Fast Atom Bombardment (FAB)-MS

Size: px
Start display at page:

Download "Fast Atom Bombardment (FAB)-MS"

Transcription

1 Fast Atom Bombardment (FAB)-MS FAB-MS was developed by Barber and coworkers in the early 1980s and soon became a very powerful technique for desorption ionization of thermally labile molecules. In FAB a beam of fast moving neutral atoms of a noble gas (Ar, Xe), is directed to strike a sample film deposited on a clean metal support: Secondary ions are emitted from the sample layer and then extracted and focused towards the mass spectrometer:

2 Source of fast atoms in FAB The Ar or Xe beam required in FAB is generated within an appropriate atom gun: Ar (or Xe) atoms are ionised by electron ionization within chamber 1 the resulting ions are then focused and accelerated by lenses when arriving in chamber 3, Ar ions exchange charge and energy with Ar neutral atoms 3 Ar ions are separated from accelerated neutral atoms, outgoing from chamber 3, by electrostatic deflectors (4): fast Ar atoms are finally directed outside, towards the sample stage.

3 A Torr Ar (or Xe) pressure is maintained inside chamber 3. Under these conditions a charge exchange without significant loss of momentum occurs in chamber 3 between incoming Ar + ions and Ar atoms. The processes occuring in the different stages of the atom gun are: 1) Ar Ar + + e - 2) Ar + Ar + (fast) 3) Ar + (fast) + Ar Ar (fast) + Ar + Kinetic energies between 3 and 10 kev can be achieved for outcoming atoms. The atom beam is usually directed at a 20 angle with respect to the sample surface:

4 When the sample surface is struck by fast atoms an intense thermal spike is produced and subsequently dissipated towards the surface outer layers through momentum transfers: FAB mechanism Analyte molecules are detached from these layers as a dense gas containing positive and negative ions, as well as neutrals, that can be eventually ionized in the plasma just above the sample surface. Actually, when a dry deposited sample is analyzed by FAB the secondary ion yield decays rapidly, due to surface damage. Much more satisfactory results can be achieved if the sample is dissolved in a liquid matrix.

5 The main requirements for a FAB matrix are: being reasonably involatile (since it must remain liquid under vacuum) having good solvating properties for the analyte Polar compounds with MW lower than 300 Da are usually adopted as matrices for FAB: Occasionally, inorganic or organic acids/bases are added to the matrix to promote protonation/deprotonation.

6 According to the more recent hypotheses the chemistry leading to ion generation during FAB takes place into special gas cavities, representing the initial interfacial region between the condensed phase (i.e. the matrix/analyte mixture) and the gas phase: The cavities are formed in the interfacial region upon sudden heating after fast atom impacts, then desorption occurs and analyte and matrix molecules move towards the so called selvedge region, where ionmolecule reactions can occur. Inorganic salts (even as impurities) can be involved in the process.

7 In the ChemicaI Ionization-like model, the following ion-molecule interactions are hypothesised to explain ion generation during FAB: Proton transfer by any organic ion (FH + ) arising from the matrix Cation/anion attachment from salts dissolved into the sample If the analyte is anionic, partial charge recombination with counterions in the sample or proton abstraction from the matrix (e.g. glycerol, G) may occur:

8 An approach proposed (1997) for the determination of reagent ions, FH +, i.e. ionic species actually involved in ionization of analyte molecules, is based on the acquisition of FAB spectra for pure glycerol samples at increasing times: Peaks related to glycerol clusters are observed, along with the GH + ion (m/z 93): G 2 H +, 185 G 3 H +, 277 G 4 H +, 369 These ions disappear progressively from the FAB spectra as time goes by, due to lower availability of glycerol molecules, required for cluster generation.

9 Relevant ions for the FAB mechanism are likely those related to m/z ratios 31, 45, 57 and 75, because they are predominant when long bombardment times, i.e. thin residual sample thicknesses, are considered. A proposed pattern for their generation is: These ions might be the actual primary ions of FAB, dominant at the interfacial region but attenuated as they pass through the selvedge region, where the following processes may occur:

10 When negative ion FAB is performed deprotonated analyte and matrix molecules are usually observed. If glycerol is used as a matrix a proton transfer can occur: Negative ion FAB spectra obtained at different times for a glycerol sample show further relevant ions, along with [G-H] - (m/z 91):

11 Glycerol clusters with the [G-H] - analysis: ion are observed in the first stage of [2G-H] -, m/z 183 [3G-H] -, m/z 275 [4G-H] -, m/z 367 Further ions, observed at longer bombardment times (m/z 71, 59, 45, 43, 25 and 17) could be the actual reagent ions and have been interpreted with the help of deuterated analogs:

12 Data available on proton affinities (PA) for the cited ions (or structural analogs) indicate that only C 2 H - and OH - can actually react exothermically with glycerol to generate the [G-H] - ion. Other species, like ions with m/z 71, 59 and 43, have slightly higher proton affinities than [G-H] - ion, thus they can deprotonate glycerol only if energy is provided by the environment, which is reasonable, during fast atom bombardment.

13 Temporal dependence of analyte signals in FAB-MS The evolution with time of matrix-related ions can influence the temporal dependence of the FAB spectrum for an analyte, as shown in the following for the cardiac glycoside digoxine (MW 780 u):

14 In this case, the quasi-molecular ion of digoxine (m/z 781) and its highmw product ions (m/z 651, 521, 391 and its de-hydrated form, 373) are almost negligible, at any acquisition time. Smaller product ions, like those with m/z 243 and 131, are very relevant already in the first spectra and then undergo a significant decrease, indicating analyte depletion from the matrix. As in experiments involving only glycerol, the [G+H] + (m/z 93) and its cluster ions (m/z 277 and 185) are initially relevant, then fragment ions of glycerol (m/z 43, 45, 57 and 75), generally indicated as F +, become prevailing in the FAB spectrum. The following processes can be generally hypothesized for glycerol/analyte interactions: Considerations related to proton affinities values lead to predict that GH + ions generate more analyte fragments than F + ones.

15 Fast atom gun preparation Practical operation of FAB-MS Before operation, the gas lines (usually Ar or Xe) connected to the fast atom gun need to be flushed out once or twice to remove any air trace. After this operation the inert gas cylinder feeding the gun is usually connected permanently to the gun. Typical gas flows during operation are 0.5 ml/min, whereas accelerating potentials of a few kv are applied. Sample stage (probe tip) preparation The stage on which the matrix/sample mixture is deposited can be made of stainless steel, copper or gold-plated copper. Preliminary cleaning is usually performed with an organic solvent but concentrated HCl can be used to remove persistent remnants of previous samples.

16 Deposition of matrix and analyte solution on the sample stage The matrix is usually deposited onto the sample stage before the sample solution, using a few L (e.g. of glycerol), that need to be properly spread out on the entire area of the stage (to avoid the presence of stage contaminant signals in the FAB spectra). Afterwards, a 1 L aliquot of the analyte solution is added, using a microsyringe, and then mixed with the matrix. Sample stage introduction The sample stage is introduced within the high vacuum region of the spectrometer using a sample probe, passing through a vacuum lock. Solvents adopted for sample dissolution (water or methanol) are quicky evacuated, thus the source pressure will stabilize at an acceptable value after a relatively short time. Afterwards, fast atom gun and extraction voltages can be switched on and the analysis can be started.

17 Intepretation of FAB-MS spectra As already evidenced, the presence of matrix peaks can complicate the interpretation of FAB-MS spectra. A clear example of matrix interference is represented by the FAB positive ion spectrum (with glycerol as matrix) for the pentapeptide methionine enkephalin (MW u): G 3 H + G 4 H + [M+H] + G 5 H + G 6 H + G7H +

18 In case one of the matrix peak has a m/z ratio close to that of the analyte, the matrix has to be changed. When traces of salts, especially Na and K salts, are present in the sample, series of Na or K-adducts with matrix clusters can be observed in the FAB spectrum, thus increasing the risk of spectral interference. In this case the spectrum may exhibit peculiar doublets with a m/z difference of 22 or 38 m/z units, corresponding to substitution of a H + with Na + or K +, respectively. As in FD, deliberate cationization can enhance FAB signals due to analyte adducts with Na + or K +, which can be useful also to recognize if adducts are actually present. When only one dominant peak is present in the FAB spectrum, apart from matrix well known peaks, the dilemma about its nature, e.g. if it corresponds to a [M+H] + or a [M+Na] + ion, can be solved by doping the sample with a K salt, thus introducing a [M+K] + peak. The dominant peak will correspond to [M+H] + or [M+Na] + if the shift observed for the new peak is 38 or 16 m/z units, respectively.

19 Limitations of FAB-MS Besides the risk of spectral interference due to matrix cluster ions, analyte and matrix adducts with alkali cation traces and even matrix-analyte cluster ions, other important limitations need to be considered for FAB-MS: high background below 200 m/z units; upper MW limit at u; difficulty in achieving a precise and accurate quantification: in many cases the analyte signal disappears because the matrix has been desorbed completely so the operator cannot be sure that the analyte has been consumed completely; high competition for ionization in mixtures, with a typical example given by a protein digest: hydrophobic peptides tend to segregate at the upper surface of the matrix/analyte coating, thus can be ionized more efficiently than hydrophilic peptides. In some cases, a single component can totally suppress the signal due to other analytes in the mixture.

20 Fast Ion Bombardment (FIB)-MS Fast ion bombardment was introduced soon after FAB-MS when it was clear that the latter could not ionize (bio)molecules with MW higher than u. The principle of FIB is using a Cs + beam instead of a Xe or Ar one, to hit the sample: Cs + ions are generated by a pellet heated by a filament and then accelerated towards the probe tip by very high potentials, up to 35 kv.

21 In its typical configuration FIB-MS is very similar to Secondary Ion Mass Spectrometry (SIMS), although applied to a liquid, instead of a solid, sample; the alternative name Liquid SIMS (LSIMS) has been then given to the technique. Although FAB and FIB mass spectra are almost identical when relative low (< 1000 u) MW analytes are considered, the higher energy of Cs + ions, compared to that of Xe or Ar fast atoms, ensures much higher sensitivities with FIB-MS. In the case of high MW compounds the difference is more striking. Good S/N mass spectra have been obtained by FIB-MS even for molecules like trypsinogen (MW u).

22 Continuous Flow Fast Atom Bombardment Continuous Flow Fast Atom Bombardment was developed in the middle 1980s to introduce samples in a flow of solvent directed towards a special probe for FAB (or FIB)-MS analysis. Two general configurations are used for CF-FAB, both adopting a special hollow direct insertion probe: Frit FAB (Ito et alt., 1985) In this case the analyte/matrixcarrying solvent flows in a fused silica tubing (~40 m internal diameter) contained in a stainless steel capillary passing through the probe. At the probe tip the silica tubing delivers the liquid through a stainless steel frit, thus a thin analyte-matrix layer is formed on the frit.

23 Continuous flow (dynamic)-fab (Caprioli et alt., 1986) The main difference with respect to Frit-FAB is that the silica tubing delivers the analyte/matrix solution directly on the probe tip, through a small hole. Copper tips were adopted only initially and then replaced by goldcoated copper tips. In both cases maximum flows of a few L/min can ensure the proper balance between delivery and vacuum/thermal desorption of solvent/analyte/matrix at the probe tip.

24 In a further design reported for CF-FAB an absorbant pad is placed just below the end of the liquid-delivering fused silica capillary: The pad absorbs excess liquid at the surface of the probe tip, thus making the sample/analyte/matrix coating thinner.

25 Operating modes for CF-FAB Three operating modes are usually adopted for CF-FAB: a) continuous sampling of the compound or reaction mixture of interest (the liquid is drawn simply by the vacuum at the opposite end of the silica capillary) b) consecutive, discrete injections of different samples through a valve (like in Flow Injection Analysis, FIA); c) coupling to a separation technique, like HPLC or Capillary Electrophoresis (CE)

26 Practical operation of CF-FAB One of the most critical points in CF-FAB is obtaining a stable flow at the probe tip, in order to obtain a steady ion beam from the source. The main parameters to check are: flow rate (usually lower than 10 L/min) flow fluctuations (a pulse-free solvent delivery pump should be used) composition of solvent-matrix mixture (usually a water/methanol or acetonitrile/glycerol 0.9 : 0.9 : 0.2 v/v/v mixture is adopted) silica capillary length (usually 1 m) and protrusion from the probe tip (1-2 mm)

27 probe tip temperature: differently from conventional FAB-MS, gentle heat is required in CF-FAB to remove excess solvent. The optimal combination between temperature and flow rate can be determined by reporting the signal stability as a function of the two parameters: A B Even small changes of one of the two parameters, if leading to cross the border of the stability region, can lead to evident modifications of the CF-FAB signal stability: A B

28 Reaction monitoring by continuous sampling-fab Continuous sampling enables the FAB-MS monitoring of reaction mixtures in real time. A typical application is following the evolution of enzymatic digests of peptides or proteins, like Substance P (a neuropeptide) digestion by carboxypeptidase Y, an enzyme that hydrolyses the peptide bond at the COOH terminus: 1217 Arg Pro Lys Pro Gln Gln Phe Phe Gly Leu Met (MW u)

29 Arg Pro Lys Pro Gln Gln Phe Phe Gly Leu Met (MW u) The sequential appearance of smaller peptides arising from carbopeptidase Y-catalyzed hydrolysis provides important information on the digestion kinetics.

30 HPLC-FAB-MS Continuous flow-fab was adopted as one of the first LC-MS interfaces before ESI became the prevailing approach. The typical FAB matrix, glycerol, can be either added already to the HPLC mobile phase (it has almost the same eluotropic strength as methanol), although at percent values not higher than 5 % (v/v), or introduced as post-column reagent, using suitable T-junctions: Further T-shaped splitters need to be usually introduced before and/or after the HPLC colum to adapt high LC flow rates (hundreds of L/min) to those suitable for CF-FAB.

31 Microbore (i.d. 1 mm) or capillary (i.d. 0.3 mm) HPLC columns can be adapted more easily to CF-FAB-MS, as their typical working flows are and 5 L/min, respectively: The post-column introduction of FAB matrix by a syringe pump is evidenced in the figure. The need for matrix can lead to an increase of flow rates, thus a subsequent Split Tee is required.

32 Applications of HPLC-FAB-MS One of the most interesting applications reported for HPLC-FAB-MS is the separation and MS analysis of peptides arising from protein enzymatic digestion. The TIC trace and a composite trace, obtained by overlap of extracted Ion Chromatograms (XIC) for selected peptides, relevant to the digest of horse heart cytochrome C with trypsine, are shown as an example: TIC XIC

Chemistry Instrumental Analysis Lecture 34. Chem 4631

Chemistry Instrumental Analysis Lecture 34. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 34 From molecular to elemental analysis there are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry

More information

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Chemistry Instrumental Analysis Lecture 37. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 37 Most analytes separated by HPLC are thermally stable and non-volatile (liquids) (unlike in GC) so not ionized easily by EI or CI techniques. MS must be at

More information

M M e M M H M M H. Ion Sources

M M e M M H M M H. Ion Sources Ion Sources Overview of Various Ion Sources After introducing samples into a mass spectrometer, the next important step is the conversion of neutral molecules or compounds to gas phase ions. The ions could

More information

Lecture 15: Introduction to mass spectrometry-i

Lecture 15: Introduction to mass spectrometry-i Lecture 15: Introduction to mass spectrometry-i Mass spectrometry (MS) is an analytical technique that measures the mass/charge ratio of charged particles in vacuum. Mass spectrometry can determine masse/charge

More information

Other Methods for Generating Ions 1. MALDI matrix assisted laser desorption ionization MS 2. Spray ionization techniques 3. Fast atom bombardment 4.

Other Methods for Generating Ions 1. MALDI matrix assisted laser desorption ionization MS 2. Spray ionization techniques 3. Fast atom bombardment 4. Other Methods for Generating Ions 1. MALDI matrix assisted laser desorption ionization MS 2. Spray ionization techniques 3. Fast atom bombardment 4. Field Desorption 5. MS MS techniques Matrix assisted

More information

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n 20.2 Ion Sources electron ionization produces an M + ion and extensive fragmentation chemical ionization produces an M +, MH +, M +, or M - ion with minimal fragmentation MALDI uses laser ablation to transfer

More information

Ionization Techniques Part IV

Ionization Techniques Part IV Ionization Techniques Part IV CU- Boulder CHEM 5181 Mass Spectrometry & Chromatography Presented by Prof. Jose L. Jimenez High Vacuum MS Interpretation Lectures Sample Inlet Ion Source Mass Analyzer Detector

More information

Mass Spectrometry. General Principles

Mass Spectrometry. General Principles General Principles Mass Spectrometer: Converts molecules to ions Separates ions (usually positively charged) on the basis of their mass/charge (m/z) ratio Quantifies how many units of each ion are formed

More information

MASS SPECTROMETRY. Topics

MASS SPECTROMETRY. Topics MASS SPECTROMETRY MALDI-TOF AND ESI-MS Topics Principle of Mass Spectrometry MALDI-TOF Determination of Mw of Proteins Structural Information by MS: Primary Sequence of a Protein 1 A. Principles Ionization:

More information

CHAPTER A2 LASER DESORPTION IONIZATION AND MALDI

CHAPTER A2 LASER DESORPTION IONIZATION AND MALDI Back to Basics Section A: Ionization Processes CHAPTER A2 LASER DESORPTION IONIZATION AND MALDI TABLE OF CONTENTS Quick Guide...27 Summary...29 The Ionization Process...31 Other Considerations on Laser

More information

Mass Spectrometry in MCAL

Mass Spectrometry in MCAL Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS GC seperates small, volatile, non-polar material MS is detection devise (Agilent 320-MS TQ Mass Spectrometer) Full scan monitoring SIM single ion monitoring

More information

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass

More information

Mass Spectrometry (MS)

Mass Spectrometry (MS) Mass Spectrometry (MS) Alternative names: Mass spectrometric (selective) detector (MSD) Spectrometry - methods based on interaction of matter and radiation Mass spectrometry - method based on formation

More information

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai 303351 Instrumental Analysis Mass Spectrometry Lecturer:! Somsak Sirichai Mass Spectrometry What is Mass spectrometry (MS)? An analytic method that employs ionization and mass analysis of compounds in

More information

TANDEM MASS SPECTROSCOPY

TANDEM MASS SPECTROSCOPY TANDEM MASS SPECTROSCOPY 1 MASS SPECTROMETER TYPES OF MASS SPECTROMETER PRINCIPLE TANDEM MASS SPECTROMETER INSTRUMENTATION QUADRAPOLE MASS ANALYZER TRIPLE QUADRAPOLE MASS ANALYZER TIME OF FLIGHT MASS ANALYSER

More information

Molecular Mass Spectrometry

Molecular Mass Spectrometry Molecular Mass Spectrometry Mass Spectrometry: capable of providing information about (1) Elemental composition of samples of matter: atomic mass (2) Structures of inorganic, organic, and biological molecules

More information

Selecting an LC/MS Interface Becky Wittrig, Ph.D.

Selecting an LC/MS Interface Becky Wittrig, Ph.D. Selecting an LC/MS Interface Becky Wittrig, Ph.D. RESTEK CORPORATION LC/MS Interfaces I. Background of LC/MS I. Historical Perspective II. Reasons for use II. Interfaces I. Transport devices II. Particle

More information

Secondary Ion Mass Spectroscopy (SIMS)

Secondary Ion Mass Spectroscopy (SIMS) Secondary Ion Mass Spectroscopy (SIMS) Analyzing Inorganic Solids * = under special conditions ** = semiconductors only + = limited number of elements or groups Analyzing Organic Solids * = under special

More information

Ionization Methods in Mass Spectrometry at the SCS Mass Spectrometry Laboratory

Ionization Methods in Mass Spectrometry at the SCS Mass Spectrometry Laboratory Ionization Methods in Mass Spectrometry at the SCS Mass Spectrometry Laboratory Steven L. Mullen, Ph.D. Associate Director SCS Mass Spectrometry Laboratory Contact Information 31 oyes Laboratory (8:00-5:00

More information

BENG 183 Trey Ideker. Protein Sequencing

BENG 183 Trey Ideker. Protein Sequencing BENG 183 Trey Ideker Protein Sequencing The following slides borrowed from Hong Li s Biochemistry Course: www.sb.fsu.edu/~hongli/4053notes Introduction to Proteins Proteins are of vital importance to biological

More information

CHROMATOGRAPHY AND MASS SPECTROMETER

CHROMATOGRAPHY AND MASS SPECTROMETER 22 CHROMATOGRAPHY AND MASS SPECTROMETER 22.1 INTRODUCTION We know that the biochemistry or biological chemistry deals with the study of molecules present in organisms. These molecules are called as biomolecules

More information

Harris: Quantitative Chemical Analysis, Eight Edition

Harris: Quantitative Chemical Analysis, Eight Edition Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 21: MASS SPECTROMETRY CHAPTER 21: Opener 21.0 Mass Spectrometry Mass Spectrometry provides information about 1) The elemental composition of

More information

1) In what pressure range are mass spectrometers normally operated?

1) In what pressure range are mass spectrometers normally operated? Exercises Ionization 1) In what pressure range are mass spectrometers normally operated? Mass spectrometers are usually operated in the high vacuum regime to ensure mean free paths significantly longer

More information

CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #22 Mass Spectrometry: Chemical Ionization (Skoog,) (Harris, Chapt.) Mercer/Goodwill CEE 772 #22

More information

Mass Spectrometry. Electron Ionization and Chemical Ionization

Mass Spectrometry. Electron Ionization and Chemical Ionization Mass Spectrometry Electron Ionization and Chemical Ionization Mass Spectrometer All Instruments Have: 1. Sample Inlet 2. Ion Source 3. Mass Analyzer 4. Detector 5. Data System http://www.asms.org Ionization

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can be obtained from a MS spectrum?

More information

LC-MS Based Metabolomics

LC-MS Based Metabolomics LC-MS Based Metabolomics Analysing the METABOLOME 1. Metabolite Extraction 2. Metabolite detection (with or without separation) 3. Data analysis Metabolite Detection GC-MS: Naturally volatile or made volatile

More information

IDENTIFICATION OF ORGANOMETALLIC COMPOUNDS USING FIELD DESORPTION IONIZATION ON THE GCT

IDENTIFICATION OF ORGANOMETALLIC COMPOUNDS USING FIELD DESORPTION IONIZATION ON THE GCT IDETIFICATIO OF ORGAOMETALLIC COMPOUDS USIG FIELD DESORPTIO IOIZATIO O THE GCT David Douce 1, Michael Jackson 1, Robert Lewis 1, Peter Hancock 1, Martin Green 1 and Stuart Warriner 2 1 Waters Corporation,

More information

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS Mass Spectrometry Hyphenated Techniques GC-MS LC-MS and MS-MS Reasons for Using Chromatography with MS Mixture analysis by MS alone is difficult Fragmentation from ionization (EI or CI) Fragments from

More information

Propose a structure for an alcohol, C4H10O, that has the following

Propose a structure for an alcohol, C4H10O, that has the following Propose a structure for an alcohol, C4H10O, that has the following 13CNMR spectral data: Broadband _ decoupled 13CNMR: 19.0, 31.7, 69.5 б DEPT _90: 31.7 б DEPT _ 135: positive peak at 19.0 & 31.7 б, negative

More information

Liquid Secondary Ion Mass Spectrometry

Liquid Secondary Ion Mass Spectrometry 380 MASS SPECTRMETRY / Liquid Secondary Ion Mass Spectrometry Liquid Secondary Ion Mass Spectrometry N J Haskins, Hants, UK & 2005, Elsevier Ltd. All Rights Reserved. Introduction Liquid secondary-ion

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene experiment CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can we get from MS spectrum?

More information

4. How can fragmentation be useful in identifying compounds? Permits identification of branching not observed in soft ionization.

4. How can fragmentation be useful in identifying compounds? Permits identification of branching not observed in soft ionization. Homework 9: Chapters 20-21 Assigned 12 April; Due 17 April 2006; Quiz on 19 April 2006 Chap. 20 (Molecular Mass Spectroscopy) Chap. 21 (Surface Analysis) 1. What are the types of ion sources in molecular

More information

Quattro Micro - How does it work?

Quattro Micro - How does it work? Quattro Micro - How does it work? 1 Introduction This document is designed to familiarise you with the principles behind how the Quattro Micro works. The level of this document is designed as Level One

More information

High Performance Liquid Chromatography

High Performance Liquid Chromatography Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN

More information

High Performance Liquid Chromatography

High Performance Liquid Chromatography Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 1 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN

More information

Molecular Mass Spectrometry

Molecular Mass Spectrometry Molecular Mass Spectrometry Mass Spectrometry: capable of providing information about (1) Elemental composition of samples of matter: atomic mass (2) Structures of inorganic, organic, and biological molecules

More information

Choosing the metabolomics platform

Choosing the metabolomics platform GBS 748 Choosing the metabolomics platform Stephen Barnes, PhD 4 7117; sbarnes@uab.edu So, I have my samples what s next? You ve collected your samples and you may have extracted them Protein precipitation

More information

Chemistry 311: Topic 3 - Mass Spectrometry

Chemistry 311: Topic 3 - Mass Spectrometry Mass Spectroscopy: A technique used to measure the mass-to-charge ratio of molecules and atoms. Often characteristic ions produced by an induced unimolecular dissociation of a molecule are measured. These

More information

Electrospray Ion Trap Mass Spectrometry. Introduction

Electrospray Ion Trap Mass Spectrometry. Introduction Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic

More information

Introduction to GC/MS

Introduction to GC/MS Why Mass Spectrometry? Introduction to GC/MS A powerful analytical technique used to: 1.Identify unknown compounds 2. Quantify known materials down to trace levels 3. Elucidate the structure of molecules

More information

Extrel Application Note

Extrel Application Note Extrel Application Note Real-Time Plasma Monitoring and Detection of Trace H 2 O and HF Species in an Argon Based Plasma Jian Wei, 575 Epsilon Drive, Pittsburgh, PA 15238. (Presented at the 191st Electrochemical

More information

Ion sources. Ionization and desorption methods

Ion sources. Ionization and desorption methods Ion sources Ionization and desorption methods 1 2 Processes in ion sources 3 Ionization/ desorption Ionization Desorption methods Electron impact ionization Chemical ionization Electro-spray ionisation

More information

MS Goals and Applications. MS Goals and Applications

MS Goals and Applications. MS Goals and Applications MS Goals and Applications 3 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis

More information

Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Q&A Transcript ---------------- Q - How do you generate multiple charges reproductively? A - If you use the same settings on the

More information

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph SCIETIFIC ISTRUMET EWS 2016 Vol. 7 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation Introduction of Hitachi Chromaster 5610 MS Detector for High Performance

More information

Chemical Aspects of Mass Spectrometry

Chemical Aspects of Mass Spectrometry Chemical Aspects of Mass Spectrometry Abraham Badu Chemistry and Biochemistry, OSU July 12, 217 Spread of MS by Discipline https://masspec.scripps.edu/ mass spectrometry 2 1 Current Challenges in Mass

More information

Questions on Electric Fields

Questions on Electric Fields Questions on Electric Fields 1. The diagram shows a positively charged oil drop held at rest between two parallel conducting plates A and B. Oil drop A B 2.50 cm The oil drop has a mass 9.79 x 10 15 kg.

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Applied Biosystems QStar Pulsar i Features of the API QSTAR Pulsar i The

More information

A process whereby an electron is either removed from or added to the atom or molecule producing an ion in its ground state.

A process whereby an electron is either removed from or added to the atom or molecule producing an ion in its ground state. 12.3 Processes and techniques 12.3.1 Ionization nomenclature Adiabatic ionization A process whereby an electron is either removed from or added to the atom or molecule producing an ion in its ground state.

More information

Auxiliary Techniques Soft ionization methods

Auxiliary Techniques Soft ionization methods Auxiliary Techniques The limitations of the structural information in the normal mass spectrum can be partly offset by special mass-spectral techniques. Although a complete description of these is beyond

More information

Chromatography. Gas Chromatography

Chromatography. Gas Chromatography Chromatography Chromatography is essentially the separation of a mixture into its component parts for qualitative and quantitative analysis. The basis of separation is the partitioning of the analyte mixture

More information

Mass spectrometry.

Mass spectrometry. Mass spectrometry Mass spectrometry provides qualitative and quantitative information about the atomic and molecular composition of inorganic and organic materials. The mass spectrometer produces charged

More information

Chromatographic Methods of Analysis Section - 4 : Ion Exchange Chrom. Prof. Tarek A. Fayed

Chromatographic Methods of Analysis Section - 4 : Ion Exchange Chrom. Prof. Tarek A. Fayed Chromatographic Methods of Analysis Section - 4 : Ion Exchange Chrom. Prof. Tarek A. Fayed Ion Exchange Chromatography (IEC) In this type of chromatography, the solid stationary phase )organic resin) is

More information

DEPOSITION OF THIN TiO 2 FILMS BY DC MAGNETRON SPUTTERING METHOD

DEPOSITION OF THIN TiO 2 FILMS BY DC MAGNETRON SPUTTERING METHOD Chapter 4 DEPOSITION OF THIN TiO 2 FILMS BY DC MAGNETRON SPUTTERING METHOD 4.1 INTRODUCTION Sputter deposition process is another old technique being used in modern semiconductor industries. Sputtering

More information

Analysis of Polar Metabolites using Mass Spectrometry

Analysis of Polar Metabolites using Mass Spectrometry Analysis of Polar Metabolites using Mass Spectrometry TransMed Course: Basics in Clinical Proteomics and Metabolomics. Oct 10-19, 2012 dd.mm.yyyy Vidya Velagapudi, Ph.D, Adjunct Professor Head of the Metabolomics

More information

15.04.jpg. Mass spectrometry. Electron impact Mass spectrometry

15.04.jpg. Mass spectrometry. Electron impact Mass spectrometry Mass spectrometry Electron impact Mass spectrometry 70 ev = 1614 kcal/mol - contrast with energy from IR (1-10 kcal/mol) or NMR (0.2 cal/mol) - typical C-C bond = 100 kcal/mol Point: lots of energy in

More information

Chapter 27: Gas Chromatography

Chapter 27: Gas Chromatography Chapter 27: Gas Chromatography Gas Chromatography Mobile phase (carrier gas): gas (He, N 2, H 2 ) - do not interact with analytes - only transport the analyte through the column Analyte: volatile liquid

More information

Introduction to SIMS Basic principles Components Techniques Drawbacks Figures of Merit Variations Resources

Introduction to SIMS Basic principles Components Techniques Drawbacks Figures of Merit Variations Resources Introduction to SIMS Basic principles Components Techniques Drawbacks Figures of Merit Variations Resources New technique for surface chemical analysis. SIMS examines the mass of ions, instead of energy

More information

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC INTRODUCTION Principle: LC/MS is a technique that combines physical separation capabilities of liquid

More information

Open Column Chromatography, GC, TLC, and HPLC

Open Column Chromatography, GC, TLC, and HPLC Open Column Chromatography, GC, TLC, and HPLC Murphy, B. (2017). Introduction to Chromatography: Lecture 1. Lecture presented at PHAR 423 Lecture in UIC College of Pharmacy, Chicago. USES OF CHROMATOGRAPHY

More information

Introduction. Chapter 1. Learning Objectives

Introduction. Chapter 1. Learning Objectives Chapter 1 Introduction Learning Objectives To understand the need to interface liquid chromatography and mass spectrometry. To understand the requirements of an interface between liquid chromatography

More information

Finnigan LCQ Advantage MAX

Finnigan LCQ Advantage MAX www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Finnigan LCQ Advantage MAX The Finnigan LCQ Advantage MAX ion trap mass spectrometer

More information

Introduction to the Q Trap LC/MS/MS System

Introduction to the Q Trap LC/MS/MS System www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap LC/MS/MS Introduction to the Q Trap LC/MS/MS System The Q Trap

More information

Gas Chromatography (GC)

Gas Chromatography (GC) Gas Chromatography (GC) Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11541 Saudi Arabia Office: AA53

More information

ELECTROMAGNETIC WAVES

ELECTROMAGNETIC WAVES VISUAL PHYSICS ONLINE MODULE 7 NATURE OF LIGHT ELECTROMAGNETIC WAVES SPECTRA PRODUCED BY DISCHARGE TUBES CATHODE RAYS (electron beams) Streams of electrons (negatively charged particles) observed in vacuum

More information

Chapter 2 The Chemistry of Life

Chapter 2 The Chemistry of Life Chapter 2 The Chemistry of Life I. Water Liquid Naturally occurring It expands liquid to solid Covers more than 75% of our surface Most abundant in living organisms most important inorganic compound for

More information

API Optimization, Source Design, and Other Considerations of LC/MS

API Optimization, Source Design, and Other Considerations of LC/MS API Optimization, Source Design, and Other Considerations of LC/MS Instrument Design Considerations: The Atmosphere-Vacuum Interface and Pumping System The hows and whys of the vacuum? We ve got to get

More information

Secondary ion mass spectrometry (SIMS)

Secondary ion mass spectrometry (SIMS) Secondary ion mass spectrometry (SIMS) ELEC-L3211 Postgraduate Course in Micro and Nanosciences Department of Micro and Nanosciences Personal motivation and experience on SIMS Offers the possibility to

More information

Every living and nonliving things is made up of matter. MATTER: anything that has mass & takes up space. What does all matter have in common?

Every living and nonliving things is made up of matter. MATTER: anything that has mass & takes up space. What does all matter have in common? the basics Every living and nonliving things is made up of matter MATTER: anything that has mass & takes up space What does all matter have in common? Smallest unit of matter ALL matter is made of particles

More information

Mass Spectrometry for Chemists and Biochemists

Mass Spectrometry for Chemists and Biochemists Erasmus Intensive Program SYNAPS Univ. of Crete - Summer 2007 Mass Spectrometry for Chemists and Biochemists Spiros A. Pergantis Assistant Professor of Analytical Chemistry Department of Chemistry University

More information

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds Atmospheric Analysis Gases Sampling and analysis of gaseous compounds Introduction - External environment (ambient air) ; global warming, acid rain, introduction of pollutants, etc - Internal environment

More information

Field-Free Atmospheric Ionization: Atmospheric Pressure Chemical Ionization & Electrospray

Field-Free Atmospheric Ionization: Atmospheric Pressure Chemical Ionization & Electrospray Field-Free Atmospheric Ionization: Atmospheric Pressure Chemical Ionization & Electrospray Edward W. Sheehan 1, Ross C. Willoughby 1, Robert Classon 2, and Dan Dodgen 2 1 Chem-Space Associates, Inc., Pittsburgh,

More information

Strategies for the Analysis of Therapeutic Peptides in Biofluids by LC-MS/MS. Lee Goodwin

Strategies for the Analysis of Therapeutic Peptides in Biofluids by LC-MS/MS. Lee Goodwin Strategies for the Analysis of Therapeutic Peptides in Biofluids by LC-MS/MS Lee Goodwin Sample Preparation Chromatography Detection General Strategies Examples New Approaches Summary Outline ABUNDANCE

More information

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography Tips and Techniques to Optimize ELSD and MS based Purification Utilizing ELSD and MS as Secondary Detectors for Prep

More information

Secondary Ion Mass Spectrometry (SIMS)

Secondary Ion Mass Spectrometry (SIMS) CHEM53200: Lecture 10 Secondary Ion Mass Spectrometry (SIMS) Major reference: Surface Analysis Edited by J. C. Vickerman (1997). 1 Primary particles may be: Secondary particles can be e s, neutral species

More information

3) In CE separation is based on what two properties of the solutes? (3 pts)

3) In CE separation is based on what two properties of the solutes? (3 pts) Final Exam Chem 311 Fall 2002 December 16 Name 1) (3 pts) In GC separation is based on the following two properties of the solutes a) polarity and size b) vapor pressure and molecular weight c) vapor pressure

More information

Liquid Chromatography

Liquid Chromatography Liquid Chromatography 1. Introduction and Column Packing Material 2. Retention Mechanisms in Liquid Chromatography 3. Method Development 4. Column Preparation 5. General Instrumental aspects 6. Detectors

More information

Secondaryionmassspectrometry

Secondaryionmassspectrometry Secondaryionmassspectrometry (SIMS) 1 Incident Ion Techniques for Surface Composition Analysis Mass spectrometric technique 1. Ionization -Electron ionization (EI) -Chemical ionization (CI) -Field ionization

More information

Mass Spectrometry. Fundamental LC-MS. Electrospray Ionisation Instrumentation

Mass Spectrometry. Fundamental LC-MS. Electrospray Ionisation Instrumentation Mass Spectrometry Fundamental LC-MS Electrospray Ionisation Instrumentation i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be

More information

CHAPTER 6: Etching. Chapter 6 1

CHAPTER 6: Etching. Chapter 6 1 Chapter 6 1 CHAPTER 6: Etching Different etching processes are selected depending upon the particular material to be removed. As shown in Figure 6.1, wet chemical processes result in isotropic etching

More information

Electronic Supplementary Information

Electronic Supplementary Information Electronic Supplementary Material (ESI) for RSC Advances. This journal is The Royal Society of Chemistry 2014 S1 Electronic Supplementary Information Flash Carboxylation: Fast Lithiation - Carboxylation

More information

Nova 600 NanoLab Dual beam Focused Ion Beam IITKanpur

Nova 600 NanoLab Dual beam Focused Ion Beam IITKanpur Nova 600 NanoLab Dual beam Focused Ion Beam system @ IITKanpur Dual Beam Nova 600 Nano Lab From FEI company (Dual Beam = SEM + FIB) SEM: The Electron Beam for SEM Field Emission Electron Gun Energy : 500

More information

CHROMATOGRAPHY. The term "chromatography" is derived from the original use of this method for separating yellow and green plant pigments.

CHROMATOGRAPHY. The term chromatography is derived from the original use of this method for separating yellow and green plant pigments. CHROMATOGRAPHY The term "chromatography" is derived from the original use of this method for separating yellow and green plant pigments. THEORY OF CHROMATOGRAPHY: Separation of two sample components in

More information

Table of Content. Mechanical Removing Techniques. Ultrasonic Machining (USM) Sputtering and Focused Ion Beam Milling (FIB)

Table of Content. Mechanical Removing Techniques. Ultrasonic Machining (USM) Sputtering and Focused Ion Beam Milling (FIB) Table of Content Mechanical Removing Techniques Ultrasonic Machining (USM) Sputtering and Focused Ion Beam Milling (FIB) Ultrasonic Machining In ultrasonic machining (USM), also called ultrasonic grinding,

More information

Mass Spectroscopy. Base peak. Molecular Ion peak. The positively charged fragments produced are separated, based on their mass/charge (m/z) ratio. M+.

Mass Spectroscopy. Base peak. Molecular Ion peak. The positively charged fragments produced are separated, based on their mass/charge (m/z) ratio. M+. Mass spectrometry is the study of systems causing the formation of gaseous ions, with or without fragmentation, which are then characteried by their mass to charge ratios (m/) and relative abundances.

More information

CHAPTER CHROMATOGRAPHIC METHODS OF SEPARATIONS

CHAPTER CHROMATOGRAPHIC METHODS OF SEPARATIONS Islamic University in Madinah Department of Chemistry CHAPTER - ----- CHROMATOGRAPHIC METHODS OF SEPARATIONS Prepared By Dr. Khalid Ahmad Shadid Chemistry Department Islamic University in Madinah TRADITIONAL

More information

3. Separation of a Mixture into Pure Substances

3. Separation of a Mixture into Pure Substances 3. Separation of a Mixture into Pure Substances Paper Chromatography of Metal Cations What you will accomplish in this experiment This third experiment provides opportunities for you to learn and practice:

More information

Chapter 3 Engineering Science for Microsystems Design and Fabrication

Chapter 3 Engineering Science for Microsystems Design and Fabrication Lectures on MEMS and MICROSYSTEMS DESIGN and MANUFACTURE Chapter 3 Engineering Science for Microsystems Design and Fabrication In this Chapter, we will present overviews of the principles of physical and

More information

Shodex TM ODP2 HP series columns

Shodex TM ODP2 HP series columns HPLC Columns Shodex TM ODP2 HP series columns Better retention of highly polar substances Technical notebook No. 6 Contents 1. Introduction 1-1. Specifications 1-2. Eluent Compatibility of ODP2 HP Series

More information

MS/MS .LQGVRI0606([SHULPHQWV

MS/MS .LQGVRI0606([SHULPHQWV 0DVV6SHFWURPHWHUV Tandem Mass Spectrometry (MS/MS) :KDWLV0606" Mass spectrometers are commonly combined with separation devices such as gas chromatographs (GC) and liquid chromatographs (LC). The GC or

More information

EEE4106Z Radiation Interactions & Detection

EEE4106Z Radiation Interactions & Detection EEE4106Z Radiation Interactions & Detection 2. Radiation Detection Dr. Steve Peterson 5.14 RW James Department of Physics University of Cape Town steve.peterson@uct.ac.za May 06, 2015 EEE4106Z :: Radiation

More information

Chapter 5. Complexation of Tholins by 18-crown-6:

Chapter 5. Complexation of Tholins by 18-crown-6: 5-1 Chapter 5. Complexation of Tholins by 18-crown-6: Identification of Primary Amines 5.1. Introduction Electrospray ionization (ESI) is an excellent technique for the ionization of complex mixtures,

More information

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA Use of High Speed/High Resolution Size-Based Chromatographic Separation of Surfactants and Oligomeric Materials with Single Quadrupole Mass Spectrometry Michael O Leary, Jennifer Gough, Tanya Tollifson

More information

Instrumental Analysis II Course Code: CH3109. Chromatographic &Thermal Methods of Analysis Part 1: General Introduction. Prof. Tarek A.

Instrumental Analysis II Course Code: CH3109. Chromatographic &Thermal Methods of Analysis Part 1: General Introduction. Prof. Tarek A. Instrumental Analysis II Course Code: CH3109 Chromatographic &Thermal Methods of Analysis Part 1: General Introduction Prof. Tarek A. Fayed What is chemical analysis? Qualitative analysis (1) Chemical

More information

( 1+ A) 2 cos2 θ Incident Ion Techniques for Surface Composition Analysis Ion Scattering Spectroscopy (ISS)

( 1+ A) 2 cos2 θ Incident Ion Techniques for Surface Composition Analysis Ion Scattering Spectroscopy (ISS) 5.16 Incident Ion Techniques for Surface Composition Analysis 5.16.1 Ion Scattering Spectroscopy (ISS) At moderate kinetic energies (few hundred ev to few kev) ion scattered from a surface in simple kinematic

More information

A) at equilibrium B) endergonic C) endothermic D) exergonic E) exothermic.

A) at equilibrium B) endergonic C) endothermic D) exergonic E) exothermic. CHEM 2770: Elements of Biochemistry Mid Term EXAMINATION VERSION A Date: October 29, 2014 Instructor: H. Perreault Location: 172 Schultz Time: 4 or 6 pm. Duration: 1 hour Instructions Please mark the Answer

More information

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB. Quantitative analysis of small molecules in biological samples 100 Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB % 0 300 500 700 900 1100 1300 1500 1700 m/z Class Overview Introduction

More information

HPLC Workshop 16 June 2009 What does this do? Chromatography Theory Review Several chromatographic techniques Even though each method utilizes different techniques to separate compounds, the principles

More information

Agilent G3212 GC-APCI Source

Agilent G3212 GC-APCI Source Agilent G3212 GC-APCI Source Quick Start Guide Where to find information 2 Getting Started 3 Step 1. Start the Data Acquisition program for the GC and the Q-TOF 3 Step 2. Prepare the GC and Q-TOF for data

More information