ANALITIČKA PROCJENA ANALIZATORA ZA MJERENJE KONCENTRACIJE ELEKTROLITA
|
|
- Alban Carr
- 6 years ago
- Views:
Transcription
1 M. Brajković-Bralić Analitička procjena analizatora za mjerenje koncentracije elektrolita Stručni članak Professional paper ANALITIČKA PROCJENA ANALIZATORA ZA MJERENJE KONCENTRACIJE ELEKTROLITA Marina Brajković Bralić * Zavod za medicinsko-laboratorijsku dijagnostiku i transfuziologiju, Opća bolnica Sveti Duh, Zagreb SAŽETAK Cilj: Prikazati analitičku procjenu analizatora za mjerenje koncentracije elektrolita provedenu prema preporukama European Committee for Clinical Laboratory Standards (ECCLS). Metode: Na uzorcima seruma, pune krvi i kontrolnim uzorcima ispitivani su parametri: ph, koncentracija natrija (Na + ), kalija (K + ), ioniziranog kalcija (Ca ++ ) i klorida (Cl - ). Analitička procjena obuhvatila je ispitivanje netočnosti i nepreciznosti instrumenta iz dana u dan, nepreciznosti unutar dana, korelaciju s drugim analizatorima (Nova Stat Profile M i Olympus AU 400). Rezultati: Provedenom analitičkom procjenom ispitivanog analizatora dobivena je zadovoljavajuća netočnost i nepreciznost mjerenja iz dana u dan i nepreciznost unutar dana, te prihvatljiv stupanj korelacije s drugim analizatorima (Nova Stat Profile M i Olympus AU 400). Zaključak: Ispitivani analizator pokazao se pouzdanim i sigurnim analizatorom za određivanje ph i koncentracije elektrolita zbog svoje zadovoljavajuće netočnosti i nepreciznosti mjerenja, te dobre korelacije s ostalim analizatorima (Nova Stat Profile M i Olympus AU 400), a, uz to, njegova jednostavnost i brzina rada čine ga prikladnim za uporabu u medicinsko-biokemijskim laboratorijima kao zamjenu za korištenje plamenih fotometara. Ključne riječi: analitička procjena, analizator za mjerenje koncentracije elektrolita, netočnost i nepreciznost iz dana u dan, nepreciznost unutar dana, usporedna određivanja. UVOD Mjerenja koncentracije (aktivnosti) elektrolita jedna su od najbrojnijih analiza koje se svakodnevno vrše u medicinsko-biokemijskim laboratorijima. U sadašnjem trenutku, ova se mjerenja provode na višekanalnim biokemijskim analizatorima ili na posebnim analizatorima koji, uz koncentracije elektrolita, mjere i ph te parcijalne tlakove plinova u krvi. Potenciometrijsko određivanje koncentracije (aktivnosti) nekog iona otopljenog u tekućini u načelu počiva na mjerenju elektromotorne sile (potencijala) galvanskog članka koji se sastoji od dvije elektrode, tj. od dva vodiča ili poluvodiča uronjena u otopinu elektrolita. Članak čine dvije Adresa za dopisivanje: Marina Brajković Bralić, dipl. inž., Zavod za medicinsko-laboratorijsku dijagnostiku i transfuziologiju, Opća bolnica Sveti Duh, Zagreb, tel.: , fax: , mbrajkovicbralic@vip.hr referentne elektrode, vanjska i unutarnja i membrana. S obje strane ionsko-selektivne membrane javlja se razlika električnog potencijala na granici faza membrana / otopina. Unutarnja strana ionsko-selektivne membrane je u dodiru s unutarnjom referentnom otopinom stalne ionske jakosti i zbog toga se nalazi na stalnom potencijalu, dok se s vanjske površine membrane električni potencijal mijenja linearno s negativnim logaritmom aktivnosti ispitivanog iona. Ovim se metodama mogu mjeriti koncentracije iona do 10-5 mol/l. Opis analizatora Namjena ispitivanog analizatora je mjerenje koncentracije (aktivnosti) natrija, kalija, klorida, ioniziranog kalcija i ph u uzorcima seruma, plazme i pune krvi, koristeći ionsko-selektivne elektrode za pojedine elektrolite, odnosno ph, BIOCHEMIA MEDICA god. 15, br. 1 2,
2 4 Analitička procjena analizatora za mjerenje koncentracije elektrolita M. Brajković-Bralić a kao način mjerenja koristi se izravno mjerenje (nije potrebno prethodno razrjeđivanje uzorka). Ispitivani analizator razvijen je uz potporu Ministarstva znanosti i tehnologije, tehnološki projekt broj TP , a može poslužiti kao zamjena za plamene fotometre, jer ne samo što je je iniji i jednostavniji za rukovanje, već pruža i veći broj mogućnosti budući da se njime može mjeriti i koncentracija klorida i ioniziranog kalcija, uz ph uzorka (Slika 1). Ekran Upravljačka ploča Igla za uzimanje uzorka Pisač S l i k a 1. Analizator za mjerenje koncentracije elektrolita s pisačem F i g u r e 1. e analyzer for measurement of electrolyte concentration with the printer Predviđeno je da analizator bude neprestano spreman za rad, vrijeme potrebno za jednu analizu je 60 sekundi, a volumen potrebnog uzorka je 250 µl. Mjerna ćelija analizatora sastoji se od 5 protočnih ionsko-selektivnih elektroda za određivanje ph, te iona natrija, kalija, kalcija i klorida (izrađene na osnovi PVC s odgovarajućim ionoforom). Elektrode su postavljene u vodoravni niz i spojene brtvama u jednu mjernu ćeliju na kraju koje se nalazi jednako spojena protočna referentna elektroda. Referentna elektroda je elektrokemijska polućelija s Ag/AgCl u kontaktu s otopinom KCl (2 mol/l). Otopina KCl služi kao solni most za uspostavljanje kontakta između referentne elektrode i uzorka. Solni most je izveden tako da je otopina KCl u kontaktu s uzorkom preko gela. Ovaj otvoreni kontakt između otopine KCl i uzorka je stabilan i pouzdan jer je time uspješno uklonjeno stvaranje proteinskih naslaga METODE Analitička procjena analizatora provođena je tijekom tri mjeseca u Zavodu za medicinsko- laboratorijsku dijagnostiku i transfuziologiju Opće bolnice Sveti Duh u Zagrebu, prema propisima Europske komisije za kliničko-laboratorijske standarde (1) i na način kako je provedena procjena i drugih analizatora (2, 3). Analitička procjena obuhvatila je ispitivanje netočnosti i nepreciznosti instrumenta iz dana u dan, nepreciznosti unutar dana, korelaciju s drugim analizatorima (Nova Stat Profile M i Olympus AU 400) i obrađene rezultate mjerenja. Tijekom rada korišteni su komercijalni kontrolni uzorci Nova Biomedical (Nova Stat Profile Control Level 1, lot No Acidosis with high electrolyte, Nova Stat Profile Control Level 2, lot No normal ph with normal electrolyte, Nova Stat Profile Control Level 3, lot No Alcalosis with low electrolyte), te uzorci pune venske krvi i seruma. Uzorci pune venske krvi i seruma analizirani su neposredno nakon njihove dnevne rutinske obrade. Za uzorak seruma krv je uzimana u epruvete zatvorenog sustava dobavljača Terumo Europe (VENOSAFE VF-O54SP, 4 ml), te je centrifugirana 10 minuta na 4000 okretaja/min. Uzorci venske krvi uzimani su u štrcaljke obložene tankim slojem Li-heparina kao antikoagulansom. Za vrijeme ispitivanja, kalibracije analizatora elektrolita provođene su otopinama vlastite izrade: standard S1 (fiziološka koncentracija elektrolita) i standard S2 (nefiziološka koncentracija elektrolita). Koncentracije elektrolita u kalibracijskim otopinama analizatora prikazane su u Tablici 1. Nakon svakog uzorka (kontrolni uzorci, serum i venska krv) mjerna ćelija analizatora je ispirana otopinom za ispiranje koja je sadržavala razrijeđenu otopinu standarda S1 uz dodatak detergenta 10 mg/l MEGA 10. T a b l i c a 1. Koncentracije elektrolita u kalibracijskim otopinama analizatora T a b l e 1. Concentrations of electrolytes in calibration solutions of the analyzer Analit Standard S1 Standard S2 ph 7,4 6,8 Na (mmol/l) 142,0 90,0 K (mmol/l) 4,0 10,0 ica (mmol/l) 1,25 3,5 Cl (mmol/l) BIOCHEMIA MEDICA god. 15, br. 1 2, 2005.
3 M. Brajković-Bralić Analitička procjena analizatora za mjerenje koncentracije elektrolita REZULTATI I RASPRAVA Netočnost i nepreciznost iz dana u dan Netočnost i nepreciznost iz dana u dan je određena mjerenjem komercijalnih kontrolnih uzoraka različitih koncentracijskih područja (koncentracija 1, koncentracija 2, koncentracija 3). Broj mjerenja za pojedine analite i dobiveni rezultati prikazani su u tablicama 2, 3 i 4. Za većinu analita, u sve tri razine kontrolnih uzoraka, rezultati srednjih vrijednosti bili su unutar dozvoljenog raspona vrijednosti korištenih kontrolnih uzoraka. Odstupanje srednje vrijednosti izvan dozvoljenog raspona bilo je prisutno pri mjerenju ph kod kontrolnih uzoraka Koncentracija 1 (x=7,259) i Koncentracija 2 (x=7,631), natrija kod kontrolnog uzorka Koncentracija 1 (x=155,8) i klorida kod kontrolnih uzoraka Koncentracija 2 (x=93,4) i Koncentracija 3(x= 78,1). Odstupanja u vrijednostima dobivenima pri mjerenju ph i natrija mogu se objasniti različitim tehničkim osobinama ph i natrijeve elektrode kod ispitivanog analizatora u odnosu na analizator za koji su kontrolni uzorci pripravljeni (Nova Stat Profile). Naime, elektrode za ph i natrij kod ispitivanog analizatora su membranske elektrode, dok analizator Nova Stat Profile koristi staklene ph i natrijeve elektrode. Kratki vijek trajanja kloridove elektrode (2 3 tjedna), a stoga i manji broj mjerenja za kloridne ione kod sva tri kontrolna uzorka u odnosu na druge analite, može biti uzrok odstupa- T a b l i c a 2. Rezultati netočnosti i nepreciznosti iz dana u dan dobiveni upotrebom kontrolnog uzorka Nova Stat Profile Control koncentracija 1 T a b l e 2. Results of between-day inaccuracy and imprecision obtained using control sample Nova Stat Profile Control Level 1 Analit N Dozvoljeni raspon x SD KV(%) ph 13 7,160-7,210 7,259 0,020 0,3 Na (mmol/l) ,5 164,5 155,8 2,5 1,6 K (mmol/l) 13 5,75 6,25 5,9 0,2 3,2 ica (mmol/l) 13 1,41 1,61 1,47 0,05 3,4 Cl (mmol/l) ,5 2,7 2,1 N = broj analiza x = srednja vrijednost SD = standardna devijacija KV = koeficijent varijacije T a b l i c a 3. Rezultati netočnosti i nepreciznosti iz dana u dan dobiveni upotrebom kontrolnog uzorka Nova Stat Profile Control koncentracija 2 T a b l e 3. Results of between-day inaccuracy and imprecision obtained using control sample Nova Stat Profile Control Level 2 Analit N Dozvoljeni raspon x SD KV(%) ph 13 7,434 7,484 7,464 0,010 0,2 Na (mmol/l) ,6 137,6 135,2 3,2 2,3 K (mmol/l) 13 4,17 4,67 4,4 0,1 3,1 ica (mmol/l) 13 1,03 1,19 1,13 0,1 6,9 Cl (mmol/l) ,4 2,0 2,2 N = broj analiza x = srednja vrijednost SD = standardna devijacija KV = koeficijent varijacije T a b l i c a 4. Rezultati netočnosti i nepreciznosti iz dana u dan dobiveni upotrebom kontrolnog uzorka Nova Stat Profile Control koncentracija 3 T a b l e 4. Results of between-day inaccuracy and imprecision obtained using control Sample Nova Stat Profile Control Level 3 Analit N Dozvoljeni raspon x SD KV(%) ph 13 7,634-7,684 7,631 0,020 0,3 Na (mmol/l) ,6-128,6 126,6 2,4 2,0 K (mmol/l) 13 1,69-2,29 2,0 0,1 3,0 ica (mmol/l) 12 0,46-0,58 0,49 0,03 5,9 Cl (mmol/l) ,1 2,0 2,54 N = broj analiza x = srednja vrijednost SD = standardna devijacija KV = koeficijent varijacije BIOCHEMIA MEDICA god. 15, br. 1 2,
4 Analitička procjena analizatora za mjerenje koncentracije elektrolita M. Brajković-Bralić nja srednjih vrijednosti klorida izvan dozvoljenog raspona. Koeficijenti varijacije za nepreciznost iz dana u dan iznosili su za većinu analita do 3,2%, osim za ionizirani kalcij koji je imao najveće koeficijente varijacije: 3,4%, 6,9% i 5,9% za koncentracije 1, 2 i 3. Najbolji rezultati su dobiveni u sve tri razine kontrolnih uzoraka za ph (KV = 0,2-0,3%). Nepreciznost unutar dana Nepreciznost unutar dana određivana je uzastopnim mjerenjem ph i koncentracija elektrolita u tri različita uzorka seruma i tri različita uzorka pune venske krvi. Broj uzastopnih mjerenja pojedinog uzorka i dobiveni rezultati sažeti su u tablicama 5 i 6. Nepreciznost unutar dana za sve analite izražena kao koeficijent varijacije kretala T a b l i c a 5. Rezultati nepreciznosti unutar dana za uzorke seruma T a b l e 5. Results of within-day imprecision for serum samples Analit Uzorak N Raspon x SD KV(%) ph 1 6 7,676 7,750 7,724 0,026 0, ,706 7,878 7,776 0,044 0, ,683 7,809 7,747 0,043 0,6 Na (mmol/l) ,4 127,5 126,2 1,2 0, ,2 148,5 145,6 1,6 1, ,4 144,8 143,6 0,9 0,0 K (mmol/l) 1 6 4,6 4,8 4,8 0,1 1, ,8 4,8 4,8 0,0 0, ,1 5,2 5,1 0,0 0,7 ica (mmol/l) 1 6 1,09 1,19 1,13 0,03 3, ,90 1,00 0,95 0,03 3, ,08 1,12 1,09 0,02 1,5 Cl (mmol/l) ,2 109,3 107,7 1,5 1, ,7 109,2 106,3 1,4 1, ,7 105,7 104,0 0,9 0,8 N = broj analiza x = srednja vrijednost SD = standardna devijacija KV = koeficijent varijacije T a b l i c a 6. Rezultati nepreciznosti unutar dana za uzorke pune krvi T a b l e 6. Results of within-day imprecision for whole blood samples Analit Uzorak N Raspon x SD KV(%) ph 1 6 7,522 7,595 7,557 0,025 0, ,507 7,619 7,563 0,045 0, ,499 7,580 7,542 0,031 0,4 Na (mmol/l) ,8 148,7 145,7 2,0 1, ,1 131,7 130,7 0,8 0, ,9 128,7 128,3 0,3 0,2 K (mmol/l) 1 6 4,7 4,9 4,9 0,1 1, ,9 4,1 4,0 0,1 2, ,3 4,4 4,4 0,1 1,3 ica (mmol/l) 1 6 1,12 1,19 1,16 0,03 2, ,10 1,12 1,11 0,01 0, ,53 0,55 0,54 0,01 1,5 Cl (mmol/l) ,2 109,2 108,0 1,2 1, ,4 106,2 105,2 0,7 0, ,1 103,0 101,5 1,0 1,0 N = broj analiza x = srednja vrijednost SD = standardna devijacija KV = koeficijent varijacije 30 BIOCHEMIA MEDICA god. 15, br. 1 2, 2005.
5 M. Brajković-Bralić Analitička procjena analizatora za mjerenje koncentracije elektrolita se, u uzorcima seruma, od 0,0 3,3%, a u uzorcima pune krvi od 0,2-2,5%. Najveće vrijednosti KV% dobivene su za ionizirani kalcij u prvom i drugom uzorku seruma (3,0 i 3,3%) te u prvom uzorku pune krvi (2,2%), dok su vrijednosti koeficijenata varijacije za ph bile najmanje i u uzorcima seruma i u uzorcima pune krvi (0,3-0,6%). Nepreciznost unutar dana za kalij je bolja u uzorcima seruma nego u uzorcima pune krvi. Usporedbom dobivenih rezultata (Tablice 2, 3, 4, 5 i 6) uočava se da je nepreciznost analizatora unutar dana bolja u odnosu na nepreciznost iz dana u dan za većinu analita. Usporedna određivanja Rezultati dobiveni na ispitivanom analizatoru uspoređivani su s rezultatima dobivenima na Nova Stat Profile M i Olympus AU 400. Usporedba je provedena s pomoću uzoraka seruma i pune venske krvi. U tablici 7. prikazani su rezultati preliminarnog ispitivanja. Standardna devijacija i ostali statistički parametri izračunati su regresijskom metodom minimalnih kvadrata (least-squares regression) tako da su u obzir uzeta odstupanja oba analizatora koja se uspoređuju. Rezultati usporedbe prikazani su grafički na slikama 1-16, neobjektivnost (engl. bias) je računana kao razlika vrijednosti očitane na ispitivanom analizatoru i vrijednosti dobivene s pomoću komercijalnog aparata. Nova Stat Profile M je acidobazni analizator namijenjen za mjerenje ph, plinova u krvi, koncentracije elektrolita i metabolita u uzorcima pune krvi. Analizator koristi nerazrijeđeni uzorak (izravan način mjerenja). Olympus AU 400 je biokemijski analizator koji može vršiti 60 različitih analiza brzinom od oko 400 analiza na sat. Od elektrolita, analizator mjeri koncentracije kalija, natrija i klorida potenciometrijskom metodom koristeći razrijeđeni uzorak (neizravan način mjerenja). Natrij Rezultati usporedbe za natrij prikazani su grafički na slikama 2-5. Korelacije za natrij, pokazale su kod oba komercijalna analizatora dobre rezultate. Međutim, razlike (bias), koje su nešto veće kod Olympusa nego kod analizatora Nova Stat Profile, - što možemo vidjeti na slikama 3 i 5, - posljedica su različite tehnike mjerenja: ispitivani analizator i Nova Stat Profile koriste izravnu metodu mjerenja, tj. ne razrjeđuje se uzorak prije mjerenja, dok Olympus koristi neizravnu metodu, tj. razrjeđuje uzorak prije mjerenja puferom stalne ionske jakosti. Razlike su, prema tome, posljedica sadržaja vode u serumu, plazmi, odnosno punoj krvi, budući da su elektroliti raspodijeljeni samo u vodenoj fazi (4). Kod natrija su razlike koje nastaju zbog ove dvije metode najizrazitije budući da su i njegove koncentracije najviše. Osim toga ispitivani analizator koristi membransku elektrodu za natrij, dok oba komercijalna analizatora imaju staklene natrijeve elektrode. Postignuta standardna devijacija kao mjera odstupanja rezultata za natrij je manja od 4 mmol/l, što je manje od dozvoljene analitičke pogreške za ovaj elektrolit (6). T a b l i c a 7. Rezultati preliminarnog ispitivanja T a b l e 7. Results of preliminary testing Analit Analizator N SD xy Nagib b Y Intercept Korelacija r Na (mmol/l) Nova Stat M 54 3,20 0,81 26,7 0,89 Olympus AU ,36 0,90 14,6 0,92 K (mmol/l) Nova Stat M 65 0,19 0,88 0,35 0,95 Olympus AU ,15 0,86 0,33 0,94 Cl (mmol/l) Nova Stat M 65 2,77 0,74 26,9 0,86 Olympus AU ,69 0,68 32,4 0,88 ph Nova Stat M 69 0,038 1,04-0,432 0,96 ica (mmol/l) Nova Stat M 78 0,05 1,04-0,037 0,98 SD xy = standardna devijacija izračunata tako da su u obzir uzeta odstupanja oba analizatora koja se uspoređuju; N = broj analiziranih uzoraka BIOCHEMIA MEDICA god. 15, br. 1 2,
6 Analitička procjena analizatora za mjerenje koncentracije elektrolita M. Brajković-Bralić Natrij (NOVA Stat) cna (NOVA Stat) mmol/l y = 0,8073x + 26,786 R 2 = 0, cna (analizator) mmol/l S l i k a 2. Usporedba vrijednosti natrija (cna) između ispitivanog analizatora i Nova Stat Profile M F i g u r e 2. Comparison of sodium (cna) between the tested analyzer and Nova Stat Profile M bias Na /mmol/l cna (NOVA Stat) mmol/l S l i k a 3. Razlike (Bias) pri određivanju c(na + ) između rezultata dobivenih s pomoću ispitivanog analizatora i Nova Stat Profile M F i g u r e 3. Bias for c(na + ) determination between results obtained using the tested analyzer and Nova Stat Profile M 32 BIOCHEMIA MEDICA god. 15, br. 1 2, 2005.
7 M. Brajković-Bralić Analitička procjena analizatora za mjerenje koncentracije elektrolita Natrij/Olympus c(olympus) mmol/l y = 0,8687x + 14,636 R 2 = 0, cna(analizator) mmol/l S l i k a 4. Usporedba vrijednosti natrija (cna) između ispitivanog analizatora i Olympus-a AU 400 F i g u r e 4. Comparison of sodium (cna) between the tested analyzer and Olympus AU 400 S l i k a 5. Razlike (Bias) pri određivanju c(na + ) između rezultata dobivenih s pomoću ispitivanog analizatora i Olympus-a AU 400 F i g u r e 5. Bias for c(na + ) determination between results obtained using the tested analyzer and Olympus AU 400 BIOCHEMIA MEDICA god. 15, br. 1 2,
8 Analitička procjena analizatora za mjerenje koncentracije elektrolita M. Brajković-Bralić Kalij Rezultati usporedbe za kalij prikazani su grafički na slikama 6-9. Ispitivani analizator daje nešto niže rezultate kalija, osobito u usporedbi s analizatorom Olympus AU 400. Međutim, korelacija je u oba slučaja dobra, što dopušta primjenu koeficijenta korelacije i tako se normaliziraju rezultati na vrijednost referentnog analizatora. Veće razlike za Olympus AU 400 su i kod kalija posljedica različitih tehnika mjerenja (izravno mjerenje prema neizravnom). Postignuta standardna devijacija oko osi x i y manja je od 0,2 mmol/l, što je dozvoljena analitička pogreška za kalij (6). S l i k a 6. Usporedba vrijednosti kalija (ck) između ispitivanog analizatora i Nova Stat Profile M F i g u r e 6. Comparison of potassium (ck) between the tested analyzer and Nova Stat Profile M S l i k a 7. Razlike (Bias) pri određivanju c(k + ) između rezultata dobivenih s pomoću ispitivanog analizatora i Nova Stat Profile M F i g u r e 7. Bias for c(k + ) determination between results obtained using the tested analyzer and Nova Stat Profile M 34 BIOCHEMIA MEDICA god. 15, br. 1 2, 2005.
9 M. Brajković-Bralić Analitička procjena analizatora za mjerenje koncentracije elektrolita S l i k a 8. Usporedba vrijednosti kalija (ck) između ispitivanog analizatora i Olympus-a AU 400 F i g u r e 8. Comparison of potassium (ck) between the tested analyzer and Olympus AU 400 S l i k a 9. Razlike (Bias) pri određivanju c(k + ) između rezultata dobivenih s pomoću ispitivanog analizatora i Olympus-a AU 400 F i g u r e 9. Bias for c(k + ) determination between results obtained using the tested analyzer and Olympus AU 400 BIOCHEMIA MEDICA god. 15, br. 1 2,
10 Analitička procjena analizatora za mjerenje koncentracije elektrolita M. Brajković-Bralić Kloridi Rezultati usporedbe klorida prikazani su na slikama Korišteni ionofor za kloride u ispitivanom analizatoru je isti kao i u mnogim komercijalnim analizatorima. Kao što se vidi na slikama, postignute su dobre korelacije, a odstupanja izražena kao standardna devijacija oko osi x i y manje su od dopuštenih analitičkih pogrešaka za ovaj elektrolit (6). S l i k a 10. Usporedba vrijednosti klorida (ccl) između ispitivanog analizatora i Nova Stat Profile M F i g u r e 10. Comparison of chloride (ccl) between the tested analyzer and Nova Stat Profile M S l i k a 11. Razlike (Bias) pri određivanju c(cl - ) između rezultata dobivenih s pomoću ispitivanog analizatora i Nova Stat Profile M F i g u r e 11. Bias for c(cl - ) determination between results obtained using the tested analyzer and Nova Stat Profile M 36 BIOCHEMIA MEDICA god. 15, br. 1 2, 2005.
11 M. Brajković-Bralić Analitička procjena analizatora za mjerenje koncentracije elektrolita S l i k a 12. Usporedba vrijednosti klorida (ccl) između ispitivanog analizatora i Olympus-a AU 400 F i g u r e 12. Comparison of chloride (ccl) between the tested analyzer and Olympus AU 400 S l i k a 13. Razlike (Bias) pri određivanju c(cl - ) između rezultata dobivenih s pomoću ispitivanog analizatora i Olympus-a AU 400 F i g u r e 13. Bias for c(cl - ) determination between results obtained using the tested analyzer and Olympus AU 400 BIOCHEMIA MEDICA god. 15, br. 1 2,
12 Analitička procjena analizatora za mjerenje koncentracije elektrolita M. Brajković-Bralić ph Za mjerenje ph upotrebljena je membranska elektroda, a rezultati korelacije s analizatorom Nova Stat Profile prikazani su grafički na slikama 14 i 15. Izmjereni rezultati ph nešto su viši nego kod komercijalnog analizatora, ali visoki koeficijent korelacije pokazuje da unatoč tome postoji dobra podudarnost. S l i k a 14. Usporedba vrijednosti ph između ispitivanog analizatora i Nova Stat Profile M F i g u r e 14. Comparison of ph between the tested analyzer and Nova Stat Profile M S l i k a 15. Razlike (Bias) pri određivanju ph između rezultata dobivenih s pomoću ispitivanog analizatora i Nova Stat Profile M F i g u r e 15. Bias for ph determination between results obtained using the tested analyzer and Nova Stat Profile M 38 BIOCHEMIA MEDICA god. 15, br. 1 2, 2005.
13 M. Brajković-Bralić Analitička procjena analizatora za mjerenje koncentracije elektrolita Ionizirani kalcij Usporedba izmjerenih vrijednosti koncentracija ioniziranog kalcija na ispitivanom analizatoru i Nova Stat Profile je prikazana na slikama 16 i 17. S l i k a 16. Usporedba vrijednosti ioniziranog kalcija c(ica) između ispitivanog analizatora i Nova Stat Profile M F i g u r e 16. Comparison of ionized calcium c(ica) between the tested analyzer and Nova Stat Profile M S l i k a 17. Razlike (Bias) pri određivanju ioniziranog kalcija c(ica) između rezultata dobivenih s pomoću ispitivanog analizatora i Nova Stat Profile M F i g u r e 17. Bias for ionized calcium c(ica) determination between results obtained using the tested analyzer and Nova Stat Profile M BIOCHEMIA MEDICA god. 15, br. 1 2,
14 Analitička procjena analizatora za mjerenje koncentracije elektrolita M. Brajković-Bralić ANALYTICAL EVALUATION OF AN ANALYZER FOR ELECTROLYTE CONCENTRATION MEASUREMENT ABSTRACT - Objective: Analytical evaluation of an analyzer for electrolyte concentration measurement according to guidelines of the European Committee for Clinical Laboratory Standards (ECCLS). Methods: e following parameters were tested on serum, whole blood and control samples: ph, sodium, potassium, calcium and chloride ions. e analytical evaluation consisted of between-day imprecision and inaccuracy, within-day imprecision and correlation with other analyzers (Nova Stat Profile M and Olympus AU 400). Results: e performed analytical evaluation of the tested analyzer showed acceptable inaccuracy and imprecision in between-day and within-day measurements and acceptable correlation with other analyzers (Nova Stat Profile M and Olympus AU 400). Conclusions: e tested analyzer was established as a reliable and safe analyzer for ph and electrolytes concentrations because of its measurement accuracy and precision and good correlation with other analyzers (Nova Stat Profile M and Olympus AU 400). Short analysis time together with the ease of use makes the tested analyzer suitable for medical biochemistry laboratory as a replacement for flame photometers. Key words: analytical evaluation, analyzer for measurement of electrolyte concentration, between-day inaccuracy and imprecision, within-day imprecision, correlation LITERATURA 1. European Committee for Clinical Laboratory Standards. Guidelines for Evaluation of Analyzers in Clinical Chemistry. Berlin, Koln: ECCLS Beuth Verlag,1986; Rogić D, Barbarić V, Matica J, Rumenjak V, Bilić A. Evaluation of the Olympus AU 200 clinical chemistry analyzer. Clinical Laboratory 1997;43: Rumenjak V, Kruhak I, Milardović S. A simple and inexpensive multichannel potentiometric blood analyzer for single and nonroutine measurements of electrolyte concentrations. Laboratory Robotics and Automation 1998; 10(4): Limmet K. Evaluation of regression procedures for method comparision studies. Clin Chem 1993;39: Hristova EN, Cecco S, Niemala JE, Rehak NN, Elin RJ. Analyzerdependent difference in results for ionized calcium, ionized magnezium, sodium and ph. Clin Chem 1995;41: Koch DD, Peters T. Selection and evaluation of methods. U: Burtis CA, Ashwood ER, urednici. Tietz fundamentals of clinical chemistry. Philadelphia: WB Saunders; 1999; BIOCHEMIA MEDICA god. 15, br. 1 2, 2005.
Determination of Electrolyte Concentration in Saliva by Potentiometrie Method
Determination of Electrolyte Concentration in Saliva by Potentiometrie Method Određivanje koncentracije elektrolita u slini potenciometrijskom metodom Summary A method for concurrent determination o f
More informationRESISTANCE PREDICTION OF SEMIPLANING TRANSOM STERN HULLS
Nenad, VARDA, University of Zagreb, Faculty of Mechanical Engineering and Naval Architecture, I. Lučića 5, 10000 Zagreb Nastia, DEGIULI, University of Zagreb, Faculty of Mechanical Engineering and Naval
More informationNAPREDNI FIZIČKI PRAKTIKUM 1 studij Matematika i fizika; smjer nastavnički MJERENJE MALIH OTPORA
NAPREDNI FIZIČKI PRAKTIKUM 1 studij Matematika i fizika; smjer nastavnički MJERENJE MALIH OTPORA studij Matematika i fizika; smjer nastavnički NFP 1 1 ZADACI 1. Mjerenjem geometrijskih dimenzija i otpora
More informationShear Modulus and Shear Strength Evaluation of Solid Wood by a Modified ISO Square-Plate Twist Method
Hiroshi Yoshihara 1 Shear Modulus and Shear Strength Evaluation of Solid Wood by a Modified ISO 1531 Square-late Twist Method rocjena smicajnog modula i smicajne čvrstoće cjelovitog drva modificiranom
More informationSTATISTICAL ANALYSIS OF WET AND DRY SPELLS IN CROATIA BY THE BINARY DARMA (1,1) MODEL
Hrvatski meteoroloπki Ëasopis Croatian Meteorological Journal, 4, 2006., 43 5. UDK: 55.577.22 Stručni rad STATISTICAL ANALYSIS OF WET AND DRY SPELLS IN CROATIA BY THE BINARY DARMA (,) MODEL Statistička
More informationDetermination of Stability Constants of Chlorocadmium Complexes in Water from Electromotive Force Measurements
CROATICA CHEMICA ACTA CCACAA 74 (1) 91 101 (2001) ISSN-0011-1643 CCA-2717 Original Scientific Paper Determination of Stability Constants of Chlorocadmium Complexes in Water from Electromotive Force Measurements
More informationPotentials in Ion-specific Electrodes
CR 0 AT IC A CHEMIC A ACT A 42 (1970) 111 CCA-578 Potentials in Ion-specific Electrodes R. H. Doremus 541.134 Conference Paper General Electric Research & Development Center, Schenectady, New York, 12309,
More informationRed veze za benzen. Slika 1.
Red veze za benzen Benzen C 6 H 6 je aromatično ciklično jedinjenje. Njegove dve rezonantne forme (ili Kekuléove structure), prema teoriji valentne veze (VB) prikazuju se uobičajeno kao na slici 1 a),
More informationPRIPADNOST RJEŠENJA KVADRATNE JEDNAČINE DANOM INTERVALU
MAT KOL Banja Luka) ISSN 0354 6969 p) ISSN 1986 58 o) Vol. XXI )015) 105 115 http://www.imvibl.org/dmbl/dmbl.htm PRIPADNOST RJEŠENJA KVADRATNE JEDNAČINE DANOM INTERVALU Bernadin Ibrahimpašić 1 Senka Ibrahimpašić
More informationComparison of flow cytometric and epifluorescent counting methods for marine heterotrophic bacteria
ISSN: 0001-5113 AADRAY ACTA ADRIAT., 48(2): 107-114, 2007 UDC:574.5 (262.3 + 261.268):57.086 Comparison of flow cytometric and epifluorescent counting methods for marine heterotrophic bacteria Danijela
More informationMode I Critical Stress Intensity Factor of Medium- Density Fiberboard Obtained by Single-Edge- Notched Bending Test
... Yoshihara, Mizuno: Mode I Critical Stress Intensity Factor of Medium-Density... Hiroshi Yoshihara, Hikaru Mizuno 1 Mode I Critical Stress Intensity Factor of Medium- Density Fiberboard Obtained by
More informationANALYTICAL AND NUMERICAL PREDICTION OF SPRINGBACK IN SHEET METAL BENDING
ANALYTICAL AND NUMERICAL PREDICTION OF SPRINGBACK IN SHEET METAL BENDING Slota Ján, Jurčišin Miroslav Department of Technologies and Materials, Faculty of Mechanical Engineering, Technical University of
More informationProjektovanje paralelnih algoritama II
Projektovanje paralelnih algoritama II Primeri paralelnih algoritama, I deo Paralelni algoritmi za množenje matrica 1 Algoritmi za množenje matrica Ovde su data tri paralelna algoritma: Direktan algoritam
More informationMathcad sa algoritmima
P R I M J E R I P R I M J E R I Mathcad sa algoritmima NAREDBE - elementarne obrade - sekvence Primjer 1 Napraviti algoritam za sabiranje dva broja. NAREDBE - elementarne obrade - sekvence Primjer 1 POČETAK
More informationATOMSKA APSORP SORPCIJSKA TROSKOP
ATOMSKA APSORP SORPCIJSKA SPEKTROS TROSKOP OPIJA Written by Bette Kreuz Produced by Ruth Dusenbery University of Michigan-Dearborn 2000 Apsorpcija i emisija svjetlosti Fizika svjetlosti Spectroskopija
More informationA COMPARATIVE EVALUATION OF SOME SOLUTION METHODS IN FREE VIBRATION ANALYSIS OF ELASTICALLY SUPPORTED BEAMS 5
Goranka Štimac Rončević 1 Original scientific paper Branimir Rončević 2 UDC 534-16 Ante Skoblar 3 Sanjin Braut 4 A COMPARATIVE EVALUATION OF SOME SOLUTION METHODS IN FREE VIBRATION ANALYSIS OF ELASTICALLY
More informationEXPERIMENTAL ANALYSIS OF THE STRENGTH OF A POLYMER PRODUCED FROM RECYCLED MATERIAL
A. Jurić et al. EXPERIMENTAL ANALYSIS OF THE STRENGTH OF A POLYMER PRODUCED FROM RECYCLED MATERIAL Aleksandar Jurić, Tihomir Štefić, Zlatko Arbanas ISSN 10-651 UDC/UDK 60.17.1/.:678.74..017 Preliminary
More informationTEORIJA SKUPOVA Zadaci
TEORIJA SKUPOVA Zadai LOGIKA 1 I. godina 1. Zapišite simbolima: ( x nije element skupa S (b) d je član skupa S () F je podskup slupa S (d) Skup S sadrži skup R 2. Neka je S { x;2x 6} = = i neka je b =
More informationPEARSONOV r koeficijent korelacije [ ]
PEARSONOV r koeficijent korelacije U prošlim vježbama obradili smo Spearmanov Ro koeficijent korelacije, a sada nas čeka Pearsonov koeficijent korelacije ili Produkt-moment koeficijent korelacije. To je
More informationUNCERTAINTY IN HULL GIRDER FATIGUE ASSESSMENT OF CONTAINERSHIP
Nenad Varda, University of Zagreb, Faculty of Mechanical Engineering and Naval Architecture I. Lučića 5, 10000 Zagreb, e-mail: nenad.varda@fsb.hr Zrinka Čižmek, University of Zagreb, Faculty of Mechanical
More informationAlgoritam za množenje ulančanih matrica. Alen Kosanović Prirodoslovno-matematički fakultet Matematički odsjek
Algoritam za množenje ulančanih matrica Alen Kosanović Prirodoslovno-matematički fakultet Matematički odsjek O problemu (1) Neka je A 1, A 2,, A n niz ulančanih matrica duljine n N, gdje su dimenzije matrice
More informationMATHEMATICAL ANALYSIS OF PERFORMANCE OF A VIBRATORY BOWL FEEDER FOR FEEDING BOTTLE CAPS
http://doi.org/10.24867/jpe-2018-02-055 JPE (2018) Vol.21 (2) Choudhary, M., Narang, R., Khanna, P. Original Scientific Paper MATHEMATICAL ANALYSIS OF PERFORMANCE OF A VIBRATORY BOWL FEEDER FOR FEEDING
More informationCyclical Surfaces Created by a Conical Helix
Professional paper Accepted 23.11.2007. TATIANA OLEJNÍKOVÁ Cyclical Surfaces Created by a Conical Helix Cyclical Surfaces Created by a Conical Helix ABSTRACT The paper describes cyclical surfaces created
More informationRELIABILITY OF GLULAM BEAMS SUBJECTED TO BENDING POUZDANOST LIJEPLJENIH LAMELIRANIH NOSAČA NA SAVIJANJE
RELIABILITY OF GLULAM BEAMS SUBJECTED TO BENDING Mario Jeleč Josip Juraj Strossmayer University of Osijek, Faculty of Civil Engineering Osijek, mag.ing.aedif. Corresponding author: mjelec@gfos.hr Damir
More informationMogućnosti utvrđivanja dodane vode i Icravfjeg mlijeka u icozije mlijel(o
N. Vahčić i sur: Mogućnosti utvrđivanja... Mljekarstvo 40 (12) 315 322, 1990 Mogućnosti utvrđivanja dodane vode i Icravfjeg mlijeka u icozije mlijel(o Mr. Nada VAHCIĆ, dr. Angelina PALIĆ, dr. Milana RITZ,
More informationFIZIKALNA KOZMOLOGIJA VII. VRLO RANI SVEMIR & INFLACIJA
FIZIKALNA KOZMOLOGIJA VII. VRLO RANI SVEMIR & INFLACIJA KOZMIČKI SAT ranog svemira Ekstra zračenje u mjerenju CMB Usporedba s rezultatima LEP-a Usporedba CMB i neutrina Vj.: Pozadinsko zračenje neutrina
More informationDEVELOPMENT OF MATHEMATICAL MODELS TO PREDICT THE EFFECT OF INPUT PARAMETERS ON FEED RATE OF A RECIPROCATORY TUBE FUNNEL FEEDER
http://doi.org/10.24867/jpe-2018-01-067 JPE (2018) Vol.21 (1) Jain, A., Bansal, P., Khanna, P. Preliminary Note DEVELOPMENT OF MATHEMATICAL MODELS TO PREDICT THE EFFECT OF INPUT PARAMETERS ON FEED RATE
More informationLINEARNI MODELI STATISTIČKI PRAKTIKUM 2 2. VJEŽBE
LINEARNI MODELI STATISTIČKI PRAKTIKUM 2 2. VJEŽBE Linearni model Promatramo jednodimenzionalni linearni model. Y = β 0 + p β k x k + ε k=1 x 1, x 2,..., x p - varijable poticaja (kontrolirane) ε - sl.
More informationKeywords: anticline, numerical integration, trapezoidal rule, Simpson s rule
Application of Simpson s and trapezoidal formulas for volume calculation of subsurface structures - recommendations 2 nd Croatian congress on geomathematics and geological terminology, 28 Original scientific
More informationPOBOLJŠANJE POSTOJEĆEG POSTUPKA ODREĐIVANJA INDIKATORA U PLINSKIM ULJIMA
D. Šeremešić et al. Određivanje indikatora u plinskim uljima... Dunja Šeremešić, Vinko Rukavina,Tatjana Tomić, Nada Uzorinac Nasipak Sažetak ISSN 0350-350X GOMABN 50, 3, 243-250 Stručni rad / Professional
More informationANALYSIS OF THE RELIABILITY OF THE "ALTERNATOR- ALTERNATOR BELT" SYSTEM
I. Mavrin, D. Kovacevic, B. Makovic: Analysis of the Reliability of the "Alternator- Alternator Belt" System IVAN MAVRIN, D.Sc. DRAZEN KOVACEVIC, B.Eng. BRANKO MAKOVIC, B.Eng. Fakultet prometnih znanosti,
More informationParametri validacije instrumentalne metode za utvrđivanje ukupnog broja bakterija u mlijeku
Parametri validacije instrumentalne metode za utvrđivanje ukupnog broja bakterija u mlijeku Nataša Mikulec, Dubravka Samaržija, Neven Antunac, Šimun Zamberlin, Zrinka Kuliš, Iva Horvat Stručni rad Professional
More informationON DERIVATING OF AN ELASTIC STABILITY MATRIX FOR A TRANSVERSELY CRACKED BEAM COLUMN BASED ON TAYLOR EXPANSION
POLYTECHNIC & DESIGN Vol. 3, No. 3, 2015. DOI: 10.19279/TVZ.PD.2015-3-3-04 ON DERIVATING OF AN ELASTIC STABILITY MATRIX FOR A TRANSVERSELY CRACKED BEAM COLUMN BASED ON TAYLOR EXPANSION IZVOD MATRICE ELASTIČNE
More informationDIPLOMSKI RAD. Usporedba metoda za određivanje aktivnosti aminotransferaza sa i bez dodatka piridoksalfosfata. Anđelka Milakara
Anđelka Milakara Usporedba metoda za određivanje aktivnosti aminotransferaza sa i bez dodatka piridoksalfosfata DIPLOMSKI RAD Predan Sveučilištu u Zagrebu Farmaceutsko-biokemijskom fakultetu Zagreb, 2017.
More informationGround vibrations level characterization through the geological strength index (GSI)
1 Ground vibrations level characterization through the geological strength index (GSI) The Mining-Geology-Petroleum Engineering Bulletin UDC: 553.5:622.2 DOI: 10.17794/rgn.2017.1.1 Original scientific
More informationVIŠESTRUKO USPOREĐIVANJE
SVEUČILIŠTE U ZAGREBU PRIRODOSLOVNO-MATEMATIČKI FAKULTET MATEMATIČKI ODSJEK Almeida Hasić VIŠESTRUKO USPOREĐIVANJE Diplomski rad Zagreb, 2014. SVEUČILIŠTE U ZAGREBU PRIRODOSLOVNO-MATEMATIČKI FAKULTET MATEMATIČKI
More informationUsing the Energy Balance Method in Estimation of Overhead Transmission Line Aeolian Vibrations
Strojarstvo 50 (5) 69-76 (008) H. WOLF et. al., Using the Energy Balance Method in Estimation... 69 CODEN STJSAO ISSN 056-887 ZX470/35 UDK 6(05)=86=0=30 Using the Energy Balance Method in Estimation of
More informationCOMPARISON OF THREE CALCULATION METHODS OF ENERGY PERFORMANCE CERTIFICATES IN SLOVENIA
10 Oригинални научни рад Research paper doi 10.7251/STP1813169K ISSN 2566-4484 POREĐENJE TRI METODE PRORAČUNA ENERGETSKIH CERTIFIKATA U SLOVENIJI Wadie Kidess, wadie.kidess@gmail.com Marko Pinterić, marko.pinteric@um.si,
More informationSveučilište J.J. Strossmayera u Osijeku. Odjel za kemiju. Sveučilišni preddiplomski studij kemije. Ana Brođanac
Sveučilište J.J. Strossmayera u Osijeku Odjel za kemiju Sveučilišni preddiplomski studij kemije Ana Brođanac Utjecaj neionskih tenzida na određivanje anionskih tenzida koristeći elektrodu s tekućom membranom
More informationSOUND SOURCE INFLUENCE TO THE ROOM ACOUSTICS QUALITY MEASUREMENT
ISSN 1330-3651 (Print), ISSN 1848-6339 (Online) DOI: 10.17559/TV-20150324110051 SOUND SOURCE INFLUENCE TO THE ROOM ACOUSTICS QUALITY MEASUREMENT Siniša Fajt, Miljenko Krhen, Marin Milković Original scientific
More informationUvod u relacione baze podataka
Uvod u relacione baze podataka Ana Spasić 2. čas 1 Mala studentska baza dosije (indeks, ime, prezime, datum rodjenja, mesto rodjenja, datum upisa) predmet (id predmeta, sifra, naziv, bodovi) ispitni rok
More informationKarakterizacija pseudokapacitivnih svojstava kemijski priređenog MnO 2 te kompozita MnO 2 /polipirol
N. ŠEŠELJ et al.: Karakterizacija pseudokapacitivnih svojstava kemijski priređenog MnO 2, Kem. Ind. 65 (3-4) (2016) 127 136 127 Karakterizacija pseudokapacitivnih svojstava kemijski priređenog MnO 2 te
More informationCASTOR A PROPULSION SHAFTLINE TORSIONAL VIBRATION ASSESSMENT TOOL
Gojko MAGAZINOVIĆ, University of Split, FESB, R. Boškovića 32, 21000 Split, Croatia E-mail: gmag@fesb.hr CASTOR A PROPULSION SHAFTLINE TORSIONAL VIBRATION ASSESSMENT TOOL Summary Castor (Computer Assessment
More informationPotassium Enzymatic Assay Kit (Liquid Stable) Catalog Number: BQ 010-EAEL
Note: Serum specimens and all materials coming in contact with them should be handled and disposed as if capable of transmitting infection. Avoid contact with skin by wearing gloves and proper laboratory
More informationSTRESSES AND DEFORMABILITY OF ROCK MASS UPON OPEN PIT EXPLOITATION OF DIMENSION STONE. Siniša DUNDA, Petar HRŽENJAK and Trpimir KUJUNDŽIĆ
Rudarsko-geološko-naftni zbornik Vol. 15 str. 49-55 Zagreb, 2003. UDC 691.2:502.2.08 UDK 691.2:502.2.08 Preliminary communication Prethodno priopćenje STRESSES AND DEFORMABILITY OF ROCK MASS UPON OPEN
More informationAPPLICATION OF NIR TECHNOLOGY IN THE ANIMAL FOOD INDUSTRY
Biotechnology in Animal Husbandry 27 (4), p 1811-1817, 2011 ISSN 1450-9156 Publisher: Institute for Animal Husbandry, Belgrade-Zemun UDC 636.085 DOI: 10.2298/BAH1104811M APPLICATION OF NIR TECHNOLOGY IN
More informationMETHODS FOR ESTIMATION OF SHIPS ADDED RESISTANCE IN REGULAR WAVES
Dunja Matulja, Faculty of Engineering, Rijeka Marco Sportelli, Instituto Superior Técnico, Lisbon, Portugal Jasna Prpić-Oršić, Faculty of Engineering, Rijeka Carlos Guedes Soares, Instituto Superior Técnico,
More informationINFLUENCE OF SELECTED METAL IONS ON STABILITY OF HYDROGEN PEROXIDE IN PRESENCE OF EDTA
Works of the Faculty of Forestry University of Sarajevo No. 2, 2011 (1-14) UDK 546.215:546.4/.8 INFLUENCE OF SELECTED METAL IONS ON STABILITY OF HYDROGEN PEROXIDE IN PRESENCE OF EDTA Utjecaj selektiranih
More informationThe Prediction of. Key words: LD converter, slopping, acoustic pressure, Fourier transformation, prediction, evaluation
K. Kostúr, J. et Futó al.: The Prediction of Metal Slopping in LD Coerter on Base an Acoustic ISSN 0543-5846... METABK 45 (2) 97-101 (2006) UDC - UDK 669.184.224.66:534.6=111 The Prediction of Metal Slopping
More informationPerformance of Deming regression analysis in case of misspecified analytical error ratio in method comparison studies
Clinical Chemistry 44:5 1024 1031 (1998) Laboratory Management Performance of Deming regression analysis in case of misspecified analytical error ratio in method comparison studies Kristian Linnet Application
More informationYou created this PDF from an application that is not licensed to print to novapdf printer (http://www.novapdf.com)
HG XV SRPSKI SIMPOZIJUM O HIDROGEOLOGIJI ZBORNIK RADOVA CIP Na XV PRIMERU ANALIZE RADIOAKTIVNIH OSOBINA PODZEMNIH VODA APPLICATION OF NONPARAMETRIC STATISTICAL METHODS: EXAMINATION OF RADIOACTIVE CHARACTERISTICS
More informationEXPERIMENTAL ANALYSIS OF COMBINED ACTION OF BENDING, SHEAR AND TORSION ON TIMBER BEAMS
Eksperimentalna analiza zajedničkog djelovanja savijanja, posmika i torzije drvenih nosača EXPERIMENTAL ANALYSIS OF COMBINED ACTION OF BENDING, SHEAR AND TORSION ON TIMBER BEAMS Tihomir Štefić, Aleksandar
More information01 English - Ref.: 95. Intended use. End point reaction system for determination of calcium in blood and urine.
Intended use. End point reaction system for determination of calcium in blood and urine. Test principle. Calcium reacts with arsenazo III in a solution a little acid yielding the blue colored complex calcium-arsenazo
More informationKLASIFIKACIJA NAIVNI BAJES. NIKOLA MILIKIĆ URL:
KLASIFIKACIJA NAIVNI BAJES NIKOLA MILIKIĆ EMAIL: nikola.milikic@fon.bg.ac.rs URL: http://nikola.milikic.info ŠTA JE KLASIFIKACIJA? Zadatak određivanja klase kojoj neka instanca pripada instanca je opisana
More informationPARAMETRIC OPTIMIZATION OF EDM USING MULTI-RESPONSE SIGNAL-TO- NOISE RATIO TECHNIQUE
JPE (2016) Vol.19 (2) Payal, H., Maheshwari, S., Bharti, S.P. Original Scientific Paper PARAMETRIC OPTIMIZATION OF EDM USING MULTI-RESPONSE SIGNAL-TO- NOISE RATIO TECHNIQUE Received: 31 October 2016 /
More informationMODELLING AND INVESTIGATING THE EFFECT OF INPUT PARAMETERS ON SURFACE ROUGHNESS IN ELECTRICAL DISCHARGE MACHINING OF CK45
S. Daneshmand i dr. Modeliranje i ispitivanje učinka ulaznih parametara na površinsku hrapavost u obradi CK električnim pražnjenjem ISSN 330-3 (Print), ISSN -339 (Online) DOI:.79/TV-009 MODELLING AND INVESTIGATING
More informationCOMPARISON OF LINEAR SEAKEEPING TOOLS FOR CONTAINERSHIPS USPOREDBA PROGRAMSKIH ALATA ZA LINEARNU ANALIZU POMORSTVENOSTI KONTEJNERSKIH BRODOVA
Ana Đigaš, Sveučilište u Zagrebu, Fakultet strojarstva i brodogradnje Maro Ćorak, Sveučilište u Zagrebu, Fakultet strojarstva i brodogradnje Joško Parunov, Sveučilište u Zagrebu, Fakultet strojarstva i
More informationInternational Journal of Pharma and Bio Sciences
Research Article Biochemistry International Journal of Pharma and Bio Sciences ISSN 0975-6299 CONCORDANCE OF SERUM ELECTROLYTES ON TWO DIFFERENT LAB EQUIPMENTS DRY CHEMISTRY VS. ABG AND ITS IMPORTANCE
More informationEASYLYTE Na/K/Cl/Ca/Li SOLUTIONS PACK REF mL
EASYLYTE Na/K/Cl/Ca/Li SOLUTIONS PACK REF 2124 800mL INTENDED USE The EasyLyte Na/K/Cl/Ca/Li Solutions Pack is intended for the quantitative determination of sodium (Na + ), potassium (K + ), chloride
More informationA Device for a Vibrating Dropping Mercury Electrode
CROATCA CHEMCA ACTA CCACAA 53 (1) 101-105 (1980) CCA-1193 YU SSN 0011-1643 UDC 541.135.52 Preliminary Communication A Device for a Vibrating Dropping Mercury Electrode T. Magjer, M. Lovric, and M. Branica
More informationModified Zagreb M 2 Index Comparison with the Randi} Connectivity Index for Benzenoid Systems
CROATICA CHEMICA ACTA CCACAA 7 (2) 83 87 (2003) ISSN-00-3 CCA-2870 Note Modified Zagreb M 2 Index Comparison with the Randi} Connectivity Index for Benzenoid Systems Damir Vuki~evi} a, * and Nenad Trinajsti}
More informationVALIDACIJA I MERNA NESIGURNOST POTENCIOMETRIJSKE METODE ZA ODREĐIVANJE SADRŽAJA 2,4 D KISELINE KAO AKTIVNE MATERIJE U PESTICIDIMA
H-14 Izvod VALIDACIJA I MERNA NESIGURNOST POTENCIOMETRIJSKE METODE ZA ODREĐIVANJE SADRŽAJA 2,4 D KISELINE KAO AKTIVNE MATERIJE U PESTICIDIMA VALIDATION AND ESTEMATE OF UNCERTAINTY OF MEASURMENT OF POTENTIOMETRIC
More informationCHARACTERIZATION OF COAL FLY ASH AND ITS ABILITY TO REMOVE Ni 2+ IONS FROM AQUEOUS SOLUTION
ISSN 1848-0071 662.613+546.74=111 Recieved: 2013-06-24 Accepted: 2013-09-06 Original scientific paper CHARACTERIZATION OF COAL FLY ASH AND ITS ABILITY TO REMOVE Ni 2+ IONS FROM AQUEOUS SOLUTION AJAY K.
More information2. Reagent Store at 2-8ºC. 3. Standard - - Calcium 10 mg/dl - Store at 2-30ºC. CALCIUM Liquiform Insert. 01 English - Ref.: 90. Ref.
CALCIUM Liquiform Insert Ref.:90 Intended use. End point reaction system for determination of calcium in blood and urine. Professional Use. [For in vitro diagnostic only] Test principle. Calcium reacts
More informationCOMPARISON OF MODULY OF ELASTICITY OBTAINED BY NON- DESTRUCTIVE AND DESTRUCTIVE TESTS OF TIMBER SAMPLES
COMPARISON OF MODULY OF ELASTICITY OBTAINED BY NON- DESTRUCTIVE AND DESTRUCTIVE TESTS OF TIMBER SAMPLES Marta Čukman University of Zagreb, Faculty of Civil Engineering, student Corresponding author: marta.cukman@gmail.com
More informationZANIMLJIV NAČIN IZRAČUNAVANJA NEKIH GRANIČNIH VRIJEDNOSTI FUNKCIJA. Šefket Arslanagić, Sarajevo, BiH
MAT-KOL (Banja Luka) XXIII ()(7), -7 http://wwwimviblorg/dmbl/dmblhtm DOI: 75/МК7A ISSN 5-6969 (o) ISSN 986-588 (o) ZANIMLJIV NAČIN IZRAČUNAVANJA NEKIH GRANIČNIH VRIJEDNOSTI FUNKCIJA Šefket Arslanagić,
More informationNAPREDNI FIZIČKI PRAKTIKUM II studij Geofizika POLARIZACIJA SVJETLOSTI
NAPREDNI FIZIČKI PRAKTIKUM II studij Geofizika POLARIZACIJA SVJETLOSTI studij Geofizika NFP II 1 ZADACI 1. Izmjerite ovisnost intenziteta linearno polarizirane svjetlosti o kutu jednog analizatora. Na
More informationBENDING-SHEAR INTERACTION OF LONGITUDINALLY STIFFENED GIRDERS
Broj 3, godina 211 Stranice 97-112 BENDING-SHEAR INTERACTION OF LONGITUDINALLY STIFFENED GIRDERS Darko Beg University of Ljubljana, Faculty of Civil and Geodetic Engineering, Slovenia, University Professor
More informationMjerenje snage. Na kraju sata student treba biti u stanju: Spojevi za jednofazno izmjenično mjerenje snage. Ak. god. 2008/2009
Mjerenje snae Ak. od. 008/009 1 Na kraju sata student treba biti u stanju: Opisati i analizirati metode mjerenja snae na niskim i visokim frekvencijama Odabrati optimalnu metodu mjerenja snae Analizirati
More informationPeriodi i oblici titranja uobičajenih okvirnih AB građevina
DOI: https://doi.org/10.1456/jce.1774.016 Građevinar /018 Primljen / Received: 30.7.016. Ispravljen / Corrected: 19..017. Prihvaćen / Accepted: 8..017. Dostupno online / Available online: 10.3.018. Periodi
More informationINVESTIGATION OF UPSETTING OF CYLINDER BY CONICAL DIES
INVESTIGATION OF UPSETTING OF CYLINDER BY CONICAL DIES D. Vilotic 1, M. Plancak M 1, A. Bramley 2 and F. Osman 2 1 University of Novi Sad, Yugoslavia; 2 University of Bath, England ABSTRACT Process of
More informationHAA je potpisnica multilateralnog sporazuma s Europskom organizacijom za akreditaciju (EA) HAA is a signatory of the European co-operation for
HAA je potpisnica multilateralnog sporazuma s Europskom organizacijom za akreditaciju (EA) HAA is a signatory of the European co-operation for Accreditation (EA) Multilateral Agreement PRILOG POTVRDI O
More informationZAVRŠNI ZADATAK. U radu je potrebno navesti korištenu literaturu i eventualno dobivenu pomoć.
SVEUČILIŠTE U ZAGREBU FAKULTET STROJARSTVA I BRODOGRADNJE Središnje povjerenstvo za završne i diplomske ispite Povjerenstvo za završne i diplomske ispite studija zrakoplovstva ZAVRŠNI ZADATAK Sveučilište
More informationSTRESS OF ANGLE SECTION SUBJECTED TO TRANSVERSAL LOADING ACTING OUT OF THE SHEAR CENTER
STRESS OF ANGLE SECTION SUBJECTED TO TRANSVERSAL LOADING ACTING OUT OF THE SHEAR CENTER Filip Anić Josip Juraj Strossmayer University of Osijek, Faculty of Civil Engineering Osijek, Student Davorin Penava
More informationESTIMATION OF UNIAXIAL COMPRESSIVE AND TENSILE STRENGTH OF ROCK MATERIAL FROM GYPSUM DEPOSITS IN THE KNIN AREA
ISSN 1330-3651 (Print), ISSN 1848-6339 (Online) DOI: 10.17559/TV-20150130152828 ESTIMATION OF UNIAXIAL COMPRESSIVE AND TENSILE STRENGTH OF ROCK MATERIAL FROM GYPSUM DEPOSITS IN THE KNIN AREA Zlatko Briševac,
More information1. Biuret - Reagent (ready to use) - Store at ºC. 2. Standard g/dl - Store at ºC. TOTAL PROTEINS Insert. 01 English - Ref.
TOTAL PROTEINS Insert Ref.:99 Intended use. End point reaction system for colorimetric measurement of total proteins in blood, pleural, synovial, and ascitic fluids. Professional use. [For in vitro diagnostic
More informationANALYSIS OF INFLUENCE OF PARAMETERS ON TRANSFER FUNCTIONS OF APERIODIC MECHANISMS UDC Života Živković, Miloš Milošević, Ivan Ivanov
UNIVERSITY OF NIŠ The scientific journal FACTA UNIVERSITATIS Series: Mechanical Engineering Vol.1, N o 6, 1999 pp. 675-681 Editor of series: Nenad Radojković, e-mail: radojkovic@ni.ac.yu Address: Univerzitetski
More informationDETERMINATION OF THE EFFECTIVE STRAIN FLOW IN COLD FORMED MATERIAL
DETERMINATION OF THE EFFECTIVE STRAIN FLOW IN COLD FORMED MATERIAL Leo Gusel University of Maribor, Faculty of Mechanical Engineering Smetanova 17, SI 000 Maribor, Slovenia ABSTRACT In the article the
More informationINFLUENCE OF SOUND SOURCE LOCATION AND ABSORPTION ON DETERMINATION OF SOUND POWER LEVELS
ORIGINAL SCIENTIFIC PAPER D. Varžić, J. Prezelj* INFLUENCE OF SOUND SOURCE LOCATION AND ABSORPTION ON DETERMINATION OF SOUND POWER LEVELS UDK 628.517.2 RECEIVED: 2014-03-24 ACCEPTED: 2015-08-17 SUMMARY:
More informationSveučilište Josipa Jurja Strossmayera u Osijeku. Odjel za kemiju. Sveučilišni preddiplomski studij kemije. Sanja Gazić
Sveučilište Josipa Jurja Strossmayera u Osijeku Odjel za kemiju Sveučilišni preddiplomski studij kemije Sanja Gazić OdreĎivanje anionskih tenzida u otpadnim vodama koristeći elektrodu s tekućom membranom
More informationU X. 1. Multivarijantna statistička analiza 1
. Multivarijantna statistička analiza Standardizovana (normalizovana) vrednost obeležja Normalizovano odstupanje je mera varijacije koja pokazuje algebarsko odstupanje jedne vrednosti obeležja od aritmetičke
More informationSveučilište Josipa Jurja Strossmayera u Osijeku. Odjel za kemiju. Diplomski sveučilišni studij kemija; istraživački smjer.
Sveučilište Josipa Jurja Strossmayera u Osijeku Odjel za kemiju Diplomski sveučilišni studij kemija; istraživački smjer Antonia Matić Određivanje kationskih tenzida u sredstvima za dezinfekciju elektrodom
More informationAS 117. Traceability to the primary standards at Radiometer. Author: Hans Bjarne Kristensen. January Contents
AS 117 Author: Traceability to the primary standards at Radiometer Hans Bjarne Kristensen January 2006 The AS series of scientific and technical articles from Radiometer is named after Professor Poul Astrup,
More informationNON-SPECIFIC METHODS FOR DETECTING RESIDUES OF CLEANING AGENTS DURING CLEANING VALIDATION
Available on line at Association of the Chemical Engineers AChE www.ache.org.rs/ciceq Chemical Industry & Chemical Engineering Quarterly 17 (1) 39 44 (2011) CI&CEQ DRAGAN M. MILENOVIĆ 1 DRAGAN S. PEŠIĆ
More informationA NEW THREE-DIMENSIONAL CHAOTIC SYSTEM WITHOUT EQUILIBRIUM POINTS, ITS DYNAMICAL ANALYSES AND ELECTRONIC CIRCUIT APPLICATION
A. Akgul, I. Pehlivan Novi trodimenzijski kaotični sustav bez točaka ekvilibrija, njegove dinamičke analize i primjena elektroničkih krugova ISSN 1-61 (Print), ISSN 1848-69 (Online) DOI: 1.179/TV-1411194
More informationMetode praćenja planova
Metode praćenja planova Klasična metoda praćenja Suvremene metode praćenja gantogram mrežni dijagram Metoda vrednovanja funkcionalnosti sustava Gantogram VREMENSKO TRAJANJE AKTIVNOSTI A K T I V N O S T
More informationADAPTIVE NEURO-FUZZY MODELING OF THERMAL VOLTAGE PARAMETERS FOR TOOL LIFE ASSESSMENT IN FACE MILLING
http://doi.org/10.24867/jpe-2017-01-016 JPE (2017) Vol.20 (1) Original Scientific Paper Kovač, P., Rodić, D., Gostimirović, M., Savković, B., Ješić. D. ADAPTIVE NEURO-FUZZY MODELING OF THERMAL VOLTAGE
More informationINNOVATION OF PARAMETER α cc FOR DESIGN RESISTANCE OF HIGH-STRENGTH CONCRETE COLUMNS
R. Sinđić-Grebović, Ž. Radovanović Inoviranje parametra α cc za proračun otpornosti stubova od betona visoke čvrstoće ISSN 1330-3651 (Print), ISSN 1848-6339 (Online) https://doi.org/10.17559/tv-20150125105926
More informationOracle Spatial Koordinatni sustavi, projekcije i transformacije. Dalibor Kušić, mag. ing. listopad 2010.
Oracle Spatial Koordinatni sustavi, projekcije i transformacije Dalibor Kušić, mag. ing. listopad 2010. Pregled Uvod Koordinatni sustavi Transformacije Projekcije Modeliranje 00:25 Oracle Spatial 2 Uvod
More informationD. Bonchev* D epartment of Physicai Chemistry, Higher Schooi of Chemicai Technoiogy, 8015 Burgas, Buigaria. Received May 12, 1976
CROATCA CHEMCA ACTA CCACAA 49 (1) 83-88 (1977) CCA-978 CND/2 Study of C2H2 + H2 System D. Bonchev* YU SSN 11-1643 516.11 Note D epartment of Physicai Chemistry, Higher Schooi of Chemicai Technoiogy, 815
More informationVELOCITY PROFILES AT THE OUTLET OF THE DIFFERENT DESIGNED DIES FOR ALUMINIUM EXTRUSION
VELOCITY PROFILES AT THE OUTLET OF THE DIFFERENT DESIGNED DIES FOR ALUMINIUM EXTRUSION J.Caloska, J. Lazarev, Faculty of Mechanical Engineering, University Cyril and Methodius, Skopje, Republic of Macedonia
More informationTHE EFFECT OF HARMFUL ZEOLITE WASTE AS REPLACEMENT ADDITION TO CEMENT ON HYDRATION AND PROPERTIES OF HARDENED CEMENT COMPOSITES
THE EFFECT OF HARMFUL ZEOLITE WASTE AS REPLACEMENT ADDITION TO CEMENT ON HYDRATION AND PROPERTIES OF HARDENED CEMENT COMPOSITES Petar Krolo, Pero Dabić, Damir Barbir Faculty of Chemical Technology in Split,
More informationLINEAR AND NON-LINEAR ANALYSES OF CABLE-STAYED STEEL FRAME SUBJECTED TO SEISMIC ACTIONS
LINEAR AND NON-LINEAR ANALYSES OF CABLE-STAYED STEEL FRAME SUBJECTED TO SEISMIC ACTIONS Marko Đuran Plan Plus d.o.o., Zagreb, MSc Ivan Lukačević University of Zagreb, Faculty of Civil Engineering, Ph.D
More informationPRECONCENTRATION OF COPPER FROM SYNTHETIC AQUEOUS SOLUTIONS USING MULTIWALLED CARBON NANOTUBES
ISSN 1848-0071 669.334+661.666=111 Recieved: 2014-07-11 Accepted: 2014-08-14 Original scientific paper PRECONCENTRATION OF COPPER FROM SYNTHETIC AQUEOUS SOLUTIONS USING MULTIWALLED CARBON NANOTUBES JANA
More informationAutomatizirana detekcija hemolize i njezina primjena u rutinskoj praksi
Automatizirana detekcija hemolize i njezina primjena u rutinskoj praksi mag. Jasna Jonjić, spec. med. biokemije Odjel za kliničku laboratorijsku dijagnostiku Hemoliza Oslobađanje intracelularnog sadržaja
More informationUPUTE ZA OBLIKOVANJE DIPLOMSKOG RADA
1 UPUTE ZA OBLIKOVANJE DIPLOMSKOG RADA Opseg je diplomskog rada ograničen na 30 stranica teksta (broje se i arapskim brojevima označavaju stranice od početka Uvoda do kraja rada). Veličina je stranice
More informationPrimjena tehnika mekog računalstva za predviđanje svojstva zrakopropusnosti tkanina
S. ALTAŞ: Primjena tehnika mekog računalstva za predviđanje svojstva zrakopropusnosti tkanina, Tekstil 66 (1-2) 1-8 (2017.) 1 Primjena tehnika mekog računalstva za predviđanje svojstva zrakopropusnosti
More informationThe Bond Number Relationship for the O-H... O Systems
CROATICA CHEMICA ACTA CCACAA 61 (4) 815-819 (1988) CCA-1828 YU ISSN 0011-1643 UDC 541.571.9 Original Scientific Paper The Bond Number Relationship for the O-H... O Systems Slawomir J. Grabowski Institute
More informationKontrolni uređaji s vremenskom odgodom za rasvjetu i klimu
KOTROI SKOPOVI ZA RASVJETU I KIMA UREĐAJE Kontrolni i s vremenskom odgodom za rasvjetu i klimu Modularni dizajn, slobodna izmjena konfiguracije Sigurno. iski napon V Efikasno čuvanje energije Sigurnost.
More information