ALLOWAY. Method Outline. Standard Laboratory Method: Total Fluoride via Distillation. Distillation then Specific Ion. Date Issued: 01/04/82

Size: px
Start display at page:

Download "ALLOWAY. Method Outline. Standard Laboratory Method: Total Fluoride via Distillation. Distillation then Specific Ion. Date Issued: 01/04/82"

Transcription

1 ALLOWAY Method Outline Standard Laboratory Method: SM4500-F - B and SM4500-F-C Parameter: Total Fluoride via Distillation Method: Distillation then Specific Ion Date Issued: 01/04/82 Date Revised: 04/09/02, 08/12/03, 01/31/05, 04/07/2008 Optimum Concentration Range: mg/l Sensitivity: N/A Detection Limit: 0.10 mg/l Reference: Standard Methods 20 th edition, 1998 Alloway Page 1 of 9

2 Total Fluoride 1.0 Scope and Application: 1.1 This method is applicable to the measurement of fluoride in drinking, surface and saline waters, and domestic and industrial wastes. 1.2 The applicable range of this method is from 0.10 to 10 mg/l. Higher levels may be measured by sample dilution. 2.0 Summary of Method: 2.1 Fluoride is determined potentiometrically using a fluoride electrode in conjunction with a standard single-junction sleeve type reference electrode and a ph meter having an expanded millivolt scale. 2.2 The fluoride elctrode has a laser-type toped lanthanum fluoride crystal across which a potential is established by fluoride solution at different concentrations. The crystal contacts the sample at one face and an internal reference solution at the other. The cell is represented Ag/AgCl, Cl (0.3M), F - (0.001M)\LaF 3 \Test Solution\reference solution. 2.3 The fluoride ion activity depends on total ionic strength and ph of the solution. Fluoride concentration is measured by adding buffer TISABII which provides a nearly uniform ionic background, adjusts ph, and breaks up complexs. 2.4 Distillation will separate fluoride from other nonvolatile constituents in water. 3.0 Definitions 3.1 Accuracy-The degree of agreement between an observed value and an accepted reference value. Accuracy includes a combination of random error (precision) and systematic error (bias) components which are due to sampling and analytical operations; a data quality indicator. 3.2 Aliquot - A portion of the field sample to be processed through the analytical steps 3.3 Analytical Batch Composed of prepared environmental samples (extracts, digestates, or concentrates) which are analyzed together as a group. An analytical batch can include prepared samples originating from various environmental matrices. 3.4 Calibration To determine, by measurement or comparison with a standard, the correct value of each scale reading on a meter, instrument, or other device. The levels of the applied calibration standard must bracket the range of planned or expected sample measurements. 3.5 Calibration Curve The graphical relationship between the known values, such as concentrations, of a series of calibration standards and their instrument response. Alloway Page 2 of 9

3 3.6 Continuing Calibration Check or Verification Standard - check standard analyzed periodically to evaluate whether the analytical process remains in control. These results are compared to the initial calibration. Noted as ICV for initial and CCV for continuing check. 3.7 Duplicate Analyses- The analyses or measurements of the variable of interest performed identically on two sub-samples of the same sample. The results from duplicate analyses are used to evaluate analytical or measurement precision but not the precision of sampling, preservation or storage internal to the laboratory. 3.8 Homogeneous Sample - A sample which has an even distribution of material. 3.9 Instrument Detection Limit - The minimum concentration of an analyte that can be measured with 99% confidence that the analyte concentration is greater than zero. Determine the IDL with seven replicates of standard solution at the instrument with no specific analytical preparation Method Detection Limit - The minimum concentration of an analyte that can be measured with 99% confidence that the analyte concentration is greater than zero. Determine the MDL with seven replicates of standard solution taken through the preparative and analytical procedures Initial Calibration - A series of standards analyzed to produce a calibration curve in order to develop a linear system Internal Standard A known amount of standard added to a test portion of a sample as a reference for evaluating and controlling the precision and bias of the applied analytical method Laboratory Control Sample A sample matrix, free from the analytes of interest, spiked with verified known amounts of analytes from a source independent of the calibration standards or a material containing known and verified amounts of analytes. It is generally used to establish intra-laboratory or analyst/technician specific precision and bias or to assess the performance of all or a portion of the measurement system. LCSA denotes aqueous samples and LCSS denotes solid samples Matrix- The component or substrate that contains the analyte of interest Matrix Spike A sample prepared by adding a known mass of target analyte to a specified amount of matrix sample for which an independent estimate of target analyte is available. Matrix spikes are used, for example, to determine the effect of the matrix on a method s recovery efficiency Matrix Spike Duplicate A second replicate matrix spike prepared in the laboratory and analyzed to obtain a measure of the precision of the recovery for each analyte. Alloway Page 3 of 9

4 3.17 Quality Assurance An integrated system of activities involving planning, quality control, quality assessment, reporting and quality improvement to ensure that a product or service meets defined standards of quality with a stated level of confidence Quality Control The overall system of technical activities whose purpose is to measure and control the quality of a product or service so that it meets the needs of users Reagent Blank or Method Blank A sample consisting of reagent(s), without the target analyte or sample matrix, introduced into the analytical procedure at the appropriate point and carried through all subsequent steps to determine the contribution of the reagents and of the involved analytical steps Reporting Limit - The practical quantitation level for an analyte, greater than the MDL, accounting for dilutions and matrix specific concerns Sample Duplicate Two samples taken from and representative of the same population and carried through all steps of the sampling and analytical procedures in an identical manner. Duplicate samples are used to assess variance of the total method including sampling and analysis Reporting Limit Check (RDL) quality control sample analyzed as part of the analytical run at the reporting limit to determine accuracy and precision. 4.0 Interferences 4.1 Extremes of ph interfere; sample ph should be between 5 and 9 s.u.. Polyvalent cations of Si +4, Fe+3, and Al+3 interfere by forming complexes with fluoride. The degree of interference depends upon the concentration of complexing cations, the concentration of fluoride and the ph of the sample. The addition of a ph 5.0 buffer (described below) containing a strong chelating agent preferentially complexes aluminum (the most common interference), silicon and iron and eliminates the ph problem. Adding ph buffer TISAB II elimates most interferences. 4.2 Fluoride forms complexs with several polyvalent cations, especially aluminum and iron. The extent of complexation depends on solution ph, the relative levels of fluoride, and nature of the complexing species The ingredient CDTA (Cyclohexylenediaminetetraacetic acid) of the buffer TISAB complexes interfering cations to release free fluoride ions. 4.3 In acidic solutions, the HF HF complex forms. If solution is alkaline the Hydroxide ion interferes with electrode response if the hydroxide concentration is greater than 1/10, the concentration of fluoride. Alloway Page 4 of 9

5 4.3.1 The TISAB buffer maintains ph between 5 and 9 s.u. so that these interferences do not occur. 5.0 Safety 5.1 The toxicity or carcinogenicity of each chemical and reagent used in this method has not been precisely defined. However, each one must be treated as a potential health hazard, and exposure to these chemicals should be minimized. Each analyst is responsible for adherence to the procedures outlined in the Chemical Hygiene Plan. Material Safety Data Sheets are kept in the laboratory. 5.2 Some method analytes have been tentatively classified as known or suspected human or mammalian carcinogens. Pure standard materials and stock standard solutions of these compounds should be handled with suitable protection to skin, eyes, etc. 5.3 Safety glasses and lab coats must be worn when handling samples and solvents. Safety glasses must be worn when handling glassware. 6.0 Equipment and Supplies 6.1 ph meter with expanded millivolt scale, ph meter with direct concentration reading or a selective ion meter such as the Orion 400 series, or equivalent. 6.2 Fluoride Ion Activity Electrode, such as Orion No , or equivalent. 6.3 Reference electrode, single junction, sleeve type, such as Orion No 90-01, or equivalent. Do not use fiber-tip reference electrodes because they behave erractically in very dilute solutions. 6.4 Magnetic stirrer, thermally insulated, and teflon-coated stirring bar. 6.5 Distillation apparatus: A 1-liter round bottom, long-necked pyrex boiling flask, connecting tube, efficient condenser, thermometer adapter and thermometer reading to 200 C. All connections should be ground glass. 7.0 Reagents and Calibration Standards 7.1 Buffer solution, ph : To approximately 500 ml of distilled water in a 1 liter beaker add 57 ml of glacial acetic acid, 58 g of sodium chloride and 2 g of CDTA (2). Stir to dissolve and cool to room temperature. Adjust ph of solution to between 5.0 and 5.5 with 5 N sodium hydroxide (about 150 ml will be required). Transfer solution to a 1liter volumetric flask and dilute to the mark with distilled water. For work with brines, additional NaCl should be added to raise the chloride level to twice the highest expected level of chloride in the sample. This is generally purchased in concentrated form. (Orion TISAB II). 7.2 Fluoride 100 mg/l - dissolve g anhydrous sodium fluoride in distilled water and dilute to 1 liter with distilled water mg/l std. 10 ml of the 100 mg/l std up to 100 ml in volumetric flask mg/l std. 10 ml of the 10 mg/l std up to 100 ml in volumetric flask. Alloway Page 5 of 9

6 mg/l std. 5 ml of the 10 mg/l std up to 100 ml in volumetric flask. 7.3 Second source fluoride standard (1000 mg/l), purchased from a commercial vendor mg/l ICV/CCV = 1 ml of the second source standard up to 500 ml 7.4 Sulfuric Acid, concentrated. 8.0 Sample Collection, Preservation, and Handling 8.1 Sample gathered for fluoride procedure can be collected in a plastic or glass non preserved container. 8.2 Sample for fluoride analysis can degrade significantly between collection and analysis. Therefore the sample should be analyzed promptly, and stored at or below 4 o C until analysis is started. 8.3 Holding time is 28 days from date of collection. 9.0 Calibration and Standardization 9.1 From 100 mg/l fluoride standard prepare 10 mg/l, 1 mg/l and 0.5 mg/l for calibration curve. (7.2) 9.2 A second source check standard is prepared from a fluoride standard purchased from a commercial vendor. (7.3) 10.0 Quality Control 10.1 One sample of each batch of ten samples is to be analyzed in duplicate and spiked with a known amount of the fluoride standard. The duplicates must be within 10% of each other and the spikes must be % of the true value. If the duplicates or spikes fall outside the acceptance range, they must be reanalyzed or flagged with a comment on the report A Initial Calibration Blank (ICB) must be analyzed after the calibration. A Continuing Calibration Blank (CCB) must be analyzed after each batch of ten samples and at the end of the analytical run. The ICB and CCB must read less than the 0.10 mg/l reporting limit A Reporting Detection Limit standard (RDL) must be analyzed following the calibration. The RDL must read % of the true value. If the RDL falls outside the acceptance range, it must be reanalyzed prior to continuing the run A Initial Calibration Verification standard (ICV) prepared from the second source standard (9.2) must be analyzed following calibration. A Continuing Calibration Verification standard (CCV) prepared from the second source standard (9.2) must be analyzed after each batch of ten samples and at the end of the analytical run. The ICV and CCV must read % of the true value If the ICV falls outside the acceptance range, the instrument must be recalibrated and the ICV must be reanalyzed prior to analyzing samples. If the CCV falls outside the acceptance range, it must be reanalyzed along with all of the samples analyzed from the last acceptable CCV A Laboratory Control Standard (LCSA) must be analyzed after calibration. Alloway Page 6 of 9

7 The LCSA must read % of the true value. If the LCSA falls outside the acceptance range, it must be reanalyzed prior to continuing the analysis The calibration Slope is calculated by the instrument after the calibration is performed. The slope must be 54 to 60 mv/dec. or the instrument must be recalibrated A preparation blank and preparation standard must be prepared with every batch of distillates In addition, a duplicate and spike is prepped with every batch of distillates Analytical Procedures Preliminary Distillation 11.1 Place 400 ml distilled water in the distilling flask Carefully add 200 ml concentrated sulfuric acid and swirl until contents are homogeneous Add glass beads and stir bar; connect the apparatus making sure all joints are tight Turn stir plate on until a continuous slow motion is achieved Turn on the water to the condenser Heat slowly at first, then as rapidly as the efficiency of the condenser will permit until the temperature of the flask contents reaches exactly 175 degrees C. Discard the distillate. This process removes fluoride contamination and adjusts the acid-water ratio for subsequent distillations Cool to 80 degrees C or below Add 200 ml sample or in the case of solid or semi-solid samples, weight 10 g of wet sample and add 100 ml distilled water, mix thoroughly, distill until temperature reaches 175 degrees C but not exceeding 180 degrees C to prevent sulfate carryover. Initial sample volume may need to be adjusted due to total sample volume being limited Use the sulfuric acid solution in the flask repeatedly until the contaminants from the samples accumulate to such an extent that recovery is affected or interferences appear in the distillate. Check periodically by distilling standard fluoride samples High fluoride samples may require that the still be flushed by using distilled water and combining distillates. Analytical Procedure 11.1 Connect fluoride specific ion electrode and a reference electrode to ph meter. (Orion Single Junction reference Electrode and Filling Station) Measure 25 ml of sample or standrds and transfer to disposable beakers Add 25 ml TISAB II into each beaker before analyzing Calibrate meter with BLK, 0.5, 1.0, and 10 mg/l standards. Alloway Page 7 of 9

8 11.5 For each batch of 10 samples analyze a blank, check standard, duplicate and spiked sample Calculations and Quantitative Analysis 12.1 The electrode output is in concentration units, mg/l. No further calculations are necessary unless accounting for dilutions. The software for the Orion EA 940 converts the concentration in millivolts to mg/l In situations when a result must be generated from the data, the following procedure will be utilized: 1. The template entitled fluoride.xls is used and located in \conifer\excel\template folder 2. Go to the template and put the cursor inside the existing graph, right mouse click to remove the graph. 3. Using column A record the samples IDS for your analytical run. 4. Column B rows 2-4, contain the calibration standard concentrations. 5. In Column C rows 2-4, record the average of the millivolt readings for each calibration standard. 6. Highlight the six cells that contain the calibration standards and the millivolt readings. 7. Click on the Chart Wizard located in the tool bar 8. Highlight XY Scatter 9. Click on NEXT three (3) times. 10. Click on Finish 11. Click on CHART located on the tool bar 12. Scroll down and click on ADD TRENDLINE 13. Click on LOGARITHMIC 14. Click on OPTIONS 15. Click on DISPLAY EQUATION ON CHART 16. Click on OK 17. Enter the millivolt reading for each sample analyzed 12.3 Calculations will need to be added for non-aqueous samples. This results in dividing by the dry weight in the grams of the original sample Data Assessment 13.1 The primary analyst bears the responsibility of producing accurate data. A documented review is performed that includes a check of the following parameters: Calibration verification, QC criteria, calculation checks, Data entry into the LIMS 13.2 The initial data review is documented on the worksheet or benchsheet and dated by the primary analyst. An individual trained in the analytical procedure must perform a secondary review Data qualifiers are used whenever deviations occur while analyzing the samples. The qualifiers are included on the Certificate of Analysis that is presented to the client. Qualifiers include: samples that were not preserved, Alloway Page 8 of 9

9 failed LFM recovery, qualified due to LRB contamination, estimated or elevated reporting limit due to sample matrix interference, etc Report Generation 14.1 A Certificate of Analysis is generated after each analysis run is complete. The report includes the EPA Method Number, sample identification, date sampled, date logged in to the LIMS, quantitative result, worksheet number, analysis date, and analyst initials Waste Management and Pollution Prevention 15.1 Samples are disposed of according to the SOP on Sample Disposal. For those samples which are deemed to be hazardous materials, these must be lab packed and removed by a third party firm For information about pollution prevention that may be applicable to laboratory practices consult "Less is Better for Laboratory Chemical Management for Waste Reduction" available from the American Chemical Society Department of Government Relations and Science Policy. Address: th St. N.W. Washington, DC Reference: 16.1 US EPA Method 340.2; Measurement of Fluoride by ISE SM 4500 F-B, C Standard Methods 20 th edition, 1998 and previous editions Alloway Page 9 of 9

10 Alloway Page 10 of 9

11 Alloway Page 11 of 9

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method: ALLOWAY METHOD OUTLINE Standard Laboratory Method: SM Parameter: Methylene Blue Method: Colorimetric Reporting Level: Reference: 0.05 mg/l Standard Methods for the Examination Of Water and Wastewater;

More information

ALLOWAY METHOD OUTLINE

ALLOWAY METHOD OUTLINE ALLOWAY METHOD OUTLINE Standard Laboratory Method SM4500-Cl -G Parameter Residual Chlorine & Free Chlorine Method DPD Colorimetric Test Kit Date Issued Originator: Section Supervisor: QA Manager Date:

More information

METHOD #: Approved for NPDES and SDWA (Ed. Rev. 1974, 1978) Fluoride, Total (Colorimetric, SPADNS with Bellack Distillation)

METHOD #: Approved for NPDES and SDWA (Ed. Rev. 1974, 1978) Fluoride, Total (Colorimetric, SPADNS with Bellack Distillation) METHOD #: 340.1 Approved for NPDES and SDWA (Ed. Rev. 1974, 1978) TITLE: Fluoride, Total (Colorimetric, SPADNS with Bellack Distillation) ANALYTE: CAS # F Fluoride 7782-41-4 INSTRUMENTATION: Spectrophotometer

More information

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE 1.0 SCOPE AND APPLICATION 1.1 This method can be used for measuring total solubilized nitrate in drinking

More information

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Method 1061 Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Company Method 1061 Revision 1. December 015 Organic Carbon in Finished

More information

Method Abstract. Flow Solution Fluoride, USGS by ISE and FIA

Method Abstract. Flow Solution Fluoride, USGS by ISE and FIA Scope Summary Interferences Fluoride, USGS by ISE and FIA This method is used for the determination of fluoride in drinking water, surface water, and domestic and industrial wastes according to USGS Method

More information

Laboratory Techniques 100: Back To Basics. Carol Injasoulian Lab Manager City of Bay City April 29,2015

Laboratory Techniques 100: Back To Basics. Carol Injasoulian Lab Manager City of Bay City April 29,2015 Laboratory Techniques 100: Back To Basics Carol Injasoulian Lab Manager City of Bay City April 29,2015 QA/QC for Environmental Measurement What is quality assurance (QA)? Broad plan to maintain quality

More information

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water Hach Method 1041 Spectrophotometric Measurement of Free Chlorine (Cl ) in Finished Drinking Water Hach Company Method 1041 Revision 1. November 015 Spectrophotometric Measurement of Free Cl in Finished

More information

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Nitrate Method 10206 Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Method 10206 Revision 2.2 January 15, 2013 Spectrophotometric

More information

METHOD DETERMINATION OF SULFATE BY AUTOMATED COLORIMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

METHOD DETERMINATION OF SULFATE BY AUTOMATED COLORIMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division METHOD 375.2 DETERMINATION OF SULFATE BY AUTOMATED COLORIMETRY Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division Revision 2.0 August 1993 ENVIRONMENTAL MONITORING SYSTEMS

More information

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater. Nitrogen, Ammonia DOC316.53.01078 USEPA 1 Nessler Method 2 Method 8038 0.02 to 2.50 mg/l NH 3 N Reagent Solution Scope and application: For water, wastewater and seawater. Distillation is required for

More information

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT METHOD 9040B ph ELECTROMETRIC MEASUREMENT 1.0 SCOPE AND APPLICATION 1.1 Method 9040 is used to measure the ph of aqueous wastes and those multiphase wastes where the aqueous phase constitutes at least

More information

Standard Test Methods for Fluoride Ion in Water 1

Standard Test Methods for Fluoride Ion in Water 1 Designation: D 1179 99 An American National Standard Standard Test Methods for Fluoride Ion in Water 1 This standard is issued under the fixed designation D 1179; the number immediately following the designation

More information

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample. METHOD 9200 NITRATE 1.0 SCOPE AND APPLICATION 1.1 This method is applicable to the analysis of ground water, drinking, surface, and saline waters, and domestic and industrial wastes. Modification can be

More information

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure Troy, NY 12180 Date: 12/12/2017 Page 1 of 7 USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure 1. Scope and Application 1.1 Analytes Dissolved inorganic

More information

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University. Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared

More information

DR/4000 PROCEDURE. IRON, Total

DR/4000 PROCEDURE. IRON, Total Method 8365 Powder Pillows DR/4000 PROCEDURE FerroMo Method* (0 to 1.800 mg/l) Scope and Application: For cooling water containing molybdate-based treatment; digestion is required for determining total

More information

ALLOWAY STANDARD OPERATING PROCEDURES FOR METHOD 504.1/8011

ALLOWAY STANDARD OPERATING PROCEDURES FOR METHOD 504.1/8011 ALLOWAY STANDARD OPERATING PROCEDURES FOR METHOD 504.1/8011 1,2-DIBROMOETHANE (EDB) AND 1,2-DIBROMO-3-CHLOROPROPANE (DBCP) IN WATER BY MICROEXTRACTION AND GAS CHROMATOGRAPHY SOP #: 504.1 Revision 1.1,

More information

METHOD 13B - DETERMINATION OF TOTAL FLUORIDE EMISSIONS FROM STATIONARY SOURCES (SPECIFIC ION ELECTRODE METHOD)

METHOD 13B - DETERMINATION OF TOTAL FLUORIDE EMISSIONS FROM STATIONARY SOURCES (SPECIFIC ION ELECTRODE METHOD) 874 METHOD 13B - DETERMINATION OF TOTAL FLUORIDE EMISSIONS FROM STATIONARY SOURCES (SPECIFIC ION ELECTRODE METHOD) NOTE: This method does not include all of the specifications (e.g., equipment and supplies)

More information

LOW-RANGE TOTAL CHLORINE

LOW-RANGE TOTAL CHLORINE LOW-RANGE TOTAL CHLORINE Thermo Orion Method AC2072 Revision 4 3/14/2002 Accepted by U.S. EPA on March 20, 2002 for wastewater compliance monitoring Accepted by U.S. EPA on September 31, 2002 for drinking

More information

Approximate Volatile Acids by Titration

Approximate Volatile Acids by Titration SOP AMBL-101-A Page 1 of 5 Standard Operating Procedure AMBL-101-A Prepared: April 12, 2006 Revised: July 16, 2014 Prepared by: Terry E. Baxter Reviewed by: Approximate Volatile Acids by Titration METHOD

More information

METHOD DETERMINATION OF AMMONIA NITROGEN BY SEMI-AUTOMATED COLORIMETRY

METHOD DETERMINATION OF AMMONIA NITROGEN BY SEMI-AUTOMATED COLORIMETRY METHOD 350.1 DETERMINATION OF AMMONIA NITROGEN BY SEMI-AUTOMATED COLORIMETRY Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division Revision 2.0 August 1993 ENVIRONMENTAL MONITORING

More information

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste Experiment ISE: Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste 67 You have been hired by the government to check the fluoride concentration labelling on some major

More information

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions.

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions. Cole-Parmer Ammonium Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS Introduction The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium

More information

Titrimetric Determination of EASTMAN Color Developing Agent, CD-3, in Process ECN-2 Developer with Sulfato Cerate

Titrimetric Determination of EASTMAN Color Developing Agent, CD-3, in Process ECN-2 Developer with Sulfato Cerate Titrimetric Determination of EASTMAN Color Developing Agent, CD-3, in Process ECN-2 Developer Sulfato Cerate ECN-0003/1 Process ECN-2 ECP-2B VNF-1/LC RVNP Formulas SD-49 INTRODUCTION This method describes

More information

Total Dissolved Residue (Total Dissolved Solids, TDS) in Aqueous Matrices

Total Dissolved Residue (Total Dissolved Solids, TDS) in Aqueous Matrices Total Dissolved Residue (Total Dissolved Solids, TDS) in Aqueous Matrices Environmental Express 2345A Charleston Regional Parkway Charleston, SC 29492 800-343-5319 Table of Contents 1. Scope and Application...

More information

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2

Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2 Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall Corvallis,

More information

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

DR/4000 PROCEDURE NITRATE. Using Powder Pillows DR/4000 PROCEDURE Method 8171 Powder Pillows or AccuVac Ampuls Cadmium Reduction Method MR (0 to 5.0 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit

More information

SECTION D.2 AMMONIA NITROGEN

SECTION D.2 AMMONIA NITROGEN SECTION D.2 AMMONIA NITROGEN CEDR Method Code: NH4F L01 a) Scope and Application i) This method describes the determination of low-level ammonia nitrogen concentrations in filtered samples taken from fresh

More information

SOP-C-120 Determination of ph

SOP-C-120 Determination of ph i. Identification of the method a. SM 4500-H + (approved 2011) and EPA 9045C (approved 1997) ii. Applicable matrix or matrices a. Waters or aqueous solutions and suspensions iii. Limits of detection and

More information

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1 Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall Corvallis,

More information

STANDARD OPERATING PROCEDURE No. 52 STATIC ACID GENERATION (NAG) TEST

STANDARD OPERATING PROCEDURE No. 52 STATIC ACID GENERATION (NAG) TEST Questa Rock Pile Stability Study SOP 52v6 Page 1 STANDARD OPERATING PROCEDURE No. 52 STATIC ACID GENERATION (NAG) TEST REVISION LOG Revision Number Description Date 52v0 Original SOP by STM 6/9/2004 52v1

More information

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC) METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC) 1.0 SCOPE AND APPLICATION 1.1 Method 7196 is used to determine the concentration of dissolved hexavalent chromium [Cr(VI)] in EP/TCLP characteristic extracts

More information

DR/4000 PROCEDURE NITRATE. 2. The display will show:

DR/4000 PROCEDURE NITRATE. 2. The display will show: Method 8192 Powder Pillows DR/4000 PROCEDURE Cadmium Reduction Method LR (0 to 0.50 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit for program number

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical Oxygen Demand, Chemical DOC316.53.01103 USEPA Reactor Digestion Method Method 10211 1 to 60 mg/l COD (ULR) TNTplus 820 Scope and application: For wastewater, process water, surface water, and cooling water.

More information

Test Method: CPSC-CH-E

Test Method: CPSC-CH-E UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 10901 DARNESTOWN RD GAITHERSBURG, MD 20878 Test Method: CPSC-CH-E1001-08 Standard Operating Procedure

More information

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3 , MR, 8171 DOC316.53.01069 Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3 N) Powder Pillows or AccuVac Ampuls Scope and Application: For water, wastewater and seawater Test preparation

More information

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0) 150mV 300mV Mono/Reference Combination S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) 1376 340 456 Fax: +44 (0) 1376 340 453 Email:

More information

Timberline Ammonia-001. Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis

Timberline Ammonia-001. Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis Timberline Ammonia-001 Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis Prepared By: Dr. Edward F. Askew Date: June 24, 2011 Table of Contents 1. SCOPE

More information

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ ) DR/4000 PROCEDURE Method 8023 1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to 0.700 mg/l Cr 6+ ) Scope and Application: For water and wastewater;usepa accepted for reporting for

More information

Procedure for Body Fluid Unit Quality Control

Procedure for Body Fluid Unit Quality Control Procedure for Body Fluid Unit Quality Control 1.0 Purpose - This procedure specifies the required elements needed for the preparation and quality control measures for reagents used for body fluid identification.

More information

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) 1.0 SCOPE AND APPLICATION 1.1 Method 7060 is an atomic absorption procedure approved for determining the concentration of arsenic in wastes,

More information

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

Scope and application: For wastewater, seawater, drinking water, surface water and process water. Cyanide DOC316.53.01495 USEPA 1 Pyridine Barbituric Acid Method 2 Method 10265 0.01 to 0.60 mg/l CN TNTplus 862 Scope and application: For wastewater, seawater, drinking water, surface water and process

More information

Spectrophotometric Determination of Ferrocyanide in Effluents

Spectrophotometric Determination of Ferrocyanide in Effluents Spectrophotometric Determination of Ferrocyanide in Effluents ECN-0025-1 INTRODUCTION This method is used to determine the concentration of ferrocyanide ion in photoprocessing solution effluents. The ion

More information

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows , 8326 Eriochrome Cyanine R Method 1 Scope and Application: For water 1 Adapted from Standard Methods for the Examination of Water and Wastewater. DOC316.53.01003 Method 8326 (0.006 to 0.250 mg/l Al 3+

More information

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water. Method 8337 Powder Pillows DR/4000 TT PROCEDURE QUATERNARY AMMONIUM COMPOUNDS Direct Binary Complex Method (0 to 5.00 mg/l as CTAB) Scope and Application: For cooling tower and pool/spa water. The estimated

More information

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7 Troy, NY 12180 Date: 03/16/2012 Page 1 of 7 USGS District Laboratory, Troy, NY Inductively Coupled Plasma Optical Emission Spectrometry Standard Operating Procedure 1. Scope and Application 1.1 Analytes

More information

Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES Version 2.

Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES Version 2. Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES 701.0 Version 2.3 July 2018 1. SCOPE AND APPLICATION This method is used for the determination

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical Oxygen Demand, Chemical DOC316.53.01104 USEPA Reactor Digestion Method Method 10212 250 to 15,000 mg/l COD (UHR) TNTplus 823 Scope and application: For wastewater and process waters; digestion is required.

More information

METHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE)

METHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE) METHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE) 1.0 SCOPE AND APPLICATION 1.1 This automated method is applicable to ground water, drinking and surface waters, and domestic and industrial wastes

More information

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,: sop-c-102 Determination of Chemical Oxygen Demand Revision 6 Approval: Laborato ry Manager/LQAO/RS O Date Z/zs Date Z-z* I A Renewat date: 7lt:,: Effective date: 3, tnitiats: wu\ Texas lnstitute for Applied

More information

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows Nitrate DOC316.53.01067 Cadmium Reduction Method Method 8192 0.01 to 0.50 mg/l NO 3 N (LR) Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7 Troy, NY 12180 Date: 03/14/2012 Page 1 of 7 USGS New York Water Science Center Soil and Low Ionic Strength Water Quality Laboratory Electrometric Measurements and Titration Standard Operating Procedure

More information

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows Silver DOC316.53.01134 Colorimetric Method Method 8120 0.02 to 0.70 mg/l Ag Powder Pillows Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows

More information

1) A ph/mv meter or an ion meter, either line operated or portable.

1) A ph/mv meter or an ion meter, either line operated or portable. COLE-PARMER Sodium Ion Electrodes Instruction Manual GENERAL INSTRUCTIONS The Cole-Parmer Sodium Ion Electrodes are used to quickly, simply, accurately, and economically measure sodium ion concentrations

More information

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge Standard Operating Procedure for GLNPO Specific Conductance: Marvin Palmer United States Environmental Protection Agency Great Lakes National Program Office Region 5 77 West Jackson 9th Floor Chicago,

More information

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right. Aluminum DOC316.53.01003 Eriochrome Cyanine R Method 1 Method 8326 0.006 to 0.250 mg/l Al 3+ Powder Pillows Scope and application: For water. 1 Adapted from Standard Methods for the Examination of Water

More information

Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and

Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared by: OEWRI Laboratory Manager

More information

Potentiometry: Ion-Selective Electrodes

Potentiometry: Ion-Selective Electrodes Potentiometry: Ion-Selective Electrodes Determination of the sodium monofluorophosphate content of a toothpaste sample using a fluoride ion-selective electrode. 1. Aims The aims of the experiment are to:

More information

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Nitrate, MR DOC316.53.01069 Cadmium Reduction Method Method 8171 0.1 to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Scope and application: For water, wastewater

More information

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction II. Quality Control A. Introduction ENCO adheres to strict quality control practices in order to assure our clients that the data provided are accurate and reliable. We are required by the EPA to analyze

More information

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation Phosphorus, Total DOC316.53.01121 USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method 8190 0.06 to 3.50 mg/l PO 3 4 (0.02 to 1.10 mg/l P) Test N Tube Vials Scope and application: For water,

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES STANDARD OPERATING PROCEDURES for Spectrophotometric Measurement Chl a, Chl b, and Bulk Carotenoids using Methanol Solvent Prepared by EcoAnalysts, Inc. 1420 South Blaine Street, Suite 14 Moscow, ID 83843

More information

Glossary of Common Laboratory Terms

Glossary of Common Laboratory Terms Accuracy A measure of how close a measured value is to the true value. Assessed by means of percent recovery of spikes and standards. Aerobic Atmospheric or dissolved oxygen is available. Aliquot A measured

More information

METHOD DETERMINATION OF TOTAL CYANIDE BY SEMI-AUTOMATED COLORIMETRY

METHOD DETERMINATION OF TOTAL CYANIDE BY SEMI-AUTOMATED COLORIMETRY METHOD 335.4 DETERMINATION OF TOTAL CYANIDE BY SEMI-AUTOMATED COLORIMETRY Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division Revision 1.0 August 1993 ENVIRONMENTAL MONITORING

More information

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ).

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ). Method 8194 DR/4000 PROCEDURE Diaminobenzidine Method* (0 to 1.000 mg/l) Scope and Application: For water and wastewater; distillation is required for determining total selenium. See the Distillation procedure

More information

Operating Instructions for Ammonium ISE Specifications

Operating Instructions for Ammonium ISE Specifications Operating Instructions for Ammonium ISE Specifications Range: The Ammonium Electrode responds to uncomplexed ion activity over the range 1 x 10-1M to 1 x 10-6M. Linear detection limit is about 10-5M Interference's:

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Ammonium Ion Selective Electrode is used to quickly, simply, accurately, and economically measure potassium in aqueous solutions. Required

More information

DPD Test N Tube Method *

DPD Test N Tube Method * CHLORINE, FREE (0 to 5.00 mg/l) DPD Test N Tube Method * Method 10102 For water, wastewater, and seawater 1. Enter the stored program number for Test N Tube free chlorine (Cl 2 ). Press: PRGM The display

More information

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY.

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY. 1 REVISOR 4740.2100 4740.2100 QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY. Subpart 1. Scope. This part applies to laboratories performing testing under the inorganic chemistry, metals,

More information

Acid Volatile Sulfides and Simultaneously Extracted Metals in Sediments

Acid Volatile Sulfides and Simultaneously Extracted Metals in Sediments Title: AVS-SEM Page 1 of 8 Acid Volatile Sulfides and Simultaneously Extracted Metals in Sediments Reference: Determination of acid-volatile sulfide and simultaneously-extracted metals in sediments using

More information

CH2MHILL GILROY/MORGAN HILL LABORATORY LABORATORY STANDARD OPERATING PROCEDURE NO TOTAL KJELDAHL NITROGEN - TKN SM 4500-N org.

CH2MHILL GILROY/MORGAN HILL LABORATORY LABORATORY STANDARD OPERATING PROCEDURE NO TOTAL KJELDAHL NITROGEN - TKN SM 4500-N org. CH2MHILL GILROY/MORGAN HILL LABORATORY LABORATORY STANDARD OPERATING PROCEDURE NO. 709 TOTAL KJELDAHL NITROGEN - TKN SM 4500-N org. B I. SCOPE AND APPLICATIONS A. The kjeldahl method determines nitrogen

More information

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials Nitrogen, Total DOC316.53.01086 Persulfate Digestion Method Method 10071 0.5 to 25.0 mg/l N (LR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific

More information

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials Nitrogen, Total DOC316.53.01085 Persulfate Digestion Method Method 10072 2 to 150 mg/l N (HR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific information

More information

METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE)

METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE) METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE) 1.0 SCOPE AND APPLICATION 1.1 This method is applicable to ground water, drinking, surface, and saline waters, and domestic and industrial wastes.

More information

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2 Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State

More information

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters) Aluminum DOC316.53.01002 Aluminon Method 1 Method 8012 0.008 to 0.800 mg/l Al 3+ (spectrophotometers) Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters) Scope and application: For water and wastewater.

More information

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA , 10255 DOC316.53.01322 Direct Measurement ISE Method Method 10255 3.55 g/l to 35 g/l Cl Powder Pillow ISA Scope and Application: For the determination of high concentrations (1 M) of chloride in brine

More information

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8030 is used to determine the concentration of the following volatile organic compounds: Compound Name

More information

COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

Method to 0.50 mg/l NH 3 N Powder Pillows

Method to 0.50 mg/l NH 3 N Powder Pillows , 8155 Salicylate Method 1 Scope and Application: For water, wastewater and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) DOC316.53.01077 Method 8155 0.01 to 0.50 mg/l NH 3 N Powder Pillows

More information

Chloramine (Mono) and Nitrogen, Free Ammonia

Chloramine (Mono) and Nitrogen, Free Ammonia Chloramine (Mono) and Nitrogen, Free Ammonia DOC316.53.01016 Indophenol Method 1 Method 10200 0.04 to 4.50 mg/l Cl 2 0.01 to 0.50 mg/l NH 3 N Powder Pillows Scope and application: For the determination

More information

(b) Write the chemical equation for the dissolution of NaOH in water.

(b) Write the chemical equation for the dissolution of NaOH in water. Making a Solution and Measuring ph Prelab Assignment Read the entire lab. Write an objective and any hazards associated with this lab in your laboratory notebook. Answer the following 6 questions in your

More information

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls) Method 8171 NITRATE, Mid Range (0 to 5.0 mg/l NO 3- -N) For water, wastewater and seawater* Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls) Using Powder Pillows 1. Enter the stored program

More information

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

Standard Operating Procedure for the Analysis of Nitrate/Nitrite and Nitrite in Fresh Waters CCAL 31B.1

Standard Operating Procedure for the Analysis of Nitrate/Nitrite and Nitrite in Fresh Waters CCAL 31B.1 Standard Operating Procedure for the Analysis of Nitrate/Nitrite and Nitrite in Fresh Waters CCAL 31B.1 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson

More information

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters) Chromium, Total DOC316.53.01034 Alkaline Hypobromite Oxidation Method 1 Method 8024 0.01 to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters) Scope and application: For water and wastewater.

More information

METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV)

METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV) METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV) 1.0 SCOPE AND APPLICATION 1.1 Method 9012 is used to determine the concentration of inorganic cyanide in an aqueous waste or leachate.

More information

Method to 0.50 mg/l NH 3 N Powder Pillows

Method to 0.50 mg/l NH 3 N Powder Pillows , 8155 Salicylate Method 1 Scope and Application: For water, wastewater and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) DOC316.53.01077 Method 8155 0.01 to 0.50 mg/l NH 3 N Powder Pillows

More information

perfection Guidebook perfection Combination Fluoride Electrode Successful Ion Measurement

perfection Guidebook perfection Combination Fluoride Electrode Successful Ion Measurement perfection Guidebook perfection Combination Fluoride Electrode Successful Ion Measurement Contents Table of Contents 1. Introduction 1 2. Required Equipment 2 3. Electrode and Measurement Setup 4 Electrode

More information

USGS Troy WSC Laboratory Ion Chromatography SOP EPA 300.0, Rev Jordan Road SOP. No. 1 Rev. No. 1.7 Troy, NY Date 01/25/18 Page 1 of 7

USGS Troy WSC Laboratory Ion Chromatography SOP EPA 300.0, Rev Jordan Road SOP. No. 1 Rev. No. 1.7 Troy, NY Date 01/25/18 Page 1 of 7 Troy, NY 12180 Date 01/25/18 Page 1 of 7 USGS Water Science Center Laboratory, Troy, NY Ion Chromatography Standard Operating Procedure 1. Scope and Application 1.1 Analytes Chloride, Nitrate, and Sulfate

More information

NaOH (aq) + HCl (aq) NaCl (aq) + H 2 O (l)

NaOH (aq) + HCl (aq) NaCl (aq) + H 2 O (l) EXPERIMENT 21 Molarity of a Hydrochloric Acid Solution by Titration INTRODUCTION Volumetric analysis is a general term meaning any method in which a volume measurement is the critical operation; however,

More information

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID) Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID) Table of Contents Section 1: Summary of Method Section 2: Scope and Application Section

More information

ACZ Laboratories, Inc Downhill Drive Steamboat Springs, CO (800)

ACZ Laboratories, Inc Downhill Drive Steamboat Springs, CO (800) 2773 Downhill Drive Steamboat Springs, CO 80487 (800) 334-5493 Analytical Report November 19, 2004 Report to: Debbie Furlow 602 Lehmberg Road Columbus, MS 39702 Bill to: Accounts Payable PO Box 2487 Columbus,

More information

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION METHOD 80 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION 1.0 SCOPE AND APPLICATION 1.1 Method 80 may be used to determine the concentration of acetonitrile (CAS No. 75-05-8) in aqueous

More information

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Organics Revision Date: July 19, 2017 Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Parameter Perfluorinated Alkyl Acids (Perfluorobutane Sulphonate (PFBS), Perflourooctane Sulphonate (PFOS),

More information

Standard Operating Procedure VUSP M

Standard Operating Procedure VUSP M Villanova University Date Nov 2009 Page 1 of 17 Villanova University Villanova Urban Stormwater Partnership Watersheds Laboratory Standard Operating Procedure VUSP M Nitrate by Discrete Analysis Systea

More information