DR/4000 PROCEDURE LEAD. 2. The display will show: 3. Fill a 250-mL graduated cylinder to the 250-mL mark with sample.

Size: px
Start display at page:

Download "DR/4000 PROCEDURE LEAD. 2. The display will show: 3. Fill a 250-mL graduated cylinder to the 250-mL mark with sample."

Transcription

1 Method 8033 DR/4000 PROCEDURE Dithizone Method* (0 to 300 µg/l) Scope and Application: For water and wastewater; USEPA accepted for reporting for wastewater analysis (digestion is required).** See Section 1 for digestion procedure. The estimated detection limit for program number 2200 is 3 µg/l Pb. * Adapted from Snyder, L. J., Analytical Chemistry, (1947) ** Procedure is equivalent to Standard Method 3500-Pb D for wastewater analysis. HACH PROGRAM HACH PROGRAM: 2200 Lead, Dithizone ENTER 1. Press the soft key under HACH PROGRAM. Select the stored program number for lead (Pb) by pressing 2200 with the numeric keys. Press: ENTER Note: If sample cannot be analyzed immediately, see Sample Collection, Storage and Preservationfollowing these steps. Adjust the ph of preserved samples before analysis. Note: The Flow Cell and Sipper Modules cannot be used with this procedure. 2. The display will show: HACH PROGRAM: 2200 Lead, Dithizone The wavelength (λ), 515 nm, is automatically selected. Note: For best results, determine a reagent blank for each new lot of reagent as follows. Prepare a reagent blank by repeating steps 3 through 14, using deionized water as the sample. Zero the instrument on chloroform by pressing the soft key under ZERO. Insert the reagent blank and the blank value will be displayed. Correct for the reagent blank by pressing the soft keys under OPTIONS, (MORE), and then BLANK:OFF. Enter the reagent blank value and press ENTER. Repeat for each new lot of reagent. 3. Fill a 250-mL graduated cylinder to the 250-mL mark with sample. Note: Clean all glassware with a 1:1 Nitric Acid Solution. Rinse with deionized water. Note: Cloudy and turbid samples may require filtering before running test. Results should be reported as soluble lead. Use a glass membrane filter to avoid loss of lead by adsorption on filter paper. Note: For proof of accuracy, use a 200 µg/l lead standard solution (preparation given in the Accuracy Check section) in place of the sample. 4. Transfer the sample into 500-mL separatory funnel. Lead_PP_Other_DTZ_Eng_4000.fm Page 1 of 8

2 , continued 5. Add the contents of one Buffer Powder Pillow, citrate type for heavy metals. Stopper the funnel. Shake to dissolve. Note: Spilled reagent will affect test accuracy and is hazardous to skin and other materials. 6. Add 50 ml of chloroform to a 50-mL graduated mixing cylinder. Add the contents of one DithiVer Metals Reagent Powder Pillow. Stopper. Invert repeatedly to mix. Pour 30 ml of this dithizone solution into a second graduated cylinder. Note: Use adequate ventilation (i.e. a fume hood). The DithiVer Powder will not all dissolve in the chloroform. For further notes see Dithiver Solution Preparation, Storage and Reagent Blank. 7. Add the 30 ml of the dithizone solution to the separatory funnel. Stopper. Invert. Open stopcock to vent. Close the stopcock and add 5 ml of 5.0 N Sodium Hydroxide Standard Solution. Stopper. Invert. Open stopcock to vent. Close the stopcock and shake the funnel once or twice and vent again. Note: Add a few drops of 5.25 N Sulfuric Acid Standard Solution if the solution turns orange on shaking. The blue-green color will reappear. To avoid higher blanks, repeat procedure on new sample and use less sodium hydroxide in this step. 8. Continue adding 5.0 N Sodium Hydroxide Standard Solution dropwise and shaking the funnel after every few drops until the color of the solution being shaken changes from blue-green to orange. Then add 5 more drops of 5.0 N Sodium Hydroxide Standard Solution. Note: A pink color in the bottom (chloroform) layer at this point does not necessarily indicate lead is present. Only after shaking with potassium cyanide in next step will a pink color in the chloroform layer confirm the presence of lead. Note: Large amounts of zinc cause the color transition at the end point to be indistinct. Note: For more accurate results, adjust the sample to ph using a ph meter, omitting the five additional drops of Sodium Hydroxide Standard Solution. Page 2 of 8 Lead_PP_Other_DTZ_Eng_4000.fm

3 , continued 9. Add 2 heaping 1.0-g scoops of potassium cyanide to the funnel. Stopper. Shake vigorously until the potassium cyanide is all dissolved (about 15 seconds). Note: Wait one minute for the layers to separate. The bottom (chloroform) layer will be pink if lead is present. Warning: Cyanide is a deadly poison. Use a fume hood. Maintain cyanide solutions at ph 11 or greater to prevent formation of cyanide gas. 10. Insert a cotton plug the size of a pea into the delivery tube of the funnel. Slowly drain the bottom (chloroform) layer into a dry 25-mL sample cell. Stopper the sample cell. Note: The lead-dithizone complex is stable for at least 30 minutes if the sample cell is kept tightly capped and out of direct sunlight. 11. Fill another sample cell with chloroform (the blank). Stopper. 12. Place the blank into the cell holder. Close the light shield. ZERO 13. Press the soft key under ZERO. The display will show: 0.00 µg/l Pb Note: If you are using a reagent blank correction, the display will show the correction. Note: For alternate concentration units, press the soft key under OPTIONS. Then press the soft key under UNITS to scroll through the available options. Press ENTER to return to the read screen. 14. Place the prepared sample into the cell holder. Close the light shield. The results in µg/l lead (or chosen units) will be displayed. Note: See Dithiver Solution Preparation, Storage and Reagent Blank for information on preparing a reagent blank. Note: See Pollution Prevention and Waste Management for proper disposal of chloroform solutions. Lead_PP_Other_DTZ_Eng_4000.fm Page 3 of 8

4 , continued Interference The following do not interfere: Aluminum Calcium Magnesium Antimony Chromium Manganese Arsenic Cobalt Nickel Cadmium Iron Zinc Interfering Substance Bismuth Copper Mercury Silver Tin Highly buffered samples or extreme sample ph Interference Levels and Treatments May exceed the buffering capacity of the reagents and require sample pretreatment; see Section ph Interference. Eliminate interference from the metals listed above by the following treatment, beginning after procedure Step 6. a. Measure about 5 ml of the prepared dithizone solution into the separatory funnel. Stopper the funnel, invert and open the stopcock to vent. Close the stopcock and shake the solution vigorously for 15 seconds. Allow the funnel to stand undisturbed until the layers separate (about 30 seconds). A yellow, red, or bronze color in the bottom (chloroform) layer confirms the presence of interfering metals. Draw off and collect the bottom (chloroform) layer for proper disposal. b. Repeat extraction with fresh 5-mL portions of prepared dithizone solution (collecting the bottom layer each time in appropriate waste collection vessel) until the bottom layer shows a pure dark green color for three successive extracts. Extractions can be repeated a number of times without appreciably affecting the amount of lead in the sample. c. Extract the solution with several 2 or 3 ml portions of pure chloroform to remove any remaining dithizone, again collecting the bottom layer each time for proper disposal. d. Continue the procedure, substituting 28.5 ml of prepared dithizone solution for the 30 ml in Step 7. Dithiver Solution Preparation, Storage and Reagent Blank Store DithiVer Powder Pillows away from light and heat. A convenient way to prepare this solution is to add the contents of 10 DithiVer Metals Reagent Powder Pillows to a 500-mL bottle of chloroform and invert several times until well mixed (carrier powder may not dissolve). Store dithizone solution in an amber glass bottle. This solution is stable for 24 hours. A reagent blank using deionized water should be carried out through the entire method to obtain the most accurate results. The amount of reagent blank determined on each lot of DithiVer Metals Reagent Powder Pillow can be subtracted automatically from the sample reading by pressing the soft keys under Page 4 of 8 Lead_PP_Other_DTZ_Eng_4000.fm

5 , continued OPTIONS, (MORE) and then BLANK: (OFF). Enter the blank value and press ENTER to return to the read screen. Sample Collection, Storage and Preservation Collect samples in acid-cleaned glass or plastic containers. Adjust the ph to 2 or less with nitric acid (about 2 ml per liter). Preserved samples can be stored up to six months at room temperature. Before analysis, adjust the ph to with 5.0 N Sodium Hydroxide. Correct the test result for volume additions; see Section Correcting for Volume Additions. Accuracy Check Standard Additions Method a. Leave the unspiked sample in the sample cell compartment. Verify that the units displayed are in µg/l. Select standard additions mode by pressing the soft keys under OPTIONS, (MORE) and then STD ADD. b. Press ENTER to accept the default sample volume (ml), 250. c. Press ENTER to accept the default standard concentration (µg/l), 50,000 (equivalent to 50-mg/L). d. Press the soft key under ENTRY DONE. e. Snap the neck off a Lead Voluette Ampule Standard, 50-mg/L Pb. f. Use the TenSette Pipet (do not use a glass pipet) to add 0.1 ml, 0.2 ml and 0.3 ml of standard, respectively to three 250-mL samples and mix each thoroughly. g. Analyze each standard addition sample as described above. Accept the standard additions reading by pressing the soft key under READ each time. Each addition should reflect approximately 100% recovery. h. After completing the sequence, the display will show the extrapolated concentration value and the best-fit line through the standard additions data points, accounting for matrix interferences. i. See Section Standard Additions for more information. Standard Solution Method Prepare a 10-mg/L lead standard solution by pipetting ml of Lead Standard Solution, 100-mg/L, into a 100-mL volumetric flask. Add 0.2 ml of concentrated nitric acid using a TenSette Pipet to prevent the adsorption of lead onto the container walls. Dilute to the mark with deionized water and mix thoroughly. To make a 200-µg/L standard, pipet 5.00 ml of the 10.0-mg/L standard into 245 ml of deionized water in the 500-mL separatory funnel in Step 4 of the Dithizone procedure. Prepare these solutions daily. Perform the lead procedure as described above. To adjust the calibration curve using the reading obtained with the 200-µg/L standard solution, press the soft keys under OPTIONS, MORE then STD: OFF. Press ENTER to accept the displayed concentration, the value of which depends on the selected units. If an alternate concentration is used, enter the actual concentration and press enter to return to the read screen. See Section Adjusting the Standard Curve for more information. Lead_PP_Other_DTZ_Eng_4000.fm Page 5 of 8

6 , continued Method Performance Precision Standard: 200 µg/lpb Program For more information on determining precision data and method detection limits, refer to Section 1.5. Estimated Detection Limit For more information on derivation and use of Hach s estimated detection limit, see Section To determine a method detection limit (MDL) as defined by the 40 CFR part 136, appendix B, see Section Sensitivity Program Number: % Confidence Limits µg/l Pb Program EDL µg/l Pb Portion of Curve Abs Concentration Abs µg/l 150 µg/l µg/l 270 µg/l µg/l See Section Sensitivity Explained for more information. Calibration Standard Preparation To perform a lead calibration using the Dithizone method, prepare a 10,000-µg/L lead stock solution by pipetting ml of a 100-mg/L Lead Standard Solution (Cat. No ) into a 100-mL volumetric flask using Class A glassware. Add 0.2 ml of concentrated nitric acid using a TenSette pipet to prevent the adsorption of lead onto the container walls. Dilute to the mark with deionized water and mix thoroughly. Prepare calibration standards containing 40, 80, 120, 160, 200, 240, 280, and 300 µg/l lead as follows: a. Prepare the standards in 500-mL separatory funnels to minimize lead loss in intermediate glassware due to adsorption. b. Into eight different 500-mL separatory funnels, pipet 1, 2, 3, 4, 5, 6, 7, and 8 ml of the 10,000-µg/L lead stock solution using Class A glassware. Add enough deionized water to bring the final volume to exactly 250 ml. For example, use 3 ml of the 10,000-µg/L lead stock solution and 247 ml of deionized water to prepare the 120-µg/L standard. c. Using the Dithizone method and the calibration procedure described in the User-Entered Programs section of the DR/4000 Spectrophotometer Instrument Manual, generate a calibration curve from the standards prepared above. Page 6 of 8 Lead_PP_Other_DTZ_Eng_4000.fm

7 , continued Summary of Method The DithiVer Metals Reagent is a stable powder form of dithizone. Lead ions in basic solution react with dithizone to form a pink to red lead-dithizonate complex, which is extracted with chloroform. Safety Good safety habits and laboratory techniques should be used throughout the procedure. Consult the Material Safety Data Sheet for information specific to the reagents used. For additional information, refer to Section 1. Pollution Prevention and Waste Management Both chloroform (D022) and cyanide (D003) solutions are regulated as hazardous wastes by the Federal RCRA. Do not pour these solutions down the drain. Chloroform solutions and the cotton plug used in the delivery tube of the separatory funnel should be collected for disposal with laboratory solvent waste. Collect the cyanide solution as a reactive waste. Be sure that cyanide solutions are stored in a caustic solution with a ph >11 to prevent potential release of hydrogen cyanide gas. See Section 3 for more information on proper disposal of these materials. REQUIRED REAGENTS AND STANDARDS Cat. No. Lead Reagent Set (100 Tests) Includes: (1) , (2) , (1) , (2) , (1) , (2) , (1) Quantity Required Description per test Unit Cat. No. Buffer Powder Pillows, Citrate... 1 pillow /pkg Chloroform ml...4 L DithiVer Metals Reagent Powder Pillows... 1 pillow /pkg Potassium Cyanide, ACS... 2 g g Sodium Hydroxide Solution, 5.0 N... 5 ml ml Sodium Hydroxide Standard Solution, 5.0 N...varies ml DB REQUIRED EQUIPMENT AND SUPPLIES QuantityRequired Description per test Unit Cat. No. Clippers, for opening powder pillows each Cotton balls, absorbent /pkg Cylinder, graduated mixing, 50-mL each Cylinder, graduated, 5-mL each Cylinder, graduated, 50-mL each Cylinder, graduated, 250-mL each DR/ Inch Cell Adapter each Funnel, separatory, 500-mL each ph Meter, sension 1, portable each Ring, support, 4-inch each Sample Cells, matched pair, 1-inch, glass, with stoppers pair Spoon, measuring, 1.0-g each Support, Ring Stand, 5 x 8 in. base each Lead_PP_Other_DTZ_Eng_4000.fm Page 7 of 8

8 , continued OPTIONAL REAGENTS AND STANDARDS Description Unit Cat. No. Chloroform, ACS ml Lead Standard Solution, 100 mg/l Pb ml Lead Standard Solution, 10-mL Voluette Ampules, 50-mg/L Pb... 16/pkg Nitric Acid Solution, 1: ml Nitric Acid, ACS ml Sodium Hydroxide Standard Solution, 5.0 N ml SCDB Sodium Hydroxide Standard Solution, 5.0 N ml MDB Sulfuric Acid, 5.25 N ml MDB Water, deionized... 4 liters OPTIONAL EQUIPMENT AND SUPPLIES Ampule Breaker Kit... each DR/4000 Carousel Module Kit... each Filter Discs, glass membrane, 47-mm /pkg Filter Holder, glass, 47-mm... each Flask, Erlenmeyer, 500-mL... each Flask, filtering, 500-mL... each Flask, volumetric, Class A, 100-mL... each ph Paper, ph 1.0 to rolls/pkg Pipet Filler, safety bulb... each Pipet, TenSette, 0.1 to 1.0 ml... each Pipet Tips, for TenSette Pipet... 50/pkg Pipet, serological, 2-mL... each Pipet, volumetric, Class A, 1.00-mL... each Pipet, volumetric, Class A, 2.00-mL... each Pipet, volumetric, Class A, 3.00-mL... each Pipet, volumetric, Class A, 4.00-mL... each Pipet, volumetric, Class A, 5.00-mL... each Pipet, volumetric, Class A, 6.00-mL... each Pipet, volumetric, Class A, 7.00-mL... each Pipet, volumetric, Class A, 8.00-mL... each Pipet, volumetric, Class A, mL... each FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web techhelp@hach.com HACH COMPANY WORLD HEADQUARTERS Telephone: (970) FAX: (970) Hach Company, All rights reserved. Printed in the U.S.A. 6/03 11ed Page 8 of 8 Lead_PP_Other_DTZ_Eng_4000.fm

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ ) DR/4000 PROCEDURE Method 8023 1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to 0.700 mg/l Cr 6+ ) Scope and Application: For water and wastewater;usepa accepted for reporting for

More information

DR/4000 PROCEDURE. IRON, Total

DR/4000 PROCEDURE. IRON, Total Method 8365 Powder Pillows DR/4000 PROCEDURE FerroMo Method* (0 to 1.800 mg/l) Scope and Application: For cooling water containing molybdate-based treatment; digestion is required for determining total

More information

DR/4000 PROCEDURE NITRATE. 2. The display will show:

DR/4000 PROCEDURE NITRATE. 2. The display will show: Method 8192 Powder Pillows DR/4000 PROCEDURE Cadmium Reduction Method LR (0 to 0.50 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit for program number

More information

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

DR/4000 PROCEDURE NITRATE. Using Powder Pillows DR/4000 PROCEDURE Method 8171 Powder Pillows or AccuVac Ampuls Cadmium Reduction Method MR (0 to 5.0 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit

More information

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ).

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ). Method 8194 DR/4000 PROCEDURE Diaminobenzidine Method* (0 to 1.000 mg/l) Scope and Application: For water and wastewater; distillation is required for determining total selenium. See the Distillation procedure

More information

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water. Method 8337 Powder Pillows DR/4000 TT PROCEDURE QUATERNARY AMMONIUM COMPOUNDS Direct Binary Complex Method (0 to 5.00 mg/l as CTAB) Scope and Application: For cooling tower and pool/spa water. The estimated

More information

DR/4000 PROCEDURE. HARDNESS, Total

DR/4000 PROCEDURE. HARDNESS, Total Method 8374 DR/4000 PROCEDURE Calcium and Magnesium; Chlorophosphonazo Colorimetric Method ULR (0 to 1,000 µg/l Ca & Mg as CaCO 3 ) Scope and Application: For ultra pure water. The estimated detection

More information

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows Silver DOC316.53.01134 Colorimetric Method Method 8120 0.02 to 0.70 mg/l Ag Powder Pillows Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows

More information

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Nitrate, MR DOC316.53.01069 Cadmium Reduction Method Method 8171 0.1 to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Scope and application: For water, wastewater

More information

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows Nitrate DOC316.53.01067 Cadmium Reduction Method Method 8192 0.01 to 0.50 mg/l NO 3 N (LR) Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows Potassium DOC316.53.01127 Tetraphenylborate Method Method 8049 0.1 to 7.0 mg/l K Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific information

More information

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right. Aluminum DOC316.53.01003 Eriochrome Cyanine R Method 1 Method 8326 0.006 to 0.250 mg/l Al 3+ Powder Pillows Scope and application: For water. 1 Adapted from Standard Methods for the Examination of Water

More information

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation Phosphorus, Total DOC316.53.01121 USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method 8190 0.06 to 3.50 mg/l PO 3 4 (0.02 to 1.10 mg/l P) Test N Tube Vials Scope and application: For water,

More information

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters) Aluminum DOC316.53.01002 Aluminon Method 1 Method 8012 0.008 to 0.800 mg/l Al 3+ (spectrophotometers) Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters) Scope and application: For water and wastewater.

More information

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters) Chromium, Total DOC316.53.01034 Alkaline Hypobromite Oxidation Method 1 Method 8024 0.01 to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters) Scope and application: For water and wastewater.

More information

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater. Nitrogen, Ammonia DOC316.53.01078 USEPA 1 Nessler Method 2 Method 8038 0.02 to 2.50 mg/l NH 3 N Reagent Solution Scope and application: For water, wastewater and seawater. Distillation is required for

More information

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3 , MR, 8171 DOC316.53.01069 Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3 N) Powder Pillows or AccuVac Ampuls Scope and Application: For water, wastewater and seawater Test preparation

More information

Mercaptoacetic Acid Method*

Mercaptoacetic Acid Method* Method 8036 MOLYBDENUM, MOLYBDATE, High Range (0 to 40.0 mg/l) Mercaptoacetic Acid Method* Using Powder Pillows For water and wastewater 1. Enter the stored program number for high range molybdenumpowder

More information

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls) Method 8171 NITRATE, Mid Range (0 to 5.0 mg/l NO 3- -N) For water, wastewater and seawater* Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls) Using Powder Pillows 1. Enter the stored program

More information

Method to 0.50 mg/l NH 3 N Powder Pillows

Method to 0.50 mg/l NH 3 N Powder Pillows , 8155 Salicylate Method 1 Scope and Application: For water, wastewater and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) DOC316.53.01077 Method 8155 0.01 to 0.50 mg/l NH 3 N Powder Pillows

More information

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right. Barium DOC316.53.01315 Turbidimetric Method 1 Method 10251 2 to 100, 20 to 1000, 200 to 10,000 mg/l Ba (spectrophotometers) Powder Pillows 2 to 80, 20 to 800, 200 to 8000 mg/l Ba (colorimeters) Scope and

More information

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials Nitrate, HR DOC316.53.01068 Chromotropic Acid Method Method 10020 0.2 to 30.0 mg/l NO 3 N (HR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific information

More information

Method to 0.50 mg/l NH 3 N Powder Pillows

Method to 0.50 mg/l NH 3 N Powder Pillows , 8155 Salicylate Method 1 Scope and Application: For water, wastewater and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) DOC316.53.01077 Method 8155 0.01 to 0.50 mg/l NH 3 N Powder Pillows

More information

DR/4000 PROCEDURE CYANIDE. 1. Press the soft key under HACH PROGRAM. Select the stored program for cyanide by pressing 1750 with the numeric keys.

DR/4000 PROCEDURE CYANIDE. 1. Press the soft key under HACH PROGRAM. Select the stored program for cyanide by pressing 1750 with the numeric keys. DR/4000 PROCEDURE Method 8027 Pyridine-Pyrazalone Method* Powder Pillows (0 to 0.240 mg/l CN ) Scope and Application: For water, wastewater and seawater. The estimated detection limit for program number

More information

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows , 8326 Eriochrome Cyanine R Method 1 Scope and Application: For water 1 Adapted from Standard Methods for the Examination of Water and Wastewater. DOC316.53.01003 Method 8326 (0.006 to 0.250 mg/l Al 3+

More information

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials Nitrogen, Total DOC316.53.01086 Persulfate Digestion Method Method 10071 0.5 to 25.0 mg/l N (LR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific

More information

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials Nitrogen, Total DOC316.53.01085 Persulfate Digestion Method Method 10072 2 to 150 mg/l N (HR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific information

More information

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution Chloride DOC316.53.01017 Mercuric Thiocyanate Method Method 8113 0.1 to 25.0 mg/l Cl Reagent Solution Scope and application: For water and wastewater. Test preparation Instrument-specific information Table

More information

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows FILL LINE Method 8155 Salicylate Method 1 Powder Pillows Scope and Application: For water, wastewater, and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) (0.01 to 0.50 mg/l NH 3 N) Test Preparation

More information

CHLORINE, TOTAL (0 to 4.00 mg/l)

CHLORINE, TOTAL (0 to 4.00 mg/l) CHLORINE, TOTAL (0 to 4.00 mg/l) DOC316.53.01261 For water, wastewater, and seawater Method 10250 DPD Method Powder Pillows USEPA accepted for reporting water and wastewater analyses * Note: This product

More information

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

Scope and application: For wastewater, seawater, drinking water, surface water and process water. Cyanide DOC316.53.01495 USEPA 1 Pyridine Barbituric Acid Method 2 Method 10265 0.01 to 0.60 mg/l CN TNTplus 862 Scope and application: For wastewater, seawater, drinking water, surface water and process

More information

Nitrogen, Total Inorganic

Nitrogen, Total Inorganic Nitrogen, Total Inorganic DOC316.53.01090 Titanium Trichloride Reduction Method Method 10021 0.2 to 25.0 mg/l N Test N Tube Vials Scope and application: For water, wastewater and seawater. Test preparation

More information

Chloramine (Mono) and Nitrogen, Free Ammonia

Chloramine (Mono) and Nitrogen, Free Ammonia Chloramine (Mono) and Nitrogen, Free Ammonia DOC316.53.01016 Indophenol Method 1 Method 10200 0.04 to 4.50 mg/l Cl 2 0.01 to 0.50 mg/l NH 3 N Powder Pillows Scope and application: For the determination

More information

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856 Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows

More information

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

Cadion Method Method to 0.30 mg/l Cd TNTplus 852 Cadmium DOC316.53.01013 Cadion Method Method 10217 0.02 to 0.30 mg/l Cd TNTplus 852 Scope and application: For wastewater and process control. Test preparation Instrument-specific information Table 1 shows

More information

Chlorine, Free and Total, High Range

Chlorine, Free and Total, High Range Chlorine, Free and Total, High Range DOC316.53.01490 USEPA DPD Method 1 Method 10069 (free) 10070 (total) 0.1 to 10.0 mg/l Cl 2 (HR) Powder Pillows Scope and application: For testing higher levels of free

More information

Sulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure

Sulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure Methylene Blue Method 1 (0 0.70, 0 7.00 and 0 70.00 mg/l) DOC316.53.01320 Method 10254 Scope and Application: For oil and gas field waters. 1 Adapted from Standard Methods for the Examination of Water

More information

Chromazurol S Method Method to 0.50 mg/l Al TNTplus 848

Chromazurol S Method Method to 0.50 mg/l Al TNTplus 848 Aluminum DOC316.53.01004 Chromazurol S Method Method 10215 0.02 to 0.50 mg/l Al TNTplus 848 Scope and application: For drinking water, surface water, swimming pool water, wastewater and process analysis.

More information

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV Iron, Total DOC316.53.01338 FerroZine Method 1 Method 10264 1 to 100 µg/l Fe (10-cm cell) Reagent Solution Scope and application: For ultrapure water. 1 Adapted from Stookey, L.L., Anal. Chem., 42(7),

More information

NITROGEN, TOTAL, HR, Test N Tube

NITROGEN, TOTAL, HR, Test N Tube Method 10072 NITROGEN, TOTAL, HR, Test N Tube 10 to 150 mg/l N TNT Persulfate Digestion Method For water and wastewater Digestion is required for determining total nitrogen. The digestion procedure is

More information

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell Hardness DOC316.53.01044 Chlorphosphonazo Colorimetric Method Method 8374 8 to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow Scope and application: For boiler and ultrapure water. Test preparation

More information

Quaternary Ammonium Compounds

Quaternary Ammonium Compounds , 8337 Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB Powder Pillows Scope and Application: For cooling tower water and pool/spa water Test

More information

NITROGEN, AMMONIA, High Range, Test N Tube

NITROGEN, AMMONIA, High Range, Test N Tube NITROGEN, AMMONIA, High Range, Test N Tube Method 10031 (0 to 50 mg/l NH 3 -N) For water, wastewater, and seawater Salicylate Method * 1. Enter the stored program number for nitrogen, ammonia, high range

More information

PAR Method Method to 2.0 mg/l Pb TNTplus 850

PAR Method Method to 2.0 mg/l Pb TNTplus 850 Lead DOC316.53.01056 PAR Method Method 10216 0.1 to 2.0 mg/l Pb TNTplus 850 Scope and application: For wastewater and process control. Test preparation Instrument-specific information Table 1 shows all

More information

Chlorine, Free and Total, High Range

Chlorine, Free and Total, High Range Chlorine, Free and Total, High Range DOC316.53.01449 USEPA DPD Method 1 Method DPD 0.1 to 8.0 mg/l Cl 2 Powder Pillows Scope and application: For testing residual chlorine and chloramines in water, wastewater,

More information

PAR Method Method to 2.0 mg/l Pb TNTplus 850

PAR Method Method to 2.0 mg/l Pb TNTplus 850 , TNTplus 850, 10216 DOC316.53.01056 PAR Method Method 10216 0.1 to 2.0 mg/l Pb TNTplus 850 Scope and Application: For wastewater and process control Test preparation Before starting the test: FOR DR 2800:

More information

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters) Silica DOC316.53.01132 Heteropoly Blue Method 1 Method 8186 0.010 to 1.600 mg/l SiO 2 (LR, spectrophotometers) Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters) Scope and application: For boiler

More information

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator Alkalinity DOC316.53.01166 Phenolphthalein and Total Alkalinity Method 8203 10 to 4000 mg/l as CaCO 3 Digital Titrator Scope and application: For water, wastewater and seawater. Test preparation Before

More information

Chlorine, Free and Total, Low Range

Chlorine, Free and Total, Low Range Chlorine, Free and Total, Low Range DOC316.53.01450 USEPA DPD Method 1 Method 8021 (free) 8167 (total) 0.02 to 2.00 mg/l Cl 2 (LR) Powder Pillows or AccuVac Ampuls Scope and application: For testing residual

More information

Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 )

Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 ) Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 ) DOC316.53.01523 DPD Method Method 10290 0.10 to 10.00 mg/l PAA / 0.05 to 5.00 mg/l H 2 O 2 Powder Pillows Scope and application: For testing peracetic

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical Oxygen Demand, Chemical DOC316.53.01103 USEPA Reactor Digestion Method Method 10211 1 to 60 mg/l COD (ULR) TNTplus 820 Scope and application: For wastewater, process water, surface water, and cooling water.

More information

DPD Test N Tube Method *

DPD Test N Tube Method * CHLORINE, FREE (0 to 5.00 mg/l) DPD Test N Tube Method * Method 10102 For water, wastewater, and seawater 1. Enter the stored program number for Test N Tube free chlorine (Cl 2 ). Press: PRGM The display

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical Oxygen Demand, Chemical DOC316.53.01104 USEPA Reactor Digestion Method Method 10212 250 to 15,000 mg/l COD (UHR) TNTplus 823 Scope and application: For wastewater and process waters; digestion is required.

More information

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827 Nitrogen, Total DOC316.53.01088 Persulfate Digestion Method Method 10208 5 to 40 mg/l N (HR) TNTplus 827 Scope and application: For water and wastewater. Test preparation Instrument-specific information

More information

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right. Nitrite DOC316.53.01074 USEPA Diazotization Method 1 Method 8507 0.002 to 0.300 mg/l NO 2 N (LR, spectrophotometers) 0.005 to 0.350 mg/l NO 2 N (LR, colorimeters) Scope and application: For water, wastewater

More information

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO Nitrate DOC316.53.01070 Dimethylphenol Method Method 10206 0.23 to 13. NO 3 N or 1.00 to 60.00 mg/l NO 3 (LR) TNTplus 835 Scope and application: For wastewater, drinking water, surface water and process

More information

DR/4000 PROCEDURE MERCURY. Phase 1: Sample Digestion must be done in a fume hood! Toxic gases may be produced.

DR/4000 PROCEDURE MERCURY. Phase 1: Sample Digestion must be done in a fume hood! Toxic gases may be produced. Method 10065 DR/4000 PROCEDURE Scope and Application: For water, wastewater and seawater. * Adapted from Analytical Chemistry, 25 (9) 1363 (1953) Cold Vapor Mercury Preconcentration Method* (0.1 to 2.5

More information

NITROGEN, TOTAL, Test N Tube * Method 10071

NITROGEN, TOTAL, Test N Tube * Method 10071 NITROGEN, TOTAL, Test N Tube * Method 10071 TNT Persulfate Digestion Method (0 to 25.0 mg/l N) Scope and Application: For water and wastewater; digestion is required for determining total nitrogen. The

More information

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting. Chlorine, Total DOC316.53.01497 USEPA DPD Method 1 Method 10260 0.04 to 10.00 mg/l as Cl 2 Chemkey Reagents Scope and application: For drinking water. This product has not been evaluated to test for chlorine

More information

NITRITE, Low Range (0 to mg/l NO 2- -N)

NITRITE, Low Range (0 to mg/l NO 2- -N) NITRITE, Low Range (0 to 0.350 mg/l NO 2- -N) Method 8507 For water, wastewater, seawater Diazotization Method * (Powder Pillows or AccuVac Ampuls); USEPA approved for reporting wastewater and drinking

More information

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832 , TNTplus 832, 10205 DOC316.53.01083 Salicylate Method Method 10205 HR (2 to 47 mg/l NH 3 N) TNTplus 832 Scope and Application: For surface waters, municipal and industrial wastewaters. Test preparation

More information

CHLORINE, FREE, Ultra-high Range ( mg/l Cl 2 ) Method 10069

CHLORINE, FREE, Ultra-high Range ( mg/l Cl 2 ) Method 10069 CHLORINE, FREE, Ultra-high Range (0.0 10.0 mg/l Cl 2 ) Method 10069 DPD Method USEPA accepted for reporting drinking water analyses * For testing higher levels of free chlorine (hypochlorous acid and hypochlorite)

More information

Trihalomethane Formation Potential (THMFP)

Trihalomethane Formation Potential (THMFP) Trihalomethane Formation Potential (THMFP) DOC316.53.01147 THM Plus Method 1 Method 10224 Scope and application: To determine the potential of potable source waters that form trihalomethanes and other

More information

SILICA, Low Range (0 to 1.60 mg/l)

SILICA, Low Range (0 to 1.60 mg/l) SILICA, Low Range (0 to 1.60 mg/l) Heteropoly Blue Method * Method 8186 For water and seawater 1. Enter the stored program number for low range silica (SiO 2 ). Press: PRGM The display will show: PRGM?

More information

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By the Authority Vested By Part 5 of the United States Code 552(a) and Part 1 of the Code of Regulations 51 the attached document has

More information

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA , 10255 DOC316.53.01322 Direct Measurement ISE Method Method 10255 3.55 g/l to 35 g/l Cl Powder Pillow ISA Scope and Application: For the determination of high concentrations (1 M) of chloride in brine

More information

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows Chlorine Free, Indophenol, 10241 Indophenol 1 1 Patent pending. DOC316.53.01256 Method 10241 (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows Scope and Application: For determining residual free chlorine levels

More information

TPH (Total Petroleum Hydrocarbons)

TPH (Total Petroleum Hydrocarbons) TPH (Total Petroleum Hydrocarbons) DOC316.53.01475 Immunoassay 1 Method 10050 Scope and application: For water. 1 This test is semi-quantitative. Results are shown as more or less than the threshold value

More information

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution , drinking water, 8359 DOC316.53.01239 Direct Measurement ISE Method Method 8359 0.04 to 4.00 mg/l NO 3 N TISAB Solution Scope and Application: Drinking water Test preparation How to use instrument-specific

More information

p-dimethylaminobenzaldehyde Method*

p-dimethylaminobenzaldehyde Method* HYDRAZINE (0 to 500 µg/l) p-dimethylaminobenzaldehyde Method* Using Reagent Solution Method 8141 For boiler water/feedwater, water and seawater 1. Enter the stored program number for hydrazine (N 2 H 4

More information

Manganese III Digestion Method * (with chloride removal)

Manganese III Digestion Method * (with chloride removal) Method 10067 OXYGEN DEMAND, CHEMICAL (20 to 1,000 mg/l) For water and wastewater Manganese III Digestion Method * (with chloride removal) 1. Enter the stored program number for Manganese III COD. Press:

More information

Sodium Chloride - Analytical Standard

Sodium Chloride - Analytical Standard Sodium Chloride - Analytical Standard Determination of Arsenic Former numbering: ECSS/CN 312-1982 & ESPA/CN-E-105-1994 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes

More information

Polychlorinated Biphenyls (PCB) in Soil

Polychlorinated Biphenyls (PCB) in Soil Polychlorinated Biphenyls (PCB) in Soil DOC316.53.01107 Immunoassay 1 Method 10050 Scope and application: For soil. 1 This test is semi-quantitative. Results are shown as more or less than the threshold

More information

ALLOWAY METHOD OUTLINE

ALLOWAY METHOD OUTLINE ALLOWAY METHOD OUTLINE Standard Laboratory Method SM4500-Cl -G Parameter Residual Chlorine & Free Chlorine Method DPD Colorimetric Test Kit Date Issued Originator: Section Supervisor: QA Manager Date:

More information

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water Hach Method 1041 Spectrophotometric Measurement of Free Chlorine (Cl ) in Finished Drinking Water Hach Company Method 1041 Revision 1. November 015 Spectrophotometric Measurement of Free Cl in Finished

More information

MERCURY (0.1 to 2.5 µg/l) Method 10065

MERCURY (0.1 to 2.5 µg/l) Method 10065 MERCURY (0.1 to 2.5 µg/l) Method 10065 Cold Vapor Mercury Preconcentration Method * Scope and Application: For water, wastewater and seawater Cat. No. 26583-88 Phase 1 Sample Digestion - must be done in

More information

Ammonia Gas Sensing Combination Electrode

Ammonia Gas Sensing Combination Electrode 51940-88 Ammonia Gas Sensing Combination Electrode Model 50250 Hach Company, 1999. All rights reserved. Printed in the U.S.A. hm/dk 4/99 1ed TRADEMARKS OF HACH COMPANY 2 TABLE OF CONTENTS SPECIFICATIONS...

More information

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method: ALLOWAY METHOD OUTLINE Standard Laboratory Method: SM Parameter: Methylene Blue Method: Colorimetric Reporting Level: Reference: 0.05 mg/l Standard Methods for the Examination Of Water and Wastewater;

More information

SIDE BY SIDE COMPARISON TABLE FOR HACH METHOD NUMBER COLORIMETRIC DETERMINATION OF CYANIDE

SIDE BY SIDE COMPARISON TABLE FOR HACH METHOD NUMBER COLORIMETRIC DETERMINATION OF CYANIDE SIDE BY SIDE COMPARISON TABLE FOR HACH METHOD NUMBER 10265 COLORIMETRIC DETERMINATION OF CYANIDE TOPIC EPA METHOD NUMBER 335.2 (1980) Direct Photometric Method 0.04 0.80 mg cyanide/l SCOPE AND APPLICATION

More information

ASTM Designation: D Standard Test Method for Determination of Iodine Number of Activated Carbon

ASTM Designation: D Standard Test Method for Determination of Iodine Number of Activated Carbon ASTM Designation: D4607-94 Standard Test Method for Determination of Iodine Number of Activated Carbon 1. Scope 1.1 This test method covers the determination of the relative activation level of unused

More information

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+ Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+ Many substances absorb light. When light is absorbed, electrons in the ground state are excited to higher energy levels. Colored

More information

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Nitrate Method 10206 Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Method 10206 Revision 2.2 January 15, 2013 Spectrophotometric

More information

LEAD (Colorimetric) 2. Muffle Furnace: Equipped with pyrometer and capable of operating at controlled temperatures up to 500 C

LEAD (Colorimetric) 2. Muffle Furnace: Equipped with pyrometer and capable of operating at controlled temperatures up to 500 C LEADX.01-1 LEAD (Colorimetric) PRINCIPLE SCOPE Organic matter in the sample is destroyed by ignition in the presence of sulfuric acid. The residue is dissolved in dilute acid, and the lead is complexed

More information

Method for estimation of iodine in urine

Method for estimation of iodine in urine Method for estimation of iodine in urine Described herewith is the standard operating procedure for estimation of iodine in urine by colorimetric method after chloric acid digestion A) Principle: Urine

More information

Experiment 8 Introduction to Volumetric Techniques I. Objectives

Experiment 8 Introduction to Volumetric Techniques I. Objectives Experiment 8 Introduction to Volumetric Techniques I Objectives 1. To learn the proper technique to use a volumetric pipette. 2. To learn the proper technique to use a volumetric flask. 3. To prepare a

More information

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample. METHOD 9200 NITRATE 1.0 SCOPE AND APPLICATION 1.1 This method is applicable to the analysis of ground water, drinking, surface, and saline waters, and domestic and industrial wastes. Modification can be

More information

Introduction to Spectroscopy: Analysis of Copper Ore

Introduction to Spectroscopy: Analysis of Copper Ore Introduction to Spectroscopy: Analysis of Copper Ore Using a Buret and Volumetric Flask: 2.06 ml of solution delivered 2.47 ml of solution delivered 50.00 ml Volumetric Flask Reading a buret: Burets are

More information

Volumetric Measurement Techniques. Technique #1 Use of a Burette. Technique #2 Use of a Pipette. Technique #3 Use of a Volumetric Flask

Volumetric Measurement Techniques. Technique #1 Use of a Burette. Technique #2 Use of a Pipette. Technique #3 Use of a Volumetric Flask Volumetric Measurement Techniques Technique #1 Use of a Burette Technique #2 Use of a Pipette Technique #3 Use of a Volumetric Flask Technique #4 Use of a Bottle-Top Dispenser Last updated 12/6/2009 5:46

More information

Sample Preparation of Electronic Device Components for Hexavalent Chromium Analysis by IEC Method :2017

Sample Preparation of Electronic Device Components for Hexavalent Chromium Analysis by IEC Method :2017 Page 1 of 3 Abstract The increasing use of consumer and electronic device components worldwide has drawn increased attention to their impact on the environment. The correct disposal of these materials

More information

Ammonia Gas Sensing Combination Electrode

Ammonia Gas Sensing Combination Electrode 51927-88 Ammonia Gas Sensing Combination Electrode Model 51927-00 Hach Company, 2001. All rights reserved. Printed in the U.S.A. te/dk 08/01 1ed 2 TABLE OF CONTENTS SPECIFICATIONS...5 SAFETY PRECAUTIONS...7

More information

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY 1.0 SCOPE AND APPLICATION 1.1 This digestion procedure is used for the preparation

More information

Analysis of Hypochlorite in Bleach

Analysis of Hypochlorite in Bleach Experiment 8 Analysis of Hypochlorite in Bleach Adapted by B. D. West and S. E. Schullery of Eastern Michigan University from ANAL 119, written by Enno Wolthuis, Calvin College, published by Chemical Education

More information

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC) METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC) 1.0 SCOPE AND APPLICATION 1.1 Method 7196 is used to determine the concentration of dissolved hexavalent chromium [Cr(VI)] in EP/TCLP characteristic extracts

More information

Spectrophotometric Determination of Ferrocyanide in Effluents

Spectrophotometric Determination of Ferrocyanide in Effluents Spectrophotometric Determination of Ferrocyanide in Effluents ECN-0025-1 INTRODUCTION This method is used to determine the concentration of ferrocyanide ion in photoprocessing solution effluents. The ion

More information

EXPERIMENT 22 SOLUBILITY OF A SLIGHTLY SOLUBLE ELECTROLYTE

EXPERIMENT 22 SOLUBILITY OF A SLIGHTLY SOLUBLE ELECTROLYTE EXPERIMENT 22 SOLUBILITY OF A SLIGHTLY SOLUBLE ELECTROLYTE INTRODUCTION Electrolytes are compounds that are present in solution as ions. They are more likely to be soluble in water than in most other liquids

More information

Titration with an Acid and a Base

Titration with an Acid and a Base Skills Practice Titration with an Acid and a Base Titration is a process in which you determine the concentration of a solution by measuring what volume of that solution is needed to react completely with

More information

METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV)

METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV) METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV) 1.0 SCOPE AND APPLICATION 1.1 Method 9012 is used to determine the concentration of inorganic cyanide in an aqueous waste or leachate.

More information

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml)

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml) GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Silver/Sulfide Ion Electrode is used to measure silver or sulfide ions in aqueous solutions. The two ions are virtually never present in solution

More information

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II) : Absorption Spectroscopy of Cobalt(II) OBJECTIVES In successfully completing this lab you will: prepare a stock solution using a volumetric flask; use a UV/Visible spectrometer to measure an absorption

More information