In Compliance with USP <645> Perform Conductivity Measurements In Compliance with USP Water is the most widely used substance, raw material, or ingredient in the production, processing and formulation of compendial articles.

More information

Acid-Base Titration. Evaluation copy

Acid-Base Titration. Evaluation copy Acid-Base Titration Computer 7 A titration is a process used to determine the volume of a solution that is needed to react with a given amount of another substance. In this experiment, your goal is to

More information

DIFFUSION THROUGH MEMBRANES STANDARDS B C.4 INTRODUCTION

DIFFUSION THROUGH MEMBRANES STANDARDS B C.4 INTRODUCTION DIFFUSION THROUGH MEMBRANES STANDARDS 3.2.12.B.1 3.2.12.C.4 INTRODUCTION Westminster College Many aspects of the life of a cell depend on the fact that atoms and molecules have kinetic energy and are constantly

More information

Acid-Base Titration. Volume NaOH (ml) Figure 1

Acid-Base Titration. Volume NaOH (ml) Figure 1 LabQuest 24 A titration is a process used to determine the volume of a solution needed to react with a given amount of another substance. In this experiment, you will titrate hydrochloric acid solution,

More information

Experiment C-10 Titration of a Strong Acid and a Strong Base

Experiment C-10 Titration of a Strong Acid and a Strong Base 1 Experiment C-10 Titration of a Strong Acid and a Strong Base Objectives To study the titration process. To follow changes in the ph during the titration process while adding a strong base to a strong

More information

Lab 1 Uniform Motion - Graphing and Analyzing Motion

Lab 1 Uniform Motion - Graphing and Analyzing Motion Lab 1 Uniform Motion - Graphing and Analyzing Motion Objectives: < To observe the distance-time relation for motion at constant velocity. < To make a straight line fit to the distance-time data. < To interpret

More information

Our Drop Counter sensor now features housing for two electrode sensors, an anti-twist mechanism, an indicator LED and two cable guides.

Our Drop Counter sensor now features housing for two electrode sensors, an anti-twist mechanism, an indicator LED and two cable guides. The Drop Counter sensor is an optical sensor that accurately records the number of drops of titrant added during a titration. The Drop Counter sensor software can automatically convert the number of drops

More information

Conductometric Titration & Gravimetric Determination of a Precipitate

Conductometric Titration & Gravimetric Determination of a Precipitate Conductometric Titration & Gravimetric Determination of a Precipitate Experiment 9 In this experiment, you will monitor conductivity during the reaction between sulfuric acid, H2SO4, and barium hydroxide,

More information

A new chapter. medicacorp.com

A new chapter. medicacorp.com A new chapter. \EDIC@ medicacorp.com Same operational reliability. Reliable performance, test after test. The photometer s xenon fl ash lamp typically has a 5-year use life and is maintenance-free. No

More information

O P E R A T I N G M A N U A L

O P E R A T I N G M A N U A L OPERATING MANUAL WeatherJack OPERATING MANUAL 1-800-645-1061 The baud rate is 2400 ( 8 bits, 1 stop bit, no parity. Flow control = none) To make sure the unit is on line, send an X. the machine will respond

More information

Description of the ED library Basic Atoms

Description of the ED library Basic Atoms Description of the ED library Basic Atoms Simulation Software / Description of the ED library BASIC ATOMS Enterprise Dynamics Copyright 2010 Incontrol Simulation Software B.V. All rights reserved Papendorpseweg

More information

ACID-BASE TITRATION (MICROSCALE)

ACID-BASE TITRATION (MICROSCALE) ACID-BASE TITRATION (MICROSCALE) LAB PH 4.PALM From Science with Handhelds, Vernier Software & Technology, 2002. INTRODUCTION Acids and bases represent a major class of chemical substances. We encounter

More information

Conductimetric Titration and Gravimetric Determination of a Precipitate

Conductimetric Titration and Gravimetric Determination of a Precipitate Conductimetric Titration and Gravimetric Determination of a Precipitate Handheld 16 In this experiment, you will monitor conductivity during the reaction between sulfuric acid, H 2 SO 4, and barium hydroxide,

More information

TA Instruments ITCRun TM Software Getting Started Guide

TA Instruments ITCRun TM Software Getting Started Guide TA Instruments ITCRun TM Software Getting Started Guide Instrument Types Supported ITCRun software is used to control the operation of the Nano ITC III, Nano ITC Standard Volume, and Nano ITC Low Volume

More information

Determination of the Rydberg constant, Moseley s law, and screening constant (Item No.: P )

Determination of the Rydberg constant, Moseley s law, and screening constant (Item No.: P ) Determination of the Rydberg constant, Moseley s law, and screening constant (Item No.: P2541001) Curricular Relevance Area of Expertise: ILIAS Education Level: Physik Topic: Hochschule Subtopic: Moderne

More information

User's Guide. DISTO online. Leica Geosystems

User's Guide. DISTO online. Leica Geosystems User's Guide DISTO online Leica Geosystems Copyright 2001 by PMS Photo Mess Systeme AG. All rights reserved. This manual describes the versions 2.x of the program DISTO online. PMS PHOTO-MESS-SYSTEME AG

More information

Operating instructions Magnetic-inductive flow meter SM6x04 SM7x04 SM8x04

Operating instructions Magnetic-inductive flow meter SM6x04 SM7x04 SM8x04 Operating instructions Magnetic-inductive flow meter SM6x04 SM7x04 SM8x04 UK 80223776 / 00 11 / 2018 Contents 1 Preliminary note...3 2 Safety instructions...4 3 Functions and features...4 4 Function...5

More information

pka AND MOLAR MASS OF A WEAK ACID

pka AND MOLAR MASS OF A WEAK ACID Experiment 10 pka AND MOLAR MASS OF A WEAK ACID Adapted by the Chemistry Faculty of Eastern Michigan University from EQUL 305,written by Richard C. Bell, Lebanon Valley College, published by Chemical Education

More information

Acid-Base Titration Curves Using a ph Meter

Acid-Base Titration Curves Using a ph Meter Acid-Base Titration Curves Using a ph Meter Introduction: In this experiment you will use a ph sensor to collect volume and ph data as you titrate two acids with sodium hydroxide. You will obtain titration

More information

Related concepts Electrolyte, electrical conductance, specific conductance, ion mobility, ion conductivity, conductometry, volumetry.

Related concepts Electrolyte, electrical conductance, specific conductance, ion mobility, ion conductivity, conductometry, volumetry. Conductometric titration with Cobra4 TEC Related concepts Electrolyte, electrical conductance, specific conductance, ion mobility, ion conductivity, conductometry, volumetry. Principle The electric conductivity

More information

INSTRUCTOR RESOURCES

INSTRUCTOR RESOURCES INSTRUCTOR RESOURCES The CCLI Initiative Learning Objectives The objectives of this experiment are to... analyze an acid mixture by titration with NaOH solution. use the MicroLAB interface to gather and

More information

Mobile coulometer for water determination anywhere you want

Mobile coulometer for water determination anywhere you want 899 Coulometer Mobile coulometer for water determination anywhere you want Extremely small & compact Maximum flexibility due to independent power supply Automatic start of titration 899 Coulometer the

More information

Skin Damage Visualizer TiVi60 User Manual

Skin Damage Visualizer TiVi60 User Manual Skin Damage Visualizer TiVi60 User Manual PIONEERS IN TISSUE VIABILITY IMAGING User Manual 3.2 Version 3.2 October 2013 Dear Valued Customer! TiVi60 Skin Damage Visualizer Welcome to the WheelsBridge Skin

More information

GreenLight Model 910 Software Use and Water Sample Analysis Manual. September 23, 2013

GreenLight Model 910 Software Use and Water Sample Analysis Manual. September 23, 2013 GreenLight Model 910 Software Use and Water Sample Analysis Manual September 23, 2013 Table of Contents Introduction... 6 Instrument Set-Up... 6 Software Installation... 6 Main Start-Up Screen... 7 Administrator...

More information

Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure

Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure DATASHEET Pascal ET Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure - electrical signals (ma, mv, V, ) - temperature (TC and

More information

Acid-Base Titration. Sample

Acid-Base Titration. Sample Acid-Base Titration Computer 7 A titration is a process used to determine the volume of a solution that is needed to react with a given amount of another substance. In this experiment, your goal is to

More information

Chemistry MultiMeasure Sensor PS-2170

Chemistry MultiMeasure Sensor PS-2170 ä 5 Instruction Manual 012-09523A Chemistry MultiMeasure Sensor PS-2170 ph/ise/orp PRESSURE CHEMISTRY TEMPERATURE SENSOR VOLTAGE 1 2 3 4 PS-2170 5 6 7 cc 8 10 15 20 25 30 35 40 45 50 55 60 Included Equipment

More information

How to Make or Plot a Graph or Chart in Excel

How to Make or Plot a Graph or Chart in Excel This is a complete video tutorial on How to Make or Plot a Graph or Chart in Excel. To make complex chart like Gantt Chart, you have know the basic principles of making a chart. Though I have used Excel

More information

Acid-Base Titration Curves Using a ph Meter

Acid-Base Titration Curves Using a ph Meter Acid-Base Titration Curves Using a ph Meter Introduction: In this experiment you will use a ph sensor to collect volume and ph data as you titrate two acids with sodium hydroxide. You will obtain titration

More information

Acceleration/Velocity/Displacement VIBRATION METER

Acceleration/Velocity/Displacement VIBRATION METER Acceleration/Velocity/Displacement VIBRATION METER Model : VB-8220 Your purchase of this VIBRATION METER marks a step forward for you into the field of precision measurement. Although this METER is a complex

More information

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7 Troy, NY 12180 Date: 03/14/2012 Page 1 of 7 USGS New York Water Science Center Soil and Low Ionic Strength Water Quality Laboratory Electrometric Measurements and Titration Standard Operating Procedure

More information

PC100 Handheld ph/cond./tds/sal. Meter. Instruction Manual

PC100 Handheld ph/cond./tds/sal. Meter. Instruction Manual PC100 Handheld ph/cond./tds/sal. Meter Instruction Manual Table of Contents 1. Introduction ------------------------------------------------------------------------------------------------------------------------

More information

Standardizing a Solution of Sodium Hydroxide. Evaluation copy

Standardizing a Solution of Sodium Hydroxide. Evaluation copy Standardizing a Solution of Sodium Hydroxide Computer 6 It is often necessary to test a solution of unknown concentration with a solution of a known, precise concentration. The process of determining the

More information

Automatic Potentiometric Titrator

Automatic Potentiometric Titrator Automatic Potentiometric Titrator AT-610 Automatic Potentiometric Titrator Two Different Titrations can be Performed Simultaneously --- Allow for Space-saving. Also, a Karl Fischer Moisture Titrator can

More information

The first address for Titration. TITRONIC 300 and TitroLine 5000

The first address for Titration. TITRONIC 300 and TitroLine 5000 The first address for Titration TITRONIC 300 and TitroLine 5000 Selection table titration piston burettes TITRONIC and automatic titrators TitroLine The most important features of titrators TitroLine and

More information

The CSC Interface to Sky in Google Earth

The CSC Interface to Sky in Google Earth The CSC Interface to Sky in Google Earth CSC Threads The CSC Interface to Sky in Google Earth 1 Table of Contents The CSC Interface to Sky in Google Earth - CSC Introduction How to access CSC data with

More information

Experiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013

Experiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013 Absorbance Experiment 13 Dilutions and Data Handling in a Spreadsheet rev 1/2013 GOAL: This lab experiment will provide practice in making dilutions using pipets and introduce basic spreadsheet skills

More information

User Manuel. EurotaxForecast. Version Latest changes ( )

User Manuel. EurotaxForecast. Version Latest changes ( ) User Manuel EurotaxForecast Version 1.23.0771- Latest changes (19.07.2003) Contents Preface 5 Welcome to Eurotax Forecast...5 Using this manual 6 How to use this manual?...6 Program overview 7 General

More information

USER MANUAL. Aquamax TAN

USER MANUAL. Aquamax TAN Aquamax TAN USER MANUAL GUARANTEE The Aquamax TAN titrators are guaranteed against any manufacturing defect or component failure. GRScientific will repair or replace, free of charge, any defective component

More information

Driven Harmonic Oscillator

Driven Harmonic Oscillator Driven Harmonic Oscillator Physics 6B Lab Experiment 1 APPARATUS Computer and interface Mechanical vibrator and spring holder Stands, etc. to hold vibrator Motion sensor C-209 spring Weight holder and

More information

Titrator 3.0 Tutorial: Calcite precipitation

Titrator 3.0 Tutorial: Calcite precipitation Titrator 3.0 Tutorial: Calcite precipitation November 2008 Steve Cabaniss A. Introduction This brief tutorial is intended to acquaint you with some of the features of the program Titrator. It assumes that

More information

Technical Procedure for Glass Refractive Index Measurement System 3 (GRIM 3)

Technical Procedure for Glass Refractive Index Measurement System 3 (GRIM 3) Technical Procedure for Glass Refractive Index Measurement System 3 (GRIM 3) 1.0 Purpose - This technical procedure shall be followed for the operation of the GRIM 3. 2.0 Scope - This procedure applies

More information

Operating Instructions. METTLER TOLEDO DL37 KF Coulometer

Operating Instructions. METTLER TOLEDO DL37 KF Coulometer Operating Instructions METTLER TOLEDO DL37 KF Coulometer Contents Contents Page. Safety measures 5 2. Introduction 6 2. Measurement principle 7 3. Setting up the system 8 3. Identification of parts 8 3.2

More information

Conductimetric Titration and Gravimetric Determination of a Precipitate

Conductimetric Titration and Gravimetric Determination of a Precipitate Conductimetric Titration and Gravimetric Determination of a Precipitate LabQuest 16 In this experiment, you will monitor conductivity during the reaction between sulfuric acid, H 2 SO 4, and barium hydroxide,

More information

Agilent MassHunter Quantitative Data Analysis

Agilent MassHunter Quantitative Data Analysis Agilent MassHunter Quantitative Data Analysis Presenters: Howard Sanford Stephen Harnos MassHunter Quantitation: Batch and Method Setup Outliers, Data Review, Reporting 1 MassHunter Quantitative Analysis

More information

Table of Contents Chemicals & Fertilizers

Table of Contents Chemicals & Fertilizers Table of Contents Chemicals & Fertilizers Chemicals & Fertilizers 4 Introduction... 4 Chemical & Fertilizer Browse Table... 5 Grounds Dimension Table... 7 Editing Grounds Area Names... 8 Adding New Grounds

More information

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter Water and Lab Products, Thermo Fisher Scientific Technical Note 505 Key Words Thermo Scientific Orion

More information

Multiphysics Modeling

Multiphysics Modeling 11 Multiphysics Modeling This chapter covers the use of FEMLAB for multiphysics modeling and coupled-field analyses. It first describes the various ways of building multiphysics models. Then a step-by-step

More information

The Clinical Chemistry System that offers more so you do less!

The Clinical Chemistry System that offers more so you do less! The Clinical Chemistry System that offers more so you do less! A unique level of automation in a compact system Safe and flexible sample management Fully automatic cuvette management Cuvette management

More information

Operation Manual. SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

Operation Manual. SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Operation Manual SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC. 4.18 Foreword

More information

Chemistry with Mr. Faucher. Acid-Base Titration

Chemistry with Mr. Faucher. Acid-Base Titration Chemistry with Mr. Faucher Name Date Acid-Base Titration 24 A titration is a process used to determine the volume of a solution needed to react with a given amount of another substance. In this experiment,

More information