The certification of the mass fraction of thyroxine in a CRM intended for calibration. Certified Reference Material IRMM-468

Size: px
Start display at page:

Download "The certification of the mass fraction of thyroxine in a CRM intended for calibration. Certified Reference Material IRMM-468"

Transcription

1 The certification of the mass fraction of thyroxine in a CRM intended for calibration Certified Reference Material IRMM-468 B. Toussaint, C.L. Klein, M. Wiergowski Report EUR EN

2 The mission of IRMM is to promote a common and reliable European measurement system in support of EU policies. European Commission Directorate-General Joint Research Centre Institute for Reference Materials and Measurements Contact information European Commission Directorate-General Joint Research Centre Institute for Reference Materials and Measurements Retieseweg 111 B-2440 Geel Belgium Tel.: +32 (0) Fax: +32 (0) Legal Notice Neither the European Commission nor any person acting on behalf of the Commission is responsible for the use which might be made of the following information. A great deal of additional information on the European Union is available on the Internet. It can be accessed through the Europa server EUR Report EN Luxembourg: Office for Official Publications of the European Communities ISBN European Communities, 2006 Reproduction is authorised provided the source is acknowledged Printed in Belgium

3 European Commission IRMM information REFERENCE MATERIALS The certification of the mass fraction of thyroxine in a CRM intended for calibration Certified Reference Material IRMM-468 B. Toussaint, C.L. Klein, M. Wiergowski European Commission Directorate General Joint Research Centre Institute for Reference Materials and Measurements Reference Materials Unit Retieseweg 111 BE-2440 Geel, Belgium Report EUR EN

4 SUMMARY This report describes the preparation, homogeneity, short-term and long-term stability and characterisation studies of the certified reference material IRMM-468, consisting of thyroxine certified for its purity. A description of the analytical procedures used and all relevant data from the certification studies are presented. The certified mass fraction (expressed on dry mass basis) is: Substance Certified mass fraction 1) Uncertainty 2) thyroxine ) The certified value is the purity after taking into consideration inorganic residues, water, ethanol and organic impurities detectable by HPLC-UV and HPLC-MS. The certified value is traceable to the International System of Units (SI). 2) The certified uncertainty is the expanded uncertainty estimated in accordance with the Guide to the Expression of Uncertainty in Measurement (GUM) [1]. It is expressed with a coverage factor k = 2, corresponding to a level of confidence of about 95%. 1

5 2

6 TABLE OF CONTENTS SUMMARY... 1 TABLE OF CONTENTS... 3 GLOSSARY INTRODUCTION BACKGROUND: NEED FOR THE CRM CHOICE OF THE MATERIAL TO BE CERTIFIED DESIGN OF THE PROJECT PARTICIPANTS STABILITY AND HOMOGENEITY STUDIES CHARACTERISATION MEASUREMENTS PREPARATION OF THE MATERIAL BOTTLING DISPATCHING OF SAMPLES TESTING OF THE MATERIAL HOMOGENEITY STUDY STABILITY STUDY Short-term stability study Long-term stability study CHARACTERISATION MEASUREMENTS CALIBRANTS ANALYTICAL BLANKS DETERMINATION OF THE IMPURITIES DETECTABLE BY HPLC HPLC-UV methods for purity assessment HPLC-UV-MS method for purity assessment LC-MS methods for purity assessment Impact of the purity of the calibrants on the value assignment of T 4 mass fraction in Laboratories B, C and D DETERMINATION OF THE WATER MASS FRACTION DETERMINATION OF THE ETHANOL MASS FRACTION Method Results DETERMINATION OF THE RESIDUAL ASH MASS FRACTION Experimental Results and discussion CALCULATION OF THE CERTIFIED VALUE FOR TOTAL T 4 MASS FRACTION CALCULATION OF THE INDICATIVE VALUE FOR T 3 MASS FRACTION EVALUATION OF UNCERTAINTIES AND CERTIFIED VALUES UNCERTAINTY EVALUATION Conceptual considerations Uncertainty source homogeneity Uncertainty source stability Uncertainty source batch characterisation Uncertainty source unknown impurity Uncertainty budget CERTIFIED VALUES METROLOGICAL TRACEABILITY

7 8. INSTRUCTIONS FOR USE TRANSPORT AND STORAGE MINIMUM SAMPLE INTAKE HYGROSCOPICITY DISSOLUTION USE OF THE CERTIFIED VALUES REFERENCES ACKNOWLEDGEMENTS ANNEX A - SHORT-TERM STABILITY STUDY: EXPERIMENTAL CONDITIONS AND EXAMPLES OF HPLC-UV CHROMATOGRAMS ANNEX B SHORT-TERM STABILITY STUDY: RAW DATA OF HPLC-UV ANALYSIS OF IRMM- 468 (2 SAMPLES, 3 REPLICATES PER SAMPLE) ANNEX C LONG-TERM STABILITY STUDY: EXPERIMENTAL CONDITIONS AND EXAMPLES OF HPLC-UV CHROMATOGRAMS ANNEX D LONG-TERM STABILITY STUDY: RAW DATA OF HPLC-UV ANALYSIS OF IRMM ANNEX E DETERMINATION OF THE IMPURITIES BY HPLC IN IRMM ANNEX F - RESULTS OF THE WATER MASS FRACTION DETERMINATION OF IRMM-468 MATERIAL BY COULOMETRIC KARL FISCHER TITRATION (LABORATORY C) ANNEX G DETERMINATION OF THE ETHANOL MASS FRACTION OF IRMM-468 (LABORATORY F)

8 GLOSSARY ANOVA a w b BAM CKFT CRM DAD DIT Analysis of variance water activity slope of the regression line Bundesanstalt für Materialforschung und prüfung (DE) Coulometric Karl Fischer Titration Certified Reference Material Photodiode Array Detector 3,5-diiodotyrosine ESI+ Electrospray Ionisation (positive mode) ε molar absorptivity F Snedecor F F crit critical value of Snedecor F Fig. figure FT 3 Free 3,3,5-triiodothyronine FT 4 Free thyroxine FT-IR Fourier Transform-Infra Red GC Gas Chromatography HPLC High Performance Liquid Chromatography ISO International Organization for Standardization IRMM Institute for Reference Materials and Measurements IVD In-Vitro Diagnostic k coverage factor LC Liquid Chromatography LGC LGC Limited (GB) LOD Limit of Detection LOQ Limit of Quantification m ep endpoint mass m s mass of the sample MIT 3-iodotyrosine (mono-iodotyrosine) MS Mass Spectrometry MS between Mean of Squares between groups MS q Mean of Squares MS within m/z n N ע NIST NMR Mean of Squares within groups mass to charge number ratio number of replicates number of vials degree of freedom National Institute of Standards and Technology (US) Nuclear Magnetic Resonance PTB Physikalisch-Technische Bundesanstalt (DE) PTFE PolyTetraFluoroEthylene R² coefficient of determination RM Reference Material RP Reversed-Phase RSD Relative Standard Deviation RT Retention Time s standard deviation s bb random variation between units SI Système International d Unités SIM Single Ion Monitoring S/N Signal to Noise ratio SS Sum of Squares t 0.05, ע T 2 T 3 T 4 T 4 w HPLC t of Student test at 95 % of confidence and ע degree of freedom 3,5-diiodothyronine 3,3,5-triiodothyronine thyroxine or 3,3,5,5 -tetraiodothyronine T 4 mass fraction as determined by HPLC T 4 w* HPLC T 4 mass fraction as determined by HPLC and corrected for the impurities which were detected in HPLC by only some of the laboratories T 4 w total total T 4 mass fraction taking into account the impurities detectable in HPLC as well as the ethanol, water and residual ash. TETRAC 3,3,5,5 -tetraiodothyroacetic acid 5

9 Temp. TRIAC TSH TSP temperature 3,3,5-triiodothyroacetic acid thyrotropin (thyroid stimulating hormone) Trimethylsilyl-tetradeuteropropionic acid sodium salt TT 3 Total concentration of 3,3,5- triiodothyronine in blood, equal to the sum of the free form and the proteinbound form of 3,3,5-triiodothyronine TT 4 Total concentration of thyroxine in blood, equal to the sum of the free form and the protein-bound form of thyroxine u b standard uncertainty of the slope u bb relative uncertainty component from homogeneity u * bb u c relative uncertainty of the maximum hidden between bottle heterogeneity combined standard uncertainty of the certified value and the measurement result u char relative uncertainty component from batch characterisation u CRM standard uncertainty of the certified value U CRM expanded uncertainty of the certified value expressed with a coverage factor k = 2, corresponding to a level of confidence of about 95% u lts relative uncertainty component from long-term stability u m measurement standard uncertainty of the result u storage 4 C relative uncertainty component from storage at 4 C u sts relative uncertainty component from short-term stability u unknown impurity relative uncertainty component from unknown impurity UV Ultra Violet X average 6

10 1. INTRODUCTION 1.1 Background: need for the CRM Thyrotropin (TSH), thyroxine (3,3,5,5 -tetraiodothyronine or T 4 ) and 3,3,5-triiodothyronine (T 3 ) play a major role in the diagnosis and monitoring of thyroid diseases. T 4 and T 3 are the primary hormones secreted by the thyroid gland. Their secretion is regulated by feedback from the hypothalamus and TSH from the pituitary gland. They are of major importance in the growth, development and sexual maturation of the human body and increase the energy expenditure in tissues. They are also involved in the regulation of the heart rhythm, body mass and metabolism. Therefore, their analysis is very frequently requested in endocrinology. In the blood, T 3 and T 4 occur in a free form (FT 3, FT 4 ) and in a protein-bound form. The sum of bound and free hormones is measured as the total hormone concentration, i.e. TT 3, TT 4. Typical physiological concentrations of TT 4 and TT 3 in adults are nmol/l and nmol/l, respectively [2]. The values of TT 4 and FT 4 concentrations in serum give an indication of the functional state of the thyroid gland. Hypothyroidism (thyroid gland deficiency) or hyperthyroidism (excessive hormone secretion) can be diagnosed and abnormal findings on physical examination of the thyroid gland can be interpreted after determination of the TT 4 and FT 4 concentrations. In order to achieve comparability of results from different analytical procedures and commercial diagnostic kits, standardisation is required using reference procedures and Certified Reference Materials (CRMs) as well as quality control materials. The Directive 98/79/EC is especially aiming to ensure the quality of In-Vitro Diagnostic (IVD) medical devices through the use of higher order reference materials and methods for calibration. In this context, a new higher order reference material for TT 4 and FT 4, IRMM-468, has been produced. It can be used by IVD manufacturers and reference laboratories for calibration of measurements procedures based on chromatographic techniques. The structure of thyroxine is shown in Figure 1. THYROXINE [ ] Figure 1 - Structure of thyroxine and CAS-number 7

11 1.2 Choice of the material to be certified From three commercially available materials the thyroxine provided by Acros Organics was selected because it appeared to be the most pure using HPLC-MS, HPLC-UV and GC-MS in scan mode. Thyroxine crystalline powder was obtained from Acros Organics (Geel, Belgium; product number: 21566, lot number: A ) with a claimed purity > 98.5 % (HPLC purity according to Acros Organics). 1.3 Design of the project The thyroxine material was aliquoted at IRMM in 100 mg units. The material was tested for homogeneity and stability (short- and long-term) and was characterised (identification and purity assessment) in five different laboratories with independent methods focusing on most likely occurring impurities. The characterisation was achieved in two steps: First, the peak area fraction in percent of T 4 and the impurities detectable by HPLC were determined by HPLC-UV, HPLC-UV-MS and HPLC-MS. The T 4 mass fraction was calculated in each laboratory by comparing the T 4 peak area to the peak areas of the impurities detectable by HPLC. The sum of all peak areas was considered as 100 %. Two approaches were used in order to calculate T 4 mass fraction from the HPLC results: 1. In the long-term stability study as well as in two laboratories participating to the characterisation study, the peak area fraction in percent of T 4 and the impurities were assumed to correspond to the respective mass fraction in percent. The mass fraction of T 4 as determined by HPLC (T 4 w HPLC ) was calculated as : T 4 w HPLC (T 4 peak area / Sum of peak areas for T 4 and HPLC-detectable impurities) x 100 % Note: For the measurements performed using UV detection, the Beer-Lambert s law should be considered: A = ε l c where ε is the molar absorption coefficient (L/(mol cm)); l is the path length (cm) and c is the concentration (mol/l). Therefore, a correction factor taking into account the molar mass of the molecules should be applied to estimate mass fractions from peak area fractions determined in UV. However in practice, the difference in peak areas before and after application of the correction factor is lower than the uncertainty of the certified value for T 4 mass fraction. The impact of the molar absorptivity on the calculation of T 4 mass fraction in IRMM-468 can be considered as negligible. 8

12 2. In the homogeneity and short-term stability studies as well as in three laboratories participating to the characterisation study, the measuring system for known impurities was calibrated using commercially available calibrants and was related to the mass of IRMM-468 analysed. The mass fraction of T 4 as determined by HPLC (T 4 w HPLC ) was calculated as: T 4 w HPLC = (100 % Sum of mass fractions of impurities) Note: The use of calibration curves to quantify the HPLC-detectable impurities in IRMM-468 implies to take into account the purity of the calibrants. However, as no reference material was available to calibrate the measuring system for the known impurities in IRMM-468, commercially available calibrants with a described purity 95 % were used. As described in Section 5.3.4, due to the very low mass fraction of the impurities determined in HPLC in IRMM-468, the impact of the calibrants purity on the determination of T 4 mass fraction was found to be negligible compared to the uncertainty of the the certified value. Both approaches used in the characterisation study of IRMM-468 provided similar results. The final value is the average of the averages of the mass fractions of T 4 determined by HPLC in the five laboratories. Second, the mass fractions of impurities non detectable by HPLC, i.e. ethanol, water and residual ash, were determined by NMR, CKFT and ashing, respectively. The mass fractions of these impurities were then substracted from the average T 4 mass fraction determined by HPLC to lead to the certified T 4 mass fraction in IRMM-468 (T 4 w total ). Studies and investigations were mainly performed in the frame of the Shared Cost Action IVD Thyroid G6RD-CT

13 2. PARTICIPANTS 2.1 Stability and homogeneity studies - Institute for Reference Materials and Measurements (IRMM), Geel, BE 2.2 Characterisation measurements The laboratories who participated to the characterisation study of IRMM-468 are: - Institute for Reference Materials and Measurements (IRMM), Geel, BE: two laboratories. - Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, DE. - National Institute of Standards and Technology (NIST), Gaithersburg, US. - LGC Limited (LGC), Teddington, GB. - Bundesanstalt für Materialforschung und prüfung (BAM), Berlin, DE. In order to preserve the anonymity of the laboratories, they will be represented by a code, between A and F, in the certification report. The characterisation of IRMM-468 consisted of four different determinations: - The determination of impurities detectable by HPLC was performed by laboratory A, B, C, D and E. - The water mass fraction determination was carried out by laboratory B. - The ethanol mass fraction determination was achieved by laboratory F. - The residual ash determination was performed by laboratory E. 10

14 3. PREPARATION OF THE MATERIAL 3.1 Bottling The bottling was done at IRMM. A total number of 917 units were produced. Each unit consisted of an amber glass vial with approximately 100 mg of T 4. The bottling was carried out in a glove box under nitrogen gas. The vials were closed using a polytetrafluoroethylene (PTFE) coated rubber stopper. The outside surface of the vials was cleaned with a soft cloth. The vials were sealed using an aluminium cap. The main stock is stored between -40 and -70 C. 3.2 Dispatching of samples Samples for homogeneity and stability studies and for characterisation measurements were dispatched to participating laboratories using a courier service. The samples were shipped at 4 C and delivered within 24 hours. 11

15 4. TESTING OF THE MATERIAL 4.1 Homogeneity study The design of the short-term stability study enabled also the assessment of the homogeneity of the material. Considering the poor stability of IRMM-468 at 60 C (see Section ), only the data obtained at -70, -20 and 4 C were used for this assessment. In total, ten samples were analysed in triplicate using HPLC-UV. The sample numbers were randomly spread over the whole batch. A one-way ANOVA was carried out grouping the data by sample number (group = sample number). The results presented in Table 1 and expressed in mass fractions show that no significant difference was observed between the samples at temperatures where the material is stable (F < F crit ). Table 1. One-way ANOVA for the homogeneity study of IRMM-468 (number of replicates = 3). SUMMARY Temperature Storage time Group (=sample number) Average mass fraction Variance of the mass fraction [ C] [weeks] [% ] [%²] Average ( X ) ANOVA Source of Variation SS ע MS q F F crit Between Groups Within Groups Total

16 As MS between (0.04) < MS within (0.06), the random variation between units (s bb ) can not be estimated properly. Therefore, the maximum hidden between bottle heterogeneity (u* bb ) was evaluated as follows: u MS * within = 4 bb n N 1 2 ( n ) where N = number of vials, n = number of replicates. A u* bb of % was obtained. The individual sample averages were calculated as well as the average of all samples (99.54 %) (Table 1). With the exception of sample n 491 (-20 C, 2 weeks) where only two replicates were performed (n =2) due to experimental error in the laboratory, the individual averages of the other samples were found to be included in the interval determined by: average of all samples ± distance from value to limit (k = 2) which corresponds to X ± U CRM (k = 2) The material can thus be considered to be homogeneous at the level of the sample intake used in the HPLC-UV procedure for homogeneity testing of the material. The sample intake in that procedure being 10 mg, the material can be considered to be homogeneous at 10 mg sample intake level. 4.2 Stability study Short-term and long-term stability studies were performed. Both studies were carried out at IRMM following an isochronous set-up [3] that consists of the simultaneous analysis of reference and test samples. A defined number of samples was exposed for different periods of time to given temperatures, then brought back to the reference temperature of 70 C where no impact on the analyte content over a long time period is expected. At the end of the study, all samples were analysed for their impurities contents in one experimental run. The IRMM-468 samples were analysed by methods employing high performance liquid chromatography coupled to an ultra-violet detector (HPLC-UV). Very similar methods were used for short-term and long-term stability studies. The experimental conditions are summarised in Table 2. More details about the instrumentation, solution preparation, Limits of Detection and Quantification (LODs and LOQs) and calibration curves are presented in Annexes A and C for the short- and long-term stability, respectively. 13

17 Table 2. HPLC-UV experimental conditions for the stability study of IRMM-468 Stability study Short-term [4] Long-term HPLC column HPLC mobile phases HPLC gradient - C18 Supelco Discovery, 250 x 4.6 mm, 5 µm - Supelguard Discovery 18 guard column 20 x 4.0 mm, 5 µm A = 0.1 % diluted phosphoric acid/ acetonitrile (65+35; volume fractions) B = demineralised water Gradient: from 70 % of A and 30 % of B to 100 % of A until 5 min, held till 30 min. Flow rate = 1.0 ml/min for 5 min; 1.5 ml/min for 25 min. idem idem Temperature 25 C idem Injection volume 100 µl idem Detection 225 nm idem Gradient: from 50 % of A and 50 % of B to 100 % of A until 5 min, held till 17 min. Flow rate = 1.0 ml/min for 5 min; 1.5 ml/min for 17 min Short-term stability study Design of the short-term stability study The design of the short-term stability study is illustrated in Table 3. The study was conducted over a period of 4 weeks at 4 different temperatures (reference temperature = -70 C). Two IRMM-468 samples were analysed per time point and temperature. Each sample was analysed in triplicate. Expected organic impurities are 3-monoiodotyrosine (MIT), 3,5-diiodotyrosine (DIT), 3,5- diiodothyronine (T 2 ) and 3,3,5-triiodothyronine (T 3 ). A calibration curve was established for each impurity using commercially available calibrants from Sigma-Aldrich. The stability of IRMM-468 samples was expressed in terms of mass fraction of thyroxine (T 4 ) calculated as the difference between 100 % and the sum of the mass fractions of the impurities. Only impurities giving a signal higher than the LOQ were taken into account for the quantification of T 4. 14

18 Table 3. Sample numbers of IRMM-468 analysed in short-term stability study Sample number Temperature [ C] Time [weeks] (reference) (reference) Results of the short-term stability study Examples of HPLC chromatograms of a blank sample (methanol) (Fig. 1), calibrant mixture with 200 mg/l of each MIT, DIT, T 2, T 3 and T 4 (Figure 2) and IRMM-468 (Figure 3) sample are shown in Annex A. LODs and LOQs are presented in Annex A. They were estimated using calibration curves. No LOQ was determined for DIT. However, taking into account the similarity between DIT and T 2, the same LOQ can be used for DIT without reasonable doubt. As LODs and LOQs were given in terms of mass concentrations in mg/l, the peak area fractions of the impurities measured in IRMM-468, assumed to correspond to the mass fractions of these impurities, were converted into mass concentrations in mg/l for evaluation. Taking into account that the short-term stability was determined on IRMM-468 solutions at 200 mg/l, the traces of the different impurities expressed as mass fraction percentages were translated into mass concentrations using the following equation: Impurity concentration [mg/l] = impurity mass fraction 200 [mg/l] No MIT could be detected in IRMM-468. Traces of DIT could occasionally be detected at levels above the LOQ. T 2 was detected at levels above LOQ, between 2.5 and 5.1 %, in the samples n 473 and 482 stored at 60 C for 2 and 4 weeks, respectively. However no T 2 was detected in the samples n 23 and 32 also stored at 60 C for 2 and 4 weeks, respectively. 15

19 T 3 was detected as a constant impurity in IRMM-468 at levels above the LOQ. The mass fractions of DIT, T 2 and T 3 were calculated using the ratio of the peak areas obtained for the respective impurities in IRMM-468 and in the calibration curve. This was related to the exact initial mass of IRMM-468, used in the preparation of the sample solution, to finally yield the mass fraction of the impurity in IRMM-468. The mass fraction of T 4 was obtained as the difference between 100 % and the sum of DIT, T 2 and T 3 mass fractions. The mass fractions of T 4 in IRMM-468 short-term stability samples are summarised in Annex B. Data were evaluated with respect to the repeatability of the measurements within a sample (3 different readings of every sample) and to the repeatability of the measurements between the different samples from different temperature/time conditions. As illustrated in Figure 2, a regression line was calculated at each tested temperature by plotting T 4 mass fraction versus the storage time. Figure 2. Regression lines obtained at different storage temperatures for short-term stability study of IRMM-468 A) -20 C 100 T4 mass fraction y = 0.017x R 2 = Time [weeks] 16

20 B) 4 C 100 T4 mass fraction y = 0.038x R 2 = Time [weeks] C) 60 C 100 T4 mass fraction y = x R 2 = Time [weeks] The slope (b) of the regression line was tested for significance using the t-test. The slope is not significantly different from zero if the ratio of the slope and the uncertainty of the slope (u b ) is lower than the t-value at 95 % of confidence. Although a degradation of some samples at 60 C is obvious (see Figure 2 C), all slopes obtained at the different temperatures (Table 4) are not significantly different from zero (b/u b < t ע, 0.05 ). Due to the large scatter of results of the samples stored at 60 C, the uncertainty of 17

21 the slope observed is high, thus not leading to a significant outcome of the t-test. The observed repeatability of samples stored at -20 and 4 C corresponds to the expected repeatability of the HPLC measurements. Table 4. Test for significance of the slope: slopes (b), uncertainties (u b ) and b/u b of shortterm stability study for IRMM-468 Temperature b [ C] [week -1 ] Short-term stability study u b [week -1 ] b/u b ע, 0.05 t Thus, the stability of IRMM-468 could be demonstrated for a period of 4 weeks at 4 and -20 C whereas at 60 C the material shows clear signs of degradation. Therefore, the cooled dispatch of IRMM-468 samples at 4 C is required Long-term stability study Design of the long-term stability study The long-term stability study was conducted over a period of 12 months at 3 different temperatures (reference temperature = -70 C) as described in Table 5. Two samples were analysed per time point and temperature. Each sample was analysed in triplicate. The presence of MIT, T 2 and T 3 impurities was investigated. No calibration curve was established. The purity of IRMM-468 in long-term stability samples was determined by considering the sum of peak areas of T 4 and the impurities detected in HPLC-UV as 100 %. The peak area fractions of T 4 and the impurities were assumed to correspond to the respective mass fractions. The final results of the stability study were evaluated by regression analysis plotting the individual results against the storage time at the test temperature. 18

22 Table 5. Sample numbers of IRMM-468 analysed in long-term stability study Sample number Temperature [ C] Time [months] (reference) (reference) Results of the long-term stability study Annex C shows HPLC chromatograms of the blank sample (dissolution solvent used for the calibration solutions) (Figure 1), calibrant mixture with 50 mg/l of MIT, T 2, T 3 and T 4 (Figure 2) and IRMM-468 (Figure 3) sample. The purity of IRMM-468 material was calculated only taking impurities above the LOQ into consideration. The LOQ was determined as the mass fraction giving a signal-to-noise ratio (S/N) equal to 10 (S/N=10). LODs and LOQs of MIT, T 2, T 3 and T 4 are presented in Annex C. T 3 impurity was quantified in IRMM-468 at levels equal or higher than the LOQ. No MIT and no T 2 were detected. The mass fractions of T 4 in IRMM-468 samples are presented in Annex D. A one-way ANOVA and a trend analysis were performed on these results. The ANOVA of the data, comparing the results grouped by temperature, is presented in Table 6. No significant difference between the T 4 mass fractions in the samples stored at the reference temperature (-70 C) and the test temperatures -20 C and 4 C, respectively, was observed (F < F crit ). 19

23 Table 6. One-way ANOVA of the long-term stability data for IRMM-468 SUMMARY Groups (= storage temperature) Number of replicates Average mass fraction Variance of the mass fraction [ C] [%²] ANOVA Source of Variation SS ע MS q F F crit Between Groups Within Groups Total As illustrated in Figure 3, a trend analysis was performed by plotting the T 4 mass fraction of IRMM-468 samples versus storage time and by calculating a regression line. Figure 3. Regression lines obtained at different storage temperatures for long-term stability study of IRMM-468 (2 samples per time, 3 replicates per sample) A) -20 C 100 T 4 mass fraction y = x R 2 = Time [months] 20

24 B) 4 C 100 T 4 mass fraction y = x R 2 = Time [months] The slope of the regression line was then tested for significance. Results are presented in Table 7. Table 7. Test for significance of the slope: slopes (b), uncertainties (u b ) and b/u b of long-term stability study for IRMM-468 Long-term stability study Temperature [ C] b [months -1 ] u b b/u b ע, 0.05 t [months -1 ] No significant trend was observed at -20 C over 12 months indicating a good stability of IRMM-468 at this temperature. A shelf-life of 3 years was determined with u lts set at 0.2 %. However, at 4 C, a significant slope was obtained over time showing a degradation of IRMM-468 between 6 and 12 months. No significant slope was observed from the data between 0 and 6 months at 4 C. In order to maintain a long-term stability of the main batch, the IRMM-468 samples will be stored at IRMM between -40 and -70 C. As the validity of the certificate is 6 months after purchase, the vial will be stored between -20 and -70 C on receipt by the customer. Taking into account the intermediate storage of IRMM-468 samples at 4 C during the certification period (25 months), a standard uncertainty contribution (0.14 %) corresponding to the period of storage at 4 C was added to the uncertainty budget (Table 22). 21

25 5. CHARACTERISATION MEASUREMENTS As indicated under Section 2.2, the characterisation aiming at the detection of impurities by HPLC was performed through an intercomparison between five laboratories. Three different characterisation techniques were used: HPLC-UV in laboratory A, B and C; HPLC-UV-MS in laboratory D; HPLC-MS in laboratory E. As for the stability study, IRMM-468 was characterised in terms of purity (peak areafraction). In addition, the water, ethanol and residual ash mass fractions, respectively, of IRMM-468 were determined by three laboratories: Laboratory C, E and F. The analytical methods used for the characterisation are described in detail in Section 5.3. Each laboratory used its own optimised procedure for sample preparation, method of injection, chromatographic separation and detection. All procedures and measurements were recorded and reported by each laboratory. A minimum of 3 replicates per sample (in total 6 samples) was analysed for the HPLC characterisation. 5.1 Calibrants Each laboratory prepared separate calibration solutions from commercially available DIT, MIT, T 2, T 3 and T 4. The calibrants were from Sigma-Aldrich, Fluka and Acros. Their purity was 95 % (Table 8).The calibrants were used to identify T 4 and the impurities. They were also used in the characterisation study to build calibration curves in order to check the linearity of the detector response and to estimate LODs and LOQs. In some laboratories, the peak area fractions obtained for the respective impurities in IRMM-468 and in the calibration curves were compared in order to estimate the mass fractions of the impurities in IRMM-468. The impact of using calibrants, which purity ( 95 %) is not certified, on the mass fraction of T 4 has been evaluated and was found to be negligible (see Section 5.3.4). Table 8. Purity of the calibrants used for the certification of IRMM-468, as specified by the supplier Sigma-Aldrich Fluka Acros T4 98 > 98.5 T3 95 T2 > DIT > 97.5 MIT > Analytical blanks Analytical blanks (consisting of HPLC mobile phase injections) were performed. All solvents used were LC grade. 22

26 5.3 Determination of the impurities detectable by HPLC This section presents the results obtained in each of the five laboratories. The experimental conditions of the different methods are summarised in Table 9. Additional information about the instrumentation used and LOD/LOQ is given in Annex E. As described in Section 1.2, the T 4 mass fraction was calculated in each laboratory by comparing the T 4 peak area to the sum of the peak areas of the impurities detectable by HPLC. The sum of all peak areas, including that of T 4, was considered as 100 %. Then two approaches were used in order to calculate T 4 mass fraction from the HPLC results: 1. In laboratories A and E, the peak area fractions of T 4 and of the impurities were assumed to correspond to the respective mass fractions in %. The mass fraction of T 4 as determined by HPLC (T 4 w HPLC ) was calculated as : T 4 w HPLC (T 4 peak area / Sum of peak areas for T 4 and HPLC-detectable impurities) x 100 %. The negligible influence of the different molar absorptivity of the studied compounds on the determination of T 4 mass fraction by HPLC-UV in Laboratory A is described in Section In laboratories B, C and D, the measuring system for known impurities was calibrated using commercially available calibrants and was related to the mass of IRMM-468 analysed. The mass fraction of T 4 as determined by HPLC (T 4 w HPLC ) was calculated as: T 4 w HPLC = (100 % Sum of mass fractions of impurities). The negligible impact of the calibrants purity on the determination of T 4 mass fraction by Laboratory B, C and D is demonstrated in Section For technical reasons, some laboratories could not investigate the presence of MIT and/or DIT impurities in IRMM-468. In these cases, the average value of MIT and/or DIT mass fraction obtained by the other laboratories was substracted from the T 4 mass fraction determined by HPLC. The resulting corrected average T 4 mass fractions (T 4 w* HPLC ) are given in Table 17. For laboratories which were able to determine MIT and DIT, the values provided by each laboratory were substracted from the T 4 mass fraction determined by HPLC, respectively. The average of the HPLC results obtained in each laboratory (average T 4 w HPLC ) was calculated and is presented in Table 17. Then the water, ethanol and residual ash mass fractions of IRMM-468 were determined as described under Sections 5.4, 5.5 and 5.6, respectively. Finally, the combined standard uncertainty of the result was calculated and will be described in Section 6. 23

27 Table 9. HPLC measurement conditions Laboratory Method HPLC column HPLC mobile phases A HPLC-UV Hypersil C x 2.1 mm B HPLC-UV C18 ProntoSIL analytical column 250 x 4.6 mm, 5 µm [volume fraction] 50 % Methanol - 50 % diluted formic acid (0.1 %; v/v) 70 % Methanol 30 % diluted phosphoric acid (0.1 %) Elution conditions Isocratic; flow rate = 0.2 ml/min Isocratic; flow rate = 0.7 ml/min Temperature [ C] Injection volume [µl] nm nm Detection C HPLC-UV - C18 Supelco Discovery, 250 x 4.6 mm, 5 µm - Guard column Discovery 20 x 4.0 mm, 5 µm A = 0.1 % diluted phosphoric acid / acetonitrile (65+35) B = demineralised water Gradient [volume fractions]: from 70 % of A and 30 % of B to 100 % of A until 5 min, held till 30 min Flow rate = 1.0 ml/min for 5 min; 1.5 ml/min for 25 min nm D HPLC-UV- MS Phenomenex Hypersil 3 C18, 150 x 2 mm, 3 µm E HPLC-MS Zorbax Eclipse XDB- C18 column 2.1 x 150 mm 20 % diluted formic acid (0.1 %) 80 % methanol containing 0.1 % formic acid A = 0.1 % diluted acetic acid B = 0.1 % acetic acid in methanol Isocratic; flow-rate = 0.2 ml/min Gradient [volume fractions]: 0 min = 35 % B; 5 min = 40 % B; 10 min = 55 % B; 35 min = 55 % B; 40 min = 90 % B; 45 min = 35 % B Flow rate = 0.25 ml/min UV: 254 nm - ESI+/MS* in SIM mode 25 5 ESI+/MS* in SIM mode 24

28 HPLC-UV methods for purity assessment Procedure applied by laboratory A One sample of IRMM-468 was analysed in triplicate. Spectra taken at the retention time of the chromatographic peaks and at different retention times were further examined for any evidence of impurities at other wavelengths. The chromatograms obtained for the T4 material in mobile phase were compared to the chromatograms obtained for the standard solution of commercially available T4 (Sigma). The spectrum obtained at the retention time of T4 in the chromatogram obtained for IRMM-468 was compared to the scan made by UV spectrophotometer. Additional qualitative purity profiling of IRMM-468 samples was performed by HPLC-ESI- MS. Reconstructed ion chromatograms of IRMM-468 showed ions corresponding to [M+H] + ions of T 2 (mass to charge ratio, m/z = 536), T 3 (m/z = 652) and T 4 (m/z = 778). Finally, the identity of T 4 in IRMM-468 was confirmed using Nuclear Magnetic Resonance (NMR): 1 H-NMR and 13 C-NMR spectroscopy. The purity of the sample was calculated from the peak areas obtained in UV and assuming that the peak area fractions correspond to the mass fractions i.e.: T 4 w HPLC (T 4 peak area / Sum of peak areas) 100 % The quantitative results are presented in Table 10. Table 10. Laboratory A: Mass fraction of T 4 in IRMM-468, expressed as an area fraction (1 sample, 3 replicates) Replicate number T 4 area fraction Average s

29 Note: As Laboratory A did not investigated MIT and DIT impurities, the average mass fraction of these two impurities obtained in the other laboratories (0.08 and %, respectively) was considered. The resulting T 4 mass fraction presented in Table 17 was %. Because of the Beer-Lambert s law: A = ε l c where A is the absorption; ε is the molar absorption coefficient (L/(mol cm)); l = path length (cm); c = concentration (mol/l) a correction factor, taking into account the molar mass of the compound, should be applied to estimate mass fractions from peak area fractions. This correction factor can be expressed as follows: peak area fraction of the impurity x molar mass of the impurity [kg/mol] molar mass of T 4 [kg/mol] The mass fraction of T 4 was re-calculated using the corrected mass fractions of the impurities. The resulting T 4 mass fraction was %. The difference between the corrected and the non-corrected T 4 mass fraction is 0.12 % and is much lower than the uncertainty of the certified value for T 4 mass fraction 0.7 % determined in section 6. Therefore, the influence of the Beer-Lambert s law on the determination of T 4 mass fraction by HPLC-UV can be considered as negligible Procedures applied by laboratory B One IRMM-468 sample was used to prepare one stock solution. From the stock solution, 5 working solutions were prepared and analysed in triplicate. The 15 results were reported as a single result (average of the 15 injections). The whole procedure was repeated on three days, using new working solutions on each day. The 3 averaged results are summarised in Table 11. The presence of T 3, T 2 and MIT impurities in IRMM-468 was investigated in UV and MS detection. However, for their quantification, only individual UV peak areas were determined. Calibration curves were established in UV detection for T 3, T 2, MIT and as an additional information for T 4 using commercially available standards. The mass concentrations of T 3, T 2 and MIT in IRMM-468 were calculated for each using its ratio of the peak area obtained in IRMM-468 to the calibrator. This was related to the exact initial mass of IRMM-468 (T 4 ), used in the preparation of the sample solution, to finally yield the mass fraction of the impurity in IRMM-468. The mass fraction of T 4 was obtained by difference between 100 % and the sum of T 3, T 2 and MIT mass fractions in %. T 3, T 2 and MIT impurities were quantified in IRMM-468. One additional and unknown impurity, eluting 0.2 min before MIT, was detected in the UV chromatogram of IRMM-468 and each commercially available standard T 4, T 3, T 2 and MIT. The area fraction of this unknown impurity was estimated using the ratio of peak areas for the unknown impurity and for MIT, related to the percentage of MIT in IRMM-468. The unknown impurity represented 0.13 % of IRMM-468. This impurity could not be identified nor the LOQ determined. Therefore the mass fraction of this impurity was not taken into account to assess 26

30 the mass fraction of thyroxine in IRMM-468. However, a relative uncertainty contribution of 0.13 % was considered in the uncertainty budget of IRMM-468 (see Table 22). Table 11. Laboratory B: Purity of IRMM-468, expressed as a mass fraction (1 sample, measured on 3 days in 5 replicates): Day MIT T 2 T 3 T Average s Procedures applied by laboratory C One IRMM-468 sample was analysed in 10 replicates. The presence of MIT, T 2, T 3 impurities in IRMM-468 was investigated using the same HPLC-UV method as described for the short-term stability study (see Section ). Calibration curves were established for MIT, T 2, T 3 and as an additional information for T 4. Coefficients of determination (R²) and slopes of the curves are given in Annex E. A different slope (b) was observed for MIT (b = L/mg) compared to that of the other impurities (b between and L/mg). This could be related to the relatively low coefficient of determination obtained for MIT (R² = 0.862) or to the different chemical structure of MIT (tyrosine base molecule) and T 2, T 3, T 4 (thyronine base molecule) (Figure 4). 27

31 TYROSINE THYRONINE Figure 4 Structure of tyrosine and thyronine The results of the purity determination of IRMM-468 are presented in Table 12. Only T 3 was higher than LOQ and could be quantified as an impurity in the samples. The T 3 mass fraction was determined using the calibration curve and was related to the exact mass of IRMM-468 used to prepare the sample solution. The T 4 mass fraction was calculated by difference between 100 % and the T 3 mass fraction. Table 12. Laboratory C: Purity of IRMM-468, expressed as a mass fraction (1 sample, 10 replicates): T 3 T Average s

32 HPLC-UV-MS method for purity assessment Procedure applied by laboratory D IRMM-468 purity was investigated in two samples, each analysed in 10 replicates on three occasions. In addition to 3,5-diiodothyronine (T 2 ) and 3,3,5- triiodothyronine (T 3 ) impurities, the presence of intermediate products and by-products of synthesis of T 4 was checked. Possible target compounds were deduced from synthesis routes described in the literature [5-11]: 3,3,5,5 -tetraiodothyroacetic acid (TETRAC), 3,3,5-triiodothyro-acetic acid (TRIAC), 3,5-diiodothyroacetic acid, and 3,5-diiodotyrosine (DIT), tyrosine. A general check for impurities in T 4 material was also performed by GC-MS. No one of the intermediate products and by-products of synthesis substances except DIT could be detected by HPLC-MS and GC-MS. In addition, Fourier Transform-Infra Red (FT-IR) measurements were carried out. Spectra corresponded to expectations and the identification of T 4 was confirmed. Calibration solutions of commercially available T 2 and T 3 were prepared at the concentration expected in IRMM-468 which is a mass fraction about 0.01 % for T 2 and 0.3 % for T 3. Calibration solutions of commercially available DIT were also prepared. T 3 and T 4 measurements yielded both an MS and a UV based estimate of the respective mass fraction. No significant difference (Student s t-test, 5 % level) could be detected between the MS- and UV-based results. The results for T 3 and T 4 were the pooled mass fraction averages of the respective MS- and UV- results. In the case of DIT and T 2, only the MS results were considered as the UV-signal was lower than the detection limit. The mass fractions of T 2 and T 3 in IRMM-468 solution (averaged over the 10 runs) were calculated using the calibration curves and the initial mass of IRMM-468 sample. This was reported as a single measurement result. The same procedure was repeated on three different occasions. The three quantification results are compiled in Table 13. The mass fraction of T 4 among all components was obtained by difference between 100 % and the impurity mass fractions. The result of quantification of DIT by HPLC-MS was a mass fraction of % in IRMM-468. As a single measurement was carried out, no uncertainty statement can directly be derived from the repetition of measurements. Therefore a rectangular probability distribution was used between % and % as a rough estimate (half width of interval round the given value = %). A standard uncertainty of % for DIT quantification was then determined using the following equation [1]: u =

33 Table 13. Laboratory D: Purity of IRMM-468, expressed as a mass fraction (2 samples, 3*10 replicates): Sample T 3 T 3 Average T 3 T 2 T 2 DIT Calculated* number (MS) (UV) (MS&UV) (MS) (UV) T , ** 99, ** 99,67 Average s Sample T 3 T 3 Average T 3 T 2 T 2 DIT Calculated* number (MS) (UV) (MS&UV) (MS) (UV) T [% ] [% ] [% ] [% ] [% ] [% ] ** ** Average s * T 4 w HPLC = (100 % average T 3 (MS&UV) T 2 (MS) DIT (MS)) ** As a single measurement of DIT was performed, the same value of DIT mass fraction was used to calculate the mass fractions of T 4 in each replicate The average of T 4 mass fractions was calculated on both samples and was % with a standard deviation of 0.01 %, 30

34 LC-MS methods for purity assessment Procedure applied by laboratory E One IRMM-468 sample was analysed in triplicate. DIT, T 2, T 3 and reverse T 3 were the targeted impurities. Reverse T 3 (3,3',5'- triidodthyronine) is a biologically inactive isomer of T 3 [12] that can be separated from T 3 using reversed-phase liquid chromatography. A working solution of IRMM-468 was prepared at 100 mg/l. Calibration solutions of DIT, T 2, T 3, reverse T 3 and T 4 were prepared at the same concentration in order to estimate LODs and LOQs but were not used to quantify the mass fractions of the impurities. IRMM-468 was analysed preliminarily by full scan HPLC-MS and HPLC-MS-MS. However, due to the lack of sensitivity of quadrupole mass spectrometers operating in full scan mode, no impurity peaks were observed in the sample. A targeted analysis of possible and expected impurities in the sample (DIT, T 2, T 3, and reverse T 3 ) was carried out using HPLC-MS in the selected ion monitoring mode. No detectable T 2 and no reverse T 3 were found. DIT was detected at levels higher than LOD but lower than LOQ. Therefore DIT was not taken into account to calculate the mass fraction of T 4 in IRMM-468 but an uncertainty contribution related to DIT LOQ was added to the uncertainty budget of IRMM-468 (see Table 22). Finally T 3 was quantified as an impurity in IRMM-468 using the relative ion intensities of T 4 and T 3 in MS. The intensity fractions of T 4 and T 3 were assumed to correspond to the respective mass fractions. This approach does not take into consideration the differences in the electrospray ion yields of DIT and T 3 or T 4 (DIT contains one phenyl base molecule whereas T 3 and T 4 contain two phenyl base molecules). However, as only T 3 and T 4 were quantifiable and given their similar chemical structure, the probability of such a bias is rather low. The results of the purity assessment are presented in Table 14. Table 14. Laboratory E: Purity of IRMM-468, expressed as a mass fraction (1 sample, 3 replicates): T 4 DIT T 2 T 3 T 4 1 < LOQ < LOD < LOQ < LOD < LOQ < LOD Average s

35 Impact of the purity of the calibrants on the value assignment of T 4 mass fraction in Laboratories B, C and D As described in Table 8 (Section 5.1), the purity of the calibrants used for the determination of T 4 mass fraction in IRMM-468 was between 95 and more than 99 %. The impact of using calibrants not certified for their purity on the determination of T 4 mass fraction was evaluated by considering the less favourable case - where each calibrant is only 95 % pure (lowest purity percent described for the calibrants by the supplier) and - where the highest mass fraction of the HPLC-detectable impurities was observed by the laboratories. Table 15 presents the T 4 mass fraction obtained in the case of calibrants 100 % pure and 95 % pure, considering the maximum HPLC-detectable impurities mass fractions observed in the laboratories. The comparison of T 4 mass fraction obtained with 100 % pure calibrants and with 95 % pure calibrants shows a difference at the second digit after the coma only. The worst case of an eventual underestimation of T 4 mass fraction in IRMM-468 shows a bias from the certified value which is largely covered by the certified uncertainty of 0.7 %. Therefore, the impact of the calibrants purity can be considered as negligible in the case of IRMM-468. Table 15. Evaluation of the impact of calibrants purity on the determination of T 4 mass fraction IRMM-468 MIT DIT T 2 T 3 Maximum impurities mass fractions observed by the laboratories If the calibrants are 100 % pure: Corresponding minimum T 4 mass fraction = (100 % maximum impurities mass fractions) If the calibrants are 95 % pure: Bias: 5 % of maximum impurities mass fractions Resulting overestimation (105 %) of impurities mass fractions Resulting underestimated T 4 mass fraction = (100 % overestimated impurities mass fractions)

The Certification of the mass fraction of 3,3,5-triiodothyronine in a CRM intended for calibration. Certified Reference Material IRMM-469

The Certification of the mass fraction of 3,3,5-triiodothyronine in a CRM intended for calibration. Certified Reference Material IRMM-469 The Certification of the mass fraction of 3,3,5-triiodothyronine in a CRM intended for calibration Certified Reference Material IRMM-469 B. Toussaint, C. L. Klein, M. Wiergowski Report EUR 21893 EN The

More information

Preparation and Certification of IRMM-081a and IRMM-086 Spike Isotopic Reference Materials

Preparation and Certification of IRMM-081a and IRMM-086 Spike Isotopic Reference Materials Preparation and Certification of IRMM-081a and IRMM-086 Spike Isotopic Reference Materials A. Verbruggen, R. Eykens, F. Kehoe, H. Kühn, S. Richter, Y. Aregbe EUR 22944 EN - 2007 The mission of the IRMM

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Amiodarone-D 4, Primary Standard A-83 Page 1 of 8 Catalog Number: A-83 Lot: Expiration: Description: Packaging: Storage: Shipping: Intended Use: Instructions

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Dehydroepiandrosterone 3-sulfate, Primary Standard D-065 Page 1 of 6 Catalog Number: D-065 Lot: H 3 C Expiration: Description: February 2019 Dehydroepiandrosterone

More information

Certificate of analysis Certified reference material code: CRM-00-GTX1&4

Certificate of analysis Certified reference material code: CRM-00-GTX1&4 Certificate of analysis Certified reference material code: CRM-00-GTX1&4 Name: Gonyautoxin-1 and Gonyautoxin-4 in 0.003 M HCl aq. Batch: 15-001 Description: This Certified Reference Material (CRM) is a

More information

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard Certified Reference Material - Certificate of Analysis Methylone, Primary Standard Page 1 of 9 Catalog Number: O Lot: Expiration Date: Description: Packaging: Storage: Shipping: Intended Use: Instructions

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis Reference Material - Primary Standard Product Name: Oxycodone 1.0 mg/ml in Methanol Catalogue Number: Lot Number: 37583 LGCAMP0672.10-01 CAS Number: 76-42-6 Molecular Formula: C

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis Reference Material - Primary Standard Product Name: Metoprolol Tartrate 1.0 mg/ml in Methanol (as free base) Catalogue Number: LGCAMP0027.00-01 Lot Number: 8525 CAS Number: 56392-17-7

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis ISO GUIDE 34 ACLASS Cert# AR-1470 ISO/IEC 17025 ACLASS Cert# AT-1467 AMOXICILLIN TRIHYDRATE CERTIFIED REFERENCE MATERIAL HO O H H N H S CH 3 3H 2 O NH 2 N CH 3 COOH CERTIFIED PURITY:

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis Reference Material - Primary Standard Product Name: LSD (Lysergic Acid Diethylamide) 1.0 mg/ml in Acetonitrile Catalogue Number: Lot Number: 16905 LGCAMP1346.00-11 O CAS Number:

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis ISO GUIDE 34 ANAB Cert# AR-1470 ISO/IEC 17025 ANAB Cert# AT-1467 ATROPINE SULFATE CERTIFIED REFERENCE MATERIAL CERTIFIED PURITY: 99.9%, U crm = ±0.2% k = 2 (Mass Balance/anhydrous

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis LGC Quality ISO Guide 34:2009 DAkkS D-RM-14176-01-00 ISO/IEC 17025:2005 DAkkS D-PL-14176-01-00 ISO 9001:2008 DQS 102448 QM08 Primary Reference Standard Tilidine Hydrochloride Hemihydrate

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis LGC Quality ISO Guide 34:2009 DAkkS D-RM-14176-01-00 ISO/IEC 17025:2005 DAkkS D-PL-14176-01-00 ISO 9001:2008 DQS 102448 QM08 Primary Reference Standard Paracetamol H N Catalogue

More information

ASPIRIN (ACETYLSALICYLIC ACID) CERTIFIED REFERENCE MATERIAL

ASPIRIN (ACETYLSALICYLIC ACID) CERTIFIED REFERENCE MATERIAL ASPIRIN (ACETYLSALICYLIC ACID) CERTIFIED REFERENCE MATERIAL HO O O CH 3 O LOT #: BCBD0724V Fluka No. 01478 NOMINAL WT: 1 g CORRESPONDES TO RTC LOT: P500003 CERTIFICATE VALIDITY DATE: 20. October 2011 VERSION:

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis LGC Quality ISO Guide 34:2009 DAkkS D-RM-14176-01-00 ISO/IEC 17025:2005 DAkkS D-PL-14176-01-00 ISO 9001:2008 DQS 102448 QM08 Primary Reference Standard Granisetron Hydrochloride

More information

Certificate of analysis Certified reference material code: CRM-00-NEO

Certificate of analysis Certified reference material code: CRM-00-NEO Nº 2/PMR3 Certificate of analysis Certified reference material code: CRM--NEO Name: Neosaxitoxin dihydrochloride in.3 M HCl aq. Batch: 14-1 Description: This Certified Reference Material (CRM) is a solution

More information

Rosemary extract liquid

Rosemary extract liquid EUROPEAN COMMISSION DIRECTORATE-GENERAL JOINT RESEARCH CENTRE Institute for Reference Materials and Measurements Community Reference Laboratory - Feed Additives Authorisation Evaluation Report of the Community

More information

Period of validity: to date of issue:

Period of validity: to date of issue: Deutsche Akkreditierungsstelle GmbH Annex to the Accreditation Certificate D-RM-14176-01-00 according to the relevant clauses of ISO Guide 34: and DIN EN ISO/IEC 17025:2005 Period of validity: 11.01.2016

More information

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 126 Infinity Binary LC system with UV detection Application Note Food Author Srividya Kailasam Agilent Technologies,

More information

PHENYLEPHRINE HYDROCHLORIDE CERTIFIED REFERENCE MATERIAL

PHENYLEPHRINE HYDROCHLORIDE CERTIFIED REFERENCE MATERIAL PHENYLEPHRINE HYDROCHLORIDE CERTIFIED REFERENCE MATERIAL OH H 3 C H N OH HCl LOT #: BCBD1315V Fluka No. 11529 NOMINAL WT: 500 mg CERTIFICATE VALIDITY DATE: 12. October 2010 VERSION: BCBD1315V-1 Note: Certificates

More information

CERTIFICATION REPORT

CERTIFICATION REPORT CERTIFICATION REPORT The certification of 202 Hg enriched methyl mercury amount content and isotopic composition Hg as methyl mercury in ethanol/water solution: ERM -AE671 EUR 28214 EN - 2016 JRC-IRMM

More information

(RS)-2-(4-isobutyrylphenyl)propanoic acid Ibuprofen Impurity J

(RS)-2-(4-isobutyrylphenyl)propanoic acid Ibuprofen Impurity J Reference Material Product Information Sheet Epichem's Quality System conforms to ISO9001:2015 as certified by ECAAS Pty Ltd - Certification number 616061. Name BP Name Synonym(s) O (RS)-2-(4-isobutyrylphenyl)propanoic

More information

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer APPLICATION NOTE 72385 Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer Authors Stephan Meding, Katherine Lovejoy, Martin Ruehl Thermo Fisher Scientific,

More information

The Theory of HPLC. Quantitative and Qualitative HPLC

The Theory of HPLC. Quantitative and Qualitative HPLC The Theory of HPLC Quantitative and Qualitative HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed Date 2017.

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed Date 2017. Final Report Title Characterisation of Job No 2016/11/12-34 AS No. 1234A Client Contact Sample Author report Signed Date 2017 Easy Reach Report 2017 v2.docx 1 of 33 Contents 1. Study Summary Page 3 2.

More information

Annex to the Accreditation Certificate D-RM according to ISO Guide 34:2009 and DIN EN ISO/IEC 17025:2005

Annex to the Accreditation Certificate D-RM according to ISO Guide 34:2009 and DIN EN ISO/IEC 17025:2005 Deutsche Akkreditierungsstelle GmbH Annex to the Accreditation Certificate D-RM-14176-01-00 according to ISO Guide 34:2009 and DIN EN ISO/IEC 17025:2005 Period of validity: 21.04.2017 to 04.08.2019 date

More information

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION Andrew J. Aubin and Tanya L. Jenkins Waters Corporation, Milford, MA, USA INTRODUCTION Benzocaine (4-Aminobenzoic

More information

Chapter 4: Verification of compendial methods

Chapter 4: Verification of compendial methods Chapter 4: Verification of compendial methods Introduction In order to ensure accurate and reliable test results, the quality control laboratory (QCL) needs to use analytical methods (and accompanying

More information

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Application Note Clinical Research Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Authors Yanan Yang 1, Victor Mandragon 2, and Peter Stone 1 1

More information

Fast peaks in chromatograms of Sudan dyes

Fast peaks in chromatograms of Sudan dyes Journal of Chromatography A, 1160 (2007) 227 234 Fast peaks in chromatograms of Sudan dyes Kadi Mölder, Allan Künnapas, Koit Herodes, Ivo Leito Institute of Chemical Physics, University of Tartu, Jakobi

More information

VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual

VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual Page 1 / 11 VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual ZV-3030-0200-20 200 2-8 C Page 2 / 11 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. TEST PRINCIPLE... 3

More information

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Replaces: Dated: Author: Date: AM-No.: New New Nils Arne Jentoft 18.06.2014 0 CHANGES This procedure is new. 1 SCOPE This document describes

More information

Substance Characterisation for REACH. Dr Emma Miller Senior Chemist

Substance Characterisation for REACH. Dr Emma Miller Senior Chemist Substance Characterisation for REACH Dr Emma Miller Senior Chemist Submission of a REACH registration dossier, either as a lead registrant or a joint registrant Submission of an Inquiry to ECHA Submission

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis Reference Substance N-Desmethylclobazam (Norclobazam) Catalogue Number: Lot Number: 16595 LGCFOR0910.01 Molecular Formula: C 15 H 11 ClN 2 O 2 Molecular Weight: 286.71 CAS Number:

More information

Preparation and Certification of IRMM1027k, Large-Sized Dried (LSD) spike

Preparation and Certification of IRMM1027k, Large-Sized Dried (LSD) spike Preparation and Certification of IRMM1027k, Large-Sized Dried (LSD) spike A. Verbruggen, J Bauwens, R. Eykens, F. Kehoe, H. Kühn, U. Jacobsson, S. Richter, Y. Aregbe EUR 23539 EN - 2008 The mission of

More information

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC.

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC. 7. Stability indicating analytical method development and validation of and in capsule dosage form by HPLC. 7.1 INSTRUMENTS AND MATERIALS USED 7.1.1 INSTRUMENTS 1. Shimadzu LC-2010 CHT with liquid chromatograph

More information

Metolachlor ESA Sodium Salt Standard

Metolachlor ESA Sodium Salt Standard Analytical Reference Materials Metolachlor ESA Sodium Salt Standard Catalog # 33200 Lot # A0102104 110 Benner Circle Bellefonte, PA 16823-8812 1-814-353-1300 1-800-356-1168 www.restek.com FOR LABORATORY

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis E-903 Page 1 of 7 8-Chloro-6-pentadeuterophenyl-4H-[1,2,4]triazolo[4,3-a][1,4]benzodiazepine Catalog umber: E-903 Cl Solution Lot: Expiration Date: Solvent: August 2016 Methanol

More information

CERTIFICATE OF ANALYSIS

CERTIFICATE OF ANALYSIS in cooperation with the WG Copper of the Committee of Chemists of GDMB CERTIFICATE OF ANALYSIS ERM -EB386 Pure copper Element Certified Values Certified value 1) Uncertainty 2) Mass fraction in mg/kg Ag

More information

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection APPLICATION NOTE 72686 Determination of underivatized aflatoxins,, G2, and in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection Authors Sylvia Grosse, Mauro De Pra, Frank

More information

CERTIFICATION REPORT Preparation and Certification of Large-Sized Dried (LSD) Spike IRMM-1027r

CERTIFICATION REPORT Preparation and Certification of Large-Sized Dried (LSD) Spike IRMM-1027r CERTIFICATION REPORT Preparation and Certification of Large-Sized Dried (LSD) Spike IRMM-1027r R. Jakopič, J. Bauwens, R. Buják, C. Hennessy, F. Kehoe, U. Jacobsson, S. Richter and Y. Aregbe 2016 Report

More information

Development And Validation Of Rp-Hplc Method For Determination Of Velpatasvir In Bulk

Development And Validation Of Rp-Hplc Method For Determination Of Velpatasvir In Bulk International Journal of Engineering Science Invention (IJESI) ISSN (Online): 2319 6734, ISSN (Print): 2319 6726 www.ijesi.org PP. 36-41 Development And Validation Of Rp-Hplc Method For Determination Of

More information

Validation of an Analytical Method

Validation of an Analytical Method Validation of an Analytical Method Refer to: ICH Guideline Q2(R1), Validation of Analytical Procedures: Teaxt andmethodology. Introduction All major laboratories eg. in the industry operates with clearly

More information

Appendix II- Bioanalytical Method Development and Validation

Appendix II- Bioanalytical Method Development and Validation A2. Bioanalytical method development 1. Optimization of chromatographic conditions Method development and optimization of chromatographic parameters is of utmost important for validating a method in biological

More information

HIPPURIC ACID and o-m-p- METHYLHIPPURIC ACIDS IN URINE BY UV CODE Z06010

HIPPURIC ACID and o-m-p- METHYLHIPPURIC ACIDS IN URINE BY UV CODE Z06010 HIPPURIC ACID and o-m-p- METHYLHIPPURIC ACIDS IN URINE BY UV CODE Z06010 Biochemistry The determination of Hippuric Acid ( HPA ) in urine is important to determine the risk due to the exposition to Toluene.

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis Reference Substance Cefalexin Monohydrate Catalogue Number: Lot Number: 38530 LGCFOR0605.00 Molecular Formula: C 16 H 17 N 3 O 4 S H 2 O Molecular Weight: 365.40 CAS Number: [ 23325-78-2

More information

Measurement Uncertainty, March 2009, F. Cordeiro 1

Measurement Uncertainty, March 2009, F. Cordeiro 1 Measurement Uncertainty, March 2009, F. Cordeiro 1 Uncertainty Estimation on the Quantification of Major Proteins in Milk by Liquid Chromatography Contribution to a Chemical Reference Measurement System

More information

Development and Validation of a HPLC Method for Chlorphenamine Maleate Related Substances in Multicomponents Syrups and Tablets

Development and Validation of a HPLC Method for Chlorphenamine Maleate Related Substances in Multicomponents Syrups and Tablets Development and Validation of a HPLC Method for Chlorphenamine Maleate Related Substances in s Syrups and Tablets Larisa Alagić-Džambić*, Midhat Vehabović, Edina Čekić, Mirsad Džambić Development Department,

More information

LUMEFANTRINUM LUMEFANTRINE

LUMEFANTRINUM LUMEFANTRINE July 2008 LUMEFANTRINE: Final text for addition to The International Pharmacopoeia (July 2008) This monograph was adopted at the Forty-second WHO Expert Committee on Specifications for Pharmaceutical Preparations

More information

Quantification of Furocoumarins

Quantification of Furocoumarins Version 8 (12/04/2013) Quantification of Furocoumarins HPLC/DAD Procedure Scope The European Cosmetics Directive 76/768/EEC, which is about to be adopted in the European Cosmetic Regulation, in Annex II

More information

with diode array detection

with diode array detection Application Note Small Molecule Pharmaceuticals Analysis of Tween 8 by highperformance liquid chromatography with diode array detection Authors Jianxin Yu, Scott Citrowske, Nikki Carlson, and Jacob Strange

More information

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3)

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3) Schedule Assignment Due Date Draft Section of Lab Report Monday 6pm (Jan 27) Quiz for Lab 2 Peer Review of Draft Complete Lab Report 1 Tuesday 9:30am Wednesday 6pm Friday 6pm Summary of Paper 2 Monday

More information

CERTIFICATE OF ANALYSIS

CERTIFICATE OF ANALYSIS in cooperation with the WG Copper of the Committee of Chemists of GDMB CERTIFICATE OF ANALYSIS Element ERM -EB383 Pure copper ausverkauft / out of stock Certified Values Certified value 1) Uncertainty

More information

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 66 Triple Quadrupole LC/MS System Application Note Environmental Authors Don Noot, Matthew Noestheden, and Ralph Hindle Vogon

More information

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity Janet Hammond Waters Corporation, Wilmslow, UK APPLICATION BENEFITS High

More information

HIPPURIC ACID, o-m-p-methylhippuric ACIDS and CREATININE IN URINE by UV CODE Z07510

HIPPURIC ACID, o-m-p-methylhippuric ACIDS and CREATININE IN URINE by UV CODE Z07510 HIPPURIC ACID, o-m-p-methylhippuric ACIDS and CREATININE IN URINE by UV CODE Z07510 Biochemistry The determination of Hippuric Acid ( HPA ) in urine is important to determine the risk due to the exposition

More information

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION METHOD 80 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION 1.0 SCOPE AND APPLICATION 1.1 Method 80 may be used to determine the concentration of acetonitrile (CAS No. 75-05-8) in aqueous

More information

CERTIFICATE OF ANALYSIS

CERTIFICATE OF ANALYSIS in cooperation with the WG Copper of the Committee of Chemists of GDMB CERTIFICATE OF ANALYSIS ERM -EB389 CuNi25 Certified value 1) Uncertainty 2) Element Mass fraction in % Cu 74.3 ± 0.5 Ni 24.7 ± 0.5

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis Reference Substance Dehydronorketamine Hydrochloride Catalogue Number: Lot Number: 14579 LGCFOR0144.07 Molecular Formula: C 12 H 12 ClNO HCl Molecular Weight: 258.14 CAS Number:

More information

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis High-Resolution Sampling D-LC for Pharmaceutical Impurity Analysis Detection of Impurities Hidden Under the API Peak at Relevant Levels Application Note Small Molecule Pharmaceuticals and Generics Authors

More information

Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers.

Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers. Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers. 1 Scope This document describes a chromatographic method which allows the identification as well as the determination of

More information

Introduction to Pharmaceutical Chemical Analysis

Introduction to Pharmaceutical Chemical Analysis Introduction to Pharmaceutical Chemical Analysis Hansen, Steen ISBN-13: 9780470661222 Table of Contents Preface xv 1 Introduction to Pharmaceutical Analysis 1 1.1 Applications and Definitions 1 1.2 The

More information

On-line Monitoring of Water during Reference Material Production by AOTF-NIR Spectrometry

On-line Monitoring of Water during Reference Material Production by AOTF-NIR Spectrometry On-line Monitoring of Water during Reference Material Production by AOTF-NIR Spectrometry Håkan Emteborg, Vikram Kestens and Alexander Bernreuther Institute for Reference Materials and Measurements (IRMM)

More information

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note Authors Siji Joseph Agilent Technologies India Pvt. Ltd.

More information

BUTYL ALCOHOL in urine by GC-FID Code GC05510

BUTYL ALCOHOL in urine by GC-FID Code GC05510 BUTYL ALCOHOL in urine by GC-FID Code GC00 BIOCHEMISTRY Butanol (or n-butyl alcohol) is an alcohol which at room temperature is presented as a colorless liquid smelling of alcohol. It is an inflammatory

More information

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent Infinity LC System with Diode Array Detector Application Note Author Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany [mau]

More information

ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES

ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES Adopted from ICH Guidelines ICH Q2A: Validation of Analytical Methods: Definitions and Terminology, 27 October 1994. ICH Q2B: Validation of Analytical

More information

HPLC Reverse Phase Test Mix #1

HPLC Reverse Phase Test Mix #1 Analytical Reference Materials HPLC Reverse Phase Test Mix #1 Catalog # 35005 Lot # A083761 110 Benner Circle Bellefonte, PA 16823-8812 (814) 353-1300 FOR LABORATORY USE ONLY. READ MSDS PRIOR TO USE. RAW

More information

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System Margaret Maziarz, Sean M. McCarthy, Michael D. Jones, and Warren B. Potts Waters Corporation, Milford, MA, USA A P

More information

GAFTI Analytical method for ISO/TS 16179:2012 Detection and Determination of Organotin Compounds in Footwear and Apparel Materials by GC-MS

GAFTI Analytical method for ISO/TS 16179:2012 Detection and Determination of Organotin Compounds in Footwear and Apparel Materials by GC-MS GAFTI Analytical method for ISO/TS 16179:2012 Detection and Determination of Organotin Compounds in Footwear and Apparel Materials by GC-MS 1. Scope This method is applicable for determining the presence

More information

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 9 DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 9 Determination of drug release during

More information

2.1 2,3 Dichloro Benzoyl Cyanide (2,3 DCBC) and survey of. manufactured commonly for the bulk drug industry, few references were

2.1 2,3 Dichloro Benzoyl Cyanide (2,3 DCBC) and survey of. manufactured commonly for the bulk drug industry, few references were . Introduction.,3 Dichloro Benzoyl Cyanide (,3 DCBC) and survey of analytical methods,3-dcbc substance, is the although advanced intermediate,3-dcbc is a of lamotrigine well-known bulk drug chemical manufactured

More information

Chromatography & instrumentation in Organic Chemistry

Chromatography & instrumentation in Organic Chemistry Chromatography & instrumentation in Organic Chemistry What is Chromatography? Chromatography is a technique for separating mixtures into their components in order to analyze, identify, purify, and/or quantify

More information

Low-level Determination of 4-Hydrazino Benzoic Acid in Drug Substance by High Performance Liquid Chromatography/Mass Spectrometry

Low-level Determination of 4-Hydrazino Benzoic Acid in Drug Substance by High Performance Liquid Chromatography/Mass Spectrometry ISSN: 0973-4945; CODEN ECJHAO E- Chemistry http://www.e-journals.net 2010, 7(2), 403-408 Low-level Determination of 4-Hydrazino Benzoic Acid in Drug Substance by High Performance Liquid Chromatography/Mass

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

HPLC analysis of VFAs, Furfural and HMF

HPLC analysis of VFAs, Furfural and HMF HPLC analysis of VFAs, Furfural and HMF Introduction The following chromatographic analysis is developed in order to analyse volatile fatty acids and carbohydrate degradation products; furfural, hydroxymethylfurfural

More information

Certification Report. REIMEP 18 Inter-Laboratory Comparison for the Measurement of Uranium Isotopic Ratios in Nitric Acid Solution

Certification Report. REIMEP 18 Inter-Laboratory Comparison for the Measurement of Uranium Isotopic Ratios in Nitric Acid Solution Certification Report REIMEP 18 Inter-Laboratory Comparison for the Measurement of Uranium Isotopic Ratios in Nitric Acid Solution S. Richter, A. Alonso, J. Truyens, H. Kühn, A. Verbruggen, R. Wellum Institute

More information

VAM Project Development and Harmonisation of Measurement Uncertainty Principles

VAM Project Development and Harmonisation of Measurement Uncertainty Principles VAM Project 3.2.1 Development and Harmonisation of Measurement Uncertainty Principles Part (d): Protocol for uncertainty evaluation from validation data V J Barwick and S L R Ellison January 2000 LGC/VAM/1998/088

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis Reference Substance Atorvastatin Calcium Catalogue Number: Lot Number: 55085 LGCFOR0400.00 Molecular Formula: 2C 33 H 34 FN 2 O 5 Ca Molecular Weight: 1155.34 CAS Number: [ 134523-03-8

More information

CERTIFICATION REPORT. The certification of the mass fraction of arsenobetaine in water: ERM -AC626

CERTIFICATION REPORT. The certification of the mass fraction of arsenobetaine in water: ERM -AC626 CERTIFICATION REPORT The certification of the mass fraction of arsenobetaine in water: ERM -AC626 EUR 28669 EN - 2017 JRC-IRMM promotes a common and reliable European measurement provides reference measurements.

More information

BRIEFING. (EXC: K. Moore.) RTS C Propylparaben C 10 H 12 O Benzoic acid, 4 hydroxy, propyl ester; Propyl p hydroxybenzoate [ ].

BRIEFING. (EXC: K. Moore.) RTS C Propylparaben C 10 H 12 O Benzoic acid, 4 hydroxy, propyl ester; Propyl p hydroxybenzoate [ ]. BRIEFING Propylparaben. The European Pharmacopoeia is the coordinating pharmacopeia for the international harmonization of the compendial standards for the Propylparaben monograph, as part of the process

More information

Soils, sludges and treated bio-waste Organic constituents - LAS by HPLC with fluorescence detection (LC-FLD) and mass selective detection (LC- MSD)

Soils, sludges and treated bio-waste Organic constituents - LAS by HPLC with fluorescence detection (LC-FLD) and mass selective detection (LC- MSD) ISO 2001 All rights reserved HORIZONTAL 13.0 April 2006 Second draft Soils, sludges and treated bio-waste Organic constituents - LAS by HPLC with fluorescence detection (LC-FLD) and mass selective detection

More information

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Application Note Food Authors Jianqiu Mi, Zhengxiang Zhang, Zhixu Zhang,

More information

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes Joe DiBussolo, Marta Kozak Thermo Fisher Scientific, San Jose, CA Application

More information

Instrumental Chemical Analysis

Instrumental Chemical Analysis L1 Page1 Instrumental Chemical Analysis Dr. Ahmad Najjar Philadelphia University Faculty of Pharmacy Department of Pharmaceutical Sciences 2 nd semester, 2016/2017 L1 Page2 Course Syllabus Course title:

More information

Available online Journal of Chemical and Pharmaceutical Research, 2012, 4(6): Research Article

Available online  Journal of Chemical and Pharmaceutical Research, 2012, 4(6): Research Article Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research, 2012, 4(6):3275-3279 Research Article ISSN : 0975-7384 CODEN(USA) : JCPRC5 Development and validation of a matrix solid-phase

More information

FORMALDEHYDE IN URINE by UV Code Z65010

FORMALDEHYDE IN URINE by UV Code Z65010 FORMALDEHYDE IN URINE by UV Code Z65010 BIOCHEMISTRY The dose of urinary Formaldehyde is made after exposure to Formaldehyde and Methanol. The Formaldehyde is a pollutant so-called ubiquist, because it

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis Reference Substance Benzoic Acid Catalogue Number: Lot Number: 43480 LGCFOR0278.01 Molecular Formula: C 7 H 6 O 2 Molecular Weight: 122.12 CAS Number: [ 65-85-0 ] Long-term Storage:

More information

RESOLUTION OENO 33/2004 DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION

RESOLUTION OENO 33/2004 DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION The GENERAL ASSEMBLY, Considering Article paragraph iv of the agreement establishing the International organisation of vine and wine Upon

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis Reference Standard 4,5α-Epoxy-14-hydroxy-3-methoxymorphinan-6-one Hydrochloride (Noroxycodone Hydrochloride) Catalogue Number: MM0672.03 Molecular Formula: C 17 H 19 NO 4 HCl Molecular

More information

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Method of determination of phtalates in spirituous beverages by gaschromatography/mass OIV-MA-BS-33 Method of determination of phtalates in spirituous beverages by gas-chromatography/mass Type IV method 1. SCOPE OF APPLICATION This method applies to the detection and assay of some phthalates

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis Bromazepam 7-Bromo-1,3-dihydro-5-(2-pyridinyl)-2H-1,4-benzodiazepin-2-one Catalog Number: Br Solution Lot: Expiration Date: Solvent: August 2015 Methanol Volume per Ampule: Not

More information

ADVANTAME. Not less than 97.0% and not more than 102.0% on the anhydrous basis. Sweetener, flavour enhancer

ADVANTAME. Not less than 97.0% and not more than 102.0% on the anhydrous basis. Sweetener, flavour enhancer ADVANTAME SYNONYMS INS No. 969 Prepared at the 80 th JECFA (2015), published in FAO JECFA Monographs 17 (2015), superseding tentative specifications prepared at 77 th JECFA (2013). An ADI of 0-5 mg/kg

More information

Proficiency testing: Aqueous ethanol. Test and measurement Workshop Marcellé Archer. 20 September 2011

Proficiency testing: Aqueous ethanol. Test and measurement Workshop Marcellé Archer. 20 September 2011 Proficiency testing: Aqueous ethanol Test and measurement Workshop Marcellé Archer 20 September 2011 NMISA Ethanol PT schemes Purpose of a PT scheme Consistent good performance is the goal One bad result

More information

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 129 Infinity LC system Application Note Food and Beverages Author Srividya Kailasam

More information

CHAPTER - 3 ANALYTICAL PROFILE. 3.1 Estimation of Drug in Pharmaceutical Formulation Estimation of Drugs

CHAPTER - 3 ANALYTICAL PROFILE. 3.1 Estimation of Drug in Pharmaceutical Formulation Estimation of Drugs CHAPTER - 3 ANALYTICAL PROFILE 3.1 Estimation of Drug in Pharmaceutical Formulation 3.1.1 Estimation of Drugs ANALYTICAL PROFILE 84 3.1 ESTIMATION OF DRUG IN PHARMACEUTICAL FORMULATION. Agrawal A et al

More information

Analytische Qualitätssicherung Baden-Württemberg

Analytische Qualitätssicherung Baden-Württemberg Analytische Qualitätssicherung Baden-Württemberg Proficiency Test UKWIR SS 17 PAH in surface water with suspended solids Anthracene, Fluoranthene, Naphtalene, Benzo[a]pyrene, Benzo[b]fluoranthene, Benzo[k]fluoranthene,

More information

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors Trace-Level Analysis of Melamine in Milk Products on Agilent 789A/5975C GC/MSD Using a ew Agilent J&W DB-5ms Ultra Inert Column and SampliQ SCX Cartridges Application ote Gas Chromatography/Mass Spectrometry/Food

More information

FORMIC ACID in URINE by GC/MS-headspace - Code GC74010

FORMIC ACID in URINE by GC/MS-headspace - Code GC74010 FORMIC ACID in URINE by GC/MS-headspace - Code GC74010 BIOCHEMISTRY The Methanol is used in the production of cellulose, synthetic resins, formaldehyde, acetic acid and fuel. The route of exposure is more

More information

CYCLOSERINE Final text for addition to The International Pharmacopoeia. (November 2008) CYCLOSERINUM CYCLOSERINE

CYCLOSERINE Final text for addition to The International Pharmacopoeia. (November 2008) CYCLOSERINUM CYCLOSERINE December 2008 CYCLOSERINE Final text for addition to The International Pharmacopoeia (November 2008) This monograph was adopted at the Forty-third WHO Expert Committee on Specifications for Pharmaceutical

More information