The Performance of Laboratories Analysing α-quartz in the Workplace Analysis Scheme for Proficiency (WASP)

Size: px
Start display at page:

Download "The Performance of Laboratories Analysing α-quartz in the Workplace Analysis Scheme for Proficiency (WASP)"

Transcription

1 Ann. occup. Hyg., Vol. 47, No. 4, pp , 003 Crown Copyright 003. Reproduced with the permission of the Controller of Her Majesty s Stationery Office. Published by Oxford University Press DOI: /annhyg/meg06 The Performance of Laboratories Analysing α-quartz in the Workplace Analysis Scheme for Proficiency (WASP) PETER STACEY*, BARRY TYLEE, DELPHINE BARD and RUSSELL ATKINSON The Health and Safety Laboratory, Broad Lane, Sheffield S3 7HQ, UK Received 9 September 00; in final form 8 November 00 The Workplace Analysis Scheme for Proficiency (WASP) is a proficiency testing (PT) scheme for the analysis of occupational hygiene and environmental air samples and is operated in the United Kingdom by the Health and Safety Laboratory (HSL) on behalf of the Health & Safety Executive (HSE). One of the 6 analytes available to laboratories is silica (α-quartz) on 5 mm Gelman GLA5000 filters. This paper investigates the performance of laboratories participating in the scheme since the HSL took over the production of the samples in The average relative standard deviation (RSD) of results obtained by a laboratory is 11.5%. This is reduced to 8.5% when the values from laboratories using indirect analytical methods are excluded. Laboratories using indirect analytical methods accounted for some of the most variable data. For the on-filter analytical methods the data suggested a relationship between relative standard deviation and loading that increased gradually from ±4% at high analyte levels to ±10 15% at low levels. The average precision estimate for the on-filter analytical methods was found to be 5.6% RSD for the infrared technique and 6.7% RSD for the X-ray diffraction technique. These figures compare favourably with those reported in the published HSE methods. No significant difference was found between the average result reported by laboratories using on-filter infrared (IR) analysis and the average result reported by laboratories using on-filter X-ray diffraction (XRD) analysis. An ANOVA analysis found the repeatability estimate was just as large as the between laboratory variation for both the XRD and IR on-filter analysis techniques. When a limited number of realistic samples were included in the scheme, XRD analysis was found to perform slightly better than IR analysis. The performance of laboratories in the WASP scheme compares very favourably with other published data from a PT scheme where indirect silica analytical methods are predominately used. Keywords: quartz; workplace air; X-ray diffraction analysis; infrared analysis; analytical performance; proficiency testing INTRODUCTION Silica is one of the most abundant minerals on earth. Free silica exists in a number of crystalline and amorphous forms, the most common of which is quartz, found in a range of minerals including sandstones, grit stones and clays. The quartz content of sandstone can be up to 95% and granite stone may be as high as 30%. Many work activities, such as drilling, chiseling, grinding, polishing of glass, fettling in potteries, demolition of buildings and power saw cutting of *Author to whom correspondence should be addressed. Tel: ; fax ; peter.stacey@hsl.gov.uk stone, concrete or brick, result in the generation of respirable silica. Silicosis is commonly accepted to be one of the oldest occupational diseases and crystalline silica is now classified as a category/class 1 carcinogen by the International Agency for Research on Cancer (IARC, 1997). The lung disease is caused by inhalation and retention in the alveoli and generally ranges from nodular silicosis to fibroses of the lung tissue. The position of the Health & Safety Executive (HSE) on the classification of silica is currently under review. In the United Kingdom, the maximum permitted level for exposure to respirable crystalline silica is agreed by the UK Health and Safety Commission and is 0.3 mg/m 3 for an 8 h time- 69

2 70 P. Stacey et al. weighted average (HSE, 00). Health surveillance of workers is also recommended if exposure to crystalline silica is likely to exceed 0.1 mg/m 3 (HSE, 1997). The Workplace Analysis Scheme for Proficiency (WASP) was established in 1988 and is sponsored by the HSE. Where analytical measurements are required to assess the exposure of a worker, it is important that the analytical performance of the laboratories performing these measurements is assessed independently so that there is confidence in the data. WASP provides a quality assurance service for the chemical analysis of occupational and environmental air samples. It provides a range of 5 analytes to support most analytical techniques used in the analysis of occupational hygiene samples. In 1993, the scheme introduced α-quartz on 5 mm diameter Gelman DM800 filters using a contracted laboratory to prepare the standards, and in 1998 the Health and Safety Laboratory (HSL) took over the production of the filters. Although the laboratories in the scheme are free to choose whatever method they wish, the scheme is specifically intended to support the HSE analytical methods published in the Methods for the Determination of Hazardous Substances (MDHS) guidance documents MDHS 37 (HSE, 1987) for infrared (IR) analysis and MDHS 51/ (HSE, 1988) for the X-ray diffraction (XRD) technique are onfilter analytical methods for quartz where the sample is analysed without any preparation. MDHS 38 (HSE, 1984) is an indirect method where the quartz is recovered from the filter and prepared as a potassium bromide pellet before IR analysis. This paper discusses the analytical performance of laboratories in the scheme, the differences in analytical performance between the XRD and the IR techniques and the effect on analytical performance during the limited introduction into the scheme of a proficiency testing (PT) sample with added crystalline phases. PREPARATION OF α-quartz STANDARDS AND SAMPLE HOMOGENEITY α-quartz standard The standard UK reference material used is A9950 Sikron F-600. This reference material, from Quartzwerke (Frechen, Germany), has a respirable α-quartz content of 96.3 ± 1.4% (Jeyaratnam and Nagar, 1993) when sampled through a cyclone and analysed on a filter. The quantification of the crystalline proportion of this material is based on comparison with the US National Institute of Standards and Technology (NIST) standard: respirable α-quartz, SRM The mass median diameter of the A9950 Sikron F-600, determined by Coulter counter analysis, is.6 µm and the mass average diameter is.8 µm, with 71% of the distribution below 4.3 µm (equivalent to 7 µm unit-density spheres). Generation of replicate samples The α-quartz aerosol is produced by a modified fluidized bed aerosol generator, and then sampled using a multi-port sampler developed by the National Institute of Occupational Health in Norway (Bjørgo et al., 1980; Anglov et al., 1993). This sampler loads each filter with a similar amount of α-quartz dust. The multi-port sampler consists of a large cylindrical chamber, the lid of which contains 114 ports. Behind each port is a critical orifice that restricts the flow through the hole to 1.9 l/min (±%) when a high vacuum is applied. Each port is fitted with a Higgins Dewell cyclone sampler containing a sample filter. The filters are individually weighed and filter sets are chosen so that laboratories using IR are provided with blank filters that closely match the absorption background of the sample filters. The quartz aerosol is generated into a dust box and then sampled by the multiport sampler. Each experimental run provides one loading of α-quartz onto the filters within the range µg. All samples are screened using IR analysis and the typical relative standard deviation (RSD) of the IR measurements on all the samples produced for a single loading is ±5.6%. This also includes the analysis error which is estimated to have a RSD of ± 5% (HSE, 1987). APPROACH TO THE EVALUATION OF THE VARIANCE OF DATA Several different approaches were used to investigate the variability of the data. Within-laboratory variability For each laboratory, the variation of its results from the assigned values was calculated as a percentage relative standard deviation (% RSD). This is the standard deviation of a laboratory s results and will include contributions from the difference of a laboratory s results from the assigned values and the precision of the method or technique used. This estimate was used to compare the performance of individual laboratories and their analytical methods. Total relative standard deviation (% RSD T ) The variation of results obtained by laboratories at each analyte level was calculated as a total percentage relative standard deviation (% RSD T ). This is the standard deviation (σ) of the data at each analyte level divided by the average result (excluding extreme values) reported by laboratories. This estimate contains the variability of results due to differences between laboratories and the within laboratory variability of the different methods or techniques. It can be used to indicate the confidence interval of a result obtained at a particular measurement level, as it is a close estimate of reproducibility sometimes

3 The performance of laboratories analysing α-quartz 71 referred to as trueness. Its use applied to proficiency testing data is described by Shulman et al. (199). Reproducibility and repeatability The performance characteristics (such as reproducibility) of a particular technique or method are often obtained from duplicate results submitted by different laboratories using the same method on identical test items by using analysis of variance (ANOVA) statistical techniques as stipulated in ISO 575- (ISO, 1994). The ISO method states that the reproducibility variance or trueness is determined as: s R = s L + s r where s R is the estimate of the reproducibility variance, s L is the between-laboratory variance estimate and s r is the repeatability variance estimate. Generally, PT schemes are not able to use this method of evaluation, as they do not usually send samples with the same loading or ask laboratories to report duplicate results; consequently the repeatability variance cannot be separated from the overall variance of the data. This is because PT schemes are designed to test the ability of a laboratory to meet specific performance requirements rather than establish contributions to the overall variance. Fortunately, for α-quartz analysis in this PT scheme, each experimental run provides enough samples for more than one round. Therefore, laboratories made measurements on identical samples in different rounds. Pairing of the results from these samples will allow us to estimate the repeatability and reproducibility variances using ANOVA evaluation. However, the estimates from this approach, applied to these data, will include a between round variation because conditions between rounds (some 3 months apart) may change, e.g. due to varying calibrations, instrumental performance or analysts. Samples used in different rounds were paired if the robust averages of results for each set of samples were within 10 µg. Estimates of s R and s r were determined from an ANOVA evaluation from five pairs of loadings, chosen so they were evenly distributed across the µg analytical range. Precision An estimate of within-laboratory precision was made for the analytical range used in the scheme by assuming precision was independent of loading. The precision estimates were calculated by determining the mean error sum of squares of the standardized ratios of the four samples, for each laboratory in each round. This estimate of within-laboratory variance, s W, is comprised of several components: Table 1. Participation in round 49 UK laboratories Non-UK laboratories Type of organization No. of participants Country No. of Participants Industry 4 Australia 1 Central 1 Denmark 1 government Consultancies 1 France 1 Germany 1 Ireland 1 Netherlands 1 Spain 1 USA 1 Total 6 8 S W = p p = 1 σ rp σ Sp ( + ) p where σ Sp is the between-sample contribution and σ rp the intra-laboratory contribution for each of the p laboratories participating in each round. Any estimate of σ rp will also contain a small contribution due to sample inhomogeneity. This precision is an estimate of the random error associated with a particular method or technique. DATA COLLECTED Data quality Since the start of the scheme, the number of laboratories analysing samples of α-quartz on filters has fallen from 3 laboratories, to 14 laboratories in round 49. At the beginning of the scheme a larger proportion of laboratories were from industry and the UK; however, by round 49 most are national institutions from Europe. The types of participants in round 49 are listed in Table 1. This work examines the 680 results reported by laboratories since the HSL undertook the production of the α-quartz standards and covers 14 rounds (36 49). A round consists of four samples, each with a different amount of silica dust, sent to laboratories every 3 months. Within-laboratory variation The ratio of each result obtained by a laboratory versus its assigned value was calculated so that the variation of a laboratory s data from the assigned values could be compared with other laboratories. The assigned value used in this scheme is a robust sample average of the results for a sample level reported by laboratories, excluding those results outside prescribed limits. The variation of laboratory

4 7 P. Stacey et al. sample at 500 C and then recovering the sample by filtering a suspension of the dust through a silver filter. Laboratories B and C obtained results with the largest negative bias. This is illustrated in Fig. 1 as their ratios with the assigned values are distributed well below the ideal (1). Laboratory B used a method where the filter is ashed and prepared as a potassium bromide pellet, and laboratory C used an XRD method where the filter is dissolved with hydrofluoric acid and the dust then transferred to a lowbackground substrate. Laboratories D and E obtained satisfactory WASP performance ratings. Laboratory D had just started the scheme and at the time of writing this paper and had only submitted one set of indirect results with a RSD of %. Laboratory E used a method based on MDHS 38. This laboratory submitted results for nine rounds and the RSD of its results from the assigned values was 3.6%. If the laboratories using indirect analytical methods are excluded, the average within-laboratory RSD is reduced to 8.5% and the median within-laboratory RSD to 8.3%. Fig. 1. Variation of laboratory data compared with the assigned value. results and bias from these consensus values is shown in Fig. 1. Laboratories producing results with very little bias will have their results distributed around the ideal ratio (1). The average within-laboratory RSD was 11.5%, but several poorly performing laboratories unduly influenced this value. The median within-laboratory RSD is less influenced by extreme values, suggesting that a typical variation of results submitted by an individual laboratory was about ±9.4%. The data were examined further to investigate possible reasons for the high RSD values produced by some laboratories. The laboratory with the largest variation of results (39%) had stated it used a direct on-filter analytical method using XRD. When investigating the data it was found that, although the PT samples were analysed without any sample preparation, the amount of quartz on the filter was quantified using a calibration where the standards had been prepared by ashing in an oven at 500 C and the dust transferred as a liquid suspension onto a silver filter. Five laboratories, identified here as A, B, C, D and E, used indirect analytical methods in the rounds investigated in this paper. Three of these five laboratories had within-laboratory RSDs >9.4%. Laboratory A had the second widest spread of results (8%) and used an indirect XRD method based on ashing the Inter-laboratory variation Figure compares the RSD T for all data at each loading level, (excluding extreme results identified using a Grubbs test over the 1% significance level) with the RSD T for the direct on-filter analytical methods. Of all data (684 results), 4.86% were identified as extreme values, and for on-filter analytical methods (516 results), 1.5% were identified as extreme results. A linear trend line was drawn on the chart for all data to allow comparison. This line shows that there was no relationship for the data following the equation (% RSD T )y = x (P = 0.061) at the 5% significance level where the slope was not significantly different from zero, and the average RSD T was 11.9%. However, the trend line drawn for the on-filter analytical methods is significant (P 0.01) and suggested a slope where at the highest loading value (400 µg), the RSD T is about ±4% and at the lowest loading values (80 90 g) is about ±10 to ±15%. At levels near the United Kingdom 8 h maximum exposure limit (MEL) of 0.3 mg/m 3 (equivalent to µg on the filter assuming a flow rate of. l/min), Fig. indicates the RSD T of results reported by laboratories using on-filter analytical methods is about ±7%. The relationship where the %RSD T increases gradually as the loading decreases is expected when instrumental techniques start to approach their limit of quantification as the reduced signal and increased noise at lower analyte levels reduce the relative precision of the analysis. Any linear relationship with zero or almost zero slope shown in Fig. on the chart for all data would have suggested that the results are substantially above the limit of detection. However, there is a large scatter of values around the line for the

5 The performance of laboratories analysing α-quartz 73 Fig.. Relationship of %RSD T with loading for all data and for data using direct-on-filter methods. chart for all data, suggesting the trend line drawn on the chart line is not a good indication of the true relationship. The lower the scatter of points around the line, the better is the indication that the laboratories are successfully controlling the errors in analysis. Between-round inter-laboratory variation It was noted that higher %RSD T values occurred in rounds 4, 43 and 45. The variation of average %RSD T of the four samples in each round is shown in Fig. 3, where the average %RSD T figures in rounds 4, 43 and 45 were higher for all data than for the on-filter analysis method users. Until round 4 only one laboratory was using an indirect analytical method and this number gradually increased to three in round 45. The introduction into the scheme of a new laboratory using an indirect method of analysis may account for the higher average %RSD T shown for all data in Fig. 3 because these laboratories would not be familiar with the scheme and their performance is likely to be worse. However, there is another factor that may have adversely affected results from round 4. This round had two sample filters loaded with 100 µg of quartz where a larger variability is expected with methods involving a recovery process because a loss of a small amount of sample is more significant at low analyte levels. Trend in the average round %RSD T The data for laboratories using the direct on-filter methods showed a general reduction of the average RSD T from 13% in round 41 to 4.5% in round 49. The Spearman rank correlation coefficient of average %RSD T against round number, over all rounds, for laboratories using the direct on-filter analytical methods, was 0.71 and the probability of a relationship not existing was <0.01. This correlation coefficient indicates a significant trend in the improvement of performance for laboratories using on-filter methods from round 36. Within-laboratory precision The precision estimates for all results submitted from round 36 are shown in Table. This table gives an indication of the average and median laboratory

6 74 P. Stacey et al. Fig. 3. Between-round variation of average %RSD T. Table. Precision estimates %RSD Technique All data All on-filter a IR on-filter a XRD on-filter a Average Median a On-filter: the figures obtained exclude data from laboratories using indirect measurement methods precision over the analytical range used in the scheme. The results obtained for the on-filter methods are comparable with the precision estimates in MHDS 37 and MDHS 51/ as both of these methods report a figure of ±5% RSD for the strongest measurement signals. These data support the figures reported in these MDHS and also the assumption that the variability of silica results between samples within a round is not greatly different from the instrumental variation on a single sample reported in the MDHS. COMPARISON BETWEEN ON-FILTER INFRARED AND ON-FILTER X-RAY DIFFRACTION ANALYTICAL METHODS Difference between the two techniques The average number of laboratories in each round using IR analysis was about four and the average number using XRD was five. The average value of each sample level for each technique in each round was calculated and compared over all 14 rounds. The difference between the two techniques, compared in Fig. 4, followed the relationship y = x with a regression coefficient (r ), of 0.97, where y is the average value of XRD analysis and x is the average value of IR analysis. The average difference between the two sets of data was 6 µg and the SD was ±15 µg. A t-test at the 5% significance level indicated that the intercept of the trend line in Fig. 4 with the axis was not different from the origin and therefore a consistent difference is not present. A similar test on Fig. 4. Difference between infrared and X-ray diffraction direct on-filter methods. the slope, also at the 5% significance level, was also not significant, indicating that rotational (alternatively known as proportional) difference was not present. These data showed there was no significant difference between the average of the values reported by the IR techniques and the average of the values reported by the XRD techniques when analysing pure α-quartz standard A9950 on filters. Reproducibility estimates The values obtained from ANOVA of sets of paired data, taken from samples prepared at the same time and analysed by the same laboratories but distributed in different rounds, are shown in Fig. 5. These values suggested that the reproducibility estimates using XRD were more variable than measurements using IR analysis and its relationship with loading was less apparent than the line for IR analysis which followed the equation %RSD = x +

7 The performance of laboratories analysing α-quartz 75 Fig. 5. Reproducibility and repeatability of direct-on-filter methods (r = 0.53). However, this relationship for reproducibility is only significant at the 10% level (P = 0.094). This trend line suggested that the reproducibility between laboratories using the IR technique at 80 µg is about ±8%, and at 50 µg the variation is about ±4%. Conclusions from these data should be regarded with caution because of the small number of laboratories involved in the ANOVA evaluation and the theoretically large uncertainty of the values obtained. However, these reproducibility estimates are comparable with the %RSD T estimates shown in Fig.. Repeatability estimates The linear trend line with the best correlation for the within-laboratory repeatability (Fig. 5) and measurement level was obtained by IR analysis. This line followed the equation y = 0.03x (r = 0.97), and the probability value for a relationship not existing was <0.01. These data suggested a strong relationship between loading and repeatability of about ±8% at the lowest loadings and ±4% at the highest loadings. The repeatability estimate for IR analysis is comparable with the figure reported in MDHS 37, where the precision of IR measurement on a single sample of 50 µg is given as ±5%. At the lowest level (88 µg), the repeatability estimate is ±8%. The figure reported in the MDHS is the instrumental variation of the measurements obtained on a single sample and the small additional variability found in this paper may be due to several factors. Not all laboratories may use exactly the same instruments, quartz standard or procedures. Also, conditions between rounds (some 3 months apart) may also change (e.g. instrumental and personnel) and this may introduce additional variability. The repeatability estimate for XRD measurements were between 4 and 10%, and their relationship with loading was much more variable (P = 0.31). In all cases the repeatability estimate was just as large as the estimate of between-laboratory variation. THE EFFECT ON PERFORMANCE WHEN SAMPLES WITH ADDITIONAL COMPONENTS ARE INTRODUCED Over a period of three rounds, two different types of dust were sampled onto the filters and sent to laboratories to ascertain if the performance of laboratories was affected when more realistic samples are analysed. An artificial mixture of pure calcite and quartz was sampled to produce two sets of replicate samples. Each set had a different amount of α-quartz dust. One sample from each set, was distributed to 1 laboratories in rounds 39, 40 and 41. The dust created by cutting a concrete paving slab with a powered saw was also sampled to produced another set of replicate

8 76 P. Stacey et al. Table 3. Variability of data from samples with added crystalline dusts Round 39 Round 40 Round 41 Average % RSD T Average % RSD T Average % RSD T Calcite mixture µg µg µg 11 Calcite mixture a 65 µg µg µg 1.3 Concrete sample b 100 µg µg 0.5 a Data excludes one laboratory with consistently low results in rounds 40 and 41. b Data excludes one laboratory from round 40 with very low results. Table 4. Comparison of data Performance on WASP test Samples Calcite mixture with 11 µg quartz Calcite mixture with 68 µg quartz Concrete dust with 105 µg quartz PAT scheme at 100 µg NIOSH intercomparison % RSD T XRD IR Precision %RSD XRD IR filter samples with a different type of dust. X-ray diffraction analysis of the paving slab identified calcium hydroxide (Portlandite) Ca(OH ), calcite CaCO 3, quartz SiO, and a trace of a feldspar mineral, possibly albite NaAlSi 3 O 8. A single sample of this type was sent to 1 laboratories in rounds 40 and 41. Variability of data The values of total laboratory variability (%RSD T ) in Table 3, from laboratories analysing quartz in the calcite and quartz mixtures, were comparable with the values of %RSD T obtained from laboratories when they analysed the usual WASP samples (Fig. ). The data for the analysis of quartz in the concrete dust was more variable. One of the samples with the concrete dust had a total variability (%RSD T ) of ±0%. The data exclude the only laboratory in these rounds using an indirect analysis method. Its results for the calcite and quartz sample mixtures were about half those reported by the other techniques, but the results for the samples with the concrete dust were very close to the average results reported by other laboratories. One difference between the two types of samples is that the deposit on the filters from concrete dust was slightly coloured, unlike the dust from the artificially generated calcite and silica mixtures. This may have helped the laboratory recover the sample more effectively. Table 4 shows values of RSD T and precision RSD for the laboratories using on-filter XRD and IR analysis. When the realistic samples were analysed, the five laboratories using on-filter XRD analysis obtained slightly better RSD T and precision than the four laboratories using on-filter IR analysis. The values for RSD T and precision were also better for XRD when compared with values reported from the analysis of the usual WASP samples where only quartz reference material is present. It is expected that XRD analysis performs better when real samples are analysed, because the background noise caused by the filter is reduced when other non-amorphous dusts are also present. This improves the peak-to-background noise ratio, which is important in quantification. As expected, with a realistic sample containing a trace silicate (feldspar) mineral, the measurements with IR analysis were more variable than the measurements with XRD analysis since silicate minerals interfere with IR analysis. On-filter XRD analysis obtained a very good RSD for precision with this sample. Comparison of %RSD T with the values obtained from other PT schemes The American PT scheme for occupational hygiene analysis, the Proficiency Analytical Testing Program (PAT) managed by the American Industrial Hygiene Association (AIHA), provides samples of silica dust on filters with coal dust, calcite or magnesium silicate as an added component. The most common methods used in the USA are the National Institute for Occupational Safety and Health (NIOSH) methods 7500 for XRD analysis (NIOSH, 1998) and 760 or 7603 for IR analysis (NIOSH, 1994a,b). All these are indirect methods where the filter is removed by treatment and the sample is transferred to a material that produces a lower background than the filter. Table 4 compares the RSD T and the precision RSD for XRD and IR analysis reported by the PAT scheme (Shulman et al., 199) and a NIOSH inter-comparison (Anderson, 1983) with the values obtained from laboratories using the on-filter analytical methods in

9 The performance of laboratories analysing α-quartz 77 the WASP scheme. The RSD T values for the WASP scheme are better than those reported by PAT (Eller et al., 1999) or the NISOH laboratory inter-comparison. However, the PAT scheme involves 80 laboratories while the WASP scheme has a much smaller number. In the WASP scheme only five laboratories were using on-filter XRD and four were using onfilter IR analysis. The limited number of on-filter XRD analysis data was also more precise than the data reported by the inter-comparison organized by NIOSH to test their methods with samples simulated to represent real matrices. However, the precision for the on-filter IR analysis on the realistic WASP samples was slightly poorer. The limited amount of WASP data suggested that over the analytical range used in the WASP scheme there is little advantage in using a NIOSH type method. CONCLUSIONS The average within-laboratory RSD for laboratories using direct on-filter analytical methods was ±8.5% Some laboratories using indirect analytical methods with a sample preparation step produced the most variable data. The data suggests that for on-filter methods of analysis, there is a relationship between loading and the total variation of results between laboratories (RSD T ) that is about ±4% at analyte levels of 400 µg and about ±10 to ±15% at analyte levels of 100 µg. The total variability (RSD T ) of results between laboratories involved in the WASP scheme, using on-filter analytical methods, analysing quartz and producing results at the MEL for crystalline silica is about ±7%. Estimates of precision compared very favourably with those published in the MDHS guidance documents 51/ and 37. There is no significant difference between the average result of laboratories using direct on-filter IR analysis and the average result of those laboratories using direct on-filter XRD analysis when analysing WASP proficiency test samples. However, the data suggested that on-filter XRD analysis produced slightly more variable data than on-filter IR analysis. The performance of laboratories analysing samples loaded with a mixture of calcite and α-quartz was similar to the performance of laboratories analysing the usual WASP samples, where only α- quartz is present. The data suggests that the variation of results between laboratories (RSD T ) and precision RSD of on-filter XRD analysis is slightly better than on-filter IR analysis when real samples are analysed. The filter sample loaded with concrete dust containing a trace feldspar mineral showed that (in these circumstances) direct on filter IR results are more variable than on-filter XRD results. The data from the on-filter analytical methods compared very favourably with other published data from indirect analytical methods. REFERENCES Anderson C. (1983) Collaborative tests for determining free silica in airborne dust. Cincinnati, OH: US Department of Health and Human Services (NIOSH) Anglov J, Holst E, Dyg S, Christensen J. (1993) Iron, manganese, copper, and titanium in welding fume dust on filters for internal and external quality assurance. Fresenius J Anal Chem; 345: Bjørgo R, Fjeldstad P, Nagy K, Thommassen Y. (1980) Parallelfiltre fra arbeidsatmosfæren i aluminiumsindustrien for interlaboratoriekontroll av analyse på PAH og flurid. Oslo: Institute of Occupational Health. Eller P, Feng A, Song R, Key-Schwartz R, Esche C, Groff J. (1999) Proficiency Analytical Testing (PAT) silica variability Am Ind Hyg Assoc J; 60: HSE. (1984) Method for the determination of hazardous substances (MDHS) 38, Quartz in respirable airborne dusts: laboratory method using infra-red spectroscopy (KBr disc technique). London: HSE Books. ISBN HSE. (1987) Method for the determination of hazardous substances (MDHS) 37, Quartz in respirable airborne dusts: laboratory method using infra-red spectroscopy (direct method). London: HSE Books. ISBN HSE. (1988) Method for the determination of hazardous substances (MDHS) 51/, Quartz in respirable airborne dusts: laboratory method using X-ray diffraction (direct method). London: HSE Books. ISBN HSE. (1997) Respirable crystalline silica. Guidance Note EH59, nd edn. London: HSE Books. ISBN HSE. (00) EH40, Occupational exposure limits. Guidance Note. London: HSE Books. ISBN IARC. (1997) Monographs on the evaluation of carcinogenic risks to humans and their supplements. Lyon: IARC. ISBN Vol. 68, pp ISO. (1994) Accuracy (trueness and precision) of measurement methods and results, ISO 575: Part. Geneva: ISO. Jeyaratnam M, Nagar N. (1993) Comparison of α-quartz standard Sikron F600 (HSE A9950) with the NIST respirable α- quartz standard SRM 1878 for bulk and on-filter analysis. Ann Occup Hyg; 37: NIOSH. (1994) Silica crystalline by IR, method 760. Manual of Analytical Methods. Issue. Cincinnati, OH: NIOSH. NIOSH. (1994) Silica in coal mine dust, method Manual of Analytical Methods. Issue. Cincinnati, OH: NIOSH. NIOSH. (1998) Silica, crystalline by XRD, method Manual of Analytical Methods. Issue 3. Cincinnati, OH: NIOSH. Shulman S, Groff J. Abell M, (199) Performance of laboratories measuring silica in the proficiency analytical testing program, Am Ind Hyg Assoc J; 53:

PILOT OF AN EXTERNAL QUALITY ASSURANCE PROGRAMME FOR CRYSTALLINE SILICA

PILOT OF AN EXTERNAL QUALITY ASSURANCE PROGRAMME FOR CRYSTALLINE SILICA PILOT OF AN EXTERNAL QUALITY ASSURANCE PROGRAMME FOR CRYSTALLINE SILICA MJ Shai, National Institute for Occupational Health ABSTRACT The mining industry has committed itself to reducing work-related silicosis.

More information

A Comparison of the Performance of Samplers for Respirable Dust in Workplaces and Laboratory Analysis for Respirable Quartz

A Comparison of the Performance of Samplers for Respirable Dust in Workplaces and Laboratory Analysis for Respirable Quartz Ann. Occup. Hyg., Vol. 57, No. 1, pp. 54 62, 2013 The Author 2012. Published by Oxford University Press on behalf of the British Occupational Hygiene Society doi:10.1093/annhyg/mes038 A Comparison of the

More information

MEASURING NANOPARTICLE EXPOSURE

MEASURING NANOPARTICLE EXPOSURE MEASURING NANOPARTICLE EXPOSURE APPLICATION NOTE NSAM-001 The Nanoparticle Surface Area Monitor reports the surface area of inhaled particles deposited in the lung Nanoparticle Exposure There is increasing

More information

Method Validation. Role of Validation. Two levels. Flow of method validation. Method selection

Method Validation. Role of Validation. Two levels. Flow of method validation. Method selection Role of Validation Method Validation An overview Confirms the fitness for purpose of a particular analytical method. ISO definition: Conformation by examination and provision of objective evidence that

More information

This document is a preview generated by EVS

This document is a preview generated by EVS INTERNATIONAL STANDARD ISO 14382 First edition 2012-12-01 Workplace atmospheres Determination of toluene diisocyanate vapours using 1-(2-pyridyl) piperazine-coated glass fibre filters and analysis by high

More information

Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000)

Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000) Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000) INTRODUCTION After a number of meetings and workshops, a group of representatives from 27 organizations

More information

How to interpret sample volume limitations stated in reference methodologies published for air sampling for Occupational Hygiene purposes

How to interpret sample volume limitations stated in reference methodologies published for air sampling for Occupational Hygiene purposes How to interpret sample volume limitations stated in reference methodologies published for air sampling for Occupational Hygiene purposes At a recent OH AIA Association meeting, I was asked to explain

More information

Respirable Crystalline Silica (RCS) Understanding the Regulations, Science & Results. Marc Hugus Manager, Business Development

Respirable Crystalline Silica (RCS) Understanding the Regulations, Science & Results. Marc Hugus Manager, Business Development Respirable Crystalline Silica (RCS) Understanding the Regulations, Science & Results Marc Hugus Manager, Business Development Marc Hugus Manager, Business Development Over 30 years consultative technical

More information

Scanning Electron Microscopy Scheme

Scanning Electron Microscopy Scheme Group Report Round 7A November 2017 Scanning Electron Microscopy Scheme BACKGROUND This report covers Round 7 of the SEMS asbestos fibre counting PT scheme. The scheme is operated by HSL, in collaboration

More information

A Field Evaluation of the BGI and Cowl Type Gravimetric Samplers

A Field Evaluation of the BGI and Cowl Type Gravimetric Samplers A Field Evaluation of the BGI and Cowl Type Gravimetric Samplers By B. K. Belle, J.J.L. Du Plessis and J.W. Oberholzer. Summary After initial studies by the CSIR indicating that a discrepancy exists in

More information

Zusammenfassung der Ergebnisse SEMS 7 Scanning Electron Microscopy Scheme - Round 7

Zusammenfassung der Ergebnisse SEMS 7 Scanning Electron Microscopy Scheme - Round 7 Zusammenfassung der Ergebnisse SEMS 7 Scanning Electron Microscopy Scheme - Round 7 Ergebnisse CRB Ergebnisse Ringversuch Sample 1-7SEM1 [ fibres mm -2 ] Amphibole Chrysotile Other inorg. fibres Total

More information

Exposure to Nanomaterials:

Exposure to Nanomaterials: Exposure to Nanomaterials: Challenges and Lessons form the Workplace Charles L. Geraci, Ph.D., CIH Centers for Disease Control and Prevention National Institute for Occupational Safety and Health Toward

More information

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3)

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3) Schedule Assignment Due Date Draft Section of Lab Report Monday 6pm (Jan 27) Quiz for Lab 2 Peer Review of Draft Complete Lab Report 1 Tuesday 9:30am Wednesday 6pm Friday 6pm Summary of Paper 2 Monday

More information

The AAFCO Proficiency Testing Program Statistics and Reporting

The AAFCO Proficiency Testing Program Statistics and Reporting The AAFCO Proficiency Testing Program Statistics and Reporting Program Chair: Dr. Victoria Siegel Statistics and Reports: Dr. Andrew Crawford Contents Program Model Data Prescreening Calculating Robust

More information

NANOTECHNOLOGY: AN INDUSTRIAL HYGIENE PERSPECTIVE. Loren Anderson, CIH, ROH

NANOTECHNOLOGY: AN INDUSTRIAL HYGIENE PERSPECTIVE. Loren Anderson, CIH, ROH NANOTECHNOLOGY: AN INDUSTRIAL HYGIENE PERSPECTIVE Loren Anderson, CIH, ROH Agenda Hazard Identification: (Recognition) Defining Nanotechnology Identification of new nanotechnology in process Communication

More information

Glossary of Common Laboratory Terms

Glossary of Common Laboratory Terms Accuracy A measure of how close a measured value is to the true value. Assessed by means of percent recovery of spikes and standards. Aerobic Atmospheric or dissolved oxygen is available. Aliquot A measured

More information

1 PRODUCT AND COMPANY IDENTIFICATION. Manufacturer. HAZARDS IDENTIFICATION Classification of the Substance or Mixture

1 PRODUCT AND COMPANY IDENTIFICATION. Manufacturer. HAZARDS IDENTIFICATION Classification of the Substance or Mixture Page 1 of 5 1 PRODUCT AND COMPANY IDENTIFICATION Manufacturer 750 LAKE COOK ROAD SUITE 440 BUFFALO GROVE, IL 60089 Emergency: Contact: Phone: Fax: Email: Web: Emergency Phone: (800) 424-9300 (ChemTrec

More information

Responsibilities: Effective Date: November Revision Date: February 8, VP, Facilities and Construction Management. Issuing Authority:

Responsibilities: Effective Date: November Revision Date: February 8, VP, Facilities and Construction Management. Issuing Authority: Title: Chemical Hygiene Written Program Effective Date: November 2005 Revision Date: February 8, 2017 Issuing Authority: Responsible Officer: VP, Facilities and Construction Management Director Environmental

More information

Part 14: Water quality Sampling. Guidance on quality assurance and quality control of environmental water sampling and handling

Part 14: Water quality Sampling. Guidance on quality assurance and quality control of environmental water sampling and handling Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 5667-14 Second edition 2014-12-15 Water quality Sampling Part 14: Guidance on quality assurance and quality control of environmental water sampling

More information

1 PRODUCT AND COMPANY IDENTIFICATION. Manufacturer. Description: Natural occuring clay mineral SDS. GHS Safety Data Sheet. PROFILE Products, LLC

1 PRODUCT AND COMPANY IDENTIFICATION. Manufacturer. Description: Natural occuring clay mineral SDS. GHS Safety Data Sheet. PROFILE Products, LLC Page 1 of 5 1 PRODUCT AND COMPANY IDENTIFICATION Manufacturer 750 LAKE COOK ROAD SUITE 440 BUFFALO GROVE, IL 60089 Emergency: Contact: Phone: Fax: Email: Web: Emergency Phone: (800) 424-9300 (ChemTrec

More information

MODULE 2B INDUSTRIAL HYGIENE LABORATORY ACCREDITATION PROGRAM (IHLAP) ADDITIONAL REQUIREMENTS

MODULE 2B INDUSTRIAL HYGIENE LABORATORY ACCREDITATION PROGRAM (IHLAP) ADDITIONAL REQUIREMENTS 2B.1 SCOPE MODULE 2B INDUSTRIAL HYGIENE LABORATORY ACCREDITATION PROGRAM (IHLAP) ADDITIONAL REQUIREMENTS The AIHA - Laboratory Accreditation Programs (AIHA-LAP), LLC s Industrial Hygiene Laboratory Accreditation

More information

Art, Science or Voodoo?

Art, Science or Voodoo? Industrial Hygiene Art, Science or Voodoo? Presented by: Ed Parsons Outreach Program Coordinator Gulf Coast Safety Institute Objectives At the completion of this session, the participants will: Know what

More information

Respirable Crystalline Silica Sampling and Laboratory Analysis

Respirable Crystalline Silica Sampling and Laboratory Analysis Respirable Crystalline Silica Sampling and Laboratory Analysis March 22, 2018 Keith P. Rickabaugh, CIH 11:00-11:15 Registration Meeting Agenda 11:15-12:00 RCS Sampling and Laboratory Analysis Keith Rickabaugh

More information

AIR. Ambient, Indoor, Workplace Air and Stack Emissions Proficiency Testing Scheme. Sample Preparation Instructions Round 1

AIR. Ambient, Indoor, Workplace Air and Stack Emissions Proficiency Testing Scheme. Sample Preparation Instructions Round 1 General Instructions Ambient, Indoor, Workplace Air and Sample Instructions Round 1 Sample Storage All samples should be stored in accordance with the instructions provided on the sample labels from the

More information

A Comparison of Fibre Counting Across Three European National Proficiency Testing Schemes

A Comparison of Fibre Counting Across Three European National Proficiency Testing Schemes Ann. occup. Hyg., Vol. 49, No. 4, pp. 309 324, 2005 Published by Oxford University Press (2005) doi:10.1093/annhyg/meh100 A Comparison of Fibre Counting Across Three European National Proficiency Testing

More information

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant SWGDRUG GLOSSARY These definitions were developed and adopted by the SWGDRUG Core Committee from a variety of sources including The United Nations Glossary of Terms for Quality Assurance and Good Laboratory

More information

Schedule for a proficiency test

Schedule for a proficiency test Schedule for a proficiency test Once a laboratory haegistered the year s programme the interaction between the laboratory and Eurofins Miljø A/S will be according to the schedule described below. 1 Reminder

More information

Considerations for an Automated SEM-EDX Routine for Characterizing Respirable Coal Mine Dust

Considerations for an Automated SEM-EDX Routine for Characterizing Respirable Coal Mine Dust 15 th North American Mine Ventilation Symposium, 2015 Sarver, E., Schafrik, S., Jong, E., Luxbacher, K. 2015, Virginia Tech Department of Mining and Minerals Engineering Considerations for an Automated

More information

Total Organic Carbon Analysis of Solid Samples for Environmental and Quality Control Applications

Total Organic Carbon Analysis of Solid Samples for Environmental and Quality Control Applications Application Note 35140710 Total Organic Carbon Analysis of Solid Samples for Environmental and Quality Control Applications Keywords 1030S Solids Module Aurora 1030 TOC Analyzer High Temperature Catalytic

More information

Analytische Qualitätssicherung Baden-Württemberg

Analytische Qualitätssicherung Baden-Württemberg Analytische Qualitätssicherung Baden-Württemberg Proficiency Test UKWIR SS 17 PAH in surface water with suspended solids Anthracene, Fluoranthene, Naphtalene, Benzo[a]pyrene, Benzo[b]fluoranthene, Benzo[k]fluoranthene,

More information

Practical Statistics for the Analytical Scientist Table of Contents

Practical Statistics for the Analytical Scientist Table of Contents Practical Statistics for the Analytical Scientist Table of Contents Chapter 1 Introduction - Choosing the Correct Statistics 1.1 Introduction 1.2 Choosing the Right Statistical Procedures 1.2.1 Planning

More information

ISO/TS TECHNICAL SPECIFICATION. Water quality Guidance on analytical quality control for chemical and physicochemical water analysis

ISO/TS TECHNICAL SPECIFICATION. Water quality Guidance on analytical quality control for chemical and physicochemical water analysis TECHNICAL SPECIFICATION ISO/TS 13530 First edition 2009-03-15 Water quality Guidance on analytical quality control for chemical and physicochemical water analysis Qualité de l'eau Lignes directrices pour

More information

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003 ALCOHOLS IV 1403 (1) HOCH 2 CH 2 OCH 3 MW: (1) 76.09 CAS: (1) 109-86-4 RTECS: (1) KL5775000 (2) HOCH 2 CH 2 OCH 2 CH 3 (2) 90.12 (2) 110-80-5 (2) KK8050000 (3) HOCH 2 CH 2 O(CH 2 ) 3 CH 3 (3) 118.17 (3)

More information

Analytische Qualitätssicherung Baden-Württemberg

Analytische Qualitätssicherung Baden-Württemberg Analytische Qualitätssicherung Baden-Württemberg Proficiency Test 2/17 TW S3 alkylphenoles in drinking water Nonylphenol, Octylphenol, Bisphenol-A provided by AQS Baden-Württemberg at Institute for Sanitary

More information

ALLOWAY METHOD OUTLINE

ALLOWAY METHOD OUTLINE ALLOWAY METHOD OUTLINE Standard Laboratory Method SM4500-Cl -G Parameter Residual Chlorine & Free Chlorine Method DPD Colorimetric Test Kit Date Issued Originator: Section Supervisor: QA Manager Date:

More information

[ Care and Use Manual ]

[ Care and Use Manual ] Oasis HLB DISKS I. Introduction Oasis HLB is a Hydrophilic-Lipophilic-Balanced, water-wettable, reversed-phase sorbent for all your SPE needs. It is made from a specific ratio of two monomers, the hydrophilic

More information

Exposure to nanomaterials in Germany

Exposure to nanomaterials in Germany Exposure to nanomaterials in Germany Results of the corporate survey of the Federal Institute for Occupational Health and Safety (BAuA) and the Association of the Chemical Industry (VCI) using questionnaires

More information

ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES

ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES ASEAN GUIDELINES FOR VALIDATION OF ANALYTICAL PROCEDURES Adopted from ICH Guidelines ICH Q2A: Validation of Analytical Methods: Definitions and Terminology, 27 October 1994. ICH Q2B: Validation of Analytical

More information

ICP-OES Application Note Number 35

ICP-OES Application Note Number 35 ICP-OES Application Note Number 35 Rapid measurement of major, minor and trace levels in soils using the Varian 730-ES Vincent Calderon Varian, Inc. Introduction As part of the global strategy for sustainable

More information

Investigation of Dust Levels in Different Areas of Underground Coal Mines

Investigation of Dust Levels in Different Areas of Underground Coal Mines International Journal of Occupational Safety and Ergonomics (JOSE) 2009, Vol. 15, No. 1, 125 10 NOTES Investigation of Dust Levels in Different Areas of Underground Coal Mines Mustafa Önder Seyhan Önder

More information

Results of Proficiency Test Rubber/Compounds May 2005

Results of Proficiency Test Rubber/Compounds May 2005 Results of Proficiency Test Rubber/Compounds May 25 Organised by: Spijkenisse, the Netherlands Authors: ing. R.J. Starink Correctors: R. Agten & dr. R.G. Visser Report: iis5p2 June 25 CONTENTS 1 INTRODUCTION...

More information

VAM Project Development and Harmonisation of Measurement Uncertainty Principles

VAM Project Development and Harmonisation of Measurement Uncertainty Principles VAM Project 3.2.1 Development and Harmonisation of Measurement Uncertainty Principles Part (d): Protocol for uncertainty evaluation from validation data V J Barwick and S L R Ellison January 2000 LGC/VAM/1998/088

More information

Nanoparticle Exposure in the Workplace

Nanoparticle Exposure in the Workplace Nanoparticle Exposure in the Workplace Michele. L. Ostraat, Ph.D. RTI International September 11, 2012 Practical Workplace Issues Related to Setting & Using Engineered Nanomaterial OELs Lack of instrumentation/characterization

More information

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT Table of Contents 1. SCOPE... 2 2. REFERENCES... 2 3. TERMS AND DEFINITIONS... 2 4. BACKGROUND... 4 5. ESTIMATION OF UNCERTAINTY OF MEASUREMENT POLICY...

More information

Analysis of domestic sludge using the Agilent 4200 MP-AES

Analysis of domestic sludge using the Agilent 4200 MP-AES Analysis of domestic sludge using the Agilent 4200 MP-AES Application note Environmental Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction Managing the treatment and disposal

More information

METHOTREXATE mg/m 3 (arbitrary). There is no OSHA PEL or AGGIH TLV for

METHOTREXATE mg/m 3 (arbitrary). There is no OSHA PEL or AGGIH TLV for METHOTREXATE Matrix: Air Control no.: T-PV2146-01-8804-CH Target Concentration: 0.04 mg/m 3 (arbitrary). There is no OSHA PEL or AGGIH TLV for methotrexate. Procedure: Samples are collected by drawing

More information

Standard Practice for Heat Aging of Plastics Without Load 1

Standard Practice for Heat Aging of Plastics Without Load 1 Designation: D 3045 92 (Reapproved 2003) Standard Practice for Heat Aging of Plastics Without Load 1 This standard is issued under the fixed designation D 3045; the number immediately following the designation

More information

Peter Kearns, PhD OECD, ENV/EHS BIPM, April 2015

Peter Kearns, PhD OECD, ENV/EHS BIPM, April 2015 EXPOSURE MEASUREMENT AND EXPOSURE MITIGATION: ACTIVITIES OF OECD S WORKING PARTY ON MANUFACTURED NANOMATERIALS Peter Kearns, PhD OECD, ENV/EHS BIPM, April 2015 Areas covered by the OECD Programme Testing

More information

Chapter 4: Verification of compendial methods

Chapter 4: Verification of compendial methods Chapter 4: Verification of compendial methods Introduction In order to ensure accurate and reliable test results, the quality control laboratory (QCL) needs to use analytical methods (and accompanying

More information

CERTIFIED REFERENCE MATERIAL RRM Au014

CERTIFIED REFERENCE MATERIAL RRM Au014 CERTIFIED REFERENCE MATERIAL RRM Au014 PRODUCT: GOLD ORE CERTIFICATE OF ANALYSIS Certificate Number: RRM CRM Au01 4 REV:000 Date: 2 2 February 2017 Analyte (Unit) Assigned value Table 1 Au014 Summary of

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

Analytical Performance & Method. Validation

Analytical Performance & Method. Validation Analytical Performance & Method Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451 Saudi Arabia Building:

More information

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

2.1. Accuracy, n- how close the indication of the thermometer is to the true value. AASHTO Accreditation Policy and Guidance on Thermometer Selection and Records Revised: January 25, 2018 Revision #: 0 1. Objective 1.1. The purpose of this document is to clearly define acceptance criteria

More information

Analytische Qualitätssicherung Baden-Württemberg

Analytische Qualitätssicherung Baden-Württemberg Analytische Qualitätssicherung Baden-Württemberg Proficiency Test 2/15 TW S5 Sulfonylurea herbicides amidosulfuron, metsulfuron-methyl, nicosulfuron, thifensulfuron-methyl, triasulfuron provided by AQS

More information

Occupational Health & Safety

Occupational Health & Safety Occupational Health & Safety Recognition of Hazard Process information MSDS evaluation (toxicity, VP, OEL etc) Manner of use (intensity and duration) Symptoms Evaluation of Hazard Control or Prevention

More information

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Method 1061 Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Company Method 1061 Revision 1. December 015 Organic Carbon in Finished

More information

Hard coal and coke Mechanical sampling. Part 7: Methods for determining the precision of sampling, sample preparation and testing

Hard coal and coke Mechanical sampling. Part 7: Methods for determining the precision of sampling, sample preparation and testing INTERNATIONAL STANDARD ISO 13909-7 Second edition 016-07-01 Hard coal and coke Mechanical sampling Part 7: Methods for determining the precision of sampling, sample preparation and testing Houille et coke

More information

NEW NIOSH SAMPLING AND ANALYTICAL METHODS FOR OCCUPATIONAL EXPOSURE ASSESSMENT

NEW NIOSH SAMPLING AND ANALYTICAL METHODS FOR OCCUPATIONAL EXPOSURE ASSESSMENT NEW NIOSH SAMPLING AND ANALYTICAL METHODS FOR OCCUPATIONAL EXPOSURE ASSESSMENT Kevin Ashley, Ph.D. US Department of Health and Human Services Centers for Disease Control and Prevention National Institute

More information

(manufactured) nano aerosols

(manufactured) nano aerosols Assessment of exposure to (manufactured) nano aerosols Derk Brouwer 2 Content Needs and relevance nano particles exposure measurements Measurement and sampling methods Observations in workplace practice

More information

International Journal of PharmTech Research CODEN (USA): IJPRIF, ISSN: , ISSN(Online): Vol.9, No.7, pp , 2016

International Journal of PharmTech Research CODEN (USA): IJPRIF, ISSN: , ISSN(Online): Vol.9, No.7, pp , 2016 International Journal of PharmTech Research CODEN (USA): IJPRIF, ISSN: 097-30, ISSN(Online): 255-9563 Vol.9, No.7, pp 399-06, 2016 Analytical Quality by Design Approach for Development of UV-Spectrophotometric

More information

Test Method Development and Validation as Pertaining to Microtrac Particle Size Measuring Instruments

Test Method Development and Validation as Pertaining to Microtrac Particle Size Measuring Instruments Test Method Development and Validation as Pertaining to Microtrac Particle Size Measuring Instruments Philip E. Plantz, PhD Application Note SL AN 17 Rev B Provided By: Microtrac, Inc. Particle Size Measuring

More information

A basic introduction to reference materials. POPs Strategy

A basic introduction to reference materials. POPs Strategy A basic introduction to reference materials POPs Strategy 2009-2010+ A tutorial 16 September 2009 Angelique Botha R&D metrologist Contents Why do we need reference materials? comparability of results metrological

More information

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY Table of Contents 1. SCOPE... 2 2. REFERENCES... 2 3. TERMS AND DEFINITIONS... 2 4. BACKGROUND... 4 5. EVALUATION OF MEASUREMENT UNCERTAINTY POLICY... 5

More information

Re-characterization of Some Factors Influencing Aerosol Sampling in the Workplace: Results from Field Studies

Re-characterization of Some Factors Influencing Aerosol Sampling in the Workplace: Results from Field Studies Fengxia HU, et al.: Comparison of Aerosol Sampling Influence Factors J Occup Health 2014; 56: 351 358 351 Journal of Occupational Health Re-characterization of Some Factors Influencing Aerosol Sampling

More information

Media blank result for a single-section sampler (e.g., membrane filter). Media blank result for back section of a sampler.

Media blank result for a single-section sampler (e.g., membrane filter). Media blank result for back section of a sampler. IV. GLOSSARY OF ABBREVIATIONS, DEFINITIONS, AND SYMBOLS AAS Acceptable range (biological) Accuracy ACGIH Ashing ASV B B b B f Bias Bioaerosol Biological monitoring Blank Atomic absorption spectrophotometry.

More information

Cu-Creatinine- Metol system

Cu-Creatinine- Metol system Quantification of Creatinine in Human Serum using Metol as a Chromogenic Probe Materials and methods 6.1. Reagents 6.1.1. N-methyl-p-aminophenol sulfate N-methyl-p-aminophenol sulfate also denoted as Metol

More information

Direct reading aerosol monitors

Direct reading aerosol monitors Health and Safety Executive Direct reading aerosol monitors 27 th June 2018 Andrew Thorpe Introduction Direct reading Aerosol Monitors (DRAMs) are often used as a complementary approach to conventional

More information

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method: ALLOWAY METHOD OUTLINE Standard Laboratory Method: SM Parameter: Methylene Blue Method: Colorimetric Reporting Level: Reference: 0.05 mg/l Standard Methods for the Examination Of Water and Wastewater;

More information

What is measurement uncertainty?

What is measurement uncertainty? What is measurement uncertainty? What is measurement uncertainty? Introduction Whenever a measurement is made, the result obtained is only an estimate of the true value of the property being measured.

More information

KISS Resources for NSW Syllabuses & Australian Curriculum.

KISS Resources for NSW Syllabuses & Australian Curriculum. Discusssion / Activity 1 Structure of the Earth Student Name... 1. Outline how we think the Sun & planets formed. The solar system formed from a cloud of gas & dust. Part of the cloud collapsed under gravity

More information

CHEK Proficiency study 642. Migration of formaldehyde from melamine kitchenware. Date 7 July 2016 Version 1 Status Final Report number CHEK

CHEK Proficiency study 642. Migration of formaldehyde from melamine kitchenware. Date 7 July 2016 Version 1 Status Final Report number CHEK CHEK Proficiency study 642 Migration of formaldehyde from melamine kitchenware Date 7 July 2016 Version 1 Status Final Report number CHEK-16-642 page 1 Colophon Number 642 Name Migration of formaldehyde

More information

Precision estimated by series of analysis ISO and Approach Duplicate Approach

Precision estimated by series of analysis ISO and Approach Duplicate Approach Agenda Item 9 JOINT FAO/WHO FOOD STANDARDS PROGRAMME CODEX COMMITTEE ON METHODS OF ANALYSIS SAMPLING Thirty-seventh th Session Budapest, Hungary, 6 February 016 (Comments prepared by the German Delegation)

More information

MODULE 4.3 Atmospheric analysis of particulates

MODULE 4.3 Atmospheric analysis of particulates MODULE 4.3 Atmospheric analysis of particulates Measurement And Characterisation Of The Particulate Content 1 Total particulate concentration 1 Composition of the particulate 1 Determination of particle

More information

A Standard Characterization Methodology for Respirable Coal Mine Dust Using SEM-EDX

A Standard Characterization Methodology for Respirable Coal Mine Dust Using SEM-EDX Resources 2015, 4, 939-957; doi:10.3390/resources4040939 Project Report OPEN ACCESS resources ISSN 2079-9276 www.mdpi.com/journal/resources A Standard Characterization Methodology for Respirable Coal Mine

More information

Analytische Qualitätssicherung Baden-Württemberg

Analytische Qualitätssicherung Baden-Württemberg Analytische Qualitätssicherung Baden-Württemberg Proficiency Test 7/16 - Ions in waste water - ammonium-nitrogen, nitrate-nitrogen, nitrite-nitrogen, total phosphorous, total cyanide, cyanide (week acid

More information

Application of mathematical, statistical, graphical or symbolic methods to maximize chemical information.

Application of mathematical, statistical, graphical or symbolic methods to maximize chemical information. Application of mathematical, statistical, graphical or symbolic methods to maximize chemical information. -However, this definition can be expanded to include: biology (biometrics), environmental science

More information

Proficiency Test 2/10 TW O5 Special organic analytes in drinking water

Proficiency Test 2/10 TW O5 Special organic analytes in drinking water Analytische Qualitätssicherung Baden-Württemberg Proficiency Test 2/ TW O5 Special organic analytes in drinking water Metabolites of pesticides and Glyphosate provided by AQS Baden-Württemberg at Institute

More information

Workshop on Understanding and Evaluating Radioanalytical Measurement Uncertainty November 2007

Workshop on Understanding and Evaluating Radioanalytical Measurement Uncertainty November 2007 1833-36 Workshop on Understanding and Evaluating Radioanalytical Measurement Uncertainty 5-16 November 2007 An outline of methods for the estimation of uncertainty that include the contribution from sampling

More information

Rock & Minerals Regents Review

Rock & Minerals Regents Review Name Rock & Minerals Regents Review Base your answers to questions 1 through 3 on the diagram below, which represents a part of the rock cycle. The igneous rock granite, and the characteristics of sedimentary

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 21501-2 First edition 2007-05-15 Determination of particle size distribution Single particle light interaction methods Part 2: Light scattering liquid-borne particle counter

More information

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition INTERNATIONAL STANDARD ISO 6974-1 Second edition 2012-05-15 Natural gas Determination of composition and associated uncertainty by gas chromatography Part 1: General guidelines and calculation of composition

More information

PUBLIC EMPLOYEE HAZARDOUS CHEMICAL PROTECTION AND RIGHT TO KNOW ACT O.C.G.A

PUBLIC EMPLOYEE HAZARDOUS CHEMICAL PROTECTION AND RIGHT TO KNOW ACT O.C.G.A PUBLIC EMPLOYEE HAZARDOUS CHEMICAL PROTECTION AND RIGHT TO KNOW ACT O.C.G.A. 45-22-2 Georgia s Right to Know Law Federal regulations require that all employees be trained on the Hazard Communications Standard

More information

Spectrophotometric Determination of Ferrocyanide in Effluents

Spectrophotometric Determination of Ferrocyanide in Effluents Spectrophotometric Determination of Ferrocyanide in Effluents ECN-0025-1 INTRODUCTION This method is used to determine the concentration of ferrocyanide ion in photoprocessing solution effluents. The ion

More information

Test Report in accordance CDPH-IAQ

Test Report in accordance CDPH-IAQ Hilti Entwicklungsgesellschaft GmbH Hiltistraße 6 86916 Kaufering Germany Test Report in accordance CDPH-IAQ Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing

More information

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water Hach Method 1041 Spectrophotometric Measurement of Free Chlorine (Cl ) in Finished Drinking Water Hach Company Method 1041 Revision 1. November 015 Spectrophotometric Measurement of Free Cl in Finished

More information

OF ANALYSIS FOR DETERMINATION OF PESTICIDES RESIDUES IN FOOD (CX/PR 15/47/10) European Union Competence European Union Vote

OF ANALYSIS FOR DETERMINATION OF PESTICIDES RESIDUES IN FOOD (CX/PR 15/47/10) European Union Competence European Union Vote 1 April 2015 European Union s CODEX COMMITTEE ON PESTICIDE RESIDUES 47 th Session Beijing, China, 13 18 April 2015 AGENDA ITEM 8 PROPOSED DRAFT GUIDELINES ON PERFORMANCE CRITERIA SPECIFIC FOR METHODS OF

More information

University of Tennessee Safety Procedure

University of Tennessee Safety Procedure University of Tennessee Safety Procedure Program Subject: Chemicals Requiring Review Prior to Use Procedure Review/Revised: 12/15/08 Affected Area/Department: Laboratories at the University Date Effective:

More information

Nanotechnology and Exposure: The Pathway to a Safe and Healthy Environment. Professor Candace Tsai Colorado State University

Nanotechnology and Exposure: The Pathway to a Safe and Healthy Environment. Professor Candace Tsai Colorado State University Nanotechnology and Exposure: The Pathway to a Safe and Healthy Environment Professor Candace Tsai Colorado State University Bogotá, 14-15 FEB 2017 Nanotechnology and Nanomaterials Engineered Nanoparticles-

More information

Measurement uncertainty and legal limits in analytical measurements

Measurement uncertainty and legal limits in analytical measurements UNCERTAINTY Measurement uncertainty and legal limits in analytical measurements MIRELLA BUZOIANU, Reference Materials Laboratory, National Institute of Metrology (INM), Bucharest, Romania 1 Introduction

More information

Laboratory ID. Laboratory Name. Address. Phone. Lab Manager/Supervisor. QA Officer. Analyst(s) Auditor. Date(s) of Audit

Laboratory ID. Laboratory Name. Address. Phone. Lab Manager/Supervisor. QA Officer. Analyst(s) Auditor. Date(s) of Audit Method(s): Laboratory ID Laboratory Name Address Phone Lab Manager/Supervisor QA Officer Analyst(s) Auditor Date(s) of Audit NEW JERSEY DEPARTMENT OF ENVIRONMENTAL PROTECTION OFFICE OF QUALITY ASSURANCE

More information

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

3.1.1 The method can detect, identify, and potentially measure the amount of (quantify) an analyte(s):

3.1.1 The method can detect, identify, and potentially measure the amount of (quantify) an analyte(s): SOP No: ICLN-Validation Levels Between Networks.003.01 Page 1 of 19 1. Purpose: This document establishes guidelines for assigning validation levels to methods based on national and international standards.

More information

Material Safety Data Sheet acc. to ISO/DIS 11014

Material Safety Data Sheet acc. to ISO/DIS 11014 Page 1/5 1 Identification of substance: Catalog number: 212185 Manufacturer/Supplier: BD Diagnostic Systems 7 Loveton Circle Sparks, MD 21152 Tel: (410) 771-0100 or (800) 638-8663 Information department:

More information

Validation and Standardization of (Bio)Analytical Methods

Validation and Standardization of (Bio)Analytical Methods Validation and Standardization of (Bio)Analytical Methods Prof.dr.eng. Gabriel-Lucian RADU 21 March, Bucharest Why method validation is important? The purpose of analytical measurement is to get consistent,

More information

Standardization needs and suggestions to EURAMET for consideration in their upcoming EMPIR calls

Standardization needs and suggestions to EURAMET for consideration in their upcoming EMPIR calls Version 3 October 2017 See : ftp://ftp.cencenelec.eu/en/researchinnovation/stair/stair-empir-needs/empir_responseform.docx RESPONSE FORM for 2018 Standardization needs and suggestions to EURAMET for consideration

More information

CHEK Proficiency study 644. Solvents in adhesive. Date 15 October 2016 Version 1 Status Final Report number CHEK

CHEK Proficiency study 644. Solvents in adhesive. Date 15 October 2016 Version 1 Status Final Report number CHEK CHEK Proficiency study 644 Solvents in adhesive Date 15 October 2016 Version 1 Status Final Report number CHEK-16-644 page 0 Colophon Number 644 Name Solvents in adhesive Contact Author Netherlands Food

More information

Measurement and control of grinding circuit performance by real-time particle size measurement on individual cyclones

Measurement and control of grinding circuit performance by real-time particle size measurement on individual cyclones Measurement and control of grinding circuit performance by real-time particle size measurement on individual cyclones C. V. O Keefe CiDRA Minerals Process Proper control of grinding circuits is essential

More information

Results of the inter-laboratory comparison exercise for TC and EC measurements (ref.: OCEC ) SUMMARY 2 1 ORGANIZATION 4

Results of the inter-laboratory comparison exercise for TC and EC measurements (ref.: OCEC ) SUMMARY 2 1 ORGANIZATION 4 Results of the inter-laboratory comparison exercise for TC and EC measurements (ref.: OCEC-2017-2) SUMMARY 2 1 ORGANIZATION 4 1.1 Samples, sub-samples and sub-sample homogeneity 4 1.2 Participants 5 1.3

More information

Introduction to IH Analytical Chemistry

Introduction to IH Analytical Chemistry Introduction to IH Analytical Chemistry Sampling and Analysis Manning, Eide, Van Etten (Apr 2013) Foreword Thank you for purchasing the Introduction to IH Analytical Chemistry self-study workbook. Course

More information

Nanotechnology & Respiratory Protection Issues

Nanotechnology & Respiratory Protection Issues Background One of the emerging topics in industrial hygiene today is nanotechnology and the nanoparticles (i.e. less than 100 nanometers in diameter) it may produce - one million nanometres make up one

More information