How to develop an ATP.... and verify it has been met. Melissa Hanna-Brown Analytical R & D; Pfizer Global R+D, Sandwich UK.

Size: px
Start display at page:

Download "How to develop an ATP.... and verify it has been met. Melissa Hanna-Brown Analytical R & D; Pfizer Global R+D, Sandwich UK."

Transcription

1 How to develop an ATP... and verify it has been met Melissa Hanna-Brown Analytical R & D; Pfizer Global R+D, Sandwich UK. 06 May 2014

2 Overview Some context accuracy, precision and risk What is an ATP? What are the elements of an ATP? an example: ATP for assay an example: ATP for purity An overview of ATP verification steps 5/27/2014 2:44 PM 2

3 Some context accuracy, precision and risk 5/27/2014 2:44 PM 3

4 Back to basics some context Purpose of a method? to provide a number with an associated confidence which describes the level of an analyte (e.g. in a starting material/process intermediate/api/dp etc) Why do we need this number? to understand how the level of the analyte corresponds to pre-defined threshold(s) (specifications) How do we determine the associated confidence? by understanding the total contributions to variability How do we improve our associated confidence? by minimising variability (through in depth understanding ) 5/27/2014 2:44 PM 4

5 QbD, ICH Q8-10 quality and methods... LSL USL In broad terms, QUALITY encompasses the TOTALITY of features of a process or product (w.r.t. properties, attributes, capabilities). To be able to understand the TOTALITY of features an in-depth understanding must be achieved. LSL Analytical variation Process variation USL σ 2 (measured) = σ 2 (process) + σ 2 (method) Process variation Analytical variation 5/27/2014 2:44 PM 5

6 The importance of Accuracy and Precision criteria in context to other method performance criteria *Modified from a slide presented by Oliver Grosche, Pittcon Mar-2013 Analytical Change Control Result ATP Decision Results: Accuracy and Precision Method Performance Method Performance Data: Method Range Specificity/Peak Resolution Linearity Limit of Detection Instrument Repeatability 5/27/2014 2:44 PM 6

7 Method Performance: Bias and Precision The figure represents the distribution of 95% of the measurements given by four different hypothetical procedures having each a true bias and a true precision. If acceptance criteria is: µ(true) - µ(measured) 15% From Hubert, et al. J Pharm. Biomed. Anal. 36 (2004) 579 Which procedures fulfill this objective and which ones will the analyst retain as valid? 5/27/2014 2:44 PM 7

8 What is an ATP? 5/27/2014 2:44 PM 8

9 ATP? The ATP describes the performance characteristics of the method... such that the data the method produces will be fit for purpose...for e.g. making decisions on about whether a batch of drug substance or drug product meets the specification criteria for assay or purity. 5/27/2014 2:44 PM 9

10 elements to incorporate in an ATP The total uncertainty, expressed in terms of systematic (accuracy) and random (precision) uncertainty (variability) components. A statement which links to the level of risk associated with making incorrect decisions (probability statement). The range for which the analytical method is expected to quantify the analyte. A description of the analyte to be tested, including the sample or matrix in which it will be tested. 5/27/2014 2:44 PM 10

11 An example...atp for assay Example ATP for Drug Product Assay The procedure must be able to accurately quantify PF-1234 in film coated tablets over the range of % of the nominal concentration with accuracy, and precision such that measurements fall within ± 3.0% of the true value with 95% probability. 5/27/2014 2:44 PM 11

12 An example...atp for assay Example ATP for API Assay The procedure must be able to accurately quantify PF-1234 Hydrochloride Drug Substance over a range of 90% to 110% of the nominal concentration with accuracy, and precision such that measurements fall within ± 2.0% of the true value with at least a 95% probability. 5/27/2014 2:44 PM 12

13 Graphical Representation of Traditional Method Validation Criteria Accuracy and Precision with Independent limits/criterion At boundaries: no trade-off between increasing method bias and/or variability It is possible to accept a method with both high bias and high variability Criteria: Bias: NMT 2.0% Variability: NMT 1.25% 5/27/2014 2:44 PM 13

14 Justification of Assay ATP ATP Criterion vs unconstrained {u,sigma} The procedure must be able to accurately quantify PF-1234 Hydrochloride Drug Substance over a range of 90% to 110% of the nominal concentration with accuracy, and precision such that measurements fall within ± 2.0% of the true value at least a 95% probability. 5/27/2014 2:44 PM 14

15 ATP criterion vs Traditional {µ,sigma} sets 25% 10% 5% 5% 5% 5% 5/27/2014 2:44 PM 15

16 Justification of Assay ATP ATP Equation The expression for defining the acceptance region is the following: where, µ = true mean / accuracy; a parameter σ = true sigma / precision; a parameter e = allowable analytical window, a fixed constant y = individual assay value; a random variable with mean µ and standard deviation σ T = true analytical content, fixed target p = minimum probability for individual assay to reside within error bound e, fixed constant Φ(.)= normal density function centered at µ, with standard deviation σ 5/27/2014 2:44 PM 16

17 Interactive ATP Tool - Modeling the Trade-off between Accuracy and Precision API Specification % Process Mean = 100.0% Process std dev = 0.2% ATP criteria: ± 2.0 from the true value with a 95% probability If method bias = 0 then the method standard deviation should be 1.0% (area under the curve that is OOS is ~ 5%) 5/27/2014 2:44 PM 17

18 Interactive ATP Tool Modeling the Trade-off between Accuracy and Precision API Specification % Process Mean = 100.0% Process std dev = 0.2% ATP criteria: ± 2.0 from the true value with a 95% probability If method bias = 1.0% then the method standard deviation should be 0.6% (area under the curve that is OOS is ~ 6%) 5/27/2014 2:44 PM 18

19 Interactive ATP Tool Modeling the Trade-off between Accuracy and Precision - with Replicates API Specification % Process Mean = 100.0% Process std dev = 0.2% ATP criteria: ± 2.0 from the true value with a 95% probability Adding another replicate (n=2) reduces the probability of an OOS result from ~ 6% to 1%. 5/27/2014 2:44 PM 19

20 An example - ATP for Quantitation of Impurities Specifications: Impurity A NMT 0.15% Impurity B NMT 0.15% The procedure must be able to accurately quantify Impurity A and Impurity B relative to PF-1234 Hydrochloride in the presence of drug substance and other impurities over a range from the reporting threshold through twice the specification limit. The accuracy and precision of the procedure must be such that the measurements fall within ± 15% of the true value for impurity levels 0.15% with at least 90% probability, and within ±10% of the true value for impurity levels > 0.15% with at least 90% probability. 5/27/2014 2:44 PM 20

21 Graphical representation of the Purity ATP statement Impurities > 0.15% Impurities 0.15% 5/27/2014 2:44 PM 21

22 An overview of ATP verification steps 5/27/2014 2:44 PM 22

23 Verification Strategy Sampling Sample Preparation = Method Verification Measurement Method ATP 5/27/2014 2:44 PM 23

24 Chromatographic Performance Verification DoEs The first DoE experiment(s) are run during the method evaluation part of the method development lifecycle the important focus of these experiments are parameter interactions and parameter criticality. Driven by outcome of risk assessment Desired outcome = draft MODR (method operable design region) 5/27/2014 2:44 PM 24

25 Chromatographic Method Verification Strategy 1 Screening DoE s Chromatographic Performance Criteria Draft MODR + 2 Verification DoE (s) ATP Verified MODR = Chromatographic Verification 5/27/2014 2:44 PM 25

26 ATP Verification DoEs Following proposal of the MODR ATP verification must be considered. Here the resulting MODR from the screening DoEs is used and multiple verification conditions selected across this MODR using samples representing the range over which the method needs to be able to perform all tested against the ATP criteria. Method parameters not deemed critical from the screening DoEs can be left out of the verification DoE (s). An important consideration during this verification is intermediate precision i.e. performing the verification on different instruments/columns, on different days with different operators. These experiments can result in further refinement of the MODR each time the MODR is refined experimental verification of ATP compliance is required. 5/27/2014 2:44 PM 26

27 Visualisation - ATP Verification Lab 2 Lab 1 5/27/2014 2:44 PM 27 27

28 Conclusions ATPs should direct the design, development and continuous verification of methods throughout the lifecycle of a method. The principle of ATPs are aligned with QbD for process - akin to a QTTP ATPs are also aligned with other industries with respect to the growing recognition of the need to understand the confidence associated with a number describing a sample attribute which has been derived from a measurement method. This is especially important when this number drives a pass/fail decision. There are currently no regulatory or industry guidances on ATPs (It s challenging to prosecute AQbD in the current global regulatory environment where requirements differ across multiple markets a paradigm change via global guidance in this area would really facilitate continuous improvement in method lifecycles). The prospective setting of an ATP allows for the total acceptable variability of the method to be defined up front. Verification of the whole method as complying with the ATP requires a depth of method understanding to be generated. 5/27/2014 2:44 PM 28

29 Thank-you for your attention 5/27/2014 2:44 PM 29

30 Acknowledgements a team effort Dave de Antonis Graham Cook Loren Wrisley Kimber Barnett Brent Harrington Tim Graul Jim Morgado Steve Colgan Debbie Kraus Chuck Melucci Neil Clayton 5/27/2014 2:44 PM 30

31 References BIPM, JICM 100: Guide to the expression of uncertainty in measurement (GUM), ISO/IEC (2008). ISO 21748:2010, Guidance for the use of repeatability, reproducibility and trueness estimates in measurement uncertainty estimation (2010). S.L.R. Ellison, M.Rosslein, A. Williams. Quantifying Uncertainty in Analytical Measurement in EURACHEM/CITAC Guide*, (2 nd ed., QUAM 2000.P1). V.R. Meyer. Measurement Uncertainty. J Chromatogr. A. 1158, (2007). ISO/IEC 17025:2005, General requirements for the competence of testing and calibration laboratories (2010). W. Horwitz, R. Albert. Quality IssuesThe Concept of Uncertainty as Applied to ChemicalMeasurements Analyst. 122, (1997). M. Feinberg, B. Boulanger, W. Dewe, P. Hubert. New advances in method validation and measurement uncertainty aimed at improving the quality of chemical data Anal Bioanal Chem. 380, (2004). J. Wallace. Ten methods for calculating the uncertainty of measurement. Science & Justice. 50, (2010). 5/27/2014 2:44 PM 31

32 A case study why ATPs matter 5/27/2014 2:44 PM 32

33 Case Study : Product X What if?... Drug Product X successfully validated and transferred per ICH Q2a principles. Method globally registered and accepted. Actual Method Performance per retrospective review of validation report Method was NOT fit for purpose was found to be ROOT 1 cause of type 1 commercial batch failure. (Potency < 95.0% of claim). Method falls outside of a more restrictive ATP acceptance criterion of within 3% of the true value with 95% probability. 5/27/2014 2:44 PM 33

34 0.3 Case Study : Product X What if?... Using the described Analytical Target Profile (ATP) evaluation tool and model scenarios based on method (combined accuracy/precision) and process performance - predicts a 3-5% OOS rate with the legacy Product X method Distribution of Results Including Analytical Method Variation % LC Individual Results Mean of 1 Reps 5/27/2014 2:44 PM 34

35 Case Study : Product X What if?... A simple change to add replicate analyses (even with the existing method), would decrease risk to < 0.1% according to modeling. QbD method design principles would have foreseen the method problem and help drive continuous improvement. 0.3 Distribution of Results Including Analytical Method Variation % LC Individual Results Mean of 3 Reps 5/27/2014 2:44 PM 35

36 New Method Developed with AQbD Principles Sample Preparation Target Profile Simple straightforward sample preparation process to quantitatively extract active The extraction solvent, extraction technique, sonication, and shaking time were studied to optimize extraction. The extraction development had 2 primary targets, which were: (1) intermediate precision of less than 1% RSD (2) spiking recovery of 100%. Additionally, the working sample solvent composition/strength choice was geared to obtain symmetric peaks for all components under the chosen separation conditions. For degradation analysis, the working sample concentration was to be optimized in order to gain additional sensitivity. 5/27/2014 2:44 PM 36

37 New Method for Product X Developed with QbD Principles Combined accuracy and precision estimate lies within this ATP acceptance criterion Method continues to run well in commercial setting. Old Method New Method 5/27/2014 2:44 PM 37

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY Table of Contents 1. SCOPE... 2 2. REFERENCES... 2 3. TERMS AND DEFINITIONS... 2 4. BACKGROUND... 4 5. EVALUATION OF MEASUREMENT UNCERTAINTY POLICY... 5

More information

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT Table of Contents 1. SCOPE... 2 2. REFERENCES... 2 3. TERMS AND DEFINITIONS... 2 4. BACKGROUND... 4 5. ESTIMATION OF UNCERTAINTY OF MEASUREMENT POLICY...

More information

Analytical Method Validation: An Updated Review

Analytical Method Validation: An Updated Review INTERNATIONAL JOURNAL OF ADVANCES IN PHARMACY, BIOLOGY AND CHEMISTRY Analytical Method Validation: An Updated Review G. Geetha, Karanam Naga Ganika Raju, B. Vignesh Kumar and M. Gnana Raja* Sankaralingam

More information

Report and Manage Post Marketing Changes to an Approved NDA, ANDA and BLA. Jane Weitzel Independent Consultant

Report and Manage Post Marketing Changes to an Approved NDA, ANDA and BLA. Jane Weitzel Independent Consultant Report and Manage Post Marketing Changes to an Approved NDA, ANDA and BLA Jane Weitzel mljweitzel@msn.com Independent Consultant IVT S Analytical Procedures & Methods Validation December 2016 San Diego

More information

Application of Gauge R&R Methods for Validation of Analytical Methods in the Pharmaceutical Industry

Application of Gauge R&R Methods for Validation of Analytical Methods in the Pharmaceutical Industry Application of Gauge R&R Methods for Validation of Analytical Methods in the Pharmaceutical Industry Richard K Burdick Elion Labs QPRC Meetings June 2016 Collaborators David LeBlond, CMC Statistical Consultant

More information

serve the goal of analytical lmethod Its data reveals the quality, reliability and consistency of

serve the goal of analytical lmethod Its data reveals the quality, reliability and consistency of Analytical method validation By Juree Charoenteeraboon, Ph.D Analytical method Goal consistent, reliable and accurate data Validation analytical method serve the goal of analytical lmethod Its data reveals

More information

Method Validation Characteristics through Statistical Analysis Approaches. Jane Weitzel

Method Validation Characteristics through Statistical Analysis Approaches. Jane Weitzel Method Validation Characteristics through Statistical Analysis Approaches Jane Weitzel mljweitzel@msn.com 1:00 to 2:30 Wednesday, Dec. 9 SESSION 6 ANALYTICAL PROCEDURES AND METHOD VALIDATION mljweitzel@msn.com

More information

ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES

ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES Adopted from ICH Guidelines ICH Q2A: Validation of Analytical Methods: Definitions and Terminology, 27 October 1994. ICH Q2B: Validation of Analytical

More information

Chapter 4: Verification of compendial methods

Chapter 4: Verification of compendial methods Chapter 4: Verification of compendial methods Introduction In order to ensure accurate and reliable test results, the quality control laboratory (QCL) needs to use analytical methods (and accompanying

More information

Traceability, validation and measurement uncertainty 3 pillars for quality of measurement results. David MILDE

Traceability, validation and measurement uncertainty 3 pillars for quality of measurement results. David MILDE Traceability, validation and measurement uncertainty 3 pillars for quality of measurement results David MILDE Traceability Validation Measurement Uncerainty Outline Introduction to 3 pillars Recommended

More information

International Journal of PharmTech Research CODEN (USA): IJPRIF, ISSN: , ISSN(Online): Vol.9, No.7, pp , 2016

International Journal of PharmTech Research CODEN (USA): IJPRIF, ISSN: , ISSN(Online): Vol.9, No.7, pp , 2016 International Journal of PharmTech Research CODEN (USA): IJPRIF, ISSN: 097-30, ISSN(Online): 255-9563 Vol.9, No.7, pp 399-06, 2016 Analytical Quality by Design Approach for Development of UV-Spectrophotometric

More information

Precision estimated by series of analysis ISO and Approach Duplicate Approach

Precision estimated by series of analysis ISO and Approach Duplicate Approach Agenda Item 9 JOINT FAO/WHO FOOD STANDARDS PROGRAMME CODEX COMMITTEE ON METHODS OF ANALYSIS SAMPLING Thirty-seventh th Session Budapest, Hungary, 6 February 016 (Comments prepared by the German Delegation)

More information

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction II. Quality Control A. Introduction ENCO adheres to strict quality control practices in order to assure our clients that the data provided are accurate and reliable. We are required by the EPA to analyze

More information

Quality by Design and Analytical Methods

Quality by Design and Analytical Methods Quality by Design and Analytical Methods Isranalytica 2012 Tel Aviv, Israel 25 January 2012 Christine M. V. Moore, Ph.D. Acting Director ONDQA/CDER/FDA 1 Outline Introduction to Quality by Design (QbD)

More information

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES Introduction DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES This Technical Document shall be applied to the quantitative determination of a Threshold Substance in a Sample

More information

DEMONSTRATING CAPABILITY TO COMPLY WITH A TEST PROCEDURE: THE CONTENT UNIFORMITY AND DISSOLUTION ACCEPTANCE LIMITS (CUDAL) APPROACH

DEMONSTRATING CAPABILITY TO COMPLY WITH A TEST PROCEDURE: THE CONTENT UNIFORMITY AND DISSOLUTION ACCEPTANCE LIMITS (CUDAL) APPROACH 1 DEMONSTRATING CAPABILITY TO COMPLY WITH A TEST PROCEDURE: THE CONTENT UNIFORMITY AND DISSOLUTION ACCEPTANCE LIMITS (CUDAL) APPROACH Jim Bergum September 12, 2011 Key Responses For Batch Release 2 Potency

More information

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC.

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC. 7. Stability indicating analytical method development and validation of and in capsule dosage form by HPLC. 7.1 INSTRUMENTS AND MATERIALS USED 7.1.1 INSTRUMENTS 1. Shimadzu LC-2010 CHT with liquid chromatograph

More information

Regulatory and Alternative Analytical Procedures are defined as follows 2 :

Regulatory and Alternative Analytical Procedures are defined as follows 2 : Title: Alternative Analytical Method Validation in Pharmaceuticals: Replacing a Regulatory Analytical Method in Cleaning Validation Authors: Stephen Lawson, Will McHale and Brian Wallace This re titled

More information

Proposed Procedures for Determining the Method Detection Limit and Minimum Level

Proposed Procedures for Determining the Method Detection Limit and Minimum Level Proposed Procedures for Determining the Method Detection Limit and Minimum Level Published by: ACIL Environmental Services Section Technical Committee Revision 3.0 3/8/006 PROCEDURES These procedures set

More information

To control the consistency and quality

To control the consistency and quality Establishing Acceptance Criteria for Analytical Methods Knowing how method performance impacts out-of-specification rates may improve quality risk management and product knowledge. To control the consistency

More information

VAM Project Development and Harmonisation of Measurement Uncertainty Principles

VAM Project Development and Harmonisation of Measurement Uncertainty Principles VAM Project 3.2.1 Development and Harmonisation of Measurement Uncertainty Principles Part (d): Protocol for uncertainty evaluation from validation data V J Barwick and S L R Ellison January 2000 LGC/VAM/1998/088

More information

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES Date: 17 May, 2012 Effective Date: 1 October, 2012 Introduction DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES This Technical Document shall be applied to the quantitative

More information

Reproducibility within the Laboratory R w Control Sample Covering the Whole Analytical Process

Reproducibility within the Laboratory R w Control Sample Covering the Whole Analytical Process Flowchart for Nordtest Method (a) Specify measurand Quantify components for within lab reproducibility A control samples B possible steps, not covered by the control sample Quantify bias components Convert

More information

Quality by Design in the Development of Analytical Procedures

Quality by Design in the Development of Analytical Procedures Quality by Design in the Development of Analytical Procedures A discussion has been growing with a view to applying the Quality by Design (QbD) concept to the development of analytical procedures and promoting

More information

INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE

INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE TEXT ON VALIDATION OF ANALYTICAL PROCEDURES Recommended

More information

The Uncertainty of Reference Standards

The Uncertainty of Reference Standards 2009 Cerilliant Corporation The Uncertainty of Reference Standards A Guide to Understanding Factors Impacting Uncertainty, Uncertainty Calculations and Vendor Certifications Authors: Ning Chang PhD, Isil

More information

Review Article Analytical Quality by Design Approach to Test Method Development and Validation in Drug Substance Manufacturing

Review Article Analytical Quality by Design Approach to Test Method Development and Validation in Drug Substance Manufacturing Chemistry Volume 2015, Article ID 435129, 8 pages http://dx.doi.org/10.1155/2015/435129 Review Article Analytical Quality by Design Approach to Test Method Development and Validation in Drug Substance

More information

Validation and Standardization of (Bio)Analytical Methods

Validation and Standardization of (Bio)Analytical Methods Validation and Standardization of (Bio)Analytical Methods Prof.dr.eng. Gabriel-Lucian RADU 21 March, Bucharest Why method validation is important? The purpose of analytical measurement is to get consistent,

More information

Ivo Leito University of Tartu

Ivo Leito University of Tartu Using method validation and performance data for estimating measurement uncertainty Ivo Leito University of Tartu ivo.leito@ut.ee.05.01 1 Overview The main question of uncertainty evaluation The different

More information

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC 17025 Table of contents 1 GENERAL... 2 2 THE STANDARD SIST EN ISO/IEC 17025... 2 3 SA'S POLICY IN IMPLEMENTING

More information

Validation of TNCO, TSNAs, and PAHs

Validation of TNCO, TSNAs, and PAHs Validation of TNCO, TSNAs, and PAHs Chris Wright British American Tobacco Group R&D UK US FDA Center for Tobacco Products Scientific Workshop 30-31 July 2013 1 General Method Testing for TNCO, TSNAs and

More information

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard Certified Reference Material - Certificate of Analysis Methylone, Primary Standard Page 1 of 9 Catalog Number: O Lot: Expiration Date: Description: Packaging: Storage: Shipping: Intended Use: Instructions

More information

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION Andrew J. Aubin and Tanya L. Jenkins Waters Corporation, Milford, MA, USA INTRODUCTION Benzocaine (4-Aminobenzoic

More information

WADA Technical Document TD2018DL DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES

WADA Technical Document TD2018DL DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES Introduction This Technical Document shall be applied to the quantitative determination of a Threshold Substance in a Sample

More information

Measurement Uncertainty Principles and Implementation in QC

Measurement Uncertainty Principles and Implementation in QC Measurement Uncertainty Principles and Implementation in QC Dr. Michael Haustein CURRENTA GmbH & Co. OHG Analytics 41538 Dormagen, Germany www.analytik.currenta.de A company of Bayer and LANXESS About

More information

Analytical Performance & Method. Validation

Analytical Performance & Method. Validation Analytical Performance & Method Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451 Saudi Arabia Building:

More information

Practical and Valid Guidelines for Realistic Estimation of Measurement Uncertainty in Pesticide Multi-Residue Analysis*

Practical and Valid Guidelines for Realistic Estimation of Measurement Uncertainty in Pesticide Multi-Residue Analysis* Practical and Valid Guidelines for Realistic Estimation of Measurement Uncertainty in Pesticide Multi-Residue Analysis* Antonio Valverde Pesticide Residue Research Group University of Almería * This presentation

More information

USE OF TOLERANCES IN THE DETERMINATION OF ACTIVE INGREDIENT CONTENT IN SPECIFICATIONS FOR PLANT PROTECTION PRODUCTS

USE OF TOLERANCES IN THE DETERMINATION OF ACTIVE INGREDIENT CONTENT IN SPECIFICATIONS FOR PLANT PROTECTION PRODUCTS CropLife International USE OF TOLERANCES IN THE DETERMINATION OF ACTIVE INGREDIENT CONTENT IN SPECIFICATIONS FOR PLANT PROTECTION PRODUCTS (Adopted on March 3rd Meeting 2005, ECPA SEG) - 1 - Technical

More information

WADA Technical Document TD2019DL DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES

WADA Technical Document TD2019DL DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES 1. Introduction This Technical Document shall be applied to the quantitative determination of a Threshold Substance in a Sample

More information

Method Validation. Role of Validation. Two levels. Flow of method validation. Method selection

Method Validation. Role of Validation. Two levels. Flow of method validation. Method selection Role of Validation Method Validation An overview Confirms the fitness for purpose of a particular analytical method. ISO definition: Conformation by examination and provision of objective evidence that

More information

International Journal of Pharmacy and Pharmaceutical Sciences ISSN Vol 2, Suppl 3, 2010

International Journal of Pharmacy and Pharmaceutical Sciences ISSN Vol 2, Suppl 3, 2010 International Journal of Pharmacy and Pharmaceutical Sciences ISSN- 0975-1491 Vol 2, Suppl 3, 2010 VALIDATION OF ANALYTICAL METHODS STRATEGIES & IMPORTANCE Review Article RAVICHANDRAN V, 1 SHALINI S, 1

More information

Part IVB Quality Assurance/Validation of Analytical Methods

Part IVB Quality Assurance/Validation of Analytical Methods 1. Introduction Part IVB Quality Assurance/Validation of Analytical Methods 1.1 Definition and purpose of validation Validation is the confirmation by examination and the provision of objective evidence

More information

BASIC CONCEPTS C HAPTER 1

BASIC CONCEPTS C HAPTER 1 C HAPTER 1 BASIC CONCEPTS Statistics is the science which deals with the methods of collecting, classifying, presenting, comparing and interpreting numerical data collected on any sphere of inquiry. Knowledge

More information

Really, A Revised MDL Procedure

Really, A Revised MDL Procedure NEMC 2016 Really, A Revised MDL Procedure Richard Burrows, Ph.D. TestAmerica Inc. 1 A Revision to the Method Detection Limit EPA published a revision to the 40 CFR Part 136 MDL procedure in the Federal

More information

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 9 DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 9 Determination of drug release during

More information

NATIONAL ASSOCIATION OF TESTING AUTHORITIES (NATA) REQUIREMENTS FOR ACCREDITATION OF ICP-MS TECHNIQUES

NATIONAL ASSOCIATION OF TESTING AUTHORITIES (NATA) REQUIREMENTS FOR ACCREDITATION OF ICP-MS TECHNIQUES NATIONAL ASSOCIATION OF TESTING AUTHORITIES (NATA) REQUIREMENTS FOR ACCREDITATION OF ICP-MS TECHNIQUES The National Association of Testing Authorities (NATA) requirements for accreditation have undergone

More information

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014) SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February

More information

MEASUREMENT UNCERTAINTY PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER Prepared by MJ Mc Nerney for ENAO Assessor Calibration

MEASUREMENT UNCERTAINTY PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER Prepared by MJ Mc Nerney for ENAO Assessor Calibration MEASUREMENT PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE UNCERTAINTY OCTOBER/NOVEMBER 2012 Prepared by for ENAO Assessor Calibration B SCOPE Introduction House Rules Overview 17025 & 15189 MU Approaches

More information

VALIDATION OF ANALYTICAL METHODS FOR PHARMACEUTICAL ANALYSIS

VALIDATION OF ANALYTICAL METHODS FOR PHARMACEUTICAL ANALYSIS EXCERPT FROM: VALIDATION OF ANALYTICAL METHODS FOR PHARMACEUTICAL ANALYSIS BY OONA MCPOLIN This excerpt is provided as a preview to enable the reader to sample the content and style of this book. You can

More information

EPAs New MDL Procedure What it Means, Why it Works, and How to Comply

EPAs New MDL Procedure What it Means, Why it Works, and How to Comply EPAs New MDL Procedure What it Means, Why it Works, and How to Comply Richard Burrows TestAmerica Inc. 1 A Revision to the Method Detection Limit EPA published a revision to the 40 CFR Part 136 MDL procedure

More information

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012 Revision 1.2 October 15, 2012 PAGE II TABLE OF CONTENTS TABLE OF CONTENTS... 1 GUIDANCE...2 1.0 Introduction... 2 2.0 Uncertainty and Traceability... 2 3.0 Need for Traceability... 3 3.1 What does traceability

More information

PQRI Technical Analytical Challenges Round 2 Interlaboratory Study: Progress and Early Findings

PQRI Technical Analytical Challenges Round 2 Interlaboratory Study: Progress and Early Findings PQRI Technical Analytical Challenges Round 2 Interlaboratory Study: Progress and Early Findings November 3, 2017 Donna Seibert Perrigo James Harrington RTI International Disclaimer Any images of or references

More information

2.1 2,3 Dichloro Benzoyl Cyanide (2,3 DCBC) and survey of. manufactured commonly for the bulk drug industry, few references were

2.1 2,3 Dichloro Benzoyl Cyanide (2,3 DCBC) and survey of. manufactured commonly for the bulk drug industry, few references were . Introduction.,3 Dichloro Benzoyl Cyanide (,3 DCBC) and survey of analytical methods,3-dcbc substance, is the although advanced intermediate,3-dcbc is a of lamotrigine well-known bulk drug chemical manufactured

More information

Method Development and Validation Of Prasugrel Tablets By RP- HPLC

Method Development and Validation Of Prasugrel Tablets By RP- HPLC Method Development and Validation Of Prasugrel Tablets By RP- HPLC K.Sonia*, Ndwabe Hamunyare, K.Manikandan Department of Pharmaceutical Analysis, SRM College of Pharmacy, SRM University, Kattankulathur,

More information

3.1.1 The method can detect, identify, and potentially measure the amount of (quantify) an analyte(s):

3.1.1 The method can detect, identify, and potentially measure the amount of (quantify) an analyte(s): SOP No: ICLN-Validation Levels Between Networks.003.01 Page 1 of 19 1. Purpose: This document establishes guidelines for assigning validation levels to methods based on national and international standards.

More information

Stability-indicating HPLC determination of tolterodine tartrate in pharmaceutical dosage form

Stability-indicating HPLC determination of tolterodine tartrate in pharmaceutical dosage form Indian Journal of Chemical Technology Vol. 13, May 2006, pp. 242-246 Stability-indicating HPLC determination of tolterodine tartrate in pharmaceutical dosage form Vinay Saxena a *, Zahid Zaheer b & Mazhar

More information

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India QbD Based Method Development on an Agilent 129 Infinity UHPLC System Combined with a Seamless Method Transfer to HPLC Using Intelligent System Emulation Technology Application Note Pharmaceutical QA/QC

More information

Verification of Pharmaceutical Raw Materials Using the Spectrum Two N FT-NIR Spectrometer

Verification of Pharmaceutical Raw Materials Using the Spectrum Two N FT-NIR Spectrometer APPLICATION NOTE FT-NIR Spectroscopy Authors: Ian Robertson Jerry Sellors PerkinElmer, Inc. Seer Green, UK Verification of Pharmaceutical Raw Materials Using the Spectrum Two N FT-NIR Spectrometer Introduction

More information

EFFECT OF THE UNCERTAINTY OF THE STABILITY DATA ON THE SHELF LIFE ESTIMATION OF PHARMACEUTICAL PRODUCTS

EFFECT OF THE UNCERTAINTY OF THE STABILITY DATA ON THE SHELF LIFE ESTIMATION OF PHARMACEUTICAL PRODUCTS PERIODICA POLYTECHNICA SER. CHEM. ENG. VOL. 48, NO. 1, PP. 41 52 (2004) EFFECT OF THE UNCERTAINTY OF THE STABILITY DATA ON THE SHELF LIFE ESTIMATION OF PHARMACEUTICAL PRODUCTS Kinga KOMKA and Sándor KEMÉNY

More information

TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD

TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD APPENDIX V TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD Validating Chromatographic Methods. By David M. Bliesner Copyright 2006 John Wiley & Sons, Inc. 159 160 APPENDIX V Title: Effective: Document No:

More information

Starting Materials For Active Substances

Starting Materials For Active Substances Starting Materials For Active Substances Workshop 1 How to build a good CEP Application Prague September 19th, 2017 Kristofer Olofsson, Medical Products Agency (MPA) Sweden Redefinition of GMP-Starting

More information

Unified Approach For Performing An Analytical Methods Comparability Study IVT s LAB WEEK Presenter: Peter M. Saama, Ph.D. Bayer HealthCare LLC

Unified Approach For Performing An Analytical Methods Comparability Study IVT s LAB WEEK Presenter: Peter M. Saama, Ph.D. Bayer HealthCare LLC Unified Approach For Performing An Analytical Methods Comparability Study IVT s LAB WEEK 2015 Presenter: Peter M. Saama, Ph.D. Bayer HealthCare LLC Agenda/ Content Page 1 Analytical Methods for Performing

More information

Why is knowledge on measurement uncertainty so important in setting policies on energy efficiency? Rainer Stamminger & Christoforos Spiliotopoulos

Why is knowledge on measurement uncertainty so important in setting policies on energy efficiency? Rainer Stamminger & Christoforos Spiliotopoulos Why is knowledge on measurement uncertainty so important in setting policies on energy efficiency? Rainer Stamminger & Christoforos Spiliotopoulos Agenda What is the problem? Differentiating targets Sources

More information

Swiss Medic Training Sampling

Swiss Medic Training Sampling Swiss Medic Training Sampling Paul Sexton Sampling Preparation for Sampling Representative Sample Re-sampling Sampling Part I What to sample? Why sample? Where to sample? Who performs sampling? How to

More information

Measurement Uncertainty, March 2009, F. Cordeiro 1

Measurement Uncertainty, March 2009, F. Cordeiro 1 Measurement Uncertainty, March 2009, F. Cordeiro 1 Uncertainty Estimation on the Quantification of Major Proteins in Milk by Liquid Chromatography Contribution to a Chemical Reference Measurement System

More information

The Uncertainty of Reference Standards A Guide to Understanding Factors Impacting Uncertainty, Uncertainty Calculations, and Vendor Certifications

The Uncertainty of Reference Standards A Guide to Understanding Factors Impacting Uncertainty, Uncertainty Calculations, and Vendor Certifications The Uncertainty of Reference Standards A Guide to Understanding Factors Impacting Uncertainty, Uncertainty Calculations, and Vendor Certifications Kevin Gates, Ning Chang, Isil Dilek, Huahua Jian, Sherri

More information

Volume 6, Issue 2, January February 2011; Article-015

Volume 6, Issue 2, January February 2011; Article-015 Research Article DEVELOPMENT AND VALIDATION OF A RP-HPLC METHOD FOR THE DETERMINATION OF DAPOXETINE HYDROCHLORIDE IN PHARMACEUTICAL FORMULATION USING AN EXPERIMENTAL DESIGN Pratik Mehta*, Ujjwal Sahoo,

More information

Use of DoE to increase process understanding of a de-bromination reaction

Use of DoE to increase process understanding of a de-bromination reaction Use of DoE to increase process understanding of a de-bromination reaction Francesco Tinazzi Senior Scientist API Development and Manufacturing Aptuit Verona Evotec AG, API update: 7 June 2018 Development

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Amiodarone-D 4, Primary Standard A-83 Page 1 of 8 Catalog Number: A-83 Lot: Expiration: Description: Packaging: Storage: Shipping: Intended Use: Instructions

More information

Outline of Presentation

Outline of Presentation General Considerations for Sampling in a PAT Environment Robert P. Cogdill 28 September 2006 Heidelberg PAT Conference Outline of Presentation Introductory thoughts on sampling Sampling Level I: Interaction

More information

PQRI Stability Shelf-Life Working Group Glossary Ver 1 0.doc

PQRI Stability Shelf-Life Working Group Glossary Ver 1 0.doc Term Accelerated testing Acceptance criteria: Batch Definition Studies designed to increase the rate of chemical degradation or physical change of a drug substance or drug product by using exaggerated

More information

Independence and Dependence in Calibration: A Discussion FDA and EMA Guidelines

Independence and Dependence in Calibration: A Discussion FDA and EMA Guidelines ENGINEERING RESEARCH CENTER FOR STRUCTURED ORGANIC PARTICULATE SYSTEMS Independence and Dependence in Calibration: A Discussion FDA and EMA Guidelines Rodolfo J. Romañach, Ph.D. ERC-SOPS Puerto Rico Site

More information

VOTING DRAFT STANDARD

VOTING DRAFT STANDARD page 1 of 7 VOTING DRAFT STANDARD VOLUME 1 MODULE 4 QUALITY SYSTEMS FOR CHEMICAL TESTING SECTIONS 1.5.1 AND 1.5.2 Description This Voting Draft Standard is a proposed revision of the 2009 standard (EL-

More information

Measurement Uncertainty: A practical guide to understanding what your results really mean.

Measurement Uncertainty: A practical guide to understanding what your results really mean. Measurement Uncertainty: A practical guide to understanding what your results really mean. Overview General Factors Influencing Data Variability Measurement Uncertainty as an Indicator of Data Variability

More information

An Automated Application Template for Dissolution Studies

An Automated Application Template for Dissolution Studies An Automated Application Template for Dissolution Studies in SDMS Vision Publisher Wolfgang Lemmerz, Chris L. Stumpf, Thomas Schmidt, and Phil Kilby Waters Corporation, Milford, MA U.S.A. OVERVIEW Dissolution

More information

OF ANALYSIS FOR DETERMINATION OF PESTICIDES RESIDUES IN FOOD (CX/PR 15/47/10) European Union Competence European Union Vote

OF ANALYSIS FOR DETERMINATION OF PESTICIDES RESIDUES IN FOOD (CX/PR 15/47/10) European Union Competence European Union Vote 1 April 2015 European Union s CODEX COMMITTEE ON PESTICIDE RESIDUES 47 th Session Beijing, China, 13 18 April 2015 AGENDA ITEM 8 PROPOSED DRAFT GUIDELINES ON PERFORMANCE CRITERIA SPECIFIC FOR METHODS OF

More information

International interlaboratory study of forensic ethanol standards

International interlaboratory study of forensic ethanol standards International interlaboratory study of forensic ethanol standards Bernard King and Richard Lawn LGC, Queens Road, Teddington, Middlesex, UK TW11 0LY. E-mail: bmjk@lgc.co.uk Received 18th January 1999,

More information

ROLE AND SIGNIFICANCE OF UNCERTAINTY IN HV MEASUREMENT OF PORCELAIN INSULATORS A CASE STUDY

ROLE AND SIGNIFICANCE OF UNCERTAINTY IN HV MEASUREMENT OF PORCELAIN INSULATORS A CASE STUDY International Conference on Ceramics, Bikaner, India International Journal of Modern Physics: Conference Series Vol. (01) 8 5 World Scientific Publishing Company DOI: 10.11/S010195101019 ROLE AND SIGNIFICANCE

More information

The Role of and the Place of Method Validation in Drug Analysis Using Electroanalytical Techniques

The Role of and the Place of Method Validation in Drug Analysis Using Electroanalytical Techniques The Open Analytical Chemistry Journal, 2011, 5, 1-21 1 Open Access The Role of and the Place of Method Validation in Drug Analysis Using Electroanalytical Techniques Mehmet Gumustas 1,2 and Sibel A. Ozkan

More information

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method: ALLOWAY METHOD OUTLINE Standard Laboratory Method: SM Parameter: Methylene Blue Method: Colorimetric Reporting Level: Reference: 0.05 mg/l Standard Methods for the Examination Of Water and Wastewater;

More information

GUIDELINES ON PERFORMANCE CRITERIA FOR METHODS OF ANALYSIS FOR THE DETERMINATION OF PESTICIDE RESIDUES IN FOOD AND FEED CXG Adopted in 2017.

GUIDELINES ON PERFORMANCE CRITERIA FOR METHODS OF ANALYSIS FOR THE DETERMINATION OF PESTICIDE RESIDUES IN FOOD AND FEED CXG Adopted in 2017. GUIDELINES ON PERFORMANCE CRITERIA FOR METHODS OF ANALYSIS FOR THE DETERMINATION OF PESTICIDE RESIDUES IN FOOD AND FEED CXG 90-2017 Adopted in 2017. CAC/GL 90-2017 2 OBJECTIVE 1. The purpose of these guidelines

More information

Document No: TR 12 Issue No: 1

Document No: TR 12 Issue No: 1 ESTIMATION OF THE UNCERTAINTY OF MEASUREMENT BY CALIBRATION LABORATORIES AND SPECIFICATION OF CALIBRATION AND MEASUREMENT CAPABILITY ON SCHEDULES OF ACCREDITATION Prepared by: SADCAS Technical Manage Approved

More information

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance Michael D. Jones, Marian Twohig, Karen Haas, and Robert S. Plumb Waters Corporation,

More information

1.11 Measurement Uncertainty

1.11 Measurement Uncertainty Document #: 4968 Page 1 of 10 1.11 Measurement Uncertainty In an effort to comply with accreditation requirements, and because scientific measurements in general are subject to variability, a budget estimating

More information

Implementing principles of Quality by Design (QbD) in validation context

Implementing principles of Quality by Design (QbD) in validation context Implementing principles of Quality by Design (QbD) in validation context Cédric Hubert a, Pierre Lebrun a,b, Eric Rozet a,b and Philippe Hubert a a Laboratory of Analytical Chemistry, CIRM, Department

More information

Metolachlor ESA Sodium Salt Standard

Metolachlor ESA Sodium Salt Standard Analytical Reference Materials Metolachlor ESA Sodium Salt Standard Catalog # 33200 Lot # A0102104 110 Benner Circle Bellefonte, PA 16823-8812 1-814-353-1300 1-800-356-1168 www.restek.com FOR LABORATORY

More information

Method Verification. The How M.L. Jane Weitzel ALACC Chair

Method Verification. The How M.L. Jane Weitzel ALACC Chair Method Verification The How M.L. Jane Weitzel ALACC Chair The What Released early 2008 http://www.aoac.org/a lacc_guide_2008.pdf Meet ISO 17025 Requirement Categories of Methods The six categories of chemical

More information

TNI Standard; EL-V1M4 Sections and (Detection and Quantitation) page1 of 8. TNI Standard VOLUME 1 MODULE 4

TNI Standard; EL-V1M4 Sections and (Detection and Quantitation) page1 of 8. TNI Standard VOLUME 1 MODULE 4 page1 of 8 TNI Standard VOLUME 1 MODULE 4 QUALITY SYSTEMS FOR CHEMICAL TESTING SECTIONS 1.5.1 AND 1.5.2 January 2016 Description This TNI Standard has been taken through all of the voting stages and has

More information

Using validation data for ISO measurement uncertainty estimation Part 1. Principles of an approach using cause and effect analysis

Using validation data for ISO measurement uncertainty estimation Part 1. Principles of an approach using cause and effect analysis Analyst, June 1998, Vol. 123 (1387 1392) 1387 Using validation data for ISO measurement uncertainty estimation Part 1. Principles of an approach using cause and effect analysis Stephen L. R. Ellison *,

More information

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity Janet Hammond Waters Corporation, Wilmslow, UK APPLICATION BENEFITS High

More information

Validated HPLC Methods

Validated HPLC Methods Validated HPLC Methods To Tweak or Not to Tweak Slide 4 Adjustments of operating conditions to meet system suitability requirements may be necessary. USP 23 p1776 Slide 5 001816S1.PPT What is the line

More information

ENVIRONMENTAL LABORATORY SECTOR VOLUME 1 MANAGEMENT AND TECHNICAL REQUIREMENTS FOR LABORATORIES PERFORMING ENVIRONMENTAL ANALYSIS

ENVIRONMENTAL LABORATORY SECTOR VOLUME 1 MANAGEMENT AND TECHNICAL REQUIREMENTS FOR LABORATORIES PERFORMING ENVIRONMENTAL ANALYSIS EL-VM-0 ENVIRONMENTAL LABORATORY SECTOR VOLUME MANAGEMENT AND TECHNICAL REQUIREMENTS FOR LABORATORIES PERFORMING ENVIRONMENTAL ANALYSIS Module : Quality Systems for Radiochemical Testing Working Draft

More information

How s that *new* LOD Coming? Rick Mealy WWOA Board of Directors DNR LabCert

How s that *new* LOD Coming? Rick Mealy WWOA Board of Directors DNR LabCert How s that *new* LOD Coming? Rick Mealy WWOA Board of Directors DNR LabCert The LOD Procedure has changed Federal Register /Vol. 82, No. 165 /Monday, August 28, 2017 ACTION: Final rule. DATES: This regulation

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Dehydroepiandrosterone 3-sulfate, Primary Standard D-065 Page 1 of 6 Catalog Number: D-065 Lot: H 3 C Expiration: Description: February 2019 Dehydroepiandrosterone

More information

STUDY OF THE APPLICABILTY OF CONTENT UNIFORMITY AND DISSOLUTION VARIATION TEST ON ROPINIROLE HYDROCHLORIDE TABLETS

STUDY OF THE APPLICABILTY OF CONTENT UNIFORMITY AND DISSOLUTION VARIATION TEST ON ROPINIROLE HYDROCHLORIDE TABLETS & STUDY OF THE APPLICABILTY OF CONTENT UNIFORMITY AND DISSOLUTION VARIATION TEST ON ROPINIROLE HYDROCHLORIDE TABLETS Edina Vranić¹*, Alija Uzunović² ¹ Department of Pharmaceutical Technology, Faculty of

More information

P103d Annex: Policy on Estimating Measurement Uncertainty for Construction Materials & Geotechnical Testing Labs Date of Issue 09/13/05

P103d Annex: Policy on Estimating Measurement Uncertainty for Construction Materials & Geotechnical Testing Labs Date of Issue 09/13/05 P103d Annex: Policy on Estimating Measurement Uncertainty for Construction Materials & Geotechnical Testing Labs Date of Issue 09/13/05 This document provides guidance on the estimation of uncertainty

More information

Validation of an Analytical Method

Validation of an Analytical Method Validation of an Analytical Method Refer to: ICH Guideline Q2(R1), Validation of Analytical Procedures: Teaxt andmethodology. Introduction All major laboratories eg. in the industry operates with clearly

More information

Test Method Development and Validation as Pertaining to Microtrac Particle Size Measuring Instruments

Test Method Development and Validation as Pertaining to Microtrac Particle Size Measuring Instruments Test Method Development and Validation as Pertaining to Microtrac Particle Size Measuring Instruments Philip E. Plantz, PhD Application Note SL AN 17 Rev B Provided By: Microtrac, Inc. Particle Size Measuring

More information

ISO/TS TECHNICAL SPECIFICATION. Water quality Guidance on analytical quality control for chemical and physicochemical water analysis

ISO/TS TECHNICAL SPECIFICATION. Water quality Guidance on analytical quality control for chemical and physicochemical water analysis TECHNICAL SPECIFICATION ISO/TS 13530 First edition 2009-03-15 Water quality Guidance on analytical quality control for chemical and physicochemical water analysis Qualité de l'eau Lignes directrices pour

More information

Computer Science, Informatik 4 Communication and Distributed Systems. Simulation. Discrete-Event System Simulation. Dr.

Computer Science, Informatik 4 Communication and Distributed Systems. Simulation. Discrete-Event System Simulation. Dr. Simulation Discrete-Event System Simulation Chapter 9 Verification and Validation of Simulation Models Purpose & Overview The goal of the validation process is: To produce a model that represents true

More information