Scintillation Detectors Particle Detection via Luminescence

Size: px
Start display at page:

Download "Scintillation Detectors Particle Detection via Luminescence"

Transcription

1 Scintillation Detectors Particle Detection via Luminescence

2 Scintillators General Characteristics Principle: de/dx converted into visible light Detection via photosensor [e.g. photomultiplier, human eye...] Main Features: Sensitivity to energy Fast time response Pulse shape discrimination Plastic Scintillator BC412 Requirements High efficiency for conversion of exciting energy to fluorescent radiation Transparency to its fluorescent radiation to allow transmission of light Emission of light in a spectral range detectable for photosensors Short decay time to allow fast response

3 Scintillators Basic Counter Setup Thin window Mu Metal Shield Iron Protective Shield Light Photomultiplier [or other photosensor] PMT Base [voltage divider network etc.] Scintillator Photosensors Scintillator Types: Organic Scintillators Inorganic Crystals Photomultipliers Gases Micro-Channel Plates Hybrid Photo Diodes Visible Light Photon Counter Silicon Photo Multipliers PMT Pulse F--? Output Signal Time [ns]

4 Inorganic Crystals Materials: Sodium iodide (NaI) Cesium iodide (CsI) Barium fluoride (BaF2)... Mechanism: Energy deposition by ionization Energy transfer to impurities Radiation of scintillation photons Time constants: exciton band impurities [activation centers] scintillation [luminescence] quenching excitations conduction band electron hole traps valence band Energy bands in impurity activated crystal showing excitation, luminescence, quenching and trapping Fast: recombination from activation centers [ns... μs] Slow: recombination due to trapping [ms... s]

5 Inorganic Crystals Crystal growth Example CMS Electromagnetic Calorimeter PbW04 ingots One of the last CMS end-cap crystals

6 Inorganic Crystals Time Constants Light Output Exponential decay of scintillation can be resolved into two components... N = Ae t/τ f + Be t/τ s τ f : decay constant of fast component τ s : decay constant of slow component Time

7 Inorganic Crystals Light Output Scintillation Spectrum for NaI and CsI Intensity [a.u.] NaI(Tl) CsI(Na) CsI(Tl) Wavelength [nm] Strong Temperature Dependence [in contrast to organic scintillators]

8 Inorganic Crystals Light Output Spectral sensitivity

9 Scintillation in Liquid Nobel Gases Materials: Helium (He) Liquid Argon (LAr) Liquid Xenon (LXe)... A Excitation A* Decay time constants: Helium : τ1 =.02 μs, τ2 = 3 μs Argon : τ1.02 μs Excited molecules Collision [with other gas atoms] A2* De-excitation and dissociation A A UV LAr : 130 nm LKr : 150 nm LXe : 175 nm Ionization A* Ionized molecules A2 + A2* e Recombination

10 Inorganic Scintillators Properties Scintillator material Density [g/cm 3 ] Refractive Index Wavelength [nm] for max. emission Decay time constant [μs] Photons/MeV NaI xxx NaI(Tl) xxx CsI(Tl) xxx Bi4Ge3O xxx CsF xxx LSO xxx PbWO xxx LHe / xxx LAr * / xxx LXe * / xxx * at 170 nm

11 Inorganic Scintillators Properties Numerical examples: NaI(Tl) PBWO4 Scintillator quality: λmax = 410 nm; hν = 3 ev photons/mev = τ = 250 ns λmax = 420 nm; hν = 3 ev photons/mev = 200 τ = 6 ns Light yield εsc fraction of energy loss going into photons e.g. NaI(Tl) : photons; 3 ev/photon εsc = ev/10 6 ev = 11.3% PBWO4: 200 photons; 3 ev/photon εsc = ev/10 6 ev = 0.06% [for 1 MeV particle]

12 Organic Scintillators Aromatic hydrocarbon compounds: e.g. Naphtalene [C10H8] Antracene [C14H10] Stilbene [C14H12]... Very fast! [Decay times of O(ns)] Naphtalene Scintillation is based on electrons of the C = C bond... Antracene π bond Scintillation light arises from delocalized electrons in π-orbitals... Transitions of 'free' electrons... Two pz orbitals

13 Organic Scintillators Molecular states: Absorption in 3-4 ev range Singlet states Triplet states Fluorescence in UV range [~ 320 nm] usage of wavelength shifters Fluorescence : S1 S0 [< 10-8 s] Phosphorescence : T0 S0 [> 10-4 s]

14 Organic Scintillators Transparency requires: Shift of absorption and emission spectra... Shift due to Franck-Condon Principle Intensity Energy Stokes-Shift Absorption Emission S1 Excited State λ Excitation into higher vibrational states De-excitation from lowest vibrational state S0 Ground State Excitation time scale : s Vibrational time scale : s S1 lifetime : 10-8 s Vibrational States Nuclear distance

15 Plastic and Liquid Scintillators In practice use... solution of organic scintillators [solved in plastic or liquid] + large concentration of primary fluor + smaller concentration of secondary fluor +... Scintillator requirements: Scintillator array with light guides Solvable in base material High fluorescence yield Absorption spectrum must overlap with emission spectrum of base material LSND experiment

16 Plastic and Liquid Scintillators A Energy deposit in base material excitation Solvent Primary fluorescent - Good light yield... - Absorption spectrum matched to excited states in base material... B Secondary fluorescent C Wave length shifter Excitations S1A γa Primary Fluor S1B γb Secondary Fluor S1C γc S0A S0B S0C

17 Plastic and Liquid Scintillators Some widely used solvents and solutes POPOP Polystyrene p-terphenyl

18 Wavelength Shifting Principle: Absorption of primary scintillation light Re-emission at longer wavelength Schematics of wavelength shifting principle Adapts light to spectral sensitivity of photosensor Requirement: Good transparency for emitted light

19 Organic Scintillators Properties Scintillator material Density [g/cm 3 ] Refractive Index Wavelength [nm] for max. emission Decay time constant [ns] Photons/MeV Naphtalene xxx Antracene xxx p-terphenyl xxx NE102* xxx NE104* xxx NE110* xxx NE111* xxx BC400** xxx BC428** xxx BC443** xxx * Nuclear Enterprises, U.K. ** Bicron Corporation, USA

20 Organic Scintillators Properties

21 Organic Scintillators Properties Light yield: [without quenching] dl dx = L de 0 dx Quenching: non-linear response due to saturation of available states Birk's law: dl dx = L 0 de dx 1+kB de dx [kb needs to be determined experimentally] Also other... parameterizations... Response different... for different particle types...

22 Scintillators Comparison Inorganic Scintillators Advantages high light yield [typical; εsc 0.13] high density [e.g. PBWO4: 8.3 g/cm 3 ] good energy resolution Disadvantages complicated crystal growth large temperature dependence Expensive Organic Scintillators Advantages very fast easily shaped small temperature dependence pulse shape discrimination possible Disadvantages lower light yield [typical; εsc 0.03] radiation damage Cheap

23 Scintillation Counters Setup Scintillator light to be guided to photosensor Light guide [Plexiglas; optical fibers] Light transfer by total internal reflection [maybe combined with wavelength shifting] Liouville's Theorem: Complete light transfer impossible as Δx Δθ = const. [limits acceptance angle] Use adiabatic light guide like 'fish tail'; appreciable energy loss 'fish tail'

24 Scintillation Counters Setup Photomultiplier WS fibre Wavelength-shifting fibre ATLAS Tile Calorimeter Scintillator Steel Source tubes

25 Photon Detection Purpose : Convert light into a detectable electronic signal Principle : Use photo-electric effect to convert photons to photo-electrons (p.e.) Requirement : High Photon Detection Efficiency (PDE) or Quantum Efficiency; Q.E. = Np.e./Nphotons Available devices [Examples]: Photomultipliers [PMT] Micro Channel Plates [MCP] Photo Diodes [PD] HybridPhoto Diodes [HPD] Visible Light Photon Counters [VLPC] Silicon Photomultipliers [SiPM]

26 Photomultipliers Principle: Electron emission from photo cathode Secondary emission from dynodes; dynode gain: 3-50 [f(e)] Typical PMT Gain: > 10 6 [PMT can see single photons...] PMT Collection

27 Photomultipliers Photocathode γ-conversion via photo effect... Bialkali: SbRbCs; SbK2Cs Photon entrance window photo cathode 4-step process: Electron Electron generation via ionization Propagation through cathode Escape of electron into vacuum Q.E % [need specifically developed alloys]

28 Photomultipliers Dynode Chain Dynodes Electron Anode UB Voltage divider Multiplication process: Electrons accelerated toward dynode Further electrons produced avalanche Secondary emission coefficient: δ = #(e produced)/#(e incoming) Typical: δ = 2 10 G = δ n = 8 15 n = Gain fluctuation: δ = kud; G = a0 (kud) n dg/g = n dud/ud = n dub/ub

29 Photomultipliers Dynode Chain Optimization of PMT gain Anode isolation Linearity Transit time B-field dependence Venetian blind Box and grid Linear focused PM s are in general very sensitive to B-fields! Even to earth field (30-60 μt). μ-metal shielding required. Circular focused

30 Photomultipliers Energy Resolution Energy resolution influenced by: Linearity of PMT: at high dynode current possibly saturation by space charge effects; IA nγ for 3 orders of magnitude possible... Photoelectron statistics: given by poisson statistics. P n (n e )= nn e e n e n! n e = de dx Photons MeV η Q.E. σ n /n =1/ n e η = 0.2; Q.E. =0.25 σ n /n =0.7% Secondary electron fluctuations: with ne given by de/dx... For NaI(Tl) and 10 MeV photon; photons/mev = 40000; light collection efficiency n e =20000 P n (δ) = δn e δ n! σ n /n =1/ δ with dynode gain δ; and with N dynodes... σn n 2 = 1 δ δ N 1 δ 1 σn/<n> dominated by first dynode stage important for single photon detection

31 CuBe dynode Photomultipliers Energy Resolution For detection of single photons δ 'standard' dynodes δ Philips photonic GaP(Cs) Philips photonic Negative electron affinity (NEA) E [ev] E [kev] counts 1 p.e. counts 1 p.e. 2p.e. Large δ!... yields better energy resolution Philips photonic CuBe dynode NAE dynode Houtermanns NIM 112 (1973) 121 pulse height pulse height

32 Micro Channel Plate "2D Photomultiplier" Gain: Fast signal [time spread ~ 50 ps] B-Field tolerant [up to 0.1T] But: limited life time/rate capability

33 Chapter 2 Light Detectors ICFA the device is covered with an anti-reflecting SIO2 lay Silicon Photomultipliers Layout Silicon Photomultiplier:SiPM Geiger Mode tracks on the surface connect all pixels to the commo Principle: + n Pixelized photo diodes Pixelsinoperated in Geiger mode operated Geiger Mode (non-linear response) Single pixel works as a binary device Si* Resistor Vbias Al - conductor Si02 + Guard ring n- Photocurrent (log scale) Features: Granularity Gain Bias Voltage Efficiency - EE,, V/cm V/cm 1E5 Energy = #photons seen by summing over all pixels no amplification linear amplification 6 1E E4 10 1E3 2 1E2 10 1E1 1 1E0 non-linear Geiger mode Doping Structure of SiPM [1] * Resistor 2 Al - conductor 103Sipixels/mm : : 106 : < 100 V : ca. 30 % E, V/cm Figure 2.13: Left: Schematic view of the SiPM topology Avalanche region material on the low resistive substrate serves as a drift reg electron generated in this region will subsequently drift i a b layer where the electrical field is high enough for avalanch Breakdown Voltage electrical field in order to avoid unwanted avalanche break 1: (a) Silicon photomultiplier microphotograph, (b) impurity levels are higher. Right: Diagram of the topol electric Works atfigure room temperature! Reverse Bias Voltage Insensitive to magnetic fields tion in epitaxy layer Gain and Single Pixel Response Alexander Tadday

34 Silicon Photomultipliers HAMAMATSU MPPC 400Pixels One of the first SiPM Pulsar, Moscow

35 Silicon Photomultipliers CALICE HCAL Prototype

36 Scintillation Counters Applications Time of flight (ToF) counters Energy measurement (calorimeters) Hodoscopes; fibre trackers Trigger systems ATLAS Minimum Bias Trigger Scintillators Particle track in scintillating fibre hodoscope

37 H1 Spaghetti Calorimeter Scintillator : BICRON BCF-12 Photosensor : Photomultipliers

38 CMS Crystal Calorimeter (ECAL)

39 CMS Crystal Calorimeter (ECAL) Scintillator : PBW04 [Lead Tungsten] Photosensor : APDs [Avalanche Photodiodes] Number of crystals: ~ Light output: 4.5 photons/mev Barrel ECAL (EB) y = = Preshower (ES) z = 2.6 = 3.0 Endcap ECAL (EE)

40 ATLAS Tile Calorimeter Photomultiplier WS fibre Wavelength-shifting fibre ATLAS Tile Calorimeter Scintillator Steel Source tubes

41 CALICE Analogue HCAL Sandwich structure: 1 m 3 - Prototype 38 layers - Scintillator Tiles+WLS+SiPMs (.5 cm) - Stainless steel absorber (1.6 cm)!"#$$%&'%()(*+,%-.//(-+(0 2006/2007 CERN Testbeam [2008/09, Fermilab] 3x3 cm 2 Tile Mounted SiPM Scintillator : Plastic Photosensor : SiPMs

42 CALICE Scintillator ECAL X-Layer [1 x 4 cm 2 ] Tungsten Scintillator layers: 2 mm Tungsten layers: 3 mm Y-Layer [1 x 4 cm 2 ] X/Y-Strips: 1 x 4 cm 2 Granularity: 1 x 1 cm 2 X-Layer [1 x 4 cm 2 ] Readout: MPPC Channels: ~ 10 7 Scintillator SiPM [1600 pixels] SiPM r/o with WLS Particle WLS Fibre

Scintillation Detectors Particle Detection via Luminescence. Kolanoski, Wermes

Scintillation Detectors Particle Detection via Luminescence. Kolanoski, Wermes Scintillation Detectors Particle Detection via Luminescence Kolanoski, Wermes Scintillators General Characteristics Principle: de/dx converted into visible light Detection via photosensor [e.g. photomultiplier,

More information

Scintillators General Characteristics

Scintillators General Characteristics Scintillators General Characteristics Principle: de/dx converted into visible light Detection via photosensor [e.g. photomultiplier, human eye ] Main Features: Sensitivity to energy Fast time response

More information

Inorganic scintillators. Geometries and readout

Inorganic scintillators. Geometries and readout Topics of this lecture K K Inorganic scintillators Organic scintillators K Geometries and readout K Fiber tracking K Photo detectors Particle Detectors Christian Joram III/1 Scintillation Scintillation

More information

PHYS 3446 Lecture #12

PHYS 3446 Lecture #12 PHYS 3446 Lecture #12 Wednesday, Oct. 18, 2006 Dr. 1. Particle Detection Ionization Detectors MWPC Scintillation Counters Time of Flight 1 Announcements Next LPCC Workshop Preparation work Each group to

More information

Chapter 4 Scintillation Detectors

Chapter 4 Scintillation Detectors Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 4-1 4.1. Basic principle of the scintillator Chapter 4 Scintillation Detectors Scintillator Light sensor Ionizing radiation Light (visible,

More information

Scintillators Definitions - 1!

Scintillators Definitions - 1! Scintillators! Scintillators Definitions - 1! Luminescence: Emission of photons (visible light, UV, X ray) after absorption of energy. Energy depostion in the material by! Light Photoluminescence! Heat

More information

Lecture 16 Light transmission and optical detectors

Lecture 16 Light transmission and optical detectors Lecture 6 Light transmission and optical detectors Charged particle traversing through a material can generate signal in form of light via electromagnetic interactions with orbital electrons of the atoms

More information

Platinum resistance. also wirewound versions. eg

Platinum resistance. also wirewound versions. eg Platinum resistance Platinum resistance Very stable and reproducible, wide T range (~ -200 C to 1000 C) T coefficient ~ +0.4%/ C Bulky and expensive for some applications (~ 2-3) need wires (R) or local

More information

Scintillation Detectors

Scintillation Detectors Scintillation Detectors J.L. Tain Jose.Luis.Tain@ific.uv.es http://ific.uv.es/gamma/ Instituto de Física Corpuscular C.S.I.C - Univ. Valencia Scintillation detector: SCINTILLATION MATERIAL LIGHT-GUIDE

More information

Detector technology. Aim of this talk. Principle of a radiation detector. Interactions of gamma photons (gas) Gas-filled detectors: examples

Detector technology. Aim of this talk. Principle of a radiation detector. Interactions of gamma photons (gas) Gas-filled detectors: examples Aim of this tal Detector technology WMIC Educational Program Nuclear Imaging World Molecular Imaging Congress, Dublin, Ireland, Sep 5-8, 202 You can now the name of a bird in all the languages of the world,

More information

Scintillation Detectors

Scintillation Detectors Radiation Measurement Systems Scintillation Detectors Ho Kyung Kim Pusan National University Scintillation detector = scintillator + light sensor Scintillators Inorganic alkali halide crystals Best light

More information

General Information. Individual Study Projects. Muon Lifetime Experiment. Today s Agenda. Wednesday: We have a few more slots to fill

General Information. Individual Study Projects. Muon Lifetime Experiment. Today s Agenda. Wednesday: We have a few more slots to fill General Information Individual Study Projects We have a few more slots to fill Muon Lifetime Experiment We can use the experiment until April 18 Monday, April 8 we will go over the setup and take a look

More information

Calorimetry I Electromagnetic Calorimeters

Calorimetry I Electromagnetic Calorimeters Calorimetry I Electromagnetic Calorimeters Introduction Calorimeter: Detector for energy measurement via total absorption of particles... Also: most calorimeters are position sensitive to measure energy

More information

Content. Introduction. Interaction of Radiation with Matter. Definitions Detectors for Ionizing Particles

Content. Introduction. Interaction of Radiation with Matter. Definitions Detectors for Ionizing Particles Introduction Overview of detector systems Sources of radiation Radioactive decay Cosmic Radiation Accelerators Content Interaction of Radiation with Matter General principles Charged particles heavy charged

More information

SCINTILLATION DETECTORS AND PM TUBES

SCINTILLATION DETECTORS AND PM TUBES SCINTILLATION DETECTORS AND PM TUBES General Characteristics Introduction Luminescence Light emission without heat generation Scintillation Luminescence by radiation Scintillation detector Radiation detector

More information

Week 6: Ch. 8 Scintillation Counters

Week 6: Ch. 8 Scintillation Counters Week 6: Ch. 8 cintillation Counters Proportional Counters Principles of cintillation Counters -- organic materials --- light production -- inorganic materials --- light production -- light output, collection

More information

Energetic particles and their detection in situ (particle detectors) Part II. George Gloeckler

Energetic particles and their detection in situ (particle detectors) Part II. George Gloeckler Energetic particles and their detection in situ (particle detectors) Part II George Gloeckler University of Michigan, Ann Arbor, MI University of Maryland, College Park, MD Simple particle detectors Gas-filled

More information

Scintillators. Detectors for Particle Physics Thomas Bergauer Institute of High Energy Physics, Vienna, Austria

Scintillators. Detectors for Particle Physics Thomas Bergauer Institute of High Energy Physics, Vienna, Austria Scintillators Detectors for Particle Physics Thomas Bergauer Institute of High Energy Physics, Vienna, Austria 5 Scintillators Content 5.1 General Introduction 5.2 Inorganic Scintillators 5.2.1 Inorganic

More information

Particle Detectors A brief introduction with emphasis on high energy physics applications

Particle Detectors A brief introduction with emphasis on high energy physics applications Particle Detectors A brief introduction with emphasis on high energy physics applications TRIUMF Summer Institute 2006 July 10-21 2006 Lecture I measurement of ionization and position Lecture II scintillation

More information

Scintillation detectors

Scintillation detectors 25 de dx Scintillation detectors excitation L25.pdf P627 YK 3/14/2012 detectable photons also by UV, or molecular collisions, chem. reactions, bubbles. etc. Detector building requirements (sometimes controversial):

More information

05 - Scintillation detectors

05 - Scintillation detectors 05 - Scintillation detectors Jaroslav Adam Czech Technical University in Prague Version 2 Jaroslav Adam (CTU, Prague) DPD_05, Scintillation detectors Version 2 1 / 39 Scintillation detector principles

More information

EEE4106Z Radiation Interactions & Detection

EEE4106Z Radiation Interactions & Detection EEE4106Z Radiation Interactions & Detection 2. Radiation Detection Dr. Steve Peterson 5.14 RW James Department of Physics University of Cape Town steve.peterson@uct.ac.za May 06, 2015 EEE4106Z :: Radiation

More information

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors Radiation Detectors 1 How do we detect ionizing radiation? Indirectly, by its effects as it traverses matter? What are these effects? Ionization and excitation of the atoms and molecules Heat 2 Types of

More information

Radionuclide Imaging MII Detection of Nuclear Emission

Radionuclide Imaging MII Detection of Nuclear Emission Radionuclide Imaging MII 3073 Detection of Nuclear Emission Nuclear radiation detectors Detectors that are commonly used in nuclear medicine: 1. Gas-filled detectors 2. Scintillation detectors 3. Semiconductor

More information

PHOTODETECTORS AND SILICON PHOTO MULTIPLIER

PHOTODETECTORS AND SILICON PHOTO MULTIPLIER ESE seminar Photodetectors - Sipm, P. Jarron - F. Powolny 1 PHOTODETECTORS AND SILICON PHOTO MULTIPLIER ESE seminar Pierre Jarron, Francois Powolny OUTLINE 2 Brief history and overview of photodetectors

More information

Introduction to scintillators

Introduction to scintillators Introduction to scintillators M. Kobayashi (KEK) 17 November, 2003 1. Luminescence, fluorescence, scintillation, phosphorescence, etc. 2. Scintillation mechanism 3. Scintillation efficiency 4. Main characteristics

More information

Lecture # 3. Muhammad Irfan Asghar National Centre for Physics. First School on LHC physics

Lecture # 3. Muhammad Irfan Asghar National Centre for Physics. First School on LHC physics Lecture # 3 Muhammad Irfan Asghar National Centre for Physics Introduction Gaseous detectors Greater mobility of electrons Obvious medium Charged particles detection Particle information easily transformed

More information

DETECTORS. I. Charged Particle Detectors

DETECTORS. I. Charged Particle Detectors DETECTORS I. Charged Particle Detectors A. Scintillators B. Gas Detectors 1. Ionization Chambers 2. Proportional Counters 3. Avalanche detectors 4. Geiger-Muller counters 5. Spark detectors C. Solid State

More information

Radiation Detector 2016/17 (SPA6309)

Radiation Detector 2016/17 (SPA6309) Radiation Detector 2016/17 (SPA6309) Semiconductor detectors (Leo, Chapter 10) 2017 Teppei Katori Semiconductor detectors are used in many situations, mostly for some kind of high precision measurement.

More information

Part III. Interaction of Photons with Matter. Photon Absorption Length λ. Intensity Attenuation. Three effects are important: Scintillators

Part III. Interaction of Photons with Matter. Photon Absorption Length λ. Intensity Attenuation. Three effects are important: Scintillators Part III Interaction of Photons with Matter Scintillators Photodetectors Cherenkov detectors Transition radiation detectors Calorimeters - shower development - electromagnetic calorimeters - hadronic calorimeters

More information

Particle Energy Loss in Matter

Particle Energy Loss in Matter Particle Energy Loss in Matter Charged particles loose energy when passing through material via atomic excitation and ionization These are protons, pions, muons, The energy loss can be described for moderately

More information

Seminar talks. Overall description of CLAS12 (Jefferson Lab) MAPS. Talks on Feb. 6 th, (Contact JR) (Contact TS)

Seminar talks. Overall description of CLAS12 (Jefferson Lab) MAPS. Talks on Feb. 6 th, (Contact JR) (Contact TS) Seminar talks Overall description of CLAS12 (Jefferson Lab) (Contact JR) MAPS (Contact TS) Talks on Feb. 6 th, 2015 Review old ionization detectors: Emulsion, Cloud chambers, Ionization chambers, Spark

More information

Studies of Hadron Calorimeter

Studies of Hadron Calorimeter Studies of Hadron Calorimeter Zhigang Wang Institute of High Energy Physics 2012.10.17 in IHEP Outline 1,The Dark Matter Calorimeter 2,The Hadron Calorimeter(HCAL) 3, Summary 1,Dark Matter Calorimeter

More information

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency. Light We can use different terms to describe light: Color Wavelength Frequency Light is composed of electromagnetic waves that travel through some medium. The properties of the medium determine how light

More information

Electronic bubble chamber

Electronic bubble chamber Topics of this lecture Inorganic scintillators Organic scintillators Geometries and readout Photo detectors Fiber tracking Nuclear emulsions Electronic bubble chamber III/1 Scintillation Scintillation

More information

Gamma and X-Ray Detection

Gamma and X-Ray Detection Gamma and X-Ray Detection DETECTOR OVERVIEW The kinds of detectors commonly used can be categorized as: a. Gas-filled Detectors b. Scintillation Detectors c. Semiconductor Detectors The choice of a particular

More information

Chapter 6: Basic radiation detectors

Chapter 6: Basic radiation detectors Chapter 6: Basic radiation detectors Set of 60 slides based on the chapter authored by C.W.E. VAN EIJK Faculty of Applied Sciences, Delft University of Technology, Delft, Netherlands of the publication

More information

Single Photon detectors

Single Photon detectors Single Photon detectors Outline Motivation for single photon detection Semiconductor; general knowledge and important background Photon detectors: internal and external photoeffect Properties of semiconductor

More information

Unit 2. Instrumentation. Experts Teaching from Practical Experience

Unit 2. Instrumentation. Experts Teaching from Practical Experience Unit 2 Instrumentation Experts Teaching from Practical Experience Gas-Filled Detectors Gas-filled detectors measure the charge released when radiation interacts with the gas Three types: Ion Chambers,

More information

Scintillators 1. YEAR, I. CYCLE. AUTHOR: BPhys. Manja Ščetinec ADVISOR/MENTOR: Prof. Boštjan Golob

Scintillators 1. YEAR, I. CYCLE. AUTHOR: BPhys. Manja Ščetinec ADVISOR/MENTOR: Prof. Boštjan Golob Scintillators SEMINAR 1. YEAR, I. CYCLE AUTHOR: BPhys. Manja Ščetinec ADVISOR/MENTOR: Prof. Boštjan Golob Faculty of Mathematics and Physics, University in Ljubljana Ljubljana, October 2017 Abstract In

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement June 2008 Tom Lewellen Tkldog@u.washington.edu Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron

More information

Introduction to Radiation Monitoring

Introduction to Radiation Monitoring Introduction to Radiation Monitoring Iain Darby Honorary Research Fellow, University of Glasgow iain.darby@glasgow.ac.uk https://at.linkedin.com/in/idarby https://www.facebook.com/iain.darby.662 Outline

More information

Particle Energy Loss in Matter

Particle Energy Loss in Matter Particle Energy Loss in Matter Charged particles, except electrons, loose energy when passing through material via atomic excitation and ionization These are protons, pions, muons, The energy loss can

More information

Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection

Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection R. Kreuger, C. W. E. van Eijk, Member, IEEE, F. A. F. Fraga, M. M. Fraga, S. T. G. Fetal, R. W. Hollander, Member, IEEE, L. M.

More information

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 19 Chapter 12 There are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry X-ray Techniques include:

More information

Inorganic Scintillators

Inorganic Scintillators Inorganic Scintillators Inorganic scintillators are inorganic materials (usually crystals) that emit light in response to ionizing radiation NaI is the protypical example Scintillation mechanism is different

More information

Characterisation of Silicon Photomultipliers for the T2K Experiment

Characterisation of Silicon Photomultipliers for the T2K Experiment Characterisation of Silicon Photomultipliers for the T2K Experiment, 18th May 2010 Martin Haigh, University of Oxford Outline Brief introduction to the T2K experiment. Overall configuration and goals.

More information

Scintillation Detector

Scintillation Detector Scintillation Detector Introduction The detection of ionizing radiation by the scintillation light produced in certain materials is one of the oldest techniques on record. In Geiger and Marsden s famous

More information

Scintillation and Light Sensitive Detectors - I. Mark Chen Queen s University

Scintillation and Light Sensitive Detectors - I. Mark Chen Queen s University Scintillation and Light Sensitive Detectors - I Mark Chen Queen s University How Particles Produce Light charged particles can produce light in materials Čerenkov radiation scintillation in some materials

More information

Radioactivity. Lecture 6 Detectors and Instrumentation

Radioactivity. Lecture 6 Detectors and Instrumentation Radioactivity Lecture 6 Detectors and Instrumentation The human organs Neither humans nor animals have an organ for detecting radiation from radioactive decay! We can not hear it, smell it, feel it or

More information

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? 2. Briefly discuss dead time in a detector. What factors are important

More information

Energy Loss of Electrons. Detectors for Particle Physics Part III. Intensity Attenuation. Interaction of Photons with Matter

Energy Loss of Electrons. Detectors for Particle Physics Part III. Intensity Attenuation. Interaction of Photons with Matter Detectors for Particle Physics Part III Interaction of electrons, photons and hadrons with matter Scintillators Photodetectors Cherenov Counters Transition radiation nergy Loss of lectrons In addition

More information

PoS(TIPP2014)033. Upgrade of MEG Liquid Xenon Calorimeter. Ryu SAWADA. ICEPP, the University of Tokyo

PoS(TIPP2014)033. Upgrade of MEG Liquid Xenon Calorimeter. Ryu SAWADA. ICEPP, the University of Tokyo ICEPP, the University of Tokyo E-mail: sawada@icepp.s.u-tokyo.ac.jp The MEG experiment yielded the most stringent upper limit on the branching ratio of the flavorviolating muon decay µ + e + γ. A major

More information

Response curve measurement of the SiPM

Response curve measurement of the SiPM Response curve measurement of the SiPM The theoritical formula : N Fired = N pix. (1 (1 1/N pix ) Npe ) does not hold because: 1) crosstalk and afterpulses increases N Fired (by a factor 1 / (1 ε) for

More information

Experimental Particle Physics

Experimental Particle Physics Experimental Particle Physics Particle Interactions and Detectors Lecture 2 2nd May 2014 Fergus Wilson, RAL 1/31 How do we detect particles? Particle Types Charged (e - /K - /π - ) Photons (γ) Electromagnetic

More information

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 55, NO. 4, AUGUST /$ IEEE

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 55, NO. 4, AUGUST /$ IEEE IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 55, NO. 4, AUGUST 2008 2425 Optical and Scintillation Properties of Inorganic Scintillators in High Energy Physics Rihua Mao, Member, IEEE, Liyuan Zhang, Member,

More information

Particle Detectors Tools of High Energy and Nuclear Physics Detection of Individual Elementary Particles

Particle Detectors Tools of High Energy and Nuclear Physics Detection of Individual Elementary Particles Particle Detectors Tools of High Energy and Nuclear Physics Detection of Individual Elementary Particles Howard Fenker Jefferson Lab May 31, 2006 Outline of Talk Interactions of Particles with Matter Atomic

More information

Photon Instrumentation. First Mexican Particle Accelerator School Guanajuato Oct 6, 2011

Photon Instrumentation. First Mexican Particle Accelerator School Guanajuato Oct 6, 2011 Photon Instrumentation First Mexican Particle Accelerator School Guanajuato Oct 6, 2011 Outline The Electromagnetic Spectrum Photon Detection Interaction of Photons with Matter Photoelectric Effect Compton

More information

PHY492: Nuclear & Particle Physics. Lecture 25. Particle Detectors

PHY492: Nuclear & Particle Physics. Lecture 25. Particle Detectors PHY492: Nuclear & Particle Physics Lecture 25 Particle Detectors http://pdg.lbl.gov/2006/reviews/contents_sports.html S(T ) = dt dx nz = ρa 0 Units for energy loss Minimum ionization in thin solids Z/A

More information

Energy Loss of Electrons. Detectors for Particle Physics Part III. Intensity Attenuation. Interaction of Photons with Matter

Energy Loss of Electrons. Detectors for Particle Physics Part III. Intensity Attenuation. Interaction of Photons with Matter Detectors for Particle Physics Part III Interaction of electrons, photons and hadrons with matter Scintillators Photodetectors Cherenov Counters Transition radiation nergy Loss of lectrons In addition

More information

Nuclear Physics Laboratory. Gamma spectroscopy with scintillation detectors. M. Makek Faculty of Science Department of Physics

Nuclear Physics Laboratory. Gamma spectroscopy with scintillation detectors. M. Makek Faculty of Science Department of Physics Nuclear Physics Laboratory Gamma spectroscopy with scintillation detectors M. Makek Faculty of Science Department of Physics Zagreb, 2015 1 1 Introduction The goal of this excercise is to familiarize with

More information

Factors Affecting Detector Performance Goals and Alternative Photo-detectors

Factors Affecting Detector Performance Goals and Alternative Photo-detectors XENON Experiment - SAGENAP Factors Affecting Detector Performance Goals and Alternative Photo-detectors Department of Physics Brown University Source at http://gaitskell.brown.edu Gaitskell Review WIMP

More information

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy Experiment VI Gamma Ray Spectroscopy 1. GAMMA RAY INTERACTIONS WITH MATTER In order for gammas to be detected, they must lose energy in the detector. Since gammas are electromagnetic radiation, we must

More information

A gas-filled calorimeter for high intensity beam environments

A gas-filled calorimeter for high intensity beam environments Available online at www.sciencedirect.com Physics Procedia 37 (212 ) 364 371 TIPP 211 - Technology and Instrumentation in Particle Physics 211 A gas-filled calorimeter for high intensity beam environments

More information

Experimental Particle Physics

Experimental Particle Physics Experimental Particle Physics Particle Interactions and Detectors 20th February 2007 Fergus Wilson, RAL 1 How do we detect Particles? Particle Types Charged (e - /K - /π - ) Photons (γ) Electromagnetic

More information

Application of Birks' law of scintillator nonlinearity in Geant4. Alexander Tadday Kirchhoff Institute for Physics Heidelberg University

Application of Birks' law of scintillator nonlinearity in Geant4. Alexander Tadday Kirchhoff Institute for Physics Heidelberg University Alexander Tadday Alexander - IRTG Tadday Meeting - IRTG - Heidelberg Meeting - 05.11.20 Application of Birks' law of scintillator nonlinearity in Geant4 Alexander Tadday Kirchhoff Institute for Physics

More information

Prospects for achieving < 100 ps FWHM coincidence resolving time in time-of-flight PET

Prospects for achieving < 100 ps FWHM coincidence resolving time in time-of-flight PET Prospects for achieving < 100 ps FWHM coincidence resolving time in time-of-flight PET, 28-Feb-2012, ICTR-PHE, Geneva, Switzerland 1 Time-of-flight PET Colon cancer, left upper quadrant peritoneal node

More information

Last Lecture 1) Silicon tracking detectors 2) Reconstructing track momenta

Last Lecture 1) Silicon tracking detectors 2) Reconstructing track momenta Last Lecture 1) Silicon tracking detectors 2) Reconstructing track momenta Today s Lecture: 1) Electromagnetic and hadronic showers 2) Calorimeter design Absorber Incident particle Detector Reconstructing

More information

Radiation Dose, Biology & Risk

Radiation Dose, Biology & Risk ENGG 167 MEDICAL IMAGING Lecture 2: Sept. 27 Radiation Dosimetry & Risk References: The Essential Physics of Medical Imaging, Bushberg et al, 2 nd ed. Radiation Detection and Measurement, Knoll, 2 nd Ed.

More information

V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber

V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber CYRIC Annual Report 2001 V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber Kawata N. Baba M. Kato M.*, Miura T.**, and Yamadera A.***, Cyclotron and Radioisotope Center, Tohoku

More information

2nd-Meeting. Ionization energy loss. Multiple Coulomb scattering (plural and single scattering, too) Tracking chambers

2nd-Meeting. Ionization energy loss. Multiple Coulomb scattering (plural and single scattering, too) Tracking chambers 2nd-Meeting Ionization energy loss Multiple Coulomb scattering (plural and single scattering, too) Tracking chambers #2 -Particle Physics Experiments at High Energy Colliders John Hauptman, Kyungpook National

More information

Interaction of particles in matter

Interaction of particles in matter Interaction of particles in matter Particle lifetime : N(t) = e -t/ Particles we detect ( > 10-10 s, c > 0.03m) Charged particles e ± (stable m=0.511 MeV) μ ± (c = 659m m=0.102 GeV) ± (c = 7.8m m=0.139

More information

Time-Resolved Liquid Scintillation Counting

Time-Resolved Liquid Scintillation Counting CHAPTER 44 Time-Resolved Liquid Scintillation Counting Norbert Roessler, Robert J. Valenta, and Stat van Cauter ABSTRACT A comparison is made between standard, two-tube coincidence liquid scintillation

More information

Particle Detectors. History of Instrumentation History of Particle Physics. The Real World of Particles. Interaction of Particles with Matter

Particle Detectors. History of Instrumentation History of Particle Physics. The Real World of Particles. Interaction of Particles with Matter Particle Detectors History of Instrumentation History of Particle Physics The Real World of Particles Interaction of Particles with Matter Tracking with Gas and Solid State Detectors Calorimetry, Particle

More information

pp physics, RWTH, WS 2003/04, T.Hebbeker

pp physics, RWTH, WS 2003/04, T.Hebbeker 3. PP TH 03/04 Accelerators and Detectors 1 pp physics, RWTH, WS 2003/04, T.Hebbeker 2003-12-16 1.2.4. (Inner) tracking and vertexing As we will see, mainly three types of tracking detectors are used:

More information

Lectures Overview. Particle Detectors Detectors

Lectures Overview. Particle Detectors Detectors Lectures 12 Particle Detectors Dec 2006, Lecture 12 Nuclear Physics Lectures, Dr. Armin Reichold 1 11.1 Detectors For photons only 11.0 Overview Photomultiplier and APD For charged particles and photons

More information

7 Particle Identification. Detectors for Particle Physics Manfred Krammer Institute of High Energy Physics, Vienna, Austria

7 Particle Identification. Detectors for Particle Physics Manfred Krammer Institute of High Energy Physics, Vienna, Austria 7 Particle Identification Detectors for Particle Physics Manfred Krammer Institute of High Energy Physics, Vienna, Austria 7.0 Content 7.1 Methods for Particle Identification 7.2 Mass of Charged Particles

More information

OPTI510R: Photonics. Khanh Kieu College of Optical Sciences, University of Arizona Meinel building R.626

OPTI510R: Photonics. Khanh Kieu College of Optical Sciences, University of Arizona Meinel building R.626 OPTI510R: Photonics Khanh Kieu College of Optical Sciences, University of Arizona kkieu@optics.arizona.edu Meinel building R.626 Announcements Homework #6 is assigned, due May 1 st Final exam May 8, 10:30-12:30pm

More information

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy Topic 2b: X-ray Fluorescence Spectrometry Text: Chapter 12 Rouessac (1 week) 4.0 X-ray Fluorescence Download, read and understand EPA method 6010C ICP-OES Winter 2009 Page 1 Atomic X-ray Spectrometry Fundamental

More information

Lectures Overview Detectors (for photons only, PMT) Particle Detectors Detectors For photons only. End of Lecture 11)

Lectures Overview Detectors (for photons only, PMT) Particle Detectors Detectors For photons only. End of Lecture 11) Lectures 11 Particle Detectors 1 For photons only 11.0 Overview Photomultiplier and APD For charged particles and photons Scintillators Gas-counters Semi-conductors (GeLi, Si) 11.2 Example analysis of

More information

Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector

Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector C. Fiorini, member, IEEE, A. Gola, M. Zanchi, A. Longoni, P. Lechner, H. Soltau, L. Strüder Abstract In this work

More information

Direct WIMP Detection in Double-Phase Xenon TPCs

Direct WIMP Detection in Double-Phase Xenon TPCs Outline PMTs in the XENON dark matter experiment XENON100 and the weekly gain calibration XENON1T and candidates for the light sensors Tests of Hamamatsu R11410 2 Direct WIMP Detection in Double-Phase

More information

Experimental Particle Physics

Experimental Particle Physics Experimental Particle Physics Particle Interactions and Detectors Lecture 2 17th February 2010 Fergus Wilson, RAL 1/31 How do we detect particles? Particle Types Charged (e - /K - /π - ) Photons (γ) Electromagnetic

More information

Time of Flight measurements with MCP-PMT

Time of Flight measurements with MCP-PMT International Symposium on the Development of Detectors, 2006/4 at SLAC Time of Flight measurements with MCP-PMT - Very high resolution TOF counter - Lifetime of MCP-PMTs T.Ohshima, K.Inami, N.Kishimoto,

More information

GEM-based gaseous Photomultipliers for UV and visible photon imaging. Dirk Mörmann Amos Breskin Rachel Chechik Marcin Balcerzyk Bhartendu Singh

GEM-based gaseous Photomultipliers for UV and visible photon imaging. Dirk Mörmann Amos Breskin Rachel Chechik Marcin Balcerzyk Bhartendu Singh GEM-based gaseous Photomultipliers for UV and visible photon imaging Dirk Mörmann Amos Breskin Rachel Chechik Marcin Balcerzyk Bhartendu Singh Gaseous Photomultiplier State of the art: Advantages: large

More information

1 Introduction. KOPIO charged-particle vetos. K - RARE Meeting (Frascati) May Purpose of CPV: veto Kl

1 Introduction. KOPIO charged-particle vetos. K - RARE Meeting (Frascati) May Purpose of CPV: veto Kl Introduction - Purpose of CPV: veto Kl decay modes with a real or apparent π and a pair of charged particles - Examples of background modes: (i) K l π π + π (ii) K l π π ± eν there are always (iii) K l

More information

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors 5. Radiation Microsensors Radiation µ-sensors convert incident radiant signals into standard electrical out put signals. Radiant Signals Classification

More information

Time-of-Flight PET using Cherenkov Photons Produced in PbF 2

Time-of-Flight PET using Cherenkov Photons Produced in PbF 2 Photons Produced in PbF 2 R. Dolenec a, S. Korpar b,a, P. Križan c,a, R. Pestotnik a, A. Stanovnik d,a a, Ljubljana, Slovenia b Faculty of Chemistry and Chemical Engineering, University of Maribor, Slovenia

More information

A RICH Photon Detector Module with G-APDs

A RICH Photon Detector Module with G-APDs A RICH Photon Detector Module with G-APDs S. Korpar a,b, H. Chagani b, R. Dolenec b, P. Križan b,c, R. Pestotnik b, A. Stanovnik b,c a University of Maribor, b J. Stefan Institute, c University of Ljubljana

More information

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Module - 2 Radioisotopes Techniques Lecture - 3 GM Counting and

More information

Study of novel gaseous photomultipliers for UV and visible light

Study of novel gaseous photomultipliers for UV and visible light Study of novel gaseous photomultipliers for UV and visible light Thesis for the degree of Ph. D. in physics presented to the Scientific Council of the Weizmann Institute of Science, Rehovot, Israel by

More information

Contents. Charged Particles. Coulomb Interactions Elastic Scattering. Coulomb Interactions - Inelastic Scattering. Bremsstrahlung

Contents. Charged Particles. Coulomb Interactions Elastic Scattering. Coulomb Interactions - Inelastic Scattering. Bremsstrahlung Contents Marcel MiGLiERiNi Nuclear Medicine, Radiology and Their Metrological Aspects. Radiation in Medicine. Dosimetry 4. Diagnostics & Therapy 5. Accelerators in Medicine 6. Therapy Planning 7. Nuclear

More information

Gamma Spectroscopy. References: Objectives:

Gamma Spectroscopy. References: Objectives: Gamma Spectroscopy References: G.F. Knoll, Radiation Detection and Measurement (John Wiley & Sons, New York, 2000) W. R. Leo, Techniques for Nuclear and Particle Physics Experiments: A How-to Approach,

More information

David B. Cassidy. Department of Physics and Astronomy, University of California, Riverside, USA. Varenna, July 09

David B. Cassidy. Department of Physics and Astronomy, University of California, Riverside, USA. Varenna, July 09 Experimental production of many- positron systems: L2, techniques David B. Cassidy Department of Physics and Astronomy, University of California, Riverside, USA cassidy@physics.ucr.edu Varenna, July 09

More information

Beam diagnostics: Alignment of the beam to prevent for activation. Accelerator physics: using these sensitive particle detectors.

Beam diagnostics: Alignment of the beam to prevent for activation. Accelerator physics: using these sensitive particle detectors. Beam Loss Monitors When energetic beam particles penetrates matter, secondary particles are emitted: this can be e, γ, protons, neutrons, excited nuclei, fragmented nuclei... Spontaneous radiation and

More information

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY ABSORBED DOSE: The amount of energy absorbed, as a result of radiation passing through a material, per unit mass of material. Measured in rads (1 rad

More information

The Mössbauer Effect

The Mössbauer Effect Experimental Physics V85.0112/G85.2075 The Mössbauer Effect Spring, 2005 Tycho Sleator, David Windt, and Burton Budick Goals The main goal of this experiment is to exploit the Mössbauer effect to measure

More information

The fast light of CsI(Na) crystals

The fast light of CsI(Na) crystals CPC (HEP & NP), 2011, xx(x): 1-5 Chinese Physics C Vol. xx, No. X, Xxx, 2011 The fast light of CsI(Na) crystals Xilei Sun 1;1) Junguang Lu 1 Tao Hu 1 Li Zhou 1 Jun Cao 1 Yifang Wang 1 Liang Zhan 1 Boxiang

More information

R&D on Astroparticles Detectors. (Activity on CSN )

R&D on Astroparticles Detectors. (Activity on CSN ) R&D on Astroparticles Detectors (Activity on CSN5 2000-2003) Introduction Results obtained with the R&D activity (2000-2003) with some drift chambers prototypes are reported. With different photocathode

More information

Particle Detectors and Quantum Physics (2) Stefan Westerhoff Columbia University NYSPT Summer Institute 2002

Particle Detectors and Quantum Physics (2) Stefan Westerhoff Columbia University NYSPT Summer Institute 2002 Particle Detectors and Quantum Physics (2) Stefan Westerhoff Columbia University NYSPT Summer Institute 2002 More Quantum Physics We know now how to detect light (or photons) One possibility to detect

More information