Comparison of Different Methods of Environmental Radioactivity Measurements at the Zagreb Radiocarbon and Tritium Laboratory

Size: px
Start display at page:

Download "Comparison of Different Methods of Environmental Radioactivity Measurements at the Zagreb Radiocarbon and Tritium Laboratory"

Transcription

1 Comparison of Different Methods of Environmental Radioactivity Measurements at the Zagreb Radiocarbon and Tritium Laboratory B. Obelić, I. Krajcar Bronić, N. Horvatinčić, J. Barešić Rudjer Bošković Institute, P.O.Box 180, Zagreb, CROATIA Abstract. Three methods for low-level radiocarbon measurement and two methods for measurement of lowlevel tritium activity in environmental samples have been developed in the Zagreb Radiocarbon and Tritium Laboratory. Various sample preparation methods are used for 14 C (CH 4 synthesis, benzene synthesis, CO 2 absorption) and 3 H (CH 4 synthesis, water distillation). A gas proportional counter and a liquid scintillation counter are used for measurement. Basic parameters (sample quantity, background count rate, efficiency, figure of merit, limit of detection) of the methods are compared, as well as 14 C and 3 H activities determined by different techniques. 1. Introduction Radiocarbon and Tritium Laboratory of the Rudjer Bošković Institute in Zagreb developed gas proportional counter (GPC) techniques for radiocarbon and tritium low-activity measurements in 1968 and 1975, respectively. Since then, more than 3400 various radiocarbon samples (archaeological, geological, hydrogeological and environmental) have been measured. Most results have been regularly published as data lists in Radiocarbon [1]. Tritium activity has been measured in various natural waters: precipitation, groundwater, springs and rivers, and atmospheric water vapor [2]. Laboratory participated in different radiocarbon and tritium international intercomparison studies and proficiency tests as a part of the quality assurance and quality control system [3-5]. Liquid scintillation counter (LSC) Quantulus 1220 has been used since 2001 for low activity measurement of both isotopes. The LSC running and data acquisition was performed by using Wallac WinQ Windows Software for contolling Wallac 1220 Quantulus TM (Version 1.2) and for data processing we use Wallac EASY View Spectrum Analysis Program (Version 1.0). For radiocarbon measurement by using LSC two different sample preparation techniques have been developed: absorption of CO 2 (LSC-A) and benzene synthesis (LSC-B) methods. Here we present a comparison of basic parameters controlling both techniques of 3 H measurement (GPC and LSC) and all three techniques for 14 C measurement (GPC, LSC-A, and LSC-B), including the comparison of measured activities. 2. Tritium measurement For GPC tritium measurement, CH 4 is obtained by reaction of water with aluminium carbide at 150 C [6], purified and used as a counting gas in a multi-wire GPC. The counting energy window is set to energies between 1 kev and 10 kev to obtain the best figure of merit (FM). Gas quality control has been performed by simultaneous monitoring of the count rate above the tritium channel, i.e., above 20 kev [7]. The lowest tritium activity that can be distinguished from the background, i.e., the limit of detection (LOD) is 0.3 Bq/L. For tritium measurement by LSC, a certain amount of distilled water is mixed with HiSafe III scintillation cocktail in a 20-mL plastic vial. The scintillation mixture containing 8 ml of sample water and 12 ml of the HiSafe III cocktail resulted in better efficiency than other mixture 1

2 compositions. The tritium window, as measured by the LSC Quantulus, comprises channels from 18 to 220. However, the optimal counting parameters were determined for the counting window comprising channels from 25 to 187. In the counting window, 92% of the total tritium spectrum is recorded, while only 82% of the background is counted, and the corresponding FM is better than in the whole tritium window. The LOD for the LCS technique is 0.6 Bq/L. The basic parameters of both GPC and LSC techniques for tritium measurement are compared in Table I. The GPC technique requires larger amount of sample and shows slightly higher background count rate, but the count rate of active sample is higher which gives also better FM value and lower detection limit. Results of measurement of tritium activities by GPC and LSC in some environmental samples (precipitation, groundwater) are compared in Fig. 1. Good agreement is obtained for low tritium activities usually measured in the environment: the slope of the fitted line is (1.01 ± 0.01), the intercept is (-0.08 ± 0.15) Bq/L, and the regression coefficient 99.9%. Table I. Comparison of basic parameters of the GPC and LSC techniques for tritium activity measurement. Data for counting time of app. 1 day; cpm = counts per minute. GPC Sample quantity (ml) 50 8 LSC Background count rate, B (cpm) Active (93 Bq/L) sample net count rate, A m (cpm) Efficiency 23.6% FM = A m 2 /B LOD (Bq/L) LSC, A (Bq/L) GPC, A (Bq/L) FIG. 1. Comparison of tritium activity concentrations in environmental samples measured by two methods, GPC and LSC. Symbols: measured activities, line: linear fit. 2

3 3. Radiocarbon measurement Complete procedures for radiocarbon measurement by all three methods are shown in Fig. 2. The first step in radiocarbon measurement is preparation of CO 2 by combustion of an organic sample or by dissolving a carbonate sample in hydrochloric acid. For GPC measurement, CO 2 is then converted to CH 4 by catalytic reaction (Ru as a catalyst) with H 2 at 470 C. Purified CH 4 is stored for 2 weeks to allow radon to decay [8], and used as a counting gas [6, 9]. For LSC measurement, CO 2 can be either absorbed in an absorption mixture (LSC-A method) or used for benzene synthesis (LSC-B method). ORGANIC SAMPLE INORGANIC SAMPLE Pretreatment (A-B-A) Combustion Dissolution HCl CO 2 H 2 Li Catalytic reaction (Ru) Carbosorb + Permafluor Carbidization Absorption Li 2 C 2 CH 4 Carbamate Hydrolisis H 2 O C 2 H 2 GPC LSC-A Catalytic trimerization C 6 H 6 Butyl PBD LSC-B FIG. 2. Three procedures for radiocarbon sample preparation and measurement in the Zagreb Radiocarbon and Tritium Laboratory. 3

4 Absorption of CO 2 (LSC-A) is performed in a vacuum line directly to a 20-mL low-potassium glass vial filled with the mixture of Carbosorb E (the absorbing medium) and Permafluor E (the scintillator) [10]. The amount of CO 2 absorbed in the absorption mixture is determined by weighting the glass vial with the mixture before and after the absorption. The optimal conditions of sample preparation were determined after tests with different CO 2 flow rates and various compositions of the absorption mixture. A flow rate of about 70 ml CO 2 per minute was chosen for routine sample preparation. Under such a flow rate it takes about 25 min to obtain the absorption mixture saturated with CO 2. The best counting efficiency (E) was obtained for the mixture containing 10 ml of Carbosorb E and 10 ml of Permafluor E. The sample preparation for LSC-B method involves the reaction of CO 2 with lithium at 700ºC - 900ºC, subsequent hydrolysis of lithium carbide to acetylene, and its catalytic trimerization on the vanadium catalyst to benzene [11]. The results of tests showed that the optimal time of carbide and hydrolysis reaction is 20 min and min, respectively. Temperature of catalytic reaction is between 60 C and 90 C with duration of 1-2 hours, depending on the quantity of acetylene. The acetylene yields range from 89% to 98%, and benzene yields range from 77% to 90%, while the benzene purity is 98.9 to 99.5%. To prepare the scintillation cocktail for LSC measurements, we add 15 mg of scintillator (butyl-pbd) per 1 g of benzene and measure 14 C activity of benzene in 7-mL low-potassium glass vials. For samples prepared by both LSC-A and LSC-B methods the optimal counting parameters for measurement in LSC were determined, as well as the SQP value as a test of sample quality. (The SQP is an index of quenching, Standard Quench Parameter, which represents the end point of the external standard spectrum, i.e., the channel number beyond which 1% of the total counts are found. As external standard we use a built-in 37 kbq 152 Eu capsule.) The number of 30-minute cycles in a run was taken to be 30 (or more), resulting in 900 (or more) minutes per sample measurement. The counting efficiency (E) was determined from the ratio of measured net count rates (A m, expressed in cpm) and the known 14 C activity of the standard (A, expressed in dpm). The counting efficiencies in the so-called " 14 C windows", encompassing the channels and for LSC-A and LSC- B, respectively, are 70% and 90%, respectively. The counting window for the LSC-A technique after optimization comprises channels between 144 and 372 and it encompasses 93% of the total 14 C spectrum and 77% of the background spectrum. The total efficiency for the LSC-A is therefore 65%. The FM value has been improved from 16 to 18.4 after optimization. The dating limit of the method is 31,800 yr. Optimization of the counting window for the LSC-B technique resulted in background reduction to 64% of that in the 14 C window, while 92% of the 14 C spectrum has remained in the window. Therefore, the total efficiency of the LSC-B technique is 82.4% and the FM value is considerably improved (from 2200 to 2908). Under adopted counting condition, the 14 C dating limit is 52,160 yr (a 5-mL benzene sample counted 1 day). The limit can be improved to 54,800 yr by using larger amount of benzene, e.g., 7 ml, if the quantity of available sample allows it. Longer counting time (e.g., 2 days) will also increase the maximal age that can be determined to 57,600 yr. Comparison of some basic parameters of the measurement of 14 C in the counting windows is shown in Table II, where also data for our GPC system are shown for comparison. The comparison of conventional 14 C ages [12] of benzene samples prepared as described above (LSC- B method) and those of the same samples measured by the GPC is shown in Fig. 3. The LSC-B ages were corrected also for different quenching, by using the quenching curve. (Small amount of acetone was added to the spike benzene, and the SQP and the count rate of such samples have been measured. For our synthesized benzene samples of purity 98.9%-99.5% the efficiency correction due to quenching was less than 1%.) The agreement between the LSC-B and GPC ages is good, giving the slope of the fitted line equal to 1.01 ± No systematic difference between the two sets is observed, as shown also by the intercept of the fitted line equal to (39 ± 41) yr. 4

5 Table II. Comparison of characteristic parameters of all three radiocarbon measurement techniques for the counting time of app. 1 day. LSC-A: absorption of CO 2 and LSC measurement; LSC-B: benzene synthesis and LSC measurement; GPC: methane preparation and gas proportional counting. The data for LSC-A and LSC-B techniques correspond to the optimized counting windows. Radiocarbon activity of the standard is 100 pmc = dpm/g of carbon. LSC-A LSC-B GPC Amount of carbon (g) Spectrum area (counting window, channel from to) Real activity of standard, A (dpm) Net count rate of standard, A m (cpm) Count rate of background, B (cpm) Efficiency, E = A m /A (%) Figure of merit, FM = A 2 m /B Maximum 14 C age T max * Tmax = 8033 ln A meas t meas B 31,800 52,160 37,500 The comparison of 14 C activities expressed in pmc (percent of modern carbon, defined as 100 pmc = dpm/g of carbon = Bq/g of carbon) obtained by the GPC and LSC-A methods is presented in Fig. 4. The agreement between the two sets of data is again good (the slope of the fitted line is 1), but the uncertainties of the LSC-A results are 2-3 times larger that those of the GPC method for the same counting time (1 day per sample). Due to high background count rate and low efficiency (Table II), the maximal determinable 14 C age for LSC-A is 31,800 yr. However, these characteristics are good enough for certain applications that do not require high precision (e.g., environmental monitoring), and also for samples having high 14 C activity LSC-B, years (BP) GPC, years (BP) FIG. 3. Comparison of the conventional 14 C ages of various samples measured by LSC-B and GPC methods. 5

6 100 LSC-A, a 14 C (pmc) GPC, a 14 C (pmc) FIG. 4. Comparison of 14 C activities measured by the LSC-A and GPC methods. 4. Conclusion The presented methods of measurements of environmental isotopes 3 H and 14 C including sample preparation differ in complexity, time consume, price, and in precision of measured results. GPC technique for tritium measurement is more complex than the LSC one, but the counting characteristics are better. The GPC method is therefore preferred in applications requiring higher precision and lower detection limit, e.g., in hydrogeological studies. LSC-A method for radiocarbon measurement is quick, cheep and simple and requires less carbon (~0.5 g) than the other two methods (~4 g). It is accurate enough for certain applications, although it is not as precise as either GPC or LSC-B method. Chemical preparations of samples for GPC and LSC-B methods of 14 C measurement are more complex, but the methods give better precision, and therefore they can be applied for dating in archaeology and geochronology ( 14 C dating). To summarize, LSC measurement techniques of tritium in water and of 14 C in various samples by CO 2 absorption are suitable for quick and accurate determination of environmental contamination (they could be used for quick determination of increased environmental 14 C and 3 H contamination in case of a nuclear accident), although both are not recommendable for applications requiring higher precision. Acknowledgement Work performed under Project of the Ministry of Science, Republic of Croatia, and with the help of IAEA TC Project CRO/2/002 "Nuclear Spectroscopy Techniques and Ion Beam Analysis in Environmental and Industrial Applications ". 6

7 References 1. Obelić, B., Krajcar Bronić, I., Horvatinčić, N., Rudjer Bošković Institute radiocarbon measurements XV. Radiocarbon 44/2: , (2002), and references therein. 2. Krajcar Bronić, I., Horvatinčić, N., Obelić, B., Two decades of environmental isotope record in Croatia: reconstruction of the past and prediction of future levels. Radiocarbon 40/1-2: , (1998). 3. Horvatinčić, N., Srdoč, D., Obelić, B., Krajcar Bronić, I., Radiocarbon dating of intercomparison samples at the Zagreb Radiocarbon Laboratory. Radiocarbon 32: , (1990). 4. Krajcar Bronić, I., Horvatinčić, N., Obelić, B., Bistrović, R., Radiocarbon intercomparison studies at the Rudjer Bošković Institute. Radiocarbon 37: , (1995). 5. Obelić, B., Horvatinčić, N., Krajcar Bronić, I., in Proceedings of the IRPA Regional Congress Dubrovnik, Croatia, 2001, (CD-ROM), edited by B. Obelić, M. Ranogajec-Komor, S. Miljanić, I. Krajcar Bronić (CRPA, Zagreb, 2002), paper no. 1o Horvatinčić, N., Radiocarbon and tritium measurements in water samples and application of isotopic analyses in hydrology. Fizika 12 (S2): , (1980). 7. Krajcar Bronić, I., Obelić, B., Srdoč, D., The simultaneous measurement of tritium activity and the background count rate in a proportional counter by the Povinec method: Three years experience at the Ruđer Bošković Institute. Nuclear Instrum. Meth. in Physics Research B17: , (1986). 8. Horvatinčić, N., Obelić, B., Krajcar Bronić, I., Srdoč, D., Bistrović, R., Sources of radon contamination in C-14 dating. Radiocarbon 37: , (1995). 9. Srdoč, D., Obelić, B., Horvatinčić, N., Sliepčević, A., Measurement of the 14 C Activity of the ANU Sucrose Secondary Standard by Means of the Proportional Counter Technique. Radiocarbon 21/3: , (1979). 10. Barešić, J., Krajcar Bronić, I., Horvatinčić, N., Obelić, B., in Proceedings of the Fifth symposium of the Croatian Radiation Protection Association, Stubičke Toplice, Croatia, 2003, edited by I. Krajcar Bronić, S. Miljanić, B. Obelić (CRPA, Zagreb, 2003), p Horvatinčić, N., Barešić, J., Krajcar Bronić, I., Obelić, B., Measurement of low 14 C activities in liquid scintillation counter in the Zagreb Radiocarbon Laboratory. 18 th International Radiocarbon Conference, Wellington, New Zealand, 2003, submitted to Radiocarbon. 12. Stuiver, M., Polach, H.A., Discussion: Reporting of 14 C data. Radiocarbon 19: , (1977). 7

Advances in radiocarbon measurement of water samples

Advances in radiocarbon measurement of water samples Advances in radiocarbon measurement of water samples R. Janovics a, M. Molnár a, I. Světlík b, I. Major a, L. Wacker c a) Institute of Nuclear Research, HAS, Debrecen, Hungary (ATOMKI) janovics@atomki.hu

More information

COMPARISON OF TWO DIFFERENT METHODS FOR DETERMINATION OF BIOGENIC FRACTION IN LIQUID FUELS

COMPARISON OF TWO DIFFERENT METHODS FOR DETERMINATION OF BIOGENIC FRACTION IN LIQUID FUELS COMPARISON OF TWO DIFFERENT METHODS FOR DETERMINATION OF BIOGENIC FRACTION IN LIQUID FUELS Jovana Nikolov 1, Ines Krajcar Bronić 2, Ivana Stojković 3, Nataša Todorović 1, Jadranka Barešić 2, Matea Krmpotić

More information

DETERMINATION OF RADON BY LIQUID SCINTILLATION a/f PARTICLE SPECTROMETRY: Towards the Resolution of a 14C Dating Problem

DETERMINATION OF RADON BY LIQUID SCINTILLATION a/f PARTICLE SPECTROMETRY: Towards the Resolution of a 14C Dating Problem [RADIOCARBON, VOL 30, No. 1, 1988, P 19-24] DETERMINATION OF RADON BY LIQUID SCINTILLATION a/f PARTICLE SPECTROMETRY: Towards the Resolution of a 14C Dating Problem HENRY POLACH Australian National University,

More information

RADIOCARBON DATING BEYOND 50,000 YEARS BY LIQUID SCINTILLATION COUNTING

RADIOCARBON DATING BEYOND 50,000 YEARS BY LIQUID SCINTILLATION COUNTING RADIOCARBON DATING BEYOND 50,000 YEARS BY LIQUID SCINTILLATION COUNTING F. G. MCCORMAC', R. M. KALIN2 and AUSTIN LONGS ABSTRACT. The maximum age that can be determined by 14C dating using the LSC method

More information

1220 QUANTULUS The Ultra Low Level Liquid Scintillation Spectrometer

1220 QUANTULUS The Ultra Low Level Liquid Scintillation Spectrometer 1220 QUANTULUS The Ultra Low Level Liquid Scintillation Spectrometer PerkinElmer LAS (UK) Ltd, Chalfont Rd, Seer Green, Beaconsfield, Bucks HP9 2FX tel: 0800 896046 www.perkinelmer.com John Davies January

More information

COMPARATIVE STUDY BETWEEN EXTERNAL STANDARD METHOD AND INTERNAL STANDARD METHOD FOR LOW-LEVEL TRITIUM MEASUREMENTS

COMPARATIVE STUDY BETWEEN EXTERNAL STANDARD METHOD AND INTERNAL STANDARD METHOD FOR LOW-LEVEL TRITIUM MEASUREMENTS International Conference Nuclear Energy in Central Europe 2001 Hoteli Bernardin, Portorož, Slovenia, September 10-13, 2001 www: http://www.drustvo-js.si/port2001/ e-mail: PORT2001@ijs.si tel.:+ 386 1 588

More information

COSMIC BACKGROUND REDUCTION IN THE RADIOCARBON MEASUREMENTS BY LIQUID SCINTILLATION SPECTROMETRY AT THE UNDERGROUND LABORATORY OF GRAN SASSO

COSMIC BACKGROUND REDUCTION IN THE RADIOCARBON MEASUREMENTS BY LIQUID SCINTILLATION SPECTROMETRY AT THE UNDERGROUND LABORATORY OF GRAN SASSO COSMIC BACKGROUND REDUCTION IN THE RADIOCARBON MEASUREMENTS BY LIQUID SCINTILLATION SPECTROMETRY AT THE UNDERGROUND LABORATORY OF GRAN SASSO Wolfango Plastino Lauri Kaihola Paolo Bartolomei 3 Francesco

More information

GEMMA RAURET, J S MESTRES and J F GARCIA

GEMMA RAURET, J S MESTRES and J F GARCIA [RADIOCARBON, VOL31, No. 3, 1989, P 380-3861 OPTIMIZATION OF LIQUID SCINTILLATION COUNTING CONDITIONS WITH TWO KINDS OF VIALS AND DETECTOR SHIELDS FOR LOW-ACTIVITY RADIOCARBON MEASUREMENTS GEMMA RAURET,

More information

Depth Distribution of H-3, C-14 and Co-60 in Decommissioning of the Biological Shielding Concrete of KRR-2

Depth Distribution of H-3, C-14 and Co-60 in Decommissioning of the Biological Shielding Concrete of KRR-2 Depth Distribution of H-3, C-14 and Co-60 in Decommissioning of the Biological Shielding Concrete of KRR-2 S.B. Hong, H.R. Kim, K.H. Chung, K.H. Chung, J.H. Park Korea Atomic Energy Research Institute

More information

PERFORMANCE AND DESIGN OF 03-ML TO 10-ML SYNTHETIC SILICA LIQUID SCINTILLATION VIALS FOR LOW-LEVEL 14C DETERMINATION

PERFORMANCE AND DESIGN OF 03-ML TO 10-ML SYNTHETIC SILICA LIQUID SCINTILLATION VIALS FOR LOW-LEVEL 14C DETERMINATION PERFORMANCE AND DESIGN OF 03-ML TO 10-ML SYNTHETIC SILICA LIQUID SCINTILLATION VIALS FOR LOW-LEVEL 14C DETERMINATION A. G. HOGG Radiocarbon Dating Laboratory, University of Waikato, Hamilton, New Zealand

More information

Radiocarbon dating method as a tool for cultural heritage protection

Radiocarbon dating method as a tool for cultural heritage protection Radiocarbon dating method as a tool for cultural heritage protection Nada Horvatinčić, Ines Krajcar Bronić, Jadranka Barešić Ruđer Bošković Institute, Zagreb Department of experimental phisics Laboratory

More information

TECHNETIUM-99 IN WATER

TECHNETIUM-99 IN WATER Analytical Procedure TECHNETIUM-99 IN WATER (TEVA DISC METHOD) 1. SCOPE 1.1. This procedure describes a method to separate and measure technetium-99 in water. 1.2. This method does not address all aspects

More information

Liquid Scintillation Counter

Liquid Scintillation Counter Instrumentation & Methods: s & Tritium Richard Sheibley Pennsylvania Dept of Env Protection Principle Beta particle emission Energy transferred to Solute Energy released as UV Pulse Intensity proportional

More information

HIGH-SENSITIVITY RADIOCARBON DATING IN THE 50,000 TO 70,000 BP RANGE WITHOUT ISOTOPIC ENRICHMENT

HIGH-SENSITIVITY RADIOCARBON DATING IN THE 50,000 TO 70,000 BP RANGE WITHOUT ISOTOPIC ENRICHMENT [RADIOCARBON, VOL. 34, No. 3, 1992, P. 351-3591 HIGH-SENSITIVITY RADIOCARBON DATING IN THE 50,000 TO 70,000 BP RANGE WITHOUT ISOTOPIC ENRICHMENT AUSTIN LONG and ROBERT M. KALIN1 Environmental Radioisotope

More information

IvanaJakonić 1, Nataša Todorović 2, JovanaNikolov 2, Ines KrajcarBronić 3, BranislavaTenjović 4, MiroslavVesković 2

IvanaJakonić 1, Nataša Todorović 2, JovanaNikolov 2, Ines KrajcarBronić 3, BranislavaTenjović 4, MiroslavVesković 2 Optimization of low-levellevel LS counter Quantulus 1220 for tritium determination in water samples IvanaJakonić 1, Nataša Todorović 2, JovanaNikolov 2, Ines KrajcarBronić 3, BranislavaTenjović 4, MiroslavVesković

More information

Liquid Scintillation Counting Performance Using Glass Vials in the Wallac 1220 QuantulusTM

Liquid Scintillation Counting Performance Using Glass Vials in the Wallac 1220 QuantulusTM CHAPTER 43 Liquid Scintillation Counting Performance Using Glass Vials in the Wallac 1220 QuantulusTM Lauri Kaihola ABSTRACT Low-potassium glass vials can obtain reduced background count rates for beta

More information

content of the dissolved bicarbonates. The applicability of several models

content of the dissolved bicarbonates. The applicability of several models [RADIocARBoN, VOL 28, No. 2A, 1986, P 436-440] ON THE INITIAL ACTIVITY OF KARST AQUIFERS WITH SHORT MEAN RESIDENCE TIME INES KRAJCAR-BRONIC, NADA HORVATINCIC, DUSAN SRDOC, and BOGOMIL OBELIC Rudjer Boskovic

More information

Energy Calibration of Liquid Scintillation Counter to allow Semi-Qualitative Nuclide Identification in Water Samples

Energy Calibration of Liquid Scintillation Counter to allow Semi-Qualitative Nuclide Identification in Water Samples Energy Calibration of Liquid Scintillation Counter to allow Semi-Qualitative Nuclide Identification in Water Samples Mashaba, Machel. 1*, Kotze, Deon. 2, Tshivhase, Victor.M. 1, Faanhof, Arnaud. 1, 2 1

More information

EVALUATION OF HIGH-PURITY SYNTHETIC SILICA VIALS IN ACTIVE AND PASSIVE VIAL HOLDERS FOR LIQUID SCINTILLATION COUNTING OF BENZENE

EVALUATION OF HIGH-PURITY SYNTHETIC SILICA VIALS IN ACTIVE AND PASSIVE VIAL HOLDERS FOR LIQUID SCINTILLATION COUNTING OF BENZENE [RADIOCARBON, VOL. 34, No. 3, 1992, P. 394-401] EVALUATION OF HIGH-PURITY SYNTHETIC SILICA VIALS IN ACTIVE AND PASSIVE VIAL HOLDERS FOR LIQUID SCINTILLATION COUNTING OF BENZENE A. G. HOGG Radiocarbon Dating

More information

A New Measurement Method For Tritium Rapid Detection by LSC. IM.CAEP Xu Qinghua

A New Measurement Method For Tritium Rapid Detection by LSC. IM.CAEP Xu Qinghua A New Measurement Method For Tritium Rapid Detection by LSC IM.CAEP Xu Qinghua xuqinghuawork@sina.com Outline Background Results and discussion Materials and methods Background FOM value for channel Sample

More information

HIGH-PRECISION 14C MEASUREMENT OF GERMAN AND IRISH OAKS TO SHOW THE NATURAL 14C VARIATIONS FROM 7890 TO 5000 BC

HIGH-PRECISION 14C MEASUREMENT OF GERMAN AND IRISH OAKS TO SHOW THE NATURAL 14C VARIATIONS FROM 7890 TO 5000 BC [RADIOCARBON, VOL. 35, No. 1, 1993, P. 93-104] HIGH-PRECISION MEASUREMENT OF GERMAN AND IRISH OAKS TO SHOW THE NATURAL VARIATIONS FROM 7890 TO 5000 BC GORDON W. PEARSON', BERND BECKER2 and FLORENCE QUA'

More information

The measurement of tritium in environmental water samples

The measurement of tritium in environmental water samples The measurement of tritium in environmental water samples 1. Introduction The Hidex 300 SL automatic TDCR liquid scintillation provides an excellent solution for the determination of H-3 in drinking water.

More information

LSC-BASED APPROACH FOR RADON IN SOIL GAS MEASUREMENTS. The Marzeev Institute of Hygiene and Medical Ecology 50, Popudrenko str., Kiev 02094, Ukraine.

LSC-BASED APPROACH FOR RADON IN SOIL GAS MEASUREMENTS. The Marzeev Institute of Hygiene and Medical Ecology 50, Popudrenko str., Kiev 02094, Ukraine. LSC-BASED APPROACH FOR RADON IN SOIL GAS MEASUREMENTS Michael Buzinny 1 Victor Sakhno Maxim Romanchenko The Marzeev Institute of Hygiene and Medical Ecology 50, Popudrenko str., Kiev 02094, Ukraine. ABSTRACT.

More information

Isotope analyses of the lake sediments in the Plitvice Lakes, Croatia Research Article

Isotope analyses of the lake sediments in the Plitvice Lakes, Croatia Research Article Cent. Eur. J. Phys. 12(10) 2014 707-713 DOI: 10.2478/s11534-014-0490-7 Central European Journal of Physics Isotope analyses of the lake sediments in the Plitvice Lakes, Croatia Research Article Nada Horvatinčić

More information

DETERMINING FUNDAMENTAL PARAMETERS OF A SINGLE-PHOTOTUBE LIQUID SCINTILLATION COUNTER

DETERMINING FUNDAMENTAL PARAMETERS OF A SINGLE-PHOTOTUBE LIQUID SCINTILLATION COUNTER DETERMINING FUNDAMENTAL PARAMETERS OF A SINGLE-PHOTOTUBE LIQUID SCINTILLATION COUNTER Pall Theodórsson Science Institute, University of Iceland, Dunhaga 3, IS-107 Reykjavík, Iceland. Email: pth@raunvis.hi.is.

More information

BALANCED WINDOW METHOD IN 14 C LIQUID SCINTILLATION COUNTING. P TheodÛrsson 1 S Ingvarsdottir G I Gudjonsson

BALANCED WINDOW METHOD IN 14 C LIQUID SCINTILLATION COUNTING. P TheodÛrsson 1 S Ingvarsdottir G I Gudjonsson RADIOCARBON, Vol 45, Nr 1, 2003, p 113 122 2003 by the Arizona Board of Regents on behalf of the University of Arizona BALANCED WINDOW METHOD IN 14 C LIQUID SCINTILLATION COUNTING P TheodÛrsson 1 S Ingvarsdottir

More information

[RADIOCARBON, VOL 31, No. 3, 1989, P ]

[RADIOCARBON, VOL 31, No. 3, 1989, P ] [RADIOCARBON, VOL 31, No. 3, 1989, P 254-259] RADON ELIMINATION DURING BENZENE PREPARATION FOR RADIOCARBON DATING BY LIQUID SCINTILLATION SPECTROMETRY DARDEN HOOD, RONALD HATFIELD, CHRISTOPHER PATRICK,

More information

Effect of liquid scintillating cocktail volume on 3 H and 14 C measurement parameters using a Quantulus spectrometer

Effect of liquid scintillating cocktail volume on 3 H and 14 C measurement parameters using a Quantulus spectrometer NUKLEONIKA 2010;55(2):155 161 ORIGINAL PAPER Effect of liquid scintillating cocktail volume on 3 H and 14 C measurement parameters using a Quantulus spectrometer Andrzej Komosa, Katarzyna Ślepecka Abstract.

More information

DIRECT ABSORPTION METHOD AND LIQUID SCINTILLATION COUNTING FOR RADIOCARBON MEASUREMENTS IN ORGANIC CARBON FROM SEDIMENTS

DIRECT ABSORPTION METHOD AND LIQUID SCINTILLATION COUNTING FOR RADIOCARBON MEASUREMENTS IN ORGANIC CARBON FROM SEDIMENTS DIRECT ABSORPTION METHOD AND LIQUID SCINTILLATION COUNTING FOR RADIOCARBON MEASUREMENTS IN ORGANIC CARBON FROM SEDIMENTS Ionut Faurescu 1,2 Carmen Varlam 1 Ioan Stefanescu 1 Stela Cuna 3 Irina Vagner 1

More information

LIQUID SCINTILLATION COUNTERS. {Beta Counters}

LIQUID SCINTILLATION COUNTERS. {Beta Counters} LIQUID SCINTILLATION COUNTERS {Beta Counters} We offer a range of Alpha, Beta & Gama counters, from Hidex Oy, Finland to meet your specific Liquid Scintillation counting requirements. Triathler LSC Sense

More information

Implementation of ISO/IEC in a low level liquid s cintillation tritium laboratory

Implementation of ISO/IEC in a low level liquid s cintillation tritium laboratory Implementation of ISO/IEC 17025 in a low level liquid s cintillation tritium laboratory LSC 2017 Advances in Liquid Scintillation Spectrometry 1-5 May 2017, Copenhagen, Denmark Ll. Pujol, M.E. Pérez-Zabaleta

More information

Making sense of Limits of Detection (with respect to Tritium analysis) James Thomson Meridian Biotechnologies Ltd.

Making sense of Limits of Detection (with respect to Tritium analysis) James Thomson Meridian Biotechnologies Ltd. Making sense of Limits of Detection (with respect to Tritium analysis) James Thomson Meridian Biotechnologies Ltd. The dilemma Customer commented that the background of Gold Star had increased and was

More information

TECHNETIUM-99 IN SOIL

TECHNETIUM-99 IN SOIL Analytical Procedure TECHNETIUM-99 IN SOIL 1. SCOPE 1.1. This procedure describes a method to separate and measure technetium-99 in soil. 1.2. This method does not address all aspects of safety, quality

More information

COMBINED LSC-BASED METHOD FOR RADON IN AIR MEASUREMENTS

COMBINED LSC-BASED METHOD FOR RADON IN AIR MEASUREMENTS COMBINED LSC-BASED METHOD FOR RADON IN AIR MEASUREMENTS Michael Buzinny The Marzeev Institute of Hygiene and Medical Ecology AMS of Ukraine, 50, Popudrenko str., Kiev 02094, Ukraine. Email:mbuz@ukr.net

More information

COMBINED LSC BASED METHOD FOR RADON IN AIR MEASUREMENTS

COMBINED LSC BASED METHOD FOR RADON IN AIR MEASUREMENTS COMBINED LSC BASED METHOD FOR RADON IN AIR MEASUREMENTS Michael Buzinny Institute of Hygiene an Medical Ecology Academy of Medical Sciences of Ukraine, Kiev, Ukraine, mbuz@ukr.net Page 1 IDEA Use sensitivity

More information

Teflon Vials For Precise C-14 in Benzene Measurements by LSC Technique

Teflon Vials For Precise C-14 in Benzene Measurements by LSC Technique "Science Stays True Here" Biological and Chemical Research, 229-233 Science Signpost Publishing Teflon Vials For Precise C-14 in Benzene Measurements by LSC Technique Vadim V. Skripkin 1, Mykhailo G. Buzynnyi

More information

LSUF Update on Organically Bound Tritium Standard project. Arvic Harms. 5 May 2004

LSUF Update on Organically Bound Tritium Standard project. Arvic Harms. 5 May 2004 LSUF 2004 Update on Organically Bound Tritium Standard project Arvic Harms 5 May 2004 Contents What is tritium and what is organically bound tritium (OBT) and why is it important? Why do we need an OBT

More information

SOURCES OF RANDOM ERROR IN THE DEBRECEN RADIOCARBON LABORATORY

SOURCES OF RANDOM ERROR IN THE DEBRECEN RADIOCARBON LABORATORY [RADIOCARBON, VOL 32, No. 3, 1990, P 283-287] SOURCES OF RANDOM ERROR IN THE DEBRECEN RADIOCARBON LABORATORY EDE HERTELENDI Institute of Nuclear Research of the Hungarian Academy of Sciences PO Box S'

More information

RADIOLOGICAL CHARACTERIZATION Laboratory Procedures

RADIOLOGICAL CHARACTERIZATION Laboratory Procedures RADIOLOGICAL CHARACTERIZATION Laboratory Procedures LORNA JEAN H. PALAD Health Physics Research Unit Philippine Nuclear Research Institute Commonwealth Avenue, Quezon city Philippines 3-7 December 2007

More information

TIME-RESOLVED LIQUID SCINTILLATION COUNTING

TIME-RESOLVED LIQUID SCINTILLATION COUNTING [RADOCARBON, VOL 32, No. 3, 1990, P 381-386] TME-RESOLVED LQUD SCNTLLATON COUNTNG MCHAEL KESSLER Packard nstrument Company, One State Street, Meriden, Connecticut 06450 ABSTRACT. Historically, scientists

More information

Northwest University, Mafikeng, Centre of Applied Radiation Science & Technology (CARST), South Africa.

Northwest University, Mafikeng, Centre of Applied Radiation Science & Technology (CARST), South Africa. Gross Alpha and Beta Measurements of Water Samples from the Wonderfonteinspruit Catchment Area in the Gauteng Province (South Africa), using Liquid Scintillation Counting Machel Mashaba 1, Deon Kotze 2,

More information

THE STATISTICS OF LOW-LEVEL COUNTING USING THE NEW GENERATION OF PACKARD LIQUID SCINTILLATION COUNTERS

THE STATISTICS OF LOW-LEVEL COUNTING USING THE NEW GENERATION OF PACKARD LIQUID SCINTILLATION COUNTERS [RADIOCARBON, VOL. 34, No. 3, 1992, P. 36-365] THE STATISTICS OF LOW-LEVEL COUNTING USING THE NEW GENERATION OF PACKARD LIQUID SCINTILLATION COUNTERS G. T. COOK1, E. M. SCOTT2, E. M. WRIGHT' and ROBERT

More information

Procedure for Radio Dating of Water Using Nuclear Hydrological Isotopes Techniques

Procedure for Radio Dating of Water Using Nuclear Hydrological Isotopes Techniques IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-4861.olume 6, Issue 2 er. I (Mar-Apr. 2014), PP 49-53 Procedure for Radio Dating of Water Using Nuclear Hydrological Isotopes Techniques Muhammad

More information

Development of Continuous Inflow Tritium in Water Measurement Technology by using Electrolysis and Plastic Scintillator

Development of Continuous Inflow Tritium in Water Measurement Technology by using Electrolysis and Plastic Scintillator Development of Continuous Inflow Tritium in Water Measurement Technology by using Electrolysis and Plastic Scintillator 4, May, 2017 Jun Woo Bae, Ukjae Lee, Hee Reyoung Kim 1 Outline Introduction Material

More information

Working Copy. P /ru /,6. 8-tt tt. EffectiveDate: I ^, Operation and Calibration of the Tri-Carb Liquid Sclntillation Counter QAM-RI.106.

Working Copy. P /ru /,6. 8-tt tt. EffectiveDate: I ^, Operation and Calibration of the Tri-Carb Liquid Sclntillation Counter QAM-RI.106. QAM-RI.106 Operation and Calibration of the Tri-Carb Liquid Sclntillation Counter Revision: 0 Laboratory Man ager/lqao/rso Date Date EffectiveDate: I ^, P /ru /,6 8-tt tt Renewal Date: lnitials: Texas

More information

Environmental Isotopes in Hydrology. Woocay substituting for Walton

Environmental Isotopes in Hydrology. Woocay substituting for Walton Environmental Isotopes in Hydrology Oct 7, 2010 1 What is an Isotope? An element is defined by the number of protons (Z) in the nucleus The number of neutrons (N) defines the isotope(s) of that element

More information

DIFFICULTIES WITH 228 Ra DETERMINATION FROM THE LOW-ENERGY BETA SPECTRUM COMPONENT IN WATER SAMPLES VIA LSC. A Walencik 1 B Koz owska 2

DIFFICULTIES WITH 228 Ra DETERMINATION FROM THE LOW-ENERGY BETA SPECTRUM COMPONENT IN WATER SAMPLES VIA LSC. A Walencik 1 B Koz owska 2 DIFFICULTIES WITH 228 Ra DETERMINATION FROM THE LOW-ENERGY BETA SPECTRUM COMPONENT IN WATER SAMPLES VIA LSC A Walencik 1 B Koz owska 2 University of Silesia, Institute of Physics, Uniwersytecka 4, 40-007

More information

Measurement of the carbon 14 activity at natural level in air samples

Measurement of the carbon 14 activity at natural level in air samples Radioprotection, Suppl. 1, vol. 40 (2005) S791-S796 EDP Sciences, 2005 DOI: 10.1051/radiopro:2005s1-116 Measurement of the carbon 14 activity at natural level in air samples A. Olivier 1, L. Tenailleau

More information

Radiocarbon dating of paper and parchment

Radiocarbon dating of paper and parchment Radiocarbon dating of paper and parchment Ines Krajcar Bronić, Nada Horvatinčić, Jadranka Barešić, Andreja Sironić, Damir Borković Ruđer Bošković Institute, Zagreb krajcar@irb.hr Zagreb, 2.-3.12.2015.

More information

This document is a preview generated by EVS

This document is a preview generated by EVS INTERNATIONAL STANDARD ISO 13168 First edition 2015-07-01 Water quality Simultaneous determination of tritium and carbon 14 activities Test method using liquid scintillation counting Qualité de l eau Détermination

More information

TRITIUM DETERMINATION IN WATER

TRITIUM DETERMINATION IN WATER TRITIUM DETERMINATION IN WATER Ricardo M. Gavini 1 Autoridad Regulatoria Nuclear, Av. del Libertador 8250, 1429, Buenos Aires, Argentina. Abstract. An analytical procedure for the determination of tritium

More information

Minimizing new tritium line sensitivity by a quality process implemented at Madagascar-I.N.S.T.N

Minimizing new tritium line sensitivity by a quality process implemented at Madagascar-I.N.S.T.N Minimizing new tritium line sensitivity by a quality process implemented at Madagascar-I.N.S.T.N J. Rajaobelison, Raoelina Andriambololona, V. Ramaroson, G. Andriamiharitsoa, L.P. Fareze, F.F.A. Razafitsalama

More information

Time-Resolved Liquid Scintillation Counting

Time-Resolved Liquid Scintillation Counting CHAPTER 44 Time-Resolved Liquid Scintillation Counting Norbert Roessler, Robert J. Valenta, and Stat van Cauter ABSTRACT A comparison is made between standard, two-tube coincidence liquid scintillation

More information

GAS PROPORTIONAL VERSUS LIQUID SCINTILLATION COUNTING, RADIOMETRIC VERSUS AMS DATING PALL THEODORSSON

GAS PROPORTIONAL VERSUS LIQUID SCINTILLATION COUNTING, RADIOMETRIC VERSUS AMS DATING PALL THEODORSSON [RADIOCARBON, VOL 33, No. 1, 1991, P 9-13] GAS PROPORTIONAL VERSUS LIQUID SCINTILLATION COUNTING, RADIOMETRIC VERSUS AMS DATING PALL THEODORSSON Science Institute, University of Iceland, Dunhaga 3,107

More information

Determination of H-3 and C-14 in the frame of decommissioning projects at the Paul Scherrer Institute

Determination of H-3 and C-14 in the frame of decommissioning projects at the Paul Scherrer Institute Determination of H-3 and C-14 in the frame of decommissioning projects at the Paul Scherrer Institute Jost Eikenberg, Maya Jäggi, Max Rüthi Paul Scherrer Institute, CH-5232 Villigen, Switzerland Content

More information

LABORATORY OF COSMOCHEMISTRY RADIOCARBON MEASUREMENTS I

LABORATORY OF COSMOCHEMISTRY RADIOCARBON MEASUREMENTS I [RADIOCARBON, VOL. 19, No. 1, 1977, P. 12-18] LABORATORY OF COSMOCHEMISTRY RADIOCARBON MEASUREMENTS I A K LAVRUKHINA and V A ALEXEEV V I Vernadsky Institute of Geochemistry and Analytical Chemistry, USSR

More information

DISTRIBUTION LIST To be filled out by Document Author or person requesting document deletion

DISTRIBUTION LIST To be filled out by Document Author or person requesting document deletion DISTRIBUTION LIST To be filled out by Document Author or person requesting document deletion: Once approved, the new or revised document or the notice of deleted document(s) should be distributed to all

More information

BUREAU INTERNATIONAL DES POIDS ET MESURES

BUREAU INTERNATIONAL DES POIDS ET MESURES 1 BUREAU INTERNATIONAL DES POIDS ET MESURES International comparison of activity measurements of a solution of 3 H (January 2009) Participating laboratory: T ½ = (4 496.862 d; u = 9.131 d)* Ampoule number

More information

DETERMINATION OF BETA EMITTERS IN MATERIALS FROM RESEARCH REACTOR DECOMISSIONING

DETERMINATION OF BETA EMITTERS IN MATERIALS FROM RESEARCH REACTOR DECOMISSIONING DETERMINATION OF BETA EMITTERS IN MATERIALS FROM RESEARCH REACTOR DECOMISSIONING Andreas Vesely 1 Herbert Trombitas Helmut Lindauer Nuclear Engineering Seibersdorf, 2444 Seibersdorf, Austria. ABSTRACT.

More information

1/5 Specifications Hidex 300 SLL automatic TDCR liquid scintillation counter

1/5 Specifications Hidex 300 SLL automatic TDCR liquid scintillation counter 1/5 Specification Hidex 300 SL-SLL 100-240V AC and 24 V DC. Features Model (425-020) Super Low Level (SLL) Automatic TDCR Liquid Scintillation Counter with Guard Detector and Low Level PMTs General Description

More information

TECHNETIUM-99 IN WATER

TECHNETIUM-99 IN WATER Analytical Procedure TECHNETIUM-99 IN WATER (WITH VACUUM BOX SYSTEM) 1. SCOPE 1.1. This procedure describes a method to separate and measure technetium-99 in water. 1.2. This method does not address all

More information

RADIOCARBON DATING REPRODUCIBILITY AT THE MUSEO DE LA PLATA RADIOCARBON LABORATORY

RADIOCARBON DATING REPRODUCIBILITY AT THE MUSEO DE LA PLATA RADIOCARBON LABORATORY [RADIOCARBON, VOL 32, No. 3, 1990, P 301-305] RADIOCARBON DATING REPRODUCIBILITY AT THE MUSEO DE LA PLATA RADIOCARBON LABORATORY ANIBAL FIGINI, ROBERTO HUARTE AND JORGE CARBONARI Laboratorio de Tritio

More information

HIGH-PRECISION INTERCOMPARISON AT ISOTRACE

HIGH-PRECISION INTERCOMPARISON AT ISOTRACE [RADIOCARBON, VOL 32, No. 3, 1990, P 335-339] HIGH-PRECISION INTERCOMPARISON AT ISOTRACE ROELF P BEUKENS IsoTrace Laboratory, University of Toronto, Toronto, Ontario M5S 1A7, Canada ABSTRACT. I conducted

More information

Activity determination of 88 Y by means of 4πβ(LS)-γ coincidence counting

Activity determination of 88 Y by means of 4πβ(LS)-γ coincidence counting Activity determination of 88 Y by means of 4πβ(LS)-γ coincidence counting Justyna Marganiec-Galazka Ole J. Nähle Karsten Kossert Division 6 Ionizing Radiation Department 6.1 Radioactivity Working Group

More information

WM2018 Conference, March 18-22, 2018, Phoenix, Arizona, USA. PVT and LaBr3(Ce)-based Radon Express Analyzers 18164

WM2018 Conference, March 18-22, 2018, Phoenix, Arizona, USA. PVT and LaBr3(Ce)-based Radon Express Analyzers 18164 PVT and LaBr3(Ce)-based Radon Express Analyzers 864 Vladislav Kondrashov *, Stephen Steranka* and Glenn Paulson** * RadComm Systems Corp. 293 Portland Dr, Oakville, Ontario L6H 5S4, CANADA ** Paulson and

More information

2100TR Liquid Scintillation Counter

2100TR Liquid Scintillation Counter 2100TR Liquid Scintillation Counter Description The Tri-Carb 2100TR liquid scintillation counter is computer-controlled, bench top liquid scintillation analyzer for detecting small amounts of alpha, beta

More information

SLAC Radioanalysis Laboratory

SLAC Radioanalysis Laboratory SLAC Radioanalysis Laboratory Henry Brogonia Dosimetry and Radiological Protection Group (DREP) DOE Environmental Radiation Protection Program Review (July 23-24, 2007) Radioanalysis Laboratory Mission

More information

GUIDE TO LABORATORY SURVEYS. Introduction

GUIDE TO LABORATORY SURVEYS. Introduction APPENDIX - V GUIDE TO LABORATORY SURVEYS Introduction Routine laboratory surveys are an important part of the overall radiation safety program in a laboratory. Surveys provide a direct measure of the presence

More information

Determination of 210 Pb and 210 Po in Water Samples

Determination of 210 Pb and 210 Po in Water Samples 1 Determination of 210 Pb and 210 Po in Water Samples Marin Ayranov 1, Zornitza Tosheva 2, Antoine Kies 2 1 Institute for Nuclear Research and Nuclear Energy, 72 Tzarigradsko chaussee, BG-1784 Sofia, Bulgaria

More information

DEVELOPMENT OF A COMBUSTION SYSTEM FOR LIQUID OR GAS SAMPLES

DEVELOPMENT OF A COMBUSTION SYSTEM FOR LIQUID OR GAS SAMPLES RADIOCARBON, Vol 46, Nr 1, 2004, p 141 145 2004 by the Arizona Board of Regents on behalf of the University of Arizona DEVELOPMENT OF A COMBUSTION SYSTEM FOR LIQUID OR GAS SAMPLES J H Park C S Lee 1 Department

More information

Liquid Scintillation at NPL: Practical Applications

Liquid Scintillation at NPL: Practical Applications 2003 LS User s Forum Liquid Scintillation at NPL: Practical Applications Andy Pearce CAIR, NPL 2 nd April 2003 Introduction Overview of applications Current projects Neutron-activated thioacetamide discs

More information

Preparation of Standard Source as a Petri Dish for Plant by Using

Preparation of Standard Source as a Petri Dish for Plant by Using Preparation of Standard Source as a Petri Dish for Plant by Using 152 Eu Element Khalid H. AL-Ubaidi 1 Salam K. AL-Nasri 2 Auday T. AL-Bayati 1* 1.Department of Physics, College of Education / Ibn Al-

More information

SHERIDAN BOWMAN British Museum Research Laboratory London WC1B 3DG, England

SHERIDAN BOWMAN British Museum Research Laboratory London WC1B 3DG, England [RADIOCARBON, VOL 31, No. 3, 1989, P 393-3981 LIQUID SCINTILLATION COUNTING IN THE LONDON UNDERGROUND SHERIDAN BOWMAN British Museum Research Laboratory London WC1B 3DG, England ABSTRACT. At the 12th International

More information

PLASTIC SCINTILLATORS: A POWERFUL TOOL TO REDUCE MIXED WASTE.

PLASTIC SCINTILLATORS: A POWERFUL TOOL TO REDUCE MIXED WASTE. PLASTIC SCINTILLATORS: A POWERFUL TOOL TO REDUCE MIXED WASTE. A. Tarancón, H. Bagán, J.F. García * and G. Rauret. Departament de Química Analítica. Facultat de Química. Universitat de Barcelona. C/ Diagonal

More information

The LSC Approach to Radon Counting in Air and Water

The LSC Approach to Radon Counting in Air and Water CHAPTER 32 The LSC Approach to Radon Counting in Air and Water Charles J. Passo, Jr. and James M. Floeckher INTRODUCTION Various methods exist to monitor 222Rn in air. There are seven commonly used types

More information

Dating of ground water

Dating of ground water PART 16 Dating of ground water Introduction Why date? - to determine when recharge occurred - to determine groundwater velocities - to reconstruct regional flow patterns How to do this? - decay of radioactive

More information

Acronyms, Abbreviations, and Symbols Foreword to the First Edition Foreword to the Second Edition Preface to the First Edition Preface to the Second

Acronyms, Abbreviations, and Symbols Foreword to the First Edition Foreword to the Second Edition Preface to the First Edition Preface to the Second Contributors p. xxix Acronyms, Abbreviations, and Symbols p. xxxi Foreword to the First Edition p. xliii Foreword to the Second Edition p. xlv Preface to the First Edition p. xlvii Preface to the Second

More information

SIGURDUR A EINARSSON and PALL THEODORSSON Science Institute, University of Iceland Dunhaga 3, IS-107 Reykjavik, Iceland

SIGURDUR A EINARSSON and PALL THEODORSSON Science Institute, University of Iceland Dunhaga 3, IS-107 Reykjavik, Iceland [RADIOCARBON, VOL 31, No. 3,1989, P 342-351] STUDY OF BACKGROUND PULSE SPECTRUM OF AN LSC SYSTEM SIGURDUR A EINARSSON and PALL THEODORSSON Science Institute, University of Iceland Dunhaga 3, IS-107 Reykjavik,

More information

Jerde, Peterson and Stein (1967) indicated that high-energy radiation

Jerde, Peterson and Stein (1967) indicated that high-energy radiation [RADIOCARBON, VOL 31, No. 3, 1989, P 332-341] LOW BACKGROUND LIQUID SCINTILLATION COUNTING USING AN ACTIVE SAMPLE HOLDER AND PULSE DISCRIMINATION ELECTRONICS JOHN E NOAKES Center for Applied Isotope Studies

More information

Hou, Xiaolin; Roos, Per. Published in: Analytica Chimica Acta. Link to article, DOI: /j.aca Publication date: 2008

Hou, Xiaolin; Roos, Per. Published in: Analytica Chimica Acta. Link to article, DOI: /j.aca Publication date: 2008 Downloaded from orbit.dtu.dk on: Jul 04, 2018 Critical comparison of radiometric and mass spectrometric methods for the determination of radionuclides in environmental, biological and nuclear waste samples

More information

RESULTS YOU CAN COUNT ON

RESULTS YOU CAN COUNT ON RESULTS YOU CAN COUNT ON Liquid Scintillation Counting Principle and the Application in Biological Research 1 2009 PerkinElmer liquid scintillation counting β particle Tri-Carb family of liquid scintillation

More information

RESOLUTION OF DIFFERENCES IN CONCENTRATION OF NATURALLY OCCURRING TRITIUM IN GROUNDWATER TRACER STUDIES

RESOLUTION OF DIFFERENCES IN CONCENTRATION OF NATURALLY OCCURRING TRITIUM IN GROUNDWATER TRACER STUDIES RESOLUTION OF DIFFERENCES IN CONCENTRATION OF NATURALLY OCCURRING TRITIUM IN GROUNDWATER TRACER STUDIES Michael P. Neary AUTHOR: Senior Research Scientist, Center for Applied Isotope Studies, University

More information

SECTION 8 Part I Typical Questions

SECTION 8 Part I Typical Questions SECTION 8 Part I Typical Questions 1. For a narrow beam of photons, the relaxation length is that thickness of absorber that will result in a reduction of in the initial beam intensity. 1. 1/10. 2. 1/2.

More information

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL A. Specification Gamma-spectrometry method is used to identify and determine the activity concentration

More information

Unit 2. Instrumentation. Experts Teaching from Practical Experience

Unit 2. Instrumentation. Experts Teaching from Practical Experience Unit 2 Instrumentation Experts Teaching from Practical Experience Gas-Filled Detectors Gas-filled detectors measure the charge released when radiation interacts with the gas Three types: Ion Chambers,

More information

THE NEW NAGOYA RADIOCARBON LABORATORY

THE NEW NAGOYA RADIOCARBON LABORATORY THE NEW NAGOYA RADIOCARBON LABORATORY YASUSHI MURAKI, GRANT KOCHAROV,1 TOOR U NISHI YAMA, YUKIKO NAR USE, TAKUYA MURATA, KIMLAKI MASUDA and Kh. A. ARSLANOVZ Solar-Terrestrial Environment Laboratory, Nagoya

More information

ISO Water quality Strontium 90 and strontium 89 Test methods using liquid scintillation counting or proportional counting

ISO Water quality Strontium 90 and strontium 89 Test methods using liquid scintillation counting or proportional counting INTERNATIONAL STANDARD ISO 13160 First edition 2012-07-15 Water quality Strontium 90 and strontium 89 Test methods using liquid scintillation counting or proportional counting Qualité de l eau Strontium

More information

Primary Standardization of 152 Eu by 4πβ(LS) γ (NaI) coincidence counting and CIEMAT-NIST method

Primary Standardization of 152 Eu by 4πβ(LS) γ (NaI) coincidence counting and CIEMAT-NIST method Primary Standardization of 152 Eu by 4πβ(LS) γ (NaI) coincidence counting and CIEMAT-NIST method A Ruzzarin 1, P A L da Cruz 2, A L Ferreira Filho 2, A Iwahara 2 1 Laboratório de Instrumentação Nuclear/Programa

More information

The Norwegian Institute of Technology, Trondheim, Norway

The Norwegian Institute of Technology, Trondheim, Norway [RADIOCARBON, VOL 22, No. 2, 1980, P 470-478] AN ANALYSIS OF SHIELDING EFFICIENCY FOR 14C COUNTERS REIDAR NYDAL, STEINAR GULLIKSEN, and KNUT LOVSETH Radiological Dating Laboratory, The Norwegian Institute

More information

A Rapid Method for Measuring Strontium-90 Activity in Crops in China

A Rapid Method for Measuring Strontium-90 Activity in Crops in China A Rapid Method for Measuring Strontium-90 Activity in Crops in China Lingjing Pan Pan 1, Guobing Yu 2, Deyun Wen 2, Zhi Chen 1, Liusi Sheng 1, Chung-King Liu 1, and X. George Xu 1,3 1 University of Science

More information

Six decades of environmental radioactivity measurements. Sven P. Nielsen

Six decades of environmental radioactivity measurements. Sven P. Nielsen Six decades of environmental radioactivity measurements Sven P. Nielsen Research Establishment Risø Established by Government in 1950 s to keep Denmark in line with development of nuclear technology including

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement June 2008 Tom Lewellen Tkldog@u.washington.edu Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron

More information

The metrology of radioactivity. C. Michotte, BIPM

The metrology of radioactivity. C. Michotte, BIPM The metrology of radioactivity C. Michotte, BIPM SUMMARY Introduction Nuclear data of interest in activity measurements Activity measurement methods (primary and secondary) Monte Carlo simulations The

More information

LIQUID SCINTILLATION

LIQUID SCINTILLATION LIQUID SCINTILLATION - MEASURING PROCEDURES, NEW DEVELOPMENTS - DGFS e.v. German Society for Liquid Scintillation Spectrometry Karlsruhe Institute of Technology, 2015 Karlsruhe MOEBIUS Siegurd ERAT Stephan

More information

Determination of the activity of radionuclides

Determination of the activity of radionuclides BUREAU NATIONAL DE MÉTROLOGIE COMMISSARIAT À L'ÉNERGIE ATOMIQUE LABORATOIRE NATIONAL HENRI BECQUEREL Note technique LNHB/04-33 Determination of the activity of radionuclides contained in volume samples

More information

DETERMINATION OF NATURAL RADIONUCLIDES IN WATER FROM SLOVAKIA USING LSC

DETERMINATION OF NATURAL RADIONUCLIDES IN WATER FROM SLOVAKIA USING LSC DETERMINATION OF NATURAL RADIONUCLIDES IN WATER FROM SLOVAKIA USING LSC Alena Belanová 1 Jana Merešová Marta Vršková Water Research Institute, Nábr. L. Svobodu 5, 812 49 Bratislava, Slovakia. ABSTRACT.

More information

- \/(o-'(n)l 2. ::z.' % modern (1) N Q- n1 + o- no,..

- \/(o-'(n)l 2. ::z.' % modern (1) N Q- n1 + o- no,.. [RADIOCARBON, VOL 22, No. 2, 1980, P 417-427] LARGE-VOLUME LIQUID SCINTILLATION COUNTING OF CARBON-14 LORENZ EICHINGER, WERNER RAUERT, JOSEF SALVAMOSER, and MANFRED WOLF Institut fur Radiohydrometrie der

More information

Application Note. Abstract. Introduction. Experimental

Application Note. Abstract. Introduction. Experimental Application Note Alpha/Beta ABA-004 Quantitation of Transuranium Elements in High Activity Waste (HAW) Using Alpha/Beta Liquid Scintillation Counting and Extractive Scintillators by Jim Floeckher Abstract

More information

LSC for Quality Control of 99m TC Eluate from 99 Mo- 99m Tc Generator

LSC for Quality Control of 99m TC Eluate from 99 Mo- 99m Tc Generator LSC2017 Conference 1-5th May, 2017, Copenhagen LSC for Quality Control of 99m TC Eluate from 99 Mo- 99m Tc Generator Xiaolin Hou Technical University of Denmark, Center for Nuclear Technologies Roskilde,

More information

The detection of low energy, beta-emitting radionuclides can be

The detection of low energy, beta-emitting radionuclides can be RADIOACTIVE FLOW DETECTORS LIQUID OR SOLID SCINTILLATORS? Andrew R. Reich, Radiornatic Instruments & Chemical Co., Inc., Tampa, Florida 33611, U.S.A. The detection of low energy, beta-emitting radionuclides

More information