Nucleic Acid Quantitation in Microplates

Size: px
Start display at page:

Download "Nucleic Acid Quantitation in Microplates"

Transcription

1 PHARMACEUTICAL application note Nucleic Acid Quantitation in Microplates HansWilly Mueller, Ph.D. J. Fenton Williams, Ph.D. Introduction Many molecular biology applications require the determination of ss-, dsdna, or RNA concentration in various samples. Most commonly, UV spectroscopy or Fluorescence spectroscopy is the preferred method of choice. Most simple and most convenient is UV spectroscopy, because nucleic acids have a characteristic UV absorbance spectrum, which can be used or quantitation without further sample derivatization and the technique is non-destructive; the sample can be used again after measurement. Nucleic Acids show a typical absorbance maximum at 260 nm in the UV range of the spectrum (see Figure 1). This wavelength is therefore used for quantitation of a DNA sample. According to Beer-Lambert s law, the absorbance reading is proportional to the concentration. Typical proportionality constants are 50 for dsdna, 33 for ssdna and 40 for RNA. Absorbance readings of the nucleic acids obtained in a 1 cm cell multiplied by the respective proportionality constant will give the concentration in µg/ml. Most common Fluorescence methods use PicoGreen dye for dsdna determination, OliGreen for ssdna determination, and RiboGreen for RNA determination. Both dyes are measured in a fluorescence spectrometer with 485 nm excitation wavelength and 535 nm emission wavelength. A big advantage of Figure 1. UV Absorbance Spectrum of a 50 µg/ml dsdna sample, showing typical absorbance maximum around 260 nm and minimum at 230 nm. UV Absorbance at 260 nm fluorescence methods is their selectivity. This means that the above methods can be used for selective detection of any single nucleic acid type in the presence of one or more of the others. Fluorescence methods are also more sensitive than UV absorbance measurements. The sample, however, has to be treated with a fluorescent reagent for measurement and thus no longer behaves as a native molecule in subsequent procedures. Since fluorescence is so sensitive, however, it is usually possible to analyze only a tiny fractional portion of the entire sample. For sensitivity comparison of Fluorescence methods and UV Absorbance methods see Table 1. Fluorescence ssdna 100 ng/ml-100 µg/ml OliGreen 1 ng/ml-1 µg/ml dsdna 150 ng/ml-150 µg/ml PicoGreen 250 pg/ml-1 µg/ml RNA 120 ng/ml-120 µg/ml RiboGreen 1 ng/ml-1 µg/ml Table 1. Comparison of measurement ranges for DNA quantitation using UV Absorption spectroscopy and Fluorescence spectroscopy. UV measurement range is based on measurement at 260 nm in a 1 cm cuvette. Look to us. And see more.

2 Sample throughput in conventional UV and Fluorescence spectrometers is very limited and relatively high sample volumes, 0.3 to 3 ml, are needed for standard cuvette measurements. For measurements in 96 well microplates only µl sample is needed and higher sample throughput can be achieved. Newer applications therefore use the universally standardized 96 well format for DNA preparations (for example DNA Miniprep-Kits). Also oligonucleotide synthesis is getting more and more productive, requiring fast and simple measurement of synthesis yields. These technology advancements thus require a different approach for DNA quantitation than standard UV measurements with conventional spectrometers. UV and Fluorescence microplate readers are ideal for these higher throughput situations, because 96 low volume samples can be measured in much less than a minute. All measurements described here were obtained with the Perkin-Elmer HTS 7000 BioAssay Reader. This novel system allows simple software selection of both through-well absorbance measurements and top or bottom fluorescence measurements in any style microplate. The ability to do both types of measurements (UV absorbance and fluorescence) in one instrument has many advantages. The most obvious advantage is the flexibility to optimize different kinds of tests in one instrument. In addition it is also very useful for fluorescence measurement to know the absorbance of the fluorescent dye. The absorbance at excitation wavelength should be below 0.05A for linear fluorescence measurement. 384 (16x24) well format, with 80 µl maximum volume, is rapidly gaining acceptance and can be accommodated by the HTS 7000 BioAssay Reader. Fluorescence measurements are usually performed in white or black opaque microplates. Excitation light is directed via an optical fiber toward the sample in the well. The emitted fluorescent light from the reporter dye is captured by another fiber and directed to the detector. For special applications such as averaging multiple readings in cell function assays, epifluorescence measurement is also done from the bottom of special plates with clear well bottoms and opaque sides. UV measurements must be done with transparent microplates. Because DNA is measured at 260 nm, special plates are needed, made from an UV transmitting material. Conventional polystyrene microplates, which are commonly used for colorimetric ELISA tests, cannot be used for DNA measurements because they do not transmit UV light. For the UV measurements, special commercially available UV transparent plates were used. These plates have excellent UV transmission properties. As can be seen from Figure 2, these plates remarkably transmit 85% of 260 nm light. Converted to absorbance, this means that an empty well will give an A260 reading of only 0.07A, which is excellent for low noise DNA measurement and can be easily subtracted as a blank reading. Different than in a conventional UV/Vis spectrophotometer using a cuvette, the pathlength of the sample in a microplate will naturally vary, depending on the sample volume in the well, the slope of the sides of the well and the shape of the bottom of the well. The optical pathlength (b) in microwells usually need not be known exactly, but is precisely controlled by the typically excellent liquid handling reproducibility (see Figure 3). Since pipetting accuracy and precision is very good, the volume accuracy and precision achieved Using Microplates A wide range of multi-well microplates are available on the market. The most commonly used type is the 96 (8x12) well plate with 350 µl maximum volume, but other formats are used as well. A trend towards plates with more wells is evolving due to high sample throughput needs of drug screening assays. So the Figure 2. Transmittance spectrum of a single well of a Corning Costar UV microplate, showing usable wavelength range between 210 nm and 800 nm.transmittance of an empty well, measured at 260 nm is about 85% (= 0.07A). Typical Standard Deviations across the empty plate are < 5 ma for empty wells. Figure 3. Optical pathlength in microwell is dependent on the sample volume. Theoretical pathlength may be calculated as b = Vol, with r = radius of the well (mm) π*r 2 Vol= sample volume (µl). 2

3 by modern pipetters will directly translate to pathlength accuracy and precision. Precision of mulitichannel pipettes is typically < 0.8% for 50 µl or < 0.2% for 300 µl volume. The accuracy is about ±1.5% for 50 µl and ± 0.6% for 300 µl volume. This variable pathlength in the wells of microplates has the advantage that assay detection sensitivity can be controlled. For more concentrated samples, less volume may be used, whereas for diluted samples, the well may be filled completely. The variable pathlength thus offers a convenient means of sensitivity adjustment without requiring special cuvette-sets with various pathlengths. For practical experiments, however, it is not recommended and not necessary to calculate the pathlength geometrically from the volume. For all volumes, the pathlengths are typically less than predicted by geometry due to the meniscus shape. The meniscus shape may be different, depending on plate type, buffer or possible detergent used. Also, because well shape (wall slope and bottom shape) may vary from plate to plate a geometric calculation of pathlength is not advisable. The difference between calculated (theoretical) and optically determined pathlength is shown in Table 2. The theoretical pathlength was calculated from the dimensions (6.4 mm diameter) of a well of a 96 well plate and various volumes using the equation b = Vol π*r 2. Volume Calculated Pathlength Optically measured (µl) (mm) pathlength (mm) Table 2. Calculated pathlength compared with actually measured pathlength. Pathlength as a function of sample volume filled into the wells of a Costar UV Plate. Sample:Various volumes of 50 µg/ml DNA solution, measured at 260 nm. The experimental determination was done with a 50 µg/ml DNA solution of which the respective volumes were pipetted into the wells of a 96 well Costar UV Microplate (Catalogue Number 3635). Ten replicates were measured per volume in the HTS 7000 Bio Assay Reader at 260 nm. The same sample was measured in a standard 1 cm cell in a conventional UV spectrometer. In both cases blank correction was done with the same volume of buffer. From the Absorbance values measured in the plate reader the optical pathlength was calculated, based on the A (1 cm) readings from the conventional UV spectrometer. Calibration For practical DNA quantitation in microplates, it is not necessary to know, compute or correct for the pathlength. According to Beer- Lambert s law, A=K*c, and there is a linear relation between absorbance reading and concentration. The proportionality factor K is the product of the molar extinction coefficient and the pathlength, neither of which have to be known independently. When working with a reference solution, a simple single standard calibration is sufficient and the result will be correct, independent of the pathlength (or volume) used. A pathlength correction is not required and therefore this one standard method is very convenient and reliable. It is good laboratory practice to do a complete calibration curve once, to check the linear measurement range of the spectrometer system and to assure that the compound to be measured obeys Beer-Lambert s law. This is shown in Figure 4 and Figure 5. A 200 µl aliquot of a serial dilution of 250 µg/ml ds DNA was pipetted in five replicates into the microwells of a 96 well plate. Figure 4 shows the Figure 4. HTSoft Plate map for dsdna calibration with 11 standards, 5 replicates each. 3

4 Absorbance Absorbance dsdna Absorbance Calibration Curve y= * x d= r= dsdna Absorbance Calibration Curve y= * x d= r= ds DNA concentration (µg/ml) ds DNA concentration (µg/ml) Figure 5. Calibration curve obtained from calibration with 11 standards of dsdna (200 µl per well). Linear calibration curve fit. Calibration curves were obtained with the Plate map from Figure 4. Figure 5a: linear curve fit with all standards. Figure 5b: Linear curve fit with standards above 2.5 A excluded. Volume in Absorbance in Approximate dsdna Microplate Microplate with Absorbance in Concentration HTS cm cell 300 µl 2.5 A 2.8 A 140 µg/ml 250 µl 2.5 A 3.4 A 170 µg/ml 200 µl 2.5 A 4 A 200 µg/ml 150 µl 2.5 A 5.5 A 275 µg/ml 100 µl 2.5 A 8 A 400 µg/ml 50 µl 2.5 A 16 A 800 µg/ml Table 3. Maximum concentration of dsdna, which can be measured in a microplate with the HTS 7000 BioAssay Reader. Maximum absolute amount of dsdna is about 40 µg. For comparison the theoretical absorbance values are listed as they would be measured (if possible) in a 1 cm cell in a conventional UV Spectrometer. Usually, the range of a 1 cm cuvette-based UV spectrometers is 0 to 3A. plate map, and Figure 5 shows the resulting calibration curves. The linear regression calibration curve shows an excellent correlation coefficient of r= , although very high absorbing samples are included in the analysis. The HTSoft instrument control and evaluation software calculated this statistical value from the curve fit of the measured standards. The software will also allow optimization of the calibration curve, by offering selective exclusion of known incorrect data points and recalculating another curve without these points. The two calibration curves from Figure 5 illustrate this data editing capability. When the two highest concentrations with absorbances above 2.5A are not used, the correlation coefficient improves incrementally to r= Both curves are equally useful for any typical DNA quantitation in molecular biology labs. In a 200 µl sample, it is possible to measure dsdna up to or above 2.5 absorbance units with the HTS 7000, depending on the precision required. 2.5A linear measurement means that dsdna concentrations of 800 µg/ml can be measured with HTS 7000 BioAssay Reader, when 50 µl sample is filled in the wells. As it can be seen from Table 3 these high absorbance values correspond to very high concentrated DNA solutions which will be impossible to measure with a 1 cm cell in a conventional UV spectrometer without dilution. An undiluted 800 µg/ml dsdna solution would result in a theoretical absorbance of about 16A when a 1 cm cell is used. The maximum absolute amount of dsdna, which can be measured with the HTS 7000 BioAssay Reader without dilution is about 40 µg. Quantitative DNA Measurement in Practice When it is once proven that a measurement system gives linear response to the DNA concentration, it is not anymore required to perform extensive calibrations. A simple universal concentration factor is not 4

5 Figure 6. HTSoft Plate map for single point calibration. Blank and 200 µg/ml standard are measured with 4 replicates. recommended, because DNA samples from different origins have unique molar extinction coefficients, depending on their base composition. A quick calibration is therefore recommended. Single point calibrations are sufficient within the proven linear range of the system. As long as the sample volume per well is kept constant for a set of samples, one single standard can be applied. This single standard is quickly prepared and should contain the same type of DNA, diluted in the same buffer system as the sample. Actual measurement of this single standard should be done in 2-4 replicates (see Figure 6). HTSoft will calculate the average and the standard deviation of the replicates. The single point calibration will give accurate results, because the standard is measured under identical conditions as the samples. Especially it is a very elegant way to compensate for pathlength effects. Also, other parameters such as volume, temperature, buffer system and preparation steps etc., are the same for the standard and the samples and their effects are automatically compensated for in a single calibration point system. This way of measuring DNA concentration is much better than measurement with known pathlength. When pathlength is known, a precise extinction coefficient must be known in order to obtain correct results. The correct extinction coefficient, however, may vary with solvent, buffer salts or ph. So it really is easier and strictly most relevant to prepare and to measure one standard. Knowledge of pathlength and extinction coefficient is not required for correct results. Even when samples in groups with different volumes in one plate have to be measured, this is easily possible using one standard for each group. HTSoft allows up to 12 different experimental groups to be defined on a single 96 well plate. Once an assay has been established and reproducibility is acceptable, HTSoft can support the common requirement of needing to apply the standard curve from the plate with the single point standard to other plates with just samples. An application with increasing importance is DNA mini preparation. DNA preparation kits facilitate simple and rapid DNA preparation (Qiagen Qiaprep and Qiawell ). These kits use the 96 well format and allow preparation of plasmid DNA within less than an hour. The eluate can be collected in an UV transparent microplate and then it is ready for direct quantitation. Eluates from these preparation kits contain up to 20 micrograms of DNA in a 75 µl Figure 7. HTSoft Concentration results from DNA miniprep. 5

6 eluate, which can be measured in 30 seconds using the HTS 7000 BioAssay Reader. The pathlength at this volume is large enough to allow reproducible absorbance measurements, as long as air bubbles on the meniscus surface are avoided. Also the measurement range is ideal: With 20 µg DNA in the 75 µl eluate, the absorbance at 260 nm will be slightly above 1, which is well within the linear measurement range of the HTS The following brief protocol describes how to determine dsdna concentration in 75 µl eluates obtained directly from Qiagen DNA minipreps. 1. Preparation of standard solution. You can either use a defined dilution of the DNA type to be prepared with the Miniprep. Recommended concentration is 200 µg/ml: Dissolve 1 mg DNA in 5 ml 10 mm Tris-Cl buffer (ph 8.5) (e.g. Sigma D-3644; 1 mg preweighed DNA that is readily soluble may be used). This dsdna standard solution has a concentration of 200 µg/ml. 2. Setup of Plate reader method on the HTS This method will use absorbance endpoint mode with 260 nm as measurement wavelength and an HTSoft plate map as shown in Figure 6. This is an example where wells A1-D1 are filled with 75 µl buffer for blank measurement and wells A2-D2 are filled with 75 µl standard solution (= 4 replicates each). The rest of the plate is filled with 75 µl sample (eluate), obtained from the Qiagen miniprep columns. In this example four replicates are used for blank and the single standard. Absorbance 0.75 Single Point Calibration for dsdna Quantitation Figure 8. HTSoft single point calibration graph for concentration determination. 75µL sample was used. 3. Plate Filling. Wells A1-D1 will be filled with 75 µl buffer solution, Wells A2-D2 will be filled with 75 µl DNA standard solution. All other wells of the microplate will be filled with 75 µl sample eluate from the miniprep columns. 4. Plate measurement. Measurement of the entire plate will take about 25 seconds. After this, the results will be displayed. Concentration results (see Figure 7) are based on single point calibration in wells A2-D2. The respective calibration graph is shown in Figure 8. Conclusion For labs with many samples to process, the HTS 7000 Bio Assay Reader in the UV absorbance mode at 260 nm offers a very quick and convenient way to measure DNA concentrations on a daily basis. The [ds DNA ] µg/ml popular 96 well plate format now offers affordable UV-transparency capabilities and therefore obviates the need for more complex and cumbersome quartz cuvette sipper and autosampler systems. The universality of the microplate format assures a smooth transition for DNA samples being processed in parallel at molecular biology labs. High sample throughput, low sample volume and wide concentration detection range are the main advantages. The HTS 7000 is of special interest for DNA quantitation in DNA Minipreparations. HTSoft provides final DNA concentration results from any array of samples and blanks. As long as pipetting is precise and reproducible, knowledge of exact pathlength in the well is unimportant because a single standard calibration will produce reliable and accurate results for all samples. Visit our Website at U.S.A.,The Perkin-Elmer Corporation, 761 Main Avenue, Norwalk, CT Tel: (800) or (203) Fax: (203) Perkin-Elmer is a registered trademark of The Perkin-Elmer Corporation. D-5647 August 98 Printed in U.S.A The Perkin-Elmer Corporation Look to us. And see more.

Low-volume, High Throughput Workflow for Analysis of Nucleic Acid Samples for Biobanking

Low-volume, High Throughput Workflow for Analysis of Nucleic Acid Samples for Biobanking A p p l i c a t i o n N o t e Low-volume, High Throughput Workflow for Analysis of Nucleic Acid Samples for Biobanking Peter J. Brescia, Jr., Chris Wilson, and Peter Banks, BioTek Instruments, Inc., Winooski,

More information

BioDrop DUO dsdna Application Note

BioDrop DUO dsdna Application Note BioDrop DUO dsdna Application Note Using a BioDrop DUO spectrophotometer to measure the concentration of low volume samples of dsdna Micro-volume measurement of DNA is a routine application in many life

More information

Microplate Spectrophotometer

Microplate Spectrophotometer Microplate Spectrophotometer Features Benefits Applications Bali, October 1999 Optical System Xenon Flashing Lamp Order Sorting Filters Monochromator Data Light collection Light Transfer Why are these

More information

Optical density measurements automatically corrected to a 1-cm pathlength with PathCheck Technology

Optical density measurements automatically corrected to a 1-cm pathlength with PathCheck Technology APPLICATION NOTE Optical density measurements automatically corrected to a 1-cm pathlength with PathCheck Technology Introduction UV/VIS spectrophotometers and microplate readers differ fundamentally in

More information

PowerWaveX Select and KC4 : A Multifunctional System for Today s Laboratory Environment

PowerWaveX Select and KC4 : A Multifunctional System for Today s Laboratory Environment PowerWaveX Select and KC4 : A Multifunctional System for Today s Laboratory Environment Figure 1. PowerWaveX Select Microplate Spectrophotometer Introduction With today's requirements for high throughput,

More information

L-malic acid measurement in wines using the SpectraMax Plus 384 Microplate Reader

L-malic acid measurement in wines using the SpectraMax Plus 384 Microplate Reader APPLICATION NOTE L-malic acid measurement in wines using the SpectraMax Plus 384 Microplate Reader Introduction Analysis of malic acid, residual sugar, volatile acidity and ammonia is very important for

More information

Protein Quantification Kit (Bradford Assay)

Protein Quantification Kit (Bradford Assay) Protein Quantification Kit (Bradford Assay) Booklet Item NO. KTD3002 Product Name Protein Quantification Kit (Bradford Assay) ATTENTION For laboratory research use only. Not for clinical or diagnostic

More information

Microdetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats

Microdetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats Application Note 24 Microdetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats INTRODUCTION Most spectrophotometric

More information

Measurement of L-Malic Acid in Wines Using the SpectraMax Plus 384 Microplate Reader

Measurement of L-Malic Acid in Wines Using the SpectraMax Plus 384 Microplate Reader now part of MDS Analytical Technologies Measurement of L-Malic Acid in Wines Using the SpectraMax Plus 384 Microplate Reader SpectraMax Application Note #18 By Theresa Heredia and Gillian Sutherland from

More information

A Simple, Sensitive Spectrophotometric Determination of Mosapride in Pharmaceutical Preparations Using Novel Reagent

A Simple, Sensitive Spectrophotometric Determination of Mosapride in Pharmaceutical Preparations Using Novel Reagent ISS: 0973-4945; CODE ECJHAO E- Chemistry http://www.e-journal.net Vol. 1, o. 5, pp 267-271, October 2004 A Simple, Sensitive Spectrophotometric Determination of Mosapride in Pharmaceutical Preparations

More information

Predictor Assay Setup Guide on the BMG LABTECH CLARIOstar Microplate Readers

Predictor Assay Setup Guide on the BMG LABTECH CLARIOstar Microplate Readers Page 1 of 18 Predictor Assay Setup Guide on the BMG LABTECH CLARIOstar Microplate Readers The BMG LABTECH CLARIOstar Microplate Readers were tested for compatibility with Life Technologies' Predictor herg

More information

Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers

Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N O T E Summary The far UV region (generally covering

More information

The ROXI Colorimeter & Fluorimeter. Laboratory Application I. Colorimetric measurements via Beer s Law.

The ROXI Colorimeter & Fluorimeter. Laboratory Application I. Colorimetric measurements via Beer s Law. The ROXI Colorimeter & Fluorimeter. Laboratory Application I. Colorimetric measurements via Beer s Law. Required Supplies & Costs: RGB LED; $1.95 Light Sensors; $3.95 ea 3-way switch; $6.54 3 ohm resistor;

More information

Application Note TABLE OF CONTENTS

Application Note TABLE OF CONTENTS Application Note 25 Making optical density measurements automatically corrected to a 1 cm pathlength in the SPECTRAmax PLUS microplate spectrophotometer (The three P s of Pathcheck : Principle, Procedures

More information

The Plus for your lab!

The Plus for your lab! BioPhotometer plus NEW! analysis system The Plus for your lab! The easy, rapid and reliable BioPhotometer plus analysis system See it clearly! UVette Precision in plastics The patented* 1 UVette is the

More information

Protein assay. Absorbance Fluorescence Emission Colorimetric detection BIO/MDT 325. Absorbance

Protein assay. Absorbance Fluorescence Emission Colorimetric detection BIO/MDT 325. Absorbance Protein assay Absorbance Fluorescence Emission Colorimetric detection BIO/MDT 325 Absorbance Using A280 to Determine Protein Concentration Determination of protein concentration by measuring absorbance

More information

An ultra-sensitive, versatile, and affordable single tube luminometer for Life Science Research.

An ultra-sensitive, versatile, and affordable single tube luminometer for Life Science Research. p e r s o n a l D E T E C T I O N N o w, y o u c a n o p t i m i z e y o u r p e r s o n a l w o r k f l o w. P r o m e g a i n s t r u m e n t s a n d r e a g e n t s i n t e g r a t e e a s i l y. An

More information

NanoDrop One Viewer software NanoDrop One Website. NanoDrop One Website NanoDrop One Viewer software NanoDrop One Website Software System Update Update Update Software, Update Note OK Language Measure

More information

Skill Building Activity 2 Determining the Concentration of a Species using a Vernier Spectrometer

Skill Building Activity 2 Determining the Concentration of a Species using a Vernier Spectrometer Skill Building Activity 2 Determining the Concentration of a Species using a Vernier Spectrometer Purpose To use spectroscopy to prepare a Beer s Law plot of known dilutions of copper(ii) sulfate so that

More information

High-throughput Quantification of DNA for NGS Library Prep with the Zephyr G3 Workstation and the VICTOR Nivo Plate Reader

High-throughput Quantification of DNA for NGS Library Prep with the Zephyr G3 Workstation and the VICTOR Nivo Plate Reader TECHNICAL APPLICATION NOTE High-throughput Quantification of DNA for NGS Library Prep with the Zephyr G3 Workstation and the VICTOR Nivo Plate Reader NGS Automation Image or Color Block Area Next generation

More information

Automation of Routine Microplate Pipetting and Dispensing

Automation of Routine Microplate Pipetting and Dispensing Automation of Routine Microplate Pipetting and Dispensing Quantitative assays of all sorts are common to many applications in basic science and clinical research. Most of these assays require the movement,

More information

MOLEBIO LAB #4: Using a Spectrophotometer

MOLEBIO LAB #4: Using a Spectrophotometer Introduction: Spectrophotometry MOLEBIO LAB #4: Using a Spectrophotometer Many kinds of molecules interact with or absorb specific types of radiant energy in a predictable fashion. For example, when while

More information

Predictor Assay Setup Guide on the Tecan Safire 2 Microplate Reader

Predictor Assay Setup Guide on the Tecan Safire 2 Microplate Reader Page 1 of 16 Predictor Assay Setup Guide on the Tecan Safire 2 Microplate Reader NOTE: The Tecan Safire 2 Microplate Reader was tested for compatibility with Invitrogen's Predictor herg FP Assay (PV5365)

More information

Spectrophotometric Determination of Iron

Spectrophotometric Determination of Iron Spectrophotometric Determination of Iron INTRODUCTION Many investigations of chemical species involve the interaction between light and matter. One class of these investigations, called absorbance spectrophotometry,

More information

GE Healthcare Life Sciences. Spectrophotometry. Handbook

GE Healthcare Life Sciences. Spectrophotometry. Handbook GE Healthcare Life Sciences Spectrophotometry Handbook Contents Spectrophotometry basics 3 What is spectrophotometry? 3 Definition 3 Lambert s Law 4 Beer s Law 4 Nucleic acid applications 6 Direct UV measurement

More information

Ammonia Assay Kit Modified Berthelot (Colorimetric)

Ammonia Assay Kit Modified Berthelot (Colorimetric) ab102509 Ammonia Assay Kit Modified Berthelot (Colorimetric) Instructions for Use For the rapid, sensitive and accurate measurement of Ammonia in various samples. This product is for research use only

More information

HOOK -Psoralen-PEO-Biotin

HOOK -Psoralen-PEO-Biotin 406PR-01 G-Biosciences 1-800-628-7730 1-314-991-6034 technical@gbiosciences.com A Geno Technology, Inc. (USA) brand name HOOK -Psoralen-PEO-Biotin A photoreactive reagent for biotinylation of nucleic acids

More information

Evaluating Labsphere s new UV-2000

Evaluating Labsphere s new UV-2000 Ciba Grenzach GmbH Optical Spectroscopy Sabrina Siegle Dr. Wolfgang Schlenker Tel. +49 76 24 / 12-28 13 E-mail wolfgang.schlenker@ciba.com May 26, 9 Evaluating Labsphere s new UV- Abstract Labsphere s

More information

Compact Knowledge: Absorbance Spectrophotometry. Flexible. Reliable. Personal.

Compact Knowledge: Absorbance Spectrophotometry. Flexible. Reliable. Personal. L A B O R A T O R Y C O M P E T E N C E Compact Knowledge: Absorbance Spectrophotometry Flexible. Reliable. Personal. The interaction of light with molecules is an essential and well accepted technique

More information

Fluorescence Spectrophotometry

Fluorescence Spectrophotometry Chemistry 422L Manual Page 27 I. Introduction Fluorescence Spectrophotometry Ru(bpy) 3 2+, where bpy = 2, 2' bipyridine, has been one of the most widely studied metal complexes in recent years. Interest

More information

Caspase Substrate Kit (Colorimetric)

Caspase Substrate Kit (Colorimetric) ab102486 Caspase Substrate Kit (Colorimetric) Instructions for Use For the rapid, sensitive and accurate measurement of Caspase activity in cell culture This product is for research use only and is not

More information

QuickZyme Total Protein Assay (to be used with acid hydrolyzates)

QuickZyme Total Protein Assay (to be used with acid hydrolyzates) QuickZyme Total Protein Assay (to be used with acid hydrolyzates) August 2014 This package insert must be read in its entirety before using this product FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC

More information

ab83360 Ammonia Assay Kit

ab83360 Ammonia Assay Kit Version 8 Last updated 18 December 2017 ab83360 Ammonia Assay Kit For the measurement of total ammonia and ammonium levels in various samples This product is for research use only and is not intended for

More information

Bradford Reagent, 5x

Bradford Reagent, 5x INSTRUCTION MANUAL Bradford Reagent, 5x Reagent for protein quantification (Cat. No. 39222) SERVA Electrophoresis GmbH Carl-Benz-Str. 7 D-69115 HeidelbergPhone +49-6221-138400, Fax +49-6221-1384010 e-mail:

More information

INTERNATIONAL CHEMISTRY TESTING. 258 Main Street- Suite 311- Milford, MA Tel: Fax:

INTERNATIONAL CHEMISTRY TESTING. 258 Main Street- Suite 311- Milford, MA Tel: Fax: SOP TITLE: DETERMINATION OF TOTAL PROANTHOCYANIDINS BY REACTION WITH DIMETHYLAMINOCINNAMALDEHYDE (DMAC) SOP NUMBER: 803-02 EFFECTIVE DATE: 03/01/14 SUPERSEDES: 803-01 NEW ISSUE DATE: 3/24/20 INTRODUCTION

More information

Alcohol dehydrogenase Assay Kit

Alcohol dehydrogenase Assay Kit Alcohol dehydrogenase Assay Kit Catalog Number KA3785 100 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 General

More information

Spectroscopy. Page 1 of 8 L.Pillay (2012)

Spectroscopy. Page 1 of 8 L.Pillay (2012) Spectroscopy Electromagnetic radiation is widely used in analytical chemistry. The identification and quantification of samples using electromagnetic radiation (light) is called spectroscopy. Light has

More information

Spectrophotometer for Life Science. BioSpec-nano C101-E112D

Spectrophotometer for Life Science. BioSpec-nano C101-E112D Spectrophotometer for Life Science BioSpec-nano C11-E112D BioSpec-nano Spectrophotometer for Life Science Drop-and-Start Analysis Automatic Optical Pathlength Setting & Automatic Wiping Low Carryover Achieved

More information

Beer's- Lambert Law and Standard Curves. BCH 312 [Practical]

Beer's- Lambert Law and Standard Curves. BCH 312 [Practical] Beer's- Lambert Law and Standard Curves BCH 312 [Practical] Spectrophotometer: Spectrophotometer can be used to measure the amount of light absorbed or transmitted by a solution. It consist of two parts:

More information

Malachite Green Phosphate Detection Kit Catalog Number: DY996

Malachite Green Phosphate Detection Kit Catalog Number: DY996 Malachite Green Phosphate Detection Kit Catalog Number: DY996 This Malachite Green Phosphate Detection Kit employs a simple, sensitive, reproducible, and non-radioactive method for measuring inorganic

More information

Caspase Substrate Assay Kit (Colorimetric)

Caspase Substrate Assay Kit (Colorimetric) Caspase Substrate Assay Kit (Colorimetric) Catalog Number KA3694 7 x 25 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information...

More information

CHAPTER - 3 ANALYTICAL PROFILE. 3.1 Estimation of Drug in Pharmaceutical Formulation Estimation of Drugs

CHAPTER - 3 ANALYTICAL PROFILE. 3.1 Estimation of Drug in Pharmaceutical Formulation Estimation of Drugs CHAPTER - 3 ANALYTICAL PROFILE 3.1 Estimation of Drug in Pharmaceutical Formulation 3.1.1 Estimation of Drugs ANALYTICAL PROFILE 84 3.1 ESTIMATION OF DRUG IN PHARMACEUTICAL FORMULATION. Agrawal A et al

More information

Z -LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar Reader

Z -LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar Reader 08 Jan 15 Page 1 of 18 Z -LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar Reader The BMG LABTECH CLARIOstar Microplate Readers were tested for compatibility with the Life Technologies Z -LYTE Assay

More information

ab Maleimide Quantification Kit - Colorimetric

ab Maleimide Quantification Kit - Colorimetric ab112140 Maleimide Quantification Kit - Colorimetric Instructions for Use For detecting maleimide in physiology solutions spectrophotometrically. This product is for research use only and is not intended

More information

ab MDA Assay Kit (competitive ELISA)

ab MDA Assay Kit (competitive ELISA) Version 1 Last updated 30 August 2018 ab238537 MDA Assay Kit (competitive ELISA) For the quantitative measurement of MDA in protein samples such as purified protein, plasma, serum and cell lysate. This

More information

Chemistry 1215 Experiment #11 Spectrophotometric Analysis of an Unknown Brass Sample

Chemistry 1215 Experiment #11 Spectrophotometric Analysis of an Unknown Brass Sample Chemistry 1215 Experiment #11 Spectrophotometric Analysis of an Unknown Brass Sample Objective In this experiment you will use spectrophotometric measurements to determine the copper concentration of a

More information

Instrumental Technique: Cuvette. Md Rabiul Islam

Instrumental Technique: Cuvette. Md Rabiul Islam Instrumental Technique: Cuvette Md Rabiul Islam 16-7-2016 What is cuvette? A cuvette is a small tube of circular or square cross section, sealed at one end, made of plastic, glass, or fused quartz (for

More information

Ammonia Assay Kit. Catalog Number KA assays Version: 02. Intended for research use only.

Ammonia Assay Kit. Catalog Number KA assays Version: 02. Intended for research use only. Ammonia Assay Kit Catalog Number KA3707 100 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information... 4 Materials Supplied...

More information

Total Protein Assay Kit

Total Protein Assay Kit Total Protein Assay Kit Catalog Number KA3946 1000 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 General Information...

More information

Cobalt Quantification Kit

Cobalt Quantification Kit ab102510 Cobalt Quantification Kit Instructions for Use For the rapid, sensitive and accurate measurement of Cobalt levels in various samples. This product is for research use only and is not intended

More information

Bio 120 Lab 5: Quantitative Analysis

Bio 120 Lab 5: Quantitative Analysis Bio 120 Lab 5: Quantitative Analysis Remember from the measurement lab, concentration is the amount of a solute, also called an analyte, in a given amount of solution. We need a way to measure the concentration

More information

Spectrophotometric quantification of nucleic acids using the nanovette microliter cell accessory with a DU 730 spectrophotometer

Spectrophotometric quantification of nucleic acids using the nanovette microliter cell accessory with a DU 730 spectrophotometer Application Information Spectrophotometric quantification of nucleic acids using the nanovette microliter cell accessory with a DU 730 spectrophotometer Amy Yoder, Mary Blair, Kenny Abell Beckman Coulter,

More information

Instructor s Advance Preparation

Instructor s Advance Preparation INSTRUCTOR'S MANUAL Instructor s Advance Preparation This protocol is designed for 80 workstations of 4 students. Each group will prepare a set of standards, a blank, and 2 milk samples (can be a blind

More information

Non-Interfering Protein Assay Kit Cat. No

Non-Interfering Protein Assay Kit Cat. No Visit our interactive pathways at /pathways User Protocol 488250 Rev. 5 January 2006 RFH Page 1 of 1 Non-Interfering Protein Assay Kit Cat. No. 488250 Note that this user protocol is not lot-specific and

More information

Experiment. Quantification of Ascorbic acid by Fluorescence Spectroscopy1

Experiment. Quantification of Ascorbic acid by Fluorescence Spectroscopy1 Experiment. Quantification of Ascorbic acid by Fluorescence Spectroscopy Modified 10/2017 Experiment. Quantification of Ascorbic acid by Fluorescence Spectroscopy1 Objective: The goal of this experiment

More information

A Spectrophotometric Analysis of Calcium in Cereal

A Spectrophotometric Analysis of Calcium in Cereal CHEM 311L Quantitative Analysis Laboratory Revision 1.2 A Spectrophotometric Analysis of Calcium in Cereal In this laboratory exercise, we will determine the amount of Calium in a serving of cereal. We

More information

Concepts, Techniques. Concepts, Techniques 9/11/2012. & Beer s Law. For a simple, transparent, COLORED material, e.g. ROYGBV

Concepts, Techniques. Concepts, Techniques 9/11/2012. & Beer s Law. For a simple, transparent, COLORED material, e.g. ROYGBV 9//22 OBJECTIVES Spectrophotometry of Food Dyes & Beer s Law Last Update: 9//22 9:54 AM What is the quantitative basis for the color of substances? How is the absorption/transmission of light measured?

More information

For the rapid, sensitive and accurate measurement of Urea levels in various samples.

For the rapid, sensitive and accurate measurement of Urea levels in various samples. ab83362 Urea Assay Kit Instructions for Use For the rapid, sensitive and accurate measurement of Urea levels in various samples. This product is for research use only and is not intended for diagnostic

More information

Module 3. Establishing Standard Curve (Bradford Assay)

Module 3. Establishing Standard Curve (Bradford Assay) Module 3. Establishing Standard Curve (Bradford Assay) 1. Melissa Myoglobin and the Missing Molecules Melissa Myoglobin had a little tube filled with a clear protein solution. To do more research with

More information

Clinical Chemistry (CHE221) Professor Hicks Week 1. Statistics Made Slightly Less Boring and Introduction to Spectrophotometry. Accuracy vs Precision

Clinical Chemistry (CHE221) Professor Hicks Week 1. Statistics Made Slightly Less Boring and Introduction to Spectrophotometry. Accuracy vs Precision Clinical Chemistry (CHE221) Professor Hicks Week 1 Statistics Made Slightly Less Boring and Introduction to Spectrophotometry 3 Accuracy vs Precision Precision is the consistency of a measurement made

More information

Lecture 5. More on UV-visible Spectrophotometry: Beer s Law and Measuring Protein Concentration

Lecture 5. More on UV-visible Spectrophotometry: Beer s Law and Measuring Protein Concentration Biological Chemistry Laboratory Biology 3515/Chemistry 3515 Spring 2018 Lecture 5 More on UV-visible Spectrophotometry: Beer s Law and Measuring Protein Concentration 23 January 2018 c David P. Goldenberg

More information

For the rapid, sensitive and accurate measurement of Caspase 9 activity in cell and tissue lysates.

For the rapid, sensitive and accurate measurement of Caspase 9 activity in cell and tissue lysates. ab65608 Caspase 9 Assay Kit (Colorimetric) Instructions for Use For the rapid, sensitive and accurate measurement of Caspase 9 activity in cell and tissue lysates. This product is for research use only

More information

Using the FL600 Fluorescence Microplate Reader for Chemiluminescence Detection What s a lumin anyway?

Using the FL600 Fluorescence Microplate Reader for Chemiluminescence Detection What s a lumin anyway? Using the FL600 Fluorescence Microplate Reader for Chemiluminescence Detection What s a lumin anyway? Chemiluminescence is a highly sensitive technique that has been employed in a wide variety of applications

More information

Human Paraneoplastic pemphigus (PNP)antibody ELISA Kit. Catalog No. MBS (96 T)

Human Paraneoplastic pemphigus (PNP)antibody ELISA Kit. Catalog No. MBS (96 T) Human Paraneoplastic pemphigus (PNP)antibody ELISA Kit Catalog No. MBS704248 (96 T) This immunoassay kit allows for the in vitro quantitative determination of human Paraneoplastic pemphigus(pnp)antibody

More information

Human Mullerian Inhibiting Substance/Anti-Mullerian hormone (MIS/AMH)Elisa Kit

Human Mullerian Inhibiting Substance/Anti-Mullerian hormone (MIS/AMH)Elisa Kit Human Mullerian Inhibiting Substance/Anti-Mullerian hormone (MIS/AMH)Elisa Kit Catalog No. CSB-E12756h (96T) This immunoassay kit allows for the in vitro quantitative determination of human MIS/AMH concentrations

More information

Human von Willebrand Factor cleavingprotease(vwf-cp) ELISA Kit. Catalog No. MBS (96T)

Human von Willebrand Factor cleavingprotease(vwf-cp) ELISA Kit. Catalog No. MBS (96T) Human von Willebrand Factor cleavingprotease(vwf-cp) ELISA Kit Catalog No. MBS705447 (96T) This immunoassay kit allows for the in vitro quantitative determination of human vwf-cp concentrations in serum,

More information

1. Preliminary qualitative analysis of unknown substances (liquid or solid).

1. Preliminary qualitative analysis of unknown substances (liquid or solid). Name of Procedure: Ultraviolet Spectroscopy Suggested Uses: 1. Preliminary qualitative analysis of unknown substances (liquid or solid). 2. Quantitative analysis of known compounds. Apparatus Used to Perform

More information

BIOO FOOD AND FEED SAFETY. Histamine Enzymatic Assay Kit Manual. Catalog #: Reference #:

BIOO FOOD AND FEED SAFETY. Histamine Enzymatic Assay Kit Manual. Catalog #: Reference #: BIOO FOOD AND FEED SAFETY Histamine Enzymatic Assay Kit Manual Catalog #: 1032-05 Reference #: 1032-05 BIOO Scientific Corp. 2010 TABLE OF CONTENTS GENERAL INFORMATION... 1 Product Description... 1 Procedure

More information

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy A. Evaluation of Analytical Parameters in Atomic Absorption Spectroscopy Objective The single feature that contributes most to making atomic

More information

International Journal of Pharma and Bio Sciences

International Journal of Pharma and Bio Sciences International Journal of Pharma and Bio Sciences RESEARCH ARTICLE PHARMACEUTICAL ANALYSIS DEVELOPMENT, ESTIMATION AND VALIDATION OF BOSENTAN IN BULK AND IN ITS PHARMACEUTICAL FORMULATION BY UV-VIS SPECTROSCOPIC

More information

Chem 310 rd. 3 Homework Set Answers

Chem 310 rd. 3 Homework Set Answers -1- Chem 310 rd 3 Homework Set Answers 1. A double line labeled S 0 represents the _ground electronic_ state and the _ground vibrational_ state of a molecule in an excitation state diagram. Light absorption

More information

EXPERIMENT 12 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of a Commercial Aspirin Tablet

EXPERIMENT 12 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of a Commercial Aspirin Tablet EXPERIMENT 12 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of a Commercial Aspirin Tablet utcomes After completing this experiment, the student should be able to: 1. Prepare standard

More information

INSTRUCTION MANUAL. Lowry Assay Kit. Kit for Protein Quantification. (Cat. No )

INSTRUCTION MANUAL. Lowry Assay Kit. Kit for Protein Quantification. (Cat. No ) INSTRUCTION MANUAL Lowry Assay Kit Kit for Protein Quantification (Cat. No. 39236) SERVA Electrophoresis GmbH Carl-Benz-Str. 7 D-69115 Heidelberg Phone +49-6221-138400, Fax +49-6221-1384010 e-mail: info@serva.de

More information

RayBio Rat TNF-alpha ELISA Kit (For Lysates)

RayBio Rat TNF-alpha ELISA Kit (For Lysates) RayBio Rat TNF-alpha ELISA Kit (For Lysates) Catalog #: ELR-TNFa-CL User Manual Last revised April 15, 2016 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite

More information

Investigating Food Dyes in Sports Beverages. Sample

Investigating Food Dyes in Sports Beverages. Sample Investigating Food Dyes in Sports Beverages Investigation 1 There are many different brands of beverages that fall under the general category of sports drinks. Most of these beverages contain an FD&C food

More information

FLUORESCENCE APPLICATIONS

FLUORESCENCE APPLICATIONS FLUORESCENCE APPLICATIONS CORRECTION OF EMISSION SPECTRA USING THE PERKINELMER MODEL LS-50 LUMINESCENCE SPECTROMETER Spectra collected on luminescence spectrometers consist of true spectral information

More information

Mouse Cholecystokinin (CCK) ELISA Kit

Mouse Cholecystokinin (CCK) ELISA Kit Mouse Cholecystokinin (CCK) ELISA Kit Catalog No. CSB-E08115m (96T) This immunoassay kit allows for the in vitro quantitative determination of mouse CCK concentrations in serum, plasma and other biological

More information

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry 13A Measurement Of Transmittance and Absorbance Absorption measurements based upon ultraviolet and visible radiation

More information

Laboratory Measurements and Procedures

Laboratory Measurements and Procedures 18 Introduction Measurements of masses, volumes, and preparation of chemical solutions of known composition are essential laboratory skills. The goal of this exercise is to gain familiarity with these

More information

Validating Slope Spectroscopy Methods: A Formula for Robust Measurements

Validating Slope Spectroscopy Methods: A Formula for Robust Measurements CWP0802261A White Paper Validating Slope Spectroscopy Methods: A Formula for Robust Measurements February 26, 2008 I-Tsung Shih, PhD Mark Salerno 1.0 Abstract Demands on measurement systems are ever increasing.

More information

Developers Guide. Development of new instrument applications for D-dimer

Developers Guide. Development of new instrument applications for D-dimer Developers Guide Development of new instrument applications for D-dimer medirox d dimer application guide.doc MediRox AB www.medirox.se Phone: +46 (0)155 45 44 10 SE-611 82 Nyköping info@medirox.se Fax:

More information

CHEMICAL KINETICS E + 2B 2C + D (1)

CHEMICAL KINETICS E + 2B 2C + D (1) CHEMICAL KINETICS Chemical kinetics is the branch of chemistry that is concerned with the study of the rates and mechanisms of chemical reactions. The rate of a reaction is a measure of its speed. Consider

More information

C101-E112. BioSpec-nano. Shimadzu Spectrophotometer for Life Science

C101-E112. BioSpec-nano. Shimadzu Spectrophotometer for Life Science C101-E112 BioSpec-nano Shimadzu Spectrophotometer for Life Science Power of small. BioSpec-nano BioSpec-nano Shimadzu Spectrophotometer for Life Science Quick and Simple Nucleic Acid Quantitation Drop-and-Click

More information

Lab Investigation 4 - How could you make more of this dye?

Lab Investigation 4 - How could you make more of this dye? Lab Investigation 4 - How could you make more of this dye? USING SPECTROSCOPY TO DETERMINE SOLUTION CON- CENTRATION Guiding Question How could you make more of this dye? INTRODUCTION A solution is a homogeneous

More information

Optical Sensing for a Changing Planet. Smart Cuvettes!

Optical Sensing for a Changing Planet. Smart Cuvettes! Optical Sensing for a Changing Planet Smart Cuvettes! phuvettes for optical ph measurements foxyvette for optical po 2 measurements phoxyvettes for both ph and po 2 727.230.1697 727.230.6845 fax info@spectrecology.com

More information

ab83360 Ammonia Assay Kit

ab83360 Ammonia Assay Kit Version 9 Last updated 7 February 2019 ab83360 Ammonia Assay Kit For the measurement of total ammonia and ammonium levels in various samples View kit datasheet: www.abcam.com/ab83360 (use www.abcam.cn/ab83360

More information

Spectroscopy II Introduction: Spectrophotometry and Fluorometry

Spectroscopy II Introduction: Spectrophotometry and Fluorometry Introduction: Spectrophotometry and Fluorometry In this experiment you will use the SpectroVis Plus (a small computer controlled Spectrophotometer + Fluorometer) to: Measure the Transmission and Absorbance

More information

Alcohol Dehydrogenase Detection Kit

Alcohol Dehydrogenase Detection Kit ab102533 Alcohol Dehydrogenase Detection Kit Instructions for Use For the rapid, sensitive and accurate measurement of Alcohol Dehydrogenase activity in various samples This product is for research use

More information

Sheep myeloperoxidase(mpo) ELISA kit

Sheep myeloperoxidase(mpo) ELISA kit Sheep myeloperoxidase(mpo) ELISA kit Catalog No. CSB-EL014757SH (96T) This immunoassay kit allows for the in vitro quantitative determination of sheep MPO concentrations in serum and plasma. Expiration

More information

ADH Activity Assay Kit

ADH Activity Assay Kit ADH Activity Assay Kit Catalog Number KA3713 100 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information... 4 Materials

More information

Total NADP and NADPH Assay Kit (Colorimetric)

Total NADP and NADPH Assay Kit (Colorimetric) ab186033 Total NADP and NADPH Assay Kit (Colorimetric) Instructions for Use An optimized assay for monitoring Total NADP and NADPH. This product is for research use only and is not intended for diagnostic

More information

NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation.

NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation. NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation. SCANNING UV VISIBLE SPECTROPHOTOMETER Features SCANNING UV VISIBLE SPECTROPHOTOMETER

More information

ab Alkaline Phosphatase Assay Kit (Luminometric)

ab Alkaline Phosphatase Assay Kit (Luminometric) Version 1 Last updated 24 April 2018 ab233466 Alkaline Phosphatase Assay Kit (Luminometric) For the measurement of Alkaline Phosphatase activity in solutions and cells. This product is for research use

More information

Rat Advanced Glycation End Products(AGEs) ELISA Kit

Rat Advanced Glycation End Products(AGEs) ELISA Kit Rat Advanced Glycation End Products(AGEs) ELISA Kit Catalog No. CSB-E09413r (96 T) This immunoassay kit allows for the in vitro quantitative determination of rat AGEs concentrations in serum, plasma and

More information

Determining the Concentration of a Solution: Beer s Law

Determining the Concentration of a Solution: Beer s Law Determining the Concentration of a Solution: Beer s Law The primary objective of this experiment is to determine the concentration of an unknown cobalt (II) chloride solution. You will use a Vernier SpectroVis

More information

Acetylcholinesterase Assay Kit (Fluorometric - Red)

Acetylcholinesterase Assay Kit (Fluorometric - Red) ab138873 Acetylcholinesterase Assay Kit (Fluorometric - Red) Instructions for Use For the detection of Acetylcholinesterase activity in blood, cell extracts, and in other solutions. This product is for

More information

Porcine Immunoglobulin E (IgE)ELISA Kit

Porcine Immunoglobulin E (IgE)ELISA Kit Porcine Immunoglobulin E (IgE)ELISA Kit Catalog No. MBS703293 (96T) This immunoassay kit allows for the in vitro quantitative determination of porcine IgE concentrations in serum and plasma,. Expiration

More information

Aldehyde Quantification Assay Kit (Colorimetric)

Aldehyde Quantification Assay Kit (Colorimetric) ab112113 Aldehyde Quantification Assay Kit (Colorimetric) Instructions for Use For detecting quantifying aldehyde in a variety of applications using a one step colorimetric method This product is for research

More information

often display a deep green color due to where the SPR occurs (i.e., the wavelength of light that interacts with this specific morphology).

often display a deep green color due to where the SPR occurs (i.e., the wavelength of light that interacts with this specific morphology). Synthesis-Dependent Catalytic Properties of Gold Nanoparticles Nanoscience is the study of materials that have dimensions, intuitively, on the nanoscale, typically between 1 100 nm. This field has received

More information

Introduction to Spectroscopy: Analysis of Copper Ore

Introduction to Spectroscopy: Analysis of Copper Ore Absorbance Introduction to Spectroscopy: Analysis of Copper Ore Introduction The goal of this lab is to determine the unknown concentration of two different copper solution samples, taken from fictitious

More information