Standard Operating Procedure. Immunology

Size: px
Start display at page:

Download "Standard Operating Procedure. Immunology"

Transcription

1 Standard Operating Procedure Immunology SOP Ref No SOP title SOP-IU010 PCA Category Laboratory Version 1.0 Page 1 of 12

2 PCA Introduction The enumeration of fresh or frozen-thawed PBMC (peripheral blood mononuclear cells) is required for use of these cells in many cell-based assays. The number of PBMC in a sample may be counted with the Guava Personal Cell Analyzer (PCA) or by traditional hemacytometer methodology. The Guava PCA system provides a cell counting method that reduces the variability of manual hemacytometer counts, especially when multiple operators are involved. The Guava PCA utilizes the Guava ViaCount assay to allow rapid and reliable determination of cell counts and viability. The ViaCount reagent differentially stains viable and non-viable cells based on their permeability to DNA-binding dyes in the ViaCount reagent. The fluorescence of each dye allows quantitative assessment of viable and non-viable nucleated cells present in a suspension. The system counts the stained nucleated cells and uses the forward scatter (FSC) properties to distinguish free nuclei and cellular debris from cells to determine an accurate cell count. The system automatically performs sample analysis and displays results as the viable and total cell counts per ml and per the original sample volume. The Guava PCA may be used to count each sample in replicate. This system offers a consistent method of counting human PBMC compared to hemacytometer counts, potentially making assays more readily transferable across individuals or laboratories. Equipment and Materials Equipment 1. Guava PCA (Cat. # ) (Guava Technologies Inc., Industrial Blvd. Hayward, CA 94545) 2. Dell Latitude laptop computer (Guava ) (Guava Technologies Inc., Industrial Blvd. Hayward, CA 94545) or equivalent 3. Guava CytoSoft Software CD, Version (Guava ) 4. Vortex mixer (Fisher ), or equivalent 5. Pipetman (Rainin P1000, P200 and P20), or equivalent Page 2 of 12

3 Materials Reagents a. Guava ViaCount (Guava 1# ) (100 tests per bottle), store at 2-8 C, manufacturer's expiry b. Guava Check Kit (Guava 1# ) (50 tests), store at 2-8 C, manufacturer's expiry c. Guava Check Beads (Guava # ), store at 2-8 C, manufacturer's expiry d. Guava Check Diluent (Guava 1# ), store at 2-8 C, manufacturer's expiry e. Guava ICF (Guava ), 100 ml bottle, store at room temp, manufacturer's expiry f. Peripheral Blood Mononuclear Cells (freshly prepared from whole blood or thawed from frozen aliquots as described in SOP IU008) g. Sodium Hypoclorite (bleach), any vendor, store at room temperature h. Deionized water from "in house" water tap, or equivalent Supplies a. Sterile tips for pipetman (Costar 4864 and Costar 9032), or equivalent b. Microcentrifuge tubes, 1.5 ml, conical bottom (Fisher ), or equivalent c. Screw cap lids with O-ring (Fisher ) or equivalent Screw cap lids with O- ring (Fisher ) or equivalent note: alternatively regular 1.5ml Eppendorf tubes (Eppendorf ) or equivalent can be used, by cutting the cap with scissors before loading the tube on the GUAVA PCA loader Procedure for GUAVA PCA A. Guava PCA startup 1. Turn on Dell Latitude laptop computer Page 3 of 12

4 2. Once the computer screen shows the Guava Desktop, turn on the Guava PCA instrument at the back right switch, which activates the green light on the front of the Guava PCA. 3. Open the CytoSoft Version software application located on the desktop. This turns the laser on. Once CytoSoft has been opened, the laser will stay on until the clean and shut down at the end of the day, even if CytoSoft is closed 4. Warm up the laser for approximately 15 minutes before acquiring samples. If the PCA will not be used for an extended period (6 hours or more), turn off the laser by running Clean and Shut Down. 5. Make sure the waste vial contains bleach and is not full. If the vial is full, empty into the appropriate biohazard container for liquid waste and refill with bleach until the bottom of the vial is completely covered. B. Running Guava Check Run Guava Check at the start of each day to ensure the system is performing properly. Guava Check averages the results from three acquisitions of a standardized Guava Check bead sample to determine if the results are within the expected range. 1. Once initialization of the CytoSoft application is complete, click on the Guava Check program listed in the Main Menu. NOTE: Make sure the system has been warmed up for approximately 15 minutes before acquiring bead samples. 2. Prepare a Guava Check bead sample for acquisition. a. Pipette 950 µl of Guava Check Diluent into a microcentrifuge tube and 100 µl of Guava Check Diluent into a second microcentrifuge tube. The first tube will be used to run the Guava Check; the second tube will be used for setting the threshold for background noise. b. Thoroughly mix the Guava Check Bead reagent using a vortex mixer set at high speed, Open the tube and pipette any volume from the lid of the Guava Check Beads back into the vial. c. Pipette 50 µl of the Guava Check Bead reagent into the first microcentrifuge tube which contains 950 µl of diluent. Cap the tube and vortex at high speed to mix. Vortex immediately before acquiring the sample. 3. Guava Check Page 4 of 12

5 a. In the Sample Information panel enter the bead lot number and expiration date from the Guava Check Bead reagent vial. b. Enter or verify the Expected Particles/ml value that corresponds to the concentration of beads in the diluted bead sample in the space below the Guava Check lot number. This value is usually 50,000 if prepared at the recommended dilution (a 20-fold dilution of 1 x 10 6 particles/ml). However some Guava check beads lots are prepared to a slightly different concentration and the accurate Expected Particles/ml value can be found stated in accompanying paper within the kit. c. Load the 100 µl Diluent tube onto the sample tube holder. Click on the Adjust Settings button. The software will prompt the loading of diluent. Click "OK" to allow the system to automatically set a threshold to exclude background noise. When a message "Begin a New Data Set appears on the bottom of the screen, the setting is adjusted. d. Confirm the system calibration status by running three replicates of Guava Check bead samples. Mix the Guava Check bead sample thoroughly and load onto the Guava PCA. Click Run 1 st Replicate. The system acquires 1000 events and displays the results in the row for Replicate 1. e. Remove the sample, recap with its original cap, mix well on the vortex and reload on the Guava PCA. Click Run 2 nd Replicate. The system acquires 1000 events and displays the results in the row for Replicate 2. f. Remove the sample, recap with its original cap, mix again and reload on the Guava PCA. Click Run 3 rd Replicate. The system acquires 1000 events and' displays the results in the row for Replicate 3. g. The software automatically calculates the average and % CV for Particles/ml (bead count), FSC and Fluorescence intensities for the three replicates are displayed. If any of the Particles/ml results falls outside the +/- 10% acceptance range, the values appear in red. This calibrates the machine for the day and may take several attempts if there are air bubbles in the tubing. h. If the average of the three replicates is red or % CV is greater than 10%, click on main menu, and run Clean and Shut Down (see Section D in this SOP). Then repeat steps IV.B.3.c-f to rerun Guava Check. If values continue to fall outside the acceptance range, refer to Chapter of the Guava PCA User's Guide for Troubleshooting or call Guava Technologies for service. i. Guava Check may be aborted before all three replicates are run if one of the first two replicates is out of range. In this case run Quick Clean, Backflush or Clean and Shut Down before attempting Guava Check again. Page 5 of 12

6 j. Once Guava Check has successfully been determined, (see step IV.B.3.g-i) above. load a microcentrifuge tube filled with deionized water and click Quick Clean to rinse the fluid system. Return to the Main Menu when Quick Clean is complete. C. Running the Guava ViaCount Assay to count PBMCs 1. Create a paper batch record from Appendix A to record test sample information. Each sample may be randomly assigned a sample code within an assay to facilitate reference to all sample demographics. That sample code should be indicated for each sample in the assay. The batch record also include space to enter Guava count data and to calculate average cell count for a set of replicates and volume needed to achieve a target cell dilution for sample input in the appropriate assay. 2. Prepare samples for Counting. a. From the CytoSoft Main Menu, click Guava ViaCount for the Cell Count and Viability Assay. b. Prepare a uniform cell suspension of PBMC for counting (freshly prepared from whole blood or by thawing previously isolated and frozen PBMC as described in SOP IU008). c. Label and prepare microcentrifuge tubes for each subject. Pipette 190 µl of ViaCount reagent into each microcentrifuge tube. d. Remove 10 ul of each cell suspension and add to the corresponding, labeled microcentrifuge tube containing ViaCount Reagent for a dilution factor of 20. If more than 1.0 x10 7 cells/ml is suspected in the diluted suspension, start at a higher dilution (reference Guava ViaCount Reagent package insert for recommended dilutions). Vortex gently or flick tube with finger to mix. e. Allow the cells to stain for at least 5 minutes at room temperature. 3. Open a data file, adjust settings and enter sample information. a. Enter ViaCount Reagent lot number and expiration into the sample information screen. b. Click New Data Set to create a data file. Select hard drive location to save file and enter a file name. Click Save. c. Click Settings, then click Adjust Settings. d. A message appears prompting operator to load a stained sample. Vortex a stained sample at moderate speed. Load, and click OK. Page 6 of 12

7 NOTE: refer to the profile reported in the figure below for correct instrument settings, detailed in the following sections: e. Adjust PM1 and/or PM2 values only if necessary, to position live cells in the upper left of the Viability (PM1) vs Nucleated Cells (PM2) dot plot between 10e0 and 10e1 on the Viability (PM1) axis and between 10e2 and 10e3 on the Nucleated Cells (PM2) axis. f. Adjust the horizontal line (PM2 threshold) on the Viability (PM1) vs Nucleated Cells (PM2) dot plot to separate the viable cells in the upper left from the debris in the lower left. Set the line approximately 2-3mm below the live cell population g. Adjust the vertical line on the Viability (PM1) vs Nucleated Cells (PM2) dot plot to separate viable cells in the upper left from non-viable cells in the upper right. The angle of the line can be adjusted to better discriminate between the two populations. h. FSC threshold may also be adjusted in the FSC vs Viability (PM1) dot plot to eliminate debris, but taking care not to exclude dead cells from the analysys, or the PBMC viability will be overestimated. Page 7 of 12

8 i. When finished to adjust setting, click Next Step button and enter the data acquisition screen. j. Ensure the Enable EasyFit and Enable Apoptosis Gate buttons in the data acquisition screen are NOT clicked. k. In the Sample information control panel check that dilution factor and original volume are accurate, or modify them accordingly. 4. Run Samples in Guava ViaCount a. Counting samples requires running three replicates of each sample. An optional sample identifier may be entered for the first sample after the first replicate is acquired and before Save and Close Current Sample is clicked. Sample identifiers for all subsequent samples may be added either before or after the sample is acquired. b. Gently vortex the first stained sample on moderate speed and load on the Guava PCA. Click Acquire Next Sample. The system acquires the sample and automatically displays the results. Record Total Viable Cells in Original Sample for this first replicate in a hard copy batch record for the first subject. Click the Save and Close Current Sample button. c. Remove sample, recap with original cap, vortex again and reload on Guava PCA for second replicate. Click Acquire Next Sample. The system acquires the sample and automatically displays the results. Record Total Viable Cells in Original Sample for this second replicate in a hard copy batch record for the first subject. Click the Save and Close Current Sample button. d. Remove sample, recap with original cap, vortex again and reload on Guava PCA for third replicate. Click Acquire Next Sample. The system acquires the sample and automatically displays the results. Record Total Viable Cells in Original Sample for the third replicate in a hard copy batch record for the first. Click the Save and Close Current Sample button. e. Repeat steps b-d for each sample. NOTE: If the concentration is greater than 500 cells/ µl or less than 10 cells/ µl, an error message will appear indicating that counting accuracy may be compromised. Click Abort and remove the sample. Either further dilute or concentrate the sample for accurate counting. f. If any of the triplicate counts is different from the others by +/- 50%, count again with a new aliquot, following steps b-d. g. Run Quick Clean with deionized water in between every replicates or more often as needed to prevent build up in the Guava PCA fluid system. Page 8 of 12

9 h. When data acquisition is complete, close CytoSoft. D. Cleaning of the Guava PCA CytoSoft offers two cleaning cycles, Quick Clean and Clean and Shut Down. It also features Abort and BackFlush functions to clear blockages in the fluid system. 1. Quick Clean: Use this feature of CytoSoft after each assay, in between replicates, or as needed. a. Load a microcentrifuge tube of deionized water or Guava ICF. b. Click Quick Clean to quickly clean the fluid system. c. If the Guava ICF was used, load a microcentrifuge tube of deionized water and click Quick Clean to rinse. If deionized water was used for the Quick clean, cleaning is finished. NOTE: The Guava PCA will prompt you to load a tube of cleaning solution first, but deionized water may be used as a quick flush. d. The tube of water may be left on the Guava PCA until the system is used again or may be discarded if sample acquisition continues. 2. Clean and Shut Down: Use this feature at the end of each day to thoroughly clean the fluid system. Clean and Shut Down also turns off the laser to prepare for system shutdown. a. Click Clean and Shut Down from the main menu. The Guava Clean screen appears. b. Add 10 drops of Guava JCF into a microcentrifuge tube and load onto the Guava PCA. Click Start Cleaning. The laser turns off when Start Cleaning is selected. A message appears to prompt loading of the cleaning solution. Ensure the cleaning tube is loaded and click OK. c. When prompted, load a microcentrifuge tube of deionized water. Click OK. d. A message will appear End of Cleaning. Click Main Menu to return to the CytoSoft main menu or click Exit to close CytoSoft. e. If continuing to run samples (i.e., Clean and Shut Down was run after Guava Check failure or in between assays), allow the laser to warm up for at least 5 minutes. NOTE: Never leave Guava ICF or any cleaning agent on the Guava PCA overnight or for an extended period of time. Always leave a tube of deionized water after cleaning. Page 9 of 12

10 3. Purging the Fluid System: The BackFlush feature reverses the flow of fluid out of the flow cell. a. If a blockage or clog is suspected due to a very low cell count or no cell counts at all, click Abort. b. Click BackFlush. A message instructs to "load a tube of 20% bleach for cleaning". Load a tube of water or Guava ICF and click OK. c. If the Guava ICF was used, load a tube of deionized water and click Quick Clean to rinse any residual cleaning agent. If deionized water was used, continue sample acquisition. d. Replace the sample and click Acquire Next Sample. 4. Shut Down of the instrument and computer. a. Once sample acquisition is complete for the day and Clean and Shut Down has been performed, click Exit in the Main Menu to close the software. b. Shut down the computer and the Guava PCA. E. Analysis A. Refer to batch record created in section C. The counts for each sample and for each replica will be represented in the batch record. B. Alternatively, the GUAVA Excel Utility Program can be used to extract pertinent analysis information from a ViaCount file and create an Excel spreadsheet file. This procedure avoids the need for filling the batch record during acquisition. Counting data can be entered in the batch record at the end of the analysis. 1. Once the ViaCount program has been closed, double click on the GuavaExcelUtil.exe icon on the desktop. 2. Choose the CSV file created during acquisition from the dialog box. If a dialog box appears indicating that the document contains macros, select the Enable Macros option. 3. The utility program creates an Excel workbook with the appropriate information for listed for each sample (replicas) counted. The file name is the same as the CSV file but with the.xls extension. The original CSV file is unchanged. 4. The file can be printed and cell counts can now be entered in the batch record. Page 10 of 12

11 Calculations are performed as follows: 1. The three replicates of Total Viable Cells in Original Sample acquired for each sample must be averaged with the following formula: Average = (Replicate 1 + Replicate 2 + Replicate 3)/3. 2. Volume to resuspend cells (ml) should be calculated based on the Average of the three replicates. Volume to Resuspend Cells (ml) = Average of three replicates (cells)/desired cell concentration (cells/ml) Page 11 of 12

12 Appendix A: Batch Record of Samples Counted (make copies and use one new sheet for each assay set-up) Assay information assay date Operator Sample key sample code 1a 1b 1c 1 average R10 ml for 1 2a 2b 2c 2 average R10 ml for 2 3a 3b 3c 3 average R10 ml for 3 4a 4b 4c 4 average R10 ml for 4 subject identification clinical site Sample Timepoint PBMC date Total viable cell count (triplicate) + average R10 ml for targeted cell concentration SIGNATURE DATE Page 12 of 12

Preparing the sample for determination of Viability

Preparing the sample for determination of Viability Preparing the sample for determination of Viability I. For preparing one sample for analysis you will need: - 1 pcs CELLCHIP - 1 piece of colored Eppendorf with SOFIA GREEN lyophilized dye - 1 piece of

More information

The NGAL Test Reagent Kit

The NGAL Test Reagent Kit PERFORMANCE DATA AND APPLICATION NOTE FOR SIEMENS DIMENSION VISTA The NGAL Test Reagent Kit REF/Cat. No. ST00CA ST00CA ST00CA Product name The NGAL Test Reagent Kit The NGAL Test Calibrator Kit The NGAL

More information

Zetasizer Nano-ZS User Instructions

Zetasizer Nano-ZS User Instructions Zetasizer Nano-ZS User Instructions 1. Activate the instrument computer by logging in to CORAL. If needed, log in to the local instrument computer Username: zetasizer. Password: zetasizer. 2. Instrument

More information

Kinetics of Crystal Violet Bleaching

Kinetics of Crystal Violet Bleaching Kinetics of Crystal Violet Bleaching Authors: V. C. Dew and J. M. McCormick* From Update March 12, 2013 with revisions Nov. 29, 2016 Introduction Chemists are always interested in whether a chemical reaction

More information

Devyser Index Plate A. Instructions for Use

Devyser Index Plate A. Instructions for Use Devyser Index Plate A Art. No.: 8-A200 Instructions for Use Page 1 of 17 TABLE OF CONTENTS TABLE OF CONTENTS 2 1. INTRODUCTION TO DEVYSER INDEX PLATE A 3 1.1 Intended use 3 1.2 Kit configuration Devyser

More information

SIEMENS DIMENSION VISTA 1 The NGAL Test Reagent Kit

SIEMENS DIMENSION VISTA 1 The NGAL Test Reagent Kit PERFORMANCE DATA AND APPLICATION NOTE FOR SIEMENS DIMENSION VISTA 1 The NGAL Test Reagent Kit REF/Cat.No. ST1CA ST2CA ST3CA Product name The NGAL Test Reagent Kit The NGAL Test Calibrator Kit The NGAL

More information

Introduction to Spectroscopy: Analysis of Copper Ore

Introduction to Spectroscopy: Analysis of Copper Ore Introduction to Spectroscopy: Analysis of Copper Ore Using a Buret and Volumetric Flask: 2.06 ml of solution delivered 2.47 ml of solution delivered 50.00 ml Volumetric Flask Reading a buret: Burets are

More information

TruSight Cancer Workflow on the MiniSeq System

TruSight Cancer Workflow on the MiniSeq System TruSight Cancer Workflow on the MiniSeq System Prepare Library Sequence Analyze Data TruSight Cancer 1.5 days ~ 24 hours < 2 hours TruSight Cancer Library Prep MiniSeq System Local Run Manager Enrichment

More information

Nexcelom ViaStain Live Caspase 3/7 Detection for 2D/3D Culture

Nexcelom ViaStain Live Caspase 3/7 Detection for 2D/3D Culture Nexcelom ViaStain Live Caspase 3/7 Detection for 2D/3D Culture Product Numbers: CSK-V0002-1, CSK-V0003-1 This product is for RESEARCH USE ONLY and is not approved for diagnostic or therapeutic use. Table

More information

University of Minnesota Nano Center Standard Operating Procedure

University of Minnesota Nano Center Standard Operating Procedure University of Minnesota Nano Center Standard Operating Procedure Equipment Name: Zeta Potential Analyzer Model: Stabino Location: PAN 185 Badger Name: Not on Badger Revision Number: 0-Inital release Revisionist:

More information

mrna Isolation Kit for Blood/Bone Marrow For isolation mrna from blood or bone marrow lysates Cat. No

mrna Isolation Kit for Blood/Bone Marrow For isolation mrna from blood or bone marrow lysates Cat. No For isolation mrna from blood or bone marrow lysates Cat. No. 1 934 333 Principle Starting material Application Time required Results Key advantages The purification of mrna requires two steps: 1. Cells

More information

Component Product # Product # Cell Lysis Reagent 100 ml 500 ml Product Insert 1 1

Component Product # Product # Cell Lysis Reagent 100 ml 500 ml Product Insert 1 1 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Cell Lysis Reagent Product # 18800 (100 ml) Product # 18801 (500

More information

FOR RESEARCH USE ONLY.

FOR RESEARCH USE ONLY. MAN-10039-04 Vantage 3D DNA SNV Qualification Kit Vantage 3D DNA SNV Qualification Kit The ncounter Vantage 3D DNA SNV Qualification Kit is designed to assess whether the ncounter MAX, FLEX, or SPRINT

More information

SIEMENS DIMENSION VISTA 1 The NGAL Test Reagent Kit

SIEMENS DIMENSION VISTA 1 The NGAL Test Reagent Kit APPLICATION NOTE FOR SIEMENS DIMENSION VISTA 1 The NGAL Test Reagent Kit REF/Cat.No. ST1RA ST2RA ST3RA Product name The NGAL Test Reagent Kit The NGAL Test Calibrator Kit The NGAL Test Control Kit R1 R2

More information

camp Direct Immunoassay Kit

camp Direct Immunoassay Kit camp Direct Immunoassay Kit Catalog Number KA0886 100 assays Version: 05 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information... 4 Materials

More information

RayBio CaspGLOW TM Fluorescein Active Caspase-3 Staining Kit

RayBio CaspGLOW TM Fluorescein Active Caspase-3 Staining Kit RayBio CaspGLOW TM Fluorescein Active Caspase-3 Staining Kit User Manual Version 1.0 May 10 th, 2015 RayBio Caspase-3 Fluorometric Assay Kit Protocol (Cat#: 68FLS-Casp3-S) RayBiotech, Inc. We Provide You

More information

RayBio Glucose Dehydrogenase Activity Assay Kit

RayBio Glucose Dehydrogenase Activity Assay Kit RayBio Glucose Dehydrogenase Activity Assay Kit User Manual Version 1.0 January 23, 2015 RayBio Glucose Dehydrogenase Activity Assay (Cat#: 68AT-GluD-S100) RayBiotech, Inc. We Provide You With Excellent

More information

Simple Operation Manual

Simple Operation Manual 苦 SET UP Simple Operation Manual (1) Check the level of buffer, purified water and wash solution Left: Buffer Center: Wash solution Right: Purified water Change buffer or refill them (W.S. and P.W.) if

More information

Predictor Assay Setup Guide on the BMG LABTECH CLARIOstar Microplate Readers

Predictor Assay Setup Guide on the BMG LABTECH CLARIOstar Microplate Readers Page 1 of 18 Predictor Assay Setup Guide on the BMG LABTECH CLARIOstar Microplate Readers The BMG LABTECH CLARIOstar Microplate Readers were tested for compatibility with Life Technologies' Predictor herg

More information

RayBio MMP-3 Activity Assay Kit

RayBio MMP-3 Activity Assay Kit RayBio MMP-3 Activity Assay Kit User Manual Version 1.0 Mar 26, 2014 RayBio MMP-3 Activity Assay (Cat#: 68AT-MMP3-S100) RayBiotech, Inc. We Provide You With Excellent Support And Service Tel:(Toll Free)1-888-494-8555

More information

Ion Sphere Assay on the Qubit 3.0 Fluorometer

Ion Sphere Assay on the Qubit 3.0 Fluorometer USER GUIDE Ion Sphere Assay on the Qubit 3.0 Fluorometer for use with: Ion Sphere Quality Control Kit (Cat. No. 4468656) Publication Number MAN0016388 Revision A.0 Ion Sphere Assay overview... 2 Materials

More information

IV. EQUIPMENT a. Horiba LA 950 main unit b. Dry sample unit that has a silencer behind it c. Cell slide unit that holds

IV. EQUIPMENT a. Horiba LA 950 main unit b. Dry sample unit that has a silencer behind it c. Cell slide unit that holds STANDARD OPERATING PROCEDURE No of pages 7 No: METH 006 Version: 1 Date: 05-06- 2011 Title: Particle Size Distribution Analysis by Laser Diffraction method with HORIBA LA 950 in wet mode I. SCOPE AND APPLICATION

More information

Modified Adams Assay for Phenolics in Wine

Modified Adams Assay for Phenolics in Wine Modified Adams Assay for Phenolics in Wine 1. Total Iron-Reactive Phenolics THIS VALUE WILL DETERMINE DILUTIONS FOR TANNIN & POLYMERIC PIGMENT ANALYSES 1.1 Into a reduced volume cuvette, pipette in the

More information

Scope: Materials: Labeled 1.5 ml or 2 ml Eppendorf tubes Appropriate pipette tips (see Table 1) Labeled LC vials

Scope: Materials: Labeled 1.5 ml or 2 ml Eppendorf tubes Appropriate pipette tips (see Table 1) Labeled LC vials Scope: This SOP applies to sample protein precipitation for global metabolomics analysis by reverse phase or HILIC- HPLC-MS. Samples include but are not limited to tissue, cells, plasma, serum, and stool.

More information

Immunoassay Kit (Colorimetric)

Immunoassay Kit (Colorimetric) RayBio cgmp Direct Immunoassay Kit (Colorimetric) User Manual Version 1.0 May 25, 2014 RayBio cgmp Direct Immunoassay Kit (Colorimetric) Protocol (Cat#: 68AT-cGMP-S100) RayBiotech, Inc. We Provide You

More information

Human anti-myelin associated glycoprotein antibody (MAG) Ab ELISA Kit

Human anti-myelin associated glycoprotein antibody (MAG) Ab ELISA Kit Human anti-myelin associated glycoprotein antibody (MAG) Ab ELISA Kit Catalog Number. MBS700087 For the qualitative determination of human anti-myelin associated glycoprotein antibody (MAG) concentrations

More information

OESTREICH LAB CHIP PROTOCOL

OESTREICH LAB CHIP PROTOCOL OESTREICH LAB CHIP PROTOCOL Buffers (all starred (*) buffers must be autoclaved before use) 10X Cell Lysis Buffer*: 100mL 0.214g Potassium Acetate (KOAc) = 10mM 0.147g Magnesium Acetate (MgAc) = 15mM 10mL

More information

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7 Troy, NY 12180 Date: 03/16/2012 Page 1 of 7 USGS District Laboratory, Troy, NY Inductively Coupled Plasma Optical Emission Spectrometry Standard Operating Procedure 1. Scope and Application 1.1 Analytes

More information

NukEx Nucleic Acid Release Reagent

NukEx Nucleic Acid Release Reagent Instruction for Use NukEx Nucleic Acid Release Reagent For general laboratory use. For in vitro use only. Reagent for the enzymatic release of nucleic acid from tissue samples, ticks, insects and swabs.

More information

Mouse Retinol Binding Protein 4 ELISA

Mouse Retinol Binding Protein 4 ELISA Mouse Retinol Binding Protein 4 ELISA For the quantitative determination of RBP4 in mouse serum and plasma. Please see Appendix A for Reference Serum Information. For Research Use Only. Not For Use In

More information

Monkey Kidney injury molecule 1,Kim-1 ELISA Kit

Monkey Kidney injury molecule 1,Kim-1 ELISA Kit Monkey Kidney injury molecule 1,Kim-1 ELISA Kit Catalog No: E0785Mo 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH

More information

Technology that counts

Technology that counts Application Note No. 102 Version 1.3 Determination of Sperm Density in Ejaculates from Bulls Application This Application Note describes how sperm cell density in ejaculates from bulls can be determined

More information

Crime Prep Adem-Kit (Cat #06210) Instruction for manual protocol

Crime Prep Adem-Kit (Cat #06210) Instruction for manual protocol (Cat #06210) Instruction for manual protocol ADEMTECH SA Parc scientifique Unitec 1 4, allée du Doyen G. Brus 33600 PESSAC France Tel: +33557020201 Fax +3355702020 Visit our Web site: www.ademtech.com

More information

ab83360 Ammonia Assay Kit

ab83360 Ammonia Assay Kit Version 8 Last updated 18 December 2017 ab83360 Ammonia Assay Kit For the measurement of total ammonia and ammonium levels in various samples This product is for research use only and is not intended for

More information

Rat Ferritin ELISA. For the determination of ferritin in serum and plasma of rats. For Research Use Only. Not For Use In Diagnostic Procedures.

Rat Ferritin ELISA. For the determination of ferritin in serum and plasma of rats. For Research Use Only. Not For Use In Diagnostic Procedures. Rat Ferritin ELISA For the determination of ferritin in serum and plasma of rats. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 41-FERRT-E01 Size: 96 Wells Version: 3 L52-16

More information

STANDARD OPERATING PROCEDURE (SOP) FOR ESTIMATION OF DIETARY INTAKE OF SODIUM (DIS) IN 24 HOUR URINE USING ELECTROLYTE ANALYZER

STANDARD OPERATING PROCEDURE (SOP) FOR ESTIMATION OF DIETARY INTAKE OF SODIUM (DIS) IN 24 HOUR URINE USING ELECTROLYTE ANALYZER STANDARD OPERATING PROCEDURE (SOP) FOR ESTIMATION OF DIETARY INTAKE OF SODIUM (DIS) IN 24 HOUR URINE USING ELECTROLYTE ANALYZER Estimation of Dietary Intake of Sodium (DIS) in 24h Urine on electrolyte

More information

Mouse Glutathione S Transferase Alpha 1 (GSTa1)

Mouse Glutathione S Transferase Alpha 1 (GSTa1) Mouse Glutathione S Transferase Alpha 1 (GSTa1) For the quantitative determination of mouse (GSTa1) in serum, plasma and other biological fluids Cat. No. KT-16414 For Research Use Only. Not for use in

More information

Human rheumatoid factor (RF) antibody (IgM) ELISA Kit

Human rheumatoid factor (RF) antibody (IgM) ELISA Kit Human rheumatoid factor (RF) antibody (IgM) ELISA Kit Catalog Number. CSB-EQ027654HU For the qualitative determination of human rheumatoid factor (RF) antibody (IgM) concentrations in serum, plasma. This

More information

PROTOCOL: GCP Protein Quantification Assay and QC (QUANT)

PROTOCOL: GCP Protein Quantification Assay and QC (QUANT) PROTOCOL: GCP Protein Quantification Assay and QC (QUANT) Purpose To determine overall protein yield and histone purity after histone extraction. SDS PAGE gels are not required, only when changing variables

More information

Phosphoenolpyruvate Assay Kit

Phosphoenolpyruvate Assay Kit Phosphoenolpyruvate Assay Kit Catalog Number KA3745 100 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information... 4 Materials

More information

Canine brain natriuretic peptide,bnp ELISA Kit

Canine brain natriuretic peptide,bnp ELISA Kit Canine brain natriuretic peptide,bnp ELISA Kit Catalog No: E0541c 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH

More information

CSA: Simian Panel E. (Product Number ) User Protocol. 1.0 Introduction Kit Contents Required Materials...

CSA: Simian Panel E. (Product Number ) User Protocol. 1.0 Introduction Kit Contents Required Materials... L757 28Jun2018 1 of 9 Table of Contents 1.0 Introduction... 2 2.0 Kit Contents... 3 3.0 Required Materials... 3 4.0 Storage... 3 5.0 Safety and Handling... 4 6.0 Protocol Overview... 4 7.0 Procedure...

More information

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III) Chem 1B Saddleback College Dr. White 1 Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III) Objectives To use spectroscopy to relate the absorbance of a colored solution to its concentration. To

More information

GGT Activity Assay Kit (Colorimetric)

GGT Activity Assay Kit (Colorimetric) GGT Activity Assay Kit (Colorimetric) Catalog Number KA3746 100 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information...

More information

In-Solution Digestion: Multi-Plate v1.0 Quick Start Guide

In-Solution Digestion: Multi-Plate v1.0 Quick Start Guide In-Solution Digestion: Multi-Plate v1.0 Quick Start Guide This guide is for users who have been trained in the proper use of the AssayMAP Bravo Platform and understand the safety guidelines in the Bravo

More information

IgD ELISA. For the quantitative determination of IgD in Human serum and plasma. Please See Appendix A for Reference Serum Information

IgD ELISA. For the quantitative determination of IgD in Human serum and plasma. Please See Appendix A for Reference Serum Information IgD ELISA For the quantitative determination of IgD in Human serum and plasma. Please See Appendix A for Reference Serum Information For Research Use Only. Not for Use in Diagnostic Procedures. Catalog

More information

11,12-EET/DHET ELISA kit

11,12-EET/DHET ELISA kit 11,12-EET/DHET ELISA kit Catalog Number: DH5/DH15/DH25 Store at -20 C. FOR RESEARCH USE ONLY V.0522010 Introduction It is well known that arachidonic acid (AA) will be converted to EET by P 450 arachidonic

More information

ELISA Technical Manual and Troubleshooting Guide

ELISA Technical Manual and Troubleshooting Guide ELISA Technical Manual and Troubleshooting Guide 2014 Contents: Page Kit Storage Guidelines 3 Materials and Equipment Required 4 Performing the ELISA Assay 5-8 Sample Handling and Storage 9 Troubleshooting

More information

BIOO FOOD AND FEED SAFETY. Histamine Enzymatic Assay Kit Manual. Catalog #: Reference #:

BIOO FOOD AND FEED SAFETY. Histamine Enzymatic Assay Kit Manual. Catalog #: Reference #: BIOO FOOD AND FEED SAFETY Histamine Enzymatic Assay Kit Manual Catalog #: 1032-05 Reference #: 1032-05 BIOO Scientific Corp. 2010 TABLE OF CONTENTS GENERAL INFORMATION... 1 Product Description... 1 Procedure

More information

Human anti-deoxyribonuclease B, anti-dnase B ELISA Kit

Human anti-deoxyribonuclease B, anti-dnase B ELISA Kit Human anti-deoxyribonuclease B, anti-dnase B ELISA Kit Catalog No: E0311h 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ

More information

Introduction to Spectroscopy: Analysis of Copper Ore

Introduction to Spectroscopy: Analysis of Copper Ore Introduction to Spectroscopy: Analysis of Copper Ore Using a Buret and Volumetric Flask: 2.06 ml of solution 2.47 ml of solution 50.00 ml delivered delivered Volumetric Flask Reading a buret: Burets are

More information

IgD ELISA. For the quantitative determination of IgD in Human Sera. Please See Appendix A for Reference Serum Information

IgD ELISA. For the quantitative determination of IgD in Human Sera. Please See Appendix A for Reference Serum Information IgD ELISA For the quantitative determination of IgD in Human Sera. Please See Appendix A for Reference Serum Information For Research Use Only. Not for Use in Diagnostic Procedures Catalog Number: 41-IGDHU-E01

More information

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III) Chem 1B Dr. White 11 Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III) Objectives To use spectroscopy to relate the absorbance of a colored solution to its concentration. To prepare a Beer s

More information

ON SITE SYSTEMS Chemical Safety Assistant

ON SITE SYSTEMS Chemical Safety Assistant ON SITE SYSTEMS Chemical Safety Assistant CS ASSISTANT WEB USERS MANUAL On Site Systems 23 N. Gore Ave. Suite 200 St. Louis, MO 63119 Phone 314-963-9934 Fax 314-963-9281 Table of Contents INTRODUCTION

More information

VDL ENDOTOXIN ASSAY: ASSAY FOR LIMULUS AMEBOCYTE LYSATE

VDL ENDOTOXIN ASSAY: ASSAY FOR LIMULUS AMEBOCYTE LYSATE 1. Purpose 1.1. The purpose of this protocol is test purified viral vectors for endotoxin contamination. 1.2. This procedure is routinely performed in the Vector Development Laboratory (VDL) following

More information

Determining the Concentration of a Solution: Beer s Law

Determining the Concentration of a Solution: Beer s Law Determining the Concentration of a Solution: Beer s Law Vernier Spectrometer 1 The primary objective of this experiment is to determine the concentration of an unknown copper (II) sulfate solution. You

More information

Human papillomavirus,hpv ELISA Kit

Human papillomavirus,hpv ELISA Kit Human papillomavirus,hpv ELISA Kit Catalog No: E0787h 96 Tests Operating instructions www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH ENTIRE PROCEDURE

More information

Mouse IgE ELISA. For the quantitative determination of IgE in mouse serum or plasma. For Research use Only. Not For Use In Diagnostic Procedures.

Mouse IgE ELISA. For the quantitative determination of IgE in mouse serum or plasma. For Research use Only. Not For Use In Diagnostic Procedures. Mouse IgE ELISA For the quantitative determination of IgE in mouse serum or plasma. For Research use Only. Not For Use In Diagnostic Procedures. Please see Appendix A for Reference Serum information Catalog

More information

Instructor s Advance Preparation

Instructor s Advance Preparation INSTRUCTOR'S MANUAL Instructor s Advance Preparation This protocol is designed for 80 workstations of 4 students. Each group will prepare a set of standards, a blank, and 2 milk samples (can be a blind

More information

Cystatin C ELISA. For the quantitative determination of cystatin C in human biological samples.

Cystatin C ELISA. For the quantitative determination of cystatin C in human biological samples. Cystatin C ELISA For the quantitative determination of cystatin C in human biological samples. Please read carefully due to Critical Changes, e.g., Calibrator concentration, Standard preparation. Please

More information

Size Determination of Gold Nanoparticles using Mie Theory and Extinction Spectra

Size Determination of Gold Nanoparticles using Mie Theory and Extinction Spectra Size Determination of Gold Nanoparticles using Mie Theory and Extinction Spectra OUTLINE OF THE PROCEDURE A) Measure the extinction spectra of each of your samples. B) Model the extinction spectrum of

More information

RATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION

RATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION Rate Law Determination of Crystal Violet Hydroxylation Revised 5/22/12 RATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION Adapted from "Chemistry with Computers" Vernier Software, Portland OR, 1997

More information

Mag-Bind Soil DNA Kit. M preps M preps M preps

Mag-Bind Soil DNA Kit. M preps M preps M preps Mag-Bind Soil DNA Kit M5635-00 5 preps M5635-01 50 preps M5635-02 200 preps January 2013 Mag-Bind Soil DNA Kit Table of Contents Introduction and Overview...2 Kit Contents/Storage and Stability...3 Preparing

More information

Dog IgM ELISA. Cat. No. KT-458 K-ASSAY. For the quantitative determination of IgM in dog biological samples. For Research Use Only. 1 Rev.

Dog IgM ELISA. Cat. No. KT-458 K-ASSAY. For the quantitative determination of IgM in dog biological samples. For Research Use Only. 1 Rev. K-ASSAY Dog IgM ELISA For the quantitative determination of IgM in dog biological samples. Cat. No. KT-458 For Research Use Only. 1 Rev. 010209 K-ASSAY PRODUCT INFORMATION Dog IgM ELISA Cat. No. KT-458

More information

IgA ELISA. For the quantitative determination of IgA in human serum and plasma. Please see Appendix A for Reference Serum Information

IgA ELISA. For the quantitative determination of IgA in human serum and plasma. Please see Appendix A for Reference Serum Information IgA ELISA For the quantitative determination of IgA in human serum and plasma Please see Appendix A for Reference Serum Information For Research use Only. Not For Use In Diagnostic Procedures. Catalog

More information

VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual

VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual Page 1 / 11 VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual ZV-3030-0200-20 200 2-8 C Page 2 / 11 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. TEST PRINCIPLE... 3

More information

TOXI-SCREENING KIT. Microbiotest for ultra-rapid on-site. toxicity screening of water STANDARD OPERATIONAL PROCEDURE

TOXI-SCREENING KIT. Microbiotest for ultra-rapid on-site. toxicity screening of water STANDARD OPERATIONAL PROCEDURE TOXI-SCREENING KIT Microbiotest for ultra-rapid on-site toxicity screening of water STANDARD OPERATIONAL PROCEDURE TABLE OF CONTENTS Introduction to the Toxi-Screening microbiotest 2 - Origin 2 - Biological

More information

ADH Activity Assay Kit

ADH Activity Assay Kit ADH Activity Assay Kit Catalog Number KA3713 100 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information... 4 Materials

More information

IgM (Canine) ELISA. For the quantitative determination of IgM in canine serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

IgM (Canine) ELISA. For the quantitative determination of IgM in canine serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. IgM (Canine) ELISA For the quantitative determination of IgM in canine serum and plasma For Research Use Only. Not For Use In Diagnostic Procedures. Please See Appendix A for Reference Serum Information

More information

PREPARATION FOR CHEMISTRY LAB: FLUORIDE IN WATER

PREPARATION FOR CHEMISTRY LAB: FLUORIDE IN WATER 1 Name: Lab Instructor: PREPARATION FOR CHEMISTRY LAB: FLUORIDE IN WATER On these problems, the solvent is water and the solution is sufficiently dilute so that the density of the solution is the same

More information

Human 25-hydroxy vitamin D3 (25HVD3) ELISA Kit

Human 25-hydroxy vitamin D3 (25HVD3) ELISA Kit Human 25-hydroxy vitamin D3 (25HVD3) ELISA Kit Catalog Number. CSB-E08097h For the quantitative determination of human 25-hydroxy vitamin D3 (25HVD3) concentrations in serum. This package insert must be

More information

IgG (Canine) ELISA. Please see Appendix A for Reference Serum information. For Research Use Only. Not For Use In Diagnostic Procedures.

IgG (Canine) ELISA. Please see Appendix A for Reference Serum information. For Research Use Only. Not For Use In Diagnostic Procedures. IgG (Canine) ELISA For the quantitative determination of IgG in canine serum or plasma Please see Appendix A for Reference Serum information. For Research Use Only. Not For Use In Diagnostic Procedures.

More information

Mouse IgG2B ELISA. Cat. No. KT-405 K-ASSAY. For the quantitative determination of IgG2B in mouse biological samples. For Research Use Only.

Mouse IgG2B ELISA. Cat. No. KT-405 K-ASSAY. For the quantitative determination of IgG2B in mouse biological samples. For Research Use Only. K-ASSAY Mouse IgG2B ELISA For the quantitative determination of IgG2B in mouse biological samples Cat. No. KT-405 For Research Use Only. 1 Rev. 020808 K-ASSAY PRODUCT INFORMATION Mouse IgG2B ELISA Cat.

More information

Porcine Immunoglobulin E (IgE)ELISA Kit

Porcine Immunoglobulin E (IgE)ELISA Kit Porcine Immunoglobulin E (IgE)ELISA Kit Catalog No. MBS703293 (96T) This immunoassay kit allows for the in vitro quantitative determination of porcine IgE concentrations in serum and plasma,. Expiration

More information

IgG (Chicken) ELISA Kit

IgG (Chicken) ELISA Kit IgG (Chicken) ELISA Kit Catalog Number KA2426 96 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of the

More information

Protein extraction and quantification by Teresa Fan, University of Kentucky BUFFER PREPARATION AND STORAGE

Protein extraction and quantification by Teresa Fan, University of Kentucky BUFFER PREPARATION AND STORAGE Protein extraction and quantification by Teresa Fan, University of Kentucky Note: This procedure follows [Fan_Extract_Polar_Lipid_Prot_SOP] Step 11 (middle part, in 1.5 ml tube). BUFFER PREPARATION AND

More information

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay... Table of Contents 1. Principle of Assay... 3 2. Specimens... 3 3. Reagents and Materials (HPLC grade)... 3 4. Standards, Controls, and Solutions... 4 5. Equipment and Special Supplies... 5 6. Instrumentation...

More information

ab83360 Ammonia Assay Kit

ab83360 Ammonia Assay Kit Version 9 Last updated 7 February 2019 ab83360 Ammonia Assay Kit For the measurement of total ammonia and ammonium levels in various samples View kit datasheet: www.abcam.com/ab83360 (use www.abcam.cn/ab83360

More information

STANDARD OPERATION PROCEDURE FOR THE COMMON FUND: PROTEIN CAPTURE REAGENTS PROGRAM (IP-MS)

STANDARD OPERATION PROCEDURE FOR THE COMMON FUND: PROTEIN CAPTURE REAGENTS PROGRAM (IP-MS) STANDARD OPERATION PROCEDURE FOR THE COMMON FUND: PROTEIN CAPTURE REAGENTS PROGRAM (IP-MS) 1. PURPOSE This procedure will describe how to prepare human Fab / antigen samples or Mouse Monoclonal Antibody

More information

Serotonin (Human) ELISA Kit

Serotonin (Human) ELISA Kit Serotonin (Human) ELISA Kit Catalog Number KA1151 96 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay... 3 General

More information

Sheep myeloperoxidase(mpo) ELISA kit

Sheep myeloperoxidase(mpo) ELISA kit Sheep myeloperoxidase(mpo) ELISA kit Catalog No. CSB-EL014757SH (96T) This immunoassay kit allows for the in vitro quantitative determination of sheep MPO concentrations in serum and plasma. Expiration

More information

E.Z.N.A. MicroElute Clean-up Kits Table of Contents

E.Z.N.A. MicroElute Clean-up Kits Table of Contents E.Z.N.A. MicroElute Clean-up Kits Table of Contents Introduction... 2 Kit Contents... 3 Preparing Reagents/Storage and Stability... 4 Guideline for Vacuum Manifold... 5 MicroElute Cycle-Pure - Spin Protocol...

More information

genesig Easy q16 Validation kit

genesig Easy q16 Validation kit Primerdesign TM Ltd genesig Easy q16 Validation kit For use on the genesig q16 50 reactions For general laboratory and research use only. 1 genesig Easy: at a glance guide For each validation test Component

More information

Axiom TM 2.0 Target Prep 384 Samples

Axiom TM 2.0 Target Prep 384 Samples Quick Reference Card Axiom TM 2.0 Target Prep 384 Samples Stage 1 DNA Amplification Introduction Running the Axiom 2.0 Assay for 384 Samples Assay requires the following sets of steps: 1. Genomic DNA Prep,

More information

Bovine FSH(Follicle Stimulating Hormone) ELISA Kit

Bovine FSH(Follicle Stimulating Hormone) ELISA Kit Bovine FSH(Follicle Stimulating Hormone) ELISA Kit Catalogue No.: EB0141 Size: 48T/96T Reactivity: Bovine Detection Range: 1.563-100mIU/ml Sensitivity:

More information

RayBio Protease Activity Assay Kit

RayBio Protease Activity Assay Kit RayBio Protease Activity Assay Kit User Manual Version 1.0 Mar 25, 2013 RayBio Protease Activity Assay (Cat#: 68AT-Protease-S100) RayBiotech, Inc. We Provide You With Excellent Support And Service Tel:(Toll

More information

Instructions for use. Glycine ELISA BA E-2100

Instructions for use. Glycine ELISA BA E-2100 Instructions for use Glycine ELISA BA E-2100 Glycine Urine ELISA 1. Intended use and principle of the test Enzyme Immunoassay for the quantitative determination of Glycine in urine. After derivatization

More information

Ammonia Assay Kit. Catalog Number KA assays Version: 02. Intended for research use only.

Ammonia Assay Kit. Catalog Number KA assays Version: 02. Intended for research use only. Ammonia Assay Kit Catalog Number KA3707 100 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information... 4 Materials Supplied...

More information

Purdue-UAB Botanicals Center for Age- Related Disease

Purdue-UAB Botanicals Center for Age- Related Disease Purdue-UAB Botanicals Center for Age- Related Disease MALDI-TOF Mass Spectrometry Fingerprinting Technique Landon Wilson MALDI-TOF mass spectrometry is an advanced technique for rapid protein identification

More information

Caspase Substrate Assay Kit (Colorimetric)

Caspase Substrate Assay Kit (Colorimetric) Caspase Substrate Assay Kit (Colorimetric) Catalog Number KA3694 7 x 25 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information...

More information

Caspase 3 (active) Red Staining Kit

Caspase 3 (active) Red Staining Kit ab65617 Caspase 3 (active) Red Staining Kit Instructions for Use For the rapid, sensitive and accurate detection of activated Caspase 3 in living cells. This product is for research use only and is not

More information

CTX-II (Human) ELISA Kit

CTX-II (Human) ELISA Kit CTX-II (Human) ELISA Kit Catalog Number KA1225 96 assays Version: 15 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay... 3 General

More information

ALDH Activity Assay Kit (Colorimetric)

ALDH Activity Assay Kit (Colorimetric) ALDH Activity Assay Kit (Colorimetric) Catalog Number KA3742 100 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information...

More information

Aflatoxin M1 (AFM1) ELISA Kit

Aflatoxin M1 (AFM1) ELISA Kit Aflatoxin M1 (AFM1) ELISA Kit Catalog Number. CSB-EL027236 This immunoassay kit allows for the in vitro quantitative determination of Aflatoxin M1 concentrations in milk, milk power. This package insert

More information

Procine GHb ELISA Kit

Procine GHb ELISA Kit Procine GHb ELISA Kit For the quantitative in vitro determination of Procine Glycated hemoglobin concentrations in serum - plasma - celiac fluid - tissue homogenate - body fluid FOR LABORATORY RESEARCH

More information

Mouse Insulin ELISA KIT S-Type

Mouse Insulin ELISA KIT S-Type Mouse Insulin ELISA KIT S-Type Research Reagent Please, read this instruction carefully before use. This is an ELISA (Enzyme Linked ImmunoSorbent Assay) kit for measurement of mouse insulin with high specificity

More information

Pig Vascular Endothelial Cell Growth Factor A (VEGF-A) ELISA

Pig Vascular Endothelial Cell Growth Factor A (VEGF-A) ELISA Pig Vascular Endothelial Cell Growth Factor A (VEGF-A) ELISA For the quantitative determination of pig VEGF-A in serum, plasma and other biological fluids. Cat. No. KT-32008 For Research Use Only. Not

More information

Endotoxin ELISA Kit (Competitive EIA)

Endotoxin ELISA Kit (Competitive EIA) Endotoxin ELISA Kit (Competitive EIA) Catalog No. DEIAPY07261 It is important that you read this entire manual carefully before starting your experiment. This kit is for Research Use Only. Not for Diagnostic

More information

Protocol Particle size distribution by centrifugal sedimentation (CPS)

Protocol Particle size distribution by centrifugal sedimentation (CPS) Protocol Particle size distribution by centrifugal sedimentation (CPS) 1. Method The CPS Disc Centrifuge separates particles by size using centrifugal sedimentation in a liquid medium. The sedimentation

More information